Druck PTX1210, PMP1210 Data sheet

PTX/PMP 1200 Series
Industrial Pressure Transmitters
" ±0.25% FS accuracy
" NACE compatible wetted parts
" Low-cost, rugged construction
" Class I, Div 1, Intrinsically safe certified " Ranges from 2.5 to 10,000 psi
PTX/PMP 1200
PTX/PMP 1200 Series
Industrial Pressure Transmitters
The PTX/PMP 1200 Series is a complete range of pressure transmitters designed to optimize the relationship between price and performance while providing flexibility from a standard production device.
At the heart of the 1200 Series is an advanced, high stability pressure measurement element, micro-machined from single crystal silicon within Druck’s own Class 100 processing facility. The silicon element is mounted within a high integrity glass to metal seal and is fully isolated from thepressure media by a hastelloy isolation diaphragm, electron beam welded to the front of the glass to metal seal. Druck’s proprietary low oil volume isolation allows fast dynamic response plus minimal thermal error . The high output silicon allows for high overload capability.
Silicon wafers being loaded into an oxidation furnace.
Applications include general purpose industrial, refrigeration, oil & gas well head monitoring and control, compressor performance monitoring and static pressure measurements in flow applications.
Surface mount electronics condition the output from the silicon diaphragm, correct for thermally induced errors, and configure the output to the required 4-20 mA or 1-5 Vdc output. Advanced design features built into the electronic circuitry enable minimum sensor size with the utmost reliability. The electronics incorporate power supply regulation, reverse polarity, overvoltage and shor t circuit protection, coupled with EMC protection components.
The fully encapsulated solid state design ensures integrity of product under high levels of shock and vibration, with an ingress protection rating of NEMA 4X, depending on electrical termination selected.
A fully tested pressure core is stocked in quantity , requiring only the addition of the electrical termination. Prior to shipping, the sensor is adjusted to meet the particular pressure range and units and terminated with the electrical connection.
Some configurations allow access to zero and span potentiometers for adjustments during set up and calibration. Others are fixed and cannot be adjusted.
A range of stainless steel pressure adapters are available to modify the standard 1/4”NPT female pressure port.
The PTX/PMP 1200 Series is ideally suited to meet the rigorous demands of the Industrial, Automotive, Aerospace and Oil & Gas markets. The PMP 1200 is designed for applications where batteries and/or solar powered systems require very low power consumption. The PTX/PMP 1240 is also being qualified for Class I, Div 1 Explosion-proof service.
Computerized testing and calibration of pressure transducers.
Loading...
+ 2 hidden pages