PTX/PMP 1200 Series
Industrial Pressure Transmitters
For over 25 y ears Druck has manufactured precision
pressure sensors with a capability to meet critical
applications in industrial, aerospace, and researc h
environments.
The PTX/PMP 1200 Series is a complete range of pressure
transmitters designed to optimize the relationship between
price and performance while providing flexibility from
a standard production device.
At the heart of the 1200 Series
is an advanced, high stability
pressure measurement
element, micro-machined from
single crystal silicon within
Druck’s own Class 100
processing facility. The
silicon element is mounted
within a high integrity glass to
metal seal and is fully isolated
from thepressure media by a
hastelloy isolation diaphragm,
electron beam welded to the
front of the glass to metal seal.
Druck’s proprietary low oil volume isolation allows fast
dynamic response plus minimal thermal error . The high output
silicon allows for high overload capability.
Silicon wafers being loaded
into an oxidation furnace.
Applications include general purpose industrial,
refrigeration, oil & gas well head monitoring and
control, compressor performance monitoring
and static pressure measurements in flow
applications.
Surface mount electronics condition the output from the
silicon diaphragm, correct for thermally induced errors, and
configure the output to the required 4-20 mA or 1-5 Vdc
output. Advanced design features built into the electronic
circuitry enable minimum sensor size with the utmost
reliability. The electronics incorporate power supply
regulation, reverse polarity, overvoltage and shor t circuit
protection, coupled with EMC protection components.
The fully encapsulated solid state design ensures integrity of
product under high levels of shock and vibration, with an
ingress protection rating of NEMA 4X, depending on electrical
termination selected.
A fully tested pressure core is stocked in quantity , requiring
only the addition of the electrical termination. Prior to
shipping, the sensor is adjusted to meet the particular
pressure range and units and terminated with the electrical
connection.
Some configurations allow access to zero and span
potentiometers for adjustments during set up and
calibration. Others are fixed and cannot be adjusted.
A range of stainless steel pressure adapters are available to
modify the standard 1/4”NPT female pressure port.
The PTX/PMP 1200 Series is ideally suited to meet the
rigorous demands of the Industrial, Automotive, Aerospace
and Oil & Gas markets. The PMP 1200 is designed for
applications where batteries and/or solar powered systems
require very low power consumption. The PTX/PMP 1240 is
also being qualified for Class I, Div 1 Explosion-proof
service.
Computerized testing
and calibration of
pressure transducers.

Standard specification
SPECIFICATION
Operating Pressure Ranges
0 to 2.5, 5, 10 psig
15, 20, 30, 50, 60, 100, 200, 300, 500, 600,
750 psig or psia.
0 to 1000, 1500, 2000, 3000, 5000, 7500,
10,000, 15,000 psi sg or psia
-15 to +30 psig, 0 to -5 and 0 to -15 psig;
11.5 to 17.0 psia (Barometric)
Please refer to manufacturer for additional
ranges.
Overpressure
The rated pressure can be exceeded by the
following without degrading performance:
4x for ranges to 900 psi g
2x for ranges 1000 to 10,000 psi sg
1.5x for range 15,00 0 psi
Pressure Containment
6x Full Scale for vented gauge
2500 psi for all absolute and sealed gauge
ranges 2000 psi and below
20,000 psi for all absolute and sealed gauge
ranges above 2000 psi
Refer to manufacturer for higher
containment
Pressure Media
Fluids compatible with 316L Stainless Steel
and Hastelloy C276 (NACE compatible)
Supply Voltage
PTX 1200
9 to 30 Vdc
The minimum supply voltage (V
must appear across the transmitter
min
) that
terminals is 9 Vdc and is calculated by:
V
-(0.02xR1)
Where Vs is supply voltage in volts
min=Vs
RL is total loop resistance in ohms
PMP 1200
8-30 Vdc, <2mA
Output
PTX 1200
4-20mA (2-wire configuration)
PMP 1200
1-5 Vdc (less than 2mA)
Zero and Span Offset
±0.5% F.S.
Long term Stability
±0.2% F.S./annum
Combined Non-linearity, Hysteresis
and Repeatability
±0.25% F.S. BSL maximum
Zero and Span Accessability
Models 1210, 1225 and 1265 only
Zero ±5% FS
Span ±5% FS
Temperature Limits
Process: -400 to +2500F
Storage: -500 to +1850F
Compensated: -40 to +1800F
Operating T emperature Range
-400 to +1850F
Temperature Effects
±2% F.S. typically; ±3% FS Thermal Error
Band Maximum over compensated
temperature range. Errors increase pro rata
below 5 psi
Weight
10 oz. Nominal
Pressure Connection
1/4” NPT female to 10,000 psi
3/8” Autoclave for 10,00 0 & 15,0 0 0 psi
ranges
Electrical Connection
1210 DIN connector zero and span
accessible
1225 Cable out with zero and span
accessible
1240 1/2” NPT male conduit fitting
1260 6 pin bayonet connection
1265 6 pin bayonet zero and span
accessible
Ingress Protection
Designed to meet Nema 4X when properly
installed with conduit fitting connection
Voltage Spike Protection
Units will withstand 600V spike test to
ENV 50142 without damage applied
between excitations lines and case
Safety Classifications
UL Intrinsically Safe for Class I, Div 1,
Grps A, B, C and D, Class II, Grps E, F, G
cUL Intrinsically Safe for Class I, Div 1,
OPTIONS
Pressure port:
1/2” NPT female via adapter
1/4” NPT male via adapter
1/2” NPT male, welded adapter
Conduit Fitting:
1/2” NPT female, via adapter
CSA/FM/UL Approval
(state which logo on transducer)
Extra Cable Length
(state length in feet or meters)
NIST Room Temperature 9-pt Calibration
Certificate
Alternate Engineering Units
(equivalent to standard psi ranges)
ORDERING INFORMATION
Please state the following:
(1) Type number PTX or PMP 1200
Model
Code
2 wire, 4-20 mA current output
PTX
PMP 3-wire, 1 to 5 Vdc, Low power
Code
DIN Connector (demountable)
1210
1225
Cable (demountable)
1/2” NPT Male Conduit Fitting
1240
6-pin Bayonet Connector1260
6-pin Bayonet Connector (demountable)
1265
(2) Operating pressure range (gauge,
sealed gauge or absolute)
(3) Options (if required)
Grps A, B, C and D, Class II, Grps E, F, G
Factory Mutual Approvals
Intrinsically Safe for Class I, Div 1,
Grps A, B, C, D for hazardous locations
PTX/PMP 1240 only
Explosion-proof and intrinsically safe
for Class I, Div 1, Grps A, B, C, D;
Class I, Div 2, Grps A, B, C, D;
Class II, Grps E, F, G;
Class III for hazardous locations
Canadian Standards Association (CSA)
Approvals
Intrinsically Safe in hazardous locations
Class I, Div 1, Grps A, B, C, D
PTX/PMP 1240 only
RELATED PRODUCTS
LPM/LPX 800 0 Series Wet/Wet Low Pressure
Differential
PTX 500/60 0 Pressure Transmitters
PTX 1230/1830 Submersible Pressure
Transmitters
MDK-24/MDK-LV Lightning Arresters
DPI 610/615 Portable Pressure Calibrators
DPI 280 Series Indicators
Please refer to the manufacturer for further
information and data sheets.
Continuing development sometimes
necessitates specification changes
without notice.
Explosion-proof and intrinsically safe in
in hazardous locations
Class I, Div 1, Grps A, B, C, D;
Class I, Div 2, Grps A, B, C, D;
Class II, Grps E, F, G;
Druck is an ISO 9001
registered company.
Class III
Venting is provided by a flame-arresting
filter designed to allow the transmitter to
breathe while preventing the ingress of
fluids.
Wiring Details
Function
PTX/PMP
1225/1240
PTX/PMP
1260/1265
PTX/PMP
1210
Wire Color Pin Letter Pin Number
Supply + Red A (1)
Supply – Black B (2)
Output + (PMP White C (3)
versions only)
Case Ground Shield F ( )
ABCDE

PTX/PMP 1200 Series
Industrial Pressure Transmitters
Installation Drawings - dimensions in inches
(1210 demountable)
A full range of alternative
cable and connector
versions are availablerefer to Druck for further
information
Associated Products
(1240 fixed)
(1260 fixed)
Druck manufactures a comprehensive range of pressure
instrumentation to complement the PTX/PMP 1200 Series.
Portable pressure, temperature and electrical field calibrators
allow for local calibration reducing plant down time while
maintaining the quality requirements of ISO 9000. In addition,
the Druck range of calibration equipment is completed by
primary standard deadweight testers.
(1225 demountable)
(1265 demountable)
Left: DPI 610 Field Portable Pressure Calibrator
Center: DPI 145 Multi-Function Digital Meter
Right: DPI 605 Precision Portable Pressure Calibrator
Inset: LP Series of Low Pressure Sensors
Druck Incorporated
4 Dunham Drive
New Fairfield CT 06812
Tel: (203) 746-0400
Fax: (203) 746-2494
E-Mail:usa.sales@druck.com
http://www.druck.com
cd
http://www.pressure.com
The LP Series provides full scale absolute ranges
as low as 0.03 psia and differential ranges as low
as ±0.04 inches H2O wet/wet differential pressure.
For further information and product
data sheets- refer to Druck.
Representative
PDS- A083 9/01