Drager Caleo User manual

D
MEDICAL
Caleo
®
Neonatal incubator
Instructions for Use Software 2.n
255 37 570

How to use the Instructions for Use

The headline... specifies the subject of the main chapter
to help you find your way around rapidly.
The page... contains instructions for use of the unit
in a combination of text and illustrations. The information is translated directly into practical actions showing the user how to use the unit.
The left-hand column... contains text
explaining the unit and guiding the user clearly to its uses through concise instructions in ergonomic sequence.
Dots refer to actions,
1 Numbers refer to the illustration next to the text and to the
sequence of actions in the case of operations consisting of several steps.
The right-hand column... contains illustrations
as a reference to the text and to guide the user in handling the unit itself. The elements mentioned in the text are highlighted, and unnecessary details are omitted. The user is guided by screens confirming the various steps required for each action.
Preparation Before using for the first time
Preparation
Before using for the first time
Check that all packaging materials have been completely
removed (see packing slip in the pack).
Check that the mains power supply voltage matches the
voltage rating specified on the nameplate (see page 123).
Check that the altitude above sea level is correctly set
(see page 73).
1 Screw the extension pole fully into the base frame and
tighten firmly. Check that it is securely held in place.
2 Slide the holder over the pole and secure it at the required
working height with the clamping screw.
3 Set the control unit to the desired working position and
clamp it securely to the holder with the locking screw.
Make sure that the swivel zone is clear.
Mounting the accessory equipment
When fitting accessories, please note:
The accessory equipment must not collide with the unit
when adjusting the tilt angle and height of Caleo
The accessory equipment on the handle side must not
move with the main unit during height adjustment of
®
(page 107).
Caleo
Screw on pole 38 mm/600 (2M 50 691) or alternatively pole 38 mm/310 (2M 50 688) or alternatively pole 25 mm/600 (2M 50 689).
Remove the cover plate from the base frame.
Screw the pole fully into the base frame and tighten firmly.
Check that it is securely held in position.
Observe the maximum loads. Danger of overturning the unit!
Pole 38 mm/600 10 kg Pole 38 mm/310 10 kg Pole 25 mm/600 3 kg
Distance between the loads and the pole: max. 150 mm.
12
®
.
3
2
1
116 37 57 0
228 37 570
14737571
Contents
Contents
For Your Safety and that of Your Patients 4
Intended Use 5
What's what 8
Operating concept 9
Preparation 12
Checking readiness for operation 26
Operation 34
Disinfecting/Cleaning/Sterilising 84
Maintenance intervals 95
Troubleshooting – Error Messages 98
Troubleshooting – Faults 103
Technical Data 106
Description 112
Order list 128
Parts List 131
Index 132

For Your Safety and that of Your Patients

For Your Safety and that of Your Patients

Strictly follow the Instructions for Use

Any use of the apparatus requires full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here.

Maintenance

The apparatus must be inspected and serviced regularly by trained service personnel at six monthly intervals. Repair and general overhaul of the apparatus may only be carried out by trained service personnel. We recommend that a service contract be obtained with DrägerService and that all repairs also be carried out by them. Only authentic Dräger spare parts may be used for maintenance. Observe chapter "Maintenance Intervals".

Accessories

Do not use accessory parts other than those in the order list.

Not for use in areas of explosion hazard

This apparatus is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to occur.

Safe connection with other electrical equipment

Electrical connections to equipment which is not listed in these Instructions for Use should only be made following consultations with the respective manufacturers or an expert.

Liability for proper function or damage

The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator to the extent that the apparatus is serviced or repaired by personnel not employed or authorized by DrägerService or if the apparatus is used in a manner not conforming to its intended use. Dräger cannot be held responsible for damage caused by non-compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of Dräger are likewise not modified by the recommendations given above.
Dräger Medical AG & Co. KGaA

Intended Use

Therapy system providing a controlled supply of warmth, humidity* and O premature babies and sick neonates up to a body weight of 5 kg or a body length of 55 cm (when treating twins, the total body weight is limited to 5 kg).

Used in

clinical environment, where premature babies or neonates need controlled climate parameters.
The unit may only be used by properly trained personnel under the supervision of qualified medical staff familiar with the currently known risks and benefits of using an incubator.
2 enrichment* in the patient capsule for
Intended Use

Possibilities for nursing and therapy

— Heat therapy through control of air temperature or skin
temperature — Humidification —O2 therapy through controlled O2 enrichment — Normal or intensive care via hand ports or a large front flap — Bed with pivoting adjustment for raising and lowering the
head
With monitoring for
— Air temperature — Skin temperature — Relative humidity —O2 concentration — Weight*
* optional equipment feature
What's what
What's what
Contents
What's what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Side view, connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Top view, bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

What's what

What's what

Front view

1 Canopy 2 Hand port 3 Front flap 4 Handle for transport 5 Housing 6 Drawer (2M 50 565) 7 Pedals for height adjustment 8 Height adjustable pillar* / Housing support 9 Connection for water heater (LuerLock) 10 X-ray drawer / removable bed 11 Display 12 Control unit 13 Pillar element
*
*
*
11
10
12
13
1
2 3 4
9
8
5
6

Side view, connections

14 Central alarm indicator 15 Sensor unit, temperature connections 16 Mains power connection 17 On/off switch 18 Fresh air filter (MX 17 015) 19 Connection for O2 for adjustment 20 Water container (2M 50 040) 21 Tubing grommet (2M 50 385) 22 Side flap 23 Tubing port (2M 50 412) 24 Feeding grommet, hood (2M 50 352) 25 Trolley castor with direction lock 26 Trolley castors, brake with breaks
*
*
22 21
20
19 18
77
111 37 570
142423
15
16 17
* optional equipment feature
2526 26 26
112 37 570

Top view, bed

What's what
Operating concept
1 Spirit levels 2 Hot air duct
Operating concept

Control unit

Hard Keys

these permanently defined keys enable the user to select various functions of Caleo®:
3 Scales 4 Bed tilt 5 Menu selection/configuration 6 Changeover key: air/skin temperature control 7 Trend display 8 Suppress alarm tone 9 Lock key pad function 10 Rotary knob
*
1
1
1
D
14
15 16
2
2
223 37 570
5
6 7
11 12
Menu
Air
Skin
8 9
Visual signals indicate alarm situations 11 Red alarm LED
12 Yellow alarm LED 13 Power failure alarm

Soft Keys

These keys, with variable functions defined by different labels on the screen, guide the user through the specific routines of the unit, from preparing for use to shutting down the unit. Depending on the current menu, different soft keys with varying functions and labels are activated. Only the soft keys required for the current menu actually appear. This precaution keeps the display clear, preventing any confusion for the user. When a soft key is pressed, its function is activated and the relevant menu is displayed on the screen. In the standard screen, the soft key labels are as follows:
14 Air/skin temperature 15 Humidity 16 O2
* optional equipment feature
*
*
10
3
4
13
040 37 570
Operating concept

Rotary knob

A single rotary knob is used to select and set parameters.
Turn knob = select
Press knob = confirm
D
193 37 570

Screen

By default, the measured values are displayed as numeric values (standard screen).
— Set values and actual measured
values for air temperature/skin temperature — Set values and actual measured values for relative humidity — Set values and actual measured values for
O2 concentration — Alarms and warnings
The screen display can also include a trend graph.
*
*
000 37 571
* optional equipment feature
10
Preparation
Preparation
Contents
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Before using for the first time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Mounting the accessory equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Doors, ports and bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Checking readiness for operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Before using for the first time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Before each use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
11
Preparation Before using for the first time

Preparation

Before using for the first time
Check that all packaging materials have been completely
removed (see packing slip in the pack).
Check that the mains power supply voltage matches the
voltage rating specified on the nameplate (see page 125).
Check that the altitude above sea level is correctly set
(see page 73). 1 Screw the extension pole fully into the base frame and
tighten firmly. Check that it is securely held in place. 2 Slide the holder over the pole and secure it at the required
working height with the clamping screw. 3 Set the control unit to the desired working position and
clamp it securely to the holder with the locking screw.
Make sure that the swivel zone is clear.
3
2

Mounting the accessory equipment

When fitting accessories, please note:
The accessory equipment must not collide with the unit
when adjusting the tilt angle and height of Caleo®.
The accessory equipment on the handle side must not
move with the main unit during height adjustment of Caleo® (page 109).
Screw on pole 38 mm/600 (2M 50 691) or alternatively pole 38 mm/310 (2M 50 688) or alternatively pole 25 mm/600 (2M 50 689).
Remove the cover plate from the base frame.
Screw the pole fully into the base frame and tighten firmly.
Check that it is securely held in position.
1
116 37 570

Observe the maximum loads. Danger of overturning the unit!

Pole 38 mm/600 10 kg Pole 38 mm/310 10 kg Pole 25 mm/600 3 kg
Distance between the loads and the pole: max. 150 mm.
12
228 37 570

Moving the control unit to the opposite side

for pole, 38 mm
1 Remove the control unit = loosen the clamping screw. 2 Remove the holder = loosen the clamping screw.
Move the holder to the other pole and set to the desired
working height.
2 Fix the holder in position = tighten the clamping screw. 1 Secure the control unit to the holder = tighten the clamping
screw.
Mounting the control unit on the handle side Only technically qualified personnel may move the control
unit from the wall side to handle side or vice versa.
Preparation
Mounting the accessory equipment
1
2
024 37 570
Fit the basic pole (2M 50 680), see page 15.
Screw in the extension pole 38/600 (2M 50 691) or
pole 38/310 (2M 50 688).
Refit the control unit as specified in the Assembly
Instructions.
The cable to the control unit must not be removed from
the cable guides on the basic pole.
Make sure there is sufficient space to swivel and tilt
the unit.

Infusion support (2M 21 514)

for pole, 38 mm Max. load per hook 3 kg
256 37 570010 37 570
Push the fixing claw onto the pole.
Push the infusion holder into the fixing and secure it firmly
with the clamping screw.
13
Preparation Mounting the accessory equipment

Swivel table (2M 21 186)

for pole, 38 mm For small parts, max. load 3 kg
Place the clamp of the swivel table on the stand pillar and
tighten the clamping screw.
Make sure that the swivel zone is clear.

Compact rail (2M 85 337)

for pole, 38 mm Max. load 5 kg.
This rail must only be mounted by qualified technical staff.
020 37 570101 37 570
For holding accessories, e.g. —O
2 monitor
Adjust the height of the compact rail to the required height
of the mounted accessory. 1 Fit the compact rail to the pole = push the compact rail over
the pole and
2 fasten with the screws.

Bronchial aspiration system (2M 85 125)

Follow the separate Instructions for Use of the bronchial aspirator.
Fix the bronchial aspirator holder to the standard rail on the
wall side or handle side.
Tighten the clamping lever.
Establish the hose connections.
1
2
14
230 37 570

Basic pole (2M 50 680)

Maximum load 10 kg
This pole must only be mounted by qualified technical staff!
Mount as specified in the Installation Instructions.
For fixing accessories, e.g. — additional pole extensions, see page 15.
— swivel table (2M 21 186), see page 14. — monitor support plate (2M 50 085), see page 16.
Distance of the loads to the pole, max. 150 mm.
Preparation
Mounting the accessory equipment

Tray 3020 (M 24 678)

Maximum load must not exceed 2 kg!
Hang the tray from the standard rail on the wall or handle
side and secure in position.

Pole extensions

The following poles can be fixed to the base pole as extension: — Pole 38 mm/600 (2M 50 691) or — Pole 38 mm/310 (2M 50 688) or — Pole 25 mm/600 (2M 50 689).
Screw the pole into the base pole as far as it will go and
tighten securely. Make sure that it is securely held in place.
019 37 570021 37 570229 37 570
Observe the maximum loads: Pole 38 mm/600 5 kg
Pole 38 mm/310 5 kg Pole 25 mm/600 3 kg
Distance between the loads and the pole: max. 150 mm
15
Preparation Mounting the accessory equipment

Notebook holder (2M 22 171)

Maximum load must not exceed 3 kg!
1 Fix the holder to the handle rail of the Caleo
®
.
2 Align the holder horizontally with the clamping screw.
Check that the holder is securely fixed in position and that
the swivelling mechanism is functioning correctly by turning
and tilting the support plate.

Monitor shelf (2M 50 085)

The monitor tray can be mounted on the wall side and/or the handle side. Maximum load must not exceed 20 kg! Shelf for monitor and ventilation equipment.
This monitor shelf must only be mounted by qualified technical staff!
To fix the monitor shelf: — on the wall side, fit a second 38 mm pole (see page 12),
— on the handle side, fit a second base pole (see page 15)
with a 38 mm pole extension (see page 15).
Mount the shelf as specified in the Installation Instructions.
Fit the monitor holder = slide the shelf over both 38 mm
poles and 3 fix in place with the screws.
1
2
022 37 570
Do not exceed the maximum installation height.

Hose holder for ventilation hoses (84 11 075)

Open the front flap.
Raise the bed and pull it out of the incubator.
Push the mattress slightly to one side.
Place the hose holder in the mounting hole in the bed and
fasten from underneath with the locking screw.
Replace the bed in the incubator and close the front flap.
The hose holder can be fixed to any of the four corners of the bed.
4 Clip the ventilation hoses and cables into the clips at the
end of the ventilation hose holder.
max. 20 cm
33
max. 100 cm
023 37 570
44
008 37 570
16
Preparation
Mounting the accessory equipment

O2 enrichment with O2 control

*
1 Screw the O2 connection hose to the underside of the
incubator. Make sure that it is tightened securely.
Connect the probe to the outlet of the central O2 supply
pipeline in the "park" position (see separate Instructions for Use).
The permissible gas pressure is between 300 and 500 kPa.
O2 monitor
To monitor the O2 concentration, use an O2 monitor that has alarm limits:
Fix the O2 monitor to the handle rail with the holder.
Place the sensor capsule in Caleo
Route the sensor cable through one of the flexible tubing
®
.
ports. Where applicable, push the sensor plug into the socket of the O2 monitor (e.g. Oxydig or MiniOx 3000) until it audibly clicks into place (see separate Instructions for Use of O2 monitor).
1
011 37 570

Vacuum mattress (2M 17 909)

The contour of the vacuum mattress can be altered as required and is then maintained after air evacuation of the mattress. Extreme positions can therefore be obtained for special applications. The mattress can remain in the incubator.
Open the front flap.
Insert and preform the vacuum mattress.
Place the patient on the mattress and finally adjust the
mattress to the desired shape.
Connect the vacuum mattress to the hose of the suction
system.
2 Open the valve and evacuate the vacuum mattress. 2 Close the valve and disconnect the hose.
Close the front flap.

Fitting the drawer (2M 50 565)

Maximum load must not exceed 7 kg.
The drawer must only be mounted by qualified technical staff!
Mount as specified in the Installation Instructions.
Fit the drawer = slide the drawer box into the groove in the
base frame.
028 37 570213 37 570
2
* optional equipment feature
118 37 570
17
Preparation Doors, ports and bed

Doors, ports and bed

Hand ports

To open the hand ports: 1 Press the catch: the relevant hand port swings open.
To close the hand ports:
Push the hand ports back into place until the catch
engages.
Front flap
To open the front flap: 2 Turn the two knobs inwards to the vertical position.
The red latch becomes visible.
11
014 29 529
Lower the front flap until it hangs down vertically towards
the floor.
Avoid pinching or jamming hoses and cables in the removable double wall!
2
2
015 37 570
016 37 570091 37 570
To close the front flap:
Raise the front flap and press into position,
3 Turn the two knobs outwards to the horizontal position until
you feel them click into place.
Make sure that both knobs are engaged in position! The red latches must no longer be visible!
18
3
3
Side flap
The side flap is opened and closed in the same way as the
front flap (see page 18).

Canopy

To open the canopy: 1 Grasp the handle on the canopy
and
2 lift open (approx. 60o).
Preparation
Doors, ports and bed
231 37 570104 37 570
2
Do not tilt the canopy forwards!
3 Raise the side support prop, and lower the canopy until the
prop is fixed in the slot of the canopy.
To close the canopy:
1 Grasp the handle on the canopy and raise it slightly. 3 Fold down the prop
and
close the canopy.
The canopy can be opened from both sides.
To remove the canopy: 4 Grasp the handles on the sides of the canopy with both hands.
5 Lift the canopy horizontally off the pillar elements.
1
3
196 37 570
4
To replace the canopy: 5 Replace the canopy horizontally, so that the guide pins fit
into the holes in the pillar element.
Take care with the sensor unit!
4
55
093 37 570
19
Preparation Doors, ports and bed

Double wall

*
To remove the double wall:
Remove the canopy and lay it upside down (on a soft,
non-abrasive surface).
1 Squeeze the double wall inwards slightly, and 2 release the oblong holes of the double wall from the
retaining lugs in the canopy.
Pull the double wall out of the canopy.
To install the double wall:
Place the canopy upside down (on a soft, non-abrasive
surface).
Insert the double wall in the canopy.
3 On one longitudinal side, fit the retaining lugs of the canopy
into the oblong holes of the double wall.
Squeeze the double wall inwards slightly and fit the
retaining lugs on the second longitudinal side of the canopy
into the oblong holes of the double wall.
Turn the canopy the right way up again.
The retaining lugs of the canopy must remain seated in the
oblong holes.
2
2
1
148 37 570
3

Removing the bed

Maximum load must not exceed 5 kg
Open the front flap and fold it down.
Pull the bed out:
4 Set both knobs horizontally to the position marked , 5 Pull the bed out towards the front as far as it will go, holding
it by the recessed handle or by the knobs.
When the bed is pulled out, the patient must be monitored constantly to prevent the risk of a fall.
Do not lean or rest any weight on the bed when it is pulled out.
After completing the care procedure, push the bed back in
until it engages. Turn the knobs to the position and
close the front flap.
Push the bed all the way in! Otherwise the hot air duct will be interrupted, and the patient may be warmed or cooled excessively!
4 4
5
252 37 570
187 37 570
* optional equipment feature
20

Using the X-ray drawer

The X-ray drawer can be pulled out when the front flap is either open or closed.
To open the X-ray drawer:
1 Set both knobs vertically to the position marked , 2 pull the drawer out by the recessed handle or the knobs.
Insert or remove the X-ray cassette. Recesses are provided
in the X-ray drawer for positioning.
Do not use the X-ray drawer as a writing support or bed for the patient.
To close the X-ray drawer: 2 push the drawer inwards under the bed until it tangibly
clicks into place.
Push the X-ray drawer in fully! Otherwise the hot air duct will be interrupted, and the patient may be warmed or cooled excessively!
Preparation
Doors, ports and bed
1
2
1
017 37 570
Sealed through-holes 3 Tubing ports (2M 50 412)
4 Tubing grommets (2M 50 385) 5 Feeding grommet, hood (2M 50 352)
Route the cables or hoses through the flexible grommets
and sleeves.
To route ventilation hoses and cables through the Caleo®, use the ventilation hose holder (page 16).

Drawer (2M 50 565)

Drawer for storing items required for nursing or treatment. The drawer is accessible from both sides.
Open the drawer: 6 grasp the drawer by the handle and pull it out as far as it
will go.
Place the required material in the drawer.
5
3
44
095 37 570
Close the drawer:
6 push the drawer back in by the handle.
6
013 37 570
21
Preparation Doors, ports and bed

Trolley (base frame) with variable height adjustment

*
To use the height adjustment facility:
Switch on Caleo
®
(see page 37).
1 Press the left pedal – Caleo® is lowered. 2 Press the right pedal – Caleo® is raised.
Adjust to a comfortable working height.
When the height stops changing, the frame is at its end position. Release the pedal.
Hoses and cables must be long enough so that they do not kink, tear or become squashed!
Do not store anything under the drawer!
The height adjustment and tilt angle adjustment cannot be operated at the same time!
The height adjustment can only be operated for max. 6 minutes during one hour.

Trolley castor with direction lock

Activate the lock lever: the castor will be locked in the
lengthways direction (longitudinal axis of device).
1 2
009 37 570
The castor with direction lock does not have a brake. When parking the unit, apply the brakes of all braked castors.
The castor with direction lock is visually distinct from the castors with brakes.
257 37 570
* optional equipment feature
22

Tilting the bed

Preparation
Doors, ports and bed
To tilt the bed:
Switch on Caleo
®
(see page 37).
1 Press button = the bed will be lowered on the sensor
unit side.
2 Press button = the bed will be raised on the sensor unit
side.
Adjust the bed to the required tilt angle.
The end position is reached when the tilt angle stops changing. Release the button.
®
The entire housing of the Caleo
incubator is tilted.
Hoses and cables must be carefully routed so that they do not kink, tear or become squashed!
The height adjustment and tilt angle adjustment cannot be operated at the same time!
D
1
2
214 37 570
Do not reach in between the housing and housing support during adjustment. Danger of injury!

Setting the bed to the horizontal position

Caleo
®
must be switched on (see page 37).
1 Press button = the bed will be raised on the control unit
side.
2 Press button = the bed will be lowered on the control unit
side.
The spirit levels show whether the bed is horizontal. 3 Spirit levels for the horizontal position of Caleo
®
in the
transverse axis.
4 Spirit level for the horizontal position of Caleo® in the
longitudinal axis. When using the integrated scales (optional, see "Weighing scale" on page 78), make sure that the unit is on a flat floor before setting.
086 37 570
3
4
3
242 37 570
23
Preparation Doors, ports and bed
Using humidifier systems
[ Only use distilled and sterilised water (Aqua dest.)!
Water container (2M 50 040):
Disinfect hands.
1 Open the water container = lift up the cap.
Fill the water container with sterilised Aqua dest.
Do not use any additives! Capacity: 2.8 L (note level marks)
1 Close water container = push down the cap.
Prepare a fresh connection tube (MX 17 018).
2 Close the clamp on the connection tube. 3 Pierce the silicone nozzle of the water container with the pin
of the connection tube. 2 Open the clamp on the connection tube.
Bleed the connection tube (let distilled water drain off).
2 Close the clamp on the connection tube.
1
3 2
097 37 570
4 Connect the LuerLock connection to the water connection
pipe. 5 Open the clamp on the connection tube.
Switch on the humidification module on Caleo
®
and set the
humidity value (page 51).
Water bag: Use only sealed bags with sterilised Aqua dest. (destilled
water)! Do not use any additives! Must not be confused with infusion solutions!
Disinfect hands.
Prepare a new connection tube (MX 17 018) and a water
bag with sterilised Aqua dest. 6 Close the clamp on the connection tube.
Insert the pin of the connection tube into the connector of
the water bag. 6 Open the clamp on the connection tube.
Bleed the connection tube (let distilled water drain off).
6 Close the clamp on the connection tube. 7 Connect the LuerLock connection to the water connection
pipe. 6 Open the clamp on the connection tube.
Switch on the humidification module on Caleo
®
and set the
humidity value (page 51).
Replacing the water bag: Distilled water bag empty = the water shortage alarm on
®
Caleo
is triggered.
Disinfect hands.
Close the clamp on the connection tube.
Replace the water bag and reopen the clamp.
4
5
185 37 570
6
7
098 37 570
24

Integrated socket strip

The integrated socket strip can be used to connect — infusion pumps and — SpO
2 measuring equipment.
Connections may only be made by qualified technical staff.
Caleo does not monitor the power supply to external devices.
[ Do not exceed the maximum power input of the
connected accessories (all 4 sockets together: max. 2A). The maximum permissible total leakage current must not be exceeded. For the leakage current of Caleo® without socket strip, see "Technical Data" on page 106.
Preparation
Doors, ports and bed
099 37 570
25
Checking readiness for operation Before using for the first time

Checking readiness for operation

Before using for the first time
Check that the mains voltage supply corresponds to the
voltage rating on the nameplate (see page 125).
Check that the altitude above sea level is correctly set
(see page 73).

Before each use

Check that the equipment has been disinfected and
cleaned in accordance with the conditions laid down by the
hospital (see "Disinfecting/Cleaning/Sterilising" on
page 84).
Check that an adequate gas supply is available for the
equipment to be used.
Check that the required accessories and therapy
equipment are available and in perfect condition.
Only use parts that have been stripped down and sterilised.
Check readiness for operation in accordance with the
relevant Instructions for Use.
Check that there are no cracks or sharp, chipped edges on
the incubator canopy.
Check that the hinges and catches on the canopy are in
proper working order.
Check that the cables and hoses have been routed
correctly and safely.
If the control unit is mounted on the handle side, check that
the cable to the control unit is routed through the cable
guides on the basic pole.
Check that there is sufficient space for adjusting the tilt and
height.
Connect to the mains supply.
Check that the slits of the sensor unit are not fouled.
Do not use multiple-plug adapters for the Caleo® power supply!
The patient leakage current may rise above the permitted limits if the protective earth conductor fails when equipment is connected to the socket strip. The risk of electric shocks cannot be excluded in such cases.
Before using the unit, make sure that the following tests have been performed:
Disinfect hands before each test!
Check that the hand ports are secure 1 Open hand port = press catch
Close hand port until the catch engages.
Pull the edge of the hand port – it must not open.
If the hand port fails to remain engaged:
Call DrägerService.
26
11
014 29 529
Check that the front flap is secure
Open the front flap and fold it down (see page 18).
1 Raise and press the flap closed and turn the two knobs
outwards until they tangibly engage in the horizontal position.
Make sure that both knobs are engaged! The red latches must no longer be visible!
If the front flap fails to remain engaged or if the red latches are visible:
Call DrägerService.
Check that the side flap is secure
Open the side flap and fold it down (page 19).
2 Fold up the side flap and press it closed. Turn the two
knobs outwards until they tangibly engage in the horizontal position.
Checking readiness for operation
Before each use
1
1
090 37 570
Make sure that both knobs are engaged! The red latches must no longer be visible.
If the side flap fails to remain engaged or if the red latches are visible:
Call DrägerService.
Check that the canopy is secure 3 Grasp the handle and open the canopy.
4 Lift open the flap (approx. 60o).
5 Raise the side prop.
Lower the canopy until the prop is secured in the slot of the
canopy.
Repeat the test on the other side.
22
189 37 570
4
3
224 37 570
If the canopy fails to remain open:
Call DrägerService.
5
225 37 570
27
Checking readiness for operation Before each use
1 Grasp the handles on the sides of the canopy with both
hands.
2 Lift the canopy horizontally off the pillar elements.
1
If the canopy holders are damaged:
Call DrägerService.

Check that the double wall is securely in place

All 4 retaining lugs in the canopy must be seated in the
oblong holes of the double wall.
If the double wall or the retaining lugs in the canopy are damaged:
Call DrägerService.

Check that the trough is secure

Remove the canopy.
Remove the mattress.
Remove the bed.
3 Check the catches of the trough.
Place the bed on the trough.
Place the mattress on the bed.
Refit the canopy.
1
22
226 37 570
135 37 570135 37 570
3
3
If the catches for the trough are damaged:
Call DrägerService.

Remove the bed

Open the front flap and fold it down.
4 Set both knobs horizontally to the position marked , 5 Grasp the bed by the recessed handle or by the knobs and
pull it out towards the front as far as it will go.
5 Push the bed back until it clicks into place, 4 turn the knobs to the position.
Close the front flap.
If the bed cannot be pulled out or pushed in or if the knobs are damaged:
Call DrägerService.
If the bed is not fully pushed in, the hot air duct will be interrupted, causing the control system to malfunction! The result may be excessive cooling or overheating of the patient!
28
4
5
4
018 37 570
Activate the self-test, check audible warning tone 1 Switch on the unit = press the on/off switch until it clicks
into position.
Checking readiness for operation
Before each use
1
During the self-test, the functions of the machine are checked.
The audible signal, alarm beep sequence, screen displays and LEDs must be checked by the user.
— An audible warning signal and an alarm beep sequence are
sounded.
If the warning signal or alarm beeps are not sounded,
Call DrägerService.
— The screen and LEDs initially go dark and are then lit. If individual pixels fail to light up or screen images are burnt, or if the LEDs fail to light up,
Call DrägerService.
— The opening screen is displayed. If the opening screen is not displayed,
Call DrägerService.
110 37 570
The unit is switched on.
042 37 571
29
Checking readiness for operation Before each use
Check the bed tilting mechanism 1 Tilt the bed.
During the tilting process, the entire housing of the Caleo must move uniformly. If not:
Call DrägerService.
Do not reach in between the housing and the housing support while the unit is moving! Risk of injury!
2 Return the bed to the horizontal position (see page 23).
The spirit levels in the bed show whether the bed is
horizontal. This check is especially important when using
the built-in weighing scales (see page 78)!
®
D
1
2
132 37 570
If the spirit levels are damaged:
Call DrägerService.

Check the height adjustment mechanism

Operate both foot pedals in succession to raise and lower
*
the Caleo® (see page 22).
Then adjust to a comfortable working height.
When adjusting the height, the entire housing of the Caleo® must rise or descend uniformly. If not:
Call DrägerService.

Check the power failure alarm

Disconnect the unit from the mains.
3 The power failure LED must flash.
253 37 570
089 37 570133 37 570
D
An intermittent tone must start. The volume of this tone must remain constant for at least 30 seconds.
If the volume decreases too soon:
Leave the incubator connected to the mains and switched
on for 24 hours to recharge the NiCd battery.
Repeat the test.
If the volume again decreases too soon:
Call DrägerService.
* optional equipment feature
30
3
Check the fresh air filter
[ Use the original Caleo
®
air filter (MX 17 015) only!
Risk of fire!
Tilt the unit to remove the fresh air filter more easily
(see page 23).
Checking readiness for operation
Before each use
211 37 570124 37 570
1 Pull the holder for the fresh air filter downwards and
remove.
2 Check that the seal on the filter box is in place.
Fitting the filter 3 Separate the filter holder and
4 filter frame at the snap-in fastening. 5 Insert the filter.
Close the filter holder and filter frame.
Mark the next due date for replacement of the new filter
(2 months after installation) on the label and
6 Stick the label to the filter holder.
1
2
123 37 570
3
4
5
If a filter is already installed:
Check the condition of the filter.
Check the due date for replacement. The label is on the
filter holder.
If the filter is fouled, damaged or moist, or if the due date for replacement has been reached:
Discard the old filter.
Fit a new filter.
6
122 37 570
31
Checking readiness for operation Before each use
Before operating the unit with skin temperature sensors, it is important that the sensors must be tested before they can be used on the unit (see "Using skin temperature control" on page 42).
The incubator is ready for operation when all checks have been carried out successfully.
32
Operation
Operation
Contents
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Switching on Caleo
Using air temperature control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Using skin temperature control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Changing over between air/skin temperature control . . . . . . . . . . . . . . . . . . . . . . . . 43
Using skin temperature control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Using humidity control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Setting AUTO humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Manually adjusting the set value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Using O2 control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Adjusting the set value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Selecting menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Kangaroo mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Trend display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Trend analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Trend selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Cleaning mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Activating configuration mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Language/date/time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Setting system parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Setting alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
O2 sensor information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Viewing software information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Lock key pad functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Alarm suppression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Weighing scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Weighing without tare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Ending operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Switch off incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
33
Operation Precautions

Operation

Precautions

Before each use, check that the unit is ready for operation (see page 26).
Make sure that all hoses and cables are routed correctly and safely without obstruction! Otherwise: Risk of extubation! Danger of disconnection!
Lively patients must be observed with particular care.
When the canopy, front flap, hand ports are opened or when the bed has been pulled out or when the tubing grommets are removed, the patient must be observed constantly to make sure that he or she cannot fall out of the incubator.
Do not lean or apply weight on the bed when it has been pulled out. Maximum load 5 kg
Do not use the X-ray drawer as writing support or bed for the patient.
Allow time for the incubator to warm up before use (page 107).
Additional external heat sources, such as sunlight, heat lamps, spotlamps and electric cushions should be avoided! They cause the temperature inside the Caleo® to rise in an uncontrolled manner.
The baby’s core temperature must be regularly monitored with an independent thermometer.
The castor with direction lock does not have a brake. When parking the unit, apply the brakes of all braked castors.
When treating larger babies, the higher heater output may cause the air temperature in the Caleo this case the double wall should be removed.
The room temperature must be at least 3 oC lower than the air temperature in the Caleo® incubator.
When treating twins, pay attention to the risk of cross­infection!
Fire hazard from supply of O2
— No naked lights or smoking. Textiles, oil and plastics can
very easily catch fire and burn rapidly in an O2 enriched atmosphere!
— All fittings and seals in contact with oxygen must be kept
free of oil and grease! — Open valves on O2 cylinders slowly! — Do not use Caleo® in the presence of flammable
anaesthetic gases or disinfecting agents. Risk of explosion! — Do not use or keep flammable liquids, such as alcohol,
ether or acetone, in Caleo®! — Do not use any electrical equipment in the patient capsule,
unless this equipment is expressly designed for use in
explosion-hazard areas. — Use the original Caleo® air filter (MX 17 015) only!
Note the physiological risks from O2.
®
incubator to rise. In
The conclusions to be drawn from the measured skin temperature are the responsibility of the attending physicians.
Skin temperature control mode must not be used on babies who are in shock or who have a high temperature!
Skin temperature control mode must not be used on twins, since the Caleo® controls only the temperature for one baby. Risk of hypothermia or overheating! For twins, the air temperature control mode must be used.
Make sure that you do not confuse the positions for the skin temperature sensors. Skin temperature control is regulated by the yellow skin temperature sensor (T1). If this sensor is stuck to the wrong part of the body, the patient may be excessively warmed!
Do not place any cloths over the hot air duct. The temperature control system would be disrupted, causing a risk of burning if the air from the hot air duct is channelled directly to the patient.
34
During O2 therapy, monitor the O2 concentration with a separate O2 monitor.
The air in the incubator may only be enriched with O2 when prescribed by a doctor.
O2 enrichment must be controlled on the basis of the arterially measured oxygen partial pressure in the patient’s blood. Otherwise, there is a danger of hyperoxaemia (damage to the eyes) or hypoxaemia (brain damage).
Medicaments and similar substances must not be atomised in the patient capsule. Correct functioning of the incubator may be impaired by the precipitation of atomised substances.
Do not cover the sensor unit and do not hang anything from the slits in the sensor unit. Keep the slits of the sensor unit free of impurities.
O
In the event of a power failure, the C capsule may rise due to an insufficient supply of fresh air. Pay attention to the risk of C
O
2 poisoning!
2 level in the patient
Operation
Precautions
The central alarm light can be deactivated (see page 124). Always check the alarm LEDs on the control panel and the associated alarm beep signals.
Front flaps
When closing the front flaps, make sure that the patient is not lying in the closing path.
The front flaps are not properly shut until the red latches are no longer visible.
When opening and closing the front flaps, make sure that the hoses and cables are not caught in the moving double wall!
Side flaps
When opening and closing the side flaps, make sure that the hoses and cables are routed safely and clear of any obstructions!
The side flaps are not properly shut until the red latches are no longer visible.
Canopy
The canopy must not be used as a shelf for laying clothing, instruments etc. Before moving the canopy, make sure that nothing has been laid on top of it. When fitting and removing the canopy cover hold it firmly in your hand. The canopy installation catch must engage correctly. Do not lift the canopy when engaged. Do not tilt the canopy forwards. When closed, make sure that the canopy cover sits firmly in place!
When fitting accessories, please note:
The accessory equipment must not collide with the unit
when adjusting the tilt angle and height of Caleo®.
The accessory equipment on the handle side must not
move with the main unit during height adjustment of Caleo® (page 109).
Observe the maximum loads. Danger of overturning the unit!
Kangaroo mode
The patient’s core temperature must be monitored constantly, because it is outside the controlled climate. Particular attention must be paid to critical patients vital parameters.
When using Phototherapy
Absorption of light through the baby’s skin will supply heat, which may increase the baby’s core temperature.
For this reason: — Roughly 15 minutes before starting phototherapy,
decrease the air temperature setting for the incubator air by about 2
o
C. — Decrease the set value for humidity. — Room temperature must be at least 3 oC lower than the air
temperature in Caleo®. This value applies to Dräger Type 4000 phototherapy units. When using other phototherapy units, especially units without a built-in fan, the temperature in the Caleo® incubator may rise even more sharply.
— Phototherapy may only be used with a stand!
Double wall
To fit and remove the double wall, remove the canopy from Caleo®.
Tilting/height adjustment
Ensure that the cables and hoses are routed correctly and safely without obstruction. Hoses and/or cables risk being trapped when tilting the Caleo® and when opening and closing the front flap.
Control unit
Do not mechanically load the cable by e.g. hanging appliances from it. If the control unit is mounted on the handle side, never remove the cable from the cable guides on the basic pole. Do not place any liquid containers above the control unit.
The core temperature of the baby must be monitored with particular care during phototherapy!
The supply of fluid to the baby must be increased, e.g. by parentral infusion, to compensate for the increased loss of water during phototherapy.
The phototherapy lamp and Caleo® canopy must not be covered with cloths, aluminium foil or other materials in
order to boost the phototherapeutic effect – risk of heat build-up, as the incubator cannot be adequately cooled with ambient air.
Danger of overheating the patient.
In-house transport
An even floor surface must be ensured when moving the Caleo® incubator. The incubator must not be used outside the hospital building, as the castors may be damaged or dislodged.
®
The Caleo
may only be moved when empty.
35
Operation Precautions
High noise levels
Excessive noise levels that can disturb the patient may be caused by:
— using O
2 head boxes and delivering pressurised gases,
— wear on the bearings of the fan motor, — placing objects on the canopy.
Observe the specified maintenance intervals – see
page 95.
Cleaning mode may only be used when Caleo® is empty. After use, allow Caleo® to cool down before dismantling.
Risk of burning when touching the heater!
Electrical safety
Do not use multiway adapters to connect the power supply of the Caleo®!
The patient leakage current may rise above the permitted limits if the protective earth conductor fails while equipment is connected to the socket strip. In such cases, the risk of electric shock cannot be excluded.
Only use electromedical ancillary equipment conforming to IEC / EN 60601-1.
When using the integrated socket strip: take into account the total leakage current and total current consumption! See "Technical Data" on page 106.
Caleo® does not monitor the power supply to external devices.
Mobile telephones must not be used within 10 metres of the incubator. Mobile telephones can interfere with the functioning of electromedical equipment and therefore endanger the patient!
* Dräger medical equipment conforms to the interference immunity
*
requirements laid down in product-specific standards or in EN 60601-1-2 (IEC 60601-1-2). However, depending on the design of mobile phone and the application situation, field strengths exceeding the values laid down in the specified standards may be generated in the immediate vicinity of mobile phones, thereby causing interference and malfunctions.
36

Switching on Caleo®

Connect the unit to the mains.
1 Switch on the unit = press the on/off switch until it clicks
into place.
Operation
Switching on Caleo
1
®
An audible signal is emitted.
— The opening screen is displayed.
The incubator performs a self-test. During the self-test, the functions of the machine are checked. The audible signal and screen displays must be checked by the user, see page 29.
— After the self-test, the standard screen for air temperature
control is displayed.
— The currently activated function is always highlighted by a
light background.
110 37 570
042 37 571
With air temperature control, the unit takes 20 minutes to warm up. During this period the alarm for "Air temp. deviation above 1.5 oC" is suppressed.
043 37 571
37
Operation Using air temperature control

Using air temperature control

The baby’s core temperature must be measured
regularly with an independent thermometer!
Do not leave the canopy open for any length of time, otherwise the air temperature in the Caleo® will drop.
Adjusting the set value for air temperature
Standard set value range 28 Extended set value range 37.1
Default setting 33
o
C to 37 oC
o
C to 39 oC
o
20
C to 27.9 oC
o
C
1 Adjust the set value = press the button.
— The current actual measured value and the desired set
value are displayed on the screen both as bar graphs and as numerical values.
— The message »set value with rotary knob« appears at the
top of the screen.
D
1
2
004 37 570
2 Increase the set value = turn the rotary knob clockwise. 2 Decrease the set value = turn the rotary knob counter-
clockwise.
2 Confirm the new setting = press the rotary knob.
If you do not wish to change the settings:
press = the new settings are cancelled
The display returns to the standard screen. The previous set value is retained. Or — Wait for 7 seconds: Caleo
®
emits 4 short beeps to prompt the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
38
312 37 571
[ When using the extended set value range, the baby's
core temperature must be continuously monitored.
If the specified set value exceeds the standard set value range — The following message is displayed at the top of the screen
»confirm extended range with rotary knob«.
Confirm the extended range = press the rotary knob.
Continue increasing the set value = turn the rotary knob
clockwise.
Operation
Using air temperature control
— The advisory message »[ >37.0 oC« appears on the
screen.
— The following message appears at the top of the screen:
»set value with rotary knob«.
Confirm the set value = press the rotary knob.
If the specified set value is below the standard range — The following message is displayed at the top of the screen:
»confirm extended range with rotary knob«.
Confirm the extended range = press the rotary knob.
Continue decreasing the set value = turn the rotary knob
counter-clockwise.
045 37 571150 37 571046 37 571
39
Operation Using air temperature control
— The advisory message »[ <28.0 oC« appears on the
screen.
— The following message appears at the top of the screen
»set value with rotary knob«.
Confirm the set value = press the rotary knob.
151 37 571313 37 571030 37 570
— The display returns to the standard screen. The measured
values are displayed.
— The set value and»set: <28.0 [« are displayed alternately.
Reducing the air temperature in Caleo®
The cooling rate depends on the incubator design and can be increased by: — removing the double wall — reducing the outside temperature (if possible) — reducing the set humidity value. The cooling rate is not accelerated by: reducing the air temperature setting below the actually desired value.
In urgent cases: open the canopy, front flap or hand ports.
In this case, the baby must be monitored constantly to ensure that he or she cannot fall out of the incubator.
If you do not wish to confirm the new set value: 1 Cancel adjustment of the set value = press button. The screen returns to the standard display mode. The previous set value is retained. Or
®
— Wait for 7 seconds: Caleo
emits 4 short beeps to prompt the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
40
D
1

Alarms

Alarm limits can be changed in the configuration (see page 71).
Operation
Using air temperature control
Example: if the deviation between the set and measured air temperature exceeds 1.5
oC*
:
— The screen displays the warning message
»Air temp. deviation above 1.5 oC «,
— The alarm tone sequence (3 beeps) is sounded,
1 The central alarm indicator lights up**, 2 The measured value flashes, 3 The yellow alarm LED flashes.
The intermittent alarm tone sequence can be muted for 15 minutes. 4 Suppress intermittent alarm tone = press button or 5 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted,
1 The central alarm indicator goes out, 2 The measured value continues to flash, 3 The yellow alarm LED continues to flash.
When the measured value returns within the range ±1.5 oC: — The warning message disappears, — The intermittent alarm tone is muted,
1 The central alarm indicator goes out, 2 The measured value remains on-screen but does not flash, 3 The yellow alarm LED goes out.
D
1
216 37 570
6 3 4
2
5
127 37 570
If the air temperature is over 38 oC (or over 40 oC in the case of the extended set value range): — The screen displays the warning message:
»Air temperature too high«,
— The intermittent alarm tone sequence (5 beeps) is sounded,
1 The central alarm light is lit**, 2 The measured value flashes, 6 The red alarm LED flashes.
The alarm tone can be muted for 5 minutes.
®
Caleo
heats up if necessary to attain the desired interior air
temperature setting.
2 The measured value continues to flash, 6 The red alarm LED continues to flash.
When the air temperature again drops below the alarm value: 4 Press the button to cancel the alarm.
For other alarms, see "Troubleshooting – Error Messages" on page 98. See "Alarm description" on page 123.
* The numerical values in this description are examples.
See "Setting alarms" on page 74.
** The central alarm light can be switched off.
See "Setting system parameters" on page 73.
41
Operation Using skin temperature control

Using skin temperature control

Immediately before using the yellow skin temperature sensor or the white peripheral sensor, insert it in the yellow or white socket and wait for the measurement signal to appear on screen.
1 Measurement signal from the yellow skin temperature
sensor (T1)
2 Measurement signal from the white peripheral temperature
sensor (T2)
If no measurement signal appears, the sensor must be replaced (page 101).
Using skin temperature measurement in air temperature control mode or skin temperature control mode.
Connect the temperature sensors for measuring the skin temperature and peripheral temperature:
Plug the yellow skin temperature connector into the yellow
socket on the sensor unit (skin temperature, T1). When using skin temperature control, the control functions refer to this sensor.
Plug the connector of the white peripheral temperature
sensor into the white socket on the sensor unit (peripheral temperature, or skin temperature of the twin in the case of air temperature control mode, T2).
Route the sensor cable through one of the flexible tubing
grommets.
Remove the protective foil from the adhesive pad and place
the skin temperature sensor on the pad.
Using the adhesive pad, attach the sensor tip to the
appropriate part of the baby’s skin.
D
1
2
235 37 570
125 37 570
Positioning the skin temperature sensor (yellow): If the baby is lying on his/her back:
Attach the yellow sensor to the abdomen, near the liver.
If the baby is lying on his/her belly:
Attach the yellow sensor to the back, preferably near the
kidneys.
Positioning the peripheral temperature sensor (white):
Attach the white sensor to the extremities, preferably the
foot or arm.
Do not use the skin temperature sensor (yellow) or peripheral temperature sensor (white) to measure rectal temperature!
Regularly check that the skin temperature sensor is correctly attached to the patient’s skin. If the skin temperature sensor has fallen off, it will measure the air temperature, resulting in possible overheating of the patient
(although the air temperature would not rise above 39 oC).
When a skin temperature sensor is attached but "air temperature control" is activated, the measured skin temperature is displayed. However, in this case, the temperature is not controlled as a function of skin temperature!
42

Changing over between air/skin temperature control

Do not use skin temperature control mode on babies in shock, as their skin temperature is much lower than normal. Skin temperature control would increase the air temperature by too much and would endanger the patient. Use air temperature control – see page 38.
Do not use skin temperature control mode on babies with a high temperature, as their skin temperature is much higher than normal. The control function would overcool the incubator, resulting in the risk of hypothermia.
Do not use skin temperature control mode on twins, because Caleo® only controls the temperature for one baby. Danger of hypothermia or overheating!
Operation
Changing over between air/skin temperature control
1 Change temperature control mode = press button. 1 The LED for the control mode requiring confirmation
flashes.
— The activated control mode is displayed on the screen.
The actual and set values are displayed as bar graphs and numeric values.
— The upper part of the screen displays the message:
»set value with rotary knob«.
D
1
003 37 570
43
320 37 571
Operation Changing over between air/skin temperature control
After the new mode has been activated, you can set the desired value with the rotary knob.
1 The LED of the activated mode flashes. 2 Adjust the set value = turn the rotary knob. 2 Confirm the new set value = press the rotary knob.
D
1
— The upper part of the screen displays the advisory message
»confirm new mode with rotary knob«.
2 Confirm the new mode = press the rotary knob.
— The display returns to the standard screen.
3
2
029 37 570
044 37 571324 37 571
1 The LED of the activated mode is lit.
If you do not wish to change the settings: 3 press = the new settings are cancelled.
The display returns to the standard screen. The previous set value is retained. Or — Wait for 7 seconds: Caleo
®
emits 4 short beeps to prompt the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
44

Using skin temperature control

Do not use skin temperature control mode on babies in shock, as their skin temperature is much lower than normal. Skin temperature control would increase the air temperature by too much and would endanger the patient. Use air temperature control mode – see page 38.
Do not use skin temperature control mode on babies with a high temperature, as their skin temperature is much higher than normal. The control function would overcool the incubator, resulting in the risk of hypothermia.
Do not use skin temperature control mode on twins, because Caleo® only controls the temperature for one baby. Danger of hypothermia or overheating!
Operation
Using skin temperature control
Make sure that you do not confuse the positions for the skin temperature sensors. Skin temperature control is regulated by the yellow skin temperature sensor (T1). If this sensor is stuck to the wrong part of the body, the patient may be excessively warmed.
Do not attach the sensor under the baby, because otherwise the core temperature, not the skin temperature, will be measured and used as control parameter!
The baby’s core temperature must be measured
regularly with an independent thermometer!
Do not leave the canopy open for any length of time, because the air temperature in Caleo® will drop!
Adjusting the set value
o
Standard set value range 34 Extended set value range 37.1 Default setting 36.5
C to 37 oC
o
C to 38 oC
o
C
1 Adjust the set value = press the button.
D
1
080 37 570
45
Operation Using skin temperature control
— The actual measured value and the set value are displayed
on the screen both as bar graphs and as numerical values.
— The following message is displayed in the top part of the
screen »set value with rotary knob«.
320 37 571163 37 570
1 Increase the set value = turn the rotary knob clockwise. 1 Decrease the set value = turn the rotary knob counter-
clockwise.
1 Confirm the new setting = press the rotary knob.
If you do not wish to change the settings:
press = the new settings are cancelled.
The display returns to the standard screen. The previous set value is retained. Or — Wait for 7 seconds: Caleo® emits 4 short beeps to prompt
the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
[ When using the extended set value range, particular
care must be taken to monitor the baby’s core temperature!
If the specified set value exceeds the standard set value range, — The following message is displayed
»confirm extended range with rotary knob«.
D
1
1 Confirm the extended range = press the rotary knob. 1 Continue increasing the set value = turn the rotary knob
clockwise.
46
047 37 571
— The advisory message »[ >37.0 oC« appears on the
screen.
— The following message appears at the top of the screen:
»set value with rotary knob«.
Confirm the set value = press the rotary knob.
Operation
Using skin temperature control
170 37 571315 37 571081 37 570
— The display returns to the standard screen. The measured
values are displayed.
— The set value and »set: >37.0 [« are displayed alternately.
If you do not wish to confirm the set value:
Cancel adjustment of the set value = press button.
The screen returns to standard display mode. The previous set value is retained. Or — Wait for 7 seconds: Caleo® emits 4 short beeps to prompt
the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
D
47
Operation Using skin temperature control

Alarms

Alarm limits can be changed in the configuration
(page 74).
Example: if the deviation between the set and measured skin temperature exceeds ±0.5 oC*: — The screen displays the warning message
»Skin 1 temp. deviation above 0.5 oC «,
— The alarm tone sequence (3 beeps) is sounded,
1 The central alarm indicator lights up**, 2 The measured value flashes, 3 The yellow alarm LED flashes.
The intermittent alarm tone sequence can be muted for 5 minutes: 4 Suppress intermittent alarm tone = press button, or 5 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted,
1 The central alarm indicator light goes out, 2 The measured value continues to flash, 3 The yellow alarm LED continues to flash.
1
216 37 570
When the measured value returns within the range ±0.5 oC: — The warning message disappears, — The intermittent alarm tone is muted,
1 The central alarm indicator light goes out, 2 The measured value remains on-screen but does not flash, 3 The yellow alarm LED goes out.
If the sensor plug is disconnected: — Instead of the measured value, 3 dashes are displayed. After 3 seconds: — The screen displays the warning message
»Connect skin 1 sensor«.
— The intermittent alarm tone sequence (5 beeps) is sounded,
1 The central alarm indicator lights up**, 2 The measured value flashes, 6 The red alarm LED flashes.
Then:
Immediately plug in the sensor
If the sensor is defective: — The screen displays the warning message
»Skin 1 sensor fault«.
— The intermittent alarm tone sequence (5 beeps) is sounded,
1 The central alarm indicator lights up**, 2 The measured value flashes, 6 The red alarm LED flashes.
Then:
Change the skin temperature sensor.
D
6 3 4
2
5
186 37 570
As long as 3 dashes appear on screen, Caleo® does not heat. Danger of hypothermia for the patient!
* The numerical values in this description are examples.
See "Setting alarms" on page 74.
** The central alarm light can be switched off.
See "Setting system parameters" on page 73.
48
The alarm tone can be muted for 5 minutes: 1 Suppress intermittent alarm tone = press button or 2 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted, — The central alarm indicator light goes out,
3 The measured value continues to flash, 4 The red alarm LED continues to flash.
For other alarms, see "Troubleshooting – Error Messages" on page 98. See "Alarm description" on page 123.
D
Operation
Using skin temperature control
4 1
3
2
215 37 570
49
Operation Using humidity control

Using humidity control

Connect the humidifier system
*
(see "Using humidifier systems" on page 24).
1 Set humidity control = press button.
— The actual value and current setpoint of the humidity control
are displayed as bar graphs and numerical values.
Soft key assignments:
2 = Do not activate new settings.
3 off = Switch off humidity control.
D
1
005 37 570
D
4 manual = Activate manual humidity control mode.
5 auto = Activate automatic humidity control mode
(see page 121).
— On activating the humidity control, AUTO humidity is
proposed as default.
2
3 5
4
031 37 570
048 37 571
* optional equipment feature
50

Setting AUTO humidity

In AUTO humidity, the humidity setpoint is automatically calculated and set by the system as a function of the air temperature setting (siehe page 121). The maximum relative humidity depends on the ambient temperature and incubator air temperature at max. 75 % relative humidity.
Operation
Using humidity control
1 Switch humidity control over to AUTO humidity = press
button (soft key).
2 Activate AUTO humidity = press the rotary knob.
— The display returns to the standard screen. The measured
value and the automatic set value are displayed.
D
1
2
036 37 570

Manually adjusting the set value

Standard set value range 30 % to 99 % Default value 50 %
3 Switch humidity control over to manual mode = press
button (soft key).
4 Increase set value = turn rotary knob clockwise. 4 Decrease set value = turn rotary knob counter-clockwise. 4 Confirm set value = press rotary knob.
The maximum relative humidity attained is between 85 % and 99 % r.h., depending on the ambient temperature and the incubator air temperature.
D
050 37 571032 37 570
3
4
51
Operation Using humidity control
— The actual value and the current set value of the humidity
control are displayed as bar graphs and numerical values.
325 37 571049 37 571
— The display returns to the standard screen. The measured
values and set values are displayed.
52
Alarms In the event of water shortage
— The following warning message appears on the screen
»Water empty, please refill«,
— The alarm tone sequence (3 beeps) is sounded, — The central alarm indicator lights up
*
,
1 The measured value flashes, 2 The yellow alarm LED flashes.
Fill the water container or replace the water bag – see page 24 and the following pages.
The intermittent alarm tone sequence can be muted for 15 minutes: 3 Suppress intermittent alarm tone = press button, or 4 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted, — The central alarm indicator light goes out,
1 The measured value continues to flash, 2 The yellow alarm LED continues to flash.
Operation
Using humidity control
D
2
3
1
4
When the cause of the alarm has been remedied, — The warning message disappears, — The intermittent alarm tone is muted, — The central alarm indicator light goes out, 1 The measured value remains on the screen but does not
flash,
2 The yellow alarm LED goes out.
For other alarms, see "Troubleshooting – Error Messages" on page 98. See "Alarm description" on page 123.
152 37 570
* The central alarm light can be switched off.
See "Setting system parameters" on page 73.
53
Operation Using O2 control

Using O2 control

*
Note the physiological risks from O2!
The air in the incubator should only be enriched with O2 when prescribed by a doctor. O2 enrichment must be controlled on the basis of the arterially measured oxygen partial pressure in the patient’s blood. Otherwise, there is a danger of hyperoxaemia (damage to the eyes) or hypoxaemia (brain damage).
Connect the probe of the O2 hose to the outlet of the
central medical gas pipeline supply (page 17).
1 Set O2 control = press soft key button.
— The actual value and the current set value of the O2 control
are displayed as bar graphs and numerical values.
D
1
006 37 570
Soft key assignments:
2 = Do not activate new settings.
3 off = Switch off O2 control.
4 on = Switch on O2 control.
— After the system is switched on, the oxygen sensors need
a warm-up phase before O2 control is active.
— The following message is displayed at the top of the screen:
»set value with rotary knob«.
— After the selected O2 setting has been confirmed,
automatic O2 control is active.
The current measured value is displayed after a few seconds.
D
2
3
4
033 37 570
* optional equipment feature
54
051 37 571

Adjusting the set value

Standard set value range 21 Vol.% to 40 Vol.% Extended set value range 40.1 Vol.% to 75 Vol.% Default setting 21 %
1 Increase set value = turn rotary knob clockwise. 1 Decrease set value = turn rotary knob counter-clockwise. 1 Confirm set value = press rotary knob.
Operation
Using O2 control
D
1
164 37 570
— The current actual measured value and the set value are
displayed on the screen both as bar graphs and as numerical values.
053 37 571
55
Operation Using O2 control
[ Note the dangers of high O2 concentrations
(see page 34)!
If the set value exceeds the standard set value range
— The following message is displayed at the top of the screen:
»confirm extended range with rotary knob«.
Confirm the extended set value range = press the rotary
knob.
Continue increasing the set value = turn the rotary knob
clockwise.
— The advisory message »[ >40 %« appears on the screen. — The following message appears at the top of the screen
»set value with rotary knob«.
Confirm the set value = press the rotary knob.
— The display returns to the standard screen. The measured
values are displayed.
052 37 571153 37 571316 37 571
— The set value and »set: >40 [« are displayed alternately.
56

Alarms

Alarm limits can be changed in the configuration (see page 74).
Operation
Using O2 control
Example: if the deviation between the set and measured O
2
concentration exceeds ±5 %*: — The screen displays the warning message
»Oxygen deviation above 5 %«,
— The alarm tone sequence (5 beeps) is sounded, — The central alarm indicator lights up**,
1 The measured value flashes, 2 The red alarm LED flashes.
The intermittent alarm tone sequence can be muted for 2 minutes: 3 Suppress intermittent alarm tone = press button or 4 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted, — The central alarm indicator goes out,
1 The measured value continues to flash, 2 The red alarm LED continues to flash.
When the measured value returns within the range ±5 Vol.%: — The warning message disappears, — The intermittent alarm tone is muted, — The central alarm indicator goes out,
1 The measured value remains on-screen but does not flash, 2 The red alarm LED goes out.
D
2
3
1
4
234 37 570
For other alarms, see "Troubleshooting – Error Messages" on page 98. See "Alarm description" on page 123.
* The numerical values in this description are examples.
See "Setting alarms" on page 74.
** The central alarm light can be switched off.
See "Setting system parameters" on page 73.
57
Operation Selecting menus

Selecting menus

1 Select menu = press »menu« button.
The required mode can be selected from the menu displayed:
2 Select item = turn the rotary knob. 2 Confirm selection (activate item) = press the rotary knob.
D
1
2
039 37 570

Kangaroo mode

See page 118 for a description of Kangaroo mode.
The patient’s core temperature or skin temperature must be constantly monitored, becaused it is outside the controlled climate. Particular attention must be paid to critical patients vital parameters. Ensure that all hoses and cables are routed correctly and safely without obstruction!
In Kangaroo mode the incubator is operated in air temperature control mode.
If Caleo
®
was previously operated in air temperature mode, the set value for air will be activated in Kangaroo mode. If Caleo® was previously operated in skin temperature mode, the last air temperature value will be loaded as set value. The yellow skin temperature sensor and the peripheral temperature sensor can be used in Kangaroo mode to monitor the skin temperature of the baby. The alarm limits for monitoring must be redefined (page 74).
The previously set values for — Humidity (page 51) and
2 (page 55) are retained in Kangaroo mode.
—O
156 37 571
The previous set value for skin temperature control is stored in buffer memory.
58
Activating Kangaroo mode 1 Display the main menu = press »menu« button.
Select "Kangaroo mode" from the menu.
2 Select item = turn the rotary knob. 2 Confirm and activate item = press the rotary knob.
— The screen displays the following advisory message:
»To leave the current mode please confirm the new Kangaroo Mode with rotary knob«.
2 Confirm Kangaroo mode = press the rotary knob.
Operation
Selecting menus
D
1
2
039 37 570
— When activated, Kangaroo mode is highlighted on screen
by a light background.
— The duration of Kangaroo mode can be displayed on the
screen in minutes and seconds (mm:ss).
Alarm muting is automatically activated, i.e. all alarms occurring within the next 5 minutes for the following parameters — Air temperature deviation — Kangaroo mode —O
2 deviation
are automatically displayed as "acknowledged" for the relevant period. See "Alarm suppression" on page 77.
The tubing grommets can be removed from the pillar
elements so that the hoses and cables remain well ordered during Kangaroo mode.
171 37 571059 37 571108 37 570
59
Operation Selecting menus

Alarms

Alarm limits can be changed in the configuration (page 74).
If the skin temperature of the yellow skin temperature sensor (skin 1) falls below the alarm limit set in the configuration:
— The screen displays the warning message
»Skin 1 temperature below 36.0
o
C «*,
— The alarm tone sequence (3 beeps) is sounded,
1 The central alarm indicator lights up**, 2 The measured value flashes, 3 The yellow alarm LED flashes.
The intermittent alarm tone sequence can be muted for 15 minutes. 4 Suppress intermittent alarm tone = press button or 5 Press the rotary knob. — The warning message remains on the screen, — The intermittent alarm tone is muted,
1 The central alarm indicator goes out, 2 The measured value continues to flash, 3 The yellow alarm LED continues to flash.
When the measured value returns above the alarm limit: — The warning message disappears, — The intermittent alarm tone is muted,
1 The central alarm indicator goes out, 2 The measured value remains on-screen but does not flash, 3 The yellow alarm LED goes out.
1
216 37 570
D
3 4
2
5
For other alarms, see "Troubleshooting – Error Messages" on page 98. See "Alarm description" on page 123.
* The numerical values in this description are examples.
See "Setting alarms" on page 74.
** The central alarm light can be switched off.
See "Setting system parameters" on page 73.
139 37 570
60
Ending Kangaroo mode
1 Display menu list = press »menu« button.
In the displayed menu, select »Return to Air Mode«
*
or
»Return to Skin Mode«*.
Operation
Selecting menus
D
1
2
039 37 570
2 Select item = turn the rotary knob. 2 Activate item = press the rotary knob.
— The screen displays the advisory message
»To leave the Kangaroo mode and return to Air Mode confirm with the rotary knob«.
2 Exit Kangaroo mode = press the rotary knob.
The former operating status with the previous set values is reactivated. The display returns to the standard screen.
After quitting Kangaroo mode, reinsert the tubing grommets in the pillar element.
336 37 571172 37 571
To maintain Kangaroo mode = press » « key.
Or — Wait for 7 seconds: Caleo® emits 4 short beeps to prompt
the user to press the rotary knob. If the rotary knob is not pressed, the display automatically returns to the standard screen after 7 seconds, and the previous set value is retained.
* the option displayed depends on the previously activated mode
61
Operation Selecting menus

Trend display

The trend screen is used for the graphical and numerical display of the measurement parameters. The data window always shows the last data in the selected time interval. In addition, the current measured values and set values are numerically displayed.
Switching over between standard/trend screen 1 Display trend = press » « button.
D
1
— The trend for the air temperature over the last 3 hours is
displayed. The current air and skin temperature values are displayed on the right next to the trend.
— T1: Yellow skin temperature sensor (Skin 1) — T2: White peripheral sensor (Skin 2)
2 Select menu = press the »TREND parameter« button.
246 37 570
329 37 571035 37 570
D
62
2
Select the desired trend display from the displayed list.
Defaults: — Air temperature trend — Zoom factor 3 hours
Select trend = turn the rotary knob.
Activate trend = press the rotary knob.
Back to menu = press » « key.
Operation
Selecting menus
307 37 571038 37 570
To exit trend display mode = press » « button.
The display returns to the standard screen.
Setting the time interval (Zoom) 1 Display menu = press »TREND parameter« button.
Select the desired trend from the displayed list.
D
1
Back to menu = press » « key.
307 37 571
63
Operation Selecting menus
1 Display the zoom menu = press the »zoom« button.
Select the desired zoom from the menu.
2 Select zoom = turn the rotary knob. 2 Activate zoom = press the rotary knob.
D
1
2
227 37 570
Back to menu = press » « button (soft key).
Return to trend selection = press the »trend« button
— In the illustrated example, the trend for the air temperature
over the last 3 hours will be displayed.
— T1: Yellow skin temperature sensor (skin 1) — T2: White peripheral sensor (skin 2)
308 37 571329 37 571
64

Trend analysis

Trend analysis is used for the graphical and numerical display of measurement parameters and their associated set values. The data window can be selected at will to any period over the last 7 days. Trend analysis can therefore be used to evaluate thermo­monitoring data. The following parameters can be set: — Skin temperature 1
(yellow skin temperature sensor, skin 1, T1)
— Peripheral temperature or skin temperature 2 (for twins)
(white temperature sensor, skin 2, T2) — Air temperature — Relative humidity (rel. %) —O2 concentration (Vol %)
*
*
.
Operation
Selecting menus
While trend analysis is in progress, no current measured values (air, skin, etc.) are displayed. The numerical values displayed are values from past readings.
If no button is pressed for 2 minutes, the display automatically reverts to the standard screen.
1 To activate trend analysis = hold down button for
4 seconds.
D
1
037 37 570
* optional equipment feature
65
Operation Selecting menus
— The trend graph for the selected measured value is
displayed on the screen.
Default values: Trend 1 Skin temperature
Trend 2 Air temperature Zoom 3 hours
The selected value is displayed as a trend graph. In this graph, the trend curve of the measured value is overlaid on the corresponding set value curve.
Time cursor: The time cursor is displayed as a vertical dotted line marking a precise point of time on the graph's time axis.
To move the time curve on the time scale = turn the rotary
knob.
310 37 571236 37 570
The time marked by the cursor line is specified underneath the dotted line by the date and time. The start time and end time of the current time window are specified to the right and left underneath the trend graph. If the time cursor is moved beyond the displayed time range, the screen adapts automatically and displays the relevant time range:
Less recent time range = turn the rotary knob counter-
clockwise.
More recent time range = turn the rotary knob clockwise.
Data window: The time cursor is associated with a data window situated to the right of the displayed trend. This data window shows the numerical values valid at the time marked on the time axis by the time cursor.
The following values are displayed in the data window:
1 Name of the selected parameter 2 Set value of this parameter at the marked time 3 Measured value at the marked time
D
1 2 3
66

Trend selection

A trend graph can be selected with the soft keys.
Soft Key assignments:
1 = Cancel. New settings will not be activated.
2 trend 1 = Select trend 1.
3 trend 2 = Select trend 2.
4 zoom = Select time interval.
Operation
Selecting menus
D
Select trend 1: 2 Display the trend 1 menu = press soft key. The following parameters can be selected as "trend 1": — air — skin — humidity
*
—O2* — weight*
5 Select trend 1 = turn rotary knob. 5 Confirm (activate) selection = press rotary knob.
The newly selected trend will be displayed on the screen.
Select trend 2: 3 Display the trend 2 menu = press soft key. The following parameters can be selected as "trend 2": — air — humidity* — oxygen* — weight* — disable
3
1
2
4
5
207 37 570
330 37 571331 37 571
5 Select trend 2 = turn rotary knob. 5 Confirm (activate) selection = press rotary knob.
* optional equipment feature
67
Operation Selecting menus
The newly selected trend 2 is displayed on the screen underneath trend 1. In addition, a second data window containing the relevant parameters is opened next to trend 2. The time cursor and time range (zoom) are the same for both trend displays. The »disable« option removes trend 2 from the screen, so that only trend 1 is displayed.
Selecting the time interval (zoom) 1 Open the Zoom menu = press soft key.
The following intervals of time can be selected for the time window (zoom) function: 3; 6; 12; 24; 48 hours or 7 days.
2 Select zoom = Turn rotary knob. 2 Activate zoom = Press rotary knob.
The selected trends will then be displayed in the newly selected time window.
All data going back a maximum of 7 days can be displayed. The individual measuring points together with the associated data for the data window are displayed. When displaying the measured weight values, up to 30 values can be stored and displayed as a trend. Values between the individual measured values are interpolated.
D
309 37 571237 37 570
3
1
2
332 37 571
Ending trend analysis
To exit trend analysis
Back to menu = press » « soft key button
or 3 Terminate trend analysis = press button or if no soft key on the screen is pressed for 2 minutes, the display will automatically revert to the standard screen.
68
Operation
Selecting menus

Cleaning mode

*
Cleaning mode is only available if Caleo® is equipped with humidity control.
Cleaning mode may only be used when Caleo® is empty.
In cleaning mode the water heater is boiled empty and then cooled.
Cleaning mode lasts about 60 minutes.
1 Open menu = press the »menu« button.
Select "Cleaning Mode" from the displayed menu.
D
D
1
1
2
2
039 37 570
2 Select item = turn the rotary knob. 2 Confirm (activate) item = press the rotary knob.
— The required operating steps are specified on the screen:
»disconnect Luerlock connector from apparatus
make sure that no patient is inside
start Cleaning Mode by
pressing ▼▼▼▼ keys.
328 37 571303 37 571
* optional equipment feature
69
Operation Selecting menus
Start Cleaning Mode = press both soft keys simultaneously.
The water heater is boiled dry. The heater is then cooled.
D
208 37 570
Risk of burns on contact with the heater!
Do not disassemble Caleo® during cleaning mode.
— The screen indicates when cleaning mode is complete.
Confirm end of cleaning mode = press rotary knob.
Caleo® performs a restart.
304 37 571
After ending cleaning mode: — Disinfect and clean Caleo
Sterilising" on page 84).
70
®
(see "Disinfecting/Cleaning/
323 37 571
Configuration
In configuration mode, you can set — Language, date and time — System parameters and — Alarm parameters and you can obtain information on —O2 sensors — the software version.
Activating configuration mode 1 Activate configuration mode = hold down the
»menu« button for 4 seconds.
Operation
Configuration
D
1
The individual parameters of the configuration can now be selected.
2 Select configuration parameters = turn the rotary knob. 2 Activate the configuration parameters = press the rotary
knob.
Cancel selection = press » « key.
— The display returns to the standard screen.
2
039 37 570
062 37 571
71
Operation Configuration
Language/date/time 1 Select language/date/time = Turn and press the rotary
knob.
— The language selection box is highlighted by a bright
border.
D
1
163 37 570
1 Select language = turn the rotary knob. 1 Confirm language = press the rotary knob.
— The selected language is displayed on the screen.
Follow the same procedure to set the date format, date and time.
1 Save settings = turn the rotary knob to »back« and then
press the control button.
Cancel selection = press » « soft key button.
— The display returns to the configuration parameter
selection menu (see page 71).
064 37 571
72
Setting system parameters 1 Select the temperature unit = turn and press the rotary
knob.
— The temperature unit is highlighted by a light background.
1 Select unit = turn the rotary knob. 1 Confirm unit = press the rotary knob.
Operation
Configuration
D
1
164 37 570
— The selected unit is displayed on the screen.
The weight unit, dT/T
2 display, central alarm indicator and
altitude above sea level* can be selected by the same procedure. — The weight unit can only be set if the optional weighing
scale is integrated in Caleo®. — If using 2 skin temperature sensors, the screen can display
T1 and either T2 or the difference between T1 and T2 (dT). — If the central alarm light is deactivated, alarm situations are
only indicated by the flashing "Alarm" LED on the control
panel and the alarm beep signal. — The altitude above sea level* can only be set if O2 control is
integrated in Caleo®.
If the local altitude is entered incorrectly, the measurement
accuracy of the O2 sensors will be reduced (e.g. 1.5 %
additional error for 1000 m difference in altitude).
066 37 571
Cancel selection = press » « soft key button.
— The display returns to the configuration parameter menu
(see page 71).
* Altitude above sea level.
73
Operation Configuration
Setting alarms 1 Select alarms = turn and push the rotary knob.
— The menu that is then displayed allows the user to select
Kangaroo mode, the initial volume of the audible alarms and the alarm limits for temperature control and O
2 control.
D
1
164 37 570
1 Select menu item = turn rotary knob. 1 Confirm menu item = press rotary knob.
Cancel selection = press » « soft key button.
— The display returns to the configuration parameter
selection menu (see page 71).
Kangaroo mode:
Adjustment ranges: Skin alarm T1 min 33 Skin alarm T2 min 33 dT alarm min –2 dT alarm max 2
o
C to 37 oC and off
o
C to 37 oC and off
o
C to 2 oC and off
o
C to 5 oC and off
Default values: Same as last setting
The text items in the description that follows are examples: 1 Select skin alarm T1 min (lower alarm limit for skin
temperature) = turn and press the rotary knob.
— The T1 min alarm selection box is highlighted by a bright
border.
1 Select alarm = turn the rotary knob. 1 Confirm alarm = press the rotary knob.
The display returns to the alarm settings menu (page 74).
067 37 571070 37 571
The other alarms are set by the same method.
Cancel selection = press » « soft key button.
— The screen returns to the configuration parameters menu
(see page 71). For a description of Kangaroo mode alarms please refer to page 119.
74
Initial volume (sound intensity) of the alarms:
Adjustment range: 1 to 8
Default: 1
— The initial sound intensity is highlighted by a light
background.
Set an initial sound intensity = turn rotary knob.
Confirm initial sound intensity = press rotary knob.
The screen returns to the alarm settings menu (page 74).
Cancel selection = press » « soft key button.
— The display returns to the configuration parameter menu
(see page 71).
Operation
Configuration
Alarm limits:
Adjustment ranges: Air temperature deviation –1.5 or –2.5 Skin temperature deviation ±0.3 to 1.0 O
2 deviation* ±3 % or ±5 %
Default values: Air temperature deviation –1.5 Skin temperature deviation ±0.5 O
2 deviation* ±5 %
Select the alarm limit for air/skin temperature and O2
o o
o
C
o
C
C
C
*
=
turn and push the rotary rotary knob.
— The alarm limit setting for the deviation in air temperature is
highlighted by a bright border.
Select air temperature deviation = turn the rotary knob .
Confirm air temperature deviation = press the rotary knob.
The screen returns to the alarm settings menu (see page 74).
The other alarm limits are set in the same way.
071 37 571
Cancel selection = press » « soft key button.
— The display returns to the configuration parameter menu
(see page 71).
* optional
072 37 571
75
Operation Configuration

O2 sensor information

*
The screen displays the following information for the O2 sensors used: — Date of manufacture — Date of last calibration — Date of next calibration.

Viewing software information

— The software version and the number of operating hours
are displayed on the screen. Where applicable, this screen
contains additional information on service intervals. Further
information on this subject is provided in the service
documentation.
305 37 571075 37 571165 37 570
1 Return to configuration parameter menu = press soft key
button or 2 Press the rotary knob. — The display returns to the configuration parameter menu
(see page 71).
* optional equipment feature
76
D
1
2

Lock key pad functions

1 To lock further on-screen setting = press button. 1 The LED on the button lights up.
Operation
Lock key pad functions
D
1
— After 4 seconds, all screen functions are locked,
except for:
1 Lock key pad function 2 Alarm suppression 3 Rotary knob 4 Bed tilt.
1 The LED in the button remains lit.
Enabling
1 Change setting on screen = press button. 1 The LED in the button goes out.
— After 4 seconds, the screen functions can be changed.
The LED in the button remains off.

Alarm suppression

2 Pressing this key if an alarm is active:
— The audible alarm is muted — The central alarm indicator goes out*. 2 The LED in the button is not lit. The permitted duration of alarm suppression depends on the type of alarm. When the cause of the alarm is removed, alarm suppression is automatically ended.
130 37 570
D
2 1
3
4
131 37 570
D
2
2 Pressing this key if no alarm is active (alarm muting):
For all alarms occurring in the next 4 mintues, of the type — Air temperature deviation — Skin temperature deviation —O2 deviation, no alarm tone will be sounded, and the central alarm indicator will not be lit. However: the alarm message is displayed on the screen, and the LEDs for the measured value and alarm flash. 2 The LED in the button will be lit.
* The central alarm light can be switched off.
See "Setting system parameters" on page 73.
134 37 570
77
Operation Weighing scale

Weighing scale

*
The weighing scale is located directly underneath the bed. In the weighing process, the entire bed and the items placed on top of it are weighed with the baby. However, these extras are deducted again as soon as the baby is lifted off the bed (tare weight), so that precise weighing of the baby is possible. Before weighing, check that the bed is fully pushed in and is in the horizontal position. The scales must be regularly checked by a test weight. The accuracy specified in the technical data of the scales is only obtained if the scales have been calibrated at the installation site.
The spirit levels show whether the bed is aligned horizontally.
®
1 Spirit levels for the horizontal alignment of Caleo
in the
transverse axis.
2 Spirit level for the horizontal alignment of Caleo® in the
longitudinal axis.
To align the bed horizontally see page 23.
1
2
®
During the weighing process, Caleo
must not be exposed
to any vibrations.
During the weighing process, no objects may be placed on the bed surface.
During the weighing process, no objects may be placed been the bed and the housing.
The weighing accuracy may be adversely affected when using the ventilation hose holder.
Starting the weighing procedure
Remove the supply line or any other objects that touch the
bed and fixed environment (trough and incubator frame) at the same time.
3 Activate the weighing process = Press button.
If the scale is not ready for operation, this function cannot be activated.
1
194 37 570
D
* optional equipment feature
78
3
129 37 570
During weighing, the user is guided through the sequence of operating steps by the following prompts:
Operation
Weighing scale
Please adjust Caleo
Turn knobs to X-ray position (see page 21).
Lift baby clear of mattress.
®
to horizontal position (see page 23).
The system waits until at least 250 g have been taken off the scale and the scale has been at rest for 3 seconds. The tare is determined.
— 1 beep is sounded.
Place patient on bed.
The system waits until at least 250 g have been added to the scale and the scale has been at rest for 3 seconds. Weighing then proceeds.
077 37 571078 37 571079 37 571
— 1 beep is sounded.
The weighing procedure is complete.
— The current and last weighing results are displayed on the
screen.
After 30 seconds, the display reverts to the standard screen. or
Display the standard screen = press the rotary knob.
The screen displays the last weight as advisory text for the next 10 minutes.
Weighing process cancelled
The weighing process is cancelled if the scales are not unloaded or loaded within 30 seconds. — 3 short beeps are emitted.
79
Operation Weighing scale

Weighing without tare

If the last tare weight was obtained no longer than 60 minutes earlier, and if no objects have since been removed from the bed or placed on it, recalibration with a new tare weight is not necessary.
1 Activate the weighing process = press soft key.
®
Caleo
proposes reweighing directly without obtaining
a new tare.
2 Weigh without tare = press button.
D
2
1
206 37 570
— Lay the patient on the bed.
The system waits until the weighing scale has remained at rest for 3 seconds. Weighing then proceeds.
— 1 beep is emitted.
The weighing procedure is ended.
— The present and last weights are displayed.
After 30 seconds, the display reverts automatically to the standard screen. or
Display the standard screen = press the rotary knob.
The screen displays the last weight as advisory text for the next 10 minutes.
333 37 571334 37 571
80

Ending operation

Switch off incubator 1 Switch off the incubator = press the on/off switch.
The incubator is switched off.
Operation
Ending operation
1
If the O2 hose is connected:
*
— Disconnect the probe from the outlet of the central
O2 supply pipeline and place it in the "Park" position
(strictly follow the separate Instructions for Use).
110 37 570
* optional equipment feature
81
82
Care
Care
Contents
Disinfecting/Cleaning/Sterilising . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Stripping down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Before reusing for a patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Care list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
83

Disinfecting/Cleaning/Sterilising

Disinfecting/Cleaning/Sterilising

Risk of burns on contact with the heater
When the incubator is closed, the heater is still hot enough to cause serious burns for a long time after switching off (still approx. 70
Clean and disinfect the incubator thoroughly: — after each change of patient — at least once a week.
Clean and disinfect the accessories, such as the suction equipment and the flowmeter in accordance with their separate Instructions for Use.
o
C after 1 hour).
Disinfecting/cleaning/sterilising
Use surface disinfectants for disinfecting. For reasons of material compatibility, use disinfectants based on — aldehydes — quaternary ammonium compounds.
Due to the chemical composition and direct effects on material compatibility, the following products are only suitable to a limited extent: — Halogen-releasing compounds — Strong organic acids —O2 releasing compounds.
Use only cleaning agent in autoclaves and other cleaning and disinfecting machines. Do not use any alkaline or chlorine releasing disinfectants, due to the risk of corrosion. As a general rule, the manufacturer's recommendations must be taken into account when choosing a cleaning agent. Liability for the specified use and for any material damage that may be caused by the product rests with the manufacturer.
Do not use any disinfectants or cleaning agents containing alcohol.
We recommend the use of disinfectants based on the current DGHM list (DGHM: German Society for Hygiene and Microbiology). The DGHM list (published by mhp Verlag GmbH, Wiesbaden) also gives the composition of each disinfectant. For countries where the DGHM list is not familiar, we recommend the types of disinfectant specified above.
84
The following surface disinfectants are recommended:
Surface disinfectants Manufacturer
Dismozon® pur Bode Chemie GmbH & Co., Germany
®
Incidur
Henkel-Ecolab Deutschland GmbH Trichlorol Lysoform, Germany Virkon Tetenal, Germany Bacillol 25 Bode Chemie GmbH & Co., Germany Seculyse Paragerm (Henkel Ecolab), France Sekupoudre Paragerm (Henkel Ecolab), France Vaposeptol Paragerm (Henkel Ecolab), France Cidex Johnson & Johnson, Taiwan Habitane Zeneca Limited, Norway Kloramin Norsk Medisinal Depot A/S, Norway Sactiv Diversey Lever, Finland Viraclean Whiteley, Australia
Disinfecting/Cleaning/Sterilising
Note the manufacturer's directions for use.
Do not treat Caleo
®
and its components in formaldehyde
vapour or ethylene oxide.
The following materials are used in the patient capsule:
Components Material
Canopy, flaps Polycarbonate Support pillar units Styrene-butadiene thermo foam rubber Bed Styrene-butadiene thermo foam rubber Housing Polystyrene Caleo® SoftBed™ Polyurethane/polyester
85
Disinfecting/Cleaning/Sterilising Stripping down

Stripping down

Check the fresh air filter
[ Use the original Caleo
Risk of fire!
Tilt the unit to remove the fresh air filter more easily
(see page 23).
®
air filter (MX 17 015) only!
1 Pull the holder for the fresh air filter downwards and
remove.
2 Check that the seal on the filter box is in place.
Fitting the filter 3 Separate the filter holder and
4 filter frame at the snap-in fastening. 5 Insert the filter.
Close the filter holder and filter frame.
Mark the next due date for replacement of the new filter
(2 months after installation) on the label and
6 Stick the label to the filter holder.
211 37 570124 37 570
1
2
123 37 570
3
4
5
If a filter is already installed:
Check the condition of the filter.
Check the due date for replacement. The label is on the
filter holder.
If the filter is fouled, damaged or moist, or if the due date for replacement has been reached:
Discard the old filter.
Fit a new filter.
86
6
122 37 570
Removing the water supply
Close the clamp on the connection tube.
Remove and dispose of the water bag and connection tube
or
1 Remove the water container from the holder. 2 Detach the connection tube from the water container and
water connection pipe.
Dispose of the connection tube.
Disinfecting/Cleaning/Sterilising
Stripping down
Clean the water container in the cleaning and disinfecting
machine at 93 oC
or
sterilise at 134
oC*
*
.
Removing the water connection pipe
Tilt the incubator to make the water connection pipe easier
to remove (see page 23).
3 Turn the water connection pipe 90o clockwise and pull it
out of the guide channel.
1
2
184 37 570
3
200 37 570
Clean the water connection pipe in the cleaning and
disinfecting machine at 93 oC
or
sterilise at 134
* Observe national and international standards on
cleaning, disinfecting and sterilisation procedures (e.g. EN 285, EN 554, EN 556).
oC*
.
*
178 37 570
87
Disinfecting/Cleaning/Sterilising Stripping down
Place Caleo
Switch on cleaning mode
®
horizontal (see page 23).
*
(see page 69).
After cleaning mode* has completed or if humidity control is not provided:
Switch off the incubator and disconnect from mains.
Remove any auxiliary equipment installed (for care
instructions see the specific Instructions for Use of the equipment concerned).
Risk of burns on contact with the heater. Allow Caleo® to cool down before stripping down further.
Remove the canopy. 1 Grasp the handles on the sides of the canopy with both
hands.
2 Lift the canopy horizontally off its supports (pillar elements).
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
1
1
22
To remove the double wall:
Remove the canopy and lay it upside down (on a soft,
non-abrasive surface).
3 Squeeze the double wall inwards slightly, and 4 Release the oblong holes of the double wall from the
retaining lugs in the canopy.
Pull the double wall out of the canopy.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
202 37 570
4
4
3
149 37 570
* Only available in combination with humidity control.
88
Open the front flap: 1 Turn the two locking knobs inwards as far as they will go
and fold down the front flap.
Fold out the adjustable double walls to clean them.
Disinfecting/Cleaning/Sterilising
Stripping down
Open the side flaps in the same way.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
Remove all through-hole grommets, sleeves and liners etc.
Clean them at 93
o
C in the cleaning and disinfecting
machine*.
Remove the mattress from the bed.
2 Remove the bed. Take care not to damage the sensor unit.
1
1
179 37 570
181 37 570119 37 570
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
Remove the X-ray drawer:
Using the recessed handle, pull out the X-ray drawer
as far as it will go.
3 Tilt the drawer upwards and pull it out of the unit.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
* Observe national and international standards on
cleaning, disinfecting and sterilisation procedures (e.g. EN 285, EN 554, EN 556).
2
2
3
120 37 570
89
Disinfecting/Cleaning/Sterilising Stripping down
Remove trough:
1 Press both catches inwards and pull the trough upwards.
2 Lift out the trough. 3 Remove the fan impeller.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
Turn the trough over.
4 Lift the air guide plate to the side to disinfect and clean.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
1
2
1
203 37 570
2
3
114 37 570
4
115 37 570
90
Base frame/body
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
Remove any impurities near the slits of the sensor unit.
Do not allow any moisture to enter the sensor. Do not disinfect the sensor unit by immersion or spraying.
Disinfecting/Cleaning/Sterilising
Stripping down
Control panel
Pull off the turn-and-press rotary knob.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect (see
manufacturer's specifications), wipe the surface with a clean damp cloth and then rub dry.
Do not allow any moisture to enter the control panel. Do not disinfect the control panel by immersion or spraying.
Use only the recommended cleaning agents and disinfectants. Otherwise, there is a risk of cracks forming in the transparent parts, e.g. due to the use of alcohol.
Do not use UV radiation on the incubator, because this treatment may cause cracks in the transparent components.
183 37 570195 37 570
D
91
Disinfecting/Cleaning/Sterilising Before reusing for a patient

Before reusing for a patient

Check that the system has been cleaned and disinfected in
conformity with the regulations of the hospital.
Reassemble the equipment with disinfected hands.
Refit all equipment – see "Stripping down", page 86.
When inserting the trough, make sure that both holders click into place.
If the holders of the trough are damaged:
Call DrägerService.
Fitting the water connection pipe:
1 Check the O-ring (2M 50 346) for damage and replace
if necessary.
2 Check the locking valve for damage.
1
2
To make the water connection pipe easier to install, tilt the incubator.
Connect the incubator to the mains power supply.
Switch on the incubator = press the on/off switch until it
clicks into position (see page 37).
Tilt the bed (see page 23).
3 Push the water connection pipe into the guide duct, holding
the LuerLock connection vertically at the top.
After overcoming a spring force,
4 turn the water connection pipe counter-clockwise 90o and
secure the locking valve in the holder.
198 37 570
3
254 37 570
4
92
199 37 570
Check that the incubator is ready for operation –
see "Before each use" on page 26.
To remove possible disinfectant residue, it is recommend running the incubator in standby mode.
Switch on Caleo
Activate air temperature control (see page 38).
Run Caleo
®
(see page 37).
®
at 37 oC with opened flaps.
If using a water container: Do not refill the water container until just before placing the patient in the incubator (see page 24)!
If using a water bag: Do not connect the water bag until just before placing the patient in the incubator (see page 24)!
Disinfecting/Cleaning/Sterilising
Before reusing for a patient
93
Disinfecting/Cleaning/Sterilising Care list

Care list

What How often How
Disinfect and clean Sterilise
Reusable components Care intervals
Connection tube Replace weekly and with each
change of patient Water container Change of patient/weekly yes yes Water connection pipe Change of patient/weekly yes yes Canopy Change of patient/weekly yes Double wall Change of patient/weekly yes Front flap Change of patient/weekly yes Side flap Change of patient/weekly yes Double walls Change of patient/weekly yes Through-hole sleeves etc. Change of patient/weekly yes Bed Change of patient/weekly yes Mattress Change of patient/weekly yes X-ray drawer Change of patient/weekly yes Trough Change of patient/weekly yes Air guide plate Change of patient/weekly yes Fan impeller Change of patient/weekly yes Base frame Change of patient/weekly yes Control panel Change of patient/weekly yes Rotary knob (control panel) Change of patient/weekly yes Fresh air filter
d
Replace every 2 months
Wipe
a
Cleaning and
disinfecting
machine
93 oC
b
c
Steam
134
oC c
a. Use surface disinfectants based on aldehydes and quaternary ammonium compounds. b. Use only detergent.
Do not use disinfectants that release alkali or chlorine. Risk of corrosion!
c. Observe national and international standards on cleaning, disinfecting and sterilisation procedures
(e.g. EN 285, EN 554, EN 556).
d. Use the original Caleo® air filter (MX 17 015) only! Risk of fire!
94

Maintenance intervals

Maintenance intervals
Disinfect and clean the incubator or the relevant parts before each maintenance operation, even when returning the equipment for repair purposes!
Disconnect from the mains before each maintenance operation. Risk of electric shock.
Intervals Who?
when
necessary
Replaceable parts:
Fresh air filter X
Grommets, sleeves etc. X
b
O-ring, for water connection pipe X
Fan motor X
Lithium battery X Authorised
Skin temperature sensors
Adhesive pads X
Mattress X Medical
Maintenance:
Device servicing and maintenance
Calibration:
O2 sensors X Authorised
Scales X
a. Use the original Caleo® air filter (MX 17 015) only! b. Replace if the material becomes brittle or sticky or if strips of material have become detached c. Replace after 18000 – 20000 operating hours d. At the latest when changing patient e. At the latest when changing patient f. Measuring accuracy depends on local geological and geographical conditions.
The specified accuracy is only applicable if the scales have been calibrated at the installation site.
f
weekly every
2 months
a
d
X
e
every
6 months
once
a year
every
two years
Medical and technical personnel
Medical and technical personnel
Medical and technical personnel
c
Authorised technicians
technicians Medical
personnel Medical
personnel
personnel
X Authorised
technicians
technicians
X Authorised
technicians
95
Maintenance intervals Disposal

Disposal

Disposal of the connection tube and fresh air filter
— with household waste.

Disposal of O2 sensors and battery

— Do not throw into the fire. Risk of explosion. — Do not force open: risk of chemical burns. — Do not recharge the battery. Special waste: — Dispose of special waste in accordance with national waste
disposal regulations.
Further information can be obtained from national and local environmental and legislative authorities and suitable waste disposal companies.

Disposal of incubator

At the end of its useful life. — Dispose of the incubator in conformity with national waste
disposal regulations. or — Hand the incubator over to a suitable waste disposal
company for disposal. Further information can be obtained from national and local environmental and legislative authorities.
All plastic components weighing more than 50 g must be marked with their material code in order to ensure environmentally sound disposal.
96
Troubleshooting – Error Messages
Troubleshooting – Error Messages
Contents
Troubleshooting – Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Troubleshooting – Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
97

Troubleshooting – Error Messages

Troubleshooting – Error Messages
All error messages are displayed on the screen. They are listed in the table below in alphabetical order. See also "Alarm description" on page 123.
Message Cause Remedy
Adjust. inoperable
Yellow alarm LED lights up on the control unit. Alarm sounded (1x).
Air heater inoperable
Measured value flashes on the screen. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Air temp. deviation above 1.5 oC
(see "Configuration" on page 71) Measured value flashes on screen. Yellow alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (3x).
Air temp. sensor inoperable
3 flashing dashes on the screen. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Air temperature too high
Measured value flashes on screen. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Calibrate scale before weighing The scales are not calibrated. Call DrägerService. Connect skin 1 sensor
Three dashes flash on screen in place of the measured value. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Device Error
with number display. Continuous alarm tone, red alarm LED lights up.
Height adjustment defective. Motor tilt position defective. Servomotor drive defective.
Air heater defective. Call DrägerService. 1 min.
The specified tolerance limit for deviation from the set value has been exceeded by a value either above the maximum or below the minimum value.
Air temperature sensors are defective.
Incubator temperature above 38 oC. Incubator temperature above 40 oC.
Probe for skin temperature (yellow) is not connected.
Device fault. Switch off device and then
Call Dräger Service. 15 min.
Reduce/increase humidity. Close canopy, front flap and hand ports. Check the set value and the configuration (page 71).
Call DrägerService. 1 min.
Remove external temperature sources. Call DrägerService.
Check connection and correct if necessary.
switch on again. Call Dräger Service.
Duration audible alarm muted
15 min.
5 min.
5 min.
* The central alarm light can be switched off. See "Setting system parameters" on page 73.
98
Troubleshooting – Error Messages
Message Cause Remedy
Fan inoperable
Red alarm LED on the control panel
Fan defective. Check fan impeller.
Call DrägerService. flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Goldcap accumulator fault
Yellow alarm LED on control unit lights up. Audible alarm sounded (1x).
Heater temp. sensor inoperable
Red alarm LED on the control panel
Goldcap rechargeable battery defective. Goldcap battery connection defective.
Air heater temperature sensors defective.
Call Dräger Service. 10 min.
Call DrägerService. 1 min.
flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Humidifier inoperable
Measured value (closed-loop humidity
Water heater is defective. Switch off humidity module.
Call DrägerService. control) or set value (open-loop humidity control) flashes on the screen. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Humidity deviation above 10 % Canopy, front flap or hand
ports are open. Sensor defective.
Wait for start-up phase. Set
value <85 %.
Close canopy, front flap or
hand ports.
Check water supply
connection.
Call DrägerService.
Humidity sensor inoperable
3 flashing dashes on the screen.
Humidity sensor is defective. Switch off humidity control.
Call DrägerService. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Key pad locked
Message displayed and yellow alarm LED lights up for 2 seconds on pressing a key.
Oxygen concentration below 18 % O2 concentration less than
Button functions (set value input/weighing/ menu button) disabled.
18 Vol.%.
Activate push button functions
(page 77).
Check that the correct gas is
connected.
Call DrägerService.
Duration audible alarm muted
5 min.
1 min.
1 min.
-
1 min.
* The central alarm light can be switched off. See "Setting system parameters" on page 73.
99
Troubleshooting – Error Messages
Message Cause Remedy
Oxygen deviation above 5 %
(can be set between 3 and 5 % – see "Configuration" on page 71) Measured value flashes on screen. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Difference in measured O2 concentration is greater than 3 or 5 %.
Values above 65 % O2 have been set.
Close the canopy, front flap, X-ray drawer, hand ports and filter holder. Check O2 connection. Check O2 supply via central medical gas pipeline or O2 cylinder. Check configuration (page 71). Reduce O2 setting to 65 %. Call DrägerService.
Oxygen module inoperable
3 flashing dashes on the screen.
Controller of O2 module is defective.
Switch off oxygen module.
Call DrägerService. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Oxygen sensor deviation above 3 %
Measured value flashes on screen.
Sensor 1 or sensor 2 is defective.
Switch off oxygen module.
Call DrägerService. Red alarm LED on the control panel flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x).
Oxygen sensor 1 inoperable
3 flashing dashes on the screen. Red alarm LED on the control panel
Sensor for oxygen measurement is defective. Sensor 1 is defective.
Switch off oxygen module.
Call DrägerService.
flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x)
Oxygen sensor 2 inoperable
3 flashing dashes on the screen. Red alarm LED on the control panel
Sensor for oxygen measurement is defective. Sensor 2 is defective.
Switch off oxygen module.
Call DrägerService.
flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (5x)
Oxygen valve fault
Measured value flashes on screen.
O
2 valve defective. Switch off O2 module.
Call Dräger Service. Yellow alarm LED flashes on control unit. Central alarm indicator lights up.* Intermittent audible alarm is sounded (3x).
Scale inoperable
The scales are defective. Call DrägerService.
3 short alarm tones
Skin 1 less than 0.5 oC above skin 2
(see "Setting alarms" on page 74) The measured values for skin temperatures flash on-screen. Yellow alarm LED on the control panel
Kangaroo mode alarm: temperature difference between the skin temperature (yellow) and peripheral temperature (white) too low.
Check patient's heat exchange.
Check configuration (page 71).
flashes. Central alarm indicator lights up.* Intermittent audible alarm is sounded (3x).
Duration audible alarm muted
2 min.
1 min.
1 min.
1 min.
1 min.
1 min.
15 min.
* The central alarm light can be switched off. See "Setting system parameters" on page 73.
100
Loading...