WARNING!
For a full understanding of the
performance characteristics of this
medical device, the user should carefully
read this manual before use of the device.
The header line on each page contains the title of the
chapter. This helps you find your way quickly from
subject to subject.
Page Body
The page body in these Operating Instructions
combines text and illustrations. The information is
presented as sequential steps of action, giving the
user hands-on experience in learning how to use the
Apollo inhalation anesthesia machine.
Left-Hand Column - the Text
The text in the left-hand column provides
explanations and step-by-step instructions on the
practical use of the machine.
Bullet points indicate separate actions. Numbers are
used both to refer to relevant details in the
illustrations and to specify the sequence of actions
where several actions are described.
Right-Hand Column - the Illustrations
The illustrations provide visual reference for the text
and for locating the various parts of the equipment.
Elements mentioned in the text are highlighted.
Renderings of screen displays guide the user and
provide a way to reconfirm actions performed.
Overview
The user can configure settings on the Apollo in
Standby mode as well as during operation. Standby
configuration allows the user to save a complete set
of defaults that are invoked automatically when the
machine is switched on (see “Configuring the default
settings in Standby” below). The configuration
settings that can be made during operation are more
limited and are valid only until the machine is
switched off (see “Configuration during operation” on
page 191).
Configuring the default settings in
Standby
anesthesia machine starts with when it is switched
on.
The default settings for ventilation, fresh-gas
delivery, and monitoring can be activated while in
Standby by pressing the >Restore Default
Settings< button (1 in Figure 124) on the standby
screen.
The default settings can be configured in Standby as
follows:
1. Press the standby key > <, and confirm by
pushing the rotary knob.
2. Press the >Default Config< button
(2 in Figure 124).
The user is requested to enter a four-digit
password in order to prevent unauthorized
changes to the basic functions (see Figure 125).
The four-digit password is assigned at the
factory. If desired, the password can be changed
or the password function can be disabled
altogether by DrägerService.
3. Select and confirm the figures successively from
Part Number: 9039994, 1st edition
the line displayed using the rotary knob. The
password is represented by asterisks (****)
below the line of numbers
The menu Standby Conf. for selecting the default
values is displayed when the password has been
entered correctly, see Figure 126.
Operating Instructions Apollo SW 4.n175
Figure 124. Location of Standby Config KeyDefault settings describe settings which the
Figure 125. Password Screen
Configuration
2
1
Part Number: 9039994, 2nd edition
Typing Conventions
User controls are designated as >Control Name<,
e.g.:
>PEEP<
Screen messages and screen options are printed in
bold, e.g.:
Default Alarm Limits
Operating Instructions Apollo SW 4.n7
Introduction
Trademarks
–Apollo®
–The Dräger® name and logo
–DrägerService®
–Drägersorb®
–Vapor®
–Spirolog®
–SpiroLife®
–WaterLock®
are registered trademarks of Dräger.
–Durasensor®
–OxiMax®
are registered trademarks of Nellcor.
–Selectatec®
is a registered trademark of Datex-Ohmeda.
All other products or brand names are trademarks of
their respective owners.
Definitions
Symbols and Abbreviations
Please refer to “Symbols” on page 17 and
“Abbreviations” on page 19 for explanations.
Notice
This document is provided for customer information
only, and will not be updated or exchanged without
customer request.
WARNING!
A WARNING statement provides important
information about a potentially hazardous
situation which, if not avoided, could result in
death or serious injury.
CAUTION!
A CAUTION statement provides important information
about a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury to the
user or patient or in damage to the equipment or other
property.
Note: A NOTE provides additional information
intended to avoid inconvenience during
operation.
Part Number: 9039994, 2nd edition
8Operating Instructions Apollo SW 4.n
Introduction
For Your Safety and that of Your
Patients
Strictly follow these Operating Instructions
WARNING!
Strictly follow these Operating Instructions.
Any use of the medical device requires full
understanding and strict observation of all
portions of these instructions. The medical device
is only to be used for the purpose specified under
“Indications for Use” on page 13 and “Intended
Use” on page 13, and in conjunction with
appropriate patient monitoring (see page 14).
Strictly observe all WARNING and CAUTION
statements throughout these Operating
Instructions and all statements on medical device
labels. Non-compliance with these WARNING and
CAUTION statements constitutes a use of the
medical device which is not in accordance with its
Intended Use.
Maintenance
WARNING!
The device must be inspected and serviced
regularly by properly trained service personnel.
Repair of the device may also only be carried out
by properly trained service personnel.
We recommend that a service contract be
obtained with DrägerService and that all repairs
also be carried out by them. We recommend that
only authentic Dräger repair parts be used for
maintenance. Otherwise, the correct functioning
of the device may be compromised.
Accessories
Not for use in areas of explosion hazard
WARNING!
This medical device is neither approved nor
certified for use in areas where combustible or
explosive gas mixtures (e.g. O
environments) are likely to occur.
Safe connection with other electrical equipment
WARNING!
Risk of electric shock.
Electrical connections to equipment not listed in
these Operating Instructions should only be made
following consultations with the respective
manufacturers of the equipment involved.
WARNING!
Risk of electric shock.
A test for leakage current must be performed by
qualified biomedical engineering personnel
before use if the Apollo is interfaced with other
equipment.
WARNING!
Risk of explosion, fire.
If an oxygen leak is suspected within or near the
anesthesia machine, do not initiate operation.
Disconnect all oxygen supplies and contact a
trained service technician.
2 or agent-enriched
Part Number: 9039994, 2nd edition
WARNING!
Only the accessories indicated on the list of
accessories (8603528) have been tested and
approved for use with the medical device.
Accordingly it is strongly recommended that only
these accessories be used in conjunction with the
specific medical device. Otherwise the correct
functioning of the medical device may be
compromised.
Operating Instructions Apollo SW 4.n9
Risk of use error.
Various potentially dangerous situations may
occur which demand the attention of properly
trained personnel.
The anesthesia machine may only be used under
the supervision of qualified medical personnel so
that assistance can be provided immediately in
the event of any malfunctions.
WARNING!
Introduction
WARNING!
Risk of fire.
In order to reduce a fire hazard, explosive
anesthetics, such as ether or cyclopropane, must
not be used.
WARNING!
Risk of device failure and/or danger to patient.
Magnetic fields may negatively influence the
proper function of the anesthesia machine, thus
endangering the patient.
Apollo must not be used with magnetic resonance
imaging (MRT, NMR, NMI)!
WARNING!
Risk of fire.
To prevent a fire hazard, drugs or other
substances based on flammable solvents, such as
alcohol, must not be introduced into the patient
system. In addition oxygen must be vented away
from the surgical site when using electrosurgical
equipment.
Adequate ventilation must be ensured if highly
inflammable substances are used for disinfection.
Note on EMC/ESD risk for the device function
General information on electromagnetic compatibility
(EMC) pursuant to the international EMC standard
IEC 60601-1-2: 2001
WARNING!
Pins of connectors identified with the
ESD warning symbol shall not be
touched and not be connected unless
ESD precautionary procedures are used.
Such precautionary procedures may include antistatic clothing and shoes, the touch of a ground
stud before and during connecting the pins, or the
use of electrically isolating and anti-static gloves.
All staff involved in the above shall receive
instruction in these procedures.
WARNING!
Risk of electric shock.
Connecting devices to the auxiliary
outlets of the anesthesia machine can
cause an increase in leakage current
beyond permissible values if the
protective conductor of a device fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting a
device (or devices) increases the leakage current
to a value which exceeds the permissible value,
do not use the auxiliary outlets of the anesthesia
machine: use a separate wall socket.
The system must meet the requirements for
medical equipment in accordance with
IEC/EN 60601-1-1 and IEC/EN 60601-1-2.
Part Number: 9039994, 2nd edition
Electromedical devices are subject to special
precautionary measures concerning electromagnetic
compatibility (EMC) and must be installed and put into
operation in accordance with the EMC information
included, see page 267.
WARNING!
Portable and mobile radio communication
equipment such as cellular radio telephones can
interfere with electromedical devices (see “EMC
declaration” on page 267).
10Operating Instructions Apollo SW 4.n
Introduction
Part Number: 9039994, 2nd edition
Safe networking of computers
When networking with electrical devices, the operator
is responsible for ensuring that the resulting system
meets the requirements set forth by the following
standards:
–EN 60601-1 (IEC 60601-1)
Medical electrical equipment
Part 1: General requirements for safety
–EN 60601-1-1 (IEC 60601-1-1)
Medical electrical equipment
Part 1-1: General requirements for safety
Collateral standard: Safety requirements for
medical electrical systems
–EN 60601-1-2 (IEC 60601-1-2)
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic compatibility;
Requirements and tests
–EN 60601-1-4 (IEC 60601-1-4)
Medical electrical equipment
Part 1-4: General requirements for safety
Collateral standard: Programmable electrical
medical devices
Follow Assembly Instructions and Instructions for Use.
Patient safety
The design of the medical device, the accompanying
literature, and the labeling on the medical device take
into consideration that the purchase and use of the
medical device are restricted to trained professionals,
and that certain inherent characteristics of the medical
device are known to the trained operator. Instructions,
warnings, and caution statements are limited,
therefore, largely to the specifics of the Dräger design.
This publication excludes references to various
hazards which are obvious to a medical professional
and operator of this medical device, to the
consequences of medical device misuse, and to
potentially adverse effects in patients with abnormal
conditions. Medical device modification or misuse can
be dangerous.
Patient monitoring
The operators of the medical device must recognize
their responsibility for choosing appropriate safety
monitoring that supplies adequate information on
medical device performance and patient condition.
of medical device performance and patient condition to
simple, direct observation of clinical signs.
The responsibility for the selection of the best level of
patient monitoring lies solely with the medical device
operator.
General Warnings and Cautions
The following WARNINGS and CAUTIONS apply to
general operation of the Apollo. WARNINGS and
CAUTIONS specific to subsystems or particular
features appear with those topics in later chapters of
these Operating Instructions or in the device-specific
Instructions for Use.
Accessories in sterile packaging
Do not use accessories in sterile packaging if the
packaging has been opened, damaged or if there are
other signs that the accessories are not sterile.
Reprocessing and resterilization of single-use
accessories is not permitted.
WARNING!
Each institution and user has a duty to
independently assess, based on its, his or her
unique circumstances, what components to
include in an anesthesia system. However, Dräger,
in the interest of the patient safety, strongly
recommends the use of an oxygen analyzer,
pressure monitor, volume monitor, and end-tidal
CO
2 monitor in the breathing circuit at all times.
CAUTION!
Risk of patient injury.
An incorrect diagnosis or misinterpretation of
measured values, or other parameters, may endanger
the patient.
Do not base therapy decisions on individual
measured values or monitoring parameters only.
Patient safety may be achieved through a wide variety
of different means, ranging from electronic surveillance
Operating Instructions Apollo SW 4.n11
Introduction
WARNING!
Risk of patient injury.
If ventilation of the patient is no longer assured
due to an obvious fault in the equipment, the
patient must immediately be ventilated with a
separate emergency ventilator.
Always keep a manual ventilator at hand.
WARNING!
Risk of burns.
Conductive breathing hoses or face masks may
cause burns during HF surgery.
Do not use these types of hoses and masks in
combination with HF surgery.
CAUTION!
Risk of mechanical failure.
The shock and vibrations caused by transportation
may lead to a mechanical failure. The application of a
wall or ceiling mounting is designated for buildings.
CAUTION!
Risk of device failure.
Compressed gas supply (pipeline supply or cylinder):
To avoid damaging the device(s) attached to a gas
supply, use only medical gases. Pay particular
attention to national and international standards
regulating the use of medical gases.
Do not use the anesthesia machine for mobile
facilities such as ambulances, helicopters, or ships.
CAUTION!
Risk of physical injury.
To avoid physical injury, e.g. pinching, pay special
attention to edges, moving parts, and corners when
working with
–drawers,
–the ventilator module,
–doors,
–the writing tray,
–swivel arms for mounted devices,
–gas cylinders,
–vaporizer units,
–CLIC absorbers and CLIC adapters,
as well as other accessories.
Part Number: 9039994, 2nd edition
12Operating Instructions Apollo SW 4.n
Introduction
Part Number: 9039994, 2nd edition
Indications for Use
The Apollo is indicated as a continuous flow
anesthesia system. The Apollo may be used for
manually assisted or automatic ventilation, delivery of
gases and anesthetic vapor, and monitoring of oxygen
and CO
2 concentration, breathing pressure, respiratory
volume, and anesthetic agent concentration and
identification.
For USA: Rx only.
Caution: Federal Law restricts this device to sale by or
on the order of a physician.
Intended Use
WARNING!
Risk of device failure and/or danger to patient.
If the intended use of this anesthesia machine is
not adhered to, it may fail and/or the patient may
be endangered.
Use the anesthesia machine only as specified in
the intended use of these Operating Instructions.
The Apollo is an inhalation anesthesia machine for use
in operating, induction, and recovery rooms. It can be
used with rebreathing systems, semi-closed to virtually
closed systems with low flow and minimal flow
techniques, and non-rebreathing systems (with the
Auxiliary Common Gas Outlet).
It may be used with O
medical gas pipeline system or by externally mounted
gas cylinders. Anesthetic agent can be delivered via
vaporizers mounted on the machine.
The Apollo is equipped with a compact breathing
system, providing fresh-gas decoupling, PEEP, and
pressure limitation. It has an electrically driven and
electronically controlled ventilator.
Ventilation modes
–Volume-controlled ventilation in Vol u me Mo de.
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional)
–Pressure-controlled ventilation in Pressure Mode.
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional)
2, N2O, and Air supplied by a
–Manual Ventilation Man.
–Spontaneous Breathing Spont.
–Pressure-Assisted Spontaneous Breathing in
Pressure Support CPAP (optional)
–Volum e AF (Volume Mode AutoFlow) (optional).
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional).
The following measured values are displayed
–Peak pressure PEAK,
Mean pressure P
MEAN,
Plateau pressure PLAT,
Positive end-expiratory pressure PEEP
–Expiratory minute volume MV,
Difference between insp. and exp. minute volume
MV
LEAK,
–Patient compliance C
Tidal volume V
PAT,
T,
Breathing rate Freq.
–Inspiratory and expiratory concentration of O
N
2O, anesthetic gas, and CO2
2,
–Difference between insp. and exp. O2
concentration Δ
O2
Optional:
–Functional oxygen saturation SpO
2,
Pulse rate Pulse
The following parameters are displayed as curves
–Airway pressure P
AW
–Inspiratory and expiratory flow
–Inspiratory and expiratory concentration of O
CO
2, and anesthetic gas
2,
Optional:
–Plethysmogram
–PAW-V loops and V-Flow loops
The following are displayed as bar graphs
–Inspiratory, expiratory, and leakage tidal volume
–Volumeter
–Pressure
–Low-flow wizard for indicating fresh-gas utilization
(optional)
Trends showing the measured values over time and a
logbook are also available.
Operating Instructions Apollo SW 4.n13
Introduction
Monitoring
By means of adjustable alarm limits which can
automatically be adapted to the momentary ventilation
situation.
With monitoring for
–Airway pressure P
AW
–Expiratory minute volume MV
–Apnea
–Inspiratory and expiratory anesthetic gas
concentration
–Detection of anesthetic gas mixtures
(simultaneous detection of up to two anesthetic
agents)
–Inspiratory O
–Inspiratory and expiratory CO
2 and N2O concentrations
2 concentrations
–Special alarm response in Bypass Mode
–Automatic agent alarm activation for multiples of
MAC (xMAC)
Optional:
–Oxygen saturation
–Pulse rate Pulse
WARNING!
Risk of electric shock.
Connecting devices to the auxiliary outlets of the
anesthesia machine can cause an increase in
leakage current beyond permissible values if the
protective conductor of a device fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting a
device (or devices) increases the leakage current
to a value which exceeds the permissible value,
do not use the auxiliary outlets of the anesthesia
machine: use a separate wall socket.
The system must meet the requirements for
medical equipment in accordance with
IEC/EN 60601-1-1 and IEC/EN 60601-1-2.
The system must meet the requirements about medical
electrical equipment in accordance to IEC/EN
60601-1-1 and IEC/EN 60601-1-2.
MEDIBUS Protocol
MEDIBUS is a software protocol for use in transferring
data between the Apollo and an external medical or
non-medical device (e.g. hemodynamic monitors, data
management systems, or a Windows-based computer)
via the RS-232 interface (see MEDIBUS Instructions
for Use of the Apollo, part number 9037426, 6th edition
or higher).
WARNING!
Risk of patient injury.
Data transferred via MEDIBUS interfaces is for
information only and is not intended as a basis for
diagnosis or therapy decisions.
Part Number: 9039994, 2nd edition
14Operating Instructions Apollo SW 4.n
Accessory weight limits
g
A
The following figures specify the maximum safe weight
limits for accessories mounted to the Apollo.
In addition to the arm-mounted accessory weight limits,
the following mounted accessory weights may not be
exceeded:
–30 lbs. maximum on either side of machine for
accessories mounted close to the machine side.
–100 lbs. maximum for accessories mounted to the
top shelf.
Figure 2. Accessory Weight Limits - Apollo Left Side
Apollo Left Side
Accessory Option
Introduction
Option
Weight
30 lb. / 13.6 kg
Part Number: 9039994, 2nd edition
CAUTION: POSSIBLE TIP
OVER HAZARD IF MOUNTING
ACCESSORIES EXCEED
APPROVED LIMITS
CAUTION: Maximum weight per arm 30 lb./ 13.6 kg.
20.0 in.
50.8 cm
pproved
Mounting Limits
8.0 in.
12.0 in.
16.0 in.
40.6 cm
30.5 cm
20.3 cm
Mount Arm Length
4.0 in.
10.2 cm
25 lb. / 11.4 kg
20 lb. / 9.1 kg
15 lb. / 6.8 kg
Option
ht
Wei
10 lb. / 4.5 kg
5lb./2.3kg
0lb./0kg
OP00695
Operating Instructions Apollo SW 4.n15
Introduction
A
Figure 3. Accessory Weight Limits - Apollo Right Side
Option
Weight
30 lb. / 13.6 kg
Apollo Right Side
Accessory Option
25 lb. / 11.4 kg
20 lb. / 9.1 kg
OP00694
CAUTION: POSSIBLE TIP
15 lb. / 6.8 kg
OVER HAZARD IF MOUNTING
ACCESSORIES EXCEED
APPROVED LIMITS
CAUTION: Maximum weight per arm 30 lb./ 13.6 kg.
10 lb. / 4.5 kg
5 lb. / 2.3 kg
CAUTION!
Risk of device failure.
If the anesthesia machine is operated when tilted,
components may be damaged or may function
improperly.
Do not operate the anesthesia machine if it is tilted
more than 5°.
0lb./0kg
pproved
Mounting Limits
8.0 in.
12.0 in.
16.0 in.
4.0 in.
10.2 cm
20.3 cm
30.5 cm
40.6 cm
20.0 in.
50.8 cm
Mount Arm Length
CAUTION!
Risk of inadvertent movement.
If not properly secured, the device may move
inadvertently during operation.
Apply the brakes on the device to ensure it cannot be
moved accidentally during operation.
Part Number: 9039994, 2nd edition
CAUTION!
Risk of injury.
If mounting accessories exceed the approved limits,
the anesthesia machine may tip over.
Maximum weight per arm = 30 lbs.
16Operating Instructions Apollo SW 4.n
Symbols
The following symbols appear on the Apollo and are
defined below.
Introduction
SymbolExplanation
SymbolExplanation
Suppress alarm tone for 2 minutes; change
priority of technical alarms and acknowledge
them
Display standard screen
Display the three basic screens in
succession
Standby/operation switch
_
Access more user options/screens
Pulse rate
Action in progress
Upper and lower alarm limits
Upper alarm limit only
Lower alarm limit only
Upper and lower alarm limits disabled
Consult accompanying documents.
Exit menu, return to preceding menu
Non-rebreathing system at common gas
outlet
Remaining battery capacity
(uninterruptible power supply UPS)
Manual ventilation
Automatic ventilation
Connector for pipeline gas supply
Backup gas cylinder
UL test mark
Plug system for vapor units
Connection for halogen lamp
Surface hot; do not touch.
Part Number: 9039994, 2nd edition
>
<
Upper alarm limit disabled
Lower alarm limit disabled
Alarm limit or measuring function disabled
4-digit password entered
Protection class type BF (body floating)
Connection for potential equalization
ESD warning label
Leakage current label
Warning: Do not connect the mains of
external display(s) to AC outlets. Use
external power outlets.
Interference
Battery supply
Alarm tone suppressed for 2 minutes
Operating Instructions Apollo SW 4.n17
Introduction
SymbolExplanation
Alarm monitoring inactive
Alarm monitoring temporarily inactive
Apnea alarm disabled
Caution!
Rotary knob
System power switch
Physical injury
Mains voltage
18Operating Instructions Apollo SW 4.n
Part Number: 9039994, 2nd edition
Abbreviations
Introduction
Part Number: 9039994, 2nd edition
List of abbreviations used in the software
and on the device
Abbreviation Explanation
Agent/agent Anesthetic gas
Air/AIRCompressed air for medical use
APLAdjustable Pressure Limitation
Aux CGOAuxiliary Common Gas Outlet
BWBody weight
CALCalibration
CO
2Carbon dioxide
COM1
COM2
CPAPContinuous Positive Airway Pressure
C
PATPatient compliance
C
SYSSystem compliance
ΔO
2Difference between inspiratory and
ΔPPSDifference in pressure to PEEP in
ΔVTDifference between inspiratory and
Des.Desflurane
etCO
2End-expiratory CO2 concentration
Enf.Enflurane
exp.Expiratory
FGFresh gas
FiCO
2Fractional inspiratory CO2
FiO
2Fractional inspiratory O2 concentration
Freq./freq.Frequency
Freq
MINMandatory minimum frequency in
Hal.Halothane
HFHigh frequency
I:ERatio of inspiration time to expiration
Interfaces used as MEDIBUS
interfaces
expiratory O
2
Pressure Support mode
expiratory tidal volume
concentration
Pressure Support mode
time
Abbreviation Explanation
insp.Inspiratory
inDesInspiratory desflurane concentration
inEnfInspiratory enflurane concentration
inHalInspiratory halothane concentration
inIsoInspiratory isoflurane concentration
inSevInspiratory sevoflurane concentration
INOPInoperable
Iso.Isoflurane
Leak
SYSSystem leakage
MACMinimum Alveolar Concentration
Man.Spont.,
Manual/Spontaneous breathing
MAN/SPONT
MVExpiratory minute volume
MV
LEAKDifference between inspiratory and
expiratory minute volume
N
2ONitrous oxide
O
2Oxygen
O
2+O2 flush
PAWAirway pressure
PAW-V loopPressure-Volume Loop
PEAKPeak pressure
PEEPPositive end-expiratory pressure
P
INSPInspiratory pressure in Pressure Mode
PLATPlateau pressure
plethPlethysmogram
P
MAXPressure limitation in Volume Mode
P
MEANMean pressure
Pressure/
Press. Mode
Press.
Support/
Pressure Mode
Pressure-controlled ventilation
Pressure Support mode
Pressure-assisted ventilation
Press. Supp.
Sev.Sevoflurane
SpO
2Functional O2 saturation
Operating Instructions Apollo SW 4.n19
Introduction
Abbreviation Explanation
Standby
Conf.
Standby configuration for default
values and settings
Sync./sync.Synchronization
TIP : TINSPRatio of inspiratory pause time to
inspiration time
TINSPInspiration time
Tri ggerTrigger level
TSLOPERise time
Vent. modeVentilation mode
V-Flow loop Volume flow loop
Volu me/
Vol. Mode
Volume Mode
Volume-controlled ventilation
Volu me AFVolume Mode AutoFlow
VTTidal volume
VT
INSPMeasured inspiratory tidal volume
List of general abbreviations
Abbreviation Explanation
kgKilogram
kPaKilopascal
lbs.Pounds
MAN/AUTOManual/mechanical ventilation
mbarMillibar
mmHgMillimeter of Mercury
mLMilliliter
NiBPNon-invasive blood pressure
NTPDNormal temperature pressure dry
(68 °F [20 °C], 1013 hPa [760 mmHg],
dry)
P
EIRP“Equivalent isotropic radiated power” of
the adjacent RF transmitter
ppmParts per million
PSPressure Support
psiPounds per square inch
RFRadio frequency
Abbreviation Explanation
ACAlternating current
AGSAnesthetic gas receiving system
ATPSMeasuring conditions at ambient
temperature, current atmospheric
pressure and with saturated gas
BTPSMeasuring conditions at body
temperature, current atmospheric
pressure and with saturated gas
cmH
2OCentimeter of water
CSPipeline gas supply / Piped medical
gas supply for O
2, N2O, Air, and
vacuum
EMCElectromagnetic compatibility
ESDElectrostatic discharge
HF surgeryHigh-frequency surgery
HMEHeat and moisture exchanger
hPaHectopascal
SORCSensitive oxygen ratio controller
T
EXPExpiratory time
UPSUninterruptible power supply
VACVacuum (e.g. for secretion aspiration)
Vol.%Percentage gas rate in relation to total
gas volume
VVolt
xMACMultiple of MAC
Units
Note: Throughout these Operating Instructions:
Ventilation pressures: cmH
Supply pressures: bar = kPa x 100
2, Air, N2O15Anesthetic gas receiving system AGS (optional)
Machine Rear view
11
10
System Components
1
9
2
3
8
4
5
7
6
Part Number: 9039994, 2nd edition
1Auxiliary outlet panel
2Cylinder tank yoke bar
3Filter for fan
4Type plate
5Cylinder support bar
6Scavenging nozzle
7AC power connector
8Gas supply block
9Connector for optional halogen lamp (remove cap
before use). Use the lamp specified in the list of
accessories only.
10Connectors (3) for backup gas cylinder pressure
sensors (covered; access from behind gas supply
block)
11Interface panel
Operating Instructions Apollo SW 4.n23
System Components
Gas supply block
1
23
564
1Connection for N2O cylinder
2Connection for Air cylinder
3Connection for O
4Connection for pipeline O
5Connection for pipeline Air
6Connection for pipeline N
2 cylinder
2
2O
Part Number: 9039994, 2nd edition
24Operating Instructions Apollo SW 4.n
Interface panel
System Components
ESD
Sensitivity
!
Caution!
Consult
Accompanying
Documents
1
2
3
1IV SystemConnection for Dräger IV System*
2COM2MEDIBUS interface
3COM1MEDIBUS interface
4SpO
*not for sale in the U.S.
2Socket for SpO2 sensor (optional)
IV System
COM 2
COM 1
2
SpO
4
Part Number: 9039994, 2nd edition
Operating Instructions Apollo SW 4.n25
System Components
Vaporizers (Optional)
Note: Before operating the vaporizer, pay special
attention to the Instructions for Use of the
vaporizer being used. Note especially the
vaporizer flow limits.
The Dräger Vapor anesthetic agent vaporizers are
used to enrich the fresh gas with a precisely metered
quantity of vapor from the liquid anesthetic agent
being used, i.e. Isoflurane, Halothane, Enflurane,
Sevoflurane, or Desflurane.
When using a Desflurane vaporizer, it must be
connected to mains power. The auxiliary power outlet
(IEC/EN 60320-2-2/F) near the vaporizer exclusion
system is provided for that purpose.
The vaporizers being used must comply with
standard ISO 8835-4. If the internal gas
measurement system fails, an independent
measurement system complying with ISO 21647
must be used.
CAUTION!
Risk of patient injury.
If the vaporizer is not correctly mounted, the freshgas flow will not be supplied with anesthetic agent
and the patient will not receive the correct
anesthesia.
Always double-check the position of the vaporizer,
make sure it is correctly mounted and do not mount
the vaporizer park holder close to the operable
vaporizer.
Vaporizer exclusion systems
The exclusion systems available for the Apollo are
described below.
(Optional)
The Dräger Interlock 2 system is used to ensure that
only one of two vaporizers can be used at a time. It
has a selector lever used to select which vaporizer is
enabled.
Moving the selector lever away from the desired
vaporizer allows that vaporizer to be used and the
other to be locked out of use.
26Operating Instructions Apollo SW 4.n
1
Note that the selector lever (1 in Figure 4) is shown in
the center position. This ensures that both vaporizers
are in the locked position. Also, this is the
recommended position for the selector lever when
moving the Apollo.
Selectatec (Optional)
The interlock system for the Selectatec is built into
the vaporizers. When a vaporizer is selected for use,
the interlocking index pins will protrude from the
sides of the vaporizer thereby not allowing the
adjacent vaporizer to be opened. For more specific
information on the Selectatec, refer to the Selectatec
Vaporizer’s instruction manual.
Dräger Auto Exclusion 2-Vaporizer
Mount (Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When one of the two vaporizers is selected for use
(opened), the interlock mechanism within that
vaporizer’s mounting system is activated
automatically, preventing the other vaporizer from
being used.
System Components
Part Number: 9039994, 2nd edition
EXCLUSION” vaporizers are compatible with
the Dräger Auto Exclusion 2-Vaporizer
Mount. See Ta b le 1 for the Auto Exclusion
Vaporizer technical data.
When using a Desflurane vaporizer, it must be
plugged into the auxiliary power outlet located on the
side of the machine above the vaporizer mount.
Table 1. Dräger Auto Exclusion Vaporizer Technical DataNote: Only vaporizers labeled as “AUTO
Normal
Operating
Range
Extended
Operating
Range
≤ 10 L/minDräger Vapor 2000
Instruction for Use
Manual’s delivered
concentration accuracy
values apply.
>10 ≤ 15 L/minDräger auto exclusion
vaporizer concentration
output accuracy may be
reduced.
Operating Instructions Apollo SW 4.n27
System Components
Mount (Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When any one of the three vaporizers is selected for
use (opened), the interlock mechanism within that
vaporizer’s mounting system is activated
automatically, preventing the other two vaporizers
from being used.
Note: Only vaporizers labeled as “AUTO
EXCLUSION” vaporizers are compatible with
the Dräger Auto Exclusion 3-Vaporizer
Mount. See Ta b le 1 for the Auto Exclusion
Vaporizer technical data.
When using a Desflurane vaporizer, it must be
plugged into the auxiliary power outlet located on the
side of the machine above the vaporizer mount.
Note: The Desflurane vaporizer should be installed
in the far left position (1 in Figure 5) with the
Dräger Auto Exclusion 3-Vaporizer Mount in
order to have optimum viewing area of the
display screen.
Figure 5. Dräger Auto Exclusion 3-Vaporizer MountDräger Auto Exclusion 3-Vaporizer
1
APL valve
The APL valve has two functions. It limits the
maximum pressure during manual ventilation and
exhausts excess gas into the scavenger system
during manual and spontaneous ventilation.
The APL valve is connected to the patient airway
through the ventilator. It functions only when the
ventilator is in Manual/Spontaneous mode.
The adjustment knob (1 in Figure 6) is used to select
between spontaneous and manual modes of
ventilation. It’s labeled to indicated approximate
pressure settings.
Part Number: 9039994, 2nd edition
28Operating Instructions Apollo SW 4.n
Pressure is released for spontaneous ventilation
when the adjustment knob is rotated fully
counterclockwise, when the index mark on the knob
lines up with the index mark on the bottom of the APL
valve (2 in Figure 6). Spontaneous ventilation
eliminates both resistance to patient exhalation and
the need to readjust back pressure.
For manual ventilation:
In manual mode, the APL valve adjustment knob can
be rotated to change the approximate pressure at
which gas will flow through the valve and into the
scavenging system. Clockwise rotation of the
adjustment knob increases the pressure, and
counterclockwise rotation of the adjustment knob
decreases the pressure. Pulling up on the APL valve
knob will temporarily relieve pressure.
Note: The APL valve is automatically excluded
from the breathing circuit whenever an
automatic ventilation mode is selected. It is
suggested that even in automatic ventilation,
the APL valve is adjusted to a pressure that
is safe for the patient.
System Components
Figure 6. APL ValveFor spontaneous ventilation:
1
2
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
If the APL valve becomes blocked due to e.g.
lines or cables being caught under the knob, the
patient may be endangered.
Route all cables away from the APL valve; do
not hang lines, hoses or cables, e.g. the sample
line, on or near the APL valve.
A manually operated O2 flush valve is located on the
front of the machine (1 in Figure 7). When actuated,
the valve delivers an unmetered flow of at least
35 L/min to the breathing system and breathing bag
while bypassing the ventilator. The Apollo does not
have to be switched on to use the O
1. To operate the O
2 flush, press the O2+ button.
2 flush.
Oxygen flows into the breathing system without
anesthetic gas as long as the button is pressed
in.
Figure 7. Location of O2 FlushO2 flush
1
Operating Instructions Apollo SW 4.n29
System Components
Auxiliary oxygen flow meter
flow of pure oxygen, used, for example in the delivery
of oxygen through a nasal cannula. Auxiliary oxygen
can be used in any ventilation mode, in Standby, or
even if the machine is switched off.
CAUTION!
Risk of inadequate pressure monitoring.
The optional auxiliary outlets are not pressure
monitored.
Pressure monitoring must be ensured by the
connected device.
Figure 8. Auxiliary Oxygen flow meterThe auxiliary oxygen flow meter delivers a metered
O
2
10
8
6
4
2
Writing table
(1 in Figure 9) which can be moved left or right or
folded down completely for convenient positioning.
To fold down the writing table, support the table with
one hand, then pull up on the release knob
(2 in Figure 9) and fold down. To bring the table up
again, swing it upward until it clicks into place.
Figure 9. Writing TableThe Apollo is equipped with a writing table
This chapter provides a description of the Apollo user
interface, which enables you to view and change
monitoring, ventilation, and status information using
keys and the rotary knob.
Main screen display
The screen display is organized into four functional
areas:
–Gas measurement
–Monitoring
–Fresh-gas delivery
–Ventilation
Figure 10 illustrates the general functional areas and
identifies the following smaller screen elements:
User Interface
Part Number: 9039994, 2nd edition
Operating Instructions Apollo SW 4.n35
User Interface
Figure 10. Standard Monitor Screen Layout
Measurement
Gas
Monitoring
1
12
2
3
4
5
6
7
8
Man
Spont
Vol
Mode
Vol . AF
Mode
11
Press
Mode
deliveryFresh-gas
Status field; displays information about the current operating mode
1
Numeric field for gas and agent measurement values
2
Alarm message field; displays alarm messages
3
User-configurable graphics field for curves and bar graphs
4
Numeric field for monitored parameter values
5
Monitoring/configuration buttons
6
Standard function keys; for selecting monitoring screens and silencing alarms
7
Prompt field; displays messages for the user
8
Ventilation parameter buttons
9
Ventilation mode keys
10
User-configurable monitoring area
11
9
Press.
Supp.
10
Aux
CGO
Part Number: 9039994, 2nd edition
Ventilation
Fresh-gas bar graphs (virtual flow tubes)
12
36Operating Instructions Apollo SW 4.n
User controls
Changes to system settings and screen displays are
made using the rotary knob, “keys” (keys with
permanently defined functions), and “buttons” (keys
with variable functions). All controls are described in
the following paragraphs.
Standard function keys
Three keys for standard functions are located on the
right side of the display screen (1 in Figure 11):
Press this key to suppress audible
alarms for 2 minutes; or to change
the priority of technical alarms and
acknowledge them.
Press this key to display in
succession the three available
monitoring screens: standard, data,
and trend.
User Interface
Figure 11. Standard Function Keys
1
Part Number: 9039994, 2nd edition
Press this key to display the
standard monitoring screen.
Rotary knob
(1 in Figure 12). It is the main control used to select
and confirm all monitoring and system settings:
–turn the rotary knob to change or select a value
or parameter (clockwise rotation increases a
value; counterclockwise rotation decreases a
value).
–press the rotary knob to set a value or confirm a
selection. If the selection is not confirmed, the
value or parameter will not change.
Vol. AF
Mode
Figure 12. Location of Rotary Knob & Standby KeyThe rotary knob is located on the bottom right side
Vol. AF
Mode
2
1
Operating Instructions Apollo SW 4.n37
User Interface
Standby key
The standby key (2 in Figure 12) is used to switch
between operating modes and Standby.
1. to set the machine to Standby mode, press the
standby key ><. Then press the rotary knob
to confirm.
The standby key is also used to enter monitoring
mode while in Standby (see “Monitoring mode” on
page 153 for more information).
Flow control knobs
and O
2 flow are located below their respective virtual
flow meters in the bottom left of the display
(1 in Figure 13). They are labeled and color-coded.
The oxygen flow control is also touch-coded with a
fluted knob.
–to increase flow, turn the appropriate flow control
knob counterclockwise
–to decrease flow, turn the appropriate flow
control knob clockwise
Figure 13. Flow Control KnobsThree control knobs for the adjustment of N2O, Air,
N 2 O
Air
_
++
1
Vol. AF
Mode
O
2
_
_
+
Part Number: 9039994, 2nd edition
38Operating Instructions Apollo SW 4.n
Ventilation control keys
keys located at the bottom of the screen.
The ventilation keys (1 in Figure 14) are used
primarily to select the ventilation mode:
>Man Spont<, >Vol Mode<, >Vol. AF Mode<,
>Press Mode<, and >Press. Supp.< (optional).
The key >Aux CGO< is used to select the optional
auxiliary common gas outlet.
Selecting Ventilation Mode or Aux CGO (optional)
1. Press the appropriate ventilation key. The key’s
LED and the status field will flash.
2. Press the rotary knob to confirm the selection.
The ventilation buttons (2 in Figure 14), located
above the keys, are used to set ventilation
parameters. These keys have variable functions,
depending on the operating status or ventilation
mode.
User Interface
Figure 14. Location of Ventilation Control KeysVentilation functions are controlled using two sets of
2
Vol. AF
Mode
1
Part Number: 9039994, 2nd edition
Setting/Selecting Ventilation Parameters
Example: setting PEEP
1. Press the button >PEEP<.
2. Turn the rotary knob until the desired value is
displayed.
3. Press the rotary knob to confirm the new value. If
the new value is not confirmed within
15 seconds, it automatically reverts to the
original value.
Operating Instructions Apollo SW 4.n39
User Interface
The majority of monitoring and configuration
functions are performed using the vertical column of
buttons along the right side of the screen
(1 in Figure 15). These keys have variable functions
and their labels change according to which
monitoring screen is selected (standard, data, or
trend). An arrow () on the button label indicates
that pressing that key will bring up a second set of
buttons with further user options.
Setting/Selecting Monitoring Functions
Figure 15. Location of Monitoring/Configuration KeysMonitoring/Configuration control keys
2
1
Example: change lower alarm limit for etCO
2:
1. Press the button >Alarm Limits<
(2 in Figure 15). The alarm limits menu is
displayed on the screen.
2. Turn the rotary knob to select the low alarm limit
value for etCO
2 (see Figure 16).
4. Turn the rotary knob until the desired alarm value
is displayed.
5. Press the rotary knob to confirm the new alarm
limit value.
6. Exit the alarm limits menu by either:
– confirming the close symbol >< with the
rotary knob, or
– pressing the >< key.
Vol. AF
Mode
Figure 16. Alarm Limits Menu3. Press the rotary knob to confirm the selection.
Alarm Limits Menu
Part Number: 9039994, 2nd edition
Vol. AF
Mode
40Operating Instructions Apollo SW 4.n
System LED indicators
A number of LED indicators are located at the bottom
of the front panel. They can light up green or red, or
can remain extinguished, to indicate gas supply and
machine power status.
The pipeline supply LEDs (1 in Figure 17 ) can be
either green, which indicates that the pipeline supply
is connected and pressure is adequate, or off
(extinguished). If the pipeline pressure transducer is
inoperable, the corresponding LED will flash green.
If the backup gas cylinder is connected and pressure
is adequate, the corresponding LED (2 in Figure 17)
will be green. If the backup gas cylinder is
connected, but the pressure is inadequate and the
pipeline supply is not available, the LED will flash
red. If the backup cylinder is not connected, the LED
will be dark (extinguished).
The Battery and AC Power LEDs (3 in Figure 17)
have two states: green or off (extinguished). The
LED that is green indicates the active power source.
Figure 17. LED Indicators
132
Vol. AF
Mode
User Interface
Part Number: 9039994, 2nd edition
Screen colors
Colors are used on the screen to indicate the status
of buttons and to highlight operating sequences.
ColorMeaning
Light green
Yellow
Dark green
Gray type
Ventilation Buttons
The ventilation buttons appear dark green when
operable and turn yellow when selected. Once the
value is changed and confirmed, the button turns
back to dark green.
–can be operated, leads to another
menu or operating function
–not yet active, presettings
–selected, can be changed or set,
not yet confirmed
–active parameter, can be
operated
–current selection (configuration
menu)
–cannot be operated
Operating Instructions Apollo SW 4.n41
User Interface
Some values change automatically when another
parameter button is selected, and those values will be
displayed in yellow in addition to the selected
parameter.
Values shown in gray indicate that:
–there is a discrepancy between set and actual
values, or
–specified accuracy is not being maintained.
Monitoring/Configuration Buttons
The monitoring buttons along the right side of the
screen appear light green when operable. When
selected, their color changes to dark green. They also
change to dark green when another submenu or
function is displayed.
Colors of Parameter Settings/Values in Menus
When the user selects a menu the parameters and
values will appear on a dark green background.
Currently selected submenus are framed in an orange
border. Parameters in gray type are inactive and
cannot be selected.
This chapter of the Operating Instructions provides
information on how to set up and install all system
components needed to prepare the Apollo for use.
The setup procedure shall be followed by the
performance of the periodic manufacturer’s
procedure.
WARNING!
Risk of patient injury.
Correct preparation of the anesthesia machine
is required to minimize the general risks
associated with the anesthesia machine.
Use only clean and disinfected parts and always
strictly follow the cleaning and assembly
instructions contained in these Operating
Instructions to prevent infection of patient or
user.
System Setup
Part Number: 9039994, 2nd edition
Preparation before first use
The Apollo has to be prepared before first use
according to the cleaning and disinfection guidelines
in the chapter “Disinfecting/Cleaning/Sterilizing” on
page 212.
Charging the battery for emergency operation
Apollo has a built-in uninterruptible power supply
UPS which maintains the power supply for at least 30
minutes (up to 90 minutes, depending on the
ventilation parameters) in the event of a mains power
failure, provided that the battery is fully charged.
Switching to battery power (UPS) takes place
automatically and is indicated on the screen by the
message: POWER FAIL.
The battery recharges automatically when the
anesthesia machine is plugged into the mains, but
only up to a maximum ambient temperature of 95° F/
35° C.
Operating Instructions Apollo SW 4.n45
System Setup
using the anesthesia machine for the first time:
Figure 18. Location of AC Power LEDThe battery must be charged for 10 hours before
1. Plug the mains power plug of the Apollo
anesthesia machine into the mains outlet.
The mains voltage must correspond to that specified
on the rating plate on the back of the machine.
The green LED labeled >
(1 in Figure 18).
2. Leave the Apollo connected to the mains for 10
hours. The anesthesia machine does not have to
be switched on.
Risk of device failure.
In the event of a power failure, any devices
connected to auxiliary power outlets will not be
powered by the UPS.
Pay special attention to all power indicators of
connected devices.
Risk of battery failure.
Allowing the battery to run low can damage it.
It must be charged at least every four weeks.
N AC Power< lights up
CAUTION!
WARNING!
Vol. AF
Mode
1
Installing the breathing system and
flow sensors
Figure 19. Installing the Flow Sensors1. Press the release button on the ventilator unit
and pull it out.
2. Loosen the three sealing screws on the ventilator
(1 in Figure 19) a quarter turn counterclockwise
with the wrench supplied.
3. Pull the breathing system up and out by the
handle (2 in Figure 19).
4. Unscrew the inspiratory and expiratory ports (4 in
Figure 19) by turning them counterclockwise.
5. Insert the flow sensors (3 in Figure 19) into the
two port connections on the breathing system,
46Operating Instructions Apollo SW 4.n
1
1
2
3
4
1
3
4
Part Number: 9039994, 2nd edition
with the electric connection on each sensor
facing down in the slot.
Note: Flow sensors must be recalibrated after
replacement by performing the power-on self
test (see chapter “Pre-use Checkout”).
(4 in Figure 19) so that the key on each port lines
up with the slot. Install the ports and tighten by
turning clockwise. Carefully seat the breathing
system onto the ventilator module, and tighten
the three sealing screws (1 in Figure 20) on the
ventilator cover.
System Setup
Figure 20. Installing the Breathing System6. Orient the inspiratory and expiratory ports
1
1
Filling and installing the absorber
A reusable absorber or the disposable CLIC
absorber can be used.
Reusable absorber
(1 in Figure 21).
2. Fill the absorber canister with fresh soda lime up
to the MAX mark.
WARNING!
Risk of injury.
Absorbent is caustic and is a strong eye, skin,
and respiratory tract irritant.
Use care when handling the absorbent to avoid
spills.
➡
Figure 21. Installing the Absorber Canister1. Push the insert fully into the absorber canister
1
➡
1
➥
2
➡
Part Number: 9039994, 2nd edition
CAUTION!
Risk of device failure.
It is recommended that Drägersorb 800 + or
Drägersorb FREE are used.
Do not use powdered soda lime, as a higher dust
load may impair functionality of the Apollo
anesthesia machine.
Operating Instructions Apollo SW 4.n47
System Setup
3. Fit the canister into position below the breathing
system, and turn counterclockwise as far as
possible (2 in Figure 21).
4. Slowly push in the ventilator module until it
engages.
5. Reset the soda lime change log to current date
by pressing the >soda lime changed< button,
see page 77.
If the breathing system is not to be used within the
next 24 hours:
–Only fill with soda lime immediately before use.
48Operating Instructions Apollo SW 4.n
Part Number: 9039994, 2nd edition
Disposable CLIC absorber (optional)
personnel, e.g. DrägerService.
System Setup
Figure 22. Installing the CLIC absorberThe appropriate adapter must be installed by trained
WARNING!
The disposable absorber must be clicked into
place before switching on the Apollo. This
ensures that the absorber is included in the leak
and compliance test for the anesthesia machine.
To click the absorber into place:
1. Press the button (1 in Figure 22); the mounting
swings open.
2. Before fitting, shake the disposable absorber,
e.g. by turning it upside down several times in
order to loosen the soda lime.
3. Remove the seal from the new disposable
absorber.
4. Slide the new disposable absorber onto the
mounting (2 in Figure 22).
5. Push the absorber into the anesthesia machine
until it engages.
6. Reset the soda lime change log to current date
by pressing the >soda lime changed< button,
see page 77.
1
2
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
The soda lime loses humidity. Generally, if the
humidity falls below a minimum set point, the
following undesirable reactions can occur,
independent of the type of lime and the
inhalation anesthetic being used:
–reduced CO2 absorption;
–increased heat build-up in the absorber and
thus, an increased breathing gas
temperature;
–formation of CO;
–absorption and/or decomposition of the
inhalation anesthetic.
These reactions could pose a danger to the
patient.
If using dry gases, only briefly flush the
anesthesia system and only if necessary.
Operating Instructions Apollo SW 4.n49
System Setup
Connecting the gas supply
WARNING!
Risk of explosion, fire.
Oil and grease may combine explosively with
oxygen or nitrous oxide. For this reason, oil and
grease must never come in contact with
pipelines, cylinders, cylinder valves, gauges,
fittings, etc., which conduct oxygen or nitrous
oxide within the machine.
CAUTION!
Risk of gas supply failure.
If all gas supplies (pipeline or cylinder) are not
connected correctly, the reserve system will not be
available in the event of a gas supply failure.
Make sure that all supplies are connected according
to the engraving on the gas supply block and the
illustrations at the back of the machine. After
connecting the supplies, ensure proper functionality.
CAUTION!
Risk of device failure.
Compressed gas supply (pipeline or cylinder): To
avoid damaging the device(s) attached to a gas
supply, use only medical gases. Pay particular
attention to national and international standards
regulating the use of medical gases.
Part Number: 9039994, 2nd edition
50Operating Instructions Apollo SW 4.n
Connecting pipeline supply of N2O, Air,
and O2
System Setup
Figure 23. Pipeline supply connections
WARNING!
Risk of patient injury.
Pipeline delivery hoses used between wall
outlets and anesthesia machines have caused
accidents when, during assembly, an oxygen
fitting was placed on one end of the hose and a
nitrous oxide fitting on the other end.
Carefully check hoses each time you connect a
machine to a wall outlet to ensure that both
ends of the hose are indexed for the same gas.
CAUTION!
Risk of device failure.
In order for the inhalation anesthesia machine to
operate as specified, the supply pressures at the
machine inlet must be within a range of 2.7 and
6.9 kPa x 100.
Make sure this is the case before initiating
operation.
1. Connect the gas fitting on each pipeline supply
hose to the corresponding fitting on the gas
supply block on the rear of the machine (see
Figure 23).
2. Connect the other end of the pressure hoses to
the terminal unit.
3. Make sure that all supplies are connected
correctly and functioning properly.
N2O
pipeline supply
connection
Air
pipeline supply
connection
2
O
pipeline supply
connection
Part Number: 9039994, 2nd edition
4. Ensure that the pipeline pressures are between
50 psi and 55 psi (see “Operating data” on
page 254 for ranges) by checking that the three
pipeline supply LEDs on the front machine panel
(1 in Figure 24) are illuminated green.
Figure 24. Location of pipeline supply pressure LEDsIf the pipeline supply pressure LEDs remain
dark, it means that the pressure is below 39 psi
or that the hoses are not connected properly.
Vol. AF
Mode
1
Operating Instructions Apollo SW 4.n51
System Setup
Connecting the backup gas cylinders for
O2, N2O, and Air
CAUTION!
Risk of gas supply failure.
Should the pipeline gas supply fail, the backup gas
cylinders on the anesthesia machine will provide a
reserve gas supply.
To prevent a complete gas failure, the backup gas
cylinders should remain on the device, valves
closed, in reserve even if the anesthesia machine is
connected to pipeline gas supply.
The Apollo is equipped with ANSI standard pinindexed hanger yokes for E-size cylinders to connect
backup gas cylinders to the anesthesia machine. The
yoke for O
are optional. All cylinder yokes are located on the
back of the machine as shown in Figure 25.
2 is standard, the yokes for N2O and Air
Figure 25. Cylinder Yoke Locations
WARNING!
Risk of gas supply failure.
When attaching a cylinder, ensure that only one
washer is installed between the cylinder and the
yoke gas inlet. The use of multiple washers will
inhibit the pin-index safety system. Be sure to
verify the presence of the index pins each time a
cylinder is installed. Never attempt to override
the pin-index safety system.
WARNING!
Risk of supply failure.
If pressure reducers not having the required
sensors and measurement features are used
instead of Dräger pressure reducers, backup
gas cylinders and their fill levels will not be
subject to alarm and monitoring functionalities
during the power-on self test and operation.
Without this monitoring, in the event of a loss of
the pipeline gas supply, the backup functionality
of the backup gas cylinders may not be
available. If monitoring for the remaining
capacity of the backup gas cylinders is not
available, the user must take other equivalent
measures.
Part Number: 9039994, 2nd edition
52Operating Instructions Apollo SW 4.n
The numbers in boldface in Step 1 below refer to
Figure 26.
System Setup
1. Connect a gas cylinder (1) to its yoke as
specified below:
– Remove the old washer (2) and install a new
washer on the seat of the yoke gas inlet
connection.
– Verify that the two index pins (3) below the gas
inlet (4) are present.
– Insert the head (5) of the gas cylinder into the
yoke from below. Ensure that the gas outlet
and indexing holes on the cylinder head align
with the gas inlet and index pins of the yoke
assembly (6). Engage the indexing holes with
the index pins.
– Turn the yoke handle (7) clockwise against the
cylinder head, so that the point of the yoke
handle bolt is aligned with the indent on the
back of the cylinder head. Verify that the
washer is in place, the index pins are
engaged, and the cylinder hangs vertically.
– Tighten the yoke firmly.
Note: When required, the cylinder valve (8) is
opened using the cylinder wrench (9) that is
provided.
Figure 26. Pin-Index Cylinder Mounting
7
6
9
OP00347A
Figure 27. Cylinder Connections
Air
Cylinder
Connection
3
4
8
2
5
1
Part Number: 9039994, 2nd edition
2. Connect the hose from each cylinder to the
corresponding ports of the gas supply block on
the back of the machine (see Figure 27).
3. Open the cylinder valves.
4. To ensure that the cylinder pressures are
adequate, check that the gauges above the
cylinder yokes indicate pressures recommended
in Table 2 on the next page. Also, if the cylinder
pressures are adequate, the cylinder pressure
LEDs on the front machine panel
(1 in Figure 28) are illuminated green.
that the cylinder pressure is inadequate or that the
cylinders are not connected properly.
5. Close the cylinder valves.
N2O
Cylinder
Connection
Figure 28. Location of Cylinder Pressure LEDsIf the cylinder pressure LEDs remain dark, it means
Vol. AF
Mode
2
O
Cylinder
Connection
1
Operating Instructions Apollo SW 4.n53
System Setup
WARNING!
Risk of gas supply failure.
If the valves remain open when connected to the
pipeline gas supply, gas may be withdrawn from
the backup gas cylinders.
Close backup gas cylinder valves whenever
pipeline gas supply is sufficient.
gas at the recommended pressures shown in
Table 2. (Indicated pressures are for E-size cylinders
at 70°F/21°C.) Cylinders measuring less than the
minimum recommended pressure (PSI - MIN) should
be replaced with new, full cylinders.
Table 2. Recommended Cylinder Gas PressuresCylinders attached to the hanger yokes must contain
GAS
Air1900/1311000/69
N
2O745/51600/42
O
21900/1311000/69
PSI/bar - FULL
(typical full load)
PSI/bar - MIN
Caution when handling O2 cylinders
WARNING!
Risk of explosion.
Oil and grease may combine explosively with
oxygen.
For this reason, oil and grease must never come
in contact with pipelines, cylinders, cylinder
valves, gauges, fittings, etc., which conduct
oxygen.
Note: Follow the Instructions for Use included with
the pressure regulator.
Part Number: 9039994, 2nd edition
54Operating Instructions Apollo SW 4.n
Connecting the scavenger system
The Apollo can be equipped with one of two kinds of
scavenger systems to provide the best match with
the hospital’s waste-gas disposal system. These
scavenger systems must comply with ISO 8835-3.
Connecting the anesthetic gas receiving
system AGS (Optional)
The anesthetic gas receiving system AGS is used
with vacuum waste-gas disposal systems.
CAUTION!
Risk of increased ambient gas concentration.
Ambient air may become contaminated with
anesthetic agent if the scavenger hoses are
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
System Setup
Part Number: 9039994, 2nd edition
sliding its bracket onto the two shoulder screws
on the side of the machine.
2. Connect one end of the transfer hose to the
fitting on the receiving system (1 in Figure 29).
3. Connect the other end of the transfer hose to the
scavenger connection on the back of the
anesthesia machine (2 in Figure 29).
4. Connect the waste-gas vacuum hose to the
output connection on the receiving system
(3 in Figure 29).
5. Connect the other end of the vacuum hose to the
hospital waste-gas disposal system.
Note: Activate hospital vacuum system before
using the receiving system.
Note: During use, the float indicator in the flow
indicator should stay between the upper and
lower marks. If necessary, regulate flow
using the flow adjustment valve
(4 in Figure 29).
Figure 29. Anesthetic gas receiving system AGS1. Install the receiving system on the machine by
2
1
4
3
Operating Instructions Apollo SW 4.n55
System Setup
WARNING!
Risk of patient injury.
If the AGS manifold is blocked, negative
pressure may result in the breathing system and
the patient’s lungs.
Always make sure the manifold is not blocked.
Note the Instructions for Use of the anesthetic gas
receiving system AGS.
Connecting the passive scavenger system (Optional)
recirculating exhaust systems. It is not meant to be
used with vacuum disposal systems.
CAUTION!
Figure 30. Passive Scavenger SystemThe passive scavenger system is used only with non-
Risk of increased ambient gas concentration.
Ambient air may become contaminated with
anesthetic agent if the scavenger hoses are
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
1. Install the passive scavenger on the machine by
sliding its bracket onto the two shoulder screws
on the side of the machine.
Note: Remove the socket from the scavenger hose
before connecting.
2. Connect one end of the transfer hose to the side
fitting on the scavenger (1 in Figure 30).
3. Connect the other end of the transfer hose to the
scavenger connection on the back of the
anesthesia machine (2 in Figure 30).
4. Connect the waste-gas hose to the bottom
connection on the scavenger (3 in Figure 30).
5. Connect the other end of the hose to the hospital
waste-gas disposal system.
Part Number: 9039994, 2nd edition
For detailed information on the passive scavenger
system, refer to separate Instructions for Use.
56Operating Instructions Apollo SW 4.n
Connecting the endotracheal aspiration system (Optional)
The optional endotracheal aspiration system for the
Apollo consists of a suction regulator and a bracket
that attaches to the side of the anesthesia machine.
The bracket is used to hold the regulator and a
suction bottle assembly of the customer’s choice.
1. Attach the endotracheal aspiration system
bracket to the side rail on the left side of the
anesthesia machine.
2. Mount the suction regulator (1 in Figure 31) onto
the bracket.
3. Prepare the suction bottle assembly according
the Instructions for Use provided with the bottle.
4. Install the bottle assembly in the slide mount
(2 in Figure 31) on the bracket.
5. Make all necessary connections between the
suction bottle, suction regulator, and piped
vacuum system as specified in the Instructions
for Use provided with the endotracheal aspiration
system.
System Setup
Figure 31. Endotracheal aspiration system Bracket
and Regulator
12
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
Do not apply unregulated suction to the patient
circuit when using this device.
Installing vaporizers
Install vaporizers as directed in the appropriate
Instructions for Use supplied with the vaporizers
available for use with the Apollo.
Operating Instructions Apollo SW 4.n57
System Setup
Installing the flexible arm for the
manual breathing bag
WARNING!
Risk of patient injury.
If incompatible materials are used in the patient
circuit, metabolic products may build up.
Breathing bags used on the Apollo must comply
with current ANSI standards.
1. Slide the bag arm assembly onto the breathing
bag port on the side of the breathing system
(1 in Figure 32).
2. Tighten the two thumb screws (2 in Figure 32) to
secure.
3. Attach the 90° fitting to the end of the bag arm
(3 in Figure 32), and attach the breathing bag to
the other end of the fitting.
Installing the patient system
WARNING!
Risk of burns.
Conductive breathing hoses or face masks may
cause burns during HF surgery.
Do not use these types of hoses and masks in
combination with HF surgery.
Figure 32. Breathing Bag Arm Connection
3
2
2
1
Part Number: 9039994, 2nd edition
CAUTION!
Risk of inadequate gas concentrations.
If the patient system components are not tightly
connected, ambient air will be added to the gas
mixture.
Make sure that all patient system components are
tightly connected.
58Operating Instructions Apollo SW 4.n
Note: Apollo (without accessories) has no
components containing latex. For latex-free
use: Use latex-free breathing bag and
breathing hoses
Note: Only use original sample line - other lines
may change the technical data for the
device.
Note: For sample lines available for use with the
Apollo, see the Accessories List, P/N
8603528.
Advisory when using accessories for ventilating
adults, children, and neonates
particular patient group, use a smaller breathing bag
and a smaller breathing hose set.
When ventilating, in particular neonates and children,
it is important to set the patient age properly.
Depending on the patients’ age setting, the MAC and
xMAC algorithms, the trigger sensitivity, and the
sensitivity of the flow measurement adapt
automatically.
System Setup
Table 1. Advisory when using accessoriesWhen applying the lower tidal volume limits of a
Tidal
volume
<50 mL0.5 Lneonatal
50 mL to 200 mL1 Lpediatric
201 mL to700 mL2 Ladult
>700 mL3 Ladult
Manual breathing
bag
Hose set
(or pediatric)
Part Number: 9039994, 2nd edition
Operating Instructions Apollo SW 4.n59
System Setup
Connecting the patient circuit
not use microbial filters in the inspiratory and
expiratory ports of the breathing system.
WARNING!
Risk of patient injury.
To protect the patient from particles and dust, a
filter must be used between the inspiratory limb
of the breathing circuit and the patient, i.e. Ypiece filter or filter on inspiratory limb.
1. Connect 22 mm (0.87 in) breathing hoses to the
inspiratory and expiratory ports or to the optional
microbial filters or filters on the breathing system
(1 in Figure 33).
2. Connect the other end of the breathing hoses to
the Y-piece (2 in Figure 33), or to the optional
filter on the Y-piece.
3. Make sure the breathing bag is attached.
4. Fit the new or empty water trap into its holder on
the front of the machine (3 in Figure 33) until it
clicks into place.
5. Connect one end of the sample line to the Luer
Lock on the water trap (4 in Figure 33).
6. Connect the other end of the sample line to the
Luer Lock on the Y-piece (2 in Figure 33). Ensure
that all Luer fittings are securely connected.
7. Make sure that the sample line is guided
correctly by using the sample line clip. This clip
should be attached to the expiratory port of the
breathing system.
Figure 33. Breathing Hose and Water Trap ConnectionsNote: If you use a microbial filter on the Y-piece, do
3
4
1
1
2
Part Number: 9039994, 2nd edition
CAUTION!
Risk of gas measurement failure and device failure.
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
Sample gas lines are single-use articles and must
not be disinfected.
60Operating Instructions Apollo SW 4.n
WARNING!
Risk of gas measurement failure.
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced at least every
four weeks.
WARNING!
Risk of gas measurement failure and device
failure.
If alcohol or cleaning agents/disinfectants come
in contact with the inside of the water trap, they
can damage the diaphragm and the
measurement system may fail as a result.
Do not use these substances and do not wash,
flush, or sterilize the water trap.
System Setup
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
If the APL valve becomes blocked due to e.g.
lines or cables being caught under the knob, the
patient may be endangered.
Route all cables away from the APL valve; do
not hang lines, hoses or cables, e.g. the sample
line, on or near the APL valve.
CAUTION!
Risk of contamination of the device.
Do not put the device into operation without a water
trap.
CAUTION!
Risk of inaccurate data.
Silicone can enter the measuring cuvette and distort
the gas measurement and it may fail as a result.
Do not spray the O-rings of the water trap holder
with silicone spray.
Operating Instructions Apollo SW 4.n61
System Setup
CAUTION!
Risk of inaccurate data.
Aerosols can damage the diaphragm and the
measurement system may fail as a result.
Do not use aerosols in the breathing system. The
water trap must not be used in combination with a
medical nebulizer.
Ventilating neonates
use pediatric hoses.
2. Use a microbial filter with a low resistance and a
low compliance. At this location in the breathing
circuit (1 in Figure 34), the resistance of a
microbial filter remains constant due to less
humidity in the inspiratory gas mixture. In the
expiratory hose, no filter should be used to avoid
the possibility of building up an intrinsic PEEP in
the patient’s circuit.
3. Slip the breathing hoses with the large sleeves
onto the inspiratory and expiratory sockets and
connect the small sleeves (2 in Figure 34) to the
Y-piece.
4. Use a tube adapter (3 in Figure 34) with
connection for the sample line (4 in Figure 34).
Tube adapters with a side port for connecting a
sample line support the CO
help flush the dead space in the Y-piece and tube
adapter.
5. Connect a 0.5 L breathing bag with socket to the
large connection sleeves. Slip the breathing
hose over the angled connector.
6. Connect the sample line to the side port of the
tube adapter.
To determine the system compliance and leakage,
firmly connect the Y-piece to the cone. When doing
so, use the patient circuit which will be used for the
next case.
2 measurement and
Figure 34. Ventilating neonates1. For tidal volumes VT of less than 50 mL:
1
4
2
insp.
3
2
exp.
Part Number: 9039994, 2nd edition
62Operating Instructions Apollo SW 4.n
WARNING!
Risk of patient injury.
If filters are blocked, sample gas taken from the
patient’s lungs could quickly lead to a negative
lung pressure.
When ventilating children and neonates, do not
use HME filters or other filters at the Y-piece in
connection with a tube adapter having a
connection for a sample line in the patient
circuit.
For measurement purposes, a permanent side
stream flow of 200 mL/min runs through the sample
line to the gas measurement system. In case of a
blocked HME filter or filter in this position at the
Y-piece, the measurement system would produce
negative pressure situations in the patient’s lungs.
System Setup
Figure 35. Ventilating neonates
Part Number: 9039994, 2nd edition
Ventilating children
200 mL: use pediatric hoses.
2. Slip the breathing hoses with the large sleeves
onto the inspiratory and expiratory sockets and
connect the small sleeves to the Y-piece
(1 in Figure 36).
3. Use an HME filter (2 in Figure 36) or filter with
connection for the sample line.
4. Connect the breathing bag with socket to the
large connection sleeves. Slip the breathing
hose over the angled connector.
5. Connect the sample line (3 in Figure 36) to the
HME filter or filter and water trap.
Figure 36. Ventilating children1. For tidal volumes VT between 50 mL and
3
1
2
1
insp.
exp.
Operating Instructions Apollo SW 4.n63
System Setup
WARNING!
Risk of patient injury.
If filters are blocked, sample gas taken from the
patient’s lungs could quickly lead to a negative
lung pressure.
When ventilating children and neonates, do not
use HME filters or other filters at the Y-piece in
connection with a tube adapter having a
connection for a sample line in the patient
circuit.
For measurement purposes, a permanent side
stream flow of 200 mL/min runs through the sample
line to the gas measurement system. In case of a
blocked HME filter or filter in this position at the
Y-piece, the measurement system would produce
negative pressure situations in the patient’s lungs.
Figure 37. Ventilating neonates
64Operating Instructions Apollo SW 4.n
Part Number: 9039994, 2nd edition
Connecting AC power
System Setup
Connecting auxiliary devices
The Apollo has two auxiliary outlets on the back of
the machine (1 in Figure 38). Each outlet is rated 4
amps and is protected by circuit breakers.
1. Connect the external device to an outlet on the
back of the machine
CAUTION!
Risk of device failure.
In the event of a power failure, any devices
connected to auxiliary outlets will not be powered by
the UPS.
Pay special attention to all power indicators of
connected devices.
CAUTION!
Risk of device failure.
If HF surgical devices are connected to the auxiliary
outlets, the leakage current may influence the
electronics of the anesthesia machine causing it to
fail.
Figure 38. Location of Auxiliary Outlets on Back of
Machine
1
1
ESD
Seneltivity
!
Caution!
Consult
Accompanying
Documents
IV System
2
SpO
COM 2
COM 1
2
2
!
Part Number: 9039994, 2nd edition
Do not connect HF surgical equipment to the
anesthesia machine’s auxiliary outlets.
There is also a dedicated (2 amp) outlet for a
Desflurane vaporizer on the side of the machine,
above the vaporizer mount (1 in Figure 39). This
outlet is protected by safety fuses.
1. Install the Desflurane vaporizer in its mount and
connect it to the outlet on the side of the
machine.
WARNING!
Risk of device failure.
If additional power extension sockets are
connected to the auxiliary outlets, device
internal electronics may be overloaded.
Do not connect additional power adapter
sockets to the auxiliary outlets.
Figure 39. Location of Desflurane Outlet
1
Operating Instructions Apollo SW 4.n65
System Setup
WARNING!
Risk of electric shock.
Connecting devices to the auxiliary
outlets of the anesthesia machine
can cause an increase in leakage
current beyond permissible values if
the protective conductor of a device fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting a
device (or devices) increases the leakage
current to a value which exceeds the
permissible value, do not use the auxiliary
outlets of the anesthesia machine: use a
separate wall socket.
The system must meet the requirements for
medical equipment in accordance with
IEC/EN 60601-1-1 and IEC/EN 60601-1-2.
Fuses for auxiliary outlets
If a circuit breaker is tripped (position 0):
1. Remedy the fault.
2. Press the switch on the circuit breaker into
position 1.
The circuit breaker is active again.
In cases of a blown safety fuse:
1. Remedy the fault.
2. Have the safety fuse replaced by an electrician.
Connecting the Apollo to ground
For e.g. intracardiac or intracranial surgery.
1. Connect one end of the grounding cable to one
of the connecting pins located at the back of the
anesthesia machine.
2. Connect the other end of the grounding cable to
the specified grounding point, e.g. on the
operating table or ceiling lamp.
3. Connect potential equalization point to the
auxiliary systems.
Part Number: 9039994, 2nd edition
66Operating Instructions Apollo SW 4.n
Connecting the Apollo to mains power
on the rating plate on the back of the machine.
System Setup
Figure 40. Location of AC Power LEDThe mains voltage must correspond to that specified
1. Plug the mains power plug of the Apollo
anesthesia machine into the mains outlet. The
green LED labeled >
N AC Power< lights up
(1 in Figure 40).
Note: The main circuit breaker for the machine is
located on the back of the machine below
the pipeline supply connections and behind
the cylinder mounts (2 in Figure 38).
Information about transport within
the clinic
Transport is defined as:
–Moving the device, other than for pure calibration
purposes.
–Removing the ceiling/wall-mounted variant from
the corresponding holder.
When transporting the anesthesia device:
–Only move the device using the handles
provided for this purpose.
–The anesthesia device should only be moved by
persons who are physically able.
–Dräger recommends that the anesthesia device
should be moved by two persons. This also helps
to improve maneuverability.
–Take special care not to bump or knock the
device when moving it over uneven surfaces,
around corners or at thresholds (e.g. in doors or
elevators).
–Do not attempt to drag the device over hoses,
cables, or other obstructions on the floor.
Vol. AF
Mode
1
Part Number: 9039994, 2nd edition
WARNING!
Risk of injury.
If handled incorrectly, the anesthesia machine
may become top-heavy and tip over causing
injury to patients and/or operators.
Observe the following points to prevent this
hazard.
Operating Instructions Apollo SW 4.n67
System Setup
To increase toppling stability:
–Remove all monitors and devices from the upper
storage area.
–Dismantle any additional mounted devices on
swivel arms or on the upper side of the device
(e.g. patient monitoring, data management
systems, syringe pumps, etc.)
–Clear the writing table and fold it down
completely.
–Position the breathing bag arm close to the
device.
–Push in the ventilator module and drawers.
CAUTION!
Risk of physical injury.
To avoid physical injury, e.g. pinching, pay special
attention to edges, moving parts and corners when
working with
–drawers,
–the ventilator module,
–doors,
–the writing tray,
–swivel arms for mounted devices,
–gas cylinders,
–vaporizer units,
–CLIC absorbers and CLIC adapters,
as well as other accessories.
The pre-use checkout procedure must be performed
to ensure that the Apollo is ready for use. This is a
recommended procedure. Follow the institution’s
policies for specific procedures.
If the Apollo fails any checkout routine, do not use
the machine until corrective action is taken. If
indicated, contact an authorized representative of
DrägerService for inspection of the unit.
WARNING!
Risk of device failure and/or patient injury.
Do not insert any additional components into or
modify the Apollo after the checkout procedure
has been started.
The anesthesia machine will not meet the
specified technical data.
Pre-use Checkout
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
Inappropriate hose length affects compliance
and can result in incorrect tidal volume delivery
to the patient.
Patient hoses must be adjusted to the
appropriate lengths prior to performing the leak
and compliance tests.
Checking the Workstation according to the Check List
The pre-use checkout procedure consists of a
manual procedure performed by the user, followed
by an automated self test. The manual procedure is
summarized in the check list that is displayed after
the machine is powered on.
Prerequisites
The device has been prepared (see “Cleaning and
Maintenance” on page 201) and assembled ready for
operation.
The pipeline supply and the power supply must be
connected.
Operating Instructions Apollo SW 4.n71
Pre-use Checkout
WARNING!
Risk of explosion, fire.
If an oxygen leak is suspected within or near the
inhalation anesthesia machine, do not initiate
operation.
Disconnect all oxygen supplies and contact a
trained service technician.
CAUTION!
Risk of inadvertent movement.
If not properly secured, the device may move
inadvertently during operation.
Apply the brakes on the device to ensure it cannot
be moved accidentally.
WARNING!
Risk of electric shock.
Connect the electrical power cable to a hospitalgrade live AC receptacle that accepts and
properly grounds the power cable. Do not use
“cheater plugs”. The term “cheater plug”
implies any and all electrical plugs or other
devices that can inhibit or prohibit the proper
grounding of the anesthesia machine.
Part Number: 9039994, 2nd edition
72Operating Instructions Apollo SW 4.n
1. Power on the machine by pressing the main
power switch on the front of the machine
(1 in Figure 41). An acoustic tone sounds.
All LEDs and the loudspeakers are tested.
Note: If all LEDs do not light up upon initialization,
contact DrägerService.
The initial screen appears after about
20 seconds. Apollo now loads its software and
tests its internal memory.
Check list
After about 35 seconds, a check list for manual tests
to be performed by the user is displayed
(see Figure 42).
Pre-use Checkout
Figure 41. Location of Main Power SwitchPower on
Part Number: 9039994, 2nd edition
list on the screen and as described in this
procedure.
If the self test has to be interrupted, e.g. for a quick
start in an emergency:
2. Press the button >Cancel Test< (1 in Figure 42),
and proceed as specified in “Emergency start” on
page 87.
The self test can be canceled up to ten consecutive
times.
WARNING!
Risk of device failure and/or patient injury.
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency, carry out a complete self test as
soon as practicable.
Figure 42. Check List Screen1. Check the components as instructed in the check
1
Operating Instructions Apollo SW 4.n73
Pre-use Checkout
Check the pipeline gas supply:
1. Verify that the pipeline supply LEDs on the front
panel light up green (1 in Figure 43). The LEDs
light up green when all pipeline supplies are
available and the pressures are between 39 psi
and 100 psi.
If the LEDs remain dark, it means that the pipeline
supply pressure is less than 39 psi or that the hoses
are not connected.
Note: If accessories are connected to the optional
O
2 or Air outlets on the gas supply block,
make sure they work correctly.
Check the cylinder gas supply
CAUTION!
Risk of gas supply failure.
If the valves are open when connected to the
pipeline gas supply, gas may be withdrawn from the
backup gas cylinders.
Close cylinder valves whenever the pipeline gas
supply is sufficient.
Figure 43. Location of pipeline supply and cylinder
pressure LEDs
Vol. AF
Mode
12
1. Using the provided cylinder wrench, slowly open
the cylinder valves.
2. Verify that the cylinder pressure LEDs light up
green (2 in Figure 43).
The LEDs light up green when the cylinder pressure
for O
2 and Air is over 290 psi and the pressure for
N
2O is over 145 psi.
The cylinder pressures are shown in the Check List
screen (see Figure 42).
3. Close the cylinder valves.
Note: A flashing cylinder pressure LED indicates
that the cylinder pressure transducer on the
back of the machine is disconnected.
The gas supplies available can be selected in the
menu Standby Conf., see page 179. Only these gas
supplies will then be checked during the self test and
an alarm issued in the event of a fault during normal
operation. The external oxygen supply and the O
cylinder cannot both be configured as not present at
the same time.
Open the backup gas cylinders which have been
configured as present for the self test and then close
them.
2
Part Number: 9039994, 2nd edition
74Operating Instructions Apollo SW 4.n
Test the O 2 flush:
O
2 must be connected for the following self test.
2. Press the >O
2+< button on the front of the
machine (1 in Figure 44).
3. Verify that the breathing bag inflates with an
audible flow.
Test the auxiliary O
2 flow meter:
1. Adjust the flow knob (2 in Figure 44) and make
sure the float moves freely over the full range of
the flow meter.
Test the function of the fresh-gas flow control
knobs:
1. Adjust the flow control knob for each available
gas (3 in Figure 44) and verify that the float
moves freely over the full range of the total flow
meter (4 in Figure 44).
Breathing bag:
Pre-use Checkout
Figure 44. Component Locations1. Occlude the Y-piece firmly onto the cone.
21345
Part Number: 9039994, 2nd edition
1. Verify that the breathing bag is properly installed
and ready for operation (5 in Figure 44).
Verify that the vaporizers are installed and ready
for use:
WARNING!
Risk of patient injury.
Before operating the vaporizer, pay special
attention to the Instructions for Use of the
vaporizer being used. Note especially the
vaporizer flow limits. For example, for the Vapor
2000, flow is 0.25 to 15 L/min, or 0.25 to 10 L/min
at concentrations higher than 5%.
If incorrect vaporizer settings are used, the
patient may be injured.
The vaporizers being used must comply with
standard ISO 8835-4. If the internal gas
measurement system fails, an independent
measurement system complying with ISO 21647
must be used.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 98).
Vapor 2000 is shown and described below.
Operating Instructions Apollo SW 4.n75
Pre-use Checkout
For the Dräger Interlock 2 System:
1. Vaporizers are mounted straight and seated
securely on the mounts.
2. Locking levers point to the left = locked position
(1 in Figure 45).
3. Check the sight glass (2 in Figure 45) and ensure
an adequate filling level.
4. Handwheel set to >0< and button is engaged
(3 in Figure 45).
5. Check the interlock mechanism.
Move the selector lever (4 in Figure 45) to the left
to lock the left vaporizer. Turn the handwheel on
the right vaporizer to a position other than >0<,
and make sure that the left vaporizer remains
locked in its >0< position.
6. Repeat test for other vaporizer.
7. Turn both handwheels to >0< positions.
For the Dräger Auto Exclusion System:
1. Vaporizers are mounted straight and seated
securely on the mounts.
2. Locking levers point to the left = locked position.
Figure 45. Dräger Interlock 2 System with Vapor 2000
Vaporizers
11
0
4
3
0
3
22
3. Check that the sight glass and ensure an
adequate filling level.
4. Handwheel set to >0< and the button is
engaged.
5. Check the interlock mechanism.
Turn the handwheel on one vaporizer to a
position other than >0<, and make sure that the
other vaporizer remains locked in its >0<
position.
6. Repeat test for other vaporizer.
7. Turn both handwheels to >0< positions.
Note: For three-vaporizer mounts, perform this test
for all three vaporizers.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 98).
Part Number: 9039994, 2nd edition
76Operating Instructions Apollo SW 4.n
Check the breathing system:
1. Make sure patient hoses are securely connected,
with optional filters inserted.
2. Make sure fresh absorbent is in the canister,
without violet discoloration.
procedure, the date and time can be logged
by pressing the >Absorb. changed< key on
the Check List screen
(1 in Figure 46). The label of the key then
changes to >Undo Change<, and can be
pushed again to undo the absorbent change
information. The absorbent change
information will be logged in the system
when the automatic test is started.
Note: Drain any water that may have collected in
the ventilator diaphragm.
For diaphragm location and disassembly instructions
see page 165.
Pre-use Checkout
Figure 46. Location of Absorb. Changed KeyNote: If the absorbent is changed during this
1
Part Number: 9039994, 2nd edition
WARNING!
Risk of device failure.
The correct operation of the anesthesia
machine will be impaired if condensation enters
the breathing system and/or the ventilator
diaphragm.
If condensation is a frequent problem, install
water traps in the breathing hoses.
WARNING!
Risk of strangulation.
If not positioned with care, hoses, cables, and
similar machine components may endanger the
patient.
Take special care when connecting the patient.
Operating Instructions Apollo SW 4.n77
Pre-use Checkout
WARNING!
Risk of patient injury.
The soda lime loses humidity. Generally, if the
humidity falls below a minimum set point, the
following undesirable reactions can occur,
independent of the type of lime and the
inhalation anesthetic being used:
–reduced CO
2 absorption,
–increased heat build-up in the absorber and
thus an increased breathing gas
temperature,
–formation of CO,
–absorption and/or decomposition of the
inhalation anesthetic.
These reactions could pose a danger to the
patient.
If using dry gases, only flush the anesthesia
system briefly, and only if necessary.
WARNING!
Risk of patient injury.
If the flow controls are left open, the ensuing
flow of gas may dry out the soda lime,
endangering the patient.
The flow control valves should be closed when
the machine is in the standby mode or when it is
switched off.
Note the Instructions for Use of the Drägersorb 800 +
or Drägersorb Free soda lime.
Verify that the scavenging system is ready for
use:
1. Check that the scavenger hose between the
AGS and the scavenger connection on back of
the machine is securely connected.
2. Check that the hose between the output
connector on the scavenger and the hospital
waste-gas disposal system is securely
connected.
3. On the AGS, make sure that the float is in
between the two marks in the sight glass on the
AGS.
Part Number: 9039994, 2nd edition
78Operating Instructions Apollo SW 4.n
Emptying the water trap
If the water trap needs to be drained or replaced, see
“Emptying the water trap” on page 227 and
“Replacing the water trap” on page 228.
Prepare the Apollo for the self test as follows:
Pre-use Checkout
1. Ensure that all flow controls are closed.
2. Occlude the Y-piece by inserting it onto the plug
on the bag arm assembly (1 in Figure 47).
3. Ensure that the sample line is connected
between the water trap and the Y-piece
(2 in Figure 47).
Check the function of the APL valve:
1. Set the APL valve to 30 (3 in Figure 47).
2. Press the O
2 flush button until system pressure
stabilizes; it should not exceed 45.
3. Release the O
2 flush button.
4. Verify that pressure does not fall below 15.
Notes on the use of bacterial/viral filters,
endotracheal tubes, Y-pieces, breathing hoses,
soda lime and other accessories for breathing
systems
Figure 47. Preparing for the Self Test
2
3
1
2
Part Number: 9039994, 2nd edition
WARNING!
Risk of patient injury.
When using additional components in the
breathing system or configurations which
deviate from the standard hose system, the
inspiratory and expiratory breathing resistances
can be increased to values which exceed the
standard requirements.
If configurations of this kind are used, the user
must pay special attention to the measured
values.
With spontaneous breathing, higher breathing
resistances mean that the patient must do more
breathing work.
Operating Instructions Apollo SW 4.n79
Pre-use Checkout
During volume-controlled ventilation, an increase in
breathing resistance has a slight effect on the applied
volume during inspiration. However, the peak
pressure PEAK is increased with the plateau
pressure PLAT remaining constant. The time
constant (RC) is increased in the expiration phase as
a result. If the expiration times are too short, the lung
might not be emptied completely, resulting in a
dynamic overfilling of the lungs (air trapping).
During pressure-controlled ventilation, an increase in
the airway resistance can reduce the inspiratory or
expiratory volume.
Before the self test is performed, the accessories
which are to be used for the application must be
connected. The expansion hoses must be drawn out
to the intended user length. This is the only way of
ensuring that the compliance is determined correctly
and a correct tidal volume is applied during volumecontrolled ventilation.
If coaxial hoses are used, leaks between the inner
and outer hose are not detected during the self test/
leak test.
Self test
If all checks in the Check List are completed
successfully:
1. Press the rotary knob or the >Start Self Test<
key on the check list screen (1 in Figure 48) to
begin the Apollo automated self test.
Figure 48. Check List Screen
1
Part Number: 9039994, 2nd edition
2
80Operating Instructions Apollo SW 4.n
Pre-use Checkout
The automatic self test lasts approximately
3 minutes. The bar graph at the top of the Self Test
screen shows the progress of the test (see Figure
49).
After the self test has been started, a double tone
(speaker test ”passed”) and a single tone (speaker
test in the power supply unit “passed”) sound one
after the other with the set alarm tone volume.
Note: If no tone is sounded, contact
DrägerService.
The tests that are performed are listed on the screen.
A clock symbol >< is displayed in the small circle
next to the component that is currently being tested.
As each component test is finished, the clock symbol
is replaced by a color code that indicates the result of
the test.
Errors discovered during the self test are marked
with yellow or red behind the respective test result.
An advisory window with information on how to
remedy the problem is displayed on the screen.
Test results are color-coded:
Figure 49. Self Test Screen
Part Number: 9039994, 2nd edition
GreenTest completed successfully.
YellowA non-critical fault was detected. The
anesthesia machine can be used with
restrictions.
Functions highlighted in yellow can
be confirmed with the >Accept< button which
is then displayed (Figure 50), e.g. speaker
failure. The anesthesia machine starts
operation without this function.
RedOperation of the anesthesia machine is
impossible or not permitted. The error must be
remedied and the test must be repeated.
The self-test can no longer be canceled at this
point.
Operating Instructions Apollo SW 4.n81
Pre-use Checkout
Interruption of the test is symbolized by an
exclamation mark.
WARNING!
Risk of device failure or patient injury.
Functions coded yellow do not meet with the
specified technical data.
The error should be remedied as soon as
possible.
WARNING!
Risk of device failure or patient injury.
Functions coded red must be remedied before
starting, e.g. if there is no O
2 supply.
The device cannot be operated in this state.
WARNING!
Risk of inadequate monitoring.
If the flow sensor, oxygen sensor, or gas sensor
is not operational, adequate substitute
monitoring must be ensured before starting the
anesthesia machine!
Special attention is required if operation is
initiated.
Figure 50. Non-Critical Error During Self Test
When the self test is completed, the system switches
Figure 51. Standby Screen Following Self-Test
to Standby. The results of the self test are indicated
on the screen by a color-coded circle (1 in Figure 51).
GreenFUNCTIONAL
Every component of the system is in satis-
1
3
factory operational order.
YellowCONDITIONALLY FUNCTIONAL
2
A non-critical fault was detected. Apollo may
be used, but call DrägerService or your local
authorized service organization.
EmptyThe self test was canceled.
In addition, a message containing instructions for
further procedures, if any, appears in the middle of
the screen (2 in Figure 51).
82Operating Instructions Apollo SW 4.n
Part Number: 9039994, 2nd edition
Pre-use Checkout
More specific results can be displayed by pressing
the >Self Test Results< button on the standby
screen (3 in Figure 51). The Self Test Results screen
is displayed
(Figure 52).
The Self Test Results screen contains the
>Absorb. changed< key (1 in Figure 52). If the
absorbent is changed between cases, this key can
be pressed to log the date and time. The label of the
key then changes to >Undo Change<, and can be
pushed again to undo the absorbent change
information. The absorbent change information will
be logged in the system when the Self Test Results
screen is exited.
Cancelling the self test:
To cancel the self test before completion, for
example, for a quick start in an emergency:
1. Press the >Cancel Test< key (2 in Figure 48),
and proceed as specified in “Emergency start” on
page 87. The self test can be canceled up to ten
consecutive times.
Figure 52. Self Test Results Screen
1
Part Number: 9039994, 2nd edition
WARNING!
Risk of device failure and/or patient injury.
Canceling the self test may lead to
malfunctions. Special attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, perform a complete
self test as soon as practicable.
System compliance test
The Apollo determines the current compliance of the
patient system with any filters, hoses, and a Y-piece.
The system compliance test result is indicated on the
Self Test result screen by the red/yellow/green
indicator and by posting the compliance value.
For data of the system compliance see
“Specifications” on page 253 and the Instructions for
Use of the devices specified in the accessories list.
The compliance value is used during volumecontrolled ventilation to correct for the reduction of
tidal volume due to system compliance. For more
information, see “Compliance compensation” on
page 109.
Operating Instructions Apollo SW 4.n83
Pre-use Checkout
Leak tests
The Apollo tests for leaks in the mechanical
subsystem and in the Man/Spont system.
Leak system
–Figure 53 shows the components tested in the
mechanical ventilation branch.
–This branch is tested with positive pressure.
–Leaks are indicated on the Self Test Results
screen by the yellow/green test result indicator
and by posting the leak value in mL/min (Leak
(system) test result in Figure 52).
Leak Man/Spont
–Figure 54 shows the components tested in the
Man/Spont leak test.
–This test is performed with sub-atmospheric
pressure.
–Leaks are indicated on the Self Test Results
screen by the red/yellow/green test result
indicator and by posting the leak value if it is over
150 mL/min (Leak (Man/Spont) result in Figure
52).
The Apollo determines the current leakage of the
breathing system and breathing hoses. The system
tolerates leaks of up to 150 mL/min.
Figure 53. Mechanical Ventilation Leak Test
Figure 54. Man/Spont Ventilation Leak Test
Part Number: 9039994, 2nd edition
Note: For leaks of more than 150 mL/min:
Check the components of the breathing
system and the breathing hoses. Repair any
leaks and repeat the leak test.
84Operating Instructions Apollo SW 4.n
Part Number: 9039994, 2nd edition
Locating and eliminating leaks
The self test incorporates a leak test. If this test is not
passed, the leaks must be remedied before
continuing the test by pressing the rotary knob. A
leak test can also be carried out later in Standby with
the >Leak Test< key.
Possible causes of leaks include:
–Absorber not firmly screwed to the breathing
system.
–APL valve is not firmly fixed to the breathing
system cover (damage) or not set to
30 hPa (cmH
–Manual breathing bag, breathing hoses, Y-piece,
or microbial filter are not connected correctly or
damaged.
–Flexible arm for manual breathing bag not fitted
correctly on the breathing system, sealing ring
soiled or damaged.
–Water trap not connected.
–Sample line for gas measurement not connected
or leaky (there may be a kinked bend in the
connections).
–Connections for the sample line for gas
measurement cracked or defective.
–O-ring of the inspiratory and expiratory ports
missing, soiled, or damaged.
–Flow sensors not fitted correctly or damaged,
rear O-ring missing.
–Breathing system cover not mounted correctly,
not all five sealing screws closed.
–Visible damage on valves or seals of the
breathing system metal valve plate.
–Breathing system not mounted correctly, not all
three sealing screws closed.
–Ventilator diaphragm defective or not fitted
correctly (Dräger legend must be visible from
above).
–15 mm (0.59 in) cone for connecting the Y-piece
scratched or damaged.
–Vaporizer fill or drain connections leaky or
opened, vaporizer not mounted correctly, O-ring
missing or handwheel not set to >0<.
2O).
Pre-use Checkout
Operating Instructions Apollo SW 4.n85
Pre-use Checkout
Additional suggestions to isolate components of the breathing system for leaks:
Carry out the described measures:
Patient Sample Line Isolation:
1. Remove the sample line for gas measurement
and seal the Luer Lock connection on the
Y-piece.
2. Perform leak test.
Exclude the breathing hoses from the leak test
1. Remove the patient circuit from the breathing
system.
2. Install a knwn tight 22 mm (0.87 in) hose
between the inspiration and expiration ports. The
breathing bag must be on the bag arm.
3. Perform leak test.
Isolation of Vaporizers:
1. Remove the vaporizers from the anesthesia
machine.
2. Perform leak test.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 98).
Part Number: 9039994, 2nd edition
86Operating Instructions Apollo SW 4.n
Emergency start
WARNING!
Risk of incorrect delivery.
The leak and compliance tests are not
performed if the self test is cancelled, and no
leak and compliance information is available.
The accuracy levels specified in the chapter
“Specifications” of this operating instructions
cannot be guaranteed.
Pre-use Checkout
The emergency start procedure shortens the self test
when the Apollo must be operational immediately.
Note: To prevent abuse of this feature, the
emergency start procedure can be
performed up to ten times in succession.
After ten cancellations, the system will not
allow another cancellation and a complete
self test must be performed.
1. Power on the anesthesia machine by pressing
the main power switch on the front of the
machine (1 in Figure 55).
2. Check that all vaporizers are closed.
3. Set an appropriate fresh-gas flow using the
oxygen flow control knob (2 in Figure 55). Verify
adequate flow by checking the total flow meter
(3 in Figure 55).
4. Start manual ventilation.
5. Continue manual ventilation while the software is
internally loaded and the electronics are tested.
After about 35 seconds, the Check List screen
appears.
Figure 55. Apollo Front Panel
2
3
Part Number: 9039994, 2nd edition
Operating Instructions Apollo SW 4.n87
Pre-use Checkout
screen (1 in Figure 56).
The machine runs through a minimal self test
that lasts about 10 seconds. Manual ventilation is
interrupted during this time, but spontaneous
breathing can continue.
Figure 56. Check List Screen6. Press the >Cancel Test< key on the Check List
Apollo is ready for operation about 1 minute after
initiating. The O
2 sensor is completely calibrated after
about 2 minutes.
WARNING!
Risk of device failure and/or patient injury.
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, carry out a complete
self test as soon as practicable.
After the minimal self test, the anesthesia machine
switches to Standby.
This chapter of the Operating Instructions
summarizes basic operation of the Apollo, including
starting operation, changing patients, and ending
operation. Specific information on setting ventilation
and monitoring parameters is provided in later
chapters of the Operating Instructions.
Typical operation
Operation of the Apollo begins with the standby
screen which is displayed after the initial self tests.
This screen allows the user to restore default settings
and enter the patient parameters needed to begin a
case.
Loading default settings
Operation Summary
Part Number: 9039994, 2nd edition
The default settings for fresh-gas delivery,
ventilation, and alarms are loaded in the standby
screen and can be modified in the standard
configuration if necessary.
These default settings are valid whenever the Apollo
is switched on. They can be changed and set as
required for the specific hospital concerned, see
“Configuring the default settings in Standby” on
page 179 for complete instructions.
1. Press the >Restore Default Settings< key on
the standby screen (1 in Figure 57).
2. Press the rotary knob to confirm the restore.
Entering the patient’s age
The set age influences the calculation of the MAC
value, the volumeter scale, the vertical axis of the
loops and ventilation monitoring graphs as well as
the alarm limits for (optional) SpO
the automatic volume adjustment of the Breathing
Sound Emulator (BSE) module during operation.
2 monitoring, and
Figure 57. Standby screen
123
In addition, the trigger sensitivities and software
algorithms for suppressing artifacts are also
modified, thus influencing the quality of ventilation in
modes supporting spontaneous breathing.
1. Press the >Age< key on the standby screen
(2 in Figure 57).
Operating Instructions Apollo SW 4.n91
Operation Summary
2. Turn the rotary knob until the correct patient age
is displayed, and press the rotary knob to
confirm.
The patient age parameter is available in the
standby screen as well as in all ventilation modes.
Changing the patient’s age during operation
immediately impacts the parameters described
above.
Entering the patient’s ideal body weight
(Optional)
The patient’s ideal body weight describes that portion
of the body relevant to setting the ventilation
parameters (the patient’s body weight minus the
assumed excess fat).
The set ideal body weight influences the ventilator
default settings for tidal volume Vfreq, as well as the alarm limits for expiratory minute
volume MV during operation.
T and frequency
1. Press the >Weight< button on the standby
screen (3 in Figure 57).
2. Turn the rotary knob until the correct weight is
displayed, and press the knob to confirm.
The patient weight parameter is available in the
standby screen as well as inall ventilation modes.
Changing the patient’s weight during Volume , Volume AF, Pressure, and Press. Support has no
influence on current ventilation settings.
Adjustment ranges and default values upon
delivery
ParameterAdjustment
range
Age<1 - 120 years40
Weight1lbto240lbs.
(1 kg to 120 kg)
Defalut value
upon delivery
--
Part Number: 9039994, 2nd edition
92Operating Instructions Apollo SW 4.n
Setting the fresh-gas flow
Operation Summary
1. Set the fresh-gas flow to desired levels using the
flow control knobs on the front panel of the
machine (1 in Figure 58).
CAUTION!
Risk of patient injury.
The use of minimum flow or low flow settings may
lead to the accumulation of metabolic products in
the breathing system.
To avoid this risk, use appropriate soda lime, or set
higher fresh-gas flows, and always use the gas
measurement module provided by the anesthesia
machine.
SORC (Sensitive Oxygen Ratio Controller)
The Apollo is equipped with an O
system to avoid hypoxic gas mixtures when N
2 minimum delivery
2O is
selected as the carrier gas.
Figure 58. Flow Control Knobs
Vol. AF
Mode
N 2 O
++
Air
_
O
2
_
_
+
1
Part Number: 9039994, 2nd edition
At flow rates of 200 mL/min and above, the N
2O
concentration can be freely set between 0 and 79%.
During an O
concentration in the fresh gas, so that the O
2 shortage, the SORC limits the N2O
2
concentration does not fall below 21%. When the
N
2O flow control is open and the O2 flow control is
closed (or O
SORC prevents N
2 flow is less than 200 mL/min), the
2O flow. During N2O failure, O2 can
still be administered.
The SORC is not active when Air is selected as the
carrier gas and 100% Air can be metered throughout
the entire flow range.
Fresh-gas Failure Detection
During operation, the Apollo checks that the piston
cylinder unit has a sufficient level of fresh gas.
If a sufficient level of fresh gas is not possible, the
system first displays the message "FGAS LOW OR LEAK".
Operating Instructions Apollo SW 4.n93
Operation Summary
In addition the alarm “PINSP. NOT ACHIEVED” or “VT
NOT ACHIEVED” is displayed if the system is unable
to maintain the defined ventilation.
To ensure continued ventilation, the anesthesia
machine will use ambient air to supplement the gas
volume if it is too low. This may change the gas
composition. Carefully check the gas composition.
WARNING!
Risk of patient awareness.
If a complete gas supply failure occurs, the
anesthesia machine will continue to function
with ambient air. However, anesthestic agents
will no longer be delivered and the inspiratory
gas composition will be diluted.
Carefully monitor the gas mixture and, if
necessary, use IV anesthetics.
DrägerService can change the behavior of the device
so that it does not use ambient air for supplementing
the gas volume. The device will then ventilate with
limited V
T or PINSP. if possible.
1. Increase the fresh-gas flow.
2. Seal any possible leaks.
Setting vaporizer concentration
Note: Refer to the appropriate Instructions for Use
for the vaporizer being used. Vapor 2000 is
shown and described below.
For the Dräger Interlock 2 exclusion system:
1. Lock the unused vaporizer by moving the
selector lever completely towards it. For
example, to lock the left vaporizer, move the
lever to the left (1 in Figure 59.).
2. With the handwheel set to >T< position on the
unlocked vaporizer, press the button and engage
the handwheel at >0< (2 in Figure 59.) Wait
5 seconds for the pressure to balance.
3. Press the button and turn the handwheel
counterclockwise to set the required anesthetic
gas concentration (3 in Figure 59.).
2. With the handwheel set to >T< position, press
the button and engage the handwheel at >0<.
Wait 5 seconds for the pressure to balance.
3. Press the button and turn the handwheel
counterclockwise to set the required anesthetic
gas concentration.
Setting ventilation mode
Set the ventilation mode as described in the chapter
“Ventilation” of this Operating Instructions.
Changing patients
Follow the steps below for successive patient cases.
Switch to standby mode
Operation Summary
Part Number: 9039994, 2nd edition
1. Press the standby key ><, and confirm with
the rotary knob.
The functions of the anesthesia machine are
switched off. All the current settings from the
previous case are retained, including fresh-gas
delivery and ventilation parameters, alarm limits, and
patient age and weight.
To activate the default settings instead of using the
current settings:
1. Press the >Restore Default Settings< key on
the standby screen, and press the rotary knob.
The default settings for gas delivery, ventilation, and
alarm limits are restored.
Enter the new patient’s age and ideal body weight
(optional) as instructed on page 92, and proceed.
WARNING!
Risk of patient injury.
Restored default settings may contain settings
inappropriate for a new patient.
After default settings have been restored, make
sure the ventilation monitoring settings are
appropriate to the patient connected.
Operating Instructions Apollo SW 4.n95
Operation Summary
Changing soda lime
A reusable absorber or the disposable CLIC
absorber can be used with the Apollo. The soda lime
must be exchanged, if:
–the soda lime in the absorber has turned violet.
The color indicator can regenerate slowly and
the soda lime may revert to white, but its
absorption capacity is nevertheless spent. You
should therefore dispose of used absorbers
immediately.
–the fractional inspiratory CO
FiCO
2 exceeds 5 mmHg.
Reusable absorber
1. Press the standby key ><, and confirm with
the rotary knob.
2. Swing the writing tray out of the way.
3. Press the release button on the ventilator unit,
and pull out the unit.
4. Turn the absorber canister counterclockwise and
pull it down and off.
2 concentration
5. Empty the used soda lime and refer to the
Instructions for Use of the soda lime for waste
removal and refilling.
6. Fill the absorber canister to upper mark with
fresh soda lime.
7. Fit the canister into position below the breathing
system, and turn it clockwise as far as possible.
8. Push the breathing system inwards until it clicks
into place.
9. Reset the absorbent change log to current date
by pressing the >soda lime changed< button.
Part Number: 9039994, 2nd edition
96Operating Instructions Apollo SW 4.n
Disposable CLIC absorber (Optional)
Operation Summary
The disposable absorber can be replaced during
operation. The valve in the mounting ensures that the
breathing system remains tightly sealed when the
absorber is removed.
Note: Since a leak test cannot be performed during
operation, no leak and compliance
information on the changed absorber is
available. Greater attention is required
during operation.
Replace the disposable absorber to ensure
continuous CO
Remove the spent absorber
1. Press the button (1 in Figure 60): the absorber
swings open sealing the breathing system so
that the ventilation can continue.
2. Slide the disposable absorber off the mount
(2 in Figure 60).
3. Dispose of the spent absorber.
Refer to the Instructions for Use of the CLIC
absorber for information on disposal.
2 absorption in the breathing system.
Figure 60. CLIC absorber
1
2
Part Number: 9039994, 2nd edition
Install the new absorber
1. Before fitting, shake the disposable absorber,
e.g. by turning it upside down several times in
order to loosen the soda lime.
2. Remove seal from new disposable absorber.
3. Slide the new disposable absorber onto the
mount (2 in Figure 60)
4. Push the absorber into the machine until it
engages.
5. Reset the absorbent change log to current date
by pressing the >soda lime changed< button
(only available in Standby mode).
Operating Instructions Apollo SW 4.n97
Operation Summary
Leak test
WARNING!
Risk of patient injury.
The system will be pressurized during the leak
test.
To prevent patient injury, do not perform the
leak test with a patient connected to the
anesthesia machine.
WARNING!
Risk of misleading data.
Changing the breathing hoses, vaporizers, or
soda lime can modify the calculated leak and
compliance values of the anesthesia machine
and influence the therapy settings.
Perform a leak test after the breathing hoses,
vaporizers, or soda lime have been replaced.
1. Set the handwheel of the vaporizer being tested
to a concentration of at least 0.2 Vol.%.
2. Press the >Leak Test< button on the standby
screen (1 in Figure 61).
The following prompt is displayed:
Before starting leak test, close the Y-piece,
connect the sample line and make sure that
all flow controls are closed. If vaporizer leaks
need to be tested, open respective vaporizer
to at least 0.2 Vol.%. Press rotary knob to
start the leak test.
3. Perform the actions as instructed.
Apollo performs the leak test for Volume Mode/
Pressure Mode in about 90 seconds, then system
compliance is determined for volume correction and
the overall system is checked for leaks in the
breathing system.
Note: The breathing bag and its hose are also
tested for leaks at the same time.
Leakage is tested in the automatic (mechanical)
ventilation line (leak (system)) and in the overall
system (leak (Man/Spont)).
Figure 61. Standby screenWith the system in Standby mode:
1
Part Number: 9039994, 2nd edition
98Operating Instructions Apollo SW 4.n
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.