Drager Alcotest 7510 MBRS Instructions For Use Manual

Dräger Alcotest® 7510 MBRS
Breath Alcohol Monitor
en
3
Contents
1 Safety-related information . . . . . . . . . . . . . . . . . . 4
2 Conventions used in this document . . . . . . . . . . 4
2.1 Definitions of alert icons. . . . . . . . . . . . . . . . . . . . . . 4
3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.1 Installing the wrist strap . . . . . . . . . . . . . . . . . . . . . . 6
4.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.3 Switch on and off . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.4 Automatic shut down . . . . . . . . . . . . . . . . . . . . . . . . 8
4.5 Attaching the mouthpiece . . . . . . . . . . . . . . . . . . . . 8
5 Performing the measurement . . . . . . . . . . . . . . . . 9
5.1 Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2 Requirements of the person to be tested . . . . . . . . . 9
5.3 GPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.4 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.4.1 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4.2 Test result . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.4.3 Additional measurement . . . . . . . . . . . . . . . . . . . . 11
5.5 Removing the mouthpiece . . . . . . . . . . . . . . . . . . . 12
6 Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6.1 Menu settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6.1.1 Main user screen . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6.1.2 Set contrast/backlight. . . . . . . . . . . . . . . . . . . . . . . 12
6.1.3 Show cal. dates . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1.4 Accuracy check . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1.5 Show last test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1.6 Show statistic counter . . . . . . . . . . . . . . . . . . . . . . 13
7 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 14
8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
8.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . . . 15
8.2 Cleaning and disinfecting . . . . . . . . . . . . . . . . . . . . 16
9 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
10 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
11 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
11.1 Battery disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
12 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Dräger Alcotest 7510 MBRS 3
1 Safety-related information
!
Before using this product, carefully read these instructions
for use and those of the associated products.
The device is not to be serviced or repaired by any person
not employed or authorised by Dräger or by the Medical Bureau of Road Safety (MBRS). All repairs are to be carri­ed out by either the MBRS, Dräger or by a third Party au­thorised by either Dräger or the MBRS.
Strictly follow the instructions for use. The user must fully
understand and strictly observe the instructions. Use the product only for the purposes specified in the intended use section of this document.
Do not dispose of the instructions for use. Ensure that they
are retained and appropriately used by the product user.
Only trained and competent users are permitted to use this
product.
Comply with all local and national rules and regulations as-
sociated with this product.
Use only genuine Dräger spare parts and accessories, or
the proper functioning of the product may be impaired.
Do not use a faulty or incomplete product, and do not mo-
dify the product.
Notify Dräger in the event of any component fault or failure.
Safe coupling with electrical devices
Electrical connections to devices which are not listed in these instructions for use should only be made following consultation with the respective manufacturers or an expert.
Not for use in areas of explosion hazard
This device is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to oc­cur.
2 Conventions used in this
document
2.1 Definitions of alert icons
The following alert messages are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each alert message is as follows:
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury. It may also be used to alert against unsafe practices.
4 Dräger Alcotest 7510 MBRS
3 Description
3.1 Product overview
1 Mouthpiece receptacle 2 Display 3 -key (UP) 4 - key (ON/OFF, Confirm)
OK
5 Opening to the audible alarm 6 USB connection 7 - key (DOWN/MENU)
M
8 Mouthpiece ejector
at right or left dependent on the mouthpiece receptacle
9 LED red, yellow, and green
1
9
8
7
Alcotest 7510
M
OK
11:53:47
04.12.17
2
3
4
6
Alcotest 7510
5
00200085.eps
Dräger Alcotest 7510 MBRS 5
3.2 Intended use
Dräger Alcotest
Breath alcohol measuring device for the fast determination
of the presence of alcohol in breath.
For portable, battery-powered operation.
The Alcotest 7510 MBRS is factory-equipped with the fol­lowing options:
with GPS (Global Positioning System)
for the satellite-based determination of the device location.
Mouthpiece receptacle on the right-hand side of the "Stan-
dard" mouthpieces.
®1)
7510 MBRS
4Use
4.1 Installing the wrist strap
1. Guide the mounting strap (thin section) of the wrist strap th­rough the housing drills on the underside of the measuring device and form a loop.
2. Pull the thicker section of the wrist strap through the formed eye.
00300085.eps
1)
Alcotest® is a registered trademark of Dräger.
6 Dräger Alcotest 7510 MBRS
4.2 Symbols
According to the device status, the following symbols are dis­played in the top row, next to the time and date information (one below the other on the right side)
11:53:47
PC
PC
There is a fault
There is a warning
The device is communicating with a PC
Database is 90 % full
Database is 100 % full, no more data can be stored
Database is 100 % full and will be overwritten with new data
GPS system active (only on devices with GPS sys­tem)
04.12.2017
00533245
Keys
OK
M
Battery or charge state of rechargeable battery: full
Battery or charge state of rechargeable battery: full
Battery or charge state of rechargeable battery: full
Battery or charge state of rechargeable battery: empty
"OK" key Switch on/off key for the device. Confirms the entry of the selected menu function(s).
"DOWN / MENU" Sets values and selects menu functions. Access to the menus.
"UP" Sets values and selects menu functions.
2
/
3
1
/
3
Dräger Alcotest 7510 MBRS 7
4.3 Switch on and off
Switch on
Press and hold the key for
approx. 1 second until the start screen appears. A self-test is completed. Device information is displayed.
If the next service date has expired:
Have the device calibrated (see
"Troubleshooting" on page 14).
Switch off
Hold the key pressed.
OK
The device shuts down automatically after approx. 2 seconds.
OK
17:42:31
08.10.2018
Alcotest 7510
00500085.eps
17:42:31
08.10.2018
Alcotest 7510
shut-down in 2 sec.
00600085.eps
4.4 Automatic shut down
17:42:31
08.10.2018
In the default setting, the device shuts down automatically 4 minutes after the last operation.
The device also shuts down automatical-
Alcotest 7510
ly when the batteries are fully dischar­ged.
auto shut-down
00700085.eps
4.5 Attaching the mouthpiece
1. Remove the mouthpiece from the packaging. For hygienic reasons, leave the mouth contact area protected by the packaging, until the mouthpiece sits firmly on the device.
2. With the chequered mark (= pressure surface) facing upwards, put the "Standard" mouthpiece into the mouth­piece receptacle at an angle and push it into the mouth­piece receptacle. The mouthpiece audibly engages.
3. The inserted mouthpiece can be illuminated by the swit­ched on device.
2
1
00400085.eps
8 Dräger Alcotest 7510 MBRS
5 Performing the measurement
5.1 Prerequisites
1. Insert a new mouthpiece for each test subject.
2. Do not allow any high alcohol concentrations on the sen­sor, e.g., by rinsing the mouth with a high proof alcohol shortly before the measurement. Doing this reduces the service life of the sensor.
5.4 Preparation
Switch on the device.
After a few seconds, the start screen appears.
17:42:31
08.10.2018
Alcotest 7510
V 5.86
5.2 Requirements of the person to be tested
If less than 20 minutes has elapsed between the persons
last drink and use of device, then residual alcohol in the mouth may distort the measurement.
The person to be tested should breathe normally and stea-
The number of days until next cali­bration appears.
00800085.eps
17:42:31
08.10.2018
dily prior to the sampling. Rapid inhalation and exhalation through the mouth should be avoided.
Next Calibration
99 days
5.3 GPS
Global Positioning is active . Once the satellites have been obtained the symbol will remain static. During acquisition it will flash.
Dräger Alcotest 7510 MBRS 9
00900085.eps
The main user screen then appears
01300085.eps
17:42:31
08.10.2018
BLOW
with » Driver «, »Specified Driver « and »Menu« options.
17:42:31
08.10.2018
Driver
Specified Driver
Menu
Device then moves from
» WARMUP «, » AIR BLANK CHECK « to » READY «.
08.10.2018
Driver
READY
Test No.
31
17:42:31
01000085.eps
1. Select the type of driver - either » Driver « or » Specified Driver « using the blue arrow keys to navigate, select by
pressing the key.
2. Select a reason for the test from the » Reason code « menu using the blue arrow keys to navigate, select by pressing the key to confirm.
OK
17:42:31
08.10.2018
Reason code:
OK
MIT
RTC Alleged Offence Opinion Consumed Demo/Training
5.4.1 Procedure
Prompt the test person to blow even-
ly and continuously into the mouth­piece. A sufficient breath flow is confirmed by a continuous audible signal and the green LED.
During the delivery of the breath sample the following message appe­ars in the display: »BLOW«.
If the breath sample is sufficient (mi­nimum breath volume and the mini-
01200085.eps
mum blowing period were reached), the sampling is triggered. The audible signal stops.
01100085.eps
The following message appears in the display:
» WAIT Analyzing «
10 Dräger Alcotest 7510 MBRS
5.4.2 Test result
OK
After 3 to 20 seconds (dependent on the detected concentration) the measure­ment result appears in the display. The measurement result is displayed wi­th the type of driver at the top of the screen, the result (»PASS« or »FAIL«) in the middle and the test num­ber underneath the result.
5.4.3 Additional measurement
Press the key to confirm and re-
turn to main user screen.
Driver
PASS
Test No.
31
Confirm
17:42:31
08.10.2018
01400085.eps
Error during the delivery of the breath sample
If the volume of the breath sample is in­sufficient, the following appears in the
17:42:31
08.10.2018
display: » Insufficient volume «. A short audible signal beeps and the red LED flashes.
To repeat, press the key.
OK
Insufficient
volume
Waiting period for a new testing standby is 4 to 20 seconds. In the case of highly irregular exhalation, e.g. sudden termination or suction at the end of the breath sample, the following
Confirm
01500085.eps
appears in the display: » Negative Breath Flow «. A short audible signal beeps and the red LED flashes.
To repeat, press the key.
OK
If the subject is unable or unwilling to provide a sample a refu­sal can be recorded on the device.
1. Press the blue key to complete the test with »Refusal/ Failure « as result. This will increment the test number.
2. Press the key to confirm.
OK
Dräger Alcotest 7510 MBRS 11
5.5 Removing the mouthpiece
01600085.eps
01900085.eps
17:42:31
08.10.2018
Show cal. dates
Accuracy check
Show last test
Show statistic counter
Back to measurement
Set contrast/backlight
Push the eject button upwards.
The mouthpiece is pushed out of the mouthpiece recep­tacle.
6Menu
6.1 Menu settings
6.1.1 Main user screen
Select »Menu«.
17:42:31
08.10.2018
Driver
Specified Driver
Menu
Use blue keys to navigate and green
to accept.
6.1.2 Set contrast/backlight
Adjust the contrast and backlight.
12 Dräger Alcotest 7510 MBRS
01000085.eps
6.1.3 Show cal. dates
Display of last date the device was calibrated. Back with green key.
6.1.4 Accuracy check
1. Select » Accuracy check « and press the key.
OK
The display shows the calibration check gas to be used. »Next« is automatically highlighted at the bottom of the
screen.
2. Press the key.
OK
After a short period, » Please apply gas « is displayed.
3. Fit a clean dry mouthpiece to the device.
4. Connect the outlet from the check gas cylinder to the mouthpiece and allow gas to flow through the mouthpiece by pressing the valve on top of the gas cylinder.
A beep is heard, followed by double beep, when the device has sampled the gas and » Please stop gas flow, analyzing « is displayed momentarily.
5. Stop the flow of gas by releasing the valve at the top of the gas cylinder. The result of the accuracy check is displayed (» Accuracy
check OK « or » Accuracy check error «).
6.1.5 Show last test
Display of the last test. Back with green key.
6.1.6 Show statistic counter
Display of number of tests in calendar year.
Dräger Alcotest 7510 MBRS 13
7 Troubleshooting
Fault Cause Remedy
Device auto­matically switches off
Battery symbol empty
Date / time in­correct
Menu is auto­matically exited
Not possible to switch on the device
Insufficient volu­me / invalid sample
No continuous tone during blo­wing
Discharge the battery / batteries
No activity for 4 minutes
Discharge the battery / batteries
Discharge ba­ckup battery
The device auto­matically returns to the measure­ment function af­ter 120 seconds
Discharge the rechargeable battery / batteries
Test subject blows too weakly or non-conti­nuously
Device is not yet ready to measure
Replace batteries.
Switch on device again.
At the next opportunity, replace batteries.
Set date / time. Contact MBRS to set date and time.
Select the menu item again.
Replace batteries.
Blow more strongly and continuously.
Wait until the device is ready to measure.
Error message in the display
» No printout possible «
» Temperature out of range «
» Service date expired «
The device alerts and blocks the ope­ration, if set
Cause Remedy
Measured value exceeds display area.
Dräger Mobile Printer not recog­nised or protocol not completely processed.
Device tempera­ture is outside of the allowed tem­perature range.
The service inter­val is expired.
Use lower concent­rations.
Maintain a visual connection and close proximity bet­ween the IR inter­faces of the devices during the entire printing peri­od.
Heat or cool the device and start the measurement again.
Calibration requi­red. Return to MBRS.
14 Dräger Alcotest 7510 MBRS
8 Maintenance
01700085.eps
+
+
+
+
01800085.eps
!
8.1 Replacing the batteries
For devices with alkaline batteries or NiMH rechargeable bat­teries. Lithium-ion rechargeable batteries cannot be replaced. If the battery smybol in the display stands at » « (empty) in­sert new batteries or charged NiMH rechargeable batteries, or place the device in the charger and charge:
1. Rotate the catch on the battery cover anticlockwise 90 and remove the battery cover.
2. In order to open the battery cover, first loosen the screw. Then rotate the catch 90o and re­move the battery cover.
o
3. Remove used alkaline bat­teries and insert new alkaline batteries – obser­ve correct polarity!
4. Replace the battery cover and lock the catch with a clockwise rotation.
Required batteries:
4 units 1.5 V alkaline type (Mignon, LR6, AA)
CAUTION
Do not throw used alkaline batteries into fire, charge them again, or open them by force. Danger of explosi­on!
Dräger Alcotest 7510 MBRS 15
8.2 Cleaning and disinfecting
Do not submerge in liquids, no liquids can touch the
connectors!
For cleaning:
wipe with a disposable towel.do not use any detergents that contain alcohol or solvents.
CAUTION
!
The use of detergents with a high alcohol content can greatly reduce the service life of the internal sensor or generate an erroneous positive test result in the short­term.
9 Transport
Transport the product in its original packaging.
10 Storage
Store the product in its original packaging.
11 Disposal
This product must not be disposed of as household waste. It is therefore marked with the adjacent symbol. You can return this product to Dräger free of charge. For information please contact the national marketing organizations or Dräger.
11.1 Battery disposal
Batteries must not be disposed of as household waste. They are therefore marked with the adjacent symbol. Dispose of batteries at battery collection centers as specified by the applicable regulations.
16 Dräger Alcotest 7510 MBRS
12 Technical data
Measuring Principle
Electrochemical sensor
Measuring range (Default setting)
Breath alcohol concent­ration
Display Area
The display is identical with the measuring range. If the measurement result is outside of the measuring range, three arrows are shown in the display » ↑↑↑« = Measuring range exceeded.
Ambient conditions
during operation
during storage
Dimensions (H x W x D)
with alkaline batteries or NiMH rechargeable bat­teries with lithium-ion rechar­geable batteries
Weight
approx. 433 g (with batteries / rechargeable batteries)
Power supply
Alkaline batteries 4 units 1.5 V alkaline type (Mignon,
0.00 to 3.0 mg/L (Ethyl alcohol mass per breath vo-
lume at 34
–10 to 50
o
C and 1013 hPa.)
o
C 10 to 100 % r.H. (non-condensing) 600 to 1100 hPa
o
–40 to 70
C
approx. 185 mm x 90 mm x 44 mm
approx. 185 mm x 90 mm x 49 mm
LR6, AA) approx. 1500 measurements
CE marking
Electromagnetic compatibility (Directive 2014/30/EC)
Dräger Alcotest 7510 MBRS 17
93 00 085 - GA 4754.602 en © Dräger Safety AG & Co. KGaA Edition 01 - February 2019 Subject to alteration
Dräger Safety AG & Co. KGaA
Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882-0 Fax +49 451 882-20 80 www.draeger.com