Detcon MicroSafe™ Model PI-500 and PI-501 universal VOC sensors are non-intrusive “Smart” sensors designed to
detect and monitor for VOC & toxic gas in the ppm range. One of the primary features of the sensor is its method of
automatic calibration which guides the user through each step via instructions displayed on the backlit LCD. The sensor features LED indicators for FAULT and CAL status and is equipped with a standard analog 4-20 mA output. The
microprocessor supervised electronics are packaged as a universal plug-in transmitter module that mates to a standard
connector board. Both are housed in an explosion proof condulet that includes a glass lens. A 16 character
alpha/numeric indicator is used to display sensor readings as well as the sensor’s menu driven features via a hand-held
programming magnet.
Typical ranges of detection are 0-10ppm, 0-20ppm, 0-50ppm, (using the PI-500) and 0-100ppm, 0-500ppm, and 01,000ppm (using the PI-501). Other ranges are available and all ranges are covered by this manual. To determine sensor
model number, reference the label located on the enclosure cover. To determine primary range, reference labeling on
the sensor head.
3.0.1 Sensor Technology
The sensors are based on a plug-in replaceable miniature PID (Photo-Ionization Detector) sensor technology. The sensor is sensitive to all ambient gases that have ionization potentials of < 10.6 eV, making it highly sensitive but extremely
non-specific. The sensor responds to most all toxic VOC compounds and many other toxic gases as well. The sensor is
comprised of a UV emitting lamp that is covered by a specific optical filter which projects only radiation in the 10.6 eV
range. Target gases that diffuse into the sensor chamber with ionization potentials of < 10.6 eV, are ionized by the radiation and give up free electrons. The free electrons are captured by the high voltage collection grid and provide a current signal that is directly proportional to the concentration of the target gas.
PI-500 Toxic Gas Sensors PG.3
Page 4
3.0.2 Universal Microprocessor Control Transmitter Circuit
detcon inc.
Program Switch #2
FLTCAL
MicroSafe™ Gas Sensor
H
OUST ON, T EXA S
PGM 2
PGM 1
M
ODEL
P
I-500
CONTRAST
Fault & Cal LEDs
Program Switch #1
Menu Driven Display
Plug-in Microprocessor Control Circuit
Display Contrast Adjust
0 PPM VOC
UNIVERSAL
T
RANSMITTER
WHT
BLK
YEL
BLU
mA
VDC Power In
4-20 mA Output
Sensor
Base Connector Board
PID Sensor Head
Transmitter Electronics
in Explosion-Proof housing
he control circuit is microprocessor based and is packaged as a universal plug-in field replaceable module, facilitating
T
easy replacement and minimum down time. The universality includes the ability to set it for any range concentration
and for any gas type. These gas and range settings must be consistent with the PID Sensor Head it is mated with.
Circuit functions include a basic sensor pre-amplifier, on-board power supplies, microprocessor, back lit alpha numeric
display, fault and calibration status LED indicators, magnetic programming switches, and a linear 4-20 mA DC output.
3.0.3 Base Connector Board
The base connector board is mounted in the explosion proof enclosure and includes: the mating connector for the control circuit, reverse input and secondary transient suppression, input filter and lugless terminals for all field wiring.
3.0.4 Explosion Proof Enclosure
The transmitter electronics are packaged in a cast metal explosion proof enclosure. The enclosure is fitted with a threaded cover that has a glass lens window. Magnetic program switches located behind the transmitter module face plate are
activated through the lens window via a hand-held magnetic programming tool allowing non-intrusive operator interface
with the sensor. Calibration can be accomplished without removing the cover or declassifying the area. Electrical classification is Class I; Groups B, C, D; Division 1 (explosion proof).
PI-500 Toxic Gas Sensors PG.4
Page 5
3.1 PR
Analog 4-20 mA Out
Functional
Block
Diagram
Power In
Pre-AmpDisplay
T
emperature
C
ompensation
C
al & Fault
L
EDs
4-20mA
M
icro-
p
rocessor
Tran smit ter
Power Supply
S
ensor
Element
I
/O Circuit
Protection
INCIPLE OFOPERATION
Ionizable target gases diffuse into the PID sensor chamber through a sintered flame arrestor. These target gases are
exposed to UV radiation emitted by the PID lamp and this causes a fraction of the molecules to give up a free electron.
The free electrons are captured by the high voltage collection grid and provide a current signal that is directly proportional to the concentration of the target gas. This change in current is completely reversible and results in the continuous monitoring of ambient air conditions.
3.2 APPLICATION
3.2.1 Sensor Placement/Mounting
Sensor location should be reviewed by facility engineering and safety personnel. Area leak sources and perimeter mounting
are typically used to determine number and location of sensors. The sensors are generally located 2 - 4 feet above grade.
3.2.2 Interference Data
Detcon Model PI-500 series PID sensors are subject to interference from many gases. This interaction is shown in the
table in Section 3.2.3. The table shows most all gases of interest and the level of signal response they have relative to a
standard isobutylene reference gas. This measure is referred to as the Response Factor (RF). As a general rule, the lower
the RF value, the stronger the signal from the PID sensor. When determining a cross-interference from one gas to
another, find the RF of your target gas and then your interfering gas(es). The cross-interference will be calculated by
dividing the RF of your interfering gas by the RF of your target gas.
For example, if your target gas is benzene and you are concerned about a cross-interference to H2S then you would calculate the cross interference to be 3.3/0.50 = 6.2. This shall be interpreted as: it will take 6.2 ppm of H2S to register as
1 ppm benzene on a PID sensor calibrated for benzene.
In many cases, the user will be interested in measuring a multiple of toxic VOC compounds. In this case the sensor
will produce a signal that is a composite total of each gases’ individual response, when taking into account the corresponding response factors.
For example, if the target gases are benzene and isobutanol and your PID sensor was calibrated for benzene then the
presence of 5 ppm benzene and 5 ppm of isobutanol would each add to the total reading. In this case, the 5 ppm benzene would register as 5 ppm, but the 5 ppm isobutanol would register as the amount of cross interference of isobutanol relative to a benzene calibration. This is calculated as discussed above where you divide the RF of isobutanol by
the RF of benzene. Using the look up table this gives you 3.8/0.50 = 7.2. So it takes 7.2 ppm isobutanol to equal 1
part benzene. Since we have 5 ppm isobutanol, that will equal 0.7 ppm on the benzene scale. The total signal will be 5
+ 0.7 = 5.7 ppm.
3.2.3 Relative Response Gas Matrix (See next page)
The table shows you the response of the PID sensor to a long list of components. It includes the compound name,
synonyms/abbreviations, and chemical formula. It also lists the 10.6 eV Response Factor (the measure of how strong
the signal from the sensor is in reference to Isobutylene gas). Isobutylene gas is the standard reference used with PID
sensors, the lower the Response Factor, the stronger the signal.
NR = not reccomended (does not register)
? = measureable but no data exist
Confirmed Value = “+” means actual gas has been used to verify RF, “blank” means it is an empirical estimate
IP = is the gases ionization potential (only gases < 10.6eV will respond to sensor)
TWA/Time Weighted Average = generally accepted limit for safe 8 hour exposure (in ppm)
Viscor 120B - see Mineral Spirits - Viscor 120B Calibration Fluid
Xylene, m-C8H100.4+8.56
Xylene, o-C8H100.6+8.56
Xylene, p-C8H100.5+8.44
10.6 eV
None1
PI-500 Toxic Gas Sensors PG.13
Page 14
3.3 SPECIFICATIONS
ethod of Detection
M
Plug-in Miniature PID Sensor
Repeatability
± 2% FS
esponse Time
R
T90 < 30 seconds
Temperature Range
0-50°C; 32-122°F
Humidity Range
0-99% RH noncondensing
Output
Linear 4-20 mA DC
Input Voltage
11.5-28 VDC
Power Consumption
Normal operation = 58 mA (1.4 watts @ 24VDC); Maximum = 74 mA (1.8 watts @ 24VDC); Maximum @ 11.5VDC = 80 mA (1 watt)
Electrical Classification
Class 1; Groups B, C, D; Div. 1
Sensor Warranty
12 Months
3.4 INSTALLATION
Optimum performance of ambient air/gas sensor devices is directly relative to proper location and installation practice.
3.4.1 Field Wiring Table (4-20 mA output)
Detcon Model PI-500 toxic gas sensor assemblies require three conductor connection between power supplies and host
electronic controllers. Wiring designators are
+
(DC), –(DC) , and
mA
(sensor signal). Maximum single conductor
resistance between sensor and controller is 10 ohms. Maximum wire size for termination in the sensor assembly terminal
board is 14 gauge.
AWG
MetersFeet
20240800
183601200
166002000
149003000
Note 1:
This wiring table is based on stranded tinned copper wire and is designed to serve as a reference only.
Note 2: Shielded cable may be required in installations where cable trays or conduit runs include high voltage lines or
other sources of induced interference.
Note 3: The supply of power must be from an isolating source with over-current protection as follows:
AWG
Over-current ProtectionAWGOver-current Protection
223A1610A
205A1420A
187A1225A
3.4.2 Sensor Location
Selection of sensor location is critical to the overall safe performance of the product. Five factors play an important role
in selection of sensor locations:
(1) Density of the gas to be detected
(2) Most probable leak sources within the industrial process
(3) Ventilation or prevailing wind conditions
(4) Personnel exposure
(5) Maintenance access
PI-500 Toxic Gas Sensors PG.14
Page 15
Density - Placement of sensors relative to the density of the target gas is such that sensors for the detection of heavier than air
EYS
Seal
Fitting
Drain
“T”
Plug any unused ports.
ases should be located within 2-4 feet of grade as these heavy gases will tend to settle in low lying areas. For gases lighter than
g
air, sensor placement should be 4-8 feet above grade in open areas or in pitched areas of enclosed spaces.
Leak Sources - Most probable leak sources within an industrial process include flanges, valves, and tubing connections
of the sealed type where seals may either fail or wear. Other leak sources are best determined by facility engineers with
experience in similar processes.
Ventilation - Normal ventilation or prevailing wind conditions can dictate efficient location of gas sensors in a manner
where the migration of gas clouds is quickly detected.
Personnel Exposure - The undetected migration of gas clouds should not be allowed to approach concentrated personnel areas such as control rooms, maintenance or warehouse buildings. A more general and applicable thought toward
selecting sensor location is combining leak source and perimeter protection in the best possible configuration.
Maintenance Access
Consideration should be given to easy access by maintenance personnel as well as the consequences of close proximity
to contaminants that may foul the sensor prematurely.
Note: In all installations, the sensor element in SS housing points down relative to grade (Fig. 1). Improper sensor orientation may result in false reading and permanent sensor damage.
3.4.3 Local Electrical Codes
Sensor and transmitter assemblies should be installed in accordance with all local electrical codes. Use appropriate conduit seals. Drains & breathers are recommended. The sensor assemblies are suitable for Class I; Groups B, C, D; Div. 1
environments.
PI-500 Toxic Gas Sensors PG.15
Page 16
3.4.4 Installation Procedure
WHT
BLK
YEL
BLU
mA
VDC Power In
4-20 mA Output
Sensor
Base Connector Board
4 3/4"
3
/4" NPT
1/4" Dia.
Mounting Holes
7 1/4"
6 1/8"
5 1/2"
3
/4" NPT
Rain
Shield
2"
2 1/8"
) Securely mount the sensor junction box in accordance with recommended practice. See dimensional drawing (Fig. 2).
a
b) Remove the junction box cover and un-plug the control circuit by grasping the two thumb screws and pulling out-
ward. Observing correct polarity, connect the DC power field wiring to the terminals labeled “+” and “–”. Then
connect the 4-20 mA return wire (reference figure 3). Reinstall cover.
Figure 3
3.4.5 Remote Mounting Applications
Some sensor mounting applications require that the gas sensor head be remotely mounted away from the sensor transmitter. This is usually true in instances where the gas sensor head must be mounted in a location that is difficult to
access. Such a location creates problems for maintenance and calibration activities. Detcon provides the PI-500 sensor
in a remote-mount configuration in which the sensor (Model PI-500-RS) and the transmitter (Model PI-500-RT) are provided in their own condulet housing and are interfaced together with a four conductor cable. Reference figure 4 for
wiring diagram.
PI-500 Toxic Gas Sensors PG.16
Page 17
1234
WHT
BLK
YEL
BLU
Remote Transmitter
PI-500-RT
Remote Sensor
PI-500-RS
W
HT
B
LK
Y
EL
BLU
Plug unused port
with 3/4 NPT plug.
Figure 4
3.5 START UP
Upon completion of all mechanical mounting and termination of all field wiring, apply system power and observe the
following normal conditions:
a) PI-500 “Fault” LED is off.
b) A temporary upscale reading will occur as the sensor powers up. This upscale reading should clear to “0” ppm with-
in approximately 5-10 minutes of turn-on, assuming there is no gas in the area of the sensor.
NOTE 1:
NOTE 2:
3.5.1 Initial Operational Tests
After a warm up period has been allowed for, the sensor should be checked to verify sensitivity to its target gas.
Material Requirements
*Detcon PN 943-000006-132 Calibration Adapter
*
If the display contrast needs adjustment, refer to section 3.11.
If the sensor does not clear to zero after 15 minutes of warm-up, there may be target VOC gases present in the area.
Span gas containing isobutylene in air. It is recommended that the target gas concentration be 50% of scale at a controlled flow rate of 200 cc/min. For example, a Model PI-500 sensor in the range 0-100ppm would require a test gas of
50ppm isobutylene. For a sensor with a range of 0-10ppm a test gas of 5ppm is recommended, etc. Other concentrations are acceptable as long as they are between 10%-90% of full-scale range.
a) Attach the calibration adapter to the sensor housing. Apply the test gas at a controlled flow rate of 200 cc/min.
Observe that the LCD display increases to a level of ±10% of applied concentration.
b) Remove the test gas and observe that the LCD display decreases to “0 PPM”.
Initial operational tests are complete. Detcon PID gas sensors are pre-calibrated prior to shipment and will, in most
cases, not require significant adjustment on start up. However, it is recommended that complete zero and span calibrations be performed within 24 hours of installation. Refer to calibration instructions in later text.
PI-500 Toxic Gas Sensors PG.17
Page 18
3.6 OP
Operating software is menu listed with operator interface via the two magnetic program switches located under the face
plate. The two switches are referred to as “PGM 1” and “PGM 2”. The menu list consists of 3 items which include submenus as indicated below. (Note: see section 3.7 for a complete software flow chart.)
01. Normal Operation
02. Calibration Mode
03. Program Menu
3.6.1 Normal Operation
In normal operation, the display tracks the current status of the sensor and gas concentration and appears as:
“0 PPM xxx” (the “xxx” is the abbreviated gas type, ie., “0 PPM VOC”. The mA current output corresponds to the
monitoring level of 0-100% of range = 4-20 mA.
3.6.2 Calibration Mode
Calibration mode allows for sensor zero and span adjustments. “1-ZERO 2-SPAN”
ERATING
a) Current Status
a) Zero
b) Span
a) View Program Status
b) Set Calibration Level
c) Set Response Factor
d) Set Zero Offset
SO
FTWARE
& MA
GNETICINTERFACE
3.6.2.1 Zero Adjustment
Zero is set in ambient air with no target gas present or with zero gas applied to the sensor. “AUTO ZERO”
3.6.2.2 Span Adjustment
Span adjustment is performed with a target gas concentration of 50% of range in air. Span gas concentrations other than
50% of range may be used. Refer to section 3.6.3.2 for details. “AUTO SPAN”
3.6.3 Program Mode
The program mode provides a program status menu (View Program Status) to check operational parameters. It also
allows for the adjustment of the calibration gas level setting.
3.6.3.1 Program Status
The program status scrolls through a menu that displays:
* The software version number.
* Range is ###
* The calibration gas level setting. The menu item appears as: “CalLevel @ xxPPM”
* The response factor setting. The item appears as: “RespFactor = x.xx”
* The current zero offset value. The menu item appears as: “ZeroOffset = x.x PPM”
* The raw signal from the sensor head. The menu item appears as: “Raw Signal = x.xx V”
* The estimated remaining sensor life. The menu item appears as: “SENSOR LIFE 100%”
3.6.3.2 Calibration Level Adjustment
The calibration level is adjustable from 10% to 90% of range. The menu item appears as: “CalLevel @ ##PPM”
3.6.3.3 Response Factor Adjustment
The Response Factor is set according to the primary target gas being detected. The menu item appears as:
“RespFactor = x.xx”
PI-500 Toxic Gas Sensors PG.18
Page 19
3.6.3.4 Zero Offset
detcon inc.
Program Switch #2
FLTCAL
MicroSafe™ Gas Sensor
HOUS TON , TEX AS
PGM2
PGM1
MODELPI-500
CONTRAST
Fault & Cal LEDs
Program Switch #1
Menu Driven Display
Plug-in Microprocessor Control Circuit
Display Contrast Adjust
0 PPM VOC
UNIVERSAL
TRANSMITTER
Magnetic Programming Tool
he Zero Offset is settable between 0 and 10 ppm to account for residual backgrounds of active VOCs in ambient air. The
T
menu item appears as: “ZeroOffset = x.x PPM”
3.6.3.5 Raw Signal
The raw signal from the sensor head is displayed for the purpose of troubleshooting.
3.6.4 Programming Magnet Operating Instructions
Operator interface to MicroSafe™ gas detection products is via magnetic switches located behind the transmitter face
plate. DO NOT remove the glass lens cover to calibrate or change programming parameters. Two switches labeled
“PGM 1” and “PGM 2” allow for complete calibration and programming without removing the enclosure cover, thereby eliminating the need for area de-classification or the use of hot permits.
Figure 5
A magnetic programming tool (see figure 5) is used to operate the switches. Switch action is defined as momentary contact, 3 second hold, and 30 second hold. In momentary contact use, the programming magnet is waved over a switch
location. In 3 second hold, the programming magnet is held in place over a switch location for 3 or more seconds. In
30 second hold, the programming magnet is held in place over a switch location for 30 or more seconds. Three and
thirty second hold is used to enter or exit calibration and program menus while momentary contact is used to make
adjustments. The location of “PGM 1” and “PGM 2” are shown in figure 6.
NOTE: If, after entering the calibration or program menus there is no interaction with the menu items for more than
30 seconds, the sensor will return to its normal operating condition.
Figure 6
PI-500 Toxic Gas Sensors PG.19
Page 20
3.7 SO
NORMAL
OPERATION
C
ALIBRATION
PGM1 (3) PGM2 (3)
1-ZERO 2-SPAN
L
EGEND
PGM1 - program switch location #1
P
GM2 - program switch location #2
(M) - momentary pass of magnet
(3) - 3 second hold of magnet
(30) - 30 second hold of magnet
I
NC - increase
D
EC - decrease
#
- numeric value
AUTO ZERO
AUTO SPAN
G
AS RANGE
P
GM1 (3)PGM2 (M)
V
IEW PROG STATUS
PGM1 (M) PGM2 (M)
P
GM1 (3)
CalLevel @ ##PPM
INC
DEC
P
GM1 (3)PGM2 (M)
S
ET CAL LEVEL
P
GM2 (3)
P
GM2 (3)
S
ENSOR LIFE ##%
P
GM1 (3)
P
GM2 (30)
CAL LEVEL @ ##PPM
PGM1 (M) PGM2 (M)
P
GM1 (3)
RESPFACTOR = X.XX
INC
DEC
P
GM1 (3)PGM2 (M)
S
ET RESPONSE FACTOR
P
GM2 (3)
PGM1 (M) PGM2 (M)
P
GM1 (3)
ZERO OFFSET = X.X PPM
INC
DEC
P
GM1 (3)PGM2 (M)
S
ET ZERO OFFSET
P
GM2 (3)
Software Version V#.#
Z
ERO OFFSET = X.XX
R
AW SIGNAL = X.XX V
Figure 6
FTWAREFLOW
CH
ART
Figure 7
3.8 CALIBRATION
Material Requirements
*Detcon PN 943-003270-000 MicroSafe™ Programming Magnet
*Detcon PN 943-000006-132 Calibration Adapter
*Zero Air gas containing no VOC compounds
*Span gas containing isobutylene in air. The cal gas concentration is recommended at 50% of range (which is the fac-
3.8.1 Calibration Procedure - Zero
NOTE 1
tory default) at a controlled flow rate of 200 cc/min. Example: for a Model PI-500 sensor with a range of 0-100ppm,
a test gas of 50 ppm isobutylene is recommended. For a sensor with a range of 0-10 ppm a test gas of 5 ppm is recommended, etc. Other concentrations can be used as long as they fall within 10% to 90% of range. See section
3.8.2 for details. Reference section 3.9-b-1 if you do not know the sensor target gas or range of detection.
: Before performing a zero calibration, determine if there are any active VOC target gases in the area. If it can
be concluded that there are no active VOC gases in the area, then execute steps b) and c) below.
NOTE 2
a) Apply a zero air standard at 200 cc/min for approximately 2-3 minutes then proceed through steps b) and c) with the
: Assuming that there are some residual VOC target gases in the background, you will require the use of the
zero air gas standard to perform a correct zero calibration starting from step a) below.
gas under continuous flow. After applying gas, place your fingertip over the exit port of the cal adapter for 10 -30
seconds to expedite the gas purging process.
b) Enter the calibration menu by holding the programming magnet over PGM 1 (see Figure #6) for 3 seconds until the
c) Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display reads:
display reads “1-ZERO 2-SPAN”, then withdraw the magnet. Note that the Cal LED should now be illuminated.
“Setting Zero”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is complete
the display will read “ZERO COMPLETE” for 5 seconds and then return to the normal operations menu reading.
PI-500 Toxic Gas Sensors PG.20
Page 21
d) Remove the zero air gas and cal adapter and allow the sensor 3-5 minutes to rest on ambient air. If there are residual
ctive VOC gases in the area, then the sensor will read higher than 0.0 ppm. If this is the case, then you can use
a
the Zero Offset feature to correct for this residual background amount.
e) Set the Zero Offset value according to the concentration value found following the zero calibration procedure. See
section 3.8.1.1 below.
3.8.1.1 Using the Zero Offset Feature
If it is determined that there is a constant and non-negligible amount of residual active VOC gases in the background
air, the Zero Offset feature can be used to correct for this.
a) Observe the sensor’s concentration reading on air after a zero air calibration procedure. This represents the back-
ground VOC contribution that you will be offsetting.
b) Access the Zero Offset software feature by applying the magnet to PGM2 for 15 seconds. Then use the magnet to
momentarily pass over PGM1 and advance to the “Set Zero Offset” menu. Then apply the magnet to PGM1 for 3
seconds to access this menu. The menu should now read as “Zero Offset = X.X. Use PGM1 to increment this
number up to your desired offset level. When the correct offset is set, apply the magnet to PGM1 for 3 seconds to
accept the value. Then apply the magnet to PGM2 for 3 seconds to return to Normal Operation.
c) When done correctly, the unit should read 0.0 when back in Normal Operation.
3.8.2 Calibration Procedure - Span
3.8.2.1 Set Response Factor
All span calibrations are recommended to be done with a calibration standard consisting of isobutylene in an air back-
ground. If your target gas is different than the isobutylene span gas, you will be required to apply the correct Response
Factor. Look up the Response Factor for your target gas in the Table shown in Section 3.2.3.
a) Enter the programming menu by holding the magnet stationary over “PGM2” for 15 seconds until the display reads
“View Program Status”, then withdraw the magnet. At this point you can scroll through the programming menu by
momentarily waving the magnet over “PGM1” or “PGM2”. The menu options are: View Program Status, Set Cal Level,
Set Response Factor, and Set Zero Offset. Scroll to the “Set Response Factor” selection.
b) Select “Set Response Factor” by holding the magnet over “PGM1” for 3 seconds until the display reads “RespFactor
= x.xx”, then withdraw the magnet. Use the magnet to make an adjustment to “PGM1” to increase or “PGM2” to
decrease the displayed value until the value is equal to the desired “Response Factor” value from Section 3.2.3.
NOTE:
If you have multiple target gases, then select the target gas with the highest Response Factor from the Table.
This provides for the safest and earliest warning levels.
NOTE:
If you are span calibrating with the target gas, instead of isobutylene, then the response factor should be left at 1.0
3.8.2.2 Span Calibration
NOTE
: Isobutylene is the recommended calibration gas for this sensor.
CAUTION:
Verification of the correct calibration gas level setting and calibration span gas concentration is
required before “span” calibration. These two numbers must be equal.
Span calibration consists of entering the calibration function and following the menu-displayed instructions. The display will ask for the application of span gas in a specific concentration. This concentration must be equal to the calibration gas level setting. The factory default setting for span gas concentration is 50% of range. In this instance, a span gas
containing a concentration equal to 50% of range is required. If a span gas containing 50% of range is not available,
other concentrations may be used as long as they fall within 10% to 90% of range. However, any alternate span gas concentration value must be programmed via the calibration gas level menu before proceeding with span calibration.
Follow the instructions below for span calibration.
PI-500 Toxic Gas Sensors PG.21
Page 22
a) Verify the current calibration gas level setting as indicated by the programming status menu. To do this, follow the
nstructions in section 3.9 and make note of the setting found in listing number 2. The item appears as
i
“GasLevel @ xxPPM”.
b) If the calibration gas level setting is equal to your calibration span gas concentration, proceed to item “f”. If not,
adjust the calibration gas level setting so that it is equal to your calibration span gas concentration, as instructed in
items “c” through “e”.
c) Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 15 seconds until
the display reads “VIEW PROG STATUS”, then withdraw the magnet. At this point you can scroll through the
programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu
options are: View Program Status, Set Cal Level, Set Response Factor, and Set Zero Offset.
d) From the programming menu scroll to the calibration level listing. The menu item appears as: “SET CAL
LEVEL”. Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the
display reads “CalGas @ ##PPM”, then withdraw the magnet. Use the programming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired
calibration span gas concentration. Exit the programming menu by holding the programming magnet over “PGM1”
for 3 seconds.
e) Exit back to normal operation by holding the programming magnet over “PGM 2” for 3 seconds, or automatically
return to normal operation in 30 seconds.
f)From the calibration menu “1-ZERO 2-SPAN” proceed into the span adjust function by holding the program-
ming magnet stationary over “PGM 2” for 3 seconds then withdraw the programming magnet. At this point the
display will ask for the application of the target gas and concentration. The display reads “APPLY xxPPM ISO”
The x’s here will indicate the actual concentration requested.
g) Apply the calibration test gas at a flow rate of 200 cc/min. After applying the span gas, hold your fingertip (block-
ing) over the exit port of the cal adapter for 10-30 seconds. This helps to expedite the purging of the internal sensor chamber. As the sensor signal changes, the display will change to “AutoSpan xxPPM”. The “xx” part of the
reading indicates the actual gas reading which will increase until the sensor stabilizes. When the sensor signal is stable it will auto span to the correct ppm reading and the display will change to “SPAN COMPLETE” for 3 seconds, then to “SENSOR LIFE: xxx%”and then “REMOVE GAS”. Remove the gas. When the signal level has
fallen below 10% of full scale, the display will return to the normal operating mode.
NOTE 1: If there is not a minimal response to the cal gas in the first minute, the sensor will enter into the calibration fault mode which will cause the display to alternate between the sensor’s current status reading and the calibration fault screen which appears as: “SPAN FAULT #1” (see section 3.8.3).
NOTE 2: If during the auto-span function the sensor fails to meet a minimum signal stability criteria, the sensor
will enter the calibration fault mode which will cause the display to alternate between the sensor’s current status
reading and the calibration fault screen which appears as: “SPAN FAULT #2” (see section 3.8.3).
3.8.3
Additional Notes
1. Upon entering the calibration menu, the 4-20 mA signal drops to 2 mA and is held at this level until you return to
normal operation.
2. If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will
enter into the calibration fault mode which will activate the fault LED (see section 3.10) and will cause the display to
alternate between the sensor’s current status reading and the calibration fault description. In these cases, the previous
calibration points will remain in memory. If this occurs you may attempt to recalibrate by entering the calibration
menu as described in section 3.8.1-a. If the sensor fails again, defer to technical trouble shooting (see section 3.13).
3.8.4 Calibration Frequency
In most applications, monthly to quarterly calibration intervals will assure reliable detection. However, industrial environments differ. Upon initial installation and commissioning, close frequency tests should be performed, weekly to
monthly. Test results should be recorded and reviewed to determine a suitable calibration interval.
PI-500 Toxic Gas Sensors PG.22
Page 23
3.8.5 PID Plug-In Sensor Maintenance
The plug-in PID Sensor will need to be properly maintained to achieve proper long-term performance. All PID sensors
use a UV lamp that has a finite lifetime. The Detcon PID UV lamp source is expected to last a least 1 year. However,
from the time of installation a gradual loss in UV lamp strength is expected. (See Figure 8) As the UV lamp strength
decreases the sensor signal will decrease accordingly. This dictates that periodic span calibrations are required to maintain calibration accuracy.
Figure 8
To determine the present signal strength of the PID sensor, execute a valid span calibration and view the Sensor Life
from the ‘
View Program Status
’ menu. Any Sensor Life value less than 30% should result in the user’s choice of replac-
ing the plug-in sensor, cleaning the UV Lamp, or replacing the UV Lamp.
If the PID sensor seems to be losing signal strength at a rate faster than Figure 8 estimates, the sensor is most likely
experiencing contamination film build-up on the UV optical filter. This will happen when exposed to certain gases or
ambient contaminations that collect on the surafce of the UV filter. The result is a decrease in the amount of emitted
UV from the lamp source. This is known to happen with gases that can be polymerized by UV light (such as heavy
complex VOC’s), airborne oil vapors, and very fine dust. As UV Filter contamination occurs, the sensor’s signal
strength falls off in addition to the expected loss rate shown in Figure 8. This phenomenon can be reversed by disassembling the sensor and carefully cleaning the UV lamp filter using a specialized cloth.
It is also possible under certain ambient contamination conditions that the sensor’s Detector Cell can have a partially
conductive film that forms across the contact grids. This condition causes the zero background signal to gradually
increase to the point where it becomes unacceptable for the range of signal input to the transmiter electronics. When
this occurs the detector cell should be replaced. This can be checked by examining the amount of raw signal that is
produced during exposure to zero gas. Refer to the ‘View Program Status’ menu and record the Raw Signal report after
5 minutes of zero gas exposure. A value that exceds 2.65V would be evidence of this problem.
3.8.5.1 General recomendations for Sensor Maintenance
1) For normal environmental exposure and signal decay, replace the plug-in sensor every 9-12 months (especially if
there are no skilled technicians to handle proper UV lamp replacement).
2) If skilled technicians are available, replace just the UV lamp every 9-12 months.
3) For abnormally high rates of signal decay, clean the UV lamp monthly, using a Lamp Cleaning Kit, and replace the
UV lamp every 9-12 months.
4) For any proven cases where the zero baseline has drifted up, replace the detector cell.
PI-500 Toxic Gas Sensors PG.23
Page 24
3.8.5.2 PID Plug-In Sensor Maintenance Procedure
All piD Sensor Cells contain six user replacable parts.
Disassembly
1) Power down the instrument and remove the sensor cell.
2) Remove the filtercap by applying a slight upward pressure with the tip of a screwdriver or an Exacto Blade just
below the hole in the cap and between the cap and the housing.
3) With a fine tipped tweezers, remove both of the Filter Media and set aside.
4) Using an Exacto Blade, remove the spacer and set it aside
PI-500 Toxic Gas Sensors PG.24
Page 25
5) With a fine tipped tweezers, gcarefilly remove the cell assembly by prying under the cell’s edge where the connector
pins are located.
6) With a fine tipped tweezers, grasp the lamp by placing the tips in the housing notch and gently pulling it out. Be
careful not to scratch the lamp or chip the edges.
Cleaning the Lamp
Grab the lamp by the cylindrical glass body and clean the window by rubbiong it against the polishing pad. Use a cir-
cular motion and try to keep the window surface flat relative to the pad. Five seconds of rubbing should be enough in
most cases. Another indication of cleaning completeness is that about 1/16th of the pads surface is used in the
process.
PI-500 Toxic Gas Sensors PG.25
Page 26
Reassembly
1) Install the lamp into the sensor, making sure that the lamps metalized pads are aligned with the corresponding exci-
tation springs inside the lamp cavity.
2) With the end of the clean tweezers, or a clean blade of a screwdriver, press down firmly being careful not to scratch
the surface of the lamp.
3) Using fine tipped tweezers, install the cell assembly. Align the pins with the corresponding sockets on the sensor
and push down on the end with the pins. Make sure the cell assembly is flush with the lamp window.
4) Place the spacer around the assembly.
PI-500 Toxic Gas Sensors PG.26
Page 27
5) Place the filter media over the Cell Assembly centered on the top of the sensor. Make sure the filters are installed in
the correct order. Filter Media #2 first, then Filter Media #1 on top, with the shiny side up.
6) Align the Cap Key with the notch on the housing. Starting at the side opposite the notch, press down until the
Filter Cap snaps on to the housing. If the Cap Key is incorrectly aligned there will be a noticable buldge on the side of
the Cap.
3.9 STATUS OF PROGRAMMING, CALIBRATION LEVEL, AND SENSOR LIFE
The programming menu has a “View Program Status” listing that allows the operator to view the gas, range, and software version number of the program, as well as the calibration gas level setting, and estimated remaining sensor life. The
programming menu also allows the changing of the calibration gas level setting (see section 3.8.2).
The following procedure is used to view the programming status of the sensor:
a) First, enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30 seconds
until the display reads “VIEW PROG STATUS”, then withdraw the magnet. At this point you can scroll
through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The
menu options are: View Program Status, and Set Cal Level.
b) Next, scroll to the “VIEW PROG STATUS” listing and then hold the programming magnet over “PGM 1” for 3
seconds. The menu will then automatically scroll, at five second intervals, through the following information before
returning back to the “VIEW PROG STATUS” listing.
PI-500 Toxic Gas Sensors PG.27
Page 28
1 -
The software version number.
- Range is ###.
2
3 - Calibration gas level setting. The menu item appears as: “CalLevel @ xxPPM”
4 - Response Factor Setting. The menu item appears as: “RespFactor = x.xx”
5 - Zero Offset Setting. The menu item appears as: “ZeroOffset = x.x”
6 - Raw signal from sensor head. The menu item appears as: “RawSignal = x.xxV”
7 - The estimated remaining sensor life. The menu item appears as: “SENSOR LIFE 100%”
c) Exit back to normal operations by holding the programming magnet over “PGM 2” for 3 seconds, or automatically
return to normal operation in 30 seconds.
3.10 PROGRAM FEATURES
Detcon MicroSafe™ PID gas sensors incorporate a comprehensive program to accommodate easy operator interface
and fail-safe operation. Program features are detailed in this section. Each sensor is factory tested, programmed, and calibrated prior to shipment.
Over Range
When the sensor detects gas greater than 100% of range, it will cause the display to flash the highest reading of its range on
and off.
Under Range Fault(s)
If the sensor should drift below a zero baseline of -10% of range, the display will indicate a fault: “ZERO FAULT”.
This is typically fixed by performing another zero cal. When the total negative zero drift exceeds the acceptable threshold the display will indicate “SENSOR FAULT” and you will longer be able to zero calibrate.
Span Fault #1
If during span calibration the sensor circuitry is unable to attain a minimum defined response to span gas, the sensor will
enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading and the
calibration fault screen which appears as: “SPAN FAULT #1”. The previous calibration settings will remain saved in
memory. Previous span calibration is retained.
Span Fault #2
If during the span routine, the sensor circuitry is unable to attain a minimum defined stabilization point, the sensor will
enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading and the
calibration fault screen which appears as “SPAN FAULT #2”. Previous span calibration is retained.
Memory Fault
If new data points cannot successfully be stored to memory the display will indicate: “MEMORY FAULT”.
Fail-Safe/Fault Supervision
Detcon MicroSafe™ sensors are programmed for fail-safe operation. All fault conditions will illuminate the fault LED,
and cause the display to read its corresponding fault condition: “ZERO FAULT”, “SENSOR FAULT”, “SPANFAULT #1”, or “SPAN FAULT #2”. A “SENSOR FAULT” and “ZERO FAULT” will cause the mA output to
drop to zero (0) mA.
Sensor Life
The “Sensor Life” feature gauges the remaining sensor life based on signal output from the PID sensor cell. When a sensor life of 25% or less remains, the sensor cell should be replaced within a reasonable maintenance schedule.
3.11 DISPLAY CONTRAST ADJUST
Detcon MicroSafe™ sensors feature a 16 character backlit liquid crystal display. Like most LCDs, character contrast can
be affected by viewing angle and temperature. Temperature compensation circuitry included in the MicroSafe™ design
will compensate for this characteristic, however temperature extremes may still cause a shift in the contrast. Display contrast can be adjusted by the user if necessary. However, changing the contrast requires that the sensor housing be
opened, thus declassification of the area is required.
PI-500 Toxic Gas Sensors PG.28
Page 29
To adjust the display contrast, remove the enclosure cover and use a jewelers screwdriver to turn the contrast adjust screw
ocated beneath the metallic face plate. The adjustment location is marked “CONTRAST”. See figure 6 for location.
The Model PI-500 uses a universal transmitter design that allows the transmitter to be set up for any target gas and any
toxic concentration range. The original transmitter set-up is done at Detcon Inc. as part of the sensor test and calibration procedure, but it may also be changed in the field if necessary. The Universal Transmitter feature is a significant
convenience to the user because it allows hardware f lexibility and minimizes the spare parts requirements to handle
unexpected transmitter failures of different gas/ranges. It is however, absolutely critical that changes to gas/range set-up
of the Universal Transmitter be consistent with the gas type and range of the PID Sensor Head that it is connected to.
The PID sensor head will display the range it is set up for, based on the isobutylene reference calibration.
NOTE:
it is mated with.
If the Universal Transmitter needs to be changed for gas type and range follow this procedure. First, unplug the transmitter temporarily and then plug it back in. While the message “Universal Transmitter” appears, take the program magnet and swipe it over magnet PGM1. This will reveal the set-up options for gas range and gas type.
Swipe over PGM1to advance through the options for gas range which include:
1, 2, 3….10 ppm
10, 15, 20……100 ppm
100, 200, 300…..1000 ppm
1000, 2000, 3000 …..10,000 ppm
When the correct range is displayed, hold magnet over PGM1 for 3 seconds to accept the selection.
Next is your selection for the gas type that will be displayed. Note, the default gas is “VOC”. In this set-up you will
enter the alpha-numeric characters of the gas type. There is space for the chemical formula name of up to six characters.
Use PGM1 and PGM2 swipes to advance through the alphabet and numbers 0-9 selection (there is a blank space after
9). When the correct alphanumeric character is highlighted, hold the magnet over PGM1 for 3 seconds to lock it in.
This moves you to the next blank and the procedure is repeated until the chemical formula is completed. After the 6th
character is locked in the transmitter will proceed to normal operation.
NOTE 1: If the gas symbol has more than 6 characters, the symbol can be replaced by an abbreviated version of the
target gas name such as TOL or TOLUEN for Toluene which has a the symbol C6H5CH3. For epichlorohydrin (symbol C3H5OCL) you can substitute the name EPI or EPICHL etc.
If the Universal Transmitter is changed for gas type and range, it must be consistent with the PID sensor head
NOTE 2: When the Universal Transmitter is re-initialized and a new gas and range is entered, the previous customer
settings for span gas value, response factor, and zero offset are reset to default levels. This must be re-programmed back
to the customer specific settings.
3.13 TROUBLE SHOOTING
Sensor reads Over-range after Power-up
Probable Cause: Sensor requiring additional stabilization time, VOC gases present in background air, Improper zero or
span calibration.
1. Verify that there is not large amounts of target gas or interfering gases in background.
2. Redo zero and span calibrations.
3. Make sure transmitter range is consistent with PID sensor head range.
Reading Higher than Anticipated
Probable Causes: Target or Interfering gases in background, Incorrect calibration for Zero or Span.
1. Verify no target or interfering gases are present. If so, use the Zero Offset feature.
2. Redo Zero and Span calibrations with validated Zero Gas and Span Gas standards.
PI-500 Toxic Gas Sensors PG.29
Page 30
Reading Lower than Anticipated
robable Causes: Zero Calibration done before unit finished stabilizing, Incorrect Span Calibration.
P
1. Redo Zero and Span calibrations with validated Zero Gas and Span Gas standards.
2. Disengage Zero Offset feature if it is not necessary.
3. Contact Detcon to determine if target gas will not diffuse past the 316 Stainless Steel flame arrestor.
Zero Calibration Fault
Probable Causes: Target gas or Interfering gases in background during Zero Calibration, Failed PID sensor.
1. Verify no target or interfering gases are present.
2. Redo Zero calibration with validated Zero Gas standard.
3. If recovering after a start-up, give more time to stabilize.
Span Calibration Fault
Probable Causes: Failed PID sensor (failed UV lamp bulb, optical window needs cleaning), ice/mud/dust blocking protective membrane, invalid span calibration gas due to age, type, and contamination or insufficient flow rate.
1. Verify there is no ice/mud/dust blocking the sensor’s Flame Arrestor.
2. Redo Span Calibration with validated Isobutylene Span Gas standard.
3. Reinitialize unit by plugging in transmitter while holding the magnet on PGM1. Scroll through and select the cor-
rect gas type. Make sure all customer settings are re-entered after “reinitialization”.
4. Clean UV Lamp or replace UV Lamp.
5. Replace with new PID sensor.
Noisy Sensor (continuous drift) or suddenly Spiking
Probable Cause: Unstable power source, Inadequate grounding, Inadequate RFI protection.
1. Verify power Source output and stability.
2. Contact Detcon for assistance in optimizing shielding and grounding.
3. Add RFI Protection accessory available from Detcon.
LCD Difficult to Read
Probable Cause: Needs adjustment.
1. Reinitialize unit by plugging in transmitter and the swiping the magnet over PGM1 while “Universal Transmitter” is
displayed. Scroll through and select the correct gas type and range (see section 3.12 Universal Transmitter Features).
Make sure all customer specific settings are re-entered after “reinitialization”.
3.14 SPARE PARTS LIST
943-000006-132Calibration Adapter
500-005065-007Connector board
327-000000-000Programming Magnet
897-850800-0003 port enclosure less cover
897-850700-000Enclosure glass lens cover
960-202200-000Condensation prevention packet (replace annually).
925-P15480-rangePI-500 Series Universal Plug-in Control Circuit
370-P10000-000Plug-in Replaceable PID sensor for 0-20 ppm and lower ranges (PI-500 Model)
370-P20000-000Plug-in Replaceable PID sensor for =>50 ppm ranges (PI-501 Model)
975-600020-000PID 10.6eV Replacement Lamp
975-600100-000PID Detector Replacement Cell
975-520040-000PID Lamp Cleaning Kit
390-404142-range* PID sensor head assembly
* Does not include plug-in replacement sensor cell.
Specify 3 Digit Range for PID sensor head as per examples below: If greater than 999ppm, use a “K” (for 1000). If
greater than 9,900ppm use a “P” (for %).
Detcon, Inc., as manufacturer, warrants each new PID plug-in sensor cell, for a specified period under the conditions
described as follows: The warranty period begins on the date of shipment to the original purchaser and ends 6 months
thereafter. The sensor cell is warranted to be free from defects in material and workmanship. Should any sensor cell fail
to perform in accordance with published specifications within the warranty period, return the defective part to Detcon,
Inc., 3200 A-1 Research Forest Dr., The Woodlands, Texas 77381, for necessary repairs or replacement.
NOTE:
cleaning.
3.16 SERVICE POLICY
Detcon, Inc., as manufacturer, warrants under intended normal use each new PI-500 series plug-in signal transmitter
Control Circuit and PID Sensor Head circuit to be free from defects in material and workmanship for a period of two
years from the date of shipment to the original purchaser. Detcon, Inc., further provides for a five year fixed fee service
policy wherein any failed signal Transmitter shall be repaired or replaced as is deemed necessary by Detcon, Inc., for a
fixed fee of $65.00. Any failed PID Sensor Head circuit shall be repaired or replaced as is deemed necessary by Detcon,
Inc., for a fixed fee of $75.00. The fixed fee service policy shall affect any factory repair for the period following the two
year warranty and shall end five years after expiration of the warranty. All warranties and service policies are FOB the
Detcon facility located in The Woodlands, Texas.
This warranty does not cover conditions where the detector cell or lamp may be dirty and can be restored by
Mailing Address: P.O. Box 8067, The Woodlands, Texas 77387-8067