Detcon MicroSafe DM-534C Instruction Manual

Page 1
Detcon MicroSafe™
Model DM-534C
Oxygen Deficiency Sensor (0-25% O2)
Operator’s Installation and Instruction Manual
4055 Technology Forest Blvd, Suite 100,
The Woodlands, Texas 77381
Ph.281.367.4100 / Fax 281.298.2868
Page 2
DM-534C Oxygen Sensor Assembly
This page left intentionally blank
DM-534C O2 Sensor Instruction Manual ii
Page 3
DM-534C Oxygen Sensor Assembly
Table of Contents
1.0 DESCRIPTION ........................................................................................................................................ 1
1.1 Sensor Technology ............................................................................................................................... 1
1.2 Base Connector Board .......................................................................................................................... 2
1.3 Explosion Proof Enclosure ................................................................................................................... 2
2.0 PRINCIPLE OF OPERATION .............................................................................................................. 3
3.0 APPLICATION ....................................................................................................................................... 3
3.1 Sensor Placement/Mounting ................................................................................................................. 3
3.2 Interference Data .................................................................................................................................. 4
4.0 SPECIFICATIONS ................................................................................................................................. 4
5.0 OPERATING SOFTWARE ................................................................................................................... 4
5.1 Normal Operation ................................................................................................................................. 5
5.2 Calibration Mode .................................................................................................................................. 5
5.3 Program Mode ...................................................................................................................................... 5
5.3.1 Program Status ............................................................................................................................. 5
5.3.2 Calibration Level Adjustment ....................................................................................................... 5
6.0 INSTALLATION ..................................................................................................................................... 5
6.1 Field Wiring Table (4-20 mA output) .................................................................................................. 5
6.2 Sensor Location .................................................................................................................................... 6
6.3 Local Electrical Codes .......................................................................................................................... 7
6.4 Accessibility ......................................................................................................................................... 7
6.5 Installation Procedure ........................................................................................................................... 7
6.6 Remote Mounting Applications ........................................................................................................... 8
7.0 START UP ................................................................................................................................................ 9
8.0 CALIBRATION ..................................................................................................................................... 10
8.1 Calibration Procedure - Span ............................................................................................................. 11
8.2 Additional Notes ................................................................................................................................. 12
8.3 Calibration Frequency ........................................................................................................................ 12
8.4 Calibration Gas Level ......................................................................................................................... 12
8.5 Status of Programming and Sensor Life ............................................................................................. 13
8.6 Program Features ................................................................................................................................ 13
8.7 Display Contrast Adjust ..................................................................................................................... 14
9.0 TROUBLE SHOOTING ....................................................................................................................... 14
10.0 SPARE PARTS LIST ............................................................................................................................ 15
11.0 WARRANTY ......................................................................................................................................... 16
12.0 SERVICE POLICY ............................................................................................................................... 16
13.0 SOFTWARE FLOWCHART ............................................................................................................... 16
Appendix C ....................................................................................................................................................... 17
Revision History ............................................................................................................................................. 17
DM-534C O2 Sensor Instruction Manual iii
Page 4
DM-534C Oxygen Sensor Assembly
Shipping Address: 4055 technology Forest Blvd, Suite 100., The Woodlands Texas 77381
Phone: 888.367.4286, 281.367.4100 • F ax: 281.292.2860 • www.detcon.comsales@detcon.com
Table of Figures
Figure 1 Construction of Galvanic Cell ............................................................................................................. 1
Figure 2 Microprocessor Control Circuit .......................................................................................................... 1
Figure 3 Base connector board .......................................................................................................................... 2
Figure 4 Explosion proof enclosures ................................................................................................................. 3
Figure 5 Functional Block Diagram .................................................................................................................. 3
Figure 6 Typical Installation ............................................................................................................................. 7
Figure 7 Typical Outline and Mounting Dimensions ........................................................................................ 8
Figure 8 Sensor Connector PCB ....................................................................................................................... 8
Figure 9 Remote mounting application ............................................................................................................. 9
Figure 10 Magnetic Programming Tool ............................................................................................................ 10
Figure 11 Spare parts diagram ........................................................................................................................... 15
Figure 12 Software Flowchart ........................................................................................................................... 16
List of Tables
Table 1 Interference Data .................................................................................................................................. 4
Table 2 Field wiring Table ................................................................................................................................ 6
Mailing Address: P.O. Box 8067, The W oodlands Texas 77387-8067
DM-534C O2 Sensor Instruction Manual iv
Page 5
DM-534C Oxygen Sensor Assembly
1.0 DESCRIPTION
Detcon MicroSafe™ Model DM534C, oxygen defici ency sensors are non-intrusive “Smart” sensors desig ned to detect and monitor O2 in air over the range of 0-25%. One of the primary features of the sensor is its method of automatic calibration which guides the user through each step via instructions displayed on the backlit LCD. The sensor output is a standard 4-20 mA signal. The microprocessor-supervised electronics are packaged as a plug-in module that mates to a standard connector board. Both are housed in an explosion proof condulet that includes a glass lens. A 16 character alpha-numeric indicator is used to display sensor readings as well as the sensor’s menu driven features via a handheld programming magnet.
1.1 Sensor Technology
The sensor technology is of the two-electrode, gal vanic metal air battery type cell, which is ho used as a field replaceable plug–in module. The cell is diffusion limited and functions as a direct current generator proportional to the amount of oxygen adsorption. The sensors are temperature compensated and show good accuracy and stability over the operatin g temperature range -4° to +122° Fahrenheit. T he sensor is warranted for two years and has an expected service li f e of up to three years in ambient air at 20.9% oxygen.
Figure 1 Construction of Galvanic Cell
The control circuit is microprocesso r based and is packaged as a plug-in field replaceable module, facilitating easy replacement and minimum down time. Circuit functions include a basic sensor preamplifier, onboard power supplies, microprocessor, back lit alpha numeric display, calibration and fault status LED ind icators, magnetic programming switches, and a linear 4-20 mA DC output.
Figure 2 Microprocessor Control Circuit
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 1 of 17
Page 6
DM-534C Oxygen Sensor Assembly
Customer
Wiring
Wiring to O2 Sensor
1.2 Base Connector Board
The base connector board is mounted in the explosion proof enclosure and includes: the mating connector for the control circuit, reverse input and secondary transient suppression, input filter, and lugless terminals for field wiring.
Figure 3 Base connector board
1.3 Explosion Proof Enclosure
The sensors are packaged in a cast me tal explosion proof enclosure. The enclosure is fitted with a threaded cover that has a glass lens window. Mag netic program switches located behind the transmitter module face plate are activated through the lens wind ow via a han dheld magnetic programming tool allowing non-intrusive operator interface with the sensor. Calibration can be accomplished without removing the cover or declassifying the area. Electrical classification is Class I; Groups B, C, D; Division 1.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 2 of 17
Page 7
DM-534C Oxygen Sensor Assembly
Figure 4 Explosion proof enclosures
2.0 PRINCIPLE OF OPERATION
Method of detection is by a controlled rate of diffusion. Air and gas diffuse through a sintered stainless steel filter and a diffusion barrier. As oxyg en is adsorb ed into the electro ly te solutio n a current i s generated betw een the cathode and anode electrodes. This current output rises with increases in oxygen concentration and reverses with lower concentrations. The quick response of the cell results in continuous monitoring of ambient air conditions.
Figure 5 Functional Block Dia gra m
3.0 APPLICATION
Model DM-534C MicroSafe™ sensors are designed to detect and monitor oxygen deficiency in ambient air in the range of 0-25%. Minimum sensitivity and scale resolution is 0.1%. Operating temperature range is -4° F. to +122° F. While the sensor is capable of operatin g outside these temperatures, performance specificati ons are verified within the limit.
3.1 Sensor Placement/Mounting
Sensor location should be reviewed by facility engineering and safety personnel. Area leak sources and perimeter mounting are typically used to determine number and location of sensors. The sensors are gen erally located 2-4 feet above grade.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 3 of 17
Page 8
DM-534C Oxygen Sensor Assembly
Methane
100% = 0
Hydrocarbons
100% = 0
Hydrogen
100% = < 2%
Carbon Monoxide
20% = < 0.5%
3.2 Interference Data
Table 1 Interference Data
4.0 SPECIFICATIONS
Method of Detection Air battery diffusion/adsorption
Electrical Classification Class I; Groups B, C, D; Div. 1.
Response Time (T90) T90 < 20 seconds
Clearing Time 90% < 20 seconds
Repeatability ± 2% FS
Range 0-25% O2
Operating Temperature -4° to +122° F
Accuracy ± 2% FS
Linearity ± 0.5% FS
Sensor Warranty 2 year conditional
Power Consumption 60 mA @ 24 VDC
Output Linear 4-20 mA DC
Input Voltage 11.5-28 VDC
5.0 OPERATING SOFTWARE
Operating software is menu listed with operator interface via the two magnet ic program switches located under the face plate. The two switches ar e referred to as “PGM 1” and “PGM 2”. The menu list c onsists of 3 items which include submenus as indicated belo w. (Note: see the end of this manual for a comple te software flow chart.)
1. Normal Operation a) Current Status
2. Calibration Mode a) Span
3. Program Menu a) Program Status b) Set Calibration Level
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 4 of 17
Page 9
DM-534C Oxygen Sensor Assembly
5.1 Normal Operation
In normal operation, the display tracks the cu rrent status of the sensor and gas concentration and app ears as: “20.9 % O2”" The mA current output corresponds to the monitoring level of 0-25% O2 = 4-20 mA.
5.2 Calibration Mode
Calibration mode allows for sensor zero and span adjustments. “2 - SPAN
The default span adjustment is set at 20.9% which is the normal atmospheric concentration of O2. Span gas concentrations other than 20.9 % may be used. Refer to section 8.1 for details “SPAN
5.3 Program Mode
The program mode provides a program status menu (Vi ew Program Status) to check operational parameters and allows for the selection of the calibration gas level setting.
5.3.1 Program Status
The program status scrolls through a menu that displays: * The gas type, range of detection and software version number. The menu item appears as: “O2 0-25 Vxxx” * The calibration gas level setting. The menu item appears as: CalLevel @ xx.x%” * The estimated remaining sensor life. The menu item appears as: “SENSOR LIFE 100%
5.3.2 Calibration Level Adjustment
The Calibration level is adjustable from 15.0% to 25.0% O2. The menu item appears as: “CalLevel @ # #%”. Factory default setting is 20.9%.
6.0 INSTALLATION
Optimum performance of ambient air/g as sensor devices is directl y relative to proper locati on and installation practice.
6.1 Field Wiring Table (4-20 mA output)
Detcon MicroSafe™ O2 sensor assemblies req uire three conductor connection between power supplies and host electronic controllers. Wiring designators are + (DC), - (DC), and mA (sensor signal). Maximu m single conductor resistance between sensor and controller is 10 ohms. Maximum wire size fo r termination in the sensor assembly terminal board is 14 AWG.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 5 of 17
Page 10
DM-534C Oxygen Sensor Assembly
Note 1: This wiring table is based on stranded tinned copper wire and is designed to serve
as a reference only.
Note 2: Shielded cable may be requir ed in installations where cable tray s or conduit runs include high voltage lines or other sources of induced interference.
Note: In all installations, the sensor points down relative to grade (Figure 6). Improper sensor orientation may result in false read ing and permanent sensor damage.
AWG
Meters
Feet
20
240
800
18
360
1200
16
600
2000
14
900
3000
Table 2 Field wiring Table
The RS485 (if applicable) requires 24 gauge, two conductor, shielded, twisted pair cable between sensor and host PC. Use Belden part number 9841. Two sets of terminals are located on the connector board to facilitate serial loop wiring from sensor to sensor. Wiring designators are A & B (IN) and A & B (OUT).
6.2 Sensor Location
Selection of sensor location is critical to the overall safe performance of the product. Five factors play an important role in selection of sensor locations:
(1) Density of the gas to be detected (2) Most probable leak sources within the industrial process (3) Ventilation or prevailing wind conditions (4) Personnel exposure (5) Accessibility for routine maintenance
Density - Placement of sensors relative to the density of the t arget gas is such that sensors fo r the detection of heavier than air gases should be located within 2-4 feet of g rade as these heavy gases wil l tend to settle in low lying areas. For gases lighter than air, sensor placement should be 4-8 feet above grade in open areas or in pitched areas of enclosed spaces.
Leak Sources - Most probable leak sources within an indu strial process include flanges, valves, and tubing connections of the sealed type wh ere seals may either fail or we ar. Other leak sources are b est determined by facility engineers with experience in similar processes.
Ventilation - Normal ventilation or prevailing wind conditions can dictate efficient location of gas sensors in a manner where the migration of gas clouds is quickly detected.
Personnel Exposure The undetected migration of gas clouds should not be allowed to approach concentrated personnel areas such as control ro oms, maintenance or warehouse buildings. A more general an d applicable thought toward selecting sensor location is combining leak source and perimeter protection in the best possible configuration.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 6 of 17
Page 11
DM-534C Oxygen Sensor Assembly
Drain
Conduit
"T"
EYS Seal Fitting
PGM 1
PGM 2
MODEL DM-534C
HOUSTON, TEXAS
FLT CAL
MicroSafe O2 Gas Sensor
ALM ALM
TM
NOTE: For products utilizing the aluminum junction box option, the conduit seal shall be
enclosure. Crouse Hinds type EYS2, EYD2 or equivalent are suitable for this purpose.
Figure 6 Typical Installation
6.3 Local Electrical Codes
Sensor and transmitter assemblies should be installed in accordance with all local electrical codes. Use appropriate conduit seals. Drains & breathers are recommended. The sensor assembli es are designed to meet NEC and CSA requirements for Class I; Grou ps B, C, D; Div. 1 environments.
6.4 Accessibility
Consideration should be given to easy access by maintenance personnel as well as the consequences of close proximity to contaminants that may foul the sensor prematurely.
placed at the entry to the junction box (see Figure 6 as an example). For products utilizing the stainless steel junction box option, the conduit seal shall be placed within 18” of the
6.5 Installation Procedure
a) Remove the junction box cover and unplug the control circuit by grasping the two thumb screws and pulling outward. b) Securely mount the sensor junction box in accordance with recommended practice. See dimensional drawing (Figure 7).
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 7 of 17
Page 12
DM-534C Oxygen Sensor Assembly
5.5"
6.1"
5.825"
7.92"
2"
0.5"
4.65"
Wall (or other
mounting surface
Splash Guard
O2 Sensor
Cal Port
1/4" Mounting holes
8-32 tapped ground point
3/4" NPT Ports
Customer
Wiring
Wiring from O2 Sensor
BlueBlue
White
Figure 7 Typical Outline and Mounting Dimensions
c) Observing correct polarity, terminate 3-conductor field wiring to the sensor base connector board in accordance with the detail shown in Figure 8. d) Replace the plug-in transmitter circuit and replace the junction box cover.
6.6 Remote Mounting Applications
Some sensor mounting applications require that the gas sensor head be remotely mounted away from the sensor transmitter. This is usually true in in stances where the gas sensor head must be mounted in a location that is difficult to access. Such a lo cation creates problems for main tenance and calibration activities. De tcon provides the DM-534C sensor in a remote-mount configuration in which the sensor (Model DM-534C-RS) and the transmitter (Model DM-534C-RT) are provided in their own condulet housing and are interfaced together with a three conductor cable. Reference f i gure Figure 9 below for wiring diagram.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 8 of 17
Figure 8 Sensor Connector PCB
Page 13
DM-534C Oxygen Sensor Assembly
Blue
White
Blue
White
Remote Transmitter DM-534C-RT
Remote Sensor DM-534C-RS
Note 1: If the display contrast needs adjustment, refer to section 8.7.
Figure 9 Remote mounting application
7.0 START UP
Upon completion of all mechanical mounting and termination of all field wiring, apply system power and observe the following normal conditions
a) “Fault” LED is off. b) A tempor ary upscale reading may occur as the sensor powers up. This upscale read ing will clear to
about 20.9% within a few minutes of turn-on, assuming there is no oxygen deficient condition in the area of the sensor.
Initial Operational Tests
After a warm up period has been allowed f or, the sensor should b e checked to verify reliable sensitivity to O2 gas.
Material Requirements
-Detcon PN 327-000000-000 MicroSafe™ Programming Magnet
-Detcon PN 613-120000-000 Splash Guard with integral Cal Port -OR-
-Detcon PN 943-000006-132 Threaded Calibration Adapter
-Detcon PN 942-001123-000 Zero Air cal gas or use ambient air if no combustible gas is present.
a) Attach the calibration adapter to the threaded sensor housi ng. Apply the test gas at a controlled flow
rate of 500 ml/m. Observe that the LCD display decreases to a level of 3% or less.
b) b) Remove the test gas and observe that the LCD display increases back to 20.9% ±2% of scale (0.5%
O2).
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 9 of 17
Page 14
DM-534C Oxygen Sensor Assembly
NOTE: If, after entering the calibration or program menus, there is no interaction with the
menu items for more than 30 seconds, the sensor will return to its normal operating condition.
c) Initial operational tests are complete. Detcon O2 g as sensors are pre-calibrated prior to shipment and
will, in most cases, not require significant adjustment on start up. However, it is recommend ed that a complete calibration test and adjustment be performed within 24 hours of installation. Refer to calibration instructions in section 8.0.
8.0 CALIBRATION
Material Requirements
-Detcon PN 327-000000-000 MicroSafe™ Programming Magnet
-Detcon PN 613-120000-000 Splash Guard with integral Cal Port -OR-
-Detcon PN 943-000006-132 Threaded Calibration Adapter
-Detcon PN 942-001123-000 Zero Air cal gas or use ambient air if no combustible gas is present.
Programming Magnet Operating Instructions
Operator interface to MicroSafe™ gas detection products is via magnetic switches located behind the transmitter face plate. DO NOT remove the glass l ens cover to calibrate or change programming parameters. Two switches labeled “PGM 1” and “PGM 2” allow for complete calibration and alarm le vel programming without removing the enclosure cover, thereby eliminating the need for area declassification or the use of hot permits.
Figure 10 Magnetic Programming Tool
A magnetic programming tool (see Figure 10) is used to operate the switches. Switch action is defined as momentary contact, 3-second hold, and 30-second hold. In momentary contact use, the programming magnet is waved over a switch location. In 3 second hold, the programming magnet is held in place over a switch location for 3 or more seconds. In 30 second hold, the programming magnet is held in place over a switch location for 30 or more seconds. Three and thirty second hold is used to enter or exit calibration and program menus while momentary contact is used to make adjustments. The location of “PGM 1” and “PGM 2” are shown in figure 6.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 10 of 17
Page 15
DM-534C Oxygen Sensor Assembly
NOTE 1: Before performing an ambient air O2 span calibration, be sure there is no
oxygen deficient condition present.
CAUTION: Verification of the correct calibration gas level setting and calibration span gas concentration is required before “span” calibration. These two numbers must be equal.
NOTE 1: If th e circuitry is unable to adjust the span to the proper setting the sensor will
FAULT” (see section 3.7.3)..
8.1 Calibration Procedure - Span
Calibration consists of entering the calibration function and following the menu-displayed instructions. The display will ask for the application of span gas in a specific concentration. This concentration must be equal to the calibration gas level setting. The factory default setting for span gas concentration is 20.9% O2 which is the normal atmospheric concentration. Other co ncentrations may be used as long as they fall with in 15.0% to
25.0% O2. However, any alternate span gas concentration value must be programmed via the calibration gas level menu before proceeding with span calibration. Follow the instructions below for span calibration.
a) Verify the current calibration gas level setting as indicated by the programming status menu. To do
this, follow the instructions in section 3.8 and make note of the setting found in listing number 12. The item appears as “CalGas @ xx.x %”.
b) If the calibration gas level setting is equal to your calibration span gas concentration, proceed to item
“f”. If not, adjust the calibration gas level setting so that it is equal to your calibration span gas concentration, as instructed in items “c” through “e”.
c) Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30
seconds until the display reads “VIEW PR OG STATUS”, and then withdraw the magnet. At this point you can scroll through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu options are: View Program Status, and Set Cal Level.
d) From the programming menu scroll to the calibration level listing. The menu item appears as: “SET
CAL LEVEL”. Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “CalGas @ ## %”, then withdraw the magnet. Use the programming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired calibration span gas concentration. Exit to the programming menu by holding the programming magnet over “PGM1” for 3 seconds.
e) Exit back to normal operation by holding the programming magnet over “PGM 2” for 3 seconds, or
automatically return to normal operation in 30 seconds.
f) From the calibration menu “2-SPAN” proceed into the span adjust function by holding the
programming magnet stationary over “PGM 2” for 3 seconds then withdraw the programming magnet. If no change of Span level is desired, wait 30 seconds for menu to return to normal operation. At this point the display will ask for the application of the target gas and concentration. The display reads “APPLY xx.x %” The xx.x here will indicate the actual concentration requested.
g) Apply the calibration test gas at a flow rate of 500 milliliters per minute. If the calibration gas level is
set at 20.9% and ambient air is verified to be 20.9% (normal atmospheric concentration of O2) then do nothing at this point. The sensor will auto calibrate to ambient air O2 concentration. After 3 minutes the sensor will auto span to the correct reading and the display will change to “REMOVE GAS” then the display will return to the normal operating mode.
enter into the calibration fault mode which will cause the display to alternate between the sensor’s current status reading and the calibration fault screen which appears a s: “CAL
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 11 of 17
Page 16
DM-534C Oxygen Sensor Assembly
8.2 Additional Notes
1. Upon entering the calibration menu, the 4-20 mA signal drops to 2 mA and is held at this level until
you return to normal operation.
2. If during calibration the sensor circuitry is unable to attain the proper adjustment for span, the sensor
will enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading an d the calibration fault screen wh ich appears as: “CAL F AULT” If this o ccurs you may attempt to recalibrate by entering the calibration menu as described in section 3.7.1a. If the sensor fails again, defer to technical trouble shooting.
8.3 Calibration Frequency
In most applications, monthly to quarterly calibration intervals will assure reliable detection. However, industrial environments differ. Upon initial installation and commissioning, close frequency tests should be performed, weekly to monthly. Test results should be recorded and reviewed to determine a suitable calibration interval.
8.4 Calibration Gas Level
Calibration consists of entering the calibration function and following the menu-displayed instructions. The default calibration gas level is 20.9% O2. The sensor however can be programmed to other calibration gas levels as long as they fall within the range of 15.0% to 25.0%. If this is the case, during span calibration the display will ask for the application o f span gas in the same concentration as the calibration gas level sett ing. Follow the instructions below for span calibration at settings other than 20.9%.
a) Verify the current calibration gas level setting as indicated by the programming status menu. To do
this, follow the instructions in section 8.0 and make note of the setting found in listing number 2. The item appears as “CalLevel @ xx.x%
b) If the calibration gas level setting is equal to your calibration span gas concentration, proceed to item
“f”. If not, adjust the calibration gas level setting so that it is equal to your calibration span gas concentration, as instructed in items “c” through “e”.
c) Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30
seconds until the display reads VIEW PROG STATUS”, and then withdraw the magnet. At this point you can scroll through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu options are: View Program Status, and Set Cal Level.
d) From the programming menu scroll to the calibration level listing. The menu item appears as: “ SET
CAL LEVEL”
e) Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the
display reads “CalLevel @ xx.x%”, and then withdraw the magnet. Use the programming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired calibration span gas concentration. Exit to the programming menu by holding the programming magnet over “PGM1” for 3 seconds.
f) Exit back to normal operation by holding the programming magnet over “PGM 2” for 3 seconds, or
automatically return to normal operation in 30 seconds.
g) From the calibration menu “2- SPAN” (section 8.1) proceed into the span adjust function by holding
the programming magnet stationary over “PGM 2” for 3 seconds until the display reads “APPLY xx.x% O2” then withdraw the programming magnet. The x’s here indicating the gas concentration requested.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 12 of 17
Page 17
DM-534C Oxygen Sensor Assembly
h) Apply the calibratio n gas at a flow rate of 500 millili ters per minut e. As the sensor signal changes, the
display will change to “SPAN xx.x%”. The “xx.x” part of the rea ding indicates th e actual gas re ading which will increase or decrease until the sen sor stabilizes. Wh en the sensor sign al is stable it will auto span to the requested concentration and the display will change to “SPAN COMPLETE” for two seconds and then “REMOVE GAS”. Remove the gas. As the signal level returns to ambient gas conditions, the display will return to the normal operation menu, “20.9% O2
8.5 Status of Programming and Sensor Life
The programming menu has a programming status listing that allows the operator to view the gas, range, and software version number of the pro gram, the calibration gas level setting , as well as the estimated remainin g sensor life.
The following procedure is used to view the programming status of the sensor:
a) First, enter the programming menu by holding the programming magnet stationary over “PGM 2” for
30 seconds until the display reads “VIEW PROG STATUS ”, then withdraw the magnet. The menu options are: View Program Status, and Set Cal Level.
b) Next, scroll to the “VIEW PROG STATUS” listing and then hold the programming magnet over
“PGM 1” for 3 seconds. The menu will then automatically scroll, at five second intervals, through the following information before returning back to the “” listing.
1. The gas type, range of detection and software version number: The menu item appears as: “O2
0-25 VX.XXX” with the Xs indicating the current version of installed firmware.
2. The calibration gas level setting: The menu item appears as: “CalLevel @ xx.x%”.
3. The estimated remaining sensor life: The menu item appears as: “Sensor life 100%”.
c) Exit back to normal operations by holding the programming magnet over “PGM 2” for 3 seconds, or
automatically return to normal operation in 30 seconds.
8.6 Program Features
Detcon MicroSafe™ toxic gas sensors incorporate a comprehensive program to accommodate easy operator interface and failsafe operation. Program features are detailed in this section. Each sensor is factory tested, programmed, and calibrated prior to shipment.
Over Range When the sensor detects gas greater than 2 5.0% O2, it will cause the display to flash “25% O2” on and off.
Sensor Fault
If either of the wires connecting the sensor cell to the connector board should fail and cause an open circuit, the sensor will go into a fault condition. “SENSOR FAULT
Calibration Fault
If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will enter into the calibration fault mode an d cause the display to altern ate between the sensor’ s current status reading and the calibration fault screen which appears as: “CAL FAULT
Fail-safe/Fault Supervision
Detcon MicroSafe™ sensors are programmed fo r failsafe operation. Any of the following fault condition will illuminate the fault LED, and cause the display to read its corresponding fault condition: “SENSOR FAULT”, or “CAL FAULT”. A “Sensor Fault” will also cau se the mA output to drop to zero (0) mA.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 13 of 17
Page 18
DM-534C Oxygen Sensor Assembly
Sensor Life
The sensor life feature is a reference bas ed on sign al output fro m the sensor cell . When a sen sor life of 2 5% or less remains; the sensor cell should be repl aced within a reasonable maintenance schedule.
8.7 Display Contrast Adjust
Detcon MicroSafe™ sensors feature a 1 6-character backlit liquid crystal display. Like most LCDs, character contrast can be affected by viewing angle and temperature. Temperature compensation circuitry included in the MicroSafe™ design will co mpensate for this charact eristic; however temp erature extremes may still cause a shift in the contrast. Display contrast can be adjusted by the user if necessary. However, changing the contrast requires that the sensor housing be opened, thus declassification of the area is required.
To adjust the display contrast, remove th e enclosure cover an d use a jeweler’s screwd river to turn t he contrast adjust screw located beneath the metallic face p late. The adjustment location is marked “CONTRAST”. See Figure 2 for location.
9.0 TROUBLE SHOOTING
Memory or Error Reports
1. Reinitialize Sensor Unplug transmitter and replug transmitter then swipe magnet over PGM 1 in the
first 3 seconds. This will clear the processor and recover from error state. Remember to put in all customer settings for range, alarm and cal gas l evel after re-initialization.
Non-readable Display
1. If display has blue background when hot, install sunshade to reduce temperature.
2. If poor contrast, adjust contrast pot accordingly.
Nothing Displayed – Transmitter not responding
1. Verify condulet has no accumulated water or abnormal corrosion.
2. Verify required DC power is applied to correct terminals.
3. Swap with a known good transmitter to determine if transmitter is faulty.
Bad 4-20 mA Output or RS485 Output
1. Check that wiring is connected to correct terminal outputs.
2. Swap with a known good transmitter to determine if transmitter is faulty.
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 14 of 17
Page 19
DM-534C Oxygen Sensor Assembly
600-610000-000
Sensor splash guard
943-000006-132
Calibration Adapter for FP, IR, DM and O2 Sensors
500-005065-007
Connector board
327-000000-000
Programming Magnet
897-850902-010
Detcon Alum Condulet w/Window Cover
960-202200-000
Condensation prevention packet (replace annually).
370-399100-000
Plug in replacement O2 sensor cell
925-345400-025
Plug in control circuit
399-200000000
Stainless steel sensor housing assembly (does not include plug in sensor
Sensor Head Wiring Connection
Sensor Head Wiring Connection
Sensor Head Wiring Connection
Sensor Head Wiring Connection
+24VDC Power IN
Common DC Power IN
4-20 mA Output
Customer Supplied
Wiring
6/32 X 3/8 Screw (2ea.)
#6 internal Star Washer (2ea.)
O2 Plug-In Sensor Assembly
P/N 370-399100-000
Transmitter Module DM-534 P/N 925-345400-025
Standard Connector Board P/N 500-005065-007
O2 SS SENSOR HOUSING P/N 399-200000-000
N7 Aluminum condulet, part of
Assembly # 897-850901-010
Splashguard with Calibration Adapter P/N:613-120000-700
Condensation Prevention Packet
P/N:960-202200-000
STANDARD CONNECTOR BOARD
P/N:500-005065-007
O2 Plug-In Sensor Assembly
P/N 027-02384-0
O'ring 1 1/4"ID X 1 7/16OD 0.103W
P/N 173
N7 Aluminum Lid with window, part
of Assembly # 897-850901-010
10.0 SPARE PARTS LIST
Figure 11 Spare parts diagram
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 15 of 17
Page 20
DM-534C Oxygen Sensor Assembly
11.0 WARRANTY
Detcon, Inc., as manufacturer, warrants each new plug-in O2 sensor cell (PN 370-399100-000), for a two y ear period under the conditions described as follows: The warranty period begins on the date of shipment to the original purchaser and ends two years thereafter. The sensor cell is warranted to be free from defects in material and workmanship. Should the sensor cell fail to perform in accordance with published specifications within the warranty period, return the defective part to Detcon, Inc. for necessary repair or replacement.
12.0 SERVICE POLICY
Detcon, Inc., as manufacturer, warrants under intended normal use each new MicroSafe™ plug-in control circuit to be free from defects in material and workmanship for a period of two years from the date of shipment to the original purchaser. Detcon, Inc., further provides for a five y ear fixed fee service policy covering the control circuit wherein any failed part s hall be repaired for a fee of $55. 00. The fixed fee service policy shall affect any necessary factory repai r for the period following the two-year warranty period and sh all end five years after expiration of the warranty . All warranties and service po licies are FOB the Detcon facili ty located in The Woodlands, Texas.
13.0 SOFTWARE FLOWCHART
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 16 of 17
Figure 12 Software Flowchart
Page 21
DM-534C Oxygen Sensor Assembly
Revision
Date
Changes made
Approvals
1.3
02/05/10
Last release
BM
1.4
02/21/12
Removed section Calibration- Zero Span. Corrected Spare
Word format
BM
1.5
07/20/18
Updated Conduit Seal in Section 6.4
MM
Shipping Address: 4055 technology Forest Blvd, Suite 100., The Woodlands Texas 77381
Phone: 888.367.4286, 281.367.4100 • F ax: 281.292.2860 • www.detcon.comsales@detcon.com
Appendix C
Revision History
Parts List (section 10.0). Added Revision history. Converted to
Mailing Address: P.O. Box 8067, The Woodlands Texas 77387-8067
DM-534C O2 Sensor Instruction Manual Rev. 1.5 Page 17 of 17
Loading...