Detcon IR-642 User Manual

detcon inc.
Detcon MicroSafe
IR-640 Carbon Dioxide Sensor (0-5%CO2)
IR642 Carbon Dioxide Sensor (10%-100%)
Operator’s Installation & Instruction Manual
November 26, 2007 • Document #2196 • Version 6.4
CAUTION:
Before operating the Model IR-640/IR-642 sensor, read this manual thoroughly and verify that the configuration of default factory settings are appropriate and correct for your application. The settings include: relay contact outputs (section 3.5.5d), alarm settings (section 3.5.5e and
3.9), and RS-485 ID (section 3.5.5f and 3.12).
Table of Contents
3.0 Description
3.1 Principle of Operation
3.2 Application
3.3 Specifications
3.4 Operating Software
3.5 Installation
3.6 Start-up
3.7 Calibration
3.8 Status of Programming, Alarms, Calibration Level, RS-485 ID, and Sensor Life
3.9 Programming Alarms
3.10 Program Features
3.11 RS-485 Protocol
3.12 Display Contrast Adjust
3.13 Optical Sensor Replacement
3.14 Trouble Shooting Guide
3.15 Spare Parts List
3.16 Warranty
3.17 Service Policy
3.18 Software Flow Chart
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.2
3.0 DESCRIPTION
Detcon MicroSafe™ Model IR-640, carbon dioxide gas sensors are non-intrusive “Smart” sensors designed to detect and monitor CO2gas in air over the 0-5% range. The Detcon MicroSafe™ Model IR-642 CO2 gas sensor measures in the 10-100% Range. One of the primary features of the sensor is its method of automatic calibration which guides the user through each step via instructions displayed on the backlit LCD. The sensor features field adjustable, fully programmable alarms and provides relays for two alarms plus fault as standard. The sensor comes with two different outputs: analog 4-20 mA, and serial RS-485. These outputs allow for greater f lexibility in system integration and installation. The microprocessor supervised electronics are packaged as a plug-in module that mates to a standard connector board. Both are housed in an explosion proof condulet that includes a glass lens window which allows for the display of sensor readings as well as access to the sensor’s menu driven features via a hand-held programming magnet.
The sensor technology is a field proven “plug-in replaceable” non-dispersive infrared (NDIR) optical type. NDIR optical sensors show an excellent response to CO2. The NDIR type sensor is characteristically stable for both span and zero and is capable of providing reliable performance with low maintenance requirements for periods approach­ing 5 years in most industrial environments. Typical ranges of detection are 0-0.30%, 0-0.50%, 0-1.00%, 0-2.00%, 0-3.00% and 0-5% with the IR640. Typical Ranges with the IR642 are 0-10%, 0-25%, 0-50%, 0-100%. Other ranges may be accommodated and all ranges are covered by this manual. To determine the range of detection, refer to the instructions found in section 3.8.
3.0.1 Non-dispersive Infrared (NDIR) Optical Sensor
The Detcon NDIR sensor is designed as a miniature single piece “plug-in replaceable” component, which can easily be changed out in the field. The NDIR sensor consists of an infrared lamp source, two pyroelectric detectors, and an optical gas sample cavity. The lamp source produces infrared radiation which interacts with CO2as it travels through the optical gas sample cavity. The infrared radiation contacts each of two pyroelectric detectors at the completion of the optical path. The “active” pyroelectric detector is covered by a filter specific to the part of the IR spectrum where CO2absorbs light. The “reference” pyroelectric detector is covered by a filter specific to the non-absorbing part of the IR spectrum. When CO2is present, it absorbs IR radiation and the signal output from the “active” pyroelectric detec­tor decreases accordingly while the “reference” detector output remains unchanged. The ratio of the “active” and “ref­erence” detector outputs is then used to compute CO2concentration.
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.3
3.0.2 Microprocessor Control Circuit
NC
ALARM 1
WHT
BLK
YEL
BLU
MA
VDC Power In
Optional RS-485
Terminating Resistor
Use 120 ohm
NO
NC
NO
NC
NO
NO/NC
COM
NO/NC
COM
NO/NC
COM
FAULT ALM -2 ALM -1
Alarm Dry Contacts
ALARM 2
FAULT
R1
R2
A
B
A
B
4-20 mA Output
RS-485 In
RS-485 Out
Optional 4-20 mA Signal Developing Resistor Use 250 ohm 1/4w
JU
M
P
E
R
S
U
N
-U
S
E
D
RED
BRN
Jumper Programmable Alarm Outputs Normally Open or Normally Closed
Sensor
Place un-used alarm programming jumper tabs here
detcon inc.
Program Switch #2
F
LT
A
LM 1
C
AL
MicroSafe™ CO2 Gas Sensor
HOUS TON, TEX AS
PGM 2
PGM 1
A
LM 2
M
ODEL
I
R-640
CONTRAST
Alarm & Cal LEDs
Program Switch #1
Menu Driven Display
Plug-in Microprocessor Control Circuit
Display Contrast Adjust
Porous Membrane
Optical Filter
Active Detector
Lamp
Reference Detector
Optical Sample Gas Chamber
Optical Filter
CO
2
CO
2
The control circuit is microprocessor based and is packaged as a plug-in field replaceable module, facilitating easy replacement and minimum down time. Circuit functions include a basic sensor pre-amplifier, sensor temperature measurement, on-board power supplies, microprocessor, back lit alpha numeric display, alarm status LED indica­tors, magnetic programming switches, an RS-485 communication port, and a linear 4-20 mA DC output.
3.0.3 Base Connector Board
The base connector board is mounted in the explosion proof enclosure and includes: the mating connector for the control circuit, reverse input and secondary transient suppression, input filter, alarm relays, lugless terminals for all field wiring, and a terminal strip for storing unused programming jumper tabs. The alarm relays are contact rated 5 amps @ 250 VAC, 5 amps @ 30 VDC and coil rated at 24 VDC. Gold plated program jumpers are used to select either the normally open or normally closed relay contacts.
3.0.4 Explosion Proof Enclosure
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.4
The sensors are packaged in a cast metal explosion proof enclosure. The enclosure is fitted with a threaded cover
IR Lamp Source
Reference Detector Output
Active Detector Output
CO2 Not Present
CO2 Present
Principle of Operation
Sample
Gas
that has a glass lens window. Magnetic program switches located behind the transmitter module face plate are acti­vated through the lens window via a hand-held magnetic programming tool allowing non-intrusive operator inter­face with the sensor. All calibration and alarm level adjustments can be accomplished without removing the cover or declassifying the area. Electrical classification is Class I; Groups B, C, D; Div. 1.
3.1 PRINCIPLE OF OPERATION
The CO2diffuses through a sintered stainless steel f lame arrestor and into the volume of the sample gas optical cavity. A lamp source provides a cyclical IR radiation source which travels through the optical gas sample cavity and terminates at two pyroelectric detectors. The “active” and “reference” pyroelectric detectors each give an output which measures the intensity of the radiation contacting their surface. The “active” pyroelectric detector is covered by an optical filter specific to the part of the IR spectrum where CO2absorbs light. The “reference” pyroelectric detector is covered by a filter specific to the non-absorbing part of the IR spectrum. When present, CO2absorbs a fraction of the IR radiation and the signal output from the “active” pyroelectric detector decreases accordingly. The signal output of the “reference” detector remains unchanged in the presence of CO2. The ratio of the “active” and “reference” detector signal outputs is then used to compute CO2concentration. By using the ratio of the active/ref­erence signal outputs, measurement drift caused by changes in the intensity of the lamp source and changes in the optical path are negated.
3.1.2 Characteristics
The NDIR optical sensor element maintains strong sensitivity to CO2, as shown in the illustration below. Typical zero cali­bration intervals would be quarterly and typical span calibration intervals would be semi-annual to annual. However, actual field experience is always the best determination of appropriate calibration intervals.
The MicroSafe™ IR-640/IR-642 is factory calibrated for a specific CO2 range (see section 3.8 to determine range). The detec­tor will be factory calibrated for the range specified at time of order. If a CO2range other than that specified at time of order is desired, contact Detcon for the appropriate field-replaceable electronics/software upgrade.
The NDIR optical sensor readings can be adversely affected by dust, dirt and oil mist accumulation as well as severe corro­sion. These deposits may reduce the optical reflectivity inside the sensor, and although accurate readings are continually main­tained, excessive loss in useable signal eventually gives way to noise and unstable readings. The optical sensor may, over long periods of time (3-7 years), lose its IR lamp source filament, and in this case an optical sensor modular replacement is required. The IR-640/IR-642 has an extensive list of Fault Diagnostics to alert and pin-point operational problems. See section
3.10
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.5
3.2 APPLICATION
Functional
Block
Diagram
Functional
Block
Diagram
Analog 4-20 mA Out
Power In
Relays Out
Pre-Amp Display
Tem per atu re
Compensation
Alarm & Fault
Relays
RS-485 & 4-20mA
Micro-
processor
Tran smit ter
Power Supply
Optical Sensor
I/O Circuit
Protection
Serial RS-485 Out
0
4
8
12
16
20
2
0406080100
% of Range of Detection
mA DC Signal Output
Output Response Curve
Model IR-640/IR-642 MicroSafe™ sensors are designed to detect and monitor CO2gas in ambient air. Minimum sen­sitivity and scale resolution is 0.01%. Operating temperature range is -40° F. to +175° F. While the sensor is capable of operating outside these temperatures, performance specifications are verified within the limit.
3.2.1 Sensor Placement/Mounting
Sensor location should be reviewed by facility engineering and safety personnel.
3.3 SPECIFICATIONS
Method of Detection
NDIR Optical
Electrical Classification
Class I; Groups B, C, D; Div. 1.
Response Time
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.6
T50 < 15 seconds; T90 < 35 seconds
Clearing Time
0% < 35 seconds
9
Repeatability
± 3% FS
Range
0 - 0.3/0.5/1/2/3/5% withthe IR-640 0-10/0-25/0-50/0-100% with the IR-642
Operating Temperature
-40° to +175° F
Accuracy
± 5% FS
Sensor Warranty
5 year pro-rated
Power Consumption
Normal operation = 95 mA (2.3 watts); Full alarm = 123 mA (<3 watts)
Zero Drift
< 0.05% per year
Output
3 relays (alarm 1, alarm 2, and fault) contact rated 5 amps @ 250 VAC, 5 amps @ 30 VDC; Linear 4-20 mA DC; RS-485 Modbus™
Input Voltage
22.5-28 VDC
3.4 OPERATING SOFTWARE
Operating software is menu listed with operator interface via the two magnetic program switches located under the face plate. The two switches are referred to as “PGM 1” and “PGM 2”. The menu list consists of 3 items which include sub-menus as indicated below. (Note: see the last page of this manual for a complete software flow chart.)
01. Normal Operation
a) Current Status
02. Calibration Mode
a) Zero b) Span
03. Program Menu
a) Program Status b) Alarm 1 Level c) Alarm 2 Level d) Calibration Level
3.4.1 Normal Operation
In normal operation, the display tracks the current status of the sensor and gas concentration and appears as: “0.00 % CO2”. The mA current output corresponds to the monitoring level and range of 0-100% FS = 4-20 mA.
3.4.2 Calibration Mode
Calibration mode allows for sensor zero and span adjustments. “1-ZERO 2-SPAN”
3.4.2.1 Zero Adjustment
Zero is set in ambient air with no CO2gas present or with zero gas applied to the sensor. “AUTO ZERO”
3.4.2.2 Span Adjustment
Unless otherwise specified, span adjustment is performed at 50% of full-scale range of CO2. “AUTO SPAN”
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.7
3.4.3 Program Mode
The program mode provides a program status menu, allows for the adjustment of alarm set point levels, and the selection of the calibration gas level setting..
3.4.3.1 Program Status
The program status scrolls through a menu that displays: * The gas type, range of detection and software version number. The menu item appears as: “CO2 RANGE V6.0” * The alarm set point level of alarm 1. The menu item appears as: “ALM1 SET @ ##%” * The alarm firing direction of alarm 1. The menu item appears as: “ALM1 ASCENDING” or descending. * The alarm relay latch mode of alarm 1. The menu item appears as: “ALM1 NONLATCHING” or latching. * The alarm relay energize state of alarm 1. The menu item appears as: “ALM1 DE-ENERGIZED” or energized. * The alarm set point level of alarm 2. The menu item appears as: “ALM2 SET @ ##%” * The alarm firing direction of alarm 2. The menu item appears as: “ALM2 ASCENDING” or descending. * The alarm relay latch mode of alarm 2. The menu item appears as: “ALM2 LATCHING” or nonlatching. * The alarm relay energize state of alarm 2. The menu item appears as: “ALM2 DE-ENERGIZED” or energized. * The alarm relay latch mode of the fault alarm. The menu item appears as: “FLT NONLATCHING” or latching. * The alarm relay energize state of the fault alarm. The menu item appears as: “FLT ENERGIZED” or deenergized. * The calibration gas level setting. The menu item appears as: “CalLevel @ #.##%” * Identification of the RS-485 ID number setting. The menu item appears as: “485 ID SET @ ##” * Sensor temperature in °C appears as: “TEMPERATURE ## °C” * The estimated remaining sensor life. The menu item appears as: “OPTICS AT 100%”
3.4.3.2 Alarm 1 Level Adjustment
The alarm 1 level is adjustable over the range 10 to 90% of full-scale range. For CO2gas sensors, the level is factory set at 20% of full-scale range. The menu item appears as: “SET ALM1 @ #.##%”
3.4.3.3 Alarm 2 Level Adjustment
The alarm 2 level is also adjustable over the range 10 to 90% of full-scale range. For CO2 gas sensors, the level is factory set at 60% of full-scale range. The menu item appears as: “SET ALM2 @ #.##%”
3.4.3.4 Calibration Level Adjustment
The Calibration level is adjustable from 10% to 90% of full-scale range. The menu item appears as: “CalLevel @ #.##%”
3.5 INSTALLATION
Optimum performance of ambient air/gas sensor devices is directly relative to proper location and installation practice.
3.5.1 Field Wiring Table
Detcon Model IR-640/IR-642 CO2 sensor assemblies require three conductor connection between power supplies and host electronic controllers. Wiring designators are conductor resistance between sensor and controller is 10 ohms. Maximum wire size for termination in the sensor assembly terminal board is 14 gauge.
AWG Meters Feet
20 240 800 18 360 1200 16 600 2000 14 900 3000
Note 1:
Note 2: Shielded cable may be required in installations where cable trays or conduit runs include high voltage
lines or other sources of induced interference.
This wiring table is based on stranded tinned copper wire and is designed to serve as a reference only.
(4-20 mA output)
+
(DC), –(DC) , and mA(sensor signal). Maximum single
Note 3: The supply of power must be from an isolating source with over-current protection as follows:
AWG
Over-current Protection AWG Over-current Protection
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.8
22 3A 16 10A
EYS
Seal
Fitting
Drain
“T”
Plug any unused ports.
20 5A 14 20A 18 7A 12 25A
The RS-485 (if applicable) requires 24 gauge, two conductor, shielded, twisted pair cable between sensor and host PC. Use Belden part number 9841. Two sets of terminals are located on the connector board to facilitate serial loop wiring from sensor to sensor. Wiring designators are
3.5.2 Sensor Location
Selection of sensor location is critical to the overall safe performance of the product. Five factors play an important role in selection of sensor locations:
(1) Density of the gas to be detected (2) Most probable leak or generation sources within the industrial process (3) Ventilation or prevailing wind conditions (4) Personnel exposure (5) Accessibility for routine maintenance
A& B
(IN) and A& B(OUT).
Density
be located within 2-4 feet of grade as CO2will tend to settle in low lying areas.
Leak Sources
tions of the sealed type where seals may either fail or wear. Other leak sources are best determined by facility engi­neers with experience in similar processes.
Ventilation
manner where the migration of gas clouds is quickly detected.
Personnel Exposure
sonnel areas such as control rooms, maintenance or warehouse buildings. A more general and applicable thought toward selecting sensor location is combining leak source and perimeter protection in the best possible configuration.
Note:
- Placement of sensors relative to the density of CO2is such that sensors for the detection of CO2should
- Most probable leak sources within an industrial process include f langes, valves, and tubing connec-
- Normal ventilation or prevailing wind conditions can dictate efficient location of gas sensors in a
- The undetected migration of gas clouds should not be allowed to approach concentrated per-
In all installations, the sensor element in SS housing points down relative to grade (Fig. 1). Improper sensor
Figure #1
orientation may result in false reading and permanent sensor damage.
3.5.3 Local Electrical Codes
Sensor and transmitter assemblies should be installed in accordance with all local electrical codes. Use appropriate conduit seals. Drains & breathers are recommended. The sensor assemblies are designed to meet NEC and CSA requirements for Class I; Groups B, C, D; Div. 1 environments.
3.5.4 Accessibility
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.9
Consideration should be given to easy access by maintenance personnel as well as the consequences of close prox-
NC
ALARM 1
WHT
BLK
YEL
BLU
MA
VDC Power In
Optional RS-485
Terminating Resistor
Use 120 ohm
NO
NC
NO
NC
NO
NO/NC
COM
NO/NC
COM
NO/NC
COM
FAULT AL M- 2 A LM -1
Alarm Dry Contacts
ALARM 2
FAULT
R1
R2
A
B
A
B
4-20 mA Output
RS-485 In
RS-485 Out
Optional 4-20 mA Signal Developing Resistor Use 250 ohm 1/4w
JUMPERS
UN-USED
RED
BRN
Jumper Programmable Alarm Outputs Normally Open or Normally Closed
Sensor
Place un-used alarm programming jumper tabs here
4 3/4"
3/4" NPT
1/4" Dia.
M
ounting Holes
8 1/4"
6 1/8" 5
1/2"
3/4" NPT
Rainshield/
Splashguard
2"
2 1/8"
imity to contaminants that may foul the sensor prematurely.
Note:
An appropriate conduit seal must be located within 18" of the sensor assembly. Crouse Hinds type EYS2,
EYD2 or equivalent are suitable for this purpose.
3.5.5 Installation Procedure
Note: See section 3.5.6 for special information on remote mounting applications in which the sensor (Model IR­640-RS) is remotely mounted away from the transmitter (Model IR-640-RT).
Figure #2
a) Remove the junction box cover and un-plug the control circuit by grasping the two thumb screws and pulling outward. b) Securely mount the sensor junction box in accordance with recommended practice. See dimensional drawing (Fig. 2). c) Observing correct polarity, terminate 3 conductor field wiring, RS-485 wiring, and applicable alarm wiring to the
sensor base connector board in accordance with the detail shown in Figure 3. Normally open and normally closed Form C dry contacts (rated 5 amp @ 120VAC; 5 amp @ 30VDC) are provided for Fault, Alarm 1, and Alarm 2.
d) Position gold plated jumper tabs located on the connector board in accordance with desired Form C dry con-
Figure #3
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.10
tact outputs: NO = Normally Open; NC = Normally closed (see figure 3).
Preamp Board - Side View
RS-485 ID Set Dip Switches
Control Circuit - Side View
Preamp Board
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
SW2SW1
FAULT
ALARM 1
Latch
Energize
Latch
Ascending
Energize
ALARM 2
Latch
Ascending
Energize
CPU Board - Top View
Alarm Programming Jumpers
Control Circuit - Side View
CPU Board
Note
: If a voltage signal output is desired in place of the 4-20mA output, a 1/4 watt resistor must be installed in position R2 of the terminal board. A 250Ω resistor will provide a 1-5V output (– to mA). A 100Ω resistor will provide a .4-2V output, etc. This linear signal corresponds to 0-100% of full-scale (see figure 3).
e) Program the alarms via the gold plated jumper tab positions located on the CPU board (see figure 4). Alarm 1
and Alarm 2 have three jumper programmable functions: latching/non-latching relays, normally energized/nor­mally de-energized relays, and ascending/descending alarm set points. The fault alarm has two jumper program­mable functions: latching/non-latching relay, and normally energized/normally de-energized relay. The default settings of the alarms (jumpers removed) are normally de-energized relays, non-latching relays, and alarm points that activate during descending gas conditions.
If a jumper tab is installed in the latch position, that alarm relay will be in the latching mode. The latching mode will latch the alarm after alarm conditions have cleared until the alarm reset function is activated. The non­latching mode (jumper removed) will allow alarms to de-activate automatically once alarm conditions have cleared.
If a jumper tab is installed in the energize position, that alarm relay will be in the energized mode. The ener­gized mode will energize or activate the alarm relay when there is no alarm condition and de-energize or de-acti­vate the alarm relay when there is an alarm condition. The de-energized mode (jumper removed) will energize or activate the alarm relay during an alarm condition and de-energize or de-activate the alarm relay when there is no alarm condition.
If a jumper tab is installed in the ascending position, that alarm relay will be in the ascending mode. The
Figure #4
ascending mode will cause an alarm to fire when the gas concentration detected is greater than or equal to the alarm set point. The descending mode (jumper removed) will cause an alarm to fire when the gas concentration detected is lesser than or equal to the alarm set point. Except in special applications, CO2 monitoring will require alarms to fire in
“ASCENDING”
Any unused jumper tabs should be stored on the connector board on the terminal strip labeled “Unused Jumpers” (see figure 3).
f) If applicable, set the RS-485 ID number via the two rotary dip switches located on the preamp board (see figure
5). There are 256 different ID numbers available which are based on the hexadecimal numbering system. If RS­485 communications are used, each sensor must have its own unique ID number. Use a jewelers screwdriver to set the rotary dip switches according to the hexadecimal table listed below. If RS-485 communications are not used, leave the dip switches in the default position which is zero/zero (0)-(0).
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.11
gas conditions.
Figure #5
Hexadecimal Table
ID# SW1 SW2
none 00
101 202 303 404 505 606 707 808
909 10 0A 11 0B 12 0C 13 0D 14 0E 15 0F 16 10 17 11 18 12 19 13 20 14 21 15 22 16 23 17 24 18 25 19 26 1A 27 1B 28 1C 29 1D 30 1E 31 1F 32 20 33 21 34 22 35 23 36 24 37 25 38 26 39 27 40 28 41 29 42 2A
ID#
SW1 SW2
43 2B 44 2C 45 2D 46 2E 47 2F 48 30 49 31 50 32 51 33 52 34 53 35 54 36 55 37 56 38 57 39 58 3A 59 3B 60 3C 61 3D 62 3E 63 3F 64 40 65 41 66 42 67 43 68 44 69 45 70 46 71 47 72 48 73 49 74 4A 75 4B 76 4C 77 4D 78 4E 79 4F 80 50 81 51 82 52 83 53 84 54 85 55
ID#
SW1 SW2
86 56 87 57 88 58 89 59 90 5A 91 5B 92 5C 93 5D 94 5E 95 5F 96 60 97 61 98 62
99 63 100 64 101 65 102 66 103 67 104 68 105 69 106 6A 107 6B 108 6C 109 6D 110 6E 111 6F 112 70 113 71 114 72 115 73 116 74 117 75 118 76 119 77 120 78 121 79 122 7A 123 7B 124 7C 125 7D 126 7E 127 7F 128 80
ID#
SW1 SW2
129 81 130 82 131 83 132 84 133 85 134 86 135 87 136 88 137 89 138 8A 139 8B 140 8C 141 8D 142 8E 143 8F 144 90 145 91 146 92 147 93 148 94 149 95 150 96 151 97 152 98 153 99 154 9A 155 9B 156 9C 157 9D 158 9E 159 9F 160 A0 161 A1 162 A2 163 A3 164 A4 165 A5 166 A6 167 A7 168 A8 169 A9 170 AA 171 AB
ID#
SW1 SW2
172 AC 173 AD 174 AE 175 AF 176 B0 177 B1 178 B2 179 B3 180 B4 181 B5 182 B6 183 B7 184 B8 185 B9 186 BA 187 BB 188 BC 189 BD 190 BE 191 BF 192 C0 193 C1 194 C2 195 C3 196 C4 197 C5 198 C6 199 C7 200 C8 201 C9 202 CA 203 CB 204 CC 205 CD 206 CE 207 CF 208 D0 209 D1 210 D2 211 D3 212 D4 213 D5 214 D6
ID#
SW1 SW2
215 D7 216 D8 217 D9 218 DA 219 DB 220 DC 221 DD 222 DE 223 EF 224 E0 225 E1 226 E2 227 E3 228 E4 229 E5 230 E6 231 E7 232 E8 233 E9 234 EA 235 EB 236 EC 237 ED 238 EE 239 FF 240 F0 241 F1 242 F2 243 F3 244 F4 245 F5 246 F6 247 F7 248 F8 249 F9 250 FA 251 FB 252 FC 253 FD 254 FE 255 FF
g) Replace the plug-in control circuit and replace the junction box cover.
3.5.6 Remote Mounting Applications
Some sensor mounting applications require that the gas sensor head be remotely mounted away from the sensor transmitter. This is usually true in instances where the gas sensor head must be mounted in a location that is diffi­cult to access. Such a location creates problems for maintenance and calibration activities. Detcon provides the IR­640/IR-642 sensor in a remote-mount configuration in which the sensor (Model IR-640-RS) and the transmitter (Model IR-640-RT) are provided in their own condulet housing and are interfaced together with a six conductor cable. Shielded cable is required and must be installed in its own (unshared) conduit (use Alpha Wire Company #6342 cable). A maximum of 100 feet of separation is allowed. Reference figure 5A for wiring diagram.
3.6 START UP
Upon completion of all mechanical mounting and termination of all field wiring, apply system power and observe the following normal conditions:
a) IR-640/IR-642 “Fault” LED is off. b) A reading of 0.00% CO2 should be indicated upon conclusion of a 12 second “warming up” cycle.
NOTE: All alarms will be disabled for 1 minute after power up. In the event of power failure, the alarm disable peri-
od will begin again once power has been restored.
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.12
Magnetic Programming Tool
1234
RED
BRN
WHT
BLK
Remote Transmitter
IR-640-RT
Remote Sensor
IR-640-RS
WHT
BLK
YEL
BLU
RED
BRN
56
YEL
BLU
Figure #5A
Material Requirements
* Detcon PN 3270 MicroSafe™ Programming Magnet * Digital volt/ohm meter.
3.6.2 Programming Magnet Operating Instructions
Operator interface to MicroSafe™ gas detection products is via magnetic switches located behind the transmitter face plate. DO NOT remove the glass lens cover to calibrate or change programming parameters. Two switches labeled “PGM 1” and “PGM 2” allow for complete calibration and alarm level programming without removing the enclosure cover, thereby eliminating the need for area de-classification or the use of hot permits.
Figure #6
A magnetic programming tool (see figure 6) is used to operate the switches. Switch action is defined as momentary contact, 3 second hold, and 30 second hold. In momentary contact use, the programming magnet is waved over a switch location. In 3 second hold, the programming magnet is held in place over a switch location for 3 or more seconds. In 30 second hold, the programming magnet is held in place over a switch location for 30 or more sec­onds. Three and thirty second hold is used to enter or exit calibration and program menus while momentary con­tact is used to make adjustments. The location of “PGM 1” and “PGM 2” are shown in figure 7.
NOTE: If, after entering the calibration or program menus, there is no interaction with the menu items for more
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.13
detcon inc.
Program Switch #2
F
LT
ALM
1
C
AL
M
icroSafe™ CO
2
Gas Sensor
HOUS TON , TEX AS
P
GM 2
PGM 1
ALM
2
MODEL IR-640
CONTRAST
Alarm & Cal LEDs
P
rogram Switch #1
Menu Driven Display
Plug-in Microprocessor Control Circuit
Display Contrast Adjust
Figure #7
than 30 seconds, the sensor will return to its normal operating condition.
3.6.3 Initial Operational Tests
After a warm up period has been allowed for, the sensor should be checked to verify sensitivity to carbon dioxide gas.
Material Requirements
* Detcon PN 6132 Threaded Calibration Adapter * Span Gas 50% of full-scale range CO2in air at a controlled flow rate of 200ml/min.
a) Attach the calibration adapter to the threaded sensor housing. Apply the test gas at a controlled f low rate of
200 ml/m. Observe that the LCD display increases to a level of 20% of full-scale range or higher.
b) Remove the test gas and observe that the LCD display decreases to “0.00 % CO2” c) If alarms are activated during the test, and have been programmed for latching operation, reset them according to
the instructions in section 3.9.2.
Initial operational tests are complete. Detcon CO2 sensors are pre-calibrated prior to shipment and will, in most cases, not require significant adjustment on start up. However, it is recommended that a complete calibration test and adjustment be performed within 24 hours of installation. Refer to calibration instructions in later text.
.
3.7 CALIBRATION
Material Requirements
* Detcon PN 3270 MicroSafe™ Programming Magnet * Detcon PN 6132 Threaded Calibration Adapter * Span Gas containing the applicable calibration gas in air. Span gas concentration is recommended at 50% of
full-scale range (which is the factory default) at a controlled f low rate of 200 ml/min. Other concentrations can be used as long as they fall within 10% to 90% of range. See section 3.7.2 for details.
3.7.1 Calibration Procedure - Zero
NOTE: Before performing a zero calibration, be sure there is no CO2gas present.
a) Enter the calibration menu by holding the programming magnet stationary over “PGM 1” (see figure 7) for 3 sec-
b) Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display
onds until the display reads “1-ZERO 2-SPAN”, then withdraw the magnet. Note that the “CAL” LED is on.
reads: “ZERO 0%”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is complete the display will read “ZERO COMPLETE” for 5 seconds and then return to the normal operations menu, 0.00 % CO2”. NOTE 1: If the circuitry is unable to adjust the zero to the proper setting the sensor will enter a calibration
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.14
fault mode which will cause the display to alternate between the sensor’s current status reading and the calibra­tion fault screen which appears as: “CAL FAULT” (see section 3.7.3).
NOTE 2: When a “cal fault” occurs, the sensor microprocessor retains its previous calibration references. Zero calibration is complete.
3.7.2 Calibration Procedure - Span
CAUTION:
Verification of the correct calibration gas level setting and calibration span gas concentration is
required before “span” calibration. These two numbers must be equal.
Calibration consists of entering the calibration function and following the menu-displayed instructions. The display will ask for the application of span gas in a specific concentration. This concentration is equal to the calibration gas level setting. The factory default setting for span gas concentration is 50% of full-scale range. In this instance, a span gas containing a concentration equal to 50% of full-scale range is required. If a span gas containing 50% of range is not available, other concentrations may be used as long as they fall within 10% to 90% of range. However, any alternate span gas concentration value must be programmed via the calibration gas level menu before proceed­ing with span calibration. Follow the instructions below for span calibration. a) Verify the current calibration gas level setting as indicated by the programming status menu. To do this, follow
the instructions in section 3.8 and make note of the setting found in listing number 14. The item appears as
“CalLevel @ x.xx%”.
b) If the calibration gas level setting is equal to your calibration span gas concentration, proceed to item “f”. If
not, adjust the calibration gas level setting so that it is equal to your calibration span gas concentration, as instructed in items “c” through “e”.
c) Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30 seconds
until the display reads “VIEW PROG STATUS”, then withdraw the magnet. At this point you can scroll through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu options are: View Program Status, Set Alarm 1 Level, Set Alarm 2 Level, Set Gas Factor, Set Cal Factor, and Set Cal Level.
d) From the programming menu scroll to the calibration level listing. The menu item appears as: “SET CAL
LEVEL”. Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “CalLevel @ ##%”, then withdraw the magnet. Use the programming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired calibration span gas concentration. Exit to the programming menu by holding the programming magnet over “PGM1” for 3 seconds.
e) Exit back to normal operation by holding the programming magnet over “PGM 2” for 3 seconds, or automati-
cally return to normal operation in 30 seconds.
f) From the calibration menu “1-ZERO 2-SPAN” (section 3.7.1-a) proceed into the span adjust function by
holding the programming magnet stationary over “PGM 2” for 3 seconds until the display reads “APPLY x.xx% CO2” then withdraw the programming magnet. The x’s here indication the gas concentration requested.
g) Apply the calibration gas at a flow rate of 200 milliliters per minute. As the sensor signal changes, the display
will change to “SPAN XX%”. The “XX” part of the reading indicates the actual gas reading which will increase until the sensor stabilizes. When the sensor signal is stable it will auto span to the request concentration and the display will change to “SPAN COMPLETE” for two seconds and then “REMOVE GAS”. Remove the gas. When the signal level has fallen below 10% of full scale, the display will return to the normal operation menu, 0.00 % CO2”.
NOTE 1: If the circuitry is unable to adjust the span to the proper setting the sensor will enter into the calibra­tion fault mode which will cause the display to alternate between the sensor’s current status reading and the cal­ibration fault screen which appears as: “CAL FAULT” (see section 3.7.3).
NOTE 2: If, after entering the span function, more than one minute elapses before calibration gas is applied, the sensor will enter the calibration fault mode which will cause the display to alternate between the sensor’s current status reading and the calibration fault screen which appears as: “CAL FAULT” (see section 3.7.3).
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.15
Span calibration is complete.
3.7.3
1. Upon entering the calibration menu, the 4-20 mA signal drops to 2 mA and is held at this level until you
2. If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will
3.7.4 Calibration Frequency
In most applications, quarterly zero calibration and annual span calibration intervals will assure reliable detection. However, industrial environments differ. Upon initial installation and commissioning, close frequency tests should be performed monthly. Test results should be recorded and reviewed to determine a suitable calibration interval.
Additional Notes
return to normal operation.
enter into the calibration fault mode which will activate fault alarm functions (see section 3.10) and cause the dis­play to alternate between the sensor’s current status reading and the calibration fault screen which appears as: “CAL FAULT”. If this occurs you may attempt to recalibrate by entering the calibration menu as described in section 3.8.1-a. If the sensor fails again, defer to technical trouble shooting.
3.8 STATUS OF PROGRAMMING, ALARMS, CALIBRATION LEVEL, RS-485 ID, AND SENSOR LIFE
The programming menu has a programming status listing that allows the operator to view the gas, range, and soft­ware version number of the program, as well as the current alarm settings, calibration gas level setting, RS-485 ID number, sensor temperature, and estimated remaining sensor life. The programming menu also allows the calibra­tion gas level setting (see section 3.72), and alarm levels (see section 3.9).
The following procedure is used to view the programming status of the sensor:
a) First, enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30 sec-
onds until the display reads “VIEW PROG STATUS”, then withdraw the magnet. At this point you can scroll through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu options are: View Program Status, Set Alarm 1 Level, Set Alarm 2 Level, and Set Cal Level.
b) Next, scroll to the “VIEW PROG STATUS” listing and then hold the programming magnet over “PGM 1”
for 3 seconds. The menu will then automatically scroll, at five second intervals, through the following informa­tion before returning back to the “VIEW PROG STATUS” listing.
The gas type, range of detection and software version number. The menu item appears as:
1 ­2 - The alarm set point level of alarm 1. The menu item appears as: “ALM1 SET @ x.xx%” 3 - The alarm firing direction of alarm 1. The menu item appears as: “ALM1 ASCENDING” 4 - The alarm relay latch mode of alarm 1. The menu item appears as: “ALM1 NONLATCHING” 5 - The alarm relay energize state of alarm 1. The menu item appears as: “ALM1 DE-ENERGIZED” 6 - The alarm set point level of alarm 2. The menu item appears as: “ALM2 SET @ x.xx%” 7 - The alarm firing direction of alarm 2. The menu item appears as: “ALM2 ASCENDING” 8 - The alarm relay latch mode of alarm 2. The menu item appears as: “ALM2 LATCHING” 9 - The alarm relay energize state of alarm 2. The menu item appears as: “ALM2 DE-ENERGIZED” 10 - The alarm relay latch mode of the fault alarm. The menu item appears as: “FLT NONLATCHING” 11 - The alarm relay energize state of the fault alarm. The menu item appears as: “FLT ENERGIZED” 12 - The calibration gas level setting. The menu item appears as: “CalLevel @ xx%” 13 - Identification of the RS-485 ID number setting. The menu item appears as: “485 ID SET @ 1” 14 - Sensor temperature in °C appears as: “TEMPERATURE xx °C” 15 - The estimated remaining sensor life. The menu item appears as: “OPTICS AT 100%”
c) Exit back to normal operations by holding the programming magnet over “PGM 2” for 3 seconds, or automati-
cally return to normal operation in 30 seconds.
3.9 PROGRAMMING ALARMS
“CO2 RANGE V6.0”
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.16
3.9.1 Alarm Levels
Both alarm 1 and alarm 2 levels are factory set prior to shipment. Alarm 1 is set at 20% of full-scale range; alarm 2 at 60% of full-scale range. Both alarms can be set in 0.01% increments from 10 to 90% of full-scale range. The fol­lowing procedure is used to change alarm set points:
a) First, enter the programming menu by holding the programming magnet stationary over “PGM 2” for 30 sec-
onds until the display reads “VIEW PROG STATUS”, then withdraw the magnet. At this point you can scroll through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”. The menu options are: View Program Status, Set Alarm 1 Level, Set Alarm 2 Level, and Set Cal Level.
b) ALARM 1 LEVEL
“SET ALARM 1 LEVEL”. Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “SET ALM1 @ x.xx%”, then withdraw the magnet. Use the program­ming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired alarm set point. Exit to the programming menu by holding the programming magnet over “PGM1” for 3 seconds, or automatically return to the programming menu in 30 seconds.
c) ALARM 2 LEVEL
“SET ALARM 2 LEVEL”. Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “SET ALM2 @ x.xx%”, then withdraw the magnet. Use the program­ming magnet to make an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to the desired alarm set point. Exit to the programming menu by holding the programming magnet over “PGM1” for 3 seconds, or automatically return to the programming menu in 30 seconds.
d) Exit back to normal operations by holding the programming magnet over “PGM 2” for 3 seconds, or automati-
cally return to normal operation in 30 seconds.
3.9.2 Alarm Reset
An alarm condition will cause the applicable alarm to activate its corresponding relay and LED. If alarm 1, alarm 2, or fault alarms have been programmed for latching relays, an alarm reset function must be activated to reset the alarms after an alarm condition has cleared. To reset the alarms, simply wave the programming magnet over either “PGM 1” or “PGM 2”, momentarily, while in normal operations mode and note that the corresponding alarm LED(s) turn off.
3.9.3 Other Alarm Functions
Alarms are factory programmed to be non-latching, de-energized; and to fire under ascending gas conditions. The fault alarm relay is programmed as normally energized which is useful for detecting a 24VDC power source failure. All alarm functions are programmable via jumper tabs. Changing alarm functions requires the sensor housing to be opened, thus declassification of the area is required. See section 3.5.5-e for details.
From the programming menu scroll to the alarm 1 level listing. The menu item appears as:
From the programming menu scroll to the alarm 2 level listing. The menu item appears as:
3.10 PROGRAM FEATURES
Model IR-640 MicroSafe™ Sensors incorporate a comprehensive program to accommodate easy operator interface and fail-safe operation. Program features are detailed in this section. Each sensor is factory tested, programmed, and calibrated prior to shipment.
Over Range
When the sensor detects gas greater than 100% of the full-scale range, it will cause the display to flash “OVER­RANGE” on and off.
Optics Life
The Optics Life feature is a reference based on the signal output from the optical sensor. When an optical sensor life of 25% or less remains, the optical sensor should be replaced within a reasonable maintenance schedule.
Calibration Fault
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.17
If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading and the calibration fault screen which appears as: “CAL FAULT.2”.
The following conditions will cause a calibration fault: 1 - Zero calibration cannot converge. 2 - Auto span cannot converge (too noisy or too unstable). 3 - Span gas is not applied before 1 minute elapses.
Fail-Safe/Fault Supervision
Model IR-640/IR-642 MicroSafe™ sensors are programmed for fail-safe operation. All of the fault conditions listed below will activate the fault relay, illuminate the fault LED, cause the mA output to drop to zero (0) mA, and cause the display to read its corresponding fault condition.
Memory Error
If the processor can’t save values to memory, the display will indicate: “MEMORY ERROR”.
Zero Fault
If the sensor should drift below –0.02%, the display will indicate: “ZERO FAULT”.
Lamp Fault
If the lamp signal is lost, the display will indicate: “LAMP FAULT.2”.
Reference Peak High Fault
If the reference peak signal is too high (>3600), the display will indicate: “SIGNAL FAULT.31”.
Active Peak High Fault
If the active peak signal is too high (>3600), the display will indicate: “SIGNAL FAULT.32”.
Reference Peak Low Fault
If the reference peak signal is too low (<500), the display will indicate: “SIGNAL FAULT.41”.
Active Peak Low Fault
If the active peak signal is too low (<500), the display will indicate: “SIGNAL FAULT.42”.
Reference Peak to Peak Low Fault
If the reference peak to peak signal is too low (<200), the display will indicate: “SIGNAL FAULT.51”.
Active Peak to Peak Low Fault
If the active peak to peak signal is too low (<200), the display will indicate: “SIGNAL FAULT.52”.
3.11 RS-485 PROTOCOL
Detcon MicroSafe™ sensors feature Modbus™ compatible communications protocol and are addressable via rotary dip switches for multi-point communications. Other protocols are available. Contact the Detcon factory for specific protocol requirements. Communication is two wire, half duplex 485, 9600 baud, 8 data bits, 1 stop bit, no parity, with the sensor set up as a slave device. A master controller up to 4000 feet away can theoretically poll up to 256 different sensors. This number may not be realistic in harsh environments where noise and/or wiring conditions would make it impractical to place so many devices on the same pair of wires. If a multi-point system is being uti­lized, each sensor should be set for a different address. Typical address settings are: 01, 02, 03, 04, 05, 06, 07, 08, 09, 0A, 0B, 0C, 0D, 0E, 0F, 10, 11, etc.
In most instances, RS-485 ID numbers are factory set or set during installation before commissioning. If required, the RS-485 ID number can be set via rotary dip switches located on the preamp circuit board. However, any change
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.18
to the RS-485 ID number would require the sensor housing to be opened, thus declassification of the area would be required. See section 3.5.4-f for details on changing the RS-485 ID number.
The following section explains the details of the Modbus™ protocol that the IR-622 MicroSafe™ sensor supports.
Code 03 - Read Holding Registers, is the only code supported by the transmitter. Each transmitter contains 6 hold­ing registers which reflect its current status.
Register # High Byte Low Byte 40000 Gas type Sensor Life
Gas type is one of the following: 01=CO, 02=H2S, 03=SO2, 04=H2, 05=HCN, 06=CL2, 07=NO2, 08=NO, 09=HCL, 10=NH3, 11=LEL, 12=O2
Sensor life is an estimated remaining use of the sensor head, between 0% and 100% Example: 85=85% sensor life
Register #
High Byte Low Byte
40001 Detectable Range
i.e. 100 for 0-100 ppm, 50 for 0-50% LEL, etc.
Register #
High Byte Low Byte
40002 Current Gas Reading
The current gas reading as a whole number. If the reading is displayed as 23.5 on the display, this register would contain the number 235.
Register #
High Byte Low Byte 40003 Alarm 1 Setpoint This is the trip point for the first alarm.
Register # High Byte Low Byte 40004 Alarm 2 Setpoint
This is the trip point for the second alarm.
Register #
High Byte Low Byte 40005 Status Bits Status Bits
High Byte Bit 7 Not used, always 0 Bit 6 Not used, always 0 Bit 5 Not used, always 0 Bit 4 Not used, always 0 Bit 3 1-Unit is in calibration 0-Normal operation Bit 2 1-Alarm 2 is ascending 0-Alarm 2 is descending Bit 1 1-Alarm 2 is normally energized 0-Alarm 2 is normally de-energized Bit 0 1-Alarm 2 is latching 0-Alarm 2 is non-latching
Low Byte Bit 7 1-Alarm 2 Relay is energized 0-Alarm 2 Relay is not energized
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.19
Bit 6 1-Alarm 1 is ascending 0-Alarm 1 is descending Bit 5 1-Alarm 1 is normally energized 0-Alarm 1 is normally de-energized Bit 4 1-Alarm 1 is latching 0-Alarm 1 is non-latching Bit 3 1-Alarm 1 Relay is energized 0-Alarm 1 Relay is not energized Bit 2 1-Fault is normally energized 0-Fault is normally de-energized Bit 1 1-Fault is latching 0-Fault is non-latching Bit 0 1-Fault Relay is energized 0-Fault Relay is not energized
The following is a typical Master Query for device # 8:
Field
Name HEX DEC RTU Slave Address 08 8 0000 1000 Function 03 3 0000 0011 Start Address Hi 00 0 0000 0000 Start Address Lo 00 0 0000 0000 No. of Registers Hi 00 0 0000 0000 No. of Registers Lo 06 6 0000 0110 CRC ## #### #### CRC ## #### ####
The following is a typical Slave Response from device # 8:
Field
Name HEX DEC RTU Slave Address 08 8 0000 1000 Function 03 3 0000 0011 Byte Count 0C 12 0000 1100 Reg40000 Data Hi 02 2 0000 0010 Reg40000 Data Lo 64 100 0110 0100 Reg40001 Data Hi 00 0 0000 0000 Reg40001 Data Lo 64 100 0110 0100 Reg40002 Data Hi 00 0 0000 0000 Reg40002 Data Lo 07 7 0000 0111 Reg40003 Data Hi 00 0 0000 0000 Reg40003 Data Lo 0A 10 0000 1010 Reg40004 Data Hi 00 0 0000 0000 Reg40004 Data Lo 14 20 0001 0100 Reg40005 Data Hi 05 5 0000 0101 Reg40005 Data Lo 50 80 0101 0000 CRC ## #### #### CRC ## #### ####
Additional Notes: The calibration LED will light when the transmitter is sending a response to a Master Query. Communications are 9600 baud, 8 data bits, 1 stop bit, No parity, half duplex 485.
3.12 DISPLAY CONTRAST ADJUST
Model IR-640/IR-642 MicroSafe™ sensors feature a 16 character backlit liquid crystal display. Like most LCDs, character contrast can be affected by viewing angle and temperature. Temperature compensation circuitry included in the MicroSafe™ design will compensate for this characteristic, however temperature extremes may still cause a shift in the contrast. Display contrast can be adjusted by the user if necessary. However, changing the contrast requires that the sensor housing be opened, thus declassif ication of the area is required.
To adjust the display contrast, remove the enclosure cover and use a jewelers screwdriver to turn the contrast
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.20
adjust screw located beneath the metallic face plate. The adjustment location is marked “CONTRAST”. See figure 7 for location.
3.13 OPTICAL SENSOR REPLACEMENT PROCEDURE
Should the optical gas sensor element (part number 370-365878-111 or 370-287724-232) require replacement, use the fol­lowing procedure: 1 - (A) If the sensor is mounted in a classified area, system power to the transmitter must first be removed before
proceeding further.
(B) If in an unclassified area, remove front enclosure cover and unplug transmitter module. 2 - Remove lower half of sensor housing using an alan wrench (3 screws). 3 - Remove existing optical sensor and replace with new optical sensor (part number 370-365878-111
).
232
or 370-287724-
4 - Re-install lower half of sensor housing. 5 - Restore system power (if classified) or plug in transmitter module and replace enclosure cover (if unclassified). 6 - As the unit reports “WARMING UP” message, use the magnetic programming tool and swipe across PGM1 or
PGM2. This will take the unit into a one-time gain setting mode which takes 1 minute to complete. 7 - Perform a new zero calibration followed by a new span calibration (see section 3.7).
3.14 TROUBLE SHOOTING GUIDE
Calibration Faults (CalFault.2) Calibration Faults (Zero or Span) can be cleared by successfully repeating the calibration attempt. Refer to the appropriate sec­tion in the Operator Manual and follow procedure.
Clearing Faults (Memory Error, LampFault.2, and Signal Faults.XX) If a Fault Diagnostic occurs (Memory Error, Lamp Fault, or Signal Fault.XX) it may be a permanent problem or temporary. If it is a temporary fault, it may be cleared using the following 1) and 2) actions. If it can not be cleared then it is a permanent problem and advanced troubleshooting may be necessary.
1) Unplug and replug transmitters to cycle power. Determine if fault has cleared.
2) Re-initialize unit by unplugging and re-plugging transmitter and then immediately swiping your magnet over PGM1. The unit will go into an initializing mode and reset the optical gain settings. Observe unit as the gain setting (GN) is counted down and Ia or Ir is counted up to 2700. Record the gain set value last reported as Ia or Ir reaches 2700. Next record the final Ia and Ir values which are displayed next. Finally record the temperature which is displayed last. Determine if fault has cleared.
Normal Readings are defined below and should be recorded to discuss with Detcon Technical Service Staff:
Gain Set:55-130 Ir: 2000-2700 Ia: 1800-2700 Temp: 32C 40C (approximately 10C higher than ambient temperature)
3) Contact Detcon for service and repair if above mentioned troubleshooting actions have not solved problem.
Zero Fault Zero Fault is an indication that the zero level has drifted below –10% of range. Zero Fault can be cleared by re-calibrating zero in target gas “free” air.
Non-readable display
1. If display has blue background when hot, install sunshade to reduce temperature.
2. If poor contrast, adjust contrast pot accordingly. Nothing Displayed – Transmitter not responding
1. Verify condulet has no accumulated water or abnormal corrosion.
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.21
2. Verify required DC power is applied to correct terminals.
. Swap with a known-good transmitter to determine if transmitter is faulty.
3
Bad 4-20 mA output or RS485 Output
1. Check that wiring is connected to correct terminal outputs.
2. Swap with a known-good transmitter to determine if transmitter is faulty
Unstable output/Sudden Spiking/Nuisance Alarms
1. Check condulet for accumulated water.
2. Check transmitter and Terminal PCB for abnormal corrosion.
3. Determine if problem correlates with condensation cycles.
4. Add/change Detcon condensation prevention packet.
5. Check for unstable power supply.
6. Check for inadequate grounding.
7. If correlates with radio communications then use Detcon RFI filter accessory.
8. Contact Detcon for assistance in optimizing shielding, grounding, and RFI protection.
Excessive Span Drift or Slow Response
1. Verify correct cal gas flow rate and proper use of the cal gas adapter.
2. Check validity of cal gas via the expiration date and use pull tube if necessary.
3. Check for obstructions through s/s sinter element (including being wet).
4. Replace plug-in sensor if Optics Life is <25%.
Drifting Zero It may be the correct reading if there are real gas leaks or the sensor was zero calibrated when actual gas was present and subse­quently cleared.
1. Recalibrate zero using target gas “free” sample.
2. Replace plug-in sensor if Optics Life is <25%.
3.15 SPARE PARTS LIST
613-010000-000 Sensor rain shield 613-120000-000 Sensor splash guard 943-000006-132 Threaded Calibration Adapter 390-000088-000 IR sensor housing assembly (plug-in detector PN370-365878-111 or 370-287724-232,
not included) 370-365878-111 Field replaceable optical plug in sensor (for .3 to 5% range) 370-287724-232 Field replaceable optical plug in sensor (for 10 to 100% range) 926-405500-RANGE IR-640 Plug-in control circuit 926-425500-RANGE IR-640 Plug-in control circuit 500-002042-000 IR Connector board 327-000000-000 Programming Magnet 897-850800-000 3 port enclosure less cover 897-850700-000 Enclosure glass lens cover 960-202200-000 Condensation prevention packet (replace annually).
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.22
Spash Guard
Rain Shield
Enclosure less cover
I
R Connector Board
C
ali br at io n Ad ap te r
E
nclosure glass lens cover
Plug-in control circuit
Field replaceable plug-in optical sensor
IR Sensor Housing Assembly
P
rogramming Magnet
C
ondensation
P
revention Packet
(
replace annually)
Figure #8
3.16 WARRANTY
Detcon, Inc., as manufacturer, warrants each NDIR optical plug in sensor (part no. 370-365878-111 or 370-287724-
232), for a pro-rated five year period under the conditions described as follows: The warranty period begins on the date of shipment to the original purchaser and ends five years thereafter. The sensor element is warranted to be free from defects in material and workmanship. Should any sensor fail to perform in accordance with published specif i­cations within the warranty period, return the defective part to Detcon, Inc., 3200 A-1 Research Forest Dr., The Woodlands, Texas 77381, for necessary repairs or replacement.
3.17 SERVICE POLICY
Detcon, Inc., as manufacturer, warrants under intended normal use each new MicroSafe™ plug-in control circuit to be free from defects in material and workmanship for a period of two years from the date of shipment to the origi­nal purchaser. Detcon, Inc., further provides for a five year fixed fee service policy wherein any failed transmitter shall be repaired or replaced as is deemed necessary by Detcon, Inc., for a fixed fee of $65.00. The fixed fee service policy shall affect any factory repair for the period following the two year warranty and shall end five years after expiration of the warranty. All warranties and service policies are FOB the Detcon facility located in The Woodlands, Texas.
3.18 SOFTWARE FLOW CHART
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.23
NORMAL
OPERATION
CALIBRATION
PGM1 (3) PGM2 (3)
PGM1 (M) PGM2 (M)
PGM1 (3)
SET ALM1 @ ##%
INC
DEC
PGM1 (3) PGM2 (M)
1-ZERO 2-SPAN
LEGEND
PGM1 - program switch location #1
PGM2 - program switch location #2
(M) - momentary pass of magnet
(3) - 3 second hold of magnet
(30) - 30 second hold of magnet
INC - increase
DEC - decrease
# - numeric value
AUTO ZERO
AUTO SPAN
S
ET ALARM 1 LEVEL
CO2 RANGE V#.#
PGM1 (3) PGM2 (M)
V
IEW PROG STATUS
ALM1 SET @ ##%
ALM2 SET @ ##%
PGM1 (M) PGM2 (M)
PGM1 (3)
SET ALM2 @ ##%
INC
DEC
PGM1 (3) PGM2 (M)
S
ET ALARM 2 LEVEL
PGM2 (3) PGM2 (3) PGM2 (3)
CAL LEVEL @ ##%
ALM1 (Firing Direction)
ALM1 (Latch State)
ALM1 (Energize State)
ALM2 (Firing Direction)
ALM2 (Latch State)
ALM2 (Energize State)
FLT (Latch State)
FLT (Energize State)
PGM1 (3) PGM1 (M)
PGM2 (30) PGM2 (M)
ALARM RESET
ALARM RESET
485 ID SET @ #
OPT IC S AT ##%
TEMPERATURE ##C
PGM1 (M) PGM2 (M)
PGM1 (3)
CAL LEVEL @ ##%
INC
DEC
PGM1 (3) PGM2 (M)
S
ET CAL LEVEL
PGM2 (3)
phone 888-367-4286, 713-559-9200 • fax 281-292-2860 • www.detcon.com • sales@detcon.com
Shipping Address: 3200 A-1 Research Forest Dr., The Woodlands, Texas 7381
Mailing Address: P.O. Box 8067, The Woodlands, Texas 77387-8067
Detcon Model IR-640/IR-642 Carbon Dioxide Sensor PG.24
Loading...