Datex-Ohmeda Tec 6 Plus User manual

5 (1)
Tec 6 Plus Vaporizer
Operation and Maintenance Manual
U
ibilit
w
ser Respons
y
This Product will perform in conformity with the description thereof contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Field Service Support center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex­Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda’s Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
Caution U.S. Federal and Canadian law restrict this device to sale by or on the order of
a licensed medical practitioner. Outside the U.S.A. and Canada, check local laws for any restriction that may apply
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code. The year of manufacture and a sequential unit number for identification.
AAA A 12345
This alpha character indicates the year of product
manufacture and when the serial number was assigned; “C” = 1999, “D” = 2000, “E” = 2001, etc. “I” and “O” are not used.
1107-0641-000 2/26/01
1
Introduction
In this section
What this manual includes ..............................................................1-2
Precautions...............................................................................1-3
Symbols ..........................................................................................1-4
1107-0641-000 2/16/01 1-1
Tec 6 Plus Vaporizer
What this manual includes
Important
This manual contains the information required to install, operate and maintain both the Datex-Ohmeda and the North American Dräger vari­ants of the Tec 6 Plus Vaporizer. All of the graphics illustrated in this manual show the Datex-Ohmeda variant, unless the procedure differs for the North American Dräger variant.
Section 1 tells you how to use this manual and supplies information
about the symbols used on the equipment.
Section 2 provides functional descriptions and theory of operation for
the major components.
Section 3 describes all the controls and indicators on the vaporizer. Section 4 tells you how to mount the vaporizer on a manifold. Section 5 provides operating instructions, including a preoperative
checkout and instructions for filling the vaporizer.
Section 6 provides maintenance instructions, including cleaning the
vaporizer and a list of spare parts. Maintenance procedures are re­stricted to those detailed in this manual.
Section 7 provides the theory of operation for the alarm and indicator
display and vaporizer delivery system.
Section 8 provides information for troubleshooting the vaporizer. Section 9 contains specifications for the vaporizer. Section 10 contains the warranty. Section 11 contains Datex-Ohmeda’s servicing policy.
Requests for servicing facilities, advice or assistance must be ad­dressed either to a local Datex-Ohmeda office or to a Datex-Ohmeda Authorized Distributor.
Datex-Ohmeda strongly recommends that you keep all relevant docu­mentation, including this manual and accompanying labels, immedi­ately available to all users.
1-2 2/16/01 1107-0641-000
Introduction
Precautions
ww
ww
WARNING Do not fill the vaporizer with any substance other than
Suprane™ (desflurane) as specified on the front label. If any substance other than Suprane™ (desflurane) is used, patient injury could occur.
U.S. Federal and Canadian law restrict this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A. and Canada, check local laws for any restrictions that may apply.
Before using the Tec 6 Plus Vaporizer, you must be familiar with the Suprane™ (desflurane) package insert. Failure to conform to the recommendations on the insert can result in patient injury.
Important:
Do not attempt to use a vaporizer that has been dropped. A dropped vaporizer MUST be sent to the nearest Datex­Ohmeda Field Operations Unit for servicing. See Section 11 for complete shipping instructions.
Do not use malfunctioning equipment. Make all necessary repairs or have the equipment serviced by an authorized Datex-Ohmeda service center. After repair, test the equipment to ensure that it is functioning properly, in accordance with the manufacturer’s published specifications.
European Standard EN 740 -
ules
requires that an appropriate gas monitor is used to monitor the concentration of anesthetic agent vapor in the inspiratory gas when the vaporizer is in operation in order to provide protection against haz­ardous output in the event of a device malfunction.
Datex-Ohmeda strongly recommends the use of anesthesia gas mon­itoring with this equipment. Refer to local standards for mandatory monitoring.
Anesthetic Workstations and Their Mod-
1107-0641-000 2/26/01 1-3
Tec 6 Plus Vaporizer
Symbols
ww
ww
WARNING Warnings tell about a condition that can cause injury to the
ww
ww
CAUTION Cautions tell about a condition that can cause damage to
Symbol Location Meaning
Warnings and Cautions tell you about dangerous conditions that can occur if you do not follow all instructions in this manual.
operator or the patient.
the equipment. Read and follow all warnings and cautions.
Control Dial Standby setting
Interlock Block Setting mark - dial marking alignment point for required setting
Interlock Block Equi-potential stud - connection for minimizing electrical potential difference
Front Panel Auditory alarm mute button
Display Panel 240 ml refill mark - indicates that an additional 240 ml of agent can be added
Display Panel Symbolizes nominal agent level display when 390 ml of agent in sump (max.)
Display Panel Symbolizes agent level display when 60 ml or more of agent in sump (min.)
Rear Label
Rear Label The degree of protection against electric shock is Type B
Top Plate Label Direction of flow
Various Warning
Various Caution
AC electrical supply must be single phase sinusoidal alternating current. The voltage requirement for the vaporizer appears on this label.
Systems with this mark agree with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance manuals. The xxxx is the certifica­tion number of the Notified Body used by Datex-Ohmeda's Quality Systems.
1-4 2/16/01 1107-0641-000
1Description
In this section
What is a Tec 6 Plus Vaporizer?.....................................................2-2
Components....................................................................................2-6
Power supply unit .....................................................................2-6
Filler assembly..........................................................................2-6
Drain plug .................................................................................2-7
Power cord and battery.............................................................2-7
Equi-potential stud....................................................................2-8
Sump ........................................................................................2-8
Heater elements .......................................................................2-8
Sensors.....................................................................................2-8
Agent level sensor....................................................................2-9
Tilt switch..................................................................................2-9
1107-0641-000 2/16/01 2-1
Tec 6 Plus Vaporizer
What is a Tec 6 Plus Vaporizer?
The Tec 6 Plus Vaporizer adds Suprane™ (desflurane) inhalant anes­thetic vapor to the medical gasses supplied to a patient. The Datex­Ohmeda vaporizer should be used in anesthesia systems equipped with a Selectatec
The North American Dräger vaporizer should be used in anesthesia systems equipped with an exclusion manifold. Both variants of the Tec 6 Plus Vaporizer must be supplied with dry medical gas and be con­nected to a suitable AC power supply.
The Tec 6 Plus Vaporizer is designed to meet UL 2601-1, EN 60601­1 and IEC 601-2-13.
®
Series Manifold only.
Warning
• Keep upright at all times.
• Do not autoclave.
• Not for use in the prescence of flammable anaesthetic agents.
• Only to be used in conlunction with an O monitor.
Warning
• Keep upright at all times.
• Do not autoclave.
Lock
before
use
• Not for use in the prescence of flammable anaesthetic agents.
• Only to be used in conlunction with an O monitor.
See manual for
• Operational Instructions.
• Factors affecting performance.
• Service and maintenance recommendations.
18
16
14
12
Next service date
10
8
7
6
5
1
4
23
Warning
High vaporizer settings with low gas flows may
lead to hypoxic mixtures within the breathing
circuit. Ohmeda strongly recommends the use of
oxygen monitoring.
Refer to manual before use.
Calibrated for use at flows of 0.2 to 10 L/min. and
at ambient temperatures of 18C to 30C at 1013mb.
Tec 6 Plus
USE ONLY DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM UP
ALARM BATTERY LOW
Use only Desflurane
See manual for
• Operational Instructions.
• Factors affecting performance.
• Service and maintenance recommendations.
18
16
14
12
10
Next service date
8
7
6
5
1
4
23
Warning
High vaporizer settings with low gas flows may
lead to hypoxic mixtures within the breathing
circuit. Ohmeda strongly recommends the use of
oxygen monitoring.
Refer to manual before use.
Calibrated for use at flows of 0.2 to 10 L/min. and
at ambient temperatures of 18C to 30C at 1013mb.
Tec 6 Plus
USE ONLY DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM–UP
ALARM BATTERY LOW
Use only Desflurane
Figure 1-1 • Tec 6 Plus Vaporizer, Datex-Ohmeda variant (left), and North American Dräger variant (right)
2-2 2/16/01 1107-0641-000
Description
Mechanical and electrical interlocks incorporated into the vaporizer help to ensure that:
The vaporizer dial can only be turned from n (Standby) when the vaporizer is locked onto the manifold and the green OPERATIONAL light is illuminated to indicate that the vaporizer has attained the correct operating temperature.
Only one vaporizer can be operated at a time.
The gas flow only enters the selected vaporizer.
Unwanted anesthetic trace vapor is minimized after the vaporizer is turned to n (Standby).
Important:
European Standard EN 740 -
ules
requires that an appropriate gas monitor is used to monitor the concentration of anesthetic agent vapor in the inspiratory gas when the vaporizer is in operation in order to provide protection against haz­ardous output in the event of a device malfunction.
Datex-Ohmeda strongly recommends the use of anesthesia gas mon­itoring with this equipment. Refer to local standards for mandatory monitoring.
Anesthetic Workstations and Their Mod-
1107-0641-000 2/16/01 2-3
Tec 6 Plus Vaporizer
ww
ww
WARNING This manual and its associated documentation must be
studied before any attempt is made to install, operate or clean any part of the Tec 6 Plus Vaporizer.
The performance of the anesthesia machine and vaporizer can be degraded if the machine and vaporizer are mis­matched.
To avoid explosive hazards, flammable anesthetic agents such as Ether and Cyclopropane must not be used in this vaporizer. Only anesthetic agents which comply with the requirements for non-flammable anesthetic agents in the IEC 601-2-13 Standard, particular requirements for the safety of anesthesia machines, are suitable for use in the presence of this vaporizer.
As this vaporizer is not suitable for use with flammable anesthetic agents such as Ether or Cyclopropane, the use of antistatic breathing tubes and face masks is not necessary. The use of antistatic or electrically conductive breathing tubes when utilizing high frequency electric surgery equipment may cause burns and is therefore not recommended in any application of this vaporizer.
The functioning of this vaporizer can be adversely effected by the operation of equipment such as high frequency surgical apparatus or short-wave therapy equipment in the vicinity.
Do not operate the vaporizer other than with dry medical gases.
The Tec 6 Plus Vaporizer must not be used in a Magnetic Resonance Imaging environment.
The functioning of this Tec 6 Plus Vaporizer can be adversely affected by electromagnetic interference exceeding the levels specified in EN 60601-1-2.
This product contains a small quantity of Mercury in the tilt switch mechanism and should be disposed of according to local regulations.
Do not attempt to use a vaporizer that has been dropped. A dropped vaporizer MUST be sent to the nearest Datex­Ohmeda Field Operations Unit for servicing. See Section 11 for complete shipping instructions.
2-4 2/16/01 1107-0641-000
Description
ww
ww
CAUTION If the vaporizer is not likely to be used for a period of 12
months, remove the battery. Failure to do so can result in damage to the vaporizer. The battery must be placed back into the vaporizer before use.
A 'No Output' alarm could be caused by downstream high frequency ventilation at greater than 60 breaths/min with an airway pressure of 100cm H
O or pulsing the O
2
faster than 1 pulse/sec over a time period of greater than 10 seconds.
flush
2
1107-0641-000 2/16/01 2-5
Tec 6 Plus Vaporizer
Components
Power supply unit
Filler assembly
The Tec 6 Plus Vaporizer operates from a 100 to 120VAC 50/60 Hz or 220 to 240V AC 50/60 Hz electrical supply. A label on the rear panel of the vaporizer states the voltage requirement for the vaporizer.
The integral power supply unit consists of a transformer and an AC to DC converter that provides the 12VDC and 5VDC supplies for the electrical system.
The filler assembly uses an agent specific fitting to ensure that only a Suprane™ (desflurane) specific bottle equipped with a Saf-T-Fill™ valve can be inserted to fill the vaporizer.
FILLER ASSEMBLY
FILLER PORT
FILLER
Figure 1-2 • Filler assembly
2-6 2/16/01 1107-0641-000
Use Only
Desflurane
AA.43.037
Description
Drain plug
ww
ww
CAUTION Do not remove the drain plug. Draining the vaporizer can
The drain plug is located at the base of the vaporizer (see Figure 2-3).
result in rapid loss of pressure and/or agent, which could lead to injury to the operator. It must only be removed at a Datex-Ohmeda Authorized Service Center.
DRAIN PLUG
Figure 1-3 • Vaporizer base
Power cord and
battery
If the AC power supply fails, a Duracell MN1604 or VARTA Energy 2000 battery incorporated in the base of the vaporizer provides power for the auditory and visual alarms only. The battery must be replaced annually.
POWER CORD
D
ghc hjv mn 623r
DURACELL
DURACELL
DURACELL MN 1604 OR VARTA ENERGY 2000 BATTERY
Figure 1-4 • Power cord and battery
1107-0641-000 2/16/01 2-7
Tec 6 Plus Vaporizer
Equi-potential stud
Sump
The equi-potential stud connected, it minimizes any electrical potential difference between the vaporizer casing and the anesthesia system.
Figure 1-5 • Equi-potential stud
The sump contains the agent and includes the agent filling port, the drain, the heaters and the agent level sensor.
The sump has a total nominal capacity of 390 milliliters (ml) which con­sists of a nominal 330 ml indicated volume and a nominal 60 ml reserve.
Y
is located on the top of the vaporizer. When
EQUI-POTENTIAL STUD
Heater elements
Sensors
Two 100W heater elements, located in the base of the sump, heat the agent to a nominal 39 °C (102 °F) to generate the working pressure.
Two 100W heater elements located in the upper part of the vaporizer help prevent condensation of the anesthetic agent in the vaporizer.
The current supplied to power the heaters alternates between the two heaters in the sump and the two heaters in the upper part of the vapor­izer. This minimizes the current requirement.
While it is connected to the electrical supply, the casing of the vapor­izer is normally warm to the touch.
Internal electro/mechanical sensors monitor and maintain the fresh gas/agent vapor pressure balance and the agent volume.
The internal sensors monitor and display operational status and acti­vate the appropriate alarm indication.
2-8 2/16/01 1107-0641-000
Description
Agent level sensor
Tilt switch
An agent level sensing probe measures the agent level. The power must be on for the agent level sensor to activate. If the sump contains between 60 and 390 ml of agent, the LCD level displays the agent lev­el in the sump. For more information, see section 3, Controls and Indi­cators.
If the vaporizer is tilted for 10 seconds while the dial is on and it is in operational mode and delivering vapor, vapor delivery will be stopped, as indicated by the flashing NO OUTPUT light and auditory alarm.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.042
Figure 1-6 • Tilt indicator (NO OUTPUT light)
1107-0641-000 2/16/01 2-9
Tec 6 Plus Vaporizer
This page left blank intentionally.
2-10 2/16/01 1107-0641-000
1Controls and Indicators
In this section
Controls...........................................................................................3-2
Dial and dial release.................................................................3-3
Lock lever .................................................................................3-4
Auditory alarm mute button.......................................................3-4
Auditory and visual indicators...................................................3-5
Alarm priority.............................................................................3-6
Status display..................................................................................3-7
Warm-up...................................................................................3-7
Operational...............................................................................3-8
Agent level display....................................................................3-8
Low agent level.........................................................................3-9
No output................................................................................3-10
Alarm battery low....................................................................3-11
1107-0641-000 2/16/01 3-1
Tec 6 Plus Vaporizer
Controls
LOCK
LEVER
DIAL RELEASE
CONCENTRATION DIAL
Note that only the Datex-Ohmeda vaporizer has the lock lever as shown.
EQUI-POTENTIAL STUD
Warning
• Keep upright at all times.
• Do not autoclave.
Lock
before
use
• Not for use in the prescence of flammable anaesthetic agents.
• Only to be used in conlunction with an O monitor.
See manual for
• Operational Instructions.
• Factors affecting performance.
• Service and maintenance recommendations.
18
16
14
12
10
Next service date
8
7
6
5
1
4
23
Warning
High vaporizer settings with low gas flows may
lead to hypoxic mixtures within the breathing
circuit. Ohmeda strongly recommends the use of
oxygen monitoring.
Refer to manual before use.
Calibrated for use at flows of 0.2 to 10 L/min. and
at ambient temperatures of 18C to 30C at 1013mb.
STANDBY SETTING
BATTERY
COVER
POWER CORD
Tec 6 Plus
AUDITORY ALARM MUTE BUTTON
USE ONLY DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM UP
ALARM BATTERY LOW
Use only
Desflurane
INDICATORS
FILLER PORT CONTROLS
D
POWER CORD
RETAINING PLATE
POWER CORD CHANNEL
Figure 3-1 • Tec 6 Plus Vaporizer controls (Datex-Ohmeda variant shown)
3-2 2/16/01 1107-0641-000
Controls and Indicators
Dial and dial release When the operational light is on, push the dial release and turn the dial
to set the agent concentration. The dial cannot be turned from the (standby) setting as long as the WARM-UP light is on.
Åben
16
14
12
10
9
8
7
6
5
1
4
23
D
Tec 6
Figure 3-2 • Concentration dial
You can select the level of agent concentration from 1% to 18%. The dial is marked at intervals of 1% from 1% to 10% and at intervals of 2% from 10% to 18%.
Use the dial release again to turn the dial from the (standby) setting and to turn the dial from 12% to a higher percent setting. All other set­tings can be made without operating the dial release.
A solenoid interlock mechanism ensures that the dial can be turned from the (standby) setting only when the green OPERATIONAL light is illuminated.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.040
Figure 3-3 • Operational light
1107-0641-000 2/16/01 3-3
Tec 6 Plus Vaporizer
Lock lever Only the Datex-Ohmeda vaporizer has a lock lever. Turn the lever fully
clockwise to lock the vaporizer in position
Figure 3-4 • Lock lever
Auditory alarm mute
button
Push to silence alarms. Hold for three seconds to start the alarm test (speaker sounds and indicators light up).
> 3 sec = Alarm test < 3 sec = Alarm silence
Figure 3-5 • Mute button
3-4 2/16/01 1107-0641-000
Auditory and visual indicators
T
The status of the vaporizer is indicated by both visual and auditory alarms (see
Press the auditory alarm mute button , located above the display panel, to mute an alarm (see
low
).
The LCD agent level display indicates the level of agent. The range is between 60 ml and 390 ml. When the indicator is below the 240 ml refill mark , you can add an additional 240 ml of agent.
AUDITORY ALARM MUTE BUTTON
Status display
).
Agent level display
Controls and Indicators
and
Alarm battery
AA.43.038
GREEN
AMBER/RED
AMBER
AMBER
AMBER
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
Figure 3-6 • Front panel display
390 ml NOMINAL
LCD AGEN LEVEL DISPLAY
240 ml REFILL
MARK
60 ml NOMINAL
1107-0641-000 2/16/01 3-5
Tec 6 Plus Vaporizer
Alarm Priority
The Tec 6 Plus Vaporizer has three distinct alarm priorities. For each priority there is a different set of visual and auditory signals:
Alarm Priority Visual Signal Auditory Signal
High Rapid red flashing
light
Medium Amber flashing light One set of three au-
Low Continuous amber
light
Note:
An internal processor failure may also present an intermittent red No Output LED and alarm tone.
Vap. Condition Low Battery Low Agent No Output
Two sets of five audio beeps every 10 seconds
dio beeps every 25 seconds.
One audio beep only
Dial ON Medium
Priority
Dial ON, Alarm Silence Pressed
Dial turned to
Low Priority Audio
Low Priority Low Priority
Figure 3-7 • Alarm priorities
Medium Priority
silenced for 120 seconds, then audio active
(dial locked until condition corrected)
High Priority
High Priority (pressing Si­lence has no effect)
Low Priority (dial locked until condition corrected)
3-6 2/16/01 1107-0641-000
Status display
Controls and Indicators
If the dial is at when an alarm occurs, the auditory alarm does not repeat.
Warm-up
The amber WARM-UP light is continuously illuminated when the va­porizer is in warm-up mode. The dial cannot be turned from when the vaporizer is warming up.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.039
Figure 3-8 • Warm-up
1107-0641-000 2/16/01 3-7
Tec 6 Plus Vaporizer
Operational
Agent level display
The green OPERATIONAL light is continuously illuminated when the vaporizer is in operational mode. The dial can be turned from when the vaporizer is operational.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.040
Figure 3-9 • Operational
The LCD agent level display consists of a series of bars that indicate the nominal volume of agent in the vaporizer sump.
The symbol means there is approximately 390 ml of agent in the sump. The symbol means there is 60 ml or more of agent in the sump. No bar displayed means there is 60 ml or less of agent in the sump.
As the sump fills, the first agent level indicating bar appears in the LCD volume display when there is between 60 and 80 ml of agent in the sump. As the level increases, more bars are displayed, as illustrated in Figure 3-9, and as the level decreases, fewer bars are displayed.
When the bars are either level with or below the symbol, the sump can accommodate 240 ml of agent, which is one full bottle. When there are no bars visible, as illustrated in Figure 3-10, the LOW AGENT level alarm is activated (see
Low agent level
).
Note:
The agent level display is calibrated to be most accurate when the vaporizer is in a perfectly level position. Deviations from a level po­sition affect the accuracy of the agent level display and may acti­vate the LOW AGENT level alarm.
3-8 2/16/01 1107-0641-000
Controls and Indicators
Low agent level
This alarm indicates that there is less than 60 ml of agent remaining in the sump and agent should be added.
Pressing mutes the auditory alarm for a period of 120 seconds to allow time to fill the vaporizer to above 60 ml.
If a LOW AGENT condition occurs during normal operation, the alarm is activated but the green OPERATIONAL light remains illuminated.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.041
Figure 3-10 • Low agent level
1107-0641-000 2/16/01 3-9
Tec 6 Plus Vaporizer
No output
This alarm indicates that the vaporizer has shut down and is no longer delivering vapor.
The NO OUTPUT alarm is activated if the dial is in the ON position and any one of the following occurs:
The agent level decreases to below 20 ml.
The vaporizer is tilted for more than 10 seconds.
A power failure of longer than 10 seconds duration occurs.
An internal malfunction is detected.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.042
Figure 3-11 • No output
If the NO OUTPUT alarm occurs while the vaporizer is in use, the va­porizer dial must be turned to .
This action de-escalates the alarm to a low priority and illuminates the amber light continuously. Either an alternative vaporizer or an alterna­tive method of anesthesia must be used.
If the NO OUTPUT alarm is activated as a result of a power failure and the dial is at when the power supply is restored, the alarm and dis­play test is activated and the vaporizer returns to either the WARM-UP or the OPERATIONAL mode.
If the dial is not at when the power supply is restored, the NO OUT­PUT alarm continues and can only be cancelled by turning the dial to . In this condition, the alarm and display test is not activated and the vaporizer returns to either the WARM-UP or the OPERATIONAL mode.
3-10 2/16/01 1107-0641-000
Controls and Indicators
Alarm battery low
This alarm indicates that the alarm battery voltage is low. A new bat­tery must be installed as soon as practical to support the NO OUTPUT alarm in the event of a subsequent external power failure.
If an ALARM BATTERY LOW condition occurs during normal opera­tion, the alarm is activated but the green OPERATIONAL light remains illuminated.
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
AA.43.043
Figure 3-12 • Alarm battery low
1107-0641-000 2/16/01 3-11