Manual Control No. : BY8-2300-0E0
Name of Product : CXDI-60G/60C
Distribution Control No.
Issued on
Service Manual Introduction
This service manual belongs to a series of after-service guides Canon Inc. publishes as part of
its comprehensive product quality guarantee program.
This service manual consists of nine chapters; “General”, “Installation Manual”, “Functions”,
“Repair Guide”, “Parts Catalog”, “Troubleshooting”, “Service Manual Report”, “Tools” and
“Appendix”. It describes an overview of the product, its functions, product configuration,
installation procedures, dimensions, specifications, and notes.
If the product undergoes a large modification, a revised edition of the service manual will be
sent to you. In other cases, a service manual report will be sent to you to update the manual.
Note 1:
This service manual is published by Canon Inc. in accordance with Article 6 (Furnishing
the Referring Materials) of the Service Assignment Contract it has concluded with your
company.
Note 2:
This service manual is the property of Canon Inc. and the company may seek to have it
returned, depending on the circumstances. You are expected to keep it until then.
Note 3:
You inquiries, suggestions, etc. about the contents of this service manual should be
addressed to:
Medical Equipment Technical Service Dept.
Canon Inc. Headquarters
30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo 146-8501, Japan
Page 3
Caution Regarding Service
This product was precisely assembled under strict manufacturing process control. There
are several hazardous locations inside of this product. Careless work while the cover is
removed can result in the pinching of fingers or electrical shock. Please perform the work
with the following important points in mind:
1. Setup, Repair, and Maintenance
In order to ensure safety, the best performance, setup, repair, and maintenance work can only
be performed by technicians who have received service training specified by Canon Inc. If
there are order required certificates or restrictions specified by the law or ordinances, those
regulations of the country must be observed.
2. Removing the external cover
When removing the cover during maintenance, repair, etc., perform the work after switching
the power off. Never touch the device with wet hands, as there is a risk of electric shock.
3. Fuse
When replacing the fuse, first resolve the reason for its failure and then replace the fuse with
the specified type. Never use a fuse other than the specified type.
4. Connecting the grounding wire
The provided ground wire must be connected to the ground terminal indoors. Make sure that
the device is properly grounded.
5. Alternation prohibition
Never modify the medical device in any way.
6. Waste control
The service provider is responsible for the disposal of used service parts, packing material,
etc. resulting from the setup, repair, or maintenance of the medical device. However, the
customer is responsible for the disposal of the medical device. Disposal activities must
follow the regulations (especially controlled industrial waste) of the country where the
device is used.
Page 4
VORSICHT
Befolgen Sie die unten angegebenen Sicherheitsanweisungen.
Mißachtung kann zu erletzungenoder Unfällen führen.
1.Zerlegung, Zusammenbau, Einstellung und Wartung
Zerlegung, Zusammenbau, Einstellung und Wartung dürfen nur von einem Wartungstechniker
durchgeführt werden, der an einem von Canon vorgeschriebenen Wartungslehrgang teilgenommen hat.
2.Entfernen von Abdeckungen
Schalten Sie unbedingt die Stromversorgung des Instruments aus, bevor Sie die Abdeckungen zwecks
Wartung und Reparatur entfernen.
Vermeiden Sie auch eine Berührung des Instruments mit nassen Händen.
Anderenfalls können Sie einen elektrischen Schlag erleiden, der zum Tod oder schwerer Verletzung
führen kann.
3.Sicherung
Wenn die Sicherung ausgewechselt werden muß, schalten Sie unbedingt die Stromversorgung des
Instruments aus, und beheben Sie die Ursache für das Durchbrennen der Sicherung.
Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag kommen.
4.Erdleiter
Erden Sie das Instrument unbedingt an einer Schukosteckdose.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag durch Leckstrom kommen.
5.Umbau
Jeder Umbau des Produktes ist strengstens untersagt, da dies zu einem Brand oder elektrischen Schlag
führen kann.
According to “IEC60601-1-1:2000”, devices installed in the patient environment are restricted
to “electric medical devices conforming to IEC60601-1”.
The Control PC and operation unit are classified under the data processing device standard
(IEC60950), therefore these items should not be installed in the patient environment.
The patient environment described below is an example cited from “IEC60601-1-1:2000” – the
measurements are only guidelines. However, the “IEC60601-1-1:2000” example must be treated
as the standard.
Therefore, the Control PC and operation unit must be installed in a location further than the
measurements below (outside of the patient environment).
*Areas where the patient moves (not only during imaging but when entering and leaving
the room, etc.) are also considered as part of the patient environment, therefore the
installation location should be determined upon consultation with the user regarding
areas outside of the patient environment.
CXDI-60G/60C represents a compact cassette model that enhances user convenience in cassettebased digital radiography. It is used in combination with Control computer (CXDI-C3, CXDI-C3S,
FC-24VE, FC-E21A or the general computers with equal performance).
The sensor cable is organized into two parts. The one is a sensor cable S150-60 (imaging unit side)
and the other is a sensor cable P630 (power box side). Two cables are connected with the connectors.
It adopts the detachable connector, which is compatible with the CXDI-55G/55C, allowing the use
of two sizes portable flat panel detector.
The vertical scanning drive (Drv-IC) and the data read out (Amp-IC) are located in one side of
LANMIT. Image data is read out from one direction.
The appearance of the CXDI-60G is almost the same as that of the CXDI-60C, but their fluorescence
substances that convert X-ray into visible light are different.
CXDI-60G
CXDI Software Ver.7.0 and later
Serial number of CXDI-60G imaging unit 100001 and later
Serial number of CXDI-60G imaging unit (nonlogo model)
CXDI-60C
CXDI Software Ver.7.2 and later
Serial number of CXDI-60C imaging unit 100001 and later
Notes on usage
From the view of risk management, guarantee is not made for the waterproofing for blood and
chemicals, hygienic safety in operating room, usage with a defibrillation device in ICU, outdoor
usage or application to animals. If the imaging unit needs to be used under such condition, the
system integrator should be responsible for the operation and understanding of the tolerance of the
product specification.
300001 and later
1
Page 9
2 CXDI System Block Diagram
AC100/120V/230V
CA1
X-Ray Generator
Symbol Description Remarks
CA1 AC Cable 3m
CA2 Sensor Cable 6.3m
CA3 X-Ray I/F Cable 20m
CA4 LAN Cable (Category 5)
CA5 LAN Cable (Category 5)
CA6 Serial Cable (Touch Panel)
CA7 VGA Cable
CA8 AC Cable 3m
Power Box
AC Power IN
X-Ray
I/F
CA3
OUT PUT 1
Control/Signal/
Power
OUT PUT 2
Control/Signal
Remote Switch
CA2
CA4
Printer
Image diagnosis
W/S
Image file device
1. General
Network
Ethernet
(100/10base-T)
CA5
Imaging Unit
(CXDI-60G/60C)
Control/Signal
/Power
LAN2
LAN1
AC100V~240V±10% 50/60Hz 2.2A
Control
computer
SERIA1(COM1)
VGA
SERIAL(COM2)
MOUSE
KEY BOARD
AC Power IN
CA8
CA6
CA7
RS232C
VGA
Mouse
Key Board
Operation
Unit
2
Page 10
3 System Diagram
3.1 Standalone System
Standalone System Block Diagram
Patient Circumstances or Medical Room
CXDI-60G/60C
AC or DC IN
Protective
Grounding
Power Box
Power Supply
1st / 2nd
AC250V
Insulation with
reinforcement
Basic Insulation
General HUB
AC250V
1. General
Control/Signal
Ethernet
(1)
*
SELV
Control/Signal
Control
Control/Signal
Connector
Imaging Unit
(2)
Remote Switch
X-ray Generator
(601Compatible)
* SELV = Safety Extra-Low Voltage
Grid Unit
FC-24VE/FC-E21A
Recommended general computer
or
3.2 Total System
It can be connect to (1) and (2) of the system where the existing products have already been
connected. Extend the ethernet port by general switching HUB or ethernet card. The maximum
number of connections is limited to four by the control software specification.
Total System Block Diagram (Example)
Patient Circumstances or Medical Room
CXDI-60G/60C
AC or DC IN
Protective
Grounding
CXDI-40E
Imaging Unit
Power Box
Power Supply
1st/2nd
AC250V
Insulation
with rei nforcemen t
Power Box
AC250V
Basic Insulation
Control/Signal
Ethernet
SELV
Control/Signal
Control
Control/Signal
Power Box
*
Imaging Unit
Connector
Remote Switch
CXDI-50
Imaging Unit
X-ray generator
(601 Compatible)
Grid Unit
Control/Signal
Ethernet
General HUB
FC-24VE/FC-E21A
Recommended general computer
or
Ethernet
Control/Signal
Ethernet
*SELV = Safety Extra-Low Voltagej
Image Examination WS
Image File Equipment
Printer etc.
3
Page 11
4 CXDI Image Processing
1. General
4.1 Proccess Flow
Born image
Raw image
Original image
QA image
DICOM output
Diagnosis image
Processed image
Correction
processing
QA processing
Offset correction
Gain correction
Defect correction
dtstore
Characteristic extraction
• Sensitivity correction
• Sharpening
• DEP
• Gradation processing
MTF enhancement
•
(Frequency enhancement)
• Grid stripe reduction
• High density clipping
request
4.2 Image Types
(1) BORN IMAGE
The image obtained with LANMIT before any correction is made.
Outside distribution of these images is prohibited, including dtstore images.
(2) RAW IMAGE
Born image after offset processing, gain correction.
This is the image with LANMIT specific characteristics corrected.
(3) ORIGINAL IMAGE
Raw image after preprocessing.
(4) QA IMAGE
Original image after gradation processing, sharpening, and other processing.
The CXDI performs image processing up to this point.
(5) DIAGNOSIS IMAGE
QA image after further image processing necessary for diagnosis.
Image processed by the user for diagnostic purposes.
(6) PROCESSED IMAGE
Diagnosis image after post-processing.
Image modified by the user or the default processed image.
4
Page 12
1. General
5 Specifications
The CXDI-60G/60C (Imaging Unit/Power Box) is the Digital Cassette that has the mobility and
can be used on the optional angles.
(1) Imaging Unit
This unit consists of the internal sensor, PWB-60Di, PCB-AD/REF, PWB-60LED and its
outer cover. The sensor unit converts the X-ray image to the electrical signal (O/E
Conversion) and after performing the A/D conversion, transfers its signal through the Power
Box with Ethernet cable to the Control computer.
Item CXDI-31 CXDI-60G/60C Remarks
Object Cassette (built-in) Cassette (built-in)/Mobile
Effective filming
range
Number of Pixels
Effective
Number of Pixels
Pixel pitch
Fluorescent
substance
Output
gradations
Transfer method
Imaging time ~ 1 sec.
Imaging cycle 15 sec.
Dimension
Imaging Unit
coloring
Imaging Unit
mass
(w/o grid)
Space between
surface where
patient gets in
contact (CFRP)
and sensor
surface (glass)
Heat emission
Power
Consumption
Mechanical
*
strength
Control
computer
Power Control
(ON/OFF)
Grid
attach/remove
detector
226.00 x 288.00 mm 234.24 x 284.16 mm
2260 x 2280
(About 6,500,000 pixel)
2256 x 2878 1464 x 1776
100µm x 100µm 160µm x 160µm
GOS 60G:GOS/60C:CsI
12bit (4,096 gradations)
A/D 14bit
Metal: Imaging Unit to E/O BOX
Optical fiber: E/O BOX to
Control computer (Through the
Multi Box)
324(W) x 327(D) x 20.8(H) mm
Art gray Cool white
2.8Kg
(Except the cable)
3.3Kg
(Including the cable)
5.75mm Within 5.1±0.5mm.
Average (43.2kJ/h)
Max 162kJ/h Max load mode
Average (12W)
Max 45W Max load mode
Load uniformly: 1.47kN(150kgf), 1min
Load locally: 0.98kN(100kgf),
The imaging unit is put on the plain surface with the Sensor side (Detector) is
up.
-CXDI-C2
-CXDI-C3
Interface with a Control computer
Ethernet: Imaging Unit to Control
computer (Through the Power Box)
344.4 (W) x 380* (D) x 22.5 (H) mm
Remote SW OFF
Remote SW OFF
φ
40mm, 1min
-FC-24VE/FC-E21A for CXDI Control
Station
-General computer with a performance
equal to FC-24VE/FC-E21A
-CXDI-C3/C3S
None
(Power Box: Operation with Remote
switch manually)
Yes
1476 x 1790
(About 2,600,000 pixel)
2.5Kg
(Except the cable)
2.8Kg
(including the cable)
36kJ/h or less
126kJ/h or less
1 image per 15 Sec.
10W or less
35W or less
(1 image per 15
Sec.)
Self-imposed rule
For CXDI-31, local
strength applies to
the center of the
sensor.
Ethernet card is
necessary to connect
CXDI-60G/60C with
CXDI-C3
5
Page 13
1. General
Item CXDI-31 CXDI-60G/60C Remarks
Sensor cable
From imaging unit to E/O Box:
4.5m
Up to four imaging units are
Count of
connected sensor
connectable to a single control
computer via multi box. For the same
type of imaging unit, the maximum of
connection number will be three.
Scattered
radiation
backward block
sheet
Lead cover Mo sheet (0.3 mm thick)
Environmentconscious
unleaded type
Photo timer Cannot be built in
When the internal temperature of
Imaging
restriction
(Imaging
Prohibition)
User interface
Imaging Unit is above 45 degree
Celsius, its state is changed to sleep
mode. And the Imaging prohibition
will be continued when the internal
temperature is below 44 degree
Celsius.
Indication
Color Two colors
Imaging
unit is off
Imaging
unit is on
Preparing
imaging
Imaging
preparation
complete
Error status
Up to four Imaging Units are connectable to a
single control computer via Network switch.
For the same type of imaging unit, the
maximum of connection number will be three.
Each imaging unit requires a power box.
When the internal temperature of Imaging Unit
is above 49 degree Celsius, its state is changed
to sleep mode. And the Imaging prohibition
will be continued when the internal
temperature is below 48 degree Celsius.
Indication Busy Sensor Power
Orange
*2
Color
Imaging
unit is off
Imaging
unit is on
Preparing
imaging
Imaging
preparation
complete
Error status
Communicating
Initialization
(when
startup)
Network not
set (when
startup)
*1: Green for China
*2: Turns on and off for 0.5 seconds each
*3: Turns on and off twice for 0.5 seconds,
then turns off for 0.5 seconds
*4: Fades in for 1 second and fades out for 1
second
*5: Fades in for 2 seconds, and then turns off
Green Blue
*1
Off Off Off
-- Off On
Blinking
--
-- On On
--
On -- On
--
--
*2
Blinking
*3
Blinking
*4
Blinking
*5
Network switch
should be procured
at each sales
company.
On
On
On
On
6
Page 14
1. General
(2) Power Box
This unit consists of PWB-60X-RAY, 60 Sensor Power Supply and its outer cover.
The function; the signal transition between Imaging unit and Control computer, the interface
to the X-ray generator equipment and power supply to the Imaging unit has been
implemented.
Item Content Remarks
Communication method
with Control computer
Communication method
with PWB-60XRAY
IEEEE
Asynchronous serial communication
method
*
802.3u (100BASE-TX) * Connector type: RJ45
Data length: 10bit
Data rate: 15.625 kHz
Reference
Power supply AC 100-240V 50/60Hz 1.2-0.7A
60 Sensor Power supply
Rated Voltage:
AC 100-240V (AC 85-264V)
Mass 3.7 Kg Except the external cables
Except bottom rubber parts
Dimension 358(W) x 200(D) x 65(H)* mm
(With bottom rubber parts:
75mm)
(3) Environment rated parameters
Operating
environment
Storing or
transporting
environment
Item Content Remarks
Ambient temperature
Relative humidity 30 to 75% RH
+5 to +35°C
Must be without dewing
Atmospheric pressure 700 to 1060 hPa
Ambient temperature
Relative humidity 10 to 60% RH
-30 to 50°C
Must be without dewing
Atmospheric pressure 700 to 1060hPa
*
IEEE: Institute of Electrical and Electronic Engineers
7
Page 15
M
M
C
C
Copyright(C) Canon Inc. Medical Technical Service Dept. All rights reserved.
X
2.. II
2
DII--
X
D
nsstt
n
a
a
Canon Inc. Japan
6
0
6
0
allll
a
n
u
n
u
G//
G
attii
a
all
a
6
6
Printed by Canon
Jul. 2009 Rev. 01
0
0
o
n
o
n
C
C
Page 16
Contents
1 Caution During the Installation ............................................................................................ 1
2 Restrictions on Installation ................................................................................................... 1
3 Caution on Installation ......................................................................................................... 2
Please pay attention to the followings when installing the system.
(1) If the equipment is hoisted, lowered or transported, it must be supported at both sides by a
minimum of two people so there is no danger of it falling.
(2) When installing the equipment, be sure the site meets the following criteria:
1) There must be no dripping water in the area.
2) The environment must be free of harmful elements such as humid or acidic air, air with
a saline or sulfur content, where there is poor ventilation or where air pressure or
temperature is unusual.
3) The equipment must not be placed at an angle or subjected to vibration or shock (this
includes during transportation).
4) The equipment must not be kept where chemical products are stored or where gasses are
generated.
5) The site’s power supply must be of the correct voltage and frequency for the equipment.
6) The site must be connected to a fully earthed cable with sufficient ground resistance to
meet standard values.
(3) After installation, be sure to dispose of waste product packaging with care and with full respect
for the environment.
2 Restrictions on Installation
(1) A clearance of at least 150mm must be left between a imaging unit and power box.
(2) It is forbidden to use the cables (Sensor cable, X-ray interface cable, etc.) from the power box
for moving parts. The only exception to this restriction is the sensor cable that is to be
connected to the imaging unit.
1
Page 18
2. Installation
3 Caution on Installation
(1) Do not install the imaging unit near electronic devices as noises and artifacts tend to appear on
images in the electromagnetic field.
e.g. CRT monitor, X-ray generator, and any other medical electronic devices.
(2) Follow the following steps to detach the sensor cable while the system is running.
1) To detach the sensor cable
- Check the LED of the imaging unit and the display of the control computer to make
sure the status of communication between the imaging unit and control computer is
idle. (*1)
- Turn off the main power switch of the power box or the remote switch. (*2)
- Make sure the LED on the imaging unit, switch on the power box and remote switch
are turned off. Power supply to the Imaging unit must be disconnected.
- Detach the connector of the sensor cable.
*1: Do not detach the sensor cable during the data transmission between the imaging
unit and control computer, it may cause data loss, system error or equipment
breakdown.
*2: Do not detach the sensor cable when the power is being supplied from the power box,
it may cause equipment breakdown. If you disconnect the imaging unit by improper
steps, the “Error” LED on the power box and remote switch will be turned on to alert
you the sensor cable is detached improperly. To restore from the error, turn off the main
switch of the power box or the remote switch. Then connect the sensor cable again by
following the steps described in the next column.
2) To attach the sensor cable
- Make sure the LED on the imaging unit, switch on the power box and remote switch
are turned off. Power supply to the Imaging unit must be disconnected.
- Connect the connector of the sensor cable.
- Turn on both the main switch of the power box and remote switch.
(3) Before proceeding with installation, ensure that the static accumulated in the bodies of the
installation personnel is discharged. Similarly, before touching the PCBs (when removing
them) or cable connectors, ensure that all static is discharged.
(4) With the launch of CXDI-60C, the specification of the CXDI-60G sensor cable is changed to
conform to CXDI-60C. For details, refer to “7.1.4 Interchanging a Imaging Unit”.
2 X-ray I/F cable 1 20m
3 Remote switch 1 20m
4 Power supply cable (with AC plug) 1 3m (100/120/230V)
5 Operation manual (for power box)
3) Sensor Cable (Optional)
No. Item Name Qty Remarks
1 Sensor Cable P630-PM (Panel mount type)1 6.3m
2 Sensor Cable P70-60 (imaging unit side) 1 0.7m
3 Sensor Cable SP780-60 (Straight type) 1 7.8 m
LAN cable for connecting Control Computer / Power Box and Network switch (Switching HUB)
for connecting the multiple Imaging Units shall be procured at each sales company.
with 1.5 m sensor cable
S150-60 ( imaging unit side)
- LAN cable (Over category 5)
Recommended length of the cable is 30m or less.
When Control Computer and Power Box are connected directly, Cross type is used, but when
they are connected via Network switch, Straight type is used. However, this is not applied
when Network switch has AUTO-MDI/MDI-X function*.
- Network switch (Switching HUB)
Sales companies adopt Network switch (Switching HUB) after conducting the test and the
operation check for Switching HUB that meets the general standard.
3
Page 20
2. Installation
4.2 Configuration
No. 1 No. 2
Name
Qty 1 Qty 1
Remarks
No. 3 No. 4
Name Power Box Name X-ray I/F cable
Qty 1 Qty 1
Remarks I/F and Power supply Remarks
CXDI-60G/60C Imaging Unit
Sensor cable S150-60
(imaging unit side)
Name
Remarks 6.3m
Sensor cable P630
(Power box side)
Connection with X-ray
generator
No. 5 No. 6
Name Remote switch Name Power supply cable
Qty 1 Qty 1
Remarks
Switch to turn on and off
Power Box
Remarks
For Power Box
(100/120/230V each type)
4
Page 21
4.3 Sensor Cable (Optional)
No. 1 No. 2
Name
Qty 1 Qty 1
Remarks
Sensor Cable P630-PM
(Panel mount type)
6.3m
Detachable cable connector
2. Installation
Remarks
Name
Sensor cable S70-60
(Imaging unit side)
0.7m
Detachable cable connector
Connector form: New type
(External diameter φ18)
No. 3
Name
Qty 1
Remarks
Sensor Cable SP780-60
(Straight type)
7.8 m
Non-detachable cable
Sensor cable P630-PM (panel mount type)
This sensor cable can be used in place of a standard sensor cable P630 (power box side). Only the
difference between them is the form of the connector. There is no difference in the function of the
cables.
Sensor cable S70-60 (imaging unit side)
This sensor cable can be used in place of a standard sensor cable S150-60 (imaging unit side). Only the
difference between them is the length of the cable. The sensor cable S70-60 is 0.7m in length. There is
no difference in the function of the cables.
Sensor cable SP780-60 (straight type)
This is a single straight type sensor cable with no detachable connectors. This single cable can be used
in place of a standard sensor cable S150-60 (imaging unit side) and a standard sensor cable P630
(power box side). There is no difference in the function of the cables.
5
Page 22
2. Installation
5 Packing Diagram
5.1 X-ray Digital Radiography System (CXDI-60G/60C)
(1)CXDI-60G/60C Imaging Unit Package
Sensor cable P630
(6.3m)
Place the sensor cable
S150-60 on the packet
corner of the paper tube
Paper tube
Corner pad
Positioning sheet
Cardboard sheet
Center pad
Imaging unit
Corner pad
Cardboard box for imaging unit
6
Page 23
2. Installation
(2)CXDI System II assembly package
Power box
Accessories
box
7
Page 24
Orig
abe
5.2 Grid (Optional)
2. Installation
(1) Outer packaging
Outer package
(2) Inner packaging
Protection sheet
Corner Packing
Material label
Product division label
Handle with
in l
Shipping
environment
label
care label
l
Protection board
Inner package
Grid unit
Plastic bag
8
Page 25
2. Installation
6 Installation Procedures
6.1 Lists of Tools Needed for Installation
No. Tool Name Unit Remarks
1. General Tools 1 set JIS Screwdriver Set
2. Laptop computer 1
3. LAN Card 1 For Laptop computer (as required)
4. Mouse 1 PS/2 type
5. Keyboard 1 PS/2 type
6. HUB 1 Connection between Control Computer and Laptop Computer
7. 10/100BASE-TX cable 2 Straight type (Control PC to Laptop computer)
CXDI Software version
8.
compatibility table
Mirror, oil-based marker
9.
and paper etc.
PC/AT compatible
(OS: Microsoft Windows XP Professional recommend)
-
1 For adjusting the alignment with the X-ray tube.
9
Page 26
2. Installation
6.2 System Installation Procedures
No. Step Conditions and Checkpoints Reference Section
1 Unpacking and checking the
product’s constituent parts
2 Connect the Imaging Unit
and the Power Box
3 Connect the Power Box and
the Control PC
4 Connect the Power Box and
the X-ray generators
5 Check date and time - Date and time must be changed according to
6 Check the software
program’s version
7 Identifying the Imaging Unit
(input the sensor serial
numbers)
8 Enter control computer serial
number.
(9) Adjusting the timing with
X-ray generator
10 Calibration -No error must be displayed.
11 Setting the Fixed ROI Areas If necessary, set the ROI area.
12 Set exposure parameter table -Set it in consultation with the technician.
13 Set annotation -Set it in consultation with the technician.
14 Connect the network and set
the output destination
15 Startup settings
16 Exposure testing -The data must be sent to the printer and
17 Check the linearity of the
transferred image density.
18 Operation unit Gamma
correction
19 Body parts settings -The engineer in charge must be consulted
20 Check and set the system
settings.
21 Total adjustments and delete
the unnecessary data.
22 Cleaning
-There must be no missing parts, damage,
dents, etc.
-There must no color changes in the shock
sensor.
-Handle the instrument carefully, as it may be
damaged if something is hit against it,
dropped or receives the strong jolt.
-The cable must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
-The cable must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
-The manufacturer of the X-ray generators
must be asked to handle the connections with
the generators.
the area where the instrument is installed.
-The compatibility of the imaging unit and
the Control PC must be checked on the
compatibility list, and the software program
must be installed or upgrade as required.
-No required usually.
storage and the image quality must be
checked.
prior to perform these settings.
-The engineer in charge must be consulted
prior to perform these settings.
-Conform according to the check sheet.
-Delete the unnecessary data.
“(1) Checking and
Setting the Date and
Time” in section 7.6.
“(2) Checking the
Firmware Version” in
section 7.6.
“(6) Identifying the
Imaging Units” in
section 7.6.
“(7) Entering Control
Computer Serial
Number” in section 7.6.
Operation Manual
“(8) Table Setup
Setting” in section 7.6.
“(9) Performing the
Annotation Setting” in
section 7.6.
“(10) Network
Connections” in section
7.6.
“(5) Set Up Startup
Menu” in section 7.6.
Section 5.7 Image
Quality.
“(11) Linearity Check
Image Density” in
section 7.6.
“(12) Operation Unit
Gamma Correction” in
section 7.6.
Operation Manual
Each section in section
7.6. Settings.
Section 7.9
Post-installation check.
10
Page 27
2. Installation
No. Step Conditions and Checkpoints Reference Section
23 Explain the operation to the
user
24 Final parameter adjustment -The engineer in charge must be consulted
prior narrowing down the adjustments to the
final values.
25 Inserting the backup floppy
disk.
-It must be confirmed at re-start that backup
files have been made.
-No necessary for the system installed in
vehicles.
26 Back up valuable data -Copy the CCR folder to the removal drive.
Operation Manual
Operation Manual
“(15) Backing up
Setting Data to FD” in
section 7.6.
“(14) Backing Up when
Installing” in section
7.6.
11
Page 28
2. Installation Manual
7 Installation
7.1 Connection to Each Unit
7.1.1 Connection Diagram
CXDI-60G/60C
Control
computer
J1
J2
J4
PWB-XRAY
POWER SUPPLY
J3
Remote Switch
POWER BOX
12
Page 29
2. Installation
7.1.2 Connecting to the Power Box
(1) Removing the cover
1) Remove the 5 screws from the back of the power box and the 2 screws on each side at the
bottom of the power box to remove the top cover.
Different types of screw are used for backside and lateral side. Make sure to use the proper type
of screw respectively on installation.
Also remove screws on
the far side
2) Before connecting the X-ray I/F cable and remote cable, remove the LAN cable connector
guard to avoid injury.
LAN cable connector guard
13
Page 30
(2) Cable connections
1) Loosen the skin top spiral and lock nut of the remote cable and X-ray I/F cable. Do not
remove the insulation lock from the X-ray I/F cable as it is put on the cable to prevent the
cable from falling off.
*The X-ray I/F cable is 20m in length. If the cable is too long, you may need to make it shorter.
Refer to “7.1.7. Adjusting the length of X-ray I/F cable” for details.
Skin top spiral
2)Get the connector and lock nut of the remote cable through the hole in a power box.
2. Installation
Remote Cable
Lock nut
Insulation lock
X-ray I/F Cable
Adjust the direction of the connector
to the notch of a hole.
Get the lock nut through
the hole.
14
Page 31
2. Installation
gap
3) Connect the connector of the remote cable to the power supply and joint the skin top spiral
and lock nut temporarily. Adjust the length of the cable with some margin in length for
wiring. After adjusting the length properly, fasten the skin top spiral tightly.
Skin top spiral
Lock nut
Connector
Note: To avoid the risk of damage when the cables are removed with very large force,
check skin top spiral of the bush (refer the following figure).
Gap 1.0 to 1.5mm
No
<Remote cable> <X-RAY I/F cable>
4) Fasten the skin top spiral and lock nut tightly using 15mm screw wrench.
15
Page 32
2. Installation
5) Connect the connector of X-RAY I/F cable to the power supply and joint the skin top spiral
and lock nut temporarily. Adjust the length of the cable so that the clearance of the skin top
spiral and insulation lock is 5mm. After adjusting the length, fasten the skin top spiral tightly.
Refer to the note in step 3.
Skin top spiral
Lock nut
Insulation lock
Connector
About 5mm clearances
6) Tighten the skin top spiral and lock nut using 15mm screw wrench.
7) After connecting the cables, attach the cover of power box and LAN cable connector guard.
*Different types of screw are used for backside and lateral side. Make sure to use proper
type of screw respectively on installation.
8) Connect the sensor cable, LAN cable, AC cable to the backside of the power box.
*1 Push the connector of the sensor cable into the connector terminal and fasten it tightly
until the lock ring stops.
*2 Only the AC cable attached to the product is allowed to use.
Sensor cable
AC Power cable
LAN cable
Lock ring
Fasten the sensor
cable while applying
the pressure
Gap
No gap
[Improper connection]
16
[Proper connection]
Page 33
2. Installation
[
9) Align the red marks on the connector of the sensor cable on both sides and joint the
connectors together until it locks. It snaps when it locks.
*1 Turn off the power of the power box before connecting or disconnecting the sensor
cable.
*2 Be careful not to drop the connector of the cable, it may cause injury or get the things
damaged as it is heavy.
Connector on the
imaging unit side
Connector on the
power box side
Gap
No gap
[Improper connection]
Proper connection]
17
Page 34
2. Installation
10) To disconnect the sensor cable, push the connector of the imaging unit side cable outward to
release the lock and pull out the connector of the imaging unit side straightforward. Do not grab
the cable itself.
*1 Turn off the power of the power box before connecting or disconnecting the sensor
cable.
*2 Be careful not to drop the connector of the cable, it may cause injury or get the things
damaged as it is heavy.
Release the lock
18
Page 35
2. Installation Manual
7.1.3 Using the Imaging Unit in the Multiple Rooms
Users can use one imaging unit in the multiple rooms with the CXDI SYSTEM II installed by
the service engineer. The steps to connect and disconnect the detachable connector of the
sensor cable are described in the operation manual.
Imaging
Unit
X-Ray R oom 1
Sensor Cable
Power Box
Control
Computer
Generator
X-Ray R oom 2
Sensor Cable
Power Box
Control
Com puter
Generator
19
Page 36
2. Installation
7.1.4 Interchanging a Imaging Unit
The detachable connector enables you to replace the imaging unit with the CXDI-55G/55C
imaging unit. As for the CXDI-60G having the old type sensor cable, plug and socket of the
connector does not match. Therefore, you need to replace it with the new type sensor cable.
With the launch of CXDI-60C, the new type sensor cable, which is conform to CXDI-60C, has
been adopted to CXDI-60G.
Operation unit
Control
Computer
Ethernet
CXDI-55G/55C
imaging unit
X-ray tube
X-ray
generator
Power box
Ethernet
CXDI-60G/60C
imaging unit
20
Page 37
2. Installation
Detachable connector of the CXDI-55G/55C sensor cable
CXDI-55G/55C
Imaging Unit
Power Box
Detachable connector of the CXDI-60G sensor cable (New type) and the CXDI-60C sensor
cable
CXDI-60G/60C
Imaging Unit
Power Box
Detachable connector of the old type sensor cable on the CXDI-60G imaging unit side
For the CXDI-60G imaging unit having the old type sensor cable, plug and socket of the
connector does not match. Refer to the Service Manual report CXDI-60G 09-002 for details.
CXDI-60G
Imaging Unit
(Old type sensor cable)
Power Box
21
Page 38
2. Installation
7.1.5 Sensor Cable (Optional)
(1) Sensor cable S70-60 (imaging unit side)
You can replace the sensor cable on the imaging unit side with the shorter type of 0.7meters
long sensor cable.
(2) Sensor cable SP780-60 (Straight type)
The sensor cable can be replaced with the straight type sensor cable with no detachable
connectors.
Operation unit
Control
Computer
Ethernet
Power box
Ethernet
(3) Sensor cable P630-PM (Panel mount type)
This sensor cable can be used in place of a standard sensor cable P630 (power box side). The
connector of the sensor cable P630-PM(panel mount type) can be tightened to a stand or a table.
Note that an insertion hole of 24.1mm in diameter is necessary for a stand or a table to fasten the
connector.
X-ray tube
X-ray
generator
CXDI-60G/60C
Imaging unit
22
Page 39
2. Installation
7.1.6 Connection Diagram for Control Computer Rear Panel
FC-E21A
VGA Port
(For Operation Unit)
COM2*
RS-232C Cable
(For Operation Unit)
Mouse
Power Supply
*Connect the RS-232C cable to the COM port, COM 1 or COM2, specified as the port in
the driver for Operation Unit.
LAN 2
For connecting to a power box
Keyboard
LAN 1
For connecting to a network
COM1*
RS-232C Port
(For Operation Unit)
23
Page 40
2. Installation
7.1.7 Adjusting the Length of X-ray I/F Cable
X-ray I/F cable is 20m in length. It may be too long for some installation sites and you
may need to make it short. Using the following parts is necessary and the output and input
condition should be satisfied.
Output
Relay: Voltage AC250V max/DC30V max, Current 10mA to 2A
Retardation
Retardation time: 20msec or less
Input
Electrical specification
Photo coupler: Current loop with the resistance 100 Ω or less
Retardation
Retardation time: 1msec or less
3 KE2-2101-000 Insulation Tube, L=1 METER F2 (Z) 6X0.25 (SUMITOMO)
Use the specified tool (YNT-2622/JST) to press the splice terminal. For details, refer to
the instruction manual for splice terminal tool.
V0.5-4 Clear/ JST
Tool number: YNT-2622 / JST
VS-2 / HAGITEC
24
Page 41
2. Installation
7.2 Starting up and Shutting Down the System
Perform the following sequences when starting up and shutting down the system.
7.2.1 Sequence for Starting up the System
Perform the following sequence when turning the system power on.
If you do not perform the correct sequence, the imaging unit cannot be recognized, resulting in
an error. (This is because the system communicates with the imaging unit when turning the
system on.)
The power box cannot be turned on in conjunction with turning on the control computer.
Since the power box is equipped with a remote switch that turns on/off the secondary output,
you caninstall the switch on your side to turn it on/off.
1) Turn on the main power of the power box.
2) Turn on the remote switch of the power box.
3) Turn on the control computer.
Note:
Ccrstart.bat should be registered in Windows Startup.
7.2.2 Sequence for Turning the Power off (Shutdown)
1) From OPU, select SYSTEM → [SHUTDOWN] or [SHUTDOWN after transfer]
The control computer automatically turns off.
2) Turn off the remote switch of the power box.
3) Turn off the main power of the power box.
Note:
Turn off the main power of the power box and OPU power when not using the system for
a long period.
25
Page 42
2. Installation
g
(
(
7.3 X-ray Controller Interface
7.3.1 Interface Signal Description
X-ray Generator
X-ray emission signal
1st SW level ON)
X-ray emission signal
2nd SW level ON)
Start X-ray emission
Generator
Setup timer
End X-ray emission
X-ray emission signal OFF
RX_REQ
RX_COM
CXDI
Detect X-ray emission
nal ON
si
Imaging possible?
Enable X-ray emission
ON
Image Data read
Enable X-ray emission
OFF
26
Page 43
•
During normal imaging
2. Installation
• During timeout due to RX_REQ not negating
Timeout: 1100msec
27
Page 44
2. Installation
7.3.2 Signal Names and Functions in the Connection with the X-ray Generator
<X-ray Sync Signal>
Signal name Functions
X-ray exposure signal
Indicates that an X-ray exposure is ordered at the X-ray generator side.
RX_REQ
RX_COM
This signal needs to be retained at least for T
X-ray release signal (RX_COM) is not output if the T
period. It takes about 1 second in the worst case scenario before the operation can be
resumed when RX_REQ is negated in this period.
Time required to assert RX_COM after receiving the RX_REQ from the X-ray
generator
Since a captured image is read from the sensor when RX_REQ is negated (or RX_COM
is time out) as a trigger, reading action is delayed if RX_COM does not negate and time
out is used as the trigger, resulting in delay of image display timing.
We recommend using a configuration in which RX_REQ is negated.
Time required from X-ray exposure completion to negating RX_REQ
* Image display timing is delayed if this is not fulfilled.
X-ray release signal
Checks whether or not imaging is ready at the CXDI side after receiving X-ray exposure
signal (RX_REQ) from the X-ray generator. This signal is output to the X-ray generator
side when imaging is ready.
Time required from asserting RX_COM to exposing X-ray
* Problems such as shading could occur on images if this is not fulfilled.
PREP
period.
PREP
is less than the necessary
28
Page 45
2. Installation
POWER BOX
X-ray Generator
+5V(isolated)
Photo coupler
RX_REQ
PC1
TLP281(GR/F)
Relay
Connector
J4
#2
#1
#4
YEL
WHT
GRY
A2
A1
X-ray request signal
*1:Impedance between A1-A2 < 100ƒ ¶
B2
X-ray permission signal
RX_COM
*2:Permitted Vold AC250Vmax/DC30Vmax
Permitted Current 10mA-2A
ORG
B1
RL1
NC2D-P-DC5V
#3
X-ray I/F Cable
The connection with CXDI-60G/60C and X-ray generator equipment
Connection conditions
1. The X-ray exposure signal line (including the switching function) must be insulated, and its total
impedance must be 100 ohms or less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V and
DC 30V, and its current ranges from 10mA to 2A.
Only the insulated secondary power supply can be connected.
3. Protective grounding for X-ray generator should be equipotent with the system.
29
Page 46
2. Installation
7.3.3 Rating and Characteristics for Relay and Photo Coupler (on PWB60XRAY)
(1) RL1 (Power Relay/Plug-in terminal type)
1) Rating (Operational coil)
Rated
Rated
voltage
(V)
DC5V 72 69.4 69.5 86.0
curre
nt
(mA) (Ω)
Coil
resistance
Coil Inductance
(mH)
Armature
OFF
Armature
ON
Pick-up
voltage
(V) (V) (V) (mW)
below
80%
Dropout
voltage
below
10%
Maximu
m voltage
135%
(at 50)
Power
consumption
Approx 360
2) Rating (Switch/Contact)
types Single stable
arrangement 2 Form C
Contact material Au-clad AgNi type
Relating capacity
AC250 5A
DC30V 5A
Max. switching power 1250VA 150W
Max. switching voltage 250V AC
Max. switching current 5A
Min. switching capacity
100µA 1V DC
3) Characteristics
Item Content
Operate time Max. 20ms
Reset time Max. 10ms
Mechanical 18,000 times/hour Maximum open/close
frequency
Withstand voltage
Life
Ambient temperature
Maximum operating frequency 50 times/Sec.
Rated load 1,800 times/hour
Between coil
contacts
Between same poles 1,000 Vrms
Mechanical 5x10
Electrical
2,000 Vrms
7
times
5
10
at 5A 250V AC
5
5x10
at 5A 30V DC
-40°C to +70°C (no freezing nor condensation)
30
Page 47
(2) PC1 (Photo-coupler)
1) Maximum Ratings (Ta = 25°C)
CHARACTERISTIC SYMBOL RATING
Forward Current IF(RMS) 50 mA
Forward Current
LED
Pulse forward current1 I
Reverse Voltage VR 5 V
Collector-Emitter Voltage V
Emitter-Collector Voltage V
Collector Current IC 50 mA
2. Installation
/°C -0.7(Ta≥53°C)
∆IF
1 A
FP
80 V
CEO
7 V
ECO
Collector Power Dissipation
DETECTOR
(1 Circuit)
Total Package Power Dissipation
(1 Circuit)
Isolation Voltage2 BVs 2500 Vms
Operation temperature
Storage temperature
2) Electrical Characteristics (Ta = 25°C)
CHARACTERISTIC SYMBOL
Forward Voltage VF I
LED
DETECTOR
SWITCH
Reverse Current IR V
Capacitance Ct V=0, f=1 MHz - 30 - pF
Collector-Emitter
Breakdown Voltage
Emitter-Collector
Breakdown Voltage
V
V
Collector Dark Current I
I
(BR)CEO
I
(BR)ECO
CEO
Rise Time tr - 2 -
Fall Time tf - 3 -
Turn-On Time tON - 3 -
Turn-Off Time t
Off
Turn-On Time tON - 2 -
CHARACTRISTIC
Storage Time tS - 25 -
Turn-Off Time t
OFF
150 mW
P
C
200 mW
P
T
T
-55 to 100
opr
T
-55 to 125
stg
TEST
CONDITION
=10 mA 1.0 1.15 1.3 V
F
=5 V - - 10 µA
R
=0.5 mA 80 - - V
C
=0.1 mA 7 - - V
E
VCE=48 V
Ambient Light
Below (100lx)
=48 V, Ta=85
V
CE
Ambient Light
Below (100lx)
V
=10 V
CC
I
=2 mA
C
R
=100Ω
L
MIN. TYP MAX. UNIT
-
-
0.01
(2)
2
(4)
0.1
(10)
50
(50)
- 3 -
V
=5 V
CC
I
=16 mA
F
R
=1.9 kΩ
L
- 40 -
µA
µs
µs
Note: Because of the construction, leak current might be increased by ambient light. Please use
photo-coupler with less ambient light
1
pulse amplitude 100µs, frequency 100Hz
2
AC, 1min R.H.≤60 LED side pins shorted together and DETECTOR side pins shorted together
31
Page 48
2. Installation
7.4 Network Settings
1. Objective
The imaging part communicates with the control computer by using Ethernet [IEEE802.3u
(100Base-TX)] to transfer X-ray images.
The control computer performs DICOM transfer in order to use Ethernet to transfer the obtained
images to the printer and storage device.
This section describes how to set up the TCP/IP that is necessary for the network connection.
Set up the following three items:
1) TCP/IP setting for the control computer network card
2) Network setting for Screwcap.ini
3) Network setting stored in the imaging unit.
2. Preparation
• Keyboard
• Mouse
Connect the keyboard and mouse to the back of the control computer.
Check if the system is connected, and then turn the system on.
3. Setup method
3.1 TCP/IP setting for the control computer network card
Perform the set up by referring to “FC-24VE (or FC-E21A) for CXDI Control Station Service
Manual” -> the chapter “System Manual” -> “Network Setup”.
Default values
IP Address:192.168.100.10
SubnetMask:255.255.255.0
32
Page 49
2. Installation
3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through Screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
imaging unit such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the imaging unit in the previous item, “3.1 TCP/IP setting for
the control computer network card”.
Screwcap.ini is located in the following directory:
Sensor IP address.*1
This must be the same as the
IP address stored in the
imaging part.
1
*
Imaging Unit IP address: This address must be the same as the IP address stored in
the next item, “3.3 Network setting stored in the sensor”.
33
Page 50
2. Installation
3.3 Network setting stored in the sensor
The factory default setting is shown in the table below.
This setting is not necessary
Item to be set Factory default value
Imaging unit IP address 192.168.100.11
Subnet mask 255.255.255.0
Host IP address 192.168.100.10
Port number for command 12121
Port number for data 12122
If you change the setting, refer to “Tool Software Operation Manual for Ethernet” → “Imaging
Part IP Address Setting”.
unless you have changed the setting.
34
Page 51
2. Installation
7.5 Setting the Fixed ROI Areas
1) Purpose
Set the fixed ROI area on the sensor to expose by the fixed ROI area because user can
not get the proper image by the Auto ROI area.
2) Setting method
2-1) Investigate the actual size and position of the ROI that is required.
2-2) Designate the SIZE, POSITION, and NUMBER (max 5)* on the sensor.
* However, in the case of using the new function “Display of AEC (Automated Exposure Control)
Field in Preview Screen” added from CXDI System Software Ver.6.2, NUMBER that can be
specified is max 3
.
(0,0)
Y
X
W
H
Hs=284mm
Sensor
Ws=234mm
2-3) Convert the size and position of the ROIs in 2) to pixel values. The pixel size of the
sensor is 160µm. For multiple values, use X’, Y’, W’, H’, X”, Y”, W”, and H” for
calculations.
X/160 µm Let this value equal A
Y/160 µm Let this value equal B
W/160 µm Let this value equal C
H/160 µm Let this value equal D
2-4) Open the “DENSITY ADJUSTMENT CONTROL” screen when the normal
imaging screen is active.
SYSTEMÆSETUP MENUÆSYSTEM SETTINGSÆ DENSITY
ADJUSTMENT CONTROL
35
Page 52
2. Installation
2-5) The “DENSITY ADJUSTMENT CONTROL” screen appears. Confirm that the
sensor is set with a sensor switch button. Press the [Fixed ROI Area] key.
2-6) Press the [Area setting] key.
Area setting button
2-7) The fixed ROI 1 setting screen appears. Input values A to D from step 3) into the
edit box, and press [ENABLE]. To set multiple fixed ROIs, input A’ to D’ and A”
to D” into fixed ROI 2 settings and fixed ROI 3 settings respectively.
-
B-B”
C-C”
D-D”
”
36
Page 53
2. Installation
2-8) The display returns to the “DENSITY ADJUSTMENT CONTROL” screen.
Confirm that POSITION and SIZE fields not set in step 5) to step 7) are disabled
(dimmed). If they are not dimmed, press the [Area setting] key, and press
[DISABLE] in fixed ROI * settings.
2-9) After Confirming all settings, and press [EXIT].
2-10) The display returns to the system settings screen. Press [OK].
2-11) “Change settings?” appears. Press [OK]. Be careful, because if [CANCEL] is
pressed, all changes made to the settings are deleted.
2-12) Return to the normal imaging screen, and turn off the power to
the CXDI.
37
Page 54
2. Installation
7.6 Settings
(1) Checking and Setting the Date and Time
Description about CXDI software in ‘Setting’ may change to some degree depending on
the versions of software. For CXDI software, see “New Function Descriptions” issued
for every version if necessary.
1) Purpose
The date and time is set to Japan standard time at factory shipment.
Reset the date and time to your local value as necessary.
2) Procedure
2-1) When CXDI application start, open the ADMINISTRATOR SETUP MENU.
SYSTEM → SETUP MENU → ADMINISTRATOR SETUP
2-2) When the “ADMINISTRATOR SETUP MENU” appears, and presses the [DATE]
button.
2-3) The dialog (Date / Time Properties) appears, and set the value properly each of the
fields which the tab sheet (Date&Time and Time Zone sheet) has. And then press
[OK].
38
Page 55
2. Installation
(2) Checking the Firmware Version
1) Purpose
1-1) Failing to use the proper versions of the firmware and PLD code with the CXDI
software can result in an error, and system operation cannot be guaranteed. Therefore,
the versions of the firmware must be checked to ensure that they are correct.
2) Notes
2-1) This check should always be performed at installation, and if necessary, the firmware
versions should be upgraded.
2-2) This check cannot be performed with only the control computer. Connect the imaging
units and other equipment, and start up in the normal imaging status.
3) Procedure
3-1) Checking the firmware alone
a. Start up the CXDI software.
b. Display the version information from the user mode.
SYSTEM > SETUP MENU > VERSION INFORMATION
c. Confirm the firmware version.
[System Module: Version Information]
BC
A
A. Hardware version
Product type: Product type is identified from Sensor serial No. that is set to the Imaging
unit. 0b.** indicates 60G and 0f.** indicates 60C.
B. Firmware initialization code version
This is the version of the initialization code written on the PWB-60Di. Initialization code
will be downloaded and settings will be reset to the default (factory) settings by turning
ON the power while pressing the initialization switch on the power box.
The display “3.0.08” on the screen indicates the version 3.00.08.
C. Firmware normal code version
This is the version of the normal code installed on the PWB-60Di. Usually the system
operates with this code. It must be updated as required.
The display “3.0.08” on the screen indicates the version 3.00.08.
39
Page 56
3-2) Checking the firmware and PLD code
(1) Connect the keyboard and mouse.
(2) Start up the CXDI system.
(3) Close the CXDI software if it starts up.
(4) Connect Telnet by referring to “Telnet Connection” in the Tool Software Operation
Manual for Ethernet.
(5) Check the versions of the firmware and PLD code on the screen displayed after the
login.
login: canon
Password: ********
CXDI-60 firmware Ver.3.00.08
build as May 19 2008 10:39:06
normal boot
HUB60 Ver.6004 prod code:0bc0
pwb-60di>
(6) After you finish checking, close HyperTerminal.
2. Installation
Firmware version
Hardware version
PLD code version
40
Page 57
2. Installation
(3) Installing Firmware and PLD Code
1) Purpose
Write exposure code and PLD code into the Flash ROM of the PWB-60Di in the imaging
unit.
2) Notes
Be sure to check that the CXDI is connected to the system.
3) Procedure
3-1) Installing the firmware
Write the firmware by referring to “Firm Write Tool Software (Firmwrite.exe)” in the Tool
Software Operation Manual for Ethernet.
Where to write: PWB-60Di
3-2) Installing PLD code
Write PLD code by referring to “PLD Write Tool Software (pldwrite.exe)” in the Tool
Software Operation Manual for Ethernet.
Where to write: PWB-60Di.
41
Page 58
2. Installation Manual
(4) Checking the Sensor Serial No.
1) Purpose
If the sensor serial number of the imaging unit differs from the sensor serial number stored
in the HD of the Control computer due to replacing the imaging unit, the connected imaging
unit cannot be detected after the CXDI software is launched and an error message appears.
In that case you need to register the sensor serial number.
2) Notes
2-1) Check the sensor serial number whenever the imaging unit is replaced.
2-2) This checking procedure must be performed with the Control computer, Imaging Unit
and all the other equipments connected and started up.
3) Sensor serial number registration
Register the sensor serial No. from CCR Console Menu. Refer to “Identifying the Sensor
Units” for details.
42
Page 59
2. Installation
(5) Set Up Startup Menu
1) Purposes
1-1) Register the CXDI software to the “Startup Group”.
The CXDI software is started automatically when the CXDI system is turned on.
1-2) Change the window view size
Hide the other application screen view except the CXDI software.
1-3) Delete the CXDI software from the “Startup Group”.
The CXDX software is not started when the CXDI system is turned on.
2) Notes
2-1) The CXDI software is not registered in the “Startup Group” at the factory setting.
Therefore register the CXDI software to the “Startup Group” after the system
installation.
2-2) The window view size of the program registered in “Startup Group” has one own size
with the each short-cut icon. Be sure to set the window view size of CXDI software at
the same time with the register to the “Startup Group”.
3) Register the CXDI software to the “Startup Group” procedure.
3-1) Connect keyboard and mouse to the control computer.
3-2) Turn the all CXDI system power on after the all installation finished. And after that
Windows XP starts.
3-3) Open the “Taskbar and Start Menu” from the Start Menu.
Start⇒Settings⇒Taskbar and Start Menu
3-4) “Taskbar and Start Menu Properties” appears. Click “Start Menu” tab, and then click
Taskbar and Start Menu Properties⇒Start Menu⇒Classic Start⇒Menu Customize
3-5) Click Add, and Create Shortcut appears. Click Browse.
3-6) Browse appears. Find a file named “ccrstart.bat” in drive [D:\ccr] and click OK.
3-7) D:\ccr\ccrstart.bat appears in the Command line. Click Next.
3-8) Select Program Folder appears. Select Startup folder and click [Next].
3-9) Select a name for the shortcut appears. Type ccrstart.bat. Click [Finish].
3-10) Close the Taskbar [Start], and login again to Windows XP.
Start⇒Shut Down⇒Log off cxdi.
3-11) After login the computer, make sure that the CXDI software starts up.
43
Page 60
2. Installation
4) Change the window view size
4-1) After the CXDI software starts, press [Alt] + [Tab] key to show the “CCR Console
Menu” prompt screen.
4-2) After the command prompt screen appears, click the icon (called System icon) where
is in right-top of its window.
4-3) System icon menu appears. Select Properties from the menu. [Fig 1]
Select the Properties item.
[Fig 1]
4-4) Click the “Font” tab from the “ccrstart.bat” properties and change its size to “6 x 13”.
4-5) Click the “Layout” tab and change the “Height” of the “Screen Buffer Size” to 5000.
Click [OK].
4-6) The “Apply Properties to Shortcut” appears and check the item of the “Modify
shortcut which started this window”. Click [OK].
5) Delete CXDI software from the “Startup Menu Group”.
5-1) Connect the keyboard and the mouse to the control computer.
5-2) Turns the all CXDI system power on, Windows XP start.
5-3) After the CXDI software start, press [Alt] + [Tab] key to show the “CCR Console
Menu” prompt screen.
5-4) Select “8 – Exit” to close the CXDI software on the “Welcome to CCR”.
5-5) After the CXDI application software closed and Window XP Desktop appear, open
the “Taskbar & Start Menu…” with “Start Menu” tab clicking.
Start>Settings>Taskbar and Start Menu
44
Page 61
2. Installation
5-6) “Taskbar and Start Menu Properties” appears. Click “Start Menu” tab, and then click
Taskbar and Start Menu Properties⇒Start Menu⇒Classic Start⇒Menu Customize
5-7) The “Remove Shortcuts/Folders” dialog box appears after click the “Remove” button.
And double-click the “Startup folder”
5-8) Remove the “ccrstart.bat” item from it.
5-9) After “Remove” button clicked, the confirmation of deleting file appears. If you are
going to remove it, click “Yes” button.
5-10) After confirm that the “ccrstart.bat” item is removed from “Startup Group”, close all
the application on the desktop and re-login to Windows XP.
5-11) Make sure that the CXDI software will not start automatically after login to Windows
XP. And then shutdown Windows XP, turn the CXDI system power off.
* When the CXDI software is deleted from the Start menu due to repair or other
reasons, be sure to always perform the procedures outlined in “Adding CXDI
software onto the Start menu” and “Changing the window size” when the repair is
complete.
45
Page 62
2. Installation
(6) Identifying the Imaging Units
1) Purpose
In order for the control computer to identify the imaging units connected, the sensor serial
number of each imaging unit is input to the Control computer.
2) Notes
2-1) These operations must always be implemented at the installation stage and when any
of the Imaging Units (sensor) or Control computer (hard disk) has been replaced or
when the combination of equipment has been changed.
2-2) The sensor serial numbers must always be input. If the serial numbers of the imaging
unit and Control computer do not match, “Sensor Unit: Detect Error (-5100)” will be
displayed on starting up the CXDI software. These numbers are the same as what is
input to the PWB-60Di. (Refer to “ Checking the sensor serial numbers”.)
3) Preparations (What to have ready)
Tool keyboard, tool mouse
4) Procedure
4-1) Start the CXDI software.
4-2) Once the normal sensor screen has appeared on the operation unit, use the keyboard to
enter the debugging mode (Use [ALT] + [TAB].).
4-3) “Welcome to CCR” appears. Select “1 Set-Up...”
Field of View Rotation (0:No 1:Yes) [0=0x0]:0
Constant for Exposure Index [-1.000000]: -1.000000
---- Need to re-start program to validate this change.
4-8) When “CCR SETUP MENU” appears, press the [Esc] key to return to “Welcome to
CCR.”
4-9) Select the command “8 - Exit” from “Welcome to CCR” menu to exit the CXDI
software.
4-10) After “Windows XP desktop” screen appears, start the CXDI software again.
* Restart the CXDI software. The screen displays the following message: Alert System Info
Error (-6) A/D board info is updated. Click “OK”.
*CXDI-60G sensor serial number: 1300****
*CXDI-60C sensor serial number: 1710****
46
Page 63
2. Installation
(7) Entering Control Computer Serial Number
1) Purpose
Set the product serial number (Control computer) to the “Device Serial Number” of the
“DICOM header”.
2) Procedure
2-1) Start up the CXDI system.
2-2) After the exposure screen appears on the operation unit, use the keyboard to enter
Debug mode. (Use [Alt] + [Tab].)
2-3) “Welcome to CCR” screen appears. Select the command “1. Set-Up…”
2-4) The “Setting Mode (0: Normal, 1: Expert) [0=0×0]:” is prompted. Select “0: Normal”
2-5) “CCR SETUP MENU” appears. Select the command ”1. System Setup”.
2-6) “CCR Serial Number [0=0×0] : ” appears. Enter the six-digit number indicated on the
naming label of the control computer unit. Press [Enter] key until “CCR SETUP
MENU” appears. [Fig 1]
Enter here
[Fig 1]
2-7) Press [Esc] key after “CCR SETUP MENU” appears to return to “Welcome to CCR”
screen.
2-8) Select “8-Exit”to exit CXDI software.
2-9) This returns you to the Windows NT desktop. Restart the CXDI software, and perform
the procedure from steps 2) to 5). Check that the serial number for the “CCR Serial
Number” item was entered correctly in step 6).
47
Page 64
2. Installation
(8) Table Setup Settings
1) Purpose
Adjust the CXDI operation unit's TABLE SETUP to match the exposure conditions (Xray tube voltage, X-ray tube current, msec or mAs value) of the X-ray generator.
2) Procedure
2-1) Start the CXDI software.
2-2) Open the TABLE SETUP Change window from the Normal Exposure window.
System ⇒ SETUP MENU ⇒ SYS. SETUP ⇒ TABLE SETUP
2-3) Select the tabs to be changed and change the X-ray tube voltage, X-ray tube current,
and msec value data to match the exposure conditions of the X-ray generator.
* See the operation manual for the details of settings.
2-4) After finishing the changes, return to the Normal Exposure window and check that the
TABLE SETUP has been changed.
48
Page 65
2. Installation
(9) Performing the Annotation Settings
1) Purpose
The settings for imprinting the annotation onto the film and the settings of the
characters used for the annotation are performed.
2) Procedure
2-1) Once the normal radiographic screen has started, open the annotation setting screen.
SYSTEM → SETUP MENU → SYS. SETUP → ANNOTATION
2-2) The annotation setting screen now appears. Proceed with the settings that will make it
possible to put the data desired by the user.
* See the operation manual for the details of settings.
49
Page 66
2. Installation
(10) Network Connections
Network settings
1) Purpose
These settings are for connecting the CXDI to the network.
1-1) Set the CXDI’s IP address, subnet mask and default gateway in Windows XP.
1-2) Set the printer and storage output destinations and parameters on the user screen.
2) Checkpoints
2-1) This item involves checking the details of the checks performed on network setting
parameters among the pre-installation inspection details and setting these parameters.
* Refer to “Appendix: Investigation Report” for the pre-installation investigation
details.
2-2) Perform the settings of this item carefully since any errors made in these settings will
make it impossible for connection to be made to the network or the images to be
transmitted properly, etc.
3) Windows XP settings
3-1) Connect the keyboard and mouse to the control computer.
3-2) After turning on the Operation unit’s power and then the Control computer’s power,
start Windows XP.
3-3) The Windows XP desktop screen appears. Right-click the [My Network] icon, and
select My Network Places from the menu.
3-4) When [Network Connection] appears, double click on Local Area Connection (Intel
(R)
PRO/1000 MT Network Connection).
3-5) When Local Area Connection Properties appears, click on the General tab, select
[Internet Protocol (TCP/IP)], and click Properties.
3-6) Based on the pre-install of inspection details set the IP address, subnet mask and
default gateway.
[Fig 1]
50
Page 67
2. Installation
3-7) Upon completion of the setting, restart the Windows XP.
3-8) Check the communication test in the sequence below to verify whether the CXDI is
now part of the network. To check the connections at the TCP/IP level, use the “ping”
command from the command prompt.
Start Æ Programs Æ Command Prompt
When the IP address of the connection destination is “173.17.7.123,” for instance, the
following messages will be repeated.
y If the CXDI has been connected properly:
Pic:>ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=2550
• If the CXDI has not been connected properly:
Pic:> ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Request time out
Request time out
Request time out
Request time out
51
Page 68
2. Installation
4) Set the printer and storage device which serves as the external output destinations.
In this case, one printer and one storage device are set.
4-1) Printer settings
A. Open the output destination setting screen from the user menu.
System Æ SETUP MENU Æ DESTINATION Æ PRINTER
* Up to four printers (2 of which can be used for output at the same time) can be
set.
B. Press the “Printer1” button, and input the following items based on the pre-
installation investigation details.
a. Printer host name (IP address) b. Port number c. Transmission destination title
a
b
c
[Fig 2]
C. Press the “SET” button, and input the parameters of the printer to be connected
based on the pre-installation inspection details. (Refer to another sheet for details
of the parameters.)
* A space delimiter must be input between each of the parameters.
By pressing the “Override” button, you can select a printer from all the registered
printers. In this case, basically you do not have to enter parameters. However, if
“?” is displayed within the parameters, you may have to enter the required
parameter at the user’s site.
Input parameters
[Fig 3]
52
Page 69
4-2) Storage settings
A) Open the output destination setting dialog from the user menu.
System Æ SETUP MENU Æ DESTINATION Æ STORAGE
* Up to four storage units (2 of which can be used for output at the same
time) can be set.
B) Press the “Storage1” button, and input the following items based on the pre-
installation investigation details.
a. Storage host name (IP address) b. Port number c. Transmission destination
title
2. Installation
a
b
c
[Fig 4]
C) Press the “SET” button, and input the parameters.
(Normally, the parameters need not be set. They must be input only when the
need arises.)
* A space delimiter must be input between each of the parameters.
Input parameters
as needed
[Fig 5]
53
Page 70
2. Installation Manual
D) After setting the output destinations, follow the procedure below to check
whether images can actually be transmitted. Return to the user menu, capture
a sample image (one X-ray image), and transmit the image to the printer and
storage. There are two errors that may result if the image cannot be
transmitted:
a. “DICOM Connect Error. Cannot connect to the target. Check network or port
number setting. Retry?”
b. “DICOM Transfer Error. Error occurred during the association. Retry?”
Message (a) indicates that connection at the TCP/IP level is not possible and that
the physical connections or the subnet mask and other settings must be checked
again.
Message (b) indicates that communication at the TCP/IP level is problem-free but
that DICOM level communication has failed. In this case, check again that
AE_TITLE of CXDI has been sent properly to the transmission destination and
that the IP address, port number and AE_TITLE of the transmission destination
which are set with CXDI have been set properly.
* “AE_TITLE” of the transmission destination is case sensitive fields.
(Permit upper-case letter or lower case letter, etc)
54
Page 71
Parameter List (Separate Document 1)
2. Installation
DICOM storage device
The parameter settings for DICOM Storage transfer used in CXDI are described below.
Parameter Meaning Description
-m maxPDU
Maximum PDU value in
byte units
-t calledTitle
Called App Entity Title
-c callingtitle
calling App Entity Title
-s SOPName
(for reference)
This parameter designates
whether class be connected
for performing association
at the beginning of
transfer.(CR/T/MR/NM/S
C/US)
-I
A-RELEASE-RES is
ignored.
-d FAC
This parameter dumps a
specific facility log.
(DCM/DUL/SRV)
* The CXDI automatically
uses 131072 internally for
operation.
* Designating a specific value
allows overwriting of the
above value.
* The AE Title setting field is
automatically applied to this
setting.
* Designating a specific value
allows overwriting of the
above value.
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* Designating a specific value
allows overwriting of the
above value.
* This is not used in the
CXDI.
* This parameter is used
simply as “-I”
* This parameter is used
simply as “-d”
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This is used when the operator who
manages the connected storage device
requests a size change.
* In DICOM printing, note that the
argument title changes to -u.
(Æ See the printing parameters.)
* Note that the meaning is opposite of the
DICOM printing argument –t.
(ÆSee the printing parameters.)
* The entry for the OPU output device
title is used here.
* Note that the meaning is opposite of the
DICOM printing argument -c.
(ÆSee the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
* This is used when the error message
“130012 Peer aborted Association (or
never connected)” occurs even though
the DICOM data transfer was
successful.
Æ This is used differently based on the
connected storage devices.
* This parameter does not affect DICOM
data transfer.
* This parameter outputs the CXDI log
based on Windows NT.
55
Page 72
2. Installation
Parameter Meaning Description
-v
This parameter dumps the
transfer log.
-jn
This is the time to take
timeout.
-k
level=0
1
2
3
* This parameter is used
simply as “-v”.
* DUL and SRV are dumped.
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* Sets the time to take timeout
in seconds.
* A variety of specifications
have since been needed in
conjunction with DICOM
modality LUT support.
* This parameter does not affect DICOM
data transfer.
* This parameter is to be changed when
taking timeout.
If DICOM Modality LUT OD is enabled,
set appropriate options to suit each
output destination.
Argument: Values necessary for executing a function, subroutine, procedure, or other operation is
passed to them. Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (x) is x.
PDU: Protocol data unit
The types of PDU’s include get-request, get-next-request, get-response, set-request, nd trap.
Note
For details about the “-v” parameter, see “Checking the Error Log”.
Note
The parameters “-v” and “-d” put the log on the console. Therefore, be sure to always erase
these parameters before operation by the user.
If –k:DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
level = 0: Do not delete (default when not specified)
1: Delete Window Center/Width
2: Delete Window Center/Width and Rescale Intercept/Slope/Type
3: Delete Rescale Intercept/Slope/Type (compatible with releases up to Ver.4.20)
Other than a loadable LUT or γ=1.0 has been specified with the output destinationspecific LUT function.
For storage: Specify -k3(or -k2).
If IMG Rescale Type = 0D is unidentifiable to storage, resulting in an error: Specify -k3.
(If only one storage is connected, simply set DICOM Modality LUT OD to "Disabled.")
The implementation allows Window Center/Width with Rescale Type = 0D specified to
be interpreted as "optical densityx1000."
For storage: Specify -k1.
The CXDI Window Center/ Width output value (implementation) is fixed at 2048/4096.)
Note
With AGFA impax Ver. 4.5.0, the specification of -k0 demonstrated a successful densityintensity conversion. But because the corresponding text in the DICOM specifications
document is ambiguously written such that the status of implementation by other
manufacturers is unknown, please be advised to consult the storage manufacturer for each
connection destination or work out on a trial and error basis. Also note that an external
storage option specification (if (DICOM modality LUT OD is set to "Enabled," OD tags
<0028, 1052 - 1054> are assigned and Level: 0 is assumed) is not supported.
56
Page 73
Parameter List (Separate Document 2)
2. Installation
DICOM printer
In the CXDI, DICOM printers are administered separately according to printer product.
The parameter settings for DICOM printer used in CXDI are described below.
Parameter Meaning Description
-C copies
This parameter uses a
number to designate the
number of copies.(1/2/...)
-y priority
Priority in the DICOM
printer
(HIGH/MED/LOW)
-D destination
Film destination
(MAGAZINE/PROCESS
OR/BIN_i)
-F film type
Film media type (“BLUE
FILM” / “CLEAR FILM”
/ “PAPER”)
* This parameter is used in the
DICOM Basic Film Session
(2000, 0010).
* When the number of copies
is designated, film sheets are
printed in the quantity
specified in a single printing
operation.
* This parameter is necessary
when printing multiple
sheets for a single data
transfer operation.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0020).
* This parameter determines
where this transfer image is
inserted into the queue in the
DICOM printer. At HIGH,
the image is printed first
among the queued images.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0040).
* Film is sent to the output
device designated by
RECEIVE MAGAZINE or
the automatic developer.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0030).
* Film is printed as the
designated film type.
* This parameter is set according to the
user’s requirement.
* In the DICOM library TYPE3, the value
is transferred together with the Tag.
However, if the value is unknown, the
value is either is transferred as a
character string with length 0, or the
element itself is not transferred.
Æ The printer default values are used if
this parameter is not entered.
* This parameter is set according to the
user’s requirement.
* In the DICOM library TYPE3, the value
is transferred together with the Tag.
However, if the value is unknown, the
value is either transferred as a character
string with length 0, or the element itself
is not transferred.
Æ The printer default values are used if
this parameter is not entered.
* Note this parameter does not determine
where this transfer image is inserted into
the CXDI queue.
Æ The printer default values are used if
this parameter is not entered.
* The film is usually discharged to the
default output device.
Æ The printer default values are used if
this parameter is not entered.
* Although many types of films cannot be
detected, the film type can be selected in
the KELP2180.
* In this parameter, be sure to put
quotation marks (“ ”) around 0x20 since
it comes between BLUE and FILM.
57
Page 74
-L session Label
Film session label
(character string)
-f films
Number of film box to be
printed
–i Format
Format at print
-l FilmSizeID
Film size
14 inch x 17 inch / 17 inch
x 14 inch / 11 inch x 14
inch / -l FilmSizeID / 10
inch x 14 inch
10 inch x 12 inch / 24 cm
x 24 cm / 24 cm x 30 cm
Æ The parameter is not transferred over
DICOM if it is not designated.
* This parameter may be displayed in
some form or another depending on the
installed printer. For example, it may be
displayed in the Control Panel for the
printer or in the corner of the film.
* If this parameter is not designated,
transfer software uses automatically
STANDARD1 1, for reason this
parameter must be transferred in the
DICOM.
Æ The parameter is not transferred over
DICOM if it is not designated. In this
case, the printer default values are used.
When this parameter is not transferred,
problems can occur since unsuitable
default values may be used.
* Some printers do not print until a supply
magazine of the designated size is
loaded, and others print even though the
designated size is different from the
currently loaded supply magazine.
Æ The printer default values are used if
this parameter is not entered. When this
parameter is not transferred, problems can
occur since unsuitable default values may
be used.
Æ The parameter is not transferred over
DICOM if it is not designated. In this
case, the printer default values are used.
When this parameter is not transferred,
problems can occur since unsuitable
default values may be used.
* This parameter is determined by asking
the printer engineer or by viewing the
conformance statement.
58
Page 75
-S configuration
Adjustment information
(character string)
-O Orientation
Film orientation
(PORTRAIT/LANDSCAP
E)
-A max_density
Maximum density
(Dx100)
–a min_density
Minimum density (Dx100)
2. Installation
* The designation method
varies according to the
printer. For example, the
MLP190 uses -m
NORMAL.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0150).
* This parameter sets the
printer (image quality)
adjustment from the SCU
side.
* In the DICOM standards,
value to be transferred is not
predetermined.
* The designation method
varies according to the
printer.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0040).
* In versions before 2.0,
printers must operate based
on this parameter.
* When using image cutout
from 17 x 17 inch size in the
CXDI, this parameter is set
and transferred
automatically.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0130).
* This parameter designates
the density of the digital
value for 0 (4095 for reverse
display) of the CXDI
transfer data image pixels.
(In the CXDI, 0 indicates
black.)
* In the CXDI, this parameter
is used to adjust the density.
Therefore, be sure to always
check that it is operating.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0120).
* This parameter designates
the density of the digital
value for 4095 (0 for reverse
display) of the CXDI
transfer data image pixels.
(In the CXDI, 4095
indicates white.)
* This parameter is not
transferred in many cases
since the minimum density
cannot be increased in most
printers.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
* This parameter is determined by asking
the printer engineer or by viewing the
conformance statement.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used.
* Starting from version 2.0, the image can
be rotated from the CXDI side without
using this parameter.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
59
Page 76
-B border_density
Border density (Dx100)
(BLACK/WHITE/D x
100)
-G empty_image_density
Empty image density
(BLACK/WHITE/D x
100)
-T trim
Trimming (NO/YES)
-P polarity
Polarity
(NORMAL/REVERSE)
-r pixel_pitch
Transfer pixel pitch for
designating the request
image size (Pixel Pitch in
um)
-N annoFmt
Annotation position
(1/2/3)
-n annotation
Annotation (character
string)
2. Installation
* This parameter is used in the
DICOM Basic Film Box (2010,
0100).
* This parameter determines the
area density around the image
on the film.
* This parameter is used in the
DICOM Basic Film Box (2010,
0110).
This parameter designates the
density of the empty image
area during multi-formatting.
* This parameter is used in the
DICOM Basic Film Box (2010,
0140).
* This parameter adds lines
around the image.
* The CXDI is normally adjusted
so that the trimming does not
appear.
* This parameter is used in the
DICOM Basic Image Box
(2020, 0020).
* Reverse image density
* This parameter is used in the
DICOM Basic Image Box
(2020, 0010).
* Position of the image on film
* The cumulative value for the
horizontal size of the image
(raw) at the designated pixel
pitch is used for the request
image size.
* The CXDI automatically uses
131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0010).
* This parameter designates the
position of the character string
to be annotated.
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0020).
* This parameter designates the
character string to be
annotated.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
* For example, the default value for
this parameter in Agfa printers is
YES. Therefore, the NO option
needs to be specifically designated if
it is desired.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the type of image that is
printed depends on the settings at the
printer side.
* If using annotation, always be sure to
transfer the annotation position.
* If this parameter is not designated, it
is not transferred over DICOM. In
this case, the printer default values
are used.
* Also, in this case, the type of image
that is printed depends on the
settings at the printer side.
60
Page 77
-u maxPDU
Maximum PDU value in
byte units
-c callingTitle
Called App Entity Title
-t callingTitle
Calling App Entity Title
-g
N-GET Printer
compatibility mode
-s
Silent mode
-p
This parameter dumps the
association parameter.
-v
This parameter dumps the
transfer log.
2. Installation
* The CXDI automatically
uses 131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
* The AE Title setting field is
automatically used in this
setting.
* The above value can be
overwritten by designating a
specific value.
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* The above value can be
overwritten by designating a
specific value.
*This parameter is used
simply as “-g”.
* In the CXDI default settings,
the printer information is not
designated. In this case, the
printer side sends all the
information that it has
(DICOM official
specifications).
* When the -g option is added,
the essential information
only is collected. This
information includes the
Printer Status and Printer
Status Info. (To prevent
installation when the printer
does not satisfy the above
DICOM specifications.)
* This parameter is used
simply as “-s”.
* This parameter is used to
prevent the transfer software
from displaying the
debugging character string
on the console.
* This parameter is used
simply as “-p”.
* This parameter is used to set
the transfer software so that
the debugging character
string is displayed on the
console.
* This parameter is used
simply as “-v”
* This parameter is used to set
the transfer software so that
the debugging character
string is displayed on the
console.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This parameter is used when the
operator of the connected storage device
requests a size change.
* In DICOM storage devices, note that the
argument title changes to -m.
(ÆSee the storage device parameters.)
* Note that the meaning is opposite of the
argument -c for DICOM storage
devices.
(ÆSee the storage device parameters.)
* The entry for the OPU output device
title is used here.
* Note that the meaning is opposite of the
argument -t for DICOM storage devices.
(ÆSee the storage device parameters.)
Æ Normally, this option is not used.
This parameter has been provided as a
remedy when a printer error occurs when
optional devices are not used.
* This parameter does not affect DICOM
data transfer.
Æ Silent mode does not need to be
designated since the CXDI automatically
makes the setting internally.
* This parameter does not affect DICOM
data transfer.
* This parameter does not affect DICOM
data transfer.
61
Page 78
2. Installation
* Both the -p and -v
parameters should be used.
These settings override the -
s parameter.
-V filename
This parameter dumps the
transfer log.
-I
A-RELEASE-RES is
ignored.
-jn
This is the time to take
timeout.
-k
level = 0
1
2
* The parameter is used for
analysis after the transfer
software saves the
debugging character string
displayed on the console to a
file with a designated
filename.
It is used only when
problems occur.
* This parameter is used
simply as “-I”
* Sets the time to take timeout
in seconds.
* The -k option has been
implemented to normalize
DICOM headers
* This parameter does not affect DICOM
data transfer.
* This is used when the error message
[130012 Peer aborted Association (or
never connected)]occurs even though the
DICOM data transfer was successful.
Æ This is used based on the connected
printers.
* This parameter is to be changed when
taking timeout.
Note
The parameters -p, -v, and -V filename display the log on the console. Therefore, be sure
to always erase these parameters before operation by the user.
In the past, -k2 was used to fix troubles, but it has now been set as the default has been changed
to -k2 to ensure precise compliance with the DICOM code. At sites wishing to adhere to their
existing window values, -k1 should be used.
level: 0:Do not delete DICOM tags. (Transfer all headers similar to storage.)
1:Delete Groups 0008, 0010, 0018, 0019 and 0020.
2:Delete Groups 0008, 0010, 0018, 0019, 0020 and Elements (0028,0030), (0028,1050),
Image Pixel Spacing (0028,0030)
Window Center (0028,1050)
Window Width (0028,1051)
Rescale Intercept (0028,1052)
Rescale slope (0028,1053)
Rescale Type (0028,1054)
For groups 0008, 0010, 0018, 0019, 0020, refer to DICOM manifesto.
62
Page 79
2. Installation
Examples of parameters used with different makers and types of printers (reference)
Name Default Parameters Significance
Kodak MLP190 -A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
80 (entered in pixelPitch
member)
14 x 17
4096 (entered in W member)
5120 (entered in H member)
Kodak KELP2180 +
Kodak Print Spooler
Model 100
Agfa DryStar 3000 -A 320 -T NO -M CUBIC -m
Kodak Imation DryView
8700 + Pacs LINK IMN
9410
Kodak Imation DryView
8700 + GW
Kodak Imation DryView
8700+8800
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
79 (entered in pixelPitch
member)
14 x 17
4090 (entered in W member)
5120 (entered in H member)
11 x 14
3194 (entered in W member)
4096 (entered in H member)
140 -S
“PERCEPTION_LUT=200”
(entered in param member)
80 (entered in pixelPitch
member)
14x17
4256 (entered in W member)
5174 (entered in H member)
-A 310 -T NO -M CUBIC -m
? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
-A 310 -T NO -M CUBIC -m
? -S LUT=?, ? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
• Trimming OFF
• Cubic spline interpolation
• Smoothing: normal
• Maximum density: 3.20
• Curve shape 0 (density linear) as Config
Info
• Trimming OFF
• Cubic spline interpolation
• Smoothing: normal
• Maximum density: 3.20
• Curve shape 0 (density linear) as Config
Info
• Trimming OFF
• Cubic spline interpolation
• Smoothing: slightly sharp (edges
emphasized)
• Maximum density: 3.20
• S
“PERCEPTION_LUT=200(LINEAR)”
(If the output fails to be linear with
“LINEAR”, on-site adjustments with the
printer manufacturer must be performed.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
63
Page 80
2. Installation
• The Kodak service engineer is
responsible for setting the smoothing
type on-site since it can be changed with
each printer.
• The rest is done by the printer itself.
Kodak Imation DryView
8700+9440
Kodak Imation Dry View
8500+
Nishimoto EL2000N -A 320 -T NO -M CUBIC -S
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
The model 8800 has a rotation
function but we understand
that Imation has not publicly
acknowledged the use of this
function.
-A 310 -T NO -M CUBIC -m
? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
3388 (entered in W member)
4277 (entered in H member)
15
(entered in param member)
80 (entered in pixelPitch
member)
4444 (entered in portraitW
member)
5296 (entered in portraitH
member)
5296 (entered in landscapeW
member)
4444 (entered in landscapeH
member)
“Prepare images using CXDI”
: Yes
A simple calculation yields a
resolution of 4444 x 5400 for
the display area of the model
EL2000. However, 5376@80
µm is set in the perpendicular
direction of the model EL2000
since the maximum size of the
CXDI images is 2688@160
µm. In this case, a small nonimage area should be output as
the border in the up/down
direction on the film according
to the calculation. In actual
fact, however, the image will
protrude in some cases
depending on how the
transport speed and other
factors have been adjusted.
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• The Kodak service engineer is
responsible for setting the smoothing
type on-site since it can be changed with
each printer.
• The rest is done by the printer itself.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• -S
• LUT = m, n is designated as the Config
Info but m and n are adjusted on-site by
the Kodak service engineer. Basically,
adjustment is performed to achieve a
linear output.
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• “15” in Config Info is linear.
• The rest is done by the printer itself.
Up to 5376 pixels can be set for H.
64
Page 81
2. Installation
The specifications are as
follows although they cannot
be set at the present time:
14" x 14" : W=4444, H=4444
(for portraits)
11" x 14" : W=4444, H=3660
(for portraits)
Fuji CR-DPL/LPD/FMDPL + FN-PS551
Konica Drypro 722 +
Printlink
Konica Li-62P + Printlink -A 320 -T NO -M CUBIC -m
-A 300 -T NO -M CUBIC -m
MEDIUM -S –P NORMAL –
B BLACK –k 2 –S ? (Site
Dependent)
(entered in param member)
14x17
3520 (entered in W member)
4280 (entered in H member)
14x14
3520 (entered in W member)
3490 (entered in H member)
11x14
2540 (entered in W member)
3600 (entered in H member)
-A 320 -T NO -M CUBIC -m
2 -S “KC_LUT=1” –O
PORTRAIT –P NORMAL –B
BLACK
(entered in param member)
80 (entered in pixel pitch
member)
14x17
4424 (entered in W member)
5324 (entered in H member)
14x14
4424 (entered in W member)
4372 (entered in H member)
11x14
3436 (entered in W member)
4424 (entered in H member)
2 -S “KC_LUT=1” –O
PORTRAIT –P NORMAL –B
BLACK
80 (entered in pixel pitch
member)
14x17
4268 (entered in W member)
5108 (entered in H member)
14x14
4268 (entered in W member)
4104 (entered in H member)
11x14
3204 (entered in W member)
4268 (entered in H member)
• Trimming OFF
• Cubic spline interpolation
• SHARP, MEDIUM or SMOOTH can be
selected from among the presettings as
the smoothing type. An AVR of 0.8 or
so is appropriate. The setting is
performed for each printer on-site.
• -S should be adjusted at the user’s site.
• LUT can be selected from among the
eight presettings 1 through 8 using
Config Info. The setting is performed
for each printer on-site.
• With -k 2, the Window Center/Level for
DICOM TAG (0028,1050) and (0028,
1051) are also deleted.
• Maximum density: 3.00
A density of 3.20 cannot be designated.
For this reason, a non-linear LUT is
required.
• Trimming OFF
• Cubic spline interpolation
• Smoothing type
1: BILINEAR
2: Sharp by spline interpolation
3: Slightly weak by spline interpolation
4: Weaker by spline interpolation
• Maximum density: 3.20
Maximum density 3.20 could not be
achieved before.
• Trimming OFF
• Cubic spline interpolation
• Smoothing type
1: BILINEAR
2: Sharp by spline interpolation
3: Slightly weak by spline interpolation
4: Weaker by spline interpolation
• Maximum density: 3.20
Before, maximum density 3.20 could
not be achieved.
65
Page 82
2. Installation
Printer Model Specifications (Reference)
Name Specifications
Kodak MLP190 80µm x 4096 x 5120 2048 x 2560 (@160 µm)
Kodak
KELP2180 +
Kodak Print
Spooler Model
100
79 µm x 4090 x 5120 (value after passing through the
print spooler)
• The above settings are the size of the effective area
when the image passes through the print spooler and
the image is plotted up to the annotation area. In other
words, these settings do not display an annotation area,
instead handling it as an image area.
• If the data is transferred without setting [Image creation
in CXDI], the Requested Image Size setting is used in
DICOM. In this case, the annotation area is
automatically displayed in the 2180 printer. As a result,
a maximum image area of 79 µm x 4090 x 4996 must
be designated. In this case, the CXDI relies on the 2180
for image rotation (Film Orientation), but images larger
than 1.7 MB cannot be rotated by the 2180. Therefore,
operation without the setting for [Image creation in
CXDI] cannot be performed in the 2180. (Although
operation is possible by setting Requested Image Size
only for using DICOM without the setting for [Image
creation in CXDI], this option is not installed in the
CXDI.)
• Although the resolution of the printer itself is 79 µm x
4090 x 5260, this complete resolution cannot be used
when the image passes through the spooler.
When “_” is used in AE Title, the association is
rejected.
Use the Disable function for N-EVENT-REPORT to
disable this setting.
The FilmSize parameter can be used. The MediaType
(BLUE, CLEAR) parameter is also supported.
• Although the Film Orientation parameter is supported
up to 1.7 MB, in actuality, DR images cannot be
rotated. Like the DryView8700, the images must be
rotated by the CXDI side.
• When Requested Image Size is expanded, the
maximum plotting size is limited (79 um x 4090 x
4996) so that the annotation area can be obtained.
When a Requested Image Size expansion error occurs,
the image is interpolated and printed at the suitable
size. In this case, the error does not return to the CXDI
side. For example, the image is printed at 310 mm even
if 326 mm is designated.
• If an expansion error occurs in the spooler when Multi
Display Format is used, the print queue cannot be
processed.
• The system is in a critical state when a Failure status is
indicated. A user message is displayed indicating this
state, and images are no longer transferred. (Fully
installed)
• During the Warning status, image transfer is performed
while the user message is displayed (Fully installed).
• If operation is aborted due to an error, a new
association could not be established when the data was
resent from the CXDI. GW was reset to recover the
error.
Maximum equivalent
area in CXDI
2018 x 2528(@160 µm)
66
Page 83
Agfa DryStar
3000
Imation DryView
8700+8800
Fuji CR-DPL
/FM-DPL + FMPS551
2. Installation
• Inserting annotations in the image can lead to problems
at the hospital. In the QCW, use annotations that are
outside of the image.
• The designated film size is 11 x 14 inch film, and
automatic selection of the magazine and printing has
been confirmed.
• Annotations in the image are problematic in the US and
EU. Thus, although DICOM annotation was used, it
was not printed. Although annotation can be transferred
without any errors in DICOM, an annotation error
occurred in the log when transferring from the gateway
to the 2180, and printing was not performed. Data was
transferred from KCR to the validation tool, and the
DICOM transfer method was compared to CXDI.
However, the only differences were in the image size,
aspect ratio, and annotation position.
• Although the annotation position is 1 in terms of
conformance, it is 0 as sent from KCR. The result did
not change even after the change.
80 µm x 4256 x 5174
• In the standard Agfa system, the annotation option is
selected. Therefore, the full image area will not be
printed correctly unless you ask the serviceman to
deselect the annotation option.
78 µm x 4096 x 5220
Note: The 8700 printer cannot display in landscape
orientation. Also, the maximum density is 3.1. If the 8800
box is not added, this printer cannot be used by the
CXDI. However, it can be used starting from CXDI
version 2.0.
• If there is an Imager Pixel Spacing tag (0018, 2264), the
imager will fail.
100 µm x 3520 x 4280 (value after passing through the
print spooler)
• The above settings are the allowable area size in a
configuration not using annotation.
• Annotation will be supported from the next version.
Annotation in currently possible in US-ASCII only. IDs
are designated 1 to 6 and correspond to the top left, top
center, top right, bottom left, bottom center, and bottom
right, respectively. The maximum area size with
annotation support is 3500 x 4170 for 35 cm x 43 cm
and 2538 x 3522 for B4.
• Use the Disable function for N-EVENT-REPORT to
disable this setting.
• A function is provided for disabling the returning of
warning messages.
0107 (Attribute list error)
Return/Not return
0116 (Attribute Value out of Range)
Return/Not return
B604 (Image has been demagnified)
Return/Not return
• The Film Size parameter can be used. The Media Type
(BLUE, CLEAR) parameter can also be used.
• The Film Orientation parameter is fully supported.
• The following presets are made so that LUT has the
2128 x 2587 (@160 µm)
1996 x 2544 (@160 µm)
2200 x 2675 (@160 µm)
67
Page 84
2. Installation
DMAX = 3.2D equivalent curve required by CXDI.
Gamma type #17 (SAR system)
Density Shift Contrast
Point 1
1.57 0.10 1.00
Point 2
2.29 0.15 1.00
68
Page 85
2. Installation
Error Return Values and Log Output for print_stuff (Reference)
The log output in this case is shown below.
18-135933[d2]ERR: d0012 Attempt to connect to unknown host: test
18-135933[d2]ERR: 130012 Peer aborted Association
(or never connected)
18-135933[d2]ERR: 180012 Failed to establish association
Return value: 0x18
CXDI description : DICOM response warning
(CCRTRANS_WRN_RESP)
The transfer process was successful, but a warning was returned from the server.
The log output in this case is shown below.
Other errors Return value: Values other than those above
Return value: 0x28
CXDI description: DICOM printer status warning
(CCRTRANS_WRN_PRN_STATUS)
The transfer process was successful, but a warning was returned as the printer
status.
The log output in this case is shown below.
CXDI description : DICOM communications error
(CCRTRANS_ERR_DICOM_TRANSE)
The log output in this case depends on the specific error. A typical example is
shown below.
Transmitting the 0019 shadow group causes a failure, and the group is not received
properly with the default. Its reception is enabled by setting the strictValidation
parameter to Off in Miil.
y Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series under the Dr.ABLE specifications, the switching operations are a
hassle. The user will find it more convenient if it is at all possible to change the
setting to “1 study 1 series/1 series multi image.” (These unusual data
specifications were requested with the full understanding of their unusualness.)
y With DICOM, the body parts (such as the abdomen and head) and their directions
(such as PA and AP) belong to the series information. It therefore follows that a
different series is required for a different body part or body part direction. This
aspect is restricted by the DICOM standard rather than by the installation and
other steps taken by us. To put it the other way around, multiple images with
different body parts and their directions cannot be put together as a series. To
remedy this problem, devising a way of enabling the viewers to reference
different series at the same time at some future point in time will be helpful.
y Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series, the switching operations are a hassle. The user will find it more
convenient if it is at all possible to change the setting to “1 study 1 series/1 series
multi image.”
y This problem arises with the Fujitsu equipment as well. Refer to the section on
Fujitsu.
send_image error return values and log output (reference)
Example of
Return value and log output
error
Successful Return value: 0x00
No log output at this time
Invalid parameter
exists.
Server fails to start. Return value: 0x00180012
As a result of
providing the
server with a
command request,
an error was
returned as
response.
• Image adjustment parameters that can be changed by users
Density : This can be set up to the maximum density of 3.1D.
Contrast : This can be set from 1 to 15.
• Image adjustment parameters that cannot be changed by users
Lookup tables called TFTs (Transfer Function Tables) are provided internally, and changes
can be made only in the service mode.
Fifteen types of characteristic curves are registered in one of these TFTs, and users can change
one of these curves as the contrast.
Over 30 TFTs are registered in the printer, and a name is allocated to each one.
For instance, 15 types of linear straight lines are registered in the TFT called “WRKST2A.”
By setting this WRKST2A TFT, adjustments can be made by combining 15 types of linear
straight lines (which cannot be changed by users) with 15 types of characteristic curves
(which can be changed by users).
Concerning connections
Two types of the Dry View 8700 are available.
• Dry View 8700 Plus
The 8700 Plus is a printer which can be connected to two diagnostic units. When used in
combination with the 8800 multi input manager, it can be connected to up to eight diagnostic
units. Images are processed as described above.
• Dry View 8700 Dual
It is possible to connect two 8700 Dual units to the 8800 multi input manager. By using these
in combination, up to seven diagnostic units can be connected. The 8700 Dual does not come
with image processing functions
Modality connection I/F
This I/F is the external interface which connects “Dry View 8700” with each modality. Select
the following item depending on the modality which would be connected.
• Digital Signal: DEIB (Digital External Interface Box)
• Video Signal: EVEIB (Enhanced Video External Interface Box)
• Keypad, Auto Filming: UKEIB (Universal Keypad External Interface Box)
73
Page 90
2. Installation
Dry View 8700 Plus
Up to 2 units can
be connected.
Dry View 8700 Plus + 8800 Multi Input Manager
Up to 8 units can
be connected.
Dry View 8700 Dual + 8800 Multi Input Manager
Up to 2 printers can be connected
(only with Dry View 8700 Dual)
Up to 7 units can
be connected.
74
Page 91
2. Installation
(11) Linearity Check of Transfer Image Density
1) Purpose
An SMPTE image is used to check whether the density linearity of the image printed
out by the printer and the image displayed on the high-definition monitor matches the
density linearity of the image transferred by the CXDI.
2) Notes
2-1) This checking procedure should be performed before the procedure in the section of
“Operation Unit Gamma Correction”.
2-2) The adjustment and checking procedures below should be completed before
performing this procedure.
A) The printer and high-definition monitor connections and setting adjustments
should be completed. The printer and monitor image output settings should be set
to LINEAR.
For example, if the KODAK MLP190 is connected, the printer parameter
“-S configuration” must be set to “-S CS000”. For other printers, refer to
“Printers and Parameter Examples Reference” in “Network Connections”, and set
so that the curve shape is 0 (density linear).
B) Be sure to calibrate the printer and high-definition monitor units separately before
performing this procedure.
C) When the imaging screen is “tray type”, change it to “category type” by selecting:
System Æ SETUP MENU Æ CUSTOMIZE DISPLAY.
3) Rough adjustment
3-1) Start up the CXDI system.
3-2) Using the two knobs at the rear of the operation unit, adjust the brightness and
contrast of the touch panel screen for optimum visibility.
3-3) On the exposure screen, select the exposure mode “SMPTE” and wait until “READY”
appears. [Fig. 1]
[Fig. 1]
75
Page 92
2. Installation
3-4) Press the exposure switch on the X-ray generator, and after the exposure, press the
END STUDY button. Transfer the SMPTE pattern image to the printer or the highdefinition monitor. [Fig. 2]
[Fig. 2]
3-5) Measure the densities of the 11 locations (0% to 100%) of test image grayscale on the
film or on the monitor. [Fig. 3]
* Measure the SMPTE image density on the film is measured using a densitometer.
Measure the SMPTE image density on the high-definition monitor using the
gradation analysis software.
* The data for the SMPTE test image grayscale transferred by the CXDI are the values
for the maximum density (3.20 in the case of the MLP 190) in the printer settings
which have been changed in 11 uniform steps.
[Fig. 3]
76
Page 93
2. Installation
3-6) Create the graph below based on the data measured in step 5).
3.50
3.00
Max setting
density
2.50
2.00
1.50
Density
1.00
0.50
0.00
Film base
density
1234567891011
Step
As shown in the above graph, the measurement values need only to nearly form a straight line
from the minimum density to the maximum density.
The important point here is that the measurement values make a straight line and a maximum
density corresponding to the settings is output.
If the measurement values deviate too far from the straight line and a maximum density
corresponding to the settings is not output, printer and monitor output linearity settings,
calibration, and other adjustments are necessary.
* The above graph is an example of measurements when the maximum density is set to 3.2 D
and the printed film is measured with a densitometer. Refer to the data below when the
maximum density is set to 3.1 D, or when gradation analysis software is used to take
measurements on the monitor.
* The LINEAR output cannot be set for some printer models.
In this case, try to select an output setting as close as possible to LINEAR.
Step1234567891011
3.2 D00.320.640.961.281.601.922.242.562.883.20
3.1 D00.310.620.931.241.551.862.172.482.793.10
12 Bit041081912291638204824572867327636864095
8 Bit0265177102128153179204230255
Ideal density values in LINEAR LINE
Possible causes of non-linearity in measurement values:
* Inadequate calibration of printer and high-definition monitor
* Inadequate settings for CXDI printer parameters
* Faults in printer or high-definition monitor
77
Page 94
2. Installation
4) Fine adjustment
4-1) As the step (3)-3) above, make the system “READY” by selecting the exposure mode
“SMPTE” on the exposure screen. [Fig. 4]
[Fig. 4]
4-2) On the X-ray generator press the exposure button. After the exposure, adjust the trim
so that the gray scale may be located in the center of the image. [Fig. 5]
* For correct density measurement of the 32-step chart, the gray scale must be printed
in the center of an image to eliminate the shading feature.
* Data may not from a straight line near the minimum and maximum densities due to
characteristics of the printer. Rotate the image on the QA screen, and reprint or retake
measurements.
[Fig. 5]
78
Page 95
2. Installation
4-3) Select the END STUDY and transfer the SMPTE pattern image to the printer or the
high-definition monitor. [Fig. 6]
[Fig. 6]
4-4) On a printed film or on the monitor, measure the density of 32 steps of the grayscale
on the test image. [Fig. 7]
*1: Measure the SMPTE image density on the film using a densitometer. The SMPTE
image density on the high-definition monitor is measured using the gradation
analysis software.
*2: The data for the SMPTE test image grayscale transferred by the CXDI are the
values for the maximum density (3.20 in the case of the MLP 190) in the printer
settings which have been changed into 32 uniform steps.
Step1
Step32
[Fig. 7]
79
Page 96
2. Installation
y
4-5) As the step (3)-6), create a graph based on the data measured in step 4), and make sure
that the data from the minimum density to the maximum density nearly form a straight
line.
Characteristics of the printer may prevent the data from forming a straight line near
the minimum and maximum densities. Rotate the image on the QA screen, and reprint
or retake measurements.
In the data and graph below, the data was obtained by setting the Kodak MLP 190 to curve
shape 0 (density linear), a maximum density of 3.20, the SMPTE test image was printed out,
and the image data was measured.
As shown in the graph, the data from the minimum density (film base density) to the
maximum density nearly form a straight line.
* For improper parameters, the printer parameter “-S CS000” was not entered.
* For the inadequate calibration, the printer calibration data was set too low.
82
Page 99
2. Installation
(12) Operation Unit Gamma Correction
1) Purpose
This procedure is performed so that the image that is printed out or displayed on a highdefinition monitor conforms exactly to the exposure image on the operation unit.
2) Notes
2-1) The procedure in “Linearity Check of Transfer Image Density” must be completed.
2-2) If image adjustment for the printer or high-definition monitor has not been made, use
the “Gamma Correction Calculation Tool” in order to correct the gamma of high
definition monitor image to be a same as printer image.
2-3) Gamma correction is an image correction process for monitors and film. It is different
from the contrast setting or grayscale setting.
This procedure is simply a visual adjustment. As a result, differences may occur
depending on the operator performing the procedure. Therefore, be sure to consult
with the responsible technician before performing this adjustment.
2-4) To make the gamma of operation unit adjust in detail, use the “Gamma Correction
Calculation Tool”.
3) Preparation
The Option button used in gamma correction is normally hidden. Edit the MenuPara.ini
file to display.
3-1) Open MenuPara.ini file.
3-2) Make the following changes, and overwrite the file.
Use Search in Edit to find OptionDlgBth.
Select Customize.
OptionDlgBtn = 0 Í Change to 1.
(0: Don’t display; 1: Display)
83
Page 100
2. Installation
4) Comparison of Operation Unit Image and Print Image or Monitor Image.
4-1) Start up the CXDI system.
4-2) Use the two adjustment knobs at the rear of the operation unit to adjust the brightness
and contrast of the touch panel screen for optimum visibility.
4-3) On the exposure screen, select the exposure mode “SMPTE” and wait until “READY”
appears. [Fig. 1].
[Fig. 1]
4-4) On the X-ray generator, press the exposure button, and after the exposure, select
“END STUDY”. Transfer the SMPTE pattern image to the printer or the highdefinition monitor. [Fig. 2]
[Fig. 2]
4-5) Take the SMPTE pattern image again. On the QA screen, compare the image
displayed on the operation unit screen with the film image printed in Step 4) above or
with the image on the high-definition monitor. Make sure there is no difference in
contrast and gradation between those images. Check both preview and magnify
images.
If there are any differences between these images, perform the procedure described in
“(4) Operation Unit Image Gamma Correction” on the next page.
If there are no differences between these images, the steps are complete.
84
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.