Manual Control No. : BY8-2257-0E5
Name of Product : CXDI-50G/50C
Distribution Control No.
Issued on
Service Manual Introduction
This service manual belongs to a series of after-service guides Canon Inc. publishes as part of
its comprehensive product quality guarantee program.
This service manual consists of seven chapters; General Information, Installation Guide, Feature
Information, Repair Guide, Parts Catalog, Troubleshooting and Service Manual Report.
If the product undergoes a large modification, a new service manual of revised edition will be
sent to you. In other cases, service manual report will be sent to you updates the manual.
Note 1:
This service manual is published by Canon Inc. in accordance with Article 6 (Furnishing
the Referent Materials) of the Service Assignment Contract it has concluded with your
company.
Note 2:
This service manual property of Canon Inc. and the company may seek to have it returned,
depending on circumstances. You are expected to keep it until then.
Note 3:
You inquiries, suggestions etc. about the contents of this service manual should be
addressed to:
Medical Products Technical Service Dept.
Canon Inc. Headquarters
30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo 146-8501, Japan
Page 3
Caution Regarding Service
This product was precisely assembled under strict manufacturing process control. There
are several hazardous locations inside of this product. Careless work while the cover is
removed can result in pinching fingers or cause electrical shock. Please perform the work
with the following important points in mind:
1. Setup, Repair and Maintenance
In order to ensure safety, the best performance, setup, repair and maintenance work can only
be performed by the technicians received the service training specified by Canon Inc. If there
are order required certificates or restrictions specified by the law or ordinances, those
regulations of the country must be observed.
2. Removing the external cover
When removing the cover during maintenance, repair, etc., perform the work after switching
the power off. Never touch the device with wet hands, as there is a risk of electric shock.
3. Fuse
When replacing the fuse, first, resolve the reason of failure and then replace the fuse with the
specified type. Never use a fuse other than the specified type.
4. Connecting the grounding wire
The provided ground wire must be connected to the ground terminal indoors. make sure that
the device is properly grounded.
5. Alternation prohibition
Never modify the medical device in any way.
6. Waste control
The service provider is responsible for the disposal of used service parts, packing material,
etc. resulting from the setup, repair or maintenance of the medical device However, the
customer is responsible for the disposal of the medical device. Disposal activities must
follow the regulations (=specially controlled industrial waste) of the country where the
device is used.
Page 4
VORSICHT
Befolgen Sie die unten angegebenen Sicherheitsanweisungen.
Mißachtung kann zu erletzungenoder Unfällen führen.
1.Zerlegung, Zusammenbau, Einstellung und Wartung
Zerlegung, Zusammenbau, Einstellung und Wartung dürfen nur von einem Wartungstechniker
durchgeführt werden, der an einem von Canon vorgeschriebenen Wartungslehrgang teilgenommen hat.
2.Entfernen von Abdeckungen
Schalten Sie unbedingt die Stromversorgung des Instruments aus, bevor Sie die Abdeckungen zwecks
Wartung und Reparatur entfernen.
Vermeiden Sie auch eine Berührung des Instruments mit nassen Händen.
Anderenfalls können Sie einen elektrischen Schlag erleiden, der zum Tod oder schwerer Verletzung
führen kann.
3.Sicherung
Wenn die Sicherung ausgewechselt werden muß, schalten Sie unbedingt die Stromversorgung des
Instruments aus, und beheben Sie die Ursache für das Durchbrennen der Sicherung.
Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag kommen.
4.Erdleiter
Erden Sie das Instrument unbedingt an einer Schukosteckdose.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag durch Leckstrom kommen.
5.Umbau
Jeder Umbau des Produktes ist strengstens untersagt, da dies zu einem Brand oder elektrischen Schlag
führen kann.
6.Bewegliche Teile
Dieses Instrument enthält bewegliche Teile.
Führen Sie während der Bewegung der Teile keine unachtsame Tätigkeit aus.
Anderenfalls können Sie verletzt werden.
7.Schnittstellenanschluß
Wenn andere Geräte über den Schnittstellenanschluß an das Instrument angeschlossen werden, prüfen Sie
nach dem Anschluß, daß der Leckstrom innerhal b des z ulässigen
Bereichs liegt.
8.Lithiumbatterie
Ersetzen Sie die Lithiumbatterie nur durch den vorgeschriebenen Typ.
Verbrauchte Batterien dürfen nicht ins Feuer geworfen und weder zerlegt noch geladen werden.
Entsorgen Sie verbrauchte Batterien umweltschonend gemäß den Gesetzen
oder Vorschriften des Landes, in dem das Instrument benutzt wird.
According to the “IEC60601-1-1:2000”, devices installed in the patient environment are
restricted to “electric medical devices conforming to IEC60601-1”.
The Control PC, operation unit, and the magnetic card reader, etc. options that are parts of the
CXDI-C3S are classified under the data processing device standard (IEC60950), therefore these
items should not be installed in the patient environment. Otherwise the Control PC is only
classified in CXDI-C3S.
The patient environment described below is an example cited from the “IEC60601-1-1:2000” –
the measurements are only guidelines. However, the “IEC60601-1-1:2000” example must be
treated as the standard.
Therefore, the CXDI-C3S must be installed in a location further than the measurements below
(outside of the patient environment).
*Areas where the patient moves (not only during imaging but when entering and leaving
the room, etc.) are also considered as part of the patient environment, therefore the
installation location should be determined upon consultation with the user regarding
areas outside of the patient environment.
Change of the image read-out method from the sensor was made to the former CXDI-50G. In the
former CXDI-50G, Aux. Shift Register Driver (PCB-D-EP), Scanner with Main Scanner Driver
(PCB-50A) and A/D Conversion (PCB-50AD) are located in the both sides of LANMIT. One panel
is considered as two virtual panels, and image data was read out from both sides. However, in new
CXDI-50G, Aux. Shift Register Driver (PCB-D-EP), Signal Read-out Circuit (PCB-50A) and the
A/D Conversion (PCB-50AD) are located in one side of LANMIT and the image data was read from
one side. Since the appearance of new CXDI-50G is the same as that of the former CXDI-50G, they
are identified
by changing the serial number of the main unit.
CXDI System Software Ver.6.4 and later
Clinic Software Ver.1.04 and later
Serial number of main units
CXDI-50C
Fluorescent substance of CXDI-50G has been changed to CsI from GOS in CXDI-50C so that the
CXDI-50C can be ranked as the more superior performance model (high sensitivity model) than
CXDI-50G. The read-out method of image data is read-out from both sides of LANMIT which is
same as CXDI-50G.
Symbol Description Remarks
CA1 AC Cable 3m
CA2 50Sensor Cable 3m/7m
CA3 X-Ray I/F Cable 20m
CA4 LAN Cable (Category 5)
CA5 LAN Cable (Category 5) 10m
CA6 Serial Cable (Touch Panel) 2m
CA7 VGA Cable 1.5m
CA8 AC Cable 3m
(CXDI-50G/50C)
CA4
Printer
Remo.
Imaging Unit
Control/Signal
Network
(100/10bese-T)
Ethernet
CA5
SERIAL(COM1)
Control PC
(CXDI-C3S)
LAN1
LAN2
AC120V±10% 60Hz 2A
AC230V±10% 50/60Hz 1.5A
SERIAL(COM2)
MOUSE
KEY BOARD
SERIAL(COM3)
SERIAL(COM4)
C Power IN
VGA
SCSI
CA8
CA6
CA7
RS232C
VGA
RS232C
Operation
Unit
ID Unit
Mouse
Key Board
2
Page 10
1. General
pply
ote switc
et
ging
22 SSyysstteemm DDiiaaggrraamm
2.1 Standalone System
CXDI-50G/50C
2.2Total System
Total System Block Diagram (Example)
Patient Circumstances or Medical Room
CXDI-50G/50C
Protective
Grounding
CXDI-40E
Imaging Unit
Ethernet
POWER BOX
50 Power
Su
1st/2nd
AC230V
Insulation
Power Box
Basic Insulation
Network switch
AC230V
Ethern
X-ray generator
(601 compatible)
Rem
Imaging Unit
h
CXDI-40G/
CXDI-31 etc.
Ima
Unit
Grid
Unit
Contorol PC
C3
Multi Box 2
Ethernet
Image Examination WS
Image File Equipment
Printer etc.
3
Page 11
1. General
33 CCXXDDII IImmaaggee PPrroocceessssiinngg
3.1 Proccess Flow
Original image
Processed image
Born image
Raw image
QA image
DICOM output
Diag. image
Correction
processing
QA processing
Offset correction
Gain correction
Defect correction
dtstore
Characteristic extraction
• Sharpening
• DEP
• Gradation processing
MTF enhancement
•
(Frequency enhancement)
• Grid stripe reduction
• High density clipping
request
3.2 Image Types
(1) BORN IMAGE
The image obtained with LANMIT before any correction is made.
Outside distribution of these images is prohibited, including dtstore images.
(2) RAW IMAGE
Born image after offset processing, gain correction.
This is the image with LANMIT specific characteristics corrected.
(3) ORIGINAL IMAGE
Raw image after preprocessing.
(4) QA IMAGE
Original image after gradation processing, sharpening, and other processing.
The CXDI performs image processing up to this point.
(5) DIAG. IMAGE
QA image after further image processing necessary for diagnosis.
Image processed by the user for diagnostic purposes.
(6) PROCESSED IMAGE
Diagnosis image after post-processing.
Image modified by the user or the default processed image.
4
Page 12
1. General
44 SSppeecciiffiiccaattiioonnss
The CXDI-50G/50C (Imaging Unit/Power Box) is the Digital Cassette that has the mobility and
can be used on the optional angles.
(1) Imaging Unit
This unit consists of the internal sensor, 50Di Board, 50AD Board, 50LED Board and its
outer cover. The sensor unit converts the X-ray image to the electrical signal (O/E
Conversion) and after performs the A/D conversion, transfer its signal through the Power
Box with Ethernet cable to the Control PC.
Item 50G 50C Remarks
Object General Shooting Cassette
Effective filming range 353 x 430mm
Number of Pixels 2214 x 2700
Effective Number of Pixels 2208 x 2688
Pixel pitch
Fluorescent substance GOS Fluorescent screen CsI one panel
Output gradations
Transfer method
Imaging cycle 15 sec. (standard)
Dimension
Imaging Unit coloring Cool white
Imaging Unit mass
(except Grid)
Space between surface where
patient gets in contact (CFRP)
and sensor surface (glass)
Remote SW OFF
Heat
emission
Power
Consumpt
-ion
Mechanical strength*
Control PC
Power Control
(ON/OFF)
Grid attach/remove detector Yes
Cable for Imaging Unit 3/7m 2types
Count of connected sensor
Scattered radiation backward
block sheet
Environment-conscious
unleaded type
Photo timer Cannot be built in
Sleep 19kcal/h
Max load mode 24kcal/h 30kcal/h
Remote SW OFF
Sleep 22W
Max load mode 28W 35W
160µm x 160µm
12bit (4,096 gradations)
A/D 14bit
Ethernet: Imaging Unit to
Control PC
(Through the Power Box)
491 (W) x 477 (D) x 23 (H)
mm
4.8Kg 4.9kg Except the cable
5.7Kg 5.8kg With 7m cable
5.2Kg 5.3kg With 3m cable
Within 5.0±0.5mm. Within 4.9±0.5mm
9.5kcal/h
11W
Cassette with resisting the
strength
(Original specification)
CXDI-C3S
General PC in market
None
(Power Box: Operation
with Remote switch
manually)
Multiple Imaging Units are
connectable to a single
control PC via Network
switch. (Limitation of the
number of connectable
Imaging Unit depends on the
specification of control
software in control PC)
Mo sheet (0.3 mm thick)
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
Å
(Mobile/Desktop PC)
Including handle portion
427 mm (excluding handle portion)
Including the Power Box
Including the Power Box
Including the Power Box
1 image per 15 Sec.
Including the Power Box
Including the Power Box
Including the Power Box
1 image per 15 Sec.
Load uniformly: 150Kg
Load partly: 100Kg/
The Imaging unit is put on the plain surface
with the Sensor side (Detector) is up.
Distinguish with Product order.
Network switch should be procured at each
sales company.
φ
40mm
5
Page 13
Imaging restriction
(Imaging Prohibition)
1. General
Item 50G 50C Remarks
When the internal
temperature of Imaging Unit
is above 49 degree Celsius,
its state is changed to sleep
User interface
mode. And the Imaging
prohibition will be continued
when the internal
temperature is below 48
degree Celsius.
Single type LED
Off: Imaging unit power is
off
On in orange: Imaging unit
power is on (unable to
perform imaging)
Blinking in green: Preparing
imaging/error status
On in green: Imaging
preparation is complete
Å
Å
INDICATION BUSY SENSOR POWER
Color Orange Green Blue
Imaging unit
is off
Imaging unit
is on
Preparing
imaging
Imaging
preparation
complete
Error status -- Blinking
Communicating
Initialization
(when
startup)
Network not
set (when
startup)
*1: Turns on and off for 0.5 seconds each
*2: Turns on and off twice for 0.5 seconds, then
turns off for 0.5 seconds
*3: Turns on and off randomly
*4: Fades in for 1 second and fades out for 1
second
*5: Fades in for 2 seconds, and then turns off
Off Off Off
-- Off On
-- Blinking
-- Off On
*3
Blinking
-- On
-- Blinking
-- Blinking
*1
On
*2
On
*4
On
*5
On
6
Page 14
1. General
(2)Power Box
This unit consists of 40XRAY Board, 50Power Supply and its outer cover.
The function; the signal transition between Imaging unit and Control PC, the interface to the
X-ray generator equipment and power supply to the Imaging unit has been implemented.
Item Content Remarks
Communication
interface standard
Communication method
IEEEE
Asynchronous serial communication
method
*
802.3u (100BASE-TX) Connector type: RJ45
Data length: 10bit
Data rate: 15.625 kHz
Reference
CXDI-50G Power supply
Rated Voltage:
AC 100-120V (AC 85-132V)
Power supply
AC 100-120V±10% 50/60Hz 1.5A
AC 200-240V±10% 50/60Hz 0.8A
AC 200-240V (AC170-264V)
Mass 4.2 Kg
Except bottom rubber parts
Dimension 358(W) x 200(D) x 65(H)* mm
(With bottom rubber parts:
75mm)
(3) Environment rated parameters
Item Content Remarks
Operating temperature
+5 to +35°C
Operating humidity 30 to 75% RH
Keeping or
Transporting
Temperature: -30 to +50°C
Humidity: 10 to 60% RH
Atmospheric pressure: 700 to 1060 hPa
Must be without dewing
*
IEEE: Institute of Electrical and Electronic Engineers
7
Page 15
Page 16
X
5
5
C
C
22.. IIn
CCooppyyrriigghhtt ((CC)) CCaannoonn IInncc.. MMeeddiiccaall T
Please pay attention to the followings when installing the system.
(1) If the equipment is hoisted, lowered or transport, it must be supported at both sides by a
minimum of two people so there is no danger of it falling.
(2) If a forklift, etc. is used to transport the equipment, make sure there is nothing that could
impede the forklift on its route to the final destination.
(3) When installing the equipment, be sure the site meets the following criteria:
1) There must be no dripping water in the area.
2) The environment must be free of harmful elements such as humid or acidic air, air with a
saline or sulfur content, where there is poor ventilation or where air pressure or
temperature is unusual.
3) The equipment must not be placed at an angle or subjected to vibration or shock (this
includes during transportation).
4) The equipment must not be kept where chemical products are stored or where gasses are
generated.
5) The site’s power supply must be of the correct voltage and frequency for the equipment.
6) The site must be connected to a fully earthed cable with sufficient ground resistance to
meet standard values.
(4) After installation, be sure to dispose of waste product packaging with care and with full
respect for the environment.
- 1 -
Page 19
2. Installation Manual
22 PPrroodduucctt CCoonnffiigguurraattiioonn
2.1 Product Configuration List
No. Item Name Qty Remarks
1 CXDI-50G/50C Imaging Unit 1 With Sensor cable (3/7m)
2 Power Box 1 100 - 120/230V
3 X-ray I/F cable 1 20m
4 Remote switch 1 20m
5 Cable clump 1 For fixing the sensor cable
6 Screw (M4 x 6 mm) XB1-1400-603 1 For fixing the cable clump
7 Sensor Information File (FD) 1
8 Power supply cable (with AC plug) 1 3m (100/120/230V)
9 Attached documents for medical - (100V)
LAN cable for connecting Control PC / Power Box and Network switch (Switching HUB) for
connecting the multiple Imaging Units shall be procured at each sales company.
- (100V)
- LAN cable (Over category 5)\
When Control PC and Power Box are connected directly, Cross type is used, but when they are
connected via Network switch, Straight type is used. However, this is not applied when
Network switch has AUTO-MDI/MDI-X function*.
- Network switch (Switching HUB)
Sales companies adopt Network switch (Switching HUB) after conducting the test and the
operation check for Switching HUB that meets the general standard.
* AUTO-MDI/MDI-X :
One of the functions of Network switch (Switching HUB) and Broadband-router.
They can detect Communication port of other side automatically and connect in proper procedure.
MDI: Network-card, etc.
MDI-X: Network switch (Switching HUB) and Broadband-router, etc.
Since AUTO-MDI/AUTO-MDI-X functions can detect automatically in proper procedure, they do not
require consideration to types of LAN cable (Straight of Cross cable).
- 2 -
Page 20
2. Installation Manual
2.2 Configuration
No. 1 No. 2
Name
Qty 1 Qty 1
Remarks 3/7m Remarks I/F and Power supply (3/7m)
CXDI-50G/50CImaging Unit
Name Power Box
No. 3 No. 4
Name X-ray I/F cable Name Remote switch
Qty 1 Qty 1
Remarks
No. 5 No. 6
Name Cable clump Name Screw (M4 x 6mm)
Qty 1 Qty 1
Remarks For fixing the sensor cable Remarks For fixing the cable clump
Connection with X-ray
generator
Remarks
Power Box power source
ON/OFF
- 3 -
Page 21
2. Installation Manual
No. 7 No. 8
Name Sensor Information file FD Name Power supply cable
Qty 1 Qty 1
Remarks Remarks
For Power Box
(100/120/230V each type)
- 4 -
Page 22
2. Installation Manual
33 PPaacckkiinngg DDiiaaggrraamm
3.1 X-ray Digital Camera System (CXDI-50G/50C)
(1) CXDI-50G/50C assemble package
(2) CXDI-50G/50C Imaging Unit package
- 5 -
Page 23
(3) Power Box assemble package
2. Installation Manual
(4) Grid (optional)
- 6 -
Page 24
2. Installation Manual
44 IInnssttaallllaattiioonn PPrroocceedduurreess
4.1 Lists of Tool Needed for Installation
Tool needed for new installation.
No. Tool Name Unit Remarks
1. General Tools 1 set
2. Note PC 1
3. LAN Card 1 For Note PC (as required)
4. Mouse 1 PS/2 type
5. Keyboard 1 PS/2 type
6. HUB 1 Connection between Control PC and Note PC
7. 10/100BASE-TX cable 2 Straight type (Control PC to Note PC)
CXDI Software version
8.
compatibility table
Mirror, oil-based marker
9.
and paper etc.
PC/AT compatible
(OS: Microsoft Windows XP Professional recommend)
-
1 For adjusting the alignment with the X-ray tube.
- 7 -
Page 25
2. Installation Manual
4.2 System Installation Procedures
No. Step Conditions and Checkpoints Reference Section
1 Unpacking and checking the
product’s constituent parts
2 Connect the Imaging Unit
and the Power Box
3 Connect the Power Box and
the Control PC
4 Connect the Power Box and
the X-ray generators
5 Check date and time - Date and time must be changed according
6 Check the software
program’s version
7 Identifying the Imaging Unit
(input the sensor serial
numbers)
8 Enter control PC serial
number.
(9) Adjusting the timing with X-
ray generator
10 Calibration -No error must be displayed.
11 Setting the Fixed ROI Areas If necessary, set the ROI area.
12 Set exposure parameter table -Set it in consultation with the technician.
13 Set annotation -Set it in consultation with the technician.
14 Connect the network and set
the output destination
15 Startup settings
16 Exposure testing -The data must be sent to the printer and
17 Check the linearity of the
transferred image density.
18 Operation unit Gamma
correction
19 Body parts settings -The engineer in charge must be consulted
20 Check and set the system
settings.
21 Total adjustments and delete
the unnecessary data.
22 Cleaning
-There must be no missing parts, damage,
dents, etc.
-There must no color changes in the shock
sensor.
-Handle the instrument carefully, as it may
be damaged if something is hit against it,
dropped or receives the strong jolt.
-The cable must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
-The cable must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
-The manufacturer of the X-ray generators
must be asked to handle the connections
with the generators.
to the area where the instrument is installed.
-The compatibility of the sensor unit and the
Control PC must be checked on the
compatibility list, and the software program
must be installed or upgrade as required.
-No required usually.
storage and the image quality must be
checked.
prior to perform these settings.
-The engineer in charge must be consulted
prior to perform these settings.
-Conform according to the check sheet.
-Delete the unnecessary data.
“(1) Checking and
Setting the Date and
Time” in section 5.6.
“(2) Checking the
Firmware Version” in
section 5.6.
“(6) Identifying the
Sensor Units” in section
5.6.
“(7) Entering Control
PC Serial Number” in
section 5.6.
Operation Manual
“(8) Table Setup
Setting” in section 5.6.
“(9) Performing the
Annotation Setting” in
section 5.6.
“(10) Network
Connections” in section
5.6.
“(5) Set Up Startup
Menu” in section 5.6.
Section 5.7 Image
Quality.
“(11) Linearity Check
Image Density” in
section 5.6.
“(12) Operation Unit
Gamma Correction” in
section 5.6.
Operation Manual
Each section in section
5.6 Settings.
Section 5.8 Postinstallation check.
- 8 -
Page 26
2. Installation Manual
23 Explain the operation to the
user
24 Final parameter adjustment -The engineer in charge must be consulted
prior narrowing down the adjustments to the
final values.
25 Inserting the backup floppy
disk.
-It must be confirmed at re-start that backup
files have been made.
-No necessary for the system installed in
vehicles.
26 Back up valuable data
Operation Manual
Operation Manual
“(15) Backing up
Setting Data to FD” in
section 5.6.
“(14) Backing Up when
Installing” in section
5.6.
- 9 -
Page 27
2. Installation Manual
55 IInnssttaallllaattiioonn
5.1 Connection to each unit
5.1.1 Connection diagram
CXDI-50G/50C
CXDI-50G
Control PC
Remote switch
POWER SUPPLY
PCBXRAY
POWER BOX
- 10 -
Page 28
2. Installation Manual
5.1.2 Connecting to the Power Box
(1) Removing the cover
Remove the 5 screws from the back of the power box and the 2 screws on each side at the
bottom of the power box.
Also remove screws on the
(2) Check the jumper pin settings on the PWB-40XRAY board in the power box. The power
supplies for 7m and 3m are not compatible due to having different output voltages. This
short pin is for determining the power, not for changing the voltage.
50 power unit setting
50 Power(7m): Short between JP1 6-8p
50 power(3m): Short between JP1 7-8p
This setting is done when shipping
With changing the jumper pin settings, the signal sent to PLD (IC1) of the PWB-40XRAY
board has either state High or Low. By using this state, the PBIF50XRAY PLD code which
written in the PLD can distinguish whether the power supply is for 3m or 7m.
The Imaging unit determines the cable length based on the dip switch setting (SW1-7:
OFF) on the PWB-50Di board. If cable connector P4 #1-#2 is open, cable length is 7m. If
#1-#2 is connected, cable length is 3m.
- 11 -
Page 29
2. Installation Manual
p
N
(3) Cable connections
1) Loosen the lock nut for each cable and connect the cables to the power box. Then fix
the cables by tightening the lock nuts.
Sensor cable
X-ray I/F cable
Lock nut
Remote cable
Note: To avoid the risk of damage when the cables are removed with very large force, Check
tightening torque of the bush (refer the following figure).
Gap 1.0 to 1.5mm
o gap
<Sensor cable> <X-RAY I/F cable,
Remote cable>
2) Attach the cable clamp to the sensor cable, and then fix it to the power box using the
screw (M4 x 6 mm).
Clamp the cable’s
ortion
shield
Cable clamp
Screw (M4×6mm)
- 12 -
Page 30
3) Connecting each cable connector.
2. Installation Manual
Connected to
Sensor cable
Connected to
Remote cable
Connected to
4) After completing the connections, attach the power box cover.
5) Connect the LAN cable to the back of the power box.
X-ray I/F cable
- 13 -
LAN cable
Page 31
2. Installation Manual
6) Connect the power cable to the back of the power box.
Power cable
Power switch
Power cable
X-ray I/F cable
Remote cable
CXDI-50G/50C sensor cable
LAN cable
- 14 -
Page 32
2. Installation Manual
5.1.3 Connection diagram for Control PC rear panel
Mouse
Keyboard
Operation unit
RS-232C cable
LAN cable (straight cable)
(Between Control PC and network)
Operation unit
VGA cable (Connector 1)
LAN cable
(Between Control PC and Power Box)
Power cable
- 15 -
Page 33
2. Installation Manual
5.2 Starting up and shutting down the System
Perform the following sequences when starting up and shutting down the system.
5.2.1 Sequence for Starting up the System
Perform the following sequence when turning the system power on.
If you do not perform the correct sequence, the imaging unit cannot be recognized, resulting in
an error. (This is because the system communicates with the imaging unit when turning the
system on.)
The power box cannot be turned on in conjunction with turning on the control PC.
Since the power box is equipped with a remote switch that turns on/off the secondary output,
you caninstall the switch on your side to turn it on/off.
1) Turn on the main power of the power box.
2) Turn on the remote switch of the power box.
3) Turn on the control PC.
Note:
Ccrstart.bat should be registered in Windows Startup.
5.2.2 Sequence for Turning the Power off (Shutdown)
1) From OPU, select SYSTEM → [SHUTDOWN] or [SHUTDOWN after transfer]
The control PC automatically turns off.
2) Turn off the remote switch of the power box.
3) Turn off the main power of the power box.
Note:
Turn off the main power of the power box and OPU power when not using the system for
a long period.
- 16 -
Page 34
5.3 X-ray Controller Interface
g
(
5.3.1 Interface Signal Description
X-ray Generator
X-ray emission signal
1st SW level ON)
X-ray emission signal
(2nd SW level ON)
Start X-ray emission
Generator
Setup timer
End X-ray emission
X-ray emission signal OFF
2. Installation Manual
RX_REQ
RX_COM
CXDI
Detect X-ray emission
nal ON
si
Imaging possible?
Enable X-ray emission
ON
Image Data read
Enable X-ray emission
OFF
- 17 -
Page 35
• When normal imaging
2. Installation Manual
• When timeout due to RX_REQ not negating
- 18 -
Page 36
2. Installation Manual
5.3.2 Signal names and functions in the connection with the X-ray generator
<X-ray Sync Signal>
Signal name Functions
X-ray exposure signal
Indicates that an X-ray exposure is ordered at the X-ray generator side.
RX_REQ
RX_COM
This signal needs to be retained at least for T
X-ray release signal (RX_COM) is not output if the T
period. It takes about 1 second in the worst case scenario before the operation can be
resumed when RX_REQ is negated in this period.
Time required to assert RX_COM after receiving the RX_REQ from the X-ray
generator
T
....................................................................... min.300 ms
PREP
Since a captured image is read from the sensor when RX_REQ is negated (or RX_COM
is time out) as a trigger, reading action is delayed if RX_COM does not negate and time
out is used as the trigger, resulting in delay of image display timing.
We recommend using a configuration in which RX_REQ is negated.
Time required from X-ray exposure completion to negating RX_REQ
* Image display timing is delayed if this is not fulfilled.
X-ray release signal
Checks whether or not imaging is ready at the CXDI side after receiving X-ray exposure
signal (RX_REQ) from the X-ray generator. This signal is output to the X-ray generator
side when imaging is ready.
Time required from asserting RX_COM to exposing X-ray
* Shading could occur on images if this is not fulfilled.
PREP
period.
PREP
is less than the necessary
- 19 -
Page 37
2. Installation Manual
The connection with CXDI-50G/50C and X-ray generator equipment
Connection conditions
1. The X-ray exposure signal line must be insulated, and its total impedance must be 100 ohms or
less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V and
DC 30V, and its current ranges from 10mA to 2A.
Only the secondary power supply can be connected.
- 20 -
Page 38
2. Installation Manual
5.3.3 Rating and Characteristics for Relay and Photo coupler (on PWB-40XRAY Board)
(1) RL1 (Power Relay/Plug-in terminal type)
1) Rating (Operational coil)
Rated
Rated
voltage
(V)
DC5V 72 69.4 69.5 86.0
curre
nt
(mA) (Ω)
Coil
resistance
Coil Inductance
(mH)
Armature
OFF
Armature
ON
Pick-up
voltage
(V) (V) (V) (mW)
below
80%
Dropout
voltage
below
10%
Maximu
m voltage
135%
(at 50)
Power
consumption
Approx 360
2) Rating (Switch/Contact)
types Single stable
arrangement 2 Form C
Contact material Au-clad AgNi type
Relating capacity
AC250 5A
DC30V 5A
Max. switching power 1250VA 150W
Max. switching voltage 250V AC
Max. switching current 5A
Min. switching capacity
100µA 1V DC
3) Characteristics
Item Content
Operate time Max. 20ms
Reset time Max. 10ms
Mechanical 18,000 times/hour Maximum open/close
frequency
Withstand voltage
life
Ambient temperature
Maximum operating frequency 50 times/Sec.
Rated load 1,800 times/hour
Between coil
contacts
Between same poles1,000 Vrms
Mechanical 5x10
Electrical
2,000 Vrms
7
times
5
10
at 5A 250V AC
5
5x10
at 5A 30V DC
-40°C to +70°C (no freezing nor condensation)
- 21 -
Page 39
(2) PCI (Photo-coupler)
1) Maximum Ratings (Ta = 25°C)
CHARACTERISTIC SYMBOL RATING
Forward Current IF(RMS) 50 mA
Forward Current
LED
Pulse forward current1 I
Reverse Voltage VR 5 V
Collector-Emitter Voltage V
Emitter-Collector Voltage V
Collector Current IC 50 mA
2. Installation Manual
/°C -0.7(Ta≥53°C)
∆IF
1 A
FP
80 V
CEO
7 V
ECO
Collector Power Dissipation
DETECTOR
(1 Circuit)
Total Package Power Dissipation
(1 Circuit)
Isolation Voltage2 BVs 2500 Vms
2) Maximum Ratings (Ta = 25°C)
CHARACTERISTIC SYMBOL
Forward Voltage VF I
Reverse Current IR V
LED
DETECTOR
SWITCH
CHARACTRISTIC
Capacitance Ct V=0, f=1 MHz - 30 - pF
Max. Forward current VFM I
Collector-Emitter
Breakdown Voltage
Emitter-Collector
Breakdown Voltage
Collector Dark Current I
V
(BR)CEO
V
(BR)ECO
I
I
CEO
V
Rise Time tr - 2 -
Fall Time tf - 3 -
Turn-On Time tON - 3 -
Turn-Off Time t
Off
Turn-On Time tON - 2 -
Storage Time tS - 25 -
Turn-Off Time t
OFF
150 mW
P
C
P
200 mW
T
TEST
CONDITION
=10 mA 1.0 1.15 1.3 V
F
=5 V - - 10 µA
R
=0.5 A - 3 4 V
FM
=0.5 mA 80 - - V
C
=0.1 mA 7 - - V
E
VCE=48 V
Ambient Light
Below (100lx)
=48 V, Ta=85
CE
Ambient Light
Below (100lx)
V
=10 V
CC
I
=2 mA
C
R
=100Ω
L
MIN. TYP MAX. UNIT
-
-
0.01
(2)
2
(4)
0.1
(10)
50
(50)
- 3 -
VCC=5 V
I
=16 mA
F
R
=1.9 kΩ
L
- 40 -
µA
µs
µs
Note: Because of the construction, leak current might be increased by ambient light. Please use
photo-coupler with less ambient light
1
pulse amplitude 100µs, frequency 100Hz
2
AC, 1min R.H.≤60 LED side pins shorted together and DETECTOR side pins shorted together
- 22 -
Page 40
2. Installation Manual
5.4 Network Settings 1
1. Objective
The CXDI-50G/50C imaging part communicates with the control PC by using Ethernet
[IEEE802.3u (100Base-TX)] to transfer X-ray images.
The control PC performs DICOM transfer in order to use Ethernet to transfer the obtained
images to the printer and storage device.
This section describes how to set up the TCP/IP that is necessary for the network connection.
Set up the following three items:
1) TCP/IP setting for the control PC network card
2) Network setting for Screwcap.ini
3) Network setting stored in the CXDI-50G/50C sensor
2. Preparation
• Keyboard
• Mouse
Connect the keyboard and mouse to the back of the control PC.
Check if the system is connected, and then turn the system on.
3. Setup method
3.1 TCP/IP setting for the control PC network card
Perform the set up by referring to “C3S Service Manual” -> the chapter “System Manual” ->
“Network Setup”.
Default values
IP Address:192.168.100.10
SubnetMask:255.255.255.0
- 23 -
Page 41
2. Installation Manual
3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
CXDI-50G/50C and control PC such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the CXDI-50G/50C imaging part in the previous item, “3.1
TCP/IP setting for the control PC network card”.
Screwcap.ini is located in the following directory:
CXDI-50G/50C IP address.*1
This must be the same as the
IP address stored in the
imaging part.
1
*
CXDI-50G/50C’s IP address: This address must be the same as the IP address stored in
the next item, “3.3 Network setting stored in the CXDI50G/50C sensor”.
- 24 -
Page 42
2. Installation Manual
3.3 Network setting stored in the CXDI-50G/50C sensor
The factory default setting is shown in the table below.
This setting is not necessary
Item to be set Factory default value
Sensor IP address 192.168.100.11
Subnet mask 255.255.255.0
Host IP address 192.168.100.10
Port number for command 12121
Port number for data 12122
If you change the setting, refer to “Tool Software Operation Manual for Ethernet” → “Imaging
Part IP Address Setting”.
unless you have changed the setting.
- 25 -
Page 43
2. Installation Manual
(
5.5 Setting the Fixed ROI Areas
1) Purpose
Set the fixed ROI area on the sensor to expose by the fixed ROI area because user can
not get the proper image by the Auto ROI area.
2) Setting method
2-1) Investigate the actual size and position of the ROI that is required.
2-2) Designate the SIZE, POSITION, and NUMBER (max 5)* on the sensor.
* However, in the case of using the new function “Display of AEC (Automated Exposure Control)
Field in Preview Screen” added from CXDI System Software Ver.6.2, NUMBER that can be
specified is max 3
.
0,0)
X
W
Y
H
Sensor
Ws=350mm
Hs=430mm
2-3) Convert the size and position of the ROIs in 2) to pixel values. The pixel size of the
sensor is 160µm. For multiple values, use X’, Y’, W’, H’, X”, Y”, W”, and H” for
calculations.
X/160 µm Let this value equal A
Y/160 µm Let this value equal B
W/160 µm Let this value equal C
H/160 µm Let this value equal D
2-4) Open the “DENSITY ADJUSTMENT CONTROL” screen when the normal
imaging screen is active.
SYSTEMÆSETUP MENUÆSYSTEM SETTINGSÆ DENSITY
ADJUSTMENT CONTROL
- 26 -
Page 44
2. Installation Manual
2-5) The “DENSITY ADJUSTMENT CONTROL” screen appears. Confirm that the
sensor is set with a sensor switch button. Press the [Fixed ROI Area] key.
2-6) Press the [Area setting] key.
2-7) The fixed ROI 1 setting screen appears. Input values A to D from step 3) into the
edit box, and press [ENABLE]. To set multiple fixed ROIs, input A’ to D’ and A”
to D” into fixed ROI 2 settings and fixed ROI 3 settings respectively.
Area setting button
- 27 -
Page 45
2. Installation Manual
2-8) The display returns to the “DENSITY ADJUSTMENT CONTROL” screen.
Confirm that POSITION and SIZE fields not set in step 5) to step 7) are disabled
(dimmed). If they are not dimmed, press the [Area setting] key, and press
[DISABLE] in fixed ROI * settings.
Check to make sure the displays are dimmed.
2-9) After Confirming all settings, and press [EXIT].
2-10) The display returns to the system settings screen. Press [OK].
2-11) “Change settings?” appears. Press [OK]. Be careful, because if [CANCEL] is
pressed, all changes made to the settings are deleted.
“Change settings?” appeared.
2-12) Return to the normal imaging screen, and turn off the power to
the CXDI.
- 28 -
Page 46
2. Installation Manual
5.6 Adjusting the photo timer
<Outline>
The photo timer installed in the imaging unit is adjusted so that the exposure time of the X-rays
generated by the X-ray generators is to OFF using the optimal value.
* This work necessitates performing some adjustments inside the X-ray generators. In order to
ensure that the work will proceed smoothly, discuss the schedule and other details with the
representative of the manufacturer of the X-ray generators.
<Preparations>
(1) Start up the CXDI system. (Normal radiography mode)
(2) Perform calibration.
(3) Change parameters of VPT button as follows.
* This operation must be done by the “VPT” button which uses “fixed ROI”. The adjustment
cannot be performed properly by the body parts buttons which use auto ROI.
1) From SYSTEM > EDIT EXPOSURE MODE, select VPT button.
2) Press the NEXT PAGE button to enter the parameter editing screen and then set the following
parameters.
(A) GENERATOR PARAMETER SCREEN
Set only the center of the fixed ROI area to ON.
(During the adjustment, the acrylic sheet may be used to cover the fixed ROI area selected
3) Press the OK button to save the parameters you changed.
- 29 -
Page 47
2. Installation Manual
<Adjustment>
It is advisable to leave the photo timer adjustments in the hands of the individual representing the
manufacturer of the X-ray generators.
Basically, it is essential that the manufacturer’s representatives perform these adjustments at the
same time as the film screen and/or screen system adjustments.
(1) Have ready the 5 cm, 10cm, and 20cm photo timer adjustment acrylics sheet, proceed to X-ray at
70k V, 100kV and 120kV, respectively, and adjust the photo timer in such a way that the Rex
value as in the list below. (At this step, it is mainly the sensitivity difference that is based on the
radiation quality of the photo timer which is adjusted.)
Imaging Unit Rex value
CXDI-40EG, CXDI-50G650
CXDI-40EC, CXDI-50C325
(2) Adjust the radiation dosage gradually using the option buttons (such as the H.S button, L.S
button, “+” (plus) and “-“ (minus) button) on the X-ray generator, and decide on the final dosage.
For instance, the L.S button is for providing about 1.5 times the default dosage.
If, in overall terms, the sensitivity has shifted slightly or it differs slightly from one body part to
another, use the H.S, L.S, “+” and “-“ buttons on the X-ray generator to adjust the sensitivity as
required.
Reference: If the Rex value is set as in the list when the acrylic sheet was X-rayed under the
conditions set by the above “Adjustment button”, this value will be in the order of 300
to 350 when the “Chest front” button is used and the chest front of a person is X-rayed
under the following conditions.
Automated ROI#3(CHEST PA)/ with back ground
CURVE SHAPE: CHEST
BRIGHTNESS: 16
CONTRAST: 10
Imaging unit Rex value
CXDI-40EG, CXDI-50GApprox. 300 - 350
CXDI-40EC, CXDI-50CApprox. 150 -175
WWiitthh aauuttoo RROOII,, aallll tthhee
aarreeaass aarree ttuurrnneedd ooffff..
- 30 -
Page 48
2. Installation Manual
5.7 Settings
(1) Checking and Setting the Date and Time
Description about CCR application in ‘Setting’ may change to some degree depending on the versions of
application. For CCR application, see “New Function Descriptions” issued for every version if necessary.
1) Purpose
The date and time is set to Japan standard time at factory shipment.
Reset the date and time to your local value as necessary.
2) Procedure
2-1) When CXDI application start, open the ADMINISTRATOR SETUP MENU.
SYSTEM → SETUP MENU → ADMINISTRATOR SETUP
2-2) When the “ADMINISTRATOR SETUP MENU” appears, and presses the [DATE]
button.
2-3) The dialog (Date / Time Properties) appears, and set the value properly each of the
fields which the tab sheet (Date&Time and Time Zone sheet) has. And then press
[OK].
- 31 -
Page 49
2. Installation Manual
(2) Checking the Firmware Version
1) Purpose
1-1) Failing to use the proper versions of the firmware and PLD code with the CXDI
application can result in an error, and system operation cannot be guaranteed.
Therefore, the versions of the firmware must be checked to ensure that they are
correct. (The combination of this firmware refers to CXDI Software Combination
List.)
2) Notes
2-1) This check should always be performed at installation, and if necessary, the firmware
versions should be upgraded.
2-2) This check cannot be performed with only the control PC. Connect the imaging units
and other equipment, and start up in the normal imaging status.
3) Procedure
3-1) Checking the firmware alone
a. Start up the CXDI system.
b. Display the version information from the user mode.
SYSTEM > SETUP MENU > VERSION INFORMATION
c. Confirm the firmware version.
A BC
A. Hardware version
Product type and Sub No.: Product type is identified from Sensor serial No. is set to the
Imaging unit (07=50G and 0a=50C). Sub No. indicates the states of the dipswitch2-1 to 4.
B. Firmware initialization code version
This is the version of the initialization code written on the PWB-50Di. Initialization code
will be downloaded and settings will be reset to the default (factory) settings by turning
ON the power while pressing the initialization switch on the power box.
Display “1.1.0” on the screen indicates version 1.01.00.
C. Firmware normal code version
This is the version of the normal code installed on the PWB-50Di. Usually the system
operates with this code. It must be updated as required.
Display “1.1.00” on the screen indicates version 1.01.00.
- 32 -
Page 50
3-2) Checking the firmware and PLD code
(1) Connect the keyboard and mouse.
(2) Start up the CXDI system.
(3) Close the CXDI host software if it starts up.
(4) Connect Telnet by referring to “Telnet Connection” in the Tool Software Operation
Manual for Ethernet.
(5) Check the versions of the firmware and PLD code on the screen displayed after the
login.
login: canon
Password: ********
CXDI-50G firmware Ver.1.01.00
build as Jul 2 2003 16:13:45
normal boot
HUB50 Ver.5010 product code:0701
pcb-50di>
(6) After you finish checking, close HyperTerminal.
2. Installation Manual
Firmware version
Hardware version
PLD code version
- 33 -
Page 51
2. Installation Manual
(3) Installing Firmware and PLD Code
1) Purpose
Write exposure code and PLD code into the Flash ROM of the PWB-50Di in the imaging
unit.
2) Notes
Be sure to check that the CXDI is connected to the system.
3) Procedure
3-1) Installing the firmware
Write the firmware by referring to “Firm Write Tool Software (Firmwrite.exe)” in the Tool
Software Operation Manual for Ethernet.
Where to write: IC1 (Flash ROM) on PWB-50Di
3-2) Installing PLD code
Write PLD code by referring to “PLD Write Tool Software (pldwrite.exe)” in the Tool
Software Operation Manual for Ethernet.
Where to write: IC11 (PLD) on PWB-50Di.
- 34 -
Page 52
2. Installation Manual
(4) Checking the Sensor Serial No.
1) Purpose
If the sensor serial number and the sensor serial number stored in the flash ROM of the
PWB-50Di differ from the image data file name stored in the hard drive of the Control PC
due to replacing the PWB-50Di or imaging unit, the connected sensor can not be detected
after the CXDI application is launched. In that case an error message appears.
2) Notes
2-1) Check the sensor serial No. whenever:
a. PWB-50Di is replaced.
b. Sensor in the imaging unit is replaced.
2-2) This checking procedure must be performed with the Control PC, Imaging Unit and
all the other equipments connected and started up.
3) Procedure
3-1) Check the sensor serial No. by referring to “Sensor serial number setting” in the
Tool Software Operation Manual for Ethernet, and write the number as required.
- 35 -
Page 53
2. Installation Manual
(5) Set Up Startup Menu
1) Purposes
1-1) Register the CXDI application software to the “Startup Group”.
The CXDI application software is scheduled to start automatically at the CXDI system
starting
1-2) Change the window view size
Hide the other application screen view except the CXDI application software.
1-3) Delete the CXDI application software from the “Startup Group”.
The CXDX application software is not started at the CXDI system starting.
2) Notes
2-1) The CXDI application software is not registered in the “Startup Group” at the factory
setting.
Therefore register the CXDI application software to the “Startup Group” after the
system installation.
2-2) The window view size of the program registered in “Startup Group” has one own size
with the each short-cut icon. Be sure to set the window view size of CXDI application
software at the same time with the register to the “Startup Group”.
3) Register the CXDI application software to the “Startup Group” procedure.
3-1) Connect keyboard and mouse to the control PC.
3-2) Turns the all CXDI system power on after the all installation finished. And after that
Windows XP starts.
3-3) Open the “Taskbar and Start Menu” from the Start Menu.
Start⇒Settings⇒Taskbar and Start Menu
3-4) “Taskbar and Start Menu Properties” appears. Click “Start Menu” tab, and then click
Taskbar and Start Menu Properties⇒Start Menu⇒Classic Start⇒Menu Customize
3-5) Click Add, and Create Shortcut appears. Click Browse.
3-6) Browse appears. Find a file named “ccrstart.bat” in drive [D:\ccr] and click OK.
3-7) D:\ccr\ccrstart.bat appears in the Command line. Click Next.
3-8) Select Program Folder appears. Select Startup folder and click [Next].
3-9) Select a name for the shortcut appears. Type ccrstart.bat. Click [Finish].
3-10) Close the Taskbar [Start], and login again to Windows XP.
Start⇒Shut Down⇒Log off cxdi.
3-11) After login the computer, make sure that the CXDI application starts up.
- 36 -
Page 54
2. Installation Manual
4) Change the window view size
4-1) After the CXDI application software start, press [Alt] + [Tab] key to show the “Debug
mode” prompt screen.
4-2) After the command prompt screen appears, click the icon (called System icon) where
is in right-top of its window.
4-3) System icon menu appears. Select Properties from the menu. [Fig 1]
Select the Properties item.
[Fig 1]
4-4) Click the “Font” tab from the “ccrstart.bat” properties and change its size to “6 x 13”.
4-5) Click the “Layout” tab and change the “Height” of the “Screen Buffer Size” to 5000.
Click [OK].
4-6) The “Apply Properties to Shortcut” appears and check the item of the “Modify
shortcut which started this window”. Click [OK].
5) Delete CXDI application software from the “Startup Menu Group”.
5-1) Connect the keyboard and the mouse to the control PC.
5-2) Turns the CXDI system power on, Windows XP start.
5-3) After the CXDI application software start, press [Alt] + [Tab] key to show the “Debug
mode” prompt screen.
5-4) Select “8 – Exit” to close the CXDI application software on the “Welcome to CCR”.
5-5) After the CXDI application software closed and Window XP Desktop appear, open
the “Taskbar & Start Menu…” with “Start Menu” tab clicking.
Start>Settings>Taskbar and Start Menu
- 37 -
Page 55
2. Installation Manual
5-6) “Taskbar and Start Menu Properties” appears. Click “Start Menu” tab, and then click
Taskbar and Start Menu Properties⇒Start Menu⇒Classic Start⇒Menu Customize
5-7) The “Remove Shortcuts/Folders” dialog box appears after click the “Remove” button.
And double-click the “Startup folder”
5-8) Remove the “ccrstart.bat” item from it.
5-9) After “Remove” button clicked, the confirmation of deleting file appears. If you are
going to remove it, click “Yes” button.
5-10) After confirm that the “ccrstart.bat” item is removed from “Startup Group”, close all
the application on the desktop and re-login to Windows XP.
5-11) Make sure that the CXDI application software will not start automatically after login
to Windows XP. And then shutdown Windows XP, turn the CXDI system power off.
* When the CXDI application is deleted from the Start menu due to repair or other
reasons, be sure to always perform the procedures outlined in “Adding CXDI
application software onto the Start menu” and “Changing the window size” when
the repair is complete.
- 38 -
Page 56
2. Installation Manual
(6) Identifying the Sensor Units
1) Purpose
In order for the control PC to identify the sensor units connected, the sensor serial number
of each sensor unit is input to the Control PC.
2) Notes
2-1) These operations must always be implemented at the installation stage and when any
of the Imaging Units (sensor) or Control PC (hard disk) has been replaced or when the
combination of equipment has been changed.
2-2) The sensor serial numbers must always be input. If the serial numbers of the sensor
unit and Control PC do not match, “Sensor Unit: Detect Error (-5100)” will be
displayed on starting up the system. These numbers are the same as what is input to
the PWB-50Di. (Refer to “ Checking the sensor serial numbers”.)
3) Preparations (What to have ready)
Tool keyboard, tool mouse
4) Procedure
4-1) Start up the CXDI unit.
4-2) Once the normal sensor screen has appeared on the operation unit, use the keyboard to
enter the debugging mode (Use [ALT] + [TAB].).
4-3) “Welcome to CCR” appears. Select “1 Set-Up...”
-------A/D Board Serial Number 0-0 –> 50G: 0x10000004
A/D Board Serial Number for SensorID#1 [0x10000004=268435460]:010000004
Custom Type [0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE]
[0=0x0]:0
Field of View Rotation (0:No 1:Yes) [0=0x0]:0
EXI constant [0.000000]: 0.000000
---- Need to re-start program to validate this change.
4-8) When “CCR SETUP MENU” appears, press the [Esc] key to return to “Welcome to
CCR.”
4-9) Select the command “8 - Exit” from “Welcome to CCR” menu to exit the CXDI
application.
4-10) After “Windows XP desktop” screen appears, start the CXDI application again.
(7) Entering Control PC Serial Number
1) Purpose
Set the product serial number (Control PC) to the “Device Serial Number” of the
“DICOM header”.
- 39 -
Page 57
2. Installation Manual
2) Procedure
2-1) Start up the CXDI system.
2-2) After the exposure screen appears on the operation unit, use the keyboard to enter
Debug mode. (Use [Alt] + [Tab].)
2-3) “Welcome to CCR” screen appears. Select the command “1. Set-Up…”
2-4) The “Setting Mode (0: Normal, 1: Expert) [0=0×0]:” is prompted. Select “0: Normal”
2-5) “CCR SETUP MENU” appears. Select the command ”1. System Setup”.
2-6) “CCR Serial Number [0=0×0] : ” appears. Enter the six-digit number indicated on the
naming label of the control PC unit. Press [Enter] key until “CCR SETUP MENU”
appears. [Fig 1]
Enter here
[Fig 1]
2-7) Press [Esc] key after “CCR SETUP MENU” appears to return to “Welcome to CCR”
2-9) This returns you to the Windows NT desktop. Restart the CXDI application, and
perform the procedure from steps 2) to 5). Check that the serial number for the “CCR
Serial Number” item was entered correctly in step 6).
* Restart the CXDI application. The screen displays the following message:
Alert System Info Error (-6) A/D board info is updated. Click “OK”
- 40 -
Page 58
2. Installation Manual
(8) Table Setup Settings
1) Purpose
Adjust the CXDI operation unit's TABLE SETUP to match the exposure conditions (Xray tube voltage, X-ray tube current, msec or mAs value) of the X-ray generator.
2) Procedure
2-1) Start the CXDI system.
2-2) Open the TABLE SETUP Change window from the Normal Exposure window.
System ⇒ SETUP MENU ⇒ SYS. SETUP ⇒ TABLE SETUP
2-3) Select the tabs to be changed and change the X-ray tube voltage, X-ray tube current,
and msec value data to match the exposure conditions of the X-ray generator.
* See the operation manual for the details of settings.
2-4) After finishing the changes, return to the Normal Exposure window and check that the
TABLE SETUP has been changed.
- 41 -
Page 59
2. Installation Manual
(9) Performing the Annotation Settings
1) Purpose
The settings for imprinting the annotation onto the film and the settings of the
characters used for the annotation are performed.
2) Procedure
2-1) Once the normal radiographic screen has started, open the annotation setting screen.
SYSTEM → SETUP MENU → SYS. SETUP → ANNOTATION
2-2) The annotation setting screen now appears. Proceed with the settings that will make it
possible to put the data desired by the user.
* See the operation manual for the details of settings.
- 42 -
Page 60
2. Installation Manual
(10) Network Connections
Network settings
1) Purpose
These settings are for connecting the CXDI to the network.
1-1) Set the CXDI’s IP address, subnet mask and default gateway in Windows XP.
1-2) Set the printer and storage output destinations and parameters on the user screen.
2) Checkpoints
2-1) This item involves checking the details of the checks performed on network setting
parameters among the pre-installation inspection details and setting these parameters.
* Refer to “Appendix: Investigation Report” for the pre-installation investigation
details.
2-2) Perform the settings of this item carefully since any errors made in these settings will
make it impossible for connection to be made to the network or the images to be
transmitted properly, etc.
3) Windows XP settings
3-1) Connect the keyboard and mouse to the control PC.
3-2) After turning on the Operation unit’s power and then the Control PC’s power, start
Windows XP.
3-3) The Windows XP desktop screen appears. Right-click the [My Network] icon, and
select My Network Places from the menu.
3-4) When [Network Connection] appears, double click on Local Area Connection (Intel
(R)
PRO/100VE Network Connection).
3-5) When Local Area Connection Properties appears, click on the General tab, select
[Internet Protocol (TCP/IP)], and click Properties.
3-6) Based on the pre-install of inspection details set the IP address, subnet mask and
default gateway.
[Fig 1]
- 43 -
Page 61
2. Installation Manual
3-7) Upon completion of the setting, restart the Windows XP.
3-8) Check the communication test in the sequence below to verify whether the CXDI is
now part of the network. To check the connections at the TCP/IP level, use the “ping”
command from the command prompt.
Start Æ Programs Æ Command Prompt
When the IP address of the connection destination is “173.17.7.123,” for instance, the
following messages will be repeated.
y If the CXDI has been connected properly:
Pic:>ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=2550
• If the CXDI has not been connected properly:
Pic:> ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Request time out
Request time out
Request time out
Request time out
- 44 -
Page 62
2. Installation Manual
4) Set the printer and storage device which serves as the external output destinations.
In this case, one printer and one storage device are set.
4-1) Printer settings
A. Open the output destination setting screen from the user menu.
System Æ SETUP MENU Æ DESTINATION Æ PRINTER
* Up to four printers (2 of which can be used for output at the same time) can be
set.
B. Press the “Printer1” button, and input the following items based on the pre-
installation investigation details.
a. Printer host name (IP address) b. Port number c. Transmission destination title
a
b
c
[Fig 2]
C. Press the “SET” button, and input the parameters of the printer to be connected
based on the pre-installation inspection details. (Refer to another sheet for details
of the parameters.)
* A space delimiter must be input between each of the parameters.
By pressing the “Override” button, you can select a printer from all the registered
printers. In this case, basically you do not have to enter parameters. However, if
“?” is displayed within the parameters, you may have to enter the required
parameter at the user’s site.
- 45 -
Page 63
2. Installation Manual
Input parameters
[Fig 3]
- 46 -
Page 64
2. Installation Manual
4-2) Storage settings
A) Open the output destination setting dialog from the user menu.
System Æ SETUP MENU Æ DESTINATION Æ STORAGE
* Up to four storage units (2 of which can be used for output at the same
time) can be set.
B) Press the “Storage1” button, and input the following items based on the pre-
installation investigation details.
a. Storage host name (IP address) b. Port number c. Transmission destination
title
a
b
c
[Fig 4]
C. Press the “SET” button, and input the parameters.
(Normally, the parameters need not be set. They must be input only when the
need arises.)
* A space delimiter must be input between each of the parameters.
[Fig 5]
- 47 -
Page 65
2. Installation Manual
D. After setting the output destinations, follow the procedure below to check
whether images can actually be transmitted. Return to the user menu, capture a
sample image (one X-ray image), and transmit the image to the printer and
storage. There are two errors that may result if the image cannot be transmitted:
a. “DICOM Connect Error. Cannot connect to the target. Check network or port
number setting. Retry?”
b. “DICOM Transfer Error. Error occurred during the association. Retry?”
Message (a) indicates that connection at the TCP/IP level is not possible and that the
physical connections or the subnet mask and other settings must be checked again.
Message (b) indicates that communication at the TCP/IP level is problem-free but
that DICOM level communication has failed. In this case, check again that
AE_TITLE of CXDI has been sent properly to the transmission destination and that
the IP address, port number and AE_TITLE of the transmission destination which are
set with CXDI have been set properly.
* “AE_TITLE” of the transmission destination is case sensitive fields.
(Permit upper-case letter or lower case letter, etc)
- 48 -
Page 66
2. Installation Manual
Parameter List (Separate Document 1)
DICOM storage device
In the CXDI, DICOM data transfer is performed using the transfer software module
“send_image” .The settings for these parameters are described below.
Parameter Meaning Description
-m maxPDU
Maximum PDU value in
byte units
-t calledTitle
Called App Entity Title
-c callingtitle
calling App Entity Title
-s SOPName
(for reference)
This parameter designates
whether class be connected
for performing association
at the beginning of
transfer.(CR/T/MR/NM/S
C/US)
-I
A-RELEASE-RES is
ignored.
-d FAC
This parameter dumps a
specific facility log.
(DCM/DUL/SRV)
* The CXDI automatically
uses 131072 internally for
operation.
* Designating a specific value
allows overwriting of the
above value.
* The AE Title setting field is
automatically applied to this
setting.
* Designating a specific value
allows overwriting of the
above value.
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* Designating a specific value
allows overwriting of the
above value.
* This is not used in the
CXDI.
* This parameter is used
simply as “-I”
* This parameter is used
simply as “-d”
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This is used when the operator who
manages the connected storage device
requests a size change.
* In DICOM printing, note that the
argument title changes to -u.
(Æ See the printing parameters.)
* Note that the meaning is opposite of the
DICOM printing argument -c.
(ÆSee the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
* Note that the meaning is opposite of the
DICOM printing argument -c.
(ÆSee the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
* This is used when the error message
“30012 Peer aborted Association (or
never connected)” occurs even though
the DICOM data transfer was
successful.
Æ This is used differently based on the
connected storage devices.
* This parameter does not affect DICOM
data transfer.
* This parameter outputs the CXDI log
based on Windows NT.
- 49 -
Page 67
2. Installation Manual
Parameter Meaning Description
-v
This parameter dumps the
transfer log.
-jn
This is the time to take
timeout.
-k
level=0
1
2
3
* This parameter is used
simply as “-v”.
* DUL and SRV are dumped.
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* Sets the time to take timeout
in seconds.
* A variety of specifications
have since been needed in
conjunction with DICOM
modality LUT support.
* This parameter does not affect DICOM
data transfer.
* This parameter is to be changed when
taking timeout.
If DICOM Modality LUT OD is enabled,
set appropriate options to suit each
output destination.
Argument: Values necessary for executing a function, subroutine, procedure, or other operation is
passed to them. Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (x) is x.
PDU: Protocol data unit
The types of PDU’s include get-request, get-next-request, get-response, set-request, nd trap.
Note
For details about the “-v” parameter, see “Checking the Error Log”.
Note
The parameters “-v” and “-d” put the log on the console. Therefore, be sure to always erase
these parameters before operation by the user.
If –k:DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
level = 0: Do not delete (default when not specified)
1: Delete Window Center/Width
2: Delete Window Center/Width and Rescale Intercept/Slope/Type
3: Delete Rescale Intercept/Slope/Type (compatible with releases up to Ver.4.20)
Other than a loadable LUT or γ=1.0 has been specified with the output destinationspecific LUT function.
For storage: Specify -k3(or -k2).
If IMG Rescale Type = 0D is unidentifiable to storage, resulting in an error: Specify -k3.
(If only one storage is connected, simply set DICOM Modality LUT OD to "Disabled.")
The implementation allows Window Center/Width with Rescale Type = 0D specified to
be interpreted as "optical densityx1000."
For storage: Specify -k1.
The CXDI Window Center/ Width output value (implementation) is fixed at 2048/4096.)
Note
With AGFA impax Ver. 4.5.0, the specification of -k0 demonstrated a successful densityintensity conversion. But because the corresponding text in the DICOM specifications
document is ambiguously written such that the status of implementation by other
manufacturers is unknown, please be advised to consult the storage manufacturer for each
connection destination or work out on a trial and error basis. Also note that an external
storage option specification (if (DICOM modality LUT OD is set to "Enabled," OD tags
<0028, 1052 - 1054> are assigned and Level: 0 is assumed) is not supported.
- 50 -
Page 68
2. Installation Manual
Parameter List (Separate Document 2)
DICOM printer
In the CXDI, DICOM printers are administered separately according to printer product.
The transfer software module is “print_stuff”. The settings for these parameters are described below.
Parameter Meaning Description
-C copies
This parameter uses a
number to designate the
number of copies.(1/2/...)
-y priority
Priority in the DICOM
printer
(HIGH/MED/LOW)
-D destination
Film destination
(MAGAZINE/PROCESS
OR/BIN_i)
-F film type
Film media type (“BLUE
FILM” / “CLEAR FILM”
/ “PAPER”)
* This parameter is used in the
DICOM Basic Film Session
(2000, 0010).
* When the number of copies
is designated, film sheets are
printed in the quantity
specified in a single printing
operation.
* This parameter is necessary
when printing multiple
sheets for a single data
transfer operation.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0030).
* This parameter determines
where this transfer image is
inserted into the queue in the
DICOM printer. At HIGH,
the image is printed first
among the queued images.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0040).
* Film is sent to the output
device designated by
RECEIVE MAGAZINE or
the automatic developer.
* This parameter is used in the
DICOM Basic Film Session
(2000, 0030).
* Film is printed as the
designated film type.
* This parameter is set according to the
user’s requirement.
* In the DICOM library TYPE3, the value
is transferred together with the Tag.
However, if the value is unknown, the
value is either is transferred as a
character string with length 0, or the
element itself is not transferred.
Æ The printer default values are used if
this parameter is not entered.
* This parameter is set according to the
user’s requirement.
* In the DICOM library TYPE3, the value
is transferred together with the Tag.
However, if the value is unknown, the
value is either transferred as a character
string with length 0, or the element itself
is not transferred.
Æ The printer default values are used if
this parameter is not entered.
* Note this parameter does not determine
where this transfer image is inserted into
the CXDI queue.
Æ The printer default values are used if
this parameter is not entered.
* The film is usually discharged to the
default output device.
Æ The printer default values are used if
this parameter is not entered.
* Although many types of films cannot be
detected, the film type can be selected in
the KELP2180.
* In this parameter, be sure to put
quotation marks (“ ”) around 0x20 since
it comes between BLUE and FILM.
- 51 -
Page 69
-L sessionLabel
Film session label
(character string)
-f films
Number of film box to be
printed
–i Format
Format at print
-l FilmSizeID
Film size
14 inch x 17 inch / 17 inch
x 14 inch / 11 inch x 14
inch / -l FilmSizeID / 10
inch x 14 inch
10 inch x 12 inch / 24 cm
x 24 cm / 24 cm x 30 cm
Æ The parameter is not transferred over
DICOM if it is not designated.
* This parameter may be displayed in
some form or another depending on the
installed printer. For example, it may be
displayed in the Control Panel for the
printer or in the corner of the film.
* If this parameter is not designated,
transfer software uses automatically
STANDARD1 1, for reason this
parameter must be transferred in the
DICOM.
Æ The parameter is not transferred over
DICOM if it is not designated. In this
case, the printer default values are used.
When this parameter is not transferred,
problems can occur since unsuitable
default values may be used.
* Some printers do not print until a supply
magazine of the designated size is
loaded, and others print even though the
designated size is different from the
currently loaded supply magazine.
Æ The printer default values are used if
this parameter is not entered. When this
parameter is not transferred, problems can
occur since unsuitable default values may
be used.
Æ The parameter is not transferred over
DICOM if it is not designated. In this
case, the printer default values are used.
When this parameter is not transferred,
problems can occur since unsuitable
default values may be used.
* This parameter is determined by asking
the printer engineer or by viewing the
conformance statement.
- 52 -
Page 70
-S configuration
Adjustment information
(character string)
-O Orientation
Film orientation
(PORTRAIT/LANDSCAP
E)
-A max_density
Maximum density
(Dx100)
–a min_density
Minimum density (Dx100)
2. Installation Manual
* The designation method
varies according to the
printer. For example, the
MLP190 uses -m
NORMAL.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0150).
* This parameter sets the
printer (image quality)
adjustment from the SCU
side.
* In the DICOM standards,
value to be transferred is not
predetermined.
* The designation method
varies according to the
printer.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0040).
* In versions before 2.0,
printers must operate based
on this parameter.
* When using image cutout
from 17 x 17 inch size in the
CXDI, this parameter is set
and transferred
automatically.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0130).
* This parameter designates
the density of the digital
value for 0 (4095 for reverse
display) of the CXDI
transfer data image pixels.
(In the CXDI, 0 indicates
black.)
* In the CXDI, this parameter
is used to adjust the density.
Therefore, be sure to always
check that it is operating.
* This parameter is used in the
DICOM Basic Film Box
(2010, 0120).
* This parameter designates
the density of the digital
value for 4095 (0 for reverse
display) of the CXDI
transfer data image pixels.
(In the CXDI, 4095
indicates white.)
* This parameter is not
transferred in many cases
since the minimum density
cannot be increased in most
printers.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
* This parameter is determined by asking
the printer engineer or by viewing the
conformance statement.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used.
* Starting from version 2.0, the image can
be rotated from the CXDI side without
using this parameter.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
Æ If this parameter is not designated, it is
not transferred over DICOM. In this case,
the printer default values are used. When
this parameter is not transferred, problems
can occur since unsuitable default values
may be used.
- 53 -
Page 71
-B border_density
Border density (Dx100)
(BLACK/WHITE/D x
100)
-G empty_image_density
Empty image density
(BLACK/WHITE/D x
100)
-T trim
Trimming (NO/YES)
-P polarity
Polarity
(NORMAL/REVERSE)
-r pixel_pitch
Transfer pixel pitch for
designating the request
image size (Pixel Pitch in
um)
-N annoFmt
Annotation position
(1/2/3)
-n annotation
Annotation (character
string)
2. Installation Manual
* This parameter is used in the
DICOM Basic Film Box (2010,
0100).
* This parameter determines the
area density around the image
on the film.
* This parameter is used in the
DICOM Basic Film Box (2010,
0110).
This parameter designates the
density of the empty image
area during multi-formatting.
* This parameter is used in the
DICOM Basic Film Box (2010,
0140).
* This parameter adds lines
around the image.
* The CXDI is normally adjusted
so that the trimming does not
appear.
* This parameter is used in the
DICOM Basic Image Box
(2020, 0020).
* Reverse image density
* This parameter is used in the
DICOM Basic Image Box
(2020, 0020).
* Reverse image density
* This parameter is used in the
DICOM Basic Image Box
(2020, 0010).
* Position of the image on film
* The cumulative value for the
horizontal size of the image
(raw) at the designated pixel
pitch is used for the request
image size.
* The CXDI automatically uses
131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0010).
* This parameter designates the
position of the character string
to be annotated.
* This parameter is used in the
DICOM Basic Annotation Box
(2030, 0020).
* This parameter designates the
character string to be
annotated.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
* For example, the default value for
this parameter in Agfa printers is
YES. Therefore, the NO option
needs to be specifically designated if
it is desired.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the printer default values are
used.
Æ If this parameter is not designated,
it is not transferred over DICOM. In
this case, the type of image that is
printed depends on the settings at the
printer side.
* If using annotation, always be sure to
transfer the annotation position.
* If this parameter is not designated, it
is not transferred over DICOM. In
this case, the printer default values
are used.
* Also, in this case, the type of image
that is printed depends on the
settings at the printer side.
- 54 -
Page 72
-u maxPDU
Maximum PDU value in
byte units
-t callingTitle
Called App Entity Title
-t callingTitle
Calling App Entity Title
-g
N-GET Printer
compatibility mode
-S
Silent mode
-p
This parameter dumps the
association parameter.
-v
This parameter dumps the
transfer log.
2. Installation Manual
* The CXDI automatically
uses 131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
* The AE Title setting field is
automatically used in this
setting.
* The above value can be
overwritten by designating a
specific value.
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* The above value can be
overwritten by designating a
specific value.
*This parameter is used
simply as “-g”.
* In the CXDI default settings,
the printer information is not
designated. In this case, the
printer side sends all the
information that it has
(DICOM official
specifications).
* When the -g option is added,
the essential information
only is collected. This
information includes the
Printer Status and Printer
Status Info. (To prevent
installation when the printer
does not satisfy the above
DICOM specifications.)
* This parameter is used
simply as “-s”.
* This parameter is used to
prevent the transfer software
from displaying the
debugging character string
on the console.
* This parameter is used
simply as “-p”.
* This parameter is used to set
the transfer software so that
the debugging character
string is displayed on the
console.
* This parameter is used
simply as “-v”
* This parameter is used to set
the transfer software so that
the debugging character
string is displayed on the
console.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This parameter is used when the
operator of the connected storage device
requests a size change.
* In DICOM storage devices, note that the
argument title changes to -m.
(ÆSee the storage device parameters.)
* Note that the meaning is opposite of the
argument -c for DICOM storage
devices.
(ÆSee the storage device parameters.)
* The entry for the OPU output device
title is used here.
* Note that the meaning is opposite of the
argument -t for DICOM storage devices.
(ÆSee the storage device parameters.)
Æ Normally, this option is not used.
This parameter has been provided as a
remedy when a printer error occurs when
optional devices are not used.
* This parameter does not affect DICOM
data transfer.
Æ Silent mode does not need to be
designated since the CXDI automatically
makes the setting internally.
* This parameter does not affect DICOM
data transfer.
* This parameter does not affect DICOM
data transfer.
- 55 -
Page 73
2. Installation Manual
* Both the -p and -v
parameters should be used.
These settings override the -
s parameter.
-V filename
This parameter dumps the
transfer log.
-I
A-RELEASE-RES is
ignored.
-jn
This is the time to take
timeout.
-k
level = 0
1
2
* The parameter is used for
analysis after the transfer
software saves the
debugging character string
displayed on the console to a
file with a designated
filename.
It is used only when
problems occur.
* This parameter is used
simply as “-I”
* Sets the time to take timeout
in seconds.
* The -k option has been
implemented to normalize
DICOM headers
* This parameter does not affect DICOM
data transfer.
* This is used when the error message
[30012 Peer aborted Association (or
never connected)]occurs even though the
DICOM data transfer was successful.
Æ This is used based on the connected
printers.
* This parameter is to be changed when
taking timeout.
Note
The parameters -p, -v, and -V filename display the log on the console. Therefore, be sure
to always erase these parameters before operation by the user.
In the past, -k2 was used to fix troubles, but it has now been set as the default has been changed
to -k2 to ensure precise compliance with the DICOM code. At sites wishing to adhere to their
existing window values, -k1 should be used.
level: 0:Do not delete DICOM tags. (Transfer all headers similar to storage.)
1:Delete Groups 0008, 0010, 0018, 0019 and 0020.
2:Delete Groups 0008, 0010, 0018, 0019, 0020 and Elements (0028,0030), (0028,1050),
In Ver.4.21, attach -k2 expressly to produce similar output. The previous option had -k1
as its default. The k option defaults to -k2, if k option is not specified.
Image Pixel Spacing (0028,0030)
Window Center (0028,1050)
Window Width (0028,1051)
Rescale Intercept (0028,1052)
Rescale slope (0028,1053)
Rescale Type (0028,1054)
For groups 0008, 0010, 0018, 0019, 0020, refer to DICOM manifesto.
- 56 -
Page 74
2. Installation Manual
Examples of parameters used with different makers and types of printers (reference)
Name Default Parameters Significance
Kodak MLP190 -A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
80 (entered in pixelPitch
member)
14 x 17
4096 (entered in W member)
5120 (entered in H member)
Kodak KELP2180 +
Kodak Print Spooler
Model 100
Agfa DryStar 3000 -A 320 -T NO -M CUBIC -m
Kodak Imation DryView
8700 + Pacs LINK IMN
9410
Kodak Imation DryView
8700 + GW
Kodak Imation DryView
8700+8800
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
79 (entered in pixelPitch
member)
14 x 17
4090 (entered in W member)
5120 (entered in H member)
11 x 14
3194 (entered in W member)
4096 (entered in H member)
140 -S
“PERCEPTION_LUT=200”
(entered in param member)
80 (entered in pixelPitch
member)
14x17
4256 (entered in W member)
5174 (entered in H member)
-A 310 -T NO -M CUBIC m ? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
-A 310 -T NO -M CUBIC m ? -S LUT=?, ? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
• Trimming OFF
• Cubic spline interpolation
• Smoothing: normal
• Maximum density: 3.20
• Curve shape 0 (density linear) as Config
Info
• Trimming OFF
• Cubic spline interpolation
• Smoothing: normal
• Maximum density: 3.20
• Curve shape 0 (density linear) as Config
Info
• Trimming OFF
• Cubic spline interpolation
• Smoothing: slightly sharp (edges
emphasized)
• Maximum density: 3.20
• S
“PERCEPTION_LUT=200(LINEAR)”
(If the output fails to be linear with
“LINEAR”, on-site adjustments with the
printer manufacturer must be performed.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
- 57 -
Page 75
2. Installation Manual
p
• The Kodak service engineer is
responsible for setting the smoothing
type on-site since it can be changed with
each printer.
• The rest is done by the printer itself.
Kodak Imation DryView
8700+9440
Kodak Imation Dry View
8500+
Nishimoto EL2000N -A 320 -T NO -M CUBIC -S
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
The model 8800 has a rotation
function but we understand
that Imation has not publicly
acknowledged the use of this
function.
-A 310 -T NO -M CUBIC m ? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
3388 (entered in W member)
4277 (entered in H member)
15
(entered in param member)
80 (entered in pixelPitch
member)
4444 (entered in portraitW
member)
5296 (entered in portraitH
member)
5296 (entered in landscapeW
member)
4444 (entered in landscapeH
member)
“Prepare images using
CXDI” : Yes
A simple calculation yields a
resolution of 4444 x 5400 for
the display area of the model
EL2000. However, 5376@80
µm is set in the perpendicular
direction of the model EL2000
since the maximum size of the
CXDI images is 2688@160
µm. In this case, a small nonimage area should be output as
the border in the up/down
direction on the film according
to the calculation. In actual
fact, however, the image will
protrude in some cases
depending on how the
trans
ort speed and other
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• S LUT = m, n is designated as the
Config Info but m and n are adjusted onsite by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
• The Kodak service engineer is
responsible for setting the smoothing
type on-site since it can be changed with
each printer.
• The rest is done by the printer itself.
• Trimming OFF
• Cubic spline interpolation
• Smoothing must be adjusted at the user’s
site.
• Maximum density: 3.10
• -S
• LUT = m, n is designated as the Config
Info but m and n are adjusted on-site by
the Kodak service engineer. Basically,
adjustment is performed to achieve a
linear output.
• Trimming OFF
• Cubic spline interpolation
• Maximum density: 3.20
• “15” in Config Info is linear.
• The rest is done by the printer itself.
Up to 5376 pixels can be set for H.
- 58 -
Page 76
2. Installation Manual
factors have been adjusted.
The specifications are as
follows although they cannot
be set at the present time:
14" x 14" : W=4444, H=4444
(for portraits)
11" x 14" : W=4444, H=3660
(for portraits)
Fuji CR-DPL/LPD/FMDPL + FN-PS551
Konica Drypro 722 +
Printlink
Konica Li-62P + Printlink -A 320 -T NO -M CUBIC -m
-A 300 -T NO -M CUBIC -m
MEDIUM -S –P NORMAL –
B BLACK –k 2 –S ? (Site
Dependent)
(entered in param member)
14x17
3520 (entered in W member)
4280 (entered in H member)
14x14
3520 (entered in W member)
3490 (entered in H member)
11x14
2540 (entered in W member)
3600 (entered in H member)
-A 320 -T NO -M CUBIC -m
2 -S “KC_LUT=1” –O
PORTRAIT –P NORMAL –B
BLACK
(entered in param member)
80 (entered in pixel pitch
member)
14x17
4424 (entered in W member)
5324 (entered in H member)
14x14
4424 (entered in W member)
4372 (entered in H member)
11x14
3436 (entered in W member)
4424 (entered in H member)
2 -S “KC_LUT=1” –O
PORTRAIT –P NORMAL –B
BLACK
80 (entered in pixel pitch
member)
14x17
4268 (entered in W member)
5108 (entered in H member)
14x14
4268 (entered in W member)
4104 (entered in H member)
11x14
3204 (entered in W member)
4268 (entered in H member)
• Trimming OFF
• Cubic spline interpolation
• SHARP, MEDIUM or SMOOTH can be
selected from among the presettings as
the smoothing type. An AVR of 0.8 or
so is appropriate. The setting is
performed for each printer on-site.
• -S should be adjusted at the user’s site.
• LUT can be selected from among the
eight presettings 1 through 8 using
Config Info. The setting is performed
for each printer on-site.
• With -k 2, the Window Center/Level for
DICOM TAG (0028,1050) and (0028,
1051) are also deleted.
• Maximum density: 3.00
A density of 3.20 cannot be designated.
For this reason, a non-linear LUT is
required.
• Trimming OFF
• Cubic spline interpolation
• Smoothing type
1: BILINEAR
2: Sharp by spline interpolation
3: Slightly weak by spline interpolation
4: Weaker by spline interpolation
• Maximum density: 3.20
Maximum density 3.20 could not be
achieved before.
• Trimming OFF
• Cubic spline interpolation
• Smoothing type
1: BILINEAR
2: Sharp by spline interpolation
3: Slightly weak by spline interpolation
4: Weaker by spline interpolation
• Maximum density: 3.20
Before, maximum density 3.20 could
not be achieved.
- 59 -
Page 77
2. Installation Manual
Printer Model Specifications (Reference)
Name Specifications
Kodak MLP190 80µm x 4096 x 5120 2048 x 2560 (@160 µm)
Kodak
KELP2180 +
Kodak Print
Spooler Model
100
79 µm x 4090 x 5120 (value after passing through the
print spooler)
• The above settings are the size of the effective area
when the image passes through the print spooler and
the image is plotted up to the annotation area. In other
words, these settings do not display an annotation area,
instead handling it as an image area.
• If the data is transferred without setting [Image creation
in CXDI], the Requested Image Size setting is used in
DICOM. In this case, the annotation area is
automatically displayed in the 2180 printer. As a result,
a maximum image area of 79 µm x 4090 x 4996 must
be designated. In this case, the CXDI relies on the 2180
for image rotation (Film Orientation), but images larger
than 1.7 MB cannot be rotated by the 2180. Therefore,
operation without the setting for [Image creation in
CXDI] cannot be performed in the 2180. (Although
operation is possible by setting Requested Image Size
only for using DICOM without the setting for [Image
creation in CXDI], this option is not installed in the
CXDI.)
• Although the resolution of the printer itself is 79 µm x
4090 x 5260, this complete resolution cannot be used
when the image passes through the spooler.
When “_” is used in AE Title, the association is
rejected.
Use the Disable function for N-EVENT-REPORT to
disable this setting.
The FilmSize parameter can be used. The MediaType
(BLUE, CLEAR) parameter is also supported.
• Although the Film Orientation parameter is supported
up to 1.7 MB, in actuality, DR images cannot be
rotated. Like the DryView8700, the images must be
rotated by the CXDI side.
• When Requested Image Size is expanded, the
maximum plotting size is limited (79 um x 4090 x
4996) so that the annotation area can be obtained.
When a Requested Image Size expansion error occurs,
the image is interpolated and printed at the suitable
size. In this case, the error does not return to the CXDI
side. For example, the image is printed at 310 mm even
if 326 mm is designated.
• If an expansion error occurs in the spooler when Multi
Display Format is used, the print queue cannot be
processed.
• The system is in a critical state when a Failure status is
indicated. A user message is displayed indicating this
state, and images are no longer transferred. (Fully
installed)
• During the Warning status, image transfer is performed
while the user message is displayed (Fully installed).
• If operation is aborted due to an error, a new
association could not be established when the data was
resent from the CXDI. GW was reset to recover the
error.
Maximum equivalent
area in CXDI
2018 x 2528(@160 µm)
- 60 -
Page 78
Agfa DryStar
g p
3000
Imation DryView
8700+8800
Fuji CR-DPL
/FM-DPL + FMPS551
2. Installation Manual
• Inserting annotations in the image can lead to problems
at the hospital. In the QCW, use annotations that are
outside of the image.
• The designated film size is 11 x 14 inch film, and
automatic selection of the magazine and printing has
been confirmed.
• Annotations in the image are problematic in the US and
EU. Thus, although DICOM annotation was used, it
was not printed. Although annotation can be transferred
without any errors in DICOM, an annotation error
occurred in the log when transferring from the gateway
to the 2180, and printing was not performed. Data was
transferred from KCR to the validation tool, and the
DICOM transfer method was compared to CXDI.
However, the only differences were in the image size,
aspect ratio, and annotation position.
• Although the annotation position is 1 in terms of
conformance, it is 0 as sent from KCR. The result did
not change even after the change.
80 µm x 4256 x 5174
• In the standard Agfa system, the annotation option is
selected. Therefore, the full image area will not be
printed correctly unless you ask the serviceman to
deselect the annotation option.
78 µm x 4096 x 5220
Note: The 8700 printer cannot display in landscape
orientation. Also, the maximum density is 3.1. If the 8800
box is not added, this printer cannot be used by the
CXDI. However, it can be used starting from CXDI
version 2.0.
• If there is an Imager Pixel Spacing tag (0018, 2264), the
imager will fail.
100 µm x 3520 x 4280 (value after passing through the
print spooler)
• The above settings are the allowable area size in a
configuration not using annotation.
• Annotation will be supported from the next version.
Annotation in currently possible in US-ASCII only. IDs
are designated 1 to 6 and correspond to the top left, top
center, top right, bottom left, bottom center, and bottom
right, respectively. The maximum area size with
annotation support is 3500 x 4170 for 35 cm x 43 cm
and 2538 x 3522 for B4.
• Use the Disable function for N-EVENT-REPORT to
disable this setting.
• A function is provided for disabling the returning of
warning messages.
0107 (Attribute list error)
Return/Not return
0116 (Attribute Value out of Range)
Return/Not return
B604 (Image has been demagnified)
Return/Not return
• The Film Size parameter can be used. The Media Type
(BLUE, CLEAR) parameter can also be used.
• The Film Orientation parameter is fully supported.
• The followin
resets are made so that LUT has the
2128 x 2587 (@160 µm)
1996 x 2544 (@160 µm)
2200 x 2675 (@160 µm)
- 61 -
Page 79
2. Installation Manual
DMAX = 3.2D equivalent curve required by CXDI.
Gamma type #17 (SAR system)
Density
Shift
Contrast
Point 1
1.57
0.10
1.00
Point 2
2.29
0.15
1.00
- 62 -
Page 80
2. Installation Manual
Error Return Values and Log Output for print_stuff (Reference)
Error example Return value and log output
Success Return value: 0x00
There is no log output in this case.
Invalid parameter Return value: 0x00000001
CXDI description : DICOM connection error
(CCRTRANS_ERR_DICOMPARAM)
The log output in this case is shown below.
----------------------------------------------------------------------------------** -s Silent mode; do not print results of all print commands
** -v Use verbose mode for DUL and SRV facilities
** x Canon Hidden Special Mode
node The host name that is running a print server
port TCP/IP port number of print server
file One or more files that contain preformatted images for printing
-----------------------------------------------------------------------------------The server has not
started up.
After a command
request was sent to
the server, an error
was returned in
response.
After a command
request was sent to
the server, a
warning was
returned in
response.
The log output in this case is shown below.
18-135933[d2]ERR: d0012 Attempt to connect to unknown host: test
18-135933[d2]ERR: 130012 Peer aborted Association
(or never connected)
18-135933[d2]ERR: 180012 Failed to establish association
Return value: 0x18
CXDI description : DICOM response warning
(CCRTRANS_WRN_RESP)
The transfer process was successful, but a warning was returned from the server.
The log output in this case is shown below.
Other errors Return value: Values other than those above
Return value: 0x28
CXDI description: DICOM printer status warning
(CCRTRANS_WRN_PRN_STATUS)
The transfer process was successful, but a warning was returned as the printer
status.
The log output in this case is shown below.
CXDI description : DICOM communications error
(CCRTRANS_ERR_DICOM_TRANSE)
The log output in this case depends on the specific error. A typical example is
shown below.
Transmitting the 0019 shadow group causes a failure, and the group is not received
properly with the default. Its reception is enabled by setting the strictValidation
parameter to Off in Miil.
y Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series under the Dr.ABLE specifications, the switching operations are a
hassle. The user will find it more convenient if it is at all possible to change the
setting to “1 study 1 series/1 series multi image.” (These unusual data
specifications were requested with the full understanding of their unusualness.)
y With DICOM, the body parts (such as the abdomen and head) and their directions
(such as PA and AP) belong to the series information. It therefore follows that a
different series is required for a different body part or body part direction. This
aspect is restricted by the DICOM standard rather than by the installation and
other steps taken by us. To put it the other way around, multiple images with
different body parts and their directions cannot be put together as a series. To
remedy this problem, devising a way of enabling the viewers to reference
different series at the same time at some future point in time will be helpful.
y Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series, the switching operations are a hassle. The user will find it more
convenient if it is at all possible to change the setting to “1 study 1 series/1 series
multi image.”
y This problem arises with the Fujitsu equipment as well. Refer to the section on
Fujitsu.
send_image error return values and log output (reference)
Example of
Return value and log output
error
Successful Return value: 0x00
No log output at this time
Invalid parameter
exists.
Server fails to start. Return value: 0x00180012
As a result of
providing the
server with a
command request,
an error was
returned as
response.
• Image adjustment parameters that can be changed by users
Density : This can be set up to the maximum density of 3.1D.
Contrast : This can be set from 1 to 15.
• Image adjustment parameters that cannot be changed by users
Lookup tables called TFTs (Transfer Function Tables) are provided internally, and changes
can be made only in the service mode.
Fifteen types of characteristic curves are registered in one of these TFTs, and users can change
one of these curves as the contrast.
Over 30 TFTs are registered in the printer, and a name is allocated to each one.
For instance, 15 types of linear straight lines are registered in the TFT called “WRKST2A.”
By setting this WRKST2A TFT, adjustments can be made by combining 15 types of linear
straight lines (which cannot be changed by users) with 15 types of characteristic curves
(which can be changed by users).
Concerning connections
Two types of the Dry View 8700 are available.
• Dry View 8700 Plus
The 8700 Plus is a printer which can be connected to two diagnostic units. When used in
combination with the 8800 multi input manager, it can be connected to up to eight diagnostic
units. Images are processed as described above.
• Dry View 8700 Dual
It is possible to connect two 8700 Dual units to the 8800 multi input manager. By using these
in combination, up to seven diagnostic units can be connected. The 8700 Dual does not come
with image processing functions
Modality connection I/F
This I/F is the external interface which connects “Dry View 8700” with each modality. Select
the following item depending on the modality which would be connected.
• Digital Signal: DEIB (Digital External Interface Box)
• Video Signal: EVEIB (Enhanced Video External Interface Box)
• Keypad, Auto Filming: UKEIB (Universal Keypad External Interface Box)
- 67 -
Page 85
2. Installation Manual
Dry View 8700 Plus
Up to 2 units can
be connected.
Dry View 8700 Plus + 8800 Multi Input Manager
Up to 8 units can
be connected.
Dry View 8700 Dual + 8800 Multi Input Manager
Up to 2 printers can be connected
(only with Dry View 8700 Dual)
Up to 7 units can
be connected.
- 68 -
Page 86
2. Installation Manual
(11) Linearity Check of Transfer Image Density
1) Purpose
An SMPTE image is used to check whether the density linearity of the image printed
out by the printer and the image displayed on the high-definition monitor matches the
density linearity of the image transferred by the CXDI.
2) Notes
2-1) This checking procedure should be performed before the procedure in the section of
“Operation Unit Gamma Correction”.
2-2) The adjustment and checking procedures below should be completed before
performing this procedure.
A) The printer and high-definition monitor connections and setting adjustments
should be completed. The printer and monitor image output settings should be set
to LINEAR.
For example, if the KODAK MLP190 is connected, the printer parameter
“-S configuration” must be set to “-S CS000”. For other printers, refer to
“Printers and Parameter Examples Reference” in “Network Connections”, and set
so that the curve shape is 0 (density linear).
B) Be sure to calibrate the printer and high-definition monitor units separately before
performing this procedure.
C) When the imaging screen is “tray type”, change it to “category type” by selecting:
System Æ SETUP MENU Æ CUSTOMIZE DISPLAY.
3) Rough adjustment
3-1) Start up the CXDI system.
3-2) Using the two knobs at the rear of the operation unit, adjust the brightness and
contrast of the touch panel screen for optimum visibility.
3-3) On the exposure screen, select the exposure mode “SMPTE” and wait until “READY”
appears. [Fig. 1]
[Fig. 1]
- 69 -
Page 87
2. Installation Manual
3-4) Press the exposure switch on the X-ray generator, and after the exposure, press the
END STUDY button. Transfer the SMPTE pattern image to the printer or the highdefinition monitor. [Fig. 2]
[Fig. 2]
3-5) Measure the densities of the 11 locations (0% to 100%) of test image grayscale on the
film or on the monitor. [Fig. 3]
* Measure the SMPTE image density on the film is measured using a densitometer.
Measure the SMPTE image density on the high-definition monitor using the
gradation analysis software.
* The data for the SMPTE test image grayscale transferred by the CXDI are the values
for the maximum density (3.20 in the case of the MLP 190) in the printer settings
which have been changed in 11 uniform steps.
[Fig. 3]
- 70 -
Page 88
2. Installation Manual
3-6) Create the graph below based on the data measured in step 5).
3.50
3.00
Max setting
density
2.50
2.00
1.50
Density
1.00
0.50
0.00
Film base
density
1234567891011
Step
As shown in the above graph, the measurement values need only to nearly form a straight line
from the minimum density to the maximum density.
The important point here is that the measurement values make a straight line and a maximum
density corresponding to the settings is output.
If the measurement values deviate too far from the straight line and a maximum density
corresponding to the settings is not output, printer and monitor output linearity settings,
calibration, and other adjustments are necessary.
* The above graph is an example of measurements when the maximum density is set to 3.2 D
and the printed film is measured with a densitometer. Refer to the data below when the
maximum density is set to 3.1 D, or when gradation analysis software is used to take
measurements on the monitor.
* The LINEAR output cannot be set for some printer models.
In this case, try to select an output setting as close as possible to LINEAR.
Step1234567891011
3.2 D00.320.640.961.281.601.922.242.562.883.20
3.1 D00.310.620.931.241.551.862.172.482.793.10
12 Bit041081912291638204824572867327636864095
8 Bit0265177102128153179204230255
Ideal density values in LINEAR LINE
Possible causes of non-linearity in measurement values:
* Inadequate calibration of printer and high-definition monitor
* Inadequate settings for CXDI printer parameters
* Faults in printer or high-definition monitor
- 71 -
Page 89
2. Installation Manual
4) Fine adjustment
4-1) As the step (3)-3) above, make the system “READY” by selecting the exposure mode
“SMPTE” on the exposure screen. [Fig. 4]
[Fig. 4]
4-2) On the X-ray generator press the exposure button. After the exposure, adjust the trim
so that the gray scale may be located in the center of the image. [Fig. 5]
* For correct density measurement of the 32-step chart, the gray scale must be printed
in the center of an image to eliminate the shading feature.
* Data may not from a straight line near the minimum and maximum densities due to
characteristics of the printer. Rotate the image on the QA screen, and reprint or retake
measurements.
[Fig. 5]
- 72 -
Page 90
2. Installation Manual
4-3) Select the END STUDY and transfer the SMPTE pattern image to the printer or the
high-definition monitor. [Fig. 6]
[Fig. 6]
4-4) On a printed film or on the monitor, measure the density of 32 steps of the grayscale
on the test image. [Fig. 7]
*1: Measure the SMPTE image density on the film using a densitometer. The SMPTE
image density on the high-definition monitor is measured using the gradation
analysis software.
*2: The data for the SMPTE test image grayscale transferred by the CXDI are the
values for the maximum density (3.20 in the case of the MLP 190) in the printer
settings which have been changed into 32 uniform steps.
Step1
Step32
[Fig. 7]
- 73 -
Page 91
2. Installation Manual
y
4-5) As the step (3)-6), create a graph based on the data measured in step 4), and make sure
that the data from the minimum density to the maximum density nearly form a straight
line.
Characteristics of the printer may prevent the data from forming a straight line near
the minimum and maximum densities. Rotate the image on the QA screen, and reprint
or retake measurements.
In the data and graph below, the data was obtained by setting the Kodak MLP 190 to curve
shape 0 (density linear), a maximum density of 3.20, the SMPTE test image was printed out,
and the image data was measured.
As shown in the graph, the data from the minimum density (film base density) to the
maximum density nearly form a straight line.
* For improper parameters, the printer parameter “-S CS000” was not entered.
* For the inadequate calibration, the printer calibration data was set too low.
4 567891011
Step
- 76 -
Page 94
2. Installation Manual
(12) Operation Unit Gamma Correction
1) Purpose
This procedure is performed so that the image that is printed out or displayed on a highdefinition monitor conforms exactly to the exposure image on the operation unit.
2) Notes
2-1) The procedure in “Linearity Check of Transfer Image Density” must be completed.
2-2) If image adjustment for the printer or high-definition monitor has not been made, use
the “Gamma Correction Calculation Tool” in order to correct the gamma of high
definition monitor image to be a same as printer image.
2-3) Gamma correction is an image correction process for monitors and film. It is different
from the contrast setting or grayscale setting.
This procedure is simply a visual adjustment. As a result, differences may occur
depending on the operator performing the procedure. Therefore, be sure to consult
with the responsible technician before performing this adjustment.
2-4) To make the gamma of operation unit adjust in detail, use the “Gamma Correction
Calculation Tool”.
3) Preparation
The Option button used in gamma correction is normally hidden. Edit the MenuPara.ini
file to display.
3-1) Open MenuPara.ini file.
3-2) Make the following changes, and overwrite the file.
Use Search in Edit to find OptionDlgBth.
Select Customize.
OptionDlgBtn = 0 Í Change to 1.
(0: Don’t display; 1: Display)
- 77 -
Page 95
2. Installation Manual
4) Comparison of operation unit image and print image or monitor image.
4-1) Start up the CXDI system.
4-2) Use the two adjustment knobs at the rear of the operation unit to adjust the brightness
and contrast of the touch panel screen for optimum visibility.
4-3) On the exposure screen, select the exposure mode “SMPTE” and wait until “READY”
appears. [Fig. 1].
[Fig. 1]
4-4) On the X-ray generator, press the exposure button, and after the exposure, select
“END STUDY”. Transfer the SMPTE pattern image to the printer or the highdefinition monitor. [Fig. 2]
[Fig. 2]
4-5) Take the SMPTE pattern image again. On the QA screen, compare the image
displayed on the operation unit screen with the film image printed in Step 4) above or
with the image on the high-definition monitor. Make sure there is no difference in
contrast and gradation between those images. Check both preview and magnify
images.
If there are any differences between these images, perform the procedure described in
“(4) Operation Unit Image Gamma Correction” on the next page.
If there are no differences between these images, the steps are complete.
- 78 -
Page 96
2. Installation Manual
5) Operation Unit Image Gamma Correction
5-1) On the QA screen displayed on Step (4) 5) above, select the “Option” tab and the
“Gamma Adjustment” button is appeared, and then press this button. [Fig. 3]
[Fig. 3]
5-2) The “Gamma Adjustment’ window appears. Change the value for the PREVIEW
IMAGE, and press OK. The gamma correction for the operation unit screen is
performed. [Fig. 4]
(Make sure the gamma value for the test image has changed on the operation unit
screen.)
The gamma value is adjustable from 1.00 to 2.50.
When the image on the operation unit is lighter (whiter) than the film image, increase
the value. On the contrary, if the image on the operation unit is darker (blacker) than
the film image, decrease the value. The default gamma value is 1.60.
[Fig. 4]
5-3) When the gamma value for the preview image is adjusted, then adjust the gamma
value for the magnify image. The default gamma value for the magnify image is 1.60.
5-4) The correction steps are complete.
- 79 -
Page 97
2. Installation Manual
(13) Changing the Total Image Count
1) Purpose
When the imaging unit is replaced (including the replacement of the LANMIT) for
servicing, the total image count displayed on the user screen can be returned to “0”if
necessary.
2) Notes
2-1) The CXDI is connected by the system.
2-2) Set the CXDI application so that it does not start up.
2-3) Files are overwritten, and so be careful when performing the procedure.
2-4) As an extra precaution, write down the numerical values before overwriting them.
2-5) The overwritten counter becomes valid the next time that the CXDI is started up.
3) Procedure
3-1) Turn on the control computer, and then start up Windows XP.
3-2) Right click on My Computer on the Desktop screen of Windows XP, and open
Explorer.
3-3) A file called “ExpResult.ini” is contained in the CCR folder.
Open this file. (D:\CCR\ExpResult.ini) [Fig 1]
[Fig 1]
- 80 -
Page 98
2. Installation Manual
3-4) Changing the items below contained in the file allows you to change the settings for
Total number of images obtained
with sensor unit 1.
Total number of images obtained
with sensor unit 2.
Total number of images obtained
with sensor unit 3.
Total number of images obtained
with sensor unit 4.
Can be overwritten at the
user screen
Can be overwritten at the
user screen
3-5) After overwriting the values, overwrite the file “ExpResult.ini” and save.
3-6) Close all windows that are open on the Desktop, and then start up the CXDI
application.
3-7) Open the “System Information” screen, and check that the changed items have been
set correctly. [Fig 2]
“TOTAL IMAGES” indicates the total number of images obtained with all sensors
automatically.
[Fig 2]
- 81 -
Page 99
2. Installation Manual
(14) Backing Up When Installing
1) Purpose
In case of re-installing the CXDI application, the necessary files ex) the exposure
position and other parameters must be backed up so that can be restored at the status of
first installation.
2) Necessary items
2-1) Removable drive such as MO drive or external HDD that can connect to USB 2.0
3) Notes
3-1) Before performing backup procedure, delete any “BodyPart” and image data exposed
for tests.
Deleting image data: Refer to the “Deleting Data” item.
Deleting “BodyPart”: Refer to the CXDI Series Operation Manual.
3-2) Backup should be performed immediately before handing over the product to the
customer only for new installations.
4) Connections
4-1) Check that all equipment is turned off.
4-2) Connect the keyboard and mouse removal drive to the control PC.
5) Settings
5-1) Turn on the operation unit power and then the control PC power.
5-2) The CXDI application starts. Press the [Alt] + [Tab] keys to switch the program to the
Command Prompt screen.
5-3) The message “Welcome to Canon CXDI” appears. Input [8] and press the [Enter] key.
(Select “8 Exit”.)
5-4) The Windows XP desktop screen appears.
5-5) Right click on My Computer on the Desktop screen, and select Explorer from Menu.
5-6) Explorer appears. Copy the CCR folder onto Removable Drive.
- 82 -
Page 100
2. Installation Manual
(15) Backing up Setting Data to FD
1) Purpose
“Important setting data (setting information which differs for each customer.)” is
backed up to floppy disks and hard disks in consideration of possible setting data loss,
hard disk corruption or other data errors.
In the event that setting data is lost or the hard disk is corrupted, this “Important setting
data” can be quickly restored to the condition before the trouble occurred by copying
from the backup data.
2) Notes
2-1) Performing this backup work means that the CXDI system will be used with the
floppy disk inserted in the floppy disk drive.
Be sure to eject the floppy disk from floppy disk drive in order to avoid damaging it,
in case of changing the layout or moving the control PC.
Likewise, when mounting the system in a vehicle, vibrations may cause damage to the
floppy disk drive.
Therefore, after backing up the latest data to the floppy disk, be sure to eject the
floppy disk.
2-2) Based on the reason in note 1) above, do not perform backups in an environment that
is exposed to vibrations. Therefore, never perform backups when the control PC is
loaded in a car.
In V4.0 and later versions, the default FD-Buck Up setting is ON. When backup to a
floppy disk is not allowed, set it to OFF.
2-3) In the product default settings, “Boot/1. Removable Device” in the system BIOS
settings is set to “Disabled”. However, as a precaution, check that “Boot/1.
Removable Device” is actually set to “Disabled”. If the setting is changed to “Legacy
Floppy”, the control PC may not start properly when a floppy disk is inserted.
For more information, see “System BIOS Settings”
2-4) Be sure to always format the floppy disk that you are using before performing
backups.
2-5) The backup procedure described here covers the case when backing up for the first
time after installation. For the second and subsequent times, data is automatically
backed up to the floppy disk whenever the user changes the exposure mode buttons or
other settings.
2-6) This backup procedure cannot be performed with just the control PC. Connect the
imaging units and other equipment, and start up in the normal exposure status.
3) Procedure
3-1) Remove the left side cover of the control PC.
3-2) Insert a formatted floppy disk (1.44 MB) in the floppy disk drive.
* Make sure the write protect of the floppy disk is unlocked at this time.
3-3) Reattach the left side cover of the control PC.
3-4) Start up the CXDI system. If you have changed the parameters including the exposure
mode button, back up all “ini.files” to a floppy disk when you restart the system again.
* When you first back up the files, the back up operation may take some time because
there are many files to be copied.
3-5) Make sure that all files have been copied and switch off the CXDI system.
The files copied to a floppy disk are the “C:\ccrbup”, and they are the latest backup
data. If the data stored in the drive D is damaged and there is no trouble in the drive C,
restore the CXDI setting data using the data in “C:\ccrbup”.
- 83 -
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.