Burkert 8202 Operating Instructions Manual

Page 1
Type 8202
ELEMENT
pH- or redox-meter pH- oder Redox-Messgerät pH- ou redox-mètre
Operating Instructions
Bedienungsanleitung Manuel d‘utilisation
Page 2
© Bürkert SAS, 2008 - 2016
Operating Instructions 1603/4_EU-ML 00560329 / Original_FR
Page 3
Typ 8202 ELEMENT
Contents
1 ABOUT THIS MANUAL .....................................................................................................................................................................7
1.1 Definition of the word "device" .......................................................................................................................................7
1.2 Validity of the manual ..........................................................................................................................................................7
1.3 Symbols used ..........................................................................................................................................................................7
2 INTENDED USE ....................................................................................................................................................................................8
3 BASIC SAFETY INFORMATION ....................................................................................................................................................9
4 GENERAL INFORMATION .............................................................................................................................................................10
4.1 Manufacturer's address and international contacts .........................................................................................10
4.2 Warranty conditions ...........................................................................................................................................................10
4.3 Information on the Internet ............................................................................................................................................10
5 DESCRIPTION ....................................................................................................................................................................................11
5.1 Area of application .............................................................................................................................................................11
5.2 General description ...........................................................................................................................................................11
5.2.1 Construction of the 8202 .................................................................................................................11
5.2.2 pH or Redox ("ORP") probe ............................................................................................................11
5.3 Description of the name plate .....................................................................................................................................12
5.4 Available versions ...............................................................................................................................................................12
6 TECHNICAL DATA .............................................................................................................................................................................13
6.1 Conditions of use ................................................................................................................................................................13
6.2 Conformity to standards and directives .................................................................................................................13
6.3 Mechanical data ...................................................................................................................................................................14
6.4 Dimensions of the 8202 ..................................................................................................................................................14
6.5 Fluid data .................................................................................................................................................................................15
6.6 Electrical data .......................................................................................................................................................................17
6.7 Data of connectors and cables....................................................................................................................................18
6.8 pH/Redox ("ORP") probe ................................................................................................................................................18
English
3
Page 4
Typ 8202 ELEMENT
7 ASSEMBLY ...........................................................................................................................................................................................19
7.1 Safety instructions .............................................................................................................................................................19
7.2 Removing the cover ...........................................................................................................................................................19
7.3 Mounting the cover ............................................................................................................................................................20
7.4 Mounting the display module .......................................................................................................................................20
7.5 Removing the display module .....................................................................................................................................21
7.6 Mounting the probe into the holder (without fluid) ..........................................................................................21
7.7 Mounting the electronic module to the sensor holder (without fluid) ...................................................22
8 INSTALLATION AND WIRING ......................................................................................................................................................23
8.1 Safety instructions .............................................................................................................................................................23
8.2 Installation onto the pipe ................................................................................................................................................24
8.3 Electrical wiring ....................................................................................................................................................................26
8.3.1 Assembling the male or female connector (accessories) ......................................................... 27
8.3.2 Making the installation equipotential ..............................................................................................27
8.3.3 Wiring a version with a single M12 fixed connector ..................................................................28
8.3.4 Wiring a version with 2 M12 fixed connectors ............................................................................ 30
9 ADJUSTMENT AND COMMISSIONING .................................................................................................................................33
9.1 Safety instructions .............................................................................................................................................................33
9.2 Knowing the operating levels .......................................................................................................................................33
9.3 Using the navigation button ..........................................................................................................................................34
9.4 Using the dynamic functions ........................................................................................................................................36
9.5 Entering a numerical value (example) .....................................................................................................................36
9.6 Browsing in a menu (example) ...................................................................................................................................36
9.7 Knowing the display ..........................................................................................................................................................37
9.7.1 Knowing the icons and LEDs ..........................................................................................................37
9.7.2 Knowing the display at the power-up of the device ...................................................................38
9.8 Knowing the Process level ............................................................................................................................................38
9.9 Accessing the Configuration level .............................................................................................................................39
9.10 Knowing the structure of the menus on the Configuration level ..............................................................40
4
English
Page 5
Typ 8202 ELEMENT
9.11 Knowing the Parameters Menu ...................................................................................................................................43
9.11.1 Transferring data from one device to another .............................................................................. 43
9.11.2 Setting the date and time .................................................................................................................44
9.11.3 Modifying the PARAM menu access code ...................................................................................44
9.11.4 Restoring the default parameters of the Process level and the outputs ............................... 45
9.11.5 Setting the data displayed in the Process level ..........................................................................45
9.11.6 Displaying the lowest and highest values measured ................................................................. 46
9.11.7 Setting the display contrast and brightness ................................................................................47
9.11.8 Choosing the output wiring mode ..................................................................................................47
9.11.9 Setting the parameters of the current outputs ............................................................................48
9.11.10 Setting the parameters of the transistor outputs ........................................................................49
9.11.11 Setting the sensor parameters ........................................................................................................51
9.12 Knowing the Calibration menu ...................................................................................................................................52
9.12.1 Activating/deactivating the Hold function ..................................................................................... 52
9.12.2 Modifying the Calibration menu access code .............................................................................52
9.12.3 Adjusting the current outputs .......................................................................................................... 53
9.12.4 Calibrating the pH sensor ................................................................................................................53
9.12.5 Calibrating the Redox- ("ORP"-) sensor ....................................................................................... 57
9.12.6 Entering an offset for the temperature measurement ................................................................60
9.13 Knowing the Diagnostic menu ....................................................................................................................................60
9.13.1 Modifying the Diagnostic menu access code ..............................................................................60
9.13.2 Monitoring the condition of the probe ...........................................................................................60
9.13.3 Monitoring the fluid temperature ..................................................................................................... 62
9.14 Knowing the Test menu .................................................................................................................................................63
9.14.1 Modifying the Test menu access code .......................................................................................... 63
9.14.2 Checking the outputs functions ......................................................................................................63
9.14.3 Checking the outputs behaviour .....................................................................................................64
9.15 Knowing the Information menu ...................................................................................................................................64
9.15.1 Reading the cause of events linked to icons ...............................................................................64
9.15.2 Reading the software versions ........................................................................................................65
9.15.3 Reading some identification informations of the device ............................................................65
English
5
Page 6
Typ 8202 ELEMENT
10 MAINTENANCE AND TROUBLESHOOTING .......................................................................................................................66
10.1 Safety instructions .............................................................................................................................................................66
10.2 Cleaning the device ...........................................................................................................................................................66
10.2.1 Cleaning the pH/Redox ("ORP") probe ........................................................................................67
10.3 Replacing the probe ..........................................................................................................................................................67
10.4 Replacing the seal of the sensor holder ................................................................................................................68
10.5 Solving a problem ..............................................................................................................................................................70
11 SPARE PARTS AND ACCESSORIES ......................................................................................................................................75
12 PACKAGING, TRANSPORT ..........................................................................................................................................................76
13 STORAGE ..............................................................................................................................................................................................76
14 DISPOSAL OF THE PRODUCT ..................................................................................................................................................76
6
English
Page 7
Typ 8202 ELEMENT
About this manual

1 ABOUT THIS MANUAL

This manual describes the entire lifecycle of the device. Please keep this manual in a safe place, accessible to all users and any new owners.
This manual contains important safety information.
Failure to comply with these instructions can lead to hazardous situations. Pay attention in particular to the chapters "Basic safety information" and "Intended use".
▶ Whatever the version of the device, this manual must be read and understood.

1.1 Definition of the word "device"

The word "device" used within this manual refers to the pH or "ORP" (Redox) transmitter type 8202 ELEMENT.

1.2 Validity of the manual

The manual is valid for the devices from the version V2.
These informations are available on the name plate, see chap. 5.3.

1.3 Symbols used

DANGER
Warns against an imminent danger.
▶ Failure to observe this warning can result in death or in serious injury.
WARNING
Warns against a potentially dangerous situation.
▶ Failure to observe this warning can result in serious injury or even death.
ATTENTION
Warns against a possible risk.
▶ Failure to observe this warning can result in substantial or minor injuries.
7
English
Page 8
Typ 8202 ELEMENT
About this manual
NOTE
Warns against material damage.
▶ Failure to observe this warning may result in damage to the device or system.
Indicates additional information, advice or important recommendations.
Refers to information contained in this manual or in other documents.
▶ Indicates an instruction to be carried out to avoid a danger, a warning or a possible risk.
→ Indicates a procedure to be carried out.
Indicates the result of a specific instruction.

2 INTENDED USE

Use of this device that does not comply with the instructions could present risks to people, nearby installations and the environment.
The device is intended solely for the measurement of:
- the pH in clean liquids or liquids containing solids, sulphides or proteins.
- or the oxidation reduction potential in clean liquids or liquids containing solids, sulphides or proteins which may present low conductivity.
▶ Use this device in compliance with the characteristics and commissioning and use conditions specified in the
contractual documents and in the user manual.
▶ Never use this device for security applications.
▶ Protect this device against electromagnetic interference, ultraviolet rays and, when installed outdoors, the
effects of climatic conditions.
▶ Use this device only if in perfect working order.
▶ Requirements for the safe and proper operation of the device are proper transport, storage and installation, as
well as careful operation and maintenance.
▶ Only use the device as intended.
8
English
Page 9
Typ 8202 ELEMENT
Basic safety information

3 BASIC SAFETY INFORMATION

This safety information does not take into account:
• any contingencies or occurrences that may arise during assembly, use and maintenance of the device.
• the local safety regulations that the operator must ensure the staff in charge of installation and maintenance observe.
Danger due to electrical voltage.
▶ Disconnect the electrical power for all the conductors and isolate it before carrying out work on the system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
Danger due to high pressure in the installation.
▶ Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Danger due to high temperatures of the fluid.
▶ Use safety gloves to handle the device.
▶ Stop the circulation of fluid and drain the pipe before loosening the process connections.
Danger due to the nature of the fluid.
▶ Respect the prevailing regulations on accident prevention and safety relating to the use of aggressive fluids.
Various dangerous situations.
To avoid injury take care:
▶ not to use the device in explosive atmospheres.
▶ not to use the device in an environment incompatible with the materials it is made of.
▶ not to use fluid that is incompatible with the materials the device is made of.
▶ not to subject the device to mechanical loads.
▶ not to make any modifications to the device. ▶ to prevent any unintentional power supply switch-on.
▶ to carry out the installation and maintenance work by qualified and skilled staff with the appropriate tools.
▶ to guarantee a defined or controlled restarting of the process, after a power supply interruption. ▶ to observe the general technical rules when installing and using the device.
English
9
Page 10
Typ 8202 ELEMENT
Basic safety information
NOTE
The device may be damaged by the fluid in contact with.
▶ Systematically check the chemical compatibility of the component materials of the device and the fluids likely
to come into contact with it (for example: alcohols, strong or concentrated acids, aldehydes, alkaline com­pounds, esters, aliphatic compounds, ketones, halogenated aromatics or hydrocarbons, oxidants and chlorin­ated agents).
NOTE
Elements / Components sensitive to electrostatic discharges
▶ This device contains electronic components sensitive to electrostatic discharges. They may be damaged if
they are touched by an electrostatically charged person or object. In the worst case scenario, these compo­nents are instantly destroyed or go out of order as soon as they are activated.
▶ To minimise or even avoid all damage due to an electrostatic discharge, take all the precautions described in
the EN 61340-5-1 norm.
▶ Also ensure that you do not touch any of the live electrical components.

4 GENERAL INFORMATION

4.1 Manufacturer's address and international contacts

To contact the manufacturer of the device, use following address:
Bürkert SAS
Rue du Giessen
BP 21
F-67220 TRIEMBACH-AU-VAL
You may also contact your local Bürkert sales office.
The addresses of our international sales offices are available on the internet at: www.burkert.com

4.2 Warranty conditions

The condition governing the legal warranty is the conforming use of the device in observance of the operating conditions specified in this manual.

4.3 Information on the Internet

10
You can find the user manuals and technical data sheets regarding the transmitter type 8202 at:
www.burkert.com
English
Page 11
Typ 8202 ELEMENT
Description

5 DESCRIPTION

5.1 Area of application

The device is intended solely for the measurement of:
• the pH in clean liquids or liquids containing solids, sulphides or proteins.
• or the oxidation reduction potential ("ORP") in clean liquids or liquids containing solids, sulphides or proteins which may present low conductivity.
Thanks to two fully adjustable transistor outputs, the transmitter can be used to switch a solenoid valve, activate an alarm and, thanks to one or two 4-20-mA current outputs, establish one or two control loops.

5.2 General description

5.2.1 Construction of the 8202

The device comprises:
• a module for measuring process values, comprising:
- a pH or Redox ("ORP") sensor measuring a potential difference (PD) in mV
- a Pt1000 temperature sensor built in the holder of the pH or Redox ("ORP") sensor, measuring a resistance.
• an acquisition / conversion module for the process values measured:
- measured PD acquisition in mV
- conversion of the measured PD into pH units (for a transmitter with pH sensor only)
- acquisition of the resistance measured and conversion into temperature
• an electrical housing which can include a display module. The display module has a navigation button used to read and/or configure the parameters of the device. The display module is available as an accessory (see chap. 11).
One version of the device with two transistor outputs and a 4-20 mA output operates on a 2-wire system and requires a power supply of 14-36 V DC. For such a version, electrical connection is done via an M12, 5-point, male fixed connector.
One version of the device with two transistor outputs and two 4-20 mA outputs operates on a 3-wire system and requires a power supply of 12-36 V DC. For such a version, electrical connection is done via an M12, 5-point, male fixed connector and an M12, 5-point, female fixed connector.

5.2.2 pH or Redox ("ORP") probe

The device can be fitted with a standard probe 120 mm long, measuring the pH or the oxidation reduction potential ("ORP").
It is screwed into a holder with the built-in Pt1000 temperature probe.
• The pH probe is a glass membrane with variable sensitivity according to the pH. When the pH probe is immersed in a solution, a difference in potential is formed, due to the hydrogen ions (H+), between the glass membrane and the solution. This difference in potential, measured in relation to a reference electrode, is directly proportional to the pH value (59.16 mV per pH unit at 25°C).
English
11
Page 12
8202 pH/ORP Transmitter V2
1
8
10
11
Typ 8202 ELEMENT
Description
• When a Redox ("ORP") probe is immersed in a solution, an exchange of electrons occurs based on the oxi­dizing and reducing effects of an electrolyte. The resulting voltage is the oxidation reduction potential ("ORP").

5.3 Description of the name plate

Supply: 14-36V 40W Max Output: 1x 4-20mA 2x Trans 1A Max Cell: pH -2/16 ORP +-2V 120mm PG13.5 Process: Temp 0/50°C limited by cell PN 16 Bar limited by cell
Made in France
IP65 -IP67 W41MT
9
Fig. 1 Name plate on the transmitter type 8202
S-N:3054
00559630
2:NPN/PNP1
3:0V
4:NPN/PNP2
1:V+
1. Type of device, parameter measured and
2 3 4 5 6
version
2. Electrical power supply and power consumption
3. Output specifications
7
4. Sensor specifications
5. Temperature ra n ge of t h e f luid, without se n sor
6. Nominal pressure of the fluid, without sensor
7. Allocation of the pins on the M12 fixed connectors
8. Construction code
9. Order code
10. Serial number
11. Protection rating
12

5.4 Available versions

The following versions of the device are available. These references include the electronic module and the sensor holder including the Pt1000.
The pH/Redox ("ORP") sensor, the display module and the fitting for connection to the process should be ordered separately. For the pH/Redox ("ORP") sensor and the display module, see the list of accessories in Chapter 11.
Voltage supply Outputs
14-36 V DC 2 transistors +
1 x 4-20 mA
12-36 V DC 2 transistors +
2 x 4-20 mA
Electrical connection
Male 5-pin M12 fixed connector
Male 5-pin M12 fixed connector + female 5-pin M12 fixed connector
Sensor
without PVC no
without PVC no
Nut material
UL Order code
yes
PVDF no
yes
yes
PVDF no
yes
559 630
559 634
559 632
559 636
559 631
559 635
559 633
559 637
English
Page 13
Typ 8202 ELEMENT
Technical data

6 TECHNICAL DATA

6.1 Conditions of use

Ambient temperature -10 to +60 °C ( without pH or Redox ("ORP") probe) Air humidity < 85%, non condensated Protection rating according to
EN 60529

6.2 Conformity to standards and directives

The device conforms to the CE directives through the following standards:
• EMC: EN 61000-6-2, EN 61000-6-3
IP65 and IP67 with connectors plugged in and tightened and electronic module cover fully sealed
• Vibration: EN 60068-2-6
• Shock: EN 60068-2-27
• Pressure: complying with article 3§3 of the pressure equipment directive 97/23/CE. According to this directive, the device can only be used in the following cases (depending on max. pressure, pipe diameter and fluid):
Type of fluid Conditions
Fluid group 1, §1.3.a only DN25 Fluid group 2, § 1.3.a DN32
or DN > 32 and PNxDN1000
Fluid group 1, § 1.3.b DN25
or DN > 25 and PNxDN2000
Fluid group 2, § 1.3.b DN125
The UL devices conform to the following standards:
• UL 61010-1
• CAN/CSA-C22.2 n° 61010-1
English
13
Page 14

6.3 Mechanical data

Typ 8202 ELEMENT
Technical data
Nickel-plated brass
(or stainless steel)
Stainless steel
EPDM
PVDF
PC
silicone
PPS
EPDM
PPS
pH/Redox ("ORP") probe
PVDF
PVC or PVDF
Stainless steel
1.4571 (316Ti)
Part Material
Box / seals stainless steel, PPS /
EPDM
Cover / seal PC / silicone
Display module PC / PBT
M12 fixed connector nickel-plated brass
(stainless steel on request)
Fixed connector holder
Screws stainless steel
Nut PVC or PVDF
Sensor holder / seal (in contact with the fluid)
pH or Redox ("ORP") probe
stainless steel 1.4404 (316L)
PVDF, stainless steel
1.4571 (316Ti) / EPDM
refer to the related manual
14
Fig. 2 Materials used in the transmitter type 8202 (without the probe)

6.4 Dimensions of the 8202

→ Please refer to the technical data sheets regarding the transmitter type 8202, available at: www.burkert.com
English
Page 15
Typ 8202 ELEMENT
Technical data

6.5 Fluid data

Pipe diameter
Type of fitting
Nut between the 8202 and the fitting
Max. fluid temperature
• with a PVDF nut (see also Fig. 3 and Fig. 5) • -20 to +130 °C
• with a PVC nut (see also Fig. 3 and Fig. 4) • 0 to +50 °C
Max. fluid pressure
pH measurement
• Measurement range • -2 to 16 pH or -580 to +580 mV
• Resolution • 0.001 pH or 0.1 mV
• Measurement deviation • ±0.02 pH or 0.5 mV
• Recommended min. divergence of the pH range associated to the 4-20 mA signal
DN25 to DN110 (DN15 to DN20 under specific conditions)
Adapter S022
G 1 1/2'' internal thread
The fluid temperature may be restricted by the probe used (see the related instruction manual), by the pressure of the fluid and the material of the adapter S022
PN16
The fluid pressure may be restricted by the probe used (see the related instruction manual), by the temperature of the fluid and the material of the adapter S022 (see Fig. 3, Fig. 4 and Fig. 5)
• 0.5 pH unit or 30 mV (eg: range 6,7 to 7,2 pH or -20 mV to +10 mV associated to the 4-20 mA output current)
Redox potential ("ORP") measurement
• Measurement range • -2000 mV to +2000 mV
• Resolution • 1 mV
• Measurement deviation • ±3 mV
• Recommended min. divergence of the redox potential ("ORP") range associated to the 4-20 mA signa
Temperature measurement
• Measurement range • -40 °C to +130 °C
• Resolution • 0.1 °C
• Measurement deviation • ±1 °C
• Recommended min. divergence of the temperature range associated to the 4-20 mA signal
Temperature probe
Temperature compensation
• 50 mV (eg: range 1550 to 1600 mV associated to the 4-20 mA output current)
• 10 °C (eg: range 10 to 20 °C associated to the 4-20 mA output current)
Pt1000 integrated in the sensor holder
Automatic (integrated Pt1000), Reference temperature = 25°C
English
15
Page 16
Typ 8202 ELEMENT
Technical data
• A: application range of a 8202 with a PVDF nut
• B: application range of a 8202 with a PVC nut
The measures have been made at an ambient temperature of 60 °C
P = Fluid pressure
T = Fluid temperature
P (bar)
16 15 14 13 12 11 10
B
PVDF
PVC
9 8 7 6 5 4 3 2 1 0
-20 0 +20 +40 +60 +80 +100 +120
A
PVDF
T ( ° C)
P (psi) 232
217.6 203
188.6 174
159.6 145
130.6 116
101.6 87
72.5 58
43.5 29 14 0
+140
Fig. 3 Fluid temperature / pressure dependency of the 8202 (without the probe) with a PVC or PVDF nut
16
P (bar)
15 14 13 12 11 10
9 8 7 6 5 4 3 2 1 0
PVC + PP
-20
Metal
0 +20 +40 +60
232
217.6 203
188.6 174
159.6 145
130.6 116
101.6 87
72.5 58
43.5 29 14 0
T (°C)
P (psi)
P = Fluid pressure
T = Fluid temperature
16
Fig. 4 Fluid temperature / pressure dependency of the 8202 (without the probe) with a PVC nut and a metal, PVC or PP
S022 adapter
P (bar)
16 15 14 13 12 11 10
9 8
PVC + PP 7 6 5 4 3 2 1 0
-20 0 +20 +40 +60 +80 +100 +140
PVC
PP
Metal
+120
232
217.6 203
188.6 174
159.6 145
130.6 116
101.6 87
72.5 58
43.5 29 14 0
P (psi)
P = Fluid pressure
T = Fluid temperature
T (°C)
Fig. 5 Fluid temperature / pressure dependency of the 8202 (without the probe) with a PVDF nut and a metal, PVC or
PP S022 adapter
English
Page 17
Typ 8202 ELEMENT
Technical data

6.6 Electrical data

Power supply
• Version with 3 outputs • 14-36 V DC, filtered and regulated
• Version with 4 outputs • 12-36 V DC, filtered and regulated
Characteristics of the power source (not supplied) of the UL versions
Current consumption
• Version with 3 outputs • 25 mA max. (at 14 V DC)
• Version with 4 outputs • 5 mA max. (at 12 V DC)
Current consumption, with loads on the transistors
Power consumption
Protection against polarity reversal
Protection against voltage spikes
Protection against short circuits
Transistor output
• limited energy source (in accordance to UL 61010-1, paragraph 9.3)
• or Class 2 source (in accordance to standards 1310/1585 and 60950-1)
1 A max.
40 W max.
yes
yes
yes, transistor outputs
NPN (/sink) or PNP (/source) (depending on software setting), open collector, 700 mA max., 0.5 A max. per transistor if both transistor outputs are wired. NPN output: 0.2-36 V DC PNP output: supply voltage
Current output
• Response time (10 % - 90 %) • 150 ms (default value)
• Version with 1 current output • max. loop impedance: 1100 W at 36 V DC, 610 W at
• Version with 2 current outputs • max. loop impedance: 1100 W at 36 V DC, 610 W at
4-20 mA, sink ("NPN sink") or source ("PNP source") (depending on software setting)
24 V DC, 180 W at 14 V DC
24 V DC, 100 W at 12 V DC
17
English
Page 18
Typ 8202 ELEMENT
Technical data

6.7 Data of connectors and cables

Number of fixed connectors Type of connectors
1 male M12 fixed connector 5-pin female M12 connector (not supplied).
For the M12 connector with order code 917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm
1 male M12 fixed connector and 1 female M12 fixed connector
5-pin female M12 connector (not supplied) and 5-pin male M12 connector (not supplied).
For the M12 connector with order code 917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm

6.8 pH/Redox ("ORP") probe

The pH or redox ("ORP") probe must satisfies the following specifications:
• combined probe;
• length: 120 mm;
• with PG 13.5 head;
• with an S7/S8 fixed connector;
• without temperature probe.
2
2
18
The specifications of the probe can be found in the manual of the probe used.
English
Page 19
1
Typ 8202 ELEMENT
Assembly

7 ASSEMBLY

7.1 Safety instructions

DANGER
Risk of injury due to electrical voltage.
▶ Shut down and isolate the electrical power source before carrying out work on the system.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
WARNING
Risk of injury due to non-conforming assembly.
▶ The device must only be assembled by qualified and skilled staff with the appropriate tools.
Risk of injury due to unintentional switch on of power supply or uncontrolled restarting of the installation.
▶ Take appropriate measures to avoid unintentional activation of the installation.
▶ Guarantee a defined or controlled restarting of the process after any intervention on the device.

7.2 Removing the cover

NOTE
The tightness of the device is not guaranteed when the cover is removed.
▶ Once the cover is removed, prevent ingress of liquid inside the housing.
The device may be damaged if a metal component comes into contact with the electronics.
▶ Prevent contact of the electronics with a metal component.
2
→ [1] Turn the cover counterclockwise with
an angle of about 15° to unlock it.
→ [2] Remove the cover.
If the cover grips to the housing:
→ Use an appropriate tool to unlock the
cover, taking care not to scratch the glass.
→ Insert an apropriate tool into the groove of
the housing.
→ Lever the cover up.
Fig. 6 Removing the cover
English
19
Page 20

7.3 Mounting the cover

1
b)
d)
→ Check that there is a seal on the housing and that it is not damaged. Replace
it if necessary.
2
→ Grease the seal if necessary, using a component compatible with the seal
material.
→ [1] Set the cover to ensure that the 4 grooves of the cover match with the 4
pins of the housing.
→ [2] Turn the cover clockwise with an angle of about 15° to lock it.
Fig. 7 Closing the cover

7.4 Mounting the display module

Typ 8202 ELEMENT
Assembly
20°
a)
Fig. 8 Mounting the display module
→ Remove the cover (see chap. 7.2).
→ Set the display module at an angle of about 20° in relation to
the desired position.
→ The module can be mounted in 4 different positions, at 90°
intervals.
c)
→ Fully push in the module and turn clockwise to lock it.
20
English
Page 21
Typ 8202 ELEMENT
Assembly

7.5 Removing the display module

→ Remove the cover if necessary (see chap. 7.2).
→ Turn the module by ca. 20° counter clockwise.
Once unlocked, the module is raised slightly by the spring action.
20°
→ Remove the module from its housing.
Fig. 9 Removing the display module

7.6 Mounting the probe into the holder (without fluid)

probe head
H
compression washer
seal
A
Following instructions are valid for a Bürkert probe.
If you use a probe from another supplier, respect the related instructions.
→ Remove the protective plugs
→ Check that dimension H on the probe is between 34 and 46 mm. If nec-
essary, adjust the height of the compression washer.
→ At first use, apply water or soapy water on the "A" seal.
→ Insert the seal into the groove on the holder.
→ Insert the probe with its seal into the holder from above.
→ Tighten the probe head using a suitable wrench.
Fig. 10 Mounting the probe into the holder (without fluid)
21
English
Page 22
Typ 8202 ELEMENT
Assembly

7.7 Mounting the electronic module to the sensor holder (without fluid)

→ Check that the probe is mounted into the sensor holder (see chap. 7.6).
→ Check that seal "A" on the holder is in good condition. Replace it if
necessary (see chap. 11)
→ Clean connectors "B" and "C" for connection of the pH/redox ("ORP")
probe with alcohol to avoid measurement errors.
→ Insert the electronic module into the holder, making sure the polarising
slots are correctly positioned.
→ Apply slight vertical pressure to engage the seal.
C
B
A
→ Fasten the electronic module and the holder together by tightening the
G2'' nut.
→ Tighten the G2'' nut by hand only, until it stops turning, to ensure good
electrical contact.
Fig. 11 Mounting the electronic module to the holder (without fluid)
→ Mount the display module (see chap. 7.4) to calibrate the transmitter.
→ Calibrate the transmitter (see chap. 9.12.4).
22
English
Page 23
Typ 8202 ELEMENT
Installation and wiring

8 INSTALLATION AND WIRING

8.1 Safety instructions

DANGER
Risk of injury due to electrical voltage.
▶ Disconnect the electrical power for all the conductors and isolate it before carrying out work on the system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high pressure in the installation.
▶ Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Risk of injury due to high fluid temperatures.
▶ Use safety gloves to handle the device.
▶ Stop the circulation of fluid and drain the pipes before loosening the process connections.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
WARNING
Risk of injury due to non-conforming installation.
▶ The electrical and fluid installation can only be carried out by qualified and skilled staff with the appropriate
tools.
▶ Install appropriate safety devices (correctly rated fuse and/or circuit-breaker).
▶ Respect the assembly instructions for the fitting used.
Risk of injury due to unintentional switch on of power supply or uncontrolled restarting of the installation.
▶ Take appropriate measures to avoid unintentional activation of the installation.
▶ Guarantee a defined or controlled restarting of the process after any intervention on the device.
English
23
Page 24
Typ 8202 ELEMENT
Installation and wiring
WARNING
Risk of injury due to non-conforming installation.
Non-conforming commissioning could lead to injuries and damage the device and its surroundings.
▶ Before commissioning, make sure that the staff in charge have read and fully understood the contents of the
manual.
▶ In particular, observe the safety recommendations and intended use. ▶ The device/installation must only be commissioned by suitably trained staff.
Protect this device against electromagnetic interference, ultraviolet rays and, when installed outdoors, the effects of the climatic conditions.

8.2 Installation onto the pipe

DANGER
Risk of injury due to high pressure in the installation.
▶ Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
If a pH/redox ("ORP") probe (with PG 13.5 head, 120 mm long and without temperature probe) from a supplier other than Bürkert is used, follow the relevant instructions on installation in the pipe.
→ Choose an appropriate position in the pipe to install
the fitting.
→ Fit the pipe with a fitting with G 1” ½ external
threaded sensor connection with respect to the instructions delivered with the fitting.
Direction of the fluid
24
Fig. 12 Mounting positions in the pipe
75°
75°
Fig. 13 Angle to the vertical
English
→ Fit the fitting at an angle of ±75° max. to the vertical
in order to ensure the good operation of the pH/redox ("ORP") probe.
Page 25
Typ 8202 ELEMENT
Installation and wiring
The probe must always be immersed in the fluid to prevent it drying out.
→ Once the transmitter has been calibrated, remove the electronic module from the sensor holder as shown
Fig. 14.
→ Unscrew the nut between the electronic module and the
sensor holder.
→ Remove the electronic module by pulling it straight out. There
may be a resistance due to the seal.
Fig. 14 Removing the electronic module from the sensor holder
→ Install the holder with its probe on the fitting as shown in Fig. 15.
→ Check the presence and the condition of seal B on the fitting.
Replace the seal if necessary.
→ Insert the holder with its probe carefully into the fitting.
B
→ Tighten the nut on the fitting by hand.
→ Charge the pipe to check the tightness of the assembly.
Fig. 15 Installing the sensor holder with its probe on a fitting
25
English
Page 26
Typ 8202 ELEMENT
Installation and wiring
→ If the sensor holder is tight, insert the electronic module back onto the sensor holder as shown in Fig. 16.
→ Check that the electrical contacts are in good condition and
clean them with a brush if necessary.
electrical contacts
→ Check that seal "A" on the holder is in good condition.
Replace it if necessary (see chap. 11 and chap. 10.4).
→ Insert the electronic module into the holder, making sure the
polarising slots are correctly positioned.
polarising slots
A
→ Apply slight vertical pressure to engage the seal.
→ Fasten the electronic module and the holder together by
tightening the G2'' nut.
→ Tighten the G2'' nut, until it stops turning, to ensure good
electrical contact with the temperature probe.
Fig. 16 Mounting the electronic module to the sensor holder, after installation of the holder on a fitting

8.3 Electrical wiring

DANGER
Risk of injury due to electrical voltage.
▶ Shut down the electrical power source of all the conductors and isolate it before carrying out work on the
system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
26
• Use a high-quality electrical power supply (filtered and regulated).
• Make sure the installation is equipotential (see chap. 8.3.2).
English
Page 27
5,5
Typ 8202 ELEMENT
Installation and wiring

8.3.1 Assembling the male or female connector (accessories)

4 3 2 1
→ Unscrew the nut [1] on the body [4].
→ Insert the cable into the nut [1], the cable clamp [2] and the seal [3], and then into
the body [4].
5
→ Strip 20 mm of the cable.
→ Cut the central wire (earth) so that its length is equal to 11.5 mm.
11,5
5,5
20
→ Expose 5.5 mm of the wires on the stripped cable.
→ Insert each wire into the appropriate pin on the terminal block [5] (see chap. 8.3.3
and 8.3.4).
→ Tighten the terminal block [5] wired to the body [4].
→ Tighten the connector nut [1].
Fig. 17 M12 multi-pin connector (not provided)

8.3.2 Making the installation equipotential

To ensure the equipotentiality of the installation (power supply - device - medium):
→ Connect together the various earth spots in the installation to eliminate the potential differences that may
occur between different earthes.
→ Observe faultless grounding of the shield of the power supply cable (see Fig. 18 and Fig. 19).
→ Special attention has to be paid if the device is installed on plastic pipes because there is no direct earthing
possible. Proper earthing is performed by earthing together the metallic devices such as pumps or valves, that are as close as possible to the device (see Fig. 19).
Power supply
Fig. 18 Equipotentiality skeleton diagram with pipes in metal
­+
12-36 V DC
27
English
Page 28
Power supply
Typ 8202 ELEMENT
Installation and wiring
Devices such as valves, pumps,...
­+
12-36 V DC
Pipes in plastic
Fig. 19 Equipotentiality skeleton diagram with pipes in plastic

8.3.3 Wiring a version with a single M12 fixed connector

Transistor output 1
2
0V
3
4
Transistor output 2
Fig. 20 Pin assignment of the fixed connector on a version with a single M12 fixed connector
Pin for the female M12 connector available as an accessory (order code 438680) Colour of the wire
1 brown 2 white 3 4 black 5
V+ (14-36 V DC)
1
white
blue
grey
Load 1 (solenoid valve for instance)
28
2
4
blue
1
brown
grey
+ -
Power supply
(solenoid valve for instance)
Load 2
black
3
14-36 V DC
Fig. 21 NPN wiring of both transistor outputs (software setting "NPN/sink", see chap. 9.11.8), of a version with 1 M12
fixed connector
English
Page 29
Typ 8202 ELEMENT
Installation and wiring
Load 1 (solenoid valve for instance)
(solenoid valve for instance)
Load 2
black
white
2
brown
1
3
4
blue
grey
+ -
Power supply
14-36 V DC
Fig. 22 PNP wiring of both transistor outputs (software setting "PNP/source", see chap. 9.11.8), of a version with 1 M12
fixed connector
4-20mA input at external device
+ -
2
blue
3
Power supply
brown
1
grey
4
+ -
14-36 V DC
3
blue
4-20mA input at external device
+ -
2
brown
1
4
grey
+ -
Power supply
14-36 V DC
Fig. 23 Possible wirings of the current output (whatever the software setting, "NPN/sink" or "PNP/source", see
chap. 9.11.8), of a version with 1 M12 fixed connector
Load 1
Load 2
black
white
3
blue
2
brown
1
4
grey
4-20mA input at external device
+ -
+ -
Power supply
14-36 V DC
Fig. 24 NPN wiring of both transistor outputs and wiring the current output in sinking mode (software setting "NPN/sink",
see chap. 9.11.8), of a version with 1 M12 fixed connector
Load 1
Load 2
black
white
blue
2
brown
3
1
4
grey
4-20mA input at external device
+ -
+ -
Power supply
14-36 V DC
Fig. 25 PNP wiring of both transistor outputs and wiring the current output in sourcing mode (software setting " PNP/
source", see chap. 9.11.8), of a version with 1 M12 fixed connector
English
29
Page 30
Typ 8202 ELEMENT
Installation and wiring

8.3.4 Wiring a version with 2 M12 fixed connectors

Transistor output 1
3
0V
Current output 1
Male fixed connector
2
1
4
V+
(12-36 V DC)
V+
(12-36 V DC)
Female fixed connector
Transistor output 2
2
1
3
4
Current output 2
0V
Fig. 26 Pin assignment of the male and female M12 fixed connectors
Connect the power supply for the transmitter to the male fixed connector; the supply is then transferred internally to pins 1 and 3 of the female fixed connector in order to ease wiring of the load to the female fixed connector.
Pin of the M12 female cable available as an accessory (order code 438680) Colour of the wire
1 brown 2 white 3 blue 4 black 5
grey
Pin of the M12 male cable available as an accessory (order code 559177) Colour of the wire
1 brown 2 white 3 blue 4 black 5
grey
30
Load 1 (solenoid
valve for instance)
white
3
2
blue
white
brown
1
grey
4
+ -
2
1
3
4
Power supply
brown
Load 2 (solenoid valve for instance)
12-36 V DC
Fig. 27 NPN wiring of both transistor outputs of a version with 2 fixed connectors (software setting "NPN/sink", see
chap. 9.11.8)
English
Page 31
Typ 8202 ELEMENT
Installation and wiring
Load 1 (solenoid valve for
instance)
white
3
2
blue
white
brown
1
grey
4
+ -
2
1
3
4
Power supply
blue
Load 2 (solenoid valve for instance)
12-36 V DC
Fig. 28 PNP wiring of both transistor outputs of a version with 2 fixed connectors (software setting "PNP/source", see
chap. 9.11.8)
st
4-20mA input at external device
1
nd
4-20mA input at external
2 device
+ - + -
black
2
brown
3
1
grey
4
blue
+ -
brown
black
Power supply
2
3
1
4
12-36 V DC
Fig. 29 Wiring of both current outputs in sinking mode, on a version with 2 fixed connectors (software setting "NPN/sink",
see chap. 9.11.8)
st
4-20mA input at exter-
1
nal device
nd
2
4-20mA input at
external device
+ -+ -
black
2
brown
3
1
grey
4
blue
+ -
2
3
1
4
black
Power supply
blue
12-36 V DC
Fig. 30 Wiring of both current outputs in sourcing mode, on a version with 2 fixed connectors (software setting "PNP/
source", see chap. 9.11.8)
English
31
Page 32
Typ 8202 ELEMENT
Installation and wiring
st
1
4-20mA input at external
device
Load 1
white
2
blue
3
black
+ - + -
brown
1
4
grey
brown
1
+ -
12-36 V DC
Power supply
Load 2
white
2
3
4
black
nd
4-20mA input at external
2 device
Fig. 31 NPN wiring of both transistor outputs and wiring of both current outputs in sinking mode, on a version with 2 fixed
connectors (software setting "NPN/sink", see chap. 9.11.8)
st
1
4-20mA input at external
device
Load 1
white
2
blue
black
+ - + -
brown
3
1
4
grey
1
+ -
12-36 V DC
4
2
black
Load 2
white
blue
3
nd
2
4-20mA input at external
device
32
Power supply
Fig. 32 PNP wiring of both transistor outputs and wiring of both current outputs in sourcing mode, on a version with 2
fixed connectors (software setting "PNP/source", see chap. 9.11.8)
English
Page 33
Typ 8202 ELEMENT
Adjustment and commissioning

9 ADJUSTMENT AND COMMISSIONING

• The settings can only be done on a device with a display module.
• Do not remove the display module while making the settings on the device.

9.1 Safety instructions

WARNING
Risk of injury due to non-conforming adjustment.
Non-conforming adjustment could lead to injuries and damage the device and its surroundings.
▶ The operators in charge of adjustment must have read and understood the contents of this manual.
▶ In particular, observe the safety recommendations and intended use. ▶ The device/installation must only be adjusted by suitably trained staff.
WARNING
Risk of injury due to non-conforming commissioning.
Non-conforming commissioning could lead to injuries and damage the device and its surroundings.
▶ Before commissioning, calibrate the pH/Redox ("ORP") sensor (see chap. 9.12.4).
▶ Make sure that the staff in charge have read and fully understood the contents of the manual.
▶ In particular, observe the safety recommendations and intended use. ▶ The device / the installation must only be commissioned by suitably trained staff.
Protect this device against electromagnetic interference, ultraviolet rays and, when installed outdoors, the effects of the climatic conditions.

9.2 Knowing the operating levels

The device has 2 operating levels:
Process level
This level is used:
• to read the measured values of 2 process values selected in the Parameters menu,
• to read both the lowest and highest values of the chosen measurable variable, that have been measured by the device since the power-up of the device or since the latest reset (this feature is not active by default),
• to reset both the lowest and highest values of the chosen process value, if the feature has been activated,
• to read the current values emitted on the 4-20 mA outputs.
• to get the state of the device and of the pH/Redox ("ORP") sensor with the icons.
English
33
Page 34
Configuration level
This level comprises 5 menus:
Menu title Relevant icon "Param": see chap. 9.11
This is when the device is be­ing parame-
tered............
....................
"Calib": see chap. 9.12
"Diagnostic": see chap. 9.13
"Test": see chap. 9.14
"Info": see chap. 9.15
Typ 8202 ELEMENT
Adjustment and commissioning

9.3 Using the navigation button

Symbolised by in
this manual
Symbolised by in
this manual
Fig. 33 Using the navigation button
Symbolised by
this manual
Symbolised by in
in
Symbolised by in
this manual
this manual
34
You want to... Press...
...browse in Process level
• next screen:
• previous screen:
English
Page 35
Typ 8202 ELEMENT
Adjustment and commissioning
You want to... Press...
• ...access the Configuration level
• ...display the Param menu
...browse in the menus of the Configuration level
...access the menu displayed
...browse in the menu functions
...select the highlighted function
...browse in the dynamic functions bar (MEAS, BACK, ABORT, OK, YES, NO)
...confirm the highlighted dynamic function
for at least 2 sec., from any screen of the Read
level
• next menu:
• previous menu:
• next function:
• previous function:
• next function:
• previous function:
...modify a numerical value
- increment the figure selected
- decrement the figure selected
- select the previous figure
- select the next figure
- allocate the "+" or "-" sign to the numerical value
- move the decimal point
-
-
-
-
­then
­then
place
to the extreme left of the numerical value
until the desired sign is displayed
to the extreme right of the numerical value
until the decimal point is in the desired
English
35
Page 36
Typ 8202 ELEMENT
Adjustment and commissioning

9.4 Using the dynamic functions

You want to... Choose...
...go back to the Process level, without validating the modifica­tions made
...validate the input dynamic function "OK" ...go back to the parent menu dynamic function "BACK" ... abort the current operation and go back to the parent menu dynamic function "ABORT" ...answer the question asked dynamic function "YES" or "NO"
dynamic function "MEAS"

9.5 Entering a numerical value (example)

Modify each digit of the numerical value using:
to increase the digit
­selected,
Select the digit at the extreme left of the numerical value
with allocate the "+" or "-" sign to the numerical value with
then
Calib.Temp
+0.000°C
MEAS ABORT OK
.
Dynamic functions (accessible through and ): see chap. 9.4

9.6 Browsing in a menu (example)

Title of the current menu, sub-menu or
The icon identifies the
current menu
This is when the device is be­ing parame-
tered............
....................
Highlighted
function
Param
Line1
Line2
Contrast MEAS ABORT OK
function.
The arrow indicates that some more functions are available which can be
displayed by using
The arrow indicates that some more functions are available which can be
displayed by using
to decrease the digit
­selected.
Select the digit at the extreme right of the numerical value with
then move the decimal point
with
.
36
Dynamic functions (accessible through and ): see chap. 9.4
English
Page 37
HOLD
T
Typ 8202 ELEMENT
Adjustment and commissioning

9.7 Knowing the display

The display module is only equipped on some versions of the device. It can be ordered as an accessory.

9.7.1 Knowing the icons and LEDs

Red LED:
shows an error;
see chap. 10.5
Yellow LED: shows that transis-
tor 1 is switched
ERR
N
E
P
O
mV_pH
TempC
L
O
C
K
1423 mV
23.8 °C
not used
Yellow LED: shows that transistor 2 is switched
Yellow LED: shows that
transistor 1 is switched
Yellow LED: shows that
transistor 2 is switched
Green LED: shows that the device is energized
Red LED: shows an error; see chap. 10.5
Fig. 34 Position of the icons and description of the LEDs with or without display module
The LEDs of the display module are duplicated on the electronic board that is located under the display module: these LEDs can only be seen if the device has no dispaly module.
Icon Possible cause and alternatives
Probe in good condition and fluid temperature within the set range.
If the monitoring of the impedance on the electrodes and/or the fluid temperature has been acti­vated, the alternative icons in this position are:
, associated with : see chap. 9.13.2, 9.13.3, 9.15.1 and 10.5
, associated with
ERR
: see chap. 9.13.2, 9.13.3, 9.15.1 and 10.5
The device is measuring.
The alternative icons in this position are:
!
flashing: HOLD mode activated (see chap. 9.12.1)
: running check that the outputs are working and behaving correctly (see chap. 9.14.2 and
9.14.3)
"maintenance" message; see chap. 9.12.4, 9.15.1 and 10.5
"warning" message; see chap. 9.11.10, 9.12.4 9.13.2, 9.13.3, 9.15.1 and 10.5
ERR
"error" message; see chap. 9.13.2, 9.13.3, 9.15.1 and 10.5
37
English
Page 38
Typ 8202 ELEMENT
Adjustment and commissioning

9.7.2 Knowing the display at the power-up of the device

When the device is switched on or the display module mounted on the electronic module, the display indicates the software version of the display.
The display then shows the first screen in the Process level:
N
E
P
O
mV_pH
TempC
Fig. 35 Display indications after power-up of the device
L
O
C
1423 mV
23.8 °C
K
See chap. 9.11.5 and 9.11.6 to choose the data to be displayed in the Process level.

9.8 Knowing the Process level

A
mV
1)
1)
Zoom on the value of the
First view of the Process
level.
Zoom on the value in the
first line.
mV_pH
1423 mV
TempC
mV_pH
1423
55 °C
B
Display of the current
outputs.
first current output.
AC1
AC2
AC1
18.3 mA
7.5 mA
38
18.3 mA
Zoom on the value in the
second line.
TempC
55
°C
1)
Zoom on the value of the
second current output.
AC2
7.5 mA
Display of the highest and
lowest values of the physical
quantity chosen
Max
Min
1450mV
1200mV
B
1)
To choose the process values to be displayed see chap. 9.11.5
2)
The display of the lowest and highest values in the Process level is deactivated by default. To activate the feature and
choose the process values see chap. 9.11.3
2)
Reset Yes/No
A
English
Page 39
Typ 8202 ELEMENT
Adjustment and commissioning

9.9 Accessing the Configuration level

Any display of the
Process level
> 2s
This is
This is when the device is be­ing parame-
tered............
....................
when the device is be­ing parame-
tered............
....................
Param
Parameters Menu
This is when the device is be­ing parame-
tered............
....................
Calib
Calibration Menu
This is when the device is be­ing parame-
tered............
....................
Diagnostic
Diagnostic Menu
Wrong
code
Code
"Param"
1)
OK
Code
"Calib"
1)
OK
Code
"Diagnostic"
1)
OK
This is when the device is be­ing parame-
tered............
Param
....................
System Display
Outputs
MEAS BACK
This is when the device is be­ing parame-
tered............
Param
....................
Display Outputs
Sensor
MEAS BACK
Calibration
System
Outputs Sensor
MEAS BACK
Diagnostic
System
Sensor
MEAS BACK
Test
System
Outputs Sensor
MEAS BACK
Test Menu
Test
Code "Test" OK
1)
Info
Error
Info
Information Menu
1)
Only if the access code to the menu has been customized. See chap. 9.11.3, 9.12.2, 9.13.1 and 9.14.1.
Warning
Maintenance
MEAS BACK
Info
Smiley Software
Product
MEAS BACK
→ See chap. 9.10 for the detailed menu functions.
English
39
Page 40
Typ 8202 ELEMENT
Adjustment and commissioning

9.10 Knowing the structure of the menus on the Configuration level

See chap. 9.9 to access the Configuration level.
Param
This is when the device is be­ing parame-
tered............
....................
System
This is when the device is be­ing parame-
tered............
....................
Up/Download Download
Upload
Date
Time
Code
Factory reset
YYYY/MM/DD
HH:MM
0*** Confirm code 0***
Reset Yes/No
Downl. Yes/No
Upload Yes/No
ss
If an “upload” has been done with this module
Display
This is when the device is be­ing parame-
tered............
....................
Line1 / Line2:
Min/Max: EnabledStatus:
PVar:
Unit:
Filter:
PVar:
EnabledLine1 / Line2:
Disabled
pH
mV_pH
mV_ORP
TempF
TempC
pH
mV
°F
°C
None
Fast
Slow
Disabled
pH
mV_pH
mV_ORP
TempF
TempC
If PVar = "pH"
If PVar = "mV_pH" or "mV_ORP"
If PVar = "TempF"
If PVar = "TempC"
40
English
Contrast
Backlight
Unit:
xx%
xx%
pH
mV
°F
°C
If PVar = "pH"
If PVar = "mV_pH" or "mV_ORP"
If PVar = "TempF"
If PVar = "TempC"
Page 41
Typ 8202 ELEMENT
Adjustment and commissioning
Param
This is when the device is be­ing parame-
tered............
....................
Outputs HWMode sink/NPN
This is when the device is be­ing parame-
tered............
....................
AC1/AC2
source/PNP
PVar:
4mA:
20mA:
Filter:
Mode Diag.: None
TR1 / TR2
PVar: pH
pH
mV_pH
mV_ORP
TempF
TempC
INPUT
INPUT
None
Fast
Slow
22mA
mV_pH
mV_ORP
TempF
TempC
warning
Sensor Type: pH
This is when the device is be­ing parame-
tered............
....................
Mode:
Mains Fcy:
Calib System Hold
Code 0*** Confirm code 0***
Outputs AC1/AC2
Mode:
Low:
High:
Contact:
Delay:
ORP
Symmetric
Asymmetric
50Hz
60Hz
Hold:Disabled
Hold:Enabled
4mA
20mA
Hysteresis
Window
INPUT
INPUT
Normally open
Normal. closed
INPUT
INPUT
INPUT
If PVar "warning"
If PVar "warning"
"warning"
If PVar
English
41
Page 42
Typ 8202 ELEMENT
Adjustment and commissioning
Calib
Diagnostic System
Sensor
Probe
Temperature
Code 0*** Confirm code 0***
Calib. Temp.
Calibration
Probe constant
Calib. interval
INPUT
NO
Auto
Constant
1st point
2nd point?
YES
Rinse
2nd point
Cal.Result
Offset:
Span:
Last cal. date
Interval
If pH probe
INPUT
INPUT
VALUE
INPUT
Glass electrodeSensor
Ref. electrode
Temperature
Activate: Yes/No
Impedance:
Warn high: INPUT
Warn low: INPUT
Err high: INPUT
Err low: INPUT
Activate: Yes/No
Temperature:
Warn high: INPUT
Warn low: INPUT
Err high: INPUT
Err low: INPUT
READ
READ
42
English
Page 43
Typ 8202 ELEMENT
Adjustment and commissioning
Test
Info Error
System
Outputs
Sensor
Software
Product READ
Code 0*** Confirm code 0***
AC1: INPUT
AC2:
TR1: OFF/ON
TR2: OFF/ON
PVar: pH
Value:
MESSAGE
MESSAGEWarning
MESSAGEMaintenance
MESSAGESmiley
Main
Sensor
INPUT
mV_pH
mV_ORP
TempF
TempC
INPUT
READ
READ

9.11 Knowing the Parameters Menu

9.11.1 Transferring data from one device to another

See chap. 9.9 to access the Parameters menu.
The function is only possible with a display module with software version V2 and a transmitter with a V2 software version of the acquisition / conversion module for the measured process values.
• On the transmitter, check the sofware version of the acquisition / conversion module for the measured process values in the menu Info -> Software -> Main.
• The software version of the display module is displayed when the display module is energized.
• The “DOWNLOAD” function is only available if an UPLOAD has been successfully performed.
• Never interrupt an upload or download procedure else the transmitter may be damaged.
43
English
Page 44
Typ 8202 ELEMENT
Adjustment and commissioning
This is when the device is be­ing parame-
tered............
....................
SystemParam
This is when the device is be­ing parame-
tered............
....................
Up/Download Download
Upload
Downl. Yes/No
Upload Yes/No
The following data can be transferred from one device to another device of the same type:
• user settings in the menu PARAM (except the date, the time, the contrast level and the brightness of the display),
• user settings in the menu DIAGNOSTIC,
• the periodicity of calibrations set in the menu CALIB
• the access codes to the menus.
DOWNLOAD: transfer the data previously uploaded into the display module using the UPLOAD function.
The parameters transferred are used by the device as soon as the message “Download OK” is displayed.
UPLOAD: upload data from the transmitter to the display module.

9.11.2 Setting the date and time

See chap. 9.9 to access the Parameters menu.
Param YYYY/MM/DD
This is when the device is be­ing parame-
tered............
....................
System Date
This is when the device is be­ing parame-
tered............
....................
Time
HH:MM
ss
DAT E: set the date (input format: year/month/day in the form YYYY/MM/DD)
seconds
TIME: set the time (input format: hours:minutes
)

9.11.3 Modifying the PARAM menu access code

See chap. 9.9 to access the Parameters menu.
Param
This is when the device is be­ing parame-
tered............
....................
If the default code (0000) is entered, the code will not be requested to access the menu.
System
This is when the device is be­ing parame-
tered............
....................
Code
0*** Confirm code 0***
Enter the new code
Confirm the new code
44
English
Page 45
Typ 8202 ELEMENT
Adjustment and commissioning
9.11.4 Restoring the default parameters of the Process level and
the outputs
See chap. 9.9 to access the Parameters menu.
The following data can be restored to their default values:
• user settings in the menu PARAM (except the date, the time, the contrast level and the brightness of the display),
• user settings in the menu DIAGNOSTIC,
• the periodicity of calibrations set in the menu CALIB
• the access codes to the menus.
Reset Yes/No
This is when the device is be­ing parame-
tered............
....................
SystemParam
This is when the device is be­ing parame-
tered............
....................
Factory Set
Execute
→ Choose “Yes” to restore the default values.
→ Choose “No” to keep the set values.

9.11.5 Setting the data displayed in the Process level

See chap. 9.9 to access the Parameters menu.
WARNING
Risk of injury due to wrong adjustment.
▶ Before setting the parameters for the display, choose the type of probe mounted on the transmitter
(see chap. 9.11.11).
Param
This is when the device is be­ing parame-
tered............
....................
Display
This is when the device is be­ing parame-
tered............
....................
Line1 / Line2:
PVar:
EnabledLine1 / Line2:
Disabled
pH
mV_pH
mV_ORP
TempF
TempC
Unit:
Filter:
PVAR: choose the process value to be displayed in the line selected.
UNIT: choose the unit for the process value displayed.
pH
mV
°F
°C
None
Fast
Slow
If PVar = "pH"
If PVar = "mV_pH" or "mV_ORP"
If PVar = "TempF"
If PVar = "TempC"
45
English
Page 46
Typ 8202 ELEMENT
Adjustment and commissioning
FILTER: choose the filter level for the measurement values displayed on the line selected. Three filter levels are proposed: "slow", "fast" or "none". Fig. 36 shows the 3 filter curves.
30 s
t
6 s
t
150 ms
t
"slow" "fast" "none"
Fig. 36 Filter curves

9.11.6 Displaying the lowest and highest values measured

See chap. 9.9 to access the Parameters menu.
WARNING
Risk of injury due to wrong adjustment.
▶ Before setting the parameters for the display, choose the type of probe mounted on the transmitter
(see chap. 9.11.11).
Param
This is when the device is be­ing parame-
tered............
....................
Display
This is when the device is be­ing parame-
tered............
....................
Min/Max: EnabledStatus:
PVar:
Disabled
pH
mV_pH
mV_ORP
TempF
TempC
46
Unit:
pH
mV
°F
°C
If PVar = "pH"
If PVar = "mV_pH" or "mV_ORP"
If PVar = "TempF"
If PVar = "TempC"
STATUS: choose to display (choice “Enabled”) or not display (choice “Disabled”) the highest and lowest measured values (of the process value chosen in PVAR hereafter) since the latest reset or the power-up of the device.
PVAR: choose the process value which highest and lowest measured values are displayed in the Process level.
UNIT: choose the preferred unit in which the lowest and highest measured values are displayed.
English
Page 47
Typ 8202 ELEMENT
Adjustment and commissioning

9.11.7 Setting the display contrast and brightness

See chap. 9.9 to access the Parameters menu.
On a version with a single M12 fixed connector and if the power supply is lower than 16 V, do not increase the backlight over 14%, to not influence the 4-20mA current output.
Param
This is when the device is be­ing parame-
tered............
....................
Display
This is when the device is be­ing parame-
tered............
....................
Contrast xx%
Backlight xx%
→ Set each percentage using and .
CONTRAST: Choose the display contrast level (as a %).
BACKLIGHT: Choose the brightness of the display (as a %).
These settings only affect the display module. They are not factored in during a device data UPLOAD (see chap. 9.11.1).

9.11.8 Choosing the output wiring mode

See chap. 9.9 to access the Parameters menu.
Param
This is when the device is be­ing parame-
tered............
....................
Outputs
This is when the device is be­ing parame-
tered............
....................
HWMode
The setting has no effect on a version with one fixed connector, if the sole current output is wired (see Fig. 23 in chap. 8.3.3).
source/PNP
sink/NPN
The wiring mode is the same for all outputs.
→ If "sink/ NPN" is set, wire the current outputs in sinking mode and the transistor outputs in NPN.
→ If "source/ PNP" is set, wire the current outputs in sourcing mode and the transistor outputs in PNP.
See the wiring for the outputs in chap. 8.3.
47
English
Page 48
Typ 8202 ELEMENT
Adjustment and commissioning

9.11.9 Setting the parameters of the current outputs

See chap. 9.9 to access the Parameters menu.
WARNING
Risk of injury due to wrong adjustment.
▶ Before setting the parameters for the display, choose the type of probe mounted on the transmitter
(see chap. 9.11.11).
The 2nd current output "AC2" is only available on a version with 2 fixed connectors.
Param
This is when the device is be­ing parame-
tered............
....................
Outputs
This is when the device is be­ing parame-
tered............
....................
AC1/AC2
PVar:
4mA:
20mA:
Filter:
Mode Diag.: None
pH
mV_pH
mV_ORP
TempF
TempC
INPUT
INPUT
None
Fast
Slow
22mA
PVAR: choose a process value associated with current output 1 or current output 2 respectively.
Functions "4mA" and "20mA" are used to define the measurement range for the process value associated with the current on the 4-20 mA output.
P1 and P2 are the values associated with a current of 4 mA or 20 mA respectively:
is higher than P2, the signal is inverted and the range P1-P2 corresponds to the range for the 20-4 mA
If P
1
current.
mA
20
48
4
0
P
1
P
Process value chosen
2
Fig. 37 4-20 mA current depending on the process value selected
4mA: choose the value of the process value (previously selected), associated with a current of 4 mA, for each
current output.
20mA: choose the value of the process value (previously selected), associated with a current of 20 mA, for each current output.
English
Page 49
Typ 8202 ELEMENT
Adjustment and commissioning
FILTER: choose the level of damping for the fluctuations of the current value for each current output. Three damping levels are proposed: "slow", "fast" or "none". The damping for the current outputs is similar to the damping of the display (see Fig. 36).
MODE DIAG: choose to emit a current of 22 mA on the current output selected when an "error" event related to diagnostics (see chap. 9.13.2 and 9.13.3) is generated by the transmitter or allow the current output to operate normally (choose "none").
An "error" event linked to a malfunction of the device is always indicated by the generation of a 22 mA current, whatever the adjustment made in the function "MODE DIAG.".
See also chap. "10.5 Solving a problem".

9.11.10 Setting the parameters of the transistor outputs

See chap. 9.9 to access the Parameters menu.
WARNING
Risk of injury due to wrong adjustment.
▶ Before setting the parameters for the display, choose the type of probe mounted on the transmitter
(see chap. 9.11.11).
Param
This is when the device is be­ing parame-
tered............
....................
Outputs
This is when the device is be­ing parame-
tered............
....................
TR1 / TR2
PVar: pH
Mode:
Low:
High:
Contact:
Delay:
Hysteresis
Normally open
Normal. closed
mV_pH
mV_ORP
TempF
TempC
warning
Window
INPUT
INPUT
INPUT
If PVar "warning"
"warning"
If PVar
"warning"
If PVar
PVAR: choose a process value associated with transistor output 1 or transistor output 2 respectively or associate the generation of a "warning" event (see chap. 9.12.4, 9.13.2, 9.13.3 and 9.15.1) to transistor output 1 or tran­sistor output 2 respectively.
If the selected transistor output is linked to the "warning" event, the transistor switches as soon as such an event is generated by the transmitter.
See also chap. "10.5 Solving a problem".
English
49
Page 50
OFF
OFF
OFF
OFF
Typ 8202 ELEMENT
Adjustment and commissioning
MODE: choose the operating mode for transistor output 1 or transistor output 2 (see Fig. 38 and Fig. 39).
LOW: enter the low switching threshold value for transistor output 1 or transistor output 2 (see Fig. 38 and
Fig. 39).
HIGH: enter the high switching threshold value for transistor output 1 or transistor output 2 (see Fig. 38 and Fig. 39).
CONTACT: choose the type of off-position (normally open, NO, or normally closed, NC) for transistor output 1 or transistor output 2 (see Fig. 38 and Fig. 39).
DELAY: choose the value of the time delay prior to switching for each transistor output.
Switching only occurs if one of the thresholds, high or low (functions "High" or "Low"), is exceeded for a duration longer than this time delay. The time delay before switching is applicable to both output thresholds.
Hysteresis operating
The change of status is done when a threshold is detected (increasing measured value: threshold high (function High) to be detected; decreasing measured value: threshold low (function Low) to be detected).
NO = Normally open NC = Normally closed
contact
ON
process value
Low High
Fig. 38 Hysteresis operating
contact
ON
process value
Low High
Window operating
The change of status occurs whenever one of the thresholds is detected.
NO = Normally open NC = Normally closed
contact
ON
process value
Low High
Fig. 39 Window operating
contact
ON
process value
Low High
50
English
Page 51
Typ 8202 ELEMENT
Adjustment and commissioning

9.11.11 Setting the sensor parameters

The monitoring (see chap. 9.13.2) of the redox ("ORP") sensor is impossible if the measurement mode is set to "asymmetrical".
See chap. 9.9 to access the Parameters menu.
Param
This is when the device is be­ing parame-
tered............
....................
Sensor Type: pH
This is when the device is be­ing parame-
tered............
....................
Mode:
ORP
Symmetric
Asymmetric
Mains Fcy:
50Hz
60Hz
TYPE: choose the type of sensor used, pH or Redox ("ORP" choice)
MODE: choose the type of measurement, symmetrical (differential) or asymmetrical.
Symmetrical measurement
The symmetrical measurement is a differential measurement: in this type of measurement, the stainless steel ring on the sensor holder is used as a reference.
(2)
(1)
(1): Measurement electrode
(3)
(2): Reference electrode
(3): Stainless steel ring
Fig. 40 Schematic diagram of symmetrical measurement
Asymmetrical measurement
In the asymmetrical measurement, the measurement is done in relation to the reference electrode.
(2)
(1)
(1): Measurement electrode
(2): Reference electrode
Fig. 41 Schematic diagram of asymmetrical measurement
MAINS FCY: choose the frequency of your mains electricity, 50 or 60 Hz. This frequency is filtered by the trans-
mitter to ensure stable measurements.
English
51
Page 52

9.12 Knowing the Calibration menu

9.12.1 Activating/deactivating the Hold function

See chap. 9.9 to access the Calibration menu.
Typ 8202 ELEMENT
Adjustment and commissioning
Calib System Hold
Hold:Disable
Hold:Enable
If the mode "Hold" is activated and if there is a power interruption, then, when the device restarts, the mode "Hold" is automatically deactivated.
The mode "Hold" is used to carry out maintenance work without interrupting the process.
To activate the mode HOLD:
→ access the "HOLD" function;
→ choose "enabled" and confirm by "OK".
To deactivate the mode HOLD:
→ access the "HOLD" function;
→ choose "disabled" and confirm by "OK".
In practice, when the device is in mode "Hold":
52
• the
!
icon is displayed in place of the icon;
HOLD
• the current emitted on each 4-20 mA output is fixed at the value of the last measurement of the physical parameter associated with each output;
• each transistor output is fixed at the status acquired at the moment the Hold function is activated;
• the device is in Hold mode until the HOLD function is deactivated.

9.12.2 Modifying the Calibration menu access code

See chap. 9.9 to access the Calibration menu.
Code 0*** Confirm code 0***Calib System
Enter the new code
If the default code (0000) is entered, the code will not be requested to access the menu.
Confirm the new code
English
Page 53
Typ 8202 ELEMENT
Adjustment and commissioning

9.12.3 Adjusting the current outputs

See chap. 9.9 to access the Calibration menu.
On a version with a single M12 fixed connector and if the power supply is lower than 16 V, before adjusting the current output, make sure the backlight is lower as 14% (see chap. 9.11.7).
Calib
4mA: adjust the current output 1 or current output 2 for 4 mA.
When the "4mA" function is selected, the transmitter generates a current of 4 mA: measure the current emitted by the 4-20 mA output using a multimeter and enter the value given by the multimeter in the function "AC1.4mA" or "AC2.4mA".
20mA: adjust the current output 1 or current output 2 for 20 mA.
When the "20mA" function is selected, the transmitter generates a current of 20 mA: measure the current emitted by the 4-20 mA output using a multimeter and enter the value given by the multimeter in the function "AC1.20mA" or "AC2.20mA".
Outputs AC1 / AC2
4mA:
20mA:
INPUT
INPUT

9.12.4 Calibrating the pH sensor

DANGER
Risk of injury due to electrical voltage
▶ Observe all applicable accident protection and safety guidelines for electrical equipment.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
See chap. 9.9 to access the Calibration menu.
Before calibrating the sensor, choose the type of probe mounted on the transmitter (see chap. 9.11.11).
English
53
Page 54
Typ 8202 ELEMENT
Adjustment and commissioning
Calib
Sensor
Probe
Temperature
CalibTemperat.
Calibration
Probe constant
Calib. interval
INPUT
NO
Auto
Constant
1st point
2nd point?
YES
Rinse
2nd point
Cal.Result
Offset:
Span:
Last cal. date
Interval
if pH probe
INPUT
INPUT
VALUE
INPUT
CALIB TEMPERAT.: choose the type of temperature compensation for the calibration process: either the tem­perature measured (select "Auto") or a fixed value (select "constant" then enter reference temperature).
→ Calibrate the sensor using one of the following methods:
- CALIBRATION: calibrate the pH sensor in 1 or 2 points (see details hereafter). A calibration process updates the last calibration date ("Last cal. date" function of the CALIB INTERVAL sub-menu hereafter).
- PROBE CONSTANTS: enter the offset and/or span values indicated on the pH sensor certificate, if sup- plied. This input replaces a calibration made by the "Calibration" function above but does not update the last calibration date ("Last cal. date" function of the CALIB INTERVAL sub-menu hereafter).
CALIB INTERVAL: read the date of the last calibration (function "Last cal. date") and set the periodicity of cali­brations, in days (function "Interval"): the transmitter generates a "maintenance" event by displaying the
icon
and a "warning" event, each time a calibration is due. Set function "Interval" to "0000 days" to ignore the function.
• The "warning" event may be associated with one or other or both transistor outputs (see chap. 9.11.10).
• See also chap. "10.5 Solving a problem".
Calibrate the pH sensor ("Calibration" function in the "Probe" menu)
• The pH sensor can be calibrated according to a 1-point or a 2-point procedure.
54
English
Page 55
Typ 8202 ELEMENT
Adjustment and commissioning
• In order not to interrupt the process, activate the HOLD function (see chap. 9.12.1).
• Before each calibration, correctly clean the electrode with a suitable product.
• In a 2-point calibration, the buffer solutions used must be at the same temperature.
• Set the periodicity of calibrations in the "Interval" function in the sub-menu CALIB INTERVAL (see above): each time a calibration is due, the transmitter generates a "maintenance" event and a "warning" event.
1-point calibration
The 1-point calibration procedure is used for quick calibration by adjusting the offset of the measurement graph with a buffer respectively a calibration solution with a known pH (to calibrate a pH sensor: see page 56)
2-point calibration
The 2-point calibration procedure is used for the precise calibration of the offset and the gradient ("span") of the sensor measurement graph.
This operation requires 2 buffer solutions: in general a first solution with a pH of 7 and a second solution with a pH very close to that of the process value to be measured (see page 56).
English
55
Page 56
Detailed procedure for the 1- or 2-points calibration of the pH probe
Calib Sensor CalibrationProbe
→ Immerse the clean sensor in the first buffer solution; the transmitter alternately
displays:
• the measured pH of the solution
• the measured temperature of the solution
1st point
7.100 pH
Typ 8202 ELEMENT
Adjustment and commissioning
Rinse
6.153 pH
25.00 °C
2nd point
10.03 pH
10.01 pH
OK
7.000 pH
OK
YES
Do you want to
NO
calibrate a 2nd point?
→ Rinse the sensor. When the pH
reaches the desired value, validate rinsing with "OK"
→ Immerse the clean sensor in the second buffer
solution; the transmitter alternately displays:
• the measured pH of the solution
• the measured temperature of the solution
→ Enter the pH of the buffer
solution.
→ Enter the pH of the buffer
solution.
56
Probe
"Error: Span/ offset out of
OK
range"
See Table 1, for the possible cause of the error message.
English
Cal. Result
Offset:-10.70 mV Span:0.0mV/pH
OK
Cal. Result Save: Yes/No
Save or not the calibration result by choosing "Yes" or "No".
The transmitter displays the calibration result.
OK
"Warning: Span/offset out of range"
See Table 1, for the possible cause of the warning message.
Page 57
Typ 8202 ELEMENT
Adjustment and commissioning
At the end of calibration of the pH sensor, two types of messages may be displayed:
Message "span" value "offset" value
"Warning:Span/ offset out of range"
"Error:Span/ offset out of range"
Table 1: Warning and error messages during pH sensor calibration
50 mV/pH < span < 53 mV/pH
or
63 mV/pH < span < 65 mV/pH
< 50 mV/pH
or
> 65 mV/pH
-60 mV < Offset < -35 mV
or
35 mV < Offset < 60 mV
< -60 mV
or
> 60 mV

9.12.5 Calibrating the Redox- ("ORP"-) sensor

Possible cause
Error in the buffer solution.
The probe has reached halfway in its lifespan
The probe must be replaced.
Recommended action
→ Use the correct
buffer solution.
→ The values can
be saved or not.
→ Replace the
probe.
→ Calibrate the
transmitter again when the new probe is in place
DANGER
Risk of injury due to electrical voltage
▶ Observe all applicable accident protection and safety guidelines for electrical equipment.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
See chap. 9.9 to access the Calibration menu.
Before calibrating the sensor, choose the type of probe mounted on the transmitter (see chap. 9.11.11).
Calib
Sensor
Probe
CalibTemperat.
Calibration
Probe constant
Calib. interval
Auto
Constant
1st point
Cal.Result
Offset:
Span:
Last cal. date
Interval
INPUT
INPUT
VALUE
INPUT
Temperature
INPUT
57
English
Page 58
Typ 8202 ELEMENT
Adjustment and commissioning
CALIB TEMPERAT.: choose the type of temperature compensation for the calibration process: either the tem­perature measured (select "Auto") or a fixed value (select "constant" then enter reference temperature).
→ Calibrate the sensor using one of the following methods:
- CALIBRATION: calibrate the Redox ("ORP") sensor in 1 point (see details hereafter). A calibration process updates the last calibration date ("Last cal. date" function of the CALIB INTERVAL sub-menu hereafter).
- PROBE CONSTANTS: enter the offset and/or span values indicated on the Redox ("ORP") sensor cer- tificate, if supplied. This input replaces a calibration made by the "Calibration" function above but does not update the last calibration date ("Last cal. date" function of the CALIB INTERVAL sub-menu hereafter).
CALIB INTERVAL: read the date of the last calibration (function "Last cal. date") and set the periodicity of cali­brations, in days (function "Interval"): the transmitter generates a "maintenance" event by displaying the
and a "warning" event, each time a calibration is due. Set function "Interval" to "0000 days" to ignore the function.
• The "warning" event may be associated with one or other or both transistor outputs (see chap. 9.11.10).
icon
• See also chap. "10.5 Solving a problem".
Calibrate the Redox ("ORP") sensor ("Calibration" function in the "Probe" menu)
• The Redox ("ORP") sensor can be calibrated according to a 1-point procedure.
• In order not to interrupt the process, activate the HOLD function (see chap. 9.12.1).
• Before each calibration, correctly clean the electrode with a suitable product.
• Set the periodicity of calibrations in the "Interval" function in the sub-menu CALIB INTERVAL (see here­tofore): each time a calibration is due, the transmitter generates a "maintenance" event and a "warning" event.
1-point calibration
The 1-point calibration procedure is used for quick calibration by adjusting the offset of the measurement graph with a buffer respectively a calibration solution with a known oxidation reduction potential ("ORP").
58
English
Page 59
Typ 8202 ELEMENT
Adjustment and commissioning
Detailed procedure for the 1-point calibration of the oxidation reduction potential ("ORP") sensor
Calib Sensor CalibrationProbe
→ Immerse the clean sensor in the buffer solution; the transmitter alter-
nately displays:
• the measured potential difference of the solution
• the measured temperature of the solution
Calibration
465.0 mV
475.0 mV
OK
Cal. Result
Probe
OK
"Error: Span/ offset out of range""
see Table 2 for the possible cause of the error message.
At the end of calibration of the oxidation reduction potential ("ORP") probe, two types of message may be displayed:
Message '"offset" value Possible cause Recommended action warning:
"Warning:Span/ offset out of range"
-60 mV < Offset < -35 mV
or
35 mV < Offset < 60 mV
Offset:-55.60mV
OK
Cal. Result Save: Yes/No
Save or not the calibration result by choosing "Yes" or "No".
Error in the buffer solution.
The probe has reached halfway in its lifespan
→ Enter the potential difference
of the reference solution (indicated on the bottle).
The transmitter displays the calibration result.
"Warning:
OK
Span/offset out of range"
see Table 2 for the possible cause of the warning message.
→ Use the correct buffer
solution.
→ The values can be saved or
not.
error: "Error:Span/ offset out of range"
Table 2: Warning and error messages when calibrating an oxidation reduction potential ("ORP") probe
< -60 mV
or
> 60 mV
The probe must be replaced.
→ Replace the probe.
→ Calibrate the transmitter
again when the new probe is in place
English
59
Page 60
Typ 8202 ELEMENT
Adjustment and commissioning

9.12.6 Entering an offset for the temperature measurement

See chap. 9.9 to access the Calibration menu.
The temperature transmitted by the Pt1000 probe may be corrected. This correction value is the temperature offset.
Calib Sensor Temperature INPUT
Enter the temperature offset (input limits: ±5 °C)

9.13 Knowing the Diagnostic menu

9.13.1 Modifying the Diagnostic menu access code

See chap. 9.9 to access the Diagnostic menu.
Code 0*** Confirm code 0***Diagnostic System
Enter the new code Confirm the new code
If the default code (0000) is entered, the code will not be requested to access the menu.

9.13.2 Monitoring the condition of the probe

See chap. 9.9 to access the Diagnostic menu.
Diagnostic
Too low or too high an impedance value of one or the other electrode of the measuring probe may indicate that the probe is broken or used.
Measurement mode (see chap. 9.11.11)
Glass electrode Reference
Symmetrical Possible Possible Impossible Possible Asymmetrical Possible Impossible Impossible Impossible
Table 3: Possible combinations of electrode monitoring depending on the measurement mode, symmetrical or
asymmetrical
Glass electrodeSensor
Ref. electrode
Activate: Yes/No
Impedance
Warn hi: INPUT
Warn lo: INPUT
Err hi: INPUT
Err lo: INPUT
READ
Monitoring of the impedance
pH probe Redox ("ORP") probe
Glass electrode Reference electrode
electrode
60
English
Page 61
ERR
Typ 8202 ELEMENT
Adjustment and commissioning
To be warned when an electrode has too low or too high an impedance:
→ activate monitoring on this electrode in the function "activate", then
→ set an impedance range (in MW for the pH electrode and in kW for the reference electrode of the pH/redox
("ORP") probe) outside of which the transmitter generates a "warning" event and displays the icons.
→ set an impedance range (in MW for the pH electrode and in kW for the reference electrode of the pH/redox
("ORP") probe) outside of which the transmitter generates an "error" event and displays the
When the transmitter generates a "warning" or "error" event:
and
→ go into the "Info" menu to read the cause of the event generation.
→ and/or go into the "Sensor" function of the Diagnostic menu to read the impedance values for each electrode
in order to identify the cause of an out of range impedance.
and
icons.
→ if necessary, clean then recalibrate the measurement probe or replace it.
• The "warning" event may also be associated with one or other or both transistor outputs. See chap. 9.11.10, function "Output.TR1" or "Output.TR2".
• The "error" event may also be associated with one or other or both current outputs. See chap. 9.11.9, function "Output.AC1" or "Output.AC2".
• See also chap. "10.5 Solving a problem"..
ACTIVATE: choose whether or not to activate monitoring of the impedance of the selected electrode.
IMPEDANCE: read the impedance measured of the selected electrode.
WARN HI: enter the impedance value above which a "warning" event is generated.
WARN LO: enter the impedance value below which a "warning" event is generated.
ERR HI: enter the impedance value above which an "error" event is generated.
ERR LO: enter the impedance value below which an "error" event is generated.
61
English
Page 62
Typ 8202 ELEMENT
Adjustment and commissioning

9.13.3 Monitoring the fluid temperature

See chap. 9.9 to access the Diagnostic menu.
The function allows for monitoring the fluid temperature and configure the behaviour of the device if the param­etered ranges are exceeded.
A malfunction in your process or the built-in temperature probe may be indicated either by too low or too high a fluid temperature or by an incorrect temperature measurement.
Diagnostic
Sensor
Temperature
Activate: Yes/No
Temperature
Warn hi: INPUT
Warn lo: INPUT
Err hi: INPUT
Err lo: INPUT
READ
To be warned when the fluid temperature is too low or too high:
→ activate monitoring of the fluid temperature in the function "activate", then
→ set a temperature range (in °C) outside of which the transmitter generates a "warning" event and displays the
and icons.
→ set a temperature range (in °C) outside of which the transmitter generates an "error" event and displays the
ERR
and
icons.
When the transmitter generates a "warning" or "error" event:
→ go into the "Info" menu to read the cause of the event generation.
→ and/or go into the "Sensor" function of the Diagnostic menu to read the measured temperature value.
→ then check whether the built-in Pt1000 is working correctly by measuring a fluid with a known temperature. If
the Pt1000 is faulty, return the device to Bürkert.
62
→ if the Pt1000 is not the cause of the problem, check the process.
• The "warning" event may also be associated with one or other or both transistor outputs. See chap. 9.11.10, function "Output.TR1" or "Output.TR2".
• The "error" event may also be associated with one or other or both current outputs. See chap. 9.11.9, function "Output.AC1" or "Output.AC2".
• See also chap. "10.5 Solving a problem".
ACTIVATE: choose whether or not to activate monitoring of the fluid temperature.
TEMPERATURE: read the fluid temperature measured in real time through the built-in Pt1000.
WARN HI: enter the fluid temperature value above which a "warning" event is generated.
WARN LO: enter the fluid temperature value below which a "warning" event is generated.
ERR HI: enter the fluid temperature value above which an "error" event is generated.
ERR LO: enter the fluid temperature value below which an "error" event is generated.
English
Page 63
T
Typ 8202 ELEMENT
Adjustment and commissioning

9.14 Knowing the Test menu

9.14.1 Modifying the Test menu access code

See chap. 9.9 to access the Test menu.
Test
If the default code (0000) is entered, the code will not be requested to access the menu.

9.14.2 Checking the outputs functions

See chap. 9.9 to access the Test menu.
• Make sure that the mode "Hold" is deactivated (see chap. 9.12.1).
System
Code 0*** Confirm code 0***
Enter the new code
Confirm the new code
• The
icon is displayed in place of the icon as soon as the check for the correct working of an output has started. During the check the related output does not react according to the measured physical value.
Test
Outputs
AC1: INPUT
AC2:
TR1: OFF/ON
TR2: OFF/ON
INPUT
AC1: check that current output 1 is working correctly by entering a current value and then selecting "OK".
AC2: check that current output 2 is working correctly by entering a current value and then selecting "OK".
TR1: check that transistor output 1 is working correctly by selecting the status of the transistor ("ON" or "OFF")
then "OK".
TR2: check that transistor output 2 is working correctly by selecting the status of the transistor ("ON" or "OFF") then "OK".
63
English
Page 64

9.14.3 Checking the outputs behaviour

T
See chap. 9.9 to access the Test menu.
• Make sure that the mode "Hold" is deactivated (see chap. 9.12.1).
Typ 8202 ELEMENT
Adjustment and commissioning
• The
icon is displayed in place of the icon as soon as the check for the correct working of an output has started. During the check the related output does not react according to the measured physical value.
The feature allows for simulating the measurement of the process value to check if the outputs are correctly configured.
Test
Sensor
PVar: pH
mV_pH
mV_ORP
TempF
TempC
Value:
INPUT
PVAR: choose the process value to be tested.
VALUE: enter a process value selected from the "PVAR" function above to check output behaviour.

9.15 Knowing the Information menu

9.15.1 Reading the cause of events linked to icons

See chap. 9.9 to access the Info menu.
Info Error
Software
Product READ
MESSAGE
MESSAGEWarning
MESSAGEMaintenance
MESSAGESmiley
Main
Sensor
READ
READ
The function allows for reading a short description of the reason why the following icons are displayed by the transmitter:
- ERROR:
ERR
- WARNING:
- MAINTENANCE:
- SMILEY: or
See also chap. "10.5 Solving a problem".
64
English
Page 65
Typ 8202 ELEMENT
Adjustment and commissioning

9.15.2 Reading the software versions

See chap. 9.9 to access the Info menu.
Info
Software
Main
Sensor
READ
READ
The function allows for reading:
• the software version of the acquisition / conversion board ("Main") for the measurable variables,
• the software version of the sensor ("Sensor").

9.15.3 Reading some identification informations of the device

See chap. 9.9 to access the Info menu.
Info
The function allows for reading some of the informations that are on the name plate of the device:
• the type of the device,
• the serial number,
• the order code.
Product READ
English
65
Page 66
Typ 8202 ELEMENT
Maintenance and troubleshooting

10 MAINTENANCE AND TROUBLESHOOTING

10.1 Safety instructions

DANGER
Risk of injury due to electrical voltage.
▶ Disconnect the electrical power for all the conductors and isolate it before carrying out work on the system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high pressure in the installation.
▶ Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Risk of injury due to high fluid temperatures.
▶ Use safety gloves to handle the device.
▶ Stop the circulation of fluid and drain the pipes before loosening the process connections.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
WARNING
Risk of injury due to non-conforming maintenance.
▶ Maintenance must only be carried out by qualified and skilled staff with the appropriate tools. ▶ Ensure that the restart of the installation is controlled after any interventions.

10.2 Cleaning the device

• Activate the HOLD function in the Calibration menu in order not to interrupt the process during cleaning (see chap. 9.12.1).
66
• Always use a cleaning product compatible with the materials from which the device is made.
→ The device can be cleaned with a cloth dampened with water or a detergent compatible with the materials the
device is made of.
Please feel free to contact your Bürkert supplier for any additional information.
English
Page 67
Typ 8202 ELEMENT
Maintenance and troubleshooting

10.2.1 Cleaning the pH/Redox ("ORP") probe

Refer to the instruction manual delivered with the probe used.

10.3 Replacing the probe

DANGER
Risk of injury due to electrical voltage.
▶ Disconnect the electrical power for all the conductors and isolate it before carrying out work on the system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high pressure in the installation.
▶ Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Risk of injury due to high fluid temperatures.
▶ Use safety gloves to handle the device.
▶ Stop the circulation of fluid and drain the pipes before loosening the process connections.
Risk of injury due to the nature of the fluid.
▶ Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
NOTE
The device loses the IP65 and IP67 protection ratings when dismounted.
▶ Protect the inside of the dismounted device.
→ Remove the probe as shown below.
→ Unscrew the nut between the sensor holder and the elec-
tronic module.
English
67
Page 68
Typ 8202 ELEMENT
Maintenance and troubleshooting
→ Remove the electronic module by pulling it straight out. There
may be a slight resistance due to the seal.
→ Unscrew the probe using a suitable wrench.
→ Remove it carefully from the holder.
→ Clean any fluid projections on the electrical contacts on the
holder.
68
Fig. 42 Removing the probe from the holder
→ Fit a new probe into the holder as shown in chap. 7.6
→ Charge the pipe to check that the mounting is tight.
→ Refit the electronic module to the holder as shown in chap. 7.7

10.4 Replacing the seal of the sensor holder

DANGER
Risk of injury due to electrical voltage.
▶ Disconnect the electrical power for all the conductors and isolate it before carrying out work on the system.
▶ If a 12-36 V DC or a 14-36 V DC powered version is installed either in a wet environment or outdoors, all the
electrical voltages must be of max. 35 V DC.
▶ All equipment connected to the 8619 shall be double insulated with respect to the mains according to the
standard IEC 61010-1:2010.
▶ Observe all applicable accident protection and safety regulations for electrical equipment.
The seal can be replaced without dismounting the holder from the pipe.
English
Page 69
A
Typ 8202 ELEMENT
Maintenance and troubleshooting
→ Unscrew the nut between the sensor holder and the electronic module.
→ Remove the electronic module by pulling it straight out. There may be a slight
resistance due to the seal.
→ Remove the used seal "A" from the holder.
→ Place the new seal "A" into the groove of the holder.
Fig. 43 Replacing the seal of the sensor holder
→ Charge the pipe to check that the mounting is tight.
→ Refit the electronic module to the holder as shown in Fig. 16 of chap. 8.2.
English
69
Page 70

10.5 Solving a problem

Red
Current
LED
output
ON 22 mA depending
ON 22 mA depending
Transistor output
on thresholds
on thresholds
Icon Message dis-
played in the Info menu
"E:Sat. ORP Stage"
ERR
+
"E:Sat. pH Stage"
"Sensor not found" The connection to
ERR
+
Typ 8202 ELEMENT
Maintenance and troubleshooting
Possible cause Recommended action
The pH or redox ("ORP") value is
→ Check the correct wiring of
the earth points. not being correctly measured due to the saturation of the
→ Check the equipotential
surfaces of the installation. input stage of the
measuring board.
→ Switch the power supply
the measurement module is interrupted.
off then on again.
→ If the error persists, return
the device to Bürkert.
ON 22 mA depending
on thresholds
ON 22 mA depending
on thresholds
ON 22 mA depending
on thresholds
ON 22 mA depending
on thresholds
"S EE Fact Read" Factory data is lost.
ERR
+
The device continues to measure but with a bad accuracy.
"S EE User Read" User data for the
ERR
+
sensor is lost (eg. type of sensor).
"S EE User Write"
"S PT Missing" The connection to
ERR
+
the Pt1000 probe is lost.
The temperature is displayed in the Process level with "+++++°C/°F".
"S PT Regulation" The fluid temperature
ERR
+
is not being correctly measured.
The process is stopped.
→ Switch the power supply
off then on again.
→ If the error persists, return
the device to Bürkert.
→ Switch the power supply
off then on again.
→ Check the sensor param-
eters in all the "Sensor"
menus then save them
again.
→ If the error persists, return
the device to Bürkert.
→ Check that the nut between
the sensor holder and the
electronic module is cor-
rectly screwed.
→ If the error persists, return
the device to Bürkert.
→ Switch the power supply
off then on again.
→ If the error persists, return
the device to Bürkert.
70
English
Page 71
Typ 8202 ELEMENT
Maintenance and troubleshooting
Red
Current
LED
output
ON 22 mA depending
Transistor output
on thresholds
Icon Message dis-
+
Possible cause Recommended action played in the Info menu
"TR COM Measure" The acquisition/con-
ERR
version module of
the process values is
faulty.
The process is
stopped.
→ Switch the power supply
off then on again.
→ If the error persists, return
the device to Bürkert.
ON 22 mA depending
on thresholds
ON 22 mA depending
on thresholds
OFF 4-20 mA depending
on thresholds
OFF 4-20 mA depending
on thresholds
"TR EE Fact Read" Parameter reading
ERR
+
"TR EE User Read"
"TR EE UserWrite" Parameter saving
ERR
+
"S Diff pH ORP"
+
"S Buff Diff pH"
"S Buff Diff ORP"
"S RTC Reinit" The date and time
+
error.
error.
The pH or redox
("ORP") value is
not being correctly
measured due to the
saturation of the input
stage of the meas-
uring board (common
mode).
are lost because the
device has not been
powered up for at
least 5 days.
The message is only
displayed at the first
powering up.
→ Switch the power supply
off then on again.
→ If the error persists, set the
device back to the default settings (chap. 9.11.4).
→ If the error persists, return
the device to Bürkert.
→ Switch the power supply
off then on again.
→ Save the settings again.
→ If the error persists, set the
device back to the default settings (chap. 9.11.4).
→ If the error persists, return
the device to Bürkert.
→ Check the wiring.
→ Check the equipotential
surfaces of the installation.
→ Set the date and time
(see chap. 9.11.2).
→ Energize the transmitter for
at least 4 hours so that the date and time are battery fed for 5 days.
English
71
Page 72
Red
Current
LED
output
ON 22 mA
ON 22 mA
Transistor output
1)
depending on thresholds
1)
depending on thresholds
Icon Message displayed
Possible cause Recommended action
in the Info menu
"E:Impedance Ref" The impedance of the
ERR
+
reference electrode is out of range.
The message is dis­played if the monitoring of the impedance of the reference electrode has been activated, depending on the set thresholds ERR LO and ERR HI (see chap. 9.13.2).
"E:Imped.Glass The impedance of the
ERR
+
pH electrode is out of range.
The message is dis­played if the monitoring of the impedance of the pH electrode has been activated, depending on the set thresholds ERR LO and ERR HI (see chap. 9.13.2).
Typ 8202 ELEMENT
Maintenance and troubleshooting
→ Go into the "Sensor"
function of the Diag­nostic menu to read the impedance value of the reference electrode (chap. 9.13.2).
→ If necessary, clean
then recalibrate the measurement sensor or replace it.
→ Go into the "Sensor"
function of the Diag­nostic menu to read the impedance value of the pH electrode (chap. 9.13.2).
→ If necessary, clean
then recalibrate the measurement sensor or replace it.
ON 22 mA
depending on thresholds
"E:Temperature" The fluid temperature is
ERR
+
out of range.
The message is dis­played if the monitoring of the fluid tem­perature has been activated, depending on the set thresholds ERR LO and ERR HI (see chap. 9.13.3).
→ Go into the "Sensor"
function of the Diag­nostic menu to read the measured fluid temper­ature (chap. 9.13.3).
→ If necessary, check
whether the built-in Pt1000 is working correctly by measuring a fluid with a known
1)
temperature.
→ If the Pt1000 is faulty,
return the device to Bürkert.
→ If the Pt1000 is not the
cause of the problem, check the process.
1)
if the MODE DIAG function of the "Output.AC1" or "Output.AC2" menu is set to "22 mA" (see chap. 9.11.9); else, the
current output delivers a standard current between 4 and 20 mA
72
English
Page 73
Typ 8202 ELEMENT
Maintenance and troubleshooting
Red
Current
LED
output
OFF 4-20 mA Switched
Transistor output
Icon Message displayed in
2)
+
OFF 4-20 mA Switched
2)
+
OFF 4-20 mA Switched
2)
+
Possible cause Recommended action
the Info menu
"W:Impedance Ref" The impedance of the
reference electrode is out of range.
The message is dis­played if the monitoring of the impedance of the reference electrode has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.2).
"W:Imped. Glass" The impedance of the
pH electrode is out of range.
The message is dis­played if the monitoring of the impedance of the pH electrode has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.3).
"W:Temperature" The fluid temperature is
out of range.
The message is dis­played if the monitoring of the fluid temperature has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.3).
→ Go into the "Sensor"
function of the Diag­nostic menu to read the impedance value of the reference elec­trode (chap. 9.13.2).
→ If necessary, clean then
recalibrate the meas­urement sensor or replace it.
→ Go into the "Sensor"
function of the Diag­nostic menu to read the impedance value of the pH electrode (chap. 9.13.3).
→ If necessary, clean then
recalibrate the meas­urement sensor or replace it.
→ Go into the "Sensor"
function of the Diagnostic menu to read the measured fluid temperature (chap. 9.13.3).
→ If necessary, check
whether the built-in Pt1000 is working correctly by meas­uring a fluid with a known temperature.
→ If the Pt1000 is faulty,
return the device to Bürkert.
→ If the Pt1000 is not
the cause of the problem, check the process.
2)
If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see chap. 9.11.10); else, the
transistor outputs are operating depending on the set thresholds.
English
73
Page 74
Typ 8202 ELEMENT
Maintenance and troubleshooting
Red LED
Current output
Transistor output
Icon Message dis-
played in the
Possible cause Recommended action
Info menu
OFF 4-20 mA Switched
2)
"M:Calib. Date" A calibration is due.
The periodicity of the cal-
→ Calibrate the sensor
(chap. 9.12.4).
ibrations is set within the "INTERVAL" function of the "CALIB INTERVAL" menu (see chap. 9.12.4).
2)
If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see chap. 9.11.10); else, the
transistor outputs are operating depending on the set thresholds.
74
English
Page 75
Typ 8202 ELEMENT
Spare parts and accessories

11 SPARE PARTS AND ACCESSORIES

CAUTION
Risk of injury and/or damage caused by the use of unsuitable parts.
Incorrect accessories and unsuitable replacement parts may cause injuries and damage the device and the sur­rounding area.
▶ Use only original accessories and original replacement parts from Bürkert.
Spare part Orde code
Seal in EPDM, Ø 46x2 mm, for the sensor holder 559 169
Accessory Order code
Display module 559 168 Set with 2 opaque covers, with seals:
- 1 screw cover with 1 EPDM seal
- 1 quarter turn closing cover with 1 silicone seal Set with 2 transparent covers, with seals:
- 1 screw cover with 1 EPDM seal
- 1 quarter turn closing cover with 1 silicone seal pH probe, 0...14 pH, 0...+80 °C, 0...6 bar, FLATRODE pH 120 mm 561 025 pH probe, 0...14 pH, -10...+60 °C, 0...6 bar, LOGOTRODE pH 120 mm 427 114 pH probe, 0...14 pH, 0...+130 °C, 0...6 bar, UNITRODE PLUS pH 120 mm 560 376 pH probe, 0...14 pH, 0...+130 °C, 0...16 bar, CERATRODE pH 120 mm 418 319 pH probe, 0...14 pH, -10...+40 °C, 0...6 bar, PLASTRODE pH 120 mm 560 377 ORP probe, -2000...+2000 mV, 0...+80 °C, 0...6 bar, FLATRODE Redox 120 mm 561 027
ORP probe, -2000...+2000 mV, -10...+50 °C, 0...6 bar, LOGOTRODE Redox 120 mm
ORP probe, -2000...+2000 mV, 0...+130 °C, 0...6 bar, UNITRODE Redox 120 mm 560 378 Storage solution for pH/ORP probe (KCl 3M), 500 ml 418 557 Buffer solution, 500 ml, pH = 4.01 418 540 Buffer solution, 500 ml, pH = 7 418 541 Buffer solution, 500 ml, pH = 10.01 418 543 Reference solution, ORP = 475 mV, 500 ml 418 555 Set of Cleaning solution for pH/ORP probes, 3x500 ml 560 949 5-pin female M12 connector, to be wired 917 116 5-pin female M12 female connector, moulded on shielded cable (2 m) 438 680 5-pin male M12 connector, to be wired 560 946 5-pin male M12 connector, moulded on shielded cable (2 m) 559 177
560 948
561 843
560 379
English
75
Page 76
Typ 8202 ELEMENT
Packaging, Transport

12 PACKAGING, TRANSPORT

NOTE
Damage due to transport
Transport may damage an insufficiently protected device.
▶ Transport the device in shock-resistant packaging and away from humidity and dirt.
▶ Do not expose the device to temperatures that may exceed the admissible storage temperature range. ▶ Protect the electrical interfaces using protective plugs.

13 STORAGE

NOTE
Poor storage can damage the device.
▶ Store the device in a dry place away from dust. ▶ Storage temperature of the device without probe: -10 to +60°C.
▶ Storage temperature of the device with probe: refer to the instruction manual delivered with the probe used.

14 DISPOSAL OF THE PRODUCT

→ Dispose of the device and its packaging in an environmentally-friendly way.
NOTE
Damage to the environment caused by parts contaminated by the fluid.
▶ Comply with the national and/or local regulations which concern the area of waste disposal.
76
English
Page 77
Page 78
www.burkert.com
Loading...