Easy
Click
Date Time
SYS
mmHg
DIA
mmHg
Puls
1/min
memo
distance
2 cm
H
C
B
G
3
English
What you should know about blood pressure
Blood pressure constantly changes throughout the day. It rises sharply in the early morning and declines during the late morning. Blood pressure rises again in the afternoon and finally drops to a low level at night. Also, it may vary in a short period of time. Therefore, readings from successive measurements can fluctuate.
SYS
mmHG
DIA
mmHG
140
120
100
80
60
65 h12 18 0
Blood pressure readings from a healthy 31-year old male, measured at 5-minute intervals.
Blood pressure measured in a doctor’s office only provides a momentary value. Repeated measurements at home better reflect one's actual blood pressure values under everyday conditions.
Moreover, many people have a different blood pressure when they measure at home, because they tend to be more relaxed than when in the doctor's office. Regular blood pressure measurements taken at home can provide your doctor with valuable information on your normal blood pressure values under actual «everyday» conditions.
The World Health Organisation (WHO) has set up the following standard blood pressure values when measured at resting pulse.
Blood pressure (mmHg)
SYS = systole
Normal values
up to 140
Mild hypertension
140 – 180
Severe hypertension
over 180
= upper value
DIA = diastole
up to 90
90 – 110
over 110
= lower value
4
What you should know about this blood pressure monitor
For accurate measurements, all wrist blood pressure monitors must be positioned at heart level. Example: If the monitor is positioned 20 cm below heart level as shown on the left (e.g. by putting the arm on a table), the
20 cm
reading you will obtain is approximately 15 mmHg too high.
Therefore, the Braun SensorControl has an active positioning system (A): it guides you to hold your measuring arm in the right position at heart level.
Additionally, your blood pressure monitor features an EasyClick cuff which automatically fastens correctly to your wrist to ensure an accurate fit for an accurate reading.
To ensure accurate measuring results, carefully
read the complete use instructions.
• This product is intended for household use only. Keep product and batteries away from children.
• People suffering from cardiac arrhythmia, vascular constriction, arteriosclerosis, diabetes or users of cardiac pacemakers should consult their doctor before measuring their blood pressure themselves, since deviations in blood pressure values may occur in such cases.
• Please consult your doctor in case of any doubt, especially if you are under medical treatment or receiving medication.
• The use of this blood pressure monitor is not intended as a substitute for consultation with your doctor.
5
Product description
A Positioning system B On/off button C Select button D Memory button («memo») E Cuff closure F Release button G Battery compartment H Distance aid
Inserting batteries
• Use alkaline batteries, type LR 03 (AAA) (e.g. Duracell) as supplied with this product (or rechargeable batteries).
LR 06
• Remove the battery compartment cover and insert
LR 03
two batteries with correct polarity (see symbol in the battery compartment (G)).
• New batteries (Duracell) will provide approximately 300 measurements.
• Only discard empty batteries. They should not be disposed of in the household waste, but at appropriate collection sites or at your dealer.
Key rules for accurate blood pressure measurements
Correct handling is key for obtaining accurate measuring results!
• Always take readings at the same time of day, ideally in the morning, under the same conditions.
• Do not measure within 30 minutes after smoking, coffee or tea consumption.
• Take off wristwatch and jewelry before fitting the appliance on the measuring arm.
• Position the monitor 2 cm
• While taking a measurement, sit down, relax, keep still, especially the hand and fingers of the measuring arm, do not move or speak.
• Hold the appliance at heart level during the measurement.
• Wait approximately 3 minutes before repeating a measurement.
below the palm line.
6
Measuring
For a reliable blood pressure measurement, please follow the instructions below.
To open the cuff, press the release button (F) underneath the cuff closure.
2 cm
distance
2 cm
memo
click!
1. Place the monitor housing on the slim side of your bare forearm with the monitor facing up. The distance aid (H) helps you to keep the required 2 cm
distance to
your palm line.
2. To close the cuff, hold the appliance against your body. Then take hold of the cuff closure, pull out the cuff and insert the cuff closure in the monitor housing until it locks into place. The cuff will adjust automatically to length needed for your wrist. Once cuff is closed, do not move the appliance anymore. If you want to change the monitor position on your arm, open and close the cuff once again.
3. Press the red on/off button (B).
4. Put the hand with the appliance on the other upper arm as shown. Slowly move the hand upward along the upper arm. Watch the positioning system (A) and move your hand up or down according to the direction of the arrow shown.
7
5. The positioning system (A) shown on the left and a beep confirm that you have reached the right measuring position. The measurement will start automatically. When the measurement is complete, a beep will sound again, and the reading is displayed:
Upper (systolic) value Lower (diastolic) value Pulse
Please note:
• In case of system errors or handling mistakes, the display will show an error code, instead of giving you inaccurate results (for details, please see section «What to do if…»).
• 15 seconds after switching on the appliance, the measuring process will start automatically, even if the appropriate measuring position has not been reached.
Turning off
Press the on/off button to turn the product off. Otherwise the product will turn off automatically after 2 minutes.
Memory mode
Your blood pressure monitor can store the readings of 50 measurements. Storing is done automatically after each completed measurement. The memory is non-volatile. This means that you will not loose stored data when changing batteries. The memory cannot be cleared. Once the memory is full, the oldest values will be overwritten.
Recalling stored readings
To recall stored readings, first turn the product off, then press the memory button (D). A single push (m will recall the latest reading. The lower display line successively shows memory index and pulse on the one hand and date and time of the respective measurement on the other hand.
Repeatedly pressing the memory button (D) will call up one by one all stored readings:
m (–) towards older readings n (+) towards younger readings
8
or
n)
Scrolling Keep one side of the memory button pressed for scrolling through the memory.
Press the on/off button (B) to exit the memory mode.
Select mode
Setting date and time
First turn off the appliance. Push the select button (C) for longer than 3 seconds to enter the select mode. Wait for a confirmation beep. The day position of the displayed date starts flashing. Using the memory button (D), you can adjust the date. Press the select button again to make the month position flash, adjust it using the memory button. Press the select button again and proceed the same way to adjust the time (hour/minute).
After having changed date and time, press the on/off button (B) to confirm. The device will then turn off.
Deactivating the positioning system
Should you wish to deactivate the positioning system (e.g. because you want to measure while lying down), proceed as follows: Turn off the product. Push the select button (C) for longer than 3 seconds and wait for the confirmation beep. The display shows the date and time settings. Then repeatedly press the select button until the positioning system symbol ( ) will be displayed next to a flashing «ON».
Press one side of the memory button to change the setting from «ON» (positioning system activated) to «OFF» (positioning system deactivated).
To confirm (B).
Please note:
Unless you confirm a modified setting by pressing the on/off button, the appliance will turn off automatically after 2 minutes without adapting the modified setting.
the change of setting, press the on/off button
9
Storage and cleaning
Clean the product with a soft, slightly damp cloth. Stains on the cuff can be carefully removed with damp cloth. Do not wash or dry clean the cuff. Do not store the product in a place exposed to direct sunlight or high humidity.
Calibration
This product has been calibrated at the time of manu­facture. If used according to the use instructions, per­iodic re-calibration is not required. If at any time, you question the measuring accuracy, please contact a Braun Service Centre.
Manufacturing date is given by the LOT number on the back of the product. The first number after LOT represents the last digit of the year of manufacture. The next three digits give the day of the year of manufacture. An example: LOT 4116 xx xx – this product has been manufactured on the 116th day of the year 2004.
10
What to do if ...
Problem Reason Solution
Batteries are flat. Memory readings can be called up. Measurements are not
(flashing)
possible.
e.g. Error codes 01-29:
................
Device may not be correctly positioned on the wrist.
e.g. Error codes 30 and
................
Measurement or inflation does not
higher: System error
Right measuring position was not reached.
start instantly.
Readings from repeated meas­urements differ considerably.
Blood pressure is a fluctuating value. For healthy adults, deviations of about 10 to 20 mmHg are possible.
Cuff is not properly fitted.
Readings were not con­sistently taken at heart level.
Talking, coughing, laughing, moving etc. when measuring will influence the reading.
Prior to a mea­surement, the
Positioning system (A) is
deactivated. positioning system (A) on the display does not work.
Insert new batteries.
Refit device and repeat measurement as shown and described under «Measuring».
Repeat measurement. If problem persists, have device checked at a Braun Service Centre.
Repeat measurement as shown and described under «Measuring» or de­activate the positioning system (A).
none
Ensure that the appliance is fitted as follows: – on top of the slim side of the forearm – with the display on top – with the battery compartment facing the user’s body
– approx. 2 cm away from the thumb root
For each measurement, proceed exactly as shown and described under «Measuring».
While measuring, relax, keep still, do not move, or speak.
Refer to the chapter «Select mode» to activate the positioning system (A).
11
Specifications
Model: BP 3510
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0 – 320 mmHg (cuff pressure)
Blood pressure 30 mmHg (minimum diastolic value) measurement: 250 mmHg (maximum systolic value)
Pulse rate: 30 – 150 / min
Laboratory accuracy: ± 3 mmHg (cuff pressure)
Inflation: Automatic, soft, adapted to blood pressure
Power supply: 2 batteries, type LR 03 (AAA)
Operating temperature: + 10 °C to + 40 °C
Storage temperature: – 20 °C to + 60 °C
Humidity: Up to 85% r.h.
Cuff: Suitable for wrist circumference of
If device is not used within specified temperature and humidity ranges the technical accuracy can not be ensured.
12
± 5 % of reading (pulse rate)
(e.g. Duracell)
13.5 to 20 cm
Type BF equipment
Subject to change without notice.
This appliance conforms to the following standards: DIN EN 60601-1: 3/96 «Medical electrical equipment» – Part 1: General requirements for safety DIN EN 1060-1: 12/95, AMD 1 09/02 «Non-invasive sphygmomanometers» – Part 1: General requirements DIN EN 1060-3: 9/97 «Non-invasive sphygmomanometers» – Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
This product conforms to the provisions of the EC directive 93 /42 / EEC (Medical Device Directive).
0297
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact the Braun Infoline (see page 2).
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household waste at the end of its useful life. Disposal can take place at a Braun Service Centre or at appropriate collection points provided in your country.
Guarantee
We grant 2 years guarantee on the product commencing on the date of purchase. Within the guarantee period we will eliminate, free of charge, any defects in the appliance resulting from faults in materials or workmanship, either by repairing or replacing the complete appliance as we may choose.
This guarantee extends to every country where this appliance is supplied by Braun or its appointed distributor.This guarantee does not cover: damage due to improper use, normal wear or use as well as defects that have a negligible effect on the value or operation of the appliance. The guarantee becomes void if repairs are undertaken by unauthorised persons and if original Braun parts are not used.
To obtain service within the guarantee period, hand in or send the complete appliance with your sales receipt to an authorised Braun Customer Service Centre.
For UK only:
This guarantee in no way affects your rights under statutory law.
13
Türkçe
Tansiyon hakk∂nda bilmeniz gerekenler
Tansiyon, gün boyunca sürekli olarak deπiµir. Sabah erken saatlerde aniden yükselir ve sabah∂n geç saatlerinde düµer. Öπleden sonra tekrar yükselir ve gece düµük seviyelerde seyreder. K∂sa zaman aral∂klar∂ içinde bile çeµitlilik gösterebilir. Dolay∂s∂yla, tansiyon ölçümleri farkl∂ sonuçlar verebilir.
SYS
mmHG
DIA
mmHG
140
120
100
80
60
65 h12 18 0
5 dakikal∂k aral∂klarla ölçülmüµ 31 yaµ∂ndaki saπl∂kl∂ bir erkeπin tansiyonu.
Doktor muayenehanesinde ölçülmüµ tansiyon sadece anl∂k bir deπer verir. Evde, günlük koµullarda, tekrarlanarak yap∂lm∂µ ölçümler daha doπru deπerleri yans∂t∂rlar.
Çoπu insan∂n tansiyon deπerleri evde ölçüldüπünde doktor muayenehanesinde yap∂lan ölçümlerden farkl∂ sonuç verir çünkü evde doktor muayenehanesinde olduklar∂ndan daha rahat hissederler. Evde günlük koµullar alt∂nda yap∂lan düzenli tansiyon ölçümleri, doktorunuza tansiyonunuz hakk∂nda önemli bilgiler saπlar ve takibini kolaylaµt∂r∂r.
Dünya Saπl∂k Organizasyonu taraf∂ndan dinlenme pozisyonunda saptanan standart tansiyon deπerleri aµaπ∂daki gibidir:
Tansiyon (mmHg) Normal Orta Yüksek
SYS = büyük tansiyon
140 a kadar
140 – 180
180 üzeri
= üst deπer
DIA = küçük tansiyon
= alt deπer
90 a kadar
90 – 110
110 üzeri
14
Bu tansiyon ölçer hakk∂nda bilmeniz gerekenler
Doπru ölçüm için bütün tansiyon ölçerler kalp seviyesinde pozisyonland∂r∂lmal∂d∂r. Örnek: Eπer alet, solda görüldüπü gibi kalp seviyesinden 20 cm aµaπ∂da pozisyonland∂r∂l∂rsa (mesela kolunuzu masan∂n üzerine koyarak), elde
20 cm
edeceπiniz ölçüm yaklaµ∂k olarak 15 mmHg yüksektir.
Böylelikle, Braun SensorControl aktif konumlandırma sistemine sahiptir (A): ölçüm yaptıπınız kolu kalp hizasında doπru pozisyonda tutmanız için sizi yönlendirir.
Ayr∂ca, tansiyon ölçeriniz EasyClick bileklik özelliπi sayesinde bileπinizin çevresine otomatik olarak doπru bir biçimde oturarak doπru ölçüm saπlar.
• Doπru ölçüm sonuçları almak için, tüm kullanma talimatlarını dikkatlice okuyunuz.
• Bu ürün sadece evde kullan∂m için üretilmiµtir. Aletinizi ve pillerini çocuklardan uzak tutun.
• Kalpte ritm bozukluπundan, µeker hastal∂π∂ndan, damar sertliπinden µikayeti olanlar ve kalp pili kullananlar∂n, bu gibi durumlarda ölçüm deπerlerinde sapmalar olabileceπinden ölçüm yapmadan önce doktorlar∂na dan∂µmalar∂ tavsiye edilir.
• Ωüphelendiπiniz herhangi bir durumda doktorunuzla temasa geçin. (Özellikle tedavi görüyor ya da ilaç al∂yorsan∂z.)
Bu tansiyon ölçerin amac∂ doktor konsültasyonunun yerne geçmek deπildir.
15
Ürün tan∂m∂
A Konumlandırma Sistemi B Açma/Kapama düπmesi C Seçim (select) düπmesi D Haf∂za düπmesi («memo») E Bileklik kapama ucu F Serbest b∂rakma düπmesi G Pil bölmesi H Mesafe yardımı
Pilleri yerleµtirme
• Ürünün içindeki gibi LR 03 (AAA) tipi alkalin piller (örn. Duracell) (ya da µarj edilebilir piller) kullan∂n.
• Pil bölmesi kapaπ∂n∂ ç∂kar∂n ve pilleri kutuplar∂na uygun
LR 06
LR 03
olarak yerleµtirin (pil bölmesindeki iµaretlere bakabilirsiniz (G)).
• Yeni piller (Duracell) yaklaµ∂k 300 ölçüm yapman∂z için yeterlidir.
Sadece boµ pilleri at∂n. Piller ev çöplüπüne at∂lmamal∂d∂r. Özel pil çöplerine at∂lmal∂d∂r.
Doπru ölçüm için önemli kurallar
Kesin ölçüm sonuçları almak için doπru tutuµ çok önemlidir.
• Ölçümü her zaman günün ayn∂ saatinde ve ayn∂ koµullar alt∂nda yap∂n∂z.
• Çay, kahve ve sigara tüketiminden en az yar∂m saat sonra ölçüm yap∂n∂z.
• Tansiyon ölçerken kolunuzun üzerinde saat veya herhangi bir mücevherat olmamal∂d∂r.
• Cihazı avuç baµlangıcı çizgisinden 2 cm
• Ölçüm alırken, oturun, rahatlayın, hareket etmeyin, özellikle ölçüm yaptıπınız elinizi ve parmaklarınızı kıpırdatmayın ve konuµmayın.
• Ölçüm sırasında cihazı kalp hizasında tutun.
∑ki ölçüm aras∂nda en az 3 dakika bekleyiniz.
geriye takınız.
16
Loading...
+ 37 hidden pages