Bio-Rad Aminex Organic Acid and Alcohol Analysis Columns User Manual

Page 1
Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547
M1250131 Rev B
Standard
Cartridge Holder and
Deashing Holder
Instruction Manual
Page 2
Warranty
Introduction
The Standard Cartridge Holder is a convenient, reusable holder that will accommodate any of Bio-Rad’s 30 x
4.6 mm HPLC cartridges. Its function is to connect the cartridge to the HPLC system.
The Deashing Holder accommodates two 30 x 4.6 mm cartridges in series. It is typically used with the Cation H
+
and
CO
3
-
carbohydrate deashing cartridges.
Installation
The inlet and outlet connections on the holders are zero dead volume connec­tions which use compression type reverse nuts and ferrules for attachment to the HPLC system (see Figure 1). Nuts and ferrules are provided with the holders. To achieve the best results, new connections should be swaged. This is mandatory when replacing cartridge holders manu­factured prior to March 1987. If tubing
1
Page 3
should be set at 0.1-0.2 ml/minute while the cartridge is being inserted. To fit the cartridge into the holder, remove one end of the holder. Loosen the other end a few turns. Insert the cartridge into the holder with firm pressure. The cartridge will fit into the holder securely. Fit the free end of the holder onto the cartridge and screw it down. With the Deashing Holder, the cen­ter fitting should be placed between the two cartridges in the holder. The holder ends need only be screwed down finger­tight. Further tightening is unnecessary, and may damage the seal or holder. Never use tools to tighten this seal.
Temperature and Pressure Limits
The maximum temperature for the holder is 60 °C. System pressure should not exceed 4,000 psi.
Leakage
If the cartridge holder leaks at less than its rated pressure limit, do not try to force a seal. A leak indicates that the seals are either dirty, scratched, or damaged. If the cartridge holder leaks, the seals should be cleaned or replaced (Catalog number 125-0147).
3
must be cut, file or grind the tubing end as necessary so that the end cut is perpendic­ular to the tubing axis, and so that the internal diameter has not been constricted.
Fig. 1. Connections to the holder.
Use 1/16" 316 stainless steel tubing,
0.010" ID, to connect the holder. See Figure 1. Tighten the assembly of tubing, ferrule, and nut finger-tight. Be sure the tubing end bottoms squarely in the holder fitting. Using a wrench, tighten the nut 3/4 turn past finger-tight. The ferrule will be permanently set against the tubing. The connection only needs to be tight enough to seal, and should not be over-tightened. Repeat this procedure for the other end of the cartridge holder.
Cartridge Installation
Prior to inserting the cartridge, all air should be removed from the tubing. After all air bubbles are removed, the flow rate
2
Nut Ferrule
Holder
1/16" stainless steel tubing
Page 4
When using a buffer as mobile phase, rinse the holder’s end fittings with water after removing the cartridge. This will prevent salt precipitation which would scratch or damage the seals.
Specifications
Materials in contact with the mobile phase are 316 stainless steel and Kel-F
®
fluorocarbon.
Ordering Information
Catalog Number Product Description
125-0131 Standard Cartridge Holder 125-0139 Deashing Holder 125-0028 1/16” Reverse Nuts with Ferrules, Parker
type (2 each)
125-0046 1/16” Reverse Nuts with Ferrules, Valco
type (2 each)
125-0027 HPLC Connector, 1/16” LDV Reverse Nut
Union (1)
125-0040 Stainless Steel Tubing, 1/16” x .010” ID x 3”
length, pkg. 5
125-0041 Stainless Steel Tubing, 1/16” x 0.10” ID x 6”
length, pkg. 5 125-0147 Seal Replacement Kit 125-0148 Replacement Seals