![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg1.png)
Bio-Scale™Mini
Profinity™GST
Cartridges, 1 and 5 ml
Instruction Manual
Catalog #
732-4620
732-4622
732-4624
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg2.png)
Table of Contents
Section 1...Introduction ........................................1
Section 2 Product Information............................2
Section 3 Connection to Low-Pressure
Chromatography Systems..................6
Section 4 Connection to Medium- and High-
Pressure Chromatography Systems .10
Section 5...Buffers and Methods........................12
Section 6 Quick Solubility Screening
Protocols..........................................13
Section 7 Preparation of E. coli Lysates ...........17
Section 8 Preparing a Cartridge, and
Subsequent Purification....................19
Section 9 Scaling Up........................................23
Section 10 Regenerating, Cleaning,
Sanitizing, and Storage.....................24
Section 11 Troubleshooting Guide......................26
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg3.png)
Section 12 Ordering Information.........................29
Section 13 References .......................................31
Section 14 Legal Notices ...................................32
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg4.png)
Section 1
Introduction
Bio-Scale Mini GST cartridges are convenient,
disposable, prepacked low-pressure chromatographic
cartridges. Bio-Scale Mini cartridges offer both
increased run-to-run reproducibility and high purity
of protein through a patent-pending column design
and novel resin technology. Compatible with aqueous
buffers most commonly used for protein purification,
Bio-Scale Mini cartridges offer improved performance for protein separation needs.
Profinity glutathione support is based on Bio-Rad's
proprietary UNOsphere
™
technology (US patent
6,423,666) for capture and purification of glutathione
S-transferase (GST)-tagged proteins. Its ligand
density has been optimized for maximum capture of
target proteins. Ideal for scale-up, the Profinity
glutathione support’s open pore structure is ideal for
purifying proteins of a wide molecular weight range.
(Smith DB et al., 1988)
Bio-Scale Mini GST cartridges are packed with
1
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg5.png)
Bio-Rad's innovative Profinity GST resin. Structural
characteristics such as the polymeric nature,
optimized ligand density, and open pore structure of
the Profinity GST bead result in superb mechanical
strength and performance of the prepacked
cartridges.
Section 2
Product Information
Bio-Scale Mini cartridges are disposable, easy-to-use,
prepacked chromatographic cartridges supplied
ready for use in convenient 1 ml and 5 ml sizes.
Cartridges are available for a variety of
chromatographic techniques, including desalting,
ion exchange (IEX), and affinity (AC) chromatography.
See Ordering Information for a listing of the complete
Bio-Scale Mini cartridge product line.
Bio-Scale Mini cartridges are quickly connected to
liquid chromatography systems or luer-end syringes.
The cartridges can be used with any liquid
chromatography system capable of setting a high
pressure limit of 45 psi (equivalent to 3 bar or
2
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg6.png)
300 kPa). Alternatively, luer fittings offer convenient
connection directly to a Luer-Lok syringe for quick,
one-step purification.
Table 1 Bio-Scale Mini GST cartridge
specifications
Sizes 1 ml and 5 ml bed volumes
Dimensions 1 ml: 40 mm length x 5.6 mm inner
diameter
5 ml: 40 mm length x 12.6 mm
inner diameter
Maximum pressure tolerance 45 psi
Recommended flow rates 1 ml: 1.0–2.0 ml/min (240–480 cm/hr)
5 ml: 5–10 ml/min (240–480 cm/hr)
Maximum flow rate 1 ml: 6 ml/min (1,440 cm/hr)
5 ml: 20 ml/min (963 cm/hr)
Fittings Female luer inlet and
male luer outlet
Column material Polypropylene
Frit material Polyethylene (HDPE)
Shipping conditions 20% ethanol
Storage recommendations 20% ethanol or 2% Benzyl Alcohol
Autoclavability Not autoclavable
3
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg7.png)
Table 2. Profinity GST Resin Specifications
Functional ligand Glutathione Derivative
Base bead UNOsphere
Particle size range 45–90 µm
Mean particle size 70 µm
Functional group density ≥60 µmol/g
Dynamic binding capacity* >11 mg/ml
Recommended linear flow rate <600 cm/hr at 25°C
Maximum operating pressure ≥43 psi
Chemical compatibility See table 3
Storage 4–8°C
Shelf life in 20% EtOH >1 year at 4–8°C
Operational temperature 4–40°C
* 60% Breakthrough for a purified 51 kD GST fusion protein at 0.5 ml/min.
Note: Dynamic binding capacity is a function of a number of factors
including pH, flow rate, and sample temperature.
Profinity GST cartridges are compatible with aqueous
buffers most commonly used with GST purification
techniques.
4
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg8.png)
Table 3. Buffer and Chemical Compatibilities for
Profinity GST cartridges
Reagent Stability
Buffer reagents
Tris 50 mM
HEPES 50 mM
MOPS 50 mM
Sodium or potassium phosphate 50 mM
Chelating agents
EDTA, EGTA 5 mM
Sulfhydryl reagents
β−Mercaptoethanol 30 mM
DTT 5 mM
TCEP 10 mM
Detergents
Nonionic detergents (Triton, Tween, NP-40) 5%
Cationic detergents (CTAB) 1%
Zwitterionic detergents (CHAPS, CHAPSO) 5%
Anionic detergents (SDS, Sarkosyl) 1%
Denaturing agents (for cleaning only)
Guanidine-HCl 6 M
Urea 8 M
Other additives
NaCl 2 M
MgCl
2
100 mM
CaCl
2
10 mM
Glycerol 20%
Ethanol 20%
Citrate 80 mM
5
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bg9.png)
Section 3
Connection to
Low-Pressure
Chromatography Systems
Bio-Scale Mini cartridges are ideal for use with
Bio-Rad's BioLogic™LP chromatography system,
Econo™Gradient pump, the patented* Model EP-1
Econo pump, and all low-pressure chromatography
instruments. Bio-Scale Mini cartridges can be
conveniently connected directly to the system using
the luer fittings on the cartridge.
1. Install 1.6 mm inner diameter (ID) tubing in the
pumphead. Adjust the platen pressure screw
(on pumphead) using a screwdriver or coin.
Turn the screw counterclockwise as far as it will
go, then turn clockwise three full turns.
Assemble with fittings and lock rings as shown
in Figure 1.
* US patent 5,135,658
6
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bga.png)
(Use orange lock rings and medium size barb fittings
with 1.6 mm tubing.)
Fig. 1. BioLogic LP system setup.
2. To maximize gradient accuracy and to apply
samples efficiently, install 1.6 mm ID tubing
from the pump to the MV-6 sample inject valve
(if available). If using the MV-6 sample inject
valve, turn the knob counterclockwise as far as
it will go so it will now correspond to the printed
diagram on the valve (see Figure 2).
Platen pressure
screw
Lock-ring
Tubing
Luer fittimg
See detail
23
1
9
456
78
0
C .
Alarm
![](/html/8d/8dc3/8dc3a707992d75bc805231199550bcc1ef724290a3ec82fb95992177b648bbc7/bgb.png)
Fig. 2. Connecting to an MV-6 valve.
3. Connect the inlet of the cartridge to the male
luer fitting on the MV-6 sample inject valve (see
Figure 2). If not using the MV-6 sample inject
valve, connect a barb to male luer fitting on the
1.6 mm ID tubing, then connect to the top of
the female luer on the Bio-Scale Mini cartridge.
For optimum performance, a cartridge should
be mounted vertically with the arrow on the
cartridge pointing downward (see Figure 3).
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
SAMPLE LOOP
INJECT
PORT
MV-6
TO
INJECT
VALVE
FILL
TO
COLUMN
WASTE
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
TO
MV-6
INJECT
VALVE
SAMPLE LOOP
INJECT
PORT
FILL
WASTE
TO
COLUMN
"INJECT" POSITION"FILL" POSITION