Beurer BM 26 User Manual [ru, de, en, fr, es, it, pl]

BM 26
D
Blutdruckmessgerät
Gebrauchsanweisung .......................................... 2 – 11
G
Blood pressure monitor
Instructions for use ............................................12 – 20
F
Tensiométre
E
Tensiómetro
Manual de instrucciones ................................... 30 – 38
I
Misuratore di pressione
Istruzioni per l’uso .............................................. 39 – 48
T
Bilgisayarli tansiyon ölçer
Kullanım kılavuzu ............................................... 49 – 57
r
Прибор для измерения кровяного
давления в плечевой артерии
Инструкция по применению ...........................58 – 68
Q
Ciśnieniomierz
Instrukcja obsługi ............................................... 69 – 77
Electromagnetic Compatibility Information ...........78 – 80
BEURER GmbH • Söfl inger Str. 218 • 89077 Ulm (Germany) Tel.: +49 (0) 731 / 39 89-144 • Fax: +49 (0) 731 / 39 89-255 www.beurer.de • Mail: kd@beurer.de
0483
DEUTSCH
Sehr geehrte Kundin, sehr geehrter Kunde,
wir freuen uns, dass Sie sich für ein Produkt unseres Sorti­mentes entschieden haben. Unser Name steht für hochwertige und eingehend geprüfte Qualitätsprodukte aus den Bereichen Wärme, Gewicht, Blutdruck, Körpertemperatur, Puls, Sanfte Therapie, Massage und Luft. Bitte lesen Sie diese Gebrauchsanweisung aufmerksam durch, bewahren Sie sie für späteren Gebrauch auf, machen Sie sie anderen Benutzern zugänglich und beachten Sie die Hinweise.
Mit freundlicher Empfehlung Ihr Beurer-Team
1. Kennenlernen
Das Oberarm-Blutdruckmessgerät dient zur nichtinvasiven Messung und Überwachung arterieller Blutdruckwerte von erwachsenen Menschen. Sie können damit schnell und einfach Ihren Blutdruck messen, die Messwerte abspeichern und sich den Verlauf und Durch­schnitt der Messwerte anzeigen lassen. Bei eventuell vorhandenen Herzrhythmusstörungen werden Sie gewarnt. Die ermittelten Werte werden nach WHO-Richtlinien eingestuft und grafisch beurteilt. Bewahren Sie diese Gebrauchsanweisung für weitere Be­nutzung auf und machen Sie diese auch anderen Benutzern zugänglich.
2. Wichtige Hinweise
Zeichenerklärung
In der Gebrauchsanweisung, auf der Verpackung und auf dem Typschild des Geräts und des Zubehörs werden folgende Sym­bole verwendet:
Vorsicht
Hinweis Hinweis auf wichtige Informationen
Gebrauchsanweisung beachten
Anwendungsteil Typ BF
Gleichstrom Entsorgung gemäß Elektro- und Elektronik-
Altgeräte EG-Richtlinie 2002/96/EC – WEEE (Waste Electrical and Electronic Equipment)
Hersteller
2
Storage
-20°C
Operating
10°C
Zulässige Lagerungstemperatur und
55°C
- luftfeuchtigkeit
RH ≤85%
Zulässige Betriebstemperatur und
40°C
- luftfeuchtigkeit
RH ≤85%
Vor Nässe schützen
SN Seriennummer
Die CE-Kennzeichnung bescheinigt die Konformität mit den grundlegenden An­forderungen der Richtlinie 93/42/EWG für
0483
Medizinprodukte
Hinweise zur Anwendung
Um eine Vergleichbarkeit der Werte zu gewährleisten, mes-
sen Sie Ihren Blutdruck immer zu gleichen Tageszeiten.
Ruhen Sie sich vor jeder Messung ca. 5 Minuten aus!
Wenn Sie mehrere Messungen an einer Person durchführen
möchten, warten Sie zwischen den einzelnen Messungen jeweils 5 Minuten.
Mindestens 30 Minuten vor der Messung sollten Sie nicht
essen, trinken, rauchen oder sich körperlich betätigen.
Wiederholen Sie die Messung im Falle zweifelhaft gemesse-
ner Werte.
Die von Ihnen selbst ermittelten Messwerte können nur zu
Ihrer Information dienen – sie ersetzen keine ärztliche Un­tersuchung! Besprechen Sie Ihre Messwerte mit dem Arzt,
begründen Sie daraus auf keinen Fall eigene medizinische Entscheidungen (z.B. Medikamente und deren Dosierungen)!
Verwenden Sie das Blutdruckmessgerät nicht bei Neugebo-
renen, Schwangeren und Präeklampsie-Patientinnen.
Erkrankungen des Herz-Kreislaufsystems können zu Fehl-
messungen bzw. zu Beeinträchtigungen der Mess genauig­keit führen. Ebenso der Fall ist dies bei sehr niedrigem Blut­druck, Diabetes, Durchblutungs- und Rhythmusstörungen sowie bei Schüttelfrost oder Zittern.
Das Blutdruckmessgerät darf nicht im Zusammenhang mit
einem Hochfrequenz-Chirurgiegerät verwendet werden.
Verwenden Sie das Gerät nur bei Personen mit dem für das
Gerät angegebenen Umfangbereich des Oberarmes.
Beachten Sie, dass es während des Aufpumpens zu einer
Funktionsbeeinträchtigung des betroffenen Gliedmaßes kom­men kann.
Die Blutzirkulation darf durch die Blutdruckmessung nicht
unnötig lange unterbunden werden. Bei einer Fehlfunktion des Gerätes nehmen Sie die Manschette vom Arm ab.
Vermeiden Sie das mechanische Einengen, Zusammendrü-
cken oder Abknicken des Manschettenschlauches.
Verhindern Sie einen anhaltenden Druck in der Manschette
sowie häufige Messungen. Eine dadurch resultierende Beein­trächtigung des Blutflusses kann zu Verletzungen führen.
Achten Sie darauf, dass die Manschette nicht an einem Arm
angelegt wird, dessen Arterien oder Venen in medizinischer Behandlung sind, z.B. intravaskulärer Zugang bzw. eine intravaskuläre Therapie oder ein arteriovenöser (A-V-) Neben­schluss.
Legen Sie die Manschette nicht bei Personen an, die eine
Brustamputation hatten.
3
Legen Sie die Manschette nicht über Wunden an, da dies zu
weiteren Verletzungen führen kann.
Sie können das Blutdruckmessgerät ausschließlich mit Bat-
terien betreiben.
Die Abschaltautomatik schaltet das Blutdruckmessgerät zur
Schonung der Batterien aus, wenn innerhalb 3 Minuten keine Taste betätigt wird.
Das Gerät ist nur für den in dieser Gebrauchsanweisung
beschriebenen Zweck vorgesehen. Der Hersteller haftet nicht für Schäden, die durch unsachgemäßen oder falschen Ge­brauch verursacht wurden.
Hinweise zur Aufbewahrung und Pflege
• Das Blutdruckmessgerät besteht aus Präzisions- und Elek­tronik-Bauteilen. Die Genauigkeit der Messwerte und Lebens­dauer des Gerätes hängt ab vom sorgfältigen Umgang: – Schützen Sie das Gerät vor Stößen, Feuchtigkeit, Schmutz,
starken Temperaturschwankungen und direkter Sonnenein-
strahlung. – Lassen Sie das Gerät nicht fallen. – Benutzen Sie das Gerät nicht in der Nähe von starken
elektro magnetischen Feldern, halten Sie es fern von Funk-
anlagen oder Mobiltelefonen. – Verwenden Sie nur die mitgelieferte oder originale Ersatz-
Manschetten. Ansonsten werden falsche Messwerte ermit-
telt.
Drücken Sie nicht auf Tasten, solange die Manschette nicht
angelegt ist.
Falls das Gerät über einen längeren Zeitraum nicht benutzt
wird, wird empfohlen die Batterien zu entfernen.
Hinweise zu Batterien
Batterien können bei Verschlucken lebensgefährlich sein.
Bewahren Sie deshalb Batterien und Produkte für Kleinkinder unerreichbar auf. Wurde eine Batterie verschluckt, muss sofort medizinische Hilfe in Anspruch genommen werden.
Batterien dürfen nicht geladen oder mit anderen Mitteln
reakti viert, nicht auseinander genommen, in Feuer geworfen oder kurzgeschlossen werden.
Nehmen Sie die Batterien aus dem Gerät heraus, wenn diese
verbraucht sind oder Sie das Gerät länger nicht benutzen. So vermeiden Sie Schäden, die durch Auslaufen entstehen können. Ersetzen Sie immer alle Batterien gleichzeitig.
Benutzen Sie keine verschiedenen Batterie-Typen,
Batterie-Marken oder Batterien mit unterschiedlicher Kapazi­tät. Verwenden Sie vorzugsweise Alkaline-Batterien.
Hinweise zu Reparatur und Entsorgung
Batterien gehören nicht in den Hausmüll. Bitte entsorgen
Sie die verbrauchten Batterien an den dafür vorgesehenen Sammel stellen.
Öffnen Sie nicht das Gerät. Bei nicht beachten erlischt die
Garantie.
Das Gerät darf nicht selbst repariert oder justiert werden.
Eine einwandfreie Funktion ist in diesem Fall nicht mehr ge­währleistet.
Reparaturen dürfen nur vom Kundenservice oder autori-
sierten Händlern durchgeführt werden. Prüfen Sie jedoch vor jeder Reklamation zuerst die Batterien und tauschen Sie diese gegebenenfalls aus.
4
Bitte entsorgen Sie das Gerät gemäß der Elektro- und
Elektronik Altgeräte EG-Richtlinie 2002/96/EC – WEEE (Waste Electrical and Electronic Equipment). Bei Rück fragen wenden Sie sich bitte an die für die Entsorgung zuständige kommunale Behörde.
3. Gerätebeschreibung
2
1
1. Manschette
2. Manschettenschlauch
3. Manschettenstecker
4. Display
5. START/STOPP-Taste
6. Speichertaste M+
7. Taste SET
8. Funktionstasten -/+
9. WHO-Skala
4
3
WHO
9
+
SET
7
8
6
SYS
mmHg
DIA
mmHg
PULSE
/
min
M
+
Anzeigen auf dem Display:
1. Systolischer Druck
2.
Einheit mmHg
3. Diastolischer Druck
4. Symbol Puls
5. Uhrzeit und Datum
und ermittelter Pulswert
6. Symbol Herzrhythmusstörungen
7. Durchschnittswerte der letzten 3 Messungen AVG
8. Aufpumpen, Luft ablassen (Pfeil)
9. WHO-Einstufung
10. Symbol Batteriewechsel
11. Symbol für Benutzer 1, 2, 3, 4
5
11
1
10
9 8
2
3
7 6
5
4
5
4. Messung vorbereiten
V
Batterie einlegen
Entfernen Sie den Deckel des Bat-
teriefaches auf der Rück seite des Gerätes.
Legen Sie vier Batterien vom
Typ 1,5 V AA (Alkaline Type LR06) ein. Achten Sie unbedingt darauf, dass die Batterien entsprechend der Kennzeichnung mit korrekter Polung eingelegt werden. Verwenden Sie keine wiederauf ladbaren Akkus.
Schließen Sie den Batteriefachdeckel wieder sorgfältig.
Wenn 4 akustische Warntöne ausgegeben werden und gleichzeitig das Symbol
auf dem Display erscheint, ist keine Messung mehr möglich und Sie müssen alle Batterien erneuern. Sobald die Batterien aus dem Gerät entfernt werden, muss die Uhrzeit neu eingestellt werden.Verbrauchte Batterien gehören nicht in den Hausmüll. Entsorgen Sie diese über Ihren Elektrofachhändler oder Ihre örtliche Wertstoff-Sammelstelle. Dazu sind Sie gesetzlich verpflichtet. Hinweis: Diese Zeichen finden Sie auf schadstoff­haltigen Batterien: Pb: Batterie enthält Blei, Cd: Batterie enthält Cadmium, Hg: Batterie enthält Quecksilber.
Datum und Uhrzeit einstellen
Sie sollten Datum und Uhrzeit unbedingt einstellen. Nur so können Sie Ihre Messwerte korrekt mit Datum und Uhrzeit speichern und später abrufen. Die Uhrzeit wird im 12-Stunden­Format dargestellt, d. h. die Uhrzeit ab 13:00 Uhr wird mit 01:00 PM angezeigt.
4 x AA (LR06) 1,5
Zur Einstellung von Datum und Uhrzeit gehen Sie wie folgt vor:
Bestätigen Sie 2 x die Taste SET.
Stellen Sie mit den Tasten + oder - zuerst den Monat 1–12
ein und bestätigen Sie mit der Taste SET.
Stellen Sie Tag/Stunde/Minute ein und bestätigen Sie jeweils
mit der Taste SET.
Drücken Sie die Taste
um den Einstellungsmodus zu ver-
lassen.
5. Blutdruck messen
Manschette anlegen
Legen Sie die Manschette am entblößten linken Oberarm an. Die Durchblutung des Arms darf nicht durch zu enge Kleidungsstücke oder Ähnliches eingeengt sein.
Die Manschette ist am Oberarm so zu platzieren, dass der untere Rand 2 – 3 cm über der Ellenbeuge und über der Arterie liegt. Der Schlauch weist zur Handflächenmitte.
Legen Sie nun das freie Ende der Man schette eng, aber nicht zu stramm um den Arm und schließen Sie den Klettverschluss. Die Manschette sollte so stramm angelegt sein, dass noch zwei Finger unter die Manschette passen.
6
Stecken Sie nun den Manschetten ­schlauch in den Anschluss für den Man schetten stecker.
Achtung: Das Gerät darf nur mit der Original-Manschette betrieben werden. Die Manschette ist für einen Armumfang von 22 bis 35 cm geeignet. Unter der Bestellnummer 162.973 ist eine größere Manschette für Oberarmumfänge von 30 bis 42 cm beim Fachhandel oder bei der Serviceadresse erhältlich.
Richtige Körperhaltung einnehmen
Ruhen Sie sich vor jeder Messung ca. 5 Minuten aus!
Ansonsten kann es zu Abweichungen kommen.
Sie können die Messung im Sitzen oder im Liegen durch-
führen. Achten Sie in jedem Falle darauf, dass sich die Man­schette in Herzhöhe befindet.
Sitzen Sie zur Blutdruckmessung bequem. Lehnen Sie Rü-
cken und Arme an. Kreuzen Sie die Beine nicht. Stellen Sie die Füße flach auf den Boden.
Um das Messergebnis nicht zu verfälschen, ist es wichtig,
sich während der Messung ruhig zu verhalten und nicht zu sprechen.
Speicher wählen
Drücken Sie die Taste Speicherplatz durch drücken der Funktionstasten
SET
. Wählen Sie den gewünschten
-/+
. Sie haben 4 Speicher á 30 Speicherplätze um die Messergebnisse von 4 verschiedenen Personen getrennt voneinander abzuspeichern oder Messungen morgens und abends separat abzuspeichern. Bestätigen Sie Ihre Wahl mit der Taste .
Blutdruckmessung durchführen
Legen Sie, wie zuvor beschrieben, die Manschette an und
nehmen Sie die Haltung ein, in der Sie die Messung durch­führen wollen.
• Starten Sie den Messvorgang durch längeres drücken der
Taste .
Nach dem Prüfen des Displays, wobei alle Ziffern leuchten, pumpt sich die Manschette automatisch auf. Wäh­rend des Aufpumpens ermittelt das Gerät bereits Messwerte, die zur Abschätzung des nötigen Aufpumpdruckes dienen. Sollte dieser Druck nicht ausreichen, pumpt das Gerät auto­matisch nach.
Dann wird der Druck in der Manschette langsam abgelassen
und der Puls erfasst.
Wenn die Messung beendet ist, wird der restliche Luftdruck
sehr schnell abgelassen. Der Puls, der systolische und der diastolische Blutdruck werden angezeigt.
Sie können die Messung jederzeit durch das drücken der
Taste abbrechen.
Wird grundsätzlich ein höherer Aufpumpdruck benötigt, so
können Sie das Nachpumpen umgehen, indem Sie kurz nach Beginn des Aufpumpvorganges die Taste M+ gedrückt hal­ten, bis der gewünschte Manschettendruck erreicht ist. Die­ser sollte ca. 30 mmHg über dem systolischen Wert liegen.
7
Das Symbol Err erscheint, wenn die Messung nicht ord-
nungsgemäß durchgeführt werden konnte. Beachten Sie das Kapitel Fehlermeldung/Fehlerbehebung in dieser Gebrauchs­anweisung und wiederholen Sie die Messung.
Das Gerät schaltet nach ca. 3 Minuten automatisch ab.
Warten Sie vor einer erneuten Messung mindestens 5 Minuten!
Ergebnisse beurteilen Herzrhythmusstörungen:
Dieses Gerät kann während der Messung eventuelle Störungen des Herzrhythmus identifizieren und weist gegebenenfalls nach der Messung mit dem Symbol
darauf hin. Dies kann ein Indikator für eine Arrhythmie sein. Arrhythmie ist eine Krankheit, bei der der Herzrhythmus aufgrund von Fehlern im bioelektrischen System, das den Herzschlag steuert, anormal ist. Die Symptome (ausgelassene oder vorzeitige Herzschläge, langsamer oder zu schneller Puls) können u.a. von Herzerkran­kungen, Alter, körperlicher Veranlagung, Genussmitteln im Über­maß, Stress oder Mangel an Schlaf herrühren. Arrhythmie kann nur durch eine Untersuchung bei Ihrem Arzt festgestellt werden.
Wiederholen Sie die Messung, wenn das Symbol nach der Messung auf dem Display angezeigt wird. Bitte achten Sie da­rauf, dass Sie sich 5 Minuten ausruhen und während der Mes­sung nicht sprechen oder bewegen. Sollte das Symbol oft erscheinen, wenden Sie sich bitte an Ihren Arzt. Selbstdiagnose und -behandlung aufgrund der Messergebnisse können gefähr­lich sein. Befolgen Sie unbedingt die Anweisun gen Ihres Arztes.
WHO-Einstufung:
Gemäß den Richtlinien/Definitionen der Weltgesundheitsor­ganisation (WHO) und neuester Erkenntnisse lassen sich die Messergebnisse gemäß nachfolgender Tabelle einstufen und beurteilen. Diese Standardwerte dienen jedoch lediglich als allgemeine Richtlinie, da der individuelle Blutdruck bei verschiedenen Per­sonen und unterschiedlichen Altersgruppen usw. abweicht. Es ist wichtig, dass Sie Ihren Arzt in regelmäßigen Abständen zu Rate ziehen. Ihr Arzt teilt Ihnen Ihre individuellen Werte für einen normalen Blutdruck sowie den Wert mit, ab dem die Höhe des Blutdrucks als gefährlich einzustufen ist.
Bereich der Blutdruckwerte Systole (in mmHg) Diastole (in mmHg) Maßnahme
Stufe 3: starke Hypertonie > = 180 > = 110 einen Arzt aufsuchen Stufe 2: mittlere Hypertonie 160 – 179 100 – 109 einen Arzt aufsuchen Stufe 1: leichte Hypertonie 140 – 159 90 – 99 regelmäßige Kontrolle beim Arzt Hoch normal 130 – 139 85 – 89 regelmäßige Kontrolle beim Arzt Normal 120 – 129 80 – 84 Selbstkontrolle Optimal < 120 < 80 Selbstkontrolle
Quelle: WHO, 1999
8
Die Balkengrafik im Display und die Skala auf dem Gerät geben an, in welchem Bereich sich der ermittelte Blutdruck befindet. Sollte sich der Wert von Systole und Diastole in zwei unter­schiedlichen WHO-Bereichen befinden (z.B. Systole im Bereich „Hoch Normal“ und Diastole im Bereich „Normal“) dann zeigt Ihnen die graphische WHO Einteilung auf dem Gerät immer den höheren Bereich an, im beschriebenen Beispiel „Hoch Normal“.
6. Messwerte speichern, abrufen und löschen
• Die Ergebnisse jeder erfolgreichen Messung werden zusam-
men mit Datum und Uhrzeit abgespeichert. Bei mehr als 30 Messdaten gehen die jeweils ältesten Messdaten verloren.
Wählen Sie mit der Taste SET und danach mit den Tasten -/+
den gewünschten Benutzerspeicher und bestätigen Sie mit der Taste schnittswert AVG der letzten 3 gespeicherten Messwerte des Benutzerspeichers angezeigt. Bei Betätigung der Taste + werden die jüngsten Messergebnisse angezeigt, bei der Taste - die ältesten.
Der Durchschnittswert AVG kann erst ab 3 gespeicherten
Werten des jeweiligen Benutzerspeichers angezeigt werden.
Um den Speicher zu löschen, wählen Sie zunächst einen Be-
nutzerspeicher aus. Halten Sie die Taste SET ca. 3 Sekunden gedrückt. Alle Werte des gegenwärtigen Benutzerspeichers werden nach 3 kurzen Signaltönen gelöscht.
Möchten Sie den Benutzerspeicher ändern, beachten Sie
Kapitel „Speicher wählen“.
. Durch drücken der Taste M+ wird der Durch-
7. Gerät reinigen und aufbewahren
Reinigen Sie Gerät und Manschette vorsichtig nur mit einem
leicht angefeuchteten Tuch.
Verwenden Sie keine Reinigungs- oder Lösungsmittel.
Sie dürfen das Gerät auf keinen Fall unter Wasser halten, da
sonst Flüssigkeit eindringen kann und das Gerät beschädigt.
Wenn Sie das Gerät aufbewahren, dürfen keine schweren
Gegenstände auf dem Gerät stehen. Entnehmen Sie die Batterien. Der Manschettenschlauch darf nicht scharf abge­knickt werden.
8. Fehlermeldung/Fehlerbehebung
Bei Fehlern erscheint auf dem Display die Fehlermeldung Err. Fehlermeldungen können auftreten, wenn 1
. der Aufpumpdruck höher als 300 mmHg ist;
2. der Blutdruckmesswert außergewöhnlich hoch oder niedrig ist; 3
. Sie sich während der Messung bewegen oder sprechen.
Wiederholen Sie in diesen Fällen die Messung. Achten Sie darauf, dass der Manschettenschlauch ordnungsgemäß einge­steckt ist und Sie sich nicht bewegen oder sprechen. Setzen Sie gegebenenfalls die Batterien neu ein oder ersetzen Sie diese.
9. Technische Angaben
Modell-Nr. BM 26 Messmethode Oszillometrisch, nicht invasive Blutdruck-
Messbereich Manschettendruck 0 – 300 mmHg,
9
messung am Oberarm
systolisch 60 – 280 mmHg, diastolisch 30 – 200 mmHg, Puls 30 – 180 Schläge /Minute
Genauigkeit der Anzeige
Messunsicherheit Max. zulässige Standardabweichung gemäß
Speicher 4 x 30 Speicherplätze Abmessungen L 155 mm x B 110 mm x H 70 mm Gewicht Ungefähr 395 g (ohne Batterien) Manschettengröße 22 bis 35 cm Zul. Betriebs-
bedingungen Zul. Aufbewah-
rungsbedingungen Stromversorgung
Batterie­Lebensdauer
Zubehör
Klassifikation Interne Versorgung, IPX0, kein AP oder
Änderungen der technischen Angaben ohne Benachrichtigung sind aus Aktualisierungsgründen vorbehalten.
• Dieses Gerät entspricht der europäischen Norm EN60601-1-2
und unterliegt besonderen Vorsichtsmaßnahmen hinsichtlich der elektromagnetischen Verträglichkeit. Bitte beachten Sie
Systolisch ± 3 mmHg, diastolisch ± 3 mmHg, Puls ± 5 % des angezeigten Wertes
klinischer Prüfung: systolisch 8 mmHg / diastolisch 8 mmHg
+10 °C bis +40 °C, ≤ 85 % relative Luft­feuchte (nicht kondensierend)
- 20 °C bis + 55 °C, ≤ 85 % relative Luft­feuchte, 800 –1060 hPa Umgebungsdruck
4 x 1,5 V Type LR06)
Für ca. 180 Messungen, je nach Höhe des Blutdrucks bzw. Aufpumpdruck
Gebrauchsanweisung, 4 x 1,5 V AA Batterien (Alkaline Type LR06), Aufbewahrungstasche
APG, Dauerbetrieb, Anwendungsteil Typ BF
AA Batterien (Alkaline
10
dabei, dass tragbare und mobile HF-Kommunikationseinrich­tungen dieses Gerät beeinflussen können. Genauere Angaben können Sie unter der angegebenen Kundenservice-Adresse anfordern oder am Ende der Gebrauchsanweisung nachlesen.
• Das Gerät entspricht der EU-Richtlinie für Medizinproduk­te 93/42/EC, dem Medizinproduktegesetz und den Normen EN1060-1 (nicht invasive Blutdruckmessgeräte Teil 1: Allgemei­ne Anforderungen), EN1060-3 (nicht invasive Blutdruckmess­geräte Teil 3: Ergänzende Anforderungen für elektromechani­sche Blutdruckmesssysteme) und IEC80601-2-30 (Medizinische elektrische Geräte Teil 2 – 30: Besondere Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale von automatisierten nicht invasiven Blutdruckmessgeräten).
• Die Genauigkeit dieses Blutdruckmessgerätes wurde sorgfältig geprüft und wurde im Hinblick auf eine lange nutzbare Lebens­dauer entwickelt. Bei Verwendung des Gerätes in der Heilkunde sind Messtechnische Kontrollen mit geeigneten Mitteln durch­zuführen. Genaue Angaben zur Überprüfung der Genauigkeit können unter der Service-Adresse angefragt werden.
10. Garantie
Wir leisten 3 Jahre Garantie für Material- und Fabrikationsfehler des Produktes. Die Garantie gilt nicht:
• im Falle von Schäden, die auf unsachgemäßer Bedienung beruhen,
• für Verschleißteile,
• für Mängel, die dem Kunden bereits beim Kauf bekannt waren,
• bei Eigenverschulden des Kunden.
Die gesetzlichen Gewährleistungen des Kunden bleiben durch die Garantie unberührt. Für Geltendmachung eines Garantie falles
innerhalb der Garantiezeit ist durch den Kunden der Nachweis des Kaufes zu führen. Die Garantie ist innerhalb eines Zeitraumes von 3 Jahren ab Kaufdatum gegenüber der Beurer GmbH, Söf­linger Straße 218, 89077 Ulm, Germany, geltend zu machen. Der Kunde hat im Garantiefall das Recht zur Reparatur der Ware bei unseren eigenen oder bei von uns autorisierten Werkstätten. Wei­tergehende Rechte werden dem Kunden (aufgrund der Garan tie) nicht eingeräumt.
11
ENGLISH
Dear Customer,
Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for the applications in the areas of heat, weight, blood pressure, body temperature, pulse, gentle therapy, massage and air. Please read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.
Best regards, Your Beurer Team
1. Getting to know your instrument
The upper arm blood pressure monitor is used for non-invasive measurement and monitoring of adults’ arterial blood pressure. You can use it to measure your blood pressure quickly and easily, storing the results and displaying the progression of readings together with the average. A warning is issued for anyone suffering from cardiac arrhyth­mia. The values determined are classified and graphically evaluated according to WHO guidelines. Keep these instructions carefully for further use and also let other users have access to them.
12
2. Important information
Signs and symbols
The following symbols are used in these instructions for use, on the packaging and on the type plate for the device and acces­sories:
Caution
Note Note on important information
Follow instructions for use
Type BF applied part
Direct current Disposal in accordance with EC Directive
2002/96/EC – WEEE (Waste Electrical and Electronic Equipment)
Manufacturer
Storage
-20°C
Permissible storage temperature and
55°C
humidity
RH ≤85%
Operating
10°C
Permissible operating temperature and
40°C
humidity
RH ≤85%
Keep dry
SN Serial number
The CE labelling certifies that the product complies with the essential requirements of
0483
Directive 93/42/EEC on medical products
Advice on use
In order to ensure comparable values, always measure your
blood pressure at the same time of day.
Before every measurement, relax for about five minutes.
If you want to perform several measurements on the same
person, wait five minutes between each measurement.
Do not take a measurement within 30 minutes after eating,
drinking, smoking or exercising.
Repeat the measurement if you are unsure of the measured
value.
The measurements taken by you are for your information
only – they are not a substitute for a medical examination! Discuss the measurements with your doctor, and never base any medical decisions on them (e.g. medicines and their administration)!
Do not use the blood pressure monitor on newborns, preg-
nant women or patients with preeclampsia.
Cardiovascular diseases may lead to incorrect measure-
ments or have a detrimental effect on measurement ac-
curacy. The same also applies to very low blood pressure, diabetes, circulatory disorders and arrhythmias as well as chills or shaking.
The blood pressure monitor must not be used in connection
with a high-frequency surgical unit.
Only use the device on people who have the specified upper
arm measurement for the device.
Please note that when inflating, the functions of the limb in
question may be impaired.
During the blood pressure measurement, blood circulation
must not be stopped for an unnecessarily long time. If the device malfunctions, remove the cuff from the arm.
Avoid any mechanical restriction, compression or bending of
the cuff line.
Do not allow sustained pressure in the cuff or frequent meas-
urements. The resulting restriction of the blood flow may cause injury.
Ensure that the cuff is not placed on an arm in which the
arteries or veins are undergoing medical treatment, e.g. intra­vascular access or therapy, or an arteriovenous (AV) shunt.
Do not use the cuff on people who have undergone a mas-
tectomy.
Do not place the cuff over wounds as this may cause further
injury.
The blood pressure monitor can only be operated with bat-
teries.
To conserve the batteries, the monitor switches off automati-
cally if no buttons are pressed for 3 minutes.
The device is only intended for the purpose described in
these instructions for use. The manufacturer is not liable for damage resulting from improper or careless use.
13
Storage and Care
The blood pressure monitor is made up of precision electron-
ic components. Accuracy of readings and the instrument’s service life depend on careful handling.
– You should protect the device from impact, moisture, dirt,
major temperature fluctuations and direct exposure to the
sun’s rays. – Never drop the device. – Do not use near strong electromagnetic fields, i.e. keep it
away from any radio systems and mobile phones. – Only ever use the cuffs provided with the monitor or origi-
nal replacement cuffs. Otherwise erroneous results will be
recorded.
Do not press any buttons until the cuff is in position.
If the instrument is not used for any length of time, we
recommend removing the batteries.
Advice on batteries
Batteries can be fatal if swallowed. You should therefore
store the batteries and products where they are inaccessible to small children. If a battery has been swallowed, call a doc­tor immediately.
Batteries should not be charged or reactivated with any other
means, nor should they be taken apart, thrown in the fire or short-circuited.
Remove the batteries from the instrument if they are worn
out or if you are not going to use the instrument for any length of time. This prevents any damage as a result of leak­age. Always replace all the batteries at the same time.
Never use different types of battery, battery brands or bat-
teries with different capacities. You should preferably use alkaline batteries.
Repair and disposal
Batteries do not belong in domestic refuse. Used batteries
should be disposed of at the collection points provided.
Never open the instrument. If these instructions are not
heeded, the warranty will be null and void.
Never attempt to repair the instrument or adjust it yourself.
We can no longer guarantee perfect functioning if you do.
Repairs may only be performed by Customer Service or
authorized dealers. However, always check the batteries and replace them if necessary prior to making any complaint.
The appliance should be disposed of according to
Regulation 2002/96/EC-WEEE (Waste Electrical and Electronic Equipment). In case of queries, please con­tact the municipal authorities responsible for waste disposal in your area.
14
3. Unit description
2
1
Icons in the display:
4
3
1. Systolic pressure
2. Unit in mmHg
3. Diastolic pressure
WHO
9
+
SET
SYS
mmHg
DIA
mmHg
PULSE
/
min
M
+
5
7
8
6
4. Pulse symbol
5. Time and date
6. Cardiac arrhythmia symbol
7. Average values of the 3 most recent measurements – AVG
8. Pump up, release air (arrow)
9. WHO classification
10. Battery replacement symbol
11. Symbol for users 1, 2, 3 and 4
and calculated pulse value
1. Cuff
2. Cuff line
3. Cuff connector
4. Display
5. START/STOP button
6. Memory button M+
7. SET button
8. Function buttons -/+
9. WHO scale
11
1
10
9 8
2
3
7 6
5
4
15
4. Prepare measurement
V
Inserting battery
Remove the battery cover from the
back of the monitor.
Insert four AA 1.5 V alkaline batter-
ies. Making absolutely sure that you insert them with the correct polarity as marked. Never use rechargeable bat teries.
Replace the battery cover carefully.
If 4 warning tones have been sounded and the symbol appears simultaneously on the display, it is no longer possible to perform a measurement and all batteries must be replaced. Once the batteries have been removed from the device, the time must be set again. Used batteries should not be disposed of in normal household waste. Dispose of them via your elec­tronics retailer or your local recycling point. You are legally obligated to do so. Note: Batteries containing pollutant substances are marked as follows: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury.
Setting date and time
You should set the date and time without fail. Otherwise, you will not be able to save your measured values correctly with a date and time and to access them again later. The time is shown in 12-hour format i.e. the time from 13:00 is displayed as 01:00 PM.
4 x AA (LR06) 1,5
Proceed as follows to set the date and time:
Confirm twice using the SET button.
Use the + or - buttons to first set the month 1–12, and then
confirm using the SET button.
Set the day/hour/minute and confirm each using the SET
button.
Press the button to exit the settings mode.
5. Measuring blood pressure
Positioning cuff
Fit the cuff round your bare left upper arm. Blood circulation in the arm should not be restricted by tight clothing or other objects.
The cuff should be placed on the upper arm so that the lower edge is 2 to 3 cm above the bend of the elbow and above the artery. The tube should be in line with the centre of the palm.
Now place the free end of the cuff snugly, but not too tightly, around the arm, and fix it with the Velcro fastener. The cuff should be fitted tight enough to allow just two fingers to fit beneath the cuff. Insert the cuff tubing into the socket for the cuff attachment.
16
Important: The instrument should only be operated with the original cuff. The cuff is suitable for an arm circumference of 22 to 35 cm. A larger cuff for upper arm circumferences from 30 to 42 cm is available from retailers or the service address under order no.
162.973.
Correct posture
Rest for approx. 5 minutes before each measurement. Other-
wise there may be divergences.
• You can perform the measurement either sitting or lying down.
Always make sure that the cuff is on a level with your heart.
To carry out a blood pressure measurement, make sure you
are sitting comfortably with your arms and back leaning on something. Do not cross your legs. Place your feet flat on the ground.
In order not to distort the result, it is important to keep still
during the measurement and not talk.
Select memory
Press the SET button. Select the desired storage space by pressing the -/+ function buttons. You have 4 memories with 30 spaces each to store the measurements of 4 different peo­ple separately or to separate measurements taken in the morn­ing and at night. Confirm your selection with the button
.
17
Measuring blood pressure
Put on the cuff as described previously and assume the posi-
tion in which you want to carry out the measurement.
Start the measurement by holding the button . After check-
ing the display, during which all numbers light up, the cuff pumps up automatically. Whilst the cuff is pumping up, the device already calculates measurements for estimating the necessary pump pressure. If this pressure is insufficient, the device automatically pumps more pressure.
Then the pressure in the cuff is slowly released and the pulse
is measured.
When the measurement has been completed, the remain ing
air pressure is released very quickly. The pulse, the systolic and the diastolic blood pressure are displayed.
Measuring can be cancelled at any time by pressing the but-
ton .
If a higher pump pressure is always required, you can bypass
the additional pumping process by holding the M+ button shortly after the pumping process starts until the desired cuff pressure is reached. This should be approx. 30 mmHg above the systolic value.
The Err icon appears if it has not been possible to perform
the measurement properly. Observe the section in these in­structions on error messages/troubleshooting and repeat the measurement.
The device switches off automatically after 3 minutes.
Wait at least 5 minutes before taking another measurement!
Evaluating results Cardiac arrhythmia:
This instrument can identify possible cardiac arrhythmia disorders during measurement and if necessary indicates the measurement with the flashing icon
. This may be an indicator for arrhythmia. Arrhythmia is a condi­tion where the heart rhythm is abnormal as a result of defects in the bioelectrical system controlling the heart beat. The symptoms (omitted or premature heart beats, slow or exces­sively fast heart rate) may be caused, among other things, by heart disease, age, physical predisposition, excessive use of stimulants, stress or lack of sleep. Arrhythmia can only be as­certained through examination by your doctor. Repeat the measurement if the flashing icon is displayed after the measurement. Please note that you should rest for 5 minutes between measurements and not talk or move dur-
Range of blood pressure values
Systolic (in mmHg)
ing the measurement. If the icon appears often, please contact your doctor. Any self-diagnosis and treatment based on the test results may be dangerous. It is vital to follow your doctor’s instructions.
WHO classification:
In accordance with the guidelines/definitions of the World Health Organization and the latest findings, the measurements can be classified and assessed according to the following table. However, these standard values serve only as a general guide­line, as the individual blood pressure varies in different people and different age groups etc. It is important to consult your doctor regularly for advice. Your doctor will tell you your individual values for normal blood pres­sure as well as the value above which your blood pressure is classified as dangerous.
Diastolic (in mmHg)
Measure
Grade 3: Severe hypertension > = 180 > = 110 Seek medical advice Grade 2: Moderate hypertension 160 – 179 100 – 109 Seek medical advice Grade 1: Mild hypertension 140 – 159 90 – 99 Have it checked regularly by doctor High-normal 130 – 139 85 – 89 Have it checked regularly by doctor Normal 120 – 129 80 – 84 Check it yourself Optimal < 120 < 80 Check it yourself
Source: WHO, 1999
The bar graph in the display and the scale on the unit indicate the range of the blood pressure which has been recorded. If the values for systolic and diastolic pressure are in two dif-
diastolic pressure in the normal range) the graphic WHO clas­sification on the unit indicates the higher range (high-normal in the example described).
ferent WHO ranges (e.g. systolic in the high-normal range and
18
6. Saving, retrieving and deleting results
The results of each successful measurement are stored
together with date and time. With more than 30 items of measured data, the earliest items of data measured are lost.
Use the SET button and then the -/+ buttons to select the
desired user memory and confirm using the button . Press­ing the M+ button displays the average value AVG of the 3 most recent measurements stored in the user memory. By pressing the + button, the most recent measurements are displayed; by pressing the - button, the oldest are displayed.
The average value AVG can only be displayed if there are 3
saved values in the respective user memory.
To clear the memory, you must first select a user memory.
Hold the SET button pressed for approx. 3 seconds. All the values in the current user memory are deleted after 3 brief beeps are output.
If you wish to change the user memory, see “Select memory”.
7. Cleaning and storing the instrument
Clean your device and cuff carefully only with a slightly mois-
tened cloth.
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube should not have any sharp kinks.
8. Error messages / troubleshooting
In case of faults, the Err message appears in the display. Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement. In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not move or talk. Re-insert the batteries if necessary, or else replace them.
9. Specifications
Model no. BM 26 Measurement
method Measurement range Cuff pressure 0 – 300 mmHg,
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
Measurement inaccuracy
Memory 4 x 30 memory spaces Dimensions L 155 mm x W 110 mm x H 70 mm Weight Approx. 395 g (without batteries) Cuff size 22 to 35 cm
19
Oscillometric, non-invasive blood pres­sure measurement on the upper arm
systolic 60 – 280 mmHg, diastolic 30 – 200 mmHg, Pulse 30 – 180 beats/minute
pulse ± 5 % of the value shown Max. permissible standard deviation
according to clinical testing: systolic 8 mmHg /diastolic 8 mmHg
Permissible operating conditions
Permissible storage conditions
Power supply
+10 °C to +40 °C, ≤ 85 % relative air humidity (non-condensing)
- 20 °C to + 55 °C, ≤ 85 % relative air hu­midity, 800 –1060 hPa ambient pressure
4 x 1,5 V
AA alkaline batteries.
Battery life For approx. 180 measurements,
depending on the blood pressure level and/or pump pressure
Accessories Instruction for use, 4 x 1.5 V AA alkaline
batteries., storage pouch
Classification Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied part
Technical information is subject to change without notification to allow for updates.
• This unit is in line with European Standard EN 60601-1-2 and
is subject to particular precautions with regard to electromag­netic compatibility (EMC). Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Service ad­dress or found at the end of the instructions for use.
• This device is in line with the EU Medical Devices Directive
93/42/EC, the “Medizinproduktegesetz” (German Medical De­vices Act) and the standards EN1060-1 (non-invasive sphyg­momanometers, Part 1: General requirements), EN1060-3 (non-invasive sphygmomanometers, Part 3: Supplemen­tary requirements for electro-mechanical blood pressure measuring systems) and IEC80601-2-30 (Medical electrical
equipment – Part 2 – 30: Particular requirements for the safety and essential performance of automated non
-invasive blood
pressure monitors).
The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful life. If using the device for commercial medical purposes, it must be regularly tested for accuracy by appropriate means. Precise instructions for checking accuracy may be requested from the service address.
20
FRANÇAIS
Chère cliente, cher client,
Nous sommes heureux que vous ayez choisi un produit de notre assortiment. Notre nom est synonyme de produits de qualité haut de gamme ayant subi des vérifications appro­fondies, ils trouvent leur application dans le domaine de la chaleur, du contrôle du poids, de la pression artérielle, de la mesure de température du corps et du pouls, des thérapies douces, des massages et de l’air. Lisez attentivement ce mode d’emploi, conservez-le pour un usage ultérieur, mettez-le à la disposition des autres utilisa­teurs et suivez les consignes.
Avec nos sentiments dévoués Beurer et son équipe
1. Premières expériences
Le lecteur de tension artérielle au bras sert à la mesure non invasive et au contrôle de la tension artérielle chez l’adulte. Vous pouvez ainsi mesurer votre tension artérielle de manière simple et rapide, enregistrer les valeurs mesurées et afficher la courbe et la moyenne des valeurs mesurées. L’appareil vous prévient en cas d’arythmie cardiaque éven­tuelle. Les valeurs obtenues sont classées conformément aux direc­tives de l’OMS et évaluées sur le plan graphique. Conservez ce mode d’emploi pour pouvoir vous y référer ultérieurement et faites en sorte qu’il soit accessible aux autres utilisateurs.
2. Remarques importantes
Symboles utilisés
Les symboles suivants sont utilisés dans le mode d’emploi, sur l’emballage et sur la plaque signalétique de l’appareil et des accessoires:
Attention
Remarque Ce symbole indique des informations impor­tantes
Respectez les consignes du mode d’emploi
Appareil de type BF
Courant continu Élimination conformément à la directive
européenne 2002/96/CE – DEEE relative aux déchets d’équipements électriques et électroniques
Fabricant
21
Storage
-20°C
Operating
10°C
Température et taux d’humidité de stockage
55°C
admissibles
RH ≤85%
Température et taux d’humidité d’utilisation
40°C
admissibles
RH ≤85%
Protéger contre l’humidité
SN Numéro de série
Le sigle CE atteste de la conformité aux exi­gences fondamentales de la directive 93/42/
0483
CEE relative aux dispositifs médicaux
Remarques relatives à l’utilisation
Mesurez toujours votre tension au même moment de la jour-
née afin que les valeurs soient comparables.
Avant toute mesure, reposez-vous environ 5minutes!
Lorsque vous devez effectuer plusieurs mesures sur une
personne, patientez à chaque fois 5minutes entre chaque mesure.
Évitez de manger, boire, fumer ou d’exercer des activités
physiques pendant au moins 30minutes avant la mesure.
Effectuez une nouvelle mesure si vous avez un doute sur les
valeurs mesurées.
• Les mesures que vous avez établies servent juste à vous tenir
informé de votre état- elles ne remplacent pas un examen mé­dical! Parlez-en avec votre médecin, vous ne devez prendre aucune décision d’ordre médical sur la base de ces seules mesures (par ex. choix de médicaments et de leurs dosages)!
• N’utilisez pas le tensiomètre sur des nouveaux-nés, des femmes enceintes et les patientes atteintes de pré-éclampsie.
Les maladies cardio-vasculaires peuvent entraîner des er-
reurs de mesure, plus précisément des mesures imprécises. Ce problème se pose aussi en cas de tension très basse, de diabète, de troubles de la circulation et du rythme cardiaque et de frissons de fièvre ou de tremblements.
Le tensiomètre ne doit pas être utilisé parallèlement à un
appareil chirurgical haute fréquence.
Utilisez uniquement l’appareil sur des personnes dont le péri-
mètre du bras correspond à celui indiqué pour l’appareil.
Veuillez noter que la fonction du membre concerné peut être
entravée lors du gonflage.
Il ne faut pas bloquer la circulation sanguine plus longtemps
que nécessaire au cours de la prise de tension. Si l’appareil ne fonctionne pas bien, retirez le brassard du bras.
Évitez de presser, d’aplatir ou de plier le tuyau du brassard
en le manipulant.
Évitez des mesures trop fréquentes ou une pression continue
du brassard. Elles entraînent une réduction de la circulation sanguine et constituent un risque de blessure.
Veillez à ne pas placer le brassard sur un bras, dont les
artères ou les veines sont soumises à un traitement médical, par exemple en présence d’un dispositif d’accès intravas­culaire destiné à un traitement intravasculaire ou en cas de shunt artérioveineux.
N’utilisez pas le brassard sur des personnes qui ont subi une
mastectomie.
Ne placez pas le brassard sur des plaies, son utilisation peut
les aggraver.
Vous ne pouvez utiliser le tensiomètre qu’avec des piles.
22
L’arrêt automatique permet de faire passer le tensiomètre en
mode économie d’énergie lorsqu’aucune touche n’est mani­pulée pendant un délai de 3 minutes.
L’appareil est conçu pour l’utilisation décrite dans ce mode
d’emploi. Le fabricant ne peut être tenu pour responsable des dommages causés par une utilisation inappropriée ou non conforme.
Remarques relatives à la conservation et à l’entretien
L’appareil de mesure de la tension artérielle est constitué
de pièces électroniques, de grande précision. L’appareil doit être conservé dans un environnement approprié afin de garantir la précision des valeurs et d’optimiser la durée de vie du produit : – Protégez l’appareil des chocs et conservez-le à l’abri de
l’humidité, de la poussière, des variations thermiques et
d’une exposition directe au soleil. – Ne laissez pas tomber l’appareil. – N’utilisez pas l’appareil à proximité de forts champs élec-
tromagnétique. Eloignez-le des radios ou des télé phones
mobiles. – Utilisez uniquement les brassards de rechange fournis
ou d’origine. Dans le cas contraire, vous obtiendrez des
valeurs mesurées erronées.
N’appuyez pas sur les touches tant que vous n’avez pas mis
le brassard.
Au cas où vous ne vous servez pas de l’appareil pendant
une longue période, nous vous recommandons de retirer les piles.
Remarques relatives aux piles
L’ingestion de piles peut se révéler mortelle. Laissez par
conséquent les piles et les produits hors de portée des jeunes enfants. Au cas où une pile a été avalée, faites immé­diatement appel à un médecin.
Les piles ne doivent être ni rechargées ni réactivées par
d’autres méthodes ni démontées ni jetées dans le feu ni court-circuitées.
Lorsqu’elles sont usagées ou si l’appareil ne doit pas être
utilisé avant longtemps, retirez les piles de l’appareil. Vous éviterez ainsi les dommages liés aux fuites. Remplacez tou­jours toutes les piles en même temps.
N’utilisez pas des types ou des marques de piles différents
et n’utilisez pas des piles d’une capacité différente. Utilisez de préférence des piles alcalines.
Remarques relatives à la réparation et à la mise au rebut
Les piles ne sont pas des ordures ménagères. Veuillez jeter
les piles usagées dans les conteneurs prévus à cet effet.
N’ouvrez pas l’appareil. Le non-respect de cette consigne
annulera la garantie.
Vous ne devez en aucun cas réparer ou ajuster l’appareil
vous-même. Le cas contraire, aucun fonctionnement irrépro­chable n’est garanti.
Les réparations doivent être effectuées uniquement par le
service après-vente ou des revendeurs agréés. Cependant avant de faire une réclamation, contrôlez d’abord les piles et changez-les, le cas échéant.
23
Pour éliminer l’appareil, conformez-vous à la directive
sur les appareils électriques et électroniques 2002/96/ CE – WEEE (Waste Electrical and Electronic Equipment). Pour toute question, adressez-vous aux collectivités locales responsables de l’élimination de ces déchets.
3. Description de l’appareil
2
1
1. Brassard
2. Tuyau de brassard
3. Connexion au brassard
4. Écran
5. Touche START/STOP
6. Touche mémoire M+
7. Touche SET
8. Touches de fonction -/+
9. Échelle de l’OMS
4
3
WHO
9
+
SET
SYS
mmHg
DIA
mmHg
PULSE
/
min
M
+
5
7
8
6
Affichages à l’écran :
1. Pression systolique
2. Unité mmHg
3. Pression diastolique
4. Symbole pouls
et valeur du pouls donnée
5. Heure et date
6. Symbole troubles du rythme cardiaque
7. Moyennes des 3 dernières mesures AVG
8. Pompage, dégonflage (flèche)
9. Classement OMS
10. Symbole changement des piles
11. Symbole utilisateur 1, 2, 3, 4
11
10
9 8
7 6
5
24
1
2
3
4
Loading...