Rev.02.15.07
NFS40 Medical
2 of 4
Embedded Power for
Business-Critical Continuity
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
Specifications
EMC CHARACTERISTICS
Conducted emissions EN55022, FCC part 15 Level A
Radiated emissions EN55022, FCC part 15 Level A
ESD air EN61000-4-2, level 3 Perf. criteria 1
ESD contact EN61000-4-2, level 4 Perf. criteria 1
Surge EN61000-4-5, level 3 Perf. criteria 1
Fast transients EN61000-4-4, level 3 Perf. criteria 1
Radiated immunity EN61000-4-3, level 3 Perf. criteria 2
Conducted immunity EN61000-4-6, level 3 Perf. criteria 1
GENERAL SPECIFICATIONS
Hold-up time 110 Vac 18 ms
230 Vac 132 ms
Efficiency 110 Vac, 230 Vac 70% typical
Isolation voltage Input/output 4000 Vac
Input/chassis 1500 Vac
Switching frequency 20-10 kHz
Approvals and VDE0750, IEC601
standards EN60601-1, UL2601
(See Note 12) CSA C22.2 No. 125
Weight 270 g (9.6 oz)
MTBF MIL-HDBK-217E 170,000 hours
ENVIRONMENTAL SPECIFICATIONS
Thermal performance Operating, see curve 0 °C to +70°C
Non-operating -40
°C to +85°C
0 ¡C to 50 ¡C ambient temp., 40 W
convection cooled
0 ¡C to 50 ¡C ambient, 50 W
20 CFM Forced air
50 ¡C to 70 ¡C ambient Derate linearly
to 50% load
Peak (30 seconds) 60 W
Relative humidity Non-condensing 5% to 95% RH
Altitude Operating 10,000 feet max.
Non-operating 40,000 feet max.
Vibration (See Note 11) 5-500 Hz 0.75 G peak
OUTPUT SPECIFICATIONS
Voltage adjustability +5 V output on triples ±5.0%
Vout on singles ±5.0%
Line regulation LL to HL, FL main output ±0.2%
LL to HL, FL auxiliary output ±1.0%
Overshoot/undershoot At turn-on 0%
Transient response +5 V ±120 mV max dev.
(1.5 A to 3 A step) 500 μs recovery
Temperature coefficient All outputs ±0.02%/°C
Overvoltage protection + 5V output 6.25 ±0.75 Vout
Minimum output current (See Note 10) 0 A
Output power limit Primary power limited 90 W input
power limit
Short circuit protection Single Continuous
Multiple Short term
INPUT SPECIFICATIONS
Input voltage range 85-264 Vac
120-370 Vdc
Input frequency range 47-440 Hz
Input surge current Cold start 110 Vac, 60 Hz 10 A
Cold start 230 Vac, 50 Hz 22 A
Safety ground 110 Vac, 60 Hz 18 μA max.
leakage current 230 Vac, 50 Hz 28 μA max.