NETWORK ANALYZERS
214 www.anritsu.com For product ordering information, see pages 4-7.
Measurement
Frequency
Measurement items
Display
Display format
Resolution
Frequency accuracy
Range
Internal reference
oscillator
S parameter: S
11
, S
21
Level characteristics: R, TA, TB
Group delay characteristics: GPDLY (any aperture)
Time domain characteristics: displays impulse and step
response of above characteristics
S
11: LOG MAG, PHASE, LIN MAG, REAL IMAG, POLAR
(M/P), VSWR, IMPD (Z∠PHASE, Q/D, R
S/CS, LS,
R+jX), ADMT (Y∠PHASE, Q/D, R
P/CP, LP, G+jB)
S
21: LOG MAG, PHASE, LIN MAG, REAL IMAG, POLAR
(M/P), HSDLY
GPDLY: REAL
LEVEL: LOG MAG
Time domain (band pass, low pass, impulse/step
response): LOG MAG, PHASE, LIN MAG, REAL, IMAG
S parameter: S
11, S21, S12, S22
Level characteristics: R, TA, TB
Time domain characteristics: displays impulse and step
response of above characteristics
S
11/S22
: LOG MAG, PHASE, LIN MAG, REAL IMAG,
POLAR (M/P), VSWR, IMPD (Z∠PHASE, Q/D,
R
S/CS
, LS, R+ jX), ADMT (Y∠PHASE, Q/D,
R
P/CP
, LP, G+ jB)
S
21/S12
: LOG MAG, PHASE, LIN MAG, REAL IMAG,
POLAR (M/P), HSDLY
LEVEL: LOG MAG
Time domain (band pass, low pass, impulse/step
response): LOG MAG, PHASE, LIN MAG, REAL, IMAG
Model
• Filter pass/fail judgment using limit tests
The segment limit test permits pass/fail testing for complex characteristics with subtle fluctuations in slope not detected as ripples.
• Filter analysis using target data search
This function is useful in filter 3 dB bandwidth measurements and
ripple searches.
Specifications
MS4661A/E
1 to 2 screens (front/back, split)
100 kHz to 3 GHz
Minimum resolution: 0.1 Hz
Same as internal reference oscillator
Standard
Aging rate: ≤±1 x 10
–6
/day (compared to after 15 minutes warm-up)
Temperature characteristics: ≤±5 x 10
–6
(0˚ to 50˚C)
Option 01
Aging rate: ≤±2 x 10
–8
/day (compared to after 24 hours warm-up)
Temperature characteristics: ≤±5 x 10
–8
(0˚ to 50˚C)
MS4662A
Continued on next page
• SAW filter measurement with time gate function
Time domain gate function is very useful for eliminating test fixture
leakage signals.