VectorStar MS4640A Series
Microwave Vector Network Analyzers
(VNAs)
MS4642A, 10 MHz to 20 GHz, K Connectors
MS4644A, 10 MHz to 40 GHz, K Connectors
MS4645A, 10 MHz to 50 GHz, V Connectors
MS4647A, 10 MHz to 70 GHz, V Connectors
Anritsu Company
490 Jarvis Drive
Morgan Hill, CA 95037-2809
USA
http://www.anritsu.com
Part Number: 10410-00266
Copyright 2015 Anritsu Company
Revision: L
Printed: April 2015
Warranty
The Anritsu product(s) listed on the title page is (are) warranted against defects in materials and workmanship for
three years from the date of shipment.
Anritsu’s obligation covers repairing or replacing products which prove to be defective during the warranty period.
Buyers shall prepay transportation charges for equipment returned to Anr itsu for warranty repairs. Obligation is
limited to the original purchaser. Anritsu is not liable for consequential damages.
LIMITATION OF WARRANTY
The foregoing warranty does not apply to Anritsu connectors that have failed due to normal w ear. Als o, the war rant y
does not apply to defects resulting from improper or inadequate maintenance, unauthorized modification or misuse,
or operation outside of the environmental specifications of the product. No other warranty is expressed or implied,
and the remedies provided herein are the Buyer’s sole and exclusive remedies.
DISCLAIMER OF WARRANTY
DISCLAIMER OF WARRANTIES. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, ANRITSU
COMPANY AND ITS SUPPLIERS DISCLAIM ALL WARRANTIES, EITHER EXPRESSED OR IMPLIED,
INCLUDING, BUT NOT LIMITED TO, IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE, WITH REGARD TO THE PRODUCT. THE USER ASSUMES THE ENTIRE RISK OF
USING THE PRODUCT. ANY LIABILITY OF PROVIDER OR MANUFACTURER WILL BE LIMITED
EXCLUSIVELY TO PRODUCT REPLACEMENT.
NO LIABILITY FOR CONSEQUENTIAL DAMAGES. TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, IN NO EVENT SHALL ANRITSU COMPANY OR ITS SUPPLIERS BE LIABLE FOR ANY
SPECIAL, INCIDENTAL, INDIRECT, OR CONSEQUENTIAL DAMAGES WHATSOEVER (INCLUDING,
WITHOUT LIMITATION, DAMAGES FOR LOSS OF BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS
OF BUSINESS INFORMATION, OR ANY OTHER PECUNIARY LOSS) ARISING OUT OF THE USE OF OR
INABILITY TO USE THE PRODUCT, EVEN IF ANRITSU COMPANY HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. BECAUSE SOME STATES AND JURISDICTIONS DO NOT ALLOW THE
EXCLUSION OR LIMITATION OF LIABILITY FOR CONSEQUENTIAL OR INCIDENTAL DAMAGES, THE
ABOVE LIMITATION MAY NOT APPLY TO YOU.
Trademark Acknowledgements
Anritsu, VectorStar, K-Connector, V-Connector, and W1-Connector are trademarks of Anritsu Company, Morgan Hill,
California, USA.
Acrobat Reader is a registered trademark of Adobe Corporation of San Jose, California, USA.
Hampshire is a registered trademark and TSHARC a trademark of Microchip Technology, Chandler, Arizona, USA.
NI is a registered trademark of the National Instruments Corporation, Austin, Texas, USA.
Smith (Smith Chart) is a registered trademark of Analog Instruments Company, New Providence, New Jersey, USA.
Teflon is a registered trademark and brand name of the DuPont Company, for products made from three types of
fluorine-containing polymers (flurorpolymers).
Windows, Windows 7, Windows XP, and ToolTips are registered trademarks of Microsoft Corporation, Redmond,
Washington, USA.
All other trademarks mentioned are the property of their respective companies.
Notice
Anritsu Company has prepared this manual for use by Anritsu Company personnel and customers as a guide for the
proper installation, operation and maintenance of Anritsu Company equipment and computer programs. The
drawings, specifications, and information contained herein are the property of Anritsu Company, and any
unauthorized use or disclosure of these drawings, specifications, and information is prohibited; they shall not be
reproduced, copied, or used in whole or in part as the basis for manufacture or sale of the equipment or software
programs without the prior written consent of Anritsu Company.
UPDATES
Updates, if any, can be downloaded from the Documents area of the Anritsu Website at:
http://www.anritsu.com
For the latest service and sales contact information in your area, pleas e visit:
http://www.anritsu.com/contact.asp
MS4640A Series VNA OMPN: 10410-00266 Rev: LTitle-3
End-User License Agreement for Anritsu Software
IMPORTANT-READ CAREFULLY: This End-User License Agreement (“EULA”) is a legal agreement between you
(either an individual or a single entity) a nd Anritsu for the MS4640A Series VNA software product identifi ed above,
which includes computer software and associated media and printed materials, and may include “online” or electronic
documentation (“SOFTWARE PRODUCT” or “SOFTWARE”). By receiving or otherwise using the SOFTWARE
PRODUCT, you agree to be bound by the terms of this EULA.
Software Product License
The SOFTWARE PRODUCT is protected by copyright laws and international copyright treaties, as well as other
intellectual property laws and treaties. The SOFTWARE PRODUCT is licensed, not sold.
1. GRANT OF LICENSE. This EULA grants you the following rights:
a. You may use ONE copy of the Software Product identified above only on the hardware product (MS4640A Series
VNA and its internal computer) which it was originally installed. The SOFTWARE is in “use” on a computer when it
is loaded into temporary memory (for example, RAM) or installed into permanent memory (for example, hard disk,
CD-ROM, or other storage device) of that computer. However, installation on a network server for the sole purpose of
internal distribution to one or more other computer(s) shall not constitute “use.”
b. Solely with respect to electronic documents included with the SOFTWARE, you may make an unlimited number of
copies (either in hardcopy or electronic form), provided that such copies shall be used only for internal purposes and
are not republished or distributed to any third party.
2. OWNERSHIP. Except as expressly licensed to you in this Agreement, Anritsu retains all right, title, and interest in
and to the SOFTWARE PRODUCT; provided, however, that, subject to the license grant in Section 1.a and Anritsu's
ownership of the underlying SOFTWARE PRODUCT, you shall own all right, title and interest in and to any
Derivative Technology of the Product created by or for you.
3. COPYRIGHT. All title and copyrights in and to the SOFTWARE PRODUCT (including but not limited to any
images, photographs, animations, video, audio, music, text, and “applets” incorporated into the SOFTWARE
PRODUCT), the accompanying printed materials, and any copies of the SOFTWARE PRODUCT are owned by
Anritsu or its suppliers. The SOFTWARE PRODUCT is protected by copyright laws and international treaty
provisions. Therefore, you must treat the SOFTWARE PRODUCT like any other copyrighted material except that you
may make one copy of the SOFTWARE PRODUCT solely for backup or archival purposes. You may not copy any
printed materials accompanying the SOFTWARE PRODUCT.
4. DESCRIPTION OF OTHER RIGHTS AND LIMITATIONS.
a. Limitations on Reverse Engineering, Decompilation, and Disassembly. You may not reverse engineer, decompile, or
disassemble the SOFTWARE, except and only to the extent that such activity is expressly permitted by applicable law
notwithstanding this limitation.
b. Rental. You may not rent or lease the SOFTWARE PRODUCT.
c. Software Transfer. You may permanently transfer al l of your rights under th is EULA, provided that you retain n o
copies, you transfer all of the SOFTWARE PRODUCT (including the MS4640A Series VNA, al l component parts, the
media and printed materials, any upgrades, this EULA, and, if applicable, the Certificate of Authenticity), and the
recipient agrees to the terms of this EULA.
d. Termination. Without prejudice to any other rights, Anritsu may terminate this EULA if you fail to comply with the
terms and conditions of this EULA. In such event, you must destroy all copies of the SOFTWARE PRODUCT.
5. U.S. GOVERNMENT RESTRICTED RIGHTS. THE SOFTWARE PRODUCT AND DOCUMENTATION ARE
PROVIDED WITH RESTRICTED RIGHTS. USE, DUPLICATION, OR DISCLOSURE BY THE GOVERNMENT IS
SUBJECT TO RESTRICTIONS AS SET FORTH IN SUBPARAGRAPH (C)(1)(II) OF THE RIGHTS IN TECHNICAL
DATA AND COMPUTER SOFTWARE CLAUSE AT DFARS 252.227-7013 OR SUBPARAGRAPHS (C)(1) AND (2) OF
THE COMMERCIAL COMPUTER SOFTWARE-RESTRICTED RIGHTS AT 48 CFR 52.227-1 9, AS APPLICABLE.
MANUFACTURER IS ANRITSU COMPANY, 490 JARVIS DRIVE, MORGAN HILL, CALIFORNIA 95037-2809.
The MS4640A Series VNA software is copyright 2015, Anritsu Company. All rights are reserved by all parties.
Title-4PN: 10410-00266 Rev: LMS4640A Series VNA OM
MS4640A Series VNA OMPN: 10410-00266 Rev: LTitle-5
CE Conformity Marking
Anritsu affixes the CE Conformity marking onto its conforming products in accordance with Council Directives
of The Council Of The European Communities in order to indicat e that these products conform to the EMC and
LVD directive of the European Union (EU).
C-tick Conformity Marking
Anritsu affixes the C-tick marking onto its conforming products in accordance with the electromagnetic
compliance regulations of Australia and New Zealand in order to indicate that these products conform to the
EMC regulations of Australia and New Zealand.
Notes On Export Management
This product and its manuals may require an Export License or approval by the government of the product
country of origin for re-export from your country.
Before you export this product or any of its manuals, please contact Anritsu Company to confirm whether or
not these items are export-controlled.
When disposing of export-controlled items, the products and manuals need to be broken or shredded to such a
degree that they cannot be unlawfully used for military purposes.
Mercury Notification
This product uses an LCD backlight lamp that contains mercury. Disposal may be regulated due to
environmental considerations. Please contact your local authorities or, within the United States, the
Electronics Industries Alliance (www.eiae.org) for disposal or recycling information.
Perchlorate Notification
This product uses a small Lithium battery that may contain perchlorate installed internally on the circuit
board. Disposal may be regulated due to environmental considerations. Please contact your local authorities
for disposal or recycling information.
Title-6PN: 10410-00266 Rev: LMS4640A Series VNA OM
Chinese RoHS Statement
EPC WEEE Directive 2002/96/EC
MS4640A Series VNA OMPN: 10410-00266 Rev: LTitle-7
Title-8PN: 10410-00266 Rev: LMS4640A Series VNA OM
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Company uses the
following symbols to indicate safety-related information. For your own safety, please read the information
carefully before operating the equipment.
Symbols Used in Manuals
Danger
This indicates a very dangerous procedure that could result in serious
injury or death, and possible loss related to equipment malfunction, if not
performed properly.
Warning
Caution
This indicates a hazardous procedure that could result in light-to-severe
injury or loss related to equipment malfunction, if proper precautions are
not taken.
This indicates a hazardous procedure that could result in loss related to
equipment malfunction if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manuals
The following safety symbols are used inside or on the equipment near operation locations to provide
information about safety items and operation precautions. Ensure that you clearly understand the meanings of
the symbols and take the necessary precautions before operating the equipment. Some or all of the following
five symbols may or may not be used on all Anritsu equipment. In addition, there may be other labels attached
to products that are not shown in the diagrams in this manual.
This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near
the barred circle.
This indicates a compulsory safety precaution. The required operation is indicated symbolically in or
near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or near the triangle.
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
MS4640A Series VNA OMPN: 10410-00266 Rev: LSafety-1
For Safety
Warning
Warning
Warning
Always refer to the operation manual when working near locations at which
the alert mark, shown on the left, is attached. If the operation, etc., is
performed without heeding the advice in the operation manual, there is a
risk of personal injury. In addition, the equipment performance may be
reduced.
Moreover, this alert mark is sometimes used with other marks and
descriptions indicating other dangers.
When supplying power to this equipment, connect the accessory 3-pin
power cord to a 3-pin grounded power outlet. If power is supplied without
grounding the equipment, there is a risk of receiving a severe or fatal
electric shock.
This equipment can not be repaired by the operator. Do not attempt to
remove the equipment covers or to disassemble internal components.
Only qualified service technicians with a knowledge of electrical fire and
shock hazards should service this equipment. There are hig h-voltage p arts
in this equipment presenting a risk of severe injury or fatal electric shock to
untrained personnel. In addition, there is a risk of damage to precision
components.
Warning
Caution
Use two or more people to lift and move this equipment, or use an
equipment cart. There is a risk of back injury if this equipment is lifted by
one person.
Electrostatic Discharge (ESD) can damage the highly sensitive circuits in
the instrument. ESD is most likely to occur as test devices are being
connected to, or disconnected from, the instrument’s front and rear panel
ports and connectors. You can protect the instrument and test devices by
wearing a static-discharge wristband. Alternatively, you can ground
yourself to discharge any static charge by touching the outer chassis of the
grounded instrument before touching the instrume nt’s front and rear panel
ports and connectors. Avoid touching the test port center conductors
unless you are properly grounded and have eliminated the possibility of
static discharge.
Repair of damage that is found to be caused by electrostatic discharge is
not covered under warranty.
Safety-2PN: 10410-00266 Rev: LMS4640A Series VNA OM
Chapter Descriptions
Chapter 1 — Overview
This chapter provides an overview of the VectorStar MS4640A Series Vector Network Analyzer (VNA)
and a description of its major functions and available documentation. A summary of available precision
component kits including AutoCal calibration kits, mechanical calibration kits, and verification kits is
included.
Chapter 2 — Installation
This chapter provides information for the initial inspection and preparation for use of the MS4640A Series
VNA and includes information on instrument installation, required operating environment, power
requirements, initial inspection, and optional rack mounting procedures. After power up, the various
power modes are described with general warm-up and calibration time intervals. The preventive
maintenance section includes information on cleaning and fuse replacement along with preparation for
storage or shipment. General set up procedures a re prov ided for remo te programming control over GPIB
and Ethernet or USB only networks.
Chapter 3 — Front and Rear Panels
The chapter provides an overview of the MS4640A Series VNA hardware user interface in clu din g fro nt
panel buttons, front panel connectors, and back panel connectors. Included are photographs of the front
and rear panels. Each front panel key is described with its function and a cross reference to the menu it
activates on the user interface display. Each port and connector is described with its connector type, its
function, input/output limits, and a cross-reference to a detailed connector pin-out diagram.
Chapter 4 — User Interface Display
The chapter describes the general display options of the MS4640A Series VNA and provides a general
description of the MENU BAR, the ICON TOOLBAR, and the right-side function menus. General
descriptions and procedures are provided for channel setup, trace graph setup, marker setup, and limit
line setup. The MS4640A Series VNA displays measurement data using a “Channel Concept” where each
channel can display both a different S-Parameter (or another specialt y para meter) and a di fferent graph
type. As each channel is selected, the previously selected trace graph types, scaling, reference delay, SParameter, and other parameters associated with that channel appear on the screen. In addition, the
same S-Parameter can be displayed simultaneously on two or more channels using the same or different
trace graph types.
Chapter 5 — Main Menu
This user interface (UI) reference chapter summary operational information about the UI and detailed
overview of the primary controls on the top-of-screen Menu Bar and Icon Bar, and the right-side application
Main Menu. The Main Menu is available for VectorStar VNAs. The Menu Bar provides drop-down menu
interface to major instrument functions. The Icon Bar provides single-click access many instrument
functions and is user-configurable to meet individual needs. The description of the Main Menu (MAIN)
provides top level descriptions of each major function with links to detailed descriptions of menus and
dialog boxes in subsequent chapters.
Appendix A — Vector Network Analyzer Primer
This chapter describes the basic functions of a Vector Network Analyzer (VNA) and how it measures
magnitude and phase characteristics of networks, amplifiers, attenuators, and antennas. Scattering
parameters (S-parameters) are defined.
Appendix B — Initial Checkout Procedures
The initial checkout procedure chapter describes the detailed procedures for initial operational checkout
(such as those performed by incoming quality inspectors) to ensure that the MS4640A Series VNA is full y
operational. A quick check procedure is described along with how to determine the instrument’s installed
options, current operating system, application software, and firmware versions. Other procedures
provided are self-test, non-ratio power, high-level signal, and system noise floor test.
MS4640A Series VNA OMPN: 10410-00266 Rev: LContents-1
Chapter Descriptions (Continued}
Appendix C — Abbreviation Glossary
This glossary defines the abbreviations and terms that appear on the connectors, hard keys, menus, and
buttons of the MS4640A Series VNA. In some cases, due to space limitations, multiple abbreviations are
used for the same term or the same abbreviation is used with different punctuation.
Contents-2PN: 10410-00266 Rev: LMS4640A Series VNA OM
MS4640A Series VNA OMPN: 10410-00266 Rev: LContents-7
Table of Contents (Continued}
Contents-8PN: 10410-00266 Rev: LMS4640A Series VNA OM
Chapter 1 — Overview
1-1Introduction
This chapter provides an overview of the VectorStar MS4640A Series Vector Network Analyzer (VNA) and a
description of its major functions and available documentation. A summary of available precision component
kits including AutoCal calibration kits, mechani c al calibration kits, and verification kits is included.
MS4640A Series VNA OMPN: 10410-00266 Rev: L1-1
1-2 VectorStar MS4640A Series VNA DescriptionOverview
1-2VectorStar MS4640A Series VNA Description
The MS4640A Series VNA is an instrument system that contains a built-in source, test set, and analyzer. Test
results are displayed real time on a front panel touch screen or also to a separate video monitor. Screen
captures can easily be printed or saved in common graphic file formats.
Depending on the model and installed options, the MS4640A Series VNAs provide a maximum frequency range
from 70 kHz to 70 GHz for standalone VNAs. Through its front panel user interface, the instrument can be
quickly configured to provide either:
•25,000 Points
MS4640A Series VNAs have up to 25,000 total test points available, with up to 16 channels, and each
channel with up to 16 trace display graphs where each channel is configured as a virtual separate VNA.
Each trace can have up to 12 standard markers and one reference marker.
•100,000 Points
The VNA can be configured with up to 100,000 total test points, limited to one channel with up to 16
trace display graphs. Each trace can have up to 12 standard markers and one reference marker.
Figure 1-1.VectorStar MS4647A VNA Front Panel - Equipped with Option 051
1-2PN: 10410-00266 Rev: LMS4640A Series VNA OM
Overview1-3 VectorStar MS4640A Series VNA Models
1-3 VectorStar MS4640A Series VNA Models
The VectorStar VNA is available in four basic models as shown in Table 1-1.
Table 1-1.VectorStar MS4640A Series VNA Models
VNA Model
NumberNameSpecificationsTest Port Connectors
MS4642AMicrowave Vector Network Ana lyzer10 MHz to 20 GHzK (m) Connector Test Ports
MS4644AMicrowave Vector Network Ana lyzer10 MHz to 40 GHzK (m) Connector Test Ports
MS4645AMicrowave Vector Network Ana lyzer10 MHz to 50 GHzV (m) Connector Test Ports
MS4647AMicrowave Vector Network Ana lyzer10 MHz to 70 GHzV (m) Connector Test Ports
For additional technical specifications and configuratio n data, see:
•VectorStar MS4640A Series VNA Technical Data Sheet and Configuration Guide – 11410-00432
MS4640A Series VNA OMPN: 10410-00266 Rev: L1-3
1-4 VectorStar VNA OptionsOverview
1-4 VectorStar VNA Options
The available options depend on the VNA instrument model number as shown in Table 1-2 below.
Table 1-2.VectorStar MS4640A Series VNA Options (1 of 2)
Option
NumberOption Name and FeaturesVNA Model and Option Compatibility
Option 001 Rack Mount Option
• Adds 19” Rack Mounting Handles
Option 002 Time Domain Option
• Displays all S-parameters and
overlays with Frequency Domain,
Low-pass Mode with added
harmonics frequency list flexibility,
Band-pass Mode, Phasor Impulse
Mode, Windowing, Gating (passband or reject-band), and Frequency
with Time Gate.
Option 004 Additional Solid State Drive (SSD)
• Plug and play Serial ATA SSD
equipped with O/S and VectorStar
Application.
• Recommended for security
operations.
Option 007 Receiver Offset
• Provides Multiple Source, which
includes mixer, IMD, and other
measurements where the source and
receiver operate at different
frequencies.
• Independent Source/Receive
Functions: Allows independent
source and receive functions for
Mixer, Harmonics, IMD, and other
measurements, where the source
and receive frequencies are offset.
• Multiple Source Control Mode:
Provides independent control the
frequencies of up to four external
sources, in addition to the internal
source, and the receiver, in a
synchronized manner.
• NxN Frequency-Translated Devices:
Provides calibration and
measurements capability for NxN
Frequency-translated devices.
• For accurate and absolute magnitude
and phase measurements of match,
gain/loss, and group delay of devices
such as mixers and converters.
Option 041 Noise FigureCan be equipped on all MS4640A VNAs.
• CD with technical publications and manuals in PDF format and miscellaneous software utilities
• DVD with a back-up image of the instrument’s solid state drive
• USB optical mouse
• Power cord
1-6PN: 10410-00266 Rev: LMS4640A Series VNA OM
Overview1-5 VectorStar ME7838 Modular BB/mm-Wave VNA System
1-5VectorStar ME7838 Modular BB/mm-Wave VNA System
The modular broadband/millimeter-wave (BB/mm-Wave) system is available in two basic configurations.
Additional configuration and configuration information is available in the relevant technical data sheet.
Table 1-3.VectorStar ME7838 Series Modular Broadband/Millimeter-Wave VNA System (1 of 2)
Part NumberNameSpecifications
Standard ME7838 Configuration Components
MS4647AVectorStar MS4647A Vector Network Analyzer
(VNA)
MS4640A-007Receiver Offset Option
MS4640A-002Time Domain Option
MS4640A-07070 kHz Low End Frequency Extension
MS4640A-08XBroadband SystemSelect one:
3739ABroadband/Millimeter-Wave Test SetWith front and rear panel interface cables
3739CBroadband/Millimeter-Wave Test Set
(Required for use with MA25300A 145 GHz
Broadband modules)
3743AMillimeter-Wave Module70 GHz to 110+ GHz
MA25300AMillimeter-Wave Module54 GHz to 145 GHz
ME7838A-SS020On Site System Assembly and Verification
Phase Stable Cable Option – Select One (1)
10 MHz to 70 GHz
V (m) Connector Test Ports
• Use MS4647A-080 if VNA Option 051,
061, or 062 is NOT selected below.
• Use MS4647A-081 if VNA Option 051,
061, or 062 IS selected below.
With front and rear panel interface cables
• 2 each with mounting bracket
• 2 each with mounting bracket
806-2061.85 mm Phase St able Interconnect Cable70 cm (24 inches)
• V (m) to V (f)
•2 cables
806-2091.85 mm Phase St able Interconnect Cable91 cm (36 inches)
• V (m) to V (f)
•2 cables
MS4647A VNA Front Panel Options – Select None or One (1)
MS4647A-051Front Panel LoopsSix (6) Front Panel Loops
• Provides front panel loops for b 1, a1, Por t
1 Source, Port 2 Source, a2, and b2
• If Option 051 is selected, use Option
MS4647A-081 above.
Overview1-6 VectorStar ME7828A Standard BB/mm-Wave VNA System
1-6 VectorStar ME7828A Standard BB/mm-Wave VNA System
In its standard broadband/millimeter-wave configuration, the VectorStar MS46 40A Seri es VNAs are avail able
in two basic configurations as described below. Additional configurations are described in the relevant
technical data sheet.
Table 1-4.VectorS tar ME7828A Series BB/mm-Wave VNA Models and Options
Part Number NameSpecificationsConnectors
VectorStar ME7828A Series Broadband/Millimeter-Wave VNA Systems, consisting of:
MS4647AMicrowave Vector Network Analyzer10 MHz to 70 GHzV (m) Test Ports
MS4640A-001 VNA Rack Mount Option. Adds handles and removes feet for shelf-mounting into a 19”.
MS4640A-007 VNA Receiver Offset Option
MS464xA-05x VNA Direct Access Loop Option
3738ABroadband VNA Test SetWith Option 002
MG37022AFast Switching Signal Generator2 each; 2 to 20 GHzWith Option 001
3742A-EWMillimeter Module with Attenuator2 each
66670-3MUX Coupler , Left70 kHz internal bias tee
66671-3MUX Coupler, Right70 kHz internal bias tee
806-206Phase Stable Interconnect Cable70 cm (24 inches) V (m) to V (f)
806-207Phase Stable Interconnect Cable70 cm (24 inches) V (m) to V (m)
3700C3Floor Console
MS464xAAny MS4640A Series VNAMS4642A, 10 MHz to 20 GHz
MS4644A, 10 MHz to 40 GHz
MS4645A, 10 MHz to 50 GHz
MS4647A, 10 MHz to 70 GHz
3738ABroadband VNA Test SetWith Option 002
MG37022AFast Switching Signal Generator2 each; 2 to 20 GHzWith Option 001
3740A-X
or
3741A-X
3700C3Floor Console
Any pair of Millimeter Wave Modules
See table below.
2 required
K connectors
K connectors
V Connectors
V Connectors
MS4640A Series VNA OMPN: 10410-00266 Rev: L1-9
1-6 VectorStar ME7828A Standard BB/mm-Wave VNA SystemOverview
Millimeter-Wave Module Options
The following millimeter-wave modules are available for the ME7828A Millimeter-Wave VNA Measurement
System. If a millimeter-wave system is required, at least one pair of modules, matched for a band, are required.
Multiple module pairs can be equipped with any VNA system. The general installation and connection
requirements for each module pair are identical.
Table 1-5.Millimeter-Wave Module General Specifications and Part Numbers
Band / Module NameFrequencyAnritsu Part NumberMeasurement Type
3740A-V, 2 eachFull 2-port measurements.
V Band
WR-15 mmW Module
E Band
WR-12 mmW Module
Extended-E Band
WR-12 mmW Module
W Band
WR-10 mmW Module
Extended W Band
WR-10 mmW Module
50 to 75 GHz
60 to 90 GHz
56 to 94 GHz
75 to 110 GHz
65 to 110 GHz
3740A-VForward measurements only.
3741A-V
3740A-E, 2 eachFull 2-port measurements.
3740A-E Forward measurements only.
3740A-E
3740A-EE, 2 eachFull 2-port measurements.
3740A-EEForward measurements only.
3741A-EE
3740A-W, 2 eachFull 2-port measurements.
3741A-WForward measurements only.
3740A-W
3740A-EW, 2 eachFull 2-port measurements.
3740A-EWForward measurements only.
3741A-EWForward measurements only.
3742A-EWFull 2-port measurements.
Transmission/Reflection.
Transmission.
Transmission/Reflection.
Transmission.
Transmission/Reflection.
Transmission.
Transmission/Reflection.
Transmission.
Transmission/Reflection.
Transmission/Reflection.
Transmission.
Transmission/Reflection, with
Attenuator.
For additional information, consult the following documents:
•VectorStar ME7828A Broadband/Millimeter-Wave Technical Data Sheet – 11410-00452
Overview1-7 VectorStar MN4690B Series Multiport VNA System
1-7 VectorSt ar MN4690B Series Multiport VNA System
In its multiport (or 4-Port) configuration, the VectorStar VNA is available in K connector and V connec tor
configurations that support all VectorStar MS4640A Series VNAs. Additional configurations and options are
described in the relevant technical data sheet.
The upper frequency limit depends on the test set and the VNA. The MN4694B Test Set upper limit 40 GHz
and the MN4697B upper limit is 70 GHz. The system usable upper frequency limit is determined by the
attached VNA.
The lower frequency limit for the MN4690B Test Sets is 10 kHz, but as above, the usable lower frequency limit
is determined by the attached VNA and its installed options. For MS4640A Series VNAs equipped with Option
070 – 70 kHz Low End Frequency Extension, the lower frequency limit is specified as 70 kHz but allowed to go
to 40 kHz.
For VNAs not equipped with Option 070, the lower limit is 10 MHz.