For safety and warning information, please read this
•
g information is
provided within the MS2690A/MS2691A/MS2692A
(Mainframe
A Signal Analyzer Operation
A Signal
to this
•
and
MS2830A/MS2840A
/MS2850A
Signal Analyzer
Operation Manual
Noise Figure Measurement Function
Ninth Edition
manual before attempting to use the equipment.
Additional safety and warnin
Operation
Signal Analyzer Operation Manual
Operation), MS2830
Manual (Mainframe Operation), MS2840
Analyzer Operation Manual (Mainframe Operation),
or MS2850A Signal Analyzer Operation Manual
(Mainframe Operation). Please also refer
document before using the equipment.
Keep this manual with the equipment.
ANRITSU CORPORATION
Document No.: M-W3655AE-9.0
Page 2
ii
DANGER
WARNING
CAUTION
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the
following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of
the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
This indicates a very dangerous procedure that could result in serious injury or
death if not performed properly.
This indicates a hazardous procedure that could result in serious injury or death if
not performed properly.
This indicates a hazardous procedure or danger that could result in light-to-severe
injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and
take the necessary precautions BEFORE using the equipment.
This indicates an obligatory safety precaution. The obligatory operation is
This indicates a warning or caution. The contents are indicated symbolically in or
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
This indicates a prohibited operation. The prohibited operation is indicated
symbolically in or near the barred circle.
indicated symbolically in or near the circle.
near the triangle.
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer
Operation Manual Noise Figure Measurement Function Operation
23 October 2012 (First Edition)
20 February 2019 (Ninth Edition)
Anritsu Corporation guarantees that this equipment was inspected at
shipment and meets the published specifications.
Anritsu Warranty
During the warranty period, Anritsu Corporation will repair or exchange this
•
software free-of-charge if it proves defective when used as described in the
operation manual.
The warranty period is 6 months from the purchase date.
•
The warranty period after repair or exchange will remain 6 months from
•
the original purchase date, or 30 days from the date of repair or
exchange, depending on whichever is longer.
This warranty does not cover damage to this software caused by Acts of
•
God, natural disasters, and misuse or mishandling by the customer.
In addition, this warranty is valid only for the original equipment purchaser. It
is not transferable if the equipment is resold.
Anritsu Corporation shall assume no liability for injury or financial loss of the
customer due to the use of or a failure to be able to use this equipment.
Anritsu Corporation Contact
In the event that this equipment malfunctions, contact an Anritsu Service and
Sales office. Contact information is available in a separate file (for the PDF
version), and on the last page of this manual (for the printed version).
Page 4
iv
This product and its manuals may require an Export License/Approval by
the Governme
export from your
country.
Before re-exporting the product or manuals, please contact us to confirm
whet
When you dispose of export
ls need
to be broken/shredded so as not to be unlawfully used for military purpose.
Only files that have been provided directly from Anritsu or generated
All other required files should be transferred by means of USB or
Do not download or install software that has not been specifically
virus security protection in
place.
Notes On Export Management
nt of the product's country of origin for re-
her they are export-controlled items or not.
-controlled items, the products/manua
Cautions against computer virus infection
• Copying files and data
using Anritsu equipment should be copied to the instrument.
CompactFlash media after undergoing a thorough virus check.
• Adding software
recommended or licensed by Anritsu.
• Network connections
Ensure that the network has sufficient anti-
Page 5
v
Software End-User License Agreement (EULA)
Please read this Software End-User License Agreement (hereafter this EULA) carefully before using
(includes executing, copying, registering, etc.) this software (includes programs, databases, scenarios,
etc., used to operate, set, etc., Anritsu electronic equipment). By reading this EULA and using this
software, you are agreeing to be bound by the terms of its contents and Anritsu Corporation (hereafter
Anritsu) hereby grants you the right to use this Software with the Anritsu-specified equipment
(hereafter Equipment) for the purposes set out in this EULA.
1. Grant of License and Limitations
1. Regardless of whether this Software was
purchased from or provided free-of-charge by
Anritsu, you agree not to rent, lease, lend, or
otherwise distribute this Software to third
parties and further agree not to disassemble,
recompile, reverse engineer, modify, or create
derivative works of this Software.
2. You may make one copy of this Software for
backup purposes only.
3. You are not permitted to reverse engineer
this software.
4. This EULA allows you to install one copy of
this Software on one piece of Equipment.
2. Disclaimers
To the extent not prohibited by law, in no
event shall Anritsu be liable for personal
injury, or any incidental, special, indirect or
consequential damages whatsoever,
including, without limitation, damages for
loss of profits, loss of data, business
interruption or any other commercial
damages or losses, arising out of or related
to your use or inability to use this Software.
3. Limitation of Liability
a. If a fault (bug) is discovered in this Software,
preventing operation as described in the
operation manual or specifications whether
or not the customer uses this software as
described in the manual, Anritsu shall at its
own discretion, fix the bug, or exchange the
software, or suggest a workaround,
free-of-charge. However, notwithstanding
the above, the following items shall be
excluded from repair and warranty.
i) If this Software is deemed to be used for
purposes not described in the operation
manual or specifications.
ii) If this Software is used in conjunction with
other non-Anritsu-approved software.
iii) Recovery of lost or damaged data.
iv) If this Software or the Equipment has been
modified, repaired, or otherwise altered
without Anritsu's prior approval.
v) For any other reasons out of Anritsu's direct
control and responsibility, such as but not
limited to, natural disasters, software virus
infections, etc.
b. Expenses incurred for transport, hotel, daily
allowance, etc., for on-site repairs by Anritsu
engineers necessitated by the above faults
shall be borne by you.
c. The warranty period for faults listed in
article 3a above covered by this EULA shall
be either 6 months from the date of purchase
of this Software or 30 days after the date of
repair, whichever is longer.
Page 6
vi
4. Export Restrictions
Yo u m ay n ot use or otherwise export or
re-export directly or indirectly this Software
except as authorized by Japanese and
United States law. In particular, this
software may not be exported or re-exported
(a) into any Japanese or US embargoed
countries or (b) to anyone on the Japanese or
US Treasury Department's list of Specially
Designated Nationals or the US Department
of Commerce Denied Persons List or Entity
List. By using this Software, you warrant
that you are not located in any such country
or on any such list. You also agree that you
will not use this Software for any purposes
prohibited by Japanese and US law,
including, without limitation, the
development, design and manufacture or
production of missiles or nuclear, chemical
or biological weapons of mass destruction.
5. Termination
Anritsu shall deem this EULA terminated if
you violate any conditions described herein.
This EULA shall also be terminated if the
conditions herein cannot be continued for
any good reason, such as violation of
copyrights, patents, or other laws and
ordinances.
6. Reparations
If Anritsu suffers any loss, financial or
otherwise, due to your violation of the terms
of this EULA, Anritsu shall have the right to
seek proportional damages from you.
7. Responsibility after Termination
Upon termination of this EULA in
accordance with item 5, you shall cease all
use of this Software immediately and shall
as directed by Anritsu either destroy or
return this Software and any backup copies,
full or partial, to Anritsu.
8. Dispute Resolution
If matters of dispute or items not covered by
this EULA arise, they shall be resolved by
negotiations in good faith between you and
Anritsu.
9. Court of Jurisdiction
This EULA shall be interpreted in
accordance with Japanese law and any
disputes that cannot be resolved by
negotiation described in Article 8 shall be
settled by the Japanese courts.
Page 7
About This Manual
(
)
Associated Documents
The operation manual configuration of the MS2690A/MS2691A/MS2692A,
MS2830A, MS2840A and MS2850A Signal Analyzer is shown below.
MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual
Main Frame Operation
MS2830A Signal Analyzer Operation Manual (Main Frame Operation)
Or
MS2840A Signal Analyzer Operation Manual (Main Frame Operation)
MS2850A Signal Analyzer Operation Manual (Main Frame Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual (Main Frame Remote Control)
MS2690A/MS2691A/MS2692A
Signal Analyzer Operation Manual
(Signal Analyzer Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual (Signal Analyzer Function Remote Control)
MS2690A/MS2691A/MS2692A
Spectrum Analy zer Operation Manual
(Spectrum An alyzer Function Operation)
Or
Or
MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual
(Signal Analyz er Function O peration)
MS2830A/MS2840A/MS2850A
Spectru m Analyzer Operation Manual
(Spectrum Analyzer Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual (Spectrum Analyzer Function Remote Control)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual (Noise Figure Measurement Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A
Signal Analyzer Operation Manual (Noise Figure Measurement Function Remote Control)
I
Page 8
• Signal Analyzer Oper ation Ma nual (Mainf rame)
• Signal Analyzer Operation Ma nual (Mainfra me Remote Control )
Description of basic operations, maintenance procedures, common
functions and common remote functions of the mainframe
• Signal Analyzer Operation Manual (Signal Analyzer Function)
• Signal Analyzer Ope ration Manual (S ignal Analyzer Function Remote Control)
Description of basic operations, common functions and common remote
functions of the signal analyzer
• Signal Analyzer Operation Manual (Spectrum Analyzer Function)
• Signal Analyzer Oper ation Ma nual (Spectrum Analyzer Function Remote
Control)
Description of basic operations, common functions and common remote
functions of the spectrum analyzer
• Signal Analyzer Oper ation Ma nual (Noise Figure Measurement Function)
<This document>
• Signal Analyzer Oper ation Ma nual (Noise Figure Measurement Function
Remote Control)
Description of basic operations, common functions and common remote
functions of the Phase Noise Measurement function
Convention Used in This Manual
Throughout this document, the use of MS269x Series is assumed unless
otherwise specified.
If using MS2830A, MS2840A or MS2850A, change MS269xA to read
MS2830A, MS2840A or MS2850A.
In this document,
indicates a panel key.
II
Page 9
Table of Contents
1
About This Manual................................................. I
The MS2690/MS2691/MS2692A or MS2830A/MS2840A/MS2850A Signal
Analyzer enables high-speed, high-accuracy, and simple measurements of
transmission characteristics of base stations and mobile stations for
various types of mobile communications.
The Noise Figure Measurement Function (hereinafter, this application) is
to measure Noise Figure (hereinafter, NF) which has a huge effect on
noise in systems that handle minute signals such as satellite
communications and radars. It can be used in various fields such as
evaluations of transceivers and devices and troubleshooting.
NF is measured with the measurement method of Y-factor method which
uses a Noise Source.
The NC346 Series noise sources by Noisecom company are supported.
1-2
Page 13
1.2 Product Composition
1.2.1 Option / Software
Table 1.2.1-1 list the Options and Software’s.
Table 1.2.1-1 Configuration of Application
Model/Symbol Product name Remarks
1.2 Product Composition
1
Overview
MS2690A-017
MS2691A-017
MS2692A-017
MS2690A-117
MS2691A-117
MS2692A-117
MS2830A-017 Noise Figure Measurement Function
MS2830A-117 Noise Figure Measurement Function
MS2840A-017 Noise Figure Measurement Function
MS2840A-117 Noise Figure Measurement Function
MS2850A-017 Noise Figure Measurement Function
MS2850A-117 Noise Figure Measurement Function
1.2.2 Noise Source
Noise Figure Measurement Function
Noise Figure Measurement Function
Retrofit
Retrofit
Retrofit
Retrofit
The MS269xA-017/117 and MS269xA-030 W-CDMA RNC simulator are
exclusive options.
The NC346 Series noise sources by Noisecom company are supported. The
models and specifications outlines of NC346 Series are as follows. For
detailed specifications, refer to the catalog and data sheet for the NC346
Series.
Table 1.2.2-1 Specifications of NC346 Series
Model DC Offset Frequency [GHz] Output ENR [dB]
NC346A – 0.01 to 18.0 5 to 7
NC346B – 0.01 to 18.0 14 to 16
NC346D – 0.01 to 18.0 19 to 25
NC346C
NC346E
NC346Ka
0.01 to 26.5 13 to 17
0.01 to 26.5 19 to 25
0.1 to 40.0 10 to 17
1-3
Page 14
Chapter 1 Overview
CAUTION
Some noise sources output DC power. Therefore, when
selecting a model from the following models of the RF
input specification “0V_DC_max”, install a DC block:
• MS2690A-017/-117
• MS2691A-017/-117
• MS2692A-017/-117
• MS2830A-044, -017/-117
• MS2830A-045, -017/-117
• MS2840A-044, -017/-117
• MS2840A-046, -017/-117
• MS2850A-047, -017/-117
• MS2850A-046, -017/-117
1-4
Page 15
1.3 Specifications
Table 1.3-1 show the specifications.
1.3 Specifications
1
Table 1.3-1 Specifications for this Application
Item Specification
Frequency Frequency
range
MS2850A-047: 30 MHz to 32 GHz
Frequency
Setting range
MS2850A-047: 10 MHz to 32 GHz
MS2690A: 30 MHz to 6 GHz
MS2691A: 30 MHz to 6 GHz
MS2692A: 30 MHz to 6 GHz
MS2830A-040: 30 MHz to 3.6 GHz
MS2830A-041: 30 MHz to 6 GHz
MS2830A-043: 30 MHz to 13.5 GHz
MS2830A-044: 30 MHz to 26.5 GHz
MS2830A-045: 30 MHz to 40 GHz
MS2840A-040: 30 MHz to 3.6 GHz
MS2840A-041: 30 MHz to 6 GHz
MS2840A-044: 30 MHz to 26.5 GHz
MS2840A-046: 30 MHz to 40 GHz
MS2850A-046: 30 MHz to 40 GHz
MS2690A: 10 MHz to 6 GHz
MS2691A: 10 MHz to 13.5 GHz
MS2692A: 10 MHz to 26.5 GHz
MS2830A-040: 10 MHz to 3.6 GHz
MS2830A-041: 10 MHz to 6 GHz
MS2830A-043: 10 MHz to 13.5 GHz
MS2830A-044: 10 MHz to 26.5 GHz
MS2830A-045: 10 MHz to 43 GHz
MS2840A-040: 10 MHz to 3.6 GHz
MS2840A-041: 10 MHz to 6 GHz
MS2840A-044: 10 MHz to 26.5 GHz
MS2840A-046: 10 MHz to 44.5 GHz
MS2850A-046: 10 MHz to 44.5 GHz
Overview
1-5
Page 16
Chapter 1 Overview
Item Specification
Table 1.3-1 Specifications for this Application (Cont'd)
This section describes the panel keys for operating the instrument and
connectors used to connect external devices. For general points of caution,
refer to the
This section describes the front-panel keys and connectors.
9
8
7
MS2690A/MS2691A/MS2692A Signal Analyzer Operation
MS2830A Signal Analyzer Operation
or
MS2850A Signal Analyzer Operation
.
10
14
12
13
6
5
4
3
2
1
19
17
Figure 2.1.1-1 MS269x series front panel
18
16
15
11
2-2
Page 19
2.1 Part Names
9
8
7
6
5
4
17
20
18
22
3 2 1
Figure 2.1.1-2 MS2830A/MS2840A/MS2850A front panel (MS2830A Example)
Power Switch
1
Press to switch between the standby state (AC power supplied) and
power-on state. The Power lamp
Power On. Press the power switch for about 2 seconds.
2
Hard disk access lamp (MS269x series, MS2830A)
Lights up when accessing the internal hard disk
102112 14
13
2
Preparation
15
11
19 16
lights orange at Standby and green at
SSD
SSD access lamp (MS2840A, MS2850A)
Lights up when accessing the internal SSD
3
4
5
6
Copy key
Press to capture display screen and save to file.
Recall key
Press to recall parameter file.
Save key
Press to save parameter file.
Cal key
Press to display the Calibration menu.
2-3
Page 20
Chapter 2 Preparation
7
8
9
10
Local key
Press to return to local operation from remote control via GPIB, Ethernet,
or USB (B), and enable panel settings.
Remote lamp
Lights when in remote-control state.
Preset key
Resets parameters to initial settings.
Function keys
Selects or configures function menu displayed on the right of the screen.
The function menu is provided in multiple pages and layers.
Press
is displayed at the bottom of the function menu, as in “1 of 2”.
Sub-menus may be displayed when a function menu is pressed. Press
to go back to the previous menu. Press
to fetch next function menu page. The current page number
to go back to the top menu.
2-4
Page 21
2.1 Part Names
11
12
Main function keys 1
Press to set or execute main functions.
Executable functions vary with the current application. When nothing
happens with the press, it indicates that the application in use does not
support the key.
Press to set frequency parameters.
Press to set level parameters.
Press to set span parameters.
No function is assigned to this key.
Press to set BW parameters.
Press to set measurement item parameters.
Main function keys 2
Press to set or execute main functions.
Executable functions vary with the current application. When nothing
happens with the press, it indicates that the application in use does not
support the key.
2
Preparation
Press to switch application.
Press to display Configuration screen.
Press to set the trace items or to switch the operation window.
Press to set measurement item parameters.
Use when switching graph marker operation.
Press to set parameters related to the peak search function.
Press to start single measurement.
Press to start continuous measurements.
2-5
Page 22
Chapter 2 Preparation
13
14
15
Rotary knob/Cursor key/Enter key/Cancel key
The rotary knob and cursor keys select display items or change settings.
Press
Press
Shift key
Operates keys with functions in blue characters on panel. Press the Shift
key so the key lamp is green and then press the target key.
Numeric keypad
Enters numbers on parameter setup screens.
to set the entered or selected data.
to cancel input or selected data.
16
17
Press
[A] to [F] can be entered by pressing keys
key lamp
RF Input connector
Inputs RF signal. This is an N type input connector.
This is a K type input connector when MS2830A-045, MS2840A-046 is
installed or MS2850A.
RF Output Control key
(When MS269xA-020/120, MS2830A-020/120/021/121,
MS2840A-020/120/021/121 installed)
When the Vector Signal Generator option is installed, the RF output
control key lamp lights orange in the RF signal output On state.
This is not available when the MS2830A-044/045, MS2840A-044/046 is
installed or MS2850A.
Outputs RF signal, when the Vector Signal Generator option is installed.
This is an N type output connector.
This is not installed when the MS2830A-044/045, MS2840A-044/046 is
installed or MS2850A.
USB connector (type A)
Connect the accessory USB keyboard, mouse or USB memory.
Modulation control key (When MS2830A-020/120/021/121,
MS2840A-020/120/021/121 installed)
When the Vector Signal Generator option is installed, the lamp on the
key lights up in green in the modulation On state.
This is not installed when the MS2830A-044/045, MS2840A-044/046 is
installed or MS2850A.
Application key (MS2830A, MS2840A, MS2850A)
Press to switch between applications.
Press to display the Spectrum Analyzer main screen.
2
Preparation
Press to display the Signal Analyzer main screen, when
MS2830A-005/105/007/006/106/009/109/077/078 or
MS2840A-005/105/006/106/009/109/077/177/078/178 is installed
or MS2850A.
Press to display the Signal Analyzer main screen, when Vector
Signal Generator option is installed. (MS2830A, MS2840A)
This is a blank key. Not used. (MS2830A, MS2840A)
Displays the main screen of the application that is selected
using the Application Switch (Auto), or displays that of the
pre-selected application (Manual).
For details, refer to 3.5.4 Changing application layout in
MS2830A Signal Analyzer Operation Manual (Mainframe
1st Local Output connector (MS2830A, MS2840A), (MS2850A: Future Extensions)
22
This is available when the MS2830A-044/045, MS2840A-044/046 is
installed.
Supplies local signal and bias current to the external mixer, and receives
the IF signal with its frequency converted.
Ref Input connector (reference frequency signal input connector)
Inputs external reference frequency signal. It is for inputting reference
frequency signals with higher accuracy than the instrument’s internal
reference signal, or for synchronizing the frequency of the
MS2690A/MS2691A/MS2692A or MS2830A/MS2840A to that of other
equipment. The following frequencies are supported:
Buffer Out connector (reference frequency signal output conn ector)
Outputs the internal reference frequency signal (10 MHz). It is for
synchronizing frequencies between other equipment and the
MS2690A/MS2691A/MS2692A or MS2830A/MS2840A.
Trigger Input connector (MS269x series only)
Inputs trigger signal from external device.
Sweep Status Out connector
Outputs signal when internal measurement is performed or
measurement data is obtained.
2-10
Page 27
2.1 Part Names
5
6
7
8
9
10
11
12
GP-IB
Monitor Out
IF Out connector (MS269x series only)
Not used.
AUX connector
Not used.
GPIB connector
Used to remote-control this unit via GPIB from external controller. For
external control via GPIB.
USB connector (type B)
Used to remote-control this unit via USB from external controller. For
external control via USB.
Ethernet connector
Connects PC or Ethernet network.
USB connector (type A)
Used to connect a USB keyboard or mouse or the USB memory supplied.
Monitor Out connector
Connects external display.
AC inlet
Supplies power.
2
Preparation
2-11
Page 28
Chapter 2 Preparation
13
14
15 HDD or
Primary HDD/SSD
16 HDD(Opt) or
Secondary HDD/SSD
17
AF Input connector (unbalanced)
18
AF Input connector (balanced)
19
20
SA Trigger Input connector (MS2830A, MS2840A, MS2850A)
This is a BNC connector for inputting external trigger signal (TTL) for
SPA and SA applications.
SG Trigger Input connector (MS2830A, MS2840A)
This is a BNC connector for inputting external trigger signal (TTL) for
Vector Signal Generator option.
HDD slot (MS2830A)
SSD slot (MS2840A, MS2850A)
HDD slot (MS2830A)
SSD slot (MS2840A, MS2850A)
IF output connector (MS2830A, MS2840A, MS2850A)
Monitor output of the internal IF signal.
This is installed when the MS2830A-044/045, MS2840A-044/046 is
installed or MS2850A.
BNC connector to input unbalanced AF signal from external device.
This is available when the MS2830A-018/118 is installed.
Standard phone jack (3 poles, 6.3 mm) connector to input balanced AF
signal from external device.
This is available when the MS2830A-018/118 is installed.
Noise Source connector
Supply (+28V) of the Noise Source Drive.
This is available when the Option 017/117 is installed.
This is a standard HDD slot.
This is a standard SSD slot.
This is a HDD slot for the options.
This is a SSD slot for the options.
2-12
Page 29
General Input/Output (Audio Function) connector
21
D-Sub 15 connector for audio function input/output from/to the external
device.
This is available when the MS2830A-018/118 is installed.
PTT Control connector
22
Banana plug jack connector to control PTT (Push to Talk).
This is available when the MS2830A-018/118 is installed.
AF Output connector (unbalanced)
23
BNC connector to output unbalanced AF signal to external device.
This is available when the MS2830A-018/118 is installed.
AF Output connector (balanced)
24
Standard phone jack (3 poles, 6.3 mm) connector to output balanced AF
signal to external device.
This is available when the MS2830A-018/118 is installed.
Demodulation Output connector
25
BNC connector to output demodulated AF signal to external device.
This is available when the MS2830A-018/118 is installed.
Headphone Output connector
26
Mini jack connector to output demodulated AF audio signal to external
device.
This is available when the MS2830A-018/118 is installed.
CAL Port connector (Future extensions) (MS2850A)
27
2.1 Part Names
2
Preparation
2-13
Page 30
Chapter 2 Preparation
Trigger Input 2 connector (MS2850A)
28
Input the trigger signal (3.3 V LVCMOS) for SPA and SA applications.
Trigger Output 1 connector (MS285 0A)
29
Output the trigger signal (3.3 V LVCMOS).
Trigger Output 2 connector (MS2850A)
30
Output the trigger signal (3.3 V LVCMOS).
USB 3.0 connector (Future extensions) (MS2850A)
31
PCIe X8 connector (Future extensions) (MS2850A)
32
2-14
Page 31
2.2 Signal Path Setup
As shown in Figure 2.2-1, connect the instrument and the DUT using an
RF cable, so that the signal to be tested is input to the RF Input connector.
To prevent an excessive level signal from being input, do not input the
signal before setting the input level using this application.
2.2 Signal Path Setup
2
Preparation
Noise Source
RF Input
DUT
Figure 2.2-1 Signal path setup example
Connect the Noise Source to the Noise Source connector.
Set the reference signal paths from external sources, as required.
Noise Source
Reference frequency signal
Figure 2.2-2 External signal input
2-15
Page 32
Chapter 2 Preparation
2.3 Application Startup and Selection
To use this application, it is necessary to load (start up) and select the
application.
2.3.1 Launching application
The application startup procedure is described below.
Note:
The XXX indicates the application name currently in use.
<Procedure>
1. Press
to display the Configuration screen.
2. Press
Switch Registration screen.
3. Press
in the Unloaded Applications list.
If “XXX” is displayed in the Loaded Applications list, this means
that the application is already loaded.
If “XXX” appears in neither the Loaded Applications nor Unloaded
Applications list, this means that the application has not been
installed.
4. Press
Loaded Applications list, this means that the application is already
loaded.
2.3.2 Selecting application
The selection procedure is described below.
<Procedure>
1. Press
2. Press the menu function key displaying “XXX”.
(Application Switch Settings) to display the Application
(Load Application Select), and move the cursor to “XXX”
(Set) to load the application. If “XXX” is displayed in the
to display the Application Switch menu.
2-16
The application can also be selected with mouse, by clicking “XXX” on the
task bar.
Page 33
2.4 Initialization and Calibration
This section describes the parameter settings and the preparations
required before starting measurement.
2.4 Initialization and Calibration
2.4.1 Initialization
After selecting this application, first perform initialization. Initialization
returns the settable parameters to their default value in order to clear the
measurement status and measurement results.
Note:
The initialization procedure is as follows.
<Procedure>
1. Press
2. Press
2.4.2 Calibration
Perform calibration before performing measurement. Calibration sets the
level accuracy frequency characteristics for the input level to flat, and
adjusts level accuracy deviation caused by internal temperature
fluctuations. Calibration should be performed when first performing
measurement after turning on power, or if beginning measurement when
there is a difference in ambient temperature from the last time
calibration was performed.
2
Preparation
When another software application is switched to or this
application is unloaded (ended), the application keeps the
parameter settings at that time. The parameter values that were
last set will be applied when this application is selected next time.
to display the Preset function menu.
(Preset).
<Procedure>
1. Press
2. Press
For details on calibration functionality only executable with this
instrument, refer to the
This section describes the screen layout of the Noise Figure measurement
function.
[1]
[2]
[3]
[5]
[4]
Figure 3.1.1-1 Screen layout
[1] Measurement parameter
Displays the specified parameter. For details about measurement
results, refer to Section 3.8 “Measurement Results.”
[2] Measurement Status Area
Displays alarms which indicate measurement errors. For details,
refer to Section 3.8 “Measurement Results.”
[3] Measurement Result Window
Displays the measurement results. One of List display, Graph display,
and Spot display is displayed depending on the setting of Frequency
Mode and Layout. For details, refer to Section 3.8 “Measurement
Results.”
[4] Status Area
Displays the Measurement Status.
3-2
[5] Function menu
Displays the functions executable with function keys.
Page 37
3.1.2 Main function menu
This section describes the main function menu on the main screen.
F1
F2
F3
F4
F5
F6
3.1 Basic Operation
3
Measurement
Function
Key
F1
F2
F3
F4 Measure
F5 Marker
F6
F8 Accessary
Menu Display Function
Frequency
Amplitude
Common Setting
Peak Search
F7
F8
Figure 3.1.2-1 Main Function Menu
Table 3.1.2-1 Main Function Menu
Sets the frequency.
Sets the level.
Sets the common items.
3.4 Setting Common Items
Opens the Measure function.
3.5 Setting the Measure Function
Opens the Marker function.
3.6 Setting the Marker Function
Opens the Peak Search function.
3.7 Setting the Peak Search Function
Performs settings for other functions.
Chapter 4 Other Functions
3.2 Setting Frequency
3.3 Setting Level
3-3
Page 38
Chapter 3 Measurement
3.1.3 Performing measurement
There are two measurement modes: single and continuous. Measurement
is performed once in the single measurement mode, and continuously in
the continuous measurement mode.
Single
Items are measured only for the measurement count (Storage Count)
before measurement is stopped.
<Procedure>
Press
Continuous
Measurement will continue even if parameters are changed or the
window display is changed. Measurement will be stopped if another
application is selected, or DUT Mode is changed.
.
3.1.4
<Procedure>
Press
.
Restrictions on File Input/Output
This application can save the measurement results and the loss
compensation tables, etc. in a file or load them from a file.
The files are input/output in csv format, and saved in different folders by
item. One folder can store up to 100 files.
For saving destination and format by item, refer to the description of each
item.
3-4
Page 39
3.2 Setting Frequency
This section describes the frequency-related settings. Press
(Frequency) on the main function menu to display the Frequency function
menu.
Table 3.2-1 Frequency Function Menu
3.2 Setting Frequency
Function
Key
F1 Frequency Mode
F2 Fixed Setting
F3 List Setting
F4 Sweep Setting
Menu Display Function
Note:
When DUT Mode is not Amplifier, the setting frequency changes
according to LO Mode as in Table 3.2-2.
Table 3.2-2 Setting Frequency when DUT is other than Amplifier
LO Mode Setting Frequency
Fixed IF (Output from DUT)Frequency
Variable RF (Input to DUT)Frequency
Opens the Frequency Mode Function Menu.
3.2.1 Frequency Mode
Opens the Fixed Setting Function Menu.
3.2.2 Fixed Setting
Opens the List Setting Function Menu.
3.2.3 List Setting
Opens the Sweep Setting Function Menu.
3.2.4 Sweep Setting
3
Measurement
3-5
Page 40
Chapter 3 Measurement
3.2.1 Frequency Mode
Press
to display the Frequency Mode Function menu.
After each mode is set, the setting screen is displayed.
Table 3.2.1-1 Frequency Mode Function Menu
(Frequency Mode) on the Frequency function menu or
Function
Key
—
F1 Fixed
F2 List
F3 Sweep
3.2.2 Fixed Setting
Function
Key
F1 Position
F7
F8
Menu
Display
Frequency Pressing [Frequency] Displays this function.
Sets the Frequency Mode to Fixed.
Setting Fixed opens the Fixed Setting Function menu.
Sets the Frequency Mode to List.
Setting Fixed opens the List Setting Function menu.
Sets the Frequency Mode to Sweep.
Setting Fixed opens the Sweep Setting Function menu.
This section describes the setup to be performed when the Frequency
Mode is set to Fixed.
Table 3.2.2-1 Fixed Setting Function Menu
Menu
Display
Same as the Frequency List Edit function menu
“Position”.
Storage
Mode
Storage
Count
Same as the Storage function menu “Mode”.
Same as the Storage function menu “Count”.
Function
3.2.2 Fixed Setting
3.2.3 List Setting
3.2.4 Sweep Setting
Function
3.2.3.1 Frequency List Edit
3.5.2 Storage
3.5.2 Storage
3-6
Page 41
3.2.3 List Setting
3.2 Setting Frequency
This section describes the setup to be performed when the Frequency
Mode is set to List.
Table 3.2.3-1 List Setting Function Menu
Function
Key
F1 Device Specifies the Table file storage destination drive.
F2
F3
F6 Edit
Menu
Display
Save
Frequency
List
Recall
Frequency
List
The file format for Frequency List Table is as follows:
Press (Edit) on the List Setting function menu to display the
Frequency List Edit Function menu.
This section describes the Frequency List Edit Function Menu that is
used to edit a Frequency List Table.
[1][2][3]
[4]
Figure 3.2.3.1-1 Frequency List Table
[1] Frequency: Setting Frequency
[2] Noise Figure: Displays Result Type set to Trace1.
[3] Gain: Displays Result Type set to Trace2.
[4] Cursor: Selects a line in the list, and can be moved by
operating the cursor keys .
3-8
Page 43
3.2 Setting Frequency
Table 3.2.3.1-1 Frequency List Edit Function Menu
Function
Key
F1 Position
F2 Frequency
F5
F6 Delete PointDeletes Point of the selected Position.
F7 Sort Table Sorts Table in ascending order of frequency.
F8 Clear Table
Menu
Display
Selects Position from List on the Table screen. Various
settings can be performed for items of the selected
Position (number).
Setting Range: 1 to (Total Point +1)
Default: 1
Maximum: 501
Sets the frequency for the selected Position
Default: Refer to Table 3.2.3.1-2
Unit: GHz/MHz/kHz/Hz
Resolution: 1 Hz
Insert Point
Below
Table 3.2.3.1-2 Setting Range of Frequency
Adds Point next to the selected Position. The added
Point reflects the current Position value as it is.
A confirmation message is displayed before clearing and
pressing Enter executes the clearing.
Function
3
Measurement
DUT Mode LO Mode Setting Range
Amplifier — Max.: The maximum frequency of the main unit or of
Other than
Amplifier
1
Fixed*
Variable*2 Max.: The maximum frequency of Noise Source*3
*1: When LO Mode is Fixed, set as IF frequency (output from DUT).
*2: When LO Mode is Variable, set as RF frequency (input from DUT).
*3: For frequency range, refer to Section 3.4.5.1 “Noise Source Select”.
Note:
Min.: The minimum frequency of the main unit or of
Min.: The minimum frequency of Noise Source*
When DUT Mode is Down Converter or Up Converter, the
frequency input range is limited according to the fixed frequency
(Local Freq or IF Freq) and Sideband Mode setting.
Noise Source*
Noise Source*
3
, whichever smaller.
3
, whichever bigger.
3
3-9
Page 44
Chapter 3 Measurement
3.2.4 Sweep Setting
This section describes the setup to be performed when the Frequency
Mode is set to Sweep.
Note:
Frequency List Table is edited by setting in Sweep Mode.
Table 3.2.4-1 Sweep Mode Function Menu
Function
Key
F1 Center
F2 Start
F3 Stop
F4 Span
F5
F8
Menu
Display
Full
Span
Sweep
Point
Function
Sets the center frequency of the frequency range for sweep.
Setting Range: Refer to Table 3.2.3.1-2
Unit: GHz/MHz/kHz/Hz
Resolution: 1 Hz
Default: 1.805 GHz
Sets the Start frequency of the frequency range for sweep.
Setting Range: Refer to Table 3.2.3.1-2
Unit: GHz/MHz/kHz/Hz
Resolution: 2 Hz
Default: 10 MHz
Sets the Stop frequency of the frequency range for sweep.
Setting Range: Refer to Table 3.2.3.1-2
Unit: GHz/MHz/kHz/Hz
Resolution: 2 Hz
Default: 3.6 GHz
Sets the frequency span range for sweep.
Setting Range: Maximum Stop Frequency
Sets the number of points for sweep. The same item with
Total Point.
Setting range: 3 to 501*
Default: 11
*: When the span is 500 Hz or less, the Sweep Point
maximum value is span+1
3-10
Page 45
3.3 Setting Level
3.3 Setting Level
This section describes the level- related settings. Press
on the main function menu or
menu.
Table 3.3-1 Amplitude Function Menu
Function
Key
— Amplitude Pressing [Amplitude] displays this function.
F1 Attenuator
F2 Pre-Amp
F5 Trace Select
F7
F8 Scale/Div
Menu Display Function
Sets the Input Attenuator.
Sets Pre-Amp On/Off.
Same as the Trace function menu “Trace Select”.
Same as the Trace function menu “Reference
Reference
Value”.
Same as the Trace function menu “Scale/Div
Value”.
to display the Amplitude function
(Amplitude)
3.3.1 Attenuator
3.3.2 Pre-Amp
3.5.1 Trace
3.5.1 Trace
3.5.1 Trace
3
Measurement
3-11
Page 46
Chapter 3 Measurement
3.3.1 Attenuator
This configures the input attenuator settings. The setting range and
resolution are as follows:
Table 3.3.1-1 Setting Parameter of Input Attenuator
Item Setting Parameter
Maximum 60 dB
Minimum 0 dB
Unit dB
Resolution MS2830A-045 Installed: 10 dB
MS2840A-046 Installed
without MS2840A-019/119: 10 dB
Others: 2 dB
MS2850A 2 dB
Default 0 dB
3.3.2 Pre-Amp
This configures the Pre-Amp On/Off.
(The function is available when MS269xA-008/108,
MS2830A-008/108/068/168, MS2840A-008/108/068/168/069/169, or
MS2850A-068/168 is installed. This function is fixed to Off if not
installed.)
■Options
On Enables the Pre-Amp function. (Default)
Off Disables the Pre-Amp function.
3-12
Page 47
3.4 Setting Common Items
This section describes the settings for the common items. Press
(Common Setting) on the main function menu to display the Common
Setting function menu.
Table 3.4-1 Common Setting Function Menu
3.4 Setting Common Items
Function
Key
F1 DUT Mode
F2 Convert Setup
F3
F4 Loss Comp
F5 ENR
F8 Cal Setup
Menu Display Function
Opens the DUT Mode Function menu.
Opens the Convert Setup Function Menu.
Enable when DUT Mode is not Amplifier.
External LO
Setup
Opens the External LO Setup Function Menu.
Enable when DUT Mode is not Amplifier.
Opens the Loss Comp Menu.
Opens the ENR Function Menu.
Opens the Cal Setup Function Menu.
3.4.1 DUT Mode
3.4.2 Convert Setup
3.4.3 External Lo Setup
3.4.4 Loss Comp
3.4.5 ENR
3.4.6 Cal Setup
3
Measurement
3-13
Page 48
Chapter 3 Measurement
3.4.1 DUT Mode
This section describes the DUT Mode function menu.
Table 3.4.1-1 DUT Mode Function Menu
Function Key Menu Display Function
F1 Amplifier Sets the DUT Mode to the Amplifier.
F2 Down Converter Sets the DUT Mode to Down Converter.
F3 Up Converter Sets the DUT Mode to Up Converter.
3-14
Page 49
3.4.2 Convert Setup
This section describes how to configure Convert Setup.
3.4 Setting Common Items
Table 3.4.2-1 Convert Setup Function
Function
Key
F2 LO Mode
F3 Local Freq
F4 IF Freq
F5 LO Power
F8
Menu DisplayFunction
Sideband
Mode
Sets local frequency mode at measurement.
Fixed: Fixes local frequency and performs
Sets IF frequency to fix when LO Mode is variable.
Enable when LO Mode is Variable.
Setting Range: 1 Hz to the maximum frequency of
the main unit
Resolution: 1 Hz
Default: 30000000 Hz (30 MHz)
Sets the output level of Local signal.
Enable when LO Control is On.
Setting Range: –100 to 100 dBm
Resolution: 0.01 dB
Default: 0.00 dBm
Sets sideband to measure.
When DUT Mode is Up Converter, DSB is unavailable.
LSB: Lower Sideband (Default)
USB: Upper Sideband
DSB: Double Sideband
3
Measurement
3-15
Page 50
Chapter 3 Measurement
3.4.3 External Lo Setup
This section describes how to configure External LO Setup.
Table 3.4.3-1 External Lo Setup Function
Function
Key
F1 LO Control
F2 GPIB Address
F3 LO Select
F4
F5 Command
F6 Settling Time
F7
F8
Menu DisplayFunction
Command
Select
LO MIN
Frequency
LO MAX
Frequency
Sets whether to control inputting external local
signal from external device.
On: Controls
Off: Does not control (Default)
Sets GPIB address of signal source externally
connected.
Setting Range: 0 to 30
Default: 18
Sets signal source of control target.
Setting Range: Actual machines list of VISA
Remote Vector SG (built-in SG option
installed)
Sets control command type of external local signal.
Enable when LO Select is not Vector SG.
SCPI: SCPI command (Default)
Custom: Arbitrary setting command
Opens Command Function Menu.
Enable when LO Select is not Vector SG.
Sets waiting time for completing external local signal
setting.
Setting Range: 0 to 5 s
Resolution: 1 ms
Default: 0 ms
Sets minimum frequency of external local signal.
Sets maximum frequency of external local signal.
3.4.3.1 Command
3-16
Note:
The Function F2 to F8 on the menu are unavailable when LO
Control is OFF.
Page 51
3.4.3.1 Command
3.4 Setting Common Items
This section describes how to configure command settings.
Table 3.4.3.1-1 Command Function
Function
Key
F1 LO Auxiliary
F2
F3 LO Freq Suffix
F4
F5
Menu DisplayFunction
LO Freq
Prefix
LO Power
Prefix
LO Power
Suffix
Sets SG RF Output remote command character
string to control external SG.
Setting Range: Supports ASCII.
Maximum character number: 79
Default: “OUTP:STAT ON”
Sets SG Frequency remote command character
string to control external SG.
Setting Range: Supports ASCII.
Maximum character number: 79
Default: “FREQ”
Sets SG Frequency remote command unit character
string to control external SG.
Setting Range: Supports ASCII.
Maximum character number: 79
Default: “HZ”
Sets SG Output Level remote command character
string to control external SG.
Setting Range: Supports ASCII.
Maximum character number: 79
Default: “POW”
Sets SG Output Level remote command unit
character string to control external SG.
Setting Range: Supports ASCII.
Maximum character number: 79
Default: “DBM”
3
Measurement
3-17
Page 52
Chapter 3 Measurement
3.4.4 Loss Comp
This section describes how to configure the loss compensation settings.
Table 3.4.4-1 Loss Comp Function Menu
Function
Key
F1 Before DUT
F2
F3
F5 After DUT
F6
F7
Menu DisplayFunction
Before DUT
Fixed
Before DUT
Table
After DUT
Fixed
After DUT
Table
Sets the loss compensation before the DUT input
port. Select from the following three options.
Off: Does not compensate the loss before the DUT
input port.
Fixed: Compensates the loss with the value of
Before DUT Fixed.
Table: Compensates the loss with the value of
Before DUT Table.*
Sets the compensation value of Before DUT Fixed.
Enable when Before DUT is Fixed.
Setting Range: –99.999999 to 99.999999 dB
Resolution: 0.000001 dB
Default: 0 dB
Opens the Before DUT Table Function Menu.
Enable when Before DUT is Table.
3.4.4.1 Before DUT Table
Sets the loss compensation after the DUT output
port.
Select from the following three options.
Off: Does not compensate the loss after the DUT
input port.
Fixed: Compensates the loss with the value of After
DUT Fixed.
Table: Compensates the loss with the value of After
DUT Table.
Sets the compensation value of After DUT Fixed.
Enable when After DUT is Fixed.
Setting Range: –99.999999 to 99.999999 dB
Resolution: 0.000001 dB
Default: 0 dB
Opens the After DUT Table Function Menu.
Enable when After DUT is Table.
3.4.4.2 After DUT Table
3-18
Page 53
3.4 Setting Common Items
*: When considering the frequency range to which the corrected value is
input is from Fa to Fb, a corrected value is applied instead if the
displayed frequency is out of the frequency range from Fa to Fb:
If the displayed frequency range is lower than Fa, then the
corrected value La is applied.
If the displayed frequency range is higher than Fb, then the
corrected value Lb is applied.
The value between corrected values is log-interpolated.
Level (dBm)
3
Measurement
Lb
La
Fa
Corrected value input range
Figure 3.4.4-1 Operation When Correction Values Are Entered
Fb
Frequency (Hz)
3-19
Page 54
,
Chapter 3 Measurement
3.4.4.1 Before DUT Table
This menu is used to configure settings for the loss compensation before
the DUT input port.
Table 3.4.4.1-1 Before DUT Table Function Menu
Function
Key
F1 Device Specifies the Table file storage destination drive.
F2
F3
F6 Edit
Menu Display Function
Save Loss Comp
Before DUT Table
Recall Loss Comp
Before DUT Table
The file format for Before DUT Table is as follows:
Directory to output files to: Anritsu Corporation\Signal Analyzer\User
Default file name for saving: LossCompBeforeDUTYYYYMMDD_n.csv
[3] Describe Frequency and Loss Value in each single line,
separating with a comma.
Setting Range of After DUT Table is as follows:
Frequency: 0 to 100 GHz
Loss Value: –99.999999 to 99.999999 dB
2.111111
3-23
Page 58
Chapter 3 Measurement
3.4.5 ENR
This menu is used to configure the ENR (Excess Noise Ratio) function
settings.
Table 3.4.5-1 ENR Function Menu
Function
Key
F1
F2
F3 ENR Mode
F4 Meas Table
F5
F6 Cal Table
F7 Spot
F8 T cold
Menu
Display
Noise Source
Select
Noise Source
Settling
Time
Use Table
for Cal
Function
Selects the Noise Source.
Sets the measurement waiting time when Noise
Source On/Off is switched.
Setting range: 0 to 5 s
Resolution: 1 ms
Default: 0 ms
Selects the ENR Mode (Table or Spot).
Default: Table
Opens the Meas Table menu.
This is available only when ENR Mode is Table.
Select the ENR Table for Cal from Meas Table
/Cal Table.
Default: Meas Table
Opens Cal Table function menu.
Enable when Use Table for Cal is Cal Table.
Opens the Spot menu.
This menu is available only when the ENR Mode
is Spot.
Sets the temperature when Noise Source is Cold.
Setting range: 0 K to 29650000 K
Setting unit: K, °C, °F
Resolution: 0.01
Default: 296.50 K
3.4.5.1 Noise Source Select
3.4.5.2 Meas Table
3.4.5.3 Cal Table
3.4.5.4 Spot
3-24
Page 59
3.4.5.1 Noise Source Select
The Noise Source Select function selects Noise Source to be used with the
measurement. This software supports the NC346 Series noise sources by
Noisecom. For detailed specifications, refer to the catalog and data sheet
for the NC346 Series.
3.4 Setting Common Items
Table 3.4.5.1-1 Noise Source Select Function
Function
Key
F1 NC346A
F2 NC346B
F3 NC346C
F4 NC346D
F5 NC346E
F6 NC346Ka
F8 User Arbitrary noise source
Menu DisplayFunction
NC346A (Option1) by Noisecom
Frequency range: 0.01 to 18.0 GHz
Output ENR 5 to 7 dB
NC346B (Option1) by Noisecom
Frequency range: 0.01 to 18.0 GHz
Output ENR 14 to 16 dB
NC346C by Noisecom
Frequency range: 0.01 to 26.5 GHz
Output ENR 13 to 17 dB
NC346D (Option1) by Noisecom
Frequency range: 0.01 to 18.0 GHz
Output ENR 19 to 25 dB
NC346E by Noisecom
Frequency range: 0.01 to 26.5 GHz
Output ENR 19 to 25 dB
NC346Ka by Noisecom
Frequency range: 0.1 to 40.0 GHz
Output ENR 10 to 17 dB
3
Measurement
3-25
Page 60
,
Chapter 3 Measurement
3.4.5.2 Meas Table
This function is used to configure the Meas Table function settings. For
description of function keys, refer to Table 3.4.5.2-1.
Table 3.4.5.2-1 Meas Table Function Menu
Function
Key
F1 Device Specifies the Table file storage destination drive.
F2
F3
F6 Edit
Menu DisplayFunction
Save Meas
Table
Recall Meas
Table
The file format for Meas Table is as follows:
Directory to output files to: Anritsu Corporation\Signal Analyzer\
Opens the Cal Table dialog box and Cal Table Edit
function menu. Setting procedure is same as the
Meas Table Edit.
3.4.5.2.1 Meas Table Edit
User Data\CalTable\
15.0111
3-28
[1] File Type: Describe the file type.
Describe ENR for Cal Table.
[2] Describe the version of the file format.
[3] Describe Frequency and ENR Value in each single line,
separating with a comma.
The setting range of Cal Table is as follows:
Frequency: 0 to 100 GHz
Loss Value: –17 to 50 dB
Page 63
3.4.5.4 Spot
3.4 Setting Common Items
This section describes how to configure the Spot mode. For description of
function keys, refer to Table 3.4.5.4-1.
Table 3.4.5.4-1 Spot Function Menu
Function
Key
F1 Spot Mode
F2 Spot ENR
F3 Spot T hot
Menu DisplayFunction
Switches Spot Mode to ENR/T hot.
Default: ENR
Sets Spot ENR Value.
This is available only when Spot Mode is ENR.
Setting range: –17 to 50 dB
Resolution: 0.001 dB
Default: 15.200 dB
Sets Spot T hot Value.
This is available only when Spot Mode is T hot.
Setting Range: 0 to 29650000 K
Unit: K, °C, °F
Resolution: 0.01
Default: 9982.80 K
3
Measurement
3-29
Page 64
Chapter 3 Measurement
3.4.6 Cal Setup
This section describes how to configure the various settings related NF
Calibration. Press
the Measure function menu to display the Cal Setup function menu.
Note:
Do not execute the CAL function of the main unit after NF
Calibration is executed.
If it is executed, correct measurement results cannot be obtained.
Table 3.4.6-1 Cal Setup Function Menu
on the Common Setting function menu or on
Function
Key
F1 Min ATT
F2 Max ATT
Menu
Display
Sets the attenuator value to start NF
Calibration. Changing Min ATT value clears NF
Calibration data.
Setting Range: 0 to 40 dB
Resolution:
[MS269xA] 2 dB
[MS2830A] 2 dB (Other than Option 045)
10 dB (Option 045)
[MS2840A] 2 dB (except below)
10 dB (Option 046
[MS2850A] 2 dB
Default: 0 dB
Sets the attenuator value to Stop NF
Calibration. Changing Max ATT value clears NF
Calibration data.
Setting Range: 0 to 40 dB
Resolution:
[MS269xA] 2 dB
[MS2830A] 2 dB (Other than Option 045)
10 dB (Option 045)
[MS2840A] 2 dB (except below)
10 dB (Option 046
[MS2850A] 2 dB
Default:
[MS269xA] 2 dB
[MS2830A] 2 dB (Other than Option 045)
10 dB (Option 045)
[MS2840A] 2 dB (except below)
10 dB (Option 046
[MS2850A] 2 dB
Function
without 019/119)
without 019/119)
without 019/119)
3-30
Page 65
3.4 Setting Common Items
Table 3.4.6-1 Cal Setup Function Menu (Cont’d)
Function
Key
F4
F6
F8
Menu
Display
Switches whether the NF Calibration result is
Apply
Calibration
Clear Cal
Data
Calibration
Now
Table 3.4.6-2 NF Calibration Function Menu
enabled or disabled.
On: Enabled.
Off: Disabled.
Default: On
Clears the NF Calibration data and saved file. If
the NF Calibration data does not exist, it cannot
be measured correctly.
Executes NF Calibration.
The NF Calibration menu (Refer to Table
3.4.6-2.) is displayed during Calibration is being
executed.
Note:
When the below settings are changed, Cal
Status becomes Uncal. The measurement
cannot be performed properly in Uncal state.
Make sure to execute NF Calibration.
Frequency List Table
DUT Mode
LO Mode
Local Freq
IF Freq
Sideband Mode
Noise Source Select
ENR
Min ATT
Max ATT
Function
3
Measurement
Function
Key
F8 Cancel
Menu
Display
Function
Stops NF Calibration and returns to the Cal
Setup menu. In this case, the previous NF
Calibration data is not cleared but held.
3-31
Page 66
Chapter 3 Measurement
3.5 Setting the Measure Function
This section describes how to set the Measure menu settings. Press
(Measure) on the main function menu or
function menu.
Table 3.5-1 Measure Function Menu
to display the Measure
Function
Key
F1 Trace
F2 Storage
F3 Layout
F5 BW
F6
F7
F8 Cal Setup
Menu Display Function
Analysis Time
Mode
Analysis Time
(Ave. Time)
Opens the Trace menu.
3.5.1 Trace
Opens the Storage menu.
3.5.2 Storage
Switches the screen layout mode between Table
and Graph.
Sets the filter bandwidth when a Gauss filter is
applied.
This is the same function as Gauss, the filter
setting of the Signal Analyzer main unit function.
For details, refer to the Operation Manual
(Mainframe Operation).
(In this application, the maximum is 8 MHz and
the minimum is 100 kHz.)
Switches Analysis Time automatic setting
ON/OFF.
Auto: Performs the automatic setting.
Manual: Sets manually.
Default: Auto
The analysis time length necessary for NF
measurement errors to converge into 2σ (±0.034
dB) is set automatically according to the BW
setting when Auto is set.
Sets the Analysis time.
Setting range: 2 s to 200 s
(The range is limited by BW.)
Setting unit: s, ms, s, ns
Resolution: 1 us to 500 s
(The resolution is limited by
BW.)
Default: 16.19 ms
Opens the Cal Setup menu.
3.4.6 Cal Setup
3-32
Page 67
3.5.1 Trace
3.5 Setting the Measure Function
This section describes the Trace Function Menu.
Table 3.5.1-1 Description of Trace Functions Menu
Function
Key
F1 Trace Select
F2 Result Type
F7 Reference
F8 Scale / Div
Function
Key
Menu Display Function
Table 3.5.1-2 Description of Result Type Menu Functions
Menu Display Function
Switches the active trace between Trace1 and
Trace2.
1: Sets the Trace to 1.
2: Sets the Trace to 2.
Default: 1
Select the measurement result to be displayed
for each trace. For details, refer to Table
3.5.1-2.
Sets the median of the vertical axis of the
graph for each Result Type. For the setting
parameters by Result Type, refer to Table
3.5.1-3.
Sets the scale of the vertical axis of the graph
for each Result Type. For the setting
parameters by Result Type, refer to Table
3.5.1-4.
3
Measurement
F1
F2
F3 Gain Displays the gain.
F4 Y - Factor
F5 T effective Displays the effective noise temperature.
F6 P Hot
F7 P Cold
Noise Figure
(dB)
Noise Factor
(Linear)
Displays the noise figure (NF).
Displays the noise factor.
Displays Y-factor (noise power ratio when
Noise Source is turned ON/OFF)
Displays the Power value measured when
Noise Source is On.
Displays the Power value measured when
Noise Source is Off.
3-33
Page 68
Chapter 3 Measurement
Table 3.5.1-3 Reference Setting Parameter List
Result Type
Maximum 100 dB 1E9 100 dB 100 dB 100 000 000 K 100 dB 100 dB
Minimum –100 dB 0 –100 dB –100 dB –100 000 000 K –100 dB –100 dB
Setting unit dB None dB dB K, °C, °F dB dB
Resolution 0.01 0.01 0.01 0.01 0.01 0.01 0.01
Default 4.0 dB 2.5 15.0 dB 5.0 dB 1000.0 K 5.0 dB 0 dB
Result Type
Maximum 20 dB 100 20 dB 20 dB 20 000 000 K 20 dB 20 dB
Minimum 0.001 dB 0.001 0.001 dB0.001 dB 0.1 K 0.001 dB 0.001 dB
Setting unit dB None dB dB K, °C, °F dB dB
Resolution 0.001 0.001 0.001 0.001 0.001 0.001 0.001
Default 1.0 dB 0.715 5.0 dB 1.0 dB 200.0 K 1.0 dB 1.0 dB
Noise
Figure
Noise
Figure
Noise
Factor
Table 3.5.1-4 Scale/Div Setting Parameter List
Noise
Factor
Gain Y-FactorT effective P hot P cold
Gain Y-FactorT effective P hot P cold
3-34
Page 69
3.5.2 Storage
3.5 Setting the Measure Function
This section describes the Storage Function Menu.
Table 3.5.2-1 Description of Storage Function Menu
Function
Key
F1 Storage Mode
F2 Storage Count
Menu Display Function
Switches Storage to off/Average.
Off: Does not perform storage.
Average: Turns on the averaging function.
(Storage is performed for the
measurement results.)
Default: Off
Sets the count when Storage Mode is Average.
Maximum: 999
Minimum: 2
Resolution: 1
Default: 2
3
Measurement
3-35
Page 70
Chapter 3 Measurement
3.6 Setting the Marker Function
Press in the main function menu to display the Marker function
menu.
Table 3.6-1 Description of Marker Function Menus
Function
Key
Page 1 Marker —
F1 Active Marker
F2 Normal Displays the frequency (time) and level.
F3 Delta
F4 Off Turns Off the Marker display.
F7 Relative To
Page 2 Marker —
F8 All Marker Off Turns Off all the markers.
Marker is intended for both Trace1 and Trace2. Additionally, Marker is
valid only to the measurement point.
Menu Display Function
Sets the Active Marker.
Setting range: 1 to 4
(This software does not support Marker
5 and later.)
The Marker frequency and level display
are values relative to the reference point
(Marker set with Relative To)
Sets the reference marker when Active
Maker is Delta.
3-36
Page 71
3.7 Setting the Peak Search Function
3.7 Setting the Peak Search Function
Press in the main function menu to display the Peak Search function
menu.
Table 3.7-1 Description of Peak Search Function Menu
Function
Key
F1 Peak Search
F2 Next Peak
F5 Min Search
F6 Next Min
Menu Display Function
Performs a peak search on the active
trace according to the type specified by
Peak Criteria.
Performs a next search on the active
trace according to the type specified by
Peak Criteria.
Performs a minimum level search on the
active trace, and then moves the marker
to it.
Performs a level search on the active
trace, and then moves the marker to a
minimum level point larger than the
current position.
3
Measurement
3-37
Page 72
Chapter 3 Measurement
Table 3.7-1 Description of Peak Search Function Menu (Cont’d)
Function
Key
F8 Peak Criteria
Menu Display Function
Search performed by Peak Search and
Next Peak depends on the Peak Criteria
setting as shown in the table below.
Peak
Criteria
Max
Min
Peak to
Peak
Default Max
The detail of Search types are as follows.
Max Search
Searches the maximum level point of the
active trace and moves the Marker.
Next Max
Searches the level of the active trace and
moves the marker to the maximum level
point which level is lower than the
current position.
Min Search
Searches the minimum level point of the
active trace and moves the marker.
Next Min
Searches the level of the active trace and
moves the marker to the minimum level
point which level is higher than the
current position.
Peak to Peak Search
Searches the level of the active trace,
moves the active marker to the
maximum level point, and moves the
maker set with Relative to of the active
marker to the minimum level point.
Peak
Search
Max
Search
Min
Search
Peak to
Peak
Next
Peak
Next
Max
Next Min
Does not
perform
a search.
3-38
Page 73
3.8 Measurement Results
Measurement results are displayed. The measurement result display
depends on the settings of Frequency Mode (Refer to Section 3.2.1
“Frequency Mode”) and Layout (Refer to Section 3.5 “Setting the Measure
Function”).
3.8.1 List display
The measurement result is displayed with List display when Frequency
Mode and Layout are as follows:
3.8 Measurement Results
3
Frequency Mode: List or Sweep
Layout: Table
[2]
[1]
[3]
Measurement
Figure 3.8.1-1 Appearance of the Screen in the List Display Mode
[1] Parameter Area
[2] Measurement Status Area
[3] List Area
3.8.1.1 Parameter Area
3.8.1.2 Measurement Status Area
3.8.1.3 List Area
3-39
Page 74
Chapter 3 Measurement
3.8.1.1 Parameter Area
This section describes the detail of the parameter area. The parameter
area is common for all of List display, Gr a ph d isp l ay, a n d S p ot d i sp l ay.
[1]
[2]
[3]
[4]
Figure 3.8.1.1-1 Close Up of the Parameter Area
[1] BW: Displays the BW value.
[2] Start Frequency: Displays the value of Start Frequency only when
[3] Stop Frequency: Displays the Stop Frequency value only when the
[4] Total Point: Displays the value of Total Point only when the
[5] ATT: Displays the Attenuator value.
[6] DUT: Displays the DUT mode.
[7] T Cold: Displays the T Cold value.
[8] Loss Status: Displays the value of Loss Status, depending on the
[5]
[6]
[7]
[11]
the Frequency mode is set to Sweep.
Frequency mode is set to Sweep.
Storage Mode is set to Fixed.
Before DUT and After DUT settings.
[8]
[9]
[10]
Before: Off After: Off Off Off
Before: Fixed After: Off Fixed Off
Before: Off After: Fixed Off Fixed
Before: Fixed After: Fixed Fixed Fixed
Before: Table After: Off Table Off
Before: Off After: Table Off Table
Before: Table After: Table Table Table
Table 3.8.1.1-1 Loss Status Display List
Loss Status Display Before DUT After DUT
Note:
When Before DUT or After DUT is set to Table, if Table loading
fails, Loss Status is displayed in red. Performing measurements as
it is does not provide correct measurement results. After Table
loading is completed, perform the measurement.
3-40
Page 75
[9] Cal Status Displays the Cal Status.
Table 3.8.1.1-2 Cal Status Display List
3.8 Measurement Results
Cal Status
Display
OK or Uncal On
Off Off
Note:
Even when Apply Calibration is On, NF Calibration data is
disabled if Cal Status becomes Uncal. When Uncal is displayed,
correct measurement results cannot be obtained. Re-execute NF
Calibration.
[10] ENR Status Displays the ENR Status.
Table 3.8.1.1-3 ENR Status Display List
ENR Status ENR Mode
Spot Spot
Table Table
Note:
When ENR Mode is Table, ENR Status may be displayed in red.
This occurs because Meas Table is not set.
Apply
Calibration
3
Measurement
Check the setting of Meas Table when ENR Status is displayed in
red.
[11] Lo Freq / IF Freq:
Displays local frequency or IF frequency that is
fixed according to LO Mode when DUT Mode is not
Amplifier.
Table 3.8.1.1-4 Displayed fixed frequency
LO Mode Displayed fixed frequency
Fixed LO Freq: Local Frequency
Variable IF Freq: IF Frequency
3-41
Page 76
Chapter 3 Measurement
3.8.1.2 Measurement Status Area
The measurement status is displayed. For details, refer to Table 3.8.1.2-1.
No. Parameter Description
[1] Warm Up Indicates that the level and frequency are not stable
[2] Level Over Displayed when the input signal level exceeds the input
[3] Measure “Measure” display blinks while the measurement is being
[4] Storage Displays Storage Mode off/Average and Count for
[1] [2][3][4]
Figure 3.8.1.2-1 Measurement Status Area
Table 3.8.1.2-1 List of displays the Measurement Status Area
when the power is turned on.
range.
executed.
Average.
3-42
Page 77
3.8.1.3 List Area
3.8 Measurement Results
This section describes the details of the List Area.
[1]
[2]
[3]
3
Measurement
Figure 3.8.1.3-1 Close Up of the List Area
[1] Result Type display area:
Displays Frequency, Result Types of Trace1 and
Trace2.
(Result Type: Refer to Section 3.5.1 “Trace”.)
Note:
Frequency display changes to IF Frequency or RF Frequency
according to LO Mode setting (Fixed / Variable) when DUT Mode is
not Amplifier.
[2] Measurement result display area:
Displays frequencies and measurement results
corresponding to the Result Type of Trace1 and
Trace2 each.
[3] Minimum/maximum frequency display area:
Displays the minimum/maximum frequencies of
the measurement range.
3-43
Page 78
Chapter 3 Measurement
3.8.2 Graph Display
The measurement results are displayed as a graph when Frequency Mode
and Layout are set as follows:
Frequency Mode: Sweep or List
Layout: Graph
[1]
[2]
[3]
[4]
Figure 3.8.2-1 Appearance of the Screen in the Graph Display Mode
[1] Parameter Area
[2] Measurement Status Area
[3] Graph Area Upper: Trace1, Lower: Trace2
[4] Marker List Area
3.8.2.1 Parameter Area
The parameter area is common for all of List display, Graph display, and
Spot display. Refer to Section 3.8.1.1.
3.8.2.2 Measurement Status Area
The measurement status area is the same for List display, Graph display,
and Spot display. Refer to 3.8.1.2.
3.8.2.1 Parameter Area
3.8.2.2 Measurement Status Area
3.8.2.3 Graph Area
3.8.2.4 Marker List Area
3-44
Page 79
3.8.2.3 Graph Area
3.8 Measurement Results
This paragraph describes the Graph Area.
Trace1
[3]
[4]
Trace2
[3]
[4]
The graph is divided into two. Trace1 result is shown in the upper field
and Trace2 result in the lower field.
[1] Result Type display area:
Displays Result Types of Trace1 and Trace2.
[2]
Figure 3.8.2.3-1 Detail of Graph Area
(Result Type: Refer to Section 3.5.1 “Trace”.)
[1]
[5]
[6]
[1][5] [2]
[6]
3
Measurement
[2] Scale/ DIV, Reference display area:
Displays the value of Scale/Div and Reference.
[3] Scale display area:
Displays the scale where the Reference value is in
the center.
[4] Graph display area:
Consists of the following items.
Vertical axis maximum value:
Reference + (Scale / div) × 5
Vertical axis minimum value:
Reference – (Scale / div) × 5
Horizontal axis maximum value:
Maximum value of Measure List Parameter
“Frequency” in the Measurement List Table
Horizontal axis maximum value:
Minimum value of Measure List Parameter
“Frequency” in the Measurement List Table
3-45
Page 80
Chapter 3 Measurement
3.8.2.4 Marker List Area
[5] Active Marker Information display area:
Displays the frequency and level of the marker
specified with Active Marker.
[6] Marker: Displays the marker icon at the frequency and level
position of the marker specified with Active
Marker.
This paragraph describes the Marker List Area.
[4]
[1]
Figure 3.8.2.4-1 Detail of Marker List Area
[1] Marker Mode display area:
The display changes according to Marker Mode.
Marker Mode: Normal
Displays the Marker number as it is.
Marker Mode: Delta
Displays with the marker number form specified
Marker Mode: Off
Deletes all displays of corresponding rows.
[2] Marker Frequency display area:
[2]
with Marker number Δ Relative to.
[3]
Trace1
Trace2
3-46
Displays the frequency of the Marker position.
[3] Level display area:
Displays the level at the Marker position of Trace1
(left side) and Trace2 (right side) each.
[4] Minimum/Maximum Frequency display area:
Displays the minimum/maximum frequencies of
the measurement range.
Page 81
3.8.3 Spot Display
3.8 Measurement Results
The measurement results are displayed with Spot display when
Frequency Mode is Fixed.
[1]
[2]
[3]
3
Measurement
Figure 3.8.3-1 Appearance of the Screen in the Spot Display Mode
[1] Parameter Area
[2] Measurement Status Area
[3] Spot Area
3.8.3.1 Parameter Area
The parameter area is the same for List display, Graph display, and Spot
display. Refer to 3.8.1.1.
3.8.3.2 Measurement Status Area
The measurement status area is the same for List display, Graph display,
and Spot display. Refer to 3.8.1.2.
3.8.3.1 Parameter Area
3.8.3.2 Measurement Status Area
3.8.3.3 Spot Area
3-47
Page 82
Chapter 3 Measurement
3.8.3.3 Spot Area
This paragraph describes the Spot Area.
[1]
[4]
Figure 3.8.3.3-1 Close Up of the Spot Area
[2][3]
[1] Measurement Frequency:
Displays the Measurement Frequency.
Switches between IF Frequency and RF Frequency
according to LO Mode (Fixed / Variable) when DUT
Mode is not Amplifier.
[2] Trace Level (Trace1):
Displays the trace level (measurement result)
according to the Result Type of Trace1.
[3] Trace Level (Trace2):
Displays the trace level (measurement result)
according to the Result Type of Trace2.
(Result Type: Refer to Section 3.5.1 “Trace”)
[4] Noise Figure: Displays the measurement results of Noise Figure.
NF Current Displays the latest measurement value of Noise
Figure.
NF Average: Displays the average value of Noise Figure.
NF Max Displays the maximum value of Noise Figure.
NF Min Displays the minimum value of Noise Figure.
3-48
NF Max to Min Displays the difference of the maximum and
minimum values of Noise Figure.
Page 83
3.8.4 Saving Measurement Results
This section describes how to save the measurement results. Press
in the NF Measurement screen, to display the Save menu.
Table 3.8.4-1 Description of the Save Function Menu
3.8 Measurement Results
Function
Key
F1 Device
F6 Save All Results
F7 Save Application Saves all the Application settings.
F8 Close Closes the Save menu.
<Procedure>
1. Press
2. When the Setting window is displayed, select the target drive and
then press
3. With the save destination set, press
The file format for the measurement results is as follows:
Displays (On) or hides (Off) the title
character string.
Erase Warm Up
Message
Erases the warmup message
display.
4.1 Selecting Other Functions
Pressing (Accessory) on the main function menu displays the
Accessory function menu.
Table 4.1-1 Accessory function menu
4.2 Setting Title
Function
Keys
F2 Title (On/Off)
F4
A title of up to 32 characters can be displayed on the screen. (Character
strings of up to 17 characters can be displayed on a function menu. The
maximum number of characters to be displayed on the top of the function
menu varies according to character string.)
<Procedure>
1. Press
2. Press
a character using the rotary knob, and enter it by pressing
Enter the title by repeating this operation. When the title is entered,
Menu Display Function
(Accessory) on the main function menu.
(Title) to display the character string input screen. Select
.
press
3. Press
(Set).
(Title) and then select “Off ” to hide the title.
4.3 Erasing Warmup Message
The warmup message (), which is displayed upon power-on and
indicates that the level and frequency are not stable, can be deleted.
<Procedure>
1. Press
2. Press
(Accessory) on the main function menu.
(Erase Warm Up Message) to erase the warmup message.
Page 89
5-1
Performance Test
Performance Test
Chapter 5
This chapter describes measurement devices, setup methods, and
performance test procedures required for performing performance tests as
preventive maintenance.
5.1
Overview of Performance Test ..................................... 5-2
5.1.1 Performance test .............................................. 5-2
5.1.2 Performance test items and instruments used . 5-3
5.2 Power Port Operation Check ........................................ 5-4
5
Page 90
Chapter 5 Performance Test
5-2
5.1 Overview of Performance Test
5.1.1 Performance test
Performance tests are performed as part of preventive maintenance in
order to prevent the performance degradation before it occurs.
Use performance tests when required for acceptance inspection, routine
inspection and performance verification after repairs.
If items that do not meet the required level are detected during
performance testing, contact an Anritsu Service and Sales office.
CAUTION
Warm up the subject testing device and measuring
instruments for at least 30 minutes, in order to stabilize
them sufficiently before running performance tests.
Demonstrating maximum measurement accuracy requires,
in addition to the above, conducting performance tests
under ambient temperatures, little AC power supply
voltage fluctuations (100 to 120 VAC, 200 to 240 VAC), as
well as the absence of noise, vibrations, dust, humidity or
other problems.
Page 91
5.1 Overview of Performance Test
5-3
Performance Test
Voltage check on operation
Function
5.1.2 Performance test items and instruments used
Table 5.1.2-1 lists the performance test items and measuring instruments
used for testing each of these test items.
Table 5.1.2-1 List of performance test items and measuring instruments
Test Items Summary
(On/Off)/non-operation and
Supply
Voltage
Power port
operation check
Perform items deemed critical at regular intervals as preventive
maintenance. A recommended cycle for routine tests of once or twice a
year is desirable.
measurement of rising/falling
time of the power port for the
Noise Figure Measurement
Main Instruments Used
(Anritsu Model Name)
Oscilloscope
5
Page 92
Chapter 5 Performance Test
5-4
MS269xA/MS2830A/MS2840A/MS2850A
Oscilloscope
Connector for
Noise Source: +28 V Pulsed
BNC
Application = Noise Figure Measurement Function
Spot ENR = 15 dB
X scale: 1 ms/div
5.2 Power Port Operation Check
Perform voltage check on operation/non-operation and measurement of
rising/falling time of the power port for the Noise Figure Measurement
Function.
Test specifications
Supply Voltage Operating On: 28 V±0.5 V
Operating Off: 0 V±0.5 V
Not Operating: 0 V±0.5 V
Turn On Time: ≤1 ms
Turn Off Time: ≤1 ms
Table 5.2-1 Test Conditions and Common Parameters List
Frequency Mode = Fixed
Fixed Frequency = 1000 MHz
RBW = 4 MHz(Default)
ENR Mode = Spot
Figure 5.2-1 Power Port Operation Check
Y scale: 5 V/div
Input Impedance: 1 MΩ
Trigger: Positive
Trigger Level: 10 V
Page 93
5.2 Power Port Operation Check
5-5
Performance Test
Time
Voltage
Turn On Time
Turn Off Time
On state Voltage
Voltage of not
Operating
Off state Voltage
Test procedure
1. Execute common parameters (Table 5.2-1) setting for the
oscilloscope.
2. After preset, execute common parameters setting (Table 5.2-1) for the
Noise Figure Measurement Function of the MS269xA, MS2830A,
MS2840A, or MS2850A.
3. Execute Single measurement and monitor the voltage (voltage for
On-state and voltage for Off-state) that arises in a terminal for the
Noise Source with the oscilloscope.
4. Check the voltage rising time.
5. Change Trigger of the oscilloscope to Negative.
6. Execute Single measurement and monitor the voltage (voltage for
On-state and voltage for Off-state) that arises in a terminal for the
Noise Source with the oscilloscope.
7. Check the voltage falling time.
8. Confirm that the rising time, falling time, voltage for On-state,
voltage for Off-state, and voltage (voltage on non-operation) after
Single measurement execution satisfy the specifications.
5
Figure 5.2-2 Power Port Operation Check Reference Figure
Page 94
Chapter 5 Performance Test
5-6.
Page 95
A-1
Appendix A
Appendix A Error Message
Out of range
The settable range is exceeded.
Not available when Frequency Mode is not
Fixed
You can set only when the Frequency Mode is set to
Fixed.
You can set only when the Frequency Mode is set to
List.
Not available when Frequency Mode is not
Sweep
You can set only when the Frequency Mode is set to
Sweep.
You cannot set when the Frequency Mode is set to
Fixed.
You cannot set when the DUT Mode is set to
Amplifier.
Not available when LO Control is Off
You cannot set when the LO Control is set to Off.
Not available when LO Mode is Variable
You cannot set when the LO Mode is set to Variable.
Not available when LO Mode is Fixed
You cannot set when the LO Mode is set to Fixed.
Remote search result : No device
Local Oscillator is not found.
You cannot set when the LO Select is set to Vector
SG.
VISA driver is not installed
VISA driver is not installed.
The local oscillator is not connected
The local oscillator is not externally connected.
Error when LO Mode is Fixed
Source.
Error when LO Mode is Fixed
Noise Source.
Error when LO Mode is Fixed
main unit.
Error when LO Mode is Variable
Source.
Error when LO Mode is Variable
Noise Source.
Error when LO Mode is Variable
main unit.
Not available when Loss Comp before DUT is
not Fixed
You can set only when the Loss Comp before DUT is
set to Fixed.
Table A-1 Message Processing and Status
Message Description
Not available when Frequency Mode is not List
Not available when Frequency Mode is Fixed
Not available when DUT Mode is Amplifier
Not available when LO Select is Vector SG
Local Frequency Error
Noise Source Maximum Frequency
Local Frequency Error
Noise Source Minimum Frequency
Local Frequency Error
SA Maximum Frequency
IF Frequency Error
Noise Source Maximum Frequency
IF Frequency Error
Noise Source Minimum Frequency
IF Frequency Error
SA Maximum Frequency
The setting exceeds the maximum frequency of Noise
The setting falls below the minimum frequency of
The setting exceeds the maximum frequency of the
The setting exceeds the maximum frequency of Noise
The setting falls below the minimum frequency of
The setting exceeds the maximum frequency of the
Page 96
Appendix A Error Message
A-2.
Not available when Loss Comp before DUT is
not Table
You can set only when the Loss Comp before DUT is
set to Table.
Not available when Loss Comp after DUT is
not Fixed
You can set only when the Loss Comp after DUT is
set to Fixed.
Not available when Loss Comp after DUT is
not Table
You can set only when the Loss Comp after DUT is
set to Table.
Not available when ENR Mode is not Table
You can set only when the ENR Mode is set to Table.
Not available when ENR Mode is not Spot
You can set only when the ENR Mode is set to Spot.
Not available when Use Table for Cal is not
Cal Table
You can set only when the Use Table for Cal is set to
Cal Table.
Not available when Spot Mode is not ENR
You can set only when the Spot Mode is set to ENR.
Not available when Spot Mode is not T hot
You can set only when the Spot Mode is set to T hot.
Not available when Peak Criteria is Peak to
Peak
You cannot set when the Peak Criteria is set to Peak
to Peak.
The measurement does not start unless the
frequency is set correctly.
The calibration does not start unless the frequency is
set correctly.
Not available during Calibration
The operation is unavailable during NF Calibration.
No file to read
No file to read is found.
File read error
File read error occurred.
File format error
File format error occurred.
Empty Title Name
The title name field is empty.
Empty File Name
The file name field is empty.
Invalid character
Invalid characters are input.
Table A-1 Message Processing and Status (Cont’d)
Message Description
Measurement disable
Calibration disable
Page 97
B-1
Appendix B
Appendix B Default Value List
<Frequency>
Frequency Mode Sweep
Sweep Setting
Center Frequency 1.805 GHz
Start Frequency 10 MHz
Stop Frequency 3.6 GHz
Span 3.59 GHz
Sweep Point 11
LO Mode Fixed
Local Freq 10.00 GHz
IF Freq 30.00 MHz
LO Power 0.00 dBm
Sideband Mode LSB
External LO Setup
LO Control Off
GPIB Address 18
LO Select Vector SG
(with built-in option SG installed)
Command Select SCPI
Settling Time 0.000 s
Page 98
Appendix B Default Value List
B-2.
Loss Comp
ENR
Cal Setup
<Measure>
Trace
Storage
Layout Graph
BW 4.000 MHz
Analysis Time Mode Auto
Analysis Time Mode (Ave. Time) 16.189 ms
<Peak Search>
Peak Criteria Max
Before DUT Off
Before DUT Fixed 0 dB
After DUT Off
After DUT Fixed 0 dB
Noise Source Select User
Noise Source Settling Time 0 ms
ENR Mode Table
Meas Table 15.200 dB
Use Table for Cal Meas Table
Cal Table 15.200 dB
T cold 296.50 K
Spot
Spot Mode ENR
Spot ENR 15.200 dB
Spot T Hot 9892.80 K
Min ATT 0 dB
Max ATT 2 dB (MS269xA)
2 dB (MS2830A-040/041/043/044)
10 dB (MS2830A-045)
2 dB (MS2840A-040/041/044)
10 dB (MS2840A-046)
2 dB (MS2850A)
Apply Calibration On
Trace Select Trace1
Result Type Noise Figure
Reference 4.00 dB
Scale/Div 1.000 dB
Storage Mode Off
Storage Count 2
Page 99
Index
Index-1
Index
References are to page numbers.
1
1st Local Output connector ............................. 2-8
A
Erase Warm Up Message ............................. 4-2