ANRITSU MS2690A, MS2691A, MS2692A Operation Manual

Page 1
For safety and warning information, please read this
MS2690A/MS2691A/MS2692A
Signal Analyzer
Operation Manual
Spectrum Analyzer Function
Operation
46th Edition
Additional safety and warning information is provided
within the MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual (Mainframe Operation). Please also refer to it before using the equipment.
Keep this manual with the equipment.
ANRITSU CORPORATION
Document No.: M-W2854AE-46.0
Page 2
DANGER
WARNING
CAUTION

Safety Symbols

To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the
following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of
the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
This indicates a very dangerous procedure that could result in serious injury or death if not performed properly.
This indicates a hazardous procedure that could result in serious injury or death if not performed properly.
This indicates a hazardous procedure or danger that could result in light-to-severe injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and
take the necessary precautions BEFORE using the equipment.
This indicates an obligatory safety precaution. The obligatory operation is indicated symbolically in or near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near the barred circle.
near the triangle.
MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual Spectrum Analyzer Function Operation
24 April 2007 (First Edition) 31 October 2019 (46th Edition)
Copyright © 2007-2019, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the publisher. The contents of this manual may be changed without prior notice. Printed in Japan
ii
Page 3
This product and its manuals may require an Export License/Approv al by
the Government of the product's country of origi
export from your
country.
Before re-exporting the product or manuals, please contact us to confirm
whet
When you dispose of export-controlled items, the products/manuals need to
be broken/shredded so as not to be unlawfully used for military purpose.
Only files that have been provided directly from Anritsu or generated
memory
Do not download or install software that has not been specifically
virus security protection in
Use antivirus software.
Notes On Export Management
n for re-
her they are export-controlled items or not.
Cautions Against Computer Virus Infection
Copying files and data
using Anritsu equipment should be copied to the instrument.
All other required files should be transferred by means of USB
stick or CompactFlash media after undergoing a thorough virus check.
Adding software
recommended or licensed by Anritsu.
Network connections
Ensure that the network has sufficient anti-
place.
Protection against malware (malicious software such as viruses).
This equipment runs on Windows Operating System.
To connect This equipment to network, the following is advised.
Activate Firewall.
Install important updates of Windows.
iii
Page 4
iv
Page 5
I

About This Manual

MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual (Main Frame Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Phase Noise Measurement Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Phase Noise Measurement Function Remote Control)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Main Frame Remote Control)
MS2850A Signal Analyzer Operation Manual (Main Frame Operation)
Or
MS2830A Signal Analyzer Operation Manual (Main Frame Operation)
MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual (Signal Analyzer Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Signal Analyzer Function Remote Control)
MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Signal Analyzer Function Operation)
Or
MS2690A/MS2691A/MS2692A and MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Spectrum Analyzer Function Remote Control)
MS2830A/MS2840A/MS2850A Signal Analyzer Operation Manual (Spectrum Analyzer Function Operation)
Or
MS2840A Signal Analyzer Operation Manual (Main Frame Operation)
MS2690A/MS2691A/MS2692A Signal Analyzer Operation Manual (Spectrum Analyzer Function Operation)
Operation manual configuration
The operation manual configuration of the MS2690A/MS2691A/MS2692A Signal Analyzer is shown below.
Page 6
II
Signal Analyzer Operation Manual (Mainframe)
Signal Analyzer Operation Manual (Mainframe Remote Control)
Description of basic operations, maintenance procedures, common functions and common remote functions of the mainframe
Signal Analyzer Operation Manual (Signal Analyzer Function)
Signal Analyzer Operation Manual (Signal Analyzer Function Remote
Control)
Description of basic operations, functions and remote functions of the signal analyzer
Signal Analyzer Operation Manual (Spectrum Analyzer Function) <This
document>
Signal Analyzer Operation Manual (Spectrum Analyzer Function Remote
Control)
Description of basic operations, functions and remote functions of the spectrum analyzer
Signal Analyzer Operation Manual (Phase Noise Measurement Function
Operation)
Signal Analyzer Operation Manual (Phase Noise Measurement Function
Remote Control)
Description of basic operations, functions and remote functions of the phase noise measurement function.
In this document,
indicates a panel key.
Page 7
III

Table of Contents

About This Manual ....................................... I
Chapter 1 Overview ................................... 1-1
1.1 Overview of Spectrum Analyzer ................................ 1-2
1.2 Features of Spectrum Analyzer ................................. 1-3
Chapter 2 Basic Operation ........................ 2-1
2.1 Display Description .................................................... 2-2
2.2 Single/Continuous Sweep .......................................... 2-4
2.3 Setting Frequency ...................................................... 2-5
2.4 Setting Level .............................................................. 2-15
2.5 Setting RBW/VBW ..................................................... 2-30
1
2
3
4
5
Chapter 3 Display Mode Selection ........... 3-1
3.1 Setting Storage Mode ................................................ 3-2
3.2 Setting Limit Line ....................................................... 3-10
3.2 Setting Time/Sweep ................................................... 3-28
3.4 Save on Event............................................................ 3-32
Chapter 4 Marker Function ....................... 4-1
4.1 Setting Marker............................................................ 4-2
4.2 Setting Peak Search Function ................................... 4-13
4.3 Displaying Marker List ............................................... 4-23
Chapter 5 Trigger Function and
Gate Function ........................... 5-1
5.1 Setting Trigger Function ............................................ 5-2
5.2 Setting Gate Function ................................................ 5-7
6
7
8
Appendix
Index
Chapter 6 Preselector Tuning Function ... 6-1
6.1 Preselector Tuning Function ...................................... 6-2
Page 8
IV.
Chapter 7 Measure Function .................... 7-1
7.1 Selecting Measure Function ...................................... 7-2
7.2 Adjacent Channel Leakage Power Measurement
(ACP) ......................................................................... 7-4
7.3 Burst Average Power Measurement .......................... 7-12
7.4 Channel Power Measurement (Channel Power) ....... 7-15
7.5 Occupied Bandwidth Measurement
(Occupied Bandwidth) ............................................... 7-17
7.6 Spectrum Emission Mask Measurement
(Spectrum Emission Mask) ........................................ 7-20
7.7 Spurious Emission Measurement .............................. 7-31
7.8 Two-tone Third Order Intermodulation Distortion ...... 7-48
Chapter 8 Other Functions ....................... 8-1
8.1 Selecting Other Function ........................................... 8-2
8.2 Setting Title ................................................................ 8-3
8.3 Deleting Warmup Message ....................................... 8-4
8.4 Setting Uncal Message .............................................. 8-5
8.5 Adjusting Internal Reference Frequency Signal ........ 8-5
8.6 Input Source for Reference Frequency Signal .......... 8-7
8.7
Pre-Amp On/Off Display ............................................ 8-8
Appendix A Error Messages ...................... A-1
Appendix B Default Value List ................... B-1
Appendix C Standard Parameter List ........ C-1
Index .......................................................... Index-1
Page 9

Chapter 1 Overview

Overview
This chapter provides an overview of the Spectrum Analyzer function.
Overview of Spectrum Analyzer .................................. 1-2
1.1
1.2 Features of Spectrum Analyzer ................................... 1-3
1
1-1
Page 10
Chapter 1 Overview

1.1 Overview of Spectrum Analyzer

The MS2690A/MS2691A/MS2692A Signal Analyzer (hereinafter, referred to as “MS2690A/MS2691A/MS2692A”) is a device used for the quick, highly accurate, and simple measurement of the transmission characteristics of various wireless devices and systems.
The spectrum analyzer function enables full-span analysis of a broad spectrum, which is a characteristic of conventional sweep type spectrum analyzers, and realizes quick and highly accurate signal processing through the use of a digital IF block. These features make the Spectrum Analyzer function ideal for various applications ranging from R&D to manufacturing.
The Spectrum Analyzer function has the following features.
Frequency band (6 GHz/13.5 GHz/26.5 GHz)
Broad analysis bandwidth (20 MHz)
High dynamic range (180 dB)
High-speed and high-accuracy signal analysis through digital IF
1-2
Page 11
Overview

1.2 Features of Spectrum Analyzer

2f
0
f
0
Spectrum analyzer waveform
Oscilloscope waveform
Frequency
Time
Amplitude
A spectrum analyzer is a measuring device for analyzing the frequency components of an input signal. The relative values of the various frequency components are displayed in graph form, with the horizontal axis representing the frequency and the vertical axis the signal level.
Figure 1.2-1 shows a schematic representation of a signal that includes a harmonic distortion component as observed with an oscilloscope and a spectrum analyzer. Observation of a signal that includes harmonic components with an oscilloscope yields a waveform with distortion, not a sine wave. This indicates that the signal includes harmonic components, but the magnitudes of the various harmonic components cannot be quantitatively measured. Using a spectrum analyzer enables the quantitative grasp of signal components by dividing the signal components into their various frequency components and graphing each frequency and amplitude. In the example shown in Figure 1.2-1, the signal can be seen to consist of a fundamental wave (f0) and a second harmonic component (2 from this figure.
f0), and their respective amplitudes can be read
×
1.2 Features of Spectrum Analyzer
1
Figure 1.2-1 Comparison of spectrum analyzer and oscilloscope
Spectrum analyzer principle
A spectrum analyzer separates the input signal down to the level of single frequencies and obtains their respective amplitudes. Concretely, following extraction of single frequencies using a narrow-band filter(s), their amplitudes are obtained through wave detection. Separation into the various signal components through the use of a filter(s) can be done using two different methods. One method consists in using multiple filters to obtain the various frequency components in one time (real-time method), while the other method consists in obtaining the various frequency components by changing the center frequency of a single filter over time (sweep method).
1-3
Page 12
Frequency-variable knob
Input signal
Input signal
Sweep-type spectrum analyzer
Real-time-type spectrum analyzer
Bandpass filter
(Waveform-variable) bandpass filter
Display device
Display device
Attenuator
Pre-selector
Mixer
Local oscillat or
Sweep s ignal
IF amp.
IF filter
Logarit hmic amp.
Wave
Video filter
Display
Y axis
X axis
RF input
Chapter 1 Overview
The real-time method, which is characterized by a small time lag, is suitable for measurement in a narrow frequency range of several tens of MHz maximum. Compared to the real-time method, the sweep method requires a long time to draw a screen, but since it allows observation of a wide frequency range on the order of several GHz at one time, it is a powerful method for high-order harmonic level measurement and the detection of unknown frequency components.
Figure 1.2-2 Spectrum analyzer principle
det ec t or
generator
Figure 1.2-3 Configuration of typical sweep-type spectrum analyzer
1-4
Page 13
1.2 Features of Spectrum Analyzer
Overview
Figure 1.2-3 shows the configuration of a typical sweep-type spectrum analyzer. The super heterodyne method, which is one of the methods that can be used for implementing a sweep-type spectrum analyzer, is described below. The input signal from external sources passes through an input circuit consisting of an attenuator and a preselector and is mixed with the signal from the local oscillator by a mixer. Here, the input signal is converted into a signal of a given frequency (IF). After being processed through an IF stage, this IF signal determines the vertical axis (level) on the display based on its level. The local oscillator, which operates on the principle that the oscillation voltage changes according to the input voltage, is controlled by the signal from the sweep signal generator (sawtooth wave generator) and determines the horizontal axis (frequency) of the display. As a result, a level that corresponds to the input frequency is displayed as a graph.
The input circuit consists of an attenuator for input signal level adjustment, and a pre-selector that prevents erroneous measurement by reducing responses other than that of the reception signal. The IF stage consists of an IF filter used to separate only the target frequency, an IF amplifier with a gain stabilized across a wide level range, and a logarithmic amplifier with excellent linearity. The MS2690A/MS2691A/ MS2692A employs a digital IF method in its IF stage to obtain superior linearity and selectivity.
1
1-5
Page 14
Chapter 1 Overview
1-6.
Page 15

Chapter 2 Basic Operation

This chapter describes the basic operation of the Spectrum Analyzer function.
Display Description ....................................................... 2-2
2.1
2.2 Single/Continuous Sweep ............................................. 2-4
2.3 Setting Frequency ......................................................... 2-5
2.3.1 Setting center frequency ................................ 2-10
2.3.2 Setting frequency span ................................... 2-11
2.3.3 Setting start frequency.................................... 2-12
2.3.4 Setting stop frequency .................................... 2-12
2.3.5 Setting frequency offset .................................. 2-13
2.3.6 Changing frequency band .............................. 2-13
2.3.7 Setting Whether To
Couple Time Domain Parameters and
Frequency Domain Parameters ..................... 2-14
2.3.8 Setting step size ............................................. 2 -1 4
2.4 Setting Level ............................................................... 2-15
2.4.1 Setting reference level .................................... 2-20
2.4.2 Setting input attenuator .................................. 2-21
2.4.3 Setting scale ................................................... 2-24
2.4.4 Setting reference level unit ............................. 2-26
2.4.5 Setting reference level offset .......................... 2-27
2.4.6 Pre-amp .......................................................... 2-28
2.4.7 Setting impedance .......................................... 2-28
2.4.8 Microwave Preselector Bypass ...................... 2-29
2.5 Setting RBW/VBW ...................................................... 2-30
2.5.1 Setting resolution bandwidth (RBW) .............. 2-32
2.5.2 Setting video bandwidth (VBW) ...................... 2-33
2.5.3 Setting Video VBW/Power VBW .................... 2-34
2
Basic Operation
2-1
Page 16
Chapter 2 Basic Operation

2.1 Display Description

This section describes the main screen and main function menus of the Spectrum Analyzer function.
Figure 2.1-1 Main screen of Spectrum Analyzer function
2-2
Page 17
2.1 Display Description
Pressing the when the Spectrum Analyzer function is selected in the Application Switch menu displays the main function menu.
The main function menu consists of two pages, which can be toggled by pressing
.
Table 2.1-1 Main function menu
Function
Key
Page1 Spectrum Analyzer Press Spectrum Analyzer to display this page.
F1 Frequency
F2 Span
F3 Amplitude
F4 BW
F5 Marker
F6 Trace
F7 Trigger/Gate
F8 Time/Sweep
Page2 Spectrum Analyzer
F1 Peak Search
F2 Measure
F7 Save on Event
F8 Accessory
Menu Display Function
Executes frequency setting and preselector auto tuning.
2.3 “Setting Frequency”
Sets the frequency span and frequency band.
2.3 “Setting Frequency”
Sets the level.
Sets the resolution bandwidth and video bandwidth.
2.5 “Setting RBW/VBW”
Sets markers.
4.1 “Setting Marker”
Sets the trace.
3.1 “Setting Storage Mode”
Sets the trigger and gate.
Chapter 5 “Trigger Function and Gate Fu
Sets the sweep time and trace points.
3.2 “Setting Time/Sweep”
Press Spectrum Analyzer and then press to display page 2.
Sets the peak search execution and conditions.
4.2 “Setting Peak Search Function”
Configures settings related to the Measure function.
Chapter 7 “Measure Function”
Sets the Save on Event function that saves Trace data upon occurrence of an event.
3.4 “Save on Event”
Sets the other functions.
Chapter 8 “Other Functions”
2.4 “Setting Level”
nction”
2
Basic Operation
2-3
Page 18
Chapter 2 Basic Operation

2.2 Single/Continuous Sweep

There are two types of sweep, single and continuous.
(1) Single sweep
Press function is on, only 1 sweep is executed when the sweep start conditions
are met.
(2) Continuous sweep
Press gate function is on, sweep is executed every time the sweep start
conditions are met.
to execute sweep once. When the trigger function or gate
to execute continuous sweep. When the trigger function or
Figure 2.2-1 Single key and Continuous key
2-4
Page 19
2.3 Setting Frequency
The Spectrum Analyzer function can set the following four measurement frequencies.
Center frequency
Frequency span
Start frequency
Stop frequency
Pressing pressing

2.3 Setting Frequency

2
Basic Operation
(Frequency) on page 1 of the main function menu, or
displays the Frequency function menu.
Pressing
displays the Span function menu.
(Span) on page 1 of the main function menu, or pressing
Figure 2.3-1 Frequency key and Span key
2-5
Page 20
Chapter 2 Basic Operation
Frequency function menu
F1
F2
F3
F4
F5
F6
F7
F8
Figure 2.3-2 Frequency function menu
Table 2.3-1 Frequency function menu
Menu Display Function
Frequency Page 1 Press Frequency to display this menu.
Center
Start
Stop
Preselector Auto Tune
Offset (On/Off)
Offset Value
Step Size
Sets the center frequency.
2.3.1 “Setting center frequency”
Sets the start frequency.
2.3.3 “Setting start frequency”
Sets the stop frequency.
2.3.4 “Setting stop frequency”
Perform the preselector auto tuning.
6.1.1 “Preselector auto tuning”
Switches on/off the frequency offset function.
2.3.5 “Setting frequency offset”
Sets the frequency offset value.
2.3.5 “Setting frequency offset”
Sets the step size of the center/start/stop frequencies.
2.3.8 “Setting step size”
2-6
Page 21
2.3 Setting Frequency
Table 2.3-1 Frequency function menu (Cont’d)
Menu Display Function
Frequency Page 2
Micro Wave Preselector Bypass
Press Frequency, and then press this menu.
Enables/Disables the Micro Wave Preselector Bypass function. This function is available with MS2692A-067/167.
2.4.8 “Microwave Preselector Bypass”
1.3.7 “MS2692A-067/167” in the
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
to display
2
Basic Operation
2-7
Page 22
Chapter 2 Basic Operation
Span function menu
F1
F2
F3
F4
F5
F6
F7
F8
Figure 2.3-3 Span function menu
Table 2.3-2 Span function menu
Menu Display Function
Span
Full Span
Zero Span
Couple Time/Frequency Domain
Frequency Band Mode
Sets the frequency span.
2.3.2 “Setting frequency span”
Sets full span.
2.3.2 “Setting frequency span”
Sets zero span.
2.3.2 “Setting frequency span”
Sets whether to couple the time domain parameters and frequency domain parameters.
2.3.7 “Setting Whether To Couple Time
Domain Parameters and Frequency Domain
Sets the frequency band mode.
2.3.6 “Changing frequency band”
Parameters”
2-8
Page 23
2.3 Setting Frequency
The display items related to the frequency parameters are described below.
2
Basic Operation
[1]
Figure 2.3-4 Display items related to frequency parameters
Table 2.3-3 Display items related to frequency parameters
No. Display Description
[1] Center or Start
[2] Span or Stop
Displays the center frequency or start frequency.
Displays the frequency span or stop frequency.
[2]
2-9
Page 24
Chapter 2 Basic Operation

2.3.1 Setting center frequency

The center frequency of the frequency range to be swept can be set.
Setting range and resolution for center frequency
Setting range: 100 MHz to 6.05 GHz (MS2690A)
Resolution: 1 Hz
Example: To set the center frequency to 1 MHz
<Procedure>
1. Press
.
*
100 MHz to 13.6 GHz (MS2691A)
100 MHz to 26.6 GHz (MS2692A)
*
*
*: The setting range is limited by the other
frequency settings.
2. After pressing
, press (MHz) to set the center frequency.
2-10
Page 25

2.3.2 Setting frequency span

The frequency span to be swept can be set. The mode when the frequency span is 0 Hz higher than is called the frequency domain mode (horizontal axis = frequency), and if it is 0 Hz, it is called the time domain mode (horizontal axis = time).
Setting range and resolution for frequency span
Setting range: 300 Hz to 6.15 GHz (MS2690A)
0 Hz (time domain mode)
Resolution:
Example: To set the frequency span to 100 MHz
<Procedure>
1. Press
.
2.3 Setting Frequency
*
300 Hz to 13.7 GHz (MS2691A) 300 Hz to 26.7 GHz (MS2692A)*
*: The setting range is limited by the other
frequency settings.
2 Hz
*
2
Basic Operation
2. After pressing span.
Example: To set the frequency span to its maximum value
<Procedure>
1. Press
2. Press
Example: To set the frequency span to 0 Hz (time domain mode)
<Procedure 1>
1. Press
2. Press
<Procedure 2>
1. Press
2. After pressing Hz.
.
(Full Span).
.
(Zero Span).
.
, press (MHz) to set the frequency
, press (MHz) and set the frequency span to 0
2-11
Page 26
Chapter 2 Basic Operation

2.3.3 Setting start frequency

The start frequency of the frequency range to be swept can be set.
Setting range and resolution for start frequency
Setting range:
Resolution: 2 Hz
Example: To set the start frequency to 10 MHz
<Procedure>
1. Press
*
100 MHz to 6.0499997 GHz (MS2690A)
100 MHz to 13.5999997 GHz (MS2691A)
100 MHz to 26.5999997 GHz (MS2692A)
*
*
*: The setting range is limited by the other
frequency settings.
.
2. Press
3. After pressing

2.3.4 Setting stop frequency

The stop frequency of the frequency range to be swept can be set.
Setting range and resolution for stop frequency
Setting range:
Resolution: 2 Hz
Example: To set the stop frequency to 1 GHz
<Procedure>
1. Press
2. Press
3. After pressing
(Start).
.
(Stop).
, press (MHz) to set the start frequency.
*
99.9997 MHz to 6.05 GHz (MS2690A)
99.9997 MHz to 13.6 GHz (MS2691A)
99.9997 MHz to 26.6 GHz (MS2692A)
*
*
*: The setting range is limited by the other
frequency settings.
, press (GHz) to set the stop frequency.
2-12
Page 27

2.3.5 Setting frequency offset

The offset value for the frequency display can be set.
Setting range and resolution for frequency offset
Setting range: 100 to 100 GHz Resolution: 1 Hz
Example: To set the frequency offset to 1 GHz
<Procedure>
1. Press
2.3 Setting Frequency
2
Basic Operation
.
2. Press
3. After pressing
(Offset Value).

2.3.6 Changing frequency band

Note:
This function can be set only when Option 003 Extension of Preselector Lower Limit to 3 GHz is installed.
The passing lower limit frequency of the preselector can be changed from 6 GHz to 3 GHz by changing the frequency band mode.
Frequency Band Mode Preselector Passing Frequency
Normal > 6.0 GHz Spurious
Example: To set the passing lower limit frequency of the preselector to 3
GHz
<Procedure>
1. Press
, press (GHz) and set the frequency offset.
Table 2.3.6-1 Frequency band
3.0 GHz
.
2. Press
3. Press the preselector to 3 GHz.
(Frequency Band Mode).
(Spurious) and set the passing lower limit frequency of
2-13
Page 28
Chapter 2 Basic Operation
2.3.7 Setting Whether To Couple Time Domain Parameters and
Frequency Domain Parameters
Sets whether to couple the time domain parameters and frequency domain parameters. If coupling is enabled, the RBW, VBW, detection mode, and trace point are coupled (shared) between the time domain and frequency domain. When coupling is enabled, the currently-selected domain parameter is used as a shared setting.
Table 2.3.7-1 Couple Time/Frequency Domain
Couple
Time/Frequency
Domain
The RBW, VBW, detection mode, and trace
On
Off
Example: To disable coupling.
<Procedure>
1. Press
.
point are coupled (shared) between the time domain and frequency domain.
The RBW, VBW, detection mode, and trace point for the time domain are separate from those for the frequency domain.
Descriptions
2. Press

2.3.8 Setting step size

The step size of the center, start and stop frequency can be set.
Setting range and resolution for step size
Setting range: 1 Hz to 6.0 GHz (MS2690A) 1 Hz to 13.5 GHz (MS2691A)
Resolution: 1 Hz
Example: To set step size to 1 MHz
<Procedure>
1. Press
2. Press
3. After pressing
(Couple Time/Frequency Domain) to set Couple
Time/Frequency Domain to Off.
1 Hz to 26.5 GHz (MS2692A)
.
(Step Size).
, press (MHz) to set step size.
2-14
Page 29
2.4 Setting Level
Pressing (Amplitude) on the main function menu, or pressing displays the Amplitude function menu.
The Marker function menu consists of 2 pages that are toggled by pressing
.

2.4 Setting Level

2
Figure 2.4-1 Amplitude key
Basic Operation
2-15
Page 30
Chapter 2 Basic Operation
F1
F2
F3
F4
F5
F6
F7
F8
Page1 Page2
Figure 2.4-2 Amplitude function menu
Table 2.4-1 Amplitude function menu
Menu Display Function
Amplitude Page 1 Press Amplitude to display this menu.
Reference Level
Attenuator (Auto/Manual)
Attenuator
Pre-amp
Log Scale Unit
Scale
Offset (On/Off)
Offset Value
Sets the reference level.
2.4.1 “Setting reference level”
Selects automatic setting or manual setting of the input attenuator.
2.4.2 “Setting input attenuator”
Sets the input attenuator.
2.4.2 “Setting input attenuator”
Sets Pre-amp to On/Off.
2.4.6 “Pre-amp”
Sets the level display unit to be used during log scale.
2.4.4 “Setting reference level unit”
Switches log level display and linear level display.
2.4.3 “Setting scale”
Switches On/Off the reference level offset function.
2.4.5 “Setting reference level offset”
Sets the reference level offset value.
2.4.5 “Setting reference level offset”
2-16
Page 31
Table 2.4-1 Amplitude function menu (Cont’d)
Menu Display Function
2.4 Setting Level
Amplitude Page 2
Impedance
Micro Wave Preselector Bypass
The display items related to the level parameters are described below.
Press Amplitude, and then press this menu.
Sets 50 /75 of impedance.
2.4.7 “Setting impedance”
Enables/Disables the Micro Wave Preselector Bypass function. This function is available with MS2692A-067/167.
2.4.8 “Microwave Preselector Bypass”
1.3.7 “MS2692A-067/167” in the
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
[2] [1]
to display
[3]
2
Basic Operation
Figure 2.4-3 Display items related to level parameters
Table 2.4-2 Display items related to level parameters
No. Display Description
[1] Reference level Displays the reference level.
Reference level
[2]
offset Input
[3]
attenuator
Displays the reference level addition offset value.
Displays the input attenuator value.
2-17
Page 32
Chapter 2 Basic Operation
The reference level (upper end of amplitude scale) range is described below.
Setting range and resolution for reference level
Setting range: Refer to Tables 2.4-3 and 2.4-4. Resolution: 0.01 dB (dB unit system) Effective number of digits: 3 (W unit) Effective number of digits: 3 (V unit)
Table 2.4-3 Reference level setting range
(When Pre-amp is set to Off).
Scale Mode Unit Reference Level Range
*1
dBm
dBV*2
dBmV*3
120 to 50 dBm
13.01 to 156.99 dBV
(Impedance = 50 )
11.25 to +158.75 dBV
(Impedance = 75 )
73.01 to 96.99 dBmV
(Impedance = 50 )
–71.25 to +98.75 dBmV
Log scale
(
Impedance = 75 )
0.224 V to 70.7 V
V
W 1 fW to 100 W
dBV (emf)*4
(Impedance = 50 )
0.274 V to +86.6 V (Impedance = 75 )
6.99 to 163.01 dBV (emf)
(Impedance = 50 )
–5.23 to +164.77 dBV (emf)
dBV/m*5
(
Impedance = 75 )
13.01 to 156.99 dBV/m
(Impedance = 50 )
–11.25 to +158.75 dBV/m
Linear scale (in dBm)
V
(
Impedance = 75 )
22.4 V to 70.7 V (Impedance = 50 )
27.4 V to 86.6 V (Impedance = 50 )
(80 to 50 dBm)
2-18
Page 33
2.4 Setting Level
Table 2.4-4 Reference level setting range (When Pre-amp is set to On).
Scale Mode Unit Reference Level Range
*1
dBm
dBV*2
120 to +30 dBm
13.01 to +136.99 dBV
(Impedance = 50 )
–11.25 to +138.75 dBV
dBmV*3
(
Impedance = 75 )
73.01 to +76.99 dBmV
(Impedance = 50 )
–71.25 to +78.75 dBmV
Log scale
(
Impedance = 75 )
0.224 V to 7.07 V
V
W 1 fW to 1 W
dBV (emf)*4
(Impedance = 50 )
0.274 V to 8.66 V (Impedance = 75 )
6.99 to +143.01 dBV (emf)
(Impedance = 50 )
–5.23 to +144.77 dBV (emf)
Linear scale (in dBm)
dBV/m*5
V
(
Impedance = 75 )
13.01 to +136.99 dBV/m
(Impedance = 50 )
11.25 to +138.75 dBV/m
(
Impedance = 75 )
2.24 V to 7.07 V (Impedance = 50 )
2.74 V to 8.66 V
(Impedance = 50 ) (100 to +30 dBm)
2
Basic Operation
*1: Unit system where 1 mW is regarded as 0 dBm
*2: Unit system where 1 V is regarded as 0 dBV. Indicated by
termination voltage with 50  terminator.
When impedance is set to 75 , it is displayed as the terminal voltage of 75 .
*3: Unit system where 1 mV is regarded as 0 dBmV. Indicated by
termination voltage with 50  terminator.
When impedance is set to 75 , it is displayed as the terminal voltage of 75 .
*4: Unit system in dBV, indicated by open voltage. The value is “dBV
+ 6.02 dB”.
*5: Unit system indicating electric field intensity. The displayed value
depends on the antenna factor.
2-19
Page 34
Chapter 2 Basic Operation

2.4.1 Setting reference level

The reference level (upper end of amplitude scale) can be set.
Example: To set the reference level to 10 dBm
<Procedure>
1. Press
.
2. Press
3. After pressing level.
(Reference Level).
, press (dBm) to set the reference
2-20
Page 35

2.4.2 Setting input attenuator

(1) Auto mode
The input attenuator is automatically set according to the set reference level. When a signal of the same level as the reference level is input, the input level to the internal mixer is set so as become 10 dBm or lower.
Tables 2.4.2-1 and 2.4.2-2 show the settings in the Auto mode.
Table 2.4.2-1 Input attenuators set in Auto mode
N
2.4 Setting Level
2
Basic Operation
(When Pre-amp is set to Off)
= Reference Level (dBm) Attenuator Auto (dB)
120
0
2
4
6
8
10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48
0 10
N
2 12
N
4 14
N
6 16
N
8 18
N
10 20
N
12 22
N
14 24
N
16 26
N
18 28
N
20 30
N
22 32
N
24 34
N
26 36
N
28 38
N
30 40
N
32 42
N
34 44
N
36 46
N
38 48
N
40 50
N
42 52
N
44 54
N
46 56
N
48 58
N
50 60
N
2-21
Page 36
Chapter 2 Basic Operation
Table 2.4.2-2 Input attenuators set in Auto mode
(When Pre-amp is set to On)
= Reference Level (dBm) Attenuator Auto (dB)
N
120
 20N 18N 16N 14N 12N 10N 8N 6N 4N 2N
0
N
2
N
4
N
6
N
8
N
10
N
12
N
14
N
16
N
18
N
20
N
22
N
24
N
26
N
28
N
20 10
N
18 12 16 14 14 16 12 18 10 20
8 22 6 24 4 26 2 28
0 30
2 32
4 34
6 36
8 38
10 40
12 42
14 44
16 46
18 48
20 50
22 52
24 54
26 56
28 58
30 60
2-22
Page 37
2.4 Setting Level
(2) Manual mode
To measure extremely low level signals such as non-harmonic spurious signals or spurious signals adjustment to the signal, set the input attenuator manually.
The setting range of the input attenuator in Manual setting is as follows.
Setting range and resolution for input attenuator Input attenuator setting range: Refer to Tables 2.4.2-3 and 2.4.2-4. Input attenuator minimum resolution: 2 dB
Table 2.4.2-3 Input attenuator setting range
(When Pre-amp is set to Off)
Attenuator Manual
Lower limit Upper limit
Logic* ( = 0,  = 1,  = 2) The minimum value is 0 dB.
Table 2.4.2-4 Input attenuator setting range
(When Pre-amp is set to On)
60 dB
2
Basic Operation
Attenuator Manual
Lower limit Upper limit
Logic* ( = 20,  = 21,  = 22) The minimum value is 0 dB
*: The following rules apply:
<1> If the reference level is 0 or if it is divisible by 2. Attenuator (dB) = RL* <2> Not <1>, and INT (RL)* Attenuator (dB) = INT (RL)* <3> Not <1>, and INT (RL)* Attenuator (dB) = INT (RL)*
*1: Reference level (dBm)
*2: Maximum integer not exceeding reference level.
60 dB
1
+ 
2
is an odd number.
2
+ 
2
is an even number.
2
+ 
2-23
Page 38
Chapter 2 Basic Operation
Example: To set the input attenuator to 20 dB
<Procedure>
1. Press
.

2.4.3 Setting scale

2. Press
3. After pressing
For measurement of second and third harmonic spurious, the mixer input level must be lowered to eliminate the effect of internal distortion. Internal distortion is below is
30 dBm, so to measure harmonic spurious up to –75 dB, the mixer
input level must be below is Auto, the attenuator value is too small.
Pressing (Scale) on the Amplitude function menu displays the Scale function menu.
(Attenuator).
, press (dB) to set the input attenuator.
75 dB (at 1 GHz) when the mixer input level
30 dBm. In this case, if the attenuator setting
F1
F2
F3
F4
F5
F6
F7
F8
When Log is
selected
Figure 2.4.3-1 Scale function menu
When Lin is
selected
2-24
Page 39
Table 2.4.3-1 Scale function menu
Menu Display Function
2.4 Setting Level
Scale (Log/Lin)
Log Scale Division Sets the scale range (Log scale range). Lin Scale Division Sets the scale range (Lin scale range). Log Scale Line
(10/12)
(1) Setting the Log scale
The log scale can be selected from the following:
0.1, 0.2, 0.5, 1.0, 2.0, 5.0, 10.0, 20.0 dB/Div
Example: To set Log Scale Division to 20 dB/Div and number of scale lines
to 12
<Procedure>
1. Press
2. Press
3. Press
4. Press
5. After pressing Division.
.
(Scale).
(Scale) and select Log.
(Log Scale Division).
Selects the scale mode (Log/Lin).
Selects the number of scale lines during log scale.
, press (dB/Div) to set the Log Scale
2
Basic Operation
6. Press selecting 12.
(2) Setting the Lin scale
The Lin scale can be selected from the following:
1, 2, 5, 10%/Div
Example: To set Lin Scale Division to 5%/Div
<Procedure>
1. Press
2. Press
3. Press
4. Press
5. After pressing
(Log Scale Line) and set the number of scale lines by
.
(Scale).
(Scale) and select Lin.
(Lin Scale Division).
, press (%/Div) to set Lin Scale Division.
2-25
Page 40
Chapter 2 Basic Operation

2.4.4 Setting reference level unit

The level display unit for the Log scale mode can be set. In the case of Lin scale, the level display unit is fixed to V.
Reference level display unit
dBm, dBV, d Bm V, V, W, d BV (emf), dBV/m
Example: To set the reference level to 10 dBmV
<Procedure>
1. Press
.
2. Press
3. Press
4. Press
5. Press
6. After pressing unit.
If V (W) is selected and a measurement result is more than 99.999 GV(GW), 99.999 GV(GW) is displayed.
(Scale).
(Scale) and select Log.
to return to the previous menu.
(Log Scale Unit) and then press (dBmV).
, press (dBmV) to set the reference level
2-26
Page 41

2.4.5 Setting reference level offset

The waveform trace can be displayed with an arbitrary offset value added.
Setting range and resolution for reference level offset
Setting range: Resolution: 0.01 dB
100 to 100 dB
2.4 Setting Level
2
Basic Operation
Set offset value to +10 dB
Figure 2.4.5-1 Adding an offset value
Example: To set the reference offset value to 10 dB
<Procedure>
1. Press
.
2. Press
3. Press
4. After pressing offset value.
(Offset) and select On.
(Offset Value).
, press (dB) to set the reference level
2-27
Page 42
Chapter 2 Basic Operation

2.4.6 Pre-amp

Note:
This function can be set only when Option 008 6GHz Pre-amp is installed.
The level sensitivity can be increased by setting Pre-amp to On.
Example: To set Pre-amp to On
<Procedure>
1. Press .
2. Press

2.4.7 Setting impedance

Sets 50 /75 of input impedance. The default value is 50 .
This setting is effective when the vertical axis is voltage units (dBV, dBmV, V, dBV (emf), dBV/m); it is not effective when the units are power (dBm, W). The actual input impedance of the equipment is 50 . If the impedance is set to 75 , numeric conversion is performed for the 75- termination and it is reflected in the voltage-units display.
When using a 75/50  converter, set the impedance to 75 . The converter loss can be corrected using the Correction function and reference level offset setting.
Example: To set impedance to 75 
<Procedure>
1. Press
(Pre-Amp) to set to On.
.
2-28
2. Press press
. After page 2 of the Amplitude function menu is displayed,
(Impedance) to set to 75 .
Page 43

2.4.8 Microwave Preselector Bypass

Note:
This function is available when installing Option 067 Microwave Preselector Bypass.
2.4 Setting Level
The frequency characteristics can be improved at 6 GHz or more of the
preselector band
Example: To set Preselector Bypass to On
<Procedure>
1. Press
2. Pressing function menu.
3. Pressing Bypass to On.
when setting preselector Bypass to On.
or .
displays the page 2 of Frequency or Amplitude
(Micro Wave Preselector Bypass) sets Preselector
2
Basic Operation
2-29
Page 44
Chapter 2 Basic Operation

2.5 Setting RBW/VBW

The resolution bandwidth (RBW) and video bandwidth (VBW) can be selected either automatically or arbitrarily.
Pressing displays the BW function menu.
(BW) on the main function menu, or pressing
Figure 2.5-1 BW key
F1
F2
2-30
F3
F4
F5
F6
F7
F8
Figure 2.5-2 BW function menu
Page 45
2.5 Setting RBW/VBW
Table 2.5-1 BW function menu
Menu Display Function
RBW (Auto/Manual)
RBW Value
VBW (Auto/Manual)
VBW Value
VBWMode (Video/Power)
Selects automatic or manual setting of the resolution bandwidth (RBW).
2.5.1 “Setting resolution bandwidth (RBW)”
Selects the resolution bandwidth (RBW).
2.5.1 “Setting resolution bandwidth (RBW)”
Selects automatic or manual setting of the video bandwidth (VBW).
2.5.2 “Setting video bandwidth (VBW)”
Sets the video bandwidth (VBW).
2.5.2 “Setting video bandwidth (VBW)”
Selects display level smoothing through the video bandwidth (VBW).
2.5.3 “Setting Video VBW/Power VBW”
2
Basic Operation
2-31
Page 46
Chapter 2 Basic Operation

2.5.1 Setting resolution bandwidth (RBW)

Auto or Manual can be selected for setting resolution bandwidth (RBW).
Auto
The resolution bandwidth is automatically set according to the frequency span setting.
Manual
Select an arbitrary resolution bandwidth (RBW). When monitoring two adjacent signals, the frequency resolution can be increased by reducing the resolution bandwidth (RBW). This has also the effect of reducing the noise level.
The resolution bandwidth (RBW) value can be selected from among the following:
30 Hz, 100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 50 kHz, 100 kHz, 300 kHz, 1 MHz, 3 MHz, 5 MHz, 10 MHz, 20 MHz, 31.25 MHz (31.25 MHz is available only with 0 span.)
Note:
For RBW 31.25 MHz, it is not a Gauss filter but a flat-top characteristic filter. If Couple Time/Frequency Domain is set to Off, this parameter specifies a separate value for frequency-axis measurement and time-axis measurement.
When Couple Time/Frequency Domain is set to Off, each parameter is specified in accordance with the already-specified frequency axis or time axis measurement.
Resolution bandwidth (RBW) setting
2-32
Example: To set the resolution bandwidth (RBW) to 10 kHz.
<Procedure>
1. Press
2. Press
3. After pressing bandwidth (RBW).
.
(RBW Value).
, press (kHz) and set the resolution
Page 47

2.5.2 Setting video bandwidth (VBW)

Auto or Manual can be selected for setting video bandwidth (VBW).
When the detection mode (Detection) is “RMS,” VBW (Auto/Manual) is disabled.
2.5 Setting RBW/VBW
Auto
The video bandwidth is set to the same as the resolution bandwidth (RBW) or the nearest value.
Manual
By setting the video bandwidth to a value that is smaller than the resolution bandwidth (RBW), noise averaging is made more efficient.
The video bandwidth (VBW) value can be selected from the following:
1 Hz, 3 Hz, 10 Hz, 30 Hz, 100 Hz, 300 Hz, 1 kHz, 3 kHz, 5 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz, 3 MHz, 10 MHz, Off
When Couple Time/Frequency Domain is set to Off, each parameter is specified in accordance with the already-specified frequency axis or time axis measurement.
Video bandwidth (VBW) setting
Example: To set the video bandwidth (VBW) to 10 kHz
<Procedure>
1. Press
2. Press
3. After pressing
.
(VBW Value).
, press (kHz) and set VBW.
2
Basic Operation
2-33
Page 48
Chapter 2 Basic Operation

2.5.3 Setting Video VBW/Power VBW

The Spectrum Analyzer function allows selection of either Video VBW or Power VBW as the video filter for the better and smooth screen display.
When the detection mode (Detection) is “RMS,” “Power VBW” is performed regardless of VBW Mode (Video/Power).
Differences between Two VBW Modes
Video VBW
Performs smoothing and averaging with a video filter after logarithmic compression (logarithmic mean).
Power VBW
Performs smoothing and averaging with a video filter before logarithmic compression (antilog mean).
Video VBW indicates similar measurement values as those obtained with a conventional spectrum analyzer, but in the case of signals that fluctuate importantly over time, such as burst waves, a large error margin results. When measuring such signals, select Power for the VBW mode.
V
Input signal
P
1/2 (3 dB)
0
~
~
Figure 2.5.3-1 Power VBW and Video VBW
Example: To set VBW Mode to Power
<Procedure>
1. Press
.
t
Power VBW (Antilog mean)
Video VBW (Logarithmic mean)
2-34.
2. Press
(VBW Mode) and select Power.
Page 49

Chapter 3 Display Mode Selection

This chapter describes the trace, storage mode, and wave detection mode.
Setting Storage Mode ................................................... 3-2
3.1
3.1.1 Storage mode types ......................................... 3-4
3.1.2 Averaging function ............................................ 3- 5
3.1.3 Max Hold/Min Hold function ............................. 3-6
3.1.4 Setting wave detection mode ........................... 3-7
3.1.5 Displaying and updating Trace ......................... 3-9
3.2 Setting Limit Line ........................................................ 3-10
3.2.1 Setting Limit function parameters ................... 3-11
3.2.2 Setting Limits .................................................. 3-12
3.2.3 Setting Limit1 to 6 ........................................... 3- 13
3.2.4 Edit parameters .............................................. 3-16
3.2.5 Limit Envelope Function ................................. 3-21
3.2.6 Saving a Limit file ........................................... 3-2 4
3.2.7 Loading a Limit file .......................................... 3-26
3.3 Setting Time/Sweep .................................................... 3-28
3.3.1 Setting sweep time ......................................... 3-30
3.3.2 Setting trace point ........................................... 3-31
3.4 Save on Event............................................................. 3-32
3.4.1 Setting Save on Event .................................... 3-32
3.4.2 Waveform files for Save on Event .................. 3-34
3.4.3 Deleting Waveform files for Save on Event .... 3-35
3
Display Mode Selection
3-1
Page 50
Chapter 3 Display Mode Selection

3.1 Setting Storage Mode

Pressing (Trace) on the main function menu, or pressing displays the Trace function menu.
Figure 3.1-1 Trace key
F1
F2
F3
F4
F5
F6
F7
F8
3-2
Figure 3.1-2 Trace function menu
Page 51
3.1 Setting Storage Mode
Table 3.1-1 Trace function menu
Menu Display Function
Active Trace
Trace Type
Storage Mode
Storage Count
Limits
Detection
Selects trace to be set. Trace Type and Storage Mode can be set for each trace.
Used for setting related to update and display of trace data.
3.1.5 Displaying/Updating Trace
Sets the storage mode type.
3.1.1 Storage Mode Types
Sets the storage count.
3.1.2 Averaging Function
Sets the limit line.
3.2 Setting Limit Line
Sets the detection mode to Pos&Neg, Positive, Sample, Negative, or RMS.
3.1.4 Setting wave detection mode
3
Display Mode Selection
3-3
Page 52
Chapter 3 Display Mode Selection

3.1.1 Storage mode types

The following storage modes can be selected for trace A, trace B, trace C, trace D, trace E, and trace F.
Table 3.1.1-1 Storage Mode Types
Mode Display Description Display Example
In the Log display mode,
Lin Average
Max Hold
Lin
Max
averaging is executed for linear values, and the result is displayed logarithmically.
At each sweep, the previous and new trace data of each horizontal axis point are compared and the larger of the two is displayed.
Min Hold
Average
Off
Min
Avg
Off
At each sweep, the previous and new trace data of each horizontal axis point are compared and the smaller of the two is displayed.
At each sweep, averaging calculations are done for each horizontal axis point, and the results are displayed. These results are used for improvement of the S/N ratio.
At each sweep, the trace data are updated and displayed. These data are used for normal measurement.
For storage mode, see Figure 3.1.4-2. It is displayed in Trace Parameter.
3-4
Page 53

3.1.2 Averaging function

This function displays the averaged trace results.
Example: To set trace A to Average and the storage count to 100
<Procedure>
1. Press
3.1 Setting Storage Mode
.
2. Press
3. Press
4. Press
5. After pressing
Table 3.1.2-1 shows the calculation method for averaging.
Sweep
Count
1 M (1)
2 M (2)
3 M (3)
… …
N  1 M (N  1)
Measurement
(Active Trace) and then press (A).
(Storage Mode) and select (Average).
(Storage Count).
, press (Set) to set the storage count.
Table 3.1.2-1 When averaging exponent = N
Value
M (n)
Displayed Value
Y (1) = M (1)
Y (2)
Y (3)
)(1Y
=
(2)Y
=
2)(NY)1(NY
)(1Y)(2M
2
3
(2)Y(3)M
3
Display Mode Selection
)2(NY)1(NM
1N
1)(NY(N)M
)1(NY(N)Y
N M (N)
N + 1 M (N + 1)
N + 2 M (N + 2)
For storage times, see Figure 3.1.4-2. It is displayed in Trace Parameter.
Display except for Spectrum Emission Mask
Display in % for Spectrum Emission Mask
(N)Y1)(NY
1)(NY2)(NY
N
N
)(NY1)(NM
N
1)(NY2)(NM
3-5
Page 54
Chapter 3 Display Mode Selection

3.1.3 Max Hold/Min Hold function

This function holds and displays the maximum value (Max Hold) or minimum value (Min Hold) of the trace data.
Example: To set trace A to Max Hold, trace B to Min Hold, and simultaneous
display
<Procedure>
1. Press .
2. Press
3. Press
4. Press
5. Press
6. Press
7. Press
(Active Trace) and then press (A).
(Trace Type) and then press (Write).
(Storage Mode) and then press (Max Hold).
(Active Trace) and then press (B).
(Trace Type) and then press (Write).
(Storage Mode) and then press (Min Hold).
3-6
Page 55

3.1.4 Setting wave detection mode

The wave detection mode can be selected from the modes listed in Table
3.1.4-1. When Couple Time/Frequency Domain is set to Off, each
parameter is specified in accordance with the already-specified frequency axis or time axis measurement.
Table 3.1.4-1 Wave detection modes
Wave
Detection
Mode
Pos&Neg
Positive Displays maximum value between sample points. Negative Displays minimum value between sample points. Sample Displays instantaneous value between sample points.
RMS
Displays line linking maximum value and minimum value between sample points.
Displays power of 2 average value (effective value) between sample points.
3.1 Setting Storage Mode
Description
3
Display Mode Selection
Positive
Negative
Sample
Sample points Screen display
Pos & Neg
Screen display
Screen display
Sample points
Figure 3.1.4-1 Screen display differences according to wave detection mode
3-7
Page 56
Chapter 3 Display Mode Selection
Example: To set the wave detection mode to Negative
<Procedure>
1. Press .
2. Press
(Detection) and then press (Negative).
Wave detection mode
Trace Parameter
Figure 3.1.4-2 Wave detection mode selection
3-8
Page 57

3.1.5 Displaying and updating Trace

You can configure Trace Type display settings here. Table 3.1.5-1 lists the available options.
Table 3.1.5-1 Trace Type Writing Modes
3.1 Setting Storage Mode
Trace Type
Writing Modes
Write
View
Blank No trace is displayed.
For trace type writing mode, see Figure 3.1.4-2. It is displayed in Trace Parameter.
Display Description
W
Updates the display per measurement.
V
Keeps the current display; it does not update the display per measurement.
3
Display Mode Selection
3-9
Page 58
Chapter 3 Display Mode Selection

3.2 Setting Limit Line

You can evaluate trace data by drawing Limit Line.
Notes:
The limit line function is not available when Measure function
is set to ON.
Limit Test (evaluation of trace data) is not performed when, on
the Amplitude function menu, the Scale menu is set to Lin. The result of Limit Test is displayed as “*** LIMITn”.
[1]
[2]
Figure 3.2-1 Display items for the Limit function
Table 3.2-1 Display items for the Limit function
No. Display Description
Displays a Pass/Fail judgment by
[1]
Limit Test
[2]
Limit Line Displays Limit Line.
determining whether or not data is within Limit Line.
3-10
Page 59

3.2.1 Setting Limit function parameters

Pressing displays the Trace function menu. Selecting function menu displays the Limit function menu.
Example: To draw –20 dBm Limit Line between 1 and 2 GHz in order to
<Procedure>
1. Press .
2. Press (Limits).
(Trace) in the main function menu or pressing
determine data as Fail when it exceeds Limit Line
3.2 Setting Limit Line
(Limits) on the Trace
3
3. Press (Limit1).
4. Press to display page 2. Press (Limit Line Type (Frequency)) and select Abs.
5. Press
6. Press
7. Press
8. Press (Limit Test) to set it to On.
9. Press (Edit) and then press (Point).
10. Press to set Point to Point1.
11. When (Frequency) is not displayed, press (Add Point).
12. Press
13. After pressing
14. Press (Amplitude).
15. Press and then press (dBm) to set Amplitude.
16. Press (Add Point).
17. Press (Frequency).
18. After pressing
(Limit Line Type (Amplitude)) and select Abs.
to display page 1. Press (Type) to set it to Upper.
(Limit Display) to set it to On.
(Frequency).
, press (GHz) to set Frequency.
, press (GHz) to set Frequency.
Display Mode Selection
19. Press (Amplitude).
20. Press
21. Press Point1 and Point2 with a line.
and then press (dBm) to set Amplitude.
(Connected to Previous Pt) and set it to On to connect
3-11
Page 60
Chapter 3 Display Mode Selection

3.2.2 Setting Limits

Selecting (Limits) on the Trace function menu displays the Limits function menu. You can set six types of Limit Line.
Table 3.2.2-1 Limits function menu
Function
key
F1 Limit1
F2 Limit2
F3 Limit3
F4 Limit4
F5 Limit5
F6 Limit6
F8
Menu Display Function
Sets Limit Line 1.
Sets Limit Line 2.
Sets Limit Line 3
Sets Limit Line 4
Sets Limit Line 5
Sets Limit Line 6
Delete All Limits
Deletes all Point values of Limit Line 1 to 6.
3.2.3 “Setting Limit1 to 6”
3.2.3 “Setting Limit1 to 6”
3.2.3 “Setting Limit1 to 6”
3.2.3 “Setting Limit1 to 6”
3.2.3 “Setting Limit1 to 6”
3.2.3 “Setting Limit1 to 6”
3-12
Page 61

3.2.3 Setting Limit1 to 6

Selecting displays the Limit1 to Limit6 function menu.
3.2 Setting Limit Line
(
Limit1) to (Limit6) on the Limits function menu
Table 3.2.3-1 Limit1 to 6 function menus
Function
key
Page 1 Limit1 to Limit6
F1 Type
F2 Limit Display
F3 Limit Test
F4 Margin (On/Off)
F5 Margin Value
F6 Edit
F7 Envelope
F8 Test Trace*
Menu Display Function
*: When one of the Limit1 to Limit6 function menus is selected, the
setting of the Active Trace (F1) on the Trace function menu is set automatically.
Press menu.
Sets the Limit Line type. Upper: Upper limit (Default)
Lower: Lower limit Sets whether to show or hide Limit Line. On: Display
Off: Hide (Default) Sets whether or not to make a judgment by Limit
Line. On: Makes a judgment.
Off: Does not make a judgment. (Default) Sets whether to display the Margin line. This line is
determined by adding Offset to the Limit Line. On: Display
Off: Hide (Default) Sets the Offset value to be added with the Margin
line. Range: –40.00 to 0.00 (Upper)
0.00 to 40.00 (Lower) Resolution: 0.01 dB Rotary knob resolution: 0.1 dB Step key resolution: 1 dB Default: 0 dB
Sets individual Points of Limit Line.
Displays the Envelope-Limit x function menu.
Sets the test target Trace. Options: A,B,C,D,E,F
Default: A
(Limit1) to (Limit6) on the Limits function
3.2.4 “Edit Parameter”
3.2.5 “Limit Envelope Function”
3
Display Mode Selection
3-13
Page 62
Chapter 3 Display Mode Selection
Table 3.2.3-1 Limit1 to 6 function menus (Cont’d)
Function
key
Page 2 Limit1 to Limit6
F1
F2
F4 Save Limit x
F5 Recall Limit x
F7 Mirror Limit
Menu Display Function
Limit Line Type (Frequency)
Limit Line Type (Amplitude)
Press menu and then press to
Sets the mode of setting the frequency axis of the Limit Line.
Abs: Absolute value of the frequency Rel: Relative value with respect to the Center Frequency (Default)
Sets the mode of setting the level axis of the Limit
Abs: Absolute level (dBm) (Default) Rel: Relative value with respect to the Reference
Level (dB) Displays the Save Limit function menu.
Displays the Recall Limit1 to 6 function menu.
Sets whether to mirror (On) or not (Off: Default) the settings for the right half to the left half. This menu is available only when the Limit Line Type (Frequency) menu is set to Rel.
(Limit1) to (Limit6) on the Limits function
display page 2.
Line.
Table 3.2.6-2 Save Limit function menu
Table 3.2.7-3 Recall Limit1 to 6 function menu
(1) Setting a display format of Limit Line
Example: To display the judgment by Limit Line by displaying Limit1 as Upper
Limit and displaying Limit Line
<Procedure>
1. Press .
2. Press
3. Press
4. Press
5. Press
6. Press
Details for Limit Test
When Type is Upper and data exceeds Limit Line, the test result will
be Fail.
When Type is Lower and data is below Limit Line, the test result will
be Fail.
(2) Margin function
(Limits).
(Limit1).
(Type) to set it to Upper.
(Limit Display) to set it to On.
(Limit Test) to set it to On.
3-14
Example: To display a Margin line by adding the Limit Line with –20 dB
<Procedure>
Page 63
1. Press .
3.2 Setting Limit Line
2. Press
3. Press
4. Press
5. Press
6. Press
(Limits).
(Limit1).
(Margin) to set it to On.
(Margin Value).
and then press (dB) to set Margin Value.
3
Display Mode Selection
3-15
Page 64
Chapter 3 Display Mode Selection

3.2.4 Edit parameters

Selecting (Edit) on the Limit1 to 6 function menu displays the Edit-Limit x function menu.
Table 3.2.4-1 Edit-Limit x function menu
Function
key
F1 Point Sets the Limit Point you want to edit.
F2 Add Point
F3 Frequency* Sets the Frequency of Limit Point. F4 Amplitude* Sets the level of Limit Point.
F5
F6
F7 Delete Point Deletes the selected Limit Point. F8 Delete Limit Deletes all Limit Points of Limit x.
Menu Display Function
Adds a Limit Point with the same frequency and level as the selected Limit Point, on the right of the selected Limit Point.
Sets whether to turn on or off linear interpolation
Connected to Previous Pt(On/Off)*
Previous Pt Level Offset*
*: The F3 to F6 function menus are not displayed, when the number of
Limit Points is 0.
between Pts (Limit Points). Pass/Fail evaluation is performed by making linear interpolation between the current Limit Point selected and the other Limit Point that is located on the nearest left of the current Limit Point.
Sets the offset value from the current Limit point for connecting Pts (Limit Points) with a line.
3-16
Page 65
3.2 Setting Limit Line
(1) Setting Point
Point Setting range: 1 to Limit Points (Max: 100) Minimum resolution: 1 Rotary knob resolution: 1 Step key resolution: 1
(2) Setting Frequency
Example: To set Frequency to 1 GHz
<Procedure>
1. Press .
2. Press
3. Press
4. Press
5. Press
6. After pressing
Frequency When the Limit Line Type (Frequency) menu is set to Abs.
Setting range: 0 to 325 GHz Default: Start Frequency Minimum resolution: 1 Hz Rotary knob resolution: 1/10000 of the frequency span Step key resolution: Step Size
When the Limit Line Type (Frequency) menu is set to Rel.
Setting range: –100 to 100 GHz Default: 0 Hz Minimum resolution: 1 Hz Rotary knob resolution: 1/10000 of the frequency span Step key resolution: Step Size
Notes:
When a point's Frequency value is higher than the other points'
value and this relationship reverses, the order of points automatically changes.
(Limits).
(Limit1).
(Edit).
(Frequency Offset).
, press (GHz) to set Frequency Offset.
2.3.2 “Setting frequency span”
2.3.8 “Setting step size”
3
Display Mode Selection
When the Limit Line Type (Frequency) is switched, a value
exceeding the setting range may be read.
3-17
Page 66
Chapter 3 Display Mode Selection
(3) Setting Amplitude
Example: To set Amplitude to –10 dBm
<Procedure>
1. Press .
2. Press
3. Press
4. Press
5. Press
6. Press
Amplitude When Limit Line Type (Amplitude) is Abs.
Setting range: –300 to 300 dBm Default: Reference Level Minimum resolution: 0.01 dBm Rotary knob resolution: 1 dBm Step key resolution: 10 dBm
When Limit Line Type (Amplitude) is Rel:
Setting range: –300 to 300 dB Default: 0 dB Minimum resolution: 0.01 dB Rotary knob resolution: 1 dB Step key resolution: 10 dB
(Limits).
(Limit1).
(Edit).
(Amplitude).
and then press (dBm) to set Amplitude.
3-18
Note:
When the Limit Line Type (Amplitude) is switched, a value exceeding the setting range may be read.
Page 67
3.2 Setting Limit Line
(4) Connected to Previous Pt
Pass/Fail evaluation between Limit Points is performed by making
linear interpolation between the current Limit Point and the other Limit Point that is located on the nearest left of the current Limit Point. Also, linear interpolation can be made with a margin added to.
Linear Interpolation: On
Current Limit Point
(5) Previous Pt Level Offset
Linear interpolation between Limit Points is performed with respect
to “previous Limit Point” and “Offset Point (○) from the current Limit Point”. Limit evaluation will be made according to the Limit Points (●), without including the Offset Point (○).
Linear Interpolation: Off
Current Limit Point
Linear Interpolation With Offset of 3 dBLinear Interpolation Without Offset
3
Display Mode Selection
Current Limit Point
Previous Pt Level Offset
Setting range: –300.00 to 300.00 dB Default: 0 dB Minimum resolution: 0.01 dB Rotary knob resolution: 1 dB Step key resolution: 10 dB
Offset Point
3 dB
Current Limit Point
3-19
Page 68
Chapter 3 Display Mode Selection
(6) Creating a new Limit Line
<Procedure>
1. Press (Edit).
2. Press
3. Set the Frequency.
4. Set the Amplitude.
5. Press
6. Set the Previous Pt Level Offset.
7. Repeat steps 3 to 6 until the Limit Points are set.
(7) Adding a new Limit Point to an existing Limit Line
<Procedure>
1. Press (Edit).
2. Press
3. Rotate the rotary knob until the point to add a Limit Point is selected.
4. Press
5. Press
6. Press
7. Press
(Add Point).
(Connected to Previous Pt) and set to On/Off.
(Point).
(Add Point).
(Frequency) and set the frequency.
(Amplitude) and set the level.
(Connected to Previous Pt) and set to On.
8. Press
(Previous Pt Level Offset) and set the offset.
3-20
Page 69

3.2.5 Limit Envelope Function

Selecting (Envelope) on the Limit1 to 6 function menu displays the Envelope-Limit x function menu.
3.2 Setting Limit Line
Set the values of the following: (Shape). Then, clicking Limit Point(s) and Limit Line from the Trace data of the measurement results.
Note:
Pressing (Create Envelope) or (Update Envelope Amplitude) clears all Limit Points set in 3.2.4 “Edit parameters”. Then, Limit Points are arranged at regular intervals in the on-screen span, according to the number and shape that are respectively set at
Table 3.2.5-1 Envelope-Limit x function menu
Function
key
F1
F2
F4
F5
F6
Menu Display
Create Envelope
Update Envelope Amplitude
Points
Offset
Shape
Automatically creates a Limit Line from the current Trace data.
This menu is unavailable when the Mirror Limit menu is set to On.
Automatically updates a Limit Point of the Peak Level that includes a Trace Point evaluated as “Fail”, referring to the current trace data.
This menu is unavailable when the Mirror Limit menu is set to On. This menu is unavailable after editing the Limit Line when Shape is set to Square.
Sets the number of Envelope Points to be arranged when automatically creating a Limit Line.
When the Create Envelope menu is executed, the set value applies.
Sets the offset level (from the current peak Trace Point to the Limit Point).
When the Create Envelope or Update Envelope menu is executed, the set value applies.
Sets the shape of the Limit Line. When the Create Envelope menu is executed, the set value
applies. This menu is unavailable when the Mirror Limit menu is set to On.
Slope: Connects Limit Points with straight lines.
Square: Connects Limit Points with vertical and
(Create Envelope) automatically creates
(Default)
horizontal lines.
(Points), (Offset) and
3
Display Mode Selection
(Points) and (Shape).
Function
3-21
Page 70
Chapter 3 Display Mode Selection
Create Envelope
Example: When executing the Create Envelope menu with the following
The following figure shows a Limit Line (—) and Limit Points () that are automatically created with respect to Trace Points (), peak values for Trace Point values (←→) and offset values (↑).
menu settings: Type: Upper, Trace Points: 15, Points: 3, Offset: 5 dB, Shape: Slope
Offset
5 dB
Limit Line
Limit Point
Offset
5 dB
Offset
5 dB
Trace Point Peak Value
Update Envelope Amplitude
Example: When a Trace Point () is evaluated as “Fail” by performing
the measurement with the following menu settings: Type: Upper, Trace Points: 15, Points: 3, Offset: 5 dB, Shape: Slope
Fail
5 dB
5 dB
5 dB
The level of the Limit Point including a Fail Trace Point is
recalculated by executing the Update Envelope Amplitude menu.
3-22
5 dB
5 dB
5 dB
Page 71
3.2 Setting Limit Line
Points
Setting range: When Shape is set to Slope:
2 to (Trace Points–2)×2+2, (Max: 100)
When Shape is set to Square:
2 to Trace Points, (Max: 100) Default: 41 Resolution: 1 (Slope), 2 (Square)
Shape
When Shape is set to Slope:
The number of Limit Points () is equal to that of Envelope Points set by by straight lines. Example: When the number of Envelope Points is 5
(Points). Limit Line (—) is created by connecting Limit Points (○)
3 dB
3 dB
3 dB
3 dB
3 dB
3
Display Mode Selection
When Shape is set to Square:
The following figure shows the number of Limit Points () that depends on the number of Envelope Points () set by (—) is created by connecting Envelope Points() by straight lines. Example:
When the number of Envelope Points is an odd number, the number of Limit Points is obtained by the following formula: (The number of Envelope Points+1)/2 When the number of Envelope Points is an even number, the number of Limit Points is determined by the following formula: (The number of Envelope Points)/2
3 dB
3 dB
(Points). Limit Line
3 dB
3-23
Page 72
Chapter 3 Display Mode Selection

3.2.6 Saving a Limit file

The Save function menu is displayed by pressing when the Spectrum Analyzer screen is active. For details on the functions, refer to Table 3.2.6-1.
Table 3.2.6-1 Save function menu
Function
key
F1 Device Sets a save destination drive.
F3 Save Limit
F4 Save on Event
F5
F7 Save Application
F8 Close Returns to the previous function menu.
Menu Display Function
Displays the Save Limit function menu.
Table 3.2.6-2 Save Limit function menu
Displays the Save on Event function menu where you can specify whether or not to save Trace data upon occurrence of an event.
This function is not available when Measure function is set to ON.
Table 3.4-1 Save on Event function menu
Saves the displayed waveform data (Trace).
Save Waveform CSV DATA
Saves the parameter setting conditions of all the launched applications.
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
Chapter 3 Common Operations
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
Chapter 3 Common Operations
3-24
Page 73
3.2 Setting Limit Line
On the Save Limit function menu, you can save settings of Limit Line to a file.
Table 3.2.6-2 Save Limit function menu
Function
key
F1 Limit1
F2 Limit2
F3 Limit3
F4 Limit4
F5 Limit5
F6 Limit6
Menu Display Function
Limit1 to Limit6
Pressing the function key displays the dialog box where you can save the Limit Line settings to a file with a specified name.
Save destination for Limit Line files
< D:¥Anritsu Corporation¥Signal Analyzer¥User Data¥Limit>
Limit Line file Name
Limit“Date”_“Sequential Number”.xml
Saves the settings for Limit Line1 to a file with a specified name.
Saves the settings for Limit Line2 to a file with a specified name.
Saves the settings for Limit Line3 to a file with a specified name.
Saves the settings for Limit Line4 to a file with a specified name.
Saves the settings for Limit Line5 to a file with a specified name.
Saves the settings for Limit Line6 to a file with a specified name.
(Sequential number from 000 to 999. Up to 1000 files can be saved in the folder.)
3
Display Mode Selection
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
Chapter 3 Common Operations
3-25
Page 74
Chapter 3 Display Mode Selection

3.2.7 Loading a Limit file

The Recall function menu is displayed by pressing when the Spectrum Analyzer screen is active. For details on the functions, refer to Table 3.2.7-1. On the Recall function menu, you can load and delete Limit Line files and delete Waveform files.
Table 3.2.7-1 Recall function menu
Function
key
F1 Device Specifies the drive where files are stored.
F3 Recall Limit
F4
F6
F7
F8 Close Returns to the previous function menu.
Menu Display Function
Displays the Recall Limit function menu.
Table 3.2.7-2 Recall Limit function menu
Recall Waveform File
Recall Current Application
Recall all Application
On the Recall Limit function menu, you can select a Limit Line you want to load or delete Limit Line file(s).
Displays the Recall Waveform File function menu.
Table 3.4.3-1 Recall Waveform File function
Displays the Parameter Save Data List. This function is used when recalling the parameter
setting conditions to only the target application.
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
Chapter 3 Common Operations
Displays the Parameter Save Data List. This function is used when recalling the parameter
setting conditions to all the loaded applications.
MS2690A/MS2691A/MS2692A Signal Analyzer
Operation Manual (Mainframe Operation)
Chapter 3 Common Operations
menu
3-26
Table 3.2.7-2 Recall Limit function menu
Function
key
F1 Limit1 Displays the Recall Limit1 function menu. F2 Limit2 Displays the Recall Limit2 function menu. F3 Limit3 Displays the Recall Limit3 function menu. F4 Limit4 Displays the Recall Limit4 function menu. F5 Limit5 Displays the Recall Limit5 function menu. F6 Limit6 Displays the Recall Limit6 function menu.
Menu Display Function
Page 75
3.2 Setting Limit Line
Load source directory of Limit Line file
< D:¥Anritsu Corporation¥Signal Analyzer¥User Data¥Limit> or < D:¥Anritsu Corporation¥Signal Analyzer¥User Data¥Parameter Setting>
Limit Line file formats that can be recalled
csv, lim, xml
On the Recall Limit1 to 6 function menu, you can select Limit Line file(s) you want to load or delete.
3
Table 3.2.7-3 Recall Limit1 to 6 function menu
Function
key
F1 Device
F3 Delete Files
F8 Recall
Function
key
F3 Select All Selects all files. F7 Delete Deletes the selected file(s).
F8 Cancel
Menu Display Function
Selects a device that resides in the folder where the target file is stored.
Displays the Delete Files function menu.
Table 3.2.7-4 Delete Files function menu
Displays the Recall function menu.
Table 3.2.7-5 Recall function menu
On the Delete Files function menu, you can select and delete Limit Line file(s).
Table 3.2.7-4 Delete Files function menu
Menu Display Function
Deselects the file(s) and returns to the Recall Limit1 to 6 function menu.
On the Recall function menu, you can load the selected Limit Line file.
Display Mode Selection
Table 3.2.7-5 Recall function menu
Function
key
F7 Set Loads the selected file.
F8 Cancel
Menu Display Function
Deselects the file and returns to the Recall Limit1 to 6 function menu.
3-27
Page 76
Chapter 3 Display Mode Selection

3.3 Setting Time/Sweep

Pressing (Time/Sweep) on the main function menu, or pressing
displays the Time/Sweep function menu.
Figure 3.3-1 Time/Sweep key
3-28
Page 77
F1
F2
F3
F4
F5
F6
F7
F8
3.3 Setting Time/Sweep
3
Display Mode Selection
Figure 3.3-2 Time/Sweep function menu
Table 3.3-1 Time/Sweep function menu
Menu Display Function
Sweep Time (Auto/Manual)
Sweep Time
Auto Sweep Time Select (Normal/Fast)
Trace Points
Sets the Auto mode/Manual mode for the sweep time.
Sets the sweep time.
Sets the Normal mode/Fast mode when the sweep time is set to Auto.
Sets the number of trace points.
3.3.1 “Setting sweep time”
3.3.1 “Setting sweep time”
3.3.1 “Setting sweep time”
3.3.2 “Setting trace point”
3-29
Page 78
Chapter 3 Display Mode Selection

3.3.1 Setting sweep time

Set the sweep time. When Auto is set, the optimum value is set automatically. Each parameter is specified in accordance with the already-specified frequency axis or time axis measurement.
Setting range and resolution for Sweep Time
Setting range:
Resolution
1 ms (2 to 999 ms)
0.1 s (1 to 9.9 s) 1 s (10 to 1000 s)
2 ms to 1000 s (in frequency domain) 1 s to 1000 s (in time domain) In Frequency Domain
1
0.1 ms (1 to 9.9 ms) 1 ms (1 to 999 ms)
0.1 s (1 to 9.9 s) 1 s (1 to 1000 s)
Example: To set Sweep Time value to 20 s
<Procedure>
1. Press .
2. Press
3. After pressing
When the sweep time is set to Auto, Normal mode (normal sweep mode (fast sweep) can be set.
Example: To set Fast mode by setting Sweep Time to Auto
<Procedure>
1. Press .
(Sweep Time).
In Time Domain
s (1 to 999 s)
, press (s) to set the Sweep Time.
)/
Fast
3-30
2. Press
3. Press mode.
(Sweep Time) and select Auto.
(Auto Sweep Time Select) and select Fast to set the Fast
Page 79

3.3.2 Setting trace point

Trace point refers to the number of waveform data points. When Couple Time/Frequency Domain is set to Off, each parameter is specified in accordance with the already-specified frequency axis or time axis measurement. The following numbers of trace points can be selected.
3.3 Setting Time/Sweep
Table 3.3.2-1 Trace Point Setting Range
Condition Range
Span > 500 MHz 1001 to 30001 100 MHz < Span 500 MHz 101 to 30001 300 Hz Span 100 MHz
and Sweep Time > 10 s 300 Hz Span 100 MHz
and Sweep Time 10 s Span = 0 Hz and Sweep Time > 10 s 101 to 30001 Span = 0 Hz and Sweep Time 10 s 11 to 30001
Resolution: 1 point
Example: To set the number of trace points to 2001.
<Procedure>
1. Press
2. Press
3. Press of trace points.
.
(Trace Points).
and then press (Set) to set the number
101 to 30001
11 to 30001
3
Display Mode Selection
3-31
Page 80
Chapter 3 Display Mode Selection

3.4 Save on Event

The Save on Event function menu is displayed by pressing and (Save on Event) when the Spectrum Analyzer screen is active. For details on the functions, refer to Table 3.4.1-1.

3.4.1 Setting Save on Event

The Save on Event function saves a waveform upon occurrence of any of the trigger events set by operations can be performed.
(Event Type). Continuous file save
Notes:
Function
key
F1
F2 Event Type
F3
F5
Menu Display
Save on Event
Save then Stop
File Name
When the Storage Mode is set to other than Off, trace data is
saved upon occurrence of one of the events specified by the Event Type menu after the number of storages reaches the Storage Count.
Save operation stops when the free HDD space falls below 1 MB.
Table 3.4.1-1 Save on Event function menu
Sets whether to turn on or off the Save on Event function. On: Turns on the function.
Off: Turns off the function. (Default) Sets events that trigger a file save operation. Limit Fail: Saves a waveform if a Limit evaluation
Limit Pass: Saves a waveform if a Limit evaluation
Margin Fail: Saves a waveform if an evaluation
Margin Pass: Saves a waveform if an evaluation
Sweep Complete: Saves a waveform whenever
Sets whether to perform the Save on Event function only once or perform it continuously.
On: Performs only once. Off: Continuously performs. (Default)
Sets the Filename that is internal data of a Folder.csv file to be saved.
Default: LIM
Function
result is Fail. (Default)
result is Pass.
result including Margin is Fail.
result including Margin is Pass.
measurement is performed, regardless of the Limit evaluation result.
3-32
Page 81
3.4 Save on Event
File Name
This menu is used to set the Filename (underlined part of the following) that is internal data of a Waveform file to be saved as Folder.csv when performing the Save on Event function.
Filename,Date,999.500001 MHz,999.501819 MHz,…,1000.500001 MHz LIM20140416163123000_R.spa, 4/16/2014 4:31:23 PM, -98.292000, -98.268000,…,-103.104000 LIM20140416163134000_R.spa,4/16/2014 4:31:34 PM,-102.664000,-102.688000,…,-106.108000
If omitted, the following is automatically set. When Event Type is other than Sweep Complete: LIM When Event Type is Sweep Complete: EOS
3
Display Mode Selection
3-33
Page 82
Chapter 3 Display Mode Selection

3.4.2 Waveform files for Save on Event

This section describes Waveform files to be saved when the Save on Event function is turned on.
Save destination and load source for Waveform files
After the parameters are set, a folder is created by the name including a timestamp of measurement start. Then, a csv file containing multiple data sets is created and stored in the folder.
Save destination for Waveform files
Example: When starting measurement on April 16 at 15:29, a file save
destination is as follows:
< D:¥Anritsu Corporation¥Signal Analyzer¥User Data¥Waveform¥04161529_1>
The number of the end of "04161529_1" is incremented automatically in order of "04161529_1", "04161529_2", and "04161529_3." It will be set to "04161530_1" if it will become 15:30.
Waveform file name
Folder.csv
Waveform file format
Example: Event Type: Other than Sweep Complete, Detection: RMS Filename,Date,999.500001 MHz,999.501819 MHz,…,1000.500001 MHz LIM20140416163123000_R.spa, 4/16/2014 4:31:23 PM, -98.292000, -98.268000,…,-103.104000 LIM20140416163134000_R.spa,4/16/2014 4:31:34 PM,-102.664000,-102.688000,…,-106.108000
LIM20140416164324000_R.spa,4/16/2014 4:43:24 PM,-101.460000,-103.128000,…,-106.500000
Example: Event Type: Other than Sweep Complete, Detection: Pos&Neg Filename, Date, 999.500001 MHz, 999.501819 MHz,…, 1000.500001 MHz LIM20140416163123000_P.spa,4/16/2014 4:31:23 PM,-98.292000,-98.268000,…,-103.104000 LIM20140416163123000_N.spa,4/16/2014 4:31:23 PM,-98.294000,-98.269000,…,-103.105000 LIM20140416163134000_P.spa,4/16/2014 4:31:34 PM,-102.664000,-102.688000,…,-106.108000 LIM20140416163134000_N.spa,4/16/2014 4:31:34 PM,-102.666000,-102.689000,…,-106.109000
LIM20140416164324000_P.spa, 4/16/2014 4:43:24 PM, -101.460000, -103.128000,…,-106.500000 LIM20140416164324000_N.spa,4/16/2014 4:43:24 PM,-101.480000, -103.129000,…,-106.700000
Example: Event Type: Sweep Complete, Detection: RMS Filename,Date,999.500001 MHz,999.501819 MHz,…,1000.500001 MHz EOS20140416163123000_R.spa,4/16/2014 4:31:23 PM,-98.292000,-98.268000,…,-103.104000 EOS20140416163134000_R.spa,4/16/2014 4:31:34 PM,-102.664000,-102.688000,…,-106.108000
EOS20140416164324000_R.spa,4/16/2014 4:43:24 PM,-101.460000,-103.128000,…,-106.500000
3-34
Page 83
Meaning of description in file format
_P: Positive detection, _N: Negative detection, _R: RMS detection, _S: Sample detection

3.4.3 Deleting Waveform files for Save on Event

The Recall function menu is displayed by pressing when the Spectrum Analyzer screen is active. The table 3.4.3-1 Recall Waveform File function menu is displayed by pressing in Recall function menu.
On the Recall Waveform File function menu, you can select Save on Event Waveform files you want to delete.
Table 3.4.3-1 Recall Waveform File function menu
3.4 Save on Event
(Recall Waveform File)
3
Display Mode Selection
Function
key
F1 Device
F3 Delete Files
Function
key
F3 Select All Selects all files. F7 Delete Deletes the selected file(s).
F8 Cancel
Menu Display Function
Selects a device that resides in the folder where the target file is stored.
Displays the list of Waveform files stored and displays the Delete Files function menu.
Table 3.4.3-2 Delete Files function menu
On the Delete Files function menu, you can select and delete Waveform file(s).
Table 3.4.3-2 Delete Files function menu
Menu Display Function
Deselects the file(s) and returns to the Recall Waveform File function menu.
3-35
Page 84
Chapter 3 Display Mode Selection
3-36.
Page 85

Chapter 4 Marker Function

This chapter describes the marker functions.
Setting Marker ............................................................... 4-2
4.1
4.1.1 Setting normal marker ...................................... 4-5
4.1.2 Setting delta marker ......................................... 4-5
4.1.3 Setting zone marker ......................................... 4-6
4.1.4 Setting marker trace ......................................... 4-7
4.1.5 Setting active marker ........................................ 4-8
4.1.6 Setting Fixed marker ........................................ 4-8
4.1.7 Setting Type of Result Display ......................... 4-9
4.1.8 Setting Freq. Count ........................................ 4-11
4.2 Setting Peak Search Function .................................... 4-13
4.2.1 Executing Peak search ................................... 4-17
4.2.2 Executing Next Peak search .......................... 4-17
4.2.3 Setting search resolution ................................ 4-17
4.2.4 Setting search threshold value ....................... 4-18
4.2.5 Executing Marker to Center Freq./
Marker to Ref. Level ....................................... 4-19
4.2.6 Executing Marker Search Function ................ 4-20
4.2.7 Executing Power Peak search ....................... 4-21
4.2.8 Executing Next Power Peak search ............... 4-21
4.2.9 Executing Min search ..................................... 4-22
4.2.10 Executing Next Min search ............................. 4-22
4.2.11 Marker tracking settings ................................. 4-22
4.3 Displaying Marker List ................................................ 4-23
4
Marker Function
4-1
Page 86
Chapter 4 Marker Function

4.1 Setting Marker

Pressing (Marker) on the main function menu, or pressing displays the Marker function menu.
The Marker function menu consists of two pages, which can be toggled by pressing
.
F1
F2
F3
F4
F5
F6
F7
F8
Figure 4.1-1 Marker key
Page1 Page2
Pag e3
4-2
Figure 4.1-2 Marker function menu
Page 87
4.1 Setting Marker
Table 4.1-1 Marker function menu
Menu Function
Active Marker
Normal
Delta
Fixed
Off
Zone Width
Relative To
Next Peak
Marker List (On/Off)
Marker Result
Marker Trace
Marker to Center Freq.
Marker to Ref. Level
All Marker Off
Marker Tracking
Sets the active marker.
4.1.5 “Setting active marker”
Sets the marker mode to Normal.
4.1.1 “Setting normal marker”
Sets the marker mode to Delta.
4.1.2 “Setting delta marker”
Sets the marker mode to Fixed
4.1.6 “Setting fixed marker”
Sets the marker display to Off.
4.1.1 “Setting normal marker”
Opens the Zone Width function menu. Set the frequency width for the zone marker.
4.1.3 “Setting zone marker”
Sets the reference marker when the active marker is set to Delta.
4.1.2 “Setting delta marker”
Searches for the second biggest peak after the active marker, and moves the marker so that the marker becomes the center frequency of the zone marker.
4.2.2 “Setting Next Peak search”
Sets the marker list display to On/Off.
4.3 “Displaying Marker List”
Opens the Marker Result function menu. Set the display type of the marker value.
4.1.7 “Setting Type of Result Display”
Selects the trace to display the marker.
4.1.4 “Setting marker trace”
Sets the presently set marker frequency to the center frequency of the trace.
4.2.5 “Executing Marker to Center Freq.
Sets the presently set marker level to the reference level of the trace.
4.2.5 “Executing Marker to Center Freq.
Sets all the markers to Off.
4.1.1 “Setting normal marker”
Conducts sweep after adjusting the center frequency of trace to that of the active marker frequency.
4.2.11 “Marker tracking settings”
.
/Marker to Ref. Level”
/Marker to Ref. Level”
4
Marker Function
4-3
Page 88
Chapter 4 Marker Function
Zone Width function menu
Pressing (Zone Width) on the Marker function menu displays the Zone Width function menu.
F1
F2
F3
F4
F5
F6
F7
F8
Menu Function
Type (Zone/Spot)
Zone Width
Couple Zone (On/Off)
Spot Line (On/Off)
Figure 4.1-3 Zone Width function menu
Table 4.1-2 Zone Width function menu
Switches between the spot marker and the zone marker.
4.1.3 “Setting zone marker”
Sets the zone marker width of the frequency domain.
4.1.3 “Setting zone marker”
Sets the Zone Width common setting On/Off. When enabled, the Type and Zone Width settings are shared among all markers. When disabled, a unique setting is possible for each marker.
4.1.3 “Setting zone marker”
Sets the line display of the spot marker to On/Off.
4.1.3 “Setting zone marker”
4-4
Page 89

4.1.1 Setting normal marker

Normal markers are displayed as ▼ and display the frequency (time) and level on the screen.
<Procedure>
1. Press
4.1 Setting Marker
.
2. Press
3. Press
Example: To set all the markers to Off.
<Procedure>
1. Press
2. Press and then press

4.1.2 Setting delta marker

The frequency and level at the marker are displayed as values relative to the reference point (marker set by Relative To). If the marker that is set by Relative To is set to Off when this function is executed, the marker is set to a Fixed marker and is displayed as  on the trace. After that, the relative values of Current marker () and Fixed marker () are displayed.
<Procedure>
1. Press
(Normal) to display the normal markers.
(Off) to stop displaying the markers.
.
to display the second page of the Marker function menu
(ALL Marker Off) to set all the markers to Off.
.
4
Marker Function
2. Press
Example: To set Relative To of marker 1 to marker 3.
<Procedure>
1. Press
2. Press active marker to 1.
3. Press Relative To to 3.
(Delta) and set delta marker.
.
(Active Marker) and then press (Marker1) to set the
(Relative To) and then press (Marker3) to set
4-5
Page 90
Chapter 4 Marker Function

4.1.3 Setting zone marker

The area enclosed by the dashed lines in Figure 4.1.3-1 is called a zone marker. The current marker exists within this zone marker and moves to the maximum level point within the range indicated by the zone marker. The frequency (or time) and level at the current marker are displayed on the top part of the screen.
The zone marker can be moved by pressing rotary knob or step key. The center frequency of the zone marker can also be directly specified from the numeric keypad.
In the frequency domain mode, the zone marker width can be set to an arbitrary value. Spot marker without width can be also set. In the time domain mode, only spot markers can be set.
then operating the
Current Marker
Zone Marker
4-6
Zone Width
Figure 4.1.3-1 Zone marker and current marker
Zone Center Frequency
Page 91
Example: To set the zone marker width to 10 MHz
<Procedure>
1. Press
.
4.1 Setting Marker
2. Press
3. After pressing width.
Example: To set the spot marker.
<Procedure>
1. Press
2. Press
3. Press point of the center frequency of the zone marker.
Note:
When Marker Result (see 4.1.7) is set to Integration or Density, Type is fixed to Zone, Spot cannot be selected.
Example: To set the line display of the spot marker to Off.
1. Press
2. Press
3. Press
Example: To set Couple Zone to On.
1. Press
(Zone Width) to open the Zone Width function menu.
, press (MHz) and set the zone marker
.
(Zone Width) to open the Zone Width function menu.
(Type) to select Spot. A spot marker will be set at the
.
(Zone Width) to open the Zone Width function menu.
(Spot Line) to set Off.
.
4
Marker Function
2. Press
3. Press lump sum.

4.1.4 Setting marker trace

The Spectrum Analyzer function can display multiple traces simultaneously. The traces for marker display can be selected by selecting a marker trace.
Example: To set the marker to trace C
<Procedure>
1. Press
2. Press
(Zone Width) to open the Zone Width function menu.
(Couple Zone) and select On to set the marker widths in a
.
(Marker Trace) and select C.
4-7
Page 92
Chapter 4 Marker Function

4.1.5 Setting active marker

Select an active marker. You can set the position of the active marker by using the rotary knob and step key.
Example: To set the active marker to 2.
<Procedure>
1. Press
.
2. Press
3. Press

4.1.6 Setting Fixed marker

The Fixed marker is displayed as  on the screen. It is fixed on the screen and has a fixed value.
Example: To set marker 3 to Fixed marker.
<Procedure>
1. Press
2. Press marker to 3.
3. Press
(Active Marker).
(Marker 2) to set the active marker to 2.
.
(Active Marker) and then press to set the active
(Fixed) to set to the Fixed marker.
4-8
Page 93

4.1.7 Setting Type of Result Display

The Marker Result function is used to select a type of a marker result.
4.1 Setting Marker
Marker Result function menu
Press (Marker Result) on page 2 of the Marker function menu to display the Marker Result function menu.
F1
F2
F3
F4
F5
F6
F7
F8
4
Marker Function
Integration
Density
Peak
At frequency domain
Figure 4.1.7-1 Marker Result function menu
Table. 4.1.7-1 Marker Result function menu
Menu Function
Displays the total power in the zone band. It is displayed in the bandwidth of
power/marker. Displays the power per 1 Hz in the zone band. It is displayed in power/Hz units. Displays the peak power in the zone. Displays the peak power indicated by the
current marker.
At time domain
4-9
Page 94
Chapter 4 Marker Function
This function allows you to select the power (total power, total density, and peak power) displayed in the zone marker.
Example: To display the power per 1 Hz in the zone band.
<Procedure>
1. Press
.
2. Press press
3. Press
to move to page 2 of the Marker function menu, and then
(Marker Result).
(Density) to set to Density.
4-10
Figure 4.1.7-2 Displaying Power Density
Page 95

4.1.8 Setting Freq. Count

The Freq.Count function menu is used to set up frequency counter measurement.
Freq. Count function menu
Press Freq.Count function menu.
4.1 Setting Marker
(Freq.Count) in page 2 of Marker function menu to display the
Function
keys
F1
F2
F3
F4
F5
F6
F7
F8
Figure 4.1.8-1 Freq. Count function menu
Table 4.1.8-1 Freq. Count function menu
Menu Display Function
4
Marker Function
F1
F2 Gate Time
Freq. Count (On Off)
Sets whether to enable or disable frequency counter measurement.
Sets how long to perform frequency counter measurement.
4-11
Page 96
Chapter 4 Marker Function
Frequency counter
Figure 4.1.8-2 Frequency counter
4-12
Page 97

4.2 Setting Peak Search Function

Four types of peak search function are available, Peak, Next Peak, Min, and Next Min search.
Peak Search function menu
Pressing (Peak Search) on page 2 of the main function menu, or pressing
displays the Peak Search function menu.
4.2 Setting Peak Search Function
4
Marker Function
Figure 4.2-1 Peak Search key
F1
F2
F3
F4
F5
F6
F7
F8
Page 1
Figure 4.2-2 Peak Search function menu
Page 2
4-13
Page 98
Chapter 4 Marker Function
Table 4.2-1 Peak Search function menu
Menu Display Function
Moves the marker so that the maximum level
Peak Search
Next Peak
Marker Search Function
Resolution
Threshold
Marker to Center Freq.
Marker to Ref. Level
Power Peak Search
Next Power Peak
Min Search
Next Min
point in the measurement band becomes the center frequency of the zone marker.
4.2.1 “Executing Peak search”
Searches for the next peak of the active marker and moves the marker so that it becomes the center frequency of the zone marker.
4.2.2 “Setting next Peak search”
Opens the Marker Search function menu. Set the markers in the order of frequency or level .
4.2.6 “Executing Marker Search Function”
Sets the resolution for Next Peak and Next Min search.
4.2.3 “Setting search resolution”
Sets a threshold value to limit the level points to be searched for.
4.2.4 “Setting search threshold value”
Sets the active marker frequency to the center frequency.
4.2.5 “Executing Marker to Center
Sets the active marker level to the reference level.
4.2.5 “Executing Marker to Center
Moves the active marker to the position where the peak power of the zone width of the active marker becomes maximum in the measurement band.
4.2.7 “Executing Power Peak search”
Searches for the next largest peak power in the zone width compared to the total power of the zone width of the active marker in the measurement band and moves the active marker
4.2.8 “Executing Next Power Peak search”
Moves the marker so that the minimum level point in the measurement band becomes the center frequency of the zone marker.
Searches for the next dip of the active marker and moves the marker so that it becomes the center frequency of the zone marker.
4.2.10 “Executing Next Min search”
Freq./Marker to Ref. Level”
Freq./Marker to Ref. Level”
4.2.9 “Executing Min search”
4-14
Page 99
4.2 Setting Peak Search Function
Threshold function menu
Pressing (Threshold) on the Peak Search function menu displays the Threshold function menu.
F1
F2
F3
F4
F5
F6
F7
F8
Figure 4.2-3 Threshold function menu
Table 4.2-2 Threshold function menu
Menu Display Function
4
Marker Function
Threshold (On/Off)
Threshold (Above/Below)
Threshold Level
Switches on/off the detection threshold value function for Peak, Next Peak, Min, and Next Min search.
4.2.4 “Setting search threshold value”
Selects Above (upward search)/Below (downward search) for Peak, Next Peak, Min, and Next Min search.
4.2.4 “Setting search threshold value”
Sets the threshold value for Peak, Next Peak, Min, and Next Min search.
4.2.4 “Setting search threshold value”
4-15
Page 100
Chapter 4 Marker Function
Marker Search Function menu
Press (Marker Search Function) on the Peak Search function menu to display the Marker Search function menu.
F1
F2
F3
F4
F5
F6
F7
F8
Figure 4.2-4 Marker Search function menu
Table 4.2-3 Marker Search function menu
Menu Function
Search Peaks Sort Y
Search Peaks Sort X
Search Peaks Number
Resolution
Threshold
Sorts the markers by level in relation to the peaks on the trace of the number specified in Search Peaks Number.
4.2.6 “Executing Marker Search Function”
Sorts the markers by frequency (time) in relation to the peaks on the trace of the number specified in Search Peaks Number.
4.2.6 “Executing Marker Search Function”
Sets the number of searches when Search Peaks Sort Y/X is executed.
4.2.6 “Executing Marker Search Function”
Sets the resolution of the search.
4.2.3 “Setting search resolution”
Sets the threshold of the search.
4.2.4 “Setting search threshold value”
4-16
Loading...