GENERAL INFORMATION
1-10
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.3 Refresh Cycle (Continued)
temperature and takes an average of 20
hours to complete (refer to Figure 1-1).
Two ways to initiate a manual refresh cycle
are:
• Disconnect from AC, unplug battery, and
then press power switch. Lower LCD
will go blank. Reconnect battery and
then reconnect instrument to AC.
• Enter Diagnostics Mode, go to page D2,
Select Battery Status, and enter 0.0 Ah
for battery-rated capacity. Press Enter
and ok. Turn instrument off and back on,
and then reconnect to AC. Once the fan
turns on (indicating start of a refresh
cycle), return to page D2 and reset
battery rated capacity to 1.3 Ah.
1.6.4 Battery Gauge
The battery gauge provides an indication of
the approximate amount of charge
remaining on the battery. It will usually
indicate less charge than is actually
available with a new battery. This is
displayed in a bar graph format on the
lower LCD Display and is active as long as
the battery is connected to the instrument.
(Refer to Figure 1-2, Lower LCD Display
Layout.)
The battery gauge circuitry measures the
current flow into and out of the battery and
maintains a record of the state of charge of
the battery. This record is reset each time
the instrument completes a full refresh
cycle. The battery gauge uses this record,
together with a measurement of the present
power requirements of the instrument, to
estimate the charge available on battery
power at the current infusion rate.
NOTES:
• The instrument label and battery gauge
are always displayed, even when the
instrument is turned off; however, the
battery gauge does not represent the
battery charge remaining when the
instrument is turned off.
• To ensure a more accurate battery gauge
reading, review the battery gauge five
minutes after starting an infusion. The
gauge updates for each program change
while infusing. Battery run time may be
affected by the operating mode, rate,
monitoring options and back pressure.
1.6.5 Power On/Off
The Battery Manager provides the interface
between the power on/off switch(es) and
the main processor. When the instrument
is off, the Battery Manager interprets either
power switch as a turn on command and
applies power to the rest of the instrument,
informing the main processor which switch
was pressed. Once power is on, further
presses of a power switch are passed on to
the main processor which determines the
appropriate response under the existing
conditions. If the response is to turn the
power off, the main processor requests that
the Battery Manager remove power from
the rest of the instrument.
If an error has been detected which causes
the watchdog to be in alarm, pushing the
power switch (or for dual channel
instruments, either power switch)
immediately causes the power to be turned
off, without intervention by the Battery
Manager.
Signature Edition
®
GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual