NOTE: When the instrument is turned off, the gauge will
indicate approximate run time for infusions of 125 ml/hr.
NOTE: In the future, ALARIS Medical Systems may provide
different battery packs. The replacement battery may have a
different rated capacity. Therefore, the Battery Manager of
the instrument needs to know that a new battery has been
added and the rated capacity has changed. Refer to Section
5.3 “Replacing Battery” and Section 6.4.6, Changing Rated
Capacity of Battery.
NOTE: The restart counter is reset whenever the clinician
presses Run/Hold, if the pump or channel is turned off, or if
an alarm occurs.
NOTE: When infusion is started, the resistance may be other
than 0%, depending on solution viscosity, catheter/tubing
size and filters.
NOTE: The instrument’s configuration information is not lost
when disconnected from AC power and battery power.
However, error history and infusion program settings may
be lost. If you want to save this information be sure to
record before disconnecting power.
NOTE: It is strongly recommended that you review Table 2-1
for a complete list of defaults before selecting factory
defaults.
NOTE: Setting the Maximum Rate below the preset KVO
Rate will lower the KVO Rate. The KVO rate will not exceed
the Maximum Rate.
NOTE: The maximum rate setting applies to all infusion
modes.
NOTE: The KVO rate will not exceed the present Maximum
Rate.
NOTE: If Display is off then the alert feature will
automatically be turned off.
NOTE: Resistance trend graph data is lost when:
1) Clear is pressed, 2) Pump is off for more than 6 hours.
NOTE: Resistance measurement is restarted at 0% when: 1)
Run/Hold is pressed to put on hold and again to start, 2)
Dose ends in Multi-Dose mode, 3) Checking line message
appears.
NOTE: Checking line alert applies to pressure, resistance and
upstream occlusion with one tone at the beginning and a
flashing popup display (on for 4 seconds, off for 6 seconds).
No nurse call activation.
NOTE: Resistance alert will give an alarm tone every thirty
seconds, with a popup display (on for 4 seconds, off for 6
seconds). Nurse call activated. This occurs when the
resistance measurement is above the alert threshold or at
100% even when checking line.
NOTE: You will see a 4 digit alpha/numeric code next to the
configuration name upon entering configuration mode.
Refer to Section 2.4.1 “Entering Configuration Mode” . This
code is only a hexadecimal reflection of your instrument’s
configuration name.
NOTE: If you have changed the configuration but not the
name, the pump will display the option to rename before
turning off or proceeding to the teach mode. Refer to
Section 2.5.2 “Pop-Up Displays” for further explanation.
NOTE: Must have 4 characters in configuration name, use
space (-) symbol to fill in any open character.
NOTE: Only the configuration settings will be transferred.
The instrument ID number and the periodic maintenance
settings from the diagnostics mode will not be transferred
through LEARN/TEACH.
NOTE: Repeat allows you to re-attempt or teach the next
pump.
NOTE: The Learn/Teach function will not work if revision
level is not the same on both pumps or if the pumps are not
the same model (ie 710x/720x to 7000A).
NOTE: Do not use 70RCS more than 40 times.
NOTE: Due to the Dynamic Monitoring feature, the rate is
varied during operation. For this reason, ALARIS MEDICAL
SYSTEMS does not recommend using automatic testers to
check rate accuracy. Generally, these devices collect small
samples and may cause the results to be incorrect even
though the instrument is accurate.
NOTE: The main PCB board can only be replaced in pumps
with main software Rev. 2.02 or higher.
NOTE: Do not use 70RCS more than 40 times.
NOTE: X’s appear once pressure calibration is started or if
pressure calibration is required.
NOTE: The main PCB board can only be replaced in pumps
with main software Rev. 2.02 or higher.
NOTE: The alarm history is not stored in the EEPROM.
NOTE: Refer to Figure 4-5 “Battery Monitor” and Figure 4-6
“Main Power Supply” when following Sections 4.5.2
through 4.5.17.
NOTE: The charger will turn off if the battery gets too hot
(>37º C) to let the battery cool down. Cool down time is not
included in the 180 minute charge time.
NOTE: The instrument’s operation, when the battery is
disconnected, is the same as the Battery Manager
generating the SHUT-DOWN* signal and requires the
instrument to be plugged into AC after the battery has been
reconnected.
NOTE: The tests to be performed on a just-repaired
instrument depend on the level of repairs made to the
pump.
NOTE: Instrument configuration will not be lost when
disconnecting power. However, history and infusion
program settings may be lost. If you want to save event log,
record before proceeding.
NOTE: The instrument will not run with battery
disconnected.