Agilent Technologies E4374A, E4371A, E4370A User Manual

USER’S GUIDE
Multi-Cell Charger/Discharger
Agilent Model E4370A
Powerbus Load
Agilent Model E4371A
64-Channel Charger/Discharger
Agilent Model E4374A
Agilent Part No. 5964-8138
Microfiche No. 5964-8139
Warranty Information
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. further certifies that its calibration measurements are traceable to the United States National Bureau of Standards, to the extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other International Standards Organization members.
WARRANT Y
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of one year from date of delivery. Agilent Technologies software and firmware products, which are designated by Agilent Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent Technologies does not warrant that the operation for the software firmware, or hardware shall be uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent Technologies . Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned to Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies’ product price, only if Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies .
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Warranty options, extended support contacts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technologies Sales and Service office for further information on Agilent Technologies’ full line of Support Programs.
2
Safety Summary
y
Agilent Technologies
y
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safet standards of design, manufacture, and intended use of the instrument. for the customer’s failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.
BEFORE APPLYING POWER
Verify that all safety precautions are taken. Note the instrument’s external markings described under "Safety Symbols".
assumes no liabilit
GROUND THE INSTRUMENT
To minimize shock hazard, the Agilent MCCD Mainframe chassis and cover must be connected to an electrical ground. The mainfr ame must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
The Agilent Powerbus Load does not connect to ac mains. Connect the ground terminal of the load to the ground terminal of the external dc source. Use a #14 AWG wire as a minimum.
ATTENTION: Un circuit de terre continu est essentiel en vue du fonctionnement sécuritaire de l’appareil.
Ne jamais mettre l'appareil en marche lorsque le conducteur de mise … la terre est d‚branch‚.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
3
Safety Symbols
SAFETY SYMBOLS
Direct current Caution, risk of electric shock
Earth (ground) terminal Caution, hot surface
Protective earth (ground) terminal (Intended for connection to external protective conductor.)
On - power (Indicates connection to the ac mains.)
Off - power (Indicates disconnection from the ac mains.)
Caution (Refer to accompanying documents.)
On - equipment (Identifies the on condition of part of the equipment.)
Off - equipment (Identifies the off condition of part of the equipment.)
Document Scope
This document describes and specifies the “standard” version of the Agilent Multi-Cell Charger/Discharger System. It contains installation instructions, connection information, programming information, example programs, and specifications. Information about the Agilent MCCD User Interface is provided online. System options are described on a separate option sheet that is shipped with this manual. All information is this manual is subject to change. Updated editions will be identified by a new printing date.
Notice
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies. The information contained in this document is subject to change without notice.
Copyright 1999, 2000 Agilent Technologies, Inc.
4
Table of Contents
Warranty Information 2 Safety Summary 3 Document Scope 4
Notice 4
Table of Contents 5
1 - GENERAL INFORMATION 9
Agilent MCCD System Capabilities 9
Basic Functions 10 Additional Features 10
Hardware Description 10
Agilent E4370A/E4374A MCCD 10 Agilent E4371A Powerbus Load 12 External Power Source 13 Multiple Agilent MCCD Configuration 14
Measurement Capability 15
Voltage Measurements 15 Current Measurements 15 Capacity Measurements 16 Cell Resistance 16 Probe Resistance 17
Data Logging 17 Protection Features 18
Internal Protection Functions 18 External Digital I/O Protection Functions 19 If AC Power Fails 19
Remote Programming Interface 19
Application Programming Interface (API) 20 Web Accessible Agilent MCCD User Interface 20
Example of a Cell Forming Process 20
2 - INSTALLATION 23
Inspection 23 Parts and Accessories 23 Location 25
Agilent E4370A MCCD Mainframe 25 Agilent E4371A Powerbus Load 25
Channel Connections 25
Voltage Drops and Wire Resistance 26 Remote Sense Connections 27
Power Bus Connections 28
Power Bus Wiring Information 28
Digital Connections 31
General Purpose I/O 31 Special Functions 32 Wiring Guidelines 32
RS-232 Connections 34 Auxiliary Output Connection 35 Installing the API Library and Measurement Log Utility 36
Visual C++ Configuration 36
5
3 - CONFIGURATION 37
Configuring the LAN 37
1. Configure the HyperTerminal program 37
2. Connect the Agilent E4370A MCCD to the COM port on the PC 38
3. Fill Out the Agilent MCCD Configuration Screens 38 Network Configuration 39 Identification Configuration 40 Miscellaneous Configuration 41
Configuring the Digital I/O 41
Mixed Configuratio n Example 44
Accessing Calibration 44
4 - AGILENT MCCD USER INTERFACE 45
Description 45
PC Requirements 45 Browser Settings 45 Security 45 Localization 46 Access 46
Using the Interface 46 Using the Agilent MCCD Measurement Log Utility 47
5 - PROGRAMMING OVERVIEW 49
A Cell Forming Overview 49
Cell Forming Example 50
Function Call Overview 53
Cell Grouping 53 Grouping Functions 54 Step/Test Functions 54 Sequence Control 55 Output Configuration 56 Instrument Protection 57 Power Fail Operation 58 Instrument State Storage 58 Status 59 Measurement Log 60 Time Stamp Function 61 Output Measurements 61 Direct output control 62 General Server functions 62 Selftest 63 Calibration 63 Serial port 64 Digital port 64 Probe check 65
6 - LANGUAGE DICTIONARY 67
API Usage Guidelines 67 API Function Summary 68 API Function Definitions 70
cfAbort 70 cfCal 70 cfCalStandard 70
6
cfCalTransfer 71 cfClose 71 cfDeleteGroup 71 cfGetCellStatus 72 cfGetCellStatusString 72 cfGetCurrent 72 cfGetDigitalConfig 73 cfGetDigitalPort 73 cfGetGroups 73 cfGetInstIdentify 74 cfGetInstStatus 74 cfGetMeasLogInterval 75 cfGetOutputConfig 75 cfGetOutputProbeTest 75 cfGetOutputState 76 cfGetRunState 76 cfGetSense 76 cfGetSenseProbeTest 77 cfGetSeqStep 77 cfGetSeqTest 77 cfGetSeqTestAnd 78 cfGetSeqTime 78 cfGetSerialConfig 78 cfGetSerialStatus 78 cfGetShutdownDelay 79 cfGetShutdownMode 79 cfGetStepNumber 79 cfGetTrigSource 79 cfGetUserIdentify 79 cfGetVoltage 80 cfInitiate 80 cfMeasACResistance 80 cfMeasCapacityAS 80 cfMeasCapacityWS 81 cfMeasCurrent 81 cfMeasDCResistance 81 cfMeasOutputProbeResistance 81 cfMeasProbeContinuity 82 cfMeasSenseProbeResistance 82 cfMeasVoltage 83 cfOpen 83 cfOpenGroup 83 cfProtect 84 cfProtectClear 84 cfReadMeasLog 84 cfReadSerial 86 cfReadTestLog 87 cfReset 87 cfResetSeq 87 cfRestart 88 cfSaveOutputConfig 88 cfSelftest 88 cfSetAutoConnect 89 cfSetCurrent 89 cfSetDigitalConfig 90
7
cfSetDigitalPort 92 cfSetErrorFunction 92 cfSetGroup 93 cfSetMeasLogInterval 93 cfSetOutputConfig 93 cfSetOutputProbeTest 94 cfSetOutputState 94 cfSetSense 95 cfSetSenseProbeTest 95 cfSetSeqStep 95 cfSetSeqTest 97 cfSetSeqTestAnd 99 cfSetSerialConfig 99 cfSetServerTimeout 99 cfSetShutdownDelay 100 cfSetShutdownMode 100 cfSetTimeout 100 cfSetTrigSource 100 cfSetVoltage 101 cfShutdown 101 cfStateDelete 101 cfStateList 102 cfStateRecall 102 cfStateSave 102 cfTrigger 102 cfWriteSerial 103
7 - C PROGRAM EXAMPLES 105
Example 1 105 Example 2 107 Example 3 112
A - SPECIFICATIONS 115
Hardware Specifications 115
B - CALIBRATION 119
Calibration Types 119
Full Calibration 119 Transfer Calibration 120 Mainframe Reference Calibration 120 Calibration Connections 120
Accessing Calibration 122 Calibration Error Messages 123
C - DIMENSION DRAWINGS 125 D - SENSE AND POWER CONNECTOR PINOUTS 127 E - IN CASE OF TROUBLE 135
Introduction 135
Fault LEDs (see Figure 1-2) 135
Selftest Error Messages 136
INDEX 137
8
1
General Information
Agilent MCCD System Capabilities
The Agilent Multi-Cell Charger/Discharger (MCCD) System has been designed to address the unique requirements and needs of lithium-ion cell manufacturing. The Agilent MCCD System can accurately charge, discharge, and measure lithium ion cells. It consists of an Agilent E4370A Multi-Cell Charger/Discharger mainframe with up to four Agilent E4374A 64-Channel Charger/Discharger cards. When fully loaded each mainframe has 256 input/output channels. Mainframes and modules can be combined in different configurations to form a low cost, high performance cell charge/discharge station in a cell manufacturing process.
The following figure is a simplified block diagram of the Agilent MCCD System. It is followed by a brief
description of the system’s basic as well as advanced features.
10 Base T Ethernet to remote monitoring and control
Powerbus
Digital I/O to outside world
Remote
Rail power
source
Powerbus
Load
Multi-cell
Local
charger /
discharger
Digital I/O
Fixture
control and
local
start/stop
Local
controls
Power
Multiple cell
tray
Sense
Serial
Local
terminal
Figure 1-1. Block Diagram of Agilent MCCD System
Serial
Local
barcode
reader
9
1 - General Informat ion
Basic Functions
Charger – The Agilent MCCD can deliver accurately controlled current and voltage into a cell for
proper forming. Each cell is independently paced through the cell forming sequence. This means that some cells can be charging and others discharging if they are at different points in the sequence.
Discharger – The Agilent MCCD can draw accurately controlled current from a cell for both
forming and capacity measurement.
Measurement – The Agilent MCCD can monitor several parameters of the cell while charging,
discharging, and resting. Measurements include voltage, current, time, internal resistance, ampere­hours, and watt-hours. These measurements are used to adjust the cell forming sequence for safety, reliability, and or proper cell forming.
Digital I/O control – The Agilent MCCD can monitor and stimulate digital I/O connected to it. This
simplifies wiring, allows ease of expansion, and is more reliable than a centralized control system. Its high-speed capability is ideal for fast fault detection and system shutdown.
RS-232 control – The Agilent MCCD can support peripherals connected to its serial ports for adding
printers, bar code readers, local terminals, robots and other types of local additional hardware via pass-through control from the host computer.
Equipment Protection – The Agilent MCCD has extensive safety features to protect both the cells
under formation and the hardware from equipment failure, programming errors, cell failures and other types of external faults.
Additional Features
LAN 10 base-T control using a web-server graphical user interface and an application programming
interface (API).
Comprehensive data storage capability and remote data collection. Easily removable charger/discharger cards for minimum downtime if repair is required. Charge/discharge sequences that can be modified in software, allowing for simple, rapid changes to
the manufacturing process without changes to system hardware.
Define and configure groups of contiguous blocks of cells or channels. This lets you simultaneously
run different sequences on groups of cells.
Continuous calibration is performed on the programming circuits during the entire charge/discharge
sequence to eliminate errors due to temperature drift.
Bi-directional power transfer and reuse of energy by using energy from discharging cells to provide
energy to charging cells.
Hardware Description
Agilent E4370A/E4374A MCCD
The Agilent E4370A MCCD mainframe is a full-width rack box that has 4 slots to hold the E4374A 64­Channel Charger/Discharger cards. The Agilent E4374A 64-Channel Charger/Discharger cards contain the circuitry that independently charges and discharges each cell at up to 5V and 2A.
NOTE: Each output channel has a maximum available compliance voltage of 5.5V. Compliance
voltage is defined as the voltage required at the cell plus any fixture/wiring voltage drops. Having this higher compliance voltage allows the full 5 V to be applied directly to the cell with a maximum of 0.5 volt loss in the wiring.
10
General Information - 1
E4370A MULTICELL CHARGER/DISCHARGER
SYSTEM
Power
Ready
Active
FAULT
External
Internal
LINE
LINE
1
Ready
Fault
2
Ready
Fault
3
Ready
Fault
4
On
Ready
Fault
Off
E4374A CHARGER/DISCHARGER
E4374A CHARGER/DISCHARGER
E4374A CHARGER/DISCHARGER
E4374A CHARGER/DISCHARGER
1
2
1
2
1
2
1
2
Applies and removes ac power from the Agilent MCCD. Relays inside the unit that connect the power bus are disengaged when power is off, so the power bus is also disconnected from the unit by this switch.
SYSTEM Power Ready
When lit, indicates that the mainframe is powered on. When lit, indicates that the unit is ready for operation.
When off, indicates that the external power bus voltage is either too high or too low.
Active
When lit, indicates that data communication is present on the LAN cable. When flashing, indicates that LAN communication is in progress.
FAULT (Refer to Appendix E to clear any fault conditions) External
When lit, indicates an external fault such as:
External digital fault signal received, Power fail shutdown signal received, High power bus voltage after power on, Low power bus voltage after power-on. Overtemperature
Internal
When lit, indicates an internal hardware fault such as:
Selftest failure, Calibration error, Hardware error.
1, 2, 3, 4 Ready Fault
Indicates the card is powered up and ready to be used When lit, indicates an internal hardware fault such as:
Selftest failure, Calibration error, Hardware error.
3
4
3
4
3
4
3
4
5
6
5
6
5
6
5
6
7
8
7
8
7
8
7
8
Figure 1-2. Agilent E4370A/E4374A MCCD Front Panel Controls and Indicators
11
1 - General Informat ion
A
+ and - Power bus connectors
A
(- bus bar is connected to chassis ground)
B
Calibration status LEDs
C
Configuration switches
D
Transfer Calibration switch
E
Digital I/O connectors
F
LAN connection
B C D E
RS-232
PORT A
RS-232
PORT B
K J
RS-232 connectors (ports A and B)
G
AC line connection (a universal AC input for line
H
voltages from 87 Vac to 250 Vac, 50/60 Hz.) Auxiliary output connection
J
Calibration port
K
F
G H
Figure 1-3. Agilent E4370A/E4374A MCCD Rear Panel Connections
Agilent E4371A Powerbus Load
For the discharging cycle, an Agilent E4371A Powerbus Load is required to dissipate excess power from discharging cells. The load operates in constant voltage mode only and sequentially switches internal resistors on and off to regulate the voltage on the power bus around a midpoint of 26.75 volts. The number of load units required depends on the number of Agilent MCCD mainframes in your system. Each Agilent E4371A Powerbus Load is capable of the full power from two 256-channel Agilent E4370A MCCD mainframes.
The Agilent E4371A Powerbus Load has a + and a power bus connector on its rear panel. There is also a ground connection. To meet safety requirements, connect the ground terminal of the Agilent Powerbus load to the ground terminal of the external dc source. The load receives its operating power from the power bus. If the dc voltage on the power bus drops below 23.8 volts, or if there is no power available on the power bus, the load will not operate. Note that the load is not programmable. It is set at the factory for the correct operating voltage and does not require calibration.
The On/Off switch on the load simply connects or disconnects the load from the power bus. Note that the internal fans draw approximately 1.5 amperes of current from the power bus.
CAUTION: When discharging its maximum rated power, the Agilent E4371A Powerbus Load
becomes hot to the touch.
12
E
4372A
POWERBUS LOAD
General Information - 1
Figure 1-4. Agilent E4371A Powerbus Load Front Panel
Figure 1-5. Agilent E4371A Powerbus Load Rear Panel
External Power Source
For the charging cycle, each Agilent MCCD mainframe requires an external dc power source to power the cells. The external power source connects to the power bus terminals on the back of the mainframe. It must be rated at 24 volts and be able to source 125% of the required cell charging power. For example, to provide the cell charging power for a 256-channel system at 5 volts, 2 amperes per channel (or 2.56 kW), the dc power source must deliver approximately 3.2 kW to each Agilent MCCD mainframe (24 V @ 133 A).
The current rating of the power source may be reduced if the charging current is reduced accordingly. For example, to provide a maximum output current of 1 ampere per cell in the previously described system, a source rated at least 63 amperes may be used.
13
1 - General Informat ion
Additionally, a single supply of sufficient amperage may be shared among multiple mainframes that are connected to a common power bus - provided that the total current can be supplied while meeting the voltage specification at the power bus terminals at the rear of the Agilent MCCD.
NOTE: If the external dc power source has an overvoltage protection circuit, it must be set
higher than 30 volts to avoid the possibility of shutting itself down during the discharge cycle.
Multiple Agilent MCCD Configuration
The following figure illustrates an Agilent E4370A MCCD system with eight fully loaded mainframes.
Agilent E4371A
Powerbus Load
25.6 kW
Power Source
(24 V @
1067A)
Agilent E4370A
MCCD
(256 channels)
Agilent E4370A
MCCD
(256 channels)
Agilent E4370A
MCCD
(256 channels)
Agilent E4371A
Powerbus Load
Agilent E4371A
Powerbus Load
Agilent E4371A
Powerbus Load
Agilent E4370A
MCCD
(256 channels)
POWERBUS
Agilent E4370A
MCCD
(256 channels)
Agilent E4370A
MCCD
(256 channels)
14
Agilent E4370A
MCCD
(256 channels)
Agilent E4370A
MCCD
(256 channels)
Figure 1-6. Maximum System Block Diagram
General Information - 1
The maximum power required for such a system is 25.6 kilowatts. A single power source of sufficient total amperage may be shared among multiple mainframes connected to the power bus, provided the total current can be provided while meeting the 24 volt dc input requirement at the power bus terminals on the rear of each mainframe. Multiple paralleled 24 volt dc sources may be used in place of the single 24 volt,
25.6 kilowatt dc source shown in the figure.
To achieve improvements in energy efficiency, the Agilent E4370A MCCD system can re-use discharge energy to supplement the energy provided by an external power source when charging other cells in a multi-unit system. This is possible because of the bi-directional power transfer capability between charging and discharging cells when connected to a common power bus. To take advantage of this energy transfer requires that some mainframes in the system must be operating in discharge mode at the same time that others are operating in charging mode.
No special control system is required for this configuration. The regulation circuits of the 24 volt dc power source, the Agilent E4370A MCCD, and Agilent E4371A Powerbus Load will operate properly without any special hardware control lines or additional software being required.
NOTE: Adequate size power bus wiring is required to carry high currents. Refer to Table 2-5.
Measurement Capability
The Agilent MCCD mainframe and charger/discharger cards have a high speed scanning system that makes voltage and current measurements on all channels. Refer to Appendix A for technical data about the measurement system. The following measurements are available:
Voltage Measurements
The Agilent MCCD measures the voltage of each channel using a calibrated internal measurement circuit. In local sensing mode, the voltage measurement is made at the power connector. In remote sensing mode, the voltage is measured at the end of the remote sense leads. The advantage of remote sensing over local sensing is that when the remote sense leads are connected to the cell, the actual voltage of the cell will be measured. Any voltage drops in the load leads will not affect the measurement. Refer to chapter 2 under Remote Sensing for more information.
NOTE: If your Agilent MCCD system is configured for local sensing, the measured output
voltage may not reflect the actual voltage at the cell. This is because any voltage drops in the wires due to wire resistance, probe resistance, connector resistance, etc. will reduce the available voltage at the cell.
Current Measurements
The Agilent MCCD measures actual current in the output current path for each channel using a calibrated internal measurement circuit.
15
1 - General Informat ion
Capacity Measurements
Amp-hour capacity - the Agilent MCCD determines amp-hour cell capacity by making calculations based on continuous current measurements.
During charge, every time the Agilent MCCD makes a measurement, it calculates the actual incremental amp-hours put into the cell during each measurement interval by multiplying the measured current times the measurement interval. It then adds this incremental amount to the accumulated amp-hour value to determine the total amp-hours delivered into the cell. Amp-hour capacity will be positive during charge. Thus, accurate amp-hour capacity measurements can be made even when charge current is not constant, such as during constant voltage charging.
During discharge, every time the Agilent MCCD makes a measurement, it calculates the actual incremental amp-hours taken out of the cell by multiplying the measured current times the measurement interval. It then adds this incremental amount to the accumulated amp-hour value to determine the total amp-hours removed from the cell. Amp-hour capacity will be negative during discharge. Thus, accurate amp-hour capacity measurements can be made even when discharge current is not constant.
Watt-hour capacity - the Agilent MCCD determines watt-hour cell capacity by making calculations based on continuous current and voltage measurements.
During charge, every time the Agilent MCCD makes a measurement, it calculates the actual incremental watt-hours put into the cell during each measurement interval by multiplying the measured current times the measured voltage times the measurement interval. It then adds this incremental amount to the accumulated watt-hour value to determine the total watt-hours delivered into the cell. Watt-hour capacity will be positive during charge. Thus, accurate watt-hour capacity measurements can be made even when charge current and voltage is varying.
During discharge, every time the Agilent MCCD makes a measurement, it calculates the actual incremental watt-hours taken from the cell during each measurement interval by multiplying the measured current times the measured voltage times the measurement interval. It then adds this incremental amount to the accumulated watt-hour value to determine the total watt-hours taken from the cell. Watt-hour capacity will be negative during discharge. Thus, accurate watt-hour capacity measurements can be made even when discharge current and voltage is varying.
Cell Resistance
In addition to continuous voltage, current, and capacity measurements, the Agilent MCCD can also measure ac and dc cell resistance. This measurement is available on command when a sequence is not running, or as its own step in the forming sequence.
The Agilent MCCD measures the ac cell resistance by first disconnecting the charge/discharge circuits from all cells. An ac waveform generator in the Agilent MCCD mainframe is connected sequentially to each cell. The ac waveform generator momentarily passes a small excitation current through each cell
while the measurement system measures the cell’s output voltage and current. By using a narrow band tuned filter and computing the magnitude and phase angle of voltage relative to current, an ac resistance measurement of the cell can be made. This method is very similar to the method used by LCR meters. Since this measurement happens sequentially for each channel, the other channels stay at rest during this test.
16
General Information - 1
The Agilent MCCD measures the dc cell resistance by first disconnecting the charge/discharge circuits from all cells. A pulse generator in the Agilent MCCD mainframe is connected sequentially to each cell. The pulse generator passes a short-duration pulsed current through each cell while the measurement system digitizes the cell voltage and current using a high accuracy, high-speed A/D converter. Using proprietary algorithms to calculate the change in voltage relative to the change in pulsed current, a dc (or pulse) resistance measurement of the cell can be made. Since this measurement happens sequentially for each channel, the other channels stay at rest during this test.
Probe Resistance
Probe resistance measurements can also be performed. The Agilent MCCD uses the remote sense to measure the resistance of both the power and sense probes. Probe resistance measurements can be made on command when a sequence is not running.
The measured probe resistance is the total resistance in the signal path, which includes wiring resistance, probe resistance, and the resistance of any connectors in the signal path. For the sense probe measurement, the resistance measurement includes the internal scanner resistance, which is typically 1000 ohms. The power and sense probe measurements return the actual measured value in ohms.
In addition to the on-command probe resistance measurements, the probes are continuously checked
while the sequence is running. See chapter 5 under “Probe Check” for more information about probe check verification.
Data Logging
During a charge/discharge sequence, the Agilent MCCD is constantly making voltage, current, and capacity measurements. Instead of logging each and every measurement into a data buffer, the data logging can be controlled so that only critical measurements are logged to the data buffer. This is called event-based data logging, which means that whenever an important event occurs, a data log record will be written into the data buffer. Buffer memory is used most efficiently when only critical measurements are stored.
The following events can be used to trigger critical measurements:
Change in voltage (V)
Change in current (I)
Change in time (t)
The acceptable range of values for ∆V, ∆I and ∆t are 0 to infinity. Setting the value to 0 or near 0 will cause all readings to be logged in the buffer, because every reading will exceed the ∆V, ∆I or ∆t value of zero. This will fill up the measurement log very quickly. Setting the value to a high number or to infinity will cause no readings to be logged in the buffer because no reading will exceed the ∆V, ∆I or t value.
If the trigger is V, a data log record will be written to the buffer when a user­specified voltage change is exceeded. If V is set to 100 mV, then each time the voltage reading changes by more than 100 mV, a record is written to the buffer.
If the trigger is I, a data log record will be written to the buffer when a user­specified current change is exceeded. If I is set to 100 mA, then each time the current reading changes by more than 100 mA a record is written to the buffer.
If the trigger is t, a data log record will be written to the buffer when a user­specified time interval is exceeded. If t is set to 1 second, then every second a record is written to the buffer. t is effectively a clock-driven data log.
17
1 - General Informat ion
The comparison test to see if the ∆V, ∆I, and ∆t values have been exceeded is done at the end of each measurement interval, so the fastest rate at which records can be written into the data buffer is the measurement rate of the Agilent MCCD. Any combination of events can be specified, so that a data log record is written into the data buffer when any of the events occur.
Each record in the data buffer contains the following information: status (including CV/CC and step number), elapsed time, voltage, current, amp-hours, and watt-hours. The total number of readings that can be stored is given in the specification table. The data log is a circular queue, which lets you continuously log data into the data buffer. When the data buffer is full, the oldest data in the buffer will be overwritten by new data. To avoid data loss, the controller must read the data from the buffer before it is overwritten. Data can be read out of the data buffer at any time during the test sequence.
NOTE: Information in the data buffer is lost when an ac power failure occurs. To prevent data
loss in the event of a power failure, use the cfShutdown function to save the data in non­volatile memory. Refer to Power Fail Operation in chapter 5 for more information. To allow the Agilent E4370A to ride through temporary ac power interruptions, connect the mainframe to a 600 VA uninterruptible power supply (UPS).
A measurement log utility is included in the software that is provided with the Agilent E4373A Documentation package. You can use this utility to read the data log and place the information in a file on your PC. For information on how to use the Agilent MCCD Measurement Log Utility, refer to chapter
4.
Protection Features
The Agilent MCCD provides extensive capability to protect both the hardware and the individual cells being formed from catastrophic damage. The Agilent MCCD can also communicate its protection status to other parts of the manufacturing system for more sophisticated forms of protection.
Internal Protection Functions
There are internal relays between the power bus and the Agilent E4374A Charge/Discharge cards. These relays protect the Agilent MCCD from overvoltage and undervoltage conditions on the power bus. They also protect the Agilent MCCD if an external fault condition is detected. Output regulators include several features to protect the cell from failures in the hardware. Internal circuits connected in series with each channel protect the system from reverse cell polarity, cell failure, and regulator failure. Internal thermal sensors check for maximum heat rise to avoid failures due to excessive temperature excursions. A fan keeps the internal temperature at an acceptable level.
Finally, the Agilent MCCD has an extra level of safety - a built-in hardware watchdog timer. The hardware watchdog timer is independent of CPU, software, or firmware activities. If, due to some internal firmware or software fault, the CPU in the Agilent MCCD should stop functioning for more than a few seconds, the hardware watchdog timer will reset the Agilent MCCD to the power-on state. In this state, the channels outputs are disconnected from the cells.
NOTE: Overvoltage and overcurrent tests can be included as part of a test sequence to
implement overvoltage and overcurrent protection (see chapter 5).
18
General Information - 1
External Digital I/O Protection Functions
The Digital I/O subsystem on the Agilent MCCD can be configured to provide protection capabilities. These digital I/O signals operate independently, so that if there is a problem with the computer or the LAN connection the protection functions of the Agilent MCCD are not compromised. As explained in chapter 2, the 16 digital I/O signals can be individually configured to provide one of the following protection functions:
External Fault Input
External Fault Output
External Interlock
External Trigger
In addition to protection capabilities, the digital I/O can also be used as general purpose I/O. When configured as a general purpose I/O, the input or output signals on the digital connector are directly controlled with API programming commands over the LAN.
This function can be used to stop the cell forming sequence if an external fault condition sets the input true.
This function can be used to signal external circuitry or another Agilent MCCD that either an external fault condition or an internal fault condition has occurred.
This function can be used to stop the cell forming sequence for reasons other than an external fault condition.
This function can be used to start a cell forming sequence.
If AC Power Fails
Should the ac line fail, the CPU in the Agilent MCCD will shut down. Any charging and discharging activity will stop, and the current sequence, test data, and programmed settings will be lost.
Note: A 600 VA uninterruptible power supply (UPS) can be used to provide ac power to the
Agilent E4370A MCCD mainframe to prevent any data loss during a power failure.
When power fails, the power bus is also disconnected from the Agilent MCCD because of the bias powered relays inside the Agilent MCCD. Thus, should a power failure occur which causes the Agilent MCCD to lose ac power, in order to provide for safety, these internal relays would be disengaged and any further charging or discharging would stop, even if the power bus were still powered and active.
Also, should a power failure occur which does not effect the Agilent MCCD but which causes the power bus to drop in voltage, this will be detected by the Agilent MCCD as a power bus undervoltage condition and the relays will open, thus preventing any further charging or discharging of connected cells.
Remote Programming Interface
The remote programming interface to the Agilent MCCD is through a LAN-based TCP/IP communication protocol. The connection to the LAN is through a standard 8-pin 10Base-T connector on the rear panel, which must first be configured according to the directions in chapter 3. The LAN communication protocol is implemented in two ways:
19
1 - General Informat ion
Application Programming Interface (API)
The application programming interface runs under Windows 95 or Windows NT 4.0 using supplied C­language function calls. These function calls are documented in chapters 5 and 6, and provide the most comprehensive method of controlling the Agilent MCCD. The API interface is the preferred method of control when the Agilent MCCD is connected to a remote computer as part of an automated manufacturing process.
Web Accessible Agilent MCCD User Interface
The Agilent MCCD has a built-in web server with a graphical user interface that is accessed through standard web browsers such as Netscape Navigator version 3.03 or Microsoft Internet Explorer version
3.02. This Agilent MCCD User Interface allows monitoring of individual cell state, measuring cell voltages and currents while the test is running, and also complete monitoring and control of test status. The Agilent MCCD User Interface is the preferred method of control when evaluating the test system, prototyping a process, or debugging a program.
Example of a Cell Forming Process
The Agilent E4370A MCCD is designed to be the integral part of a complete cell forming process as shown in Figure 1-7. As shown in the figure, many of the previously mentioned protection and external signal capabilities of the Agilent E4370A MCCD are implemented using the digital I/O connections. The serial ports on the back of the Agilent MCCD are used to control local peripherals directly from the host computer. The remote programming interface to the Agilent MCCD lets you seamlessly integrate all of these capabilities into the cell forming process.
&RQWURO3&
)L[WXUH
/$1
Ready
7UD\RIFHOOV
'LJ,2IRUILUHVPRNHGHWHFWRU
'LJ,2IRUIL[WXUH
RSHQFORVH
H
'LJ,2IRU
EXWWRQVDQG
LQGLFDWRUV
Test Start
Figure 1-7. Typical Cell Forming Station
3RZHU6HQVHOLQHV
%DU&RGH6FDQQHU
FRPPXQLFDWLRQV
3RZHUEXV
0&&'
6HULDO
The following cell forming example describes how an Agilent E4370A MCCD may be used to run a semi-automated process where the only human actions required are: entering data with a barcode
20
General Information - 1
scanner, loading and unloading a test fixture, and manually starting the cell forming process. Chapters 5 and 6 describe all of the function calls that are available to implement a cell forming process.
The control PC sends a signal via the LAN to the digital I/O to turn on the Ready light on the test
fixture. This tells the operator that the system is ready for another tray of cells. The control PC also begins polling for serial data on the RS-232 buffer of the Agilent MCCD.
The operator scans the bar code on the tray of cells sitting on the conveyor belt. The operator then
loads the tray into the test fixture and closes the fixture.
After detecting that data is available on the RS-232 buffer, the control PC reads the bar code data.
Based on the data, it downloads the correct forming sequence into the Agilent MCCD. It also downloads setup information such as which channel outputs to enable, probe check settings, trigger source, etc.
The control PC then polls the digital I/O lines for the Start button. When the operator presses Start, the control PC detects it and polls the digital I/O lines to make sure
the fixture is closed. It sends a signal to turn off the Ready light and turn on the Test light, indicating to the operator that the cell forming sequence has started.
The control PC then sends a trigger to the Agilent MCCD to start the forming sequence. It also starts
polling the instrument status for the completion of the test sequence.
The cell forming sequence runs. The test sequence automatically applies a stimulus to the cells,
monitors cell parameters to determine if a cell passes or fails, and stores the test results. During the test sequence, the Agilent MCCD monitors the dedicated digital I/O lines that are connected to the fire and smoke detectors. This allows rapid response in case of a problem.
When the instrument status in the Agilent MCCD shows that the sequence is complete, the control
PC sends commands to the Agilent MCCD to measure the internal resistance of all cells and then upload all measurement data.
Finally, the control PC sends a signal to turn off the Test light and light the Ready light. The
operator knows that it is now safe to remove the tray from the fixture and start another batch.
Chapter 7 contains several programming examples written in C. The purpose of these examples is to show you how to implement the various functions of the Agilent MCCD so that you can develop your own application programs. Program #2 matches the example described here.
21
2
Installation
Inspection
When you receive your equipment, inspect it for any obvious damage that may have occurred during shipment. If there is damage, notify the shipping carrier and the nearest Agilent Sales and Support Office immediately. The list of Agilent Technologies Sales and Support Offices is at the back of this guide. Warranty information is printed in the front of this guide.
Until you have checked out the Agilent MCCD, save the shipping carton and packing materials in case the unit has to be returned. If you return the Agilent MCCD for service, attach a tag identifying the model number, serial number, and the owner. Also include a brief description of the problem.
Parts and Accessories
Table 2-1 lists items that are included with your Agilent E4370A/E4374A MCCD.
Table 2-2 lists accessory items that are not included with the Agilent E4370A/E4374A MCCD, but must
be purchased separately. Except for the User’s Guide, all of these items are required to make connections from the Agilent MCCD to either the computer, test fixture, or external devices that will be controlled by the Agilent MCCD.
You can either order these items by ordering the appropriate kit, or order them directly from the manufacturer. Table 2-3 lists the addresses of the manufacturers of the connector parts.
Table 2-1. Supplied Items
Item Part Number Description
Power Cord (1) Contact your
Agilent Sales and Support office
Table 2-2. Accessories
Item Manufacturer’s
Part Number
Digital connectors (2) Phoenix
MSTB-2.5/10-STF
Calibration connector (1) Auxiliary bias connector (1)
Phoenix MSTB-2.5/4-ST
A power cord appropriate for your location.
Description
10-pin terminal plugs that connect to the digital connectors on the back of the unit.
4-pin terminal plugs that connect to the calibration and auxiliary connectors on the back of the unit.
23
2 - Installation
Table 2-2. Accessories (continued)
Item Manufacturer’s Part
Description
Number
Documentation Package Agilent E4373A Contains user documentation, software
drivers, and utility programs .
Serial cable Agilent 34398A RS-232 null-modem cable for port A or B.
(see figure 2-4 for schematic)
37-pin D-sub connector
AMP 205210-2 Mating connector for Agilent E4374A front
panel channel connectors. Eight connectors are required for each Agilent E4374A card. (Connector pins on Agilent E4374A cards are rated at 5 A maximum.)
Connector hood for 37­pin connector
AMP 749916-2 Eight connector hoods are required for each
Agilent E4374A card.
Crimp style contacts for 37-pin connector
AMP 66506-9 Crimp contact for 37 pin connector
16 contacts are required for each connector. (Pins only accept wires sized 20-24 AWG.)
Crimp tool for crimp
AMP 58448-2 Hand crimp tool
style contacts
Solder style contacts for 37-pin connector
AMP 66570-2 Crimp contact for 37 pin connector
16 contacts are required for each connector. No tooling is required. (Pins only accept wires sized 18 AWG.)
Front Panel Filler Panel Agilent p/n 5002-1505 One blank filler panel is required for every
empty slot in Agilent MCCD mainframes.
Rack mount Flange Kit Agilent p/n 5062-3979 Includes 2 flanges, fasteners, and mounting
screws
Rack mount Flange Kit with Handles
Agilent p/n 5062-3985 Includes 2 handles, 2 flanges, fasteners, and
mounting screws
Table 2-3. Manufacturer’s Addresses
Company Address Contact
Phoenix Contact P.O. Box 4100
Harrisburg, PA 17111-0100
Phone:717-944-1300 Fax: 717-944-1625
http://www.phoenixcontact.com/index.html AMP Harrisburg, PA 17111 http://www.amp.com/ Agilent Technologies See list at back of this manual http://www.agilent.com/
24
Installation - 2
Location
Agilent E4370A MCCD Mainframe
The outline diagrams in Appendix C give the dimensions of your Agilent MCCD mainframe. The mainframe may be installed free-standing, but must be located with sufficient space at the sides and back of the unit for adequate air circulation. You can rack mount the mainframe in standard 600 mm (23.8 in.) width system cabinets. This provides sufficient clearance for airflow. Support rails are also required when rack mounting the mainframe. These are usually ordered along with the cabinet.
A fan cools the Agilent MCCD mainframe by drawing air in on the left side of the unit and discharging it through the back and side. Minimum clearance is 9 cm (3.5 inches) along the sides. Minimum clearance behind the mainframe is 23 cm (9 inches). Do not block the fan exhaust at the rear or the side.
NOTE: To ensure proper cooling of the Agilent MCCD mainframe, there should be no open slots
in the front of the mainframe. If an Agilent E4374A Charger/Discharger Card is either not installed or has been removed from a slot, a blank filler panel must be installed in the opening. Refer to Table 2-2.
Agilent E4371A Powerbus Load
CAUTION: To ensure adequate airflow to cool the Agilent Powerbus Load requires you to leave 0.6
meters (2 feet) of open space in front of the load and directly behind the load. If you are rack-mounting the load, leave the rack door off.
When discharging its maximum rated power, the Agilent E4371A Powerbus Load becomes hot to the touch.
The outline diagrams in Appendix C give the dimensions of your Agilent Powerbus Load. The unit may be installed free-standing, but must be located with sufficient space at the front and back of the unit for adequate air circulation. Fans cool the unit by drawing air in on front and discharging it through the back. Maximum airflow is 10 cubic meters per minute (350 cubic feet per minute).
You can rack mount the Agilent E4371A Powerbus Load in standard 600 mm (23.8 in.) width system cabinets, provided that you remove the rear door. This provides sufficient clearance for airflow. Rack mount kits are described in Table 2-2. Support rails are required when rack mounting the unit. To meet safety requirements, connect the ground terminal of the Agilent Powerbus load to the ground terminal of the external dc source.
Channel Connections
Each Agilent E4370A MCCD mainframe can control up to 256 individual charge/discharge cells when four Agilent E4374A Charger/Discharger cards are installed. Each Agilent E4374A Charger/Discharger contains 64 channels. Note that in the programming sections of this manual, channels are also referred to as outputs. When fully loaded, the 256 charge/discharge channels are configured as follows:
25
2 - Installation
Table 2-4. Channel Configuration
Card Connector Number
Number 1 2 3 4 5 6 7 8
1 2 3 4
1 - 8 9 - 16 17 - 24 25 - 32 33 - 40 41 - 48 49 - 56 57 - 64
65 - 72 73 - 80 81 - 88 89 - 96 97 - 104 105 - 112 113 - 120 121 - 128 129 - 136 137 - 144 145 - 152 153 - 160 161 - 168 169 - 176 177 - 184 185 - 192 193 - 200 201 - 208 209 - 216 217 - 224 225- 232 233 - 240 241 - 248 249 - 256
Power connections on each Agilent E4374A card are through eight 37 pin D-subminiature connectors. These connectors allow for shielding and strain relief. Corresponding sense connections are also available on the connectors. Refer to Table 2-2 for information about ordering the mating connectors. As indicated in he table, mating connectors accept wire sizes from AWG 24 up to AWG 18, depending on the type of connector that you are using. You must wire up the mating connector to make your wire connections. Install the mating connector on the front of the Agilent E4374A card when complete. Refer to Appendix D for detailed pinout assignments of the front panel connectors.
If specific channels are not being used, you can configure them to be inactive. Inactive channels are open-circuited. Note that there are two ways to configure the channel outputs, each having different effects when the unit is powered on.
If you configure the channel outputs using the cfSetOutputConfig() function (see chapter 6), the
settings are NOT saved in non-volatile memory. Each time you power up the unit, you must reprogram the settings.
If you configure the channel outputs using the Sequence setup page in the Agilent MCCD User
interface (see chapter 4), the settings ARE saved in non-volatile memory. The unit will wake up with those settings when it powered up.
NOTE: If the mainframe has empty card slots, the channels that are normally reserved for those
card slots will be treated as inactive channels.
Voltage Drops and Wire Resistance
NOTE: Each channel has a maximum of 5.5V and 2A available at the power connector.
At the rated output, the Agilent E4374A Charger/Discharger will tolerate up to a 0.5 volt drop in the load leads due to wire resistance, probe resistance, connector resistance, etc. Higher voltage drops will reduce the available voltage at the cell. Proper wiring design including using larger gauge wires and low-resistance fixture contacts can minimize voltage losses in the wiring and maximize the available voltage for charging the cells.
The length of the leads from the power connector to the cells is determined by how much voltage drop your system can tolerate. The voltage drop is directly determined by the wire, connector, and probe resistance (see table 2-5). Refer to Remote Sense Connections for more information.
To optimize performance and minimize the possibility of output instability and output noise, please observe the following guidelines:
26
Installation - 2
It is good engineering practice to either twist or shield the sense and power wires. Twist the power wires together and keep them as short as possible. Twist the sense wires together but do not twist them together with the power wires. If possible, shield the sense wires. Connect the shield to the case. Keep the total cable length as short as possible. Use low resistance fixture contacts.
Remote Sense Connections
The sense connections provide remote sense capability at the fixture. Sense connections on each card are through the same connectors that house the power connections.
Remote sensing allows the output voltages to be sensed at the cell, thus compensating for any losses in the wiring. On the Agilent E4374A cards, the compliance voltage (the voltage that the Agilent MCCD can provide in excess of the programmable rating) can be up to 5.5 volts to compensate for any IR voltage drop in the wiring between the channel output and the cell connections. This higher compliance voltage allows the full 5 V to be applied directly to the cell with a maximum of 0.5 volt loss in the wiring. If the charging voltage at the lithium ion cell is between 4.0 and 4.1 volts, the higher compliance voltage can compensate for a maximum of 1.4 volt to 1.5 volt loss in the wiring.
The following table gives the resistance values of various wire sizes so that you can calculate the voltage drops for various wire lengths and diameters. Larger and shorter wires result in lower voltage drops. The table also gives the maximum wire that limit the voltage drop to 1.4 volts with a maximum current of 2A. (1.4 volts is the difference between the 5.5 compliance voltage available at the power connector and the
4.1 volts required to charge a typical lithium ion cell.)
Table 2-5. Resistance of Stranded Copper Conductors
AWG No. mm
18 0.825 0.022 0.0066 30 20 0.519 0.034 0.0105 20 22 0.324 0.055 0.0169 12 24 0.205 0.087 0.0267 8
2
Resistance (at 20 deg. C) /m /ft
Maximum length in meters to limit
Voltage drop to 1.4 V @ 2A
(total length of + and - leads)
As an example, assume that you are using AWG #24 wire for your power connections and your charging voltage is 4.1 volts at 2 amperes. Using this diameter wire and assuming a maximum current of 2 amperes, the maximum distance from the power connector to the cell is limited to about 4 meters. This is because with a total wire length of 8 meters for both the + and power leads, the maximum voltage drop in the wiring is 1.4 volts (2A X 0.7
). With a charging voltage of 4.1 volts required at the cell and a
compliance voltage of 5.5 volts, this is the maximum voltage drop that the Agilent MCCD can tolerate.
NOTE: This example does not account for any additional lead path resistance that may be
present such as fixture contact resistance, or fixture relays. If additional resistance is present, lead length must be reduced yet further.
27
2 - Installation
Power Bus Connections
CAUTION: Observe polarity when making the power bus connections to both the Agilent MCCD
mainframe and the Agilent Powerbus Load. Reversed polarity connections will result in damage to both the Agilent MCCD mainframe and the Agilent Powerbus load. The
negative () bus bar on the Agilent MCCD mainframe is connected to chassis ground.
Connections to the power bus are made via + and bus bars on the back of the Agilent E4370A MCCD mainframe and Agilent E4371A Powerbus Load units. These bus bars let you interconnect multiple mainframes, external power sources, and other loads. Bus bars have mounting holes that accept 7 mm diameter bolts.
NOTE: Fasten a suitable terminal lug to each power bus cable. Do not connect bare wires
directly to the bus bars. Stranded cables with more and smaller diameter wires are easier to work with than cables with fewer and large diameter wires.
When making your power connections you can use discrete terminated wires, bus bars, or combinations of both. For proper operation all power bus configurations should have minimum loop area for low magnetic radiation and should be kept away from CRTs. The following guidelines may be helpful in deciding whether to use wires or bus bars.
Discrete terminated wires:
Are the better solution for connecting to individual units and the total current carrying requirements
are 120 A or less.
Have minimal alignment, insulation or routing problems. Are preferred for small cell charging systems.
Bus bars:
Are the better solution for high current carrying requirements. Can be custom designed or purchased; can use standard high current building parts. Use nuts and bolts or self tapped holes for connections. Require careful surface preparation and cleaning at connection points.
WARNING ENERGY HAZARD. If high current power bus connections touch, severe arcing
may occur - resulting in burns, ignition, or welding of parts. Do not attempt to make any connections to the power bus when the power bus is live.
Power Bus Wiring Information
The following table provides information about the resistance and ampacity of several standard wire sizes that may be suitable for power bus connections. This information is important because the resistance of the power bus wiring will cause a voltage drop in the power bus wires. If the voltage drop is large enough, it may prevent the Agilent E4370A MCCD mainframe from operating correctly in charging mode, or the Agilent E4371A Powerbus Load from operating correctly in discharging mode.
28
Installation - 2
Table 2-6. Ampacity and Resistance of Stranded Copper Conductors
AWG No. Area
in mm
10
8 6 4
2 1/0 2/0 3/0 4/0
5.26
8.36
13.3
21.1
33.6
53.5
67.4
85.0 107
Ampacity Resistance
2
40 60
80 105 140 195 225 260 300
in /meter
0.00327
0.00206
0.00129
0.00081
0.00051
0.00032
0.00025
0.00020
0.00016
Resistance
in /feet
0.00099
0.00062
0.00039
0.00025
0.000156
0.000098
0.000078
0.000062
0.000049
Notes
1. Wire ampacities are based on 30° C ambient temperature with conductor rated at 60° C.
2. Resistance is nominal at 20° C wire temperature.
Figure 2-1 illustrates two typical power bus configurations consisting of two Agilent E4370A MCCD mainframes connected to one Agilent E4371A Powerbus Load and two external dc power supplies. As shown in the figure, current requirements may vary widely based on the way the equipment is connected to the power bus.
+ -
charging = 133A
+
P
o
r
S
w
e
o
u
r
c
(24 V
@
_
133 A
e
)
gile
A
(256 channels)
A
(256 channels)
gilen
nt E
MCCD
t E
MCCD
4370A
4370A
+
_
+
3
3
1
_
Flexible Wires
133
/
A
charging = 133A
maximum charging current = 266A
+
P
o
w
e
r
S
u
o
r
c
e
133 A
@
(24 V
_
A
3
3
1
A
/87
A
+
P
r
e
w
+
133A
_
7A
8
T
e
r
m
l
i
a
n
B
l
c
o
k
o
(24 V @ 133 A)
_
1
7
4
A
+
A
gi
Powerbus Load
_
lent E
S
u
o
r
c
4371A
)
discharging = 174A
e
maximum discharging current = 174A
charging = 133A discharging = 87A
charging = 133A discharging = 87A
+
P
w
e
o
r
S
(24 V @ 133 A)
_
+
A
gile
nt E
Powerbus Load
_
+
A
gilent E
MCCD
(256 channels)
_
+
A
gilen
t E
MCCD
(256 channels)
_
u
o
r
4371A
4370A
4370A
c
e
STAR CONFIGURATION
Figure 2-1. Typical Power Bus Configurations
Rigid Bars
Flexible Wires
BUS BAR CONFIGURATION
29
2 - Installation
The star configuration on the left is designed so that each section of the power bus carries no more current than the rating of the equipment that it is connected to. This configuration lets you use longer lead lengths because the voltage drop in each lead is directly related to the amount of current flowing in the lead. However, this configuration requires you to run separate leads from each Agilent MCCD mainframe to the load as well as the power supply, thus increasing the total amount of wiring required.
The bus bar configuration on the right is designed to minimize the amount of wiring between the equipment. However this requires larger diameter wires or bus bars. This is because the leads from the power supplies as well as the leads to the load are required to carry the full charging and discharging current for two Agilent E4370A MCCD mainframes. Larger currents result in larger voltage drops in the wiring, which may prove unacceptable with long lead lengths.
Charging Mode Guidelines:
Power bus wires must be capable of handing the full charging current requirements of all Agilent E4370A MCCD units connected to the power bus. In the example that follows, the calculations are for worst case current requirements. Calculate the input current requirement of one fully loaded Agilent E4370A MCCD as follows:
1. Multiply the power used by one cell times the number of cells in the Agilent MCCD. Divide the result by the efficiency of the unit to determine the total input power required for that mainframe. The efficiency of the unit in charging mode is assumed to be 80%, which is a worst-case value as far as calculating the total power required by the mainframe.
#_of_cells × power_per_cell
0.8
= Max_power_in
2. Divide the input power requirements of the Agilent MCCD by the minimum voltage required at the input terminals of the Agilent MCCD. The result will be the maximum charging current required by the Agilent MCCD. (Double this current if you are simultaneously charging two Agilent MCCD mainframes as illustrated in Figure 2-1.)
Max_ power_in Power_source_voltage
= Max_powerbus_current
3. Determine the voltage drop that the maximum current will produce in the power bus leads using the resistance values in Table 2-6.
4. Add this voltage drop to the minimum voltage required at the input terminals of the Agilent MCCD to determine the output voltage setting of the dc power supply.
5. The voltage at the input terminals of the Agilent MCCD during charging mode must be between 25.2 and 22.8 volts. If the sum of the voltage drops in both the + and power bus leads causes the voltage at the mainframe power terminals to drop below 22.8 volts, the Agilent E4370A MCCD will shut down due to an undervoltage condition. Use a larger size wire to reduce the voltage drop.
Discharging Mode Guidelines:
Power bus wires must also capable of handing the full discharging current requirements of all Agilent E4370A MCCD units connected to the power bus. In the example that follows, the calculations are also for worst case current requirements. Calculate the output current of one fully loaded Agilent E4370A MCCD as follows:
30
Loading...
+ 111 hidden pages