The Agilent E3633A and Agilent E3634A are high performance 200 watt singleoutput dual range programmable DC power supplies with both GPIB and RS232 interfaces. The combination of bench-top and system features in these
power supplies provides versatile solutions for your design and test
requirements.
1 8V/20A* or 25V/7A** range selection key Selects the 8V/20A or 25V/7A
range and allows the full rated output to 8V/20A or 25V/7A.
2 20V/10A* or 50V/4A** range selection key Selects the 20V/10A or
50V/4A range and allows the full rated output to 20V/10A or 50V/4A.
3 Overvoltage protection key Enables or disables the overvoltage protection
function, sets trip voltage level, and clears the overvoltage condition.
4 Overcurrent protection key Enables or disables the overcurrent protection
function, sets trip current level, and clears the overcurrent condition.
5 Display limit key Shows voltage and current limit values on the display and
allows knob adjustment for setting limit values.
6 Recall operating state key Recalls a previously stored operating state from
location ‘‘1’’, ‘‘2’’, or ‘‘3’’.
7 Store operating state / Local key
1
Stores an operating state in location ‘‘1’’,
‘‘2’’, or ‘‘3’’ / or returns the power supply to local mode from remote interface
mode.
8 E r ror / Cal ibrate k e y
2
Displays error codes generated during operation, selftest and calibration / or enables calibration mode (the power supply must be
unsecured before performing calibration). See Service Guide for more details
on calibration.
9 I/O Configuration / Secure key
3
Configures the power supply for remote
interfaces / or secure or unsecure the power supply for calibration. See
Service Guide for more details on how to secure or unsecure the power supply.
10 Output On/Off key Enables or disables the power supply output. This key
toggles between on and off.
11 Control knob Increases or decreases the value of the blinking digit by turning
clockwise or counter clockwise.
12 Resolution selection keys Move the blinking digit to the right or left.
13 Voltage/current adjust selection key Selects the knob control function for
voltage or current adjustment.
1
The key can be used as the ‘‘Local’’ key when the power supply is in the remote
interface mode.
2
You can enable the ‘‘calibration mode’’ by holding down this key when you
turn on the power supply.
3
You can use it as the ‘‘Secure’’ or ‘‘Unsecure’’ key when the power supply is
in the calibration mode.
*For Agilent E3633A Model **For Agilent E3634A Model
3
Front-Panel Voltage and Current Limit Settings
You can set the voltage and current limit values from the front panel using the
following method.
Use the voltage/current adjust selection key, the resolution selection keys,
and the control knob to change the voltage and current limit values.
1 Select the desired range using the range selection keys after turning on the
power supply.
2 Press the
3 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired voltage limit
by turning the control knob. If the display limit times out, press the key
again.
4 Set the knob to current control mode by pressing the key.
5 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired current limit
by turning the control knob.
6 Press the
will go to output monitoring mode automatically to display the voltage and
current at the output or the display will go to output monitoring mode
immediately by pressing the
Display
key to show the limit values on the display.
Limit
Volt ag e
Current
Output
key to enable the output. After about 5 seconds, the display
On/Off
Output
key again.
On/Off
Display
Limit
Note All front panel keys and controls can be disabled with remote interface commands.
The Agilent E3633A and Agilent E3634A must be in "Local" mode for the front panel
keys and controls to function.
4
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
8VShows the 8V/20A range is selected. (Agilent E3633A model)
20VShows the 20V/10A range is selected. (Agilent E3633A model)
25VShows the 25V/7A range is selected. (Agilent E3634A model)
50VShows the 50V/4A range is selected. (Agilent E3634A model)
OVPThe overvoltage protection function is enabled when the
annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
OCPThe overcurrent protection function is enabled when the
annunciator turns on or the overcurrent protection circuit has
caused the power supply to shutdown when the annunciator blinks.
CALThe power supply is in calibration mode.
LimitThe display shows the limit values of voltage and current.
ERROR Hardware or remote interface command errors are detected and
the error bit has not been cleared.
OFFThe output of the power supply is disabled (For more information,
see page 52 in the User’s Guide).
UnregThe output of the power supply is unregulated (output is neither CV
nor CC).
CVThe power supply is in constant voltage mode.
CCThe power supply is in constant current mode.
To review the display annunciators, hold down key as you turn on
Display
Limit
the power supply.
5
The Rear Panel at a Glance
1 Power-line voltage setting
2 Power-line fuse-holder assembly
3 AC inlet
• Select the GPIB or RS-232 interface (see chapter 3 in User’s Guide).
• Set the GPIB bus address (see chapter 3 in User’s Guide).
• Set the RS-232 baud rate and parity (see chapter 3 in User’s Guide).
6
In This Book
This is the Service Guide for your Agilent E3633A and E3634A DC power
supplies. Unless otherwise stated, the information in this manual applies to
both two models.
Specifications Chapter 1 lists the power supply’s specifications and
describes how to interpret these specifications.
Quick Start Chapter 2 prepares the power supply for use and helps you get
familiar with the front panel features.
Calibration Procedures Chapter 3 provides performance verification and
calibration procedures.
Theory of Operation Chapter 4 describes block and circuit level theory
related to the operation of the power supply.
Service Chapter 5 provides guidelines for returning your power supply to
Agilent Technologies for servicing, or for servicing it yourself.
Replaceable Parts Chapter 6 contains a detailed parts list of the power
supply.
Backdating Chapter 7 describes the difference between this manual and
older issues of this manual.
Schematics Chapter 8 co ntains th e po wer sup ply ’s sch ematic s, disasse mbly
drawings, and component locator drawings.
If you have questions relating to the operation of the power supply, call
1-800-452-4844 in the United States, or contact your nearest Agilent
Technologies Sales Office.
If your Agilent E3633A or Agilent E3634A fails within three years of purchase,
Agilent will repair or replace it free of charge. Call 1-800-258-5165 ("Express Exchange") in the United States, or contact your nearest Agilent Technologies
Sales Office.
To Prepare the Power Supply for Use - - - - - - - - - - - - - - - - - - - - - 23
To Check the Rated Voltages of the Power Supply - - - - - - - - - - - 25
To Check the Rated Currents of the Power Supply - - - - - - - - - - 26
To Use the Power Supply in Constant Voltage Mode - - - - - - - - - 28
To Use the Power Supply in Constant Current Mode - - - - - - - - - 30
To Store and Recall the Instrument State - - - - - - - - - - - - - - - - - - 32
To Program Overvoltage Protection - - - - - - - - - - - - - - - - - - - - - - 34
The performance specifications are listed in the following pages.
Specifications are warranted in the temperature range of 0 to 40°C with a
resistive load. Supplemental characteristics, which are not warranted but
are descriptions of performance determined either by design or testing.
Chapter 3 ‘‘Calibration Procedures’’ contains procedures for verifying the
performance specifications.
Less than 50 msec for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Command Processing Time
Average time for output voltage to begin to change after receipt of digital data
when the power supply is connected directly to the GPIB or RS-232 is less than
100 msec
OVP and OCP Accuracy, ±(% of output + offset)
OVP0.5% + 0.5 V
OCP0.5% + 0.5 A
Activation time : Average time for output to start to drop after OVP or OCP
condition occurs.
OVP<1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
OCP<10 msec
16
Chapter 1 Specifications
Supplemental Characteristics
Supplemental Characteristics
Table 1-2. Supplemental Characteristics
ParameterAgilent E3633AAgilent E3634A
Output Programming Range
(maximum programmable values)
Voltage Programming
Speed: Maximum time required
for output voltage to settle within
1% of its total excursion (for
resistive load). Excludes
command processing time.
Low Range 0 to +8.24 V/
0 to 20.6 A
High Range 0 to +20.6 V/
0 to 10.3 A
OVP 1 V to 22 V 1 V to 55 V
OCP 0 A to 22 A 0 A to 7.5 A
Full Load No Load Full Load No Load
Up95 msec45 msec80 msec100 msec
Down30 msec450 msec 30 msec450 msec
0 to +25.75 V/
0 to 7.21 A
0 to +51.5V/
0 to 4.12 A
1
Remote Sensing Capability
Voltage dropUp to 0.7 V per each lead
Load regulationAdd 5 mV to spec for each 1-volt change in the + output
lead due to load current changes.
Load voltageSubtract voltage drop in load leads from specified output
voltage rating.
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage0.01% + 3 mV
Current0.02% + 3 mA
Stability, ±(% of output + offset)
Following 1 hour warm-up, change in output over 8 hours under constant load,
line, and ambient temperature
Voltage0.02% + 1 mV
Current0.1% + 1 mA
17
Chapter 1 Specifications
Supplemental Characteristics
Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed
1 V if the output control is set to less than 1 V. If the output control is set to
1 V or higher, there is no overshoot.
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Three (3) user-configurable stored states
Recommended Calibration Interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the
(+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output
to the (+) sense and the (-) output and the (-) sense terminals.
AC Input Ratings (selectable via rear panel selector)
std115 Vac ± 10%, 47 to 63 Hz
opt 0E3230 Vac ± 10%, 47 to 63 Hz
opt 0E9100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
700 VA with full load
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for full rated output. At higher temperatures, the output current is
derated linearly to 50% at 55 °C maximum temperature.
18
Chapter 1 Specifications
Supplemental Characteristics
Storage Temperature
-20 to 70 °C for storage environment.
Environmental Conditions
Designed for indoor use in an installation category II, pollution degree 2
environment. Designed to operate at a maximum relative humidity of 95 %
and at altitudes of up to 2000 meters.
Dimensions*
213 mmW x 133 mmH x 348 mmD (8.4 x 5.2 x 13.7 in)
*See below for detailed information.
Weight
Net9.5 kg (21 lb)
Shipping12 kg (26 lb)
1
Figure 8-1. Dimensions of Agilent E3633A and E3634A Power Supplies
19
Chapter 1 Specifications
Supplemental Characteristics
20
2
Quick Start
Quick Start
One of the first things you will want to do with your power supply is to become
acquainted with its front panel. Written procedures in this chapter prepare the
power supply for use and familiarize you with most front-panel operations.
• The power supply is shipped from the factory configured in the front-panel
operation mode. At power-on, the power supply is automatically set to
operate in the front-panel operation mode. When in this mode, the frontpanel keys can be used. When the power supply is in remote operation mode,
you can return to front-panel operation mode at any time by pressing the
Store
(Local) key if you did not previously send the front-panel lockout
Local
command. A change between front-panel and remote operation modes will
not result in a change in the output parameters.
• The power supply has two output ranges. This feature allows more voltage
at a lower current or more current at a lower voltage. The desired output
range is selected from the front panel or over the remote interfaces. The
20V for the E3633A and 25V or 50V for the E3634A annunciator indicates
or
the presently selected range.
• When you press key (the
Display
Limit
Limit annunciator flashes), the display of
the power supply goes to the limit mode and the present limit values will
be displayed. In this mode, you can also observe the change of the limit
values when adjusting the knob. If you press the key again or let the
Display
Limit
display time-out after several seconds, the power supply will return the
display to the meter mode (the
Limit annunciator turns off). In this mode,
the actual output voltage and current will be displayed.
• The output of the power supply can be enabled or disabled from the front
panel by pressing key. When the output is off, the
Output
On/Off
OFF annunciator
turns on and the output is disabled.
• The display provides the present operating status of the power supply with
annunciators and also informs the user of error codes. For example, the
power supply is operating in CV mode in the 8V/20A* or 25V/7A** range and
controlled from the front panel, then the
CV and 8V* or 25V** annunciators
will turn on. If, however, the power supply is remotely controlled, the
annunciator will also turn on, and when the power supply is being addressed
over GPIB interface, the
Adrs annunciator will turn on. See “Display
Annunciators’’ on page 5 for more information.
8V
Rmt
*For Agilent E3633A Model **For Agilent E3634A Model
22
Chapter 2 Quick Start
To Prepare the Power Supply for Use
To Prepare the Power Supply for Use
The following steps help you verify that the power supply is ready for use.
Power
Output
On/Off
1 Check the list of supplied items.
Verify that you have received the following items with your power supply. If
anything is missing, contact your nearest Agilent Technologies Sales Office.
One power cord for your location.
One User’s Guide.
This Service Guide.
Certificate of Calibration.
2 Verify that the correct power-line voltage setting is selected and that
the correct power-line fuse is installed.
The line voltage is set to 100, 115 or 230 Vac from the factory according to the
input power option selected when you ordered the power supply. Change the
voltage setting if it is not correct for your location (see the next page for
detailed information). For 100 or 115 Vac operation, the correct fuse is 6.3 AT
(Agilent part number 2110-1030) and for 230 Vac operation, the correct fuse is
3.15 AT (Agilent part number 2110-1031).
3 Connect the power cord and turn on the power supply.
A power-on self-test occurs automatically when you turn on the power supply.
The front-panel display will light up while the power supply performs its
power-on self- test. After performing the self-test, the power supply will go into
the power-on / reset state; the output is disabled (the
on); the 8V/20A* or 25V/7A** range is selected (the
OFF annunciator turns
8V* or 25V** annunciator
turns on); the knob is selected for voltage control.
Notice that the
OVP and OCP annunciators also turn on.
4 Enable the outputs.
Output
Press key to enable the outputs. The
On/Off
8V* or 25**, OVP, OCP, and CV annunciators are lit. The flashing digit can be
OFF annunciator turns off and the
adjusted by turning the knob. Notice that the display is in the meter mode.
‘‘Meter mode’’ means that the display shows the actual output voltage and
current.
2
*For Agilent E3633A Model **For Agilent E3634A Model
23
Chapter 2 Quick Start
To Prepare the Power Supply for Use
1
Remove the power cord. Remove the
fuse-holder assembly with a flat-blade
screwdriver from the rear panel.
2
Install the correct line fuse. Remove
the power-line voltage selector from the
power-line module.
100 or 115 Vac, 6.3 AT fuse
230 Vac, 3.15 AT fuse
3
Rotate the power-line voltage selector
until the correct voltage appears.
4
Replace the power-line voltage selector
and the fuse-holder assembly in the rear
panel.
100, 115, or 230 Vac
Install the correct fuse and verify that the correct line voltage appears in the
window.
24
Power
Output
On/Off
Chapter 2 Quick Start
To Check the Rated Voltages of the Power Supply
To Check the Rated Voltages of the Power Supply
The following procedures check to ensure that the power supply produces its
rated voltage output with no load and properly responds to operation from the
front panel.
For each step, use the keys shown on the left margins.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
(the
8V* or 25V** annunciator turns on); and the knob is selected for voltage control.
2 Enable the outputs.
OFF annunciator turns off and the 8V* or 25V**, OVP, OCP, and CV
The
annunciators are lit. The flashing digit can be adjusted by turning the knob.
Notice that the display is in the meter mode. “Meter mode’’ means that the
display shows the actual output voltage and current.
2
3 Check that the front-panel voltmeter properly responds to knob
control.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero.
1
4 Ensure that the voltage can be adjusted from zero to the maximum
rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob
until the voltmeter indicates ‘‘8.0 volts’’* or ‘‘25.0 volts’’**.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the voltage.
*For Agilent E3633A Model **For Agilent E3634A Model
25
Power
Output
On/Off
Display
Limit
Volt ag e
Current
Chapter 2 Quick Start
To Check the Rated Currents of the Power Supply
To Check the Rated Currents of the Power Supply
The following procedures check to ensure that the power supply produces its
rated current outputs with a short and properly responds to operation from
the front panel.
For each step, use the keys shown on the left margins.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
(the
8V* or 25V** annunciator turns on); and the knob is selected for voltage
control.
2 Connect a short across (+) and (-) output terminals with an insulated
test lead.
3 Enable the outputs.
OFF annunciator turns off and the 8V* or 25V**, OVP, and OCP
The
annunciators turn on. The
resistance of the test lead. The flashing digit can be adjusted by turning the
knob. Notice that the display is in the meter mode. “Meter mode’’ means that
the display shows the actual output voltage and current.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
the voltage limit to 1.0 volt to assure CC operation. The
light.
5 Check that the front-panel ammeter properly responds to knob control.
Set the knob to the current control, and turn the knob clockwise or counter
clockwise when the display is in the meter mode (the
Check that the ammeter responds to knob control and the voltmeter indicates
nearly zero (actually, the voltmeter will show the voltage drop caused by the
test lead).
CV or CC annunciator is lit depending on the
Lmt annunciator will be flashing). Adjust
CC annunciator will
Lmt annunciator is off).
*For Agilent E3633A Model **For Agilent E3634A Model
26
Chapter 2 Quick Start
To Check the Rated Currents of the Power Supply
1
6 Ensure that the current can be adjusted from zero to the maximum
rated value.
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter
indicates ‘‘20.0 amps’’* or ‘‘7.0 amps’’**.
Note If an error has been detected during the output checkout procedures, the ERROR
annunciator will turn on. See “Error Messages’’ for more information, starting on
page 123 in chapter 5 of the User’s Guide.
2
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the voltage.
*For Agilent E3633A Model **For Agilent E3634A Model
27
Power
Display
Limit
Chapter 2 Quick Start
To Use the Power Supply in Constant Voltage Mode
To Use the Power Supply in Constant Voltage Mode
To set up the power supply for constant voltage (CV) operation, proceed as
follows.
For each step, use the keys shown on the left margin.
1 Connect a load to the desired output terminals.
With power-off, connect a load to the desired output terminals.
2 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
(the
8V* or 25V** annunciator turns on); and the knob is selected for voltage control.
20V,10A50V,4A
Press * or ** key to operate the power supply in the 20V/10A* or
50V/4A** range before proceeding to the next step. The
annunciator turns on.
3 Set the display to the limit mode.
Notice that the
Limit annunciator flashes, indicating that the display is in the
limit mode. When the disp lay is in the limit mode, you can see the voltage and
current limit values of the power supply.
20V* or 50V**
Volt ag e
Current
In constant voltage mode, the voltage values between the meter and
limit modes are the same, but the current values are not. Moreover, if the
display is in the meter mode, you cannot see the change of current limit
value when adjusting the knob. We recommend that you should set the
display to “limit” mode to see the change of current limit value in the
constant voltage mode whenever adjusting the knob.
1
4 Adjust the knob for the desired current limit.
Check that the
Limit annunciator still flashes. Set the knob for current control.
The second digit of the ammeter will be flashing. The flashing digit can be
changed using the resolution selection keys and the flashing digit can be
adjusted by turning the knob. Adjust the knob to the desired current limit.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting current.
*For Agilent E3633A Model **For Agilent E3634A Model
28
Chapter 2 Quick Start
To Use the Power Supply in Constant Voltage Mode
Volt ag e
Current
1
5 Adjust the knob for the desired output voltage.
Check that the
Limit annunciator still flashes. Set the knob for voltage control.
The second digit of the voltmeter will be flashing. Change the flashing digit
using the resolution selection keys and adjust the knob to the desired output
voltage.
Display
Limit
6 Return to the meter mode.
Display
Press key or let the display time-out after several seconds to return to
Limit
the meter mode. Notice that the
Limit annunciator turns off and the display
shows “OUTPUT OFF” message.
Output
On/Off
7 Enable the output.
OFF annunciator turns off and the 8V* (or 25V**) or 20V* (or 50V**), OVP,
The
OCP and CV annunciators are lit. Notice that the display is in the meter mode.
In the meter mode, the display shows the actual output voltage and current.
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify that
CV annunciator is lit. If the CC annunciator is lit, choose a higher current
the
limit.
Note During actual CV operation, if a load change causes the current limit to be exceeded,
the power supply will automatically crossover to the constant current mode at the
preset current limit and the output voltage will drop proportionately.
2
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
*For Agilent E3633A Model **For Agilent E3634A Model
29
Loading...
+ 106 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.