Seattle Sound and Vibration, inc.
Print Date: December, 2000
ãSeattle Sound and Vibration, inc. 1996, 2000.
All rights reserved.
Agilent Part number 35670-90059
Special PermissionAmoco Corporation and Agilent Technologies, Inc are granted
full editing, duplication and distribution privileges.
AGILENT 35670A
Supplemental Operator’s Guide
SECTION 1 : TURNING ON THE AGILENT 35670A .......................................................................................... 1
SECTION 2 : MEASUREMENT STATE AFTER TURN-ON............................................................................... 2
SECTION 3 : MEASURING WITH TRANSDUCERS ...........................................................................................3
U
SING ICP-TYPE TRANSDUCERS WITH THE AGILENT 35670A .............................................................................. 3
U
SING TRANSDUCERS WITH EXTERNAL PREAMPLIFIERS .......................................................................................... 3
S
PECIFYING TRANSDUCER SENSITIVITY AND UNITS ................................................................................................. 4
SECTION 4 : MEASURING A SINGLE CHANNEL POWER SPECTRUM - DISPLACEMENT................... 5
S
ELECTING SINGLE CHANNEL OPERATION ............................................................................................................... 5
ETTING UP TRANSDUCER UNITS FOR DISPLACEMENT............................................................................................. 5
S
S
ELECTING FREQUENCY SPAN .................................................................................................................................. 5
Q
UANTIFYING POWER SPECTRUM RESULTS.............................................................................................................. 7
REQUENCY ZOOMING TO INCREASE RESOLUTION................................................................................................... 8
M
AXIMIZING DYNAMIC RANGE ................................................................................................................................8
VERAGING TO REDUCE MEASUREMENT VARIANCE................................................................................................ 9
A
SECTION 6 : MEASURING A SINGLE CHANNEL TIME WAVEFORM ...................................................... 11
S
ELECTING SINGLE CHANNEL OPERATION ............................................................................................................. 11
S
ELECTING TIME RECORD LENGTH......................................................................................................................... 12
D
ISPLAYING TIME WAVEFORMS ............................................................................................................................. 12
U
SING MANUAL ARM TO CAPTURE A SINGLE TIME WAVEFORM............................................................................ 12
RESETTING THE ANALYZER................................................................................................................................... 15
ELECTING FREQUENCY SPAN ................................................................................................................................15
S
D
ISPLAYING DUAL-CHANNEL POWER SPECTRUM MEASUREMENTS....................................................................... 15
U
SING MARKERS WITH DUAL CHANNEL MEASUREMENTS ..................................................................................... 16
AXIMIZING DYNAMIC RANGE .............................................................................................................................. 18
ELECTING TIME RECORD LENGTH......................................................................................................................... 20
D
ISPLAYING DUAL TIME WAVEFORMS ................................................................................................................... 20
RMING DUAL CHANNEL TIME MEASUREMENTS ................................................................................................... 21
A
T
RIGGER ING A DUAL-CHANNEL TIME MEASUREMENT........................................................................................... 21
SING MARKERS WITH DUAL CHANNEL TIME MEASUREMENTS ............................................................................ 23
SECTION 10 : MEASURING DUAL CHANNEL SPECTRA AND TIME WAVEFORMS.......................... 23
AGILENT 35670A
Supplemental Operator’s Guide
SECTION 11 : MODAL TESTING USING A HAMMER AND ACCELEROMETER ................................. 25
S
ETTING UP THE TRANSDUCER PARAMETERS ......................................................................................................... 25
ETTING UP INPUT RANGE ...................................................................................................................................... 26
S
C
HOOSING A PRELIMINARY FREQUENCY SPAN....................................................................................................... 26
ETTING UP TIME DISPLAYS ................................................................................................................................... 27
U
SING FORCE/RESPONSE WINDOWS ....................................................................................................................... 28
D
ISPLAYING HAMMER TEST RESULTS .................................................................................................................... 29
HANGING FREQUENCY SPAN ................................................................................................................................31
SECTION 12 : HAMMER TEST SETUP WITHOUT A FORCE TRANSDUCER........................................ 32
S
ETTING UP THE HAMMER TEST WITHOUT A FORCE TRANSDUCER ....................................................................... 32
SECTION 13 : ORDER DOMAIN RESULTS IN LIST MODE........................................................................ 33
ISPLAYING DUAL-CHANNEL SPECTRA COMPARED WITH RECALLED DATA ......................................................... 35
R
ECALLING THE SPECTRA FROM DISK .................................................................................................................... 36
R
ECALLING THE SPECTRA FROM NON-VOLATILE RAM ......................................................................................... 36
S
CALING THE DISPLAYS .......................................................................................................................................... 37
U
SING MARKERS TO COMPARE............................................................................................................................... 38
SECTION 15 : WATERFALL SPECTRA AT TIME INTERVALS ................................................................ 39
S
ELECTING SINGLE CHANNEL OPERATION ............................................................................................................. 39
ETTING UP THE WATERFALL DISPLAY.................................................................................................................. 39
S
ETTING UP TIME STEP ARM .................................................................................................................................. 40
SING SLICE MARKERS WITH WATERFALL DATA .................................................................................................. 40
U
C
OLLECTING A SPECIFIED NUMBER OF SPECTRA DURING A TIME INTERVAL ......................................................... 41
SECTION 16 : WATERFALL SPECTRA AT RPM INTERVALS .................................................................. 42
S
ETTING UP THE TACHOMETER .............................................................................................................................. 42
S
PECIFYING THE START RPM ................................................................................................................................. 42
C
OLLECTING A SPECIFIED NUMBER OF SPECTRA DURING AN RPM INTERVAL....................................................... 42
TARTING AND PAUS ING A MEASUREMENT ............................................................................................................ 43
S
S
CALING THE DISPLAY............................................................................................................................................ 43
SECTION 19 : TWO SPECTRAL TRACES SHOWING MILS AND IPS WHILE EU IS G........................ 56
SECTION 20 : PEAK HOLD DURING A MACHINE RUN-UP AND COAST-DOWN ................................ 58
AGILENT 35670A
Supplemental Operator’s Guide
SECTION 21 : USING AN EXTERNAL TRIGGER FOR TIME AVERAGING ........................................... 60
S
PECIFY USER LEVELS FOR TRIGGERING ................................................................................................................ 60
HARACTERIZING THE EXTERNAL TRIGGER ........................................................................................................... 60
C
S
ETTING UP EXTERNAL TRIGGER ........................................................................................................................... 61
M
EASURING TIME AVERAGED SPECTRUMS WITH EXTERNAL TRIGGERING............................................................. 62
SECTION 22 : SAVING TRACE DATA TO A 3.5” DISK OR NON-VOLATILE RAM .............................. 63
S
AVING TRACE TO 3.5” DISK .................................................................................................................................. 63
C
ONFIRMING CONTENTS OF 3.5” DISK FILE ............................................................................................................ 63
AVING TRACE TO NON-VOLATILE RAM (NV-RAM)........................................................................................... 64
S
C
ONFIRMING CONTENTS OF NV-RAM ................................................................................................................... 64
SECTION 23 : RECALLING TRACE DATA FROM 3.5” DISK OR NON-VOLATILE MEMORY.......... 65
R
ECALL TRACE FROM 3.5” DISK ............................................................................................................................ 65
R
ECALL TRACE DATA FROM NON-VOLATILE RAM (NV-RAM)............................................................................ 65
SECTION 24 : PLOTTING AND PRINTING TRACE DATA ......................................................................... 67
G
ENERATING OUTPUT WITH THE AGILENT 35670A............................................................................................. 67
P
LOTTING THE DISPLAY .......................................................................................................................................... 67
P
RINTING THE DISPLAY........................................................................................................................................... 68
SECTION 25 : IMPORTING PLOTS INTO MICROSOFT WORD................................................................ 69
D
ETERMINING IF MS WORD HAS HP-GL GRAPHICS IMPORT FILTER..................................................................... 69
I
NSTALLING THE MS WORD HP-GL GRAPHICS IMPORT FILTER............................................................................. 69
LOT TO A FILE USING THE AGILENT 35670A ..................................................................................................... 70
P
I
NSERTING PICTURE OF AN HP-GL PLOT FILE INTO A MS WORD DOCUMENT. ...................................................... 70
SECTION 26 : RUNNING A SINGLE CALIBRATION TEST ON COMMAND .......................................... 71
SECTION 27 : RETURNING THE AGILENT 35670A TO A PRESET CONDITION.................................. 71
SECTION 28 : PRECAUTIONS TO PREVENT LOSS OF DATA .................................................................. 71
SECTION 29 : TRANSDUCER UNIT CONVERSION WITH THE AGILENT 35670A............................... 72
ATH FUNCTIONS AND XDCR UNIT CONVERT................................................................................................. 73
C
ONVERTING FREQUENCY RESPONSE UNITS TO COMPLIANCE ............................................................................... 73
SECTION 30 : USING THE AGILENT DEMO UNIT WITH THE AGILENT 35670A................................ 75
Section 1 :Turning On the AGILENT 35670A
A
The entire turn-on procedure for the AGILENT 35670A takes 40 seconds. In
that time the analyzer completes a hardware self-test, boots its operating system,
completes a front end self-calibration and autoranges all input channels. The
default measurement state of the analyzer is a two-channel FFT measurement
with a span of 51.2 kHz, front ends autoranging, input low floating, and DC
coupled, averaging OFF and with a power spectrum display.
: CH1 Pwr Spec X:0 Hz Y:40.9385 uVrms
10
mVrms
LogMag
decades
nVrms
6
10
0Hz51.2kHz
AGILENT 35670A
Supplemental Operator’s Guide
FFT
ANALYSIS
OCTAVE
ANALYSIS
ORDER
ANALYSIS
SWEPT
SINE
CORRELATN
ANALYSIS
HISTOGRAM
/ TIME
CHANNELS
1
4
2
REF CHANS
1,3
1
TIME
CAPTURE
Typical Display After Turn-on Sequence
1
AGILENT 35670A
Supplemental Operator’s Guide
Section 2 :Measurement State After Turn-on
Inst Mode FFTChannels 2
Meas Data Data APWR SPEC1Data BPWR SPEC2
Trac Coord Coord A MAG-LOGCoord B MAG-LOG
FreqStart0 HzStop51.2 kHz
WindowCH1 Type FLATTOPBW488.88 Hz
AvgStatusOFFTypeRMS
TriggerTypeFREE-RUNLevel0 V
Measurement SourceINPUT
Data CTIME1Data DTIME2
Coord C REALCoord D REAL
Center25.6 kHzSpan51.2 kHz
Rec Len 7.8125 msResoltn 400
CH2 Type FLATTOPBW488.88 Hz
CH3 Type FLATTOPBW488.88 Hz
CH4 Type FLATTOPBW488.88 Hz
Number10Overlap 0 %
Fast Avg OFFUpdate5
Rept Avg FREEZEOvld Rej OFF
SlopePOSITIVE
Ext Lvl 1.48 V
Ext Fltr OFF
CH1 Dlay 0 sCH2 Dlay 0 s
CH3 Dlay 0 sCH4 Dlay 0 s
ArmIMMEDIATEStrt RPM 600 rpm
RPM Step 60 rpmTime Stp 500 ms
Wtrfall 15RPM Dir UP
SourceStatusOFF
InputCHANNEL 1CHANNEL 2CHANNEL 3CHANNEL 4
StatusONONOFFOFF
Range3.9858 mVp 3.9858 mVp 3.9858 mVp 3.9858 mVp
Auto Rng OFFOFFOFFOFF
Coupling DCDCDCDC
InputLow FLOATFLOATFLOATFLOAT
AliasFlt ONONONON
A WtFltr OFFOFFOFFOFF
ICPOFFOFFOFFOFF
EngrUnit OFFOFFOFFOFF
EU Label USERUSERUSERUSER
EU Mult1 V/EU1 V/EU1 V/EU1 V/EU
--- TACHOMETER ---
Puls/Rev 1
Level0 V
RangeLOW
SlopePOSITIVE
Holdoff 0 s
Level0 VpkDC Offst 0 V
TypeSINUSOIDSine Frq 10.24 kHz
Arb Fltr OFFArb Reg D1
Arb Rept ON
2
Section 3 :Measuring with Transducers
The AGILENT 35670A can be used with a broad range of transducers. Transducers
can use an external preamplifier to convert the raw signal into a voltage that is
compatible with the AGILENT 35670A. Some transducers have internal circuitry
which provides preamplification, however these internal circuits must be powered
externally. The AGILENT 35670A can provide the necessary constant-current
supply for transducers that are of type ICP, Isotron, Piezotron, DelaTron, Low
Impedance, etc.
Using ICP-Type Transducers with the AGILENT 35670A
To set up Channel 1 for an ICP-type transducer:
• Press the [Input] hardkey
• Press [FRONT END CH1 SETUP] softkey (F7).
• Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
Note that the analyzer automatically switches the front end to AC Coupling in order
to reject the large DC signal that is generated by the constant-current supply.
To set up Channel 2 for an ICP transducer:
• Press the [Input] hardkey
• Press [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
• Press [FRONT END CH2 SETUP] softkey (F7).
• Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.
AGILENT 35670A
Supplemental Operator’s Guide
For AGILENT 35670A analyzers with four channels, these same procedures can be
repeated for Channels 3 and 4.
Using Transducers with External Preamplifiers
No special front end setup is required to use transducers with external preamplifiers
except if the preamplifier’s DC component is too high, resulting in a loss of dynamic
range. This can be easily minimized by switching to AC coupling as follows:
• Press the [Input] hardkey
• Press [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
• Press [FRONT END CH1 SETUP] softkey (F7).
• Press [COUPLING AC DC] softkey (F5) until AC is highlighted.
A transducer with an external preamplifier will convert mechanical units (EU) to
volts. Care must be taken to observe gain settings on the external preamplifier,
because these gain settings directly affect the resulting transducer sensitivity.
3
AGILENT 35670A
Supplemental Operator’s Guide
For example, if an accelerometer has am internal sensitivity of 25 pC/g, and a charge
amplifier has a gain setting of 10 mV/pC, then the resulting transducer sensitivity is
250 mV/g. However is the gain setting on the charge amplifier is changed to 1
mV/pC, then the resulting sensitivity would then be 25 mV/g.
Specifying Transducer Sensitivity and Units
The AGILENT 35670A provides the ability to convert electrical signals into
Engineering Units. Units can be specified as either conventional units or User units.
Conventional units of motion available are:
To set up transducer properties for an accelerometer attached to Channel 2 with a
sensitivity of 9.7 pC/g, follow these steps:
• Press [Input] hardkey.
• Press [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
• Press [XDCR UNIT CH2 SETUP] softkey (F8)
• Press the [XDCR SENSITVTY] softkey (F6).
• Enter 9.7, then press the [mV/EU] softkey (F2).
• Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
• Press the [XDCR UNIT LABEL] softkey (F7).
• Select the [g] softkey (F2).
Likewise if a force transducer is attached to Channel 1 with a sensitivity of 12.2
mV/lb., do the following:
• Press [Input] hardkey.
• Press [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
• Press [XDCR UNIT CH1 SETUP] softkey (F8)
• Press the [XDCR SENSITVTY] softkey (F6).
• Enter 12.2, then press the [mV/EU] softkey (F2).
• Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.
• Press the [XDCR UNIT LABEL] softkey (F7).
• Press the [MORE CHOICES] softkey (F9).
• Select the [lb] softkey (F6).
4
AGILENT 35670A
Supplemental Operator’s Guide
Section 4 :Measuring a Single Channel Power Spectrum - Displacement
This section contains procedures for setting up and measuring a single channel
power spectrum of displacement using an AGILENT 35670A Dynamic Signal
Analyzer. Included are steps to select single channel operation, set up transducer
parameters, frequency span, choose input range, select windowing, and turn on
averaging in order to improve measurement accuracy. Instructions will also be
presented to set up trace coordinates and trace scaling.
Selecting Single Channel Operation
Specify the single-channel instrument mode to measure one signal. However it
should be noted that it is acceptable to measure a single channel in the default two-
channel instrument mode and simply ignore the second channel. Nevertheless
instructions will be given here to measure in single channel mode.
To measure a single-channel spectrum, turn on the AGILENT 35670A and then:
• Press the [Inst Mode] hardkey
• Press [CHANNELS 1 24 ] softkey (F7) until 1 is highlighted.
Setting Up Transducer Units for Displacement
To measure power spectrums using a displacement transducer with a sensitivity of
200 mV/mil, follow these steps:
• Press the [Input] hardkey.
• Enter 800 and then press the [Hz] softkey (F2).
Any frequency below 51.2 kHz may of course be entered, however the analyzer will
select the lowest span that contains the desired frequency. For example, if 1200 Hz
is entered, the analyzer is switched to a span of 1600 Hz.
Selecting Frequency Span
The default frequency span after turn-on is 51.2 kHz, which is usually too high for
most transducer measurements. To reduce the span to 800 Hz, perform the following
steps:
• Press the [Freq] hardkey.
• Enter 800 and then press the [Hz] softkey (F2).
5
AGILENT 35670A
Supplemental Operator’s Guide
Any frequency below 51.2 kHz may of course be entered, however the analyzer will
select the lowest span that contains the desired frequency. For example, if 1200 Hz
is entered, the analyzer is switched to a span of 1600 Hz.
* Available in single-channel mode when anti-alias filter is turned off.
To begin data acquisition,
• Press the [Start] hardkey.
To automatically scale the display to fit the data,
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
6
AGILENT 35670A
Supplemental Operator’s Guide
InputCH1 Rng: 501.7819 mVpk CH2 Rng: 631.706 mVpk
Date: 10-02-96 Time: 08:42:00 PM
A: CH1 Pwr Spec X:160 Hz Y:169.558 mVrms
Vrms
LogMag
decades
uVrms
CH3 Rng: 2.5149 Vpk CH4 Rng: 2.5149 Vpk
1
6
1
0Hz800Hz
Quantifying Power Spectrum Results
Use the marker knob to read the amplitude at any measured frequency. Rotate the
knob on the AGILENT 35670A front panel until the marker indicator (the small
diamond following the spectral trace) is placed at the frequency of interest. The
frequency and amplitude at this bin is presented as X and Y values at the top of the
trace. Units of [Hz] and [Vrms] or [grms] identify the results.
Special marker features such as peak tracking can speed analysis. To always track
the peak amplitude and frequency, follow these steps:
• Press the [Marker] hardkey.
• Press [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.
The frequency and amplitude of the peak bin in every trace will be determined and
displayed.
7
AGILENT 35670A
Supplemental Operator’s Guide
Section 5 :Improving Measurement Results
Frequency Zooming to Increase Resolution
The AGILENT 35670A has the ability to zoom in on a signal to show detail with
very fine resolution. After choosing a center frequency, it is possible to decrease the
span and therefore increase the resolution all the way down to 61 µHz. To zoom in
on a signal, move the marker to the desired center frequency, then:
• Press the [Freq] hardkey.
• Press the [CENTER] softkey (F2).
• Press the [Mkr Value] hardkey.
To set a new, lower span,
• Press the [Freq] hardkey.
• Press the [Down Arrow] hardkey
• Press the yellow [Start] hardkey to begin the zoomed measurement.
To return to baseband measurements,
• Press the [Freq] hardkey.
• Press the [ZERO START] softkey (F5).
Maximizing Dynamic Range
The default setting for input range is CH1 AUTO RANGE. For steady-state signals
this works well because the front end will find the input range that just exceeds the
signal level, and therefore maximize the dynamic range of the measurement. When a
front end is set up to autorange and the signal either overloads or becomes too faint
for the current range setting, the analyzer automatically searches for a new range and
then re-starts the measurement. To select CH1 AUTO RANGE:
• Select the [CH1 AUTO RANGE] softkey (F6)
If a signal is not steady, but has periods of high and low levels, then CH1 AUTO
RANGE may not set the front end at the most useful range. Instead, try CH1 AUTO
UP ONLY. This causes the input range to ratchet up when an overload is detected,
however when the signal is quiet, the front end range is not reduced. This prevents
an overload if another high level signal is expected. CH1 AUTO UP ONLY works
well with transient signals such as those expected during impact testing. To set up
this condition:
• Press the [Input] hardkey.
• Select the [CH1 AUTO UP ONLY] softkey (F5).
8
AGILENT 35670A
Supplemental Operator’s Guide
If continuous real-time measurements over a long duration are required, and
absolutely no overload conditions are desired, then a fixed input range can be set that
is greater than any signal expected. During periods of low signal level, dynamic
range will not be optimal, however the entire measurement can be completed without
overloads. To set up a fixed range of 1 volt:
• Press the [Input] hardkey.
• Select the [CHANNEL 1 RANGE] softkey (F3), enter 1, then press the
[Vpk] softkey (F1).
If engineering units are turned on ([XDCR UNIT ON OFF]) then the range can be
specified in EU. For example, if an accelerometer is connected to channel 1 with
XDCR UNIT turned on and XDCR UNIT LABEL specified as [g], then the input
range can be specified as 4 “g” as per the following:
• Press the [Input] hardkey.
• Select the [CHANNEL 1 RANGE] softkey (F3)
• Enter 4, then press the [EU]softkey (F8).
Averaging to Reduce Measurement Variance
Spectral measurements of steady signals give results which have a statistical
variation over time. Averaging several results will help reduce this variation. The
AGILENT 35670A has several averaging techniques available. For general
spectrum measurements, RMS or power averaging is commonly used. To average a
series of ten spectrum measurements using RMS averaging, follow these steps:
• Press the [Avg] hardkey.
• Press the [NUMBER AVERAGES] softkey (F2), enter 10, then press the
[ENTER] softkey (F1).
• Press the [AVERAGE TYPE] softkey (F3)
• Press the [RMS] softkey (F1)
• Press the [Avg] hardkey.
• Turn on averaging by pressing the [AVERAGE ON OFF] softkey (F1)
until ON is highlighted.
9
AGILENT 35670A
Supplemental Operator’s Guide
•
Systm Util Date: 10-02-96
Time: 08:40 PM
Date: 01-02-04 Time: 12:44:00 PM
A: CH1 Pwr Spec X:160 Hz Y:174.015 mVrms
1
Vrms
LogMag
6
decades
1
uVrms
0Hz800HzAVG: 5
10
Section 6 :Measuring a Single Channel Time Waveform
This section contains procedures for setting up and measuring a single channel time
waveform using an AGILENT 35670A Dynamic Signal Analyzer. Included are
steps to select single channel operation, set up the trace time length, and set up
triggering to capture a transient event. Instructions will also be presented to scale the
time trace.
Selecting Single Channel Operation
To measure a single-channel time waveform, turn on the AGILENT 35670A and
then:
• Press the [Inst Mode] hardkey
• Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
Note: if the AGILENT 35670A was already turned on and in an unknown
state, then preset the analyzer by doing the following:
• Press the green [Preset] hardkey
• Press [DO PRESET] softkey (F1).
AGILENT 35670A
Supplemental Operator’s Guide
Then set the instrument mode to single channel as above.
0.001953 s0.003906 s0.007813 s0.015625 s 0.03125 s
0.003906 s0.007813 s0.015625 s0.03125 s0.0625 s
0.007813 s0.015625 s0.03125 s0.0625 s0.125 s
0.015625 s0.03125 s0.0625 s0.125 s0.25 s
0.03125 s0.0625 s0.125 s0.25 s0.5 s
0.0625 s0.125 s0.25 s0.5 s1 s
0.125 s0.25 s0.5 s1 s2 s
0.25 s0.5 s1 s2 s4 s
0.5 s1 s2 s4 s8 s
1 s2 s4 s8 s16 s
2 s4 s8 s16 s32 s
4 s8 s16 s32 s64 s
8 s16 s32 s64 s128 s
16 s32 s64 s128 s256 s
32 s64 s128 s256 s512 s
64 s128 s256 s512 s1024 s
128 s256 s512 s1024 s2048 s
256 s512 s1024 s2048 s4096 s
512 s1024 s2048 s4096 s8192 s
11
AGILENT 35670A
Supplemental Operator’s Guide
Selecting Time Record Length
The default time record length at a frequency span of 51.2 kHz is only 7.8 ms, which
is usually not enough time to capture a mechanical event or observe a transducer’s
time domain signal. To increase the time record length to 0.25 s, perform the
following steps:
• Press the [Freq] hardkey
• Press the [RECORD LENGTH] softkey (F8)
• Enter 0.25 and then press the [S] softkey (F1)
The analyzer selects a new frequency span of 1.6 kHz, which at the default
resolution of 400 lines, takes 0.25 seconds to capture. The following table outlines
the time record lengths for all possible frequency spans and resolutions:
Displaying Time Waveforms
Time waveforms are displayed in a time trace. To set up the analyzer to display a
single time trace of channel 1, follow these steps:
• Press the [Disp Format] hardkey.
• Press the [SINGLE] softkey (F1)
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1)
• Press the [Meas Data] hardkey.
• Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
• Press the [TIME CHANNEL 1] softkey (F5)
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Using Manual Arm to Capture a Single Time Waveform
Often it is desirable to observe just one time waveform in detail. To acquire just one
record length at a time, follow these instructions:
• Press the [Trigger] hardkey.
• Select the [ARM SETUP] softkey (F7)
• Select the [MANUAL ARM] softkey (F2)
• Press the yellow [Start] hardkey to begin the measurement.
• Press the [Trigger] hardkey again.
Then to arm a single time record,
• Press the [ARM] softkey (F9)
One time record will be captured and displayed. To capture another time record,
12
AGILENT 35670A
Supplemental Operator’s Guide
• Press the [ARM] softkey (F9) again.
Event Triggering
The AGILENT 35670A has the ability to trigger the acquisition of a time waveform
based on the slope and level of the measured signal. To set up triggering properly,
the polarity of the measured pulse must be known in advance: will the event start
with a negative slope or positive slope? Will the event be mostly positive or mostly
negative? Use trial measurements to determine polarity if unknown.
Proper triggering is also affected by the input range. Fortunately the AGILENT
35670A has the ability to set the trigger point to be a percentage of the input range,
so even with CH1 AUTO UP ONLY, successful triggering should be possible
automatically.
To specify trigger parameters for a “positive” event on channel 1 where the trigger
point should have a positive slope, follow these steps:
• Press the [Trigger] hardkey.
• Press the [CHANNEL 1 234] softkey (F3) until 1 is highlighted.
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [CHANNEL LEVEL] softkey (F1)
• Enter 5 then press the [PERCENT (%)] softkey (F4)
• Press the [SLOPE POS NEG] softkey (F5) until POS is highlighted.
Use pre-triggering to acquire some data before the trigger point. To set up a 0.01s
pre-trigger condition, follow these steps:
• Press the [Trigger] hardkey.
• Press the [TRIGGER SETUP] softkey (F6)
• Press the [ALL CHANNELS] softkey (F8)
• Press the [CHANNEL * DELAY] softkey (F9)
• Enter -.01 then press the [S] softkey (F1)
To start the data acquisition,
• Press the yellow [Start] hardkey.
The analyzer should show in highlighted text above the trace:
WAITING FOR CH 1 TRIGGER
13
AGILENT 35670A
Supplemental Operator’s Guide
InputCH1 Rng: 1.0012 Vpk CH2 Rng: 631.706 mVpk
Date: 10-02-96 Time: 08:47:00 PM
A: CH1 Time X:38.0859 ms Y:1.06837 V
Real
/div
-2.5
CH3 Rng: 2.5149 Vpk CH4 Rng: 2.5149 Vpk
2.5
V
500
mV
V
-9.7656ms489.7461ms
Quantifying Time Trace Results
Use the marker knob to read the amplitude at any time during the trace. Rotate the
knob on the AGILENT 35670A front panel until the marker indicator (the small
diamond following the spectral trace) is placed at the desired time. The time and
amplitude at this bin is presented as X and Y values at the top of the trace. Units of
[s] and [Vrms] or [grms] identify the results.
Peak tracking is very useful for determining maximum amplitude of positive peaks
in the time domain:
• Press the [Marker] hardkey.
• Press [PEAK TRK ON OFF] softkey (F6) until ON is highlighted.
Care must be taken with negative peaks, however, because PEAK TRK ON only
looks for the maximum value, and large negative peaks will be ignored.
14
Section 7 :Measuring Dual Channel Spectra
This section contains procedures for setting up and measuring a dual channel
spectrum using an AGILENT 35670A Dynamic Signal Analyzer. Procedures are
similar to those in the section “Measuring Single Channel Spectrum,” however
display setup and results are different and will be described here.
To set up the analyzer for a dual channel spectrum measurement, proceed as follows:
Presetting the Analyzer
• Press the green [Preset] hardkey
• Press [DO PRESET] softkey (F1).
• Press the [Pause/Cont] hardkey if desired.
Selecting Dual Channel Operation
Measure two signals with the default two-channel instrument mode. If the analyzer
has been changed to 1 or 4 channel operation, then follow these steps:
• Press the [Inst Mode] hardkey
• Press [CHANNELS 1 2 4 ] softkey (F7) until 2 is highlighted.
AGILENT 35670A
Supplemental Operator’s Guide
Selecting Frequency Span
Reduce the span to 800 Hz as follows:
• Press the [Freq] hardkey
• Enter 800 and then press the [Hz] softkey (F2)
Displaying Dual-Channel Power Spectrum Measurements
To see the results of a dual-channel measurement, the analyzer’s display format must
be changed to display two traces. Proceed as follows:
• Press the [Disp Format] hardkey.
• Press the [UPPER/LOWER] softkey (F3)
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Meas Data] hardkey.
• Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
• Press the [PWR SPEC CHANNEL 1] softkey (F3)
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
• Press the [Active Trace] hardkey again.
• Press the [B] softkey (F2).
15
AGILENT 35670A
Supplemental Operator’s Guide
• Press the [Meas Data] hardkey.
• Press the [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Note: many of these keystrokes are not necessary if performed in the given order
because the analyzer defaults to many of these settings.
To acquire dual channel spectra,
• Press the yellow [Start] hardkey.
Meas Data
[FFT]
Date: 10-02-96 Time: 08:49:00 PM
A: CH1 Pwr Spec X:60 Hz Y:147.382 mVrms
Vrms
LogMag
decades
uVrms
B: CH2 Pwr Spec
Vrms
LogMag
decades
uVrms
A: CH1 Pwr Spec B: CH2 Pwr Spec
C: CH1 Time D: CH2 Time
1
6
1
0Hz800Hz
X:60 Hz Y:80.4887 mVrms
1
4
100
0Hz800Hz
Using Markers with Dual Channel Measurements
Each trace has independent marker control. All functions available for a single
marker are possible with both markers. To set up individual, independent markers
on each channel, follow these steps:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Marker] hardkey.
• Toggle the [MARKER ON OFF] softkey (F1) until ON is highlighted.
16
AGILENT 35670A
Supplemental Operator’s Guide
Use the marker knob to read the amplitude at any measured frequency for Channel 1
in Trace A.
To set up the marker for Channel 2 in Trace B,
• Press the [Active Trace] hardkey.
• Press the [B] softkey (F2).
• Press the [Marker] hardkey.
• Toggle the [MARKER ON OFF] softkey (F1) until ON is highlighted.
The marker knob is now active for Trace B, which is Channel 2. To gain control of
the marker on Trace A again, make it the active trace as follows:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
Coupled Markers
Markers on multiple traces can be coupled so that one marker will always track the
frequency specified in another trace. This makes comparisons of level differences
across channels simple. To set up coupled markers so that Trace B will follow the
marker frequency specified in Trace A, do the following:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Marker] hardkey.
• Toggle the [COUPLED ON OFF] softkey (F1) until ON is highlighted.
Move the knob to move the marker in Trace A. Notice that the marker in Trace B
moves to the same frequency.
Coupled Markers with Peak Tracking
Coupled markers can be further used to track a peak frequency in Trace A and
determine the response amplitude at the peak frequency in Trace B. To do this, leave
coupled markers on, as above, then set up peak tracking in Trace A as follows:
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1).
• Press the [Marker] hardkey.
• Toggle the [PEAK TRK ON OFF] softkey (F1) until ON is highlighted.
• Press the yellow [Start] hardkey.
Notice that the marker in Trace A moves to the peak value, whereas the marker in
Trace B moves to the same frequency as determined in Trace A.
Autoranging to maximize the dynamic range of measurements in the AGILENT
35670A can function on all channels simultaneously. Therefore all capability
previously mentioned for single-channel front-end ranging is available for dualchannel operation.
To set up autorange on all channels, follow these steps:
• Press the [Input] hardkey.
• Press [ALL CHANNELS] softkey (F2).
• Select the [CH* AUTO RANGE] softkey (F6)
For signals that are not steady, use CH* AUTO UP ONLY. CH1 AUTO UP ONLY
can be used for all channels to set up modal impact testing. To set up this condition:
• Press the [Input] hardkey.
• Select the [CH* AUTO UP ONLY] softkey (F5).
To set up the same fixed range on all channels:
• Press the [Input] hardkey.
• Select the [CHANNEL * RANGE] softkey (F3)
• Enter the range, either in units of g or Volts.
Dual Channel Averaging
Averaging to reduce measurement variance applies to all channels being measured.
Dual channel averaging is set up the same as single channel averaging:
• Press the [Avg] hardkey.
• Press the [NUMBER AVERAGES] softkey (F2), enter 10, then press the
[ENTER] softkey (F1).
• Press the [AVERAGE TYPE] softkey (F3)
• Press the [RMS] softkey (F1)
• Press the [Avg] hardkey.
• Turn on averaging by pressing the [AVERAGE ON OFF] softkey (F1)
until ON is highlighted.
18
AGILENT 35670A
Supplemental Operator’s Guide
Avg
[FFT]
Date: 10-02-96 Time: 08:51:00 PM
A: CH1 Pwr Spec X:60 Hz Y:112.185 mVrms
Vrms
LogMag
decades
uVrms
B: CH2 Pwr Spec
mVrms
LogMag
decades
uVrms
Type: RMS Number: 10
Update Rt: 5 Overlap: 0 %
1
5
10
0Hz800HzAVG: 10
X:60 Hz Y:95.0385 mVrms
100
4
10
0Hz800HzAVG: 10
19
AGILENT 35670A
Supplemental Operator’s Guide
Section 9 :Measuring Dual Channel Time Waveforms
This section contains procedures for setting up and measuring a single channel time
waveform using an AGILENT 35670A Dynamic Signal Analyzer. Included are
steps to select single channel operation, set up the trace time length, and set up
triggering to capture a transient event. Instructions will also be presented to scale the
time trace.
Selecting Dual Channel Operation
To measure a single-channel spectrum, turn on the AGILENT 35670A. The default
instrument mode is dual-channel operation.
Note: if the AGILENT 35670A was already turned on and in an unknown
state, then preset the analyzer by doing the following:
• Press the green [Preset] hardkey.
• Press [DO PRESET] softkey (F1).
This always returns the analyzer to dual-channel operation.
Selecting Time Record Length
Time record length can be entered directly or indirectly. It is entered directly under
the [Freq] hardkey as follows (this example sets up a 0.25 second time record):
• Press the [Freq] hardkey
• Press the [RECORD LENGTH] softkey (F8)
• Enter 0.25 and then press the [S] softkey (F1)
Time record length can also be specified indirectly by entering frequency and
resolution in lines. The relationship between time record length (T), lines of
resolution (N), and frequency (F) is:
N
=
T
()
F
1024.
For 400 lines of resolution (default) and 1600 Hz span, the time record length should
be 400/(1.024*1600) ≈ 0.25 seconds. The table showing the relationship between
lines, span, and time record length can be found in the section on Single Channel
Time Waveforms.
Displaying Dual Time Waveforms
Dual time waveforms can be shown in the UPPER/LOWER display format, with
AUTO SCALE turned ON, as follows:
• Press the [Disp Format] hardkey.
• Press the [UPPER/LOWER] softkey (F3).
• Press the [Active Trace] hardkey.
• Press the [A] softkey (F1)
• Press the [Meas Data] hardkey.
20
AGILENT 35670A
Supplemental Operator’s Guide
• Press the [CHANNEL 1 234] softkey (F1) until 1 is highlighted.
• Press the [TIME CHANNEL 1] softkey (F5)
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted..
• Press the [Active Trace] hardkey.
• Press the [B] softkey (F1)
• Press the [Meas Data] hardkey.
• Press the [CHANNEL 1 2 34] softkey (F1) until 2 is highlighted.
• Press the [TIME CHANNEL 2] softkey (F5)
• Press the [Scale] hardkey.
• Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
Arming Dual Channel Time Measurements
To manually arm a dual channel time measurement, proceed the same as for the
single channel case:
• Press the [Trigger] hardkey.
• Select the [ARM SETUP] softkey (F7)
• Select the [MANUAL ARM] softkey (F2)
• Press the yellow [Start] hardkey to begin the measurement.
• Press the [Trigger] hardkey.
Then to arm a single time record,
• Press the [ARM] softkey (F9)
One time record will be captured and displayed. To capture another time record,
• Press the [ARM] softkey (F9) again.
Triggering a Dual-Channel Time Measurement
Either channel may be monitored to initiate a trigger for a measurement. Setting up
the trigger event is the same as for single channel time waveforms, except that
channel 1 or channel 2 must be selected as the trigger. To specify the trigger
channel, proceed as follows:
• Press the [Trigger] hardkey.
• Press the [CHANNEL 1 234] softkey (F3) until 1 is highlighted.
This sets up the analyzer to trigger off an event on channel 1. To set up channel 2 as
the trigger channel, do the following:
• Press the [Trigger] hardkey.
• Press the [CHANNEL 1 2 34] softkey (F3) until 2 is highlighted.
21
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.