A-dec Inc. makes no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of merchantability
and fitness for a particular purpose. A-dec Inc. shall not be held liable
for any errors contained herein or any consequential or other damages
concerning the furnishing, performance or use of this material. The
information in this document is subject to change without notice. If you
find any problems in the documentation, please report them to us in
writing. A-dec Inc. does not warrant that this document is error-free.
No part of this document may be copied, reproduced, altered, or
transmitted in any form or by any means, electronic or mechanical,
including photocopying, recording, or by any information storage and
retrieval system, without prior written permission from A-dec Inc.
Trademarks and Additional Intellectual Property Rights
A-dec, the A-dec logo, A-dec 500, A-dec 300, Cascade, Cascade Master
Series, Century Plus, Continental, Decade, ICX, ICV, Performer,
Preference, Preference Collection, Preference ICC, and Radius are
trademarks of A-dec Inc. and are registered in the U.S. and other
countries. A-dec 200, Preference Slimline, and reliablecreativesolutions
are also trademarks of A-dec Inc. None of the trademarks or trade
names in this document may be reproduced, copied, or manipulated in
any manner without the express, written approval of the trademark
owner.
Certain touchpad symbols are proprietary to A-dec Inc. Any use of
these symbols, in whole or in part, without the express written consent
of A-dec Inc. is strictly prohibited.
This equipment/system is intended for diagnostic and therapeutic
treatment of dental patients by licensed health care professionals.
A dental operative unit (with or without accessories) is an AC-powered
device intended to supply power to and serve as a base for other dental
devices, such as a dental handpiece, a dental operating light, an air or
water syringe unit, an oral cavity evacuator, a suction operative unit,
and other dental devices and accessories.
Comments and Feedback
If you have any feedback or comments about this document, contact:
A-dec Inc.
Technical Communications Department
2601 Crestview Drive
Newberg OR 97132 USA
Regulatory information is provided with A-dec equipment as
mandated by agency requirements. This information is delivered in the
equipment’s Instructions for Use or the separate Regulatory Information and Specifications document. If you need this information, please go to
the Document Library at www.a-dec.com
.
Product Service
For service information, contact your local authorized A-dec dealer. To
find your local dealer, go to www.a-dec.com
The A-dec 200 Service Guide provides service information for the A-dec
200 dental system, including the chair, programming, delivery system,
cuspidor and support center, assistant’s instrumentation, utilities, and
dental light. Users of this guide should understand basic operation and
maintenance of dental and medical equipment, and use of flow diagrams.
CAUTIONPossible injury or equipment damage. Service to be
performed by trained personnel only.
Get Support
For questions not addressed in this document, contact A-dec Customer
Service using contact information for your region.
The following regulatory symbols may appear throughout the document.
NOTENotes indicate additional information, and when it is
important that instructions are followed.
TIP Tips indicate tips or tricks to make installation, use, or
maintenance easier.
WARNINGWarning indicates potential severe injury or death if
instructions are not followed properly.
CAUTIONCaution indicates when failure to follow instructions
could result in damage to product or minor injury.
DANGER Danger indicates warnings of dangerous voltage and of
certain electrical shock.
BIOHAZARD Biohazard indicates potential infection if
instructions are not properly followed.
This guide contains service, maintenance and adjustments; flow diagrams;
exploded parts breakdown of assemblies; and troubleshooting.
IMPORTANTImportant indicates areas in which to refer to or use
specific instructions.
86.0016.00 Rev A 1
A-dec 200 Service Guide
Cuspidor
(page 55)
Dental Chair
(page 7)
Assistant’s
Instrumentation
(page 63)
Foot Control
Footswitch
Solids Collector
(page 66)
Utilities
(page 73)
Support Center
(page 55)
Delivery System
(page 35)
Dental Light
(page 81)
Touchpad
(page 31)
Serial Number
Location
A-dec 200 System
A-dec 200 system comes configured as shown below:
Figure 1. A-dec 200 Systems
2 86.0016.00 Rev A
Serial and Model Numbers
1
2
Introduction
Product serial and model number information can be found on the serial/
model number labels. When you contact customer service, the serial
number helps identify the product and when it was manufactured.
Use Table 1 and Figure 2 to reference how to identify serial/model number
information.
Table 1. M o nt h I d e n t i fi c a t i o n Ta b le
LetterMonthLetterMonth
AJanuaryGJuly
BFebruaryHAugust
CMarchISeptember
DAprilJOctober
EMayKNovember
FJuneLDecember
Figure 2. Serial Number Label Example
200
(1) The REF number is the model number. (2) The first letter of the serial number
indicates the month the product was manufactured. The first digit of the serial
number is the year of manufacture (for example, L8 = December 2008).
86.0016.00 Rev A 3
A-dec 200 Service Guide
Service Tools
This table lists the types of tools available from A-dec for servicing A-dec 200 equipment and their recommended use:
Table 2. Recommended Tools
ToolTas kPart IllustrationPart Number
Drive air pressure gaugeAdjusting handpiece drive air pressure, 0-60 psi (4.13 bar)
This gauge does not fit the Borden 3-hole coupler
50.0271.00
Hemostat• Troubleshooting or repairing a unit
• Stopping air or water flow through tubing
Hex key setServicing or installing A-dec equipment (plastic case included)009.018.00
®
Loctite
O-ring toolsReplacing O-rings during quick field repairs (fits the four
Panel mount gauge• Checking air/water pressure
Installing threaded fasteners to prevent loosening060.001.00 (Red 271)
smallest O-ring sizes)
• Checking inline pressure gauge for testing purposes
009.008.00
060.002.00 (Blue 242)
009.013.00
026.118.00
4 86.0016.00 Rev A
Table 2. Recommended Tools (continued)
CAUTIONUse only A-dec Silicone lubricant or
the O-rings may be damaged.
ToolTas kPart Illustra tionPart Nu mber
A-dec Silicone lubricantLubricating internal moving parts such as O-rings, oral
evacuator valves, and bushings
Sleeve toolAiding in securing 1/4" tubing sleeves and 1/8" uni-clamps98.0072.00
98.0090.01
Introduction
Snap ring toolInstalling and removing internal and external snap rings (fits
all snap rings used in A-dec equipment)
Tubing stripperSeparating the extruded air and water lines in vinyl tubing009.035.00
Umbilical stringerRouting additional tubing or wiring through existing umbilical
assemblies (12’ [3.66 meter] stringer with threading holes on
both ends)
Valve test syringeTesting of pilot operated valves; used to apply a static
pressure of 5-75 psi (.34-5.17 bar)
009.007.00
009.015.00
98.0050.01
86.0016.00 Rev A 5
A-dec 200 Service Guide
6 86.0016.00 Rev A
DENTAL CHAIR
2
This section provides detailed information related to service, maintenance,
and adjustment of the A-dec 200 dental chair.
Contents
• Product Overview, page 8
• Service, Maintenance, and Adjustments, page 10
• Illustrated Parts Breakdown, page 17
Figure 3. A-dec 200 Chair
86.0016.00 Rev A 7
A-dec 200 Service Guide
Master Toggle
(Power On/Off)
Status Light
Chair Lift Arm
Chair Stop
Plate
Status
Light
Assistant’s
Arm
Product Overview
Chair Load Capacity
Maximum Chair Capacity:
• Patient Load 136 kg (300 lbs.)
• Module/Accessory Loads (maximum off-center) with A-dec 200 system:
68 kg (150 lbs.) @ 406 mm (16")
Power and Status
The chair and system is controlled by the master toggle on the delivery
system. The power should always be turned off for service. When the Adec logo on the touchpad or the status light on the chair lift arm are
illuminated, the system is on and ready for use. If the status light blinks,
the limit switch has been activated.
Figure 4. Power and Status
Limit Switch and Chair Lockout
If anything becomes lodged under the chair lift arm or assistant’s arm, a
limit switch stops the downward motion of the chair. Pressing the chair
stop plate or lifting up on the assistant’s arm activates the limit switches.
Use the footswitch or touchpad to raise the chair, then remove the object.
The optional lockout kit inhibits the operation of the dental chair when a
handpiece is removed from its holder and the foot control pressed. When
this happens, the chair status light blinks quickly. To resume, replace the
handpiece and use the footswitch or touchpad to raise the chair.
Figure 5. Chair Lockout Overview
8 86.0016.00 Rev A
Circuit Board Components
Table 3. LED IdentificationFigure 6. Chair Circuit Board Components
LEDStatusDescription
DS16 - AC power
LED
DS15 - Status LED OffSystem is not functioning, no power or
DS17 - Data LEDOffNo DCS communication, not connected
DS6 Stop plate
limit switch
DS4 - Chair
lockout
DS3 and DS5 -
Chair
potentiometers
DS7, DS13,
DS12, DS14
Chair relay LEDs
DS8Capacitor switch
DS9Back up limit switch
DS10Base up limit switch
DS11 off = relay is off; on = relay is on
DS1, DS2Position 3 function selection
OffNo 24 VAC power, tripped circuit
breaker, power supply turned off, no
line voltage
Green, steady24VAC at the terminal strip
circuit board has failed
Green, steadyNormal operation
to the DCS, or DCS has failed
Green, steadyDetects active DCS
Green, blinkingValid DCS Message
OffClosed, (normal)
RedOpen, (activated)
OffOpen, (normal)
RedClosed, (activated)
OffPotentiometer:
Not connected or bad connection
Moving in wrong direction
Limited range of motion, or
Cable is not on wheel
Yello w, steadyNormal operation
Yello w, fast blinkUpper end of travel
OffRelay is off
-
OnRelay is on
Dental Chair
86.0016.00 Rev A 9
A-dec 200 Service Guide
Service, Maintenance, and Adjustments
Contents
• Potentiometers, page 11
• Hydraulic System, page 13
• Solenoids, page 14
• Test the Motor Pump, page 15
• Headrest Adjustments, page 16
• Chair Speed Adjustments, page 16
Factory Default Routine
When a new circuit board is installed in the chair, factory default routine
needs to be run to learn the range of motion of the chair. The routine:
• Sets the base and back upper limits
• Calculates new presets based on actual range of motion of the chair
• Verifies that the potentiometers work
To start the factory default routine, place the “spare” jumper in the factory
default position on the P17 test points of the chair circuit board (see "Circuit
Board Components" on page 9 for reference).
When running the factory default routine the chair:
1. Moves base down
2. Moves base up
3. Moves back down
4. Moves back up
5. Moves base and back to Position 0
6. Beeps three times
NOTEThe jumper must remain in the factory default
position to complete the factory default routine. The status
LEDs on the standard and deluxe touchpads and the chair
circuit board double blink while the factory default routine
is running and after the routine is complete.
10 86.0016.00 Rev A
Potentiometers
Actuator Pin
Switch Arm
Limit Switch
Mounting
Bracket
Clamping
Screws
25"
63.5 cm
Potentiometers provide the controller with current position values for the
chair base and back. The controller saves the chair values when
programmed and compares the values with current position values for the
pre-position and auto-return functions.
Adjust the Base Up Limit Switch:
1. Remove the motor pump cover.
2. Loosen the two screws clamping the limit switch to the mounting
bracket.
Figure 7. Mounting Bracket
Dental Chair
Figure 8. Chair Position
4. Push the limit switch against the actuator on the drive gear until the
switch opens (clicks), then tighten the clamping screws. (See Figure 7).
5. Position the chair base down until the limit switch has closed, then
position the chair full base up. Check the distance between the top of
the base plate to the flat area around the threaded stud the chair adapter
mounts to. If the distance is incorrect, repeat steps 2 through 4.
3. Position the chair as shown in Figure 8.
86.0016.00 Rev A 11
A-dec 200 Service Guide
Potentiometer
Gear
Positioning
Potentiometer
Limit Switch
Mounting
Screw
Large Drive
Gear
Limit Switch
Mounting Screws
Bracket Mounting
Screws
Switch Arm
Drive Shaft
Urethane Tubing
3/8" OD
Potentiometer
Shaft
Tip
Potentiometer
Shaft
Clockwise
Counterclockwise
Adjust the Base Positioning Potentiometer
1. Remove the motor pump cover and position the chair base down.
2. Use a 3/16" hex key to remove the limit switch and mounting screw. See
Figure 9.
3. Turn the potentiometer gear counterclockwise until it stops.
4. Align the potentiometer assembly, then turn the potentiometer gear
clockwise two teeth.
5. Reinstall the limit switch and potentiometer assembly. Make sure the
potentiometer gear does not turn and the two gears mesh properly.
6. Ensure that the electrical connections to the limit switch and positioning
potentiometer are property set.
7. While observing the two gears for binding, lower the chair base.
CAUTIONDo not raise to the full base up position until after you
have checked the base up limit switch for proper adjustment. the
chair may go into hydraulic lock if not adjusted properly.
8. Reinstall the cover, and program the auto-positioning functions.
Figure 9. Base Potentiometer
Adjust the Back Potentiometer
1. Position the chair back to its full up position.
2. Disconnect the limit switch wiring harness from the limit switch.
3. Remove the limit switch mounting screws and limit switch from the
bracket. Do not bend the switch arm.
4. Remove the bracket mounting screws.
5. Remove the drive shaft from the potentiometer shaft.
6. Remove the drive shaft from the chair by moving it toward the chair
backrest, and slightly to the side to dislodge it from the holder.
Figure 10. Remove Drive Shaft
12 86.0016.00 Rev A
7. Turn the potentiometer shaft clockwise until it no longer turns, then
turn the shaft counterclockwise 1/8" of a turn
Figure 11. Adjust Back Potentiometer
8. Reinstall the shaft.
Hydraulic System
Hydraulic Cylinders
Hydraulic Fluid Reservoir
Solenoid/Manifold
Assembly
Motor Pump
The hydraulic system consists of:
• Hydraulic fluid reservoir
The fluid level in the reservoir can be seen through the sides of the
reservoir and is serviced via a top fill cap.
•Hydraulic cylinders
The hydraulic cylinders control the base lift and back functions. Springs
and gravity retract the rod during base and back down functions.
• Motor-driven hydraulic pump
The hydraulic pump and the starter capacitor supply hydraulic fluid
from the reservoir, under pressure, to the chair lift and tilt hydraulic
cylinders for back up and base up functions.
• Solenoid/manifold assembly
This assembly gates hydraulic fluid to and from the two cylinders.
Depending on the chair function called for, the controller selects which
solenoid-actuated manifold valves are opened or closed. The solenoid/
manifold assembly also includes four adjustable needle valves used to
restrict or divert the flow of hydraulic fluid to and from the lift and tilt
cylinders. These valves provide the rate of travel adjustment for chair
base and back movement.
Dental Chair
Figure 12. Hydraulic System
86.0016.00 Rev A 13
NOTE If cable ties are present in the product and you need to
remove them for servicing, make sure to replace the ties after
service is completed.
A-dec 200 Service Guide
Hydraulic
Manifold
O-ring
Poppet
Poppet
Sleeve
Washer
Solenoid
ID Washer (Solenoid
Specifications)
P10
Solenoids
Tes t th e Sol eno i d
To test the magnetic pull of the solenoid hold the tip of screwdriver near a
solenoid and activate the appropriate chair function. You should feel the
tug of the magnetic field generated around the solenoid.
Figure 13. Test the Solenoid
Remove and Replace the Solenoid
1. Lower the chair base and back to the full down position. Remove the
motor pump cover, then unplug the chair.
WARNINGThe solenoid coils are powered by line voltage (100,
120, or 240 V). Failure to unplug the chair may result in serious
injury from electrical shock.
4. Using a flat-tipped screwdriver, loosen and then remove the sleeve and
poppet from the manifold assembly.
WARNINGTo prevent the possibility of over-heating and failure,
replace the entire solenoid assembly.
5. Remove the o-ring from inside the manifold, and install a new o-ring.
6. Install a new sleeve and poppet; tighten the poppet sleeve using a flat-
tipped screwdriver.
7. Install a new coil on the plunger. Do not overtighten the retaining nut.
8. Strip approximately 1/4" of insulation from the wires cut in step 2, and
install a crimp-on butt-type connector on each wire.
9. On the new solenoid, cut the wiring to length allowing enough to reach
the crimped-on connectors. Strip approximately 1/4" of insulation from
the wires and crimp each wire into a connector.
Figure 14. Remove and Replace Solenoid
2. Using a pair of wire cutters, cut the wiring to the faulty solenoid at
3. Using a 9/16" wrench, remove the solenoid retaining nut and slide the
14 86.0016.00 Rev A
about mid point between the solenoid and connector P10.
coil off the poppet sleeve.
CAUTIONUse caution when removing and replacing the coil. The
poppet sleeve is easily bent. Even slight bending of the sleeve will
result in the malfunction of the solenoid valve.
Tes t t he Mot or Pu mp
Red Wire
White Wire
Motor Pump
This test requires the use of a current pickup probe.
• Clip the probe onto the red wire going to the motor pump.
• Use the footswitch or touchpad to raise the chair.
You should read 5 Amps (maximum) of current for 120 V motor pump, or
2.5 Amps (maximum) of current for 240 V motor pump.
Figure 15. Test Motor Pump
Dental Chair
86.0016.00 Rev A 15
A-dec 200 Service Guide
Glide Bar
Locking
Knob
Adjust chair
base speed.
Adjust chair
back speed.
Headrest Adjustments
Turn the locking knob clockwise to lock it into the desired position. Slide
the headrest and glide bar up or down to adjust the height.
WARNINGWhen the glidebar has reached its maximum
recommended working height, a warning will be visible on the
patient’s side of the glide bar. Do not use the headrest in a position
where this warning is visible.
Figure 16. Headrest Adjustments
Chair Speed Adjustments
The speed for moving the chair seat and back can be adjusted. Use a 3/32"
hex key to adjust the chair base speed and back speed on the manifold (see
Figure 17. Adjust Manifold for Chair Speed
NOTE If cable ties are present in the product and you need to
remove them for servicing, make sure to replace the ties after
service is completed.
16 86.0016.00 Rev A
Illustrated Parts Breakdown
3
1
5
4
2
Dental Chair
Part Identification
In this section, you will find serviceable components tables that correspond
to the illustrations. The tables identify all parts and kits, including those
that are not for sale. Parts that are not for sale are indicated with the symbol
shown below:
Chair Covers
Figure 18. A-dec 200 Dental Chair
†— Indicates that the individual part is not available for sale. These parts
are typically part of a kit and/or larger assembly that is for sale.
Contents
• Chair Covers, page 17
• Upper Chair Assembly, page 18
• Mid Chair Assembly, page 20
• Lower Chair Assembly, page 22
• Headrest Assembly, page 23
• Motor Pump and Capacitor Assembly, page 24
• Hydraulic Manifold Assembly, page 26
• Hydraulic Tank Assembly, page 27
• Base Limit Switch Assembly, page 27
• Back Potentiometer Assembly, page 28
• Base Position Potentiometer Assembly, page 29
ItemPart NumberDescription
161.3821.00Rear chair cover
261.3847.00Bottom cover
361.3846.00Utility cover
86.0016.00 Rev A 17
461.2239.01Lift arm cover
561.2242.00Umbilical routing bracket
A-dec 200 Service Guide
Upper Chair Assembly
ItemPart NumberDescription
1—Double articulating headrest assembly (see page 23)
261.3824.00Backrest cover
361.3791.00Armrest release button
4013.052.00Compression spring, .843 OD X 1.375 FL
561.1245.02 †Arm support, LH
6004.157.00Wave spring washer, .990 OD stainless steel
7004.126.00Flat washer
861.3827.00Armrest cover
961.3790.01Armrest pivot plate, LH
10005.123.01Button head socket screw, 10-32 X 5/8"
1161.3108.00Flanged bearing
12001.164.00Socket shoulder screw, 1/2-13 X 5/8 X 5/8
1361.3788.01Pivot bolt cover
1461.2740.00Backrest link pin
15016.131.00Flanged bearing
16007.069.00Set screw, socket 1/4-20 X 1/4
1761.3792.00Back link
1861.3789.01Armrest pivot plate, RH
19005.147.00Flat head socket screw, 1/4-20 X 1-1/4
20006.122.01Retainer nut, 10-32 zinc
2161.2743.00Brake shoe
2261.1247.02 †Arm support, RH
23001.268.00Socket screw, 1/4-20 X 3/30 mm stainless steel
24002.136.00Flat head socket screw, 1/4-20 X 1/38
† Indicates that the individual part is not available for sale
261.0460.00Flow adjust screw with o-ring5030.015.02O-ring, package of 10
3030.004.02O-ring, AS568-004 package of 10661.1333.00
61.1334.00
Figure 24. Hydraulic Manifold
Solenoid, 8-watt, 100 V, yellow wires
Solenoid, 8-watt, 120V, black wires
Solenoid, 8-watt, 240 V, red wires
Manifold assembly, hydraulic, 120 V
Manifold, assembly, hydraulic, 240V
26 86.0016.00 Rev A
Dental Chair
2
43
1
2
3
Hydraulic Tank Assembly
ItemPart NumberDescription
1006.112.00Retainer Nut, 3/4"
261.1489.00Pick-Up Tube Solder Assembly
3004.161.00Rubber Washer
461.1339.00Hydraulic Tank
Figure 25. Hydraulic Tank Assembly
Base Limit Switch Assembly
ItemPart NumberDescription
1001.021.00Socket Head Screw, 4040 X 1/2 Stainless Steel
2044.184.00Switch, 15 Amp 125 Volt
361.1315.00Mounting Bracket
Figure 26. Base Limit Switch Assembly
86.0016.00 Rev A 27
A-dec 200 Service Guide
1
2
3
4
5
Back Potentiometer Assembly
ItemPart NumberDescription
161.1225.00Helical Drive Shaft
2025.113.00Compression ring
3002.134.00Socket Head Screw, 1/4-20 X 1/4
4041.372.00Potentiometer, 5 K Ohm
561.1226.00Mounting Bracket
Figure 27. Back Potentiometer
28 86.0016.00 Rev A
Base Position Potentiometer Assembly
1
2
3
4
5
6
7
8
ItemPart NumberDescription
1002.024.00Socket Head Screw, 4-40 X 3/4
260.0080.00Mounting Plate
3041.372.00Potentiometer, 5 K Ohm
461.1223.00Base Position Mounting Bracket
5044.184.00Switch, 15 Amp, 125 Volt
661.1367.00Limit Switch Mounting Plate
7002.120.00Socket Head Patch Screw, 1/4-20 X 1 Stainless Steel
861.2625.00 †Gear, 10 Tooth, 24 Point
† Indicates that the individual part is not available for sale
Figure 28. Base Potentiometer
Dental Chair
86.0016.00 Rev A 29
A-dec 200 Service Guide
30 86.0016.00 Rev A
PROGRAMMING
Touchpad
Footswitch
3
The A-dec 200 product offers the option of having a standard touchpad.
The touchpad centralizes treatment room controls into one touch surface
including the chair, light, cuspidor controls, and auxiliary equipment.
NOTE
• The A-dec 200 product ships with a non-functioning cuspidor
board. If the system has a touchpad and cuspidor, an upgrade
to the cuspidor board is required to allow control of the
cuspidor functions with the touchpad.
• If the system has a touchpad and dental light, the dental light
circuit board is required (located in the floor box) to allow
control of the dental light from the touchpad.
Figure 29. Touchpad or Footswitch Programming
86.0016.00 Rev A 31
A-dec 200 Service Guide
Entry/Exit
Treatment 1
Treatment 2
Program Button
X-ray/Rinse
Program Button
X-ray/Rinse
Treatment 2
Entry/Exit
Treatment 1
Touchpad
To u c h p a d O p t i o n s
The A-dec logo on the touchpad indicates the system status. If it appears as
solid blue, the system is on and ready for use.
Chair Positions
The chair direction arrows on the standard and deluxe touchpads allow
you to manually move the chair base up/down and back up/down. Table
lists the icons and the action of each icon.
Figure 30. Chair Position Touchpad Button
Footswitch
Table 4. Chair Preset Positions
Footswitch/TouchpadDescription
Entry/Exit: Positions chair for patient entry/exit. On
systems with a doctor’s touchpad, it also turns the
dental light off.
Treatment 1: Positions the chair base and back down.
On systems with a doctor’s touchpad, it also turns the
dental light on.
Treatment 2: Positions the chair base down and back
up. On systems with a doctor’s touchpad, it also turns
the dental light on.
X-ray/Rinse: Moves the chair for either x-ray or rinse
position. Press again to move the chair to the previous
position. On systems with a doctor’s touchpad, also
turns the dental light off or back on.
Chair Preset Buttons
Use the program buttons to assign and save chair preset positions. To
program the chair presets Entry/Exit, Treatment 1, and Treatment 2:
1. Use the manual controls to adjust the chair position as desired.
2. Press and release the Program button. One beep indicates programming
mode is on.
3. Within four seconds, press the chair position button you wish to reset
(for example and ). Three beeps indicate the new setting is
programmed into memory.
Chair Preset Position Buttons
Chair position buttons are factory preset to automatically move the chair
(see Figure 30).
32 86.0016.00 Rev A
X-Ray/Rinse Button
The x-ray/rinse preset button moves the chair and patient into an upright
position for x-rays or cuspidor access. To change the function to an
additional programmable chair position:
1. On the touchpad or footswitch, press and hold the power button and
Dental Light
Button
LED
Indicator
the x-ray/rinse button at the same time for three seconds.
○ One beep indicates the button can be configured as Treatment 3.
○ Three beeps indicate that the X-ray/Rinse button has been
configured as the x-ray/rinse function (toggles between x-ray/rinse
and the previous position).
Dental Light
The dental light can be operated from the manual 3-position switch or the
optional touchpad. To operate the touchpad, press and hold the button to
turn the light off. Press the light button to choose between two intensities.
When the light is in the composite setting, the LED indicator on the
touchpad flashes.
2. Program the preset position as instructed under "Chair Positions" on
page 32.
TIP If the X-ray/Rinse button is changed to a preset position, it
operates the same as treatment buttons 1 and 2.
Figure 31. Dental Light Operation
Auto On/Off
The optional auto on/off feature turns on the light when the chair back
reaches a treatment position. Press or and the dental light turns off.
To disable the auto on/off feature, press and hold and
simultaneously for three seconds. One beep confirms the dental light on/
off function is off.
To enable the auto on/off feature, press and hold and
simultaneously for three seconds. Three beeps confirm the auto on/off
feature is activated.
86.0016.00 Rev A 33
A-dec 200 Service Guide
Bowl Rinse
Button
Cupfill
Button
Cuspidor Cupfill and Bowl Rinse
The cuspidor cupfill and bowl rinse functions are dependent upon your
specific configuration.
Customize Cupfill and Bowl Rinse Functions
Perform this operation with the doctor’s touchpad only.
Standard Cuspidor (no touchpad)
Press and hold the cupfill button on the cuspidor for the desired amount of
water. Water will continue to flow until the button is released.
Press the bowl rinse button on the cuspidor once for a 10-second rinse. For
continuous rinse, hold the button down. When the button is released, the
water will continue to flow for 10 additional seconds.
Cuspidor with Optional Touchpad
If your system includes a touchpad, you can use the buttons on the
touchpad or the cuspidor to operate and program bowl rinse and cupfill
functions:
Table 5. Cupfill and Bowl Rinse Functions
ButtonDescription
Cupfill Button:
• Press the Cupfill button for a timed operation. The factory preset
is a 2.5second fill.
• Press and hold the Cupfill button for manual operation.
Bowl Rinse Button:
• Press the Bowl Rinse button for a timed operation. The factory
preset is a 30 second rinse.
• Press and hold the Bowl Rinse button for manual operation.
1. Press on the touchpad or press and hold both the cupfill and bowl
rinse buttons on the cuspidor (see Figure 32). Release them when you
hear one beep.
2. Press and hold the Cupfill () or Bowl Rinse () button for the
desired amount of time.
3. Release the button. Three beeps confirm the setting.
TIP Press twice in less than two seconds to activate the
continuous operation mode. Press the button once to end the
continuous bowl rinse mode.
Figure 32. Cuspidor Tower Cupfill and Bowl Rinse Buttons
34 86.0016.00 Rev A
DELIVERY SYSTEM
4
This section provides information related to service, maintenance, and
adjustments of the A-dec 200 delivery system.
Contents
• Product Overview, page 35
• Service, Maintenance, and Adjustments, page 37
• Illustrated Parts Breakdown, page 47
Product Overview
A-dec 200 delivery system has been designed to mount on the A-dec 200
Support Center. The support center mounts to the chair using a post
mount. The A-dec 200 delivery system provides the air and water used to
operate the handpieces, syringes and accessories, and electrical power and
data control of other modules.
The A-dec 200 standard configuration has a balanced flexarm with manual
brake, three handpiece control block positions (3-position block is
standard), a control head with room to house integrated accessories, and an
autoclavable syringe.
Figure 33. A-dec 200 Delivery System
86.0016.00 Rev A 35
A-dec 200 Service Guide
Air Coolant Tubing
(Yellow)
Water Coolant Tubing
(Blue)
Drive Air Tubing
(Clear)
(Exhaust - to Oil Collector)
(Tubing From Holder)
Air
Coolant
Water
Coolant
Drive Air
(ribbed)
Water
Coolant
Air
Coolant
Exhaust Air
Exhaust
Air
Drive Air
A-dec Tubing
A-dec products use four sizes of outside diameter tubing: 1/8", 1/4", 3/8",
and 5/16". The A-dec 200 delivery system uses standard A-dec tubing and
vinyl handpiece tubing. See "Tubing and Flow Diagrams" on page 109 to
identify tubing. See "Handpiece Tubing Replacement" on page 42 for
instructions on replacing tubing.
The handpiece tubing connects to the control block using tubing connectors
and the appropriate A-dec tubing.
Figure 35. Vinyl Handpiece Tubing Control Block Connections
36 86.0016.00 Rev A
Service, Maintenance, and Adjustments
Tension Set Screw
Tu r n C lo ck wi se
Adjust Flexarm Height
Delivery System
Contents
• Flexarm Adjustments, page 37
• Instrument Holder Adjustments, page 38
• Control Block, page 39
• Handpiece Control Adjustments, page 40
• Oil Collector, page 41
• Handpiece Tubing Replacement, page 42
• Quad Voltage Intraoral Light Source (QVIOLS), page 43
• Intraoral Light Source Length and Voltage, page 45
Flexarm Adjustments
• Tension: If the control head flexarm drifts right or left, use a 3/32" hex
key to adjust the tension setscrew. Turn the screw clockwise to tighten
or counterclockwise to loosened the tension.
• Height: Turn the knob counterclockwise to disengage the flexarm brake
and adjust the height. Turn the knob clockwise to lock the position.
Figure 36. Flexarm Adjustments
86.0016.00 Rev A 37
A-dec 200 Service Guide
Screw
Instrument Holder Adjustments
Rotate the holders independently. Pull the holder slightly away from the
adjacent one, rotate to the desired position, then release.
CAUTIONTwisting the holder without pulling it away from the
adjacent one will damage the mechanism.
Figure 37. Doctor’s Holder Adjustments
Handle
To position the handle, remove the screw at the end of the handle, adjust
the handle for use, then replace and tighten the screw.
Figure 38. Handle Adjustment
38 86.0016.00 Rev A
Control Block
The control block might need to be removed to service the control head. For
example, you many need to remove the control block to change a
diaphragm, to change a cartridge, or to service o-rings.
Remove the Control Block
1. Remove the cover.
2. Loosen and remove the two screws a that secure the control block to the
control delivery system frame.
3. Lift the control block up from the base of the control center.
Figure 39. Remove Control Top
Delivery System
Figure 40. Remove Control Block on Delivery Systems
86.0016.00 Rev A 39
A-dec 200 Service Guide
Air, Water (Blue)
Coolant Keys
Drive Air Key
Handpiece
Holder
Foot Control
Wet/Dr y Toggle
Water
Coolant Key
Handpiece
Holder
Foot Control
Air Coolant
Key
Handpiece Control Adjustments
Adjust the Water Coolant
1. Turn the air coolant, water coolant, and drive air all the way down.
2. Lift a handpiece from the holder, and flip the wet/dry toggle to water
(towards the blue dot).
3. Step on the foot control.
4. Adjust the water coolant flow until there is 1 drop every 2 seconds.
Figure 41. Adjust Water Coolant
Adjust the Air Coolant
1. Lift a handpiece from the holder, and step on the foot control.
2. Adjust the air coolant flow until the spray is a fine mist.
Figure 42. Adjust Air Coolant
40 86.0016.00 Rev A
NOTEContact A-dec Customer Support for information on
servicing the foot control or syringe. See "Get Support" on page 1.
Delivery System
Drive Air Key
Holder
Pressure Gauge
Toggle and
Foot Control
Adjust the Drive Air Pressure
NOTEUse a handpiece pressure gauge attached to the handpiece
tubing for exact drive air measurement. See the manufacturer’s
handpiece documentation for the drive air pressure specification.
1. Lift a handpiece from the holder.
2. Install a pressure gauge.
3. Flip the toggle to dry, and step on the foot control.
4. Adjust the drive air pressure according to manufacturer’s
recommendations.
○ To increase flow, turn the key counterclockwise.
○ To decrease flow, turn the key clockwise.
Figure 43. Adjust Drive Air Pressure
Oil Collector
The oil collector needs to be serviced once a week for normal use and more
often for heavier use. To service:
1. Unsnap the oil collector cover located under the control head and
discard the old gauze.
CAUTIONDo not remove the foam pad located inside the oil
collector cover.
Figure 44. Oil Collector Removal
2. Fold a new gauze pad (51 mm x 51 mm [2" x 2"]) into quarters and place
inside cover.
3. Snap the oil collector cover closed.
86.0016.00 Rev A 41
A-dec 200 Service Guide
Handpiece #3
Handpiece #2
Handpiece #1
Air Coolant
Water C ool ant
Drive Air
Handpiece Tubing Replacement
1. Remove the delivery system cover.
2. Cut the handpiece tubing you are replacing from the colored A-dec
tubing/control block.
3. Pull the old handpiece tubing out of the control head.
4. Route the new handpiece tubing through the base of the control head.
5. Connect the new handpiece tubing to the control block using the
connectors and colored tubing previously used.
The A-dec colored tubing is identified by it’s color (see bullets below and
Figure 35).
○ Yellow tubing to the air coolant port
○ Blue tubing to the water coolant port
○ Clear tubing to the drive air port
NOTEVinyl tubing is not color coded. For vinyl tubing
identification, see Figure 35 on page 36.
6. Replace the delivery system cover.
Figure 45. Handpiece Barb Connection
Adjust Tubing Length
1. Adjust the length of the tubing so it drapes with syringe tubing.
2. Insert the tubing in the tubing retainers.
Figure 46. Adjust Length of Handpiece Tubing
42 86.0016.00 Rev A
1
2
3
4
5
6
7
12
11
10
9
8
13
14
15
16
18
17
Quad Voltage Intraoral Light Source (QVIOLS)
Part Number: 90.1168.00
The quad voltage intraoral light source (QVIOLS) provides four
independent fiber optic voltage outputs. Each output is adjustable from
3VDC to 7VDC at 1.5 Amps. Only one output can be on at a time.
Activating an input on the QVIOLS turns on its respective output.
Table 7. QVIOLS Circuit Board Descriptions
ItemDescription
1DS1 AC Power LED
2DS2 Status LED
3DS3 Data LED
4J1 - 24VAC Input
5J1 - 0VAC Input
6P2 - DS4 - normally closed jumper
7P1 Data Port
8J3 Switch input Common
9J3 Switch Input #1
10J2 Switch Input #2
11J2 Switch Input #3
12J2 Switch Input #4
13J4 Light Source Output #1
14J4 Light Source Output #2
15J5 Light Source Output #3
16J5 Light Source Output #4
17S1 Decrease Lamp Output
18S2 Increase Lamp Output
Delivery System
Figure 47. QVIOLS Circuit Board
NOTEOn the A-dec 200 product, a jumper should always
be located within P2, and DS4 should be On.
86.0016.00 Rev A 43
A-dec 200 Service Guide
Intraoral Light
Source Outputs
Decrease
and
Increase
Buttons
Intraoral Light Source Adjustments
The intraoral light source (IOLS) voltage adjustment on the A-dec 300
doctor’s delivery system is located on the QVIOLS circuit board. Each
output voltage is preset to 3.2VDC at the lamp terminals when the lamp is
on.
WARNING The Length and Voltage Table, page 45, is only
valid for devices rated for 3.5 VDC and 0.75 Amp 26AWG
wires. For devices drawing a different amount of current,
requiring a different voltage, or with a different wire gauge,
please contact A-dec Customer Service. (See "Get Support" on
page 2.)
1. Use a 7/64" hex key to remove the control head cover.
2. Set the voltmeter to DC voltage and place its probes on the IOLS output
terminals for the handpiece you are testing.
3. Lift the handpiece from its holder.
NOTEWhen the intraoral light source output is On, its
respective LED is On. for example, LED DS6 is On when
handpiece #2’s intraoral light source is activated.
Figure 48. Intraoral Light Source Voltage
4. Use the buttons behind the terminal to adjust the voltage according to
the Length and Voltage Table, page 45.
44 86.0016.00 Rev A
Intraoral Light Source Length and Voltage
Table 8. Length and Voltage Table
Voltage at terminal strip
Wire length in
A-dec tubing
(in)(cm)VDC +/- .02(in)(cm)VDC +/- .02
48 122 3.401082743.69
54 1373.431142903.72
60 1523.461203053.75
66 1683.491263203.78
72 1833.521323353.81
78 1983.551383513.84
842133.581443663.87
902293.611503813.90
962443.641563963.93
1022593.67
A-dec/W&H, Bien Air or other bulbs rated
at 3.5V
Length and Voltage
Wire length in
A-dec tubing
Delivery System
Voltage at terminal strip
A-dec/W&H, Bien Air or other bulbs
rated at 3.5V
NOTETable 8 pertains to fiber-optics powered with 26AWG wires, 0.75 Amp loads, and a desired bulb voltage of 3.2 VDC. For
fiber-optics powered with 26AWG wires and other ratings, use the equation:
T = (Z x 0.006 x Y) + X where:
T: Terminal strip voltage(VDC)
X: Desired voltage at lamp (VDC)
Y: Rated lamp/load current (in Amps)
Z: Length of 26AWG wire (inches) from terminal trip to lamp
86.0016.00 Rev A 45
A-dec 200 Service Guide
46 86.0016.00 Rev A
Illustrated Parts Breakdown
Delivery System
This section contains illustrated parts breakdowns specific to the A-dec 200
delivery system.
Part Identification
In this section, you will find serviceable components tables that correspond
to the illustrations. The tables identify all parts and kits, including those
that are not for sale. Parts that are not for sale are indicated with the symbol
shown below:
†— Indicates that the individual part is not available for sale. These parts
are typically part of a kit and/or larger assembly that is for sale.
238.0710.00 †Block cap8021.016.01Plug, package of 10
338.0711.01Diaphragm, package of 59004.005.02Washer, package of 10
438.1839.00Control block assembly (with barbs)10023.004.03Barb, 1/8", package of 10
538.0766.02Flow control screw, package of 51138.1775.00Performer (A-dec 200) block service kit
6001.021.00Screw12023.001.03Barb, 1/4", package of 10
13023.805.01Barb, 5/16", package of 10
† Indicates that the individual part is not available for sale
Figure 51. Control Block Assembly
52 86.0016.00 Rev A
Toggles Assemblies (Flush Toggle and Master On/Off Toggle)
33.0009.83
(Flush Toggle)
33.0173.00
(Master On/Off Toggle)
2
3
4
5
6
7
1
1
3
8
9
7
ItemPart NumberDescription
1011.038.01Pin, , package of 10
233.0050.00Valve Body
322.0040.00Spring
429.0830.00Stem With O-rings, 2 Way
533.0007.00Disk
633.0036.01Toggle, Momentary, Gray
7006.009.00Hex Nut, 15-32 X 9/16 X 3/32
829.0840.00Stem With O-rings, 3 Way
922.0462.03Toggle, Gray 3
Figure 52. Flush Toggle and Master On/Off Toggle
Delivery System
86.0016.00 Rev A 53
A-dec 200 Service Guide
1
2
3
Oil Collector
ItemPart NumberDescription
1001.016.01Socket head screw, 10-32 X 3/8"
2026.143.0Gauze pad
377.0059.01Oil collector assembly, white 2
Figure 53. Oil Collector
54 86.0016.00 Rev A
CUSPIDOR AND SUPPORT CENTER
5
This section provides detailed information related to service, maintenance,
and adjustment of the A-dec 200 cuspidor and support center.
Contents
• Product Overview, page 55
• Service, Maintenance, and Adjustments, page 56
• Illustrated Parts Breakdown, page 57
Product Overview
A-dec 200 Support Center provides chair side mounting of the A-dec 200
Delivery System, Cuspidor, A-dec 200 Dental Light, and Assistant’s
Instrumentation. The support center mounts to the A-dec 200 chair using a
post mount.
Figure 54. A-dec 200 Support Center with Cuspidor
86.0016.00 Rev A 55
A-dec 200 Service Guide
Pinch Valve
Service, Maintenance, and Adjustments
Bowl Rinse Flow Adjustment
Adjustments to the cuspidor bowl rinse flow are made inside the support
center. To adjust the flow:
1. Loosen the two thumb screws at the bottom of the support center and
carefully pull the cover out.
2. With the cuspidor bowl rinse on, tighten or loosen the pinch valve to
adjust the flow.
3. For the best rinsing action, adjust the flow pattern by rotating the bowl
rinse.
Figure 55. Pinch Valve Adjustment
Self-Contained Water System
The self-contained water system provides water to the handpieces,
syringes, and cuspidor cupfill. The system includes a 2 liter water bottle
that mounts to the support center and offers a way to ensure the quality of
treatment water. Turn the bottle counterclockwise to remove it.
WARNINGUse only A-dec self-contained water bottles. Do not use
any other bottles, including glass or plastic beverage bottles. Do
not use damaged bottles. These can pose a serious safety hazard
while pressurized. A-dec plastic water bottles cannot withstand
heat sterilization. Attempting to do so will damage the bottle and
your sterilizer.
CAUTION Use caution when using the self-contained water
system with accessories that require and uninterrupted water
supply (such as scalers) as these could get damaged without a
continuous water source. Do not use saline solutions, mouth
rinses, or any chemical solutions (not specified in this guide) in
your A-dec self-contained water system. These may damage the
system components and cause your dental unit to fail.
Figure 56. Self-Contained Water Bottle
Adjusting the Cuspidor Cupfill and Bowl Rinse
For adjustment information, see "Cuspidor Cupfill and Bowl Rinse" on
page 34.
56 86.0016.00 Rev A
Illustrated Parts Breakdown
Cuspidor
This section contains illustrated parts breakdowns specific to the A-dec 200
Support Center and Cuspidor.
Part Identification
In this section, you will find serviceable components tables that correspond
to the illustrations. The tables identify all parts and kits, including those
that are not for sale. Parts that are not for sale are indicated with the symbol
shown below:
†— Indicates that the individual part is not available for sale. These parts
are typically part of a kit and/or larger assembly that is for sale.
Contents
• A-dec 200 Support Center, page 58
• A-dec 200 Cuspidor, page 60
• Self-Contained Water Bottle, page 61
• Cupfill/Bowl Rinse Manifold, page 62
86.0016.00 Rev A 57
A-dec 200 Service Guide
A-dec 200 Support Center
ItemPart NumberDescription
1—Cuspidor assembly, (see page 60)
275.2271.00Cuspidor bushing
375.2252.00Top cover, white 2
4044.184.00Limit switch
5001.218.00Screw
6—Assistant’s telescoping arm assembly (see page 68)
7005.002.00Screw
860.0080.00Mounting plate
9003.112.00Top cover mounting screws
1075.2259.00Top plate
1175.2253.00Side cover, white 2
1275.2254.00Front cover, white 2
1338.1803.00Cupfill/bowl rinse manifold (see page 62)
1475.2270.00Chassis
15027.064.00Thumbscrew
1675.2258.00Support chassis
1724.0388.02Regulator assembly, 40 PSI preset
1890.1167.00Data line PCB
1990.1082.00Stand off, package of 5
2077.1086.00Lower bushing
2175.0150.00Post plug
2214.0486.00Water bottle assembly (see page page 61)
23006.052.00Nut
2477.1085.00Upper bushing
2577.1087.00Trim ring
2675.2255.00 Rear cover, white 2
58 86.0016.00 Rev A
Figure 57. A-dec 200 Support Center
1
Cuspidor Assembly (see page 60)
2
3
6
Assistant’s Telescoping Arm
(see page 68)
4
5
7
24
22
Self-Contained Water
Bottle Assembly,
(see page 61)
9
25
19
11
13
Cup Fill/Bowl Rinse Manifold
(see page 62)
15
7
14
17
12
23
8
18
20
21
9
10
16
7
26
Cuspidor
86.0016.00 Rev A 59
1
2
3
4
5
6
7
8
9
10
11
13
22
24
21
14
15
23
16
17
18
12
29
17
19
20
25
26
27
28
A-dec 200 Service Guide
A-dec 200 Cuspidor
ItemPart NumberDescription
177.0042.01Spout, bowl rinse, back mount
277.1112.00Spout, cupfill, cuspidor
3030.011.02O-ring, AS568-011, .301 ID x .070 W, package of 10
477.0105.00Socket, bowl rinse spout
5030.014.02O-ring, AS568-014, .489 ID x .070 W, package of 10
6030.012.02O-ring, AS568-012, .364 ID x .070 W, package of 10
777.0044.01Gasket, spout, cup cuspidor
877.0099.00Button, cuspidor fill/rinse, left
977.0100.00Button, cuspidor fill/rinse, right
1043.0010.00Switch assembly, 20"
1177.0097.01Support, cuspidor mount with barbs
12005.088.00Screw, socket head, 6-32 x 1-1/4
13001.088.00Screw, socket head, 10-32 x 5/8 Stainless Steel
1477.0108.01Top housing, cuspidor, back mount
1577.0040.00Tube, drain, cuspidor, back mount
16035.053.01O-ring, vacuum, 20 mm x 2 mm, package of 10
17002.058.00Screw, button head socket, 3/8-16 x 3/4
18004.141.00Washer, flat, STL, .261 ID
1977.0109.01Bottom housing, cuspidor, back mount
2075.2273.00Hub, cuspidor
2175.2274.00Cuspidor stop
22005.106.00Screw, 10-32 x 2 1/2, socket head
2312.0991.00Drain seal
24001.088.00Screw, socket head, 10x32 x 5/8 stainless steel
2577.0236.00Bracket, hold down, cuspidor
26006.009.00Nut, hex, 15/32-32 x 9/16 x 3/32
27004.132.00Washer, flat, stainless steel, .500 ID
28004.035.00Washer, flat, nylatron, .511 ID
2977.0038.01Bowl, cuspidor, ceramic
Figure 58. Cuspidor
60 86.0016.00 Rev A
Self-Contained Water Bottle
1
2
ItemPart NumberDescription
114.0487.00200 self-contained cap assembly
214.0486.00Water bottle
Cuspidor
Figure 59. Self-Contained Water Bottle
86.0016.00 Rev A 61
A-dec 200 Service Guide
2
3
5
6
7
5
7
4
6
1
Cupfill/Bowl Rinse Manifold
ItemPart NumberDescription
1041.660.00Solenoid, 24 vac assembly
2030.016.02O-ring, package of 10
377.0816.00 †Manifold
4001.088.00Socket head screw, 1/4 - 20 x 5/8, stainless steel
5004.005.02Washer, package of 10
6023.805.01Barb, 5/16 x 10 - 32, package of 10
7023.804.00Barb, 5/16 x 1/8 NPT
† Indicates that the individual part is not available for sale
Figure 60. Cupfill/Bowl Rinse Manifold
62 86.0016.00 Rev A
ASSISTANT’S INSTRUMENTATION
HVE
SE
6
This section provides detailed information related to service, maintenance,
and adjustment of the A-dec assistant’s instrumentation.
Contents
• Product Overview, page 63
• Service, Maintenance, and Adjustments, page 64
• Illustrated Parts Breakdown, page 67
Product Overview
The A-dec 200 assistant’s instrumentation is equipped with an autoclavable
syringe, high volume evacuator (HVE), and saliva ejector (SE). Some
configurations may include an optional dual HVE or instrumentation that
supports a chair-side or single-operatory vacuum system. Integrated into
the support center is the solids collector, which connects with the HVE and
SE to separate solids from the evacuated material.
The assistant’s vacuum instruments disconnect from the tubing for easy
cleaning, and they are fully autoclavable.
Figure 61. Assistant’s Instrumentation
86.0016.00 Rev A 63
A-dec 200 Service Guide
Service, Maintenance, and Adjustments
Auto-air Holder
Each handpiece in the holder assembly automatically activates when you
lift it from its holder.
Holders provide vacuum On/Off switching for users whose vacuum
system requires this functionality. The vacuum pump activates
automatically when you lift the HVE or SE from the holder. The vacuum
turns off when you place the HVE or SE back into the holder.
Figure 62. Auto-Air Holder
Positioning Assistant’s Holder
The independently adjustable holders rotate to allow customized
positioning for each instrument on the assembly.
To rotate holder:
1. Pull holder slightly away from the adjacent one.
2. Rotate to the desired position and release.
Figure 63. Rotating Instrument Holders
NOTEAuto-air holders rotate together. Standard holders rotate
individually
64 86.0016.00 Rev A
Assistant’s Touchpad Connections
24 VAC
Data Line
The assistant’s instrumentation uses a standard touchpad that serves as a
single touch surface for controlling the chair, dental light, and cuspidor.
The touchpad can rotate 340° for access and visibility.
Figure 64. Assistant’s Touchpad with Standard Holder Connections
Vacuum Instrumentation
To use the HVE and SE, lift the holder from the valve to activate the
vacuum.
NOTE Lifting the handpiece from the holder only activates the
vacuum with auto-air switches.
Figure 65. HVE and SE Operation
86.0016.00 Rev A 65
A-dec 200 Service Guide
Solids Collector
The solids collector aids in preventing solids from entering the central
vacuum system.
CAUTION Use appropriate gloves when handling
contaminated parts.
Replace Solids Collector Screen
1. Turn off vacuum or open the HVE control valve.
2. Remove the solids collector cap.
3. Remove the solids collector screen.
4. Discard the screen according to your local regulations.
CAUTION Do not empty the screen into the cuspidor. Doing so
could plug the drain.
5. Insert the new screen in the collector and replace the cap.
CAUTION Ensure that the solids collector cap is inserted properly.
Failure to do so will not allow for proper suction.
Figure 66. Replace Solids Collector Screen
66 86.0016.00 Rev A
Illustrated Parts Breakdown
This section contains illustrated parts breakdowns specific to the A-dec 200
Support Center and Cuspidor.
Part Identification
In this section, you will find serviceable components tables that correspond
to the illustrations. The tables identify all parts and kits, including those
that are not for sale. Parts that are not for sale are indicated with the symbol
shown below:
†— Indicates that the individual part is not available for sale. These parts
are typically part of a kit and/or larger assembly that is for sale.
Contents
• Telescoping Assistant’s Arm, page 68
• Assistant’s Touchpad Assembly, page 69
• Assistant’s Auto-Air Holder, page 70
• Assistant’s Standard Holder, page 72
86.0016.00 Rev A 67
A-dec 200 Service Guide
Assistance Touchpad Assembly
(see page 69)
1
2
3
4
4
8
5
6
7
11
Holders Assembly
(see page 70 for Auto-Air;
page 72 for Standard)
† Indicates that the individual part is not available for sale
* Part of Item 3, p/n 90.1216.00
70 86.0016.00 Rev A
Figure 69. Auto-Air Assistant’s Holders Assembly
1
2
10
6
7
8
9
12
14
15
17
18
18
21
19
23
3
4
5
11
16
22
20
13
86.0016.00 Rev A 71
A-dec 200 Service Guide
1
2
3
5
4
6
7
6
Assistant’s Standard Holder
ItemPart NumbersDescription
199.0649.01 †Tu r r et
299.0701.00Turret axle, assistant’s arm
399.0653.03
99.0650.03
99.0652.03
499.0660.00 †Axle
599.3684.00End cap assembly kit
699.0698.00Bearing
799.0699.00Inner hub assistant’s arm
† Indicates that the individual part is not available for sale
Figure 70. Assistant’s Standard Holder Components
Holder, SE/Syringe kit, White 2, 9 mm
Holder, standard HVE kit, White 2, 11 mm
Holder, assistant’s HVE kit, White 2, 15 mm
72 86.0016.00 Rev A
UTILITIES
7
This section provides detailed information related to service, maintenance,
and adjustment of the A-dec 200 utility area.
Contents
• Product Overview, page 74
• Service, Maintenance, and Adjustments, page 75
• Illustrated Parts Breakdown, page 77
Figure 71. Utilities on A-dec 200 Dental Chair
86.0016.00 Rev A 73
A-dec 200 Service Guide
Front Post
Back Post
Manual Shutoff
Valv es
Floor Box Cover
NOTE If cable ties are present in
the product and you need to
remove them for servicing, make
sure to replace the ties after service
is completed.
1
2
3
4
5
6
7
8
9
10
11
12
Product Overview
The utilities for your system are located under the chair floor box. To access
the utilities, pull the floor box cover up at the front posts, and then forward
and up to slide the cover off the back posts.
Shutoff Valves
Manual shutoff valves control the air and water to the system. To prevent
leaks, these valves should remain fully open (turned counterclockwise)
except while servicing the system.
Air and water pass through separate filters before entering the regulators.
Replace these filters when they become clogged and cause restricted flow.
Figure 72. Floorbox Cover Removal and Shutoff Valves
Gauge and Pre-Regulator
The pre-regulator controls the air and water pressure in the unit. The gauge
displays the unit air pressure.
Figure 73. Utilities Diagram
Table 9. Utilities Descriptions
ItemDescriptionItemDescription
1 Moisture separator
(optional)
2Cuspidor drain tube85/8" exhaust tube
320 mm tube to drain9Durr adapter
4Connector clips (four)1020 mm tube to liquid
5Liquid separator tank11Y connector
6Cap12Exhaust tube
7Air exhaust
separator drain
74 86.0016.00 Rev A
Service, Maintenance, and Adjustments
Filter Housing
Filter
Manual Shut
Off Valve
O-ring
Beveled Edge
Manual Air
Shutoff Valve
Filter Housing
Filter
Filter Housing
Manual Water
Shutoff Valve
Utilities
Air and Water Filter Replacement
Air and water pass through separate filters before entering the regulators.
Replace a filter when it becomes clogged and causes restricted flow.
Figure 74. Air and Water the Filter Components
.
CAUTION To ensure proper delivery system operation,
install the filter with the beveled edge facing the manifold.
NOTETurn the pre-regulator knob clockwise to increase
pressure and counterclockwise to decrease pressure. Read
the pressure gauge while adjusting. Water pressure will
increase/decrease by half of the gauge indication.
Figure 75. Replacing Filters
To replace the Filter:
1. Turn off the master toggle and close the manual shutoff valves (turn
clockwise).
2. Bleed the system of air and water pressure by operating the syringe
buttons until air and water no longer flow.
3. Using a standard screwdriver, remove the filter housing from the air or
water filter pre-regulator assembly and remove the filter.
4. Replace the filter if it is clogged or discolored. Install the filter with the
beveled edge facing the manifold.
86.0016.00 Rev A 75
A-dec 200 Service Guide
76 86.0016.00 Rev A
Illustrated Parts Breakdown
Utilities
This section contains illustrated parts breakdowns specific to the A-dec 200
utilities.
Part Identification
In this section, you will find serviceable components tables that correspond
to the illustrations. The tables identify all parts and kits, including those
that are not for sale. Parts that are not for sale are indicated with the symbol
shown below:
†— Indicates that the individual part is not available for sale. These parts
are typically part of a kit and/or larger assembly that is for sale.
1026.154.00Gauge1124.0135.00Regulator, body, white
2001.024.00Screw1222.0460.00Spring
324.0365.00Bracket1324.0132.00Piston with O-ring
424.0232.00Stud, filter/regulator manifold1422.0440.02Diaphragm, package of 10
524.0234.01Filter, package of 61524.0368.00Valve cover
6030.019.03O-ring, package of 1016004.005.02Washer, flat nylon, 187 ID
724.0229.00Filter housing17023.004.01Barb, 1/8 x 10-32
8021.042.00Adaptor18001.026.00Screw, socket head, 6-32 x 7/8"
924.0162.00Manifold, filter/regulator19006.015.00Nut, hex, 6-32 x 5/16 x 9/64"
1024.0137.01Gasket, 9-hole2024.0182.81Regulator, air assembly
21006.009.00Nut, hex, 15/32 - 32 x 9/16 x 3/32"
Figure 76. Air and Water Filter/Regulators Body Assembly Components
78 86.0016.00 Rev A
Water Filter/Regulator Body Assembly
1
2
3
4
5
6
17
7
8
9
10
11
16
12
15
14
13
18
19
ItemPart NumberDescription
124.0229.00Housing, filter
2030.019.03O-ring, package of 3
324.0234.02Filter kit, package of 2
424.0232.00Stud, filter/regulator manifold
5021.016.04Plug, hex head, 10 - 32
6004.005.02Washer
7021.042.00Adapter, 1/8" MPT to 3/8"
824.0137.019-hole gasket
924.0355.00Body, black
10013.032.00Spring
1124.0132.00Piston with o-ring
1224.0140.00Spacer, water regulator
1322.0440.02Diaphragm, package of 10
14023.004.01Barb, 1/8" x 10-32
15004.005.01Washer, flat nylon 187 ID
1624.0368.00Valve cover
17001.021.00Screw, socket head, 4-40 x 1/2", stainless steel
18001.024.00Screw, socket head, 4-40 x 3/8", stainless steel
1924.0142.00Plunger
Utilities
Figure 77. Air and Water Filter/Regulators Body Assembly Components
86.0016.00 Rev A 79
A-dec 200 Service Guide
80 86.0016.00 Rev A
DENTAL LIGHT
Light Shield
Ta b ( h id de n)
Light Shield
Pivot Housing/Diagonal
Tension Adjustment
8
This section provides detailed information related to service, maintenance,
and adjustment of the A-dec 200 Dental Light.
Contents
• Product Overview, page 82
• Service, Maintenance, and Adjustments, page 83
• Illustrated Parts Breakdown, page 89
Figure 78. A-dec 200 Dental Light
86.0016.00 Rev A 81
A-dec 200 Service Guide
Treatment
Positions
X-Ray /Rinse PositionEntry/Exit Position
Product Overview
The dental light provides three intensities at the light head and the optional
touchpad.
Dental Light Specifications
• Electrical (Transformer Output): 12.1/16/17 VAC
• Bulb: Quartz Xenon Halogen, single-end prongs, extended life
• Rating: 17V/95 watts
• Color temperature: 4800 Kelvin
• Heat output: 325 BTU/Hour
• Light Pattern: 3.9" x 6.7" at 27.6" (100 mm x 170mm at 700 mm)
Nominal Light Intensity
• Composite: 4400 lux (743 fc)
• High: 17,000 lux (2230 fc)
On/Off Button
The dental light can be operated from the assistant’s or doctor touchpad.
To turn the light on, off, or change intensity, press the dental light button
on the touchpad.
Auto On/Off Feature
The optional auto on/off feature turns the light on when the chair back
reaches a treatment position. Press or , and the dental light turns off
Figure 79. Touchpad and Footswitch Light Features
To disable the auto. Press any button that moves the chair into an operating
position. The dental light turns on when the chair reaches that position.
To activate/deactivate the auto on/off feature, press and hold the program
and dental light buttons simultaneously for three seconds. One beep
confirms the auto on/off is off. Three beeps confirm the auto on/off is on.
82 86.0016.00 Rev A
Service, Maintenance, and Adjustments
10
A
M
P
10
A
M
P
On
On
Off
LED
Indicator
Dental
Contents
• Dental Light Wire Connections on the 200 Dental Chair, page 84
• Intensity Switches, page 83
• Dental Light Wire Connections on the 200 Dental Chair, page 84
• Dental Light Relay Circuit Board, page 86
• Bulb Replacement, page 88
• Rotation Adjustments, page 87
• Bulb Replacement, page 88
Circuit Breaker Location
A circuit breaker will interrupt the flow of electricity under abnormal
conditions. If the circuit breaker should trip, inspect the wiring to ensure
there are no shorts, and reset by pushing the circuit breaker. The circuit
breaker for the dental light is located on the power supply.
Figure 80. Circuit Breaker
Intensity Switches
A-dec dental lights can be operated from the manual 3-way switch or the
optional touchpad. The dental light is always off when the 2-way switch is
in the center location. To turn the light on or off from the touchpad, press
and hold the dental light button.
The dental light features two intensities: high and composite (low). For
systems without a touchpad, flip the 3-way switch either side of center to
select the intensity. For systems with a touchpad, ensure that the 2-way
switch is in either of the On positions, and press the light button to choose
the intensity. When the light is in the composite setting, the LED indicator
on the touchpad flashes.
Figure 81. Intensity Switches
86.0016.00 Rev A 83
A-dec 200 Service Guide
Dental Light
Circuit Board
BLK/WHT
Blk
BLU
VIO
WHT
YELWHT
VIO
BLU
YEL
BLK/WHT
Blk
VIO
BLU
Blk
VIO
BLU
Blk
VIO
BLU
Blk
VIO
WHT
WHT
Firststar Light
Switch
200 PS Power Cord
To AUX Power Connector
on Chair Circuit Board
300 W Power Supply
100 V (p/n 43.0214.00)
100 - 240 V (p/n 43.0215.00)
220 - 240 V (p/n 43.0216.00)
WHT
NOTEFor field installation, cut off the existing red
connector and strip the wires.
BLU
Blk
Dental Light Wire Connections on the 200 Dental Chair
Term in al
Vol tag eTerminal LabelWire
200 Dental Light
J217/12.1 VACVIOViolet
J20 VACBLKBlack
NOTESee "Circuit Board Components" on page 9 to identify dental light connections on the chair circuit board.
NOTEThe LED indicator must be connected to the power supply for the dental light circuit board to function correctly.
Figure 82. Dental Light Wire Connections With Touchpad
84 86.0016.00 Rev A
Figure 83. Dental Light Wire Connections Without Touchpad
Dental Light
Circuit Breaker
300 W Power Supply
100 V (p/n 43.0214.00)
100 - 240 V (p/n 43.0215.00)
220 - 240 V (p/n 43.0216.00)
200 PS Power Cord
To Auxiliary Power Connector on
Chair Circuit Board
DS3 - Data LEDOffNo DCS communication, not connected to
DS4, DS5 Dental light
relay LEDs
OffNo 24 VAC power, tripped circuit breaker,
power supply turned off, no line voltage
Green, steady24 VAC at the terminal strip
OffSystem is not functioning, no power or
circuit board has failed
Green, steadyNormal operation
the DCS, or DCS has failed
Green, steadyDetects active DCS
Green, blinkingValid DCS message
DS4, DS5
Off, OffDental light off
On, OffHigh intensity
Off, OnComposite intensity
On, OnMedium intensity
86 86.0016.00 Rev A
Dental
Adjustment Nut
Loosen
Tighten
Tension Adjustment Screw
Tension Adjustment Nut
Yo k e P l u g
Flexarm Adjustments
Remove the screw and cover from the flexarm. Turn the tension adjustment
nut inside the flexarm using a 1/2" open end wrench. Tighten the nut by
turning it clockwise, if the flexarm moves too easily, or tends to drift down
by itself. Loosen the nut by turning it counterclockwise, if the arm drifts up.
NOTEThe weight of the flex arm cover affects the flex arm
counter-balance. Set cover on flex arm to test tension adjustment.
Figure 85. Adjustment
Rotation Adjustments
If the light head is difficult to position, moves too easily, or tends to slip out
of position, the rotation tension can be adjusted diagonally and vertically.
Diagonal Adjustment
Use a 5/64" hex key to adjust the setscrew on the pivot housing. Turn the
screw clockwise to increase tension; counterclockwise to decrease tension.
Figure 86. Diagonal Adjustment
Vertical Adjustment
1. Remove the yoke plug on one side of the light.
2. Using a 5/16" nut driver, turn the adjustment nut clockwise to increase
tension; counterclockwise to decrease tension.
3. Reinstall the yoke plug.
Figure 87. Vertical Adjustment
86.0016.00 Rev A 87
A-dec 200 Service Guide
Light Shield
Tab ( hidden)
Light Shield
New Bulb
Bulb Replacement
Follow these steps when replacing the dental light bulb:
1. Turn the light off and allow it to cool.
WARNINGTo avoid burning your fingers, allow the bulb to cool
before removing. Never operate the dental light with the light
shield removed. The clear shield contains UV blocking additives
and is also your protection in the unlikely event that the bulb
shatters.
2. Hold the light shield and gently squeeze one side while pulling the
shield away from the dental light. Set the light shield aside.
3. Using a gauze pad or cloth to protect your fingers, carefully pull the old
bulb from its socket. Discard the bulb.
4. Holding the new bulb in its outer wrapper, carefully insert the bulb pins
into the socket. A small section of each pin is still visible when the bulb
is fully seated.
Figure 88. Bulb Replacement
5. Remove and discard the outer wrapper. Then reinstall the light shield.
6. Verify the operation of the light by turning it on and operating it at each
88 86.0016.00 Rev A
CAUTIONTake care when handling the bulb. The bulb base is
fragile and can break under excessive pressure. Do not remove the
outer wrapper when handling the new bulb. Finger oils can affect
light performance and severely limit bulb life. If you should
inadvertently touch the bulb, gently clean it with cotton
dampened with isopropyl or ethyl alcohol.
CAUTIONDo not attempt to install the light shield if it has a
broken tab. Contact your authorized A-dec dealer for a
replacement shield and install it before operating the light.
intensity setting.
Illustrated Parts Breakdown
1
2
3
Dental
This table lists the 200 Dental light parts and part numbers that correspond
to the illustration.
200 Dental Light
ItemPart NumberDescription
1041.709.00Bulb
297.0411.00Shield
390.1304.00Light switch kit
490.1309.00Light socket kit
Figure 89. Dental Light Bulb, Shield and Light Switch Kit
86.0016.00 Rev A 89
A-dec 200 Service Guide
90 86.0016.00 Rev A
9
TROUBLESHOOTING
The troubleshooting tables contain tips and troubleshooting information to assist in diagnosing problems. This content is not intended to cover every
situation, but includes the most common problems that you may encounter.
Dental Chair
ProblemPossible CauseAction
Factory defau lt test halts
during the Base Up test,
and the circuit board beeps
one time
Chair base or back is stuck in
full up position
The limit switch is not
activated, or
Down solenoid poppet is
unable to open based on
excess hydraulic pressure
Input voltage could be low or
outside the required range.
Motor capacitor or Base Up
solenoid are defective
Potentiometer is not changing
voltage.
Hydraulic lock has occurredCorrect Hydraulic Lock
Verify that the input voltage and voltage selection resistors (100 - 120 VAC = R72) and R74 (220 - 240
VAC = R 73).
If Base limit switch is activated, verify switch operation.
If the motor thermal limiter is open, and the motor is hot, wait for the motor to cool off.
Replace the motor capacitor or Base Up solenoid.
Verify the potentiometer LED comes On when the base is moving.
Check the potentiometer mechanical drive and electrical connections.
1. Remove the motor/pump cover from the chair.
2. Fit a 5/8" open end wrench to the high pressure outlet port (either lift or tilt, whichever is in
hydrostatic lock) of the hydraulic manifold.
3. Wrap a rag around both the fitting and the end of the wrench. The rag will absorb the small
amount of fluid vented from the fitting.
4. Carefully loosen the fitting about one-half turn, counterclockwise, and re-tighten.
5. Cycle the chair a couple of times to verify it is no longer in hydraulic lock.
Factory defau lt test halts
during the Back Down test,
and the circuit board beeps
one time
86.0016.00 Rev A 91
Stop plate limit switch is
activated
Verify switch operation
A-dec 200 Service Guide
ProblemPossible CauseAction
Stop plate is jammedRemove and reinstall the stop plate.
Back Down solenoid is defectiveTest the solenoid and replace if needed.
Back is in hydrostatic lockSee Correct Hydraulic Lock, page 91
Factory default test halts
during the Back Up test
Factory default test halts
during the Base Down test
Chair moves by itself when
power is turned On
Potentiometer is not changing
voltage
Back up limit switch is activatedVerify switch operation.
Back Up solenoid is defectiveTest solenoid and replace if needed.
Back is in hydrostatic lockSee Correct Hydraulic Lock, page 91
Potentiometer is not changing
voltage.
Stop plate limit switch is
activated
Base Down solenoid is defectiveTest solenoid and replace if needed.
Base is in hydrostatic lockSee Correct Hydraulic Lock, page 91
Potentiometer is not changing
voltage
The jumper is in FACT DEFAULT
position
Short circuit in the touchpad or
footswitch
Short circuit on the circuit boardReplace the circuit board.
Verify that the potentiometer LED is On when the back is moving.
Check the potentiometer mechanical drive and electrical connections.
Verify potentiometer LED is On when base is moving.
Check potentiometer mechanical drive and electrical connections.
Verify switch operation
Verify that the potentiometer LED is On when the back is moving.
Check the potentiometer mechanical drive and electrical connections.
Verify that the jumper is in the SPARE position
Unplug the touchpad and footswitch; reset the circuit breaker. If the problem isn’t repeated, the
touchpad or footswitch may have shorted.
No power to chair or unit.
Office still has power.
92 86.0016.00 Rev A
The chair is unplugged.
The Mains On/Off button is in Off
position.
1. Verify power is available at the outlet.
2. Plug chair in to power source.
Press the Mains On/Off button.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.