3B Scientific Millisecond Counter User Manual

3B SCIENTIFIC
Millisecond Counter (230 V, 50/60 Hz) 1012832 Millisecond Counter (115 V, 50/60 Hz) 1012833
Instruction sheet
06/12 SD
®
PHYSICS
1 4-mm-socket “Start” 2 4-mm socket “Stop” 3 4-mm socket “Ground” 4 Power supply socket 5 4-digit display
1. Safety instructions
The millisecond counter conforms to the safety regulations for electrical test, control and laboratory equipment as specified in DIN EN 61010 Part 1. It is intended for use in dry rooms suitable for electrical equipment or installations.
Safe operation of the equipment is guaranteed, provided it is used correctly. However, there is no guarantee of safety if the equipment is used in an improper or careless manner.
If it is deemed that the equipment can no longer be operated without risk (e.g. visible damage has occurred), the equipment should be switched off immediately.
Use the equipment in dry rooms only.
Only operate using the supplied plug-in
power supply.
Do not exceed the maximum input voltage of 5 V for the start and stop inputs.
2. Description
The millisecond counter is a compact counter for measuring times in the millisecond range. It is particularly well suited for use in experiments with the free-fall apparatus 1000738. Counting starts with a signal to the “Start” input and stops on one to the “Stop” input. Zero calibration is carried out automatically for each start. Both inputs respond to a rising edge and feature internal pull-up resis­tors.
The millisecond counter 1012833 is for operation with a mains voltage of 115 V (±10%), and the unit 1012832 is for operation with a mains voltage of 230 V (±10%).
3. Contents
1 Counter 1 Plug-in power supply 1 Instruction sheet
1
4. Technical data
7. Operation
Inputs:
Connectors: 4-mm safety sockets
7.1 General operation
Connect a 12 V AC plug-in power to the
Internal resistance “Start” input: 2.4 kΩ “Stop” input: 5.6 kΩ
Connect both inputs (1 + 2) to ground (3).
Opening the “Start” input (1) (disconnecting
Edges for “Start” and “Stop” inputs: Rising
Trigger threshold for “Start” input: Low 0...0.5 V, high 1...5 V “Stop” input: Low 0...1 V, High 2...5 V
Display:
Display: 4-digit LED display Measuring range: 1...9999 ms Resolution: 1 ms
Measurement ceases as soon as the “Stop” input (2) is opened.
The display will go back to displaying zero as soon as both inputs are connected to ground again.
Both inputs should be connected via normally closed switches (see Fig. 1).
Precision: Quartz precision
General data:
Voltage supply: Plug-in power supply
12 V AC, 500 mA Dimensions: 100x75x35 mm³ approx Weight: 400 g approx. incl. plug-
in power supply
5. Care and maintenance
millisecond counter (4).
the ground) starts the measurement.
Before cleaning the equipment, disconnect it
from its power supply.
Use a soft, damp cloth to clean it.
6. Disposal
The packaging should be disposed of at
local recycling points.
Should you need to dispose of the equip-
ment itself, never throw it away in normal domestic waste. Local regulations for the disposal of electrical equipment will apply.
Fig. 1 Schematic diagram for connection of the
inputs
7.2 Set-up with free-fall apparatus
Additionally required: 1 Free fall apparatus 1000738
Connect the inputs (1, 2, 3) of the millisec­ond counter to the 3 sockets of the free-fall apparatus making you sure you match all the colours (see Fig. 2).
Connect a 12 V AC plug-in power to the millisecond counter (4).
Measurement starts as soon as the steel ball is released from the start console and stops when it hits the trap plate. Zero calibration is carried out as soon as the steel ball is placed on the start console. The timer is then ready to carry out another measurement.