Information contained in this document is copyright Carl Zeiss Meditec Incorporated and may not be
reproduced in full or in part by any person without prior written approval of Carl Zeiss Meditec. Its
purpose is to provide the User with adequately detailed information so as to repair, maintain, and order
spare parts for the instrument supplied. Every effort has been made to keep the information contained
in this document current and accurate as of the date of publication or revision. However, no guarantee
is given or implied that the document is error-free or that it is accurate with regard to any specification.
TRADEMARK CREDITS
Humphrey Field Analyzer is a registered trademark of Carl Zeiss Meditec Inc.
All trademarks, registered trademarks, and trade names that appear in this manual are the property of
their respective holders.
Revision C
Carl Zeiss Meditec Inc.
5160 Hacienda Drive
Dublin, CA 94568
Document:Field Analyzer Series HFA II - i Field Service Guide
Part No.:52235 - Revision C
Issued Date:November 2005
Listed at the bottom of each page is the part number of the field service guide, along with the Revision
letter and date for that page (for example, 52235C1105). Subsequent revisions to a page will be noted
by a corresponding change to the Revision letter and date.
Pages in this document are at Revision C unless noted otherwise below.
Note - The Revision “B” Humphrey Field Analyzer Series HFA II-i Field Service Guide, P/N 52235
has been revised to Revision “C”. The new revision updates all references from “Humphrey Systems”
to “Carl Zeiss Meditec Inc. (CZMI)”.
PageRevisionPageRevisionPageRevisionPageRevision
iii
Humphrey Field Analyzer II - iTable of Contents
Table of Contents
Section 1 - General Information
1.1About This Field Service Guide ................................ 1 - 3
1.1.1 General Information ................................... 1 - 3
This Service Guide is the field service reference for troubleshooting, repair, adjustment, and
calibration of the Models 720i, 740i and 750i HFA II - i Field Analyzers, manufactured by
Carl Zeiss Meditec Inc. The information presented in this Guide assumes that the reader is
already trained and experienced in operation and service of the Humphrey Field Analyzer
Series 700.
This field service guide is designed to support Level 1 Field Service, which employs modular
replacement of printed circuit boards and other assemblies that are most effectively repaired
at a central repair facility. This is the service strategy used in U.S. domestic Field Service,
and in Carl Zeiss Meditec Service training classes. See Section 1.3 for additional information
regarding Level 1 and Level 2 Repair Center service.
The procedures in this field service guide assume that the reader is familiar with operation of
the instrument. Complete operating instructions are contained in the HFA II - iUser's Guide. Information contained in the User's Guide is not repeated in this field service
guide. The User's Guide can be ordered separately by standard Carl Zeiss Meditec parts
order. Refer to Section 7 for the User's Guide parts information.
The general layout of the field service guide is shown below. For greater detail, please refer
to the Table of Contents.
Level 1 Field Service Guide Layout
Section 1 General Information
Section 2 Preventive Maintenance & System Checkout
Section 3 Parts Removal/Replacement
Section 4 Adjustment/Calibration
Section 5 Troubleshooting
Section 6 Diagrams
Section 7 Parts
Appendices
Confidential and Proprietary
52235C11051 - 3Field Service Guide
General InformationHumphrey Field Analyzer II - i
1.1.2Conventions
The following conventions apply in this manual:
PThe terms left, right, front and back of the instrument are as viewed from the
patient position, unless noted otherwise.
PDimensions are given in inches unless noted otherwise.
1.2About Service Bulletins
Field Service Bulletins are a vital element of service support. Bulletins are used to quickly
convey technical information on a variety of field service topics, including:
Pinstrument design changes
Ptechnical problems and
Service bulletins are used to issue revised pages for service manual/service guide updates.
Your service bulletins should be filed where easily accessible for quick reference.
Pservice manual revisions
Pupgrade announcements/
procedures
Psystem checkout – checklist
Pservice disclaimer forms
Psystem work sheets
NOTICE
Field Service Bulletins are Confidential and Proprietary, for the sole use of
personnel employed by Carl Zeiss Meditec, Carl Zeiss Meditec affiliates, and
authorized Carl Zeiss Meditec distributors.
Carl Zeiss Meditec has a well-deserved reputation for high quality, reliable
instruments, unsurpassed in the industry.
As a Carl Zeiss Meditec employee, affiliate, or distributor you are required to
handle your service bulletins as appropriate for proprietary and confidential
information.
Confidential and Proprietary
Field Service Guide1 - 452235C1105
Humphrey Field Analyzer II - iGeneral Information
1.3HFA II - i Service Strategy
1.3.1Two-Level Service Strategy
A two-level service strategy is used for the HFA II - i: Level 1 for on-site service; and Level 2
for Repair Center service. Level 1 service employs modular replacement, wherein faulty
circuit boards and certain other assemblies are replaced rather than repaired on-site. These
faulty assemblies are shipped to a Carl Zeiss Meditec Repair Center for repairs. There are
also certain procedures that require special equipment available only at a Repair Center.
Designated Repair Centers (currently Dublin, CA and Jena, Germany) are the second level
of service for the HFA II - i. The Repair Centers perform major circuit board troubleshooting
and repair, plus any other service action that requires special equipment or procedures not
available in the field.
Several of the circuit boards in the HFA II - i are multilayer boards and use Surface Mount
Technology (SMT) components. These boards require special equipment and techniques for
troubleshooting and repair.
All service procedures (including instrument calibration) can be performed in the field, except
for those listed below. For Carl Zeiss Meditec U.S. Domestic operations, the following
procedures must be performed at the Carl Zeiss Meditec Repair Center.
PCircuit board troubleshooting and component replacement
PRepair of floppy, tape, and hard drives.
PAlignment of projection carriage rails and first projection mirror (top turret mirror)
PRepair of power supply assembly
PRepair of camera assembly
1.3.2Three Steps to Completing an HFA II - i Service Call
The basic approach to an HFA II - i service call is outlined below. This typical process
includes collection of general instrument calibration data and light intensity data both Before
service and again After service. The process is described in detail in Section 4.8.1. For
guidelines, refer to HFA II - i Field Service Paperwork Requirements (Section 1.3.3).
1.Obtain the Before Light Intensity instrument data (4.8.1).
This step assumes that the HFA II - i is operable; that is, it will power up to the
Main Menu without error. This data gives the service representative a base from
which to evaluate the light intensity operation of the instrument, and a point of
comparison if recalibration is required.
Confidential and Proprietary
52235C11051 - 5Field Service Guide
General InformationHumphrey Field Analyzer II - i
If a repair is required to render the instrument operable, and as long as the repair
does not affect the original light intensity data, the repair can be performed and
then the Before data can be acquired.
The following repairs will affect light intensity data:
•Hard Drive replacement or initialization;
•cleaning or replacement of the ND wedges, color wheel*, or brightness
detector;
•replacement of the Motor Driver PCB.
*
Note - Cleaning or replacing the color wheel will not affect the white/white
Before light intensity data; it only affects the blue light intensity data.
2.Perform the needed instrument service.
This step includes any parts replacement, adjustments, calibration, cleaning, etc.
to repair, update and/or upgrade the instrument.
3.Obtain the After light intensity instrument data (4.8.1) if the Before data was not within
specifications, or if something was done during service that affects light intensity (see list
in step 1).
When instrument service has been fully completed (but before reinstalling the
outer covers), a final evaluation of the instrument may be required (see Section
2.1, System Checkout). During this step, the Before and After light intensity data
are compared. If necessary, a Calibration Notice is given to the customer.
1.3.3HFA II - i Field Service Paperwork Requirements
The following guide identifies the paperwork that must be completed and sent to Carl Zeiss
Meditec Customer Service following each service call or preventive maintenance visit on the
HFA II - i by U.S. Domestic service engineers. For all service engineers, this guide
identifies actions essential to properly perform various types of service calls on the HFA II - i.
For all service calls:
FSR
G
Before Foveal ‡
G
If replacing/initializing Hard Drive; cleaning/replacing brightness detector, ND wedges, or
color wheel; or replacing Motor Driver PCB:
If obtainable, include all items from list above under"For all service calls," plus items
G
listed below under "Additional if light intensity recalibration is required". If possible,
obtain Before paperwork prior to the repair.
Confidential and Proprietary
Field Service Guide1 - 652235C1105
Humphrey Field Analyzer II - iGeneral Information
Additional if upgrade to Blue-Yellow:
Before and After Light Intensity Verification Printouts
G
After Foveal ‡
G
Calibration Notice*, if needed.
G
If performing PM:
Covered by items listed above under "For all service calls."
G
Before Light Intensity Verification Printouts
G
Additional if light intensity recalibration is required:
Before and After Light Intensity Verification Printouts
G
After Foveal ‡
G
Calibration Notice*, if needed.
G
Notes -
*Formerly called "Doctor Card" or "Doctor Letter."
‡For Model 720i, see Section 4 in the HFA II - i Field Service Guide.
Confidential and Proprietary
52235C11051 - 7Field Service Guide
General InformationHumphrey Field Analyzer II - i
1.4Configuration Parameters
Configuration parameters can be entered and stored in the system by the user. This data is
stored on the hard disk. Calibration data also is stored on the hard disk. There is the
possibility that this data may become altered or erased during servicing of the instrument.
To minimize the possibility of altering the calibration values, configuration parameters or
doctor setups during service, the following practices should be observed.
•Whenever possible, when servicing a customer's instrument, backup the
calibration values on the calibration values disk. This option is available via the
Calibration Menu.
•Whenever possible, when servicing a customer's instrument, backup the customer-
selected configuration. This option is available via the Setup and Additional
Setup menus.
•When finished servicing the instrument, restore the customer's configuration
selections.
•Never intentionally alter the customer's existing doctor setups.
1.5Precautions
The following precautions should be observed whenever the HFA II - i is being installed or
serviced. Point out to the customer any potential hazard and the appropriate corrective action.
WARNING: The CRT and associated circuitry can deliver a lethal
shock. Always employ standard high-voltage safety precautions
when working around the CRT circuitry. DO NOT use metal tools
when making CRT adjustments.
General Safety Precautions
1.The instrument is equipped with a grounding-pin power plug. The instrument must be
plugged into an outlet with a properly grounded receptacle.
2.Ensure that the fuses installed in the instrument and the power table are of the proper
rating.
3.Use of an extension cord is not recommended. Doing so may compromise the safety of
the operator and/or patient.
4.Do not overload the AC outlet being used to operate the instrument.
Confidential and Proprietary
Field Service Guide1 - 852235C1105
Humphrey Field Analyzer II - iGeneral Information
5.If the power cord or plug on the instrument is damaged, a shock or fire hazard may
result. Do not allow continued operation of the instrument until the damaged cord or plug
has been replaced.
6.To prevent personal injury and damage to the instrument:
PUse only the power table recommended by Carl Zeiss Meditec.
PDo not place the instrument on uneven or sloped surfaces.
7.For stability of the power table, strictly adhere to the following guidelines:
PEnsure that the instrument is secured to the power table with the screws provided.
PDo Not place the instrument near the operator end of the table during setup,
service, or operation.
PBefore servicing the instrument, cycle the table to its lowest position and ensure that
the slider is locked in position.
8.Ensure that the instrument is installed on a stable, vibration-free surface.
9.Be cautious when you first touch the projection assembly. The projection assemblycan become extremely hot any time the projection bulb is lit for prolonged periods,
such as during sustained patient testing.
10. The bowl lamp voltage is approximately 550 volts — AVOID TOUCHING!
Instrument Precautions
1.When the instrument is being unpacked, save the shipping materials for possible future
use. Whenever the instrument is shipped, it must be properly packed to prevent damage.
Do Not place objects in the bowl during shipment of the HFA II - i.
2.When spare parts are received, save the shipping materials for returning the defective
part(s), if appropriate.
3.Handle interconnecting cables carefully. Many of these are constructed of extremely
small coax cable and are easily damaged.
4.To avoid possible damage to circuit board components, do not plug/unplug cables while
power is applied to the instrument.
5.Do Not use Windex® to clean the touch screen, as it can scratch plastics. The
recommended cleaner is a 50% solution of isopropyl alcohol and water. Otherwise, use
a commercially available anti-static plastic cleaner.
6.Use extreme care whenever working in or near the bowl to avoid causing marks or
scratches to the inner bowl surface.
Confidential and Proprietary
52235C11051 - 9Field Service Guide
General InformationHumphrey Field Analyzer II - i
7.Do not rub the inner bowl surface while attempting to clean it; rubbing will cause a
noticeable polished area in the bowl.
8.Do not touch the glass surface of the projection lamp with your bare fingers. Any oil, dirt
or grease on the lamp can shorten its effective life and diminish light output.
9.Do not place items on top of the unit; the internal mechanism of the instrument is very
close beneath the top cover.
10. Ensure that none of the ventilation openings in the instrument are blocked. Excessive
heat buildup within the instrument can cause instrument failures.
11. DO NOT flex the circuit boards. This instrument uses multilayer circuit boards. Multilayer
circuit boards are inherently susceptible to damage by excessive flexing.
12. Proper ElectroStatic Discharge (ESD) precautions must be observed whenever you are
disassembling or handling the instrument's circuitry. Many of the components are
extremely vulnerable to static discharge damage. A Field Service Static Protection Kit is
available for order and must be used for ESD protection during service of this instrument.
Refer to Appendix A for details regarding this kit.
13. An abrupt, harsh noise will ensue if the chinrest comes up against its travel limit during
chinrest movement. Moving the chinrest up against its limits for a few seconds causes no
mechanical harm.
1.6Internal Layout
The parts drawings in Section 7 of this Field Service Guide illustrate the internal physical
layout of the instrument. Diagrams in Section 6 illustrate the functional layout of the
instrument.
1.7Special Topics
1.7.1The Touch Screen
The HFA II - i uses a transparent, analog, resistive-membrane touch screen. It is constructed of
two pieces of thin, highly linear, electrically conductive film (Indium Tin Oxide). The two
pieces of film are separated by a small air gap. The air gap is maintained by small (.001"),
dielectric spacer dots.
Each film sheet has a set of parallel bus bars applied along opposite edges of the film. The two
sheets are oriented so that the bus bars on one sheet are perpendicular to those on the other
sheet. Slight pressure will cause the conductive surfaces to come into contact. The location of
the contact point can be detected by a logic circuit measuring the voltage found at that
particular point.
The analog type of touch screen gives a "voltage divider" analog response that allows
positional determination.
Confidential and Proprietary
Field Service Guide1 - 1052235C1105
Humphrey Field Analyzer II - iGeneral Information
1.7.2Gaze Tracking
The HFA II - i uses two systems for measuring patient fixation: the standard Heijl-Krakau blindspot monitoring and the IR Gaze Tracking System. Both methods can be used, either together
or alone, or they can both be turned off, as required. This description covers the IR Gaze
Tracking System.
The direction of a patient's gaze is determined in two steps: first, a reflex marker is established on
the corneal surface; and second, the location of the pupil center is determined.
Gaze tracking is initialized in the following manner when a selected test is first started:
The patient is asked to fixate on the central illumination LED. Gaze tracking turns on the reflex
gaze IR LED located just under the diamond fixation pattern and turns off eye illumination
briefly . Light from the LED is reflected off the cornea, and back to the IR sensitive camera
(Figure 1-1). The majority of the cornea appears black except for the reflected spot. This image
is digitized and stored in memory. The reflected spot is referred to as the reflex marker
(Figure 1-2). Because the corneal surface is rounded, the reflex marker will move very little even
if the patient's eye rotates, and thus the marker becomes a (relatively) stationary reference point.
Next, the system locates the pupil center by illuminating the entire eye with the two IR LEDs
located either in the bottom of the bowl, or in the trial lens holder (when in the raised position).
The iris appears bright with a dark pupil (Figure 1-3). This image is also digitized and stored in
memory. It is the relationship between the location of the reflex marker on the cornea and the
location of the pupil center that determines fixation (Figure 1-4).
Note - When gaze tracking is being initialized, it appears as repetitive "strobing" when
viewed by the operator via the video insert on the HFA II - i monitor.
During a test, each time a spot is projected into the bowl, the locations of the reflex marker and
the center of the pupil are compared to the initial images stored in memory. If the patient is
fixating correctly, the positional relationship between the reflex marker and the pupil center will
be the same as that of the stored images (Figure 1-4). If the patient is off fixation, the positional
relationship between the reflex marker and the pupil center will be different, as in Figure 1-5.
The greater the misalignment, the higher the mark on the Gaze Graph (Figure 1-6).
(continued)
Confidential and Proprietary
52235C11051 - 11Field Service Guide
General InformationHumphrey Field Analyzer II - i
Spikes that appear on the Gaze Graph (Figure 1-6) are analyzed as follows:
PUpward spikes indicate that the patient has lost fixation;
•a spike that reaches the top horizontal line (or higher) indicates 10 degrees (or
more) off fixation;
•a spike that extends halfway to the top line indicates 5 degrees off fixation.
PDownward spikes indicate as follows:
•a short spike downward indicates that the gaze at that time cannot be
determined by the software.
•a long spike downward indicates that the patient blinked at the time fixation was
checked.
The absence of marks on the graph indicates proper fixation.
Possible problems associated with the gaze tracking system are reflections from the trial lens,
fingerprints on the trial lens, an improperly aligned trial lens holder, an improperly calibrated or
aligned Gaze Tracking box, and excessive patient tearing.
FIGURE 1-1. Location of Corneal Reflex Marker
FIGURE 1-2. Corneal Reflex marker Location Digitized and Stored in Memory
Confidential and Proprietary
Field Service Guide1 - 1252235C1105
Humphrey Field Analyzer II - iGeneral Information
FIGURE 1-3. Determining the Pupil Center
FIGURE 1-4. Patient Fixating – Corneal
Reflex Marker and Pupil in Proper
Relationship
FIGURE 1-5. Patient Not Fixating –
Corneal Reflex Marker and Pupil Not in
Proper Relationship
FIGURE 1-6. Gaze Graph
Confidential and Proprietary
52235C11051 - 13Field Service Guide
General InformationHumphrey Field Analyzer II - i
1.7.3Head Tracking / Auto Pupil / Vertex Monitoring
These three features are dependent on the Gaze Tracking system. If Gaze Tracking has
successfully initialized, any of these three features can be utilized.
Head Tracking
The Head Tracking feature is designed to lessen the appearance of a trial lens artifact
image when the patient's eye is off center in relation to the center of the trial lens holder.
The intent is to reduce the possibility of inducing an arc-like defect or ring scotoma in the
patient's field test results.
Head Tracking is active when the trial lens holder is in the up position and Head Tracking
has been set to ON in the setup menu. The Head Tracking feature will track the center of
the eye in relation to the trial lens holder. If the patient's eye moves from the center of the
trial lens holder by more than 3 mm for more than one consecutive sample, the head
tracking feature will gently move the chinrest and headrest to automatically reposition the
patient's eye in the center of the trial lens holder. The tracking will stop if the eye doesn’t
follow the correction. Tracking begins when the test is started. This feature will operate
properly only if the patient properly rests on the chinrest.
Auto Pupil
When the Auto Pupil feature is set to ON in the setup menu, the gaze monitoring system
will determine the size of the patient's pupil to the nearest 0.5 mm at the beginning of each
test (during initialization of gaze tracking), and will automatically enter that information into
the Patient Data information screen, marked Auto (*).
Vertex Monitoring
The Vertex Monitoring feature is designed to lessen the appearance of a trial lens artifact
image when the patient's head moves backwards (away from) the trial lens holder. The
intent is to reduce the possibility of inducing an arc-like defect or ring scotoma in the
patient's field test results. Vertex Monitoring will alert the operator (via a double beep) if
the patient's eye moves more than 7 mm away from its original position for any one
measurement. (The sample rate is once every question.)
Vertex distance measurement is obtained during gaze initialization. During gaze
initialization, the two IR LEDs on the trial lens holder appear as two dots of light on the
corneal surface. (Refer to 1.7.2 Gaze Tracking for a complete description.) The distance
between these two dots will decrease as the head moves away from its original position. If
the distance exceeds the software limits, an alert will sound. The test continues, and a popup window appears allowing the user to reinitialize, continue without reinitializing, or turn
off vertex monitoring.
Confidential and Proprietary
Field Service Guide1 - 1452235C1105
Humphrey Field Analyzer II - iGeneral Information
1.7.4HFA II - i Light Intensity Fundamentals
The HFA II - i uses one detector mounted at the end of the turret to measure both spot and bowl
intensities. The projection lamp voltage is controlled by software to set maximum stimulus
brightness. This means that the lamp can be operated at a lower voltage when the lamp is new;
and as it ages, more voltage is applied in order to maintain the same level of brightness. This
increases lamp life expectancy and reduces power consumption.
During light intensity calibration, the projection calibration value is stored in memory and set as
close as possible to obtain 929 Ft-L or 10,000 asb of light output. The background lights are
fluorescent and the calibration values are set as close as possible to obtain 2.92 Ft-L or 31.5 asb.
During calibration, light attenuation is measured at 175 different points on each of the two ND
wedges, and the results are stored in memory. The two ND wedges are used in combination to
obtain the desired brightness of the projected spot. The duration of the spot is 200 ms and is
controlled by the software operating the shutter, located between the projection lamp and the
ND wedges.
During the power-on sequence, the bowl intensity is set to the calibrated value (2.92 Ft-L or
31.5 asb). The brightness detector is then pointed at a black patch located on the inside of the
front cover, the shutter is closed, and a measurement is made by the detector. This establishes
the zero asb reference. Next, the shutter is opened, and a spot projected on the bowl
approximately 35E above center is measured by the detector. The projection voltage is adjusted
to match the calibration value stored in memory. This measurement sets the maximum
brightness level (10,000 asb, or 0 dB). These two measurement points determine the slope of
the light from dark to maximum brightness.
If the measured intensity varies from that stored during calibration, the lamp voltage is adjusted
and measured again. This continues until the stored intensity and measured intensity match. If
the voltage is adjusted above 10 V, the test will fail and a projection lamp error message will
appear on the screen. If the test does not fail, ten different points on each ND wedge are
measured and compared with their stored values. If these values deviate more than ±.5 dB
from the stored values, the test will fail and a wedge failure error message will be displayed.
1.7.5Comparing HFA II and HFA II - i
Hardware -
The HFA II - i version differs from the prior production version, the HFA II, in that significant
upgrades have been made.
The HFA II - i uses an off-the-shelf CPU PCB featuring an Intel Celeron 433 MHz processor,
while the HFA II uses a 20 MHz Motorola 68020-based processor. The new CPU PCB
dramatically decreases the test processing time resulting in significant overall test time reduction.
The HFA II - i CPU PCB contains 64 megabytes of DIMM memory while the HFA II has 4
megabytes of memory on the current CPU. The increase is significant in the ability to support
software features planned for the future; in particular, the implementation of Ensemble. In
Confidential and Proprietary
52235C11051 - 15Field Service Guide
General InformationHumphrey Field Analyzer II - i
addition, the new board features components selected to ensure ease of parts availability,
particularly integrated circuit components.
The HFA II - i CPU PCB uses an IDE hard drive controller. The HFA II CPU PCB uses a SCSI
hard drive controller. The HFA II - i has an IDE hard drive while the HFA II uses a SCSI hard
drive. Note that the two drives are not interchangeable.
Intel Celeron Processor
Mouse/Glide Pad
Keyboard
Backplane
Board
Backplane
Board
USB
Device
USB
Device
PS2
Mouse
PS 2
Keyboard
PCI
EISA
Bus
USB
USB
System Memory 64MB
Flash Memory
(Boot Data)
NOVRAM/Clock
EIDE Controller
Floppy Controller
Serial Interface
Printer Interface
J 9
J 10
J 7
J 3
Ethernet
Port
J 4
Serial
Port
Serial
Port
Parallel
Printer
Port
Hard Drive
MO Drive
Network
Floppy
Disk
FIGURE 1. CPU PCB Functions and Interconnects
Confidential and Proprietary
Field Service Guide1 - 1652235C1105
Humphrey Field Analyzer II - iGeneral Information
Many of the unique functions that the HFA II CPU PCB supported are now included in the HFA
II - i Backplane PCB. The functions of the Backplane PCB include; the patient switch, touch
screen, camera frame grabber, and video.
FIGURE 2. Backplane PCB Functions and Interconnects
The DAT streamer tape of the HFA II has been replaced with a magneto-optical medium reader
in the HFA II - i. Besides reducing cost, the new HFA II - i drive provides greater long-term
storage of large patient files.
For EMI considerations, the HFA II - i CPU PCB and Backplane PCB are housed in a metal
enclosure. The enclosure is bolted the chassis in the location of the HFA II CPU PCB.
To accommodate the enclosure the rear cover was redesigned. The rear cover does not make
use of 1/4 turn fasteners.
Confidential and Proprietary
52235C11051 - 17Field Service Guide
General InformationHumphrey Field Analyzer II - i
The HFA II - i has a larger capacity power supply to provide the additional Vcc power required
for the HFA II - i.
The HFA II - i uses a new Motor Driver Board (MDB). The new MDB will ensure longer
component availability.
FIGURE 3. HFA II - i Interconnects
Confidential and Proprietary
Field Service Guide1 - 1852235C1105
Loading...
+ 407 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.