YSI EC300A User Manual [en, es, fr]

Operations Manual
EcoSense® EC300A
Portable
Conductivity, Salinity
and Temperature
Instrument
English Français Español
Portugu
ês
WARRANTY
The EcoSense® EC300A Instrument is warranted for one year from date of purchase by the end user against defects in materials and workmanship. EC300A probes and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by: (i) failure to install, operate or use the product in accordance with YSI’s written instructions; (ii) abuse or misuse of the product; (iii) failure to maintain the product in accordance with YSI’s written instructions or standard industry procedure; (iv) any improper repairs to the product; (v) use by you of defective or improper components or parts in servicing or repairing the product; or (vi) modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.
CONTACT INFORMATION
YSI Inc. 1725 Brannum Lane Yellow Springs OH, 45387 Tel: 800-897-4151; 937-767-7241 Fax: 937-767-1058 Email: environmental@ysi.com
Website: www.ysi.com
1
CONTENTS
WARRANTY .......................................................................................................................... 1
CONTACT INFORMATION ................................................................................................ 1
GENERAL INTRODUCTION ............................................................................................... 3
INITIAL INSPECTION .......................................................................................................... 3
THE INSTRUMENT
BATTERY INSTALLATION .................................................................................................. 3
Battery Disposal ......................................................................................................... 3
KEY FUNCTIONS OF THE MODEL EC300A .................................................................... 4
THE LCD DISPLAY ............................................................................................................... 4
MEASUREMENT MODES ................................................................................................... 4
CALIBRATION ...................................................................................................................... 5
TDS .............................................................................................................................. 5
CELL ............................................................................................................................. 5
Temperature Coefficient ........................................................................................... 5
Temperature Reference ............................................................................................ 5
Conductivity Calibration ........................................................................................... 5
CONDUCTIVITY MEASUREMENTS ................................................................................... 6
SAVING, VIEWING AND DELETING DATA
PROBE MAINTENANCE ..................................................................................................... 6
TROUBLESHOOTING ......................................................................................................... 7
SPECIFICATIONS ................................................................................................................ 8
RECOMMENDED SPARE PARTS LIST ............................................................................... 8
............................................................................................................. 3
................................................................... 6
2
GENERAL INTRODUCTION
The EC300A is a precision tool that measures conductivity, salinity and temperature. A built-in microprocessor calculates and compensates for all parameters related to conductivity and temperature determinations.
This instrument is waterproof (IP67) when the connector cap is installed. The mechanical touch keys are highly reliable with tactile and audio feedback. This instrument uses one 9V battery. Re-calibration is not required when power is restored.
The front of the instrument has a large LCD that displays temperature and either temperature compensated or non-temperature compensated conductivity, salinity or TDS simultaneously along with user prompts and mode indicators. The unit prompts users through calibration and measurement procedures.
The model EC300A is available with a single four-electrode cell. Other features include automatic conductivity ranging, automatic temperature compensation, long battery life, and 50/60 Hz AC noise rejection. This meter is universal and user-friendly for field, industrial and laboratory applications.
INITIAL INSPECTION
Carefully unpack the unit and accessories, and inspect for shipping damages. Compare received parts with materials listed on the packing list. Notify YSI immediately of any damage or missing parts. Save all packing materials until satisfactory operation is confirmed.
THE INSTRUMENT
Though the instrument is housed in a water-proof IP67 case, DO NOT use it underwater. The connector is not waterproof unless the cap is installed. In case of submersion without the cap connected, follow these steps immediately:
1. Dry the connector if necessary, and replace the conductivity probe. Rinse unit carefully
with distilled water. After rinsing and drying, inspect and clean connectors to remove all contaminants that may affect probe connections
2. Wait for the unit and probe to dry completely before resuming operation.
3. If the unit does not function correctly after steps 1 and 2, call YSI for possible repair or
replacement (see Warranty)
.
.
BATTERY INSTALLATION
An initial display of “BAT” on the LCD indicates approximately one hour of battery life for unit operation within specifications. Replace battery when “BAT” appears on the LCD. (See Figure 1.)
To replace battery, remove the two battery cover screws, battery cover and o-ring. Replace the 9V battery. Replace battery cover and o-ring (align the o-ring properly to insure a good seal) and fasten the two battery cover screws for the splash-resistant feature.
Battery Disposal
This instrument is powered by a 9 volt battery, which the user must remove and dispose of when the batteries no longer power the instrument. Disposal requirements vary by country and region, and users are expected to understand and follow the battery disposal requirements for their specific locale.
3
Figure 1.
Battery Installation
KEY FUNCTIONS OF THE MODEL EC300A
1. : Turns the unit ON or OFF. Calibration values are not erased when the unit is turned
off. When the unit is not in use, turn it off to save battery life. The instrument has a 30 minute auto shut off feature when not in use. For long-term storage, remove the battery.
2. MODE: Selects display mode. In Normal operation, press MODE to switch the display
between uncompensated conductivity, temperature compensated conductivity, salinity, total dissolved solids (TDS), Recall and Delete.In calibration mode, this key exits the current calibration and displays the next calibration parameter.
3. CAL: In normal operation, changes the mode from Normal to Calibration.
4. (Enter) : In Calibration Set-up, press this key to save the current parameter to
memory.
5. and Keys: Increases or decreases the display value as desired.
THE LCD DISPLAY
1. CONDUCTIVITY: Displays when
measuring conductivity.
2. BAT: Low battery indicator.
3. CELL: Indicates conductivity cell
constant value.
4. Main display for compensated and
uncompensated conductivity, salinity and TDS values.
5. TDS: Displays when measuring total
dissolved solids.
6. SALINITY: Displays when measuring
salinity.
7. CAL: Calibration mode indicator.
8. C: Flashes during temperature compensated conductivity measurement. During
calibration, indicates temperature reference unit.
%: Displays during calibration; indicates temperature coefficient unit.
9. ppt: Parts per thousand; indicates salinity measurement.
10. uS, mS: micro Siemens, milli Siemens; Indicates conductivity measurement.
11. g/L: grams/Liter; indicates TDS measurement.
12. C: Temperature display.
Figure 1. LCD Display
MEASUREMENT MODES
1. Temperature - Current solution temperature continually displays.
2. Temperature Compensated Conductivity - Measurement of conductivity, compensated
to 25C or another specified value between 15 and 25C. Expressed as uS/cm or mS/cm with a flashing “C”.
3. Uncompensated Conductivity – Direct measurement of conductivity, not compensated
4. Salinity – Measurement of salinity; expressed in parts per thousand (ppt).
to a specific temperature. Expressed as uS/cm or mS/cm.
4
5. TDS – Measurement of total dissolved solids (TDS); expressed in grams per liter (g/L). Carefully observe the units displayed at the far side of the LCD to determine the desired
mode.
CALIBRATION
Calibration setup contains five sections: TDS, Cell, Temperature Coefficient, Temperature reference, and Conductivity Calibration. To access these sections:
1. Connect the conductivity probe and cable assembly to the unit and turn the unit on. The
screen will display CELL and the cell constant of the conductivity probe.
2. Allow temperature readings to stabilize, then press CAL to enter the calibration mode;
CAL appears on the LCD. Press MODE to sequentially display the following sections:
Note: Press Enter ( ) to accept any values changes in each section and automatically
advance to the next section. If there are no changes, the unit accepts the current value and proceeds to the next section.
TDS
TDS is determined by multiplying conductivity (mS) by a TDS factor. The default factor value is 0.65. To change the TDS factor, use the and keys to adjust the value between 0.30 and
1.00. Press Enter ( ) to save the new value, or press MODE to cancel the change and display the CELL screen.
CELL
The second screen will display CELL and the current cell value. The default cell value is 5.00 and is displayed in the lower right of the screen. The unit allows a variance of ±0.50 before displaying an error message. The cell value cannot be adjusted at this screen; calibrating
conductivity is the only way to adjust the cell constant. Press Enter ( ) to reset the cell constant to 5.00 and display the Temperature Coefficient screen.
Note: Be certain to press Enter ( ) to reset the cell constant to 5.00. If MODE is pressed, the unit retains the previous cell constant and calibrates from a value that is already offset.
Temperature Coefficient
The unit uses the temperature coefficient to calculate temperature compensated conductivity. The default value is 1.91%. To change the temperature coefficient, use the and keys to adjust the value between 0 and 4.00%. Press Enter ( ) to save the new value, or press MODE to cancel the change and display the Temperature Reference screen.
Temperature Reference
The unit uses the temperature reference value to calculate temperature compensated conductivity. The default value is 25C. To change the temperature coefficient, use the and keys to adjust the value between 15 and 25C. Press Enter ( ) to save the new value, or press MODE to cancel the change and display the Conductivity Calibration screen.
Conductivity Calibration
1. Immerse the probe in a standard of known conductivity, preferably a standard in the
middle range of the solutions to be measured. Completely submerge the probe without touching the sides of the calibration container. Shake the probe lightly to remove any air bubbles trapped in the conductivity cell.
5
2. Allow temperature to stabilize. The message ‘rAng’ (range) may display briefly to indicate
unit auto-ranging; this is normal. After temperature stabilization, use the and keys to adjust the conductivity value to that of the conductivity standard at 25C. Press Enter ( ) to calibrate. The unit beeps twice to indicate a successful calibration, then automatically switches to normal operation mode.
CONDUCTIVITY MEASUREMENTS
1. Turn the unit on. Place the probe in the solution to be measured. Completely submerge
the probe. Shake the probe lightly to remove any trapped air bubbles in the conductivity cell.
2. Press MODE to enter the desired measurement mode. The message ‘rAng’ (range) may
appear briefly on the display indicate auto-ranging; this is normal. Allow temperature to stabilize before taking measurements.
SAVING, VIEWING AND DELETING DATA
The EC300A can save 50 data records. When in measurement mode, press to save a record. The instrument will confirm the saved data by displaying SAVE and the record number for one second. “Full” is displayed when trying to save data and the memory is full.
To view saved data, press mode until RECALL is displayed and then press . Use the Up or Down arrow keys to review different saved records. Press Mode to escape back to measurement mode.
To delete data records, press Mode while in measurement mode until DELETE is displayed. Press . “All” will be displayed and blinking. Press the Up or Down arrow key to switch between delete ‘All’ or ‘Each’ options. Select either ‘All’ or ‘Each’ by pressing while that option is displayed.
If ‘All’ is selected, all records will be deleted from memory and ‘None’ will be displayed. Press Mode twice to return to the measurement mode.
If ‘Each’ is selected, the Up and Down arrow keys will allow you to scroll through the saved data records. Press to delete the selected record. All records after the deleted record will shift up to keep the records in sequential order. For example, if record 3 is deleted, record 4 will become record 3 and record 5 will become record 4. Press Mode twice to return to the measurement mode.
PROBE MAINTENANCE
The most important requirement for accurate and reproducible conductivity measurements is a clean cell. A dirty cell changes the conductivity of a solution through contamination. Clean the cell thoroughly before storing it. To clean the conductivity cell:
1. Dip the cell in cleaning solution and agitate for two
to three minutes. Any foaming acid tile cleaner, such as Dow Chemical Bathroom Cleaner, should clean adequately. For a stronger cleaner, use a solution of 1:1 isopropyl alcohol and 1 N HCl. Remove the cell from the cleaning solution.
2. Use the nylon brush (supplied) to dislodge any
contaminants from inside the electrode chamber.
Repeat steps one and two until the cell is completely clean. Rinse the cell thoroughly in deionized, or clean tap water.
6
TROUBLESHOOTING
MAIN DISPLAY PROBLEM POSSIBLE SOLUTION
OvEr
OvEr/Undr during calibration
Conductivity is >200.0 mS Salinity is > 70.00 ppt
Cell Constant Calibration
is out of range
Completely submerge the
probe.
Allow sufficient time for the
electrode and Temp probe stabilization.
Recalibrate with correct value
for the conductivity standard.
Replace conductivity
standard.
Clean cell. Return for service.
MAIN
DISPLAY
OvEr/Undr
SECONDARY
DISPLAY
OvEr
Undr
Temperature >90.0 C
Temperature < -10.0 C
Decrease/Increase the
sample temperature.
Return for service.
7
SPECIFICATIONS
y
Display Range Accuracy Resolution
Conductivity, Auto-ranging
0.0 to 499.9 uS/cm 500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm
±1% of reading plus 2 uS/cm ±1% of reading plus 5 uS/cm
±1% of reading plus 0.05 uS/cm
±2.5% of reading plus 0.5 mS/cm
0.01 uS/cm 1 mS/cm
0.01 mS/cm
0.1 mS/cm
Salinity 0.0 to 70.0 ppt 0.2% Full Scale 0.1 ppt
Temperature
°C
-10.0 to 90 °C ±0.2 °C or ±0.4% Full Scale, whichever is greater
0.1 °C
Reference Temperature Temperature Coefficient 0.0% to 4.0% Cell Constant
TDS Constant Range 0.30 to 1.00 Power/Battery life
Calibration Back-up Yes Audio Feedback Yes, on all touch keys Instrument Case
Operating Temp. Range Operating Relative Humidity Range up to 95% Temperature Probe Dimensions (L x W x H) Weight (batteries included)
15.0 to 25.0
5.00
One 9V batter
Waterproof when connector cap installed, IP 67
0 to 50
Thermistor, 10k / 25
18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in) 270 grams (.6 lb)
°C
± 0.50
Approximately 500 hour
°C
°C
RECOMMENDED SPARE PARTS LIST
PART # DESCRIPTION
300-1 1-meter probe and cable assembly 300-4 4-meter probe and cable assembly.
300-10 10-meter probe and cable assembly.
606043 Carrying case, hard sided.
485 Carrying case, soft sided.
Item #606042REF
Revision A, July 2012
For the latest version of this manual, visit www.ysi.com
8
GARANTIE
L’appareil EcoSense® EC300A est garanti pour une période d’un an, à compter de la date d’achat par l’utilisateur final, contre tout défaut matériel et de fabrication. Les sondes et les câbles de l’ EC300A sont garantis pour une période d’un an, à compter de la date d’achat par l’utilisateur final, contre tout défaut matériel et de fabrication. Pendant la période de garantie, YSI s’engage à réparer ou à remplacer, gratuitement et à sa discrétion, tout produit qu’YSI peut établir comme étant couvert par la garantie.
Pour faire valoir cette garantie, écrivez ou appelez votre représentant YSI ou contactez le Service clientèle d’YSI à Yellow Springs, Ohio, États-Unis. Envoyez le produit et son justificatif d’achat en port payé au Centre de service homologué sélectionné par YSI. La réparation ou le remplacement seront effectués et le produit vous sera retourné en port payé. Les produits réparés ou remplacés sont garantis jusqu’à expiration de la période de garantie originale ou pour au moins 90 jours, à compter de la date de réparation ou de remplacement.
Limitation de garantie
Cette garantie ne s’applique pas aux produits YSI endommagés ou présentant des dysfonctionnements pour les raisons suivantes : (i) installation, exploitation ou utilisation du produit d’une façon non conforme aux instructions écrites d’YSI ; (ii) abus ou mésusage du produit ; (iii) manquement à l’entretien du produit conformément aux instructions écrites d’YSI ou aux procédures industrielles normales ; (iv) réparation non conforme du produit ; (v) utilisation par vous de pièces ou de composants défectueux ou non conformes lors de l’entretien ou de la réparation du produit, ou ; (vi) modification du produit d’une façon non expressément autorisée par YSI.
CETTE GARANTIE REMPLACE TOUTES LES AUTRES GARANTIES, EXPRESSES OU INDUITES, Y COMPRIS LES GARANTIES DE COMMERCIABILITÉ OU D’ADAPTATION À UN USAGE PARTICULIER. LA RESPONSABILITÉ D’YSI SELON LES TERMES DE CETTE GARANTIE SE LIMITE À LA RÉPARATION OU AU REMPLACEMENT DU PRODUIT, CONSTITUANT VOTRE SEUL ET UNIQUE RECOURS POUR TOUT PRODUIT DÉFECTUEUX COUVERT PAR CETTE GARANTIE. YSI NE POURRA EN AUCUN CAS ÊTRE TENU RESPONSABLE DE DOMMAGES SPÉCIAUX, INDIRECTS, ACCIDENTELS OU CONSÉCUTIFS RÉSULTANT DE L’UTILISATION DE TOUT PRODUIT DÉFECTUEUX COUVERT PAR CETTE GARANTIE.
COMMENT NOUS CONTACTER
YSI Inc. 1725 Brannum Lane Yellow Springs OH, 45387, États-Unis Tel : 800-897-4151 ; 937-767-7241 Télécopie : 937-767-1058 E-mail : environmental@ysi.com
Site Web : www.ysi.com
1
TABLE DES MATIÈRES
GARANTIE ............................................................................................................................ 1
COMMENT NOUS CONTACTER ...................................................................................... 1
INTRODUCTION GÉNÉRALE ............................................................................................. 3
INSPECTION INITIALE ........................................................................................................ 3
L’APPAREIL
INSTALLATION DE LA PILE ................................................................................................ 4
FONCTIONNALITÉS CLÉS DU MODÈLE EC300A .......................................................... 4
ÉCRAN À CRISTAUX LIQUIDES ......................................................................................... 4
MODES DE MESURE .......................................................................................................... 5
ÉTALONNAGE ..................................................................................................................... 5
MESURE DE LA CONDUCTIVITÉ ...................................................................................... 6
ENREGISTREMENT, AFFICHAGE ET SUPPRESSION DES DONNÉES .......................... 6
ENTRETIEN DE LA SONDE ................................................................................................ 7
DÉPANNAGE ....................................................................................................................... 8
SPÉCIFICATIONS ................................................................................................................ 9
LISTE DES PIÈCES DÉTACHÉES RECOMMANDÉES ...................................................... 9
.......................................................................................................................... 3
Mise Au Rebut De La Piles ........................................................................................ 4
TSD .............................................................................................................................. 5
CELL ............................................................................................................................. 5