The EcoSense® EC300 Instrument is warranted for one year from date of purchase by the end user
against defects in materials and workmanship. EC300 probes and cables are warranted for one year
from date of purchase by the end user against defects in material and workmanship. Within the
warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI
determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service
in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the
Authorized Service Center selected by YSI. Repair or replacement will be made and the product
returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the
original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by: (i) failure to install,
operate or use the product in accordance with YSI’s written instructions; (ii) abuse or misuse of the
product; (iii) failure to maintain the product in accordance with YSI’s written instructions or standard
industry procedure; (iv) any improper repairs to the product; (v) use by you of defective or improper
components or parts in servicing or repairing the product; or (vi) modification of the product in any way
not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR
REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE
REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT
SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS
WARRANTY.
CONTACT INFORMATION
YSI Inc. • 1725 Brannum Lane • Yellow Springs OH, 45387
800-897-4151 • 937-767-7241 • Fax: 937-767-1058
Email: ecosense@ysi.com
RECOMMENDED SPARE PARTS LIST ........................................................ 7
2
Page 4
GENERAL INTRODUCTION
The EC300 is one of three instruments in the EcoSense product line from YSI. The EC300 is a
precision tool that measures conductivity, salinity and temperature. A built-in microprocessor
calculates and compensates for all parameters related to conductivity and temperature
determinations.
This unit has a splash-resistant IP65 case. The mechanical touch keys are highly reliable with
tactile and audio feedback. This instrument uses one 9V battery. Re-calibration is not required
when power is restored.
The front of the instrument has a large LCD that displays temperature and either temperature
compensated or non-temperature compensated conductivity, salinity or TDS simultaneously
along with user prompts and mode indicators. The unit prompts users through calibration and
measurement procedures.
The model EC300 is available with a single four-electrode cell. Other features include automatic
conductivity ranging, automatic temperature compensation, long battery life, and 50/60 Hz AC
noise rejection. This meter is universal and user-friendly for field, industrial and laboratory
applications.
INITIAL INSPECTION
Carefully unpack the unit and accessories, and inspect for shipping damages. Compare
received parts with materials listed on the packing list. Notify YSI immediately of any damage or
missing parts. Save all packing materials until satisfactory operation is confirmed.
SPLASH RESISTANCE
Though the EC300 instrument is housed in a splash proof case, DO NOT use it underwater; the
connector is not waterproof. The splash-resistant case prevents permanent damage to the unit if
accidentally dropped into non-corrosive solutions. In case of submersion, follow these steps
immediately:
1. Dry the connector if necessary, and replace the conductivity probe. Rinse unit carefully with
distilled water. After rinsing and drying, inspect and clean connectors to remove all
contaminants that may affect probe connections.
2. Wait for the unit and probe to dry completely before resuming operation.
3. If the unit does not function correctly after steps 1 and 2, call YSI for possible repair or
replacement (see Warranty).
BATTERY INSTALLATION
An initial display of “BAT” on the LCD indicates approximately one hour of
battery life for unit operation within specifications. Replace battery when
“BAT” appears on the LCD. (See Figure 1.)
To replace battery, remove the two battery cover screws, battery cover
and o-ring. Replace the 9V battery. Replace battery cover and o-ring
(align the o-ring properly to insure a good seal) and fasten the two battery
cover screws for the splash-resistant feature.
3
Figure 1.
Battery Installation
Page 5
KEY FUNCTIONS OF THE MODEL EC300
1. : Turns the unit ON or OFF. Calibration values are not erased when the unit is turned off.
When the unit is not in use, turn it off to save battery life. For long-term storage, remove the
battery.
2. MODE: Selects display mode. In Normal operation, press MODE to sequentially display
uncompensated conductivity, temperature compensated conductivity, salinity and total
dissolved solids (TDS). In calibration mode, this key exits the current calibration and
displays the next calibration parameter.
3. CAL: In normal operation, changes the mode from Normal to Calibration.
4.
(Enter) : In Calibration Set-up, press this key to save the current parameter to memory.
5. Δ and ∇ Keys: Increases or decreases the display value as desired.
THE LCD DISPLAY
1. CONDUCTIVITY: Displays when measuring conductivity.
2. BAT: Low battery indicator.
3. Main display for compensated and
uncompensated conductivity, salinity
and TDS values.
8. °C: Flashes during temperature compensated conductivity measurement. During
calibration, indicates temperature reference unit.
%: Displays during calibration; indicates temperature coefficient unit.
9. ppt: Parts per thousand; indicates salinity measurement.
10. uS, mS: micro Siemens, milli Siemens; Indicates conductivity measurement.
11. g/L: grams/Liter; indicates TDS measurement.
12. °C: Temperature display.
2
3
4
Figure 1. LCD Display
51
6
7
8
9
10
11
12
MEASUREMENT MODES
1. Temperature - Current solution temperature continually displays.
2. Temperature Compensated Conductivity - Measurement of conductivity, compensated to
25°C or another specified value between 15 and 25°C. Expressed as uS/cm or mS/cm with
a flashing “°C”.
3. Uncompensated Conductivity – Direct measurement of conductivity, not compensated to
a specific temperature. Expressed as uS/cm or mS/cm.
4. Salinity – Measurement of salinity; expressed in parts per thousand (ppt).
5. TDS – Measurement of total dissolved solids (TDS); expressed in grams per liter (g/L)
Carefully observe the units displayed at the far side of the LCD to determine the desired mode.
4
Page 6
CALIBRATION
Calibration setup contains five sections: TDS, Cell, Temperature Coefficient, Temperature
reference, and Conductivity Calibration. To access these sections:
1. Connect the conductivity probe and cable assembly to the unit and turn the unit on. The
screen will display CELL and the cell constant of the conductivity probe.
2. Allow temperature readings to stabilize, then press CAL to enter the calibration mode; CAL
appears on the LCD. Press MODE to sequentially display the following sections:
Note: Press Enter (
advance to the next section. If there are no changes, the unit accepts the current value and
proceeds to the next section.
) to accept any values changes in each section and automatically
TDS
TDS is determined by multiplying conductivity (mS) by a TDS factor. The default factor value is
0.65. To change the TDS factor, use the Δ and ∇ keys to adjust the value between 0.30 and
1.00. Press Enter (
the CELL screen.
) to save the new value, or press MODE to cancel the change and display
CELL
The second screen will display CELL and the current cell value. The default cell value is 5.00
and is displayed in the lower right of the screen. The unit allows a variance of ±0.50 before
displaying an error message. The cell value cannot be adjusted at this screen; calibrating
conductivity is the only way to adjust the cell constant. Press Enter (
constant to 5.00 and display the Temperature Coefficient screen.
Note: Be certain to press Enter (
unit retains the previous cell constant and calibrates from a value that is already offset.
) to reset the cell constant to 5.00. If MODE is pressed, the
) to reset the cell
Temperature Coefficient
The unit uses the temperature coefficient to calculate temperature compensated conductivity.
The default value is 1.91%. To change the temperature coefficient, use the Δ and ∇ keys to
adjust the value between 0 and 4.00%. Press Enter (
MODE to cancel the change and display the Temperature Reference screen.
) to save the new value, or press
Temperature Reference
The unit uses the temperature reference value to calculate temperature compensated
conductivity. The default value is 25°C. To change the temperature coefficient, use the Δ and ∇
keys to adjust the value between 15 and 25°C. Press Enter (
MODE to cancel the change and display the Conductivity Calibration screen.
) to save the new value, or press
Conductivity Calibration
1. Immerse the probe in a standard of known conductivity, preferably a standard in the middle
range of the solutions to be measured. Completely submerge the probe without touching
the sides of the calibration container. Shake the probe lightly to remove any air bubbles
trapped in the conductivity cell.
2. Allow temperature to stabilize. The message ‘rAng’ (range) may display briefly to indicate
unit auto-ranging; this is normal. After temperature stabilization, use the Δ and ∇ keys to
adjust the conductivity value to that of the conductivity standard at 25°C. Press Enter (
to calibrate. The unit beeps twice to indicate a successful calibration, then automatically
switches to normal operation mode.
)
CONDUCTIVITY MEASUREMENTS
1. Turn the unit on. Place the probe in the solution to be measured. Completely submerge the
probe. Shake the probe lightly to remove any trapped air bubbles in the conductivity cell.
5
Page 7
2. Press MODE to enter the desired measurement mode. The message ‘rAng’ (range) may
appear briefly on the display indicate auto-ranging; this is normal. Allow temperature to
stabilize before taking measurements.
PROBE MAINTENANCE
The most important requirement for accurate and reproducible conductivity measurements is a
clean cell. A dirty cell changes the conductivity of a solution through contamination. Clean the
cell thoroughly before storing it. To clean the conductivity cell:
1. Dip the cell in cleaning solution and agitate for two to
three minutes. Any foaming acid tile cleaner, such as
Dow Chemical Bathroom Cleaner, should clean
adequately. For a stronger cleaner, use a solution of
1:1 isopropyl alcohol and 1 N HCl. Remove the cell
from the cleaning solution.
2. Use the nylon brush (supplied) to dislodge any
contaminants from inside the electrode chamber.
3. Repeat steps one and two until the cell is completely clean. Rinse the cell thoroughly in
deionized, or clean tap water.
TROUBLESHOOTING
MAIN DISPLAY PROBLEM POSSIBLE SOLUTION
OvEr
OvEr/Undr during calibration
• Conductivity is >200.0 mS
• Salinity is > 70.00 ppt
Cell Constant Calibration is
out of range
• Completely submerge the
probe.
• Allow sufficient time for the
electrode and Temp probe
stabilization.
• Recalibrate with correct value
for the conductivity standard.
• Replace conductivity standard.
• Clean cell.
• Return for service.
MAIN
DISPLAY
OvEr/Undr
SECONDARY
DISPLAY
OvEr
Undr
Temperature >90.0 °C
Temperature < -10.0 °C
6
• Decrease/Increase the sample
temperature.
• Return for service.
Page 8
SPECIFICATIONS
Display Range Accuracy Resolution
Conductivity,
Auto-ranging
0.0 to 499.9 uS/cm
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm
±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm
±1% of reading plus 0.05 uS/cm
±2.5% of reading plus 0.5 mS/cm
0.01 uS/cm
1 mS/cm
0.01 mS/cm
0.1 mS/cm
Salinity 0.0 to 70.0 ppt 0.2% Full Scale 0.1 ppt
Temperature °C -10.0 to 90 °C ±0.2 °C or ±0.4% Full Scale,
0.1 °C
whichever is greater
Reference Temperature
Temperature Coefficient
Cell Constant
TDS Constant Range
Power
Calibration Back-up
Audio Feedback
Water Resistance
Operating Temp. Range
Operating Relative Humidity Range
Temperature Probe
Dimensions (L x W x D)
Weight (batteries included)
15.0 to 25.0
0.0% to 4.0%
5.00
0.30 to 1.00
One 9V battery
Yes
Yes, on all touch keys
Splash-resistant, IP 65
0 to 50
up to 95%
Thermistor, 10kΩ / 25
186 mm x 70 mm x 37 mm (7.3 in x 2.8 in x 1.5 in)
430 grams (1 lb)
°C
± 0.50
°C
°C
RECOMMENDED SPARE PARTS LIST
PART # DESCRIPTION
300-4 4-meter probe and cable assembly.
300-10 10-meter probe and cable assembly.
380 Carrying case, hard sided.
480 Instrument carrying case with shoulder strap, soft sided.
For the most recent version of this manual, visit www.ysiecosense.com
7
Item #605369 • Drawing #A605369
Revision D • March 2008
Page 9
GARANTIE
L’appareil EcoSense® EC300 est garanti pour une période d’un an, à compter de la date d’achat par
l’utilisateur final, contre tout défaut matériel et de fabrication. Les sondes et les câbles de l’ EC300
sont garantis pour une période d’un an, à compter de la date d’achat par l’utilisateur final, contre tout
défaut matériel et de fabrication. Pendant la période de garantie, YSI s’engage à réparer ou à
remplacer, gratuitement et à sa discrétion, tout produit qu’YSI peut établir comme étant couvert par la
garantie.
Pour faire valoir cette garantie, écrivez ou appelez votre représentant YSI ou contactez le Service
clientèle d’YSI à Yellow Springs, Ohio, États-Unis. Envoyez le produit et son justificatif d’achat en port
payé au Centre de service homologué sélectionné par YSI. La réparation ou le remplacement seront
effectués et le produit vous sera retourné en port payé. Les produits réparés ou remplacés sont
garantis jusqu’à expiration de la période de garantie originale ou pour au moins 90 jours, à compter
de la date de réparation ou de remplacement.
Limitation de garantie
Cette garantie ne s’applique pas aux produits YSI endommagés ou présentant des
dysfonctionnements pour les raisons suivantes : (i) installation, exploitation ou utilisation du produit
d’une façon non conforme aux instructions écrites d’YSI ; (ii) abus ou mésusage du produit ; (iii)
manquement à l’entretien du produit conformément aux instructions écrites d’YSI ou aux procédures
industrielles normales ; (iv) réparation non conforme du produit ; (v) utilisation par vous de pièces ou
de composants défectueux ou non conformes lors de l’entretien ou de la réparation du produit, ou ;
(vi) modification du produit d’une façon non expressément autorisée par YSI.
CETTE GARANTIE REMPLACE TOUTES LES AUTRES GARANTIES, EXPRESSES OU
INDUITES, Y COMPRIS LES GARANTIES DE COMMERCIABILITÉ OU D’ADAPTATION À UN
USAGE PARTICULIER. LA RESPONSABILITÉ D’YSI SELON LES TERMES DE CETTE
GARANTIE SE LIMITE À LA RÉPARATION OU AU REMPLACEMENT DU PRODUIT,
CONSTITUANT VOTRE SEUL ET UNIQUE RECOURS POUR TOUT PRODUIT DÉFECTUEUX
COUVERT PAR CETTE GARANTIE. YSI NE POURRA EN AUCUN CAS ÊTRE TENU
RESPONSABLE DE DOMMAGES SPÉCIAUX, INDIRECTS, ACCIDENTELS OU
CONSÉCUTIFS RÉSULTANT DE L’UTILISATION DE TOUT PRODUIT DÉFECTUEUX
COUVERT PAR CETTE GARANTIE.
COMMENT NOUS CONTACTER
YSI Inc. • 1725 Brannum Lane • Yellow Springs OH, 45387, États-Unis
800-897-4151 • 937-767-7241 • Télécopie : 937-767-1058
E-mail : ecosense@ysi.com • Site Web : www.ysiecosense.com
LISTE DES PIÈCES DÉTACHÉES RECOMMANDÉES................................ 8
2
Page 11
INTRODUCTION GÉNÉRALE
Le EcoSense® EC300 est un des trois instruments de la ligne de produits EcoSense de YSI. Le
modèle EC300 est un outil de précision qui mesure la conductivité, la salinité et la température.
Un microprocesseur incorporé calcule et compense tous les paramètres relatifs à la
détermination de la conductivité et de la température.
L’appareil est doté d’un boîtier résistant aux éclaboussures conforme à la norme IP65. Les
touches mécaniques sont très fiables et produisent une réaction tactile et sonore. Cet appareil
utilise une pile de 9 V. Aucun réétalonnage n’est nécessaire lorsque l’alimentation électrique est
rétablie.
L’avant de l’instrument est muni d’un grand écran à cristaux liquides affichant la conductivité,
compensée ou non par la température, la salinité ou le TSD (total des solides dissous) et,
simultanément, des invites utilisateur et des indicateurs de mode. L’appareil émet des invites
destinées à l’utilisateur lors des procédures d’étalonnage et de mesure.
Le modèle EC300 est disponible avec une cellule unique à quatre électrodes. Parmi les autres
fonctionnalités, on notera le calcul automatique de la fourchette de conductivité, la
compensation automatique de la température, la longue durée de vie des piles et une
élimination du bruit de 50/60 Hz c. a. Cet appareil de mesure est convivial et particulièrement
souple dans les applications sur le terrain, industrielles et en laboratoire.
INSPECTION INITIALE
Déballez soigneusement l’appareil et les accessoires et vérifiez qu’ils n’ont pas été
endommagés lors de l’expédition. Comparez les pièces reçues aux matériaux répertoriés dans
le bordereau d’emballage. Notifiez immédiatement YSI s’il s’avère que des pièces sont
endommagées ou manquantes. Mettez de côté les matériaux d’emballage jusqu’à ce que le
fonctionnement correct de l’appareil soit confirmé.
PROTECTION CONTRE LES ÉCLABOUSSURES
Bien que le modèle EC300 soit abrité dans un boîtier résistant aux éclaboussures, ne l’utilisez
PAS sous l’eau, car son connecteur n’est pas étanche. Le boîtier résistant aux éclaboussures
prévient les dommages permanents si l’appareil est accidentellement immergé dans des solutions
non corrosives. En cas de submersion, prenez immédiatement les mesures suivantes :
1. Séchez le connecteur, le cas échéant, et remplacez la sonde de conductivité. Rincez
soigneusement l’appareil avec de l’eau distillée. Après le rinçage et le séchage, inspectez et
nettoyez les connecteurs en vue d’éliminer tout contaminant pouvant affecter les connexions de
la sonde.
2. Attendez que l’appareil et la sonde soient parfaitement secs avant de reprendre les
opérations.
3. Si l’appareil ne fonctionne pas correctement après les étapes 1 et 2, appelez YSI en vue
d’une réparation ou d’un remplacement éventuels (voir la Garantie).
3
Page 12
INSTALLATION DE LA PILE
Lorsque l’écran à cristaux liquides affiche pour la première fois « BAT », il
reste environ une heure de fonctionnement sur pile selon les
spécifications. Remplacez la pile lorsque l’indication « BAT » s’affiche sur
l’écran. (Vo ir Figure 1.)
Pour remplacer la pile, enlevez les deux vis du compartiment ainsi que le
couvercle et le joint torique. Remplacez la pile de 9 V. Replacez le
couvercle et le joint torique (alignez le joint correctement afin
d’assurer une bonne étanchéité) et revissez les deux vis du
compartiment pour conserver une bonne résistance aux
éclaboussures.
Figure 1.
Installation de la pile
FONCTIONNALITÉS CLÉS DU MODÈLE EC300
1. : Met l’appareil hors ou sous tension. Les valeurs d’étalonnage ne sont pas effacées
lorsque l’appareil est mis hors tension. Lorsque l’appareil n’est pas utilisé, mettez-le hors
tension pour économiser la pile. Enlevez la pile pour un entreposage prolongé.
2. MODE : Permet de sélectionner le mode d’affichage. En fonctionnement normal, appuyez
sur MODE pour afficher successivement la conductivité non compensée, la conductivité
compensée par la température, la salinité et le TSD (total des solides dissous). En mode
d’étalonnage, cette touche permet de quitter l’étalonnage actuel et d’afficher le paramètre
d’étalonnage suivant.
3. CAL : En fonctionnement normal, passe du mode Normal au mode Étalonnage
(Calibration).
4.
(Entrée) : Lors de la configuration de l’étalonnage, appuyez sur cette touche pour
enregistrer le paramètre actuel en mémoire.
5. Touches Δ et ∇ : Augmentent ou diminuent la valeur affichée, comme voulu.
ÉCRAN À CRISTAUX
LIQUIDES
1. CONDUCTIVITY: S’affiche lors de la mesure de la conductivité.
2. BAT: Indicateur de pile déchargée.
3. Affichage principal des valeurs de
conductivité compensée ou non, de
salinité et de TSD.
4. CAL: Indicateur de mode
d’étalonnage (Calibration).
5. TDS: S’affiche lors de la mesure du
total des solides dissous (TSD).
6. SALINITY : S’affiche lors de la
mesure de la salinité.
7. CELL: Indique la valeur constante de la cellule de conductivité.
8. °C : Clignote lors de la mesure de la conductivité compensée par la température. Lors de
l’étalonnage, indique l’unité de référence de la température.
% : S’affiche lors de l’étalonnage ; indique l’unité du coefficient de la température.
9. ppt : Parties par millier ; indique la mesure de salinité.
10. uS, mS : microsiemens, millisiemens ; indique la mesure de conductivité.
11. g/L : Grammes/litre ; indique la mesure du TDS.
2
3
4
Figure 2.Écran à cristaux liquides
51
6
7
8
9
10
11
12
4
Page 13
12. °C : Affichage de la température.
MODES DE MESURE
1. Température – La température actuelle de la solution s’affiche constamment.
2. Conductivité compensée par la température – Mesure de conductivité, compensée de
25 °C ou d’une autre valeur spécifiée entre 15 et 25 °C. Exprimée en uS/cm ou mS/cm avec
un « °C » clignotant.
3. Conductivité non compensée – Mesure directe de la conductivité, non compensée par
une température spécifique. Exprimée en uS/cm ou mS/cm.
4. Salinité – Mesure de la salinité ; exprimée en parties par millier (ppt).
5. TSD – Mesure du total des solides dissous (TSD) ; exprimée en grammes par litre (g/L).
Observez soigneusement les unités affichées à l’extrémité de l’écran à cristaux liquides pour
déterminer le mode voulu.
ÉTALONNAGE
La configuration de l’étalonnage comporte cinq sections : étalonnage du TSD, de la cellule, du
coefficient de la température, de la référence de la température et de la conductivité. Pour
accéder à ces sections :
1. Connectez la sonde de conductivité et l’assemblage du câble à l’appareil et mettez l’appareil
sous tension. L’écran affiche CELL et la constante de la cellule de la sonde de conductivité.
2. Laissez la lecture de température se stabiliser, puis appuyez sur CAL pour entrer en mode
d’étalonnage ; la mention CAL s’affiche sur l’écran à cristaux liquides. Appuyez sur MODE
pour afficher successivement les sections suivantes :
Remarque : appuyez sur Entrée (
des sections et avancer automatiquement à la section suivante. Si aucune modification n’est
apportée, l’appareil accepte la valeur actuelle et passe à la section suivante.
) pour accepter la modification des valeurs dans chacune
TSD
Le TSD est déterminé en multipliant la valeur de conductivité (mS) par un facteur de TSD. La
valeur du facteur par défaut est de 0,65. Pour modifier la valeur du facteur de TSD, utilisez les
touches Δ et ∇ pour définir la valeur entre 0,30 et 1,00. Appuyez sur Entrée (
enregistrer la nouvelle valeur ou sur MODE pour annuler la modification et afficher l’écran
CELL.
) pour
CELL
Le deuxième écran affiche CELL et la valeur actuelle de la cellule. La valeur par défaut de la
cellule, affichée dans la partie inférieure droite de l’écran, est de 5,00. L’appareil accepte une
variation de ±0,50 avant d’afficher un message d’erreur. La valeur de la cellule ne peut pas être
réglée dans cet écran ; le seul moyen de régler la constante de la cellule est d’effectuer un
étalonnage de la conductivité. Appuyez sur Entrée (
la valeur 5,00 et afficher l’écran Coefficient de la température.
Remarque : Veillez à bien appuyer sur la touche Entrée (
cellule sur la valeur par défaut 5,00. Si vous appuyez sur la touche MODE, l’appareil retient la
constante précédente de la cellule et effectue l’étalonnage d’après une valeur qui est déjà
erronée.
) pour rétablir la constante de la cellule à
) pour rétablir la constante de la
Coefficient de la température
L’appareil utilise le coefficient de la température pour calculer la conductivité compensée par la
température. La valeur par défaut est de 1,91 %. Pour modifier la valeur du coefficient de la
température, utilisez les touches Δ et ∇ pour définir la valeur entre 0 et 4,00 %. Appuyez sur
Entrée (
afficher l’écran Référence de la température.
) pour enregistrer la nouvelle valeur ou sur MODE pour annuler la modification et
5
Page 14
Référence de la température
L’appareil utilise la valeur de référence de la température pour calculer la conductivité
compensée par la température. La valeur par défaut est de 25 °C. Pour modifier la valeur de la
référence de la température, utilisez les touches Δ et ∇ pour définir la valeur entre 15 et 25 °C.
Appuyez sur Entrée (
modification et afficher l’écran Étalonnage de la conductivité.
) pour enregistrer la nouvelle valeur ou sur MODE pour annuler la
Étalonnage de la conductivité
1. Plongez la sonde dans un standard de conductivité connue, se trouvant de préférence au
milieu de la fourchette des solutions devant être mesurées. Submergez complètement la
sonde sans toucher les parois du conteneur d’étalonnage. Agitez doucement la sonde pour
enlever les bulles d’air piégées dans la cellule de conductivité.
2. Laissez la température se stabiliser. Le message « rAng » (fourchette) peut s’afficher
brièvement pour indiquer que l’appareil calcule automatiquement la fourchette. Cela est
normal. Une fois que la température est stabilisée, utilisez les touches Δ et ∇ pour régler la
valeur de conductivité sur celle du standard de conductivité à 25 °C. Appuyez sur Entrée
(
) pour étalonner. L’appareil émet deux signaux sonores pour indiquer que l’étalonnage
est réussi, puis passe automatiquement en mode de fonctionnement normal.
MESURE DE LA CONDUCTIVITÉ
1. Mettez l’appareil sous tension. Placez la sonde dans la solution devant être mesurée.
Submergez complètement la sonde. Agitez doucement la sonde pour enlever les bulles d’air
piégées dans la cellule de conductivité.
2. Appuyez sur MODE pour entrer dans le mode de mesure voulu. Le message « rAng »
(fourchette) peut s’afficher brièvement pour indiquer que l’appareil calcule automatiquement
la fourchette. Cela est normal. Laissez la température se stabiliser avant d’effectuer des
mesures.
6
Page 15
ENTRETIEN DE LA SONDE
La condition la plus importante pour obtenir des mesures de conductivité précises et
reproductibles est d’utiliser une cellule propre. Une cellule sale altère la conductivité d’une
solution par contamination. Nettoyez soigneusement la cellule avant de l’entreposer. Pour
nettoyer la cellule de conductivité :
1. Plongez la cellule dans une solution de nettoyage et
agitez pendant deux à trois minutes. Tout acide de
nettoyage moussant pour céramique, tel que le
produit de nettoyage pour salle de bain de Dow
Chemical, doit convenir à cette opération. Pour
obtenir un produit de nettoyage plus puissant, utilisez
une solution d’alcool isopropylique et de chlorure
d’hydrogène 1N à mélange égal. Retirez la cellule de
la solution de nettoyage.
2. Utilisez la brosse en nylon (fournie) pour déloger tout contaminant se trouvant à l’intérieur
de la chambre de l’électrode.
3. Répétez les étapes un et deux jusqu’à ce que la cellule soit complètement nettoyée. Rincez
soigneusement la cellule avec de l’eau désionisée ou de l’eau courante propre.
DÉPANNAGE
AFFICHAGE PRINCIPAL PROBLÈME SOLUTION POSSIBLE
OvEr
OvEr/Undr lors de
l’étalonnage
AFFICHAGE
PRINCIPAL
OvEr/Undr
AFFICHAGE
SECONDAIRE
OvEr
Undr
• Conductivité > 200,0 mS
• Salinité > 70,00 ppt (x10-3)
L’étalonnage de la constante
de la cellule est hors
limites
Température > 90,0 °C
Température < -10,0 °C
• Submergez complètement la
sonde.
• Attendez suffisamment
longtemps pour que l’électrode
et la sonde Temp se stabilisent.
• Effectuez un nouvel étalonnage
avec une valeur de standard de
conductivité correcte.
• Remplacez le standard de
conductivité.
• Nettoyez la cellule.
• Retournez l’appareil au centre
de service.
• Augmentez/Diminuez la
température de l’échantillon.
• Retournez l’appareil au centre
de service.
7
Page 16
SPÉCIFICATIONS
Affichage Fourchette Précision Résolution
Calcul
automatique de
la fourchette de
conductivité
0,0 à 499,9 uS/cm
500 à 4 999 uS/cm
5,00 à 49,99 mS/cm
50,0 à 200,0 mS/cm
±1 % de la lecture plus 2 uS/cm
±1 % de la lecture plus 5 uS/cm
±1 % de la lecture plus 0,05 uS/cm
±2,5% de la lecture plus 0,5 mS/cm
0,01 uS/cm
1 mS/cm
0,01 mS/cm
0,1 mS/cm
Salinité 0,0 à 70,0 x 10-3 0,2 % pleine échelle 0,1 x 10-3
Température °C -10,0 à 90 °C Le plus grand de ±0,2 °C ou
0,1 °C
±0,4 % pleine échelle
Référence de la température
Coefficient de la température
Constante de la cellule
Fourchette de constantes du TSD
Alimentation
Sauvegarde de l’étalonnage
Touches sonores
Étanchéité
Fourchette de températures de
fonctionnement
Limite d’humidité relative lors du
fonctionnement
Sonde de température
Dimensions (L x l x P)
Poids (avec pile)
15,0 à 25,0 °C
0,0 % à 4,0 %
5,00 ± 0,50
0,30 à 1,00
Une pile de 9 V
Oui
Oui, toutes les touches tactiles
Résistance aux éclaboussures, norme IP 65
0 à 50 °C
Jusqu’à 95 %
Thermistor, 10 kΩ / 25 °C
186 mm x 70 mm x 37 mm
430 g
LISTE DES PIÈCES DÉTACHÉES RECOMMANDÉES
N° RÉF. DESCRIPTION
300-4 Assemblage, câble de 4 mètres et sonde.
300-10 Assemblage, câble de 10 mètres et sonde.
380 Sacoche de transport, flancs durs.
480 Sacoche de transport de l’appareil avec courroie d’épaule, flancs souples.
Article n° 605369 • Dessin n° A605369
Pour obtenir la version la plus récente de ce manuel, visitez www.ysiecosense.com
8
Révision D • Mars 2008
Page 17
GARANTÍA
El medidor EcoSense® EC300 tiene un año de garantía contra defectos de materiales y fabricación,
contado a partir de la fecha de compra por el usuario final. Las sondas y cables del medidor EC300
tienen un año de garantía contra defectos de materiales y fabricación, contado a partir de la fecha de
compra por el usuario final. Durante el período de garantía, YSI reparará o reemplazará, según su
criterio, sin coste alguno, cualquier producto que YSI determine que está cubierto por esta garantía.
Para hacer valer esta garantía, escriba o llame al representante local de YSI, o comuníquese con el
Servicio de atención al cliente de YSI en Yellow Springs, Ohio, EE.UU. Envíe el producto y la factura
de compra, con el flete prepagado, al centro de servicio técnico autorizado seleccionado por YSI. Se
realizará la reparación necesaria o el reemplazo del producto y este será enviado de vuelta, con el
flete prepagado. Los productos reparados o reemplazados se garantizan durante el resto del período
de la garantía original, o al menos durante 90 días contados a partir de la fecha de reparación o
reemplazo.
Limitación de la garantía
Esta garantía no tendrá validez en caso de daños o fallos en el producto de YSI debido a lo siguiente:
(i) la instalación, funcionamiento o utilización del producto de manera contraria a las instrucciones
escritas suministradas por YSI; (ii) abuso o uso inadecuado del producto; (iii) falta de mantenimiento
del producto de acuerdo con las instrucciones escritas suministradas por YSI o con los
procedimientos estándar de la industria; (iv) cualquier reparación indebida realizada en el producto;
(v) utilización por parte del usuario de componentes o repuestos defectuosos o inadecuados para el
mantenimiento o reparación del producto; o (vi) cualquier modificación del producto no autorizada de
manera expresa por YSI.
ESTA GARANTÍA SE OTORGA EN LUGAR DE CUALQUIER OTRA GARANTÍA, EXPLÍCITA O
IMPLÍCITA, LO QUE INCLUYE TODA GARANTÍA DE COMERCIALIZACIÓN O IDONEIDAD
PARA UN PROPÓSITO ESPECÍFICO. DE CONFORMIDAD CON ESTA GARANTÍA, LA
RESPONSABILIDAD DE YSI SE LIMITA A LA REPARACIÓN O REEMPLAZO DEL
PRODUCTO, LO CUAL SERÁ LA SOLUCIÓN ÚNICA Y EXCLUSIVA QUE TENDRÁ EL
COMPRADOR POR CUALQUIER PRODUCTO DEFECTUOSO CUBIERTO POR ESTA
GARANTÍA. EN NINGÚN CASO YSI SERÁ RESPONSABLE POR NINGÚN DAÑO
CUANTIFICABLE, INDIRECTO, INCIDENTAL O CONSIGUIENTE QUE RESULTARA DE
ALGÚN PRODUCTO DEFECTUOSO CUBIERTO POR ESTA GARANTÍA.
INFORMACIÓN DE CONTACTO
YSI Inc. • 1725 Brannum Lane • Yellow Springs OH, 45387. EE.UU.
800-897-4151 • 937-767-7241 • Fax: 937-767-1058
Correo electrónico: ecosense@ysi.com • Página en Internet: www.ysiecosense.com
LISTA DE PIEZAS DE RECAMBIO RECOMENDADAS................................ 8
2
Page 19
INTRODUCCIÓN GENERAL
El modelo EC300 es uno de los tres medidores de la línea de productos EcoSense de YSI. El
EC300 es una herramienta de precisión que mide la conductividad, la salinidad y la
temperatura. Tiene un microprocesador integrado que calcula y realiza la compensación de
todos los parámetros relacionados con las determinaciones de la conductividad y de la
temperatura.
Esta unidad cuenta con un estuche IP65 resistente a salpicaduras. Las teclas de contacto
mecánico son muy confiables y al pulsarlas proporcionan una respuesta táctil y audible. Este
instrumento utiliza una pila de 9 voltios. No requiere nueva calibración cuando se restablece la
corriente.
La parte delantera del medidor tiene una pantalla grande de cristal líquido que muestra
simultáneamente la temperatura y la conductividad, salinidad o TDS con compensación de
temperatura o sin compensación de temperatura, junto con las indicaciones para el usuario y
los indicadores del modo de funcionamiento. La unidad orienta a los usuarios durante los
procedimientos de calibración y medición.
El modelo EC300 está disponible con una sola celda de cuatro electrodos. Entre otras
características se incluyen la calibración automática de la conductividad, compensación
automática de la temperatura, larga duración de la pila y rechazo de ruido de CA de 50/60 Hz.
Este medidor es universal y fácil de usar en aplicaciones in situ, industriales y de laboratorio.
INSPECCIÓN INICIAL
Saque la unidad de su embalaje con cuidado y verifique que no haya sufrido daños durante el
envío. Compare las piezas recibidas con los materiales enumerados en la lista de embalaje.
Avise inmediatamente a YSI en caso de que haya piezas faltantes o dañadas. Guarde todos los
materiales de embalaje hasta que confirme que la unidad funciona satisfactoriamente.
RESISTENCIA A LAS SALPICADURAS
Aunque el medidor EC300 está alojado en un estuche resistente a las salpicaduras, NO debe
usarse bajo el agua ya que el conector no es impermeable. Este tipo de estuche evita los daños
permanentes en la unidad en caso de que caiga accidentalmente en soluciones no corrosivas. En
caso de inmersión, siga estos pasos de inmediato:
1. Seque el conector, si es necesario, y cambie la sonda de medida de la conductividad.
Enjuague la unidad cuidadosamente con agua destilada. Después del enjuague y secado,
revise y limpie los conectores para eliminar cualquier contaminante que pueda afectar las
conexiones de la sonda.
2. Espere hasta que la unidad y la sonda se sequen por completo antes de reanudar el
funcionamiento.
3. Si la unidad no funciona correctamente después de realizar los pasos 1 y 2, comuníquese
con YSI para su posible reparación o reemplazo (consulte la garantía).
3
Page 20
INSTALACIÓN DE LA PILA
En la pantalla de cristal líquido aparecerá el mensaje de “BAT” (pila) para
indicar que queda aproximadamente una hora de carga de la pila para el
funcionamiento de la unidad según las especificaciones. Cambie la pila
cuando aparezca el mensaje “BAT” (pila) en la pantalla de cristal líquido.
(Vea la figura 1).
Para cambiar la pila, saque los dos tornillos de la tapa, luego retire la tapa
y el aro tórico. Coloque una nueva pila de 9 voltios. Vuelva a colocar la
tapa y el aro tórico (alinee este aro correctamente para garantizar un buen
sellado) y ajuste los dos tornillos de la tapa para que funcione la
protección contra salpicaduras.
FUNCIONES PRINCIPALES DEL MODELO EC300
1. : para encender y apagar la unidad. Los valores de calibración no se suprimen cuando
se apaga la unidad. Cuando la unidad no esté en uso, apáguela para ahorrar carga de la
pila. Para el almacenamiento a largo plazo debe quitar la pila.
2. MODE: selecciona el modo de la pantalla. Durante el funcionamiento normal, pulse de
manera secuencial MODE para ver la conductividad sin compensación, la conductividad,
salinidad y la cantidad de sólidos disueltos (TDS, por sus siglas en inglés) con
compensación de temperatura. En el modo de calibración, esta tecla sale de la calibración
actual y muestra el siguiente parámetro de calibración.
3. CAL: durante el funcionamiento normal, cambia el modo de Normal a Calibration
(Calibración).
4.
(Enter) : en el ajuste de la calibración, pulse esta tecla para guardar el parámetro actual
en la memoria.
5. Δ y ∇ teclas: para aumentar o disminuir el valor que aparece en la pantalla según lo desee.
PANTALLA DE CRISTAL
LÍQUIDO
1. CONDUCTIVITY: aparece al medir la conductividad.
2. BAT: indicador de pila agotada.
3. Pantalla principal para los valores
de conductividad, salinidad y TDS
compensados y no compensados.
4. CAL: indicador del modo de
calibración.
5. TDS: aparece al medir la cantidad
de sólidos disueltos.
6. SALINITY: aparece al medir la
salinidad.
7. CELL: indica el valor constante de la celda de conductividad.
8. °C: parpadea durante la medición de la conductividad con compensación de temperatura.
Durante la calibración, indica la unidad de referencia de la temperatura.
%: aparece durante la calibración; indica la unidad del coeficiente de temperatura.
9. ppt: partes por millar; indica la medición de la salinidad.
10. uS, mS: micro Siemens, milli Siemens; indica la medición de la conductividad.
11. g/L: gramos/Litro; indica la medición de TDS.
2
3
4
Figura 2. Pantalla de cristal líquido
Figura 1.
Instalación de la pila
51
6
7
8
9
10
11
12
4
Page 21
12. °C: indicador de la temperatura.
MODOS DE MEDICIÓN
1. Temperatura: la pantalla muestra constantemente la temperatura actual de la solución.
2. Conductividad con compensación de temperatura: medición de la conductividad,
compensada a 25°C o a otro valor especificado entre 15 y 25°C. Expresada en uS/cm o en
mS/cm con un “°C” parpadeante.
3. Conductividad no compensada: medición directa de la conductividad, sin compensar a
una temperatura específica. Expresada en uS/cm o en mS/cm.
4. Salinidad: medición de la salinidad; expresada en partes por millar (ppt).
5. TDS: medición de la cantidad de sólidos disueltos (TDS); expresada en gramos por litro
(g/L)
Observe detenidamente las unidades mostradas en el extremo de la pantalla de cristal líquido
para determinar el modo deseado.
CALIBRACIÓN
la configuración de calibración contiene cinco secciones: TDS, celda, coeficiente de
temperatura, referencia de la temperatura y calibración de la conductividad. Para acceder a
estas secciones:
1. Conecte el conjunto de sonda de medida de la conductividad y cables a la unidad y
enciéndala. La pantalla mostrará CELL y la constante de la celda de la sonda de medida
de la conductividad.
2. Deje que las lecturas de temperatura se estabilicen, luego pulse CAL para entrar en el
modo de calibración; CAL aparece en la pantalla de cristal líquido. Pulse MODE para ver
las siguientes secciones de manera secuencial:
Nota: Pulse Enter
automáticamente a la próxima sección. Si no hay cambios, la unidad acepta el valor actual
y procede a la próxima sección.
para aceptar cualquier cambio de valor en cada sección y pasar
TDS
La TDS se determina al multiplicar la conductividad (mS) por un factor de TDS. El valor
predeterminado del factor es 0,65. Para cambiar el factor de TDS, use las teclas Δ y ∇ para
ajustar el valor entre 0,30 y 1,00. Pulse Enter
para cancelar el cambio y mostrar la pantalla CELL.
para guardar el nuevo valor o pulse MODE
CELDA
La segunda pantalla mostrará CELL y el valor actual de la celda. El valor predeterminado de la
celda es 5,00 y se muestra en la parte inferior derecha de la pantalla. La unidad permite una
variación de ±0,50 antes de mostrar un mensaje de error. El valor de la celda no se puede
ajustar en esta pantalla; la única manera de ajustar la constante de la celda es calibrando la
conductividad. Pulse Enter
aparezca la pantalla Coeficiente de temperatura.
Nota: Asegúrese de pulsar Enter
pulsa MODE, la unidad conserva la constante anterior de la celda y calibra a partir de un valor
que ya está desfasado.
para reajustar la constante de la celda a 5,00 y para que
para restablecer la constante de la celda en 5,00. Si se
Coeficiente de temperatura
La unidad utiliza el coeficiente de temperatura para calcular la conductividad con compensación
de temperatura. El valor predeterminado es de 1,91%. Para cambiar el coeficiente de
temperatura, use las teclas Δ y ∇ para ajustar el valor entre 0 y 4,00%. Pulse Enter
guardar el nuevo valor o pulse MODE para cancelar el cambio y mostrar la pantalla Referencia de la temperatura.
5
para
Page 22
Referencia de la temperatura
La unidad utiliza el valor de referencia de la temperatura para calcular la conductividad con
compensación de temperatura. El valor predeterminado es 25°C. Para cambiar la referencia de
la temperatura, utilice las teclas Δ y ∇ para ajustar el valor entre 15 y 25°C. Pulse Enter
guardar el nuevo valor o pulse MODE para cancelar el cambio y mostrar la pantalla Calibración de la conductividad.
para
Calibración de la conductividad
1. Sumerja la sonda en una solución estándar de conductividad conocida, preferentemente
una en la escala intermedia de las soluciones que se van a medir. Sumerja por completo la
sonda sin tocar los lados del recipiente de calibración. Sacuda ligeramente la sonda para
eliminar cualquier burbuja de aire que esté atrapada en la celda de conductividad.
2. Deje que la temperatura se estabilice. Es posible que aparezca brevemente el mensaje
‘rAng’ (escala) para indicar la calibración automática de la unidad; esto es normal. Después
de la estabilización de la temperatura, utilice las teclas Δ y ∇ para ajustar el valor de
conductividad a aquél del estándar de conductividad a 25°C. Pulse Enter
La unidad emite dos tonos para indicar una calibración exitosa, luego cambia
automáticamente al modo de operación normal.
para calibrar.
MEDICIONES DE CONDUCTIVIDAD
1. Encienda la unidad. Coloque la sonda en la solución que se va a medir. Sumerja por
completo la sonda. Sacuda ligeramente la sonda para eliminar cualquier burbuja de aire
que esté atrapada en la celda de conductividad.
2. Pulse MODE para entrar en el modo de medición deseado. Es posible que aparezca
brevemente en la pantalla el mensaje ‘rAng’ (escala) para indicar la calibración automática
de la unidad; esto es normal. Deje que la temperatura se estabilice antes de tomar las
mediciones.
6
Page 23
MANTENIMIENTO DE LA SONDA
El requisito más importante para realizar mediciones de conductividad precisas y reproducibles
es que la celda esté limpia. Una celda sucia cambia la conductividad de una solución a través
de la contaminación. Limpie a fondo la celda antes de almacenarla. Para limpiar la celda de
conductividad:
1. Sumerja la celda en una solución de limpieza y agite
durante dos a tres minutos. Cualquier limpiador de
azulejos ácido y espumoso, como Dow Chemical
Bathroom Cleaner, sirve para limpiar la celda
adecuadamente. Para obtener un limpiador más
fuerte, utilice una solución 1:1 de alcohol isopropílico
y 1 N HCI. Saque la celda de la solución de limpieza.
2. Utilice el cepillo de nylon (incluido) para eliminar
cualquier contaminante del interior de la cámara de
electrodos.
3. Repita los pasos uno y dos hasta que la celda esté completamente limpia. Enjuague a
fondo la celda con agua desmineralizada o limpia de grifo.
LOCALIZACIÓN DE FALLOS
PANTALLA PRINCIPAL PROBLEMA POSIBLE SOLUCIÓN
OvEr
OvEr/Undr durante la
calibración
PANTALLA
PRINCIPAL
OvEr/Undr
PANTALLA
SECUNDARIA
OvEr
Undr
• La conductividad es
>200,0 mS
• La salinidad es > 70,00 ppt
La calibración de la
constante de la celda está
fuera de la escala
La temperatura es
>90,0 °C
o la temperatura es
< -10,0 °C
• Sumerja por completo la sonda.
• Deje transcurrir suficiente
tiempo para que se estabilicen
el electrodo y la sonda Temp.
• Recalibrar con el valor correcto
para el estándar de
conductividad.
• Cambie el estándar de
conductividad.
• Limpie la celda.
• Envíelo al servicio técnico.
• Disminuya/aumente la
temperatura de la muestra.
• Envíelo al servicio técnico.
7
Page 24
ESPECIFICACIONES
Pantalla Escala Precisión Resolución
Conductividad,
calibración
automática
0,0 a 499,9 uS/cm
500 a 4999 uS/cm
5,00 a 49,99 mS/cm
50,0 a 200,0 mS/cm
±1% de lectura más 2 uS/cm
±1% de lectura más 5 uS/cm
±1% de lectura más 0,05 uS/cm
±2,5% de lectura más 0,5 mS/cm
0,01 uS/cm
1 mS/cm
0,01 mS/cm
0,1 mS/cm
Salinidad 0,0 a 70,0 ppt 0,2% de la escala completa 0,1 ppt
Temperatura en
°C
-10,0 a 90 °C ±0,2 °C o ±0,4% de la escala
completa, lo que sea mayor
0,1° C
Temperatura de referencia
Coeficiente de temperatura
Constante de la celda
Escala de la constante TDS
Energía
Respaldo de la calibración
Respuesta audible
Resistencia al agua
Escala de temperatura de
funcionamiento
Escala de humedad relativa de
funcionamiento
Sonda de medida de la temperatura
Dimensiones (lar x anc x prof)
Peso (con la pila)
15,0 a 25,0 °C
0,0% a 4,0%
5,00 ± 0,50
0,30 a 1,00
Una pila de 9 voltios
Sí
Sí, en todas las teclas
IP 65, resistente a salpicaduras
0 a 50
Hasta 95%
Termistor, 10kΩ / 25
186 mm x 70 mm x 37 mm (7,3 pulg. X 2,8 pulg. X 1,5 pulg.)
430 gramos (1 libra)
° C
° C
LISTA DE PIEZAS DE RECAMBIO RECOMENDADAS
PIEZA N
o
DESCRIPCIÓN
300-4 Conjunto de sonda de 4 metros y cables.
300-10 Conjunto de sonda de 10 metros y cables.
380 Estuche portátil, de lados rígidos.
480 Estuche portátil para medidor con correa para el hombro, no rígido.
Artículo #605369 • Ilustración #A605369
Para la versión más reciente de este manual, visite www.ysiecosense.com
8
Revisión D • Marzo de 2008
Page 25
GARANTIE
Für das Instrument EcoSense® EC300 wird für den Zeitraum von einem (1) Jahr ab dem Datum des
Kaufs durch den Endbenutzer eine Garantie bezüglich Material- und Fabrikationsfehlern gewährt. Für
EC300-Sonden und -Kabel wird für den Zeitraum von einem (1) Jahr ab dem Datum des Kaufs durch
den Endbenutzer eine Garantie bezüglich Material- und Fabrikationsfehlern gewährt. Innerhalb der
Garantiefrist wird YSI jedes Produkt, das nach Ansicht von YSI unter diese Garantie fällt, nach
eigenem Ermessen entweder kostenlos reparieren oder ersetzen.
Setzen Sie sich schriftlich oder telefonisch mit Ihrem örtlichen YSI-Vertreter in Verbindung oder
kontaktieren Sie den YSI-Kundendienst in Yellow Springs (Ohio/USA), um diese Garantie in Anspruch
zu nehmen. Schicken Sie das Produkt mit dem Kaufnachweis unter Vorauszahlung der Frachtkosten
an das von YSI ausgewählte Vertragskundendienstzentrum. Dort wird die Reparatur oder der Ersatz
vorgenommen und das Produkt anschließend frachtfrei an Sie zurückgesandt. Für reparierte oder
ersetzte Produkte erstreckt sich die Garantie über die Restlaufzeit des ursprünglichen
Garantiezeitraums, mindestens jedoch über 90 Tage ab dem Datum der Reparatur oder des Ersatzes.
Garantiebegrenzung
Diese Garantie gilt nicht für irgendwelche Beschädigungen oder Ausfälle eines YSI-Produkts, die
durch Folgendes verursacht werden: (i) Versäumnis, das Produkt in Übereinstimmung mit den
schriftlichen Instruktionen von YSI zu installieren, in Betrieb zu nehmen oder zu benutzen; (ii)
Missbrauch oder Zweckentfremdung des Produkts; (iii) Versäumnis, das Produkt in Übereinstimmung
mit den schriftlichen Instruktionen von YSI oder einem branchenüblichen Verfahren in Stand zu
halten; (iv) unsachgemäße Reparaturen am Produkt; (v) eine von Ihnen vorgenommene Verwendung
defekter oder ungeeigneter Komponenten oder Teile bei der Wartung oder Reparatur des Produkts;
oder (vi) die Abänderung des Produkts in irgendeiner Weise, die nicht ausdrücklich von YSI
genehmigt ist.
DIESE GARANTIE TRITT AN DIE STELLE ALLER ANDEREN GEWÄHRLEISTUNGEN,
GLEICHGÜLTIG, OB VERTRAGLICH ODER GESETZLICH ZUGESICHERT,
EINSCHLIESSLICH JEGLICHER ZUSICHERUNG ALLGEMEINER
GEBRAUCHSTAUGLICHKEIT ODER EIGNUNG FÜR EINEN BESTIMMTEN ZWECK. DIE
HAFTUNG VON YSI GEMÄSS DIESER GARANTIE BESCHRÄNKT SICH AUF DIE
REPARATUR ODER DEN ERSATZ DES PRODUKTS, UND DIES STELLT IHREN EINZIGEN
UND AUSSCHLIESSLICHEN RECHTSBEHELF FÜR JEDES DEFEKTE PRODUKT DAR, FÜR
DAS DIESE GARANTIE GILT. YSI HAFTET AUF KEINEN FALL FÜR IRGENDWELCHE
BESONDEREN, MITTELBAREN, BEILÄUFIG ENTSTANDENEN ODER FOLGESCHÄDEN, DIE
AUS IRGENDEINEM DEFEKTEN PRODUKT RESULTIEREN, FÜR DAS DIESE GARANTIE
GILT.
KONTAKTINFORMATION
YSI Inc. • 1725 Brannum Lane • Yellow Springs, OH 45387, USA
800-897-4151 (gebührenfrei in den USA) • +(937)-767-7241 • Fax: +(937)-767-1058
E-Mail: ecosense@ysi.com • Website: www.ysiecosense.com
LISTE EMPFOHLENER ERSATZTEILE......................................................... 8
2
Page 27
ALLGEMEINE EINLEITUNG
Das Modell EC300 ist eines von drei Instrumenten aus der EcoSense-Produktgruppe von YSI.
Beim EC300 handelt es sich um ein Präzisionsgerät, das Leitfähigkeit, Salzgehalt und
Temperatur misst. Ein eingebauter Mikroprozessor berechnet und kompensiert für alle
Parameter, die mit der Bestimmung der Leitfähigkeit und Temperatur zusammenhängen.
Dieses Gerät besitzt ein spritzwasserfestes Gehäuse gemäß der Schutzklasse IP65. Die
mechanischen Berührungstasten mit fühlbarem und hörbarem Feedback sind äußerst
zuverlässig. Dieses Messinstrument wird von einer 9 Volt-Batterie gespeist. Es ist keine erneute
Kalibrierung notwendig, wenn die Stromversorgung wiederhergestellt wird.
Die Vorderseite des Instruments beherbergt ein großes LCD-Feld, das gleichzeitig mit
Eingabeaufforderungen für den Benutzer und Modus-Anzeigen die Temperatur und entweder
die für die Temperatur kompensierte bzw. nicht kompensierte Leitfähigkeit, die Salzhaltigkeit.
oder den Gesamtgehalt an gelösten Stoffen anzeigt. Das Gerät führt den Benutzer mit
Eingabeaufforderungen durch die Kalibrierungs- und Messverfahren.
Das Modell EC300 wird mit einer einzelnen vierpoligen Messzelle geliefert. Andere Merkmale
umfassen die automatische Leitfähigkeitsmessung, die automatische Temperaturkompensation,
eine lange Batterielebensdauer und eine 50/60 Hz AC Rauschunterdrückung. Das Instrument ist
vielseitig und benutzerfreundlich für Anwendungen in der Feldforschung, Industrie und im
Labor.
INSPEKTION VOR DEM EINSATZ
Packen Sie das Gerät und sein Zubehör sorgfältig aus und untersuchen Sie alle Teile auf
eventuelle Transportschäden. Vergleichen Sie die erhaltenen Teile mit den Gegenständen, die
auf dem Packzettel aufgeführt sind. Benachrichtigen Sie YSI unverzüglich über irgendwelche
Schäden oder fehlende Teile. Bewahren Sie das Verpackungsmaterial auf, bis der einwandfreie
Betrieb des Geräts bestätigt ist.
SPRITZWASSERSCHUTZ
Das EC300 Instrument ist in einem spritzwassergeschützten Gehäuse untergebracht. Verwenden
Sie es NICHT unter Wasser; das Anschlussteil ist nicht wasserdicht. Das spritzwassergeschützte
Gehäuse verhindert permanente Schäden am Gerät, falls es versehentlich in nichtätzende
Lösungen fällt. Falls das Gerät in Flüssigkeit eingetaucht wurde, sind sofort folgende Schritte
durchzuführen:
1. Falls notwendig, das Anschlussteil trocknen und die Leitfähigkeitssonde auswechseln. Spülen
Sie das Gerät sorgfältig mit destilliertem Wasser ab. Nach dem Abspülen und Trocknen sollten
Sie die Anschlussstücke inspizieren und säubern, um alle verunreinigenden Substanzen zu
entfernen, die die Sondenanschlüsse angreifen könnten.
2. Warten Sie, bis das Gerät und die Sonde völlig trocken sind, bevor Sie das Gerät wieder in
Betrieb nehmen.
3. Falls das Gerät nach den Schritten 1 und 2 nicht korrekt funktionieren sollte, rufen Sie YSI
wegen einer möglichen Reparatur oder eines Ersatzes an (siehe Garantie).
3
Page 28
EINSETZEN DER BATTERIE
Eine anfängliche Anzeige von „BAT“ auf dem LCD bedeutet, dass das
Gerät noch ca. 1 Stunde innerhalb der Spezifizierungen mit der Batterie
betrieben werden kann. Wechseln Sie die Batterie aus, wenn „BAT“ auf
dem LCD-Anzeigefeld erscheint. (Siehe Abbildung 1)
Zum Auswechseln der Batterie die beiden Schrauben für die
Batterieabdeckung, die Batterieabdeckung und den O-Ring entfernen.
Wechseln Sie die 9 Volt-Batterie aus. Setzen Sie die Batterieabdeckung
und den O-Ring wieder ein (achten Sie auf die korrekte Ausrichtung
des O-Rings für eine gute Abdichtung) und ziehen Sie die beiden
Schrauben für die Batterieabdeckung fest, damit das Gerät
spritzwassergeschützt bleibt.
Abbildung 1.
Einsetzen der Batterie
HAUPTFUNKTIONEN DES MODELLS EC300
1. : Schaltet das Gerät EIN oder AUS. Die Kalibrierungswerte werden nicht gelöscht, wenn
das Gerät ausgeschaltet wird. Wenn das Gerät nicht benutzt wird, schalten Sie es aus, um
die Batterie zu schonen. Entfernen Sie die Batterie, wenn das Instrument längere Zeit
gelagert werden soll.
2. MODE: Wählt den Anzeigemodus. Im normalen Betrieb MODE drücken, um nacheinander
die unkompensierte Leitfähigkeit, die für Temperatur kompensierte Leitfähigkeit, den
Salzgehalt und die Gesamtmenge der gelösten Stoffe (TDS) anzuzeigen. Im Kalibriermodus
dient diese Taste zum Verlassen der derzeitigen Kalibrierung und zum Anzeigen des
nächsten Kalibrierparameters.
3. CAL: Ändert im Normalbetrieb den Modus von Normal auf Kalibration.
4.
(Eingabe) : Bei der Kalibriereinstellung diese Taste drücken, um den aktuellen
Parameter im Speicher zu speichern.
5. Δ und ∇Tasten:Erhöhen oder verringern, wie gewünscht, den Anzeigewert.
DAS LCD-ANZEIGEFELD
1. CONDUCTIVITY: Wird beim Messen der Leitfähigkeit angezeigt.
2. BAT: Anzeige für niedrigen
Batteriestand.
3. Hauptanzeige für kompensierte und
unkompensierte Leitfähigkeit,
Salzgehalt und Werte für den
Gesamtgehalt an gelösten Stoffen
(TDS).
4. CAL: Anzeige des Kalibrier-Modus.
5. TDS: Wird beim Messen des
Gesamtgehalts an gelösten Stoffen
angezeigt.
6. SALINITY: Wird beim Messen des Salzgehalts angezeigt.
7. CELL: Zeigt den konstanten Wert des Messgefäßes für die Leitfähigkeit.
8. °C: Blinkt beim Messen der für Temperatur kompensierten Leitfähigkeit. Zeigt während des
Kalibrierens die Temperaturreferenzeinheit.
%: Wird während des Kalibrierens angezeigt; zeigt die Einheit des Temperaturkoeffizienten.
9. ppt: Teile pro Tausend; zeigt Messung des Salzgehalts an.
10. ìS, mS: Mikro-Siemens, Milli-Siemens; zeigt Messung der Leitfähigkeit an.
11. g/L: Gramm/Liter; zeigt Messung des Gesamtgehalts an gelösten Stoffen an.
1
2
3
4
Abbildung 2. LCD-Anzeigefeld
4
5
6
7
8
9
10
11
12
Page 29
12. °C: Temperaturanzeige.
MESSUNGS-MODI
1. Temperatur – Die Temperatur der aktuellen Lösung wird kontinuierlich angezeigt.
2. Für Temperatur kompensierte Leitfähigkeit – Messung der Leitfähigkeit, kompensiert auf
25 °C oder einen anderen angegebenen Wert zwischen 15 und 25 °C. Wird in µS/cm oder
mS/cm mit einem blinkenden „°C“ dargestellt.
3. Unkompensierte Leitfähigkeit – Direkte Messung der Leitfähigkeit, nicht für eine spezielle
Temperatur kompensiert. Wird in µS/cm oder mS/cm ausgedrückt.
4. Salzgehalt – Messung des Salzgehalts; ausgedrückt in Teile pro Tausend (ppt).
5. TDS – Messung des Gesamtgehalts an gelösten Stoffen; ausgedrückt in Gramm pro Liter
(g/L)
Beobachten Sie aufmerksam die an der ganz rechten Seite des LCD-Anzeigefelds angezeigten
Einheiten, um den gewünschten Modus zu bestimmen.
KALIBRIERUNG
Das Einrichten der Kalibrierung umfasst fünf Abschnitte: Kalibrieren von Gesamtzahl der
gelösten Teile (TDS), Messgefäß (Cell), Temperaturkoeffizient (Temperature Coefficient),
Temperaturreferenz (Temperature reference) und Leitfähigkeit (Conductivity). Für Zugriff auf
diese Abschnitte:
1. Die Baugruppe mit Leitfähigkeitssonde und Kabel am Gerät befestigen und das Gerät
einschalten. Der Bildschirm zeigt CELL und die Messgefäßkonstante der
Leitfähigkeitssonde an.
2. Warten Sie, bis sich die Temperaturanzeigen stabilisiert haben und drücken Sie dann CAL,
um den Kalibriermodus einzugeben; auf der LCD-Anzeige erscheint CAL. Drücken Sie
MODE, um nacheinander folgende Abschnitte anzuzeigen:
Hinweis: Drücken Sie Eingabe (
einzugeben und automatisch zum nächsten Abschnitt weiterzugehen. Wenn keine
Änderungen vorhanden sind, akzeptiert das Gerät den aktuellen Wert und geht zum
nächsten Abschnitt weiter.
), um alle Änderungen der Werte in diesem Abschnitt
GESAMTGEHALT AN GELÖSTEN STOFFEN (TDS)
Der Gesamtgehalt an gelösten Stoffen wird durch Multiplizieren der Leitfähigkeit (mS) mit einem
TDS-Faktor bestimmt. Der Vorgabewert dieses Faktors ist 0,65. Zum Ändern des TDS-Faktors
die Tasten Δ und ∇ verwenden, um den Wert zwischen 0,30 und 1,00 einzustellen. Eingabe
(
) drücken, um den neuen Wert zu speichern, oder MODE drücken, um die Änderung
aufzuheben und den Bildschirm CELL anzuzeigen.
MESSGEFÄSS (CELL)
Der zweite Bildschirm zeigt CELL und den aktuellen Messgefäßwert. Der Vorgabewert für das
Messgefäß ist 5,00; er wird an der rechten unteren Seite des Bildschirms angezeigt. Das Gerät
lässt eine Abweichung von ±0,50 zu, bevor es eine Fehlermeldung anzeigt. Der Messgefäßwert
kann nicht auf diesem Bildschirm eingestellt werden; das Kalibrieren der Leitfähigkeit ist die
einzige Möglichkeit zum Justieren der Messgefäßkonstante. Eingabe (
Messgefäßkonstante auf 5,00 zurückzustellen und den Bildschirm Temperaturekoeffizient
anzuzeigen.
Hinweis: Drücken Sie auf jeden Fall „Eingabe“ (
zurückzustellen. Wenn MODE gedrückt wird, bewahrt das Gerät die vorherige
Messgefäßkonstante und kalibriert von einem bereits ausgeglichenen Wert.
), um die Messgefäßkonstante auf 5,00
) drücken, um die
5
Page 30
Temperaturkoeffizient
Das Gerät verwendet den Temperaturkoeffizienten, um die für die Temperatur kompensierte
Leitfähigkeit zu berechnen. Der Vorgabewert beträgt 1,91%. Zum Ändern des
Temperaturkoeffizienten die Tasten Δ und ∇ verwenden, um den Wert zwischen 0 und 4,00%
einzustellen. Eingabe (
die Änderung aufzuheben und den Bildschirm Temperaturreferenz anzuzeigen.
) drücken, um den neuen Wert zu speichern, oder MODE drücken, um
Temperaturreferenz
Das Gerät verwendet den Temperaturreferenzwert zum Berechnen der für die Temperatur
kompensierten Leitfähigkeit. Der Vorgabewert beträgt 25 °C. Zum Ändern
derTemperaturreferenz die Tasten Δ und ∇ verwenden, um den Wert zwischen 15 und 25 °C
einzustellen. Eingabe (
die Änderung aufzuheben und den Bildschirm Kalibrieren der Leitfähigkeit anzuzeigen.
) drücken, um den neuen Wert zu speichern, oder MODE drücken, um
Kalibrieren der Leitfähigkeit
1. Die Sonde in einen Standard bekannter Leitfähigkeit eintauchen, bevorzugt in einen
Standard im mittleren Bereich der zu messenden Lösungen. Die Sonde vollständig
eintauchen, ohne dabei die Seiten des Kalibrierbehälters zu berühren. Die Sonde leicht
schütteln, um eventuell im Leitfähigkeitsmessbehälter eingeschlossene Luftblasen zu
entfernen.
2. Warten Sie, bis sich die Temperatur stabilisiert hat. Die Meldung „rAng“ (Bereich) wird
eventuell kurz angezeigt, um anzuzeigen, dass das Gerät automatisch den Bereich
feststellt; das ist normal. Nachdem sich die Temperatur stabilisiert hat, die Tasten Δ und ∇
verwenden, um den Wert der Leitfähigkeit auf den der Standardleitfähigkeit von 25 °C
einzustellen. Eingabe (
erfolgreiche Kalibrierung anzuzeigen, und schaltet dann automatisch auf den
Normalmodus um.
) drücken, um zu kalibrieren. Das Gerät piept zweimal, um eine
MESSUNGEN DER LEITFÄHIGKEIT
1. Schalten Sie das Gerät ein. Tauchen Sie die Sonde in die zu messende Lösung. Die Sonde
vollständig eintauchen. Schütteln Sie die Sonde leicht, um eventuell im Messbehälter für
die Leitfähigkeit eingeschlossene Luftblasen zu entfernen.
2.MODE drücken, um den gewünschten Mess-Modus einzugeben. Die Meldung „rAng“
(Bereich) kann kurz auf der Anzeige erscheinen, um anzuzeigen, dass das Gerät
automatisch den Messbereich feststellt; das ist normal. Warten Sie, bis sich die Temperatur
stabilisiert hat, bevor Sie Messungen durchführen.
6
Page 31
WARTUNG DER SONDE
Die wichtigste Voraussetzung für präzise und wiederholbare Leitfähigkeitsmessungen ist ein
sauberer Messbehälter. Ein verschmutzter Behälter verändert die Leitfähigkeit einer Lösung
durch Kontaminierung. Reinigen Sie den Behälter vor dem Lagern gründlich. Zum Reinigen des
Messbehälters für Leitfähigkeit:
1. Den Behälter in Reinigungslösung tauchen und zwei
oder drei Minuten lang kräftig schütteln. Ein
schäumender säurehaltiger Fliesenreiniger, wie
beispielsweise ein chemischer Badezimmerreiniger
(Dow Chemical Bathroom Cleaner), sollte für
ausreichende Reinigung sorgen. Als stärkeres
Reinigungsmittel eine 1:1 Lösung aus
Isopropylalkohol und 1 N HCl verwenden. Den
Behälter aus der Reinigungslösung entfernen.
2. Die (mitgelieferte) Nylonbürste zum Lösen von Verschmutzungen an der Innenseite der
Elektrodenkammer verwenden.
3. Wiederholen Sie die Schritte 1 und 2, bis der Behälter vollkommen sauber ist. Spülen Sie
den Behälter gründlich mit entionisiertem oder sauberem Leitungswasser.
AUFFINDEN UND BESEITIGEN VON STÖRUNGEN
HAUPTANZEIGEPROBLEMMÖGLICHELÖSUNG
OvEr
OvEr/Undr während des
Kalibrierens
HAUPT-
ANZEIGE
OvEr/Undr
NEBEN-
ANZEIGE
OvEr
Undr
• Leitfähigkeit ist >200,0 mS
• Salzgehalt ist > 70,00 ppt
Kalibrierung der
Behälterkonstanten liegt
außerhalb des
Messbereichs
Temperatur >90,0 °C
Temperatur < -10,0 °C
• Die Sonde vollständig
eintauchen.
• Geben Sie der Elektrode und
der Temp.-Sonde genügend
Zeit, sich zu stabilisieren.
• Mit dem korrekten Wert für den
Leitfähigkeitsstandard eine
erneute Kalibrierung
durchführen.
• Den Leitfähigkeitsstandard
austauschen.
• Den Behälter reinigen.
• Schicken Sie das Gerät zur
Inspektion zurück.
• Die Temperatur der Probe
verringern/erhöhen.
• Schicken Sie das Gerät zur
Inspektion zurück.
7
Page 32
TECHNISCHE DATEN
Anzeige Bereich Genauigkeit Auflösung
Leitfähigkeit,
automatische
Messbereichsu
mschaltung
Salzgehalt 0,0 bis 70,0 ppt 0,2% des Skalenendwerts 0,1 ppt
Temperatur °C -10,0 bis 90 °C Der jeweils größere Wert von ±0,2
0,0 bis 499,9 ìS/cm
500 bis 4999 ìS/cm
5,00 bis 49,99 mS/cm
50,0 bis 200,0 mS/cm
±1% des Ablesewerts plus 2 ìS/cm
±1% des Ablesewerts plus 5 ìS/cm
±1% des Ablesewerts plus 0,05
±2,5% des Ablesewerts plus 0,5
°C oder ±0,4% des Skalenendwerts
ìS/cm
mS/cm
0,01 ìS/cm
1 mS/cm
0,01 mS/cm
0,1 mS/cm
Referenztemperatur
Temperaturkoeffizient
Behälterkonstante
Bereich der TDS-Konstante
Stromquelle
Sicherung der Kalibrierungsdaten
Audio-Feedback
Wasserfestigkeit
Betriebstemp.- Bereich
Relative Luftfeuchtigkeit
Betriebsbereich
Temperatursonde
Abmessungen (L x B x T)
Gewicht (mit Batterie)
15,0 bis 25,0 °C
0,0% bis 4,0%
5,00 ± 0,50
0,30 bis 1,00
Eine 9 Volt-Batterie
Ja
Ja, für alle Tasten
Spritzwassergeschützt, IP 65
0 bis 50 °C
bis zu 95%
Thermistor, 10 kΩ / 25 °C
186 mm x 70 mm x 37 mm (7,3 Zoll x 2,8 Zoll x 1,5 Zoll)
430 Gramm (1 lb)
0,1 °C
LISTE EMPFOHLENER ERSATZTEILE
Artikel-
BESCHREIBUNG
nummer
300-4 4-Meter Sonden- und Kabelbaugruppe.
300-10 10-Meter Sonden- und Kabelbaugruppe.
380 Tragebehälter, Hartschale.
480 Instrumententragetasche mit Schultergurt, weich.
Die aktuellste Version dieses Handbuchs finden Sie unter www.ysiecosense.com
8
Artikel-Nr. 605369 • Zeichnung-Nr. A605369
Revision D • März 2008
Page 33
GARANZIA
Il misuratore EcoSense® EC300 è coperto da garanzia per un anno da difetti di lavorazione e dei
materiali a partire dalla data di acquisto da parte dell’utente finale. Le sonde e I cavi di EC300 sono
coperti da garanzia per un anno da difetti di lavorazione e dei materiali a partire dalla data di acquisto
acquisto da parte dell’utente finale. Durante il periodo di validità della garanzia, YSI si assume la
responsabilità di riparare o sostituire, a sua discrezione, gratuitamente, qualsiasi prodotto che YSI
ritenga coperto da garanzia.
Per esercitare il diritto alla garanzia, scrivere o contattare il rappresentante YSI di zona, oppure il
Servizio Assistenza Clienti a Yellow Springs, in Ohio. Inviare il prodotto e la prova d’acquisto, con
spese di spedizione prepagate, al Centro Assistenza autorizzato scelto da YSI. Una volta effettuata la
riparazione o la sostituzione il prodotto verrà rinviato, sempre con spese di spedizione prepagate. I
prodotti riparati o sostituti sono coperti da garanzia per il rimanente periodo di validità della garanzia
originale, o per almeno 90 giorni dalla data di riparazione o sostituzione.
Limitazione della garanzia
Questa garanzia non si applica ai prodotti YSI il cui danno o cattivo funzionamento è dovuto a: (i)
installazione, messa in funzione o utilizzo del prodotto non conformi alle istruzioni scritte di YSI; (ii)
abuso o uso improprio del prodotto; (iii) mancato rispetto delle istruzioni scritte di YSI o delle
procedure standard dell’industria; (iv) eventuali riparazioni improprie del prodotto; (v) utilizzo da parte
dell’utente di parti o componenti impropri o difettosi in fase di manutenzione o riparazione del
prodotto; o (vi) eventuali modifiche del prodotto in modalità non espressamente autorizzate da YSI.
QUESTA GARANZIA SOSTITUISCE TUTTE LE ALTRE, ESPLICITE O TACITE, COMPRESE
QUELLE DI COMMERCIABILITÀ O IDONEITÀ PER UNO SCOPO SPECIFICO. LA
RESPONSABILITÀ DI YSI SECONDO QUESTA GARANZIA SI LIMITA ESCLUSIVAMENTE
ALLA RIPARAZIONE O ALLA SOSTITUZIONE DEL PRODOTTO CHE COSTITUIRÀ L’UNICA
ED ESCLUSIVA FORMA DI RIMBORSO PER EVENTUALI DIFETTI COPERTI DA QUESTA
GARANZIA. IN NESSUN CASO YSI SARÀ RESPONSABILE DI EVENTUALI DANNI SPECIALI,
INDIRETTI, INCIDENTALI O CONSEQUENZIALI DERIVANTI DA EVENTUALI DIFETTI DEL
PRODOTTO COPERTO DA QUESTA GARANZIA.
REFERENTE
YSI Inc. • 1725 Brannum Lane • Yellow Springs, Ohio, 45387 USA
800-897-4151 • 937-767-7241 • Fax: 937-767-1058
E-mail: ecosense@ysi.com • sito Web: www.ysiecosense.com
ELENCO DEI PEZZI DI RICAMBIO RACCOMANDATI ................................. 8
2
Page 35
INTRODUZIONE GENERALE
Il modello EC300 è uno dei tre misuratori della linea di prodotti EcoSense di YSI. EC300 è uno
strumento di precisione in grado di misurare conduttività, salinità e temperatura. Un
microprocessore incorporato calcola e compensa tutti I parametri relativi alle determinazioni
della conduttività e della temperatura.
Questa unità ha un alloggiamento IP65 a tenuta stagna. I tasti a sfioramento meccanico sono
estremamente affidabili grazie alla conferma di azionamento tattile e acustica. Lo strumento
utilizza una batteria da 9V. Quando viene ripristinata l’alimentazione non è richiesta alcuna
ricalibrazione.
La parte anteriore dello strumento presenta un ampio display LCD che visualizza
simultaneamente la temperatura e la conduttività – compensata o non compensata a
temperatura – la salinità o la TDS insieme ai prompt per l’utente e agli indicatori di modalità.
L’unità fornisce messaggi agli utenti nel corso delle procedure di calibrazione e di misurazione.
Il modello EC300 è disponibile con una singola cella a quattro elettrodi. Altre caratteristiche
includono la regolazione automatica della conduttività, la compensazione automatica a
temperatura, una batteria a lunga durata e un filtro a CA 50/60 Hz. Questo misuratore è
universale e di facile utilizzo per applicazioni su campo, industriali e di laboratorio.
ISPEZIONE INIZIALE
Disimballare accuratamente l’unità e gli accessori e valutare la presenza di eventuali danni da
trasporto. Controllare se I pezzi ricevuti corrispondono ai materiali elencati nella distinta di
spedizione. Informare immediatamente YSI di eventuali parti danneggiate o mancanti.
Conservare tutti I materiali di imballaggio fino al corretto funzionamento.
TENUTA STAGNA
Sebbene il misuratore EC300 sia montato in un alloggiamento a tenuta stagna, NON utilizzarlo
sott’acqua; il connettore non è resistente all’acqua. L’alloggiamento a tenuta stagna impedisce
danni permanenti all’unità nel caso cada accidentalmente in soluzioni non corrosive. In caso di
immersione, procedere immediatamente come indicato di seguito.
1. Asciugare il connettore se necessario e riposizionare la sonda per la conduttività.
Risciacquare accuratamente l’unità con acqua distillata. Dopo il risciacquo e l’asciugatura,
ispezionare e pulire I connettori per rimuovere tutte le eventuali sostanze contaminanti che
potrebbero danneggiare I collegamenti della sonda.
2. Attendere che l’unità e la sonda siano completamente asciutte prima di riprendere
l’utilizzo.
3. Se l’unità non funziona correttamente dopo le fasi 1 e 2, rivolgersi a YSI per eventuali
riparazioni o per la sostituzione (vedere Garanzia).
INSTALLAZIONE DELLA BATTERIA
Un messaggio iniziale di “BAT” sul display indica un tempo residuo di
funzionamento a batteria pari a circa un’ora, entro le specifiche. Sostituire
la batteria quando sul display appare “BAT”. (Vedere Figura 1.)
Per la sostituzione, rimuovere le due viti sul coperchio della batteria, il
coperchio della batteria e l’o-ring. Sostituire la batteria da 9V. Rimettere a
posto il coperchio della batteria e l’o-ring (allineare l’o-ring correttamente
per garantire un’adeguata chiusura) e serrare le due viti sul
coperchio della batteria per salvaguardare l’ermeticità.
3
Figura 1
Installazione della
batteria
Page 36
FUNZIONI CHIAVE DEL MODELLO EC300
1. : Accende e spegne l’unità. I valori di calibrazione non si cancellano quando l’unità viene
spenta. Quando l’unità non è in uso, spegnerla per prolungare la durata utile della batteria.
Se si consertva l’unità per un periodo prolungato, rimuovere la batteria.
2. MODALITÀ: seleziona la modalità del display. Nel funzionamento Normale, premere MODE
per visualizzare in sequenza la conduttività non compensata, la conduttività compensata a
temperatura, la salinità e la TDS (Total Dissolved Solids, Solidi totali disciolti). In modalità
calibrazione, questo tasto termina la calibrazione corrente e visualizza il parametro della
calibrazione successiva.
3. CAL: durante il funzionamento normale, modifica la modalità da Normale a Calibrazione.
4.
(Invio): in Impostazione Calibrazione, premere questo tasto per salvare il parametro
corrente in memoria.
5. Tasti Δ e ∇:aumentano o diminuiscono il valore presente sul display come desiderato.
IL DISPLAY LCD
1. CONDUTTIVITÀ: appare durante la misurazione della conduttività.
2. BAT: indicatore di batteria scarica.
3. Display principale per i valori di
conduttività compensata e non
compensata, salinità e TDS.
2
4. CAL: indicatore della modalità di
Calibrazione.
5. TDS: appare durante la misurazione
dei solidi totali disciolti (TDS).
3
6. SALINITÀ: appare durante la
misurazione della salinità.
7. CELL: indica il valore della costante
4
di cella della conduttività.
8. °C: lampeggia durante la
misurazione della conduttività
Figura 1.Display LCD
compensata in temperatura. Durante
la calibrazione indica l’unità di
riferimento termico.
%: appare durante la calibrazione; indica l’unità del coefficiente termico.
9. ppt: parti per migliaia; indica la misurazione della salinità.
10. uS, mS: micro Siemens, milli Siemens; indica la misurazione della conduttività.
11. g/l: grammi/litro; indica la misurazione di TDS.
12. °C: display della temperatura.
51
6
7
8
9
10
11
12
4
Page 37
MODALITÀ DI MISURAZIONE
1. Temperatura – Viene visualizzata continuamente la temperatura della soluzione corrente.
2. Conduttività compensata a temperatura – Misurazione della conduttività, compensata a
25 °C o a un altro valore specificato compreso tra 15 e 25°C. Espressa come uS/cm o
mS/cm con “°C” lampeggiante.
3. Conduttività non compensata – Misurazione diretta della conduttività, non compensata a
una temperatura specifica. Espressa come uS/cm o mS/cm.
4. Salinità – Misurazione della salinità; espressa in parti per migliaia (ppt).
5. TDS – Misurazione dei solidi totali disciolti (TDS); espressa in grammi per litro (g/l)
Osservare attentamente le unità visualizzate in piccolo sul display per determinare la modalità
desiderata.
CALIBRAZIONE
L’impostazione della calibrazione contiene cinque sezioni: TDS, Cella, Coefficiente termico,
Riferimento termico e Calibrazione conduttività. Per accedere a queste sezioni procedere come
indicato di seguito.
1. Collegare la sonda per la conduttività e il gruppo cavi all’unità e attivarla. Lo schermo
visualizza CELL e la costante di cella della sonda per la conduttività.
2. Attendere la stabilizzazione delle letture della temperatura, quindi premere CAL per entrare
in modalità calibrazione; sull’LCD appare CAL. Premere MODE per la visualizzazione in
sequenza delle sezioni riportate di seguito.
Nota: premere Invio (
per passare automaticamente alla sezione successiva. Se non ci sono modifiche, l’unità
accetta il valore corrente e passa alla sezione successiva.
) per accettare ogni eventuale modifica dei valori in ciascuna sezione e
TDS
Il valore TDS viene determinato moltiplicando la conduttività (mS) per un fattore TDS. Il valore
del fattore predefinito è 0,65. Per modificare il fattore TDS, utilizzare i tasti Δ e ∇ per regolare il
valore tra 0,30 e 1,00. Premere Invio (
per annullare la modifica e visualizzare la schermata CELL.
) per salvare il nuovo valore, oppure premere MODE
CELL
La seconda schermata visualizza CELL e il valore corrente della cella. Il valore predefinito della
cella è uguale a 5,00 e viene visualizzato in basso a destra. L’unità consente una variazione di ±
0,50 prima di visualizzare un messaggio di errore. Non è possibile regolare il valore della cella in
questa schermata; la calibrazione della conduttività è l’unico modo per regolare la costante di
cella. Premere Invio (
Coefficiente termico.
Nota: premere Invio (
mantiene la costante di cella precedente ed esegue la calibrazione da un valore già offset.
) per riportare la costante di cella a 5,00 e visualizzare la schermata
) per riportare la costante di cella a 5,00. Se si preme MODE, l’unità
Coefficiente termico
L’unità utilizza il coefficiente termico per calcolare la conduttività compensata a temperatura. Il
valore predefinito è 1,91%. Per modificare il coefficiente termico utilizzare i tasti Δ e ∇ per
regolare il valore tra 0 e 4,00%. Premere Invio (
MODE per annullare la modifica e visualizzare la schermata Riferimento termico.
) per salvare il nuovo valore, oppure premere
Riferimento termico
L’unità utilizza il valore di riferimento termico per calcolare la conduttività compensata a
temperatura. Il valore predefinito è di 25°C. Per modificare il riferimento termico utilizzare i tasti
Δ e ∇ per regolare il valore tra 15 e 25 °C. Premere Invio (
5
) per salvare il nuovo valore, o
Page 38
premere MODE per annullare la modifica e visualizzare la schermata Calibrazione
conduttività.
Calibrazione conduttività
1. Immergere la sonda in una soluzione standard a conduttività nota, preferibilmente che si
trovi al centro dell’intervallo delle soluzioni da misurare. Immergere completamente la sonda
senza toccare i lati del contenitore di calibrazione. Agitare leggermente la sonda per
rimuovere le eventuali bolle d’aria intrappolate nella cella di conduttività.
2. Attendere che la temperatura si stabilizzi. È possibile che sul display appaia brevemente il
messaggio ‘rAng’ (range) ad indicare l’auto-regolazione dell’unità; ciò è normale. Dopo la
stabilizzazione della temperatura, utilizzare i tasti Δ e ∇ per regolare il valore di conduttività
a quello dello standard di conduttività a 25 °C. Premere Invio (
calibrazione. L’unità emette un doppio segnale acustico ad indicare la calibrazione riuscita,
quindi automaticamente commuta verso la modalità di funzionamento normale.
) per eseguire la
MISURAZIONI DI CONDUTTIVITÀ
1. Accendere l’unità. Posizionare la sonda nella soluzione da misurare. Immergere
completamente la sonda. Agitare leggermente la sonda per rimuovere le eventuali bolle
d’aria intrappolate nella cella di conduttività.
2. Premere MODE per attivare la modalità di misurazione desiderata. È possibile che sul
display appaia brevemente il messaggio ‘rAng’ (range) ad indicare l’auto-regolazione; ciò è
normale. Attendere che la temperatura si stabilizziprima di effettuare le misurazioni.
MANUTENZIONE DELLA SONDA
Il requisito più importante per misurazioni accurate e riproducibili della conduttività è
rappresentato da una cella pulita. Una cella sporca modifica la conduttività di una soluzione per
contaminazione. Pulire con cura la cella prima di conservarla. Per pulire la cella di conduttività
procedere come descritto di seguito.
1. Porre la cella in una soluzione detergente e agitare
per due - tre minuti. Qualsiasi sostanza detergente
schiumosa a base di acidi, quale il Dow Chemical
Bathroom Cleaner, dovrebbe garantire una pulizia
adeguata. Per un detergente più aggressivo,
utilizzare una soluzione formata da alcool isopropilico
1:1 e HCl 1 N. Rimuovere la cella dalla soluzione
detergente.
2. Utilizzare la spazzola in nylon (in dotazione) per
rimuovere le eventuali sostanze contaminanti dall’interno della camera dell’elettrodo.
3. Ripetere le fasi uno e due finché la cella è completamente pulita. Risciacquare con cura la
cella in acqua deionizzata o di rubinetto pulita.
6
Page 39
INDIVIDUAZIONE E RISOLUZIONE DEI PROBLEMI
DISPLAY PRINCIPALE PROBLEMA SOLUZIONE POSSIBILE
OvEr
OvEr/Undr durante la
calibrazione
DISPLAY
PRINCIPALE
OvEr/Undr
DISPLAY
SECONDARIO
OvEr
Undr
• La conduttività è >200,0
mS
• La salinità è > 70,00 ppt
La calibrazione della
costante di cella è fuori dal
range
Temperatura >90,0 °C
Temperatura < -10,0 °C
• Immergere completamente la
sonda.
• Consentire un tempo sufficiente
di stabilizzazione per l’elettrodo
e la sonda Temp.
• Ricalibrare con il valore corretto
per lo standard di conduttività.
• Sostituire lo standard di
conduttività.
• Pulire la cella.
• Restituire all’assistenza.
• Diminuire/Aumentare la
temperatura del campione.
• Restituire all’assistenza.
SPECIFICHE
Display Range Precisione Risoluzione
Conduttività,
Auto-
regolazione
Salinità da 0,0 a 70,0 ppt 0,2% scala intera 0,1 ppt
Temperatura °C da -10,0 a 90 °C ± 0,2 °C o ± 0,4% scala intera,
da 0,0 a 499,9 uS/cm
da 500 a 4999 uS/cm
da 5,00 a 49,99
mS/cm
da 50,0 a 200,0
mS/cm
±1% della lettura più 2 uS/cm
±1% della lettura più 5 uS/cm
±1% della lettura più 0,05 uS/cm
±2,5% della lettura più 0,5 uS/cm
quale dei due valori è maggiore
0,01 uS/cm
1 mS/cm
0,01 mS/cm
0,1 mS/cm
0,1 °C
Temperatura di riferimento
Coefficiente temperatura
Costante di cella
Intervallo costante TDS
Potenza
Back up di calibrazione
Feedback audio
Resistenza all’acqua
Intervallo temp. di funzionamento
Intervallo di umidità relativa al
funzionamento
Sonda temperatura
Dimensioni (L x P x H)
Peso (batterie incluse)
da 15,0 a 25,0
da 0,0% a 4,0%
5,00 ± 0,50
da 0,30 a 1,00
Una batteria da 9V
Sì
Sì, su tutti i tasti a sfioramento
IP 65 a tenuta stagna
0 - 50
fino al 95%
Termistore, 10kΩ / 25
186 x 70 x 37 mm (7,3 x 2,8 x 1,5 pollici)
430 grammi (1 lb)
°C
°C
°C
7
Page 40
ELENCO DEI PEZZI DI RICAMBIO RACCOMANDATI
N. PEZZO DESCRIZIONE
300-4 Sonda di 4 metri e gruppo cavi
300-10 Sonda di 10 metri e gruppo cavi
380 Contenitore per il trasporto, rigido
480 Contenitore per il trasporto del misuratore con spallina, morbido