Fig. 1-1Structure of the IQ SENSOR NET operating manual
The IQ SENSOR NET operating manual has a modular structure like the
IQ S
ENSOR NET itself. It consists of a system operating manual and the
operating manuals of all the components used.
Please file this component operat ing manual in to the ring binder of th e
system operating manual.
1 - 1
OverviewMIQ/CR3; DIQ/CR3
1.2Features of the combi output module
General characteristicsThe combi output module has three current outputs and three relay
outputs. You can link current outputs and relay out puts to sensors. The
linked current outputs and relay outputs can, for example, be used to
monitor sensors or to outpu t measurement data. Unlinked relay outputs
can be used for general monitoring functions.
With the standard MIQ module housing, the co mbi output module has
the same characteristics as all MIQ modules regarding stability,
leakproofness and weather resist ance. It also pr ovides t he same wi de
variety of installation opt ions (stacked mounting, canopy mounting,
tophat rail mounting, etc.).
Instrument typesThe MIQ/CR3 and DIQ/CR3 combi output modules differ in their
compatibility with the individual IQ S
SystemCompatible combi output module
ENSOR NET system families.
2020 XTMIQ/CR3
182DIQ/CR3
Terminal stripThe combi output module has the following electrical connections on
the terminal strip inside the housing:
3 x relay contact
3 x current output
2 x SENSORNET connection
1 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Safety instructions
2Safety instructions
This operating manual contains special instructions that must be
followed during the installation of th e combi output module. Thus, it is
essential for the operator to read this component operating manual
before carrying out any work with the sys tem. In addition to this manual,
the S
AFETY chapter of the IQ SENSOR NET system operating manual
must be followed.
Always keep this component operating manual together with the
system operating manual and all other component operating manuals
in the vicinity of the IQ S
ENSOR NET system.
Special user
qualifications
General safety
instructions
Other labels
External circuits carryi ng mains volt age must only be connect ed to the
relay contacts by a qualified electrician.
Safety instructions in this operat ing manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the danger level:
Warning
indicates instructions that must be followed precisely in order to
prevent serious dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid slight injuries to personnel or damage to the instrument or
the environment.
Note
This symbol indicates instructions that describe special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba76032e0101/2012
2 - 1
Safety instructionsMIQ/CR3; DIQ/CR3
2.1Authorized use
The authorized use of the combi output module consists of providing
relay and current outputs in the IQ S
ENSOR NET.
Please observe the technical specifications according to chapter 6
T
ECHNICALDATA. Only operation according to the i nstructions in this
operating manual is authorized.
Any other use is considered to be unauthorized. Unauthorized use
invalidates any claims with regard to the guar antee.
2.2General safety instructions
The combi output module is constructed and inspected according to the
relevant guidelines and norms for electronic instruments (see chapter
6 T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and
operational safety
The failure-free function and operational safety of the combi output
module is only guarant eed if the general ly applica ble saf ety measur es
and the special safety instruc tions in this operating manual are followed
during its use.
The failure-free function and operational safety of the combi output
module is only guaranteed under the environmental condit ions that are
specified in chapter 6 T
ECHNICALDATA.
Safe operationIf safe operation is no longer possible, the combi output module must
be taken out of operation and secured against inadvertent operation.
Safe operation is no longer possible if the combi outpu t mo dule:
has been damaged in transport
has been stored under adverse conditions for a lengthy period of
time
is visibly damaged
no longer operates as described in this manual.
If you are in any doubt, contact the supplier of your combi output
module.
2 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Installation
3Installation
3.1Scope of delivery
The scope of delivery of the combi output module is li sted in the
I
NSTALLATION chapter of the system operating manual.
3.2Installation in the IQ SENSOR NET
The IQ SENSOR NET provides a number of options for integrating the
combi output module mechanically and electric ally in the system
(stacked mounting, distributed mounting, etc.). The various types of
installation are described in detail in the I
system operating manual.
NSTALLATION chapter of the
3.2.1Software requirements of the MIQ/CR3 and IQ S
ENSOR NET
Software requirements for the use of the Cleaning function:
MIQ/CR3: from version 2.20 up
Controller: from version 2.20 up
If you use a terminal with a software version from 2.20 up, the Clean
display flashes. If you use an MIQ/T2020 terminal with a software
version lower than 2.20, the Clean display does not appear.
The measured value display of the sensor flashes.
Software requirements for use of the funct ions, PID cont roller, Manual control and Fixed current value:
MIQ/CR3: from version 2.30 up
Note
It is possible to update the software if your components have older
software versions. Please contact YSI.
ba76032e0101/2012
3 - 1
InstallationMIQ/CR3; DIQ/CR3
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm
cable gland M20
3.3Electrical connections: General instructions
Cable glandsAll electric cables are fed from below via prepared openings in the
enclosure of the module. Cable glands with different clampi ng ranges
are included with the mo dule to provid e sealing between the cable and
enclosure as well as for strain relief. Select the matching cable gland
for the respective cable diameter:
Small, c lamping range 4.5 to 1 0 mm. This cable gland is suitabl e for
all IQ S
ENSOR NET sensor cables.
Large, clamping range 7 to 13 mm. This cabl e gland is r equi red for
cable sheaths with an outside diameter of more than 10 mm and is
screwed into the enclosure via an extension piece.
Note
If necessary, you can order more large cable gla nds in a set of 4 pieces
(Model EW/1, Order No. 480 051).
3 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Installation
General installation
instructions
Observe the following points when attaching connecti ng wires to the
terminal st rip
Shorten all wires to be used to the length required for the instal lation
Always fit all the ends of the wires with wire end sleeves before
connecting them to the terminal strip
Any wires that are not used and project into the enclosure must be
cut off as closely as possible to the cable gl and.
Screw a small cable gland with sealing ring in to each remaining free
opening and close it with a blind plug.
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages which could result in life
threatening electric shock when working with t he IQ S
Always cut off any wires that are not in u se as closely as possible
to the cable gland.
ENSOR NET.
ba76032e0101/2012
3 - 3
InstallationMIQ/CR3; DIQ/CR3
3.4Connections to the relay and current outputs
Warning
If external electrical circuits that are subject to the danger of
physical contact are incorrectly connected to the relay contacts,
there may be a danger of life threaten ing electric shock. Elec trical
circuits are regarded to be subject to the danger of physical
contact when there are voltages higher than the Safety Extra Low
Voltage (SELV).
Pay attention to the following points during installation:
Electrical circuits subject to the danger of physical contact
must only be connected by a qualified electrician.
Electrical circuits subject to the danger of physical contact
must only be connected when they are voltage-free.
If electrical circuits subject to the danger of physical contact
are switched with a relay, no circuit that is not subject to this
danger (e. g. the MIQ/CHV module) may be operated in the
same output module. Use an additional output module for such
applications.
Switching voltages and switching currents on the relay
contacts must not exceed the values specified in chapter 6
T
ECHNICALDATA. Protect electrical cir cuits against currents that
are too high with an electrical fuse.
Only single-phase consumers can be switched with the relays.
Under no circumstances must multiphase consumers be
switched with the aid of several relays (example three-phase
current driven pumps). Always switch multiphase consumers
via a protective relay.
After the output module has been installed, it may only be
opened if all external voltages have been switched off
beforehand.
3 - 4
Materials required Wire end sleeves, suitabl e for the connecting wires, with suitable
crimping tool
4 x screwed cable gland with sealing ring (scope of delivery of the
combi output module).
Tools Cable stripping knife
Wire stripper
Phillips screw driver
Small screw driv e r
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Installation
1
3
2
RelayCurrent output
X18 X17
250 VAC
5AAC
R1
X16 X15
250 VAC
5AAC
R2
X14 X13
250 VAC
5AAC
R3
X12X10X8X11X9X7
0/4...20mA 0/4...20mA 0/4...20mA
+ REC -C1+ REC -C2+ REC -
C3
X6 X5 X4
SENSORNET 2
RED
SHIELD
GREEN
X3 X2 X1
SENSORNET 1
RED
SHIELD
GREEN
ON
OFF
SN TERMINATOR
Connecting lines to the
terminal strip
1Open the module.
Fig. 3-1Terminal strip with the relay and current connections
2Screw the cab le gland (pos. 1 i n Fig. 3-1) with the seal ing ring
(pos. 2) into the module housing.
3Loosen the coupling ring (pos. 3 in Fig. 3-1).
4Feed the line through the cable gland in the module housing.
5Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the l abel located under the
terminal strip.
6Tighten the coupling ring (pos. 3 in Fig. 3-1).
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that safe areas could come into
contact with dangerous voltages. This could result in life
threatening electric shock when working with t he IQ S
Always cut off any wires that are not in u se as closely as possible
to the cable gland.
ENSOR NET.
ba76032e0101/2012
3 - 5
InstallationMIQ/CR3; DIQ/CR3
7Close the module.
3 - 6
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4Settings
The combi output module has three relays outputs and thre e current
outputs.
Relay outputs operate as openers or closers.
Current outputs provide a current that depends on the measured
value.
On the terminal or Universal Transmitter, you can
assign names to the outputs (with the 2020 XT system only, see
section 4.2).
link outputs with sensors (see section 4.3)
delete links of outputs with sensors (see section 4.4)
adjust outputs (see section 4.5 and section 4.6)
check the condition of the outputs (see section 4.7)
Functions of current and
relay outputs
Basic information on how to use relay outputs is given in section 4.1.
Note
The general operating principles are given in the relevant system
operating manual or in the component operating manual of the terminal
components.
Relay output (see section 4.5)
System monitoring
Sensor monitoring
Limit indicator
Frequency controller
Pulse-width contr.
Cleaning
Sensor-controlled
Manual control
Alarm contact (with the 2020 XT system only)
ba76032e0101/2012
Current output (see section 4.6)
Recorder
PID controller
Fixed current value
4 - 1
SettingsMIQ/CR3; DIQ/CR3
4.1Basic information on relay functions
In this chapter, you will find general basic information concerning the
following relay functions:
Monitoring (see section 4.1.1)
Limit indicator (see section 4.1.2)
Proportional output (see section 4.1.3)
4.1.1Monitoring
When using a relay for monitor ing, a relay action (Open, Close) occurs
when certain states occur. This function is suitable, e.g. for the
monitoring of errors in the system.
Note
For monitoring functions, use the relay preferably as an opener (see
section 4.5.1). In the cas e of an err or, t he r elay ope ns. As a res ult, the
monitoring function operates even if, e.g. the supply voltage fails.
4 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.1.2Limit indicator
With a limit indicator, a relay switches when a speci fied limiting value is
exceeded or undercut.
Limit indicators can be used in the following way:
Monitoring a limiting value using a relay:
when a limiting value (upper o r l ower limiting value) is exceeded or
undercut, a relay switches. The Open or Close relay actions are
possible in each case (see page 4-4).
Monitoring two limiting values using two relays:
If the upper limiting v alue is exceeded or unde rcut, a relay swi tches,
and if the lower limiting value is exceeded or undercut, anothe r relay
switches. The Open or Close rel ay actions are possible in each case
(see page 4-4).
Note
If the simple monitoring function (Open, Close) with one or two relays
is not sufficient, use proportional output (see section 4.1.3).
ba76032e0101/2012
4 - 3
SettingsMIQ/CR3; DIQ/CR3
Time
Measured
value
UL
LL
Hysteresis UL
Hysteresis LL
3
5
Relay 1
Relay 2
4
1
2
t1
t1
6
t2t2
Monitoring limiting
values using one or two
relays
Fig. 4-1Switching points for relays with the function of a limit indicator
1Upper limit value (relay 1) exceeded
2Selected switching delay t1 for relay 1 expired
Relay 1 switches
3Hysteresis for upper li miting value (relay 1) undercut
4Selected switching delay t1 for relay 1 expired
6Hysteresis for lower limiting value (relay 2) exceeded
Selected switching delay t2 for relay 2 expired
Relay 2 switches back
A switching delay (t) can be set up for each relay for swit ching
processes. This is the time period for which a limiting value must be
exceeded before the relay switches. This prevents frequent switching if
the measured values are close to the limiting value.
4 - 4
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.1.3Proportional output
In the case of proporti onal output, a relay switc hes cyclic ally on and of f
in a defined measured value range (proportional range). At the same
time, the relay switches with a:
duration of operation that corresponds to the measured value
(pulse-width output, see page 4-7) or
switching frequency (frequency output , see page 4-8).
Proportional outputs can be used in the following way:
Output with one relay:
An output range is defi ned with a Start value and an End value. No
output takes place above and below the output range
(see page 4-6).
Output with two relays:
An output range is defined for each relay with a Start value and an
End value. One relay outputs in the up per output range and a further
relay in the lower output range (see page 4-6).
ba76032e0101/2012
4 - 5
SettingsMIQ/CR3; DIQ/CR3
10
100
Proportional
band
Measured value
1
2
90
0
Switching frequency f or
Pulsev
width
Proportional bands
Relay 1 and 2
Measured value
Relay 1
Relay 2
1
2
2
1
10
100
90
0
Switching frequency f or
Pulsev
width
Output with one relay
Fig. 4-2Output with one relay
Output with two relays
Fig. 4-3Output with two relays
1Start value
Measured value with a minimum pulse width or switching
frequency
2End value
Measured value with a maximum pulse width or switching
frequency
4 - 6
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
T
Time [s]
Relay
Off
On
t
On
t
Off
Pulse width outputThe output of the pulse width is used, e.g. for controlling valves.
Pulse-width regulation changes the durat ion of operation (t
) of the
on
output signal. Depending on the posi tion of the measured value in the
proportional range, the relay is operated for a longer or shorter time
period.
Fig. 4-4Relay output of the pulse-width output
The cycle duration (T) is made up of the tur n-o n and t urn-of f swit chin g
duration (t
duration (T) remains constant, the turn-on duration (t
, t
) of the relay together. While the selected cycle
on
off
) changes
on
depending on the measured value and, with it, the puls e width (v). The
pulse width can be set from 0 % to 100 %.
v = (t
v = pulse width, T = cycle duration, t
/ T) * 100 %
on
= turn-on duration
on
If the measured value is at the end of the proportional range (End
value), the turn-on duration (t
) is long, the turn-off dur ation is short.
on
This means the relay operates for a longer period.
If the measured value is at the beginning of the proportional range
(Start value ), the turn-on duration (t
) is short, and the relay
on
operates for a correspondingly shorter period.
Note
If the duration of the closi ng or open ing pul se is shorter t han 0.1 s, the
relay remains open or closed for the complete cycle duration.
ba76032e0101/2012
4 - 7
SettingsMIQ/CR3; DIQ/CR3
Time [s]
t= 0.3 s
On
Relay
Off
On
Frequency outputSwitching frequency output is used, e.g. for controlling dosing pumps.
In contrast to the pulse-width output, not the pulse width is modulated
with frequency output b ut the swi tching frequency of the out put signal .
Depending on the position of the measured value in the proportional
range, the relay is switched more often or le ss often.
Fig. 4-5Relay output of frequency output
While the selected switching duration (ton = 0.3 s) always remains
constant, the switching frequency at whic h the relay switches changes
depending on the measured value.
If the measured value is at the end of the proportional range (End
value), the switching frequency is higher.
If the measured value is at the beginning of the proportional range
(Start value), the switching frequency is low.
4 - 8
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
Measured value
Switching frequency f or
Pulsev
width
End valueStart value
Max.
Min.
Measured value
Start valueEnd value
Max.
Min.
Switching frequency f or
Pulsev
width
Characteristic curvesThrough the select ion of the Start value and En d valu e, the proportional
output can be operated with a p ositive or negative char acteristic curve.
Positive cha r a cteristic curv e :
Select the End value to be greater than the Start value.
The turn-on duration or frequency increases wit h an increasing
measured value (see page 4-10).
Negative characteristic curve:
Select the End value to be smaller than the St art value.
The turn-on duration or frequency decreases with an increasing
measured value (see page 4-11).
The maximum values for the pulse width or switching frequency are
assigned to the End value and the minimum values f or turn-on duration
or frequency are assigned to the Start value.
ba76032e0101/2012
Fig. 4-6Positive characteristic curve
Fig. 4-7Negative characteristic curve
4 - 9
SettingsMIQ/CR3; DIQ/CR3
0
10
50
90
100
Proportional
band
t
On
t
Of f
50%50%
10%
10%
90%
90%
Cycle
duration T
Time
Measured value
21
Pulsev [%]
width
0
60
120
Proportional
band
Time
Measured value
21
50 % of maximum
switching frequency
Maximum
switching frequency
Switching frequency f [1/min]
Minimum
switching frequency
Positive characteristic
curve
The proportional output range begins above the initial va lue. If the
proportional range is undercut or exceeded, the selected behavior
comes into force.
Fig. 4-8Pulse width output
4 - 10
Fig. 4-9Frequency output
1Start value
Measured value with a minimum pulse width or switching
frequency
2End value
Measured value with a maximum pulse width or switching
frequency
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
0
10
50
90
100
Proportional
band
t
On
t
Of f
50%50%
10%
10%
90%
90%
Cycle
duration T
Time
Measured value
12
Pulsev [%]
width
0
50
100
Measured value
Proportional
band
1
50 % of maximum
switching frequency
Maximum
switching frequency
Time
2
Switching frequency f [1/min]
Minimum
switching frequency
Negative characteristic
curve
The proportional output range begins below the initial value. If the
proportional range is undercut or exceeded, the selected behavior
comes into forc e .
Fig. 4-10 Pulse width output
ba76032e0101/2012
Fig. 4-11 Frequency output
1Start value
Measured value with a minimum pulse width or switching
frequency
2End value
Measured value with a maximum pulse width or switching
frequency
4 - 11
SettingsMIQ/CR3; DIQ/CR3
4.2Entering / editing the name of an output
For easier identificati on of the outputs, an individual name can be given
to each output in the Edit list of outputs overview. This option is
available with the 2020 XT system only.
1Open the Settings menu with s.
2Using
settings -> Edit list of outputs.
The Edit list of outputs displ ay opens.
3Highlight a name in t he Name column with d and confirm with
d and g, select and confirm the menu item, System
g.
Fig. 4-12 130 - Edit list of outputs
4 - 12
4Select a letter, a numeral or a special character with d and
confirm with
5Complete the name of the output and confirm with
g.
g.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.3Linking the output with a sensor
The following steps describe oper ation wi th the 202 0 XT system. With
the System 182 XT, the outputs of the DIQ/CR3 can be found directly
under the Settings menu (see system operating manual of the DIQ/
S 182 XT).
1Open the Settings menu with s.
2Using
d and g, select and confirm the menu item, System
settings -> Settings of outputs and li nks.
The Settings of outputs and links display opens.
3Highlight the column with d and confirm with g.
4Highlight an output with
d and confirm with g. The Link
with... display opens.
The display shows a list of the sensors to which a link is
possible.
ba76032e0101/2012
Fig. 4-13 150 - Settings of outputs and links:Link with...
5Select a sensor with d and confirm with g.
The output is linked with the selected sensor.
Note
Outputs that are linked with sensors can be identified in the Ser. no.
field of the Settings of outputs and links overview by the specif ication of
the linked sensor.
4 - 13
SettingsMIQ/CR3; DIQ/CR3
4.4Deleting a link with an output
If a link from a current or relay output with a sensor is no longer
required, the link can be deleted.
The following steps describe operati on with the 20 20 XT system. With
the System 182 XT, the outputs of the DIQ/CR3 can be found directly
under the Settings menu (see system operati ng man ual of the DIQ/
S 182 XT).
1Open the Settings menu with s.
2Using
settings -> Settings of outputs and links.
The Settings of outputs and links display opens.
3Highlight the col umn & with d and confirm with g.
4Highlight a linked output with
Fig. 4-14 150 - Settings of outputs and links:Erase link
d and g, select and confirm the menu item, System
d and confirm with g.
4 - 14
5Select Erase li nk with d and confirm with g.
A security prompt appears.
6Select Erase link with
The link is deleted.
d and confirm with g.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.5Setting the relay outputs
The following steps describe oper ation wi th the 202 0 XT system. With
the System 182 XT, the outputs of the DIQ/CR3 can be found directly
under the Settings menu (see system operating manual of the DIQ/
S 182 XT).
1Call up the measured value display with m.
2Open the Settings menu with
3Highlight the Settings of outputs and links menu item with d
and confirm with
appears.
4Highlight the Feature column with
5Highlight a line for a relay output (Rx) in the Feature column
with
d and confirm with g.
The Settings of outputs and links display opens.
6Highlight the Relay function menu item with
with
g.
g. The Settings of outputs and links display
s.
d and confirm with g.
d and confirm
ba76032e0101/2012
Fig. 4-15 150 - Settings of outputs and links
7Select one of the functions in the following list with d and
confirm with
g.
4 - 15
SettingsMIQ/CR3; DIQ/CR3
FunctionDescription
No functionThe relay output is not used.
System monitoringsee section 4.5.2
Sensor monitoringsee section 4.5.3
Limit indicatorsee section 4.5.4
Frequency controllersee section 4.5.5
Pulse-width contr.see section 4.5.6
Cleaningsee section 4.5.7
Sensor-controlledsee section 4.5.8
Manual controlsee section 4.5.9
Alarm contactsee section 4.5.10
8Carry out the setti ngs for the relay outputs with d and g.
The settings include the relay action (see section 4.5.1) and
sensor-dependent settings.
9Highlight and confirm Save and quit with
d and g.
The new settings are stored.
As soon as a function is selected for a relay output, a relay action can
be selected (see section 4.5.1).
4.5.1Relay action
The following actions of the relay can be laid down in the Action setting:
SettingsExplanations
OpenThe relay should open if any event occurs.
CloseThe relay should close if any event occurs.
Note
Preferably set up the relay output as normally closed for monitoring
functions (Action Open).
4 - 16
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.5.2System monitoring
FunctionThe System monitoring enables to monitor system errors.
In order to set up the System monitoring function for a relay output, the
relay output must not be linked with a sensor (see section 4.3).
It can be used to monitor the following system errors.
Settings
SettingsSelectionExplanations
Power failureOn
Off
The Power failureOn function
monitors the supply voltage in
the IQ S
ENSOR NET on the
combi output module. If the
voltage falls below the critical
value, the relay switches.
CommunicationOn
Off
The Communication On function
monitors the functioning of the
controller and the
communication to the combi
output module.
Collective error
message
On
Off
The Collective error message
On function simultaneously
monitors the proper functioning
of all sensors and the
functioning of the monitoring
combi output module. (for
details see below this table)
ActionOpenFor all functions of the System
monitoring, the relay action is set to Open.
Collective error
message
ba76032e0101/2012
In the case of a Collective error message, the relay opens i f one of the
following malfunctions occurs:
One of the sensors properly registered at the controller does not
supply a valid main measured value
One of the sensors properly registered at the controller does not
supply a valid secondary measured value
The monitoring combi output modul e has not received any new data
from the controller for two minutes.
In any case, the relay remains open for 10 seconds and only closes
when the malfunction is no longer present.
In the following cases, the relay does not
open despite an invalid
measured value:
4 - 17
SettingsMIQ/CR3; DIQ/CR3
The sensor is being calibrated
The sensor is in the maintenance condition
The sensor is being cleaned with the aid of a valve module in the
system (compressed air operated cleaning system).
4.5.3Sensor monitoring
FunctionThe Sensor monitoring function enables to monitor sensor errors and
the maintenance condition.
In order to set up the Sensor monitoring function for a relay out put, the
relay output must be linked with a sensor (see section 4.3).
Settings
SettingSelectionExplanation
ErrorsSpecialSpecial sensor errors are
monitored and can prompt a
relay action.
AllAll sensor errors (special ones
and general ones) are
monitored and can prompt a
relay action.
OffSensor errors are not
monitored.
Mainten. ConditionOn
Off
Switching on and off the
maintenance condition (see
section 4.9) is monitored and
can prompt a relay action.
ActionOpen
Relay action (see section 4.5.1)
Close
Note
Preferably set up the relay output as opener for mon it o ring functions
(ActionOpen, see section 4.5.1).
4 - 18
Special
sensor errors
Sensor messages include error s and informatio n that are regi stered by
the sensor.
The special sensor errors are sensor-dependent. Details of this are
given in the component operating manual of the respective sensor.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
General
sensor errors
Initcan prompt a relay action for a short time, depending
on the starting behavior of the system
----Invalid measured value, or defective sensor
ErrorCommunication with sensor interrupted
OFLMeasuring range undercut or exceeded (overflow)
4.5.4Limit indicator
FunctionThe characteristic of the limit indicator is laid down in the Lim it value
UL, Limit value LL, Hysteresis UL and Hysteresis LL settings. The
fundamentals of the function are described in the introductory chapter
(see section 4.1.2).
In order to set up the Li mit indicator function for a relay output, the relay
output must be linked with a sensor (see section 4.3).
Settings
SettingSelection/ValuesExplanation
Limit valuesUL main variable
LL main variable
UL adjoining var.
LL adjoining var.
Limit value ULAny upper and lower
limiting value
Limit value LL
within the measuring
range (sensordependent)
Hysteresis UL0 - 5% of the
Hysteresis LL
measuring range
Main variable
designates the actual
measured parameter
of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).
Minimum spacing
between the upper
and lower limiting
value:
5 % of the measuring
range
Hysteresis for Limit
value UL and Limit
value LL.
ba76032e0101/2012
4 - 19
SettingsMIQ/CR3; DIQ/CR3
SettingSelection/ValuesExplanation
Behavior at errorOpen
Close
Unchanged
The relay opens,
closes, or remains
unchanged in case of
system errors or
sensor errors (see
page 4-39).
ActionOpen
Close
Relay action (see
section 4.5.1)
Switching delay0 ... 3600 sThe time period for
which a limit ing val ue
must be exceeded
before the relay
operates.
Prevents frequent
switching for
measured values that
are close to the
limiting value.
4 - 20
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.5.5Frequency controller
FunctionThe characteristic of the frequency output is laid down in the Start
value, End value, Frequency (f) min. and Frequency (f) max. settings.
The fundamentals of the function are described in the intr oductory
chapter (see section 4.1.3).
In order to set up the Frequency controller function for a relay output,
the relay output must be linked with a sensor (see section 4.3).
Settings
SettingSelection/ValuesExplanation
Measured variableMain variable
Adjoining variable
Main variable
designates the
actual measured
parameter of the
sensor (e.g . pH,
oxygen, etc.).
Adjoining variable
designates an
additional
measured
parameter (e.g.
temperature).
Start valuewithin the
End value
measuring range
(sensor-
Minimum spacing:
5 % of the
measuring range
dependent)
Frequency (f) min.0 to 120 1/minMinimum spacing:
Frequency (f) max.
10 1/min
Frequency with error0 to 120 1/minIn case of system
errors or sensor
errors (see page 4-
39), the relay
switches with the
frequency
specified.
ActionOpen
Close
Relay action (see
section 4.5.1)
Characteristic curveIf a value is entered for End value that is greater than the Start value,
the output has a positive characteristic curve.
In order to obtain a negative characterist ic curve, a value must be
entered for End value that is smaller than the value fo r Start value.
ba76032e0101/2012
4 - 21
SettingsMIQ/CR3; DIQ/CR3
4.5.6Pulse-width contr.
FunctionThe characteristic of the pulse width output is laid down in the Start
value, End value, Pulse width (v) min. and Pulse width (v) max.
settings. The fundamentals of the function are described in the
introductory chapter (see section 4.1.3).
In order to set up the Pul se-widt h c ontr. func ti on fo r a r elay o utput , the
relay output must be linked with a sensor (see section 4.3).
Settings
SettingSelection/ValuesExplanation
Measured variableMain variable
Adjoining variable
Main variable
designates the
actual measured
parameter of the
sensor (e.g. pH,
oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).
Start valuewithin the
measuring range
End value
(sensor-
Minimum spacing:
5 % of the
measuring range
dependent)
Pulse width (v) min.0 ... 100 %Minimum spacing:
10 % of the Cycle
Pulse width (v) max.
duration (T)
Cycle duration (T)5 ... 100 sLength of the
switching period T
T = (t
on
+ t
off
)
Error pulse width0 ... 100 %In case of system
errors or sensor
errors (see page 4-
39), the relay
switches with the
pulse-width
specified.
ActionOpen
Close
Relay action (see
section 4.5.1)
Characteristic curveYou can specify the minimum and maximum pulse width (v). This
determines the steepness of the characteristic curve of the output.
4 - 22
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.5.7Cleaning
Note
When using the MIQ/CHV PLUS valve module, it is best to set the
Cleaning function directly at the valve output (V) of the MIQ/CHV PLUS
(see operating manual of the MIQ/CHV PLUS). The relay can then be
used for other purposes.
FunctionThe Cleaning f unction enables the time controlled automatic sta rt of the
sensor cleaning function with the aid of a relay of the combi output
module.
The relay controls the MIQ/CHV valve module and switches on or off
the compressed air for the CH sensor cleaning head.
In order to set up the Cleaning function for a relay output, the relay
output must be linked with a sensor (see section 4.3).
Testing the
operativeness
The relay assigned to the combi output module always works as a
closer.
The cleaning cycle consists of Cleaning dura tion and Adjustment time.
During the cleaning cycle the Clean display f lashes. The outputs linked
with this sensor are frozen. The maintenance condition (see
section 4.9) is active.
After the Cleaning duration, the relay i s opened. During the following
Adjustment time the outputs remain blocked.
The outputs linked with this sensor are only releas ed when the cleaning
cycle is completed. The Clean display disappears. The maintenance
condition is finished.
You can test the operativeness of the cleaning system as follows:
manually open or close the relay with the Manual contr ol f unction (see
section 4.5.9) and, while doing so, check the behavior of the clea ning
system.
Alternatively, you can test the operativeness of the cleaning system by
checking the performance of the function at the start time set up
(reference time ± interval). To perfor m a test immediat ely , you can set
the reference time so that the next cleaning will start in a few minutes
(settings: see following table) .
ba76032e0101/2012
4 - 23
SettingsMIQ/CR3; DIQ/CR3
Settings
SettingSelection/ValuesExplanation
Reference time (h)0 ... 23 hTime at which a
cleaning cycle is
Reference time
(min)
0 ... 60 min
started. Further
cleaning cycles will
be performed at the
times specified by the
cleaning interval.
Interval unit1 .. 7 d
1 .. 24 h
5 .. 60 min
Cleaning interval1/2/3/4/5/6/7 d
Selection of range
and unit for the
Cleaning interval.
Repeat interval for
the cleaning function:
or:
1/2/3/4/6/8/12/24 h
Time between the
start time of a
cleaning cycle and
or:
5/10/15/20/30/60 min
the start time of the
next cleaning cycle*.
Cleaning duration0 ... 300 sDuration of the
cleaning
Adjustme n t time0 ... 900 sTime extension to
allow the sensor to
adjust to the test
sample after the
cleaning.
* With short cleaning intervals, the adjustable values for the Cleaning duration and Adjustment time are limited. The following values apply:
Cleaning intervalCleaning durationAdjustment time
≤ 10 minmax. 60 smax. 120 s
≤ 20 minmax. 180 smax. 300 s
Note
With this, the cleaning times are fixed. They on ly change when the
reference time is changed.
Note
The reference time and all furt her cleaning times r elate to the date and
time of the system clock. How to set the system clock is described in
the system operating manual.
4 - 24
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
relay
condition
open
closed
00:00
Time
3
1
2
t1t1t1
t1a t1b
44
t2t2t2t2
1
04:0012:0024:0020:00
Example
SettingResult
Reference time (h):
Reference time
(min):
Interval unit:
12
0
Hours (h)
8 h
Reference time: 12:00 hours
This specifies the following
start times :
04:00, 12:00 and 20:00 hours
Cleaning interval:
ba76032e0101/2012
Fig. 4-16 C leaning cycle
1Reference time
Start of a cleaning cycle (t1)
Start of the specified Cleaning duration (t1a)
2End of the specified Cleaning duration (t1a)
Start of the specified Adjustment time (t1b)
3End of the specified Adjustment time (t1b)
End of the cleaning cycle (t1)
4Reference time ± Cleaning interval (t2)
Start of a cleaning cycle
t1Cleaning cycle =
Cleaning duration (t1a) Adjustment time (t1b)
Linked outputs are frozen.
t2Cleaning interval
4 - 25
SettingsMIQ/CR3; DIQ/CR3
Canceling the cleaningA running cleaning cycle is canceled:
Automatically
– If the sensor switches to the inactive condi tion during the cleaning
cycle
Manually
– By pressing
c
– By switching on the maintenance condition
Each time the cleaning cycle is canc eled, the relay opens immediately.
If the cleaning cycle i s canceled aut omatically, the o utputs linked to the
sensor are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the
maintenance condition. The linked outputs are only released after the
maintenance condition was terminated manually.
The next cleaning cycle will be performed at the time set up.
Note
In case of a power failure, all relays open. The cleaning cycle is
canceled. The outputs linked with the sensor change to the non-act ive
condition (see section 4.8.2). As soon as the power is available, the
outputs are released again. The next cleaning cycle will be perform ed
at the time set up.
4 - 26
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.5.8Sensor-controlled
With the Sensor-controlled function, the relay is controlled by a linked
sensor.
Requirements For the 2020 XT system, a controller with a software version from
2.80 up is required
Sensor that transmits signals to tr igger a cleaning cycle, e.g. UV/VIS
sensor
Settings
SettingSelection/ValuesExplanation
Pulse lengthAutomaticThe durati on of the air
cleaning process is
programmed in the
sensor.
The relay aut o m a tically takes over the
cleaning duration
from the sensor.
0,5 s
1 s
2 s
The relay finishes the
cleaning after the
selected interval.
3 s
Note
Settings of the cleaning proces s can be done in the Settings of sensors
and diff. sensors menu of the respective sensor.
ba76032e0101/2012
4 - 27
SettingsMIQ/CR3; DIQ/CR3
4.5.9Manual control
FunctionWith the Manual control funct ion, y ou can tes t the ope rativ eness of an
instrument that is connected to the relay. To do so, close or open the
relay manually and, while doing so, check the behavior of the
connected instrument.
The settings for other functions in the Relay function menu, as for
example Frequency controller and Pulse-width contr., are retained
while the Manual control is carried out.
4.5.10 Alarm contact
Note
The Alarm contact function is only effective with systems with alarm
function (System 2020 XT).
FunctionThe Alarm contact function triggers a relay action (close or open) if a
specified alarm event occurs. The Alarm contact function is only
available for relays that are not linked with a sensor. If necessary, an
existing link has to be deleted.
operating manual. Other possibilities of how to use the alarm function
are described there as well.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.6Setting of current outputs
The following steps describe oper ation wi th the 202 0 XT system. With
the System 182 XT, the outputs of the DIQ/CR3 can be found directly
under the Settings menu (see system operating manual of the DIQ/
S 182 XT).
1Call up the measured value display with m.
2Open the Settings menu with
3Highlight the Settings of outputs and links menu item with d
and confirm with
appears.
4Highlight the Feature column with
5Highlight a line for a curr ent output ( Cx) in the Feature col umn
with
d and confirm with g.
The Settings of outputs and links display opens.
6Highlight the Current output menu item with
with
g.
g. The Settings of outputs and links display
s.
d and confirm with g.
d and confirm
ba76032e0101/2012
Fig. 4-17 150 - Settings of outputs and links
7Select a function with d and confirm with g.
4 - 29
SettingsMIQ/CR3; DIQ/CR3
FunctionSettings
No functionThe current output is not used.
Recordersee section 4.6.1
PID controllersee section 4.6.2
Fixed current valuesee section 4.6.3
8Carry out the setti ngs for the current output with d and g.
9Highlight and confirm Save and quit with
d and g.
The new settings are stored.
4.6.1Recorder
FunctionThe measured values of the linked sensor at the current out put are set
up as current intensity in the Recorder application. The output of
measured values is laid down in the Recor der type, Start value and End value settings.
Settings
SettingSelection/ValuesExplanation
Recorder type0 to 20 mA or
4 to 20 mA
Start valuewithin the
measuring range
End value
(sensor-
Minimum spacing:
5 % of the measuring
range
dependent)
Measured variableMain variable
Adjoining variable
Main variable designates
the actual measured
parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured parameter
(e.g. temperature).
4 - 30
Attenuation0 ... 40 mA/sSpeed of change of the
output current (mA/s) in
the case of an erratic
change of the input
signal.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
I -> OFL/UFLErrorCurrent values outside
the range between Start value and End value are
considerd as errors. The
current output reacts as
specified under Behavior
at error (see below).
LimitationThe current at the output
is limited to the Start
value or End value.
Behavior at errorFixed current
value
UnchangedThe current at the output
In case of system errors
and sensor errors, and in
case of measured values
outside the range
between Star t value and
End value, the current
output supplies the
specified fixed current
value.
Possible values:
0 ... 21 mA.
remains unchanged.
ba76032e0101/2012
4 - 31
SettingsMIQ/CR3; DIQ/CR3
+++=
dt
dx
Tdtx
T
1
xKII
e
de
i
e0Controller
p
minmax
X
II
K
−
=
actualnome
xxx−=
maxControllermin
III≤≤
4.6.2PID controller
FunctionWith the PID controller function, you can use an output as a controller
output. The controller can be configured as a Proportional controller
with a switchable Integral and Derivati ve controller part ( PID controller).
The control mode of the PID controller is described by the following
equation:
with:
I
Controller
I
0
Current at the controller output at time t
Current at output if x
actual
= x
KGain
X
p
x
e
x
actual
x
nom
Proportional band
Error signal
Actual value (current measured value)
Nominal value
tiIntegral control part
tdDerivative control part
tTime
I
min
I
max
Lower current limitation
Upper current limitation
The adjustable control parameters are x
and T
(see setting table on page 4-35).
d
nominal
nominal
, I0 , Xp , I
min
, I
max
, ti
4 - 32
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
0
Imin
Imax
Current I [mA]
Measured valueNominal value
Io
Xp
By activating or deactivating the Integral (ti) and Derivative (td)
controller part, the following controller types can be configured:
Controller typetd [s]ti [s]
P controller00
PI controller01 to 9999
PD controller1 to 99990
PID controller1 to 99991 to 9999
Characteristic curve of
the proportional
controller
For a P controller, the interrelations hip between the measured value
and current I at the controller output form s the fol lowing characteristic
curve:
The proportional band Xp is limited by the measuring range of the
linked sensor. If a value is entered for the parameter Xp that is greater
than zero, the controller has a negative characteristic curve (example
Fig. 4-18). To obtain a positive characteristic curve, a negative value
must be entered for Xp .
ba76032e0101/2012
4 - 33
SettingsMIQ/CR3; DIQ/CR3
0
8
14
12
Imax
Imin
Current I [mA]
Measured value [mg/l]
428
Io
Nominal value
Xp
mg/l
mA
1
mg/l6
mA6
K==
Application example Feedback control of the D. O. concentration
Nominal value4mg/l
Xp10 % of the measuring range or 6 mg/l
Imin8mA
Imax14 mA
Io12 mA
ti0 s (no I controller part)
td0 s (no D controller part)
The control parameters form the following (negative) characteristic
curve:
Fig. 4-19 Characteristic curve for application example
4 - 34
The controller works with the following gain:
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
()
mg/lmg/l
mg/l
mA
1mA12I
Controller
54−⋅+=
()
mA11mg/l1
mg/l
mA
1mA12I
Controller
=−⋅+=
Within the proportional band, a concentration increase by 1 mg/l
causes a decrease of the current output by 1 mA. If, for example, the
measured concentration is 5 mg/l, 11 mA are output:
The highest concentration with which the controller works within the
proportional band is 8 mg/l (corresponding to Imin = 8 mA), the lowest
one is 2 mg/l (corresponding to Imax = 14 mA).
Settings
SettingSelection/ValuesExplanation
Measured variableMain variable
Adjoining variable
Nominal valuewithin the measuring
range (sensordependent)
Xp5 ... 100 %
-5 ... -100 %
of the measuring
range
Imin0 ... 20 mALower current
Main variable
designates the actual
measured parameter
of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).
Nominal value the
measured value is
regulated to
Proportional band of
the controller.
Negative values
result in a positive
characteristic curve.
limitation *
ba76032e0101/2012
Imax0 ... 20 mAUpper current
limitation *
* Note:
Spacing between
Imin and Imax:
at least 5 mA
4 - 35
SettingsMIQ/CR3; DIQ/CR3
SettingSelection/ValuesExplanation
Io0 ... 20 mACurrent value on the
output if the
measured value
equals the Nominal
value
ti0 ... 9999 sHold-back time:
Integral part of the
controller
(0 = not effective)
td0 ... 9999 sReset time :
Derivative part of the
controller
(0 = not effective)
Behavior at error
Fixed current valueIn the case of an
error, the current
output supplies the
current value defined
in the Current with error field (any in the
range 0 ... 21 mA).
UnchangedIn the case of an
error, the current on
the output remains
unchanged.
4 - 36
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.6.3Fixed current value
FunctionWith the Fixed current value function, you can test the operative ness of
the instruments connected to the outputs: output different current
values to the output and, while doing so, check the behavior of the
connected instrument.
SettingSelection/ValuesExplanation
Current outputFixed current
value
Using Save and qu it, the
nominal amperage that
was input as Inom is
output.
Inom0 ... 20 mAThe nominal amperage
that is output.
Note
The settings for other functions in the Current output menu, as for
example PID controller and Recorder, are retained while the Fixed current value is carried out.
ba76032e0101/2012
4 - 37
SettingsMIQ/CR3; DIQ/CR3
4.7Checking the status of the outputs
This function offers a simple overview of the states of all outputs of the
combi output module.
For relays, the status open or closed is displayed.
For current outputs, the current value available at the output is
displayed.
With the System 182 XT, you can recall the states of the outputs via the
measured value and status display (see system operati ng manual of
the DIQ/S 182 XT).
With the 2020 XT system, the function, 394 - Status of outpu t channels,
in the menu, Settings/Service/List of all components can be accessed
as follows:
1Call up the measured value display with m.
2Open the Settings menu with
3Highlight t he Service menu item wit h d and confirm wit h g.
The Service dialog box opens.
4Highlight the List of all components menu it em with
confirm with
The List of all components dialog box opens.
5Select the r equired component (column Model, entry MIQCR3)
with
d and confirm with g.
The 394 - Status of output channels window opens.
g.
s.
d and
4 - 38
Fig. 4-20 394 - Status of output channels
6Using m or e, exit the 394 - Status of output channels
window.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.8Behavior of linked outputs
4.8.1Behavior in case of error
For linked relay outputs or current outputs, you can determine the
behavior of the outputs in case of errors.
Depending on the use of the output, the beha vior in case of errors is set
in the following menus:
Outputmenu
Frequency controllerFrequency with error (see section 4.5.5)
Pulse-width contr.Error pulse width (see section 4.5.6)
RecorderCurrent with error (see section 4.6.1)
Error eventsThe defined behavior occurs in the case of the following events or
Freezing the states of
the outputs
Resumption of the
normal function
conditions:
The linked sensor does not supply a valid measured value
(display of Init, Error, "-----", or OFL)
Communication of the MIQ/CR3; DIQ/CR3 with the I Q SENSOR NET
controller is disturbed fo r longer than 2 minutes.
The supply voltage for the MIQ/CR3; DIQ/CR3 is too low.
In the Recorder functi on, the measured value of the li nked sensor is
outside the range between Start value and End value.
Independent of the specified behav ior at error, the followi ng conditions
lead to the freezing of the output states:
The linked sensor is in the maintenance condition
(display of Cal, Clean, or a flashing measured value).
Communication of the MIQ/CR3; DIQ/CR3 with the I Q S
ENSOR NET
controller is disturbed for a short time. After the malfunction has
lasted for 2 minutes the output changes to the specified behavior in
the case of an error.
The relay output or current output automatically returns to its normal
state as soon as all error conditions are over and all conditions causing
a freezing of outputs are cleared.
ba76032e0101/2012
4 - 39
SettingsMIQ/CR3; DIQ/CR3
4.8.2Behavior in non-operative condition
An output is non-operative when no functi on is activated for the out put.
An output becomes non-operative in case of
Power failure
(As soon as the supply voltage is sufficie nt again, the non-operati ve
condition of the outputs ends. The outputs function as speci fied by
the user again.)
Erasing a link to a sensor
Changing the Measuring mode sensor setting for a linked sensor
Changing the Measuring range sensor set ti ng for a linked sensor
Note
Before editing sensor settings a note appears on the display to inform
you that links will be erased when you change the Meas uring mode or
Measuring range sensor setting.
Settings in the non-
active condition
Current outputRelay output
Current: 0 ARelay: Open
4 - 40
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Settings
4.9Maintenance condition of sensors
The maintenance condition of sensors serves to calibrate, clean,
service and repair (remove and replace) sensors.
In the maintenance condition
the system does not react to the current measured value or the
condition of the selected sensor
linked outputs are frozen
sensor errors do not prompt changes in the conditions of linked
outputs.
The maintenance condition is automatically activat ed
during calibration. After calibration the sensor remains in the
maintenance condition until the maintenance condition is switched
off manually (see section 4.9.2)
during the cleaning cycle (see section 4.5.7)
Switch on the maintenance condition manually when you want to clean,
service or repair (remove and replace) a sensor (see section 4.9.1).
When you have finished cleaning, servicing or repairing the sensor,
switch off the maintenance condition manuall y (see section 4.9.2).
Note
When a sensor is in the maintenance condition, the measured values
or status indications of the sensor in th e measured valu e display fl ash.
ba76032e0101/2012
4 - 41
SettingsMIQ/CR3; DIQ/CR3
4.9.1Switching on the maintenance condition
1Call up the measured value display with m.
2Select the sensor you want to swit ch on the maintenance
condition for with
not flash.
3Open the 300 - Display/Opt ions menu with g.
d. The measured values of the sensor do
4Using
condition on/off or Maintenance of sens or S0x and confirm with
d, highlight the menu item, Switch maintenance
g. A window that informs you about the maintenance
condition opens up.
5Confirm Continue with g.
The selected sensor is in the maintenance condition. Link ed
outputs are frozen.
6Call up the measured value display with
The measured values of the sensor flash.
4.9.2Switching off the maintenance condition
1Call up the measured value display with m.
2Select the sensor you want to swit ch off the maintenance
condition for with
flash.
3Open the 300 - Display/Opt ions menu with g.
d. The measured values of the sensor
m.
4 - 42
4Using
condition on/off or Maintenance of sens or S0x and confirm with
d, highlight the menu item, Switch maintenance
g. A window that informs you about the maintenance
condition opens up.
5Confirm Continue with g.
The maintenance condition of the sel ect ed sensor is swit ched
off. Linked outputs are released.
6Call up the measured value display with
The measured values of the sensor do not flash.
m.
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Maintenance and cleaning
5Maintenance and cleaning
5.1Maintenance
The combi output module does not require any special maintenance
work. The general maintenance of IQ S
described in the IQ S
ENSOR NET system operating manual.
5.2Cleaning
The cleaning of IQ SENSOR NET components is described in the
IQ S
ENSOR NET system operating manual.
ENSOR NET components is
ba76032e0101/2012
5 - 1
Maintenance and cleaningMIQ/CR3; DIQ/CR3
5 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Technical data
6Technical data
Note
Electrical data
General technical data on MIQ modules are given in the T
DATA chapter of the IQ SENSOR NET system operating manual.
Nominal voltageMax. 24 VDC via the IQ SENSOR NET (for
details, see the T
the IQ S
ENSOR NET system operating
ECHNICALDATA chapter of
manual)
Power consumption3 W
Protective classII
ECHNICAL
Instrument safety
Terminal connections
Overvoltage
II
category
Applicable norms– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1
IQ SENSOR NET
connections
2
Additional connectable SENSORNET
terminator (terminating re sistor)
Number of relay
3
contacts
Number of current
3
outputs
Terminal typeScrew-type terminal strip, accessible by
opening the lid
Terminal rangesSolid wires:0.2 ... 4.0 mm
2
AWG 24 ... 12
2
2
Line cross-section of
cables carrying mains
Flexible wires:0.2 ... 2.5 mm
Europe:1.5 ... 4.0 mm
USA:AWG 14 ... 12
voltage
ba76032e0101/2012
Cable feeds4 cabl e g lands M16 x 1. 5 on t he un derside
of the module
6 - 1
Technical dataMIQ/CR3; DIQ/CR3
Relays
OutputPhysically separated from the
IQ S
ENSOR NET
Max. switching voltage250 VAC or 24 VDC
Max. switching current5 A (AC and DC)
Installation
When placing an order please have the following information availabl e:
YSI account number (if available)Name and Phone Number
Model number or brief descriptionBilling and shipping address
QuantityPurchase Order or Credit Card
:(800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
:environmental@ysi.com
1725 Brannum Lane
Yellow Springs, OH 45387
USA
:www.ysi.com
7.2Service Information
YSI has authorized service centers throughout the United States and
Internationally. For the neares t ser vice center inf ormati on, please vi si t
www.ysi.com
directly at 800-897-4151.
When returning a product for service , i nclude the Product Retur n f orm
with cleaning certification. Th e form must be completely fil led out for an
YSI Service Center to accept the instrument for service. The Product
Return form may be downloaded at www.ysi.com
‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76032e0101/2012
7 - 1
Contact InformationMIQ/CR3; DIQ/CR3
7 - 2
ba76032e0101/2012
MIQ/CR3; DIQ/CR3Indexes
8Indexes
8.1Explanation of the messages
In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S
ENSOR NET system for the MIQ/CR3 output module.
Note
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OGBOOK chapter of the IQ SENSOR NET system
operating manual.
Note
All message codes of the output module MIQ/CR3 or DIQ/CR3 end
with the number "411".
8.1.1Error messages
Message codeMessage text
EA4411Current output range undercut
* Check process
* Check settings and, if necessary, change them
EA5411Current output range exceeded
* Check process
* Check settings and, if necessary, change them
EI3411Burden resistor too large (> 500 Ohm) or current loop interrupted
* Check burden, terminal connections and connection lines
ba76032e0101/2012
8.1.2Info messages
The MIQ/CR3 or DIQ/CR3 output module does not send any info
messages.
8 - 1
IndexesMIQ/CR3; DIQ/CR3
8 - 2
ba76032e0101/2012
1725 Brannum Lane
Yellow Springs, Ohio 45387 USA
+1 937-767-7241
800-765-4974 (US)
FAX (937) 767-1058
Email: environmental@ysi.com
Internet: www.ysi.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.