YSI 600R User Manual

6-Series Sonde Quick Start Guide
is document will assist an experienced user with the steps required to make your YSI 6-Series product operational.
For full instructions, please refer to the 6-Series User Manual.
Unpacking and Inspection
Inspect the outside of the shipping container(s) for damage. If you see any damage, contact your shipping carrier immediately.
Remove the equipment from the shipping box. Some parts are loose in the box, so check the packing material carefully. Check o all items on the packing list and inspect them for damage.
Items Needed
Sonde
• Probes
• Calibration cup
• Probe guard
• Batteries
• DI water
• Flathead screwdriver
• Calibration standards
• Communication or eld cable
If using 6562 DO probe:
• 6570 maintenance kit
• 5775 membrane kit
• Razor blade
(if sonde has battery compartment)
Preparing Sonde for Use
To prepare the sonde for operation, you will need to:
• Install a new membrane on the YSI 6562 DO Probe, if you are using this item.
• Install probes (sensors) into the connectors on the sonde bulkhead.
• Provide power for the sonde through batteries or line power.
• Connect a eld cable or PC communication cable.
It is important to calibrate all
!
sensors prior to using the sonde for eld measurements. Refer to Section 2.6.2 of the 6-Series User Manual.
Installing Membrane on 6562 DO Probe
(Skip to next section if you are not using this probe)
e 6562 Rapid Pulse™ Dissolved Oxygen probe is shipped with a protective dry membrane on the sensor tip. Remove the O-ring and membrane. Handle the probe with care. It is important not to scratch or contaminate the sensor tip.
!
e DO sensor on the 600R/QS cannot be removed.
necessary, shake o electrolyte and start over.
4. Secure a membrane between your le thumb and the probe body. Touch membrane only at the ends.
5. With the thumb and forenger of your right hand, grasp the free end of the membrane and gently stretch it up, over, and down the other side of the sensor. e membrane should be stretched below the ridge where the O-ring seats so a good seal is achieved.
Open the 6562 DO Probe Kit and follow these steps:
1. Prepare the electrolyte solution as specied on the bottle.
2. Put/keep the protective cap on the connector end of the probe to protect it from contamination.
3. Hold the probe in a vertical position and apply a few drops of KCl solution to tip. e uid should completely ll the small moat around the electrodes—and create a large meniscus on the probe face. Be sure no air bubbles are stuck to the face of the sensor. If
6. Secure the end of the membrane under the forenger of your le hand.
7. Roll the O-ring over the end of the probe, being careful not to touch the membrane surface with your ngers.
8. If bubbles or wrinkles are present, remove the membrane and repeat steps 3-7.
9. Trim excess membrane with a razor blade. Rinse o any excess KCl solution, but be careful not to get any water in the connector.
Refer to Section 2.3 of the 6-Series User Manual for complete instructions and photos.
Loading...
+ 4 hidden pages