11. Accessories and Replacement Parts................................................................... 25
i
Page 4
12. Appendix A - Specifications................................................................................. 26
13. Appendix B - Health and Safety........................................................................... 27
14. Appendix C - pH Buffer Values ............................................................................ 28
ii
Page 5
1. Introduction
The YSI Model 60 Handheld pH/Temperature System is a rugged, micro-processor based, digital
meter with an attached pH and temperature probe. The pH sensor can be easily replaced in the
field.
The Model 60 has a non-detachable cable available in lengths of 10, 25, 50 or 100 feet (3, 7.6,
15.2 or 30.5 meters). The probe body has been manufactured with stainless steel to add rugged
durability and sinking weight.
The YSI Model 60 has the following features:
• Capability to measure at depths of up to 100 feet (30.5 meters)
• Microprocessor control
• Field replaceable low maintenance pH sensor
• Push-button calibration
• Simultaneous display of pH and temperature
• Automatic temperature compensation
• Data storage for 50 sets of readings with on screen recall
• Waterproof case (IP65)
The Model 60’s micro-processor allows the system to be easily calibrated with the press of a few
keys. Additionally, the micro-processor performs a self-diagnostic routine each time the
instrument is turned on. The self-diagnostic routine provides useful information about the
function of the instrument and probe.
A transport chamber, built into the instrument case, provides a convenient place to store the
probe when transporting the system. The Model 60 case is waterproof (rated to IP65) allowing
operation in the rain without damage to the instrument.
The Model 60 is powered by six AA-size alkaline batteries. A new set of alkaline batteries will
provide approximately 100 hours of continuous operation. When batteries need to be replaced,
the LCD will display a “LO BAT” message.
The YSI Model 60 is designed for use in environmental, aquaculture, and industrial applications
where accurate pH and temperature measurements are desired.
1
Page 6
2. Preparing the Meter
2.1 Unpacking
When you unpack your new YSI Model 60 Handheld pH and Temperature System for the first
time, check the packing list to make sure you have received everything you should have. If there
is anything missing or damaged, call the dealer from whom you purchased the system. If you do
not know which of our authorized dealers sold the system to you, call YSI Customer Service at
800-765-4974 or 937-767-7241, and we'll be happy to help you.
2.2 Warranty Card
Please complete the Warranty Card and return it to YSI. This will record your purchase of this
instrument in our computer system. Once your purchase is recorded, you will receive prompt,
efficient service in the event any part of your YSI Model 60 should ever need repair.
2.3 Batteries
There are a few things you must do to prepare your YSI Model 60 for use. First, locate the six
AA-size alkaline batteries and the battery cover kit which were included. Then locate the
markings inside each of the two battery-chamber sleeves that illustrate the correct way to install
the batteries. Install the batteries are shown.
NOTE: It is very important that the batteries be installed ONLY as illustrated. The instrument
will not function and may be damaged if the batteries are installed incorrectly.
Hand strap
Battery chamber
cover
Thumb screw
Attach the battery chamber cover to bottom of the instrument using the thumb screw as shown in
Figure 1. Make sure that the o-rings are in place. The battery-chamber cover is marked with the
words "OPEN" and "CLOSE."
Turn the instrument on by pressing and releasing the ON/OFF key on the front of the
instrument. The liquid crystal display (LCD) should come on. Allow a few seconds for the
instrument to complete its diagnostic routine. If the instrument does not operate, consult the
chapter entitled Troubleshooting.
Polarity marking
O-rings
Figure 1
2
Page 7
You may also want to take the instrument into a dark location and, with the instrument ON, hold
down the LIGHT key. The instrument back-light should illuminate the LCD so that the display
can be easily read.
2.4 Transport Chamber
The Model 60 has a convenient transport chamber built into the instrument’s side. This chamber
provides a storage area and protection for the pH sensor while transporting the system in the
field. Insert the round sponge (provided with the Model 60) into the bottom of the chamber. Put
6-8 drops of tap water into the sponge. The wet sponge creates a humid environment for the pH
sensor to prevent it from drying out during transport in the field (up to one week). The transport
chamber is NOT intended for long term storage of the pH sensor. See 5.3 pH Sensor Storage.
Probe transport
chamber
Figure 2
2.5 Hand Strap
The hand strap (see Figure 1 on previous page) is designed to allow comfortable operation of the
Model 60 with minimum effort. If the hand strap is adjusted correctly, it is unlikely that the
instrument will be easily dropped or bumped from your hand.
To adjust the hand strap on the back of the meter, unsnap the vinyl cover and pull the two Velcro
strips apart. Place your hand between the meter and the strap and adjust the strap length so that
your hand is snugly held in place. Press the two Velcro strips back together and snap the vinyl
cover back into place.
2.6 The Meter Case
The meter case is sealed at the factory and is not intended to be opened, except by authorized
service technicians. Do not attempt to separate the two halves of the meter case as this may
damage the instrument, break the water-resistant seal and may void the manufacturer's
warranty.
2.7 Calibration Vessels
A plastic 100 mL graduated cylinder and a plastic container are provided with the Model 60. The
graduated cylinder provides a convenient place to calibrate the pH sensor minimizing the amount
of solution needed. The plastic container should be filled with distilled water and used as a rinse
vessel while in the field. See section 4.2 Calibration for details.
3
Page 8
3. Preparing the Probe
The YSI Model 60 is shipped without the pH sensor installed. The pH sensor must be installed
before using the system (see section 3.1 Installing the Sensor, below). The sensor is shipped with
a protective bottle filled with a mixture of pH 4 buffer and KCl solution. Do not remove the
bottle until you are ready to use the instrument. Save the bottle for long term storage of the
probe.
3.1 Installing the Sensor
A pH sensor is included with the Model 60. Install the pH sensor as follows:
1. Remove the sensor from its protective packing.
2. Insert the pH sensor into the probe body (be sure to align the tabs on the sensor with the slots in
the probe body) and twist 1/4 turn to lock in place. See Figure 3.
NOTE: Once installed, leave the pH sensor attached to the probe until replacement is needed.
3. Carefully remove the protective bottle (containing pH 4 buffer/KCl solution) from the sensor.
Save the bottle and solution for long term (over one week) storage of the sensor. Seal the storage
bottle with the cap provided.
4. Rinse the sensor tip with distilled or deionized water.
5. Calibrate the system before use. See section 4.2 Calibration.
Align tabs with
slots in probe
body
Turn sensor 1/4 turn
to lock in place
Insert pH sensor
into probe body
Figure 3
4
Page 9
4. Operation
The following diagram is an overview of the operation of the Model 60. See the following
sections for details of operation.
Figure 4
5
Page 10
4.1 Turning The Instrument On
With the batteries installed correctly, press the ON/OFF key. The instrument will activate all
segments of the display for a few seconds, which will be followed by a self test procedure which
will last for several more seconds. During this power on self test sequence, the instrument’s
microprocessor is verifying that the system is working properly. If the instrument were to detect
a problem, a continuous error message would be displayed. See the chapter entitled
Troubleshooting for a list of error messages.
7.23 pH
24.8
°C
4.2 Calibration
The YSI Model 60 MUST be calibrated before making pH measurements. Calibration may be
performed at 1, 2 or 3-points (at pH 7, 4 and 10, or at pH 6.86, 4.01 and 9.18). Perform a 1-point
calibration (at pH 7 or at pH 6.86) ONLY if a previous 2 or 3-point calibration has been
performed recently. In most cases, a 2-point pH calibration will be sufficient for accurate pH
measurements, but if the general range of pH in the sample is not known, a 3-point calibration
may be necessary. 3-point calibration assures accurate pH readings regardless of the pH value of
the sample. See Section 8.1 pH for more details.
WARNING: Calibration reagents may be hazardous to your health. Refer to Appendix B -
Health and Safety for more information.
Before calibrating the YSI Model 60, complete the procedures discussed in the Preparing
the Meter and Preparing the Probe chapters of this manual.
The user can choose from two sets of pH buffer values for 3-point calibration. The first set
consists of the standard YSI pH buffer values of pH 7 (YSI 3822), pH 4 (YSI 3821) and pH 10
(YSI 3823). The second set available is the NIST pH 6.86, 4.01 and 9.18. Note that the first calibration point must be either pH 7 or pH 6.86. Calibration is performed as follows:
1. Turn the instrument on by pressing the ON/OFF key. If the instrument was already on, press
the MODE key until pH is displayed.
2. Rinse the probe with deionized or distilled water, then carefully dry the probe (or rinse it
with some of the pH buffer solution to be used for calibration).
3. Place 30 to 35 mL of the pH buffer you have chosen to calibrate the system with (pH 7 or
6.86) in the 100 mL graduated cylinder (provided with the Model 60). The graduated
cylinder minimizes the amount of solution needed. Immerse the probe making sure that both
the pH and temperature sensors are covered by the solution (see Figure 5 on the following
page).
6
Page 11
For best results:
• Calibrate as close as possible to the sample temperature.
• After storage in pH 4 buffer/KCl solution, place the pH sensor in pH 7 (6.86) buffer and
allow to acclimate before calibrating (5 to 10 minutes).
•Always give the pH and temperature sensors enough time to equilibrate with the
temperature of the buffer.
Temperature sensor
pH sensor (covered by sensor guard)
Figure 5
4. To enter the calibration menu, use two fingers to press and release both the UP ARROW
and DOWN ARROW keys at the same time. The Model 60 display will show CAL at the
bottom, STAND will be flashing and the main display will show 7.00 (the buffer to be used
to adjust the offset).
First buffer value.
Flashing
7.00 pH
STAND CAL 24.8
°C
Used to adjust the offset.
NOTE: If you will be calibrating with pH buffers of 6.86, 4.01 and 9.18 (instead of 7, 4 and
10), press both the UP ARROW and DOWN ARROW keys again. The display will change
to 6.86.
NOTE: The Model 60 automatically accounts for the fact that the true pH of the buffers
changes with temperature, therefore, the pH values displayed during calibration will vary
with temperature. For example, pH 7 buffer at 20°C (rather than 25°C) has an actual pH of
7.02 and this number (rather than 7.00) will appear on the display when the probe is placed
in the solution. See Appendix C - pH Buffer Values.
7
Page 12
5. Press the ENTER key. The Model 60 display will show CAL at the bottom, STAND will
g
stop flashing and the pH calibration value is shown with the middle decimal point flashing.
7.00 pH
STAND CAL 24.8
6. When the reading is stable (does not change by 0.01 pH in 10 seconds), the decimal point
will stop flashing. Press and hold the ENTER key to save the calibration point. The Model
60 will flash SAVE on the display along with OFS to indicate that the offset value has been
saved.
°C
Flashes until
readin
is stable
SAVE
OFS
7. SLOPE will now appear on the display and be flashing. This indicates that the slope is ready
to be set using a second pH buffer. The system is now calibrated at a single point. If you are
only performing a single point calibration, press the MODE key to return to normal
operation.
7.00 pH
SLOPE
CAL 24.8
8. Rinse the probe with deionized or distilled water, then carefully dry the probe.
STOP HERE IF PERFORMING A 1-POINT CALIBRATION.
9. If you are performing a 2-point or 3-point calibration, fill a clean container with the second
value pH buffer (pH 4 or 10, or pH 4.01 or 9.18) and immerse the probe into the solution.
Make sure that the temperature sensor is immersed.
10. Press the ENTER key. The Model 60 display should now show CAL at the bottom, SLOPE
will stop flashing and the pH calibration value (automatically sensed by the instrument) is
shown with one of the decimal points flashing.
°C
Flashing
8
Page 13
.4.01 pH
Second buffer value
SLOPE CAL 24.8
°C
Left decimal flashes (until
reading is stable) for buffers
lower than first cal point
If the second pH buffer is less than the first buffer (which was used to adjust the offset; pH 7
or pH 6.86), the left decimal point will flash as shown above. If the second pH buffer is
greater than the first, the right decimal point will flash as shown below.
Second buffer value
10.0.0 pH
SLOPE CAL 24.8
°C
Right decimal flashes (until
reading is stable) for buffers
higher than first cal point
11. When the reading is stable (does not change by 0.01 pH in 10 seconds), the decimal point
will stop flashing. Press and hold the ENTER key to save the first SLOPE. The Model 60
will flash SAVE on the display along with SLP to indicate that the first slope value has been
saved.
SAVE
SLP
12. SLOPE will start flashing again indicating that the slope is ready to be set using a third pH
buffer.
.4.01 pH
SLOPE CAL 24.8
°C
13. The system is now calibrated at two points. If you are only performing a two point
calibration, press the MODE key to return to normal operation.
14. Rinse the probe with deionized or distilled water, then carefully dry the probe.
Flashing
STOP HERE IF PERFORMING A 2-POINT CALIBRATION.
9
Page 14
15. If you are performing a 3-point calibration, fill a clean container with the third value pH
p
buffer (pH 4 or 10, or pH 4.01 or 9.18) and immerse the probe into the solution. Make sure
that the temperature sensor is immersed.
NOTE: The third buffer must not be the same as the second buffer. For example; if the
second buffer was less than pH 7, the third buffer must be greater than pH 7.
16. Press the ENTER key. The Model 60 display will now show CAL at the bottom, SLOPE
will stop flashing and the pH calibration value (automatically sensed by the instrument) is
shown with one of the decimal points flashing. If the third pH buffer is less than the first
buffer (which was used to adjust the offset; usually pH 7), the left decimal point will flash. If
the third pH buffer is greater than the first, the right decimal point will flash.
Third buffer value
1.0.0.0 pH
SLOPE
CAL 24.8
or
°C
Right decimal flashes (until
reading is stable) for buffers
higher than first cal point
Third buffer value
.4.0.1 pH
SLOPE CAL 24.8
17. When the reading is stable (does not change by 0.01 pH in 10 seconds), the decimal point
will stop flashing. Press and hold the ENTER key to save the second SLOPE. The Model 60
will flash SAVE on the display along with SLP to indicate that the second slope value has
been saved.
°C
Left decimal flashes (until
reading is stable) for buffers
lower than first cal
oint
SAVE
SLP
The system is now calibrated at three points and will return to normal operation.
18. Rinse the probe with deionized or distilled water.
10
Page 15
4.3 Making Measurements
After the system has been set-up and pH has been calibrated as described in 4.2 Calibration, it is
ready to make measurements. Simply insert the probe into the sample, shake gently to remove
any trapped air bubbles and wait for the readings to stabilize (approximately 60 seconds). The
first reading after storage in buffers may take longer to stabilize (5 to 10 minutes), therefore, the
probe should be stored in the transport chamber when making field measurements. It is
important that the probe be inserted into the sample far enough so that both the pH sensor and
temperature sensor are covered by the liquid (see Figure 5 on page 7).
The Model 60 has three modes:
¾ pH -- Displays pH and temperature (°C)
¾ Recall -- Allows previously stored data to be displayed.
¾ Erase all -- Allows ALL previously stored data to be deleted.
To change between the Model 60 modes, simply press and release the MODE key. The Model
60 will cycle through the modes as follows:
pH and
Temperature °C
NOTE: When the Model 60 is first turned on, it is in the pH mode.
RecallErase all
4.4 Saving Data
The Model 60 is equipped with non-volatile memory that is capable of storing up to 50 different
sets of readings. Non-volatile means that you do not need to worry that your data will be lost due
to a power failure or interruption, such as when the batteries are removed. Each set consists of
pH and temperature. The Model 60 will also assign a site identity number to each set of readings
to allow easy review of the data. This feature is useful in situations where transcribing data is
difficult or not available.
SAVE
01
Site identity
While pH is displayed on the screen, press the ENTER key and hold it for approximately 2
seconds. The meter will flash SAVE on the display along with the current site identity (1
through 50) being used.
When all 50 sites are full, the display will flash FULL on the screen. This message will remain
on the screen (even after power down) until a key is pushed.
11
Page 16
FULL
Once you have acknowledged the memory is full, any subsequent saved data will begin
overwriting existing data starting with site #1. No additional warning will be displayed.
4.5 Recalling Stored Data
1. To put the Model 60 into the RECALL mode, press the MODE key until “rcl” is displayed
on the screen along with the site ID number in the lower right corner.
rcl
01
2. Press the ENTER key to review the last set of data that was saved. The Model 60 will
display the pH and temperature.
3. Press the UP ARROW key to move up through the saved sets of data.
4. Press the DOWN ARROW key to move down through the saved sets of data.
5. When the correct Site ID# is displayed, press the ENTER key to display the data.
6. When you have finished recalling data, press MODE two times to return to normal
operation.
NOTE: The Model 60 will recall data as a list. When the UP ARROW is pressed the Model 60
will display the Site ID# for the previously recorded data. For example: If you are reviewing Site
ID# 5 and the UP ARROW is pressed, the Model 60 will display Site ID#4. If you are reviewing
Site ID# 5 and Site ID# 5 was the last set of data stored, the DOWN ARROW key will display
Site ID# 1.
Here is an example of the Model 60 memory.
Site ID #1
Site ID #2
Site ID #3 If the UP ARROW key was pressed the Model 60 would display Site ID #2
Site ID #4
Site ID #5
Site identity
12
Page 17
4.6 Erasing Stored Data
1. To erase the data that is stored in the Model 60’s memory, press the MODE key until the
Model 60 displays ErAS on the screen.
2. Press and hold the DOWN ARROW and ENTER keys simultaneously for approximately 5
seconds.
ErAS
3. Successful erasure is indicated by the Model 60 displaying DONE on the display for 1 to 2
seconds.
dOnE
The instrument will automatically change to normal operation after completion and the next
saved data will be stored in site ID# 1.
IMPORTANT: Data in all 50 site ID’s will be erased completely and will be lost forever. Do
not use the erase function until all recorded data has been transcribed to an archive outside the
Model 60.
4.7 Display Backlight
At times it may be necessary to take measurements with the Model 60 in dark or poorly lit areas.
To help in this situation, the Model 60 comes equipped with a backlight which will illuminate
the display so that it can be easily read. To activate the backlight, press and hold the LIGHT
key. The display will remain lit as long as the key is pressed. When you release it, the light goes
out to preserve battery life.
13
Page 18
5. Maintenance of the pH Sensor
5.1 pH Sensor Precautions
1. When making measurements or performing the calibration procedure, make certain that the
level of sample or pH buffer is high enough to cover both the pH and temperature sensors.
2. Rinse the probe with deionized water between changes of calibration buffer solutions.
3. During pH calibration, allow the sensors time to stabilize with regard to temperature
(approximately 60 seconds) before proceeding with the calibration protocol. The pH readings
after calibration are only as good as the calibration itself.
4. Clean and store the probe according to the instructions found below.
5.2 pH Sensor Cleaning
Cleaning is required whenever deposits or contaminants appear on the glass pH sensor. Unscrew
and remove the small guard that protects the pH sensor. Use tap water and a clean cloth or lens
cleaning tissue to remove all foreign material from the glass sensor.
If good pH response is not restored by the above procedure, perform the following additional
procedure:
1. Soak the probe for 10 to 15 minutes in clean water containing a few drops of commercial
dishwashing liquid.
2. GENTLY clean the glass bulb by rubbing with a cotton swab soaked in the cleaning solution.
3. Rinse the probe in clean water, wipe with a cotton swab saturated with clean water, and then
rerinse with clean water.
If good pH response is still not restored by the above procedure, perform the following
additional procedure:
1. Soak the probe for 15 to 30 seconds in one molar (1 M) hydrochloric acid (HCl).
2. GENTLY clean the glass bulb by rubbing with a cotton swab soaked in the acid.
3. Rinse the probe in clean water, wipe with a cotton swab saturated with clean water, and then
rerinse with clean water.
4. Reinstall the small guard that protects the pH sensor.
If biological contamination of the reference junction is suspected or if good response is not
restored by the above procedures, perform the following additional cleaning step:
1. Soak the probe for approximately 1 hour in a 1 to 1 dilution of commercially-available
chlorine bleach.
2. Rinse the probe with clean water and then soak for 1 hour in clean water to remove residual
bleach from the junction.
14
Page 19
5.3 pH Sensor Storage
For short term storage between measurements in the field (up to one week), place the probe in
the transport chamber in the side of the instrument case. Make sure that the sponge inside the
chamber is wet (tap water).
For long term storage (over one week), place the probe in the storage bottle (provided)
containing a mixture of 50% pH 4 buffer and 50% 1.5M KCl. This will assure the fastest
possible pH response. If this mixture is not available, storage in tap water is the next best choice.
Do NOT store the probe dry or in distilled or deionized water.
NOTE: After storage in the pH 4/KCl solution described above, place the probe in the transport
chamber in the side of the instrument case or soak the probe in pH 7 buffer for 5 to 10 minutes
allowing it to acclimate before calibrating.
If the probe has been inadvertently left in air and the reference electrode junction has dried out,
good function can usually be restored by soaking the probe in the pH 4/KCl solution described
above.
15
Page 20
6. Discussion of Measurement Errors
There are two basic types of pH errors. The first type are errors related to limitations of
instrument design and tolerances of components. The second type are errors due to basic sensor
accuracy tolerances, mainly background signal, linearity, and variations in temperature
coefficient. It is unlikely that the actual error in any measurement will be the maximum possible
error.
Errors
• Component and circuitry error: ±0.03 pH
• pH error caused by sensor accuracy and temperature compensation:
±0.1 pH for measurements at 10°C from calibration temperature
±0.2 pH for measurements at 20°C from calibration temperature
16
Page 21
7. Troubleshooting
Error Messages
The instrument performs a Power On Self Test each time it is turned on. The following error
messages are provided to facilitate troubleshooting. They appear on the LCD when an error is
detected.
Symptom Possible Cause Action
1. Instrument will not turn on
16. Instrument "locks up"
16. Instrument will not calibrate due to
unstable readings (decimal point
keeps flashing)
16. pH readings are inaccurate
16. LCD displays "LO BAT"
16. Main Display reads “nOnE”
16. Main Display reads “OVEr”
16. Main Display reads “undr”
16. Main Display reads “OVEr”
(Secondary display reads “ovr”)
16. Main Display reads “undr”
(Secondary display reads “udr”)
• Low battery voltage
• Batteries installed wrong
• Meter requires service
• Instrument has received a shock
• Batteries are low or damaged
• System requires service
• pH sensor is fouled
• pH sensor is bad
• System requires service
• pH buffers out of spec
• Cal procedure not correct
• Sample temperature is over 20°C
from cal temperature
• pH sensor is fouled or damaged
• pH Sensor is bad
• System requires service
• Batteries are low or damaged • Replace batteries (pg 2)
• During recall, no data is currently
stored in memory.
•When calibrating, pH level is over
range allowed for the buffer value
selected.
• When measuring, pH level is > 14
• When calibrating, pH level is under
range allowed for the buffer value
selected.
• When measuring, pH level is < 0
• Temperature reading is >75°C
• Temperature reading is <-5°C
• Replace batteries (pg 2)
• Check battery polarity (pg 2)
• Return system for service (pg 21)
• Remove battery lid, wait 15 seconds
for reset, replace lid. (pg 2)
• Replace batteries (pg 2)
• Return system for service (pg 21)
• Clean pH sensor (pg 14)
• Replace pH sensor (pg 4, 25)
• Return system for service (pg 21)
• Recalibrate with known good
standards (pg 6)
•Calibrate within ±20°C of sample
temp (±10°C for best results)
• Clean pH sensor (pg 14)
• Replace pH sensor (pg 4, 25)
• Return system for service (pg 21)
• Store data before attempting to recall
(pg 11)
•Recalibrate with known good
standards (pg 6)
• Clean pH sensor (pg 14)
• Replace pH sensor (pg 4, 25)
• Return system for service (pg 21)
• Recalibrate with known good
standards (pg 6)
• Clean pH sensor (pg 14)
• Replace pH sensor (pg 4, 25)
• Return system for service (pg 21)
• Measure samples at a temperature
within the range of the system.
•Measure samples at a temperature
within the range of the system.
17
Page 22
Symptom Possible Cause Action
11. Main display reads “PErr”
12. Main display reads “Err”
(Secondary display reads “ra”)
13. Main display reads “Err”
(Secondary display reads “ro”)
16. Main display reads “FAIL”
(Secondary display reads “eep”)
16. Readings on main display don’t
change
•Incorrect sequence of keystrokes. •Refer to manual section for step by
step instruction for the function you
are attempting.
•System has failed its RAM test
check procedure.
•System has failed its ROM test
check procedure.
•EEPROM has failed to respond in
time.
• Meter is in recall mode. • Press MODE key to return to Normal
• Turn instrument OFF and back ON
again.
• Return the system for service (pg 21)
• Turn instrument OFF and back ON
again.
• Return the system for service (pg 21)
• Return the system for service (pg 21)
Operation (pg 5, 11)
18
Page 23
8. Principles of Operation
8.1 pH
The YSI Model 60employs a field replaceable pH sensor for the determination of hydrogen ion
concentration. The sensor is a combination electrode consisting of a proton selective glass
reservoir filled with buffer at approximately pH 7 and a Ag/AgCl reference electrode which
utilizes gelled electrolyte. A silver wire coated with AgCl is immersed in the buffer reservoir.
Protons (H+ ions) on both sides of the glass (media and buffer reservoir) selectively interact with
the glass, setting up a potential gradient across the glass membrane. Since the hydrogen ion
concentration in the internal buffer solution is invariant, this potential difference, determined
relative to the Ag/AgCl reference electrode, is proportional to the pH of the media.
Our testing of the Model 60 pH sensor indicates that it should provide long life, good response
time and accurate readings in most environmental waters, including fresh water of low ionic
strength. No special sensor is required (nor offered) for water of low conductivity.
Calibration And Effect Of Temperature
The software of the YSI Model 60
calculates pH from the established linear relationship between
pH and the millivolt output as defined by a variation of the Nernst equation:
E = E
+ 2.3RT * pH where E = millivolts output
o
nF
E
= a constant associated with the reference electrode
o
T = temperature of measurement in degrees Kelvin
R, n, and F are invariant constants
Thus, in simplified y = mx + b form, it is (mv output) = (slope)x(pH) + (intercept). In order to
quantify this simple relationship, the instrument must be calibrated properly using buffers of
known pH values. In this procedure, the millivolt values for two standard buffer solutions are
experimentally established and used by the YSI Model 60
software to calculate the slope and
intercept of the plot of millivolts vs. pH. Once this calibration procedure has been carried out,
the millivolt output of the probe in any media can readily be converted by the YSI Model 60
software into a pH value, as long as the calibration and the reading are carried out at the same temperature. This last qualifier is almost never met in actual environmental measurements, thus,
a mechanism must be in place to compensate for temperature or, in other words, to accurately
convert the slope and intercept of the plot of pH vs. millivolts established at T
calibration) into a slope and intercept at T
(temperature of measurement). Fortunately, the
m
(temperature of
c
Nernst equation provides a basis for this conversion.
19
Page 24
According to the Nernst equation as shown above, the slope of the plot of pH vs. millivolts is
directly proportional to the absolute temperature in degrees Kelvin. Thus, if the slope of the plot
is experimentally determined to be 59 mv/pH unit at 298 K (25 C), then the slope of the plot at
313 K (40 C) must be (313/298) * 59 = 62 mv/pH unit. At 283 K (10 C), the slope is calculated
to be 56 mv/pH unit ((283/298) * 59). Determination of the slope of pH vs. mv plots at
temperatures different from T
is thus relatively simple. In order to establish the intercept of the
c
new plot, the point where plots of pH vs. mv at different temperatures intersect (the isopotential
point) must be known. Using standard pH determination protocol, the YSI Model 60 software
assigns the isopotential point as the mv reading at pH 7 and then calculates the intercept using
this assumption. Once the slope and intercept to the plot of pH vs. mv are assigned at the new
temperature, the calculation of pH under the new temperature conditions is straightforward, and
is automatically carried out by the software.
Number of Calibration Points
When calibrating the YSI Model 60, you have the choice of 1-point 2-point, or 3-point
calibration. Perform a 2 or 3 point calibration at least once per day for accurate results.
Select the 1-point option only if you are adjusting a previous calibration. If a 2-point or 3-point
calibration has been performed previously (at least once per day), you can adjust the calibration
by carrying out a 1-point calibration at pH 7 (or pH 6.86). This calibration procedure adjusts
only the pH offset and leaves the previously-determined slope unaltered.
Select the 2-point option to calibrate the pH probe using only two calibration standards. In this
procedure, the pH sensor is calibrated using a pH 7 (or pH 6.86) buffer and one additional
buffer. A two point calibration procedure (as opposed to a 3-point procedure) can save time if
the pH of the sample is known to be either basic or acidic. For example, if the pH of a sample is
known to vary between 5.5 and 7, a two point calibration with pH 7 and pH 4 buffers is
appropriate. Three point calibration with an additional pH 10 buffer will not increase the
accuracy of this measurement since the pH is not within this higher range.
Select the 3-point option to calibrate the pH probe using three calibration solutions. In this
procedure, the pH sensor is calibrated with a pH 7 (or pH 6.86) buffer and two additional
buffers. The 3-point calibration method assures maximum accuracy when the pH of the media to
be monitored cannot be anticipated.
8.2 Temperature
The YSI Model 60 system utilizes a thermistor which changes predictably in resistance with
temperature variation. The algorithm for conversion of resistance to temperature is built-in to the
Model 60 software, and accurate temperature readings, in degrees Celsius, are provided
automatically. No calibration or maintenance of the temperature sensor is required.
20
Page 25
9. Warranty and Repair
YSI Model 60 pH Meters are warranted for two years from date of purchase by the end user
against defects in materials and workmanship. YSI Model 60 probes, cables and sensors are
warranted for one year from date of purchase by the end user against defects in material and
workmanship. Breakage of pH sensors is NOT covered under warranty. Within the warranty
period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI
determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer
Service in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid,
to the Authorized Service Center selected by YSI. Repair or replacement will be made and the
product returned, transportation prepaid. Repaired or replaced products are warranted for the
balance of the original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by (i) failure to
install, operate or use the product in accordance with YSI’s written instructions, (ii) abuse or
misuse of the product, (iii) failure to maintain the product in accordance with YSI’s written
instructions or standard industry procedure, (iv) any improper repairs to the product, (v) use by
you of defective or improper components or parts in servicing or repairing the product, or (vi)
modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO
REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE
AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS
WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE
PRODUCT COVERED BY THIS WARRANTY.
YSI Authorized Service Centers
Please visit www.ysi.com or contact YSI Technical Support for the nearest authorized service center.
NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic
materials must be cleaned and disinfected. Biological contamination is presumed for any
instrument, probe, or other device that has been used with body fluids or tissues, or with waste
water. Radioactive contamination is presumed for any instrument, probe or other device that has
been used near any radioactive source.
If an instrument, probe, or other part is returned or presented for service without a Cleaning
Certificate, and if in our opinion it represents a potential biological or radioactive hazard, our
service personnel reserve the right to withhold service until appropriate cleaning,
decontamination, and certification has been completed. We will contact the sender for
instructions as to the disposition of the equipment. Disposition costs will be the responsibility of
the sender.
When service is required, either at the user's facility or at YSI, the following steps must be taken
to insure the safety of our service personnel.
1. In a manner appropriate to each device, decontaminate all exposed surfaces, including any
containers. 70% isopropyl alcohol or a solution of 1/4 cup bleach to 1 gallon tap water are
suitable for most disinfecting. Instruments used with waste water may be disinfected with
.5% Lysol if this is more convenient to the user.
2. The user shall take normal precautions to prevent radioactive contamination and must use
appropriate decontamination procedures should exposure occur.
3. If exposure has occurred, the customer must certify that decontamination has been
accomplished and that no radioactivity is detectable by survey equipment.
4. Any product being returned to the YSI Repair Center, should be packed securely to prevent
damage.
5. Cleaning must be completed and certified on any product before returning it to YSI.
22
Page 27
9.2 Packing Instructions
1. Clean and decontaminate items to insure the safety of the handler.
2. Complete and include the Cleaning Certificate.
3. Place the product in a plastic bag to keep out dirt and packing material.
4. Use a large carton, preferably the original, and surround the product completely with packing
material.
5. Insure for the replacement value of the product.
Cleaning Certificate
Organization
Department
Address
City _______________ State ______ Zip
Country __________________ Phone
Model No. of Device ______ Lot Number
Contaminant (if known)
Cleaning Agent(s) used
Radioactive Decontamination Certified?
(Answer only if there has been radioactive exposure)
___ Yes ___ No
Cleaning Certified By
Name Date
23
Page 28
10. Required Notice
The Federal Communications Commission defines this product as a computing device and
requires the following notice:
This equipment generates and uses radio frequency energy and if not installed and used properly,
may cause interference to radio and television reception. There is no guarantee that interference
will not occur in a particular installation. If this equipment does cause interference to radio or
television reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
• re-orient the receiving antenna
• relocate the computer with respect to the receiver
• move the computer away from the receiver
• plug the computer into a different outlet so that the computer and receiver are on
different branch circuits.
If necessary, the user should consult the dealer or an experienced radio/television technician for
additional suggestions. The user may find the following booklet, prepared by the Federal
Communications Commission, helpful: "How to Identify and Resolve Radio-TV Interference
Problems." This booklet is available from the U.S. Government Printing Office, Washington, DC
20402, Stock No. 0004-000-00345-4.
24
Page 29
11. Accessories and Replacement Parts
The following parts and accessories are available from YSI or any Franchise Dealer authorized
by YSI.
CAUTION - Avoid inhalation, skin contact, eye contact or ingestion. May
affect mucous membranes.
Inhalation may cause severe irritation and be harmful. Skin contact may cause irritation;
prolonged or repeated exposure may cause Dermatitis. Eye contact may cause irritation or
conjunctivitis. Ingestion may cause nausea, vomiting and diarrhea.
FIRST AID:
INHALATION - Remove victim from exposure area to fresh air immediately. If
breathing has stopped, give artificial respiration. Keep victim warm and at rest.
Seek medical attention immediately.
SKIN CONTACT - Remove contaminated clothing immediately. Wash affected
area with soap or mild detergent and large amounts of water (approx. 15-20
minutes). Seek medical attention immediately.
EYE CONTACT - Wash eyes immediately with large amounts of water (approx.
15-20 minutes), occasionally lifting upper and lower lids. Seek medical attention
immediately.
INGESTION - If victim is conscious, immediately give 2 to 4 glasses of water and
induce vomiting by touching finger to back of throat. Seek medical attention
immediately.
The following table lists the values of NIST pH buffer solutions at various temperatures.
Temperature pH 4.01 pH 6.86 pH 9.18
0°C 4.005 6.984 9.463
5°C 4.003 6.950 9.395
10°C 4.001 6.924 9.333
15°C 4.002 6.899 9.277
20°C 4.003 6.879 9.226
25°C 4.005 6.863 9.180
30°C 4.010 6.852 9.139
35°C 4.020 6.844 9.102
37°C 4.025 6.842 N/A
40°C 4.033 6.840 9.070
45°C 4.047 6.837 9.042
50°C 4.061 6.836 9.018
28
Page 33
1725 Brannum Lane
Yellow Springs, Ohio 45387 USA
937 767-7241 • 800 765-4974 • Fax 937 767-9353 A31136A - Web
environmental@ysi.com • www.YSI.com January 07