YSI 5905 Operating Manual

YSI 5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual
TABLE OF CONTENTS
I
OPERATING PRINCIPLES .......................................................2
USING THE PROBE ............................................................. 3
Changing the Membrane Cap....................................................3
Warm Up.....................................................................................5
Electrode Cleaning...................................................................... 5
OPERATING PROCEDURES AND PRECAUTIONS ............................ 7
Air Calibration ............................................................................7
Checking the Probe Zero............................................................ 7
Membrane Life............................................................................ 7
Interferences................................................................................ 8
Acids ............................................................................................ 8
Erroneous Readings.................................................................... 8
Storage......................................................................................... 8
PROBLEM SOLVING ............................................................ 9
Trouble Shooting Guide.............................................................9
SERVICING..................................................................... 11
WARRANTY AND REPAIR ..................................................... 13
Authorized Service Centers...................................................... 14
Cleaning Instructions ...............................................................14
Packing Instructions................................................................. 15
PROBE SPECIFICATIONS .....................................................17
APPENDIX A ................................................................... 18
APPENDIX B ...................................................................19
i
INTRODUCTION
The YSI Model 5905/5010/200-BOD/Pro-BOD (Biochemical Oxygen Demand) probe is used for measuring dissolved oxygen in all popular size BOD bottles. These probes feature self stirring, easily replaceable membrane caps, and a refurbishable electrode system.
The 5905 probe is designed for use with all YSI grey-box-style dissolved oxygen meters, except for the Model 55 or Model 550A. It must be used with an adapter (5011) in order to be used on the Model 5000 and 5100 bench top instruments.
The 5010 probe is specifically designed for use with the Model 5000 and 5100 only.
The 200-BOD probe is specifically designed for use with the Model DO200 only.
The Pro-BOD probe is specifically designed for use with the Pro20 and Professional Plus instruments.
This manual also applies to probes 5905-J, 5905-K, 5905-L, 5905­W (Appendix A), and probes 5010-J, 5010-K, 5010-L, and 5010­W (Appendix A)
---IMPORTANT---
Intended Use
The 5905/5010/200-BOD/Pro-BOD probe is intended for laboratory use, not field use. It is not waterproof and should never be immersed past the taper on the stem of the probe.
1
OPERATING PRINCIPLES
The 5905/5010/200-BOD/Pro-BOD probe is a polarographic dissolved oxygen sensor. An oxygen permeable membrane covers an electrolytic cell consisting of a cathode and an anode. This membrane acts as a diffusion barrier and an isolation barrier preventing the fouling of the cathode surface by impurities in the environment. The cathode is the gold electrode, and the anode is the silver electrode which completes the electrolytic cell and acts as a reference electrode.
Upon entering the cell through the membrane, oxygen is reduced at an applied potential of -0.8 V referenced to the silver electrode. The reduction current at the cathode is directly proportional to the partial pressure of oxygen in liquid (expressed as %-air saturation) which is also proportional to the concentration of dissolved oxygen (in mg/L) at a particular temperature. Thus the same partial pressure of oxygen (%-air saturation) in liquid gives different concentrations of dissolved oxygen (mg/L) at different temperatures because of the different solubilities of oxygen at different temperatures.
2
USING THE PROBE
The 5905/200-BOD/Pro-BOD must be plugged into a YSI meter and the stirrer power plug must be plugged into its power adapter. The 5010 must be plugged into a YSI meter only, as it gets its stirring power from the 5000/5100 instrument.
Changing the Membrane Cap
---IMPORTANT---
New Probe Use
New probes are shipped with a dry membrane cap to protect the electrodes. A new membrane cap must be installed before first use. For the 5905/5010, use a Teflon membrane from a 5906 Membrane Kit. For the 200-BOD/Pro-BOD, use a PE membrane from a 5908 Membrane Kit.
Follow these instructions to change membrane caps (Figure 1):
1. Remove the stir paddle from the probe by pulling it straight out.
2. Unscrew the old membrane cap from the probe. Before installing a new membrane, clean the probe tip with deionized water in order to remove any contaminants.
3. Hold the membrane cap and fill it at least ½ full with the electrolyte solution provided. Screw the membrane cap onto the probe moderately tight. A small amount of electrolyte should overflow.
4. Rinse off excess electrolyte from the probe with deionized water.
5. Reinstall the stir paddle.
3
Remove
Stir Paddle
Fill Membrane Cap
with Solution
Unscrew Cap
Replace
Stir Paddle
Screw Cap on
moderately tight
Figure 1: Membrane Installation on the 5905/5010/200-BOD/Pro­BOD .
4
Warm Up
Every time the instrument is turned on, the 5905/5010/200­BOD/Pro-BOD Probe must be allowed to warm up for about 10­20 minutes before taking the first reading. This allows the probe to use up all the oxygen retained in the cap.
---IMPORTANT---
New Probe Break In (motor off)
For best results, please allow the probe more than the usual 20 minutes to warm up after the first membrane is installed. The process may be faster if the probe is allowed to warm up in a BOD bottle filled with clean water. If possible, it’s a good idea to let new probes warm up overnight (with a new membrane and KCl) prior to first use.
Electrode Cleaning
The 5905/5010/200-BOD/Pro-BOD should be cleaned only when erratic readings occur or after about every 500 hours of use (two months). Each cleaning removes material and reduces the life of the probe, so excessive cleaning should be avoided.
Gold Cathode
For correct probe operation, the gold cathode must be textured properly (for its location, see figure 2). It can become tarnished or plated with silver after extended use. The gold cathode can be cleaned by using the sanding disc provided in the 5906 and 5908 Membrane Kits. Place the disc in the palm of your hand, then sand the gold with a twisting motion about 3 times or until all silver deposits are removed and the gold appears to have a matte finish. If the cathode remains tarnished, return the probe for service.
5
Loading...
+ 15 hidden pages