YSI 5905 Operating Manual

Page 1
YSI 5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual
Page 2
TABLE OF CONTENTS
I
OPERATING PRINCIPLES .......................................................2
USING THE PROBE ............................................................. 3
Changing the Membrane Cap....................................................3
Warm Up.....................................................................................5
Electrode Cleaning...................................................................... 5
OPERATING PROCEDURES AND PRECAUTIONS ............................ 7
Air Calibration ............................................................................7
Checking the Probe Zero............................................................ 7
Membrane Life............................................................................ 7
Interferences................................................................................ 8
Acids ............................................................................................ 8
Erroneous Readings.................................................................... 8
Storage......................................................................................... 8
PROBLEM SOLVING ............................................................ 9
Trouble Shooting Guide.............................................................9
SERVICING..................................................................... 11
WARRANTY AND REPAIR ..................................................... 13
Authorized Service Centers...................................................... 14
Cleaning Instructions ...............................................................14
Packing Instructions................................................................. 15
PROBE SPECIFICATIONS .....................................................17
APPENDIX A ................................................................... 18
APPENDIX B ...................................................................19
i
Page 3
INTRODUCTION
The YSI Model 5905/5010/200-BOD/Pro-BOD (Biochemical Oxygen Demand) probe is used for measuring dissolved oxygen in all popular size BOD bottles. These probes feature self stirring, easily replaceable membrane caps, and a refurbishable electrode system.
The 5905 probe is designed for use with all YSI grey-box-style dissolved oxygen meters, except for the Model 55 or Model 550A. It must be used with an adapter (5011) in order to be used on the Model 5000 and 5100 bench top instruments.
The 5010 probe is specifically designed for use with the Model 5000 and 5100 only.
The 200-BOD probe is specifically designed for use with the Model DO200 only.
The Pro-BOD probe is specifically designed for use with the Pro20 and Professional Plus instruments.
This manual also applies to probes 5905-J, 5905-K, 5905-L, 5905­W (Appendix A), and probes 5010-J, 5010-K, 5010-L, and 5010­W (Appendix A)
---IMPORTANT---
Intended Use
The 5905/5010/200-BOD/Pro-BOD probe is intended for laboratory use, not field use. It is not waterproof and should never be immersed past the taper on the stem of the probe.
1
Page 4
OPERATING PRINCIPLES
The 5905/5010/200-BOD/Pro-BOD probe is a polarographic dissolved oxygen sensor. An oxygen permeable membrane covers an electrolytic cell consisting of a cathode and an anode. This membrane acts as a diffusion barrier and an isolation barrier preventing the fouling of the cathode surface by impurities in the environment. The cathode is the gold electrode, and the anode is the silver electrode which completes the electrolytic cell and acts as a reference electrode.
Upon entering the cell through the membrane, oxygen is reduced at an applied potential of -0.8 V referenced to the silver electrode. The reduction current at the cathode is directly proportional to the partial pressure of oxygen in liquid (expressed as %-air saturation) which is also proportional to the concentration of dissolved oxygen (in mg/L) at a particular temperature. Thus the same partial pressure of oxygen (%-air saturation) in liquid gives different concentrations of dissolved oxygen (mg/L) at different temperatures because of the different solubilities of oxygen at different temperatures.
2
Page 5
USING THE PROBE
The 5905/200-BOD/Pro-BOD must be plugged into a YSI meter and the stirrer power plug must be plugged into its power adapter. The 5010 must be plugged into a YSI meter only, as it gets its stirring power from the 5000/5100 instrument.
Changing the Membrane Cap
---IMPORTANT---
New Probe Use
New probes are shipped with a dry membrane cap to protect the electrodes. A new membrane cap must be installed before first use. For the 5905/5010, use a Teflon membrane from a 5906 Membrane Kit. For the 200-BOD/Pro-BOD, use a PE membrane from a 5908 Membrane Kit.
Follow these instructions to change membrane caps (Figure 1):
1. Remove the stir paddle from the probe by pulling it straight out.
2. Unscrew the old membrane cap from the probe. Before installing a new membrane, clean the probe tip with deionized water in order to remove any contaminants.
3. Hold the membrane cap and fill it at least ½ full with the electrolyte solution provided. Screw the membrane cap onto the probe moderately tight. A small amount of electrolyte should overflow.
4. Rinse off excess electrolyte from the probe with deionized water.
5. Reinstall the stir paddle.
3
Page 6
Remove
Stir Paddle
Fill Membrane Cap
with Solution
Unscrew Cap
Replace
Stir Paddle
Screw Cap on
moderately tight
Figure 1: Membrane Installation on the 5905/5010/200-BOD/Pro­BOD .
4
Page 7
Warm Up
Every time the instrument is turned on, the 5905/5010/200­BOD/Pro-BOD Probe must be allowed to warm up for about 10­20 minutes before taking the first reading. This allows the probe to use up all the oxygen retained in the cap.
---IMPORTANT---
New Probe Break In (motor off)
For best results, please allow the probe more than the usual 20 minutes to warm up after the first membrane is installed. The process may be faster if the probe is allowed to warm up in a BOD bottle filled with clean water. If possible, it’s a good idea to let new probes warm up overnight (with a new membrane and KCl) prior to first use.
Electrode Cleaning
The 5905/5010/200-BOD/Pro-BOD should be cleaned only when erratic readings occur or after about every 500 hours of use (two months). Each cleaning removes material and reduces the life of the probe, so excessive cleaning should be avoided.
Gold Cathode
For correct probe operation, the gold cathode must be textured properly (for its location, see figure 2). It can become tarnished or plated with silver after extended use. The gold cathode can be cleaned by using the sanding disc provided in the 5906 and 5908 Membrane Kits. Place the disc in the palm of your hand, then sand the gold with a twisting motion about 3 times or until all silver deposits are removed and the gold appears to have a matte finish. If the cathode remains tarnished, return the probe for service.
5
Page 8
Silver Anode
It is normal for a dark layer of silver chloride to cover the silver anode (for its location, see figure 2). After prolonged use it may become necessary to clean the anode. Soak the probe in a 14% ammonium hydroxide solution for 2 to 3 minutes or overnight in a 3% ammonium hydroxide solution. Rinse with deionized water, recharge with electrolyte, and install a new membrane. The anode may also need a very light sanding with the supplied sanding discs. Sand lightly in a helical fashion (down and around), then rinse well with deionized water.
Temperature
Sensor
Anode
(Silver)
Cathode
(Gold)
Figure 2: 5905/5010/200-BOD/Pro-BOD Sensor
6
Page 9
OPERATING PROCEDURES AND PRECAUTIONS
---IMPORTANT---
Intended Use
The 5905/5010/200-BOD/Pro-BOD probe is intended for laboratory use, not field use. It is not waterproof and should never be immersed past the taper on the stem of the probe.
Air Calibration
When calibrating in air, make sure there are no water droplets on the membrane. Water droplets will cause a low calibration reading. Excess water may be removed by shaking the probe downward. If droplets persist it may be necessary to carefully remove them using a clean cloth or paper towel.
Checking the Probe Zero
The probe zero is checked by immersing the probe in a sodium sulfite solution (0.08M or 3g Na has an inert gas bubbling through it (e.g. nitrogen, argon). The meter should read less than 1% dissolved oxygen in either of these environments. If it does not, change the membrane or clean the probe.
/300mL), or in water which
2SO3
Membrane Life
Membrane life depends on use. Membranes will last longer if installed properly and treated with care during use. Erratic readings will result from loose, wrinkled, or fouled membranes or from bubbles larger than 1/8” in the electrolyte solution. If erratic readings or other evidence of membrane damage occurs, replace the membrane and KCl solution.
7
Page 10
The average replacement interval is two to four weeks. Probes in constant or heavy use may require more frequent membrane changes.
Interferences
Hydrogen sulfide, sulfur dioxide, halogens, carbon monoxide, chlorine, nitric oxide, and nitrous oxide can cause the probe to give erroneous readings. If you suspect erroneous readings, it may be necessary to determine if these are the cause.
Acids
Avoid any environment that contains substances such as concentrated acids, caustics, and strong solvents that may attack the probe. Probe materials that may be damaged by these substances include FEP Teflon or Polyethylene, EPR rubber, ABS plastic, and stainless steel.
Erroneous Readings
Erroneous readings may occur if the membrane is coated with oxygen consuming bacteria or oxygen evolving algae. Heavy residue may coat the membrane causing incorrect readings. Frequent membrane changes will eliminate this problem.
Storage
When the probe is not in use, store the probe in a BOD bottle containing at least 1 inch of water. For long term storage, remove the membrane cap, rinse the probe tip with deionized water, and install a dry membrane cap (without electrolyte solution).
8
Page 11
PROBLEM SOLVING
Trouble Shooting Guide
Use this quick reference table to correct symptoms you might be experiencing.
Symptoms Actions
Unable to calibrate
Allow approximately 20 min. and up to 30
min. for probe to polarize when first turned on
Change membrane
Clean probe (see instructions under Electrode
Cleaning)
Does not return to calibration value
Unstable readings
Allow sufficient time for temperature and
oxygen stabilization
Check for water droplets on membrane
Change membrane
Clean probe (see instructions under Electrode
Cleaning)
Allow sufficient time for temperature and
oxygen stabilization
Allow 60 sec for reading to stabilize
Change membrane
Clean probe (see instructions under Electrode
Cleaning)
9
Page 12
Symptoms Actions
Unstable readings Allow sufficient time for temperature and
oxygen stabilization
Allow 60 sec for reading to stabilize
Change membrane
Clean probe (see instructions under Electrode
Cleaning)
Motor not working
Loud motor
Membrane cap is tight
Check power supply
Manually turn the stir paddle to help start the
motor
Replace motor
Check if stir paddle is pushed all the way in
Replace motor
Use pliers to loosen the cap (turn counter
clockwise while facing sensor tip)
10
Page 13
SERVICING
The 5905/5010/200-BOD/Pro-BOD has four user replaceable parts. The motor assembly can be replaced if it becomes noisy, unbalanced, or malfunctions. The probe assembly, housing assembly, and stir paddle can also be replaced if they become damaged.
To disassemble the probe, perform the following steps while referring to figure 3:
1. Remove stir paddle by pulling it straight out.
2. Remove the two screws (A) on the outside of the housing
using a 3/32 Allen wrench.
3. Loosen the top nut of the cable strain relief (B) from the housing. This will allow the cable to feed through the housing assembly.
4. Pull the probe and motor assembly out of the housing and disconnect the two plugs (C).
5. To remove the motor assembly from the probe assembly, remove the two screws (D) using the 3/32 Allen wrench.
11
Page 14
(B)
(A)
(D)
Housing Assembly
(A)
(C)
(C)
(D)
Motor Assembly
Probe Assembly
Stir Paddle
Figure 3: Exploded View of 5905/5010/200-BOD/Pro-BOD Probe
12
Page 15
WARRANTY AND REPAIR
YSI Models 5905/5010/200-BOD/Pro-BOD Dissolved Oxygen Probes are warranted for one year from date of purchase by the end user against defects in materials and workmanship. Electrode cleaning or maintenance is not covered under warranty. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This warranty does not apply to any YSI product damage or failure caused by (i) failure to install, operate or use the product in accordance with YSI’s written instructions, (ii) abuse or misuse of the product, (iii) failure to maintain the product in accordance with YSI’s written instructions or standard industry procedure, (iv) any improper repairs to the product, (v) use by you of defective or improper components or parts in servicing or repairing the product, or (vi) modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.
13
Page 16
Authorized Service Centers
YSI has several Authorized Service Centers in the U.S. and around the world. Please refer to the YSI web site (www.YSI.com nearest Authorized Service Center.
) for your
Cleaning Instructions
NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned and disinfected. Biological contamination is presumed for any
instrument, probe, or other device that has been used with body fluids or tissues, or with wastewater. Radioactive contamination is presumed for any instrument, probe or other device that has been used near any radioactive source.
If an instrument, probe, or other part is returned or presented for service without a Cleaning Certificate, and if in our opinion it represents a potential biological or radioactive hazard, our service personnel reserve the right to withhold service until appropriate cleaning, decontamination, and certification has been completed. We will contact the sender for instructions as to the disposition of the equipment. Disposition costs will be the responsibility of the sender.
When service is required, either at the user's facility or at YSI, the following steps must be taken to insure the safety of our service personnel.
1. In a manner appropriate to each device, decontaminate all exposed surfaces, including any containers. 70% isopropyl alcohol or a solution of 1/4 cup bleach to 1 gallon tap water is suitable for most disinfecting. Instruments used with waste water may be disinfected with .5% Lysol if this is more convenient to the user.
2. The user shall take normal precautions to prevent radioactive contamination and must use appropriate decontamination procedures should exposure occur.
3. If exposure has occurred, the customer must certify that
14
Page 17
decontamination has been accomplished and that no radioactivity is detectable by survey equipment.
4. Any product being returned to the YSI Repair Center should be packed securely to prevent damage.
5. Cleaning must be completed and certified on any product before returning it to YSI.
Packing Instructions
1. Clean and decontaminate items to insure the safety of the handler.
2. Complete and include the Cleaning Certificate.
3. Place the product in a plastic bag to keep out dirt and packing
material.
4. Use a large carton, preferably the original, and surround the product completely with packing material.
Insure for the replacement value of the product.
15
Page 18
Alternatively, if you are planning to send your probe directly to YSI for servicing, please visit www.ysi.com the Product Return Form.
16
, download and fill out
Page 19
Probe Specifications
Cathode: Gold
Anode: Silver
Membrane: FEP Teflon (for 5905/5010)
Polyethylene (for 200-BOD/Pro-BOD)
Dissolved Oxygen Accuracy: ± 0.1 mg/L or ± 2% of reading,
whichever is greater
Operating Temperature Range: 0° to 45°C
Temperature Accuracy: ± 0.2°C
Polarizing Voltage: 0.8 volts
Response Time: Typical response to a step change in dissolved
oxygen at 20°C is 90% of the change in 30 seconds.
17
Page 20
APPENDIX A
The 5905 and 5010 probes come in several versions, designed to fit different bottle sizes.
5905 5010 200-BOD Pro-BOD
5905 - J 5010 - J
5905 - K 5010 - K
5905 - L 5010 - L
5905 - W 5010 - W
5° taper per side to fit U.S. Standard BOD bottles.
Includes an adapter to fit 5° taper per side smaller bottles with built - in overflow funnels.
Includes a
Includes a
19 funnel.
24 funnel.
6 long probe for use with wine bottles.
The 5010 probe will only work with the Model 5000 and 5100 Instruments. A 5905 probe will also work on these instruments by using a 5011 plug adapter. A 200-BOD will only work with a DO200 Instrument. The Pro-BOD will only work with the Pro20 and ProPlus Instruments.
18
Page 21
Pro-BOD
N/A
605306
059895
115776
110592
115777
N/A
073035
N/A
200-BOD
N / A
605306
059895
609204
110592
609210
N / A
073035
073047
5010-W
059880
N / A
059950
110593
110592
115011
N / A
N / A
N / A
5010-L
059880
N / A
059902
115904
110594
115011
059911
N / A
N / A
5010-K
059880
N / A
059902
115904
110594
115011
059933
N / A
N / A
5010-J
059880
N / A
059902
115904
110592
115011
059929
N / A
N / A
5010
059880
N / A
059895
110590
110592
115011
N / A
N / A
073035
N / A
073047
5905-W
059880
N / A
059950
110593
110592
110591
APPENDIX B
N / A
5905-L
059880
N / A
059902
115904
110594
110591
059911
073035
073047
5905-K
059880
N / A
059902
115904
110594
110591
059933
See dealer
See dealer
Accessories and Replacement Parts by Item Number
5905-J
059880
N / A
059902
115904
110592
110591
059929
073035
073047
5905
059880
N / A
059895
110590
110592
110591
N / A
073035
073047
Description
5906 Membrane Kit
5908 Membrane Kit
Stir Paddle
Probe Assembly
Motor Assembly
Housing Assembly
Funnel / Adapter
115 V Power Supply
230 V Power Supply
19
Page 22
1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA +1 937 767-7241 800 897-4151 (US) A59882L Fax +1 937 767-9353 May 09 environmental@ysi.com www.ysi.com
© 1997 YSI Incorporated
059882
20
Loading...