Yokogawa PH97 User Manual

Instruction
Installation
Calibration
Operation
Maintenance
KEY FEATURES
Instruction
Manual
Manual
Model PH87
Retraction Sensor Assembly
Yokogawa Corporation of America 2 Dart Road, Newnan, Georgia U.S.A. 30265 Tel: 1-800-258-2552 Fax: 1-770-254-0928
IM M1263JL-A
Doc. Nr. L2528
January 2012
3rd edition
2
TABLE OF CONTENTS
Overview ....................................................................................................................................... 3
Assembly Instructions-Stage #1 .................................................................................................. 4
Assembly Installation Instructions- with /B ball valve option ......................................................................8
In-Service Installation-with /B ball valve option ................................................................................ 10
Sensor Service ........................................................................................................................................... 11
Sensor Replacement .................................................................................................................................. 14
Exploded View of PH87 ............................................................................................................................. 16
Part Numbers and Descriptions ................................................................................................................. 17
PH87 Dimensions and Clearance Requirements .............................................................................. 18
IM M1263JL-A
Sensor Installation Overview
Shown with Optional Ball valve
NOTE: THE FOLLOWING INFORMATION IS INTENDED TO PROVIDE THE USER WITH A PREVIEW OF WHAT WILL BE REQUIRED TO INSTALL THE MODEL PH87 SENSOR ASSEMBLY. PLEASE REVIEW THE DIAGRAM AND GUIDELINES BEFORE PROCEEDING.
Typical Sensor Installation Diagram
Shown with Optional /B Ball Valve
3
Sensor Installation Guidelines:
ASSEMBLY
pH/ORP SENSOR
[1]. Install the model PH97 sensor in pipe or vessel between 15° and 90° from horizontal.
[2]. Connect ex conduit with adequate service loop to junction box. [3]. Connect the sensor cable to the terminals inside the junction box. [4]. Keep cable run between junction box and pH / ORP instrument as short as possible.
IM M1263JL-A
4
8
Model PH87 Retraction Sensor Assembly
Sensor Assembly Instructions-Stage1
BEGIN STAGE #1 ASSEMBLY OF THE MODEL PH87 BY FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [A] THROUGH [F] IN ORDER AS SHOWN BELOW:
[STEP B]. SLIDE
PACKING GLAND, ITEM NO.3, ONTO SHEATH, ITEM NO.4. KEEP LOCKING NUT AND FERRULES LOOSE BUT ATTACHED TO PACKING GLAND
[STEP A]. INSTALL PACKING GLAND O-RING, ITEM NO. 3d, INTO PACKING GLAND BODY, ITEM NO. 3c.
[STEP F]. THREAD SENSOR CABLE THROUGH COUPLER TUBE, ITEM NO. 1, THEN SLIDE COUPLER TUBE INTO SHEATH UNTIL IT CONTACTS SENSOR.
[STEP D]. THREAD SENSOR CABLE THROUGH SHEATH AND SLIDE THE SENSOR INTO SHEATH UNTIL IT STOPS AGAINST FLARED END
FLARED END OF SHEATH
[STEP C]. STRETCH O-RINGS, ITEM NO. 2, INTO GROOVES ON SENSOR
PLEASE NOTE
REFER TO PAGES 16 AND 17 OF THE SENSOR ASSEMBLY AND ITEMIZED PART LIST FOR ALL THESE COMPONENTS
REFER TO PAGES 15 AND 16 OF THE SENSOR ASSEMBLY AND ITEMIZED PART LIST FOR ALL THESE COMPONENTS.
[STEP E]. STRETCH O-RING, ITEM NO. 1a, INTO GROOVE ON COUPLER TUBE, ITEM NO. 1.
IM M1263JL-A
Sensor Assembly Instructions-Stage1 (Continued)
(Continued)
5
[STEP G]. THREAD SENSOR CABLE THROUGH COUPLER TUBE, ITEM NO. 1, THEN SLIDE COUPLER TUBE INTO SHEATH UNTIL IT CONTACTS SENSOR.
THE COUPLER TUBE AND SENSOR SHOULD BE THREADED TOGETHER UNTIL BOTH ARE SEATED AGAINST THE ENDS OF THE SHEATH,
IM M1263JL-A
6
Sensor Assembly Instructions—Stage1
(Continued)
Model PH87 Retraction Sensor Assembly
Sensor Assembly Instructions-Stage1 (Continued)
[STEP H]. REMOVE JUNCTION BOX
COVER AND THERAD SENSOR CABLE
[Step H]. REMOVE
THROUGH JUNCTION BOX ASSEMBLY
JUNCTION BOX COVER AND
(JUNCTION BOX AND PROVIDED
THERAD SENSOR CABLE THROUGH
PIPE NIPPLE). THREAD JUNCTION
JUNCTION BOX ASSEMBLY
BOX ASSEMBLY PIPE NIPPLE INTO
(JUNCTION BOX AND PROVIDED
SENSOR COUPLER ASSEMBLY,
PIPE NIPPLE). THREAD JUNCTION
ITEM NO. 1.INSTALL SUITABLE FLEX
BOX ASSEMBLY PIPE NIPPLE INTO
CONDUIT LONG ENOUGH TO ALLOW
SENSOR COUPLER ASSEMBLY,
FOR SERVICING OR REPLACEMENT
ITEM NO. 1.INSTALL SUITABLE FLEX
OF SENSOR.
CONDUIT LONG ENOUGH TO ALLOW FOR SERVICING OR REPLACEMENT OF SENSOR.
[STEP I]. CONNECT THE SENSOR WIRES TO THE 5-TERMINAL STRIP IN
[Step I]. CONNECT THE SENSOR
THE JUNCTION BOX IN THE
WIRES TO THE 5-TERMINAL STRIP IN
FOLLOWING ORDER. THE RED TC WIRE
THE JUNCTION BOX IN THE FOL-
CONNECTS TO TERMINAL #1 AND
LOWING ORDER. THE RED TC WIRE
THE BLACK TC WIRE TO TERMINAL
CONNECTS TO TERMINAL #1 AND
#2. CONNECT THE SENSOR
THE BLACK TC WIRE TO TERMINAL
REFERENCE (BLACK) TO #3,
#2. CONNECT THE SENSOR REFER-
THE SOLUTION GROUND (GREEN)
ENCE (BLACK) TO #3, THE SOLUTION
TO TERMINAL #4 AND PH (CLEAR) TO
GROUND (GREEN) TO TERMINAL #4
TERMINAL #5. THREAD THE EXTENSION
AND pH (CLEAR) TO TERMINAL #5.
CABLE THROUGH CONDUIT. ATTACH
THREAD THE EXTENSION CABLE
THE LABELED WIRES TO THE
THROUGH CONDUIT. ATTACH THE
TERMINALS IN THE SAME ORDER AS
LABELED WIRES TO THE TERMINALS
THE SENSOR WIRES LISTED ABOVE.
IN THE SAME ORDER AS THE SEN- SOR WIRES LISTED ABOVE. SLIDE-
REPLACE THE JUNCTION BOX COVER.
THE PROTECTIVE INSULTING BOOT OVER CONNECTORS AND REPLACE THE JUNCTION BOX COVER.
IM M1263JL-A
pH9
(Continued)
Sensor Assembly Instructions-Stage1 (Continued)
THE STAGE #1 ASSEMBLY OF THE pH97 SENSOR IS NOW COMPLETE AND SHOULD APPEAR AS ONE OF THE TWO ASSEMBLIES SHOWN BELOW
STANDARD ASSEMBLY
STANDARD ASSEMBLY
7
[STEP J]. PERFORM A 2-POINT CALIBRATION WITH PH BUFFER SOLUTIONS BEFORE PROCEEDING FURTHER WITH THE ASSEMBLY AND INSTALLATION.
NOTE: REFER TO PH 97 SENSOR INSTRUCTIONS FOR WIRING AND CALIBRATION PROCEDURES.
NOTE: REFER TO PH 97 SENSOR INSTRUCTIONS FOR WIRING AND CALIBRATION PROCEDURES.
IM M1263JL-A
8
Sensor Assembly Installation Instructions—/B Option
8/B OPTION FOR
Model PH87 Retraction Sensor Assembly
Sensor Assembly Instructions-/B Ball Valve Option
BEGIN ASSEMBLY OF THE /B BALL VALVE OPTION FOR MODEL PH87 SENSOR BY FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [A] THROUGH [D] IN ORDER AS SHOWN BELOW:
PLEASE NOTE
REFER TO PAGES 16 AND 17 OF THE
REFER TO PAGES 15 AND 16 OF THE
SENSOR ASSEMBLY AND ITEMIZED
SENSOR ASSEMBLY AND ITEMIZED PART
PART LIST FOR ALL THESE
LIST FOR ALL THESE COMPONENTS.
COMPONENTS
[STEP A]. THREAD THE REDUCER BUSHING, ITEM NO.6, USING TEFLON TAPE TO SEAL AND TIGHTEN IN PLACE
[STEP B]. THREAD THE NIPPLE, ITEM NO.7, INTO BALL VALVE, ITEM NO.6, USING TEFLON TAPE TO SEAL AND TIGHTEN IN PLACE
[STEP C]. THREAD THE NIPPLE, ITEM NO.7, ALONG WITH BALL VALVE, ITEM NO.6, AND REDUCER BUSHING, ITEM NO.5, INTO WELDALET OR AP­PROPRIATE PROCESS FITTING. USE TEFLON TAPE TO SEAL THREADS AND TIGHTEN IN PLACE
[STEP D]. INSURE THAT THE VALVE IS IN THE CLOSED POSTION AS SHOWN BELOW BEFORE PROCEEDING TO NEXT STEP.
PLEASE NOTE: /B OPTION IS REQUIRED (OR CUSTOMER SUPPLIED BALL VALUE IS
REQUIRED) FOR RETRACTION OF PH87 WHILE PROCESS IS UNDER PRESSURE
IM M1263JL-A
Sensor Assembly Installation Instructions-/B Ball Valve (Continued)
8
/B OPTION
THE ASSEMBLY OF THE PH87 SENSOR IS NOW COMPLETE AND SHOULD APPEAR AS SHOWN. DO NOT TIGHTEN PACKING GLAND LOCKING NUT OR OPEN BALL VALVE UNTIL SENSOR ASSEMBLY IS READY TO GO ON-STREAM.
COMPLETE /B BALL VALVE OPTION ASSEMBLY OF THE MODEL PH87 BY FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [E] AND [F] IN ORDER AS SHOWN BELOW:
9
[STEP E]. APPLY TEFLON TAPE TO PIPE THREADS OF PACKING GLAND FITTING AND SLIDE FITTING TO SENSOR END OF SHEATH. INSTALL STAGE #1 SUBASSEMBLY INTO VALVE TAKING CARE NOT TO FORCE THE SENSOR INTO THE BALL PORTION OF THE VALVE TO PREVENT BREAKAGE OF THE GLASS BULB.
[STEP F]. SECURELY TIGHTEN PACKING GLAND FITTING INTO VALVE ASSEMBLY TO PROVIDE A PROPER SEAL.
THE ASSEMBLY OF THE PH87 SENSOR IS NOW COMPLETE AND SHOULD APPEAR AS SHOWN. DO NOT TIGHTEN PACKING GLAND LOCKING NUT OR OPEN BALL VALVE UNTIL SENSOR ASSEMBLY IS READY TO GO ON-STREAM.
IM M1263JL-A
10
Sensor Assembly In-Service Instructions/B Option
Model PH87 Retraction Sensor Assembly
Sensor Assembly In-Service Instructions-/B Ball Valve Option
PLEASE NOTE
REFER TO PAGES 16 AND 17 OF THE
REFER TO PAGES 15 AND 16 OF THE
SENSOR ASSEMBLY AND ITEMIZED
SENSOR ASSEMBLY AND ITEMIZED PART
PART LIST FOR ALL THESE
LIST FOR ALL THESE COMPONENTS.
COMPONENTS
THE FINAL ASSEMBLY IS NOW READY TO BE PLACED INTO SERVICE, WITH PACKING GLAND LOCKING NUT ENGAGED BUT NOT TIGHTENED AND SENSOR IN THE RETRACTED POSITION.
[STEP A]. TO PLACE ASSEMBLY
[Step A]. TO PLACE ASSEMBLY
INTO SERVICE OPEN BALL
INTO SERVICE OPEN BALL
VALVE COMPLETELY.
VALVE COMPLETELY. Caution: OPENING VALVE
CAUTION: OPENING VALVE
SUBJECTS SENSOR ASSEMBLY
SUBJECTS SENSOR ASSEMBLY
TO LINE PRESSURE. IF THE
TO LINE PRESSURE. IF THE
SENSOR IS NOT RETRACTED
SENSOR IS NOT RETRACTED
COMPLETLY AGAINST STOP,
COMPLETLY AGAINST STOP,
OPENING THE VALVE COULD
OPENING THE VALVE COULD
RESULT IN THE SENSOR
RESULT IN THE SENSOR
SUDDENLY RETRACTING THE
SUDDENLY RETRACTING THE
REMAINING DISTANCE
REMAINING DISTANCE
CAUSING DAMAGE TO THE
CAUSING DAMAGE TO THE
SENSOR ASSEMBLY OR INJURY
SENSOR ASSEMBLY OR INJURY
TO THE SERVICE PERSON.
TO THE SERVICE PERSON.
[STEP B]. INSERT STAGE #1 SUBASSEMBLY INTO PROCESS.
[STEP C]. SECURE SENSOR IN PLACE BY TIGHTENING LOCKING NUT AND FERRULES OF GLAND FITTING, ITEM NO.3.
DO NOT INSERT TIP OF SENSOR BEYOND INSIDE WALL OF PROCESS PIPE TO AVOID TORQUE AND POSSIBLE BREAKAGE.
THE IN-SERVICE ASSEMBLY SHOULD APPEAR AS SHOWN BELOW.
IM M1263JL-A
11
Sensor Service Instructions
BEFORE SERVICING REMOVE THE MODEL PH97 SENSOR FROM SERVICE FOLLOWING THE
STEP-BY-STEP INSTRUCTIONS [A] AND [B] IN ORDER AS SHOWN BELOW:
Model PH87 Retraction Sensor Assembly
Sensor Service Instructions
BEFORE SERVICING REMOVE THE MODEL PH97 SENSOR FROM SERVICE FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [A] AND [B] IN ORDER AS SHOWN BELOW:
[STEP B]. FULLY
RETRACT STAGE #1 SUBASSEMBLY SHAFT.
[Step B]. FULLY RETRACT STAGE #1 SUB- ASSEMBLY SHAFT.
[STEP A]. LOOSEN LOCKING NUT AND FERRULES OF GLAND
[Step A]. LOOSEN LOCKING
FITTING, ITEM NO. 3.
NUT AND FERRULES OF GLAND
CAUTION: THE SENSOR ASSEMBLY
FITTING, ITEM NO. 3.
IS STILL UNDER PRESSURE.
Caution: THE SENSOR ASSEMBLY
LOOSENING THE LOCKING NUT
IS STILL UNDER PRESSURE. LOOSENING THE LOCKING NUT
COULD RESULT IN THE SENSOR
COULD RESULT IN THE SENSOR
RETRACTING SUDDENLY AND
RETRACTING SUDDENLY AND
WITH A GREAT DEAL OF FORCE.
WITH A GREAT DEAL OF FORCE.
THIS COULD CAUSE DAMAGE
THIS COULD CAUSE DAMAGE
TO THE SENSOR ASSEMBLY
TO THE SENSOR ASSEMBLY
OR INJURY TO THE SERVICE
OR INJURY TO THE SERVICE
PERSON.
PERSON.
IM M1263JL-A
12
Sensor Service Instructions
9
(Continued)
Model PH87 Retraction Sensor Assembly
Sensor Service Instructions (Continued)
BEFORE SERVICING REMOVE THE MODEL PH97 SENSOR FROM SERVICE FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [C] AND [D] IN ORDER AS SHOWN BELOW:
[STEP C]. CLOSE BALL VALVE
COMPLETELY.
[STEP D]. COMPLETELY UNTHREAD GLAND FITTING, ITEM NO.4, FROM VALVE ASSEMBLY
IM M1263JL-A
Sensor Service Instructions (Continued)
(Continued)
9
CONTINUE SERVICING OF THE MODEL PH97 SENSOR BY FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [E] THROUGH [G] IN ORDER AS SHOWN BELOW:
[Step G]. TO TEST SENSOR PERFORM A 2-POINT CALIBRATION WITH pH BUFFER AS SHOWN BELOW.
[STEP E]. REMOVE STAGE #1 ASSEMBLY FROM VALVE ASSEMBLY.
WITH THE STAGE #1 SUBASSEMBLY COMPLETELY REMOVED FROM THE VALVE ASSEMBLY THE SENSOR MAY NOW BE TESTED OR REPLACED.
[STEP F]. BEFORE TESTING THE SENSOR BE SURE THAT IT IS CLEAN
IF THE SENSOR TESTS WITHIN ACCEPTABLE LIMITS REINSTALL AND PLACE SENSOR BACK IN SERVICE BY FOLLOWING ASSEMBLY INSTRUCTIONS.
IF TESTING PROVES SENSOR T0 BE FAULTY CONTINUE WITH THE SERVICE INSTRUCTIONS FOR SENSOR REPLACEMENT.
13
[STEP G]. TO TEST SENSOR PERFORM A 2-POINT CALIBRATION WITH PH BUFFER AS SHOWN BELOW.
IF THE SENSOR TESTS WITHIN ACCEPTABLE LIMITS REINSTALL AND PLACE SENSOR BACK IN SERVICE BY FOLLOWING ASSEMBLY INSTRUCTIONS.
IF TESTING PROVES SENSOR T0 BE FAULTY CONTINUE WITH THE SERVICE INSTRUCTIONS FOR SENSOR REPLACEMENT.
IM M1263JL-A
14
Model PH87 Retraction Sensor Assembly
Sensor Replacement Instructions
TO DISASSEMBLE THE MODEL PH87 SENSOR WITH A SENSOR MOUNTED JUNCTION BOX,
FOLLOW THE STEP-BY-STEP INSTRUCTIONS [A] THROUGH [D] IN ORDER AS SHOWN BELOW:
Sensor Replacement Instructions
BEFORE SERVICING REMOVE THE MODEL PH87 SENSOR FROM SERVICE FOLLOWING THE STEP-BY-STEP INSTRUCTIONS [A] AND [D] IN ORDER AS SHOWN BELOW:
[Step C]. WHILE FIRMLY GRIPING SHEATH PUSH ON COUPLER TUBE -
[STEP C]. WHILE FIRMLY GRIPING SHEATH
JUNCTION BOX ASSEMBLY TO
PUSH ON COUPLER TUBE JUNCTION BOX
UNSEAT SENSOR O-RINGS
ASSEMBLY TO UNSEAT SENSOR O-RINGS
[Step A]. REMOVE JUNCTION BOX COVER AND DISCON-
[STEP A]. REMOVE
NECT ALL WIRES FROM TERMINALS.
JUNCTION BOX COVER AND DISCONNECT ALL WIRES FROM TERMINALS.
[Step B]. WHILE FIRMLY GRIPING
[STEP B]. WHILE FIRMLY GRIPING
THE SENSOR AND THE SHEATH AT
THE SENSOR AND THE SHEATH AT
THE SAME TIME UNTHREAD THEM
THE SAME TIME UNTHREAD THEM
FROM THE COUPLER ASSEMBLY BUT
FROM THE COUPLER ASSEMBLY BUT
DO NOT COMPLETELY DISENGAGE
DO NOT COMPLETELY DISENGAGE
THREADS.
THREADS.
[Step D]. COMPLETELY UNTHREAD
[STEP D]. COMPLETELY UNTHREAD
SENSOR FROM COUPLER TUBE
SENSOR FROM COUPLER TUBE
ASSEMBLY.
ASSEMBLY.
CONTINUE WITH SERVICE INSTRUCTIONS AT STEP "N".
CONTINUE WITH SERVICE INSTRUCTIONS AT STEP “N”.
IM M1263JL-A
Sensor Replacement Instructions (Continued)
(Continued)
DISASSEMBLE THE ASSEMBLY INTO THE COMPONENT PARTS AS SHOWN IN ONE OF THE ILLUSTRATIONS BELOW. TO INSTALL A NEW SENSOR AND RETURN THE MODEL PH87 TO SERVICE FOLLOW THE ASSEMBLY INSTRUCTIONS STARTING FROM STAGE #1 STEP “D” (PAGE 4)
15
FIGURE 1: STANDARD ASSEMBLY READY FOR
FIGURE 1: STANDARD
SENSOR REPLACEMENT
ASSEMBLY READY FOR SENSOR REPLACEMENT
FIGURE 2: ASSEMBLY WITH SENSOR
FIGURE 2: ASSEMBLY WITH
MOUNTED JUNCTION BOX READY FOR
SENSOR MOUNTED JUNCTION
SENSOR REPLACEMENT
BOX READY FOR SENSOR REPLACEMENT
IM M1263JL-A
16
Model PH87 Retraction Sensor Assembly
Sensor Housing Assembly
Sensor Housing Assembly
8
JUNCTION BOX
JUNCTION BOX
2 REQUIRED
9
pH SENSOR (SEE MODEL PH97 SPEC SHEET)
PH SENSOR
(SEE MODEL PH97
SPEC SHEET)
PLEASE NOTE
REFER TO PAGES 15 AND 16 OF THE SENSOR
ALL O-RINGS REQUIRE PROPER LUBRICATION PRIOR TO SERVICING. WE RECOMMEND USING PARKER O LUBE OR
ASSEMBLY AND ITEMIZED PART LIST FOR ALL
EQUIVALENT. FOLLOW THE LUBRICATION INSTRUCTIONS
THESE COMPONENTS
PROVIDED BY THE MANUFACTURER.
IM M1263JL-A
Sensor Housing Assembly
THESE ARE THE INDIVIDUAL PART NUMBERS FOR THE ITEMS SHOWN IN THE EXPLODED VIEW ON THE PREVIOUS PAGE. PART NUMBERS FOR A COMPLETE ASSEMBLY ARE LISTED ON NEXT TWO PAGES, ON THE SPECIFIC DRAWING.
Item Part Number Description
1 M1234DQ Coupler Assembly, Kynar (PVDF), N=7”
M1234TV-A Coupler Assembly, Kynai (PVDF), N=21”
These part numbers includes the o-ring
1a, 2 M1234DS O-rings, viton, sensor, size -019, quantity of 10
3 M1234DT Compression Fitting, 1” Tube x 1” MNPT
3a M1234DW Compression Fitting, Nut
3b M1234DX Compression Fitting, Ferrule, Kynar (PVDF)
3c M1234EJ Compression Fitting, Body, Titanium
3d M1234DZ Compression Fitting, O-ring, viton, size -120
4 M1234EA Sheath, Titanium, N=7”
M1234JQ-A Sheath, Titanium, N=21”
17
5 M1234WZ-A Bushing, 1.25” MNPT x 1” FNPT
6 M1234WY-A Ball Valve, Full Port, 1.25” NPT
7 M1234XA-A Nipple, 1.25” NPT
8 M1234EE Junction Box Assembly, Sensor Mount
9 See pH Sensor Specications
IM M1263JL-A
18
Scale:
Drwg Nr:
Sheet:
Date:
Drawn By:
21-July_00
Rev:
R. Ramus
Group:
1 / 1
Title:
P2254
Type:
Product Data
Part Nr:
PH87/B
PH87/B Insertion Assembly:
Dimensions & Clearance Requirements
2 Dart Road, Newnan, GA 30265
27.25" 40.05" 41.40"
MODEL NO. "N" LENGTH "R" DIMENSION "C" DIMENSION
PH87-07/B
PH87-21/B
13.25" 28.05"
29.40"
Insertion
Length
Retraction
Distance
Minimum
Clearance
"N" = FULLY INSERTED
3.72"
5.22"
1 1/4" MNPT
Minimum
Clearance
“C” DIMENSION
Distance
Retraction
“R” DIMENSION
"C" = MINIMUM CLEARANCE REQUIRED FOR SENSOR REMOVAL
"R" = FULLY RETRACTED
15.65" FULLY INSERTED
3/4" FNPT
Length
Insertion
PH87-21/B 21” 40.05” 41.40”
MODEL NO. “N” LENGTH
PH87-07/B 7” 28.05” 29.40”
PH87 With /B Ball Valve Insertion Assembly:
Dimensions & Clearance Requirements
2 Dart Road, Newnan, GA 30265
Tel.: 770-254-0400 Fax.:770-251-6427
IM M1263JL-A
27.25"
33.15" 36.90"
Scale:
Drwg Nr:
Sheet:
Date:
Drawn By:
21-July_00
Rev:
R. Ramus
Group:
1 / 1
Title:
P2254
Type:
Product Data
Part Nr:
PH87
PH87 Insertion Assembly:
Dimensions & Clearance Requirements
"N" = FULLY INSERTED
MODEL NO. "N" LENGTH "R" DIMENSION "C" DIMENSION
PH87-07
PH87-21
13.25" 19.15"
22.90"
Insertion
Length
Retraction
Distance
Minimum
Clearance
2 Dart Road, Newnan, GA 30265
3.72"
Minimum
Clearance
“C” DIMENSION
Distance
Retraction
“R” DIMENSION
19
9.65" FULLY
INSERTED
"R" = FULLY RETRACTED
"C" = MINIMUM CLEARANCE
3/4" FNPT
REQUIRED FOR SENSOR REMOVAL
1 " MNPT
Length
Insertion
MODEL NO. “N” LENGTH
PH87-21 27.25” 33.15” 36.90”
PH87-07 13.25” 19.15” 22.90”
PH87 Insertion Assembly:
Dimensions & Clearance Requirements
2 Dart Road, Newnan, GA 30265
Tel.: 770-254-0400 Fax.:770-251-6427
IM M1263JL-A
YOKOGAWA HEADQUARTERS 9-32, Nakacho 2-chome, Musashinoshi Tokyo 180 Japan
Tel. (81)-422-52-5535 Fax (81)-422-55-1202 E-mail: webinfo@mls.yokogawa.co.jp www.yokogawa.com.jp
YOKOGAWA EUROPE B.V. Euroweg 2 3825 HD Amersfoort The Netherlands
Tel. +31 (0)88-4641000 Fax +31 (0)88-4641111 E-mail: info@nl.yokogawa.com www.yokogawa.com/eu
YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 United States
Tel. (1)-770-253-7000 Fax (1)-770-251-2088 E-mail: info@yca.com www.yokogawa.com/us
YOKOGAWA ELECTRIC ASIA Pte. Ltd. 5 Bedok South Road Singapore 469270 Singapore
Tel. (65)-241-9933 Fax (65)-241-2606 E-mail: webinfo@yas.com.sg www.yokogawa.com.sg
Yokogawa has an extensive sales
and distribution network.
Please refer to the website
(www.yokogawa.com/an/index.htm) to contact your
nearest representative.
IM M1234JL-A Subject to change without notice Printed in The USA Copyright © 2008
IM M1263JL-A
Loading...