• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specied in this user’s manual, the protection provided by this instrument may be impaired.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modication of the product is strictly prohibited.
Notes on Handling User’s Manuals
n
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you nd mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
Warning and Disclaimer
n
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from
using the product or any defect of the product that YOKOGAWA can not predict in advance.
n Signal Words
The following words are used in this manual.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
IM 12B07N01-01E
After-sales Warranty
n Do not modify the product.
n During the warranty period, for repair under warranty consult the local sales representative
or service oce. Yokogawa will replace or repair any damaged parts. Before consulting for
repair under warranty, provide us with the model name and serial number and a description
of the problem. Any diagrams or data explaining the problem would also be appreciated.
• If we replace the product with a new one, we won’t provide you with a repair report.
• Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa shall conduct dened warranty service based on its standard. When the customer site is
located outside of the service area, a fee for dispatching the maintenance engineer will be
charged to the customer.
• Returned goods that have been in contact with process uids must be decontaminated and
disinfected prior to shipment. Goods should carry a certicate to this eect, for the health
and safety of our employees. Material Safety Data sheets must be included for all components of the process to which the sensor have been exposed.
nIn the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insucient maintenance by user.
• Failure due to modication, misuse or outside-of-specications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from re, earthquake, storms and oods, lightning, disturbances, riots, warfare,
radiation and other natural changes.
nYokogawa does not warrant conformance with the specic application at the user site.
Yokogawa will not bear direct/indirect responsibility for damage due to a specic application.
<INTRODUCTION>
iii
n Yokogawa Electric will not bear responsibility when the user congures the product into
systems or resells the product.
n Maintenance service and supplying repair parts will be covered for ve years after the pro-
duction ends. For repair for this product, please contact the nearest sales oce described in
this instruction manual.
Note: The temperature may be limited by the specications of the sensor.
Flow rate: 3 to 11 L/min
Note: The ow rate may be limited by the specications of the sensor.
Pressure: Atmospheric pressure to 500 kPa
Note: The pressure may be limited by the specications of the sensor.
1-1
IM 12B07N01-01E
1-2
< 1. Specications >
Utility required for cleaning unit:Utility required for cleaning unit:
TypePressure (kPa)Flow Ratee
Water jet200 to 400 + Liquid pressure5 to 20 l/min
Water brush100 to 250 + Liquid pressure20 to 30 l/min
Air jet200 to 400 + Liquid pressure 100 to 300 Nl/min
Air brush150 to 250 + Liquid pressure300 to 600 Nl/min
Note 1: Pressure and ow rate must be simultaneously satised at the holder inlet port.
Note 2: A large braid-reinforced tube of ø22 x ø15 is recommended for supply due to the ow rate.
1.1.2PH8HFF(ExplosionproofType)
The holder is used only when using Ultrasonic cleaning system in the explosionproof area.
Use PH8HF when using no cleaning, jet cleaning or brush cleaning.
Applicable sensors:
General pH Sensor PH8ERP, PH8EFP
General ORP Sensor OR8ERG, OR8EFG
Mounting: 2-inch pipe mounting vertical or horizontal, with 1 set of mounting bracket.
Note: Make sure the mounting pipe is rmly installed.
Cleaning method: Ultrasonic cleaning
Material:
Holder; Polypropylene or stainless steel (equivalent to SUS316)
O-ring; Fluoro rubber (FKM) or Peruoroelastomer (FFKM)
Mounting bracket; Stainless steel (equivalent to SUS304)
Cleaning unit (wetted parts):
Ultrasonic; Stainless steel (equivalent to SUS316), titanium or Hastelloy C
Construction:
TIIS ameproof type (for d2G4 gas) Cable entrance port of terminal box; G 3/4
Weight:
Holder; Approx. 3 to 3.2 kg (polypropylene)
Approx. 5.6 to 7.6 kg (stainless steel)
Mounting bracket; Approx. 0.5 kg
Temperature range: -5 to 80°C
Note: The temperature may be limited by the specications of the sensor.
Flow rate: 3 to 11 L/min
Note: The ow rate may be limited by the specications of the sensor.
Pressure:Atmospheric pressure to 500 kPa
Note: The pressure may be limited by the specications of the sensor.
IM 12B07N01-01E
< 1. Specications >
1.2ModelandSuxcodes
1.2.1 Flow-Through Type Holder PH8HF
1-3
O-ring
Option Code
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
................
...................
/MF1
/PF
105
100
80
60
(°C)
40
20
Temperature
Description
Flow-through type holder
Polypropylene (Refer to note below for selection)
Stainless steel
Heat-resistant PVC
Titanium
Rc1
1 NPT female thread
JIS 10K 25 FF (*6)
ANSI Class 150 1 FF ange (for polypropylene holder -PP) (*6)
ANSI Class 150 1 RF ange with serration (for SUS316 holder -S3)
None
For ultrasonic cleaning (Transducer: SUS316) (*1)
For ultrasonic cleaning (Transducer: Titanium) (*2)
For ultrasonic cleaning (Transducer: Hastelloy C) (*3)
For jet cleaning. The solenoid valve must be specied separately
For brush cleaning. The solenoid valve must be specied separately
None
1m
3m
7m
10m
15m
20m
Rc1/2
1/2 NPT
Style A
Mounting bracket (stainless steel) (*5)
Peruoroelastomer (FFKM) (*4)
Stainless Steel
Polypropylene
(or Stainless Steel)
0 100
200 300
Pressure (kPa)
400 500
ModelSuxCode
PH8HF
Material (*7)-PP
Process Connection -JPT
pH Measuring System-T................ Always -T
Cleaning System -NN
Cable Length for Ultrasonic Cleaning
Connector for Jet or Brush Cleaning
Style Code*A
Option Special Mounting
*1: General purpose (Normal pH 3 to 14)
*2: For salt water
*3: For acid (Normal pH 0 to 4)
*4: Choose Peruoroelastomer (FFKM) when this holder is used in organic solvent, high alkali or high temperature alkali.
*5: Mounting bracket is generally not required when the stainless steel holder is installed in-line in a pipe
It is required where the holder is installed in a sampling rack (in which case the U-bolt included in /MF1 in not used).
*6: Only mating dimensioms are according to ange standard.
*7: Criteria for material selection (-PP or -S3)
In general, polypropylene is recommended from the viewpoint of chemical resistance.
However stainless steel is recommend in any of the following cases:
• The liquid contains organic reagent, oxidizing agents, etc., which can attack polypropylene.
• The temperature/pressure correlation of the process condition falls within the hatched area of the diagram shown right.
• The use of polypropylene is not reasonable from a viewpoint of strength or past experience.
For stainless steel, normally a 3 to 14 pH value is recommended.
*1: General purpose (Normal pH 3 to 14)
*2: For salt water
*3: For acid (Normal pH 0 to 4)
*4: Use PH8HS for no cleaning, Jet cleaning or Brush cleaning.
*5: Choose Peruoroelastomer (FFKM) when this holder is used in organic solvent, high alkali or high temperature alkali.
*6: Mounting bracket is generally not required when the stainless steel holder is installed in-line in a pipe
It is required where the holder is installed in a sampling rack (in which case the U-bolt included in /MF1 in not used).
*7: Criteria for material selection (-PP or -S3)
In general, polypropylene is recommended from the viewpoint of chemical resistance.
However stainless steel is recommend in any of the following cases:
• The liquid contains organic reagent, oxidizing agents, etc., which can attack polypropylene.
• The temperature/pressure correlation of the process condition falls within the hatched area of the diagram shown right.
• The use of polypropylene is not reasonable from a viewpoint of strength or past experience.
For stainless steel, normally a 3 to 14 pH value is recommended.
Select the material of wetted parts with careful consideration of process characteristics.
Inappropriate selection may cause leakage of process uids, which greatly aects facilities.
Considerable care must be taken particularly in the case of strongly corrosive process uid such
as hydrochloric acid, sulfuric acid, hydrogen sulde, and sodium hypochlorite. If you have any
questions about the wetted part construction of the product, be sure to contact Yokogawa.
IM 12B07N01-01E
< 1. Specications >
1.3ExternalDimensions
1. PH8HF - PP - ¨PT - T - NN - NN, PH8HF - S3 - ¨PT - T - NN - NN UNIT: mm