YOKOGAWA PH200 Instruction Manual

IM 12B6C2-E-H
6th edition
Instruction Manual
Model PH200 2-wire pH transmitter
FAIL
HOLD
MEASURE AUT.CAL MAN.CAL DISPLAY TEMP.
YES
NO
ENT
OUTPUT SET HOLD SERVICE
*
NO MODEYES
>
ENT>
MODE
TEMP.MAN
HOLD
pH mV
OUTPUT HOLD
FLAG
MEASURED
VALUE DISPLAY
MESSAGE DISPLAY
KEY PROMPT FLAGS
SELECTION KEYS YES : Accept setting NO : Change to new setting
ADJUSTMENT KEYS > : Choose digit for adjustment ^ : Adjust digit (to decrease pass
through zero)
ENT : Enter new value
NOTE: The first digit changes from 1, -1, - to blank
SELECT MODE MEASURE/MAINTENANCE
Can be used to escape program at any time
MANUAL TEMPERATURE
COMPENSATION FLAG
FAIL FLAG
MENU POINTER FLAGS
MENU FOR MAINTE­NANCE FUNCTIONS (see chapter 5)
MENU FOR COMMIS­SIONING FUNCTIONS (see chapter 4)
SELECT MODE MEASURE/COMMISSIONING
IM 12B6C2-E-H
HOLD
pH mV
TEMP.MAN
NO MODEYES
>
YES
FAIL
NO
ENT
MODE
MEASURE AUT.CAL MAN.CAL DISPLAY TEMP. HOLD
ENT>
OUTPUT SET HOLD SERVICE
*
IM 12B6C2-H-H

CONTENTS

1. INTRODUCTION..................................................................1
1-1. Application......................................................................1
1-2. Required components for pH measurement...................1
1-3. Identification ...................................................................1
2. TECHNICAL SPECIFICATIONS...........................................2
2-1. General specifications .....................................................2
2-2. Functional description .....................................................3
3. INSTALLATION AND WIRING .............................................4
3-1. Installation and dimensions.............................................4
3-1-1. Installation site ...................................................4
3-1-2. Mounting methods.............................................4
3-2. Wiring of sensors............................................................6
3-2-1. General precautions...........................................6
3-2-2. Additional precautions for installations in
hazardous areas ................................................6
3-2-3. Liquid earth........................................................6
3-2-4. Access to terminal and cable entry....................6
3-2-5. Connection diagram for sensors........................7
3-3. Wiring of power supply...................................................8
3-3-1. General precautions...........................................8
3-3-2. Additional precautions for installations in
hazardous areas ................................................8
3-3-3. Connection of the power supply........................8
3-3-4. Switching the instrument on...............................8
4. COMMISSIONING..............................................................11
4-1. Operations overview .....................................................11
4-2. Output range adjustment.....................OUTPUT...........12
4-3. Set up the HOLD function ...................SET HOLD ........14
4-4. Sensor selection and diagnostics..................................16
4-4-1. Selection of measurement and reference
electrode..........................................................16
4-4-2. Selecting a temperature sensor .......................16
4-4-3. Off-line calibration checks................................16
4-4-4. On-line impedance chacks ..............................16
5. MAINTENANCE..................................................................17
5-1. Automatic calibration...........................AUT.CAL...........18
5-2. Manual calibration ...............................MAN.CAL .........20
5-3. Selecting a value to display .................DISPLAY ...........22
5-4. Temperature compensation.................TEMP ...............24
5-5. Hold output function.....................................................26
6. TROUBLE SHOOTING .......................................................28
6-1. Introduction ..................................................................28
6-2. Error messages and explanation...................................29
IM 12B6C2-E-H
7. SERVICE MODE.................................................................30
7-1. Introduction.................................................................30
7-2. Access to service settings...........................................31
7-3. Temperature function...................................................32
7-4. Temperature sensors...................................................32
7-5. Sensor checking..........................................................33
7-6. Display resolution ........................................................34
7-7. Signalling of fail condition ............................................34
7-8. Stabilization check .......................................................34
7-9. Auto return ..................................................................35
7-10.Buffer tables................................................................35
7-11.Temperature adjustment..............................................36
7-12. Manual adjustment of ITP, slope and
asymmetry potential....................................................36
7-13. Passwordprotection by three digit code......................37
7-14. Restore default settings ..............................................37
8. CLASSIFICATION...............................................................38
8-1. Cenelec........................................................................38
8-2. FM................................................................................39
9. CHANGE FROM PH TO ORP MEASUREMENT ..............40
9-1. How to change from pH measurement to
ORP measurement.......................................................40
9-2. Commissioning the transmitter .....................................40
9-3. Maintenance of the transmitter .....................................40
9-4. Special features............................................................40
9-5. Wiring diagram .............................................................41
9-6. Buffer values .................................................................41
APPENDIX A............................................................................42
IM 12B6C2-E-H
1-1. Application
The EXA PH200 transmitter is a 2-wire pH instrument intended to be used in industrial installations in the field. For easy maintenan­ce and operation it should be located close to the sensors. It is powered from a remote low voltage DC power supply through the 2­wire connection. The EXA PH200 is compa­tible with most commercial available pH sen­sors and fitting systems. The instrument is available in two versions:
- A general purpose version for use in safe areas.
- An intrinsically safe version for use in hazardous areas. The instrument can then be installed in Zone 1 with the sensors in Zone 1 or Zone 0.
The micro-processor is used in this instru­ment for continuous sensor diagnostics, flexible on site commissioning and fine tu­ning by advanced functions.
In general a pH loop can be set up for diffe­rent purposes:
- To be part of a total process control sys­tem
- To indicate dangerous limits of a process.
- To monitor product quality.
1-2. Required components for pH
measurement
1. A pH electrode and a reference electrode or a combined pH-reference electrode.
2. A temperature sensor Pt100 (according to IEC 751 or DIN) or Pt1000
3. A fitting for the above electrodes with accessories
4. Signal cables with or without extension boxes, etc.
5. The EXA PH200 2-wire transmitter with mounting accessories for wall or pipe mounting
6. A DC power supply (nominal 24 V DC) with cabling and optional zener barriers or an intrinsic safe power supply
7. Peripherals: e.g. strip-chart recorder, panel indicator, PID-controller.
1-3. Identification
The instrument has an identification label, which is fixed to the front plate. This serves as a reference for the full model code, power supply voltage and serial num­ber. This label also carries authorised marks to certify compliance with the current regula­tory norms.
1

1. INTRODUCTION

MODEL SERIAL NO. SUPPLY
2-1. General specifications A. Intrinsic safety
– BASEEFA : Certified by and meets the require-
ments of EEx [ia] ib IIC T4 of CENE­LEC Certificate nr. : 89C2379)
– FM : For IS CL1, DIV1, GP ABCD
T3B for TA-30 to 70 °C T4 for TA-30 to 40 °C Approval report: J.I. 1Y1A7.AX
– CSA : For Ex[ia] Class 1, Div. 1, Groups C
and D, T4A Approval file: LR 102851-1
B. Indicating range :-1 to +15 pH. C. Transmission signal : 4 - 20 mA DC; max. load 550 D. Transmission range : 0 - 14 pH.
User programmable to any range with a minimum span of 1 unit.
E. Power supply
– Model PH200G : 17 - 40 V DC,
dependent on load
– Model PH200S : 17 - 31,5 V DC,
dependent on load
F. Climatic conditions
– Ambient temperature : -10 to +55 °C – Storage temperature : -30 to +70 °C – Relative humidity : 10 to 90% – Weather protection : Rain and dust-tight IP65 (NEMA 4X)
G. Housing
– Material : Cast aluminium with chemical resi-
stant coating – Window : Flexible poly-carbonate – Colours : Moss green/off-white – Cable glands : Polyamide – Cable entries : Two
h
” glands
Hose connection optional – Earth connection : For external ground
H. Mounting possibilities
– Surface mounting : Two M6 bolts, 9 mm length – Wall/Pipe mounting : With optional mounting kit – Sunlight protection : With optional hood – Panel mounting : 154 x 172 octagonal with
30 mm corners (fixing with self-tap-
ping screws
I. Shipping details
– Dimensions : 162 x 178 x 115 mm – Package : 225 x 225 x 220 mm – Weight : ca. 2,5 kg
J. Safety and security
– Data protection : EEPROM memory – Interference test : According IEC 801 – Power down : No affect, reset to measurement

2. TECHNICAL SPECIFICATIONS

IM 12B6C2-E-H
2
IM 12B6C2-E-H
2-2. Functional description
The EXA PH200 is a real time micro-control­ler operated pH analysing system. It uses a dedicated micro-controller with exclusively developed software to control all functions necessary in such a system. The input and output functions are concentrated in the analogue section of the instrument. Even these functions are operated through special interfaces designed to give a minimum of interference problems to the digital func­tions. The digital and logic functions are designed to operate securely. The unique FAIL-function can give a warning when the micro-controller has found a fault in the measuring loop. This is accomplished by sending a discrete 22 mA signal on the out­put. The micro-controller checks the working of the analogue to digital input converter cir­cuit and watches over the conversion to the output signals. By using non-volatile memory (EEPROM) for the essential in formation the operating parameters are fixed in memory without the need for a battery. The software is designed with the user in mind. It uses a simple step by step, question and answer style to communicate with the operator by giving messages on the second line of the display and indicating which keys are to be pressed in the display too.
The user-interface is limited to a basic set of 6 keys accessible through the flexible win­dow cover. The keys are scanned conti­nuously and actions are taken immediately. An extensive system of checking va lues and parameters is implemented in the software . The 2-wire design gives maximum flexibility of operation with other industrial instruments and controlling systems. An intrinsically safe version is available for hazardous areas. The rugged housing is designed for outdoor use (IP65). In addition it is shaped in a way that indoor panel mounting is also possible. The optional universal mounting kit includes brackets for all mounting methods. The commissioning of the instrument can be performed on site or in advance. A three level operating system keeps the approach clear and distinguishes between every day operation and maintenance against one time operations like commissioning or fine-tuning. All this makes the EXA PH200 operate like a normal analogue transmitter with a number of additional functions.
These extra functions are only possible by using a micro-controller at the heart of the system:
• Sensor checking durig measurement (1)
• Simplified calibration through a semi­automatic sequence.
• Additional process temperature compen­sation can be activated.
• Simple range adaptation making it a ver­satile instrument.
• Wide selection of temperature sensors.
• A HOLD-function for the output.
• Three buffer tables can be programmed.
• Programmable isothermal point of inter­section.
• Protection against unauthorized access by a passcode.
(1) European patent No. 0241601
US patent No. 4777444
3

3. INSTALLATION AND WIRING

IM 12B6C2-E-H
3-1. Installation and dimensions
3-1-1. Installation site
As the transmitter is a rain-tight type it can be installed outdoors as well as indoors. It should, however, be installed as close as possible to the sensors avoiding long cable lengths between sensors and transmitter. The certified version can be installed in Zone 1. Select an installation site where:
- mechanical vibrations and shocks are negligible;
- no relay/power switches are in the direct environment;
- the transmitter is not mounted in direct sunlight and severe weather conditions;
- maintenance activities are possible (no corrosive atmospheres).
The ambient temperature and humidity should be within the limits of the specifica­tions .
3-1-2. Mounting methods
The EXA PH200 transmitter has universal mounting possibilities:
- panel mounting using selftapping screws
- surface mounting on a plate (by bolts from the back)
- wall mounting on a bracket (e.g. thick brick wall);
- pipe mounting using a bracket on a hori­zontal or vertical pipe (maximum diame­ter 50 mm).
4
Panel mounting
Surface mounting
DimensionsCut-outs for panel mounting
2 x M6 (Screws)
Unit: mm (“)Unit: mm (“)
154 (6.06`)
162 (6.4)
180 (7)
77
115 (4.5)
30 1.2)
30 (1.18)
172(6.77)
2x ø4
min. 203 (8)
min. 229 (9)
IM 12B6C2-E-H
5
Option /M: Universal wall/pipe mounting kit
Pipe mounting (vertical)
Pipe mounting (horizontal)
2” Pipe
2” Pipe
Wall mounting
115
(4.53)
70 (2.75)
56 (2.2)
200
(7.87)
2x ø6.6
(0.26)
4x ø10
(0.4)
77 (3.03)
IM 12B6C2-E-H
3-2. Wiring of sensors
3-2-1. General precautions
Generally, transmission of signals from pH sensors is at a very low voltage and high impedance level. Thus a lot of care must be taken to avoid interference. Before connec­ting sensor cables to the transmitter make sure that next conditions are met: – the sensor cables are not mounted in
tracks together with high voltage and or power switching cables
– only standard coaxial electrode cables or
extension cable are used
– the transmitter is mounted within the dis-
tance of the sensor cables (max. 10 m)
– the setup is kept flexible for easy inser-
tion and retraction of the sensors in the fitting.
3-2-2. Additional precautions for instal-
lations in hazardous areas
Make sure that the total of capacitances and inductances connected to the input termi­nals of the EXA PH200 do not exceed the limits given in the certificate. This sets a limit to the cable and extensions used. – Grounding:
• If the sensors are mounted in a grounded fitting, the external earth connection on the left hand side of the transmitter must be connected to "protective" earth.
• If the sensors are mounted in a fitting iso­lated to ground, it is recommended (not required) to connect the earth connection on the left hand side of the transmitter to "protective" earth.
– The intrinsic safe version of the EXA
PH200 instrument can be erected in Zone 1.
– The sensors can be installed in Zone 0 or
Zone 1 if a safety barrier according to the limits given in the system certificate is used.
3-2-3. Liquid earth
In all circumstances, the sensor side of the measuring loop must be grounded to the measuring liquid. The EXA PH200 uses advanced differential high impedance input circuits. This technique calls for a grounding to the liquid. In addition to that the sensor checking circuits also use the liquid earth for measurement of impedance of the sensors. All Yokogawa fittings have provisions for this connection. It is usually called liquid earth in all our manuals. A separate connection should be made to the terminal numbered 14 in all cases to get a proper and stable measuring loop.
3-2-4. Access to terminal and cable
entry
1. To access terminals remove the front cover of the EXA PH200 by releasing the 4 captive screws.
2. Thread the sensor cables into the con­nection space and connect the cables to the terminals as indicated in the wiring diagram. Make sure all connections are firm and do not touch each other.
3. Screw the gland securely and tighten it to keep out moisture. DO NOT use a wrench to tighten the nut.
4. The optional hose connection is used to guide the cables coming from an immer­sion fitting through a protective plastic tubing to the transmitter.
6
IM 12B6C2-E-H
7
3-2-5. Connection diagram for sensors
+
-
G 11 12 14 13 17 15
16
Rd Bl Rd Bl Rd
Bl
+
-
G 11 12 14 13 17 15
16
Rd Bl Wt RdBl
+
-
G 11 12 14 13 17 15
16
+
-
G 11 12 14 13 17 15
16
T
1
T
2
SE RE GE S
Rd=Red, Bl=Blue, Wt=White,
Single electrodes
Temperature Combined
Liquid earth
Temperature - Reference pH Pt 100 AgCl glass (Pt 1000) (other) (other)
EXA compact and retractable electrodes
pHCombined electrode
Combined electrodes
Liquid earth
Connect the numbered cable leads to the corresponding terminals
IM 12B6C2-E-H
3-3. Wiring of power supply
3-3-1. General precautions
Do not activate the power supply yet. First make sure that the DC-power supply is according to the specifications given.
WARNING: DO NOT USE ALTERNATING CURRENT OR MAINS POWER SUPPLY! !
The cable leading to the distributor (power supply) or safety barrier transports power to and output signal from the transmitter. Use a two conductor shielded cable with a size of at least l.25 mm
2
and an outside diameter of 9 to 15 mm. The cable gland supplied with the instrument accepts these diameters. The maximum length of the cable is 2000 metre. This ensures the minimum operating voltage for the instrument.
3-3-2. Additional precautions for instal-
lation in hazardous areas
1. Ensure that the total of capacitances and inductances connected to the terminals of the EXA PH200 do not exceed the limits given in the certificate of the safety barrier or distributor.
2. The cable used should preferably have a BLUE colour or marking on the outside.
3. Grounding:
• If the transmitter is mounted on a
grounded surface (e.g. a metal frame fixed in the soil) the shield of the 2-wire cable may NOT be connected to ground at the distributor.
• If the transmitter is mounted on a non-
conducting surface (e.g. a brick wall) it is recommended to ground the shield of the 2-wire cable at the distributor end.
4. Installation for EEx ia (sensors in Zone 0 or 1): Generally, the distributor with input/out­put isolation has no external earth con­nection. If there is an earth connection on the distributor and the external connec­tion of the transmitter is connected to "protective" earth, the shield of the 2-wire cable may NOT be connected to "protec­tive" earth at the distributor too.
3-3-3. Connection of the power supply
The terminal strip is accessed as was described in § 3-2-4. Use the left-hand gland to insert the 2-wire cable to the trans­mitter. Connect the supply to the terminals mar ked +, - and G as is indicated in the figures on pages 9 & 10.
3-3-4. Switching the instrument on
After all connections are made and checked, the power can be switched on from the dis­tributor. Observe the correct activation of the instrument at the display. If for any reason the display does not indicate a value, con­sult the trouble shooting section.
8
IM 12B6C2-E-H
Dependance of load to supply voltage
9
Wiring diagrams for power supply
General purpose design pH-transmitter EXA PH200G
Sensor
Distributor
Output
Supply
NOTE: The outside earth terminal should be connected to site ground by a large area conductor (e.g. a flat earth strip) for best protection against interference.
1200
1000
800
600
400
200
0
1718 20 24 40
Ohm
550
Supply voltage (V)
Load resistance ()
+ –
+ –
G
Wiring diagrams for hazardous areas
IM 12B6C2-E-H
10
Intrinsically safe design (CELENEC standard EEX ib [ia] IIC T4) pH-transmitter EXA PH200S
Intrinsically safe design (CELENEC standard EEX ib [ia] IIC T4) pH-transmitter EXA PH200S
EEx ib Certified safety barrier
EEx ib Certified distributor with input/output isolation
Electronic current limiting barrier Vmax: 31.5 V lmax: 35 mA Pmax: 1,1 W Zenerbarrier with resistor Vmax: 28 V lmax: 93.3 mA Pmax: 0.66 W
Shunt zener barrier or supply unit or isolated repeater Vmax: 22 V Imax: 85 mA
Sensor
Distributor
Output
Supply
Sensor
Output
Supply
Protective earthProtective earth
Protective earth
Hazardous area Safe area
Zone 0 or 1 Zone 1
Hazardous area Safe area
Zone 0 or 1 Zone 1
+ –
+ –
G
+ –
+ –
G
IM 12B6C2-E-H
4-1. Operations overview
NOTE: All 3 levels can be separately protected by a password (see §7-13).
11

4. COMMISSIONING

Routine Use Chapter
MAINTENANCE AUT.CAL Calibration with buffer solutions 5-1 Operation by keys through MAN.CAL Calibration in process 5-2 closed cover DISPLAY Show or fix additional values 5-3
TEMP Select automatic or manual
temperature compensation 5-4
HOLD Switch HOLD-function on/off 5-5
COMMISSIONING OUTPUT Adjusting the output function 4-2 Operation by *-key when SET HOLD Activating the HOLD-function 4-3 cover is removed
SERVICE SERVICE Finetuning the performance 7 Operation by coded entry from commissioning
Loading...
+ 33 hidden pages