Yokogawa FT5V User Manual

User's Manual
1
Thank you for purchasing the JUXTA Signal Conditioner. Please read through this manual before use for correct handling.
Network Solutions Business Division
IM 77J08T05-01E
1st Edition Jan. 2007 (YK)
CAUTIONARY NOTES FOR SAFE USE OF THE PRODUCT
This User’s Manual should be carefully read before installing and operating the product. The following symbol is used on the product and in this manual to ensure safe usage.
Th is sym bol is dis pla yed on the pr oduct w hen it is
necessary to refer to the User's Manual for information
on pe rso nal and ins trume nt safet y. Th is sy mbo l i s
displayed in the User's Manual to indicate precautions
to a void dan ge r t o the op er ator, suc h as an elect ri c
shock.
The following symbols are used only in this manual.
NOTE
D r a w s a t t e n t i o n t o e s s e n t i a l i n f o r m a t io n f o r
understanding
the operations and/or functions of the product.
CHE CKING PRODUCT SPE CIFICATION S AND PACKAGE
(1) Checking the Model and Product Specications
Check that the model and specifications indicated on the
nameplate attached to the main unit are as ordered.
(2) Packaged Items
Check that the package contains the following items:
FT5A or FT5V: 1
Mounting block: 2
Tag number and range label : 1 each
Mounting screw (M4 x 0.7): 2
User’s Manual (this manual : IM 77J08T05-01E)
GENERAL
The FT5A/FT5V is a free range type thermocouple converter that
is connected to an IEC/JIS-standard thermocouple (TC), such as
a Type K, T, E, J, R, S, B or N thermocouple to convert tempera-
ture signals into isolated
DC current or DC voltage signals.
Selection of input type (sensor type), temperature unit (°C, K), I/
O range setting, burnout setting, output adjustment, RJC on/off
and I/O monitoring can be made using the optional Parameter
Setting Tool (VJ77) or Handy Terminal (JHT200).
MODEL AND SUFFIX CODES
Model Sufx codes Description
FT5A -
 
*C /Thermocouple converter
(Free Range Type)
FT5V Input -1 K (CA)
-2 T (CC)
-3 E (CRC)
-4 J (IC)
-5 R
-6 S
-7 B (RH)
-8 N
-0
Custom order
Output A 4 to 20 mA DC FT5A
B 2 to 10 mA DC C 1 to 5 mA DC D 0 to 20 mA DC E 0 to 16 mA DC F 0 to 10 mA DC G 0 to 1mA DC Z Custom order 1 0 to 10 mV DC FT5V 2 0 to 100 mV DC 3 0 to 1 V DC 4 0 to 10 V DC 5 0 to 5 V DC 6 1 to 5 V DC 7 -10 to 10 V DC
0 Custom order Style code *C Style C Burnout /BU Up
/BD Down /BN None
Temperature unit /DF Fahrenheit (°F) display function
Models FT5A, FT5V Thermocouple Converter (Free Range Type)
2
IM 77J08T05-01E 1st Edition 2007.01.15-00
1. MOUNTING METHOD
1.1 Installing/Uninstalling the Mounting Block
The mou nt in g bloc k is necessary for rack mounting or w al l
mounting.
Installing the mounting block
Fit the mounting block into the grooves of the converter,
and slide it in the direction of an arrow as shown below-
left.
Uninstalling the mounting block
Lift the stopper using a slotted screw-driver as shown
below-right, and pull the mounting block to remove it.
Groove
(Rack mounting)
Groove (Wall mounting)
Stopper
Mounting block
1.2 Rack Mounting
Take the FRK panel and install it on mounting angle as shown
below.
Th is m etho d is sui tab l e f o r h i gh dens i ty moun t ing of the
converters on a 19-inch rack panel.
Mounting angle
Mounting angle
Mounting Dimensions
450±1
465±1
M5 screw (4 places)
Supplied spacer(50) or Wiring space(30 to 60)
FRK panel
Mounting screw (M4x0.7) (2 screws in the package)
50
25
25
50
Unit: mm
1.3 Wall Mounting
Mount the converter onto the FRK panel as shown below or
mount it directly on the wall.
FRK panel
1.4 DIN Rail Mounting
Place the converter so that the DIN rail ts into the upper part of
the DIN-rail groove at the rear of the socket, and fasten the socket
using the slide lock located underneath the socket.
To remove the converter, pull down the slide lock using a slotted
screw-driver.
Slide lock
DIN rail
DIN rail groove
1.5 Mounting Dimensions
Mounting angle
M4×0.7 Mounting screw
Support
24.524.5
24.5
24.5 24.5
Angle Mounting Dimensions
n×24.5+2
M4×0.7 Mounting screw
Unit: mm
73 90
24.524.5
24.5 24.5 24.5
Wall Mounting Dimensions
90
2. INSTALLATION LOCATION
Avoid the following environments for installation locations: Areas with vibration, corrosive gases, dust, water, oil, solvents,
direct sunlight, radiation, a strong electric eld, and/or a strong magnetic eld, altitude of more than 2000m above sea level.
If there is any risk of a surge being induced into the power line and/or signal lines due to lightning or other factors, a dedicated lightning arrester should be used as protection for both this
converter and a eld-installed device.
Operating temperature/humidity range: 0 to 50C/5 to 90%RH (no condensation)
3. EXTERNAL WIRING
WARNING
Be sure to turn OFF the power supply before wiring to avoid the risk of electric shock. Use a tester or similar device to ensure that no power is being supplied to a cable to be connected.
3
IM 77J08T05-01E 1st Edition 2007.01.15-00
M4 screw terminals are provided for the connection of external signals. Attach a crimp-on lug to each wire for connection to the terminals.
●Recommended cables: A nominal cross-sectional area of 0.5
mm2 or thicker for signal cables, and that of 1.25 mm2 or thicker for power cables.
Wiring Diagram
1
2
4 5
6 7
3
6
7
4
5
RJC
Input(RJC)
Output
Power supply
2
3
DO NOT use the terminal
NOTE
●The power line and input/output signal lines should
be ins t all ed a w ay f rom nois e-g e ner a tin g so u rce s.
Otherwise accuracy cannot be guaranteed.
●A d h e re s t ri c t l y t o th e sp e c i f ic a t i on s t o a v o id
overheating or damage. Before turning on the power, ensure the following: (a ) Po wer sup ply volt age and inp ut s ign a l v alue
applied to the product should meet the required
specications.
(b) The external wiring to the terminals and wiring to
ground are as specications.
●Do n o t op e r a t e th e p r oduc t i n th e p r e sen c e o f ammable or explosive gases or vapors.
●The product is sensitive to static electricity; exercise
ca r e wh e n in oper a tion . Bef o re y o u op e rat e the pr odu ct, tou ch a ne arb y me tal pa rt t o d isch arg e static electricity.
●C onne ct t he R JC s ens o r t o th e te r min al of the converter together with the input signal line so that the crimp-on lug of the input signal line overlaps the
RJC sensor.
●Handle the RJC sensor lead wire with care to prevent
disconnection.
Power Supply and Isolation
Power supply voltage: 24V DC ±10% (ripple: less than 5% p-p) Current consumption: FT5A 65mA, FT5V 50mA
Insulation resistance: 100MΩ at 500V DC between input, output
and power supply mutually
Withstand voltage: 1500V AC/minute between input-output,
input-power supply
500V AC/minute between output-power supply
4. SETTING PARAMETERS
The parameters are set as you specied in your order. Refer
to the following to change the default settings.
Set the parameters using a PC (VJ77 Parameter Setting Tool) or the Handy Terminal. Refer to "5. List of Parameters" in this manual and the User’s Manual for VJ77 PC-based Parameters Setting Tool (IM 77J01J77-01E) or the User’s Manual for JHT200 Ha ndy Termin al (IM JF81- 02E ). Par ame ter s a re sho wn in brackets "[ ]".
Connection of Setting Tools
VJ77 Dedicated adapter (E9789HA) [Provided with VJ77]
PCwithVJ77installed
Dedicated cable (E9786WK) [Provided with VJ77]
*BesuretousetheVJ77ofversionR1.05orlater.
JHT200 Handy Terminal
JUXTA communication cable 3 pin connector (F9182ED) [Provided with VJ77 and JHT200]
Setting Thermocouple Type
Set the thermocouple type to be connected to the input in
[B06: TYPE].
Display Measuring range Measuring span
K TYPE -270 to 1372°C
3mV or more
E TYPE -270 to 1000°C J TYPE -210 to 1200°C T TYPE -270 to 400°C R TYPE -50 to 1768°C S TYPE -50 to 1768°C B TYPE 0 to 1820°C N TYPE -270 to 1300°C
Setting Temperature Unit
Set the temperature unit of the input range in [B09: UNIT].
NOTE
Do not set the "-----" as the temperature unit, even if it
is displayed during the setting.
Otherwise it may cause the malfunction or damage.
Setting Input Range
Set the zero percent value of input range in [B10: ZERO], and
the span of input range in [B11: SPAN].
Setting Burnout Action
Set the burnout action in [B12: BURN OUT].
Set "OFF", "UP", or "DOWN."
NOTE
The wire resistance compensation should be applied
when t h e s e t t i n g of i n p u t r a n g e , burno u t a c t i o n
direction or input wiring has been changed.
4
IM 77J08T05-01E 1st Edition 2007.01.15-00
5. LIST OF PARAMETERS
Parameter Display Items A DISPLAY Display
A01 INPUT Input value A02 OUTPUT Output value A03 STATUS Status (*1) A04 REV NO Revision number.
B SET Setting
B01 TAG NO.1 Tag number 1 B02 TAG NO.2 Tag number 2 B03 COMMENT1 Comment 1 B04 COMMENT2 Comment 2 B06 TYPE Input type B09 UNIT Unit B10 ZERO Zero (0% of input range) B11 SPAN Span (Input span) B12 BURN OUT Burnout
C ADJUST Adjustment
C01 OUT 0% Output 0% adjustment C02 OUT 100% Output 100% adjustment C03 WIRING R Wire resistance compensation C04 ZERO ADJ Input zero adjustment C05 SPAN ADJ Input span adjustment C08 RJC RJC on/off (*2)
*1:
This “STATUS” is for the customer’s engineer to check the history.
*2:
The RJC returns to ON after the power is turned off and then on again.
6. MAINTENANCE
The product starts running immediately when the power is turned on; however, it needs 10 to 15 minutes of warm-up before it
meets the specied performance.
6.1 Calibration Apparatus
A voltage current source (Yokogawa 7651 or equivalent) A digital multimeter (Yokogawa 7561 or equivalent)
A precision resistor of 250 Ω ±0.01%, 1W
A setting tool for adjustment (Refer to "4. Setting Parameters" in this manual.
6.2 Calibration Procedure
1. Connect the instruments as shown below.
1
2 3
4 5
6 7
3
2
3
6
7
4
5
24VDC
Ro
DMM
Voltage current source
Input
Output
Ro: 250Ω precision resistor for
current output
Setting tool
Power supply
2. Turn off the RJC using the setting tool.
3. Use the voltage current source and apply the electromotive force equivalent to 0, 25, 50, 75, and 100% of the measuring range to the converter.
4. Verify that the corresponding output voltages are 0, 25, 50, 75,
and 100% respectively and within the specied accuracy rating.
(Ro is used for current output.) Use the setting tool (VJ77 Parameter Setting Tool or JHT200 Handy Terminal) to adjust the input/output signals.
   
Turning the RJC on/off
(1) Read the parameter [C08: RJC]. (2) Select "ON" or "OFF" in the parameter
[C08: RJC].
NOTE
The RJC returns to "ON" after the power is turned off
and then on again.
Input Adjustment Procedure
(1) Input the value equivalent to 0% value of the input range
to the converter.
(2) Read the parameter
[C04: ZERO ADJ] and check the
input value.
(3) Select “INC” or “DEC” in the parameter
[C04: ZERO
ADJ] to adjust.
INC: Increase (Adjusts the increased value of the input
[voltage] value)
DEC: Decrease (Adjusts the decreased value of the
input [voltage] value) RST: Reset of the adjustment value (4) Input the value equivalent to 100% value of the input
range to the converter. (5) Read the parameter
[C05: SPAN ADJ] and check the
input value. (6) Adjust the span in the same way as (3).
Output Adjustment Procedure
When adjusting 0% value of output: (1) Set the adjustment value 0% in the parameter [C01:
OUT 0%].
•The value equivalent to 0% of the output range will be
output, irrespective of the input. (2) Check the output value via digital multimeter, and adjust
it in the parameter [C01: OUT 0%].
•If it slips out to the (+) side, set (−) value equivalent
to slipout; if it slips out to the (−) side, set (+) value
equivalent to slipout for adjusting the output value to
0%. *: The 100% value of output can be adjusted by the same
operation as the above in [C02: OUT 100%].
Wire Resistance Compensation
(1) Short-circuit at the sensing end of the eld side.
FT5A FT5V
Short-circuit at the cable end
(2) Read the parameter [C04: ZERO ADJ] and make sure
the setting is "RESET". (If the set ti ng is " EXECU TE, sel ect "RE SET". The
compensation value will be reset.) (3) If "EXEC UTE " i s s ele cte d in the pa ram eter
[C 03:
WIRING R], the wire resistance compensation will be
done automatically.
For adjustment using a setting tool, refer to the User’s Manual for each setting tool and “5. List of Parameters” in this manual. User’s Manual for VJ77 [Document No.: IM 77J01J77-01E];
however, use the VJ77 of version R1.05 or later.
User’s Manual for JHT200 [Document No.: IM JF81-02E]