7. SPARE PARTS LIST .................................................................................................21
IM 12B6K1-01E-E
1
1. INTRODUCTION
1-1. General
The flow fittings (Model FF20) and their subassemblies (Model FS20) are used to mount
sensor tips in a piping system so that the pH and/or the ORP (Redox) potential of the liquid
flowing through it can be measured.
The flow fittings are for connection between two pipes of the piping system providing a
“flow through” path. From a practical plant aspect (for easy maintenance and calibration)
the mounting place is in a by-pass behind a sample valve. They are supplied with a ring to
hold calibration dish for this cleaning and/or calibration.
The subassemblies should be cemented or welded direct in the piping system or in a
T-piece of it. The subassemblies of stainless steel meet the requirements of DIN 11850
for mounting in sanitary constructions.
1-2. Features
- Wide choice fo construction materials.
- High degree fo standardisation reduces spare holding requirements.
- Direct mounting of sensors with DIN dimensions.
- Liquid earth sensor for stable measurements.
- High pressure and temperature specifications.
- Chemical or brush cleaning (optional in 4-hole fittings).
2. UNPACKING AND CHECKING
2-1. Unpacking and inspection
When you receive the flow fitting (Model FF20) or the subassembly flow fitting (Model FS20)
it is packed in a cardboard box. Open the box and check that the model code is the same
as on the packing list (see §2-3 for the model code). Also check that it is supplied with the
options you ordered. These options are delivered in seperate bags.
If you have any problems or questions, please contact the nearest Yokogawa service center
or sales organisation for assistance.
2-2. Mounting location
The flow fitting or subassembly flow fitting
is intended to be used for pH and/or ORP
(Redox) measurement. The location should
be in a bypass of a piping system behind a
GO O D
sample valve. The angle of the fitting with the
horizontal should be less than 15°
(see fig. 2-1).
The pressure and temperature ratings of the
electrode inside the fitting should be noticed
to determine the maximum rating of the
measurement point.
Figure 2-1
GO O D
HO R IZ ON TAL
15 º
15 º
INC OR RE CT
IM 12B6K1-01E-E
2
2-3. Specificaitons
2-3-1. General specifications
MATERIALS
- Wetted parts
a. body : refer to model code
b. O-rings : silicon rubber
c. liquid earth sensor
(not in 1-hole subassembly) : titanium (plastic designs) SS AISI 316 (SS designs)
- Non-wetted parts
a. mounting bracket : SS AISI 316 (SS designs) PVC
(plastic designs)
b. electrode mounting sets : Ryton R4
c. holder for calibration dish : SS AISI 316
d. calibration dish : PP
e. retaining nut for electrode holder : SS AISI 304
VOLUME MEASURING VESSEL
- 3-hole design : 130 ml
- 4-hole design : 250 ml
PROCESS CONNECTIONS: 1/2” NPT or flange LAP-joint (DIN or ANSI).
(fittings only) See model code.
NOMINAL PIPE SIZE FOR MOUNTING SUBASSEMBLIES
- 1-hole : DN20
- 3-hole : DN50
- 4-hole : DN80
WEIGHT* : see table:
Tabel 2-1
Material PP SS PVC PVDF
Fitting
1-hole subassembly 0,2 kg 0,1 kg
3-hole subassembly 1,2 kg 0,5 kg
4-hole subassembly 3 kg 1,4 kg
3-hole fitting 1,1 kg 2,2 kg 1,5 kg
4-hole fitting 1,4 kg 6,5 kg 1,8 kg
* The accessories are not included.
IM 12B6K1-01E-E
2-3-2. Funcitonal specifications
TEMPERATURE
- min -10 °C
- max depending on material and application (see fig. 2-2)
FLOW RATE (fittings only) : 0,1 to 10 1/min (depending on applicaiton)
-KCl-reservoir/KElectrolyte tubing (2.5 m) is included.
-Salt bridge/SFor liquid which cannot stand contamination
with KCl.
-Certificate/MMaterial certificate 3.1 according to EN 10204
for wetted metal parts only
5
IM 12B6K1-01E-E
6
2-5. DIMENSIONS
2-5-1. Flow Fittings
PP/PVDF
FF20-P22
FF20-F22
PP/PVDF
FF20-P33
FF20-F33
Ø71 (2. 80 ")
Ø71 (2.80 ")
1/2" NPT
Ø 9 2
(3.62")
1/2" NPT
1/2" NPT
1/2" NPT
IN
IN
OUT
60 (2. 36")
OUT
60 (2. 36")
Ø 8 (0,31 ")
Ø 90
(3.54")
(3.54 ")
5 (2.1 6")
Ø8 (0.31 ")
Ø 9 0
SS
52 [2.05"]
255 [10.04"]
6,5 (1.43" )
76 (10 .82" )
3
2
71 [2.80"]
½"NPT
Outlet
17.5 [0.69"]
67 [2.64"]
70.3 [2.77"]
5
½" NPT inlet
=
=
76 [2.99]
175 [6.89"]
48 [1.89"]
3x59 [50.35"]
FF20-S22
SS
71 [2.80"]
205 [8.07"]
15 + 3 (8 .46" + 0.12 ")
6,5 (1.4 3 ")
2
3
5 (2. 16")
5
67 [2.64"]
½" NPT inlet
=
=
175 [6.89"]
48 [1.89"]
76 [2.99"]
52 [2.05"]
½"NPT
Outlet
17.5 [0.69"]
FF20-S33
PP/PVDF
Ø90 (3.54")
320 (12. 6")
FF20-P43
FF20-F43
IM 12B6K1-01E-E
230 (9.0 5")
280 (11.02")
1/2"NPT
130 (5.11")
SS
150 (5.90")
1/2"NPT
1/2"NPT
4XØ 12 ( 0.47" )
70 (2 .75")
225 ( 8.86 ")
OUT
Ø90 (3.54")
")
4XØ12 (0.47
OUT
90 (3.54")
20 ( 4.72 ")
90 ( 3.54 ")
1
15 (4.52")
1
1/2"NPT
157 (0.18")
IN
80 (3 .15")
IN
265 (10.43")
Distance wall-outlet: 85 (3.35”)
FF20-S43
160 (6.30")
2-5-2. Subassamblies
SSSSPP/PVDFPP/PVC/PVDF
169 (6.65")
60 (2.36")
112 (4.41")
Ø56 (2.20")
124(4.88")
112 (4.41")
2-11,5NPT
ANSI B.20.1
FS20-S22-WEFS20-S22-TP
190(7.48")
82(3.23")
112(4.41")
2-11.5 NPT
ANSI
FS20-P22-TP
FS20-F22-TP
200(7.87")
92(3.62")
112 (4.41")
B.21.1
SSSSPP/PVDFPP/PVC/PVDF
124 (4.88")
53 (2.08")
105 (4.13")
Ø 56 (2.20")
105 (4.13")
2-11,5 NPT
ANSI B.20.1
FS20-S32-WEFS20-S32-TP
PVC
146 (5.75")
75 (2.95")
105 (4.13")
FS20-P32-TP
FS20-F32-TP
2-11,5 NPT
ANSI B.20.1
138 (5.43")
70 (2.75")
27
105
SS
Ø 63 (2.48")
FS20-P22-WE
FS20-V22-WE
FS20-F22-WE
Ø 63 (2.48")
FS20-P32-WE
FS20-V32-WE
FS20-F32-WE
189(7,44")
81.5(3.21")
136 (5.35")
66 (2.60")
26 (1.02")
105 (4.13")
89 (3.50")
45 (1.77")
106 (4.17")
Ø 88 (3 .46")
Ø 90 (3 .54")
FS20-V43-WEFS20-S43-WE
IM 12B6K1-01E-E
129 (5.08")
58 (2.28")
8
12B6K1-41
MOUNTING IN A
SHUNT LINE
12B6K1-40
MOUNTING IN A
SAMPLE LINE
3. PIPING AND INSTALLATION
3-1. General
It is important that, whatever method of mounting is used, the point of measurment is
truly representative of the entrie solution. Avoid an area where the measurement varies
significantly or the flow can be interrupted (the sensors must always be immersed in the
process liquid). The recommended type of fitting or subassembly will depend on pressure,
temperature, kind of liquid, pollution, etc. Check whether the specifications of the fitting/
subassembly and electrodes fulfil the maximum occuring process conditions. The fitting/
suassembly has several options and optional connection possibilities. Check that you
received the correct size and type.
3-2. Piping
3-2-1. Mounting in a sample line
Part of the process liquid is tapped off to
PROCESS PIPING
the flow fitting or subassembly (see fig. 3-1)
DRAIN
.
Figure 3-1
3-2-2. Mounting in a shunt line
By means of restrictor in the main line a small flow of the process liquid is lead through
the flow fitting or subassembly via the shunt line (see fig. 3-2). The pressure in the fitting or
subassembly is controlled by means of two valves.
RESTRICTOR
P1
PROCESS PIPING
P2
P2 < P1
Figure 3-2
IM 12B6K1-01E-E
3-2-3. Mounting in a sample line with extra connection for buffer liquid/cleanser
12B6K1-42
MOUNTING IN A
SAMPLE LINE
+CONNECT. FOR
BUFFER...
12B6K1-44
MOUNTING THE FITTING
12B6K1-43
MOUNTING THE FITTING
12B6K1-45
LOCKING SCREWS
The sample line has an extra input for cleaning or calibration (see fig. 3-3).
PROCESS PIPING
9
buffer
liquid
cleaner
Figure 3-3
3-3. Installation
3-3-1. Mounting the fitting
Install the fitting at a convenient location for maintenance and calibration.
Ensure that there is some place at the top of the fitting (approx. 20 cm) for mounting or
replacing the electrodes.
Mounting the fitting in a piping system is shown in fig. 3-4.
Refer to §2-4 for dimensional drawings.
DRAIN
Figure 3-4 Figure 3-5
NOTE: The process connection (liquid outlet) can be turned from right to left.
1. The plastic fittings have locking screws for the holder of the calibration dish
(see fig. 3-5). These screws must be loosened before turning.
2. The liquid outlet of the stainless steel fittings can be turned after loosening from the
mounting bracket (see exploded view at the end of this manual).
IM 12B6K1-01E-E
10
3-3-2. Mounting the subassembly
The subassembly can be cemented or welded directly in a piping system or in a T-piece of
it. Install the subassembly at a convenient location for maintenance and calibration.
Ensure that there is some place at the top of the fitting (approx. 20 cm) for mounting or
replacing electrodes.
Fig. 3-6 shows some mounting examples. Select the mounting position so that the sensors
are immersed in the process liquid during measurement.
(Model WU20) should be used (see fig. 3-7).
GLUE
GLUE
GLUE
3-hole (DN50)
4-hole (DN80)
PVC
Figure 3-6
3-4. Sensor mounting
3-4-1. General
For stable measurement the glass- and the reference electrode should be mounted in the
holes nearest to the earth connection.
When mounting sensors of Yokogawa in a fitting or subassembly the electrode cables
(Model WU20) should be used (see fig. 3-7).
screening
TRIAX-cable (model WU20-LT) for combined electrode
colour
marker
Cable nut
10 cm
COAX-cable (model WU20-PC) for single electrode
colour
marker
colour
colour
outer
marker
inner
marker
screening
core red
white
core red
blue
Figure 3-7
IM 12B6K1-01E-E
Cable nut
10 cm
screen
blue
3-4-2. Mounting in a 1-hole design
12B6K1-21
1-HOLE MOUNTING
12B6K1-22
MOUNTING IN A 3- AND
4-HOLE DESIGN
Unscrew the electrode mounting set and fix the
cable and electrode as shown in fig. 3-8.
3-4-3. Mounting in a 3- and 4-hole design
Unscrew the nut from the top end of the fitting
or subassembly. Fix the electrodes and cables
as shown in fig. 3-9.
Note: In the holes for placing the electrodes are
blanking plugs (stoppers). These plugs
can be placed in unused holes after
pushing the two O-rings over the bottom
end.
ATTENTION
To prevent fouling of the contact new
Yokogawa electrodes are delivered with a cap.
It is recommended to remove the cap just
before fixing the cable.
ENSURE THAT THERE IS NO MOISTURE OR
DUST IN THE PLUG
Figure 3-8 Figure 3-9
11
4. ACCESSORIES
4-1. General
For mounting the non-DIN sized sensors in the
fittings and subassemblies Yokogawa added a
range of accessories to the program of fittings
which can be indespensable tools for optional
measuring tools (the specifications of the
accessories are on seperate sheets, see GS
12B6J1-E-E).
See the chapters below for mounting the
accessories.
4-2. Mounting kit for BELLOMATIC
electrodes (Model FP20-S13)
Using this mounting kit a BELLOMATIC
electrode of Yokogawa can be fixed in a fitting
or subassembly (see fig. 4-1).
This mounting kit can also be used for fixing
combined electrodes.
Figure 4-1
Electrode
Ring
Ring
O-rings
(11 X 3)
IM 12B6K1-01E-E
12
RECOMMENDED ELECTRODES1)
TYPE DESCRIPTION
SR20-AC32 Reference electrode
SC29-PTC542) Combined redox/ref. electrode
1
) For specifications of the electrodes see GS 12B6J1
2
) Atmospheric conditions only
PARTS AND ACCESSORIES
ORDER NR. DESCRIPTION
K1500HC Set rubber ring (10x)
K1500GE Set O-rings (5x) for BELLOMATIC electrode
K1500GF
K1520VA
K1500GG
K1520VN
250 ml KCI-solution (1 m.)
250 ml KCI-solution (3,3 m.)
250 ml KCI-solution (1 m.) thickend
250 ml KCI-solution, (3,3 m.) thickend
4-3. Mounting kit for sterilisable
applications (Model FP20-S14)
To assure sterilisable applications and a
constant flow of electrolyte this mounting kit
can be used for fixing combined electrodes
with a large KCI-reservoir (see fig. 4-2).
For pressurising the system there is a 1/8”
nipple for connecting to each standard
pump. The pressure inside the holder may
never exceed the specifications of the
electrodes. See GS 12B6J1 for these
specifications.
Tape
ELECTRODES TO BE USED1)
TYPE DESCRIPTION
SC21-AAC54 Combined pH/ref.
electrode
O-rings
(K1500GS)
SC29-PTC54 Combined redox/ref.
electrode
1) For specifications see general
specifications sheet GS 12B6J1.
PARTS AND ACCESSORIES
ORDER NR. DESCRIPTION
K1500GS Set O-rings
K1520VN 250 ml KCI-solution,
(3,3 m.) thickend
Figure 4-2
IM 12B6K1-01E-E
Electrode
O-rings (11x3)
(K1500BZ)
4-4. Mounting kit for refillable ref.
KCI-reservoire
(K1500FU)
Silicone tubing
(K1500GA)
O-rings
(11 x 3
(K1500BZ)
)
electrodes
This mounting kit replaces the standard
electrode mouting set so that elctrodes of
Yokogawa with a long glass shaft can be fixed
in the fitting or subassembly (see fig. 4-3). This
mounting kit may be used as process pressure
up to 3 bar. Higher pressure ratings require
the standard mounting set for electrodes with
DIN dimensions or, alternately a salt bridge
(see §4-5). The refillable electrodes can be
connected to a KCI-reservoir via a silicone
tubing (see fig. 4-3). The pressure on the KCIsolution and consequently, the liquid outlet
can be increased by mounting the reservoir
above the fitting or subassembly. This prevents
penetration of the process liquid into the
electrode.
Note: The pressure on the KCI-solution must
exceed the liquid pressure. Figure 4-3
In §7 you will find the spareparts
4-5. Cleaning systems
4-5-1. General
The standardised design of the fittings and subassemblies makes it possible to mount
cleaning systems directly. The 4-hole types are especially designed for these applications.
13
Three different types of cleaning systems are available (see fig. 4-4):
- mechanical cleaning (electrically or mechanically driven)
- chemical cleaning
Detailed specifications of the cleaning systems are on seperate sheets
(see GS 12B6V1-E-E).
Figure 4-4
brush cleaning
chemical cleaning
IM 12B6K1-01E-E
14
12B6K1-39
SELECTION CRITERIA
K1547PP
O-rings
(11x3)
K1500BZ
4-5-2. Selection criteria
cleaning system mechanical chemical acoustical
brush acid base emulsifier ultra sonic
Applications:
oils fats x x
resins (wood, pulp) x x
emulsions of latex x
fibers (paper, textile) x
crystalline precipations
(carbonates) x xx x
amorpheus precipations x xx x
4-5-3. Mechanical cleaning
The brush of this cleaning system periodically strikes along the sensitive glass membrane
of the electrode, so that this part is wiped frequently preventing sediment formation on it.
Electrically or pneumatically versions are available.
ORDER NR. DESCRIPTION
WU20-PC Cable (to be used as a
K1520NG Brush holder
K1520NB Brush
supply cable for the motor)
K1520NF Motor unit
Figure 4-5
IM 12B6K1-01E-E
Motor
O-rings (11x 3
(K1500BZ)
Brush holder
)
Cable
WU20
Pneumatically driven brush cleaning
12B6K1-27
PNEUMATICAL CLEANING
(Model FC20-VP)
CONTROL
UNIT
PARTS AND ACCESSORIES
ORDER NR. DESCRIPTION
K1520NA Tubing (ø4 mm)
TUBING
K1520NA
K1520NG Brush holder
K1520NB Brush
K1520NG Protection hose
K1520NH Piston
K1520NJ Control unit
PISTON
O-RING (11x 3)
BRUSH HOLDER
Figure 4-6 Example FC20-VP
4-5-4. Chemical cleaning HCN.
The chemical cleaning system is effective in
processes where deposits can be removed
by a suitable solvent. Good cleaning effects
can be obtained from periodically spraying of
an electrode. EXA has a built-in wash timer
with programmable washtime, interval time
and relaxation time for automatic cleaning
of the electrodes, preventing pollution of
the pH sensitive parts. After washing, it is
possible to check the response time of the
electrodes. A built-in (no return) nozzle in
the spray unit prevents penetration of the
process liquid in the cleaning system. The
standardised dimensions allow mounting in
O-rings
(11x3)
K1500BZ
all flow-, insertion- and immersion fittings of
Yokogawa as well as back-end mounting
on the 4-in-one pH/Redox electrodes, like
FU20.
PROTECTION
HOSE
K1520NC
15
Features
• The EXA analyzer has a built-in timer and
HOLD circuit.
• Built-in (no return) nozzle to prevent
penetration of the process liquid into the
cleaning system.
Figure 4-7 Example HCN4
K1547PP
(spare tubing, EPDM)
IM 12B6K1-01E-E
16
Specifications
Materials
Nozzle : Hastelloy
O-rings : EPDM rubber
Mounting set : Stainless steel
Tubing : 1/4” (OD Ø) Nylon tubing
Process cond. : Max. 1 MPa (10 bar) at 100 ºC
This reference electrode/salt bridge
combination allows pH and ORP(Redox)
measurement with the normal electrodes in
those cases when:
a. excessive contamination of the ceramic
flow diaphragm is expected
- the flow of the reference liquid
through the diaphragm is increased
by pressurising the container.
Consequently, the contamination rate will
decrease.
b. the process to be measured cannot stand
the contamination with KCl
- the salt bridge can be filled with several
electrolytes.
c. measurements have to be performed at
pressures up to 10 bar and temperatures
up to 100 °C
- the container with reference liquid can
be pressurised.
FLOW TUBE (A)
Material : glass
Flow diaphragm : ceramic, PTFE or sleeve
Connector : Ryton R4
Before the electrodes can be serviced, the electrode holder should be physically separated
from the process. The fittings can be changed from the measuring position in the
maintenance position by following the reversed procedures described in §4.
5-2. Cleaning
Cleaning the fitting
When process liquid contains slurry which intends to settle in the fitting regularly, the
electrode holder must be removed for cleaning.
Cleaning the electrodes
When the sensitivity of the electrode has decreased or the response has slowed down,
the electrode should be cleaned. The 3- and 4-hole fittings are provided with a calibration
ring with a set of beakers. The electrode holder complete with electrodes and liquid earth
sensor can be placed on the ring for cleaning.
If cleaning with hot water is not sufficient, more aggressive water based agents should be
used.
- Deposits of limes, hydroxides or carbonates can be removed by immersing the electrode
in a solution containing diluted hydrochloric acid. Afterwards rinse the electrode with
water.
- Deposits of oil and fat can be removed with hot water in conjunction with a detergent.
When the results are unsatisfactory a mild (carbonate based) abrasive can be used.
- Protein (albuminous) deposits should be removed with a protein enzymatic solution. For
instance a solution containing 8.5 ml concentrated hydrochloric acid and 10 gr pepsin in
1 litre water.
IM 12B6K1-01E-E
18
ATTENTION
Avoid using non-polar solvents like tri-chloro ethylene, toluene or hexane. Even cleaning
with ethanol or acetone is not recommended. These solvents will break up the gel-layer on
the glass bulb and afterwards needs to remain soaked in water for at least 12 hours before
functioning normal. The PTFE diaphragm of the combined electrode can sometimes be
regenerated by putting it in hot (60 to 80 °C) 1 molar Potassium chloride (KCl) solution and
letting it cool to room temperature. After cleaning the probe is re-inserted into the process
by following the reverse procedure (see §3-5).
5-3. Calibration
It is recommended to start calibration with clean electrodes. Always calibrate new
electrodes. With the electrodes connected to the transmitter a calibration can take place.
Check the appropriate chapters in the instruction manual of the pH transmitter for details.
General procedure for calibration
To calibrate a pH transmitter two buffer solutions with known pH value are required. It is
recommended that one buffer solution has a value near pH 7.
Depending on the liquid to be measured the second buffer solution should either be in the
acidic or base area. Normally the buffers pH 4.01 or 9.22 are used.
Generally the procedure for calibration is as follows:
- Clean the electrodes
- Rinse the electrodes
- Immerse the electrodes in the first buffer solution (pH 7)
- Adjust the asymmetry setting of the transmitter for reading a known value
- Rinse the electrodes
- Immerse the electrodes in the second buffer (pH 4 or pH 9)
- Adjust the slope setting of the transmitter for reading a known value
- Rinse the electrodes.
During calibration the temperature compensation should be active. It is advised top
calibrate with buffer solutions at a temperature near the process temperature.
5-4. Replacing the O-rings
The exploded view (see §6) shows the position of the sealing O-rings. The O-rings used
in the wetted part are made of silicone rubber, which has superior resistance to corrosion
suitable for use with most process liquids.
Usually no periodical inspection is necessary. To prevent trouble, replace the O-ring seal
periodically, e.g. every year.
IM 12B6K1-01E-E
6. EXPLODED VIEW
12B6K1-31
ASSEMBLED FLOW
FF20-.33
12B6K1-32
ASSEMBLED FLOW
FF20-.43
12B6K1-33_New
ASSEMBLED FLOW
FF20-S33
12B6K1-34
ASSEMBLED FLOW
FF20-S43
19
FF20-.33
FF20-P22
FF20-F22
FF20-S33
FF20-.43
FF20-S22
FF20-S43
IM 12B6K1-01E-E
20
12B6K1-21
1-HOLE MOUNTING
12B6K1-35
SUBASSY FLOW
FS20-.33
12B6K1-36
SUBASSY FLOW
FS20-.43
12B6K1-37
SUBASSY
FS20-S33
12B6K1-38
SUBASSY FLOW
FS20-S43
FS20-.12
FS20-S22
FS20-P22
FS20-F22
FS20-V22
FS20-S32FS20-S43
FS20-.43FS20-.32
IM 12B6K1-01E-E
SPARE PARTS
Accesoires
Part no. Description
K1500FU KCl reservoir PVC for F*20
SB20 Salt bridge
K1500BX Grommet for watertight cable input in PG 16 gland
K1500FV Liquid earth cable (10 m)
K1500DU Liquid earth cable (25 m)
WU20-PC02 COAX-cable (2 m) for single electrode
WU20-PC05 COAX-cable (5.5 m) for single electrode
WU20-PC10 COAX-cable (10 m) for single electrode
WU20-PC15 COAX-cable (15 m) for single electrode
WU20-PC20 COAX-cable (20 m) for single electrode
WU20-PC25 COAX-cable (25 m) for single electrode
WU20-LT02 TRIAX-cable (2 m) for combined electrode
WU20-LT05 TRIAX-cable (5.5 m) for combined electrode
WU20-LT10 TRIAX-cable (10 m) for combined electrode
WU20-LT15 TRIAX-cable (15 m) for combined electrode
WU20-LT20 TRIAX-cable (20 m) for combined electrode
WU20-LT25 TRIAX-cable (25 m) for combined electrode
IM 12B6K1-01E-E
22
Spare parts FF20
Part no. Description
K1500BV O-rings EPDM 11x3 (6 Pcs.)
K1500BW Flow tube for SB20-VC
K1500BY Option /R for F*20.. (82850747)
K1500BZ O-rings Viton 11x3 (6Pcs)51250
K1500DU Liquid earth cable 25m
K1500DW Set of 12 cable nuts for WU20
K1500DX 5 m tubing for SB20
K1500DZ Nut SS, FF/S20-3* + ISC40FF/S
K1500EK O-rings viton 6.07x1.78 (5x2)
K1500EQ O-ring set EPDM FF20-S22
K1500ER O-ring set Viton FF20-S22
K1500ES O-ring set EPDM FF20P&F33
K1500ET O-ring set Viton FF20P&F33
K1500EU O-ring set EPDM FF20-S33
K1500EV O-ring set Viton FF20-S33
K1500EW O-ring set EPDM FF20P&F43
K1500EX O-ring set Viton FF20P&F43
K1500EY O-ring set EPDM FF20-S43
K1500EZ O-ring set Viton FF20-S43
K1500FK O-ring set Viton FF20P&F22
K1500FL O-ring set standard FF20P&F22
K1500FM O-ring set FF20-S22
K1500FU KCl reservoir PVC for F*20
K1500FV Liquid earth cable 10m
K1500GA 5 m tube for KCl reservoir
K1500GN O-ring set FF/FS20 3-hole SS
K1500GP O-ring set FF/FS20 3-hole P/F
K1500GR O-rings silicon 11x3 8pcs
K1500GT O-ring set silicon. FF20-.4.
K1500GZ Earthpin assy for F*20 non-S
K1500HA Calibr. dish for FF20-*33 (50)
K1500HB Calibr. dish for FF20-*43 (50)
K1500HD O-rings silicon 11x3 50pcs
K1511DP O-rings viton 21.9x2.62 (5x2)
K1520CE Spare 3-hole holder PP
K1520CF Spare 3-hole holder PVDF
K1521AD Flange adapter /FS3
K1521AE Flange adapter /FF3
K1521AF Flange adapter /FP3
K1521AG Flange adapter /FS4
K1521AH Flange adapter /FF4
K1521AJ Flange adapter /FP4
K1521AK Flange adapter /FS1
K1521AL Flange adapter /FF1
IM 12B6K1-01E-E
K1521AM Flange adapter /FP1
K1521AN Flange adapter /FS2
K1521AP Flange adapter /FF2
K1521AQ Flange adapter /FP2
K1547PA Hast. cleaning unit HCN2/3
K1547PB Hast. cleaning unit HCN4
K1547PF Nozzle and mounting HCN2/3/F
K1547PG Nozzle and mounting HCN4
K1547PH 10 m Nylon Tube and mounting
K1547PP Spare Part EPDM spraying valves
23
IM 12B6K1-01E-E
24
IM 12B6K1-01E-E
25
IM 12B6K1-01E-E
YOKOGAWA ELECTRIC CORPORATION
World Headquarters
9-32, Nakacho 2-chome, Musashino-shi
Tokyo 180-8750
Japan
www.yokogawa.com
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road
Newnan GA 30265
USA
www.yokogawa.com/us
YOKOGAWA EUROPE BV
Euroweg 2
3825 HD AMERSFOORT
The Netherlands
www.yokogawa.com/eu
YOKOGAWA ELECTRIC ASIA Pte. LTD.
5 Bedok South Road
Singapore 469270
Singapore
www.yokogawa.com/sg
YOKOGAWA CHINA CO. LTD.
3F Tower D Cartelo Crocodile Building
No.568 West Tianshan Road Changing District
Shanghai, China
www.yokogawa.com/cn
Yokogawa has an extensive sales and
distribution network.
Please refer to the European website
(www.yokogawa.com/eu) to contact your
nearest representative.