Model FLXA402
4-Wire Converter
Start-up and Safety Precautions
IM 12A01F01-01EN
This Start-up Manual explains mainly the installation and wiring of the
FLXA402.
For detailed information and other information, the User’s Manual of the
FLXA402 should be referred to.
IM 12A01F01-01EN
4th Edition
Page 2
u Introduction
Thank you for purchasing the FLXATM402 4-Wire Converter.
This Instructor’s Manual contains all essential information for the user to make full use of
FLXA402.
Please read the following respective documents before installing and using the FLXA402.
The related documents are listed as follows.
FLXA402 4-Wire Converter
Start-up and Safety Precautions
FLXA402 4-Wire Converter
Installation and Wiring
FLXA402 4-Wire Converter
Operation of Converter
FLXA402 4-Wire Converter
Operation of pH/ORP
FLXA402 4-Wire Converter
Operation of SC
FLXA402 4-Wire Converter
Operation of ISC
FLXA402 4-Wire Converter
Operation of DO
“EN” in the document number is the language code.
An exclusive User’s Manual might be attached to the products whose sux codes or option
codes contain the code “Z” (made to customers’ specications). Please read it along with this
manual.
Technical Information
ContentsDocument numberNote
FLXA402 4-Wire Converter
HART communication
FLXA402 4-Wire Converter
Modbus communication
“EN” in the document number is the language code.
You can download the latest documents from our website.
http://www.yokogawa.com/an/xa402/download/
2
IM 12A01F01-01ENPrinted manual (This manual)
IM 12A01F01-02ENOnline manual
IM 12A01F01-03ENOnline manual
IM 12A01F02-01ENOnline manual
IM 12A01F03-01ENOnline manual
IM 12A01F04-01ENOnline manual
IM 12A01F05-01ENOnline manual
TI 12A01F01-61ENOnline manual
TI 12A01F01-62ENOnline manual
n Drawing Conventions
Some drawings may be partially emphasized, simplied, or omitted, for the convenience of
description.
Some screen images depicted in the user’s manual may have dierent display positions or
character types (e.g., the upper / lower case). Also note that some of the images contained in this
user’s manual are display examples.
• Please hand over the user’s manuals to your end users so that they can keep the user’s
manuals on hand for convenient reference.
• Please read the information thoroughly before using the product.
• The purpose of these user’s manuals is not to warrant that the product is well suited to any
particular purpose but rather to describe the functional details of the product.
• No part of the user’s manuals may be transferred or reproduced without prior written
consent from YOKOGAWA.
• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at
any time, without notice or obligation.
• If you have any questions, or you nd mistakes or omissions in the user’s manuals, please
contact our sales representative or your local distributor.
n Trademark Notices
FLEXA, FLXA and SENCOM are trademarks or registered trademarks of Yokogawa Electric
Corporation.
All other company and product names mentioned in this user’s manual are trademarks or
registered trademarks of their respective companies.
We do not use TM or ® mark to indicate those trademarks or registered trademarks in this user’s
manual.
n Product Disposal
The instrument should be disposed of in accordance with local and national legislation/
regulations.
3
n Warranty and service
Yokogawa products and parts are guaranteed free from defects in workmanship and material
under normal use and service for a period of (typically) 12 months from the date of shipment from
the manufacturer.
Individual sales organizations can deviate from the typical warranty period, and the conditions of
sale relating to the original purchase order should be consulted. Damage caused by wear and
tear, inadequate maintenance, corrosion, or by the eects of chemical processes are excluded
from this warranty coverage.
In the event of warranty claim, the defective goods should be sent (freight paid) to the service
department of the relevant sales organization for repair or replacement (at Yokogawa discretion).
The following information must be included in the letter accompanying the returned goods:
• Part number, model code and serial number
• Original purchase order and date
• Length of time in service and a description of the process
• Description of the fault, and the circumstances of failure
• Process/environmental conditions that may be related to the failure of the device.
• A statement whether warranty or non-warranty service is requested
• Complete shipping and billing instructions for return of material, plus the name and phone
number of a contact person who can be reached for further information.
Returned goods that have been in contact with process uids must be decontaminated/
disinfected before shipment. Goods should carry a certicate to this eect, for the health and
safety of our employees.
Material safety data sheets should also be included for all components of the processes to which
the equipment has been exposed.
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 4
u CE marking products
n Authorized Representative in EEA
The Authorized Representative for this product in EEA is Yokogawa Europe B.V. (Euroweg 2,
3825 HD Amersfoort, The Netherlands).
n Identication Tag
This manual and the identication tag attached on packing box are essential parts of the product.
Keep them together in a safe place for future reference.
n Users
This product is designed to be used by a person with specialized knowledge.
n How to dispose the batteries:
This is an explanation about the EU Battery Directive. This directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.oce.
Do not dispose them as domestic household waste.
Battery type: Manganese dioxide lithium battery
4
Notice: The symbol (see above) means they shall be sorted out and collected as ordained in the
EU Battery Directive.
n Information of the WEEE Directive
This product is purposely designed to be used in a large scale xed installations only and,
therefore, is out of scope of the WEEE Directive. The WEEE Directive does not apply.
This product should be disposed in accordance with local and national legislation/regulations.
The WEEE Directive is only valid in the EU.
n RE Directive
When FLXA402 contains Bluetooth communication, FLXA402 is built in compliance with
requirements of RE Directive:
We, Yokogawa Electric Corporation hereby declare that this equipment, FLXA402 is in
compliance with the essential requirements and other relevant provisions of Directive 2014/53/
EU. The full text of the EU declaration of conformity is available at the following internet address:
http://www.yokogawa.com/an/xa402/download/
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 5
5
u
Control of Pollution Caused by the Product
This is an explanation for the product based on “Control of Pollution caused by Electronic
Information Products” in the People’s Republic of China.
The FLXA402 should only be used with equipment that meets the relevant IEC, American or
Canadian standards.
Yokogawa accepts no responsibility for the misuse of this unit.
Don’t install “general purpose type” instruments in the hazardous area.
Do not use an abrasive or organic solvent in cleaning the instrument.
Electrostatic discharge
The FLXA402 contains devices that can be damaged by electrostatic discharge. When servicing
this equipment, please observe proper procedures to prevent such damage. Replacement
components should be shipped in conductive packaging. Repair work should be done at
grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic
discharge.
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 6
CAUTION
The Instrument is packed carefully with shock absorbing materials, nevertheless, the instrument
may be damaged or broken if subjected to strong shock, such as if the instrument is dropped.
Handle with care.
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
When you open the front panel, make sure the screws are completely out of the screw holes, and
then open the front panel slowly in order not to damage the threaded parts on the housing. If the
threaded parts are damaged and the screws cannot be tightened, the waterproof performance
will deteriorate.
The HART communication may be inuenced by strong electromagnetic eld.
In this case another trial of the HART communication and/or operation with FLXA402 touch
screen can be carried out.
Be careful to touch the concentrated sulfuric acid.
n Safety, Protection, and Modication of the Product
• In order to protect the system controlled by the product and the product itself and ensure
safe operation, observe the safety precautions described in this user’s manual. We assume
no liability for safety if users fail to observe these instructions when operating the product.
• If this instrument is used in a manner not specied in this user’s manual, the protection
provided by this instrument may be impaired.
• If any protection or safety circuit is required for the system controlled by the product or for
the product itself, prepare it separately.
• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter
simply referred to as YOKOGAWA) when replacing parts or consumables.
• Modication of the product is strictly prohibited.
• The following safety symbols are used on the product as well as in this manual.
6
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual in order
to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes
what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order to
prevent the instrument (hardware) or software from being damaged, or a system failure from
occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Protective Ground Terminal.
n Warning and Disclaimer
The product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor
responsibility to any person or entity with respect to any direct or indirect loss or damage arising
from using the product or any defect of the product that YOKOGAWA can not predict in advance.
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 7
uCompliant Standards
Safety:
CE/Low Voltage Directive (FLXA402 without Bluetooth communication):
EN 61010-1, EN 61010-2-030
UL:
UL 61010-1, UL 61010-2-030
CSA:
CAN/CSA-C22.2 No.61010-1
CAN/CSA-C22.2 No.61010-2-030
GB30439
Russian:
TR CU 004/2011
Installation altitude: 2000 m or less
Category based on IEC 61010: I (DC model)
Category based on IEC 61010: II (AC model) (Note1)
Pollution degree based on IEC 61010: 2 (Note2)
Note1
Installation category, called over-voltage category, species impulse withstand voltage.
Equipment with “Category I” is used for connection to circuit in which measures are taken to limit transient over-voltages to an
appropriately low level.
Category II is energy-consuming equipment to be supplied from the xed installation.
Note2
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
Degree 2 is the normal indoor environment.
EMC:
CE/ EMC Directive (FLXA402 without Bluetooth communication):
EN 61326-1 Class A,Table 2 (For use in industrial locations)
Inuence of immunity environment (Criteria A): Output shift is specied within ± 10 % of F.S.
EN 61326-2-3
EN 61000-3-2
EN 61000-3-3
RCM: EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard
한국 전자파적합성 기준
Russian: TR CU 020/2011
Class A
7
Standards for Bluetooth communication:
Compliant standard: Bluetooth Ver 3.0 Class2
Applicable countries / regions; (regulations)
Japan, EU, USA, Canada, Australia, New Zealand, Singapore
CE/ RE Directive:
EN 61010-1
EN 61010-2-030
EN 62479
EN 301 489-1, EN 301 489-17
EN 61326-1 Class A, Table2
EN 61326-2-3
EN 300 328
RCM:
AS/ NZS 4268, AS/ NZS 2772.2
FCC15C
ICES-003
IMDA TS SRD
Environmental regulation:
RoHS Directive: EN50581
Waste Electrical and Electronic Equipment (WEEE) Directive:
This product is purposely designed to be used in a large scale xed installations only and,
therefore, is out of scope of the WEEE Directive. The WEEE Directive does not apply.
The WEEE Directive is only valid in the EU.
REACH: Regulation(EC) 1907/2006
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 8
FM nonincendive approval (sux code Type : -DD) :
Applicable Standard
FM Class 3600
FM Class 3611
FM Class 3810
ANSI/UL 121201
ANSI/UL 61010-1
ANSI/UL 61010-2-030
ANSI/NEMA 250
Certicate No.
FM18US0281
u Control Drawing
When selecting the nonincendive model (sux code Type : -DD), Please check the condition of
Control Drawing (NFM038-A81).
For details, refer to the User’s Manual IM 12A01F01-02EN.
8
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 9
<Start-up and Safety Precautions>
1. Instrument Check
Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged
during shipment. If damage is found, retain the original packing materials (including the outer
box) and then immediately notify the carrier and the relevant Yokogawa sales oce.
n Checking the model and sux code
Make sure the model and sux code on the nameplate axed to the left side of the housing.
Model and sux code is shown in Table 3.
-A-AB-P1-P1-A2-NR
-N-N-N-NN/U/H6
9
n Mark position of FM nonincendive
When selecting the nonincendive model (sux code Type : -DD), make sure the FM nonincedive
information on the standard nameplate axed to the right side of the housing.
n Checking accessories
The instrument is shipped with standard accessories. Optional accessories are sold separately if
necessary. Make sure the accessories in Table 1 or Table 2 are included.
*1: When FLXA402 has Ethernet communication, one of them is only for Ethernet cable.
*2: The universal mounting kit (/UM) contains the brackets for both /U and /PM options.
*3: /CB6, /CD6 or /CF6 includes adapter only for Ethernet cable.
General purpose for CE, RCM, China standard
General purpose for CSA
General purpose for KC
General purpose
General purpose for EAC with PA
General purpose for EAC
NI for FM
pH/ORP (PH)
Conductivity (SC)
Inductive conductivity (ISC)
Dissolved oxygen (DO)
Digital sensor (DO70G)
SENCOM SA
Without input
pH/ORP (PH)
Conductivity (SC)
Inductive conductivity (ISC)
Dissolved oxygen (DO)
SENCOM SA
2 x 4-20 mA Output + 1 x Contact Input
(mA1 output: with HART)
4 x 4-20 mA Output + 2 x Contact Input +
1 x 4-20 mA Input (mA1 output: with HART)
Contact outputs (Wash and Fail contact outputs)
Without Contact outputs (without Wash and Fail
contact outputs)
Without Bluetooth
Bluetooth
Without Digital communication
Modbus TCP/IP
Modbus RTU (RS-485)
Global except Japan
Japan
Universal mounting kit
Pipe and wall mounting hardware
Panel mounting hardware
Hood, stainless steel
Hood, stainless steel + urethane coating
Stainless steel tag plate
G1/2 x 4 pcs
1/2NPT x 4 pcs
M20 x 1.5 x 4 pcs
G1/2 x 3 pcs + G 1/2 for Ethernet x 1 pcs
1/2NPT x 3 pcs + 1/2 NPT for Ethernet x 1 pcs
M20 x 1.5 x 3 pcs + M20 for Ethernet x 1 pcs
10
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 11
<Start-up and Safety Precautions>
2. Wiring and Installation
Open the front panel and remove the shield cover and then install the cable glands. The shield
cover will be re-installed after the wiring is completed.
For details, refer to the User’s Manual IM 12A01F01-02EN.
2.1 Installation site
The FLXA402 is weatherproof and can be installed both inside and outside. It should, however,
be installed as close as possible to the sensor to avoid long cable runs between the instrument
and sensor.
Select an installation site where the ambient temperature and humidity are within the limits of the
instrument specications as below. If the instrument is installed outside and exposed to direct
sunlight, a sun shade hood should be used.
Ambient Operating Temperature: -20 to +55 ºC
Storage Temperature: -30 to +70 ºC
Humidity: 10 to 90% RH at 40ºC (Non-condensing)
Select an installation site that meets the following conditions.
• Mechanical vibrations and shocks are negligible
• No relay switch and power switch are installed close to the converter
• There is space for cable connection beneath the cable glands
• Not exposed to direct sunlight or severe weather conditions
• Maintenance is possible
• No corrosive atmosphere
The FLXA402 can be mounted on a wall, pipe or panel when the mounting kit is ordered. For
dimensional information please refer to the User’s Manual IM 12A01F01-02EN.
11
CAUTION
When you open the front panel, make sure the screws are completely out of the screw holes, and
then open the front panel slowly in order not to damage the threaded parts on the housing. If the
threaded parts are damaged and the screws cannot be tightened, the waterproof performance
will deteriorate.
NOTE
Be careful not to lose the four front panel screws.
2.2 Wiring
2.2.1 Preparation
Power supply and relay contact (if equipped) should be connected rst. Those terminals are
behind the shield cover. Next, connect the others. For details, refer to the User’s Manual IM
12A01F01-02EN. Read it carefully before wiring.This manual describes how to use the FLXA402
with Yokogawa’s or other companies’ sensors. Please read carefully this manual and the
instruction manuals relevant to those sensors you use before using this instrument.
WARNING
• Cables that withstand temperatures of at least 75 °C should be used for wiring.
• Wiring work should be performed to meet IP66 or higher requirements. Tighten four front
panel screws to the following torque; 1.5 to 1.6 N·m
• Use cables that comply with UL2556VW-1 or equivalent.
• Always place the shield cover over the power supply and contact terminals for safety
reasons and to avoid interference.
IM 12A01F02-01EN 4th Edition : Aug. 07, 2020-00
Page 12
<Start-up and Safety Precautions>
Shield
cover
2.2.2 Cables, Terminals, glands and conduit adapter
The FLXA402 is supplied with terminals suitable for the connection of nished wires in the size
range of 0.13 to 2.5 sq.mm. (26 to 14 AWG). The cable glands supplied will form a tight seal on
cables with an outside diameter of 6 to 12 mm (0.24 to 0.47 inches). Unused cable entry holes
must be sealed with cable glands including the close up plugs supplied.
For details, refer to the User’s Manual IM 12A01F01-02EN.
12
2.2.3 Wiring the power supply
CAUTION
Make sure the power supply is switched o. Power rating must comply with FLXA402
specication. Power voltage must match with the one indicated on the name plate.
WARNING
● You must install external power supply switch or circuit breaker for power supply.
● The external power supply switch or a circuit breaker must comply with a current rating of 5A
or IEC60947-1 or IEC60947-3
● Yokogawa recommend installing the external power supply switch, circuit breaker and
FLXA402 converter all in the same location.
● Install the external power supply switch or circuit breaker to the place where operators
access easily. To alert users, put a label on the external power switch.
● Wire cables of power supply, contact output securely with cable rack, conduit and vinyl
band. Unplugged cables are dangerous and may cause an electric shock.
You should check local safety regulation to see if you shall install external circuit breaker. Follow
the local regulation and install if necessary.
FLXA402 is protected by fuse which provides overcurrent protection of inner circuit.
Specic current and voltage ratings of fuse vary depending on power system. Use always a time-
delay fuse for 250 VAC fuse in accordance with IEC60127.
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 13
<Start-up and Safety Precautions>
WARNING
Fuse replacement should be performed only by a qualied service personnel. See Chapter 4.
MAINTENANCE, Fuse.
l Access to terminal and cable entry
The power supply terminals are behind the shield cover.
Guide the power cables through the gland. The terminals will accept wires of 2.5 mm2 (14 AWG).
Always use cable nishings. For details, refer to the User’s Manual IM 12A01F01-02EN.
l Power supply terminals
Connect power supply to the power supply terminals.
For details, refer to the User’s Manual IM 12A01F01-02EN.
AC power terminals
(AC Version)
13
DC power terminals
(DC Version)
l Grounding the converter
For the safety of the user and to protect the instrument against interference, the housing must
always be connected to ground. This has to be done by a cable with large cross-sectional area.
This cable can be xed to the rear of the housing or the internal ground connections. For details,
refer to the User’s Manual IM 12A01F01-02EN.
(M4 screw)
(M4 screw)
WARNING
The minimum cross sectional area of the protective grounding wire should be 0.75 mm2. For
CSA safety standard (Type: - AB), use cables with a cross section of 0.75 - 2.1 mm2.
IM 12A01F02-01EN 4th Edition : Aug. 07, 2020-00
Page 14
<Start-up and Safety Precautions>
2.2.4 Wiring the contact signals
l Contact outputs
The FLXA402 unit’s four contacts (switches) that can be wired and congured to suit user
requirements. Contact S4 is programmed as a fail-safe contact. Refer to the User’s Manual IM
12A01F01-02EN for contact outputs setup.
Use a cable with the rated voltage of 300 V AC or higher and with the electric current capacity
according to the load connected to contact output.
l Contact inputs
It is necessary to use screening/shielding on the input signal cables. For details, refer to the
User’s Manual IM 12A01F01-02EN.
2.2.5 Wiring the mA-input/ output signals
l mA-input signals
The mA-inputs receive current signals of 4-20 mA. Should be used the cable beyond 30 V AC.
It is necessary to use screening/shielding on the input signal cables.
For details, refer to the User’s Manual IM 12A01F01-02EN.
l mA-output signals
The output signals consist of current signals of 4-20 mA. The maximum load can be 600 ohms on
each. Should be used the cable beyond 30 V AC. It is necessary to use screening/shielding on
the output signal cables. For details, refer to the User’s Manual IM 12A01F01-02EN.
14
2.2.6 Wiring the sensor
The FLXA402 can be used with a wide range of commercially available sensor types, both from
Yokogawa and other manufacturers.
Terminal screw size is M3, and torque of screw up is 0.6 N•m.
Pin terminal, ring terminal and spade terminal can be used.
For details, refer to the User’s Manual IM 12A01F01-02EN.
2.2.7 Wiring the communication
The FLXA402 can equip Ethernet (Modbus TCP/IP) communication, RS-485 (Modbus RTU)
communication.
It is necessary to use screening/shielding on the communication cable.
For details, refer to the User’s Manual IM 12A01F01-02EN
CAUTION
Do not tighten up four front panel screws one by one.
Each front panel screw should be tightened up in two times of screwing. And, rstly the screw
at the upper left should be screwed a bit, the next is at the lower right, third is at the upper right,
and fourth is at the lower left. The second round is the same sequence again to tighten up four
screws.
Do not use an electric screwdriver with high revolutions. If an electric screwdriver is used for
these front panel screws, the revolutions of the electric screwdriver should be less than 400 rpm.
Four screws should be tightened to the following torque; 1.5 to 1.6 N•m
IM 12A01F01-01EN 4th Edition : Aug. 07, 2020-00
Page 15
<Start-up and Safety Precautions>
3. Operation
When all wiring is completed, turn on the power to the instrument. Make sure that the LCD
screen turns on, and then wait for the Home display appears. If congurations are not proper, an
error indicator may be displayed, or the measurement values displayed may be incorrect. Check
the initial settings and change them to suit your purpose.
For details, refer to the following User’s Manual.
ContentsDocument number
FLXA402 4-Wire Converter
Operation of Converter
FLXA402 4-Wire Converter
Operation of pH/ORP
FLXA402 4-Wire Converter
Operation of SC
FLXA402 4-Wire Converter
Operation of ISC
FLXA402 4-Wire Converter
Operation of DO
FLXA402 4-Wire Converter
HART communication
FLXA402 4-Wire Converter
Modbus communication
l Change language
The default language setting for the FLXA402 is English. To select a dierent language other
than English, refer to the User’s Manual IM 12A01F01-03EN.
IM 12A01F01-03EN
IM 12A01F02-01EN
IM 12A01F03-01EN
IM 12A01F04-01EN
IM 12A01F05-01EN
TI 12A01F01-61EN
TI 12A01F01-62EN
15
l Operation
According to the sensor to be used, refer to appropriate user’s manual about operation.
4. Maintenance
n Periodic maintenance
The FLXA402 requires very little periodic maintenance, except to make sure the front window
is kept clean in order to permit a clear view of the display and allow proper operation of the
touchscreen. If the window becomes soiled, clean it using a soft damp cloth or soft tissue.
To deal with more stubborn stains, a neutral detergent may be used.
When you must open the front cover and/or glands, make sure that the seals are clean and
correctly tted when the unit is re-assembled in order to maintain the housing’s weatherproof
integrity against water and water vapor.
The pH measurement uses high impedance sensors and may otherwise be prone to problems
caused by exposure of the circuitry to condensation
CAUTION
Never use harsh chemicals or solvents. In the event that the window does become heavily
stained or scratched, refer to the parts list for replacement part numbers.
l Battery
The FLXA402 contains a lithium cell (battery) to support the clock function when the power
is switched o. The cell has an expected working life of 10 years. Should this cell need to be
replaced, contact your nearest Yokogawa service center.
l Fuse
There is a circuit board mounted fuse protecting the instrument. If a fuse blows soon after the
change, contact your nearest Yokogawa service center.
IM 12A01F02-01EN 4th Edition : Aug. 07, 2020-00
Page 16
<Start-up and Safety Precautions>
Revision record
Aug. 2020/ 4th Edition
Addition of regulation in Compliant Standard.
July 2020/ 3rd Edition
Addition of “Type: -AQ and -AR”.
Apr. 2019/ 2nd Edition
Addition of “Type: -DD”, etc.
Nov. 2018/ 1st Edition
Newly released.