Yaskawa iQpump User Manual

YASKAWA iQpump
Intelligent Pump Controller
Manual Supplement
Use this supplement with User Manual TM.iQp.06 and Programming Manual TM.iQp.07. To properly use the product, read this manual thoroughly and retain for easy reference, inspection, and maintenance. Ensure the end user receives this manual.
MANUAL NO. TM.iQp.10
2 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
Table of Contents
1 PREFACE AND SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 PRODUCT OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 DRIVE CHANGE SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5 FEATURE DETAILS - REMOTE DRIVE DISABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
6 FEATURE DETAILS - LOW CITY PRESSURE ALARM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7 SUCTION CONTROL VIA CONSTANT PRESSURE W/WELL DRAW DOWN. . . . . . . . . . . . . . . . . . . . . . . 14
8 GEOTHERMAL MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
9 WATER LEVEL/SUCTION PRESSURE CONTROL IN MEMOBUS MULTIPLEX . . . . . . . . . . . . . . . . . . . . 26
10 IQPUMP MEMOBUS/NETWORK OPERATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
11 IMPROVED AND MISCELLANEOUS FUNCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
12 SPEED FOLLOWER DECELERATION TIME SWITCHOVER <0036>. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
13 TRANSDUCER WIRING USING AN EXTERNAL POWER SUPPLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
PARAMETER LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Use this addendum with User Manual TM.iQp.06 and Programming TM.iQp.07.
To properly use the product, read this manual thoroughly and retain for easy reference, inspection, and maintenance. Ensure the end user receives this manual.
.
YASK AWA TM.iQp.10 iQpump Technical Manual Supplement 3
This Page Intentionally Blank
4 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement

1 Preface and Safety

SUPPLEMENT
1 Preface and Safety
Yaskawa manufactures products used as components in a wide variety of industrial systems and equipment. The selection and application of Yaskawa products remain the responsibility of the equipment manufacturer or end user. Yaskawa accepts no responsibility for the way its products are incorporated into the final system design. Under no circumstances should any Yaskawa product be incorporated into any product or design as the exclusive or sole safety control. Without exception, all controls should be designed to detect faults dynamically and fail safely under all circumstances. All systems or equipment designed to incorporate a product manufactured by Yaskawa must be supplied to the end user with appropriate warnings and instructions as to the safe use and operation of that part. Any warnings provided by Yaskawa must be promptly provided to the end user. Yaskawa offers an express warranty only as to the quality of its products in conforming to standards and specifications published in the Yaskawa manual. NO OTHER WARRANTY, EXPRESSED OR IMPLIED, IS OFFERED. Yaskawa assumes no liability for any personal injury, property damage, losses, or claims arising from misapplication of its products.

Applicable Documentation

The following manuals are available for the iQpump Controller:
Option Supplement
Yaskawa AC Drive - iQpump Controller Manual Supplement Manual No: TM.iQp.10
Read this manual first.This document is a supplement to the iQpump Controller User Manual and Technical Manual and Simplex Cheat Sheet. This supplement lists the effect of this custom software on the parameters in the drive and function descriptions in the manual. Access this web site to obtain a PDF or additional printed copies of this supplement. U.S.: http://www.yaskawa.com
Yaskawa Drive
Yaskawa AC Drive-iQpump Controller User Manual TM.iQp.06
Yaskawa AC Drive-iQpump Controller Programming Manual TM.iQp.07
iQpump Controller Simplex Cheat Sheet TM.iQp.04
To obtain instruction manuals for Yaskawa products access these sites: U.S.: http://www.yaskawa.com Other areas: contact a Yaskawa representative.
For questions, contact the local Yaskawa sales office or the nearest Yaskawa representative.

Terminology Used in this Manual

These terms are used in this manual.
iQpump Controller: The iQpump Controller may also be referred to as “drive” in this manual.

Obtaining Support

When seeking support for a drive with custom software, it is imperative to provide the unique part number shown on the drive nameplate. The software is flashed to the control board memory and the operation of parameters, functions, and monitors are different than the standard drive software, as described herein.
Refer to the Yaskawa office location listed on the back cover of this supplement for support inquiries.
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 5

2 General Safety

NOTICE
2 General Safety
Abnormal Pump Operation Hazard
Important note when replacing an iQpump Controller.
Verify software ID drive monitor U1-14 when replacing an iQpump Controller in a multiplex network configuration. Set parameter P9-99 to a value of 0 : "A-version 30034" if (ONE or MORE) iQpump Controllers on the network have software version monitor U1-14=30034 and nameplate PRG: 0034. Otherwise do not change P9-99 setting from factory default. Failure to comply will result abnormal drive operation.

Supplemental Safety Information

Read and understand this manual and the iQpump Controller User Manual before installing, operating, or servicing this option unit. The drive must be installed according to the iQpump Controller User Manual and local codes. Observe all cautions and warnings in this document and the standard drive manuals.
Refer to the iQpump Controller Simplex Cheat Sheet TM.iQp.04 and the iQpump Controller User Manual TM.iQp.06 for safety information and installation and start-up instructions.
This document is a supplement to the standard iQpump Controller Manuals. It describes the enhanced functionality of drive software version U1-14=30035 and nameplate PRG: 0035 or greater.
• Custom software is provided to add functionality to a standard AC drive to enhance or enable use in a specific application.
6 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement

3 Product Overview

3 Product Overview

About This Product

Yaskawa has simplified the application of variable frequency drives to pumping applications. Pumps are typically used whenever water must be pumped from deep water wells or open bodies of water such as rivers, lakes, irrigation canals, and water storage facilities. The integrated pump specific software and quick setup parameters, allow the operator easy access to setup control values for a wide range of applications.
Feature Enhancement Summary
The following is a brief summary of features for drive software PRG:0035 that are added or enhanced over previous iQpump Controllers:
Initialization Function: A special initialization function is created to quickly and conveniently configure the drive as a generic speed controller. This initialization re-configures the Pump Quick Setup (Quick Start) menu to show parameters likely to require adjustment in generic applications.
b1-04 (Reverse Operation Selection): Added new parameter B1-04 (Reverse Operation Selection) which allows the motor phases to be exchanged.
FCarrier Frequency Selection (C6-02): Default has been changed to 1 (2.0 kHz).
Remote Drive Disable: This feature prohibits the drive from running when a new multifunction digital input is closed for
a set amount of time.
Low City Pressure Alarm Selection: Added a parameter that selects the Low City Pressure alarm text that will show when in a Low City Pressure condition.
Suction Control Support via the Constant Pressure with Well Draw Down Control: Suction control option has been integrated into the Water Level control.
Modified Terminal A1 Loss Detection: Different alarm/fault texts will occur when feedback loss is detected on Terminal A1 to differentiate from PI feedback loss on Terminal A2.
iQpump Memobus Multiplex: Lag Drive Speed Follower: Lag drives follow the speed of the lead drive.
iQpump Memobus Multiplex: Lag Turn Off: Lag drives can run at a fixed speed for a set time before turning off.
iQpump Memobus Multiplex: Flow Meter: New option to stage or de-stage based on Flow Rate.
iQpump Memobus Multiplex: Stop History: A new lead drive selection (P1-01) option has been added.
Geothermal Mode: The speed of the drive is determined by an analog temperature input and a temperature-speed curve.
iQpump Multiplex: Water Level/Suction Pressure Control
• Ability to run the iQpump Controller Memobus network with just 1 drive connected to a Water Level or Suction Pressure device
• Ability to automatically switch from the Analog Water Level/Suction Pressure Source to the Network Source in case of transducer loss or wire breaks.
• Pump De-stage due to the Water Level being below the Minimum Water Level.
• Pump De-stage due to the Suction Pressure being below the Minimum Suction Pressure.
• Disallow Pump Staging when PI is being influenced by the Water Level/Suction Pressure control
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 7
3 Product Overview

Applicable Models

The iQpump Controller is available in these models in Table 1.
Table 1 Applicable Models
Drive Software Version <1>
(All P7 drive models) CIMR-P7U-107
<1> See “PRG” on the drive nameplate or drive monitor U1-14 for the software version number.
Listed on the drive nameplate as PRG:0035 or PRG:0036
VSP1300035 or VSP1300036.
8 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement

4 Drive Change Summary

4 Drive Change Summary

Overview

The following is an overview of changes made the drive with software PRG: 0035 and PRG: 0036.
Drive Software Changes
Table 2 Parameter Enhancements for PRG: 0035
Parameter Description Page
A1-01 Language Selection Selection 6:Portuguese is removed. 40
A1-03 Initialization Mode Added selection 7770:General Purpose. 38
b1-01 Frequency Reference Selection
b1-03 Stop Method Default changed from: Ramp to Stop (0) to Coast to Stop (1). 41
C6-02 Carrier Frequency Default value changed to setting 1 (2.0 kHz) for all drive sizes. 41
P8-01 Water Lev./Suc. Pres. Selection
P8-09 Low Level Behavior Added selection 3:Auto-Restart (time set by P8-12). 17
P9-01 Lead Drive Selection
P9-05 Lag Drive Mode Added selection 3:Follow Lead Speed. 29
P9-08 Add Pump Mode Added selection 3:Flow Meter. 32
P9-12 Remove Pump Mode
P9-23 Max Pumps Running
P9-25 Highest Node Address
Added selection 5:Geothermal Mode. 20
Selection 1: Changed from Enabled to Water Level Control. Added selection 2:Suction Pressure Control.
Added selection 2:Stop History. 34
Added selection 3:Flow Meter. 33
Range changed from 1 - 16 to 1 - 8. Default changed from 16 to 8.
Range changed from 2 – 16 to 2 - 8. 42
16
Table 3 Digital Input H1- Enhancements for PRG: 0035
Function Description Page
72 Remote Drive Disable
Function Description Page
4A Transducer Loss <1>
57 Low Water Level <1>
58 Low Suction Level
<1> Modified from previous version.
Function Description Page
20 Geothermal Mode Temperature input for geothermal mode operation. 21
Digital Input function (normally open or normally closed) that can be used to prevent the drive from running with a programmable delay time.
Table 4 Digital Output H2- Enhancements for PRG: 0035
Closed: During a “Feedback Loss” alarm. Closed: During a “FBL – Feedback Loss” fault”. Closed: During an “A1-LOST Terminal A1 Lost” alarm. <1>New Closed: During an “A1-LOST Terminal A1 Lost” fault. <1>New
This will energize if P8-01 = 1 and the level in the well drops below the Low Level Detection Level (P8-07) for more than the Low Level Detection Delay Time (P8-08), or if there is a LOWWL – Low Water Level Fault.
This will energize if P8-01 = 2 and the suction pressure drops below the Low Suction Pressure Detection Level (P8-07) for more than the Low Suction Pressure Delay Time (P8-08), or if there is a LOSUC – Low Suction Pressure Fault.
Table 5 Added H3-09 Analog Input Function for PRG: 0035
11
36
18
18
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 9
4 Drive Change Summary
Table 6 Additional or Modified Fault Codes for PRG: 0035
Fault Code Description Page
OPE13
Terminal A1
OPE17 Run/Stp-CoastTmr
OPE18
Net Incompatible
OPE19
Geothermal Set
Tem p Lo st Geothermal Input
LOSUC Low Suction
Alarm Description Page
Net FlowMeter
Lost, Chk Source
LOSUC Low Suction
Low Water
in Tank
Temp Lo st Geothermal Input
R-DNE-S<1> Remote Drv Dis
AnalogA1 Lost
Switched to Net
Net Wtr/Suctn
Lost, Chk Source
<1> The character is a place holder for the actual digital input designation, such as S1, S2, or S3.
Displayed when terminal A1 is assigned to more than one function. OPE13 is modified from the previous software version PRG:0034.
Displayed when Run/Stop Control and Coast To Stop w/Timer are both enabled. 42
Displayed when using features specific to iQpump software version 30035+ and nameplate PRG: 0035 or greater with parameter P9-99 set to (0: A:Version 30034).
Displayed when a parameter selection is not compatible with the Geothermal Mode (b1-01 = 5) setting.
Displayed when geothermal input is not present. 23
Suction pressure is below programmed level. 18
Table 7 Additional Alarms or Messages for PRG:0035
Displayed when there is no drive on the Memobus Network with an analog Flow Meter. 28
Displayed when Low City Pressure is active and alarm selection P4-27 = 1. 13
Displayed when Low City Pressure is active and alarm selection P4-27 = 2. 13
Displayed when a geothermal input is not present. 23
Remote drive disable is active. 12
Displayed when there is a defective or broken analog input source. Analog Terminal A1 has not been detected, and the Network Water Level or Suction Pressure signal is now used.
Displayed when a valid analog source for Water Level or Suction Control Pressure can not be found on the network.
28
43
42
23
23
Table 8 Additional Alarms or Messages for PRG:0036
Alarm Description Page
OPE20
Net WL/SP Mode
Low WL/SP Drive Disabled
10 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
Incompatible Network Water Level / Suction Pressure Mode. -
Drive is unable to run because the Water Level or Suction Pressure is below the P8-05 setting. -

5 Feature Details - Remote Drive Disable

iQpump Drive StoppingiQpump Drive Enabled
Time
Remote
Drive Disable
Active
P4-25 On-Delay Time
Inactive
5 Feature Details - Remote Drive Disable

Overview

This function allows users to prohibit the iQpump Controller from running by using a multi-function digital input set to "Remote Drv Disbl" (72) in combination with a programmable on time (P4-25) and off time delay (P4-26).
The "Remote Drive Disable" input logic can be selected with parameter P4-24.
Function Operation
When the Remote Drive Disable (H1-0 = 72) input is active for the amount of time specified in P4-25 the drive will be forced into a sleep-like state. Stopping method used is defined by b1-03 (Stopping Method Selection) setting.
If the iQpump Controller is running as part of a Memobus Network (multiplex system), it will be taken out and considered unavailable to run.
Alarm "R-DNE-S" will be displayed, where S is the Terminal programmed for Remote Drive Disable (H1-0 = 72).
The iQpump Controller will restart and perform a pre-charge if needed when the Remote Drive Disable ( H1-0 = 72) is inactive for the time specified in P4-26.
Note: When the drive run is disabled using this feature, the pre-charge function is reset and will run if necessary.
P4-24 Remote Drive Disable Selection
Selects the type of pressure switch connected to the “Remote Drv Disbl” digital input (H1-0 = 72).
Figure 1
Setting Description
0
1
Normally Open (factory default) Closed indicates a "Remote Disable" condition - active
Normally Closed Open indicates a "Remote Disable" condition - active
Figure 1 Remote Disable: Inactive to Active Operation
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 11
5 Feature Details - Remote Drive Disable
iQpump Drive EnablediQpump Drive Stopped
Time
Remote
Drive Disable
Active
P4-26 Off-Delay Time
Inactive
Figure 2
Figure 2 Remote Disable: Active to Inactive Operation
P4-25 Remote Drive Disable On-Delay Time
Range Description Default
1 - 1000 sec
Sets the amount of time a “Remote Drive Disable” condition needs to be present before the drive will stop.
0 sec
P4-26 Remote Drive Disable Off-Delay Time
Range Description Default
1 - 1000 sec
Multi-function Input Setting: Remote Drive Disable Function
Setting Description
72 Remote Drv Disbl
Alarm: Remote Drive Disable
Alarm Display Description
R-DNE-Sx
Remote Drv Dis
Note: S or Sx denotes the digital input terminal programmed for this function.
Sets the amount of time a "Remote Drive Disable" condition must be absent before the drive will be allowed to run.
Remote Drive Disable: Prevents the iQpump Controller from running when active for the time set in P4-25. Must be inactive for the time set in P4-26 to allow the drive to run again. Note: Parameter P4-24 determines if this input is Normally Open or Normally Closed.
Multi-function Input Terminal S (H1-0 = 72) has been closed for the time set in P4-25 when P4-24 = 0. or Multi-function Input Terminal S (H1-0 = 72) has been open for the time set in P4-25 when P4-24 = 1.
0 sec
12 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement

6 Feature Details - Low City Pressure Alarm

6 Feature Details - Low City Pressure Alarm

Overview

The alarm function is used for stopping the iQpump Controller or preventing the iQpump Controller from running based on an external contact closure.
This function has been enhanced by parameter P4-27 which allows users to select the alarm message that will be displayed when a Low City Pressure condition is detected.
Function Operation
P4-27 Low City Pressure Alarm Text
Selects the alarm message that will be displayed when a Low City Pressure condition is detected.
Setting Description
0 Low City Pressure (factory default)
1 Low Suction Pressure
2Low Water in Tank
Alarm: Low Pressure/Low Water in Tank
Alarm Display Description
Shown when P4-27 = 1 and when the digital input has been active (closed for P4-21 = 0, or open for
Low Suction Pressure
Low Water In Tank
P4-21 = 1) for the time set in P4-22. The drive, if running, coasts-to-stop and does not run until the digital input has been inactive for the time set in P4-22.
Shown when P4-27 = 2 and when the digital input has been active (closed for P4-21 = 0, or open for P4-21 = 1) for the time set in P4-22. The drive, if running, coasts-to-stop and does not run until the digital input has been inactive for the time set in P4-22.
Related Parameters
P4-21 Low City Pressure Input Select
Setting Description
0 Normally Open (closed for the "Low City Pressure" condition)
1 Normally Closed (open for the "Low City Pressure" condition) (factory default)
P4-22 Low City Pressure Input Delay Time
Range Description Default
Sets the amount of time a "Low City Pressure" condition must be present before
1 - 1000 sec
Multi-function Input Setting H1-: Low City Pressure
Setting Description
73 Low City Pressure
the drives will stop. Also sets the amount of time that the pressure must be adequate before the drive system will restart.
Indicates that sufficient/insufficient pressure is present on the inlet to the pump. This setting is used mainly for pressure booster stations.
10 sec
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 13

7 Suction Control via Constant Pressure w/Well Draw Down

7 Suction Control via Constant Pressure w/Well Draw Down

Overview

This function enables the iQpump Controller to monitor suction pressure at the inlet of the pumps.
Note: This function is active when parameter P8-01 is set to 2 for Suction Control.
Packaged Booster Systems have a desired discharge pressure and a given suction pressure from the city water system or from a suction tank. Such systems are often designed to handle a specific pressure and flow rate.
The suction pressure may have a wide range between high and low pressures. The suction pressure typically decreases with increased fluid flow rate. In some instances, pump cavitation may occur if suction pressure falls below a certain pressure level.
In addition low or negative suction pressure can lead to damage such as pipe collapse due to external forces acting on the pipe.
The iQpump Controller will accept an analog suction transducer (Terminal A1) and can be programmed to trigger an alarm or shutdown the system when the suction pressure falls outside of a normal operating range.The iQpump Controller will automatically restart and return to normal operation once the suction pressure returns to a normal level.
When operating in multiplex mode, additional control can be programmed to de-stage any active pumps as a method to try and reduce pump loading to prevent suction pressure from continuing to drop below the shut down level.
Function Description
The iQpump Controller will regulate outlet pressure of the pump system using the standard iQpump Controller features when there is adequate suction pressure available at the inlet of the pumps and offers two options to respond to a drop in suction pressure.
1. Regulate Outlet Pressure and Suction Pressure: (P8-03 > or = P8-04) To regulate suction pressure set the suction pressure setpoint (P8-03) to a value greater or equal than the minimum suction pressure (P8-04). In this mode the iQpump Controller will try to regulate the suction pressure based on the programmed suction pressure setpoint (P8-03) level.
As the suction pressure decreases and approaches the suction pressure setpoint level (P8-03), the iQpump Controller will slow down causing the outlet pressure and flow to decrease. When the suction pressure rises above the P8-05 level for more than the P1-05 time, normal operation (outlet pressure regulation) will resume.
When the suction pressure drops below Minimum Suction Pressure (P8-04), for more than the Sleep Delay Time (P2-03), the iQpump Controller will go to sleep.
Note: The Suction Control Minimum Speed parameter (P8-06) should be set to a high enough value that will ensure flow.
2. Regulate Outlet Pressure Only: (P8-03 < P8-04) Set the suction pressure setpoint (P8-03) to a value smaller than the minimum suction pressure (P8-04) to regulate outlet pressure.
This mode allows the iQpump Controller to maintain the outlet pressure setpoint using the standard iQpump Controller features and go to sleep immediately when the suction pressure drops below the Minimum Suction Pressure (P8-04), for more than the Sleep Delay Time (P2-03).
Normal operation (outlet pressure regulation) will resume when the suction pressure returns above the P8-05 level for more than the P1-05 time.
14 YAS KAWA TM.iQp.10 iQpump Technical Manual Supplement
7 Suction Control via Constant Pressure w/Well Draw Down
Duplex System Example
A pump system consisting of a duplex domestic pressure booster system using a pressured city supply averaging 40
Example:
Pump System Settings
• All pumps have the same flow rate and will alternate
• Lag Pump will track speed of lead pump
• All drives have individual discharge transducers rated 150 psi. Upon failure, system will look to network for feedback information.
• Suction Transducer is rated 75 psi
• Below 40 psi city pressure the pump system will start to reduce speed until reaching 30 psi
• System Auto setpoint 85 psi, with a 5 psi start level.
Duplex System Example (Related Parameters)
- 60 psi. If city pressure starts to fall below 40 psi the pump system will start to slow down to reduce the chances of pump cavitation. When suction pressure (city pressure) drops to 30 psi, the pump system will shut down (sleep) and wait until city pressure returns, after which the system will automatically start and operate under normal condition.
Parameter Description Booster Pump 1 Booster Pump 2
H3-02 Terminal A1 Gain 231.3% 231.3%
H3-03 Terminal A1 Bias -25.0% -25.0%
H3-08 Terminal A2 Signal 2 2
H5-01 Drive Node Network Address 1 2
P1-01 Pump Mode 3 3
P1-03 FD Device Scaling 150 psi 150 psi
P1-04 Start Level -5.0 psi -5.0 psi
P8-01 WtrLvl/SuctionPres Selection 2 2
P8-02 Suction Transducer Scaling 75 psi 75 psi
P8-03 Suction Pressure Setpoint 40.0 psi 40.0 psi
P8-04 Minimum Suction Pressure 30.0 psi 30.0 psi
P8-05 Wakeup Suction Pressure 40.0 psi 40.0 psi
P9-02 Feedback Source 2 2
P9-05 Lag Drive Mode 2 2
P9-25 Highest Node Address 3 3
P9-50 Suction Pressure Source 2 2
U1-01 Auto Setpoint 85 psi 85 psi
Note: In multiplex mode the network signal for suction pressure can be used. Refer to Transducer Wiring using an External Power
Supply on page 48.
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 15
7 Suction Control via Constant Pressure w/Well Draw Down
Enabling Suction Pressure Control
The suction pressure control selection is added to parameter P8-01 Water Level/Suction Pressure Selection. Follow these steps for basic suction pressure control setup:
1. Enable suction control by setting parameter P8-01 to 2 (Suction Control).
2. Set suction transducer scaling (P8-03) in psi, scaling for analog input A1 (20 mA scale).
3. Set suction pressure setpoint (P8-03) in psi.
4. Select iQpump operation when a drop in suction pressure occurs:
a. Regulate outlet pressure only and go to sleep immediately when suction pressure falls below minimum suction
pressure level (P8-04). Set suction pressure setpoint (P8-03) below minimum suction pressure level (P8-04).
b. Regulate suction pressure when pressure is below suction pressure setpoint (P8-03) but above the minimum suction pressure level (P8-04). Set suction pressure setpoint (P8-03) to above the minimum suction pressure level (P8-04).
5. Set minimum suction pressure (P8-04) – Sleep Level for Suction Control.
6. Set suction pressure wake-up level (P8-05) – Wake-up level.
7. Set suction control minimum speed (P8-06) – Minimum Flow Speed.
P8-01 Water Level/Suction Pressure Selection
Sets the mode of operation for the Water Level / Suction Control function.
Setting Description
0 Disabled (factory default)
1
2 Suction Pressure Control (New)
Note: The iQpump Controller will only perform the outlet pressure and suction pressure control when "normal" PI mode is enabled
(b5-01 > 0, not jogging, not disabled via MFDI, etc.).
P8-02 Suction Transducer Scaling
Water Level Control (This function is defined as "Enabled" for iQpump software PRG: 0034)
Terminal A1 is used for the suction pressure analog input. The analog pressure sensor, mounted on the inlet side of the pump(s) will provide the required signal. Set this parameter to the full-scale pressure of the transducer. The suction pressure will then be displayed on the U1-98 monitor.
Range Description Default
5 - 500 psi
P8-03 Suction Pressure Setpoint
Range Description Default
0.0 - 1200.0 psi
P8-04 Minimum Suction Pressure
Sets the full scale (20 mA) output of the pressure transducer connected to Terminal A1.
Sets the amount of suction pressure the iQpump Controller will attempt to regulate.
100 psi
20.0 psi
Range Description Default
0.0 - 1200.0 psi
16 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
When the suction pressure drops to below this level for more than the P2-03 time, the drive will go to sleep and turn off all lag pumps.
10.0 psi
7 Suction Control via Constant Pressure w/Well Draw Down
P8-05 Wake-Up Suction Pressure
Range Description Default
If the drive is forced to sleep based upon the minimum suction pressure
0.0 - 1200.0 psi
P8-06 Suction Control Minimum Speed
Range Description Default
0.00 - 120.00 Hz
P8-07 Low Suction Pressure Detection Level <1>
Range Description Default
0.0 - 1200.0 psi
(P8-04), the suction pressure must go above this level for more than the P8-13 time in order to wake up.
This parameter sets the minimum speed the drive will run when the drive is controlling suction pressure. When the drive is controlling outlet pressure or this parameter is set less than P1-06 and P4-05, P1-06 and P4-05 will be used as the minimum speed.
When the amount of suction pressure drops below this level for more than the P8-08 time, the drive will respond depending on the P8-09 setting. A setting of
0.0 disables this detection.
30.0 psi
0.00 Hz
0.0 psi
P8-08 Low Suction Pressure Detection Time
<1>
Range Description Default
0.0- 300.0 min
Sets the amount of time delay that the suction pressure must drop below the P8-07 level before the drive will react. *Time units are defined by P8-14.
0.1 min
P8-09 Water Level/Suction Pressure Selection
Sets how the drive will respond when the water level in the well drops below the P8-07 level for more than the P8-08 time.
Setting Description
0 No Display (Digital Output Only)
1 Alarm Only (factory default)
2Fault
3 Auto-Restart (time set by P8-12) (New)
P8-10 Suction Control Proportional Gain
Range Description Default
0.00 - 25.00 Sets the proportional gain for the suction pressure control. 2.00
P8-11 Suction Control Integral Time
Range Description Default
0.0 - 360.0 sec
Sets the integral time for the suction pressure control. A setting of zero disables the suction pressure control integrator.
5.0 min
P8-12 Suction Control Auto-Restart Time
Range Description Default
0.1 - 6000.0 min
<1>
Low Level/Low Suction Detection (P8-07/P8-08): This feature is disabled when the drive is in Run-Stop control (b1-02 = 5, Timed Run).
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 17
Sets the amount of time the drive will wait before attempting an auto-restart of the "Low Suction" fault. Effective only when P8-09 = 3 and L5-01 > 0.
5.0 min
7 Suction Control via Constant Pressure w/Well Draw Down
P8-13 Suction Pressure Sleep Wake-Up Time
Range Description Default
If the iQpump Controller has been forced to sleep based upon the minimum
0 - 3600 sec
P8-14 Low Suction Pressure Detection Time Unit
suction pressure (P8-04), the pressure must go above the P8-05 level for more than this time in order to wake up.
1 sec
Defines the time unit for P8-08.
Setting Description
0 Minutes (factory default)
1 Seconds
Suction Pressure Transducer Wire-break Detection
Wire-break detection is active on Terminal A1 when PI Feedback Loss Detection is enabled (b5-12 = 1 or 2) , Suction control is enabled (P8-01 = 2) and PI control is NOT disabled via Multi-Function Digital Input.
The iQpump Controller detects a wire-break condition when the suction transducer signal on Terminal A1 falls below -
6.25 % or rises above 106.25 % for more than 1 second and will react according to parameter b5-12 (Feedback Loss Detection).
Note: Wire-break detection on Terminal A1 detection is checked after the gain/bias parameters (H3-02 and H3-03) are applied.
Fault: Low Suction
Fault Display Description
LOSUC
Low Suction
Low Suction: Suction pressure is below the P8-07 level for more than the P8-08 time.
Parameters P8-12 to P8-14 are also available for Water Level control P8-01 set to 1. The parameter titles change accordingly.
P8-12 Water Level Auto-Restart Time
Range Description Default
Sets the amount of time the iQpump Controller will wait before attempting an
0.1 - 6000.0 min
P8-13 Water Level Sleep Wake-Up Time
Range Description Default
0 - 3600 sec
P8-14 Low Water Detection Time Unit
auto-restart of the "Low Water" fault. Effective only when P8-09 = 3 and L5-01 > 0.
If the iQpump Controller is forced to sleep based upon the minimum water level (P8-04), the pressure must go above the P8-05 level for more than this time in order to wake up.
5.0 min
1 sec
Setting Description
0 Minutes (factory default)
1 Seconds
Multi-function Output Setting H2-: Low Suction (Modified)
Setting Description
This output will energize if P8-01 = 1 and the level in the well drops below the Low Level Detection Level
57 Low Water
18 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
(P8-07) for more than the Low Level Detection Delay Time (P8-08), or if there is a LOWWL - Low Water Level Fault.
7 Suction Control via Constant Pressure w/Well Draw Down
Multi-function Output Setting H2-: Low Suction (New)
Setting Description
This output will energize if P8-01 = 2 and the suction pressure drops below the Low Suction Pressure
58 Low Suction
Detection Level (P8-07) for more than the Low Suction Pressure Delay Time (P8-08), or if there is a LOSUC - Low Suction Pressure Fault.
Related Parameters
When P1-01 = 3 (Memobus Multiplex), the Water Level or Suction Pressure can be transmitted or read through the network. Refer to Water Level/Suction Pressure Control in Memobus Multiplex on page 26.
P9-50 Water Level Source (P8-01 set to 0 or 1, Water Level Control)
Defines which signal to use for Water Level Control (P8-) when P1-01=3.
Setting has no effect when P1-01≠3.
Setting Description
0 Analog Only (factory default)
1 Analog -> Network, No Alarm
2 Analog -> Network, Alarm
3 Network Only
P9-50 Suction Pressure Level Source (P8-01 set to 2 Suction Control)
Defines which signal to use for Suction Pressure Control (P8-) when P1-01=3.
Setting has no effect when P1-01≠3.
Setting Description
0 Analog Only (factory default)
1 Analog -> Network, No Alarm
2 Analog -> Network, Alarm
3 Network Only
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 19

8 Geothermal Mode

Aquifer
Heat
Exchanger
iQpump
Geothermal
Heater
Thermostat
Pool
M
Discharge Valve
Pool
Pump
Filter
M
Well
Pump
Discharge
Well
Warm Cold
Cold
Warm
Cold
Warm
Warm
Cold
Cold
8 Geothermal Mode

Overview

A geothermal well facilitates heat transfer between the earth and a known system, such as space heating, dehydration, electric power generation and food processing.
The geothermal function has the ability to regulate the speed of the iQpump Controller based on an external temperature signal following a preset temperature-speed curve.
Figure 3
Figure 3 Geothermal Pool Application

Geothermal Mode Setup Procedure

Follow these steps to setup the iQpump Controller in Geothermal Mode:
1. Set iQpump Controller parameters to factory default by setting parameter A1-03 to 2220 for "2 Wire" Initialization.
2. Disable PI Control by setting parameter b5-01 to 0 "Disabled"
3. Set multi-function analog input function H3-09 to 20 "Geothermal Temp"
4. Enable Geothermal Mode by setting parameter b1-01 to 5 "Geothermal Mode"
5. Set display monitor to selectable mode by setting o1-06 to 1 "3 Mon Selectable"
6. Select monitor display line #1 to display output frequency by setting o2-01 to 2 "Output Freq"
7. Select monitor display line #3 to display temperature by setting o1-08 to 80 "Geothermal Temp"
8. Set the temperature transducer range to properly scale the input. (P4-31 and P4-32)
9. Set the temperature-speed curve to the intended operation (P4-33 ~ P4-38)
<1>
"3330" for 3-Wire Initialization or "7770" for General Purpose Initialization can also be used.
20 YAS KAWA TM.iQp.10 iQpump Technical Manual Supplement
<1>

Related Parameters

b1-01 Frequency Reference Source Selection
Set the frequency reference source (b1-01) to 5 for Geothermal mode operation.
Setting Description
0 Operator - Digital preset speed d1-01 (factory default)
1 Terminals - Analog input Terminal A1 (or Terminal A2, see parameter H3-13)
2 Serial Com - RS-485 terminals R+, R-, S+ and S-
3 Option PCB - Option board connected at 2CN
5
Geothermal Mode - Frequency reference is dependent on temperature input (H3-09 = 20) <0035>
H3-09 Multi-function Analog Input Setting: Geothermal Temperature
Program H3-09 = 20 to use an external temperature sensor in Geothermal mode operation.
Setting Description
Geothermal Temperature
20
Analog input function for use with an external temperature sensor. Range scaling: 0V (or 4 mA) = P4-31 ~ 10 V (or 20 mA) = P4-32
8 Geothermal Mode
Geothermal Mode Frequency/Temperature Characteristic
When the iQpump Controller is set to operate in Geothermal Mode (b1-01 = 5), the drive's frequency reference is determined by the analog input Geothermal Temperature (H3-09 = 20) and the setting of parameters P4-33 ~ P4-38. The Geothermal characteristic can be programmed in normal or inverse operation.
Normal Operation: P4-33 < P4-34 and P4-35 < P4-36 < P4-37 < P4-38
Inverse Operation: P4-33 > P4-34 and P4-35 < P4-36 < P4-37 < P4-38
P4-31 Minimum Geothermal Temperature Input
Range Description Default
-110.0 ~ 440.0 °F Sets the temperature that corresponds to a 0V (or 4 mA) analog input. 0.0 °F
P4-32 Maximum Geothermal Temperature Input
Range Description Default
-110.0 ~ 450.0 °F Sets the temperature that corresponds to a 10V (or 20 mA) analog input. 150.0 °F
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 21
8 Geothermal Mode
Maximum
Geothermal Speed
㩿㪧㪋㪄㪊㪋㪀
Minimum
Geothermal Speed
㩿㪧㪋㪄㪊㪊㪀
Low Temp
@ Max
㩿㪧㪋㪄㪊㪌㪀
Low Temp
@ Min
㩿㪧㪋㪄㪊㪍㪀
Temperature (°F)
High Temp
@ Min
㩿㪧㪋㪄㪊㪎㪀
High Temp
@ Max
㩿㪧㪋㪄㪊㪏㪀
Maximum
Geothermal Speed
㩿㪧㪋㪄㪊㪋㪀
Minimum
Geothermal Speed
㩿㪧㪋㪄㪊㪊㪀
Low Temp
@ Max
㩿㪧㪋㪄㪊㪌㪀
Low Temp
@ Min
㩿㪧㪋㪄㪊㪍㪀
Temperature (°F)
High Temp
@ Min
㩿㪧㪋㪄㪊㪎㪀
High Temp
@ Max
㩿㪧㪋㪄㪊㪏㪀
Parameter No. Parameter Name Description Range Default
Sets the frequency reference characteristics based on the set
P4-33 Minimum Geothermal Speed
temperature points and the corresponding frequency.
For proper operation, P4-34 > P4-33 and P4-38 > P4-37 > P4-36 > P4-35.
0.00 ~
120.00 Hz
40.00 Hz
P4-34 Maximum Geothermal Speed
P4-35
P4-36
P4-37
P4-38
Low Temperature to Run at Maximum Geothermal Speed
Low Temperature to Run at Minimum Geothermal Speed
High Temperature to Run at Minimum Geothermal Speed
High Temperature to Run at Maximum Geothermal Speed
Geothermal Mode Frequency/Temperature Characteristic| (Normal Operation)
.
Geothermal Mode Frequency/Temperature Characteristic (Inverse Operation)
0.00 ~
120.00 Hz
-110.0 ~
450.0 °F
-110.0 ~
450.0 °F
-110.0 ~
450.0 °F
-110.0 ~
450.0 °F
40.00 Hz
55.0 °F
65.0 °F
75.0 °F
85.0 °F
Figure 4
22 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
Figure 4 Digital Operator Display Showing Geothermal Mode Operation
8 Geothermal Mode
Geothermal Mode iQpump Controller Monitors
U1-80 Geothermal Temperature Input
Unit Description
0.1 °F
Note: Only shown when b1-05 = 5 (Geothermal Mode)
Displays Geothermal temperature input after gain and bias are applied. This is the temperature used by the Geothermal Function to determine what frequency to run the iQpump Controller.
Geothermal Temperature Loss Detection
P4-39 Geothermal Temperature Loss Detection
Selects iQpump Controller action when the temperature sensor signal from Terminal A2 is below 3 mA or above 21 mA.
Setting Description
0Disabled
1 Alarm (factory default)
2Fault
Note: Only effective when H3-08 = 2 (4-20 mA) and H3-09 = 20 (Geothermal Temp).
Alarm: Temperature Sensor Lost (factory default)
Alarm Display Description
The geothermal temperature sensor is not present. Alarm occurs when:
1. b1-01 = 5, H3-09 = 20 (Geothermal Temp), H3-08 = 2 (4-20 mA), P4-39 = 1 (Alarm), and the input has either dropped below 3 mA or went above 21 mA. Check: Ensure the device connected to Terminal A2 is installed and working properly.
2. b1-01 = 5, and H3-09 20 (Geothermal Temp), and P4-39 = 1 (Alarm).
Tem p Lo s t
Geothermal Input
Alarm: Geothermal Parameters Programming Error
Alarm Display Description
Geo Params
Check P4-35~P4-38
Fault: Temperature Sensor Lost
Fault Display Description
Tem p Lo s t
Geothermal Input
Check: Terminal A2 must be assigned to Geothermal Temp (H3-09 = 20).
3. b1-01 = 5, H3-09 = 20 (Geothermal Temp), H3-08 = 2 (4-20 mA), P4-39 = 2 (Fault), the drive is either in HAND mode or has no run command, and the input is below 3 mA or above 21 mA. Check: Ensure the device connected to Terminal A2 is installed and working properly.
4. b1-01 = 5, and H3-09 20 (Geothermal Temp), P4-39 = 2 (Fault), and the drive is either in HAND mode or has no run command. Check: Terminal A2 needs to be assigned to Geothermal Temp (H3-09=20).
The drive is running at the P4-31 level because of an incorrect setting.
The temperature parameter values must be set in the following order: P4-35 < P4-36 < P4-37 < P4-38
The geothermal temperature sensor is not present. Fault occurs when:
1.b1-01 = 5, and H3-09 20 (Geothermal Temp), and P4-39 = 2 (Fault). Check: Terminal A2 must be assigned to Geothermal Temp (H3-09 = 20).
2.b1-01 = 5, H3-09 = 20 (Geothermal Temp), H3-08 = 2 (4-20 mA), P4-39 = 2 (Fault), the drive is either in HAND mode or has no run command, and the input is below 3 mA or above 21 mA. Check: Ensure the device connected to Terminal A2 is installed and working properly.
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 23
8 Geothermal Mode
60Hz
㩿㪧㪋㪄㪊㪋㪀
35Hz
㩿㪧㪋㪄㪊㪊㪀
55°F
㩿㪧㪋㪄㪊㪌㪀
65°F
㩿㪧㪋㪄㪊㪍㪀
Temperature (°F)
75°F
㩿㪧㪋㪄㪊㪎㪀
85°F
㪧㪋㪄㪊㪏㪀
Output
Temperature
Transducer
V+
A2
AC
+V
EG
iQpump
Output
V+
A2
AC
+V
EG
iQpump
Common
4-20mA
Temperature
Transducer
0-10V
Set S1-2 to OFF
O
F F
1
2
Set S1-2 to ON
* Factory Default
O
F F
1
2
Set parameter H3-08 Terminal A2 Signal to ‘0’ for 0 – 10V.
Fault: Geothermal Frequency Reference/Temperature Curve Error: OPE19
Fault Display Description
Parameter selection is not compatible with the Geothermal Mode (b1-01 = 5) setting. Fault occurs when
OPE19
Geothermal Set
b1-01 = 5 (Geothermal Mode), and one of the following is NOT set:
- b5-01 = 0 (PI Disabled)
- P1-01 = 0 (Simplex) Check: Confirm parameter settings for b1-01, b5-01, and P1-01.
Example: Geothermal Pump System Transducer Setup
Geothermal Pump system using a temperature transducer 4 - 20 mA with a temperature range of 0 - 120 °F. The
Example:
system requires a minimum pump speed of 35 Hz. The system requires increased pump speed when the temperature increases above 75 °F or when temperature falls below 65 °F.
Geothermal Mode Setup Procedure:
Refer to Geothermal Mode Setup Procedure on page 20 for steps to setup the iQpump Controller for Geothermal Mode.
24 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement
8 Geothermal Mode
Related Parameters for Geothermal Mode Operation
Parameter Description iQpump Controller
b1-01 Frequency Reference Source 5
b5-01 PI Mode 0
H3-09 Terminal A2 Function Selection 20
o1-06 User Monitor Selection Mode 1
o1-02 <1>
o1-08 Third Line User Monitor Selection 80
P1-01 Pump Mode 0
P4-31 Minimum Geothermal Temperature Input 0.0 °F
P4-32 Maximum Geothermal Temperature Input 120.0 °F
P4-33 Minimum Geothermal Speed 35.00 Hz
P4-34 Maximum Geothermal Speed 60.00 Hz
P4-35 Low Temperature to Run at Max. Geothermal speed 55.0 °F
P4-36 Low Temperature to Run at Min. Geothermal speed 65.0 °F
P4-37 High Temperature to Run at Min. Geothermal speed 75.0 °F
P4-38 High Temperature to Run at Max. Geothermal speed 85.0 °F
P4-39 Geothermal Temperature Loss Detection 1
<1> After programming is completed, cycle power to the drive, as changes to the o1-02 monitor requires a drive power-up cycle to be effective.
User Monitor Selection After Power-Up 2
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 25

9 Water Level/Suction Pressure Control in Memobus Multiplex

9 Water Level/Suction Pressure Control in Memobus Multiplex

Overview

When using Water Level or Suction Pressure Control in a Memobus Multiplex system, (see parameter P8-01). The iQpump Controller can be programmed to receive the Water Level or Suction Pressure information from another drive on the network with a Water Level or Suction Pressure input.

Parameters

Parameter representation for this function depends on the setting of parameter P8-01.
Parameter Settings
P8-01 Water Level/Suction Pressure Selection
Setting Description
0 Disabled (factory default)
1 Water Level Control (Previous version PRG: 0034 functionality for setting is "Enabled")
2 Suction Pressure Control
P9-50 Water Level Source (P8-01 set to 0 or 1, Water Level Control)
Setting Description
0 Analog Only (factory default)
1 Analog -> Network, No Alarm
2 Analog -> Network, Alarm
3Network Only
P9-50 Suction Pressure Level Source (P8-01 set to 2 Suction Pressure Control)
Setting Description
0 Analog Only (factory default)
1 Analog -> Network, No Alarm
2 Analog -> Network, Alarm
3Network Only
P9-50 = 0 (Analog Only)
This drive will transmit its Water Level/Suction Pressure signal to the network. If this signal is lost, the drive will not switch to another signal on the network. The detection action in this mode is solely on parameter b5-12.
P9-50 = 1 (Ana->Net, No Alarm)
This drive will transmit its Water Level/Suction Pressure signal to the network when the analog input is healthy. If this signal is lost, the drive will switch to another signal on the network if available. If there are no available network signals, then this drive will act according to the setting of parameter b5-12 (when = 1 or 2).
Note: Setting parameter b5-12 to 0 will disable analog feedback detection and will prevent the iQpump Controller from switching to the
Network Water Level/Suction Pressure.
P9-50 = 2 (Ana->Net, Alarm)
This drive will transmit its Water Level / Suction Pressure signal to the network when the analog input is healthy. If this signal is lost, the drive will switch to another signal on the network if available. An alarm will be displayed indicating that the analog signal has been lost and that the signal is coming from the network. If there are no available network signals, then this drive will act according to the setting of parameter b5-12 (when = 1 or 2).
Note: Setting parameter b5-12 to 0 will disable analog feedback detection and will prevent the iQpump Controller from switching to the
Network Water Level/Suction Pressure.
26 YAS KAWA TM.iQp.10 iQpump Technical Manual Supplement
9 Water Level/Suction Pressure Control in Memobus Multiplex
P9-50 = 3 (Network Only)
The drive will always use a valid network Water Level / Suction Pressure signal. If there are no available network signals, the drive will act according to the setting of b5-12 with the following differences:
Setting b5-12 to 1: Instead of an alarm, Network P8- Lost message is displayed.
Setting b5-12 to 2: Instead of a fault, Network P8- Lost message is displayed. In this condition the iQpump
Controller no longer accepts iQpump Controller Network run commands and the stopping method is fixed to coast-to­stop.
b5-12 PI Feedback Reference Missing Detection Selection
Setting Description
0Disabled
1Alarm
2 Fault (factory default)
Any drive with b5-12 set to 0 (Disabled) and P9-50 ≠ 3 will effectively have no Terminal A1 signal loss detection and will continuously send the Water Level/Suction Pressure to the iQpump Controller Network regardless of a faulty or an invalid signal.
When b5-12 set to 2 (Fault) and P9-02 = 3, iQpump Controller will display a feedback loss alarm instead of a fault when one of the following conditions active:
1. The iQpump Controller is in Hand Mode
2. There is no Lead iQpump Controller on the network
3. The iQpump Controller is not in Auto Mode
Minimum Water Level or Suction Pressure Detection Operation
In case of a minimum water level or suction pressure, the lead iQpump Controller will de-stage when one or more Lag iQpump Controllers are present, otherwise the lead iQpump Controller will go to sleep.
Note: Setting parameter b5-12 to 0 will disable analog feedback detection and will prevent the iQpump Controller from switching to the
Network Water Level/Suction Pressure.
Related Parameters
b5-12 PI Feedback Reference Missing Detection Selection
Setting Description
0 Disabled (factory default)
1Alarm
2Fault
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 27
9 Water Level/Suction Pressure Control in Memobus Multiplex
P2-03 Sleep Delay Time
Sets delay time when water level or suction pressure signal falls below the minimum level programmed (P8-04).
Range Description
Network Multiplex Mode: Water Level Control (P8-01 = 1): Parameter specifies the time delay before the lead drive de-stages when the water level (U1-98) falls below the Minimum Water Level (P8-04).
Network Multiplex Mode: Suction Control (P8-01 = 2): Parameter specifies the time delay before the
0 - 3600 sec
Default: 5 sec
lead drive de-stages when the suction pressure (U1-98) falls below the Minimum Suction Pressure (P8-
04).
Contactor Multiplex Mode (P1-01 =1 or 2): There are two contactor multiplex modes, one for Water Level and another for Suction. Contactors programmed for Multiplex (H2- = 40 and 41) will open one by one when the Water Level (U1-98) is below the Minimum Water Level (P8-04) for the time set in the Sleep Delay Time (P2-03).
IMPORTANT: Staging Restriction when in Water Level/Suction Pressure Control.
• When P1-01 = 3 (Memobus Network) and the iQpump Controller’s PI output is being influenced by the Water Level/ Suction Pressure Control, pump system staging is disabled.
• When P1-01 = 1 or 2 (Multiplex system) and the iQpump Controller’s PI output is being influenced by the Water Level/ Suction Pressure Control, Multi-function output contacts (H2- = 40 and 41) are prohibited from closing.
Alarm: Water Level/Suction Pressure Transducer Lost
Alarm Display Description
Analog Terminal A1 signal is lost and the Network Water Level or Suction Pressure signal is now used.
AnalogA1 Lost
Switched to Net
Alarm: Network Water or Suction Pressure Signal Lost
Alarm Display Description
Net Wtr/Suctn
Lost, Chk Source
Fault: Programming Error: OPE13 Terminal A1
Fault Display Description
OPE13
Terminal A1
Cause: Defective or broken analog input source.
Countermeasure: Check to ensure the Water Level or Suction Pressure source is installed and working
properly. If no signal is present, set P9-50 = 3 to have it always read from another drive's network Water Level or Suction Pressure.
Network source for Water Level or Suction Control Pressure is lost.
Cause: Valid analog source for Water Level or Suction Control Pressure can not be found on the network. Countermeasure: Check the source on drives configured as P9-50 3.
Cause: Terminal A1 is assigned to more than one of the following functions.
• Frequency Reference (b1-01 = 1)
• Dual Zone PI is enabled (b5-01 = 2)
• Flow Meter Enabled (P6-01 > 0)*
• Water Level/Suction Pressure Control Enabled (P8-01>0)*
• Hand Mode Ref Term A1 (P5-01 = 0)
Note: An OPE13 error will not be generated if b1-01 = 1 and P5-01 = 0, and none of the other conditions
listed above apply.
*When P1-01 = 3 (Memobus), a setting of P6-01 > 0 and P8-01 > 0 is allowed only if P9-40 = 1 or P9-50 = 3. Countermeasure: Reprogram b1-01, b5-01, P6-01, or P8-01.
Note: The OPE13 fault is modified in drive software version PRG:0035. Previous software versions have different functionality.
28 YASK AWA TM.iQp.10 iQpump Technical Manual Supplement

10 iQpump MEMOBUS/Network Operation

10 iQpump MEMOBUS/Network Operation

Lag Drive Speed Follower and Lag Turn Off

Certain multiplex pump systems require a common pump speed for all running pumps. To allow for this operation, a new selection "Follow Lead Spd" is added to parameter P9-05.
Function Description
P9-05 Lag Drive Mode
Setting Description
0
1
2
3
Note: The rate at which the Lag drives follow the Lead drive's output speed is dependent on the communication speed of the iQpump
Controller network (H5-02), the maximum number of iQpump Controllers on the network (P9-25) and the current number of iQpump Controllers online.
Fixed Speed (factory default) Runs at the P9-06 setting after the P9-07 time expires.
PI Regulation. Uses PI to determine speed.
Turn Off The iQpump Controller stops running when it switches to a lag drive after P9-07 time expires.
Follow Lead Spd The iQpump Controller follows the speed of the active Lead drive. Use P9-30 gain and P9-31 bias setting to adjust reference signal.
Display Message for Follow Lead Speed Operation:
When the iQpump Controller is operating in Auto Mode, Lag operation and the Follow Lead Speed function is active (P9-05 = 3) then the display will show the actual frequency reference in Hz instead of the selected system units.
YAS KA WA TM.iQp.10 iQpump Technical Manual Supplement 29
Loading...
+ 65 hidden pages