Welch Allyn ProBP-3400 User Manual

Page 1
Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
Directions for use
Page 2
© 2011 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided byWelch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Welch Allyn, SureBP Technology, and Welch Allyn FlexiPort are registered trademarks of Welch Allyn. The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Welch Allyn is under license.
Software in this product is Copyright 2011 Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, contact Welch Allyn Technical Support (www.welchallyn.com/ support) or call:
USA
+1 800 535 6663
Australia
+61 2 9638 3000
+1 315 685 4560
Canada
European Call Center
Germany
Malaysia
Singapore
United Kingdom
+1 800 561 8797 +353 46 90 67790 +49 695 098 5132
Japan
+81 4 2703 6084 +603 7875 3341 +65 6419 8100
Spain
+34 917 499 357 +44 207 365 6780
China
France
Italy
Latin America
Netherlands South Africa
Sweden
+86 21 6327 9631 +33 155 69 58 49 +39 026 968 2425 +1 305 669 9003 +31 202 061 360 +27 11 777 7555 +46 85 853 6551
103701 (CD)
DIR 80016085 Ver. B
Welch Allyn, Inc. 4341 State Street Road
Skaneateles Falls, NY 13153-0220 USA
www.welchallyn.com
Regulatory Affairs Representative
Welch Allyn, Limited
Navan Business Park
Dublin Road
Navan, County Meath
Republic of Ireland
Page 3

Contents

Initial device setup ................................................................................. 1
Introduction ............................................................................................. 3
Symbols ................................................................................................... 5
General warnings and cautions ............................................................. 9
iii
Intended use ...................................................................................................................................... 3
NIBP warnings and cautions ................................................................ 13
Contents checklist ................................................................................. 15
Controls and indicators ......................................................................... 17
Screen elements .................................................................................... 19
Connections ........................................................................................... 21
Blood pressure hose and cuff .............................................................. 23
Blood pressure cuff selection ....................................................................................................... 23
Power transformer ................................................................................ 25
Power transformer and wall plug/line cord assembly .............................................................. 25
Charge the device ........................................................................................................................... 26
Setup ...................................................................................................... 27
Insert the battery ............................................................................................................................. 27
Mount the device ............................................................................................................................ 28
Assemble power cord and wall plug/line cord .......................................................................... 28
Charge the ProBP 3400 .................................................................................................................. 28
Initial startup .................................................................................................................................... 28
Page 4
iv
Contents Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
Battery ................................................................................................... 31
Startup .................................................................................................... 33
Standby mode ....................................................................................... 35
Blood pressure measurement .............................................................. 37
Settings .................................................................................................. 39
Settings matrix ................................................................................................................................. 39
Unit of measure ............................................................................................................................... 39
Pressure presets ............................................................................................................................. 40
Bluetooth wireless technology ..................................................................................................... 40
Advanced settings .......................................................................................................................... 41
Data review ............................................................................................ 45
Review data ..................................................................................................................................... 45
Delete data ....................................................................................................................................... 45
Maintenance and service ...................................................................... 47
Calibrate the device ........................................................................................................................ 47
Change the battery ......................................................................................................................... 47
Clean the device .............................................................................................................................. 48
Specifications ........................................................................................ 51
Physical specifications .................................................................................................................. 51
Mechanical specifications ............................................................................................................ 51
Electrical specifications ................................................................................................................ 52
Environmental specifications ........................................................................................................ 52
Product disposal ............................................................................................................................. 53
Standards and compliance ................................................................... 55
General radio compliance ............................................................................................................. 55
Guidance and manufacturer's declaration .......................................... 57
Emissions and immunity information ........................................................................................... 57
Warranty ................................................................................................ 61
Troubleshooting ................................................................................... 63
Inaccurate blood pressure readings ........................................................................................... 63
Cuff inflation and deflation with no blood pressure reading displayed ................................. 64
No cuff inflation .............................................................................................................................. 64
Cuff pops off ..................................................................................................................................... 65
Cuff deflating too slowly ................................................................................................................ 65
Cuff inflates beyond SureBP inflation pressure preset while device is in use ..................... 65
Device does not turn on ................................................................................................................. 66
Bluetooth troubleshooting ............................................................................................................. 67
Page 5
Directions for use Contents v
Appendix ................................................................................................ 69
ProBP 3400 approved accessories .............................................................................................. 69
Configuration options ..................................................................................................................... 72
Page 6
Contents Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
vi
Page 7

Initial device setup

Before using ProBP 3400 for the first time, you must configure the device for use. See Setup for additional information.
1
Page 8
Initial device setup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
2
Page 9

Introduction

This Directions for use manual is a comprehensive guide designed to help you understand the capabilities and operation of the ProBP 3400 non-invasive blood pressure device. The information in this manual includes all options available with the device. Read this manual thoroughly before attempting to setup, configure, use, troubleshoot, or maintain the device.

Intended use

3
The ProBP 3400 automatically measures systolic and diastolic pressure (excluding neonates) and pulse rate, as well as calculates Mean Arterial Pressure (MAP).
The device is intended to be used by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.
Page 10
Introduction Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
4
Page 11

Symbols

Documentation symbols

Warning: The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death.
5

Power symbols

Caution: The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data.
Consult operating instructions.
(green indicator) External power present, battery charged
(amber indicator) External power present, battery is charging
No external power present Battery charge level
Battery charging
Rechargeable battery

Button symbols

Power on/power off Return to previous screen
Blood Pressure Start/Stop Select
Page 12
6
Symbols Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
Navigation (Up, Down, Left, Right)

Shipping, storing, and environment symbols

Fragile; handle with care Relative humidity
limits
Temperature limits
Recovery/Recyclable Recycle the product
This end up Keep dry
IPXØ
Equipment is not protected against the ingress of liquid

Connectivity symbols

Bluetooth® wireless technology enabled USB connection
Devices are connected via Bluetooth wireless technology
Lithium-ion battery
separate from other disposables
Bluetooth radio is disabled or not paired

Miscellaneous symbols

Meets essential requirements of European Medical Device Directive 93/42/EEC
Wireless radio alert in Europe. European Communities Class 1 radio equipment.
Intertek ETL listed Type BF applied parts
Regulatory Affairs Representative
EMC Framework of Australia
Page 13
Directions for use Symbols 7
Class II equipment Available for sale only
upon order of a physician or licensed health care provider
Serial number Manufacturer
Reorder number Call for maintenance
Non-ionizing electromagnetic radiation
Page 14
Symbols Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
8
Page 15

General warnings and cautions

WARNING The information in this directions for use is a comprehensive guide to the operation of ProBP 3400. For best results, read this directions for use thoroughly before using the device.
WARNING The device is designed for medical clinician use. Although this directions for use may illustrate medical spot-check techniques, only a trained clinician who knows how to take and interpret a patient's vital signs should use this device.
9
WARNING The device is intended for use only in environments with clinician
supervision.
WARNING The device is not intended for continuous monitoring. Do not leave the device unattended while taking measurements on a patient.
WARNING The device is not intended for use during patient transport.
WARNING Fire and explosion hazard. Do not operate the device in the presence
of a flammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen­enriched environments; or in any other potentially explosive environment.
WARNING Use only Welch Allyn approved accessories. Use of unapproved accessories with the device can affect patient and operator safety, and can reduce product performance and accuracy. To ensure patient safety and optimal product performance, use only accessories and supplies recommended for or supplied with the device, and use according to the accessory manufacturer's directions for use.
WARNING Every three months, inspect the blood pressure cuff and other accessories for fraying or other damage. Replace as necessary.
WARNING Inaccurate measurement risk. Do not use the device on patients who are connected to heart/lung machines.
WARNING Electric shock hazard. Do not open the device or attempt repairs. There are no user-serviceable parts inside ProBP 3400 other than battery replacement. Only perform routine cleaning and maintenance procedures specifically described in this directions for use. Inspection and servicing of internal parts shall only be performed by qualified service personnel.
WARNING The device complies with applicable domestic and international standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using the device in close proximity to other equipment.
Page 16
10
General warnings and cautions Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
WARNING Welch Allyn is not responsible for the integrity of any mounting installation. Welch Allyn recommends that customers contact their Biomedical Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory.
WARNING The device is not defibrillator proof.
WARNING The device may not function properly if dropped or damaged. Do not
use the device if you notice any signs of damage. Qualified service personnel must check any device that is dropped or damaged for proper operation before putting the device back into use.
WARNING Defective batteries can damage the device. If the battery shows any signs of damage, leakage, or cracking, it must be replaced immediately, and only with a battery recommended for or supplied with the device.
WARNING Improper disposal of batteries may create an explosion or contamination hazard. Never dispose of batteries in refuse containers. Do not dispose of the battery in fire. Always recycle batteries according to local regulations.
WARNING Improper handling of the battery can lead to heat generation, smoke, bursting, or fire.
WARNING Do not remove the label from the battery.
WARNING Do not disassemble, modify, or solder the battery.
WARNING Do not directly connect or short circuit the positive (+) and negative (-)
battery terminals.
WARNING To avoid short circuits, keep battery terminals away from metal objects.
WARNING Do not expose the battery to temperatures higher than 80 °C/176 °F.
WARNING If the accuracy of any measurement is in question, check the patient's
vital sign(s) with an alternate method and then check to verify the device is functioning properly.
WARNING For proper patient electrical isolation and battery charging, use only the provided external power supply to charge the device.
WARNING Electric shock hazard. Use the USB connector only to connect to devices complying with IEC 60601-1 or other IEC standards as appropriate to the device. The user is responsible for verifying that the system complies with the requirements of the system standard IEC 60601-1-1 if additional devices are connected to the ProBP 3400.
WARNING Electric shock hazard. Before cleaning the device, disconnect the power cord from the power source and the device.
Caution Take care to prevent water or other fluid from entering any connectors on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions.
Page 17
Directions for use General warnings and cautions 11
Caution The device is not heat-resistant. Do not autoclave.
Caution Use the device within stated operating temperature ranges. The device
will not meet performance specifications if used outside these temperatures ranges.
Caution Always unplug the external power source from the outlet before moving the device to a new location.
Page 18
General warnings and cautions Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
12
Page 19

NIBP warnings and cautions

WARNING ProBP 3400 is not intended to measure blood pressure on neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28 days or less of age if born at term (37 weeks gestation or more); otherwise up to 44 gestational weeks.
WARNING To ensure pediatric blood pressure accuracy and safety, the Small Child Reusable Cuff (REUSE-08) and the Small Child Disposable Cuff (SOFT-08) are the smallest cuffs approved for use with young children. The child's arm must fit within the range markings on the cuff.
13
WARNING Do not compress the blood pressure hose or cuff. This may cause
system errors or patient safety risks to occur.
WARNING NIBP readings may be inaccurate for patients experiencing moderate to severe arrhythmia.
WARNING Inaccurate measurement risk. Do not use the device on patients who are experiencing convulsions or tremors.
WARNING Use only blood pressure cuffs and hoses listed as approved accessories to ensure safe and accurate NIBP measurements.
WARNING Patient injury risk. When several blood pressure measurements are taken on the same patient, regularly check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.
WARNING Do not allow a blood pressure cuff to remain on the patient more than 3 minutes when inflated above 15 mmHg. Excessive cuff tightness may cause venous congestion, peripheral nerve injury, discoloration of the limb, and patient distress.
WARNING Inaccurate measurement risk. Do not place the cuff where it can disturb proper circulation. Do not place the cuff on any area where circulation is compromised or on any extremity used for intravenous infusions.
WARNING The blood pressure cuff must be properly positioned to ensure blood pressure accuracy and patient safety. Wrapping the cuff too loosely (preventing proper inflation) may result in inaccurate NIBP readings.
WARNING Patient injury risk. Never install Luer Lock connectors on Welch Allyn blood pressure tubing. Using these connectors on blood pressure cuff tubing creates the risk of mistakenly connecting this tubing to a patient's intravenous line and introducing air into the patient's circulatory system.
Page 20
14
NIBP warnings and cautions Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
WARNING NIBP measurements may be inaccurate in the presence of excessive motion artifact. Minimize extremity and cuff motion during blood pressure readings.
WARNING The position and physiologic condition of the subject can affect a blood pressure reading.
WARNING If the blood pressure cuff is not at heart level, note the difference in reading due to the hydrostatic effect. Add the value of .2 kPa (1.80 mmHg) to the displayed reading for every 2.5 cm (inch) above heart level. Subtract the value of .2 kPa (1.80 mmHg) from the displayed reading for every 2.5 cm (inch) below heart level.
WARNING Proper blood pressure cuff size and placement is essential to the accuracy of the blood pressure determination. See Blood pressure cuff selection for sizing information.
Page 21

Contents checklist

Unpack the ProBP 3400 and any applicable accessories and inspect for missing items. Retain the shipping materials in the event of shipping damage or for return, if necessary, to Welch Allyn for repair or warranty service. Report any signs of shipping damage to the carrier. Report any missing or damaged items to the Welch Allyn Service Center near you.
All ProBP 3400 devices include the following components:
ProBP 3400 Device. This device automatically measures systolic and diastolic pressure (excluding neonates) and pulse rate, as well as calculates Mean Arterial Pressure (MAP).
15
Directions for Use CD. Read this Directions for Use thoroughly before using the ProBP 3400.
Save this CD for reference.
Battery. Install the battery before using the device. See Battery for additional information.
Blood Pressure Cuff(s). See Blood pressure hose and cuff for additional information.
Blood Pressure Hose. Latex-free pressure hose with connectors to attach various sizes of
blood pressure cuffs to the Welch Allyn ProBP 3400 device.
USB Cable. Attaches to Power Transformer and device to provide power to the ProBP 3400 and charge the internal battery.
Power Transformer and Wall Plug/Line Cord. Power transformer and wall plug (or line cord) assembles and attaches to USB cable to provide power to the ProBP 3400 and charge the internal battery.
Welch Allyn Service Tool CD. The Welch Allyn Service Tool supports maintenance activities on select Welch Allyn products.
Startup Guide. Use the Startup Guide to setup device for first-time use.
Warranty. Complete the ProBP 3400 warranty today at www.welchallyn.com/warranty.
Page 22
Contents checklist Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
16
Page 23

Controls and indicators

17
1. Power on/power off button: controls power to the device.
2. Blood Pressure Start/Stop button: initiates a new blood pressure cycle from the Home
screen. Pressing again aborts an active blood pressure measurement. This button returns user to the Home screen from any other screen on the device.
3. Up navigation button: highlights the previous option in the Display window or increases
numeric values.
4. Return button: returns the user to the previous screen.
5. Right navigation button: highlights the Settings tab in the Display window or highlights
options to the right.
6. Charging LED: indicates when device is connected to external power and battery charging
state.
7. Down navigation button: highlights the next option in the Display window or decreases
numeric values.
8. Select button: selects list item that has been highlighted.
9. Left navigation button: highlights the Review tab in the Display window or highlights
options to the left.
Page 24
Controls and indicators Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
18
Page 25

Screen elements

The liquid crystal display (LCD) may indicate any of the following: systolic blood pressure (mmHg or kPa), diastolic blood pressure (mmHg or kPa), MAP (mmHg or kPa), pulse rate (bpm), date, time, record number, and battery charge level.
Note Your model may not contain all of these options.
19
1. NIBP display: shows systolic and diastolic values of NIBP readings.
2. Bluetooth radio status (if equipped): indicates status of Bluetooth radio.
3. USB: indicates USB is connected.
4. Battery level indicator: displays the battery charge level.
5. Date: shows the current date.
6. Clock: shows the current time.
7. Reading number identified: shows what reading the device is on. The ProBP 3400 device
retains up to 50 readings in memory.
8. MAP: shows the MAP value.
9. Settings: displays the Settings menu, when selected.
10. Review: displays the Review menu, when selected.
11. Pulse rate display: shows the pulse rate.
Page 26
Screen elements Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
20
Page 27

Connections

21
1. Mounting connection
2. Battery door screw
3. Blood pressure hose connection port
4. USB/external power cord connection port
Page 28
Connections Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
22
Page 29

Blood pressure hose and cuff

Identify and have available the ProBP 3400, blood pressure cuff, and the blood pressure hose.
1. Inspect the blood pressure hose; notice that one end has a single, gray Welch Allyn FlexiPort fitting and the other end is plain with two recessed holes.
2. Completely push the plain end of the blood pressure hose onto the two silver blood pressure ports on the device. Make sure the hose is fully seated.
3. Snap the Welch Allyn FlexiPort connector onto the blood pressure cuff.
23

Blood pressure cuff selection

Careful sizing of the cuff is important for accurate blood pressure readings. If the cuff is too small or too large, you may have false high or low readings, respectively. When there is an area of overlap for using a smaller or larger cuff, use the larger size cuff.
The device uses oscillometric technology to determine blood pressure; therefore, if the cuff extends to the antecubital fossa (bend in the elbow), you can still acquire an accurate blood pressure reading.
Measure the arm circumference (midway between the elbow and shoulder) for the correct cuff size.
Wrap the cuff around the patient’s upper arm and verify that the artery index marker falls within the two divisions that identify the “range” on the cuff to indicate a proper fit.
The following table provides measurements for Welch Allyn blood pressure cuffs.
Page 30
24
Blood pressure hose and cuff Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
Cuff Size Soft
Disposable One-Piece
Reusable One­Piece Cuff
(1 per pack)
Circumference (cm)
Circumference (in)
Cuff (5 per pack)
Small child (size 8) Soft-08 Reuse-08 12.0 – 16.0 4.7 – 6.3
Child (size 9) Soft-09 Reuse-09 15.0 – 21.0 5.9 – 8.3
Small adult (size 10) Soft-10 Reuse-10 20.0 – 26.0 7.9 – 10.2
Adult (size 11) Soft-11 Reuse-11 25.0 – 34.0 9.8 – 13.4
Adult long (size 11L) Soft-11L Reuse-11L 25.0 – 34.0 9.8 – 13.4
Large adult (size 12) Soft-12 Reuse-12 32.0 – 43.0 12.6 – 16.9
Large adult long (size 12L)
Thigh (size 13) Soft-13 Reuse-13 40.0 – 55.0 15.7 – 21.7
Soft-12L Reuse-12L 32.0 – 43.0 12.6 – 16.9
For ordering information, see the ProBP 3400 approved accessories in the Appendix.
Page 31

Power transformer

Power transformer and wall plug/line cord assembly

The power transformer and wall plug are packaged separately and must be assembled prior to use. The power transformer assembles to a line cord for the mobile stand configuration of the ProBP
3400.
1. Align the grooves in the wall plug or line cord with the casing located on the transformer.
25
Ensure that the two prong plug on the wall plug or line cord and the plug receptacle on the power transformer are aligned.
2. Slide the wall plug or line cord onto the transformer until it is firmly seated.
Page 32
Power transformer Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
26

Charge the device

Charge the ProBP 3400 with the Welch Allyn-supplied power supply.
To charge the device:
1. Insert the USB B connector into the USB/external power connection port on the back of the device.
2. Insert the other end of the USB cord into USB connector on the power transformer, then plug the transformer into the main power source.
Page 33

Setup

Before first-time use of the ProBP 3400, complete the following steps:

Insert the battery

The battery is shipped separately from the ProBP 3400 device. You must install and charge the battery for six hours before the device can be used.
27
1. Remove the screw from the battery compartment door with a Phillips screwdriver.
2. Remove the battery compartment door.
3. Remove the battery from the shipping bag.
4. Using the polarized guidelines on the battery label, insert the battery, positive (+) side first, into the battery compartment.
5. Replace the battery compartment door.
6. Secure the screw with the screwdriver.
Page 34
28
Setup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Mount the device

For mounting instructions, see the accessory's directions for use. For mounting options, see the ProBP 3400 approved accessories in the Appendix.

Assemble power cord and wall plug/line cord

1. Align the grooves and two prong plug in the wall plug or line cord with the casing and plug receptacle located on the transformer.
2. Slide the wall plug or line cord onto the transformer until it is firmly seated.

Charge the ProBP 3400

To charge the device:
1. Insert the USB B connector into the USB/external power connection port on the back of the device.
2. Insert the other end of the USB cord into USB connector on the power transformer, then plug the transformer into the main power source.

Initial startup

To complete initial configuration, you must turn on the device, and select an operating language, date, and time.
1. Press the Power on/power off button. After device power up, the display window shows the Language screen.
2. Use the Up and Down navigation buttons to highlight the correct option.
3. Press the Select button to choose a language. A confirmation window displays your language selection.

Set date and date format

After you select a language, set the device Date and Date format:
1. The Date format field is highlighted. Press the Up navigation or Down navigation buttons to access the correct format option.
2. Press the Select button to accept the date format as shown.
3. Press the Down navigation button to highlight the date.
4. Press the Select button. The first numeric field is highlighted for editing.
5. Use the Up navigation button to increase the number value; use the Down navigation button to decrease the number value.
6. Press the Right navigation button to move to the next field.
7. Repeat steps 5 and 6 to edit additional numeric values.
8. Press the Right navigation button to highlight the OK button.
9. Press the Select button to accept the date as shown.
10. Press the Down navigation button to highlight the Next button.
11. Press the Select button to move to the Time and Time format screen.
Page 35
Directions for use Setup 29

Set time and time format

1. The Time format field is highlighted. Press the Up navigation or Down navigation buttons to access the correct format option.
2. Press the Select button to accept the time format as shown.
3. Press the Down navigation button to highlight the time.
4. Press the Select button. The first numeric field is highlighted for editing.
5. Use the Up navigation button to increase the number value; use the Down navigation button to decrease the number value.
6. Press the Right navigation button to move to the next field.
7. Repeat steps 5 and 6 to edit additional numeric values.
8. Press the Right navigation button to highlight the OK button.
9. Press the Select button to accept the entry.
10. Press the Down navigation button to highlight the Next button.
11. Press the Select button to complete the Setup process. The device will display the Home screen.
Page 36
Setup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
30
Page 37

Battery

31
WARNING Defective batteries can damage the device. If the battery shows any
signs of damage, leakage, or cracking, it must be replaced immediately, and only with a battery recommended for or supplied with the device.
WARNING Improper disposal of batteries may create an explosion or contamination hazard. Never dispose of batteries in refuse containers. Do not dispose of the battery in fire. Always recycle batteries according to local regulations.
WARNING Improper handling of the battery can lead to heat generation, smoke, bursting, or fire.
WARNING Do not remove the label from the battery.
WARNING Do not disassemble, modify, or solder the battery.
WARNING Do not directly connect or short circuit the positive (+) and negative (-)
battery terminals.
WARNING To avoid short circuits, keep battery terminals away from metal objects.
WARNING Do not expose the battery to temperatures higher than 80 °C /176 °F.
WARNING Electric shock hazard. Do not open the device or attempt repairs.
There are no user-serviceable parts inside ProBP 3400 other than battery replacement. Only perform routine cleaning and maintenance procedures specifically described in this directions for use. Inspection and servicing of internal parts shall only be performed by qualified service personnel.
The ProBP 3400 device is powered by a lithium-ion rechargeable battery.
The battery is shipped separately from the ProBP 3400 device. You must install and charge the battery for six hours before using the device.
To install the battery:
1. Remove the screw from the battery compartment door with a Phillips screwdriver.
2. Remove the battery compartment door.
3. Remove the battery from the shipping bag.
4. Using the polarized guidelines on the battery label, insert the battery, positive (+) side first, into the battery compartment.
Page 38
32
Battery Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
5. Replace the battery compartment door.
6. Secure the screw with the screwdriver.
The battery charges when the ProBP 3400 is connected to an external power source. While the
ProBP 3400 is charging, the symbol displays an amber indicator and the battery charging
indicator, , displays on the Home screen. The battery charging indicator is only visible
when the device is powered on. When the battery is charged, the symbol displays a green indicator, and the battery level indicator is steady with all segments continuously shown on the Home screen. An operator can use the device while the battery is charging; however, the battery charges faster when the instrument is not in operation.
A dead battery may result if the ProBP 3400 is left uncharged or unused for a long period of a time. If this occurs, charge the battery by connecting the ProBP 3400 to an external power source. Charge the battery for at least six hours before disconnecting the device from the power supply.
If the ProBP 3400 will be unused for several months or longer, remove the battery prior to storing the device.
Page 39

Startup

33
Press the Power on/power off button to turn the device on or off. Upon each power up, the display lights up and the ProBP 3400 displays the model number. When the internal self-check completes, the display shows the Home screen with all values blank, and the device is ready for operation.
If disconnected from an external power source, the ProBP 3400 automatically powers off when not used for 30 minutes. When the device is connected to external power, the device will automatically power off after three hours of non-use.
If a system error is detected, the device displays a system fault message that contains a wrench icon and a system fault code to assist service and engineers in diagnosing the problem.
Page 40
Startup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
34
Page 41

Standby mode

The standby mode conserves battery power. The display goes black, then enters standby mode two minutes after a reading or last button pressed.
Press any button to bring the ProBP 3400 out of Standby Mode.
If disconnected from an external power source, the ProBP 3400 automatically powers off when not used for 30 minutes. When the device is connected to external power, the device will automatically power off after three hours of non-use.
35
Page 42
Standby mode Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
36
Page 43

Blood pressure measurement

The preferred blood pressure measurement site for adults and children is the upper arm. Keep the patient’s arm relaxed and motion-free during measurement(s).
WARNING Inaccurate measurement risk. Do not place the cuff where it can disturb proper circulation. Do not place the cuff on any area where circulation is compromised or on any extremity used for intravenous infusions.
37
WARNING The blood pressure cuff must be properly positioned to ensure blood
pressure accuracy and patient safety. Wrapping the cuff too loosely (preventing proper inflation) may result in inaccurate NIBP readings.
WARNING Do not allow a blood pressure cuff to remain on the patient more than 3 minutes when inflated above 15 mmHg. Excessive cuff tightness may cause venous congestion, peripheral nerve injury, discoloration of the limb, and patient distress.
WARNING ProBP 3400 is not intended to measure blood pressure on neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28 days or less of age if born at term (37 weeks gestation or more); otherwise up to 44 gestational weeks.
To initiate blood pressure measurements:
1. Press the Power on/power off button to turn the device on.
2. Properly size the blood pressure cuff and position it around the patient's bare upper arm with the artery index marker over the brachial artery. Leave room between the cuff and the arm for no more than two fingers.
3. From the Home screen, press the Blood Pressure Start/Stop button.
The ProBP 3400 inflates the blood pressure cuff to the appropriate pressure and displays the pressure as the blood pressure cuff is deflating.
If your device contains the SureBP feature (captures blood pressure on inflation): From the Home screen, press the Blood Pressure Start/Stop button. The ProBP 3400 inflates the cuff to the appropriate level, measuring the blood pressure as the cuff is inflating. The systolic display shows the pressure in the cuff as the blood pressure determination is in process. If the device is unable to determine a blood pressure while the cuff is inflating due to patient movement, excessive noise, or an arrhythmia, the device will use the Step algorithm to inflate the cuff to a higher pressure, then attempt to measure the blood pressure while deflating the cuff.
Pressing the Blood Pressure Start/Stop button at any time during a blood pressure determination aborts the measurement and rapidly deflates the cuff.
Page 44
38
Blood pressure measurement Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
When complete, the ProBP 3400 device displays systolic, diastolic, pulse rate measurements, and if enabled, MAP calculation.
Page 45

Settings

39
1. From the Home screen, press the Right navigation button. Settings is highlighted.
2. Press the Select button. The Settings menu displays.
3. Use Up or Down navigation buttons to highlight the Unit of measure, Pressure preset, Bluetoothradio, or Advanced menus.
Note The Bluetooth radio menu option is only visible for devices that
contain the Bluetooth licensed feature. The NIBP algorithm is only visible for devices that contain the SureBP licensed feature.

Settings matrix

Settings > Unit of measure

Unit of measure

To select a unit of measure:
Pressure presets
Bluetooth radio
Advanced > Cycle count
Data management
Date
Time
MAP
NIBP algorithm
Language
1. From the Settings screen, use Up or Down navigation buttons to highlight Unit of measure.
2. Press the Select button. The Unit of measure menu displays.
3. Use Up or Down navigation buttons to highlight mmHg or kPa.
4. Press the Select button to select the unit of measure.
Page 46
40
Settings Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
The default unit of measure is mmHg.

Pressure presets

To select a pressure preset:
1. From the Settings screen, use Up or Down navigation buttons to highlight Pressure presets.
2. Press the Select button. The Pressure presets menu displays.
3. Use Up or Down navigation buttons to highlight desired pressure.
4. Press the Select button to select the pressure.
The default inflation pressure is 160 mmHg or 21.3 kPa.

Bluetooth wireless technology

To wirelessly transfer data from the ProBP 3400 to another device with Bluetooth technology, you must enable Bluetooth technology in both devices, then pair and connect the devices. Pairing will create a unique wireless link between two Bluetooth wireless technology enabled devices.
The default setting for Bluetooth radio is Enable.

Enable Bluetooth radio and pair devices

When trying to pair the ProBP 3400 with a target device, keep the devices within several meters or feet of each other.
To enable the Bluetooth radio and pair your Bluetooth wireless technology­enabled devices:
1. From the Settings screen, use Up or Down navigation buttons to highlight Bluetooth radio.
2. Press the Select button.
3. Use the Up or Down navigation buttons to highlight Enable.
4. Press the Select button to confirm your selection.
5. Press the Down navigation button to highlight the Pair button.
6. Press the Select button to place the ProBP 3400 into Pairing mode.
7. Activate the Bluetooth feature on your target device and search for the ProBP 3400.
Consult your target device's user manual for device-specific instructions.
8. Select the ProBP 3400 from the list of devices found by your target device.
9. Enter the PIN (1234) to pair the ProBP 3400 to your target device.
The ProBP 3400 supports the Secure Simple Pairing feature, which allows the device to be paired to a target device without requesting a PIN. This feature is available for target devices that are compliant with Bluetooth version 2.1 or higher.
10. When pairing is complete, the ProBP 3400 will automatically connect to your target device and will attempt to reconnect to your target device each time the ProBP 3400 is powered on.
When devices are connected, the symbol appears in the left corner of the ProBP 3400 screen.
Page 47
Directions for use Settings 41

Connect the ProBP 3400 to a target device

When the ProBP 3400 is paired to a target device, it will attempt to automatically reconnect each time you turn on the ProBP 3400. If the devices do not connect within 60 seconds of turning on the ProBP 3400, the Bluetooth radio will shut down to conserve battery power.
If the ProBP 3400 does not automatically reconnect,
1. Press the Power on/power off button to turn the device off.
2. Ensure that you are within several meters or feet of the target device.
3. Press the Power on/power off button to turn the device on.

Advanced settings

To view or modify advanced settings:
1. From the Settings screen, press the Up or Down navigation buttons to highlight the Advanced menu.
2. Press the Select button. The Advanced menu displays.
3. Use Up or Down navigation buttons to highlight the Cycle count, Data management, Date, Time, MAP or Language screens.

Cycle count

To view device cycle count:
1. From the Advanced screen, use Up or Down navigation buttons to highlight Cycle count.
2. Press the Select button. The total number or readings taken on the device displays.
3. Use the Down navigation button to highlight Close.

Data management

The ProBP 3400 is able to store up to 50 readings in its internal memory. The Data management function allows you to manage how additional readings are stored.
1. From the Advanced screen, use Up or Down navigation buttons to highlight Data management.
2. Press the Select button.
3. Use the Up or Down navigation buttons to highlight Auto overwrite or Ask overwrite.
If Auto overwrite is selected, the device will overwrite previous readings, starting with Reading 1, once internal memory is full. If Ask overwrite is selected, the user will have to confirm the save of each additional reading once the memory is full.
4. Press the Select button to confirm your choice.
The default data management setting is Auto overwrite.
Page 48
42
Settings Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Select date

To modify the date shown on the Home screen:
1. From the Advanced screen, use Up or Down navigation buttons to highlight Date.
2. Press the Select button to edit the entry.
3. The Date format field is highlighted. Press the Up navigation or Down navigation buttons to access the correct format option.
4. Press the Select button to accept the date format as shown.
5. Press the Down navigation button to highlight the date.
6. Press the Select button. The first numeric field is highlighted for editing.
7. Use the Up navigation button to increase the number value; use the Down navigation button to decrease the number value.
8. Press the Right navigation button to move to the next field.
9. Repeat steps 7 and 8 to edit additional numeric values.
10. Press the Right navigation button to highlight the OK button.
11. Press the Select button to accept the entry.

Select time

MAP
To modify the time shown on the Home screen:
1. From the Advanced screen, use Up or Down navigation buttons to highlight Time.
2. Press the Select button to edit the entry.
3. The Time format field is highlighted. Press the Up navigation or Down navigation buttons to access the correct format option.
4. Press the Select button to accept the time format as shown.
5. Press the Down navigation button to highlight the time.
6. Press the Select button. The first numeric field is highlighted for editing.
7. Use the Up navigation button to increase the number value; use the Down navigation button to decrease the number value.
8. Press the Right navigation button to move to the next field.
9. Repeat steps 7 and 8 to edit additional numeric values.
10. Press the Right navigation button to highlight the OK button.
11. Press the Select button to accept the entry.
1. From the Advanced screen, use Up or Down navigation buttons to highlight MAP.
2. Press the Select button.
3. Use the Up or Down navigation button to highlight On or Off.
4. Press the Select button to confirm your choice.
The default MAP setting is On.
Page 49
Directions for use Settings 43

NIBP algorithm

To modify the NIBP algorithm used on the device:
1. From the Advanced screen, use Up or Down navigation buttons to highlight NIBP algorithm.
2. Press the Select button to edit the entry.
3. Use the Up or Down navigation buttons to highlight SureBP or Step.
4. Press the Select button to accept the entry as shown.
If SureBP is selected, the device inflates the cuff to the appropriate level, measuring the blood pressure as the cuff is inflating. The systolic display shows the pressure in the cuff as the blood pressure determination is in process. If Step is selected, the device will inflate the cuff, then measure the blood pressure while deflating the cuff. If the device is unable to determine a blood pressure while the cuff is inflating due to patient movement, excessive noise, or an arrhythmia, the device will use the Step algorithm to inflate the cuff to a higher pressure, then attempt to measure the blood pressure while deflating the cuff.
The default NIBP algorithm is SureBP.

Language

To change device language:
1. From the Advanced screen, use Up or Down navigation buttons to highlight Language.
2. Press the Select button. The Language screen displays.
3. Use the Up or Down navigation buttons to highlight the correct Language option.
4. Press the Select button to choose a language.
A popup confirmation screen will display your language selection. The default language is English.
Page 50
Settings Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
44
Page 51

Data review

Data is accessed through the Review menu. Data can be identified by number, date, time reading was taken, diastolic/systolic parameters, MAP, and pulse rate data captured at the time of the reading.

Review data

1. From the Home screen, press the Left navigation button. The Review menu option is highlighted.
2. Press the Select button. Review menu displays.
3. Use Up or Down navigation buttons to review the desired reading.
45

Delete data

1. From the Home screen, press the Left navigation arrow. The Review menu option is
2. Press the Select button. Blood pressure, pulse rate, MAP data (if enabled), date, and time
3. Use Up or Down navigation buttons to highlight desired reading.
4. Press the Select button to delete selected reading. A popup confirmation screen will appear;
5. Press the Select button to confirm deletion of selected or all readings or press the Blood
highlighted.
displays.
use Up or Down navigation buttons to highlight deletion of the selected reading, deletion of all readings, or to exit the screen without saving changes.
Pressure Start/Stop button to exit without saving your changes.
Page 52
Data review Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
46
Page 53

Maintenance and service

Calibrate the device

Welch Allyn recommends annual calibration of the ProBP 3400.
To calibrate the ProBP 3400,
Use the Welch Allyn Service Tool.
Send the device to a Welch Allyn Service Center near you.
47

Change the battery

WARNING Defective batteries can damage the device. If the battery shows any signs of damage, leakage, or cracking, it must be replaced immediately, and only with a battery recommended for or supplied with the device.
WARNING Improper disposal of batteries may create an explosion or contamination hazard. Never dispose of batteries in refuse containers. Do not dispose of the battery in fire. Always recycle batteries according to local regulations.
WARNING Improper handling of the battery can lead to heat generation, smoke, bursting, or fire.
WARNING Do not remove the label from the battery.
WARNING Do not disassemble, modify, or solder the battery.
WARNING Do not directly connect or short circuit the positive (+) and negative (-)
battery terminals.
WARNING To avoid short circuits, keep battery terminals away from metal objects.
WARNING Do not expose the battery to temperatures higher than 80 °C/176 °F.
WARNING Electric shock hazard. Do not open the device or attempt repairs.
There are no user-serviceable parts inside ProBP 3400 other than battery replacement. Only perform routine cleaning and maintenance procedures specifically described in this directions for use. Inspection and servicing of internal parts shall only be performed by qualified service personnel.
The ProBP 3400 device is powered by a lithium-ion rechargeable battery.
Page 54
Maintenance and service Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
48
To remove the battery:
1. Remove the screw from the battery compartment door with a Phillips screwdriver.
2. Remove the battery compartment door.
3. Remove the battery from the battery compartment.
4. Remove the new battery from the shipping bag.
5. Using the polarized guidelines on the battery label, insert the battery, positive (+) side first, into the battery compartment.
6. Replace the battery compartment door.
7. Secure the screw with the screwdriver.
If the ProBP 3400 will be unused for a long period of time, remove the battery prior to storing the device.

Clean the device

WARNING Electric shock hazard. Before cleaning the device, disconnect the power cord from the power source and the device.
WARNING Take care to prevent water or other fluid from entering any connectors on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions.
Caution The device is not heat-resistant. Do not autoclave.
The following agents are compatible with the device:
CaviWipes
Sani-Cloth®Plus
70 percent isopropyl alcohol
10 percent chlorine bleach/90 percent water solution
Note Clean the device on a routine basis, according to your facility's protocols and
standards or local regulations.

CaviWipes or Sani-Cloth Plus

Follow the cleaning agent manufacturer's guidelines.
1. Using CaviWipes or Sani-Cloth Plus, wipe the surface of the device to remove all debris.
2. Allow the device surface to dry for a minimum of 10 minutes before using the device.

70 percent isopropyl alcohol

Wipe the device with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
Page 55
Directions for use Maintenance and service 49

10 percent chlorine bleach/90 percent water solution

1. Wipe the device with a clean cloth slightly dampened with a 10 percent bleach and 90 percent water solution.
2. Rinse with a clean cloth slightly dampened with clean water.
3. Allow the device surface to dry for a minimum of 10 minutes before using the device.

Clean the accessories

Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent solution.
The same cleaning agents used to clean the device can be used on the stand and mounting accessories.
Note Clean the device on a routine basis, according to your facility's protocols and
standards or local regulations.
Page 56
Maintenance and service Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
50
Page 57

Specifications

Physical specifications

Performance
This section describe normal ranges for the ProBP 3400 device.
Blood Pressure Accuracy
Blood pressure accuracy meets or exceeds ANSI.AAMI SP10:2002 standards for noninvasive blood pressure accuracy (±5 mmHg mean error, 8 mmHg standard deviation). Blood pressure accuracy is validated for pressure measurement using the upper arm only.
51
Cuff pressure range 0 to 300 mmHg
Systolic range 60 to 250 mmHg
Diastolic range 30 to 160 mmHg
Mean Arterial Pressure (MAP) range MAP is a calculated reading that yields an approximate
value.
Pulse rate range 35 to 199 bpm
Pulse rate accuracy ±5.0%
Overpressure cutoff 300 mmHg ±15 mmHg

Mechanical specifications

Dimensions
Weight 450 g (0.99 lb)
40 to 190 mmHg
Height: 15.0 cm (5.91 inches)
Width: 8.0 cm (3.15 inches)
Depth: 5.6 cm (2.20 inches)
Mounting
Portability May be used as a handheld device
Custom mobile stand
Custom wall mount
Custom desk mount
Page 58
Specifications Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
52

Electrical specifications

Power requirements:
Degree of protection: Type BF applied part
Safety classification: Class II
Internally powered: Battery, lithium-ion type, 3.7V, 2100mAh, 7.8Wh
Protection against the ingress of water:
Safety mode of operation: Continuous operation
Standards:
Input: 100-240VAC, 0.18A, 50-60Hz
Output: 5VDC, 0.5A
IPXØ
This device complies with the following standards:
EN/IEC 60601-1
EN/IEC 60601-1-2
EN 1060-1:1996 Specification for non-invasive sphygmomanometers - Part 1: General requirements
EN 1060-3:1997 Specification for non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
FCC Subpart 15C
FCC ID #PI4411B, IC 1931B-BTM411

Environmental specifications

WARNING Fire and explosion hazard. Do not operate the device in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen­enriched environments; or in any other potentially explosive environment.
Caution Use the device within stated operating temperature ranges. The device will not meet performance specifications if used outside these temperatures ranges.
Operating temperature 10°C to 40° C (50° to 104° F)
Storage temperature -20°C to 50° C (-4° to 122° F)
Operating altitude -170 to 4877 m (-557 to 16,000 ft.)
Storage humidity 15 to 95% (noncondensing)
Page 59
Directions for use Specifications 53

Product disposal

Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not apply.
For more specific disposal information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service at +44 207 365 6780.
Page 60
Specifications Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
54
Page 61

Standards and compliance

General radio compliance

The Bluetooth Wireless Card must be used in strict accordance with the manufacturer’s instructions as described in the user documentation that comes with the product.
This device complies with Part 15 of the FCC rules and with the rules of the Canadian ICES-003. Operation is subject to the following two conditions: (1) This device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
55

Federal Communications Commission (FCC)

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
This device may not cause harmful interference.
This device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If not installed and used in accordance with the instructions, it may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna
Increase the distance between the equipment and the receiver
Connect the equipment to an outlet on a circuit different from that to which the receiver is connected
Consult the dealer or an experienced radio/TV technician for help
The user may find the following booklet prepared by the Federal Communications Commission helpful:
The Interference Handbook
This booklet is available from the U.S. Government Printing Office, Washington, D.C. 20402. Stock No. 004-000-0034504.
Page 62
56
Standards and compliance Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
Welch Allyn is not responsible for any radio or television interference caused by unauthorized modification of the devices included with this Welch Allyn product, or the substitution or attachment of connecting cables and equipment other than specified by Welch Allyn.
The correction of interference caused by such unauthorized modification, substitution, or attachment will be the responsibility of the user.

Industry Canada (IC)

This device complies with RSS 210 of Industry Canada.
Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of this device.
L’utilisation de ce dispositif est autorisée seulement aux conditions suivantes: (1) il ne doit pas produire de brouillage et (2) l’ utilisateur du dispositif doit étre prêt à accepter tout brouillage radioélectrique reçu, même si ce brouillage est susceptible de compromettre le fonctionnement du dispositif.
This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe B est conform à la norme NMB-003 du Canada.

European Union

The Bluetooth radio contained within this equipment complies with the essential requirements of the European Union R&TTE Directive (1999/5/EC).
Page 63

Guidance and manufacturer's declaration

Emissions and immunity information

Electromagnetic emissions
57
The device is intended for use in the electromagnetic environment specified below. The customer or user of the device should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
a
The device contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz frequency hopping spread-spectrum transmitter for the purpose of wireless communication. The radio is operated according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/ EC). The transmitter is excluded from the EMC requirements of 60601-1-2:2007, but should be considered when addressing possible interference issues between this and other devices.
Group 1 The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
Class B
Complies
The device is suitable for use in all establishments other than domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes, provided the following warning is heeded:
WARNING This equipment/system is intended for use by healthcare professionals only. This equipment/ system may cause radio interference or may disrupt the operation of nearby equipment a. It may be necessary to take mitigation measures, such as re-orienting or relocating the device or shielding the location.
Page 64
Guidance and manufacturer's declaration Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
58
Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
±6 kV contact ± 8 kV air
±2 kV for power supply lines
±1 kV for input/ output lines
± 1 kV differential mode
± 2 kV common mode
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
Compliance level Electromagnetic environment -
guidance
± 6 kV contact ± 8 kV air
±2 kV for power supply lines
±1 kV for input/ output lines
± 1 kV differential mode
± 2 kV common mode
>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5 seconds
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.
Electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 1 GHz
Compliancelevel Electromagnetic environment -
guidance
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
3 Vrms
3 V/m
d= (1.17)
d= (1.17)
d= (2.33) 800 MHz to 2,5 GHz
80 to 800 MHz
Page 65
Directions for use Guidance and manufacturer's declaration 59
Electromagnetic immunity
whereP is the maximum output power rating of the transmitter in watts (W) andd is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey a , should be less than the compliance level in each
frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated max. output power of transmitter (W)
150 kHz to 80 MHz
d= (1.17)
0.01 0.11667 0.11667 0.23333
0.1 0.36894 0.36894 0.73785
1 1.1667 1.1667 2.3333
10 3.6894 3.6894 7.3785
100 11.667 11.667 23.3333
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, whereP is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
80 MHz to 800 MHz
d= (1.17)
800 MHz to 2.5 GHz
d= (2.23)
Page 66
Guidance and manufacturer's declaration Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
60
Recommended separation distances between portable and mobile RF communications equipment and the device
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Page 67

Warranty

Welch Allyn will warranty the ProBP 3400 device and battery to be free of defects in material and workmanship and to perform in accordance with manufacturer specifications for the period of one year from the date of purchase from Welch Allyn or its authorized distributors or agents.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date specified during product registration, 3) the date of purchase of the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
61
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
The product warranty is also subject to the following terms and limitations:
Accessories are not covered by the warranty. Refer to the Directions For Use provided with individual accessories for warranty information.
Shipping cost to return a device to a Welch Allyn Service center is not included.
A service notification number must be obtained from Welch Allyn prior to returning any products or accessories to Welch Allyn's designated service centers for repair. To obtain a service notification number, contact Welch Allyn Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.
Page 68
Warranty Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
62
Page 69

Troubleshooting

Inaccurate blood pressure readings

Possible cause Corrective action and explanation
63
Incorrect cuff size
Patient’s arm position Ensure patient’s arm is at heart level.
Arm movement during blood pressure cycle
Blood pressure taken over clothing Take blood pressure on a bare arm.
Arrhythmia
Change in blood pressure between auscultatory reading and ProBP 3400 reading
Incorrect reference Use the correct Korotkoff sound to determine diastolic
Use Welch Allyn approved cuffs only. Measure patient’s arm circumference midway
between elbow and shoulder (see “Blood pressure cuff selection” to select correct cuff size).
Keep arm still during blood pressure cycle.
Movement may cause inaccuracies from artifact.
Check for regularity of heart rate (palpate pulse or check device).
Moderate to severe heart rate irregularities may make blood pressure difficult to measure accurately.
Check blood pressure immediately before ProBP 3400 reading.
Blood pressure is dynamic and changing. It is normal for blood pressure to fluctuate 5 to 10 mmHg.
blood pressure.
Many listeners incorrectly equate diastolic blood pressure with the disappearance of sound only (phase 5). ProBP 3400 was developed using the American Heart Association recommendations, which state that phase 5 be used unless sound continues to 0 mmHg, in which case the change in the quality of sound (phase 4) is to be used.
Deflate cuff no faster than 3 mmHg per second.
One of the major sources of error in auscultatory blood pressure measurement is deflating the cuff too quickly. The American Heart Association
Page 70
Troubleshooting Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
64
Possible cause Corrective action and explanation
recommends deflation no faster than 3 mmHg per second.
Only use a sphygmomanometer that is calibrated.
An uncalibrated sphygmomanometer may result in inaccurate blood pressure measurements.
Poor auscultatory sound recognition by observer Use high-quality stethoscope. Have a different
observer check patient’s blood pressure.

Cuff inflation and deflation with no blood pressure reading displayed

Possible cause Corrective action and explanation
Leak in pneumatic system
Arm movement during cycle
Cuff tubing or pressure hose movement artifact
Inflation too quick; check NIBP cuff and tubing connections may display.
Cuff pressure limits exceeded. Powering down may display.
NIBP air leak; check cuff and tubing connections may display.
Unable to determine NIBP; check inflation settings may display.
Ensure all cuff attachments are tight. Carefully check for leaks in the blood pressure cuff,
tubing, and pressure hose attached to ProBP 3400.
Unable to determine NIBP; check connections; limit patient movement may
display.
Keep arm still during blood pressure cycle. Movement may cause inaccuracies from artifact, long
cycle times, and error message.
Unable to determine NIBP; check connections and tubing for kinks may display.
Do not contact cuff tubing or pressure hose during blood pressure cycle.
Movement may cause inaccuracies from artifact.
User may have pressed the Blood Pressure
Start/Stop button

No cuff inflation

Possible cause Corrective action and explanation
Connections between device and cuff loose Check all connections.
User cancelled NIBP reading will display. Press the Blood Pressure Start/Stop button to initiate blood pressure measurement.
Page 71
Directions for use Troubleshooting 65
Possible cause Corrective action and explanation
Device is being used in inclimate environment
Internal errors or messaging error occurred

Cuff pops off

Possible cause Corrective action and explanation
Inappropriate cuff size See “Blood pressure cuff selection” to select correct
Cuff not applied securely Smooth hook and loop securely before inflating cuff.
Cuff applied inside out Re-apply cuff. Verify that the Welch Allyn label is

Cuff deflating too slowly

Ambient temperature is outside of operating range. Retry measurement may display.
Operate the device within temperature range in the Environmental specifications section.
NIBP feature not functional. Call for service
may display. Contact a Welch Allyn Service Center.
cuff size. If cuff continues to pop off, notify biomedical department or Welch Allyn Technical Support.
facing away from arm.
Possible cause Corrective action and explanation
Patient movement
Arrhythmia
Small leak in pneumatic system Check cuff tubing and pressure hose for leaks.
Have patient sit still.
Do not have arm tight against chest wall, as respiration may affect speed and accuracy of blood pressure measurement.
Check for regularity of heart rate (palpate pulse or check device).
Moderate to severe heart rate irregularities may make blood pressure difficult to measure accurately.

Cuff inflates beyond SureBP inflation pressure preset while device is in use

Possible cause Corrective action and explanation
Device switched from SureBP to Step algorithm If the device is unable to determine a blood pressure
while the cuff is inflating due to patient movement, excessive noise, or an arrhythmia, the device will use
Page 72
Troubleshooting Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
66
Possible cause Corrective action and explanation
the Step algorithm to inflate the cuff to a higher pressure, then attempt to measure the blood pressure while deflating the cuff.
Arm movement during blood pressure cycle
Blood pressure taken over clothing Take blood pressure on a bare arm.
Arrhythmia

Device does not turn on

Possible cause Corrective action and explanation
Low battery
Battery installed incorrectly
Device not powering up Unplug the device from wall receptacle and check for
Keep arm still during blood pressure cycle.
Movement may cause inaccuracies from artifact.
Check for regularity of heart rate (palpate pulse or check device).
Moderate to severe heart rate irregularities may make blood pressure difficult to measure accurately.
Plug in the device. Check connections between the ProBP 3400 and the transformer, then between the transformer and wall receptacle.
Using the polarized guidelines on the battery label, insert the battery, positive (+) side first, into the battery compartment.
breaks in the cord. If connections are secure, check electrical outlet for power. Charging indicator is on if connections are good and the device is plugged into a working outlet.
Plug the device into a known working electrical outlet.
The device may not power up if the battery is completely discharged. Connect the device to an external power source for at least 15 minutes before attempting to power up again. If device powers up, charge the battery for at least six hours before disconnecting the device from the power supply.
Replace the battery.
Notify Biomedical department or Welch Allyn Technical Support.
System error The device displays a system fault message that
contains a wrench icon and a system fault code
to assist service and engineers in diagnosing the problem.
Page 73
Directions for use Troubleshooting 67

Bluetooth troubleshooting

ProBP 3400 and the target device will not pair

Possible cause Corrective action and explanation
Bluetooth radio is not enabled on the ProBP 3400
ProBP 3400 is not in Pairing mode
There is too much distance between ProBP 3400 and the target device
The wrong PIN was entered for a Bluetooth 2.0 connection
There is a problem with the target device Refer to operating system or computer hardware
See Bluetooth wireless technology section for instructions on enablingBluetooth functionality in the device.
See Bluetooth wireless technology section for instructions on enabling pairing in the device.
Move the ProBP 3400 closer to the target device.
If a PIN is requested, enter 1234.
manuals for further troubleshooting.

ProBP 3400 and the target device are not connected

Possible cause Corrective action and explanation
ProBP 3400 and the target device have not been paired.
Follow instructions for pairing the ProBP 3400 to a target device.
If pairing was attempted:
Complete pairing process again
See troubleshooting section for ProBP 3400 and the target device will not pair.
There is too much distance between ProBP 3400 and the target device
The target device and/or software were not launched in the correct order
There is a problem with the target device Refer to operating system or computer hardware
Move the ProBP 3400 closer to the target device.
The connection process was not completed in the correct order.
1. Exit the application on the computer.
2. Power down, then power up ProBP 3400.
3. Relaunch the computer application.
manuals for further troubleshooting.
Page 74
Troubleshooting Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
68

ProBP 3400 and the target device lost connection

Possible cause Corrective action and explanation
Unexpected loss in Bluetooth communication The ProBP 3400 powered down
There is too much distance between ProBP 3400 and the target device
There is a problem with the target device Refer to operating system or computer hardware
1. Exit the target application.
2. Power down the ProBP 3400
3. Restart the ProBP 3400
4. Relaunch the target application.
Move the ProBP 3400 closer to the target device.
manuals for further troubleshooting.
Page 75

Appendix

ProBP 3400 approved accessories

Flexiport® cuffs (Latex free)

Part Number Model Description
69
Reuse-08 Reusable Cuff, reuse, SM CHILD, 2-tube
Reuse-09 Reusable Cuff, reuse, CHILD, 2-tube
Reuse-10 Reusable Cuff, reuse, SM AD, 2-tube
Reuse-11 Reusable Cuff, reuse, ADULT, 2-tube
Reuse-11L Reusable Cuff, reuse, AD LONG, 2-tube
Reuse-12 Reusable Cuff, reuse, LG AD, 2-tube
Reuse-12L Reusable Cuff, reuse, LG AD LONG, 2-tube
Reuse-13 Reusable Cuff, reuse, THIGH, 2-tube
Soft-08 Disposable Cuff, soft, SM CHILD, 2-tube (box of 5)
Soft-09 Disposable Cuff, soft, CHILD, 2-tube (box of 5)
Soft-10 Disposable Cuff, soft, SM AD, 2-tube (box of 5)
Soft-11 Disposable Cuff, soft, ADULT, 2-tube (box of 5)
Soft-11L Disposable Cuff, soft, AD LONG, 2-tube (box of 5)
Soft-12 Disposable Cuff, soft, LG AD, 2-tube (box of 5)
Soft-12L Disposable Cuff, soft, LG AD LONG, 2-tube (box of 5)
Soft-13 Disposable Cuff, soft, THIGH, 2-tube (box of 5)
Page 76
Appendix Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
70

Blood pressure accessories (Latex free)

Part Number Model Description
3400-30 ProBP 3400 Double tube blood pressure hose (5 ft)
3400-31 ProBP 3400 Double tube blood pressure hose (10 ft)

Mounting options

Part Number Description
4600-61 Mobile stand with basket and mounting kit for ProBP 3400
4601-61 Wall mount with basket for ProBP 3400
4602-61 Desk mount for ProBP 3400

Accessories and replacement parts

Part Number Description
BATT11 Lithium Ion battery, 1 Cell
3400-925 ProBP 3400 USB cable, 2.44 m (8 ft)
3400-926 ProBP 3400 USB cable, 0.30 m (16 inches)
4600-100 5W power supply mounting kit (for use with mobile stand)
3400-561 ProBP 3400 basket adapter with screw
3400-461 ProBP 3400 wall adapter with screw
PWCD-5WW-B Power cord assembly, for handheld, or desk or wall
PWCD-5WT-B Power cord assembly B, for mobile stand mounted ProBP
PWCD-5WW-2 Power cord assembly 2, for handheld, or desk or wall
PWCD-5WT-2 Power cord assembly 2, for mobile stand mounted ProBP
mounted ProBP 3400, North America
3400, North America
mounted ProBP 3400, Europe
3400, Europe
PWCD-5WW-4 Power cord assembly 4, for handheld, or desk or wall
mounted ProBP 3400, United Kingdom
PWCD-5WT-4 Power cord assembly 4, for mobile stand mounted ProBP
3400, United Kingdom
PWCD-5WW-6 Power cord assembly 6, for handheld, or desk or wall
mounted ProBP 3400, Australia/New Zealand —Orange
Page 77
Directions for use Appendix 71
Part Number Description
PWCD-5WT-6 Power cord assembly 6, for mobile stand mounted ProBP
3400, Australia/New Zealand —Orange
PWCD-5WW-C Power cord assembly C, for handheld, or desk or wall
mounted ProBP 3400, China
PWCD-5WT-C Power cord assembly C, for mobile stand mounted ProBP
3400, China
PWCD-5WW-7 Power cord assembly 7, for handheld, or desk or wall
mounted ProBP 3400, South Africa
PWCD-5WT-7 Power cord assembly 7, for mobile stand mounted ProBP
3400, South Africa
PWCD-5WT-J Power cord assembly N, for mobile stand mounted ProBP
3400, Japan
3400-100 Battery door

Service

Part Number Description
S1-3400 Comprehensive Partner Program, ProBP 3400, 1 year
S2-3400 Biomed Partner Program, ProBP 3400, 1 year
S3-3400 Investment Pro Partner Program, ProBP 3400, 1 year
S4-3400 Preventive Partner Program, ProBP 3400, 1 year
103521 Welch Allyn Service Tool CD

Licensed features

Part Number Description
3400-SUREBP SureBP activation code
3400-BT Bluetooth activation code

Literature/Documentation

Part Number Description
Directions for Use
103701 Kit, CD, Directions for Use, ProBP 3400
Page 78
Appendix Welch Allyn Connex® ProBP™ 3400 digital blood pressure device
72
Part Number Description
Startup Guide
407673 Startup Guide ProBP 3400
Directions for Use, Mounting Options
104388 Directions for use, Mounting kit instructions for ProBP 3400
mobile stand
104387 Directions for use, Mounting kit instructions for ProBP 3400
wall mount
104386 Directions for use, Mounting kit instructions for ProBP 3400
desk mount

Configuration options

Model Description
34BFHT-B* Includes Welch Allyn SureBP® technology, Bluetooth® wireless technology,
rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs, handheld configuration
34BFWT-B* Includes Welch Allyn SureBP® technology, Bluetooth® wireless technology,
rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs, wall mounted configuration
34BFST-B* Includes Welch Allyn SureBP® technology, Bluetooth® wireless technology,
rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs, mobile stand mounted configuration
34XFHT-B* Includes Welch Allyn SureBP® technology, rechargeable lithium-ion battery, size
11 and size 12 FlexiPort® cuffs, handheld configuration
34XFWT-B* Includes Welch Allyn SureBP® technology, rechargeable lithium-ion battery, size
11 and size 12 FlexiPort® cuffs, wall mounted configuration
34XFST-B* Includes Welch Allyn SureBP® technology, rechargeable lithium-ion battery, size
11 and size 12 FlexiPort® cuffs, mobile stand mounted configuration
34BXHT-B* Includes Bluetooth® wireless technology, rechargeable lithium-ion battery, size 11
and size 12 FlexiPort® cuffs, handheld configuration
34BXWT-B* Includes Bluetooth® wireless technology, rechargeable lithium-ion battery, size 11
and size 12 FlexiPort® cuffs, wall mounted configuration
34BXST-B* Includes Bluetooth® wireless technology, rechargeable lithium-ion battery, size 11
and size 12 FlexiPort® cuffs, mobile stand mounted configuration
34XXHT-B* Includes rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs,
handheld configuration
34XXWT-B* Includes rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs, wall
mounted configuration
Page 79
Directions for use Appendix 73
Model Description
34XXST-B* Includes rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs,
mobile stand mounted configuration
*Replace final digit in model number with regional code listed in the table below.

Accessories and replacement parts

Code Region
6 Australia/New Zealand - Orange
C China
2 Europe
J Japan (mobile stand version only)
B North America
7 South Africa
4 United Kingdom
Page 80