The POEM NIBP Module Developer’s Kit provides the hardware, software, and accessories
required to operate a POEM NIBP Module from a Personal Computer (PC). This allows a
developer to control the POEM Module and view its NIBP and diagnostic outputs without having
to develop hardware and software interfaces to a host system. This version of the Developer’s Kit
includes:
•POEM Module Interface Box which includes:
• POEM Module and pump mounted inside for physical protection, acoustical damping
• Interface Board mounted on top for connections to power and data communications
• External NIBP hose fitting
• Interconnect ribbon cable between Module and Interface Board (PN 660-0164-XX)
•CD with POEM PC and POEM Talk Software (for PC) and manuals.
•NIBP Cuff (adult, reusable)
•NIBP Hose (10 ft.)
•RS232 Cable for connecting Interface Board to PC (PN 660-0163-XX)
•Sync Cable for connecting source for ECG synchronization signal (PN 660-0178-XX)
•9 Vdc international power adapter (with interchangeable mains plugs) operating on
nominal 100-240 Vac, 50/60 Hz (PN 503-0122-00).
•POEM NIBP Module OEM Implementation Manual
- Provides detailed information
about the POEM NIBP Module and how to integrate it into a host system.
Information furnished by Welch Allyn OEM Technologies is believed to be accurate and reliable. However, no
responsibility is assumed by Welch Allyn OEM Technologies for its use, nor for any infringements of patents or other
rights of third parties which may result from its use. No license is granted by implication or otherwise under any
patent or patent rights of Welch Allyn OEM Technologies.
Disclaimers: This document may be wholly or partially subject to change without notice. All rights are reserved. No
one is permitted to reproduce or duplicate, in any form, the whole or part of this document without written permission
from Welch Allyn OEM Technologies.
POWER CONNECTOR (FOR POWER ADAPTER OR “BARREL
CONNECTOR PIGTAIL CABLE)
POWER CONNECTIONS
(FOR REGULATED EXTERNAL POWER SUPPLY)
CABLE TO PC
(WITH POEM PC OR
POEM TALK SOFTWARE
INSTALLED)
CONNECTION FOR OPTIONAL ECG
SYNCHRONIZATION (REQUIRED FOR SMARTCUF)
After connection, the NIBP Module is controlled from the PC using the POEM PC or POEM Talk
software program. The developer has access to all POEM Module operating modes and patient
modes.
If desired, the Developer’s Kit may be partially or totally connected to the host’s target system for
convenience in developing code in the host system. The Interface Board’s power supply inputs or
logic-level serial interface may be directly connected to the host target system.
Interface Board
All electrical connections to the POEM Module are made through the Interface Board mounted
on the top of the Interface Box. The Interface Board provides:
•Two possible power connections (only one used at a time):
• J1: 5-contact terminal strip to connect to HOST_5V, HOST_3.3V, VBUS, and GND
• J2: Barrel-style power input connector which permits powering the POEM Module
from a single, regulated or non-regulated, dc-output power adapter (or from a single
regulated dc supply)
•5V linear regulator (used only with power adapter input)
•F1: Fuse in series with VBUS
•P1: Jumpers to configure the polarity for the power adapter input
•D3: Power indicator LED lights whenever 5 V power is present
•S1: Reset switch; pressing S1 resets the POEM Module
•J4: A 0.1-inch grid pad arrangement for either soldering wires to or adding a
connector for access to all Module connections
•RS232 level translators for RXD and TXD
•J6: A female DB9 connector for direct connection to a PC’s serial port
•J5: A 1x6 connector for coupling ECG_SYNC from the host system.
The Interface Board includes safety features to protect the Interface Board and the POEM
Module from damage that might be caused by inappropriate supply voltages, polarities, or power
sequencing.
Power Requirements
Simplified power connections are described below. For additional information, refer to the POEM
NIBP Module OEM Implementation Manual.
METHOD 1 - Regulated or Non-Regulated Power Adapter (or Pigtail)
Connection at J2
The power adapter and configuration requirements are shown in Figure 2.
1 2
DC Output
6-9 V
V x A ≥ 6
POWER
ADAPTER
AC Input
P1
3 4
JMPR2 JMPR1
J2
Input Connector: Must be “barrel-type” with
0.080-inch diameter center pin, and 0.215-inch
1
1
diameter outer sleeve.
Configure P1 for the appropriate polarity of the
adapter input connector:
For negative center pin:
For positive center pin (factory default):
Configure JMPR1 and JMPR2 as shown (factory
default setting)
1 2
3 4
1 2
3 4
J1
Leave J1 unconnected
Figure 2 - Power Adapter Configuration
A universal power adapter is included in the Developer’s Kit. It meets the requirements described
in Figure 2 (with positive center pin connector, 0.080-inch center pin, 0.215-inch diameter outer
sleeve connector). One of the included mains connectors will be appropriate for almost every
country in the world. Alternatively, a barrel-connector pigtail cable may be ordered from Welch
Allyn OEM Technologies. Contact Welch Allyn OEM Technologies for more information.
CAUTIONTo avoid possible equipment damage, do not use any ac-output power
adapter, or any dc-output power adapter that does not meet the requirements
described in these instructions.