Welch Allyn PneumoCheck User manual

Service Manual
Operator Manual
Cardio Menu
PneumoCheck™
Spirometer Model # 61000,
Printer Charger #’s 76100 & 76106
Part 610132
Welch Allyn Medical Division
4341 State Street Road
P.O. Box 220
Skaneateles Falls, NY 13153-0220
.oNtraP.veRegnahCfonoitpircseD#NCEetaDdevorppA
231016D noitamrofnIgnitoohselbuorTedulcniotnwardeRdnadesiveR 09563-579tcOerbmetteS.R 231016E 2noitceSnitseThcneBregrahC/retnirPdeddA 52324-500.voNorreLnarF
DRAWINGS AND/OR ILLUSTRATIONS AND/OR PART NUMBERS CONTAINED IN THIS DOCUMENT ARE FOR REFERENCE PURPOSES ONLY! FOR CURRENT REVISIONS CALL WELCH ALLYN.
Service Manual 610132 Rev. E Pneumocheck Handle and Printer/Charger 1
2 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Contents
General Information........................................................................................................... 5
1.1 Handheld Spirometer ................................................................................................................6
1.2 Printer .........................................................................................................................................8
Service ...............................................................................................................................11
2.1 Bench Test Of Pneumocheck.................................................................................................12
2.2 Bench Test of Printer Charger................................................................................................17
2.3 Required Tools and Equipment .............................................................................................21
2.4 Test and Adjustment Procedures ...........................................................................................22
Theory of Operation ..........................................................................................................23
3.1 Spirometer Circuit Description ..............................................................................................24
3.2 Spirometer - Printer Circuit Description ...............................................................................30
3.3 Messages.................................................................................................................................33
Troubleshooting................................................................................................................37
4.1 PneumoCheck Handle Technical Troubleshooting ............................................................. 40
4.2 Customer Service Questions and Answers.........................................................................43
Disassembly and Repair...................................................................................................49
5.1 Pneumocheck Handle Disassembly Procedures ................................................................ 50
5.2 Pneumocheck Printer/Charger Disassembly Procedures ................................................. 56
Pneumocheck Handle and Printer/Charger Spare Parts ...............................................65
List of Drawings and A Specs ..........................................................................................67
Service Manual 610132 Rev. E Pneumocheck Handle and Printer/Charger 3
Help Information:
All service and repairs must be performed by authorized Welch Allyn personnel or
agents, using Welch Allyn replacement parts. Failure to do so will invalidate the product warranty.
If you have any questions about servicing the 61000 Pneumocheck or the 76100 or 76105 Printer Chargers, please call or write Welch Allyn:
Attn: Product Service Department
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220
U.S.A.
Telephone: 800-535-6663
Fax: 315-685-4653
Troubleshooting assistance is contained in Section 4 of this manual.
Please read and understand the PneumoCheck and Printer/Charger Operating
Instructions PN 610140-3.
4 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 1 - General Information
1.1 Basic Description - Handheld Spirometer
1.2 Basic Description - Printer
General Information
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 5
Section 1 - General Information
1.1 Handheld Spirometer
Refer to Owners Guide
P/N 610140-3
The handheld spirometer is a small, battery operated device incorporating a laminar flow sensing pneumatic liquid crystal display, and electronic circuitry in an injection molded case.
The front panel of the handheld spirometer includes the eight character, alphanumeric display, two push-button keys labeled TEST and DATA, an ON/OFF slide switch and a battery charge indicator light. The flow sensor is housed in the top portion of the device and is removable for cleaning and sterilization. The mouth piece is at one end of the flow sensor and fits disposable cups or disposable cardboard tubes. There are two electrical contacts on each side to the spirometers batteries. On one side of the spirometer is a DC jack for battery charging in case the operator chooses not to purchase the optional printer base.
The main function of the handheld spirometer is to gather data from two kinds of spirometry maneuvers by patients: the FVC maneuver and the MVV maneuver.
The TEST key is used to select and perform the desired test. It is normally pressed when the spirometer is displaying the word SELECT. It can also be pressed during a maneuver or during the results display to prematurely quit a test and begin another.
The DATA key is used to continue whenever the device pauses to display some information. When the word SELECT is being displayed, the DATA key can be used to show the results of the test that is currently stored in memory.
The general flow of usage of the device starts with the selection of a test. The operator then waits for an audible beep, indicating that it is ready to acquire data from the maneuver. The patient puts the mouth piece to his mouth and performs the maneuver. Upon completion indicated by another audible beep, there is a brief delay while the results are calculated then the result messages are displayed. This completes the loop, and the machine goes back to SELECT and waits for another keypress.
6 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
6 Pneumocheck Handle and Printer/Charger
Section 1 - General Information
When not in use, the handheld spirometer sits-in the cradle of the printer base (or it is connected to an outlet via a DC power adapter,) and the battery is charged. The battery indicator light remains on whenever a source of charge current is present. The spirometer becomes operational when it is lifted from the printer base or disconnected from the DC adapter. The spirometer is functional only while disconnected from the charger or printer base and the ON/OFF switch is ON. The word SELECT appears on the display and the device waits for a keypress from the operator.
When the word SELECT is on the display, you can do one to the following:
1. Press the test key which will start the sequence for doing an FVC test, and MVV test, or a calibration of the spirometer.
2. Press the DATA key which will display the test values of the best test of a test is currently stored in the spirometer’s memory.
3. Put the spirometer back into the printer cradle which will start the sequence for transmitting any test data in the spirometer’s memory to the printer’s memory, and then resume charging the battery.
4. Connect the spirometer to the DC power adapter, which will resume charging the battery.
5. Turn the unit off using the ON/OFF slide switch.
A more detailed description of the operation of the spirometer is contained in the Operator’s Manual.
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 7
Section 1 - General Information
1.2 Printer
The printer base consists of a receptacle to hold the spirometer, a battery charging circuit, a high speed thermal line printer, and associated microprocessor a power circuitry, all housed in an injection molded case.
The front panel of the printer includes 2 indicator lights, and 3 push button keys marked STOP, FEED, and PRINT. Also on the front panel are 5 numeric rotary switches: 2 digits for AGE, and 3 digits for HEIGHT. The are also 2 slide switches for selecting SEX and RACE.
There is a set of 4 option switches located on the underside of the unit.
The main function of the printer base is to print reports of spirometry tests performed by the handheld spirometer, and to charge the spirometer’s battery.
The general flow of usage is as follows: When the spirometer is placed in the cradle of the base,
it immediately begins sending data to the printer base during which time the message SENDING appears on the panel of the printer base. Once the data transfer is complete, the printer base gives a compound beep and continues to charge. The READY light on the printer base turns on solidly, indicating that there is a test stored in its memory. The operator then sets the patient demographic switches for the patient, then presses the PRINT key to print the report.
There are two indicator lights on the front panel of the printer base. One is labeled POWER and is on whenever the unit is plugged into a powered outlet. The other light is labeled READY, in the memory of the printer base. When the device is receiving data from the spirometer, this light flashes. Once the transfer is complete and test is in memory, the light remains on.
8 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 1 - General Information
The READY light is also used to indicate an error condition. Errors are numbered from one to three. When an error occurs, an alarm sounds and the READY light flashes in groups of 1, 2, or 3 flashes (at a rate of times per second), corresponding to the error that was detected. A one second pause occurs between these flashes. This continues until the error condition has been cleared.
The PRINT key is used to print a report. If there is no report to be printed and the PRINT key is pressed, a short chirp will sound and no report is printed.
The FEED key is used to feed paper out of the printer. The paper is fed as long as the key is held down. The key is not operable while a report is being printed or while the data is being transferred from the spirometer.
The STOP key aborts the current task (report printing or data transfer) and returns to the wait state of the base. It is also used to clear the machine of any error condition that it might be in. It does not clear the printer base’s memory.
The demographic switches are used to indicate the patient’s age, height, sex, and race. The AGE switch consists of two digits, each of which can range from 0 to
9. The range of valid values depends on the normal
being used.
The HEIGHT switches operate in the same manner as the AGE switch, except that it has three digits. Either inches or centimeters can be used for the height. As was the case for AGE the range of valid values depend on the normal being used.
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 9
Section 1 - General Information
If the value for either age or height fails outside of valid range, then the normal values are not calculated or printed on the report.
The SEX switch is a slide switch. The up position indicates the patient is a female, while the down position indicates that the patient is male.
The RACE switch is also a slide switch. The up position indicates race group 1 and the down position indicates race group 2. Race group 1 is used if the patient is Black, Oriental, East Indian, or Pakistani. Race group 2 is used for all other races.
There are four option switches located on the underside of the printer base. The switches are dip switches that can be in either the ON or OFF position. These switches can be changed while the printer base is on. The options specify certain aspects of the report printing. They are:
1. OFF
2. OFF Knudson, 1983
1. ON
2. OFF Knudson, 1976
1. OFF
2. ON Morris, Crapo et al
1. ON
2. ON European
3. OFF: Short report form
3. ON: Long report form
4. OFF: Flow/Volume plot
4. ON: Volume/Time plot
These options are discussed in further detail in the Operator’s Manual.
10 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service
2.1 Bench Test of Pneumocheck
2,2 Bench Test of Printer/Charger
2.3 Required Tools and Equipment
2.4 Test and Adjustment Procedures

Service

Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 11
Section 2 - Service
2.1 BENCH TEST OF PNEUMOCHECK
SECTION A
1 Remove pneumotach if necessary.
Caution: Adequate ESD protection must be worn at all times when working on the Pneumocheck, handle or printer/charger.
2 Obtain a good set of batteries and insert in
pneumocheck.
3 Replace pneumotach. Turn pneumocheck on. If
unit powers up continue testing (steps 3 & beyond). If unit doesn’t power up, take apart. Refer to troubleshooting guide.
4 Put into data mode (hold down white button,
turn unit on and release white button).
5 Observe LCD; all segments should be on.
6 Depress and release ‘test’ button. Unit should
step through a self diagnostic. Testing > ROM OK > RAM OK > EEPRM OK> ASLEEP
In ‘ASLEEP’ mode, the unit will stop and await further instruction. If unit stops before sleep mode, there is a failure. Refer to troubleshooting guide.
7 Depress & release test button: ON reading
should be displayed. It should read between 30 and 80. Continue testing whether on reading is in range or not. Calibration of this circuit will be covered in SEC B, STEP 6.
8 Depress & release test button: Unit should give
BAT reading. This should be 815 +/- 1% (807-823). Even if number is outside of range, continue on to steps 9-12. Refer to troubleshooting guide to diagnose after completing step 12.
12 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service
9 Depress & release test button: Ambient
temperature should be displayed. 20 degrees C = 68 degrees F 22 degrees C = 72 degrees F 24 degrees C = 75 degrees F 25 degrees C = 77 degrees F 27 degrees C = 81 degrees F 30 degrees C = 86 degrees F If unit displays a reading outside ambient temperature range, complete steps 10 thru 12. Then, refer to troubleshooting guide. If unit displays a reading of 390-400 degrees F, check battery voltages. If battery voltage is low, replace battery and repeat steps 4 thru
12.
10 Depress & release test button: Unit should
read FO = 55. This is a software coefficient.
11 Depress & release test button: Unit should
display a flashing ‘BAT OK’ message.
12 Depress & release test button: Unit should
display ‘DO DCV’ message.
13 Turn unit off. Depress test button and turn unit
on, release test button. Unit should read: ‘Test Mode’ >VER.3.5 >clear > select. The latest software is VER 3.5: If unit reads VER
3.2, the software must be changed. The PROM is part no 610102-535. It goes into position U4 of the LCD digital PC board.
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 13
Section 2 - Service
Disassemble Unit:
SECTION B
Wear ESD wrist strap when performing the following operations.
1 Observe backplate; notice what must be
present:
If there is a raised surface where the ‘flat’ should be and/or if there is no recessed circle, replace backplate (610015-503).Remove battery holder before discarding old backplate.
NOTE:Observe back of LCD board. If back of LCD board does not have a metal bar, you must replace this board.
2 Apply a moderate amount of force to the side
of each battery contact. Contact should hold firm. If it does not, take battery contact off and clean any dried epoxy off of metal conducting surface on PC board. Obtain a new battery contact (610067-1) and solder to PC board, noting proper positioning.
Before soldering, screw contact onto PC board with appropriate hardware.
Lastly, tighten to specifications (approx. 12-16 in/oz.) or until moderately firm. Take note that black fiber washer is not destroyed. Perform this step whether epoxy was adequate or contact had to be soldered.
14 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service
3 If required, change Red LED to orange
(230015-6).
4 If necessary, change software. You must
remove digital board from front plate to do this. PROM is location U4 on digital board.
5 If necessary, reverse side contacts. To do this,
remove the two metal contacts from the black plastic piece. Reverse the plastic piece so that metal contacts will be raised outside of plastic surface of main frame. Do this for each side.
6 Calibrate the transducer circuit: to get to ON
reading, see steps 4-7 section A. If ON reading was in spec.(see step 7, section A), set to 50 or as close as you can get (within 5 cts.), using R33 (pot at bottom of PC board).
Apply 58.8 DPA (60 mm of water with manometer) to transducer via long hose.
Adjust R36 (pot just above R33) to 844 +/- 5 cts. Return to ODPA, reading should be 50 +/­5 cts. Readjust if necessary.
If the transducer reading is outside of the 30-80 range, measure the transducer output ofpins 2 and 4 of the transducer. They are the two white wires coming from the transducer. You should see no more than approx. 300 micro volts (.3mv). If this is the case, readjust ON reading to spec. as per sec. B, step 6.
If you see more than 300 mcvolts (.3mv), replace transducer.
If you cannot build pressure in transducer, diaphragm is torn, replace transducer.
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 15
Section 2 - Service
Once transducer is replaced, calibrate according to Section B, step 6.
Once transducer is calibrated, re-assemble unit.
7 With unit assembled and good batteries in,
without the pneumotach (top piece), insert into working printer charger. With a DC ammeter, measure current between the top of each battery. You should see approx. 11 ma recharge current. If reading is outside of 10-12 ma, refer to troubleshooting guide.
8 If reading is good, assemble pneumotach onto
handle. Obtain final test form. Step through, test in ‘DATA’ mode as per Section A, steps 4-12 . Observe ‘ON’ reading . As you do this and write down in appropriate place on form. Continue to move forward until you see DO DCV.
At this point throw switch on test printer to RXTX OK position. This puts a short circuit between the RX and TX communication lines. Insert handle. The following messages should appear:
DCV SEEN > RXTX OK > CHARGE > screen blank. If they do, remove handle, the message DO DCV
should re-appear. Return switch on test printer to RX fail position.
These tests are taken from spec. A01943 Rev.F.
Get ready to calibrate w/3.0 liter syringe according to spec. A01945 Rev.D.
16 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service
1. ESD protection is required whenever working with the pneumocheck printer .
2. Apply power to the unit. It should beep twice and the “power LED” will light. The “data” light should not be lit.
3. Note and record “dip” switch settings if customer owned. These switches will have to be returned to these positions when testing is complete.
4. Check printer well to make sure there is no foam pad. If there is one, it must be removed. Make sure printer is powered down. Remove pad by walking and rolling back pad from ends. Move slowly as to not tear the pad if possible. Continue until pad is removed. Clean remaining adhesive with an adhesive remover solution. Try to avoid scratching the printer well plastic piece.
2.2 Bench Test of the
Pneumocheck Printer/charger
5. Complete the following tests: A . With a DVM set on the 200VDC range, measure 14vdc across contacts as shown below that are marked with an (**).
14 vdc
** ##
gnd
## **
B. Check voltage between ground contact in printer well and pin 8 of the J2 connector. The J2 connector is the longer of the two groups of pins in front of the printhead. Pin 1 is at the left end of the longer group, counting left to right. The voltage should be between
4.75 and 5.75 vdc.
C. Check voltage between ground contact in printer well and pin 1 of the J2 connector. The voltage should be 24vdc +/- .5 volts.
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 17
Section 2 - Service
D. Install a paper roll if one is not installed. Measure the
dc voltage between the ground contact in printer well and pin 3 of the J1 connector . The J1 connector is the smaller of the two groups of pins in front of the printhead. Pin 1 is on the left, counting left to right. The voltage should be 3.00 vdc or less.
6. Making sure the red capped lever at right of printhead is pushed all the way toward the back of the printer, hold down the feed button. The paper should advance steadily forward.
7. Hold down feed button, then press and release the stop button. The paper should automatically feed in the reverse direction. Release the feed button. Paper will stop feeding.
8. Push the print button. The unit should beep, indicating that it has no data. The data light should remain unlit.
9. Download data into printer by placing spirometer handle, in on state, with data in it, into the printer well. The data light should flash repeatedly for several seconds. After this, the data light should be lit.
10.Make sure all dip switches on bottom of printer are set to the “off” position. On top of printer, set the following: Age: 36; Height: 159; Race: 2; Sex: F .
Depress and release the print button to obtain a printout. The printout should agree with the settings. Also, the standard printed should be Knudson 1976 (dip switches 1 and 2 in off position), short form (dip switch 3 in off position), and the flow over volume graph (dip switch 4 in off position).
18 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service
1 1.Set dip switches on bottom of printer to “on” position.
On top of printer, set the following: Age: 59; Height: 96; Race: 1; Sex: M.
Depress and release the print button to obtain a printout. Observe printout. The printout should agree with the settings. Also, the standard should be Euro CCS (dip switches 1 and 2 in on position), the long form, which is a questionnaire (dip switch 3 in on position), and the volume over time graph (dip switch 4 in on position).
12.Return dip switches to position they were when unit came in from customer. If unit is a loaner , or a new unit being returned to stock, set all dip switches to off position.
13.Hold down the print button and press/release the stop button. Release the print button. A test printout should print. At right side of top line, check the version of software. It should be V3.2. If it is V3.1, it needs to be changed. The “PROM” that controls this is in position U5 on the main
pcboard. Replace existing Prom with
part number 610102-506. This Prom is software version V3.2.
14.Perform agency tests on printer. These include the ground continuity test and the leakage current test. If the main pcboard, transformer or line filter was replaced, perform the hypot test. Observe standards. They are:
Ground continuity test: .2 ohm maximum resistance with 10 amperes of current flow through the earth ground line from the wall plug to a distant metal point on the unit, probably the transformer screw opposite where the ground wire is connected;
Leakage current test: 100 microamps or less from any dead metal part on printer to earth ground, with the printer ungrounded and powered up.
Hypot test: 1200 vac across the hot and neutral (tied together) and earth ground for one second without breakdown (voltage reduction).
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 19
Section 2 - Service
15.This concludes the printer bench test. Make sure there is an ample supply of paper in printer. When loading paper, feed about 2-3 inches, put leading edge of paper through paper cutter envelope before putting paper cutter in place. Any questions call Welch Allyn Technical Service at 800-535-6663, Technical Support.
20 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 2 - Service

Required Tools:

1. Compression ring tool - Welch Allyn tool number T8608
2. Snap ring pliers (external) VACO #22
3. #1 Phillips head screwdriver
4. 1/4 inch flat head screwdriver
5. 3/32" Hex Key

Required Test Equipment

1. Merriam M131 Manometer with C34259 mount kit or Welch Allyn Pressure Applicator Box.
2.3 Tools and Test Equipment Listing
2. Pressure Source - Welch Allyn number
T9161
3. 0-7VDC Power supply with 100 mA current limit
4. 0-14VDC Power Supply
5. 0-50mADC ammeter
6. 0-14VDC voltmeter
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 21
Section 2 - Service
2.4 TEST AND
ADJUSTMENTS

PNEUMOCHECK HANDLE

A set of specification drawings have been included in Section 7 with reference made to them here. The following tests should be done before the unit is returned to the customer.
1 Calibrate and test the handle per specification
drawing A01945. “Calibrating with Welch Allyn Spirometer.
2. Perform all tests on specification drawing A01951, “Dynamic Waveform Testing”
For troubleshooting use, specification drawing A01943 has been included for testing at the PC Board Level.
PNEUMOCHECK PRINTER/
CHARGER
22 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
1. Perform Dielectric Strength Test per:
A00167- UL/CSA Models A00178- IEC Models
2. Perform Leakage Current Test per A00171
3. Perform Grounding Continuity Test per:
A00172 - UL/CSA A02242 - IEC
4. Perform all tests listed on A01958, “Base Final Testing”
For troubleshooting use, specification drawing A01946 has been included for testing at the PC Board Level.
Section 3 - Theory of Operation
3.1 Spirometer Circuit Description
3.2 Spirometer - Printer Circuit Description
3.3 Messages

Theory of Operation

Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 23
Section 3 - Theory of Operation

3.1 Spirometer Circuit Description

General
The handheld spirometer’s electronics are contained on two boards, designated the “Digital Board” and the “Analog Board”. The analog board contains the power control circuitry, as well as the analog signal amplifiers. The digital board contains the digital circuitry, as well as the front panel components.
The spirometer is designed to minimize power consumption by having four separate operational states:
a. OFF When SW1 is off only the battery charge
circuit is enabled. Only diode leakage current flows.
b. ASLEEP In this state SW1 is on and Vcc is
present but the processor is not running. The switched analog power is off.
In this state: i. CR1 is not oscillating
ii. RB is oscillating (100Khz) iii. All port bits are tri-state except for P6
which has all bits low iv. The data and address bus is passive v. The RAM is disabled (U5 pin 26 is low) vi. TP6 is low
Normally in this state, since the LCD controller is
running, there will be something on the display. The software can, however, intentionally blank the display.
c. AWAKE In this state the digital circuitry is
operating, but the analog power is off. This state is entered when the processor has some work to do.
24 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Section 3 - Theory of Operation
In this state:
i. VA+ and VA- are not present ii. U12 and U13 are not powered iii. U8 i s not operating iv . CR1 is oscillating (8Mhz) v . The CPU clock (pin 22 of U4) is running
(2Mhz)
In normal operation this state will terminate and return to the sleep state as soon as the p ro ce ss or has nothing to do, such as while it is waiting for a keypress.
d. FULL ON In this state all of the circuitry is turned
on. The charge pump may or may not
be operating.
If everything is correct the power consumption in the four states should be:
State 1: OFF: Less that 0.1ma, typ 0.001ma State 2: ASLEEP: Less than 2.5ma, typ 1.6 ma State 3: AWAKE: Less than 16ma, typ 10ma State 4: FULL ON: Less than 35ma, typ 25ma
NOTE: That achieving these levels is both a hardware and a software function. Malfunctioning software can
raise the power to as much as 5ma in State 2 and 70ma
in State 4.
The maintenance mode of the spirometer allows it to be
placed in each state for testing purposes.
2. Power Consumption
Service Manual 610132 Rev. E Pneumocheck Handle Printer/Charger 25
Section 3 - Theory of Operation
3. Spirometer Power Circuit Description
The battery charge is supplied by U14 which is a voltage regulator wired as a 12ma current source. D6 and D7 provide reverse polarity protection.
Transistor T1 is the source for the current used in the spirometer if it is turned on while plugged into the printer. Its output voltage is regulated by the zener D13 to a voltage which will reverse bias D18 while at the same time insuring that the voltage reaching U8 does not exceed its design limit of 10v. This means that the tested voltage at TP1 must lie between 7.7 and 9.6v with 14-20v applied to the DC jack.
R22 and R23 divide the zener voltage to provide the sense voltage so that the processor can tell when DC is applied. This voltage should be between 4.20 and 4.85 volts with DC applied to the charging jack. This sense voltage is taken to pin 35 of the CPU and pin 1 of the U1.
The raw power for the spirometer comes from either the battery via D18, or if DC is supplied, via T1. It goes through the ON/OFF switch SW1 to the 5v regulator
119. U9 also supplies the low battery warning if Vcc comes out of regulation. This will occur when the battery voltage drops below about 6.3 volts.
Power for the analog sampling circuitry is taken directly from the battery so that it is impossible to test patients while powering the unit from the DC jack. T5 and T6 form a software controlled switch to turn analog power on only while actually needed and thus conserve battery power. This voltage will be on when CPU pin 2 is high. The voltage at the collector of T6 will be 6 to 10 volts when on, and can be easily tested at pin 1 or U1.
This switched analog voltage is inverted by U8, which should have a visible charge pump oscillating at eight kilohertz across C18. It will develop at least -6 volts across C20.
The switched analog voltage and its inversion are regulated by U7, T3, and T4 to provide +/- 4 volts VA+ and VA-. R18 and R19 divide Vcc to provide a guide voltage. U7A and T3 use this voltage to give VA+. U7B and T4 invert VA+ to provide a matching VA-.
26 Pneumocheck Handle and Printer/Charger Service Manual 610132 Rev. E
Loading...
+ 63 hidden pages