Welch Allyn Blood Pressure Monitor User manual

ThermoScan
PRO 4000
Type 6021
1-800-327-7226
Type: 6021
I / O
6021221_PRO4000_S1 Seite 1 Freitag, 17. März 2006 11:19 11
Welch
Service Documentation
Allyn
Part Number: 701627
Rev. G
This documentation provides only technical information - technical data, calibration instruction, general information and regulations for medical products.
The function of the thermometer is described in the Operating Instructions, Welch Allyn P/N 701622
Technical data: Display: LCD-Display.
Display resolution: 0.1 °C / °F.
Displayed temperature range: 20.0 °C - 42.2 °C (68 .0 °F - 108.0 °F). Work place temperature range: 10.0 °C - 40.0 °C (5 0.0 °F - 104.0 °F). Work place humidity: up to 95 %. In the range between: 35.5 °C - 42.0 °C (95.9 °F - 107.6 °F).
the accuracy of allowable tolerance with probe cover is: ± 0.2 °C (± 0.4 °F ).
Outside the range from: 35.5 °C - 42.0 °C (95.9 °F - 107.6 °F). the accuracy of allowable tolerance with probe cover is: ± 0.3 °C (± 0.5 °F ).
Probe tip: heated up to approximately 37°C. Power supply: rechargeable Braun battery-pack with
base-station and smart plug
or 2 1.5 V type AA (LR 6) batteries. Smart plug: 100-240 V - input / 12 V – output. Base-station: 12 V - input / approx. 2.4 V – output. Battery-life time: 3 month (with *) / 1000 measurements. Rechargeable battery lifetime: 3 month / 500 measurements per charge. Rechargeable battery charging: Thermometer stored in basis-station. Rechargeable battery charging time: 14 hours. Base-station function indications: 1 LED, green for power connection
1 LED, yellow for charging indication. Electronic anti-theft *: for Pro 4000 appliances only and with
the base-station in use. Automatic turn off: If no button is pressed within 60 seconds, the
appliance will turn off automatically.
Welch Allyn
Manual turn off: Push - button for >3 seconds.
701627 REV. G Page 1 of 9
Welch
Service Documentation
Allyn
Part Number: 701627
Rev. G
General hints
Production code Near the battery storage LOT 4042.
4 = Year / 042 = Day.
Attention:
Hint:
Do not dispose of empty batteries or defective rechargeable batteries in household waste. Disposal can take place at appropriate collection sites provided in your country
Type BF equipment This appliance conforms to the following standards: DIN EN 60601-1
<<Medical electrical equipment>>-Part 1: General requirements for safety.
Hint:
Labour Group: 03
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device Directive).
0297
Welch Allyn
701627 REV. G Page 2 of 9