Welch Allyn Autorefractor Quick Reference Guide

R
R
TM
Quick Reference Guide
Pre-test Set-up:
Push any button to turn the unit on.
Position the patient so that the test can be conducted level with, and square to, patient’s eyes.
Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double beep and see the desired icon on the LCD.
Explain the test procedure to the patient:
“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”
Test Procedure:
Push the “GO” button on the unit
Check that the patient is fixated correctly throughout the test.
Position the unit at the correct distance from the patient:
When the unit is too far away, you will hear slow, low-pitched beeps.
Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———
When the unit is too close, you will hear quick, high-pitched beeps.
The crosshair will flash in synch with the tones.
Look through the peephole and align the crosshair on the pupil of the patient’s right will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °
If you are not acquiring data (steady tone without acquiring data chirps):
Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.
Ask the patient if they can see the red light.
When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not need to press any buttons for the left eye test.)
Without changing position, rotate
the unit to the left eye and align the crosshair over the pupil. Repeat test.
At the end of the test, you will hear the “tah-dah” sound again.
If the unit has not gathered enough good readings from either eye, you will hear 5 tones when the test stops. - - - - - You must re-test that eye.
To stop a test at any time, hold any button until 5 tones sound.
Hints To Insure Good Results & Rapid Test:
Ambient lighting:
Don’t perform by uncovered windows
Dimmed light can help for those with small pupils
Too dim is difficult for those with dark irises
Entry angle - straight and level
•Rotate
Fixation: “Look at the red light”
Remind and monitor
Match patient age and mode
Correct distance/Not taking readings (steady tone without chirps):
Ensure you are level and square with the patient
Scan the crosshair around the eye in an outward moving spiral until
Ask patient if they can see the red light
Aim high if testing through glasses
Make sure the patient’s eyelids are not occluding the pupil
unit to test left eye
These are noted on the print out
chirps begin, hold this location
• • • • • •
••••••
eye. While the unit is acquiring data, you
Clear and Child/Adult Calibration
• Clears readings
• Changes child/adult (hold down)
Left/Right Toggle & Print
• Selects eye to re-test
• Prints results (hold button down while aiming patient side of SureSight™ at printer until you hear “tah-dah”)
“GO” Button
Unit on: starts test
Unit off: recalls old
readings
TM
Quick Reference Guide
Pre-test Set-up:
Push any button to turn the unit on.
Position the patient so that the test can be conducted level with, and square to, patient’s eyes.
Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double beep and see the desired icon on the LCD.
Explain the test procedure to the patient:
“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”
Position yourself at eye-level and square with the face of the patient.
Test Procedure:
Push the “GO” button on the unit
Check that the patient is fixated correctly throughout the test.
Position the unit at the correct distance from the patient:
When the unit is too far away, you will hear slow, low-pitched beeps.
Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———
When the unit is too close, you will hear quick, high-pitched beeps.
The crosshair will flash in synch with the tones.
Look through the peephole and align the crosshair on the pupil of the patient’s right will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °
If you are not acquiring data (steady tone without acquiring data chirps):
Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.
Ask the patient if they can see the red light.
When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not need to press any buttons for the left eye test.)
Without changing position, rotate
the unit to the left eye and align the crosshair over the pupil. Repeat test.
At the end of the test, you will hear the “tah-dah” sound again.
If the unit has not gathered enough good readings from either eye, you will hear 5 tones when the test stops. - - - - - You must re-test that eye.
To stop a test at any time, hold any button until 5 tones sound.
Hints To Insure Good Results & Rapid Test:
Ambient lighting:
Don’t perform by uncovered windows
Dimmed light can help for those with small pupils
Too dim is difficult for those with dark irises
Entry angle - straight and level
•Rotate
Fixation: “Look at the red light”
Remind and monitor
Match patient age and mode
Correct distance/Not taking readings (steady tone without chirps):
Ensure you are level and square with the patient
Scan the crosshair around the eye in an outward moving spiral until
Ask patient if they can see the red light
Aim high if testing through glasses
Make sure the patient’s eyelids are not occluding the pupil
unit to test left eye
These are noted on the print out
chirps begin, hold this location
• • • • • •
••••••
eye. While the unit is acquiring data, you
Clear and Child/Adult Calibration
• Clears readings
• Changes child/adult (hold down)
Left/Right Toggle & Print
• Selects eye to re-test
• Prints results (hold button down while aiming patient side of SureSight™ at printer until you hear “tah-dah”)
“GO” Button
Unit on: starts test
Unit off: recalls old
readings
Results
Reliability Number indicates the number of good readings
obtained and their consistency, based on a 1 to 9 scale (higher numbers are better).
Sphere
Cylinder
Reliability Number
Results
Reliability Number indicates the number of good readings
obtained and their consistency, based on a 1 to 9 scale (higher numbers are better).
Sphere
Cylinder
Reliability Number
6 is typically acceptable
•>
Differ ence
5 is marginal / repeat the test if possible 4 is poor / repetition of the test is necessary
•<
If the patient did not appear to fixate on the unit during data acquisition, repeat the test.
is the sphere, or power of the eye measured in diopters. Negative numbers indicate myopia (near-sightedness), positive
S
numbers indicate hyperopia (far-sightedness).
is the cylinder, a measure of astigmatism, or irregular focus of the eye.
C
is the difference in the mean spherical power between the two eyes (displayed in child mode only).
D
* An asterisk on a reading in child mode indicates the reading is in the referral range for a pre-schooler.
A +9.9 or -9.9 indicates a reading outside the unit’s measurement range.
Referral Criteria
6 Months 3-5 years*
Sphere <-2.0 or >+3.5 <
Cylinder ------- > 1.0
-1.0 or > +2.0
0.5 1.00
1.5
Child/Adult Mode
+2.5+1.5-0.5 +0.5-1.5 +2.0+1.0-1.0 0
6 is typically acceptable
•>
Differ ence
5 is marginal / repeat the test if possible 4 is poor / repetition of the test is necessary
•<
If the patient did not appear to fixate on the unit during data acquisition, repeat the test.
is the sphere, or power of the eye measured in diopters. Negative numbers indicate myopia (near-sightedness), positive
S
numbers indicate hyperopia (far-sightedness).
is the cylinder, a measure of astigmatism, or irregular focus of the eye.
C
is the difference in the mean spherical power between the two eyes (displayed in child mode only).
D
* An asterisk on a reading in child mode indicates the reading is in the referral range for a pre-schooler.
A +9.9 or -9.9 indicates a reading outside the unit’s measurement range.
Referral Criteria
6 Months 3-5 years*
Sphere <-2.0 or >+3.5 <
Cylinder ------- > 1.0
-1.0 or > +2.0
0.5 1.00
1.5
Child/Adult Mode
+2.5+1.5-0.5 +0.5-1.5 +2.0+1.0-1.0 0
Difference >1.5 >1.0
1.50.5 1.00
Within Normal Limits
Outside Normal Limits
Readings outside the normal limits should be confirmed prior to referral.
It is recommended that referral criteria be agreed upon between you and the eye care specialist receiving your referrals.
Approximate
Relationship of Sphere to Visual Acuity
Negative (-)
Diopters
0.0 20/20 0.0 20/20
-0.5 20/30-40 +1.0 20/20
-0.8 20/50 +2.0 20/20
-1.0 20/60 +3.0 20/25
-1.25 20/70 +3.25 20/30
-1.75 20/100 +3.75 20/40
-2.0 20/200 +4.0 20/40-50
-3.0 20/285 +4.3 20/50
-4.0 20/420 +5.0 20/70
-5.0 20/565
Myopia
(Near-Sightedness)
Positive (+)
Diopters
Hyperopia
(Far-Sightedness)
✭✭
✭✭
✭✭
✭✭✭✭
Ages 5-15
Charging
A charged unit can test patients for approximately 3 hours (continuous use).
To charge, firmly place the unit into the stand.
With 15 minutes of on-time remaining, the low-battery indicator will begin to flash on the display. When low-battery indicator is steady, battery must be recharged.
Part Number 140357-2 Rev. G
Difference >1.5 >1.0
1.50.5 1.00
Within Normal Limits
Outside Normal Limits
Readings outside the normal limits should be confirmed prior to referral.
It is recommended that referral criteria be agreed upon between you and the eye care specialist receiving your referrals.
Approximate
Relationship of Sphere to Visual Acuity
Negative (-)
Diopters
0.0 20/20 0.0 20/20
-0.5 20/30-40 +1.0 20/20
-0.8 20/50 +2.0 20/20
-1.0 20/60 +3.0 20/25
-1.25 20/70 +3.25 20/30
-1.75 20/100 +3.75 20/40
-2.0 20/200 +4.0 20/40-50
-3.0 20/285 +4.3 20/50
-4.0 20/420 +5.0 20/70
-5.0 20/565
Myopia
(Near-Sightedness)
Positive (+)
Diopters
Hyperopia
(Far-Sightedness)
✭✭
✭✭✭✭✭✭✭✭
Ages 5-15
Charging
A charged unit can test patients for approximately 3 hours (continuous use).
To charge, firmly place the unit into the stand.
With 15 minutes of on-time remaining, the low-battery indicator will begin to flash on the display. When low-battery indicator is steady, battery must be recharged.
Part Number 140357-2 Rev. G