The information in this document is subject to change without notice.
Welch Allyn makes no warranty of any kind on this material, including but not limited to, the implied
warranties of merchantability and fitness for a particular purpose. Welch Allyn shall not be liable for
errors contained herein or for incidental or consequential damages concerning the furnishing,
performance, or use of this material.
This document contains proprietary information that is protected by copyright. All rights are reserved. No
part of this document may be photocopied, reproduced or translated to another language without prior
written consent of Welch Allyn.
Before using this instrument, read this guide and become thoroughly familiar with the contents.
Welch Allyn only considers itself responsible for any effects on safety, reliability and performance of the
equipment if:
1 assembly operations, extensions, re-adjustments, modifications or repairs are done by persons
authorized by Welch Allyn, and
2 the electrical installation of the relevant room complies with the IEC or national requirements, and
3 the instrument is used according to the instructions for use presented in this manual.
WARNING
Welch Allyn assumes no liability for failures resulting from RF interference between Welch Allyn
medical electronics and any radio frequency device at levels exceeding those established by applicable
standards.
The use of accessories other than those recommended by Welch Allyn may compromise product
performance.
United States Federal law restricts this device to sale by or on the order of a licensed health care
practitioner.
ii
Table Of Contents
1HOW THIS MANUAL WORKS.....................................................................................................................1
1.1A Q
UICK TOUR OF THE WELCH ALLYN ATLAS MONITOR...............................................................................2
2MONITORING THE PATIENT......................................................................................................................4
4CAPTURING AND DISPLAYING TREND DATA.................................................................................... 32
5USING PRINT AND WAVEFORM FREEZE............................................................................................. 34
5.1T
HE PRINTER – LOADING PAPER AND TROUBLESHOOTING ...........................................................................36
6CONNECTING TO THE PATIENT............................................................................................................. 38
6.1C
6.2C
6.3C
6.4C
6.5C
ONNECTING THE NIBP CUFF....................................................................................................................... 40
ONNECTING THE SPO2 FINGERCLIP SENSOR ................................................................................................ 43
ONNECTING THE ECG ELECTRODES ...........................................................................................................45
ONNECTING THE TEMPERATURE PROBE (MODELS 622XX & 623XX).......................................................... 49
ONNECTING THE END TIDAL CO2 SAMPLE TUBE (MODEL 623XX).............................................................. 50
7USING THE MENUS...................................................................................................................................... 51
7.1T
HE SET DATE AND TIME AND OTHER OPTIONS MENU................................................................................. 51
7.2T
HE ADVANCED CONFIGURATION MENU...................................................................................................... 53
7.2.1Advanced Configuration Menu Settings.............................................................................................. 55
8CLEANING AND MAINTAINING THE ATLAS MONITOR.................................................................. 57
9UNPACKING AND INSTALLING THE ATLAS MONITOR ..................................................................59
10APPENDIX A: TECHNICAL SPECIFICATIONS FOR THE ATLAS MONITOR................................ 61
The Welch Allyn Atlas Monitor is intended for use in a hospital or clinical environment. It should not
be used at home or in emergency transport vehicles. Monitor users should be skilled at the level of a
technician, nurse, doctor or medical specialist.
The function of the Welch Allyn Atlas Monitor is to register ECG, CO
concentration, Heart Rate,
2
Noninvasive Blood Pressure (Systolic, Diastolic, and Mean Arterial Pressure), Pulse Oximetry,
Respiration Rate and Temperature for adult and pediatric patients (over the age of 3 years), in all hospital
or clinic facilities.
To ensure patient electrical isolation, the Atlas Monitor should only be connected to other equipment
that provides patient electrical isolation. When connecting the Welch Allyn Atlas Monitor to any
instrument, verify proper operation before clinical use. Accessory equipment connected to the monitor’s
serial data interface must be certified according to IEC Standard 950 for data-processing equipment or
IEC Standard 601-1 for electromedical equipment. All combinations of equipment must be in compliance
with IEC Standard 601-2 systems requirements. Anyone who connects additional equipment to the signal
input / output port is configuring a medical system and is therefore responsible that the system comply
with the requirements of IEC Standard 601. If in doubt, consult the Welch Allyn Technical Service
Department.
The Welch Allyn Atlas Monitor and its accessories should be tested by qualified service personnel at
regular intervals to verify proper operation, according to the procedures of the user’s institution. A
Service Manual is available from the manufacturer. Other important safety information is located
throughout this manual where appropriate.
iv
Warnings, Cautions, and Notes
All operating personnel should be familiarized with the general safety information in this summary.
Specific warnings and cautions will also be found throughout the operator’s manual. Such specific
warnings and cautions may not appear here in this summary.
Conforms toIEC 60601-1
UL 2601-1
CAN/CSA C22.2 No 601-1
C US Approved to Australia AS 3200.1, Appendix Z
Defibrillator-proof, Type CF Applied Part
Attention, consult accompanying document
Handle with care
Storage temperature. Refer to technical specification for more details.
Lead Acid Battery. For disposal see the Maintenance section of this manual.
Storage humidity. Refer to technical specification for more details.
Warning – Tells you about something that could hurt the patient or hurt the operator.
Caution – Tells you about something that could damage the monitor.
Note – Tells you other important information.
v
Warnings
• The Welch Allyn Atlas Monitor is designed for use by medical clinicians. Although this manual may
illustrate medical monitoring techniques, this system should only be used by a trained clinician who
knows how to take and interpret a patient’s vital signs.
• Do not operate this product in the presence of flammable anesthetics. Explosion may result.
• WARNING – PACEMAKER PATIENTS. Rate meters may continue to count the pacemaker rate
during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance. See this manual for disclosure of the
pacemaker pulse rejection capability of this instrument.
• This device must be used in conjunction with clinical signs and symptoms. This device is only
intended to be an adjunct in patient assessment. Certain arrhythmias or pacemaker signals could
adversely affect heart rate indications or alarms.
• During defibrillation, keep the discharge paddles away from ECG and other electrodes, as well as
other conductive components in contact with the patient. Avoid contact with any accessories
connected to the Welch Allyn Atlas Monitor’s panel.
• Prolonged use or the patient’s condition may require changing the SpO
Change sensor site and check skin integrity, circulatory status, and correct alignment at least every 4
hours.
• When monitoring blood pressure over an extended period of time, or at frequent intervals, it is
recommended to check the cuff site and cuffed extremity regularly for possible ischemia, purpura
and/or neuropathy.
• Thoracic impedance respiration measurement may interfere with some pacemakers. Refer to the
pacemaker’s manual.
• To ensure patient safety, the conductive parts of the ECG electrodes (including associated
connectors) and other patient-applied parts, should not contact other conductive parts, including
earth ground, at any time.
• The safety and effectiveness of this product in the detection of apnea, particularly for infants and
neonates, has not been established.
• This equipment must not be connected to any other equipment that is not compliant with EN60601-1,
or a possibility exists that combined leakage currents could exceed safe limits.
• WARNING: Use of accessories, transducers, and cables other than those specified may result in
degraded electromagnetic compatibil ity perfo rmance of thi s devic e.
• Do not operate this product with MRI (Magnetic Resonance Imaging) equipment.
• It is the operator’s responsibili ty to set alarm lim its as approp ria te for each ind iv idua l patien t.
• Any Atlas Monitor which has been dropped or damaged should be checked by qualified service
personnel to insure proper operation prior to use.
• There are no user serviceable parts inside the Atlas Monitor other than paper replacement and
battery replacement.
• Blood pressure measurements may not be accurate for patients experiencing moderate to severe
arrhythmias.
• This Atlas Monitor should not be used on patients who are linked to heart / lung machines.
• If the integrity of the external protectiv e conduc tor in the ins tal lat ion or its arrangements is in doubt,
equipment shall be operated from its internal power source (models 622xx and 623xx).
• If an electrosurgical unit is used, place the ECG cable and wires as far as possible from the site of
the surgery and from the electrosurgical cables. This will minimize interference and the risk of burns
to the patient. Ensure that the electrosurgical return cable (neutral) is well attached and making good
contact with the patient.
sensor site periodically.
2
vi
• End tidal carbon dioxide (ETCO2) and breath rate measurement and alarm capability are active
ONLY when the second trace option is set to CO
ETCO
or ECG) the CO
and breath rate waveforms and data to another second trace selection (SpO2, Respiration
2
and breath rate monitoring and alarm capability will be disabled. This occurs
2
. Should the operator change from viewing the
2
even if the watertrap and cannula are still inserted into the Monitor.
• Impedance Respiration rate measurement and alarm capability are active ONLY when the second
trace option is set to Respiration. Should the operator change from viewing the Respiration
waveforms and breath rate to another selection (SpO
, CO2 or ECG) the Respiration rate
2
monitoring and alarm capability will be disabled. This occurs even if the ECG cable is still inserted
into the Monitor.
Cautions
• Place the Welch Allyn Atlas Monitor and accessories in locations where they cannot harm the patient
if they fall off a shelf or mount.
• Never place fluids on top of this monitor. In case of fluid spilling on the monitor, disconnect power
cord, wipe clean immediately and have the monitor serviced to ensure that no hazard exists.
• This Welch Allyn Atlas Monitor should not be stacked directly on top of other equipment, and other
equipment should not be stacked on top of this Welch Allyn Atlas Monitor. If stacking is necessary,
observe the Welch Allyn Atlas Monitor to verify normal operation in the stacked configuration in
which it will be used.
• Unplug the external power cord from the monitor before cleaning or disinfecting the monitor.
• Do not autoclave, subject to ethylene oxide sterilization, or immerse the Welch Allyn Atlas Monitor in
liquid. Sterilize accessories only according to the manufacturer’s instructions.
• Make frequent electrical and visual checks on cables and electrode wires.
• Ensure the AC rating for the device is correct for the AC voltage at your installation site before using
the monitor. The AC rating is shown on the back of the instrument. If the rating is not correct, do not
use the monitor, and contact the Welch Allyn Technical Service Department for help.
• Line isolation monitor transients may resemble actual cardiac waveforms and thus inhibit heart rate
alarms. Use care in placement of ECG electrodes and routing of cables to avoid interference and
noise.
• Electrode polarization: the type of electrode used can affect the recovery time from overload,
especially defibrillation. Electrodes of dissimilar metals should not be used.
• If the accuracy of any measurement is in question, check the patient’s vital sign(s) by an alternate
method and then check the Atlas Monitor for proper functioning.
• Extremity and cuff motion should be minimized during blood pressure determinations.
• The pulse oximeter is calibrated to determine the percentage of arterial oxygen saturation of
functional hemoglobin. Significant levels of dysfunctional hemoglobins such as carboxyhemoglobin
or methemoglobin may affect the accuracy of the measurement.
• Grounding reliability can only be achieved when equipment is connected to an equivalent receptacle
marked “Hospital Only” or “Hospital Grade”.
Notes
• Sidestream waste material and the CO2 watertrap should be treated as biohazard material.
• Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the
vii
American National Standard, Electronic or automated sphygmomanometers (SP 10).
• Blood Pressure measurements can be affected by the position of the patient, by the patient’s
physiological condition, and other factors.
• The Blood Pressure system and Temperature system may not meet specifications if operated or stored
at conditions outside the stated ranges, or subjected to excessive shock or dropping.
•The Blood Pressure system is compliant with requirements of EN 1060-3:1995 Specification for Non-
invasive sphygmomanometers.
• The Atlas Monitor is designed with protective circuitry and current isolation that eliminates any risk
to the patient from possible software errors.
Product Warranty Information
Welch Allyn warrants the Welch Allyn Atlas Monitor, when new, to be free of defects in material and
workmanship and to perform in accordance with manufacturer’s specifications for a period of two years
from the date of purchase from Welch Allyn or its authorized distributors or agents. (Pulse oximetry
sensors and temperature probes are warranteed for one year). Welch Allyn will either repair or replace
any components found to be defective or at variance from the manufacturer’s specifications within this
time at no cost to the customer. It shall be the purchaser’s responsibility to return the instrument to Welch
Allyn or an authorized distributor, agent, or service representative. This warranty does not include
breakage or failure due to tampering, misuse, neglect, accidents, modification or shipping. This warranty
is also void if the instrument is not used in accordance with manufacturer’s recommendations or if
repaired by other than Welch Allyn or an authorized agent. Purchase date determines warranty
requirements. No other express warranty is given.
Return the Instrument Registration Card
Remember to submit the instrument registration card for warranty validation. Complete the information
and mail the pre-addressed card to Welch Allyn. You may also register on-line at
<http://www/welchallyn.com/medical/support/warranty>.
Service Information: Service Policy
All service and repairs must be performed by authorized Welch Allyn personnel or agents, using
approved Welch Allyn replacement parts and approved process materials. Failure to do so will invalidate
the product warranty. Please refer to the product warranty for specific coverage.
Service Information: Technical Assistance
If you have an equipment problem that you cannot resolve, call the Welch Allyn Service Center nearest
you for assistance. Technical service support is available to you by telephone on normal business days at
the phone numbers listed below.
viii
If you are advised to return a product to Welch Allyn for service or repair, schedule the repair with the
service center nearest you.
Before returning a product for repair you must obtain authorization from Welch Allyn. An RMA
(Return Materials Authorization) number will be given to you by our service personnel. Be sure to
note this number on the outside of your shipping box. Returns without an RMA number will not be
accepted.
ix
Service Information
For Technical Support or to obtain return instructions, please contact your nearest Welch Allyn service
center listed below:
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153-0220
Phone: 1-800-535-6663
Fax: 315-685-4653
Welch Allyn, Ltd. - Canada
160 Matheson Blvd. E., Unit #3
Mississauga, Canada L4Z 1V4
Phone: 905-890-0004
Fax: 905-890-0008
Welch Allyn, Ltd. - UK
Aston Abbots, Buckinghamshire
England HP22 4ND
Phone: 011-44-1296-689905
Fax: 011-44-1296-682104
Welch Allyn, Ltd. - Australia
PO Box 864
Ground Floor, 18-20 Orion Road
Lane Cove, NSW 2066, Australia
Phone: 011-61-294-183-155
Fax: 011-61-294-183-650
CLINICAL SUPPORT
For clinical questions about Atlas Monitor call the Welch Allyn Clinical Support line at 800-769-4014
Extension 3225 or 315-685-4100 Extension 3225.
x
The CE Mark on this product indicates it has been tested to and conforms with the
provisions noted with the 93/42/EEC Medical Device Directive.
European contact for regulato ry com pliance :
European Regulator Manager
Welch Allyn LTD.
Navan, Co, Meath
Republic of Ireland
Phone 353-46-79060
Fax: 353-46-27128
Service Information: Service Manual / Spare Parts
A service manual is available by request to qualified electronics personnel. The service manual is a
comprehensive guide to troubleshooting, service and repair of the Welch Allyn Atlas Monitor.
A complete spare parts price list is available upon request. Spare parts may be ordered from your local
Welch Allyn Service Center.
Service Information: Service Loaners
Service loaners are provided, on request, when repair service is provided by a Welch Allyn Service
Center. Loaners for products repaired while under the original warranty, or while under extended
warranty or service contract, are provided free of charge and are shipped within 48 hours of notification
of need. Shipment charges to the user are paid by Welch Allyn.
For service repairs outside of warranty or contract, loaners will be available for a nominal charge and will
be shipped subject to availability. Loaners will be shipped pre-paid.
xi
1 How This Manual Works
This manual is arranged so that everything about one topic is found in a single section. The statement
immediately below the chapter title (like this one) appears in italics and presents the important points of
the topic. Most topics include an illustration or a table. The chapters are numbered so that logically
connected topics begin with the same number-- for example 2.1 and 2.2.
The Welch Allyn Atlas Monitor and this manual
are designed for ease of use. Everything you
need to know about a specific operation of the
monitor is available in one place. This means
you can see all the required information at a
glance.
Redundancy - There is some redundancy in this
manual; some step-by-step procedures are
repeated in many places wherever they are
pertinent. We did this so you would not have to
search through other pages to find what you
need to know “right now.”
For instance, setting an alarm limit is fully
explained in the section on blood pressure, again
in the section on SpO
, and in several other
2
places.
Paragraphs - The statement in italics
immediately below the chapter heading
describes what the section is about. Sometimes
just reading this and looking at the illustrat ions
will give you enough information.
Section Numbers - The double numbered pages
indicate the relationship between the main
subject and closely related topics.
The sections are organized so that what you
want to know first is put first. Like most medical
professionals who use the Atlas Monitor, you
are probably very adept at taking blood pressure
and connecting ECG leads to patients, so the
section on how the Atlas Monitor is used for
monitoring patients and what you need to know
to operate the monitor comes first. We put the
information on connecting blood pressure cuffs
and ECG leads in a later section.
This manual is not meant for reading straight
through, like a book, although you can read it
that way. If you read it like this, the built-in
redundancy may become a little tedious. When
your read a paragraph or a step-by-step
procedure with which you are already familiar,
just skip it. It is there for the person who is
doing the activity for the first time.
1
1.1 A Quick Tour of the Welch Allyn Atlas Monitor
When you turn on the power, the monitor starts with preset or default alarm levels. Waveforms are
displayed on the screen. Readings are displayed on the screen and on LEDs. You can perform the most
common operations — set and silence alarms, display trend data, print waveforms and trend data — from
the front panel without using a menu. This section gives only a brief overview of the monitor; later
sections present all the details.
AC~ indicator, located below the screen,
Which model do you have? The Welch Allyn
Atlas Monitor is a single, portable unit providing
all the measurement capability normally needed to
monitor patients under anesthesia, patients
recovering from surgery, and patients who require
bedside monitoring. The model number is
encoded into the first three digits of the serial
number on the back of the unit. There are three
models:
ModelFeatures
621SO
621SP
622SO
622SP
622NO
622NP
623SP
623NP
Power on - The Power On/Standby button is in
the lower right corner. When you firs t turn on the
monitor:
• all alarms are enabled, but no alarm will
sound until after a valid measurement value is
received.
• all alarm limits are set at their defaul t va lu es.
• all the trend data (history) is cleared.
SpO2, SpO2 waveform
Pulse Rate
NIBP: Systolic, Diastolic, MAP
ECG waveform, heart rate
Printer (optional)
All features of model 621xx,
plus:
Impedance Respiration
Patient Temperature
Battery Operation
PC Communication
Remote Nurse Call
Printer (optional)
All features of model 622xx,
plus:
End Tidal CO
from ETCO
and Breath Rate
2
2
Printer (standard)
A lit
means the unit is being powered by the wall
outlet, and that the battery is being charged
(models 622xx and 623xx).
Silencing Alarms - You can silence any alarm for
90 seconds by pressing the large blue
Silence
button on the right-hand edge of the instrument.
Silenced alarms continue to flash, as long as the
measurement is outside the alarm limits.
You can suspend an individual alarm by pressing
ALARM Off button. There are four ALARM Off
its
buttons, each controlling a different group of
measurements. The audible alarm is suspended as
long as the red LED in the button is lit. When an
alarm is suspended, the audible alarm will not
sound, but readings will still flash when the
measurement is outside the limits.
Trend Data - Trend data is captured every time
blood pressure is measured, whether this event is
automatic or manual. Trend data is also
automatically captured every 15 minutes if blood
pressure intervals are longer, or blood pressure is
not used. Push the
data. Scroll through the trend data with either
Trend button to see the tre nd
Set
button. Push Trend again to return to the
waveform display. The monitor will hold up to
144 lines of trend data, which is 36 hours if data is
captured every 15 minutes.
Printing - A printer option is available with both
models 621xx and 622xx, and is a standard model
623xx feature. Push the
Print button to print what
is on the screen. If the waveforms are displayed
on the screen, the
Print button prints 15 seconds
2
of waveforms plus all the current measurements.
The printout captures data from 9 seconds before
Print button was pressed until 6 seconds after.
the
If trend data is displayed on the screen,
Print
prints all the trend data.
If your model of the Atlas Monitor does not have
the optional printer, the
Freeze, and it freezes, or halts, the waveform
Print button is labeled
display for 10 seconds to permit studying of the
waveform.
3
2 Monitoring the Patient
The patient’s vital measurements are displayed as numeric readings and as waveforms. You can set the
measurement limit alarm levels, silence the alarms for a short period, and suspend individual alarms.
You can print waveforms and current measurements, or print all the stored trend data.
The front panel has two sides:
The left side displays
waveforms, numeric
readings, and trend
data on a CRT screen.
Each side responds to the adjacent Select and Set
The right side has
measurements
displayed in green
and red LEDs.
buttons used for setting alarm limits. Each side
also has
ALARM Off buttons, which are used to
suspend (turn off) individual alarms.
Note: The temperature measurement does not
have an alarm.
Setting alarm limits - Press the Select button to
choose which alarm limit you want to set. Each
time you press the
next alarm, shown by small
Select button, it cycles to the
HI and LO indicators,
and the measurement display flashes the current
alarm setting. The
Set button changes this alarm
limit. Press the top or bottom of the button to
change the limit up or down.
Note: the flashing alarm setting mode only lasts
10 seconds before reverting back to the normal
measurement mode. If you take too long to set
a limit, you’ll need to press the
Select button
and start over again.
Press the Select button to go to the next
measurement alarm. Press it several times to cycle
out of all the alarm settings and go back to the
normal measurement mode.
Silencing alarms - The Silence button silences
all alarms for 60, 90 or 120 seconds. This silence
period can be set to one of these three choices in
Advanced Configuration. During the silence
period, there will be no audible alarms, even for
measurements that go outside the limit range for
the first time. However, any measurement that is
outside the limits you set will flash.
Trend data - Press the Trend button to see the
trend data. The waveform display is replaced by
the first screen of trend data, starting with the
most recent measurements at the top. View the
rest of the trend data by pressing the
Set button up
or down.
The trend memory can hold up to 144 lines of
measurements, which is 36 hours of data if taken
at 15 minute intervals.
Printing - The Print button prints what you see
on the screen - the waveforms (including all
current measurements) or the screen with trend
data. When you push the button, waveforms are
printed starting from 9 seconds before you pushed
Print button until 6 seconds after you pushed
the
the button for a total printout of 15 seconds. The
other information on the printout is cap ture d at the
time that the
Freeze
the button is labeled
Print button is pressed.
– If your monitor does not have a printer,
Freeze. Pressing Freeze
stops the waveform display for 10 seconds, and
then the readout resumes.
4
Top – Model 623xx showing waveform display and alarm controls
Bottom – Model 623xx showing trend display and associated
controls
5
2.1 Monitoring Blood Pressure
Blood pressure can be measured at timed intervals which you set, or you can start the blood pressure
measuring cycle manually. Systolic and Diastolic readings are shown on the LEDs at the upper right of
the monitor. You can set the high and low alarm limits for both the systolic and diastolic measurements.
Blood pressure cycles - You can measure NIBP
at timed intervals or manually. To set a timed
interval, press the
Auto button to cycle through
the available intervals: X, 1, 3, 5, 10, 15, 30, or
60 minutes. Wait until the number for the
selected interval stops flashing; the measurement
will automatically begin 20 seconds la ter .
Note: The interval is timed from the start of one
BP cycle to the start of the next cycle.
Pressing the button one more time after the 60 is
lit will return the
Auto timing to the Off mode
(indicated by an “X”). In this mode, automatic
measurement at timed intervals will not occur.
Initially, both blood pressure displays will be
blank.
Stopping a blood pressure cycle - The BP
Start/Cancel
button does one of two things:
• If a blood pressure measurement is not in
progress, pressing
BP Start/Cancel will
start a blood pressure measurement cycle,
whether the
Auto timer is set to a specific
interval or is off.
• If a blood pressure measurement is in
progress, pressing
BP Start/Cancel will
deflate the cuff immediately and cancel the
measurement. If the
Auto button is in one of
the timed modes, the cuff will inflate again
after the selected number of minutes.
Alarms - When any of the blood pressure limits
are exceeded, an audible alarm sounds and the
affected measurement flashes. You can silence
this alarm, and all alarms, by pressing the large
Silence button at the right side of the
blue
instrument. This will silence all alarms for 60,
90, or 120 seconds, depending on the setting
selected in Advanced Configuration. However,
any measurement still outside the set limits will
flash.
To suspend the blood pressure alarms, press the
BP ALARM S Of f button so the red LED in the
button lights. A suspended alarm will still flash
if it goes outside the range of the limits, but it
will not sound the audible alarm.
Trend data – Trend data is captured at each
blood pressure cycle, whether it is started
automatically or manually. If the
Auto timing of
NIBP is off (X), or greater than 15, then trend
data is captured every 15 minutes.
Initial pressure – The initial cuff pressure can
be set in the Advanced Configuration. The Atlas
Monitor will pump up to the selected initial cuff
pressure. If this pressure is too low to measure
the systolic pulse, the system will repeatedly
increase pressure by 40 mmHg and measure
again.
Note: Canceling a blood pressure cycle does not
end automatic BP timing. If the Auto is set to
any number, the next blood pressure cycle
will start again after that number of minutes
have elapsed.
6
Atlas Monitor showing NIBP displays, controls and tubing
connector
Setting alarm limits - To set the Systolic and
Diastolic alarm limits, use the
Select and Set
buttons on the right side of the monitor to follow
these steps:
• Press the
SpO
Diastolic
• Each push of
Select button to cycle through
LO, then Systolic HI and LO, and
2
HI and LO.
Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and the
HI or LO LED will
flash, indicating which limit is being
changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the
Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
LO LEDs are lit. The instrument is now in
HI or
its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).
7
2.1.1 MAP (Mean Arterial Pressure)
Mean Arterial Pressure is calculated from the systolic and diastolic measurements. MAP may be
optionally displayed by selecting it on the Advanced Configuration menu. MAP is displayed in the upper
left corner of the screen.
MAP - (Mean Arterial Pressure) display can be
turned on and off by using the Advanced
Configuration menu. When MAP is shown, you
will see it in the upper left corner of the screen,
above the ECG waveform.
To display (or turn off) MAP, enter the Advanced
Configuration menu:
• Press the
• Press the
• Use the left
Date/Time button.
Trend button.
Select button to highlight
“MAP.”
• Press the left
• Press the
Set to choose “Yes” or “No.”
Trend button again to exit
Advanced Configuration.
If MAP is displayed on the screen, it will also
appear in the trend data and in the current
readings of a waveform printout.
Alarms - When either of the MAP limits are
exceeded, an audible alarm sounds and the
affected measurement flashes. You can silence
this alarm, and all alarms, by pressing the large
Silence button at the right side of the instru-
blue
ment. This will silence all alarms for 60, 90, or
120 seconds, depending on the setting selected in
Advanced Configuration. However, any
measurement still outs ide the set limits will flash.
measured directly.
Setting alarm limits – If MAP is displayed, you
can set the alarm limits, using the
Select and Set
buttons on the left side of the monitor to follow
these steps:
• Press the left Select button to cycle
through MAP HI and LO, Heart Rate HI and
LO, Respiration HI and LO(models 622xx
or 623xx)
LO.
, then (model 623xx) CO
HI and
2
• Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
HI or LO will flash,
indicating which limit is being changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
HI
or LO indicators are lit. The instrum ent is
now in its normal measurement mode. (If
you do not press any button for 10 seconds,
the instrument will automatically revert to
its normal measurement mode).
To suspend the MAP alarm, press the
ALARMS Off
button so the red LED in the button
BP
lights. A suspended alarm will still flash if it goes
outside the range of the limits, but it will not
sound the audible alarm.
Note: MAP is calculated mathematically from the
Systolic and Diastolic pressures; it is not
8
Atlas Monitor waveform display showing location of MAP reading
9
2.2 Monitoring SpO2, Pulse Rate and the SpO2 Waveform
The oximetry and pulse rate measurements are generally taken with the reusable fingerclip sensor
(provided), however a wide variety of SpO
volume is displayed as a vertical bar graph, called the Plethysmograph, beside the SpO
right side of the monitor. The SpO
pulse tone gives an audible indication of pulse rate and oxygen level.
2
sensors are available as accessory items. The oximetry pulse
2
% display on the
2
Pulse - A fingerclip sensor provides the source
of the light transmitted through the patient’s
finger to determine the oximetry and pulse rate
measurements. The green Pulse Rate numbers
may sometimes differ slightly from the Heart
Rate displayed over the ECG waveform, even
though they both measure beats per minute
(bpm). This is normal.
Oxygen level - The oxygen level is displayed in
red numbers as a percentage. The
Plethysmograph vertical bar graph next to the
percentage shows the strength of the
SpO
2
fingerclip sensor signal with each beat. If this
signal is low, it could indicate that the fingerclip
sensor is not placed properly, or that the patient
has poor perfusion. Pigmented skin and nail
polish can also lower the signal.
System Displays – The system will begin
displaying the Plethysmograph signal almost
immediately upon attachment of the fingerclip
sensor to the patient. The SpO
level and Pulse
2
Rate will be displayed within about 5 seconds,
after the system determines th at the rea ding is
stabilized.
Second waveform - The pulse oximetry
waveform can be selected as a second trace. If
this is chosen, the bottom line of the screen
displays the SpO
waveform. Select the Second
2
trace source from the Advanced Configuration
menu.
Pulse tone - A short SpO
tone sounds with
2
every pulse beat.
• The pulse tone timing is based on the ECG
heart rate. If ECG is not used, the pulse tone
timing is based on the SpO
measurement.
2
• The pulse tone pitch is determined by the
oxygen level, increasing in frequency (pitch)
as the percentage of oxygen elevates.
The pulse tone volume can be controlled by a
button on the lower right panel. The button is
below the SpO
icon and
display, labeled with a speaker
2
SpO
.
2
Note: The pulse tone volume can be turned
completely off with this button.
Note: If the SpO
pulse tone is in synchrony with the
SpO
2
is inactive, the timing of the
2
ECG heart rate as is normally the case, but
the pitch of the tone is steady, unvarying, and
different from the tone tracking oxygen
content.
Alarms - When the oxygen percentage falls
below the SpO
limit, an alarm sounds and the
2
affected measurement flashes. You can silence
this alarm, and all alarms, by pressing the large
Silence button at the right side of the
blue
instrument. This will silence all alarms for 60,
90, or 120 seconds, depending on the setting
selected in Advanced Configuration. However,
any measurement still outside the set limits will
flash.
To suspend the SpO
ALARMS Off button so the red LED in the
alarm, press the SpO
2
2
button lights. A suspended alarm will still flash
if it goes outside the range of the limits, but it
will not sound the audible alarm.
10
Atlas Monitor showing SpO2 displays, controls, and sensor connector
Setting alarm limits - To set the SpO2 alarm
limits, use the
Select and Set buttons on the
right side of the monitor to follow these steps:
• Press the
SpO
Diastolic
• Each push of
Select button to cycle through
LO, then Systolic HI and LO, and
2
HI and LO.
Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and the
HI or LO LED will
flash, indicating which limit is being
changed.
• Press the
Set button up or down to raise or
lower the alarm limit. (When you come to the
end of the range, the numbers will stop
changing).
• Press the
Select button again to go to the
next limit, or press it several times until none
of the measurements flash and no
HI or LO
LEDs are lit. The instrument is now in its
normal measurement mode. (If you do not
press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).
Pulse and Heart Rate alarms - There is one
case where the Heart Rate alarm receives status
information from the Oximeter pulse rate:
If ECG is inactive and the Heart Rate is shown
as dashes, the Heart Rate alarm is triggered by
the Pulse rate instead of the ECG Heart Rate.
If the Pulse rate falls outside the Heart Rate
limits, the Pulse measurement flashes and the
alarm sounds. Use the large blue
Silence button
to temporarily silence the alarm, and use the
SpO2 ALARMS Off button to suspend it.
11
Saving volume settings – You can save your
volume settings for alarms and for pulse tone
after you change them, so that they become the
initial settings every time the Atlas Monitor is
powered on. After making the alarm and pulse
volume changes, press Date/Time to display the
Other Options menu, and press Print (orFreeze) to save your settings. Press Date/Time
to return to the main screen. You may repeat this
whenever you want to change your settings.
12
2.3 Monitoring Heart Rate and the ECG Waveform
The trace displays one of the ECG leads. This lead is indicated above the right end of the trace, near the
♥
symbol and heart rate value. Choose the lead with the Lead Select button. Heart rate is displayed at
the right.
ECG displays – The ECG waveform is always
displayed on the upper half of the screen. The
ECG waveform cascades (continues) from the
upper half of the screen to also appear at the
bottom portion when other waveforms are not
selected in Advanced Configuration. The Heart
Rate is displayed above the right end of the top
waveform, near the ♥ symbol. The symbol for
the selected ECG lead is shown to the right of
the Heart Rate.
There is always a scale reference bar shown to
the left of the upper ECG waveform. This scale
bar has a height that represents a 1 mV signal.
The apparent height of the scale bar will vary
depending upon the ECG gain setting being
used, but will always correspond to a 1mV
signal.
Selecting leads - Press Lead Select to change
the lead display. The ECG function can use
either 3 wire leads - I, II, III , or 5 wire leads I, II, III, aVR, aVL, aVF, and V. 3 wire or 5
wire lead setting is selected in the Advanced
Configuration menu.
Note: When using 3 wire leads, the ECG lead
set must be set correctly. Incorrect results
and noisy waveforms can be obtained if the
system is configured for 5 wire leads when
using 3 wire leads.
Alarms - When either of the Heart Rate limits
are exceeded, an audible alarm sounds and the
affected measurement flashes. You can silence
this alarm, and all alarms, by pressing the large
Silence button at the right side of the
blue
instrument. This will silence all alarms for 60,
90, or 120 seconds, depending on the setting
selected in Advanced Configuration. However,
any measurement still outside the set limits will
flash.
To suspend the Heart Rate alarms, press the HR
ALARMS Off
button so the red LED in the
button lights. A suspended alarm will still flash
if it goes outside the range of the limits, but it
will not sound the audible alarm.
If ECG is inactive for any reason, the Heart Rate
display will be dashes “---” and the Heart Rate
alarm will respond to the pulse oximetry rate.
Setting alarm limits –To set the Heart Rate
alarm limits, use the
Select and Set buttons on
the left side of the monitor to follow these step s:
• Press the left Select button to cycle through
MAP HI and LO, Heart Rate HI and LO,
Respiration
623xx)
LO.
HI and LO (models 622xx or
, then (model 623xx) CO
HI and
2
• Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
HI or LO will flash,
indicating which limit is being changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
LO indicators are lit. The instrument is now
HI or
in its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).
13
Atlas Monitor showing ECG waveform display, controls, and ECG
connector
14
ECG settings in Advanced Configuration The five settings in the Advanced Configuration
menu associated with the ECG are listed in this
table:
Note: When using 3 wire leads, the ECG lead
set must be set correctly. Incorrect results
and noisy waveforms can be obtained if the
system is configured for 5 wire leads when
using 3 wire leads.
Advanced Configuration
ECGsettingPossible values
ECG gainautomatic |
10 mm/mV
ECG lead set3 wire | 5 wire
ECG speed6.25 | 12.5 | 25 mm/s
ECG bandwidthMonitor | Extended
Second trace
selection
You can change the ECG settings in Advanced
Configuration:
• Press the
• Press
Date/Time button.
Trend.
•Use either Select button to highlight ECG
gain, ECG lead set , ECG speed, ECG
bandwidth, or Second trace selection.
• Use the
Set button to change any of these
values.
• Press
Trend to exit Advanced Configuration
ECG gain - The height of the vertical ruler that
appears to the left of the ECG waveform
indicates a 1 mV amplitude and is 10 mm high if
10 mm/mV gain is chosen. When automatic
gain is selected the ruler height will vary, but it
will always indicate a 1 mV signal size. The
ruler size is automatically increased or decreased
to scale to a particular set of waves, but the
vertical line still indicates the same amplitude of
1 mV.
ECG | SpO2 |
Respiration | CO
2
ECG speed - The amount of ECG waveform
shown on the CRT is determined by the trace
speed. A slower trace speed means more
seconds of waveform are shown on the CRT.
ECG bandwidth - The ECG waveform can be
displayed and printed in either Monitor or
Extended bandwidth. Monitor mode allows for
a clearer picture of the waveform by filtering out
noise. Extended mode, usually used with
cardiac paced patients, shows the finer nuances
of ECG waveform, facilitating the detection of
conditions such as ischemia.
Note: Detection of ischemia is the interpretation
of the clinician only, the Atlas Monitor does
not provide automated ischemia detection.
Note: It is normal for the ECG baseline to
wander slightly in Extended bandwidth.
Pacemaker signals – The Atlas Monitor
displays pacemaker signals exactly as they are
captured. There is no option to display symbolic
pace indications. Use Extended bandwidth for
enhanced display of pacemaker signals.
ECG lead set - The Atlas Monitor allows for
both 3 wire and 5 wire ECG lead sets.
The lower trace can display the Impedance Respiration waveform if this is chosen in Advanced
Configuration. In this case, the Respiration Rate is displayed above the right side of the waveform. The
Impedance Respiration waveform is always derived from Lead I (RA-LA).
What is it? - Respiration Rate is measured with
the ECG leads. As the chest expands and contracts
during the respiration cycle, the resistance, or
impedance, between the RA-LA electrodes (
I
) changes. The result of these changes indicates
Lead
the respiration rate.
For best performance in monitoring impedance
respiration rate, change the LA and RA electrode
placement to the mid-axillary line on each side of
the chest as shown in the section on connecting
the ECG.
Where is it? - In Atlas Monitor model 622xx, the
lower trace normally shows cascad ing ECG . In
Atlas Monitor model 623xx, the lower trace
normally displays the ETCO
waveform.
2
However, the lower trace can instead show
Impedance Respiration, if it is chosen in
Advanced Configuration.
movement, making it less accurate than ETCO
2
for measuring the breath rate. For this reason,
model 623xx users often prefer to view the
ETCO
waveform and let the monitor measure
2
breath rate from this source.
Alarms - When either of the respiration rate limits
are exceeded, an audible alarm sounds and the
affected measurement flashes. You can silence
this alarm, and all alarms, by pressing the large
Silence button at the right side of the instru-
blue
ment. This will silence all alarms for 60, 90, or
120 seconds, depending on the setting selected in
Advanced Configuration. However, any
measurement still outs ide the set limits will flash.
To suspend the respiration rate alarms, press the
CO2/RE S P ALARM S Of f (RES P ALARMS Off)
button so the red LED in the button lights. A suspended alarm will still flash if it goes outside the
range of the limits, but it will not sound the
audible alarm.
How to display it – You can change the Second
trace selection settings in Advanced
Configuration:
• Press the
• Press
Date/Time button.
Trend.
•Use either Select button to highlight Second
trace selection.
• Use either
from the choices ECG, SpO
CO
2
• Press
Set button to choose Respiration
, Respiration,
2
(model 623xx).
Trend to exit Advanced Configuration.
Warning: Impedance Respiration rate
measurement and alarm capability are active
ONLY when the second trace option is set to
Respiration. Should the operator change from
viewing the Respiration waveforms and breath
rate to another selection (SpO
, CO2 or ECG)
2
the Respiration rate monitoring and alarm
capability will be disabled. This occurs even if
the ECG cable is still inserted into the Monitor.
Setting alarm limits –To set the Respiration
alarm limits, use the
Select and Setbuttons on
the left side of the monitor to follow these step s:
• Press the left Select button to cycle through
MAP HI and LO, Heart Rate HI and LO,
Respiration
only) CO
HI and LO, then (model 623xx
HI and LO.
2
• Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
HI or LO will flash,
the end of the range, the numbers will stop
changing).
• Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
LO indicators are lit. The instrument is now
HI or
in its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).
indicating which limit is being changed.
• Press the Set button up or down to raise or
lower the alarm limit. (When you come to
17
2.5 Monitoring Temperature (Models 622xx & 623xx)
The Temperature, measured on the skin surface with a skin sensor, is displayed in °F or °C, as chosen in
Advanced Configuration. There are no audible alarms for Temperature. An invalid temperature reading
is indicated by dashes “---” in the numeric display.
Temperature can be measured with a skin sensor.
No alarms – There are no temperature alarm
limits and no audible alarms for temperature. If
there is no temperature probe connected when the
monitor is first turned on, the TEMP display will
be blank.
If the probe becomes disconnected from the
patient or the monitor, the TEMP display will
show steady dashes “---”, but there will be no
alarm.
Changing the scale – The temperature display
can be in °F or °C, as selected in Advanced
Configuration.
You can change the Temperature units setting in
Advanced Configuration:
• Press the
• Press
Date/Time button.
Trend.
• Use either Select button to highlight
Temperature units.
• Use either
• Press
Set button to choose °
Trend to exit Advanced Configuration.
°F or °°°°C.
°°
Note: The temperature display is blank at power-
on until a temperature probe has been detected.
18
Atlas Monitor showing Temperature display and connector
19
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.