Welch Allyn Atlas User manual

Patient Monitor
Operator Manual
Welch Allyn Medical Division
oNtraPveR
noitpircseD#NCEetaDdevorppA
E34-0026A
launaM
ecivreSSALTAfoesaeleRweN
92404-599/01PL/SR
Drawings and/or illustrations and/or part numbers contained in this document are for reference purposes only. For current revisions call the Welch Allyn
Customer Service phone number listed in Section 1 page 2.
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS PATIENT MONITOR i

Contents PAGE

Section 1: General Information
1.1 About the Atlas ............................................................................................................ 1
1.2 T echnical Help Information, W orldwide T ech Support Phone Number........................... 2
1.3 Product Model Number Structure ................................................................................. 3
1.4 Main Menu Architecture ................................................................................................ 6
1.5 Atlas System Block Diagram ........................................................................................ 7
Section 2: Service (See User Manual 6200-42E for Atlas Specifications)
2.1 Incoming Inspection,Checklist ...................................................................................... 1
2.2 Repair Tests ............................................................................................................. 2-4
Table 2-1 :Tools Required for Service, Calibration and Software Loading ................... 5
Table 2-2:Software/Firmware Revision Levels ............................................................ 6
2.3 Calibration Procedures
BP Calibration - Section 2.3.1 ........................................................................... 7
ETCO2 Calibration (623Models only) - Section 2.3.2.........................................7&8
CO2 Reset (623Models only) - Section 2.3.3 ........................................................ 9
Printer Calibration - Section 2.3.4 ....................................................................... 10
Battery V oltage Calibration (Models 622xx and 623xx only)- Section 2.3.5 .... 10&11
T emperature Measurement Subsystem Calibration- Section 2.3.6 ...................... 12
Set Battery Charge V oltage-Section 2.3.7 .......................................................... 13
2.4 Software
Software Chart#2.4 .1.......................................................................................... 14
Downloading Operating System 2.4.................................................................... 15
2.5 Downloading NVRAM T ext files ............................................................................ 15&16
2.6 Downloading Software and NVRAM T ext files............................................................. 16
2.7 Product Numbering Structure .................................................................................... 17
Section 3: Troubleshooting (Also see Atlas Operator’s manual Appendix E)
3.1 Functional T ests/ Initial Diagnostic Steps ...................................................................... 1
3.2 T echnical T roubleshooting T ables: Complaint/Cause/Corrective Action ....................... 12
3.3 T op Level T roubleshooting Index .................................................................................. 15
3.4 Diagnostic T ests and Test Setups ............................................................................... 17
Section 4: Disassembly and Repair
About Section 4 .................................................................................................................... 1
4.1 Model 200 Dissassembly .............................................................................................. 2
4.2 Model 200 Re-assembly.............................................................................................. 13
4.3 Other Models notes ..................................................................................................... 15
ii WELCH ALL YN A TLAS P ATIENT MONITOR
SERVICE MANUAL 6200-43E REV. A
Appendix Section:
A...............................................................................................................theory of operations
B...................................................................................................................... repair parts list
C....................................................................................interconnect diagram 620396 Rev A
D...................................................................... electrical schematics and circuit board layouts
E...........................Safety T ests: Required Equipment and Procedures and Test Results Form
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS PATIENT MONITOR iii
Section 1 - General Information
The Atlas combines in one unit all the necessary mea­surements for patients under anesthesia, for surgical recovery , or bed side monitoring. See Section 1.3 for a complete listing of product models and options.
According to the standards of care for Nurse Anesthe­tists and Anesthesiologists, all patients receiving con­scious sedation are to be continuously monitored throughout the procedure and recovery phase by ECG, SpO2, and NIBP. CO2 monitoring is a requirement dur­ing gas anesthesia (when patient is ventilated).
The Atlas combines a CRT to display ECG and CO waveforms and LEDs for the other numeric values to maximize visibility and viewing angle. Although not designed to be a transport product, the monitor has an integral handle and it is small and light enough at 13 lb to be easily moved. A one hour battery enables the monitor to be moved with the patient from the surgery room to recovery room. It also maintains unit operation for up to an hour when power is interrupted.
1.1 About the Atlas Monitor
2
IMPORT ANT: for a complete description on the func­tion and use of the Atlas, as well as user safety warn­ings, cautions, and warranty information, read and understand the Atlas Operator’s Manual part number 6200-42E (English). Other languages are available.
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS PATIENT MONITOR 1
Section 1 - General Information
1.2

Help Information

Safety Warnings
All service and repairs must be performed by fully trained and properly equipped personnel, using genuine replacement parts and correct procedures. Failure to do
so will invalidate the product warranty.
Read and understand all safety warnings and service notes printed in this Service Manual and the Operator’s Manual part number 6200-42E. If in doubt about any precaution or procedure, for phone help, or to order additional copies of the Atlas Operator’s Manual, contact:
Customer Service Welch Allyn, Inc. 4341 State Street Road, PO Box 220 Skaneatles Falls, NY 13153-0220 U.S.A
Telephone 1-800-535-6663
When calling, refer to the model number on the bottom of the Atlas. The Model Number is the first three
digits of the Serial number number found on the bottom of the Atlas.
Treat all returned opened Nasal CO2 Sample Lines and watertraps as Bio Hazard material and dispose of them in an approved manner.
Troubleshooting assistance is contained in Section 3 of this manual to help determine which board is malfunctioning. This manual does not support repairing the printed circuit boards.
Year 2000 Information: The Atlas is Y2K compliant and will not encounter “Year 2000” problems.
2 WELCH ALL YN A TLAS P ATIENT MONITOR
SERVICE MANUAL 6200-43E REV . A
Section 1 - General Information
1.3 Product Model Number Structure
621S0 ECG, Nonin SpO2, NIBP
621SP ECG, Nonin SpO2, NIBP , Printer
622S0 ECG, Nonin SpO2, NIBP, T emp, Respiration, Battery , RS232
622SP ECG, Nonin SpO2 , NIBP, T emp, Respiration, Battery , RS232, Printer
622N0 ECG, Nellcor SpO2, NIBP , Temp, Respiration, Battery , RS232
622NP ECG, Nellcor SpO2, NIBP, T emp, Respiration, Battery , RS232, Printer
623SP ECG, Nonin SpO2, NIBP, ETCO2, T emp, Respiration, Battery , RS232, Printer
623NP ECG, Nellcor SpO2, NIBP, ETCO2, T emp, Respiration, Battery , RS232, Printer
SUFFIX:
Use letter designation for language localization as follows: E = English, F= French, G= German, I= Italian, S= Spanish, P= Portuguese C = Chinese, J= Japanese
Use number designation for line cord localization as follows:
1 = US, Canada, Japan V ersion 2 = European Version 4 = United Kingdom Version 6 = Australian V ersion
SERVICE MANUAL 6200-43E REV A WELCH ALLYN ATLAS PATIENT MONITOR 3
Section 1 - General Information
1.4 Main Menu Architecture
*
Also note that this is for the 623xx system.
On a 622xP system, delete the CO2 Reset box, the Calibrate CO2 item, and the CO
units item.
On a 621xP system, delete the CO2 Reset box, the Calibrate CO2 item, the CO
units item, Respiration speed item, the T emperature units item, and the Battery test item.
On systems without printers (621x0 and 622x0) delete the Printer test pattern item.
*
The Display A/D Channels also lets you press <SET> up and down to display additional sets of information, but does not change the menu page that you are on, just writes different information on the right side of the screen.
4 WELCH ALL YN A TLAS P A TIENT MONITOR SERVICE MANUAL 6200-43E REV A
2
2
Section 1 - General Information
Section 1.5 Atlas System Block Diagram
SERVICE MANUAL 6200-43E REV A WELCH ALLYN ATLAS PATIENT MONITOR 5
Section 2 - Service
o 2.1.1 Inspect shipping package and product
for damage. Make a record of possible shipping damage.
2.1.2 List accessories in box.
o power cord o SpO2 probe o ECG leads o ECG lead cable o Cuff(s) o Operator Manual o nasal C02 Line o ECG Electrode(s) o temperature probe o BP tubing o paper o loose parts, describe
2.1 Incoming Inspection
Service Intervals for Calibration and Main­tenance are listed in Atlas Operator’s Manual Appendix C.
o 2.1.3 Clean and disinfect by following the
instructions printed in the Operator Manual.
o 2.1.4 Operate the Atlas to verify the customer
complaint before making any changes to the unit. Call the customer if the complaint is unclear .
NOTE: Perform REPAIR TESTS in
Section 2.2 to fully inspect the Atlas monitor before and after servicing. Refer to Section 3 for Troubleshooting help.
2.1.5 If the unit has caused or is suspected of having caused an injury of any type:
DO NOT DISASSEMBLE OR REP AIR THE UN IT IN ANY WA Y. Contact Welch Allyn Customer Service immediately .
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 1
Section 2 - Service
2.2 Repair Tests 2.2.1 Blood Pressure System Leakage T est:
BP Leak Test
Unit must not leak more than 5 mmHg in a 15 second interval while attached to a 100 cc volume at test pressures of 50mmHg and 250mmHg.
Need:
100cc (+10/-0cc) test cavity stopwatch squeeze bulb
2.2.1.1 Connect Atlas to test 100cc volume and Setra Gage as shown in Figure 2.3.1.1
2.2.1.2 Turn Atlas ON
2.2.1.3 Press Time and Date button to access Options Menu.
2.2.1.4Press Lead Select button to select Service Mode. Select V erify Manometer
2.2.1.5Pressurize the Atlas with squeeze bulb to 50mmHg.
NOTE: Allow reading to stabilize for 15 seconds.
2.2.1.6Observe pressure for 15 additional seconds. Unit should not leak more than 5mmHg during this time.
2.2.1.7Perform this process at 250mmHg
BP Dump Test
Manometer Accuracy Test
2.2.2 Dump T est: Unit must be able to deflate a 500 cc volume from greater than 260.0 mmHg to less than
15.0 mmHg in 10 seconds or less.
2.2.3 Manometer Accuracy T est Internal temperature of the unit must be less than 32.0 degrees Celsius before performing test.
2.2.3.1 The primary transducer must be within +/-
0.75 mmHg at 0 +/-0.3 mmHg. The safety transducer must be within +/-1 mmHg at 0+/-0.3 mmHg.
2.2.3.2 The primary transducer must be within +/-
1.5 mmHg at 50 +/-5.0 mmHg. The safety transducer must be within +/-1.5 mmHg at 50 +/- 5.0 mmHg.
2.2.3.3 The primary transducer must be within +/-
1.5 mmHg at 150 +/-5.0 mmHg. The safety transducer must be within +/-4.5 mmHg at 150 +/- 5.0 mmHg.
2.2.3.4 The primary transducer must be within +/-
2 WELCH ALL YN A TLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.2.4 Deflation T est: With the unit connected to a large adult cuff and a target inflation pressure of 200 mmHg* have the unit perform a complete “normal” BP cycle. For steps two and three, the size of the step must be between 3.0 mmHg and 1 1.0 mmHg. Step 4 and all other steps above or equal to 40 mmHg, step size must be between 7.0 mmH gand 1 1.0 mmHg. All steps below 40 mmHg, except for the last step, will be between 4.5 mmHg and 10 mmHg. The last step will be between 0.01 mmHg and 10 mmHg.
2.2.5 Charge Voltage Test: Check battery charging circuit:
Specification :No Load: 6.85VDC across right pin (+) and left pin (-) of Main PCB battery charge connector .
Deflation T est
* Go to service mode screen and set initial
pressure to 200mmHg.
Charge Voltage Test
2.2.6 Hardware Fail Safe Tests
2.2.6.1 Over Pressure T est: Atlas hardware must detect over pressure on unit pneumatic system between 296.0 mmHg and 329.0 mmHg.
2.2.6.2 Over 15 mmHg T est: Atlas hardware must detect if the pneumatic system has been pressurized greater than 15 mmHg for more than 155 seconds but less than 180 seconds.
2.2.6.3 Under 15 mmHg Test: Atlas hardware must detect if the pneumatic system has been pressurized less than 15 mmHg for more than 25.0 seconds but less than 35.0 seconds before alowing new inflation cycle in “non­Stat” Auto Mode.
2.2.7 RS232 T est: The RS232 communication operation will be confirmed with successful serial transmit and receive.
Hardware Fail Safe Tests
RS232 Test
2.2.8 Printer Option T est:
Printer Option T est
The printer must be able to print out the test pattern . Inspect print quality .
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 3
Section 2 - Service
ECG Test
NOTE: Section 3 of this Service Manual contains troubleshooting steps for the ECG subsystem. These tests will help determine if the main board is faulty .
2.2.9 ECG T est: Use a calibrated simulator to check performance. NOTE: There is no calibration for the ECG or impedance Respiration subsystems. If the performance does not match up to that expected using a calibrated simulator then there could be a problem with the cable, leads, connectors, wiring or the main board itself. If the main board is faulty then replace it.
SpO2 T est
NOTE:There is no
calibration for the SpO2 and Pulse subsystem. If the performance does not match up to that expected using a calibrated simulator, or the Phantom finger set, then the subsystem board
must be replaced.
2.2.10SpO2 T ests:
Need:
Appropriate Phantom finger set SpO2 cuff and cable Or: Nellcor or Nonin (as fitted) simulator (replaces cuff to drive subsystem) Or: Calibrated SpO2 simulator that has a cuff
fitting that simulates a perfused finger
Nonin: Settings for NoninPatient Simulator 8000S are 98% O2 / 80 BPM. SpO2 board accuracy after 25 second stabilization period must be within +/-2%O2 and +/- 2BPM. Nellcor: Settings for Nellcor Patient Simulator SRC-2 are 81% O2 and 112 BPM. SpO2 board accuracy after 25 second stabilization period must be within +/-2% O2 and +/- 2 BPM.
Temperature Test
4 WELCH ALL YN A TLAS MONITOR
2.2.1 1TEMP ACCURACY VERIFICA TION
Need: Calibrated thermometer (DIGIT AL OR GLASS) small insulated container with cover for warm water
2.2.1 1.1Fill container wtih approximately 96 degree F water
2.2.1 1.2Attach temperature probe neaqr the sensing part of the thermometer and insert into the warm water .
2.2.1 1.3Accuracy must be within +/- 0.2 degrees F.
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
T able 2-2:Software/Firmware revision levels
LEDOMmetsySgnitarepOredaoltooBroclleNninoNnoyrP
OS12699/8/9,0004.10.AA99/02/6,0000.10.AA7V PS12699/8/9,0004.10.AA99/02/6,0000.10.AA7V OS22699/8/9,0004.10.AA99/02/6,0000.10.AA7V PS22699/8/9,0004.10.AA99/02/6,0000.10.AA7V ON22699/8/9,0004.10.AA99/02/6,0000.10.AA79/71/210.0.2.1V PN22699/8/9,0004.10.AA99/02/6,0000.10.AA79/71/210.0.2.1V PS32699/8/9,0004.10.AA99/02/6,0000.10.AA7V00.1VE.0 PN32699/8/9,0004.10.AA99/02/6,0000.10.AA79/71/210.0.2.1V00.1VE.0
6 WELCH ALLYN ATLAS MONITO R SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.3.1 BP Calibration : Need: 500cc vessel (approximate)
calibrated digital manometer squeeze bulb with one-way valve tubing and T fittings PC with HyperT erminal * serial cable
2.3.1.1 Connect: manometer, bulb, and 500cc vessel to BP port with “T” connectors, Atlas to PC with serial cable.
2.3.1.2 Turn AtlasON and start HyperTerminal on PC.
2.3.1.3 Enter commands on Serial interface:
Pangea> bp valve close<ENTER> Pangea> bp safety off<ENTER>
2.3.1.4 Enter command:
Pangea> bp cal 5000
Do not press <ENTER> yet!
2.3.1.5 Raise pressure with bulb to as close to
50.00mmHg as possible or slightly higher. Let the pressure bleed down toexactly
50.00mmHg. Now press <ENTER> . Take no more than 3 minutes for this step.
2.3.1.6 Enter command:
Pangea> bp cal 25000<ENTER>
2.3.1.7 Repeat step 2.3.1.5 with 250.00 as target.
Press <ENTER> when pressure deteriorates to
250.00 mmHG.
2.3.1.8 Enter command:
Pangea> nvram write<ENTER>
2.3.1.9 Disconnect serial cable and instruments and cycle power on Atlas. BP cal complete.
2.3.1.10 Verify accuracy of pressure settings by repeating step 2.3.1.3 then 2.3.1.5 and compare Atlas front panel reading with Setra readout. Do this at 50.00mm and 250 mm Hg.
2.3 Calibration
2.3.1 BP Calibration
Caution: Improper use, stor­age, handling of compressed gas vessels can cause injury or death! Follow gas manu­facturers safety processes.
500cc volume
Setra
Figure 2.3.1.1
*Access and set up HyperT erminal: 1-Start 2-Settings 3-Control Panel 4-Add/Remove Programs 5-Windows Setup 6-Communications 7-selectHyperT erminal 8-APPL Y
Settings are: 9600 Baud, 8 bit word, 1 stop bit no parity , no flow control ANSI character set Find HyperT erminal in Programs, Accessories
2.3.2 ET CO2 Calibration: 623 Models only Need: T ank of approximately 10% CO2, balance N
2
(certified) Blood Gas Mixture Tubing and T connectors Watertrap and scrubber
2.3.2.1 Insert water trap. Power up.
2.3.2.2 Place the instrument into Service Mode: Press Date/Time Lead Select button.
2.3.2.3 Select Calibrate CO
2
Message “Remove CO2 water trap” will appear.
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 7
2.3.2 ET CO2 Calibration
NOTE:Replace water trap every 6 hrs of use. Treat water trap and used NasalCO2 Sample lines as bio hazard material!
NOTE: The Scrubber looks
similar to a watertrap, but is filled with white granules. It is included with model 623XX.
Section 2 - Service
Note
approximately 2 PSI.
: Adjust regulator to
valve
2
CO
watertrap
Figure 2.3.2.11
vent
2.3.2.4 Remove water trap from water trap socket .
2.3.2.5 Message “Install scrubber” will appear .
2.3.2.6 Attach scrubber to water trap.
2.3.2.7 Insert scrubber/water trap assembly into water trap socket. Message “Enter span gas value using <Set> button” will appear.
2.3.2.8 Change the default value (10%) to the value of span gas being used (calibrated 8% to 12% concentration known to ±0.01%).
2.3.2.9 Press <BP Start/Cancel> Message: “Sampling” appears while the instrument is sampling the scrubber air.
2.3.2.10 When instructed, remove the scrubber from the CO2 water trap. Do not remove water trap.
2.3.2.1 1 Attach the calibrated source of CO2 gas to the CO2 side-stream sampling tube as per Figure 2.3.2.1 1
2.3.2.12 Adjust the CO2 regulator just enough to allow a small amount of gas to flow out of the vent (approximately 3 PSI).
NOTE:Replace water trap every 6 hrs of use. Treat water trap and used NasalCO2 Sample lines as bio hazard material!
SR856: CO2 Calibration beyond the year 2021
Important: Once positive flow is established, let the it flow for one minute before pressing the Continue button and actually sampling. After one minute, press the Continue button to let the Atlas start sampling CO
2.3.2.13 Press <BP Start/Cancel> Message: “CO2 calibration successful” or “CO2 calibration failed” will appear on th e CRT display .
(Note: Serial Communication is not the cause of a failure if the Prion Serial appears in the Service Screen. Therefore the problem is with the board or connections and not serial communication problems.)
2.3.2.14 Press Trend button to Exit.
T o calibrate the CO2 system in years 2022 or later it is necessary to reset the system clock to an earlier year , perform the calibration, and then set the clock to the correct year again. The service center should keep a record of this since the Cal display in the Service Mode screen will be incorrect.
2.
.
8 WELCH ALL YN A TLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.3.3 CO2 RESET Need: Watertrap and Scrubber
2.3.3.1 Press the Date/Time button on the lower right of the monitor. The Set Date and T ime and Other Options menu will be displayed.
2.3.3.2 Select the Reset CO2 selection by pressing the CO2/RESP ALARMS Off button. CO2 Reset screen will appear.
2.3.3.3 Follow the instructions on the screen. “Remove the CO2 watertrap”. “Install the CO2 Scrubber onto the CO watertrap.”
NOTE: The Scrubber looks similar to a watertrap,
but is filled with white granules. It is included with model 623XX.
2.3.3 CO2 RESET
2
2.3.3.4 Remove the tubing from the watertrap and attach the Scrubber to the watertrap.
2.3.3.5 Insert the watertrap/Scrubber combination into the watertrap socket.
2.3.3.6 Messages: The system will report that it is Sampling. The system will then report Reset Complete.
2.3.3.7 Remove the watertrap/Scrubber from the watertrap socket.
2.3.3.8 Detach the Scrubber from the watertrap and reattach the tubing.
2.3.3.9 Replace the watertrap in the watertrap socket.
2.3.3.10 Press Trend to return to the waveform screen.
NOTE: There may be a message indicating a 5 minute Warming Up period.
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 9
Section 2 - Service
2.3.4

PRINTER CALIBRATION

Note: Two lines are displayed: Waveform +128, Text +70 These two numbers are factory defaults and a good starting point if the system is printing very poorly or not at all. The LEFT<SET> button controls the Waveform setting, up and down. The RIGHT <SET> but­ton controls the Text setting, up and down. Make changes tothe settings as needed, where a larger number = darker printing and a smaller number = lighter printing. Make initial changes of about 10 points each time. After making an adjustment, press <TREND> to exit the Service Mode.
2.3.4 PRINTER CALIBRA TION Need: PC with HyperT erminal Serial cable Connector
2.3.4.1 Use HyperTerminal to get strobe width settings for printer normal for waveforms ­text for trend
Advanced configuration printing:
Pangea> nvram get
printer_strobe_width_normal<ENT>
(normal value 130)
Pangea> nvram get printer_strobe_width_text<ENT>
(normal value 70)
2.3.4.2 Change values - larger number for darker printing. Change with:
Pangea> nvram set printer_strobe_width_normal XX X Pangea> nvram set printer_strobe_width_text YYY
2.3.4.3 Test and reset until satisfied.
2.3.4.4 Press Trend button to exit the Service Mode.
2.3.5 BA TTER Y VOLTAGE CALIBRA TION
2.3.5 BA TTERY VOL T AGE CALIBRA TION for models 622XX and 623XX.
NOTE: CHECK BA TTERY VOL T AGE CALIBRA TION
AFTER REPLACING MAIN PCB. Need: DC power supply rated: 7 VDC at 5A Battery eliminator cable - Atlas battery plug on one end, interface to the power supply on the other. There should be access for voltmeter probes at the Atlas end of the cable when it is installed DMM / DVM with 10mV resolution on a 10V scale PC with HyperT erminal Serial cable and connector
2.3.5.1. Connect serial cable to PC and Atlas
2.3.5.2 Remove battery from Atlas
2.3.5.3 Set the power supply to 6.8V +/- 200mV
2.3.5.4 Connect the power supply to the Atlas battery connector .
2.3.5.5 Turn At las ON.
2.3.5.6 Reduce the power supply to 6.0V
10 WELCH ALL YN A TLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.3.5.7 Measure the voltage at the battery connector (at the Atlas) to the nearest 10mV .
NOTE: Do not measure at the power supply , since
cable resistance will introduce error.
2.3.5.8 At the HyperT erminal, type:
Pangea> power cal XXXX<ENTER>
(where XXXX represents the measured voltage in millivolts no decimal point.) For example, if you measured 6.010V at the battery connector , use the command “power cal 6010<ENTER>”.
2.3.5.8 The Atlas will respond:
raw = ZZZZ mV true = 6010 mV OK
(where ZZZZ is the raw uncalibrated reading that the instrument made.)
2.3.5.9 Reduce the power supply to 5.6 volts. Y ou
should soon hear the “low battery” alarm.
2.3.5.10Measure the voltage at the battery connector to the nearest 10mV .
2.3.5.1 1At the HyperT erminal, type:
Pangea> power cal XXXX<ENTER>
(where XXXX represents the measured voltage in millivolts no decimal point.) For example, if you measured 5.590V at the battery connector , use the command “power cal 5590<ENTER>”.
2.3.5.12
raw = ZZZZ mV true = 5590 mV OK
The Atlas will respond:
(where ZZZZ is the raw uncalibrated reading that the instrument made.)
2.3.5.13
Pangea> hw reset<ENTER>
2.3.5.14 Turn Atlas OFF and remove the power
supply .
2.3.5.15 Re-Install the battery .
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 11
Finish by typing: (this will re-boot A TLAS)
Section 2 - Service
2.3.6 TEMPERATURE MEASUREMENT
SUBSYSTEM
CALIBRATION
2.3.6 TEMPERA TURE MEASUREMENT SUBSYSTEM CALIBRA TION:
Need: PC Serial Cable HyperTerminal 1kOhm - 2kOhm 1% 1/2watt resistor Large RCA plug (1/4”) 2 conductor Short length of wire (optional) Soldering iron and solder
2.3.6.1Prepare Resistor/Plug assembly . Solder the resistor across RCA plug terminals.
2.3.6.2 Plug the resistor assembly into the Atlas T emperature jack.
2.3.6.3 Connect the Atlas to the PC serial port.
2.3.6.4 Turn the Atlas ON.
NOTE: T emperature display will show a valid temperature.
2.3.6.5 Start HyperTerminal on the PC.
2.3.6.6 At the Pangea prompt, type:
Pangea>temp cal xxxxxx<ENTER>
(where the xxxxxx is the value of the precision resistor in centiOhms- Specify the Ohms, tenths, and hundredths of Ohms with no decimal point- for example, a 1200.00 Ohm resistor would be: Pangea>temp cal 120000<ENTER>
2.3.6.7
Wait four seconds then type:
Pangea>temp state<ENTER>
2.3.6.8 Displayed resistance will be within 0.5 Ohms of the value that you entered. The Offset should not exceed 5.0 Ohms.
NOTE: The system will silently fail (without error) if it
is unable to calibrate properly . Therefore you must manually verify that this resistance is correct.
2.3.6.9 Turn the Atlas OFF.
2.3.6.10Unplug the resistor/plug assembly.
2.3.6.1 1Remove the serial cable.
2.3.6.12Check the accuracy of the
temperature sensing system using the process described in Section 2.2.1 1
12 WELCH ALL YN A TLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.3.7Set battery charging voltage: Specification :No Load: 6.85VDC Need: DVMM
2.3.7.1Remove battery from unit and unplug.
2.3.7.2Use DVMM to check across right pin (+)
and left pin (-) when viewed looking into the battery compartment.
2.3.7.3Adjust potentiometer R338 to obtain
6.85VDC. (It is located at the right of the battery jack. Turn it counterclockwise to increase the charging voltage.
2.3.7 Set Battery Charge Voltage
SERVICE MANUAL 6200-43E REV. A WELCH ALLYN ATLAS MONITOR 13
4X noitautiS noitidnoC noitcA sliateD decnavdA
11A
1B
1C
1D
noitarugifnoC
tesresu(
gnisusecnereferp
slortnoclenaptnorf
1E
ezilaitinI
gnisuMARVN
lanimretrepyh
1F
daolnwoD
metsysgnitarepo
swodniwgnisu
1G
1H
tarbilaCeR
?saltA
,rewoP,PB(
erutarepmet
)
OCtondna
2
1I
ecneuqeS
erawtfoS
edargdpU
Software Chart 2.4.1
22A
deliaf
33A
deliaf
44A
2B
draobniaM
3B
draobUPC
4B
onsahtinU
smelborp
2C
gnicalpertsuJ .draobniaM lanigiroehT
sidraobUPC
.KO
3C
gnicalpertsuJ .draobUPC lanigiroehT
sidraobniaM
.KO
4C
wenehtesU
ytilitudaolnwod
"exe.ld_salta"
niaMecalpeR
ezilaitini,draob
dna,marvn
.etarbilacer
UPCecalpeR
daoldnadraob
gnitarepo
.metsys
.deriuqer
2D
3D
4D
tonlanimreTrepyH
.noitarbilac
SEY
sdaol"exe.ld_salta"
llasetirwrevohcihw
sgnittesecnereferp
decnavdA(
2E
lanimreTrepyHdeeN dnamarvngnittesrof
3E
elifexe.ld_saltadeeN
4E
ON
eliftxt.nommoc_marvn
htiwsgnittessuoiverp
,riaperretfA.stluafed
resullakcehcesaelp
)uneMnoitarugifnoC sgnittesMRALAdna
yawehtmehttesdna
.deviecersawtinueht
2F
evobasaemaS..seY
SEY
3F
evobasaemaS..seY
ON
4F
SEY erawtfos
UPCni
2G
:ON gnitarepo
sawhcihwUPC
.dehcuotnu
3G
SEY
erawtfos
UPCni
4G
ON htiW
sedisernoitamrofni
2H
nisedisermetsys
3H
ON
sedisernoitamrofni
4H
exe.ld_salta ehtmargorp
noitarbilac
erastnatsnoc .degnahcton
SEY eht
ehtfoMARVN
niamwen
tonlliwdraob
tcerrocehtevah
noitarbilac
.stnatsnoc
nuR.1
exe.ld_salta
mralakcehC.2
rehtodnasgnittes
remusnoc
ecnereferp
decnavda
noitarugifnoc
.sgnittes
2I
MARVNteS-1 etarbilaceR-2
mralakcehC-3
sgnittes
3I
daolnwod-1
metsysgnitarepo
mralakcehC-2
rehtodnasgnittes
remusnoc
ecnereferp
decnavda
noitarugifnoc
sgnittes
4I
UPChtoB
niamdna
deliafdraob
Section 2 - Service
gnicalpeR
UPCehthtob
niaMdna
.sdraob
htobecalpeR
esu,sdraob
,exe.ld_salta
,erugifnocer
.etarbilac
.noitarbilac
lanimreTrepyHdeeN dnamarvngnittesrof
evobasaemaS..seY
SEY
SEY
erawtfos
UPCni
SEY eht
sedisernoitamrofni
ehtfoMARVN
niamwen
tonlliwdraob
tcerrocehtevah
noitarbilac
.stnatsnoc
sgnittes
exe.ld_saltanuR-1
marvnteS-2
etarbilaceR-3
mralates/kcehC-4
14 WELCH ALLYN ATLAS MONITOR SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.4 Downloading Operating System When required: to load latest revision software*
o na fully functioning Atlas or to reload software after replacing the MCU board. The atlas_dl.exe program loads the following files:
atlas.out.gz nvram_common.txt nvram_(model#).txt ** nvram_(language).txt **
Equipment or supplies required:
PC with Windows ‘95/NT Serial cable with connector: (COM1to PC) File : atlas_dl.exe
2.4.1 Connect serial cable between Atlas and PC COM1
2.4.2 Double click “atlas_dl.exe” explorer window
2.4.3 After downloading is complete, check alarm
and other User controlled advanced configuration settings since these are ‘reset’ by this downloading process.
2.4 Downloading Operat­ing System
(when upgrading software or
replacing CPU board)
**Note: Hyperterminal queries the Atlas to deter­mine which model number and language to down­load.
Note: If Atlas calibration was satisfactory prior to downloading software then recalibration is not required.
Downloading Operating System complete
2.5 Downloading NVRAM files with Hyperterminal*. The NVRAM resides on Main Board. Hyper terminal loads the following files:
cal_init.txt common.txt (model#).txt ** (language).txt,** printer .txt (if required), or no_printer.txt.
2.5.1Connect serial cable between Atlas and
PC COM1 as in 2.4 above.
2.5.2Open HyperT erminal program on PC:
ST ART PROGRAMS ACCESSORIES HYPERTERMINAL
2.5.3Turn Atlas ON. You should see some
version information and a prompt:
Pangea>
2.5.4 Transfer Text files:
(TRANSFER/SEND TEXT FILE) NVRAM_CAL_INIT .TXT (only if main board has been replaced)
2.5 Downloading NVRAM T ext files
after replacing Main Board
*Configure HyperT erminal
9600 Baud 8 bits 1 stop bit no parity no flow control
SERVICE MANUAL 6200-43E REV . A WELCH ALLYN ATLAS MONITOR 15
Section 2 - Service
2.5 Downloading NVRAM T ext files
when replacing Main Board
continued
2.6 Downloading Soft-
ware &
NVRAM T ext files
after replacing Main Board
AND CPU board
NVRAM_COMMON.TXT NVRAM_NO_PRINTER.TXT
or NVRAM_PRINTER.TXT NVRAM_(model 200,210,220).TXT NVRAM_<LANGUAGE>.TXT
2.5.5 After these text files are transferred, T ype the following serial commands to update the serial number:
nvram set serial___ (last 3 digits of serial#) nvram write hw reset (hardware reset restartsAtlas and
saves the new settings)
2.5.6 Verify the serial number correct
2.5.7 Recalibrate Atlas unit. After calibration is complete.
2.5.8 Check alarm settings and User selected
advanced Configurations
2.6 Downloading Software & NVRAM T ext files
2.6.1 Use atlas_dl.exe as in step 2.4. to Program ne w CPU board.
2.6.2 Use HyperTerminal as in step 2.5 to set NVRAM on new Main Board.
2.6.3 Recalibrate Atlas
2.6.4 Check alarm settings and User selected advanced Configurations
16 WELCH ALL YN A TLAS MONITOR
NOTE: Firmware Download
Software versions of OEM boards are not field upgradeable. (SpO2 boards or the CO2 board.) Replace the subsystem board with a higher (current) version if necessary .
SERVICE MANUAL 6200-43E REV. A
Section 2 - Service
2.7 Product Model Number Structure
621S0* ECG, Nonin SpO2, NIBP
621SP* ECG, Nonin SpO2, NIBP , Printer
622S0** ECG, Nonin SpO2, NIBP , Temp, Respiration, Battery, RS232, Nurse Call
622SP** ECG, Nonin SpO2, NIBP , Temp, Respiration, Battery, RS232, Printer , Nurse Call
622N0** ECG, Nellcor SpO2, NIBP , Temp, Respiration, Battery, RS232, Nurse Call
622NP** ECG, Nellcor SpO2, NIBP , Temp, Respiration, Battery, RS232, Printer , Nurse Call
623SP** ECG, Nonin SpO2, NIBP , ETCO2, T emp, Respiration, Battery , RS232, Printer , Nurse Call
623NP** ECG, Nellcor SpO2, NIBP , ETCO2, T emp, Respiration, Battery , RS232, Printer , Nurse Call
SUFFIX:
Use letter designation for language localization as follows: E = English, F= French, G= German, I= Italian, S= Spanish, P= Portuguese C = Chinese, J= Japanese
Use number designation for line cord localization as follows:
1 = US, Canada, Japan Version 2 = European Version 4 = United Kingdom Version 6 = Australian Version
Specifications: See Operator Manual
Specifications for all of the above listed models of Atlas including performance, accuracy , range, size, weight, power, environmental, are documented in an appendix to the User Manual.
* Model 200 Main Board and Schematic 200 ** Model 220 Main Board and Schematic 220
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS MONITOR 17
Section 3 - Troubleshooting
3.1 Functional Test and Initial Diagnostic
1.Review customer complaint and determine if it is safe to plug in and turn on Atlas
2.Plug in Atlas, no sensors attached.
3.Check for AC~ LED lit.
4.Install paper in printer if fitted.
5.Turn on power
6.Green light in power button.
7.Loud beep when button pressed.
8.Three dashes in SYSTOLIC.
9.Three dashes in DIASTOLIC.
10.Two dashes in SpO2 (takes several seconds after BP dashes come on).
11.Three dashes in PULSE (takes several seconds after BP dashes come on).
12.Pleth: none, or a single bar at the bottom, or two bars at the bottom.
13.No lights in TEMP.
14.No lights in ALARMS OFF buttons.
15.X lit on AUTO.
16.AC~ lit
17.CRT display comes on slowly if cold, quickly if still warm from last use.
18.May see version string in center if comes on quickly, not a problem if not seen because it comes on slowly.
19.CRT: three dashes for Heart Rate
20.Heart picture
21.Lead select symbol
22.Scale bar
23.One or two lines of dashes for waveforms – depends upon settings.
24.Three dashes for MAP – or blank, depends upon settings.
25.Error message(s) at bottom of screen?
Note: “Idle” screen is normal waveform display.
26.If 622 or 623, pull AC cable. Should be no change except AC~ unlit.
26.1. If errors of low, very low or depleted battery,
or if system dies, plug back in and repeat test in 2 hours.
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS MONITOR 1
Section 3 - Troubleshooting
3.1 Functional Test and Initial Diagnostic
Plug AC back in.
MENU TESTS
27.Press <DA TE/TIME><TREND>
28.Get Advanced Configuration menu.
29.Write down all settings for resetting to customer prefer­ence later.
30.Set language to your native tongue if necessary to allow you to write down the other settings. The top item is always the language, press either <SET> button to step through list.
31.Press <PRINT> if printer fitted.
32.Press <TREND> to return to idle screen.
33.Press <DA TE/TIME>
34.Get date/time menu.
35.V erify date/time, set if necessary .
35.1. Bad date may indicate battery problem. If
date was bad, turn off unit, pull power cable, wait 5 minutes reconnect power cable, turn on unit. Check date again.
35.1.1. If date comes back bad: Replace main board Model 200. Model 210, 220, if battery not dead replace main board.
Press Date/Time to return to waveform screen if needed.
36.Press <DATE/TIME><LEAD SELECT>
37.Get Service Mode menu.
38.Examine version/configuration data in lower half of menu, and write it all down.
39.Press <SELECT> to highlight Reset to factory de-
faults.
40.Press <BP START/CANCEL> to reset configuration.
41.Press <TREND> to return to idle screen.
42.Press <DA TE/TIME><TREND>
43.Get Advanced Configuration menu.
44.Set language to your native tongue if necessary
44.1. We have just reset to factory defaults.
Compare settings to factory defaults appropri­ate for the country – in Operator manual. Ifno t matching, indicates memory problems. Changing only the language should not change any of the other factory default settings.
45.Press <DATE/TIME><LEAD SELECT>
2 WELCH ALL YN ATLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Section 3 - Troubleshooting
46.Get Service Mode menu.
47.Press <Select> to highlight CRT test pattern.
48.Press <BP START/CANCEL> to show test pattern. Examine display .
49.Press any key to end display .
50.Press <Select> to highlight Printer test pattern (if fitted).
51.Press <BP START/CANCEL> to start test pattern.
52.Printer should print test pattern.
53.Press any key to end display .
54.Examine printout.
55.Press <Select> to highlight LED test .
56.Press <AUTO> to turn on all LEDs .
57.Press <BP START/CANCEL> to show automatic test pattern.
58.Watch for a while, look for glitches in pattern.
59.Press <Set> to go to manual mode and step through individual segments if needed to observe a problem.
3.1 Functional Test and Initial Diagnostic
60.Press <Select> to highlight Button test.
61.Press <BP START/CANCEL> to start test.
62.Press every button on system, <BP ST ART/CANCEL> last.
62.1. Verify that buttons match up with their names,
and that all buttons are functional.
62.1.1. If names don’t match, indicates memory corruption: Replace main board
63.Press <Select> to highlight Display A/D channels (three or four screens worth)
63.1. Write down all values for later review.
63.2. Press <Set> and write down all values for
each screen.
64.Press <Trend> to return to Idle screen.
65.Connect the BP port to the BP simulator.
66.Set the simulator for a normal reading (140/80, 100BPM, NSR).
67.Press <BP START/CANCEL>
67.1. System should start pump, display manom-
eter value in SYSTOLIC LED; this value should track and be very close to pressure
BP test
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS MONITOR 3
Section 3 - Troubleshooting
3.1 Functional Test and Initial Diagnostic
displayed by manometer in BP simulator (if fitted). Largest number shown in SYSTOLIC should be very close to the Initial pressure setting recorded above from Advanced Configuration.
67.2. System should step down pressure, showing step values in SYSTOLIC LED, and then dis­play correct SYSTOLIC and DIASTOLIC values. System may show MAP value de­pending upon country .
68.Press <DA TE/TIME><TREND> to get to Advanced Configuration menu.
69.Press <Select> to highlight Initial pressure.
70.Press <Set> to change Initial pressure to 280 mmHg.
71.Press <Select> to highlight MAP.
72.Press <Set> to change MAP to Yes.
73.Press <TREND> to return to idle screen.
74.Press <BP START/CANCEL>
75.System should start pump, display manometer value in SYSTOLIC LED; this value should track and be very close to pressure displayed by manometer in BP simulator (if fitted). Largest number shown in SYS­TOLIC should be very close to the Initial pressure setting of 280 mmHg.
75.1. If pressure shown exceeds 300 mmHg:
Recalibrate BP
76.System should step down pressure, showing step values in SYSTOLIC LED, and then display correct SYSTOLIC and DIASTOLIC values. System should show MAP value. MAP value should match what is shown by simulator.
77.Set simulator to highest Systolic <=250, lowest Dias­tolic >= 30, and lowest heart rate >=30.
78.Press <AUTO>
79.X goes unlit, 1 flashes for 10 seconds.
80.20 seconds after 1 stops flashing, BP measurement starts.
81.BP reading as above.
81.1. If BP does not start: Replace main board
81.2. If BP measurement incorrect recalibrate
4 WELCH ALL YN ATLAS MONITOR
SERVICE MANUAL 6200-43E REV. A
Loading...
+ 123 hidden pages