Welch Allyn AT-5 Installation Guide

AT-10
FAA Transmission Set-Up Instructions
Push the MEMORY button Select 2 Settings The following window will appear:
1 Select data for resting ECG 2 Select data for exercise ECG 3 Configuration of transmission 4 Phone No. Modem 5 Init. Modem 6 Further settings
Select 1: select data for resting ECG Make sure the unit is configured the following way:
Select records for transmission: Interpretation <1>: yes Intervals and axis <2>: No Measurements <3>: yes Average cycles (orig. data) <4>: No Lead R1: V1 (compressed data) <5>: No Lead R2: II (compressed data) <6>: No 12 leads (compressed data) <7>: No 12 leads (original data) <8>: No ECG (SCP-ECG) <9>: yes
Use the Yes + or No- keys on the AT-10 keypad to toggle between Yes and No
Use the Up & Down arrow keys on the AT-10 keypad to move from one field to the another
Press ESC to get out of this menu Select 3 Configuration of transmission Make sure the unit is configured the following way:
Kind of transmission <1>: Modem: protocol =CCITT
-----------------Mode setting RS-232/422------------------­Channel: <2>: RS-232 Baud <3>: 38400 Parity <4>: No Stop <5>: 1
Press ESC to exit this menu Select 4 Phone No. Modem Select T for Tone Enter 9 if you need an outside line and place a comma after 9 to pause the modem when dialing out Enter FAA phone number (405) 954-1971. See chart. Press Enter to save Press ESC to exit this menu
Enter one of the following options:
Long Distance: 14059541971 Long distance plus outside line:
9,14059541971
Local call plus outside line:
9,9541971
Local call: 9541971
71023-0001A Page 1 of 2
Taking an ECG for FAA
Make sure the following information is entered: Pat-Name: Firstname-Lastname. Place a – between the first and last name. Can only be a one-word entry.
44 Pat- No: Required Field. Use pilot SSN. The format can be either “xxx-xx-xxxx” or “xx xxx xxxx”
Born: Enter the date of birth. The format has to be either “mm-dd-yy” or “mm dd yy” Sex: M or F Height: in inches Weight: in lbs. BP: systolic/diastolic Med: Free text for medication
44 Rem: Required Field. Enter the doctor’s Aviation Medical Examiner Number (AME). Up to 5 digits.
Single ECG Transmission
Push the SEND DATA button after having acquired an ECG A “data is output” message will appear if the transmission was successful.
Batch Data Transmission
Enter patient demographics Take an ECG recording
Press STORE DATA after each test
To send a batch
Press MEMORY Select <1> List /Select. Press ENTER in the Data Search window A window with all patient records will open. Select 2 to transmit Select “Transmit all files” Press ENTER
Note: if you want to send selected files, highlight each file and press ENTER. A * will be placed to the left of the file indicating it has been selected. Select 2 to transmit and “transmit file(s)”. Press ENTER. The system will send one file at a time, but will re-connect automatically until all files are sent. When the transmission is done, a “data is output” message will appear above the leads V1-V6 box.
To abort transmission
Press Send Data Select <1> Quit transmit mode to quit OR Select <2> Continue transmit to resume transmitting A “Transmit Mode Ending” message will appear above the leads V1-V6 box.
Press Memory Select 1 List/select Select 2 Transmit Select “Transmit remaining files” using down arrow key Press Enter.
To resume transmission
71023-0001A
Page 2 of 2