Weinmann MEDUMAT Standard a User manual

MEDUMAT Standard a
Ventilator
Description and instructions for use
EN
Contents
1. Overview . . . . . . . . . . . . . . . . . . . . 4
1.2 Symbols used on the ventilator . . . . 7
2. Description of ventilator . . . . . . 10
2.1 Uses . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 Owner/operator and user
qualification . . . . . . . . . . . . . . . . . 11
2.3 Ventilation function . . . . . . . . . . . 11
2.4 Controlled ventilation. . . . . . . . . . 13
2.5 Assisted ventilation. . . . . . . . . . . . 13
2.6 Check ventilation curve . . . . . . . . 14
2.7 Patient valve. . . . . . . . . . . . . . . . . 14
2.8 Modules. . . . . . . . . . . . . . . . . . . . 15
3. Safety instructions . . . . . . . . . . . 16
3.1 Safety regulations. . . . . . . . . . . . . 16
4. Installation . . . . . . . . . . . . . . . . . . 20
4.1 Wall mounting for STATION
MEDUMAT. . . . . . . . . . . . . . . . . . 20
4.2 Installation kit for the wall
mounting . . . . . . . . . . . . . . . . . . . 21
4.3 Connecting up the oxygen
cylinder . . . . . . . . . . . . . . . . . . . . 21
4.4 Ventilation hose . . . . . . . . . . . . . . 22
5. Using the ventilator . . . . . . . . . . 25
5.1 Switching on/self test . . . . . . . . . . 25
5.2 Selecting the ventilation settings. . 26
5.3 Select ventilation method . . . . . . . 28
5.4 Performing ventilation . . . . . . . . . 30
5.5 Monitoring ventilation . . . . . . . . . 30
5.6 Ventilation with PEEP Valve . . . . . 31
5.7 Ventilation with HME filter . . . . . . 32
5.8 Ventilating with bacteria filter. . . . 32
5.9 Terminating ventilation. . . . . . . . . 33
5.10 Alarm signals . . . . . . . . . . . . . . . . 33
5.11 Calculation of oxygen content/
remaining operating time . . . . . . . 37
5.12 Alternative ventilation
procedures . . . . . . . . . . . . . . . . . . 38
6. Hygienic preparation . . . . . . . . . 39
6.1 MEDUMAT Standard a. . . . . . . . . 39
6.2 Patient valve . . . . . . . . . . . . . . . . 40
6.3 Hose system . . . . . . . . . . . . . . . . 41
6.4 Components and accessories . . . . 42
6.5 Fittings . . . . . . . . . . . . . . . . . . . . 43
6.6 Cleaning, disinfecting and
sterilizing . . . . . . . . . . . . . . . . . . . 44
7. Functional checks . . . . . . . . . . . . 45
7.1 Preparation for functional check . 45
7.2 Obligatory checks . . . . . . . . . . . . 46
7.3 Check for leaks in the system. . . . 47
7.4 Check of patient valve . . . . . . . . . 48
7.5 Checking the minute volume . . . . 49
7.6 Check of maximal ventilation
pressure. . . . . . . . . . . . . . . . . . . . 51
7.7 Check assisted ventilation . . . . . . 52
7.8 Check of alarm systems . . . . . . . . 54
8. Troubleshooting . . . . . . . . . . . . . 57
8.1 Batteries . . . . . . . . . . . . . . . . . . . 59
8.2 Cut-out system . . . . . . . . . . . . . . 60
8.3 Adjustment of manometer. . . . . . 61
8.4 Change valve membrane in
patient valve . . . . . . . . . . . . . . . . 61
9. Servicing . . . . . . . . . . . . . . . . . . . . 63
9.1 Intervals. . . . . . . . . . . . . . . . . . . . 63
9.2 Sending in device. . . . . . . . . . . . . 64
9.3 Storage . . . . . . . . . . . . . . . . . . . . 64
9.4 Disposal. . . . . . . . . . . . . . . . . . . . 65
10. Supply schedule . . . . . . . . . . . . . . 66
10.1 Standard supply schedule . . . . . . 66
10.2 Accessories . . . . . . . . . . . . . . . . . 66
10.3 Spare parts . . . . . . . . . . . . . . . . . 67
11. Technical data . . . . . . . . . . . . . . . 68
11.1 Device . . . . . . . . . . . . . . . . . . . . . 68
11.2 Patient’s hose system. . . . . . . . . . 70
11.3 Pneumatics . . . . . . . . . . . . . . . . . 71
11.4 Resistance to interference . . . . . . 72
11.5 O
content when using Air Mix . . 72
2
2 Contents
EN
11.6 Switching from Air Mix to
No Air Mix . . . . . . . . . . . . . . . . . . 73
12. Warranty . . . . . . . . . . . . . . . . . . . . 74
13. Declaration of conformity . . . . . 74
Contents 3
EN

1. Overview

Control panel MEDUMAT Standard a
1 Ventilation pressure gauge
7 Colour code
Toddler (yellow, 10 kg to 30 kg) Child (orange, 30 kg to 60 kg) Adult (brown, 60 kg to 110 kg)
4 Air Mix/No Air Mix switch
3 Alarm acknowledgement2 Alarm panel
8 ON/OFF switch
9 Ventilation frequency regulator
5 Minute volume regulator
10 Max. ventilation pressure
regulator
6 On/Off switch,
assisted ventilation

1.1 Device

Stenosis Disconnection < 2,7 bar O
2
No assist
MEDUMAT Standard a
Air Mix
No Air Mix
4 Overview
EN
MEDUMAT Standard a connections
12 Catch for STATION MEDUMAT
wall mounting
11 Pressure gas connection
13 Connection for ventilation tube
14 Pressure gauge hose connection
15 Relief valve
16 Dust cover
17 Mixed air filter
12 Catch for STATION MEDUMAT
wall mounting
I
Overview 5
EN
MEDUMAT Standard a device combinations
19 Ventilation hose
18 Ventilation mask
21 Filter
23 Patient valve
24 PEEP valve
22 Hose casing
25 Tube
or
20 Pressure gauge hose
Disposable hose system
6 Overview
EN

1.2 Symbols used on the ventilator

Position
> PSU< 134°
Patient valve
The symbol on the patient valve indicates that the lip and valve membranes in the expiration and spontaneous breathing arms must be changed im­mediately if they are crinkled, sticky or misshapen. Under no circumstances continue to use the patient valve for ventilation in this case, as malfunctions are likely (note “7.4 Check of patient valve“ on page 48).
The symbol indicates the correct position for insertion of the lip membrane.
When connecting the patient valve, take care to ensure that the direction of respiratory gas flow is correct.
Overview 7
EN
MEDUMAT Standard a
1
2
3
4567
1 Inlet 2,7 - 6 bar O2.
2 Tube system connection
3 Maximum pressure 100mbar
I
4
5 Servicing label: indicates when the next service is due.
8 Overview
Safety check and servicing label
Safety check label: (in Germany only) marks when the next safety check as per §6 of the German law relating to users of medical devices is required.
EN
MEDUMAT Standard a device information plate
SN
6
IP24
Serial number of device
Date of manufacture
Direct voltage
3,6 V lithium battery
Do not dispose of device in domestic waste
CE symbol (confirms that the product conforms to the applicable European directives)
Type BF application part
Degree of protection – against the ingress of solid particles – against access to hazardous parts – against the ingress of water with a harmful effect
Other markings
7 Follow instructions for use
Overview 9
EN

2. Description of ventilator

2.1 Uses

The MEDUMAT Standard a is an automatic (short­term) ventilator with the option of assisted ventila­tion.
You can use MEDUMAT Standard a:
to revive patients at the site of the emergency
on a longer term basis in more protracted emer­gencies, e.g. fires.
You can use MEDUMAT Standard a while transport­ing patients:
between the various rooms and departments of a hospital;
between the hospital and other premises;
in emergencies;
when transport over a considerable distance is planned.
10 Description of ventilator
MEDUMAT Standard a:
is designed to provide controlled ventilation to persons of approx. 10 kg body weight or more, or in the case of assisted ventilation, of approx. 15 kg body weight or more.
is used to treat respiratory arrest;
can be preset to parameters that ensure evenly balanced ventilation provided that the selected max­imum ventilation pressure P
can be supplied with additional modules for aspiration and oxygen inhalation. (N.B.
is not exceeded.
max
EN
MEDUMAT Standard a cannot be used as a ven­tilator simultaneously with these modules.)

2.2 Owner/operator and user qualification

As an owner/operator or user, you must be familiar with the operation of this medical device. Observe the legal requirements for operation and use (in Ger­many, the regulations governing owner/operators of medical devices apply in particular). Basic recom­mendation: get a person authorized by Weinmann to provide you with proper instruction about the handling, use and operation of this medical device.

2.3 Ventilation function

MEDUMAT Standard a operates within a pressure range of 2.7 to 6 bar and at a flow rate of not less than 70 l/min O
It uses high-pressure, medicinal-grade oxygen. An external pressure reducer brings this down to the re­quired operating pressure. The oxygen supply is fed in at input valve.
Both the infinitely variable ventilation frequency and the inspiration/expiration ratio of 1:1.67 in the case of controlled ventilation are regulated by internal electronic control mechanisms.
. It has an in-built power pack.
2
The gas for inspiration is routed to the patient through the ventilation tube via the patient valve and the ventilation mask or hose. A lip membrane
Description of ventilator 11
EN
Air Mix
No Air Mix
Air Mix
No Air Mix
in the patient valve ensures that the expiration gas can be exhaled through the expiration arm.
Regardless of the ventilation mode selected, the pa­tient has the option of breathing spontaneously be­tween ventilation strokes via the patient valve. In this case, the patient draws air for breathing from the ambient air.
With the Air Mix setting, in the case of mechanical ventilation, atmospheric air is admixed to give an O concentration generally of between 55 % and 85 % at 10 mbar ventilation pressure (note “11.5 O
con-
2
tent when using Air Mix“ on page 72).
In certain indications and in cases where the surround­ing atmosphere is contaminated, you can switch to
No Air Mix and ventilate with pure oxygen.
The injector unit is switched off when switching from
Air Mix to No Air Mix. This increases minute vol-
ume which can result in the set pressure limit being
exceeded and a stenosis alarm (Stenosis) being trig-
gered. In this case, set minute volume correspond­ingly
lower.
In the opposite instance, in other words when switching from
No Air Mix to Air Mix, the injector
unit is switched on. This reduces minute volume which can lead to the set pressure limit being
. In this case, set minute volume correspondingly
shot higher.
under-
2
12 Description of ventilator
EN

2.4 Controlled ventilation

Mandatory ventilation stroke: the device, not the patient, determines the time of the next breathing stroke.
After being switched on, the MEDUMAT Standard a is automatically in Controlled Ventilation mode. This administers mandatory ventilation strokes to the in­tubated patient according to the ventilation values set on the device.

2.5 Assisted ventilation

In addition to Controlled Ventilation mode, the MEDUMAT Standard a also has an Assisted Ventila­tion mode.
After you have switched on Assisted Ventilation mode by pressing the LED indicates this mode.
Triggered ventilation stroke: the patient can trig­ger a ventilation stroke by his own breathing efforts.
IPPV: intermittent positive pressure ventilation (= con­trolled ventilation).
The patient now has the option of triggering a trig­gered ventilation stroke within a time window of 40 % of expiration. To do so, the patient must generate a flow of over 6 l/min. by his own breathing efforts.
If the breathing efforts of the patient are not suffi­cient to trigger, the patient automatically receives a mandatory ventilation stroke at the end of the time window, so that the set minute volume is guaran­teed.
With this function, the ventilation strokes of the de­vice can be synchronised with the breathing efforts of the patient. Between the mandatory ventilation strokes of the device, the patient has the option of breathing in air from the surrounding atmosphere via the patient valve.
Assist key, a flashing green
Description of ventilator 13
EN
If the patient does not trigger the device, an alarm is
Respiration hose connection
Expiration tube
Connection for pressure gauge tube
Mask/tube connection
Sponta­neous breath­ing tube
triggered. The patient continues to receive con­trolled ventilation.

2.6 Check ventilation curve

The ventilation curve is checked at ventilation pressure gauge.

2.7 Patient valve

The gas for inspiration is channelled into the patient’s airways through the patient valve.
It is designed so that spontaneous breathing is pos­sible, even if the MEDUMAT Standard a fails, regard­less of which ventilation mode you have selected.
14 Description of ventilator
EN

2.8 Modules

M
OD
UL
O
x
y
g
e
n
O
2
0
5
1
0
1
5
l/min
O
2
MODUL Combi
O
2
Modules with additional functions can be attached to MEDUMAT Standard a.
Please refer to the directions for use enclosed with the modules for exact details of how to fit and oper­ate these. It is essential to read these directions care­fully before using the modules. The most important points are listed below:
Oxygen MODULE
The Oxygen MODULE enables you to apply oxygen inhalation.
Put the switch marked
O
into the “I“ position. Se-
2
lect the desired oxygen volume by turning the knob marked
l/min to a setting between 0 and 15 l/min.
You can check this setting on the volume mano­meter.
Combi MODULE
The Combi MODULE enables you to apply both oxygen inhalation and suction.
For inhalation put the switch marked position. Select the desired oxygen volume by turning the knob marked
l/min to a setting between 0 and
15 l/min. You can check this setting on the volume manometer.
For suction switch the tumbler marked position. The suction pressure is locked at -0.5 bar.
O2 into the “I
Vac to the “I
Description of ventilator 15
EN

3. Safety instructions

3.1 Safety regulations

For your own safety, the safety of your patients, and to comply with the requirements of EU Directive 93/ 42/EEC, please observe the following points:
General
Please read the directions for use carefully. They are an integral part of the ventilator and should be available for reference at all times.
Before starting to work with MEDUMAT Standard a, you must understand how to operate it.
Please comply with section “6. Hygienic prepara­tion“ on page 39 to prevent infection or bacte­rial contamination.
MEDUMAT Standard a should be used only by medically qualified personnel who have had training in ventilation techniques. Incorrect use can cause severe physical injury.
16 Safety instructions
It is advisable for you to have servicing and re­pairs carried out only by the manufacturer, Weinmann, or by qualified technicians expressly authorized by Weinmann.
If third-party items are used, functional failures may occur and fitness for use may be restricted. Biocompatibility requirements may also not be met. Please note that in such cases, any claim under warranty and liability will be voided if nei­ther the accessories nor genuine replacement
EN
parts recommended in the instructions for use are used.
MEDUMAT Standard a should be used only for the purposes for which it is designed (note “2.1 Uses“ on page 10).
MEDUMAT Standard a is not designed for use under hyperbaric conditions (pressure chamber).
MEDUMAT Standard a should never be used with flammable anaesthetics.
In the case of use in poisoned or low-oxygen atmospheres, do not operate the MEDUMAT Standard a with the „Air Mix“ setting or in Assist mode.
A back-up ventilator should always be available in case of technical failure.
Warning: Modifications may not be made to the device.
Have modifications to the device carried out only by the manufacturer, Weinmann, or by specialist staff expressly authorized by the manufacturer.
Oxygen
Highly-compressed oxygen can lead to spontaneous explosive reactions in combination with flammable substances (fat, oil, alcohol, disinfectants, etc.):
All screw connections and other components of the ventilator must be kept absolutely free of oil and grease.
Always wash your hands before starting to work on the oxygen supply.
Smoking and open flames are strictly prohibited in the vicinity of all fittings containing or trans­porting oxygen.
During assembly and when changing the oxygen cylinder, only hand pressure should be used
Safety instructions 17
EN
when tightening the screw connections to the cylinder and to the pressure reducer. Never use tools for this purpose. Excessive tightening damag­es the screw threads and seals and can cause leaks.
Protect oxygen cylinders from accidental falls. If a cylinder falls, the pressure reducer or the valve may break off and cause a violent explosion.
Important note Always open the valve of the oxygen cylinder
slowly to prevent pressure damage to the other fittings.
The oxygen cylinder should never be completely emptied as this may allow moisture-containing air to enter the cylinder and cause corrosion.
Operation
Both the patient and the ventilator must be kept under constant observation during ventilation.
Make sure that neither the expiration tube nor the spontaneous breathing tube on the patient valve is blocked or impeded in any other way, e.g. by the patient’s position.
18 Safety instructions
MEDUMAT Standard a must never be used si­multaneously with a module as this would make it impossible to work to the selected parameters.
Note Disposable hose systems WM 28110 (2 m) and
WM 28188 (3 m) are only intended to be used once.
Software
Extensive validation tests have been performed to minimize risks arising from software errors.
EN
Accessories
Please protect the silicone and rubber compo­nents against UV radiation and prolonged expo­sure to direct sunlight, as this can make them brittle and friable.
Safety instructions 19
EN

4. Installation

A permanent mounting is usually necessary only when MEDUMAT Standard a is installed as a fixture in rescue vehicles, helicopters or aircraft. In these cases either the STATION MEDUMAT or installation kits are available as accessories.
If MEDUMAT Standard a is supplied complete with carrying platform, it is ready for use and requires no further installation. Separate directions for use are supplied with the carrying platform.
Functional tests must be carried out after installation (note “7. Functional checks“ on page 45).

4.1 Wall mounting for STATION MEDUMAT

The wall mounting for the STATION MEDUMAT should be installed at an appropriate place, e.g., on a side panel inside the vehicle. Please refer to the sheet enclosed with the STATION MEDUMAT for details of dimensions and the installation procedure.
20 Installation
To place MEDUMAT Standard a in the wall mounting, first insert the slides on the underside of the ventilator into the matching grooves in the STATION MEDUMAT and then press the MEDUMAT Standard a inwards until the catch snaps into the fastening at the top of the STATION MEDUMAT.
EN

4.2 Installation kit for the wall mounting

A number of kits are available for installing a wall mounting, e.g. on a panel inside a vehicle. The kit size varies according to the number of modules at­tached to the MEDUMAT Standard a.
Please refer to the sheet enclosed with each installa­tion kit for details of dimensions and installation procedure.

4.3 Connecting up the oxygen cylinder

Wash your hands thoroughly before any work on the oxygen supply. Hydrocarbon compounds (e.g. oils, greases, alcohol for cleaning, disinfectants, hand cream or sticking plaster) can lead to explosive reac­tions if they come into contact with highly-com­pressed oxygen.
Never use wrenches or similar tools to tighten or loosen the screw connections.
Removal of empty cylinder
1. Close the valve of the oxygen cylinder.
Switch on MEDUMAT Standard a with ON/OFF switch. This exhausts any residual oxygen and depressurizes the ventilator. Wait until the pres­sure gauge on the pressure reducer shows oxygen content before uncoupling the screw connection by hand.
2. First switch off MEDUMAT Standard a again.
3. Then loosen the screw connection to the cylinder.
Installation 21
0 bar
EN
Connecting up new cylinder
I
1. First open the valve of the new oxygen cylinder and close it again quickly. This will blow out any particulate matter.
Keep the valve opening away from the body, making sure that neither yourself or other per­sons can be injured by escaping particles!
2. Next couple the pressure reducer to the valve on the oxygen cylinder with the fluted connecting nut. Tighten up the nut by hand.
3. If the pressure hose is not already connected to the exit valve of the pressure reducer, make this connection with the G 3/8 connecting nut.
4. Screw the other end of the pressure hose on to pressure gas connection on the MEDUMAT Standard a if this has not yet been done.

4.4 Ventilation hose

22 Installation
1. Slide the pressure gauge hose onto the connection.
2. Slide the ventilation tube onto the connection. Make sure that the pressure gauge hose already connected is not kinked. If necessary, turn the ventilation tube while sliding on as appropriate.
Do solely grasp the ventilation tube by its end (posi­tion of arrow in adjacent drawing). Otherwise the hose may become damaged or tear.
EN
3. Plug the other end of the ventilation tube and the
Po
s
it
i
o
n
>
PS
U
<
1
3
4
°
P
o
s
i
t
io
n
>
P
S
U
<
1
34°
pressure measurement tube onto the patient valve.
4. If the patient is intubated, insert the patient valve into the tube,
or
if a mask is being used for ventilation, insert the patient valve into the connector on the mask. (This is identical with the connector on the tube.)
HME-Filter
If a heat and moisture exchanger (HME) filter is required, this should be installed between the pa­tient-side connector on the patient valve and the tube or mask.
Always follow the manufacturer’s directions for use.
Installation 23
EN
PEEP Valve
If a PEEP valve is needed, this should be inserted into the expiration tube on the patient valve.
Always follow the manufacturer’s directions for use.
Bacteria filter
If a bacteria filter is used, fit it between the device outlet port and the ventilation tube.
Always follow the manufacturer’s directions for use.
24 Installation
EN

5. Using the ventilator

Stenosis Disconnection < 2,7 bar O
2
No assist

5.1 Switching on/self test

1. Open the valve of the oxygen cylinder slowly. The pressure gauge will now show the pressure in the cylinder.
2. Where appropriate, calculate how long the re­maining oxygen will last (note “5.11 Calculation of oxygen content/remaining operating time“ on page 37). Always change the cylinder in good time, e.g., when the pressure is lower than 50 bar, to ensure that oxygen is available for an adequate period.
3. Select the desired settings for the ventilation (note “5.2 Selecting the ventilation settings“ on page 26).
4. Switch on the MEDUMAT Standard a with ON/ OFF switch. The ventilator will then run a self test lasting approx. 2 seconds.
If no fault is detected, the five LEDs in alarm field come on and a brief alarm sounds. After that, the MEDUMAT Standard a starts working with the set ventilation values in „Controlled Ventilation“ mode.
If an error is found, the LED’s in alarm panel will start to flash. If this happens, MEDUMAT Standard a must not be used for ventilation.
Using the ventilator 25
EN

5.2 Selecting the ventilation settings

MEDUMAT Standard a must never be used
simultaneously with a module as this would make it impossible to work to the selected parameters.
Air Mix
No Air Mix
The settings can be selected either before or after the MEDUMAT Standard a is switched on. We recom­mend selection before switching on to prevent unnecessary waste of oxygen.
Air Mix/No Air Mix
In the case of a given indication, it is possible to ventilate using pure oxygen or using mixed air.
1. For mixed air, set switch to
The oxygen concentration administered will nor­mally lie somewhere between 55% and 85% at a ventilation pressure of 10 mbar. You can read off the exact figure from relevant diagram (note “11.5 O
content when using Air Mix“ on
2
page 72).
No Air Mix setting should be used when the
The surrounding atmosphere is polluted or has a low oxygen content or when the indication requires this.
Air Mix.
No Air Mix
26 Using the ventilator
Air Mix
2. Set Switch to
No Air Mix.
When the ventilator is switched from an air/oxy­gen mixture (
Air Mix) to pure oxygen (No Air Mix),
the minute volume will vary only within the set tolerances (note “11. Technical data“ on page 68).
EN
Respiratory frequency
MV (l/min)
1. The respiratory frequency can be set with regulator knob.
Minute volume
1. The minute volume can be set with regulator knob.
Recommendation for breathing frequency and minute volume in the case of controlled ventilation:
Toddler
yellow
Body weight 10 - 30 kg 30 - 60 kg 60 - 110 kg
Respiratory frequency
Minute volume 3 - 5 l/min 5 - 7 l/min 7 - 13 l/min
25 - 40 min-115 - 25 min-18 - 15 min
The values given in the table are recommendations. Deviating values are possible for certain indications.
Child
orange
Adult
brown
-1
Maximal ventilation pressure
1. The ventilation pressure can be set with regula­tor knob max. ventilation pressure.
Recommendations for the Maximal ventilation pressure:
Intubation Mask ventilation
45 mbar 20 mbar
The values given in the table are recommendations. Deviating values are possible for certain indications.
Using the ventilator 27
EN
If the set level is reached, e.g. in cases where compli-
Ventilation pressure
in mbar
Time in s
Expiration
Inspira­tion
Ventilation frequency
30
20
10
Example of a ventilation curve in Controlled Ventilation mode:
ance is inadequate, MEDUMAT Standard a sets off a stenosis alarm (note “ Stenosis alarm“ on page 34).

5.3 Select ventilation method

Ventilation in Controlled Ventilation mode
The MEDUMAT Standard a is automatically in Controlled Ventilation mode when switched on. The LED in key
The patient is supplied with air for breathing at an in­spiration/expiration ratio of 1: 1.67 according to the set ventilation parameters (note “5.2 Selecting the ventilation settings“ on page 26).
Assist does not come on.
Ventilation in Assisted Ventilation mode
28 Using the ventilator
To switch on „Assisted Ventilation“ mode, please press key
Assist.
„Assisted Ventilation“ mode is indicated by the green LED in key
Assist flashing.
EN
Triggered ventilation
Time in s
Inspira­tion
30
20
10
Trigger
window
Triggered
ventilation pulse
Expiration
Trigger
pulse
Spontaneous breath
Example of a ventilation curve in Assisted Ventilation mode:
stroke: the patient triggers a ventilation stroke through his own breathing effort.
During assisted ventilation, the patient likewise re­ceives a controlled ventilation stroke in accordance with the ventilation frequency set. In addition, the patient is given the option of trigger­ing the device himself before a controlled ventilation stroke. This synchronises the controlled ventilation strokes with the breathing efforts of the patient.
Note: different trigger points in consecutive phases can lead to a change in the in­spiration/expiration ratio. The set values for breath­ing frequency and minute volume are maintained.
The patient furthermore has the option of perform­ing a spontaneous breath via the patient valve be­tween triggered ventilation strokes. In this case, the patient draws air for breathing from the ambient air.
If the patient fails to trigger the device within the time window in two consecutive phases, i.e., is making no more breathing effort, the No Assist alarm is triggered (note “5.10 Alarm signals“ on page 33).
You end „Assisted Ventilation“ mode by pressing key
Assist. The MEDUMAT Standard a then continues
working in “Controlled Ventilation” mode. The LED in key
Assist goes out.
Using the ventilator 29
EN

5.4 Performing ventilation

Intubation
The patient will normally be intubated before the patient valve is connected to the tube.
1. Attach the patient valve to the connector of the tracheal tube.
2. Monitor the respiratory parameters during ventila­tion, e.g., with Weinmann’s CAPNOCOUNT mini WM 97144. This will indicate whether the tube is correctly positioned and ventilation is ade­quate.
Ventilation mask
1. Attach the mask to the patient valve.
2. Place the mask over the patient’s mouth and nose.
3. Stretch the mask over the head and use Esmarch’s grip to seal it hermetically.
If necessary, insert a Guedel tube to keep the airways open before putting on the mask.

5.5 Monitoring ventilation

The patient must be monitored constantly during ventilation.
Ventilation pressure gauge shows the ventilation sequence.
High airway resistance, e.g., caused by obstructions or external cardiac massage, tend to affect minute volume (exact details can be obtained from the
30 Using the ventilator
EN
manufacturer Weinmann on request). If the preset
Example for the ventilation sequence before and after the lung compliance diminishes
maximal ventilation pressure is exceeded in two suc­cessive inspiration phases (note “ Stenosis alarm“ on page 34) you should use a respirometer to check the ventilation volume actually being received by the pa­tient. The respirometer can be attached to the expi­ration tube on the patient valve with an adapter. Monitor the respiratory parameters during ventilation, e.g., with Weinmann’s CAPNOCOUNT mini WM 97144.
If lung compliance diminishes during ventilation in the
No Air Mix setting, the ventilator will react with
an increase in ventilation pressure at constant vol­ume. For details of how MEDUMAT Standard a re­acts in the “11.5 O
Air Mix setting, please refer to Section
content when using Air Mix” on page 72.
2

5.6 Ventilation with PEEP Valve

A PEEP valve can be fitted to the expiration tube on the patient valve with an adapter.
This valve makes it possible to use positive end­expiratory pressure (PEEP). Please see the PEEP valve instructions for details of settings.
Using the ventilator 31
EN

5.7 Ventilation with HME filter

A conventional heat and moisture exchange (HME) filter with standard 15/22 mm connectors can be fitted on the inspiration tube of the patient valve for hygienic purposes and to condition the inspired air. This will increase both inspiratory and expiratory resistance. You should therefore monitor ventilation pressure and ventilation volume very carefully.
A close watch must be kept for any increase in dead space, especially in children.
Always read and follow the manufacturer’s directions for use.

5.8 Ventilating with bacteria filter

To protect the patient and the environment from in­fection, you can connect the upgrade set for the de­vice outlet port filter between the device outlet port and the ventilation tube. Always monitor the exch­ange of air at the device inlet carefully.
32 Using the ventilator
Follow the instructions for use for the device outlet port filter upgrade set.
EN

5.9 Terminating ventilation

Stenosis Disconnection < 2,7 bar O
2
No assist
Important note! Never empty the oxygen cylinder completely. Return the cylinder for filling while it still contains residual pres­sure. This prevents entry of moist atmospheric air that can cause corrosion.

5.10 Alarm signals

1. Check the oxygen supply on the gauge on the pressure reducer gauge. If the pressure has dropped to 50 bar or less, the cylinder should be refilled or a reserve cylinder should be available in order to ensure that the ventilator will be able to function properly.
2. Close the valve of the oxygen cylinder.
3. Switch off MEDUMAT Standard a. ON/OFF switch must be kept pressed down for at least 2 seconds until the LEDs in the alarm panel light up. This is a safety device to prevent the ventila­tor from being switched off unintentionally.
Alarm panel signals the following alarms:
Stenosis: Stenosis or a rise to maximal ventila-
tion pressure P
in two successive
max
inspiration phases
Disconnection:Disconnection between
MEDUMAT Standard a and the patient in two successive inspiration phases
< 2,7 bar: Drop in oxygen pressure to below
2.7 bar
: Battery charge inadequate
Using the ventilator 33
EN
No Assist: in „Assisted Ventilation“ mode,
patient fails to trigger within the time window in two consecutive phases.
In addition to all the visual alarms, an acoustic alarm is triggered. Only in the case of the No Assist alarm, is the acoustic alarm triggered with a delay of 1 min­ute.
The patient valve is designed to enable spontaneous breathing in case of equipment failure.
When is the alarm set off?
An alarm signal is given as soon as any one of the aforementioned functional failures occurs. The rele­vant LED starts to flash and an acoustic signal sounds.
Simultaneous disconnection and drop in oxygen pressure will initially set off only the < 2,7 bar alarm.
Stenosis alarm
Actual ventilation pressure exceeds the level set with regulator knob max. ventilation pressure.
Up to Serial No. 1799 MEDUMAT Standard a switches to expiration immedi-
From Serial No. 1800 MEDUMAT Standard a briefly switches to expiration
34 Using the ventilator
ately if the maximum ventilation pressure is exceeded. This does not affect the set frequency. Old appliances (up to serial no.1799) that have been given a new board in the course of repairs or servicing behave in the same way as appliances from serial no. 1800 on­wards.
whenever the maximum ventilation pressure is ex­ceeded but tries to continue inspiration in the same inspiration phase.
EN
If the maximum ventilation pressure is exceeded for a second time during the same inspiration phase, the device finally switches to expiration and vents the patient tube system completely. The next inspiration starts with the following ventilation stroke according to the frequency selected. The frequency is not affected.
The alarm is set off when airway resistance is exceed­ed in two successive inspiration phases. This is to prevent false alarms, alarms caused by coughing for example.
Disconnection alarm
The rise in pressure during the inspiration phase is less than 5 mbar. This is generally due to an interrup­tion in the breathing system.
The alarm is set off when the rise in pressure is less than 5 mbar in two successive inspiration phases.
< 2,7 bar O2 alarm
Oxygen pressure at the pressure connection to MEDUMAT Standard a has dropped to less than
2.7 bar. The reason is usually an almost empty
oxygen cylinder.
In this case MEDUMAT Standard a will no longer function correctly because the operating parameters will not lie within the permissible limits.
Alarm
The battery is failing. Failure of the automatic venti­lation function must be expected. Immediate steps must be taken to provide alternative ventilation (note “5.12 Alternative ventilation procedures“ on page 38).
Using the ventilator 35
EN
No Assist Alarm
In “Assisted Ventilation” mode, patient fails to trig­ger within the time window in two consecutive phases.
The LED in alarm field flashes and the acoustic alarm is triggered after a 1-minute delay.
Cancelling acoustic alarm
The acoustic alarm can be temporarily cancelled by pressing alarm acknowledgement:
Stenosis: 30 Seconds Disconnection: 30 Seconds < 2,7 bar: 30 Seconds
: 120 Seconds
No Assist: 120 Seconds
The optical alarm will continue to flash.
If the cause of the alarm is not eliminated, the acous­tic alarm will start to sound again after a short inter­val.
Both the optical and the acoustic alarm are cancelled automatically as soon as the cause is eliminated.
36 Using the ventilator
EN
5.11 Calculation of oxygen content/remaining
Real ventilation time(min)
oxygen content (l)
MV (l/min)
----------------------------------------------
100
O
2
-concentration()
--------------------------------------------------¥=
x
Real ventilation time(min)
1000l
11l/min
---------------------
100
100%
----------------¥91min 1h31min===
x
operating time
Oxygen content of cylinder
Oxygen volume = Volume of cylinder x cylinder pressure.
Cylinder volume x cylinder pressure = oxygen content
Example 1
Example 2
Real ventilation time
Example 1:
O
-content = 1000 l; MV = 11 l/min; 100% O2 (No Air Mix).
2
This gives the following equation:
10 l x 200 bar = 2000 l
10 l x 100 bar = 1000 l
The real ventilation time is prolonged when MEDUMAT Standard a is switched to
Mix.
Using the ventilator 37
Air
EN

5.12 Alternative ventilation procedures

If MEDUMAT Standard a ceases to function during a ventilation procedure, the following alternatives can be applied:
Ventilation bags
1. Remove the patient valve from the tube or the mask.
2. Replace it with the ventilation bag, e.g. a Weinmann COMBIBAG WM 11000, and perform manual ventilation.
A ventilation aid
Alternatively, you can perform mouth/mask ventila­tion with a Weinmann LIFEWAY WM 10580.
Exhaustion of oxygen supply
In emergency situations when the oxygen supply runs out, MEDUMAT Standard a can also function with compressed air.
38 Using the ventilator
EN

6. Hygienic preparation

Whenever MEDUMAT Standard a and accessories have been used, they must undergo a hygienic preparation. Observe the instructions regarding use of disinfectant. We recommend gigasept for disinfecting by immersion and terralin for disinfecting by wiping.
Make sure you perform a functional check after every hygienic preparation (note “7. Functional checks“ on page 45).
This product may contain disposable items. Dispos-
able items are intended to be used only once. So use these items only once and do Reprocessing disposable items may impair the func­tionality and safety of the product and lead to un­foreseeable reactions as a result of ageing, embrittlement, wear, thermal load, the effects of chemical processes, etc.
®
FF (new)
®
protect
not reprocess them.

6.1 MEDUMAT Standard a

MEDUMAT Standard a’s outer casing simply needs to be wiped with a dry or damp cloth.
Never immerse MEDUMAT Standard a in disinfect­ant or other fluids. Just wipe over with disinfectant. Otherwise damage may be caused to the device, thus endangering users and patients (note “6.6 Cleaning, disinfecting and sterilizing“ on page 44).
Hygienic preparation 39
EN

6.2 Patient valve

Lip membrane
Disc membrane Expiration tube
Disc membrane Spontaneous breathing tube
Always grasp the hoses by their ends. Otherwise you might damage or tear them.
1. Disconnect the patient valve from the hoses.
2. Dismantle the patient valve as shown in the ad­jacent diagram. It is neither necessary nor per­missible to remove the membrane in the spontaneous breathing nozzle for cleaning and disinfection.
3. Clean the components under running water.
4. Brush the parts of the patient valve inside and outside thoroughly, using a normal soft labora­tory bottle brush.
5. Carry out sterilization/disinfection.
6. Make sure all internal and external surfaces are thoroughly wetted and free from bubbles. Wait for the full disinfection time to elapse.
7. Then rinse the components thoroughly inside and out with distilled water.
40 Hygienic preparation
8. Dry the components thoroughly.
9. Reassemble the patient valve.
When reassembling, make sure that the lip membrane is correctly positioned. The patient valve may not function properly if it contains re­sidual water!
Crinkled, misshapen and sticky lip and valve membranes must be replaced.
10. Always perform a functional check before reus­ing the valve (note “7.4 Check of patient valve“ on page 48).
EN

6.3 Hose system

I
Caution! Only reusable hose system WM 22520 (scope of
supply) is suitable for the hygienic preparation de­scribed here. Do WM 28110 (2 m) and WM 28188 (3 m) available as accessories to hygiene preparation. Replace it with a new one.
not subject disposable hose systems
Ventilation hose
1. Take the ventilation tube and the pressure gauge hose off both connection ports.
Warning! Take hold of the hoses at the end as
shown in the drawing, otherwise the hoses may be damaged or pulled off. Seal both ends of the pressure gauge hose.
2. Carry out sterilization/disinfection.
3. Make sure all internal and external surfaces are thoroughly wetted and free from bubbles. Wait for the full disinfection time to elapse.
4. Rinse the ventilation tube thoroughly inside and out with distilled water.
5. Dry the components thoroughly.
6. For reassembling see “4.4 Ventilation hose” on page 22.
Pressure gauge tube
To disinfect the pressure gauge tube of the ventila­tion tube, proceed as follows:
1. Connect one end of the pressure gauge tube to a sterile disposable 20-ml syringe.
Hygienic preparation 41
EN
2. Immerse the other end in the dilute disinfectant solution (for gigasept residence time 15 minutes).
3. Draw the disinfectant solution through the pressure gauge tube into the syringe until the latter is full. Do not flush through the pressure gauge tube in the opposite direction!
4. Detach the syringe from the pressure gauge tube and empty it out completely.
5. Repeat the procedure 5 more times.
6. After disinfection, the pressure gauge tube must be rinsed with distilled water at least 8 times using the same principle.
You can support the subsequent drying process with medical compressed air or medical oxygen.
Then allow the component to dry thoroughly. If any water is left in the pressure gauge tube of the venti­lation tube, the device may not function correctly!

6.4 Components and accessories

®
FF: 6% solution,
42 Hygienic preparation
Masks, hoses and all rubber components should be cleaned in a disinfectant solution:
1. Make sure all internal and external surfaces are thoroughly wetted and free from bubbles. Wait for the full disinfection time to elapse.
2. After disinfection always rinse the components thoroughly with distilled water to prevent any adverse effects from disinfectant residues.
3. Always let the rubber components dry out in the air.
EN
4. Carry out a visual check of the masks and hoses and replace any damaged components.
Silicone ventilation tubes, patient valves (see preced­ing section) and ventilation masks can also be autoclaved.

6.5 Fittings

This does WM 28110 (2 m) and WM 28188 (3 m).
In cases where external cleaning of fittings (e.g., pres­sure reducer, valve) becomes absolutely essential, use only a clean cloth which should either be dry or moistened with clean water.
Never immerse fittings in disinfectant or other fluids. Just wipe over with disinfectant. Fluids must not get into the pressure reducer. Otherwise explosions might occur.
If in exceptional cases you have no alternative but to disinfect by wiping, take particular care to prevent any fluid getting into the pressure reducer.
In addition to the risk of explosion, there is also the risk of disinfectant getting into the patient’s respira­tory tract with the oxygen and leading to injury.
not apply to disposable hose systems
Hygienic preparation 43
EN

6.6 Cleaning, disinfecting and sterilizing

The hygienic preparation should be performed on MEDUMAT Standard a and the accessories used as described in the following table.
Description of
component
MEDUMAT Standard a
Patient valve
Silicone ventilation mask
Ventilation hose
Oxygen fittings
Device outlet port filter upgrade set
Hose casing, reusable
Always follow the directions for use supplied with the disinfectant. We recommend gigasept for disinfecting by immersion and terralin
®
FF (new)
®
protect for disinfecting by wiping. You are recommended to wear suitable gloves (e.g. household or disposable gloves) during disinfection procedures.
For further information on hygiene treatment and a list of all cleaning agents and disinfectants which can be used, please see our Internet brochure at www.weinmann-emergency.com.
Cleaning Disinfection
With a dry or damp cloth
In warm water with a mild household detergent
With a dry or damp cloth
Follow the instructions for use for the device outlet port filter upgrade set
Wipe with a dry or damp cloth
Wiping Not permissible
Immerse in a weak
(1)
solution
Wiping Not permissible
Rinse cycle 30°C, no spin
Rinsing in
washing machine
Rinse program at
(2)
95 °C
Possible during cycle
Sterilization
Not permissible
Steam sterilization at up to 134 °C
(3)
Not permissible
Not permitted
44 Hygienic preparation
(1) After disinfection rinse the components thoroughly with dis-
tilled water and allow them to dry. (2) Thermal disinfection in automatic cleansing device. (3) Steam sterilization at 134 °C with devices to EN 285, dwell
time 5 minutes (or 18 minutes at 121 °C).
EN

7. Functional checks

MEDUMAT Standard a must not be used if the func­tional checks reveal defects or deviations from the selected parameters.
First try to correct the error with the help of the in­formation provided in section “8. Troubleshooting“ on page 57. If this is not possible, have the device re­paired by the manufacturer Weinmann or by special­ists explicitly authorised to do so by same.

7.1 Preparation for functional check

For the functional check, you require:
patient’s hose system – test bag – oxygen cylinder – soap-and-water solution made from non-
perfumed soap
adapter from test set WM 15357
We recommend that you hold reserve stocks of the following items:
washers for the connections;
dust filters;
lip membranes for the patient valve;
membrane for spontaneous breathing arm;
membrane for expiration arm.
1. Connect the device to the oxygen cylinder.
2. Connect the patient’s hose system to the device.
Functional checks 45
EN
Note Check the test bag before each functional
check. The balloon of the test bag must be un­damaged and firmly connected to the connec­tor. Have the test bag serviced at the same time as the device.

7.2 Obligatory checks

Before each use:
Carry out a functional check.
After each use or dismantling:
Clean, disinfect or sterilize the ventilator and its
components (note “6. Hygienic preparation“ on page 39);
•Check the:
lip membrane in the patient valve – valve membrane in the expiration arm – valve membrane in the spontaneous breath-
Carry out a functional check.
ing arm (note “7.4 Check of patient valve“ on page 48). They may be neither crinkled, sticky or misshapen.
46 Functional checks
At least every 6 months, if the ventilator has not been used in the intervening period:
Carry out a functional check.
EN

7.3 Check for leaks in the system

1. Open the valve of the oxygen cylinder slowly. You will now be able to read the pressure in the cylinder from the gauge on the pressure reducer. For example, a reading of 200 bar means that the cylinder is full, 100 bar that it is half full.
Always change the cylinder in good time, e.g., when the pressure is lower than 50 bar, to en­sure that oxygen is available for an adequate period.
2. Close the cylinder valve again.
3. Watch the needle of the gauge on the pressure reducer for approx. 1 Minute. If it stays in the same place, the system is free of leaks. If the needle drops steadily, there is a leak somewhere.
Repairing leaks
Always keep a stock of washers for the connec­tions available.
Important note! The screw connections on the oxygen supply system must be tightened by hand only.
1. Prepare a soap/water solution using non-per­fumed soap.
2. Wet all the screw and hose connections with the solution. Bubbles will form at the site of the leak.
3. Depressurise the system:
To do this, first close the oxygen cylinder. Switch on MEDUMAT Standard a briefly until the pressure gauge on the O
cylinder reads “0“.
2
Then switch MEDUMAT Standard a off again.
4. If leaks are discovered, the defective components must be changed.
5. After changing, recheck the seal.
6. If it proves impossible to eliminate the leak, the ventilator will have to be repaired.
Functional checks 47
EN

7.4 Check of patient valve

Checking reusable hose system
1. Dismantle the patient valve.
2. Carry out a visual check of all the components for cracks or other physical damage.
Crinkled, sticky or misshapen lip membranes must be changed immediately. They must never be used during ventilation as they can cause serious functional disturbances.
Also perform a visual check of the valve mem­branes in the expiration and spontaneous breathing arms. To do so, there is no need to dismantle the valve membranes. Crinkled, mis­shapen or sticky valve membranes must be re­placed, however, as they can lead to consider­able malfunctions.
3. Reassemble the patient valve.
48 Functional checks
When reassembling, make sure that the lip membrane is correctly positioned.
EN
Checking disposable hose system
Visual inspection
Check the following items by inspecting the patient hose system:
the patient valve and the connectors may not exhibit any external damage, cracks or dirt.
the hose connections must be located firmly and securely on the connecting pieces.
the patient valve and emergency air membranes may not exhibit any damage or deformation.

7.5 Checking the minute volume

Check the ventilation frequency
1. Open the valve of the oxygen cylinder slowly.
2. Switch on MEDUMAT Standard a.
3. Select the following settings. – Frequency: 8 min
-1
–MV: 5 l/min – Ventilation pressure (P
): 60 mbar
max
Air Mix: switched on.
4. Count the number of inspiration phases over a period of one minute exactly. The number should lie between 7 and 9.
5. Turn up the frequency to 40 min
-1
(right limit).
6. Count the number of inspiration phases over a period of one minute exactly. The number should lie between 38 and 42.
7. Switch MEDUMAT Standard a off again.
Functional checks 49
EN
Inspiration stroke = MV/frequency = 8/8 = 1
Check the breath volume
1. MEDUMAT Standard a must be switched off and the oxygen cylinder must be open.
2. Place the bag from test kit WM 15382 on to the patient valve.
3. Select the following settings: – Frequency: 8 min –MV: 8 l/min –P
: 60 mbar
max
–No Air Mix
Note For the test, place the test bag on a firm base.
During the expiration phase, press the test bag with the flat of your hand until the volume is completely expelled through the patient valve.
4. Switch on MEDUMAT Standard a. The test bag must be completely inflated at the end of the in­spiration phase. This ensures a breath volume of 1 litre per inspiration stroke. At all events the test bag is not sufficiently inflated if a disconnection alarm is set off. During the expiration phase you must simulate the expiration stroke of the test bag by hand.
5. Switch off MEDUMAT Standard a.
-1
50 Functional checks
Risk of injury if test bag removed incorrectly!
If the test bag is removed incorrectly, the con­nector of the test bag may remain on the pa­tient’s hose system. The increased airway resis­tance on inspiration which results may injure the patient.
When removing the test bag, always take it
off at the connector.
6. Detach the test bag from the patient valve.
7. Select the following setting:
EN
Frequency 30 min
-1
–MV 3 L/min –P
max
60 mbar
–No Air mix
8. Switch on MEDUMAT Standard a and close the patient connection to the patient valve. A steno­sis alarm should be set off.
9. Switch MEDUMAT Standard a off again.
Note Instead of the test bag you can also use a respiro-
meter (see accessories) to check the breath volume.

7.6 Check of maximal ventilation pressure

Important note! For this test, always set the tumbler switch to “No Air Mix”. For physical reasons, losses occur at the injector if the switch is set to “Air Mix”.
Important note! Use the test bag. If you try to keep the tube connector closed with your hand, the needle will swing over and it will be impossible to ob­tain an accurate reading.
1. MEDUMAT Standard a must be switched off and the oxygen cylinder must be open.
2. Check that the needle of the manometer on MEDUMAT Standard a is standing at “
0“ (note
“8.3 Adjustment of manometer“ on page 61).
3. Place the bag from test kit WM15382 on to the patient valve.
4. Select the following settings: – Frequency: 8 min
-1
–MV: 7 l/min –P
: 20 mbar
max
–No Air Mix
5. Switch on MEDUMAT Standard a.
During this test you must not assist the expira­tion stroke. The oxygen must remain unpressur­ized in the test bag. In this way the pressure gradually builds up. Between 15 and 25 mbar MEDUMAT Standard a should set off the
Stenosis alarm.
6. Switch off MEDUMAT Standard a.
Functional checks 51
EN
Risk of injury if test bag removed incorrectly!
If the test bag is removed incorrectly, the con­nector of the test bag may remain on the pa­tient’s hose system. The increased airway resis­tance on inspiration which results may injure the patient.
When removing the test bag, always take it
off at the connector.
7. Detach the test bag from the patient valve.
8. Repeat this test for a maximum ventilation pressure of 60 mbar with the following setting: – Frequency: 8 min
-1
–MV: 9 l/min –P
: 60 mbar
max
–No Air Mix
During this test you must not assist the expira­tion stroke. The oxygen must remain unpressur­ized in the test bag. In this way the pressure gradually builds up. Between 55 and 65 mbar MEDUMAT Standard a should set off the
Stenosis alarm.

7.7 Check assisted ventilation

1. The MEDUMAT Standard a must be switched off and the oxygen cylinder must be open.
2. Plug the test bag with adapter of test set WM15382 onto the patient valve.
3. Select the following setting:
frequency: 8 min – MV: 8 l/min. –P –Air Mix
52 Functional checks
: 60 mbar
max
-1
EN
4. Switch on the MEDUMAT Standard a.
Stenosis Disconnection < 2,7 bar O
2
No assist
5. Press the Assist key to switch on „Assisted Ventilation“ mode.
„Assisted Ventilation“ mode is indicated by the flashing green LED in the Assist key.
6. Wait 2 inspiration phases.
The No Assist alarm should then be triggered: the yellow No Assist LED in alarm field flashes. The acoustic alarm comes on only after 1 min­ute.
7. Simulate inspiration pulses with one hand by pressing the test bag together several times.
The pointer of ventilation pressure gauge
must go into the negative range during the simulation. Negative pressure must be at least – 0.8 mbar;
As soon as the MEDUMAT Standard a
detects the pulse within the time window, the yellow No Assist LED goes out. This en­sures that the MEDUMAT Standard a recog­nizes trigger pulses.
8. Switch the MEDUMAT Standard a off again.
Risk of injury if test bag removed incorrectly!
If the test bag is removed incorrectly, the con­nector of the test bag may remain on the pa­tient’s hose system. The increased airway resis­tance on inspiration which results may injure the patient.
When removing the test bag, always take it
off at the connector.
9. Detach the test bag from the patient valve.
1
Functional checks 53
EN

7.8 Check of alarm systems

Important note! The stenosis and disconnec­tion alarms are set off only when the cause of the alarm is repeated in two successive inspiration phases. This prevents trig­gering of the alarm by a very short-lived dysfunction.
Important note! In this test the rise in pres­sure is strong enough to make the manometer nee­dle overswing considera­bly. There are technical reasons for this and it does not indicate any malfunc­tion.
Important note! In this test the rise in pres­sure is strong enough to make the manometer nee­dle overswing considera­bly. There are technical reasons for this and it does not indicate any malfunc­tion.
Stenosis
1. Open the oxygen cylinder.
2. If necessary: Remove the tube or the ventilation mask from the patient valve.
3. Switch on MEDUMAT Standard a
4. Set max. ventilation pressure regulator at 60 mbar.
5. Keep the ventilation connector on the patient valve closed with the flat of your hand during two successive inspiration phases. The
Stenosis
alarm should be set off.
6. Switch the MEDUMAT Standard a off again.
Interruption of breathing system (Disconnection)
1. Open the oxygen cylinder.
2. If necessary: Remove the tube or the ventilation mask from the patient valve.
3. Switch on MEDUMAT Standard a
4. Set max. ventilation pressure regulator at 60 mbar.
5. Keep the ventilation connector on the patient valve closed with the flat of your hand during two successive inspiration phases. The alarm should be set off.
6. Then remove your hand. The
Stenosis alarm
should cease (LED stops flashing, acoustic alarm stops sounding).
After two successive inspiration phases the Disconnection alarm should be set off.
Stenosis
54 Functional checks
EN
7. Switch the MEDUMAT Standard a off again.
Drop in O2 pressure (<2,7 bar O2)
1. Open the valve of the oxygen cylinder slowly.
2. Switch on MEDUMAT Standard a.
3. Close the valve on the oxygen cylinder. When the oxygen pressure in the system has sunk be­low 2.7 bar, the
<2,7 bar O
alarm should be set
2
off.
4. Switch the MEDUMAT Standard a off again.
Energy supply ( )
The alarm signalling a failing battery is checked automatically in the self test that runs when MEDUMAT Standard a is switched on.
The energy supply is in order if no alarm is set off when the valve on the oxygen cylinder is opened and MEDUMAT Standard a is switched on and starts to function correctly.
Failure to trigger (No Assist)
1. The oxygen cylinder must be open.
2. Switch on the MEDUMAT Standard a.
3. To switch on „Assisted Ventilation“ mode, press
Assist.
key
„Assisted Ventilation“ mode is indicated by the flashing green LED in key
Assist.
Functional checks 55
EN
4. After the second ventilation stroke, the yellow
Stenosis Disconnection < 2,7 bar O
2
No assist
alarm message „No Assist“ must flash in alarm field. The acoustic alarm sounds after a delay of 1 minute if the fault has not been eliminated in the meantime.
5. Switch the MEDUMAT Standard a off again.
56 Functional checks
EN

8. Troubleshooting

Defect Cause of defect Elimination
MEDUMAT Standard a cannot be switched on
Stenosis alarm (excessive airway resistance)
Disconnection alarm (interruption of breathing system)
< 2,7 bar alarm (oxygen pressure too low)
Alarm
Alarm No Assist
MEDUMAT Standard a defective Arrange for repair
Battery failure Replace both batteries (8.1, page 59)
Obstruction of airways Remove obstruction.
Tube incorrectly positioned Correct tube position.
P
set too low Adapt P
max
Kink or obstruction in ventilation hose/mask/tube
MEDUMAT Standard a defective Arrange for repair
Ventilation hose leaking/slipped out
Pressure gauge hose leaking/ slipped out/not connected
MEDUMAT Standard a defective Arrange for repair
Oxygen cylinder nearly empty Change O2 cylinder (4.3, page 21)
Oxygen valve closed Open oxygen valve
Pressure reducer defective Replace pressure reducer
Kink or blockage in oxygen hose Take action to correct
Battery failing Replace both batteries (8.1, page 59)
Fuse defective Replace fuse (8.2, page 60)
Patient does not trigger device within time window
Patient does not trigger device at all
Valve membrane in spontaneous breathing arm defective or missing
max
Remove kink or obstruction; if necessary replace parts.
Check connectionsTube/mask incorrectly positioned
Adapt ventilation frequency to suit patient
Continue ventilating in Controlled Ventilation mode
Insert new valve membrane (8.4, page 61))
.
Troubleshooting 57
EN
Defect Cause of defect Elimination
Visual alarms flashing but no acoustic alarm.
Acoustic alarm but no visual alarm
Acoustic alarm and all visual alarms flashing
MEDUMAT Standard a is functioning but without any displays
MV too low
Unusually high oxygen consumption
MEDUMAT Standard a cannot be switched off
Manometer needle not standing at “0“
Test bag is not sufficiently inflated during functional check, disconnection alarm
No stenosis alarm when patient valve is closed during functional check(see “7.5 Checking the minute volume”, item 6–9)
Short-term electronic disruption
Pressure gauge hose on MEDUMAT Standard a or on patient valve slipped off
Kink in pressure gauge hose
Wrongly selected ventilation parameter
MEDUMAT Standard a defective Arrange for repair
Leak in oxygen supply Seek and eliminate leak (7.3, page 47)
User error
Manometer needle needs adjustment
Ventilation parameters wrongly selected
Patient valve not working properly
Pressure gauge hose not fitted Fit pressure gauge hose
Patient valve not working properly
Switch off and on again. If error recurs, arrange for repair.
Check pressure gauge hose
Check ventilation parameters
Keep switch depressed for at least 2 seconds.
Adjust (8.3, page 61)
Correct ventilation parameters
Check lip membrane
Check lip membrane
58 Troubleshooting
EN

8.1 Batteries

18: Button cell for auxiliary power 19: Fuse protecting against internal
short circuit
20: 3.6 V lithium battery for main
power supply to MEDUMAT Standard a
18
20
19
Caution! Batteries and rechargeable batteries do not belong in domestic waste! Every consumer is obliged by law to hand in all batteries and rechargeable batteries, re­gardless of whether they contain harmful substances or not, at a community/lo­cal collection point or at a store so that they can be disposed of in an environ­mentally-friendly way.
MEDUMAT Standard a is fitted with two batteries which should always be replaced simultaneously. A CR2430 button cell supplies auxiliary power to the electronics if the main batteries fail. This makes it possible to set off an alarm if the main batteries fail. At the same time, the MEDUMAT Standard a switches to expiration.
The battery capacity is calculated to be sufficient for power requirements under normal operating condi­tions for the full period between servicing every two years. All the batteries are replaced at every two years during servicing.
We recommend having the batteries changed only by Weinmann the manufacturer or by qualified per­sons expressly authorized by that company. Special precautions need to be taken during the change in order to prevent disruption of the electronics.
The following action should be taken in emergen­cies:
Troubleshooting 59
EN
Important note! In order to prevent damage to the electronics, avoid all contact with the printed circuit board unless ade­quate protection against electrostatic charge is available.
Important note! The 3.6 V lithium battery is specially designed for this ventilator. Only batteries supplied by Weinmann should be used.

8.2 Cut-out system

Battery replacement
1. Unscrew the back panel of MEDUMAT Standard a (6 cross-slotted screws).
2. Tilt the CR2430 button cell slightly to remove it.
3. Insert a new button cell.
4. Remove the plug from the printed circuit board and lift the 3.6 V lithium battery out of its hold­er.
5. Insert the new battery.
6. Replace the back panel of MEDUMAT Standard a.
MEDUMAT Standard a is fitted with a fusible cut-out to protect it against internal short circuits.
Important note! In order to prevent damage to the electronics, avoid all contact with the printed circuit board.
60 Troubleshooting
Changing a fuse
1. Unscrew the back panel of MEDUMAT Standard a (6 cross-slotted screws).
2. Remove the defective fuse.
3. Insert a new fuse. Use only officially approved fuses (see “11. Technical data” on page 68).
EN
4. Replace the back panel and carry out a function-
adjustment
screw
Spontaneous breathing arm
Expiration­arm
al check (see “7. Functional checks” on page 45).

8.3 Adjustment of manometer

When MEDUMAT Standard a is switched off and the valve of the oxygen cylinder is closed, the manome­ter needle should read exactly “
0“.
Proceed as follows to adjust the needle:
1. Carefully pry off the plastic cover of the adjust­ment screw.
2. Adjust the position of the needle by turning the adjustment screw with a small screwdriver (e.g. a watchmaker’s screwdriver).
3. Reinsert the plastic cover.

8.4 Change valve membrane in patient valve

If one of the valve membranes in the expiration or spontaneous breathing arms of the patient valve is crinkled, sticky or misshapen, it must be changed.
Spontaneous breathing arm
1. Take the spontaneous breathing insert out of the patient valve. To do so, push the two lock­ing lugs out of their seat, using a small screw­driver, for example.
2. Pull the defective valve membrane out of the spontaneous breathing insert using pointed tweezers.
Troubleshooting 61
EN
3. Put in a new valve membrane.
4. Push the spontaneous breathing insert back into the patient valve.
Expiration arm
1. Use pointed tweezers to pull the defective valve membrane out of the expiration arm.
2. Insert a new valve membrane.
Note:
This applies only to the reusable system.
62 Troubleshooting
EN

9. Servicing

9.1 Intervals

Have the cleaned and disinfected device serviced at regular intervals. Servicing, safety checks ([sicherhe­itstechnische Kontrollen or STKs] in accordance with
§6 of the German law governing the owners/opera­tors of medical devices - only applies to Germany) and maintenance measures such as servicing and re­pairs may only be performed by the manufacturer or by specialists expressly so authorized by the manu­facturer.
Maintain the following intervals:
Interval Parts affected
– System components: e.g. carrying
systems, tube connections Every 2 years (service and safety check)
Every 4 years
Every 10 years Steel and aluminum oxygen cylinders
– Accessories – Test bag – Oxygen fittings – Specified wear parts relevant to safety
– Oxygen fittings – Specified wear parts relevant to safety
* Disposable hose systems WM 28110 (2 m) and
WM 28188 (3 m) do not require any maintenance.
Person to carry out
Manufacturer or specialists expressly so authorized by the manufacturer
Servicing 63
EN

9.2 Sending in device

Warning!
Risk of infection from contaminated parts
during maintenance measures!
The device, components and accessories may be
contaminated and infect specialist staff with bacteria
or viruses when they are carrying out maintenance
measures.
Clean and disinfect device, components and ac-
Do not send in potentially contaminated parts.
1. Remove components and accessories.
2. Clean the device, components and accessories
3. Send device, and if necessary components and
Note If you send in obviously contaminated parts,
cessories.
(note “6. Hygienic preparation“ on page 39).
accessories, to WEINMANN Emergency or to specialist staff expressly authorized by WEINMANN Emergency.
these will be disposed of by WEINMANN Emer­gency or by specialist staff expressly authorized by WEINMANN Emergency at your expense.

9.3 Storage

64 Servicing
If you do not intend to use MEDUMAT Standard a for
a long period, we recommend the following storage
precautions:
1. Clean and disinfect the ventilator (note “6. Hygienic preparation“ on page 39).
EN
Important note! Remember that the ventilator still requires servic-

9.4 Disposal

2. Store MEDUMAT Standard a in a dry place.
ing at the stipulated intervals even when in stor­age, otherwise it cannot be used when removed from storage.
Note With disposable hose systems WM 28110 (2 m)
and WM 28188 (3 m), observe the storage tem­perature of -40 °C to 70 °C at a rel. humidity of 15 % to 95 %. These products can be stored for a maximum of 2 years.
Do not dispose of the device with domestic waste. For proper disposal of the device and its compo­nents, please contact a certified waste disposal site for electronic goods. Ask your Environmental Officer or local council for the address. The device packag­ing (cardboard and inserts) can be disposed of in pa­per recycling facilities.
Disposing of batteries/recheargable battieres
Used batteries/rechargeable batteries may not be disposed of in domestic waste. Contact Weinmann or your local authority waste disposal department.
Servicing 65
EN

10. Supply schedule

10.1 Standard supply schedule

1. MEDUMAT Standard a, complete WM 22800
consisting of:
MEDUMAT Standard a WM 22810 – Description and Operating instructions WM 16677 – Abbreviated directions for use WM 16687 – Set of mounting attachments WM 15288 – Ventilation hose and patient valve
with spontaneous breathing facility (reusable) WM 22520 – Ventilation mask for adults, size 5 WM 5074 – Testing kit for checking functions WM 15382

10.2 Accessories

The following accessories are not included in the standard supply schedule.
1. Oxygen cylinder, 2 liters WM 1822
2. Aluminium lightweight oxygen cylinder, 2 liters WM 1814
3. WM pressure reducer WM 30301
4. WM pressure hose 10 bar with G 3/8 connecting nozzle, at the other end optional G 3/8 connection nut or connector socket to oxygen supply
5. Permanent fixing kit for MEDUMAT Standard a WM 15196
6. Permanent fixing kit for MEDUMAT Standard a plus 1 module WM 15198
7. Permanent fixing kit for a second module WM 15199
8. Wall mounting for STATION MEDUMAT WM 22550
9. PEEP valve with connection socket WM 3215
66 Supply schedule
EN
10. Patient hose system and patient valve 2 m (disposable) WM 28110
11. Patient hose system and patient valve 3 m (disposable) WM 28188
12. Ventilation mask, transparent with inflatable silicone rim
for children and adolescents, size 3 WM 5082
13. Rendell-Baker silicone ventilation mask:
for children aged approx. 3 – 12 years, size 3 WM 5063
14. Device outlet port filter upgrade set WM 15780
15. Hose casing WM 8297

10.3 Spare parts

You can order replacement parts separately if required. You can obtain a current list of replacement parts either on the Internet at www.weinmann-emergency.com or through your specialist dealer.
Supply schedule 67
EN

11. Technical data

MEDUMAT Standard a
Dimensions LxBxH in mm
190x110x90 inc. connections
Weight incl. accessories
approx. 1.1 kg
Prod. category 93/42/EEC
II b
Operating parameters Temperature range Humidity Air pressure
–18 °C to +60 °C 15 % to 95 % 70 kPa to 110 kPa
Storage/transport Temperature range Humidity Air pressure
–40 °C to +70 °C 15 % to 95 % 70 kPa
(a)
to 110 kPa
Electromagnetic compatibility (EMC) following EN60601-1-2 and EN 794-3:
– Interference
suppression
– Interference immunity
EN 55011
EN 61000-4 Parts 2, 3 and 8
Control
Timing pulse, constant pressure, constant volume
Gas input Medicinal oxygen
Operating pressure 2.7 to 6.0 bar
(b)
Minimal gas volume required
70 l/min O
2
Insp-exp. ratio assisted ventilation
1:1.67 1:1 to 1:2.33 variable
Ventilation frequency
infinitely variable from 5 to 40 min
-1
Minute volume (MV)
infinitely variable from 3 to 20 l/min
Tiidal volume
75 to 4,000 ml, infinitely variable
MV tolerances: room temp. (20 °C) –18 °C to +60 °C
±20% ±20%
max. ventilation pressure
infinitely variable from 20 to 60 mbar
(c)
O2 concentration – Air Mix – No Air Mix
see page 72 100% O
2
High-pressure gas connection
External thread G 3/8
Connection to ventilation tube
External diameter 13 mm
Power supply
life expectancy max. storage
maintenance-free lithium battery 3.6 V; 5.2 Ah, > 2 years 10 years after delivery
Auxiliary energy for alarm system max. storage
Button cell CR2430
10 years after delivery
F1 fuse T500 L250V
(a.) Under normal atmospheric conditions, 70 kPa
correspond to a maximum altitude for use of
approx. 3,000 meters. (b.) 1 bar =ˆ 100 kPa (c.) 1 mbar =ˆ 1 hPa
MEDUMAT Standard a

11.1 Device

68 Technical data
EN
Subject to technical change without notice.
MEDUMAT Standard a
Degree of protection against water
IP24
Standards applied
EN 60601-1 EN 60601-1-2 EN 794-3 + A2
Alarm sound pressure
54 dB A
Manometer accuracy
Class 1.6
Patient valve resist­ance (EN 794-3): Inspiration Expiration
<6 mbar at 60 l/min <6 mbar at 60 l/min
Elasticity of breathing system
Negligible
MEDUMAT Standard a
Technical data 69
EN

11.2 Patient’s hose system

Patient’s hose system
(reusable), 2 m
WM 22520
Patient’s hose system
(disposable), 2 m
WM 28110
Patient’s hose system
Operation: Temperature range Humidity:
-18 °C to +60 °C 15% to 95%
Storage -40°C to +70°C
Patient valve – inspiration tube – mask/endotracheal tube
Patient valve – expiration tube
15 mm internal tapered connector
22 mm external tapered connector
EN 5356-1
30 mm external tapered connector
EN 5356-1
Ventilation tube connection WEINMANN Emergency-specific
Standards applied EN 794-3
Patient valve resistance (as per EN 794-3): Inspiration Expiration Spontaneous respiration
<6 mbar <6 mbar
1.5 mbar
(1)
at 60 l/min
(1)
at 60 l/min
(1)
at 30 l/min
<6mbar <6mbar
1.36 mbar
(1)
at 60 l/min
(1)
at 60 l/min
(1)
at 30 l/min
1.51 mbar
Patient valve dead space 12.8 ml 8 ml
®
Materials used PSU, silicone EVA, K-Resin
Elasticity of breathing system
(1)
1 mbar =ˆ 1 hPa
negligible
, PS, PVC (DEHP-free), silicone
(disposable), 3 m
WM 28188
(1)
<6mbar <6mbar
at 60 l/min
(1)
at 60 l/min
(1)
at 30 l/min
70 Technical data
EN

11.3 Pneumatics

input pressure
2.7 - 6 bar
frequency regulator
P/E converter
A/D converter
evaluation electronics
p
V1
pressure regulator
patient valve
exp.-insp.
P/E converter
valve rocker
V2
manometer
injector unit
V8
relief valve
V6
V3
amplifier valve
V4
exhaust valve
V9
non-return valve
V7
air inlet
V5
regulator valve
air mix valve
The input pressure at p is max. 6 bar. This is reduced by V1 to 2,7 bar dyn. This is the input pressure at V6, V2 and V3.
Inspiration/No Air Mix
Valve rocker V6 is opened and switches over V7.
An electrical impulse to V2 opens V3 and closes V4.
Oxygen flows through V5 into injector unit V9 and onwards to the patient valve.
If the ventilation pressure in the patient valve rises above 100 mbar, the relief valve will open.
Technical data 71
EN
Inspiration/Air Mix
Valve rocker V6 is closed. This closes V7. O2 flows into injector unit V9 through V5 and sucks in air through V7. The air-oxygen mixture flows to the patient valve.
Expiration/Air Mix or No Air Mix
Another electrical impulse closes V2. Exhaust valve V4 opens and exhausts injector unit V9. The patient breathes out through the patient valve.
Patient valve
The respiratory gas flows into the patient’s airways during inspiration. The expiratory pressure then switches the valve over and enables the patient to breathe out.
Inspiration
Patient
Medumat
Inspiration Expiration
Expiration

11.4 Resistance to interference

Resistance to radio interference is tested to EN 794­3 at 10 V/m. Malfunctions may occur at higher field strengths.

11.5 O2 content when using Air Mix

The following diagram shows the oxygen concentra­tion prevailing at various counter-pressures and min­ute volumes when
Air Mix is switched on.
72 Technical data
EN
100
95
90
85
80
(%)
2
75
O
70
65
60
55
50
3 5 7 9 11 14 17 20
10 mbar counter-pressure 15 mbar counter-pressure 30 mbar counter-pressure
Minute volume (l/min)

11.6 Switching from Air Mix to No Air Mix

The injector unit is switched off when switching from
Air Mix to No Air Mix. This increases minute vol-
ume which can result in the set pressure limit being
exceeded and a stenosis alarm (Stenosis) being trig-
gered. In this case, set minute volume correspond­ingly
lower.
In the opposite instance, in other words when switching from unit is switched on. This reduces minute volume which can lead to the set pressure limit being
. In this case, set minute volume correspondingly
shot higher.
No Air Mix to Air Mix, the injector
under-
Technical data 73
EN

12. Warranty

WEINMANN Emergency gives the customer a limited manufacturer warranty on new original WEINMANN Emergency products and any replacement part fitted by WEINMANN Emergency in accordance with the warranty conditions applicable to the product in question and in accordance with the warranty periods from date of purchase as listed below. The warranty conditions can be downloaded from www.weinmann-emergency.com on the Internet. We can also send you the warranty conditions on re­quest.
In the event of a claim under warranty, contact your specialist dealer.
Product
WEINMANN Emergency devices including accessories (except masks) for oxygen medicine and emergency medicine
Masks including accessories, rechargeable batteries, batteries (unless quoted differently in the technical documentation), sensors, tube systems
Disposable products None
Warranty
period
2 years
6 months

13. Declaration of conformity

WEINMANN Emergency Medical Technology GmbH + Co. KG declares herewith that the product com­plies fully with the respective regulations of the Med­ical Device Directive 93/42/EEC. The unabridged text of the Declaration of Conformity can be found on our website at www.weinmann-emergency.com
74 Warranty
Manufacturer
WEINMANN Emergency Medical Technology GmbH + Co. KG Frohbösestraße 12 22525 Hamburg GERMANY
Center for Production, Logistics, Service
WEINMANN Emergency Medical Technology GmbH + Co. KG Siebenstücken 14 24558 Henstedt-Ulzburg
GERMANY T: +49 40 88 18 96-120 E: customerservice@weinmann-emt.de
Simply Professional Made in Germany WEINMANN-Emergency.com
WM 16677m 05/2016 EN
Loading...